DIURNAL VARIATIONS OF CENTRAL AND PARACENTRAL CORNEAL THICKNESS AND CURVATURE

Size: px
Start display at page:

Download "DIURNAL VARIATIONS OF CENTRAL AND PARACENTRAL CORNEAL THICKNESS AND CURVATURE"

Transcription

1 ARCH SOC ESP OFTALMOL 2008; 83: ORIGINAL ARTICLE DIURNAL VARIATIONS OF CENTRAL AND PARACENTRAL CORNEAL THICKNESS AND CURVATURE VARIACIONES DIURNAS DE ESPESOR Y CURVATURA CORNEAL CENTRAL Y PARACENTRAL GIRÁLDEZ-FERNÁNDEZ MJ 1, DÍAZ-REY A 2, GARCÍA-RESUA C 1, YEBRA-PIMENTEL-VILAR E 3 ABSTRACT Objective: Corneal topography and thickness has clinical importance in contact lens fitting and refractive surgery, however repeated measurement of corneal thickness and curvature is complicated by the natural phenomenon of diurnal variation. Our aim was determine diurnal variations of central and paracentral corneal thickness and curvature over a period of ten hours. Materials and methods: Corneal thickness and curvature of 10 right eyes of 10 young healthy men were determined by Orbscan Topography System and EyeSys videokeratoscope respectively. Both parameters were determined for central and paracentral regions 1 mm and 2 mm from the centre of the cornea at 2-hour internals for ten hours. Results: The cornea was thickest and flattest on awakening. There was a difference in corneal thickness (ANOVA, Schfféc method p< 0.05) and curvature (ANOVA, Tamhane method p< 0.05) over time for all the corneal locations studied, with greater changes observed in the peripheral corneal data. The minimum value was 10 and 8 hours after eye opening for central and paracentral zones, respectively. Change in central and paracentral corneal RESUMEN Objetivo: La topografía y espesor de la córnea tienen una gran importancia clínica en la adaptación de lentes de contacto y en cirugía refractiva. Sin embargo, la repetibilidad en las medidas de estos parámetros se ve comprometida por el fenómeno natural de las variaciones diurnas. Nuestro objetivo es determinar las variaciones diurnas de espesor y curvatura corneal central y paracentral durante un periodo de 10 horas. Material y método: Se determinó la curvatura y espesor corneal en los ojos derechos de 10 varones jóvenes mediante Orbscan y videoqueratoscopio respectivamente. Ambos parámetros fueron determinados para la zona central y paracentral a 1 y 2 mm del centro de la córnea, cada dos horas durante un periodo de 10 horas. Resultados: La córnea presentaba su espesor máximo y curvatura mínima al abrir el ojo. Se observó una variación estadísticamente significativa de espesor (ANOVA, método de Schefféc p< 0,05) y de curvatura (ANOVA, método de Tamhane p < 0,05) a lo largo del día en todas las localizaciones estudiadas, si bien la variación y los rangos de valores en la zona paracentral eran mayores. Los valo- Received: Jan. 29, Accepted: Feb. 15, Santiago de Compostela University. Santiago de Compostela. Spain. 1 Graduate in Optics and Optometry. 2 Graduate in Medicine. Ophthalmologist. Minho University. Braga. Portugal. 3 Ph.D. in Pharmacy. Correspondence: Eva Yebra-Pimentel Vilar Escuela Universitaria de Óptica y Optometría Universidad de Santiago de Compostela. Campus Sur Santiago de Compostela Spain eyebra@usc.es

2 GIRÁLDEZ-FERNÁNDEZ MJ, et al. Ten young men took part in the study, having no history of contact lens use, ocular or systemic pathology or receiving at the time any type of topical or systemic medication. Women were not included in the study sample in order to avoid the effects of hormonal activity on the corneal thickness and curvature (15). After explaining the procedures to the patients, they were asked to sign the informed consent. Immediately upon awakening and with the purpose of eliminating the effect of tear evaporation on the size of the cornea (8), each subject covered his right eye with a patch up to the first measurement. The thickness and curvature values of the right eye cornea were determined with the Orbscan Topography System (Orbscan II version 3.0) and with the EyeSys videokeratoscope (EyeSys Corneal Analysis System version 3.1; EyeSys, INC, Houston, TX), respectively. Although previous studies have demonstrated that Orbscan yields higher values than the optical and ultrasonic pachymeters (16,17), its usefulness has been proved for measuring corneal thickness in investigation and in clinical practice with a precithickness was strongly correlated with corneal curvature, except for the 2 mm nasal and superior semi-meridians. Conclusions: These data indicate a shift in corneal thickness and curvature that can be of clinical relevance as individual changes vary greatly (Arch Soc Esp Oftalmol 2008; 83: ). Key words: Cornea, corneal thickness, corneal curvature, pachymetry, diurnal variation. res mínimos para ambos parámetros se observaron a las 10 y 8 horas después de abrir los ojos para las zonas central y paracentral respectivamente. Las variaciones de espesor y curvatura estaban altamente correlacionadas excepto para la localización nasal y superior a 2 mm del centro. Conclusiones: Estos datos evidencian un cambio en espesor y curvatura que pueden ser de relevancia clínica atendiendo a la variación individual encontrada. Palabras clave: Córnea, espesor corneal, curvatura corneal, paquimetría, variación diurna. INTRODUCTION The determination of corneal thickness has gained relevance in recent years, party due to the growing interest in the continued use of contact lenses (1,2) and refractive surgery (3-5). Corneal thickness provides valid information about its physiological condition and possible changes associated to diseases, traumas and hypoxia. The diurnal variations of corneal thickness are an obstacle for repeated pachymetric measurements. The pattern of this variation is that the cornea attains maximum thickness upon awakening after night sleep (6-8). Optical and ultrasound pachymetry have been utilized to measure corneal thickness in periods of hours (7-10). Utilizing the Haag-Streit optical pachymeter, Kiely et al (7) found a diurnal variation of 10 µm for the central thickness and 20 µm in the periphery at 40º. The corneal curvature has also been considered in relation to refractive surgery, the adaptation of contact lenses and others. Several authors have studied the daytime change of the central corneal curvature which show that the cornea is flatter upon awakening, with a slight but statistically significant increase in curvature throughout the day (6,10,11). More significant variations are observed in the superior and inferior hemi-meridians (12), in addition to gender differences (13). Kiely et al (7) studied the relationship between the curvature and corneal thickness and found a significant correlation between the diurnal thickness and curvature variations in the horizontal meridian. On the other hand, Rom et al (14) did not find a relationship between the variations of corneal thickness and topography. This study determined the central and para-central corneal thickness and curvature values (at 1 and 2 mm from the centre) during a 10-hour period, with measurements being taken at 2 hour intervals. The aim was to determine the daytime corneal thickness and curvature values in the central and para-central locations as well as to study the relationship between the thickness and curvature variations. SUBJECTS, MATERIAL AND METHOD 184 ARCH SOC ESP OFTALMOL 2008; 83:

3 Diurnal variation in corneal thickness and curvature sion in relation to gauge spheres of ±2 µm (range, 1.2 µm in the centre to 3.2 µm in the periphery) (10,16). Repeated measurements on individuals produced a mean standard deviation of 9.08 µm in the central 7 mm of the cornea (10). However, the repetitiveness and reproducibility of the Orbscan is low when measuring corneal curvature (18), for which reason the curvature values were determined by videokeratoscopy. Previous studies have shown a high repetitiveness in measurements taken with the videokeratoscope, with a standard deviation of ± 0.01 mm in humans and gauge spheres (19.20). The thickness and curvature measures were taken in nine corneal locations: central and at 1 and 2 mm from the centre in the nasal, temporal, superior and inferior half-meridians. The first measurements were taken immediately after withdrawing the eye patch and at 2 hour intervals for a 10-hour period. The statistical analysis was made utilizing SPSS 6.1 (SPSS Inc., Chicago, Illinois, USA). For comparing the daytime thickness and curvature variations in said locations an ANOVA test was applied with an Univariant General Linear Model. The corneal thickness and curvature were considered as dependent variables whereas the central location, at 1 and 2 mm from the centre, meridians and the measurement time were considered as factors. As Levene s test for equalizing variances was statistically significant for the thickness but not for the curvature, the Sheffec and Tamhane methods were utilized for thickness and curvature, respectively. The relationship between the thickness and curvature measurements were assessed by means of linear regression analysis. RESULTS Figure 1 and 2 show the corneal thickness and curvature measurements taken at the central and nasal, temporal, superior and inferior areas at 1 and 2 mm of the centre. The statistical analysis showed statistically significant corneal thickness variations in all the locations and period of the study (ANO- VA, Schfféc method, p < 0.05). The cornea exhibited its maximum thickness upon opening the eye after night sleep, which then reduced during the day in all the locations. The mean diurnal variation in central corneal thickness was of 14 µm (1.95%), with individual values in the range of 9-22 µm, and with the minimum value after 10 hours of said initial opening. The mean gradient of the pachymetric descent in the centre of the cornea was of The thickness variations corresponding to the nasal, temporal, superior and inferior at 1 and 2 mm of the centre are shown in tables I and II, the minimum value being approximately 8 hours after opening the eye. The mean gradient of the pachymetric des- Fig. 1: Mean value of central and nasal, temporal, superior and inferior corneal thickness immediately after opening the eye and at 2-hour intervals in a 10-hour period: a) at 1 mm from the centre; b) at 2 mm from the centre of the cornea. ARCH SOC ESP OFTALMOL 2008; 83:

4 GIRÁLDEZ-FERNÁNDEZ MJ, et al. Fig. 2: Mean value of central and nasal, temporal, superior and inferior corneal curvature immediately after opening the eye and at 2-hour intervals in a 10-hour period: a) at 1 mm from the centre; b) at 2 mm from the centre of the cornea. cent for 1 and 2 mm measurements in the meridians are shown in table III, where the values corresponding to the nasal, temporal, inferior and superior locations at 2 mm from the centre indicate a larger pachymetric reduction than in the centre of the cornea ( 0.686). In the case of the superior meridian, it can be seen that the mean gradient of the thickness Table I. Percentage of maximum corneal thickness diurnal variation at 1 and 2 mm of the cornea centre in the nasal, temporal, superior and inferior hemi-meridians Location Nasal Temporal Superior Inferior from the centre 1 mm (%) mm (%) Table II. Variations in corneal thickness at 1 and 2 mm of the cornea centre in the nasal, temporal, superior and inferior hemi-meridians during a 10-hour period after opening the eyes (mean values and range) Location Nasal Temporal Superior Inferior from the centre 1 mm (µm) 22 (7-32) 22 (13-31) 24 (12-27) 20 (8-26) 2 mm (µm) 19 (10-40) 19 (12-47) 23 (12-36) 16 (13-44) variation 1 mm from the centre is slightly smaller than the central one, and the value obtained at 2 mm, even though it is bigger, fits in better with the pachymetric variation in the centre of the cornea. The smallest curvature was observed with the first opening of the eye after sleep, with a statistically significant difference when compared with the values obtained throughout the day (ANOVA, method of Tamhane p < 0.05). After an initial increase of the corneal curvature, it remains virtually stable through the day in all the locations of the study. The mean variation of central curvature observed in the period of the study was of 0.07 mm, with the smallest value being at 2 hours after opening the eyes. Table IV shows the values of the para-central measurements of the study. No statistically significant differences were found between the thickness variations (ANOVA, Table III. Mean gradient of pachymetric descent in paracentral location at 1 and 2 mm of the cornea centre in the nasal, temporal, superior and inferior hemi-meridians Location Nasal Temporal Superior Inferior from the centre 1 mm (µm) mm (µm) ARCH SOC ESP OFTALMOL 2008; 83:

5 Diurnal variation in corneal thickness and curvature Schefféc method, p > 0.05) or curvature variations (ANOVA, Tamhane method p > 0.05) observed in the centre of the cornea and at 1 and 2 mm from the centre in all the hemi-meridians measured. Similarly, no statistically significant differences were found between the thickness variations (ANOVA, Schefféc method p > 0.05) or between the curvature variations (ANOVA, Tamhane method p > 0.05) which were measured in the hemi-meridians of the study. The nasal, temporal, superior and inferior thickness and curvature variations at 1 and 2 mm from the center are shown in figures 3 and 4 respectively. Table IV shows the results of the statistical analysis comparing the daytime variations in thickness and curvature in the corneal locations of the study. A high degree of correlation was observed between the variations in the center of the cornea (r= 0.983, r 2 = 0.967, p < 0.001) and in the other locations, excepting the nasal and superior areas at 2 mm from the center. DISCUSSION The purpose of this study is to contribute to the knowledge about the effect of time on the corneal thickness and curvature in the course of the day. By establishing said changes in the central and paracentral areas of the cornea, it would be possible to Table V. Correlation between the diurnal thickness and curvature variations in each location of the study r r 2 p Nasal 1 mm mm Temporal 1 mm mm Superior 1 mm mm Inferior 1 mm mm determine the time in which basal measurements should be taken. The relevance and implications of knowing the diurnal variations of corneal thickness are particularly important in the clinical evaluation of PRK patients and in candidates for other keratorefractive surgeries. In fact, the most common contraindication for LASIK surgery is reduced central corneal thickness, which is also an important factor in the characteristics of the flap made during the intervention (21,22). Corneal curvature has also been considered in relation to refractive surgery due to its relationship with the thickness and diameter of the flap (23). In this study, the maximum corneal thickness for all locations was measured immediately after opening the eyes in the morning. This finding matches those of other studies (5-10). The reduced oxygen supply during sleep leads to anaerobic metabolism which causes and increase in lactic acid production, in turn leading to an increase in osmotic pressure which facilitates the diffusion of water from the aqueous humor to the stroma. This increases the corneal thickness (24). When comparing the corneal thickness obtained during a 10-hour process, we observe a statistically significant variation in all the locations of the study. Even though the mean daytime variation of the observed central corneal thickness (14 µm) was similar to that found by other authors (7,10,25), individual variations could be clinically relevant (range of 9-22 µm). As can be seen in table III, the magnitude of thickness change and the range of values is larger in para-central data. This fact is also observed when comparing the mean gradient values of the central pachymetric reduction (-0.686) with the values of the nasal, temporal, inferior and superior paracentral data at 2 mm (table III). In the case of the superior meridian, it can be seen that the mean gradient of the thickness variation at 1 mm from the centre is slightly smaller than the central variation, and even when the value obtained at 2 mm is larger, it most closely matches the central value. This can be due to the fact that the superior cornea, both in eye clo- Table IV. Mean variation and value range of central and para-central curvature at 1 and 2 mm from the centre of the nasal, temporal, superior and inferior hemi-meridians (mm) Central Location Nasal Temporal Superior Inferior 0.07 ( ) 1 mm 0.05 ( ) 0.06 ( ) 0.07 ( ) 0.06 ( ) 2 mm 0.07 ( ) 0.07 ( ) 0.06 ( ) 0.07 ( ) ARCH SOC ESP OFTALMOL 2008; 83:

6 GIRÁLDEZ-FERNÁNDEZ MJ, et al. Fig. 3: Variation of corneal thickness at 1 and 2 mm of the centre of the nasal, temporal, superior and inferior hemimeridians. sed and open conditions, is the area with the largest hypoxic stress, which leads to a smaller thickness reduction in open eye conditions (26-28). The thickness increase in the periphery of the cornea and the greater thickness variation observed could imply changes in the regulating hydration capacity and in the specific control thereof, indicating the presence of a lateral hydration gradient extending outwardly from the apex (10). The hydration variations throughout the profile of the cornea could be a source of errors with some refractive surgery patients because the diurnal variation in pachymetry could be greater than the amount of tissue eliminated for each dioptre. In addition, many microkeratomes generate meniscus having different thicknesses, and therefore the corneal biomechanic 188 ARCH SOC ESP OFTALMOL 2008; 83:

7 Diurnal variation in corneal thickness and curvature Fig. 4: Variation of corneal curvature at 1 and 2 mm of the centre in the nasal, temporal, superior and inferior hemimeridians. stress could be compromised if in certain locations the residual stroma is very low. If the minimum corneal thickness value is not adequately determined, in theory the measures could yield a higher spurious value, leading to an overestimation which could increase the risk of developing postop kerectasia. The central and paracentral corneal thickness exhibited their lowest value 10 and 8 hours after ARCH SOC ESP OFTALMOL 2008; 83: opening the eyes. Previous studies have shown that the lowest corneal thickness values are obtained between 5 and 12 hours after opening the eye (7-9,25), in some cases obtaining a corneal thinning exceeding the base value. The differences between studies could be due to differences in the measurement intervals and frequencies. The thinning of the cornea beyond the base vale has already been stu- 189

8 GIRÁLDEZ-FERNÁNDEZ MJ, et al. died (29), and even though its presence is proven it is an aspect of corneal hydration dynamics for which no explanation has yet been found. Odenthal et al (29) suggest that the most likely possibility that the rate of the endothelial pump is not constant and that hypoxic stress may activate a functional reserve in the capacity of the ionic pump delaying its activity at a normal level. When this situation occurs, the corneal thickness could be influenced by other factors such as the blinking frequency, the quality of the lachrymal film and diurnal variation in the production of tears and intra-ocular pressure. This could explain the increased thickness occurring after the cornea reaches its minimum thickness between 8 and 10 hours after the nightly eye closure (29). In what concerns curvature, our results are consistent with those found by other researchers (7,11,13). Initially, in contrast with Kiely et al (7), we did not observe differences in the daily variations which occur in the horizontal and vertical meridians. Using the eyes in near focus and eyelid pressure could play a relevant role in the daytime changes of corneal curvature (12,30,31). For this reason, it is likely that the results of this study will differ from those found in other studied where the subjects used their eyes in near focus during the measurement period. In addition, in our study the effect of eyelids is not very probable when the measures are taken at 1 and 2 mm from the cornea centre. Read et al (12) suggest a correlation between the amount of work done in near focus and the curvature change values, with possible associated factors being corneal hydration, tear stability, blinking frequency, vertical eyelid aperture and the time elapsed between the near focus ocular activity and the measurement. Even though statistically significant differences between meridians were not observed, Figure 4 shows a curvature increase in the superior and inferior meridians 8 hours after opening the eyes. This finding matches agrees with other studies, which refer an increase in the curvature of the vertical meridian at the end of the day (11,12). It has been suggested that diurnal corneal curvature thickness variations in normal eyes could be related to changes in the cornea thickness (7,32). MacRae et al (32) found that in normal eyes and in those treated with radial keratotomy there was a relationship between the corneal thickness and changes in its shape. However, they observed that said correlation was not statistically significant when the sample was divided in 3 groups (8 patients per group). In our study, the statistical analysis shows a strong correlation between corneal thickness and curvature changes (table V). These results can be seen graphically in the similarity existing between figures 2 and 4, and match other studies which include normal and radial keratotomy corneas (32,33). On the other hand, Kwitko et al (34), did not observe said correlation considering it was due to the small number of subjects in their sample, which limited statistical analysis and the detection of statistical correlations. Surprisingly, with the same sample size of Kwitko et al (34), we obtained statistical significance with a similar tendency to that found by MacRae et al (32) in which, when the thickness of the cornea is reduced, the curvature is increased. This relationship between thickness and curvature could be due to the greater thickness reduction which takes place in the para-central area of the cornea when compared to the centre. In other words, said difference in the thickness reduction could be the cause of the observed curvature increase. REFERENCES 1. Brennan NA, Coles ML. Extended wear in perspective. Optom Vis Sci 1997; 74: Yebra-Pimentel E, Giraldez MJ, Gonzalez J, Perez MV, Parafita M. Changes in corneal thickness after daily and extended wear of hydrogel lenses: a comparison of optical and ultrasonic pachometry. Int Contact Lens Clin 1998; 25: Pallikaris IG, Kymionis GD, Astyrakakis NI. Corneal ectasia induced by laser in situ keratomileusis. J Cataract Refract Surg 2001; 27: Rao SN, Epstein RJ. Early onset ectasia following laser in situ keratomileusis: case report and literature review. J Refract Surg 2002; 18: Jonsson M, Behndig A. Pachymetric evaluation prior to laser in situ keratomileusis. J Cataract Refract Surg 2005; 31: Mertz GW. Overnight swelling of the living human cornea. J Am Optom Assoc 1980; 51: Kiely PM, Carney LG, Smith G. Diurnal variations of corneal topography and thickness. Am J Optom Physiol Opt 1982; 59: Holden BA, Mertz GW, McNally JJ. Corneal swelling response to contact lenses worn under extended wear conditions. Invest Ophthalmol Vis Sci 1983; 24: Harper CL, Boulton ME, Bennet D, Marcyniuk B, Jarvis- Evans JH, Tullos AB, et al. Diurnal variations in human corneal thickness. Br J Ophthalmol 1996; 80: Lattimore MR Jr, Kaupp S, Schallhorn S, Lewis R 4th. Orbscan pachymetry: implications of a repeated measures 190 ARCH SOC ESP OFTALMOL 2008; 83:

9 Diurnal variation in corneal thickness and curvature and diurnal variation analysis. Ophthalmology 1999; 106: Cronje S, Harris WF. Short-term keratometric variation in the human eye. Optom Vis Sci 1997; 74: Read SA, Collins MJ, Carney LG. The diurnal variation of corneal topography and aberrations. Cornea 2005; 24: Handa T, Mukuno K, Niida T, Uozato H, Tanaka S, Shimizu K. Diurnal variation of human corneal curvature in young adults. J Refract Surg 2002; 18: Rom ME, Keller WB, Meyer CJ, Meisler DM, Chern KC, Lowder CY, et al. Relationship between corneal edema and topography. CLAO J 1995; 21: Kiely PM, Carney LG, Smith G. Menstrual cycle variations of corneal topography and thickness. Am J Optom Physiol Opt 1983; 60: Marsich MW, Bullimore MA. The repeatability of corneal thickness measures. Cornea 2000; 19: Giraldez Fernandez MJ, Diaz Rey A, Cerviño A, Yebra- Pimentel E. A comparison of two pachymetric systems: silt-scanning and ultrasonic. CLAO J 2002; 28: Cho P, Lam AK, Mountford J, Ng L. The performance of four different corneal topographers on normal human corneas and its impact on orthokeratology lens fitting. Optom Vis Sci 2002; 79: Dave T, Ruston D, Fowler C. Evaluation of the EyeSys model II computerized videokeratoscope. Part I: Clinical assessment. Optom Vis Sci 1998; 75: Dave T, Ruston D, Fowler C. Evaluation of the EyeSys model II computerized videokeratoscope. Part II: The repeatability and accuracy in measuring convex aspheric surfaces. Optom Vis Sci 1998; 75: Miyata K, Tokunaga T, Nakahara M, Ohtani S, Nejima R, Kiuchi T, et al. Residual bed thickness and corneal forward shift after laser in situ keratomileusis. J Cataract Refract Surg 2004; 30: Ozdamar A, Kucuksumer Y, Devranoglu K. Flap dimensions created with the Hansatome microkeratome: correlation with Orbscan-derived anterior segment data. J Cataract Refract Surg 2005; 31: Spadea L, Cerrone L, Necozione S, Balestrazzi E. Flap measurements with the Hansatome microkeratome. J Refract Surg 2002; 18: Klyce SD. Stromal lactate accumulation can account for corenal oedema osmotically following wpithelial hypoxia in the rabbit. Journal of Physiology 1981; 321: du Toit R, Vega JA, Fonn D, Simpson T. Diurnal variation of corneal sensitivity and thickness. Cornea 2003; 22: Benjamin WJ, Rasmussen MA. Oxygen consumption of the superior cornea following eyelid closure. Acta Ophthalmol 1988; 66: Benjamin WJ, Ruben CM. Human corneal oxygen demands at superior, central and inferior sites. J Am Optom Assoc 1995; 66: Erickson P, Comstock TL, Zantos SG. Effects of hydrogel lens transmissibility profiles on local corneal swelling during eye closure. Optom Vis Sci 1996; 73: Odenthal MT, Nieuwendaal CP, Venema HW, Oosting J, Kok JHC, Kijlstra A. In vivo human corneal hydration control dynamics: a new model. Invest Ophthalmol Vis Sci 1999; 40: Buehren T, Collins MJ, Carney L. Corneal aberrations and reading. Optom Vis Sci 2003; 80: Lieberman DM, Grierson JW. The lids influence on corneal shape. Cornea 2000; 19: MacRae S, Rich L, Phillips D, Bedrossian R. Diurnal variation in vision after radial keratotomy. Am J Ophthalmol 1989; 107: Kwitko S, Garbus JJ, Hwang DG, Gauderman WJ, McDonnell PJ. Computer-assisted study of diurnal variation in corneal topography after penetrating keratoplasty. Ophthalmic Surg 1992; 23: Kwitko S, Gritz DC, Garbus JJ, Gauderman WJ, McDonnell PJ. Diurnal variation of corneal topography after radial keratotomy. Arch Ophthalmol 1992; 110: ARCH SOC ESP OFTALMOL 2008; 83:

Central corneal thickness (CCT) measurement is a critical

Central corneal thickness (CCT) measurement is a critical ARTICLE Topographic Paracentral Corneal Thickness With Pentacam and Orbscan: Effect of Acoustic Factor Javier González-Pérez, O.D., M.Sc., Ph.D., Jose M. González-Méijome, O.D., Ph.D., María T. Rodríguez

More information

Accuracy and Repeatability of a New Portable Ultrasound Pachymeter

Accuracy and Repeatability of a New Portable Ultrasound Pachymeter Accuracy and Repeatability of a New Portable Ultrasound Pachymeter Queirós A 1, González-Méijome JM 1, Fernandes P 1, Jorge J 1, Almeida J B 1, Parafita MA 2 1 Department of Physics (Optometry), School

More information

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis Han-Yin Sun 1, 2, 3, Hsiu-Wan Yang 4, I-Tsung Wu 4, Jung-Kai Tseng 2, 3 1, 5* and Shun-Fa

More information

THE RELATIONSHIPS BETWEEN OCULAR OPTICAL COMPONENTS AND IMPLICATIONS IN THE PROCESS OF EMMETROPIZATION

THE RELATIONSHIPS BETWEEN OCULAR OPTICAL COMPONENTS AND IMPLICATIONS IN THE PROCESS OF EMMETROPIZATION ARCH SOC ESP OFTALMOL 2008; 83: 307-316 ORIGINAL ARTICLE THE RELATIONSHIPS BETWEEN OCULAR OPTICAL COMPONENTS AND IMPLICATIONS IN THE PROCESS OF EMMETROPIZATION RELACIÓN ENTRE LOS COMPONENTES ÓPTICOS OCULARES

More information

ORIGINAL ARTICLE. Corneal Refractive Therapy with Different Lens Materials, Part 1: Corneal, Stromal, and Epithelial Thickness Changes

ORIGINAL ARTICLE. Corneal Refractive Therapy with Different Lens Materials, Part 1: Corneal, Stromal, and Epithelial Thickness Changes 1040-5488/07/8404-0343/0 VOL. 84, NO. 4, PP. 343 348 OPTOMETRY AND VISION SCIENCE Copyright 2007 American Academy of Optometry ORIGINAL ARTICLE Corneal Refractive Therapy with Different Lens Materials,

More information

Corneal Swelling Response to Contact Lenses Worn Under Extended Wear Conditions

Corneal Swelling Response to Contact Lenses Worn Under Extended Wear Conditions Corneal Swelling Response to Contact Lenses Worn Under Extended Wear Conditions Drien A. Holden, George W. Merrz, and John J. McNolly Corneal thickness changes were monitored in ten subjects (1) over a

More information

Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system

Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system 774 Ocular Surface and Tear Center, Bascom Palmer Eye Institute, University of Miami School of Medicine, Miami, USA Z Liu A J Huang S C Pflugfelder Zhongshan Ophthalmic Center, Sun Yat-sen University of

More information

Description of iatrogenic corneal ectasia in patients without traditional risk factors

Description of iatrogenic corneal ectasia in patients without traditional risk factors ARTICLE Description of iatrogenic corneal ectasia in patients without traditional risk factors Julio Ortega-Usobiaga, MD, PhD 1 ; Rosario Cobo-Soriano, MD, PhD 1 ; Fernando Llovet-Osuna, MD, PhD 1 ; Stephan

More information

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND,, PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including

More information

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29.

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29. NIH Public Access Author Manuscript Published in final edited form as: Cornea. 2013 December ; 32(12): 1544 1548. doi:10.1097/ico.0b013e3182a7f39d. Repeatability of corneal epithelial thickness measurements

More information

Summary Recommendations for Keratorefractive Laser Surgery June 2013

Summary Recommendations for Keratorefractive Laser Surgery June 2013 Summary Recommendations for Keratorefractive Laser Surgery June 2013 Background Laser assisted in-situ keratomileusis (LASIK) surgery is the most commonly performed keratorefractive surgery; altering the

More information

CLINICAL SCIENCES. Management of Post-LASIK Corneal Ectasia With Intacs Inserts

CLINICAL SCIENCES. Management of Post-LASIK Corneal Ectasia With Intacs Inserts Management of Post-LASIK Corneal Ectasia With Intacs Inserts One-Year Results CLINICAL SCIENCES George D. Kymionis, MD, PhD; Charalambos S. Siganos, MD, PhD; George Kounis, BSc; Nikolaos Astyrakakis, OD;

More information

Recent concerns regarding the depth of tissue ablation with

Recent concerns regarding the depth of tissue ablation with Volume Estimation of Excimer Laser Tissue Ablation for Correction of Spherical Myopia and Hyperopia Damien Gatinel, 1 Thanh Hoang-Xuan, 1 and Dimitri T. Azar 1,2 PURPOSE. To determine the theoretical volumes

More information

SMILE A Solution to Those who go in Harm s Way

SMILE A Solution to Those who go in Harm s Way SMILE A Solution to Those who go in Harm s Way Steven Schallhorn, M.D. Chief Medical Officer November 12, 2016 Operational Basis for Involvement Commander of Naval Special Warfare identified deficiencies

More information

Laser in situ keratomileusis (LASIK) has proven to be

Laser in situ keratomileusis (LASIK) has proven to be Autorefractometry after laser in situ keratomileusis Dimitrios S. Siganos, MD, PhD, Corina Popescu, MD, Nikolaos Bessis, DOpt, Georgios Papastergiou, MD Purpose: To correlate cycloplegic subjective refraction

More information

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation Mohammad Naser Hashemian, MD 1 Mahdi AliZadeh, MD 2 Hassan Hashemi, MD 1,3 Firoozeh Rahimi, MD 4 Abstract Purpose: To present

More information

Simultaneous Topography-guided Surface Ablation with Collagen Cross-linking for Keratoconus

Simultaneous Topography-guided Surface Ablation with Collagen Cross-linking for Keratoconus IJKECD Case series Simultaneous Topography-guided Surface Ablation with Collagen 10.5005/jp-journals-10025-1124 Cross-linking for Keratoconus Simultaneous Topography-guided Surface Ablation with Collagen

More information

The two currently accepted methods for correcting

The two currently accepted methods for correcting New Technique Therapeutic Alloplastic Laser in situ Keratomileusis for Myopia Arturo Maldonado-Bas, MD; Ruben Pulido-Garcia, MD ABSTRACT BACKGROUND: A new technique, therapeutic alloplastic laser in situ

More information

Corneal Hysteresis, Resistance Factor, Topography, and Pachymetry After Corneal Lamellar Flap

Corneal Hysteresis, Resistance Factor, Topography, and Pachymetry After Corneal Lamellar Flap Corneal Hysteresis, Resistance Factor, Topography, and Pachymetry After Corneal Lamellar Flap Damien Gatinel, MD, PhD; Slim Chaabouni, MD; Pierre-Alexandre Adam, MD; Jacques Munck, OD; Michel Puech, MD;

More information

Corneal Topography Pattern in Healthy Volunteers Coming to the Ophthalmology Department Hayatabad Medical Complex, Peshawar as Attendants

Corneal Topography Pattern in Healthy Volunteers Coming to the Ophthalmology Department Hayatabad Medical Complex, Peshawar as Attendants Original Article Corneal Topography Pattern in Healthy Volunteers Coming to the Ophthalmology Department Hayatabad Medical Complex, Peshawar as Attendants Naz Jehangir, Sofia Iqbal, Mushtaq Ahmad Pak J

More information

Cataract Surgery in the Patient with a History of LASIK or PRK

Cataract Surgery in the Patient with a History of LASIK or PRK Cataract Surgery in the Patient with a History of LASIK or PRK #56996-RS April 2018 Sebastian Lesniak, MD Matossian Eye Associates None Disclosures Bio Matossian Eye Associates, Hopewell NJ, 7/2015 Present

More information

Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter

Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter Abbas-Ali Yekta, PhD 1 Hassan Hashemi, MD 2,3 Mehdi KhabazKhoob, MSc 3 Asghar Dostdar, MSc 4 Shiva

More information

Prof.Paolo Vinciguerra, M.D. 1, 2 Riccardo Vinciguerra, M.D Humanitas University 1. Humanitas Clinical and Research Center IRCS 2

Prof.Paolo Vinciguerra, M.D. 1, 2 Riccardo Vinciguerra, M.D Humanitas University 1. Humanitas Clinical and Research Center IRCS 2 Prof.Paolo Vinciguerra, M.D. 1, 2 Riccardo Vinciguerra, M.D. 1-3 1 Humanitas University 1 Humanitas Clinical and Research Center IRCS 2 Columbus, Ohio State University 3 University of Insubria, Varese

More information

Retrospective Testing of the Score for the Detection of Ectasia Susceptibility: A Case Report of Ectasia 7 Years after LASIK

Retrospective Testing of the Score for the Detection of Ectasia Susceptibility: A Case Report of Ectasia 7 Years after LASIK IJKECD 10.5005/jp-journals-10025-1055 Case Retrospective Report Testing of the Score for the Detection of Ectasia Susceptibility: Case Report of Ectasia 7 Years after LSIK Retrospective Testing of the

More information

Scanning-slit topography (SST) analysis of corneal thickness

Scanning-slit topography (SST) analysis of corneal thickness Reliability of Noncontact Pachymetry after Laser In Situ Keratomileusis Miguel J. Maldonado, 1,2 Alberto López-Miguel, 1 Juan C. Nieto, 1 Juan Cano-Parra, 3 Begoña Calvo, 4,5 and Jorge L. Alió 6 PURPOSE.

More information

Anterior segment imaging

Anterior segment imaging CET CONTINUING Sponsored by 1 CET POINT Anterior segment imaging Sundeep Vaswani, BSc (Hons), MCOptom 39 The anterior segment of the eye encompasses all structures from the front surface of the cornea

More information

The early detection of keratoconus is essential

The early detection of keratoconus is essential REFRCTIVE SURGERY Identifying Keratoconus Suspects With Wavefront Preoperative wavefront analysis may help surgeons prevent erratic LSIK results and postoperative complications. Y EUGENIO M. CNDL, MD;

More information

Analysis of eye movements during myopic laser in situ keratomileusis

Analysis of eye movements during myopic laser in situ keratomileusis 15th International SCHWIND User Meeting, Vancouver 2014 Analysis of eye movements during myopic laser in situ keratomileusis Thomas Kohnen Department of Ophthalmology Goethe-University, Frankfurt, Germany

More information

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Jack T. Holladay, MD, MSEE, FACS; Warren E. Hill, MD, FACS; Andreas Steinmueller, MSc ABSTRACT PURPOSE:

More information

Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy PATIENTS AND METHODS. Patients

Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy PATIENTS AND METHODS. Patients Special Issue Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy Jack X. Ma, 1 Maolong Tang, 2 Li Wang, 3 Mitchell P. Weikert, 3 David Huang, 2 and Douglas D. Koch

More information

Personal data. Curriculum Vitae. Experience. Date of birth 18 December 1969

Personal data. Curriculum Vitae. Experience. Date of birth 18 December 1969 1 Curriculum Vitae Dr. Sophia I. Panagopoulou University of Crete Medical School, IVO 71003 Heraklion Greece Phone: +302810394645 FAX: +302810394653 e-mail: spanagop@med.uoc.gr Personal data Date of birth

More information

Inter-examiner agreement of the AS-OCT Visante corneal thickness

Inter-examiner agreement of the AS-OCT Visante corneal thickness Peer-reviewed Journal of the Editorial Changes in Editorial Board José Manuel González-Méijome Original articles Spanish General Council of Optometry Task oriented visual satisfaction and wearing success

More information

Intrastromal corneal ring

Intrastromal corneal ring Intrastromal corneal ring Kyriakidou Nantia M.D. Diathlasis Day Care Unit Scienti1ic Workshop of Diathlasis Day Care Unit 18-19 November, 2016 The Met Hotel Thessaloniki, Greece DAY CARE UNIT DIATHLASIS,

More information

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by Pentacam as A New Imaging Technique Mohammad Ali Zare, MD 1 Hassan

More information

LASIK has been the primary type of corneal refractive surgery

LASIK has been the primary type of corneal refractive surgery Cornea Higher-Order Aberrations of Anterior and Posterior Corneal Surfaces in Patients With Keratectasia After LASIK Naoyuki Maeda, 1 Tomoya Nakagawa, 1 Ryo Kosaki, 1 Shizuka Koh, 1 Makoto Saika, 2 Takashi

More information

Refractive Surgery Dilemma

Refractive Surgery Dilemma Refractive Surgery Dilemma Section Editor: lireza aradaran-rafii, MD CSE PRESENTTION 33-year-old man seeking refractive surgery presented with refractive error of -1.75-4.0 20 in the right and -0.75-2.5

More information

Contact lens extended wear (EW) creates a stressful environment

Contact lens extended wear (EW) creates a stressful environment Impact of Rigid Gas-Permeable Contact Lens Extended Wear on Corneal Epithelial Barrier Function Meng C. Lin, 1 Andrew D. Graham, 1 Robert E. Fusaro, 2 and Kenneth A. Polse 1 From the 1 School of Optometry

More information

A procedure for Present and Future

A procedure for Present and Future A procedure for Present and Future MOSTAFA M. SALAH MD PROFESSOR Head of Refractive Unit RESEARCH INSTITUTE OF OPHTHALMOGY CAIRO -EGYPT 1 THIN FLAP LASIK DEF. Creation of Intended regular thin flap less

More information

CATARACT SURGERY AFTER RADIAL KERATOTOMY

CATARACT SURGERY AFTER RADIAL KERATOTOMY AFTER RADIAL KERATOTOMY How to avoid disappointment. BY BERNARD MATHYS, MD CATARACT SURGERY Radial keratotomy (RK; Figure 1) was a popular refractive surgical procedure to correct myopia in the 1970s and

More information

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification Original Article Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification with Foldable IOL Irum Abbas, Atif Mansoor Ahmad, Tahir Mahmood Pak J Ophthalmol 2009, Vol. 25 No.

More information

Original Article High myopia as a risk factor for post-lasik ectasia: a case report

Original Article High myopia as a risk factor for post-lasik ectasia: a case report Original Article High myopia as a risk factor for post-lasik ectasia: a case report Mona Harissi-Dagher, MD, a,b Sonja A. F. Frimmel, c and Samir Melki, MD, PhD a,d Author affiliations: a Massachusetts

More information

BY MICHAEL W. BELIN, MD

BY MICHAEL W. BELIN, MD Point/Counterpoint: The Pentacam Ve The Pentacam s rotating imaging provides a more accurate picture of the cornea. BY MICHAEL W. BELIN, MD It was not long ago that discussions concerned whether topography

More information

Effect of refractive surgery on corneal properties and visual acuity

Effect of refractive surgery on corneal properties and visual acuity Quarterly of the Horizon of Medical Sciences Vol. 19, No. 1, Spr 2013 Pages: 11-15 Effect of refractive surgery on corneal properties and visual acuity Mahjoub M.* BSc, Momeni Moghadam H. 1 PhD, Rohani

More information

Effect of Refractive Surgery on Corneal Properties and Visual Acuity

Effect of Refractive Surgery on Corneal Properties and Visual Acuity 2013 DAZCo, Danesh Afarand Zistagen Co., IRAN ISSN: 1735-0344 Quarterly of the Horizon of Medical Sciences 2013; 19(1): 11-15 Effect of Refractive Surgery on Corneal Properties and Visual Acuity A R T

More information

Gas Stress Test for Assessment of Corneal Endothelial Function

Gas Stress Test for Assessment of Corneal Endothelial Function LABORATORY INVESTIGATIONS Gas Stress Test for Assessment of Corneal Endothelial Function Nobuyuki Ohguro,* Mamoru Matsuda,* Masakatsu Fukuda,* Shigeru Kinoshita and Yasuo Tano* *Department of Ophthalmology,

More information

Doctors of Optometry Course Notes

Doctors of Optometry Course Notes Doctors of Optometry Course Notes OD22 1 CE Contact Lens Management of the Irregular Cornea Monday, February 19, 2018 9:05 am 10:00 am Plaza A 2 nd Fl Presenter: Dr. Maria Walker Dr. Maria K. Walker earned

More information

9/25/2017 CASE. 67 years old On 2 topical meds since 3 years. Rx: +3.0 RE LE

9/25/2017 CASE. 67 years old On 2 topical meds since 3 years. Rx: +3.0 RE LE CASE 67 years old On 2 topical meds since 3 years Rx: +3.0 /-0.5@65 RE +2.5/-0.5@115 LE IOP : 17 RE 19 LE CD: 0.5 RE 0.6 LE 1 67 years old On 2 topical meds since 3 years Rx: +3.0 /-0.5@65 RE +2.5/-0.5@115

More information

Redalyc. Disponible en:

Redalyc. Disponible en: Redalyc Sistema de Información Científica Red de Revistas Científicas de América Latina, el Caribe, España y Portugal Núñez, María Ximena;Blanco, Claudia Efficacy of Orbscan II and Pentacam topographers

More information

PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET

PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET 616.365.5775 www.keillasik.com PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET Please read this entire booklet. Discuss its contents with your doctor so that questions are answered to your

More information

Nature and Science 2017;15(11) Mohamed Elmoddather. MD

Nature and Science 2017;15(11)   Mohamed Elmoddather. MD Outcome of PRK in Management of Post LISIK Residual Myopia and Myopic Astigmatism Mohamed Elmoddather. MD Ophthalmology Faculty of Medicine, Al-Azhar University, Assuit, Egypt shahdmsaleh@hotmail.com Abstract:

More information

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Learn Connect Succeed. JCAHPO Regional Meetings 2017 Learn Connect Succeed JCAHPO Regional Meetings 2017 Faculty Biometry and IOL Calculations ASCRS and ASOA Symposium and Congress Los Angeles, CA Daniel H. Chang, M.D. - Empire Eye and Laser Center Bakersfield,

More information

POST-LASIK ECTASIA MANAGEMENT

POST-LASIK ECTASIA MANAGEMENT POST-LASIK ECTASIA MANAGEMENT A. John Kanellopoulos MD 1,2 1: Laservision.gr Clinical & Research Eye Institute, Athens, Greece 2: NYU Medical School Department of Ophthalmology, NY, NY Financial interests:

More information

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page 7574-7579 Biomechanical Corneal Changes Post LASIK with Mechanical Microkeratome Flap versus Femtosecond Flap Department of Ophthalmology,

More information

Colleen Riley, O.D., M.S., F.A.A.O., Graeme Young, M.Phil., Ph.D., F.C.Optom., D.C.L.P., F.A.A.O., and Robin Chalmers, O.D., F.A.A.O.

Colleen Riley, O.D., M.S., F.A.A.O., Graeme Young, M.Phil., Ph.D., F.C.Optom., D.C.L.P., F.A.A.O., and Robin Chalmers, O.D., F.A.A.O. Eye & Contact Lens 32(6): 281 286, 2006 2006 Contact Lens Association of Ophthalmologists, Inc. Prevalence of Ocular Surface Symptoms, Signs, and Uncomfortable Hours of Wear in Contact Lens Wearers: The

More information

Abdel Rahman ElSebaey, MD, PhD.

Abdel Rahman ElSebaey, MD, PhD. Surface Ablation Refractive Surgery Abdel Rahman ElSebaey, MD, PhD. Menoufia University History Correction of optical defects of human eye started 1200 AD. Spherical error corrected by spectacle on 13

More information

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma CLINICAL SCIENCES Clinical Significance of Central Corneal Thickness in the Management of Glaucoma Carolyn Y. Shih, MD; Joshua S. Graff Zivin, PhD; Stephen L. Trokel, MD; James C. Tsai, MD Objective: To

More information

IntraLASIK Correction Of Nearsightedness, Farsightedness and Astigmatism Using IntraLase TM Technology

IntraLASIK Correction Of Nearsightedness, Farsightedness and Astigmatism Using IntraLase TM Technology IntraLASIK Correction Of Nearsightedness, Farsightedness and Astigmatism Using IntraLase TM Technology INDICATIONS AND PROCEDURE This information is being provided to you so that you can make an informed

More information

Why Cataract and Refractive Surgeons Need the Pentacam Four surgeons discuss the device s value, including new software additions.

Why Cataract and Refractive Surgeons Need the Pentacam Four surgeons discuss the device s value, including new software additions. Why Cataract and Refractive Surgeons Need the Pentacam Four surgeons discuss the device s value, including new software additions. Produced under an unrestricted educational grant from Oculus, Inc.. Extremely

More information

Cornea and Contact Lens Institute of Minnesota. Specialty Contact Lenses and Vision Management

Cornea and Contact Lens Institute of Minnesota. Specialty Contact Lenses and Vision Management Cornea and Contact Lens Institute of Minnesota Specialty Contact Lenses and Vision Management We focus on specialty contact lenses. is a leading national resource for specialized contact lenses and eye

More information

Gender and wavefront higher order aberrations: Do the genders see the world differently?

Gender and wavefront higher order aberrations: Do the genders see the world differently? Original article : Do the genders see the world differently? Reilly CD, Blair MA 59 MDW/MCST, Dept Ophthal, Wilford Hall Medical Center, Lackland AFB, TX, USA Abstract Introduction: Wavefront sensing technology

More information

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus 273 This is an Open Access article licensed under the terms of the Creative Commons Attribution- NonCommercial-NoDerivs 3.0 License (www.karger.com/oa-license), applicable to the online version of the

More information

FEP Medical Policy Manual

FEP Medical Policy Manual FEP Medical Manual 9.03.05 Corneal Topography/Computer-Assisted Corneal Topography/ Photokeratoscopy Last Review: September 2016 Next Review: September 2017 Related Policies 9.03.28 Corneal Collagen Cross-linking

More information

Fitting Keratoconus and Other Complicated Corneas

Fitting Keratoconus and Other Complicated Corneas Fitting Keratoconus and Other Complicated Corneas Christine W Sindt OD FAAO Professor, Clinical Ophthalmology Director, Contact Lens Service University of Iowa Disclosure Consultant: ALCON Vision Care

More information

Bilateral Keratectasia 34 Years after Corneal Transplant

Bilateral Keratectasia 34 Years after Corneal Transplant 24 Bilateral Keratectasia 34 Years after Corneal Transplant Xavier Valldeperas a, b Martina Angi b, c Vito Romano d Mario R. Romano b, e a Department of Ophthalmology, Hospital Universitari Germans Trias

More information

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1 European Journal of Ophthalmology / Vol. 19 no. 2, 2009 / pp. 180-187 Central corneal thickness measurements with optical coherence tomography and ultrasound pachymetry in healthy subjects and in patients

More information

Original Article Comparison of different femtosecond laser refractive surgeries on higher-order aberrations

Original Article Comparison of different femtosecond laser refractive surgeries on higher-order aberrations Int J Clin Exp Med 2016;9(11):22375-22380 www.ijcem.com /ISSN:1940-5901/IJCEM0034556 Original Article Comparison of different femtosecond laser refractive surgeries on higher-order aberrations Xinli Zhang

More information

Pardon the Objection: Cases Marc R. Bloomenstein OD, FAAO Scot Morris, OD Derek Cunningham, OD Kathy Mastrota, OD

Pardon the Objection: Cases Marc R. Bloomenstein OD, FAAO Scot Morris, OD Derek Cunningham, OD Kathy Mastrota, OD I. Introduction a. Treating the anterior segment different for each patient b. How can new technology help your patients? c. The anterior segment is going to be the hot bed for new technology II. Lipiflow

More information

Clinical Outcomes after Topography-based Corneal Laser Surgery with the. WaveLight Oculyzer and Topolyzer Platforms

Clinical Outcomes after Topography-based Corneal Laser Surgery with the. WaveLight Oculyzer and Topolyzer Platforms 1 Clinical Outcomes after Topography-based Corneal Laser Surgery with the WaveLight Oculyzer and Topolyzer Platforms By Arthur B. Cummings, M.D. 1 and Nadia Mascharka, MSc 2 Corresponding Author: Arthur

More information

Interpretation of corneal tomography

Interpretation of corneal tomography Interpretation of corneal tomography Presented by Chameen Samarawickrama - Westmead Hospital - Liverpool Hospital - University of Sydney - University of New South Wales The University of Sydney Page 1

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Pachymetry Page 1 of 8 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Pachymetry Professional Institutional Original Effective Date: March 11, 2004 Original Effective

More information

SynergEyes PS Practitioner Training

SynergEyes PS Practitioner Training SynergEyes PS Practitioner Training Post-Surgical Market For 30 years, visual scientists from around the world have struggled with the challenge of surgically correcting human refractive error. While great

More information

Ocular Jeopardy Marc R. Bloomenstein OD, FAAO

Ocular Jeopardy Marc R. Bloomenstein OD, FAAO I. Introduction a. Introduction of topics b. Various ocular related topics II. Lipiflow a. Meibomian gland dysfunction i. Define the disease ii. Prevalence iii. Current treatment options 1. Lid scrubs

More information

Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population

Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population Original article: Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population Dr Nikhil Mahajan*, Prof. Swati Tomar** **Professor,*Resident Department of Ophthalmology,

More information

CLINICAL STUDY. AC-M Wong, C-C Wong, NS-Y Yuen and S-P Hui

CLINICAL STUDY. AC-M Wong, C-C Wong, NS-Y Yuen and S-P Hui (2002) 16, 715 721 2002 Nature Publishing Group All rights reserved 0950-222X/02 $25.00 www.nature.com/eye Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical

More information

Informed IntraLASIK Consent

Informed IntraLASIK Consent 9201 Sunset Boulevard Suite 709 West Hollywood, CA 90069 Informed Intra Consent Phone 310. 275. 5533 Fax 310. 275. 5523 info@benjamineye.com www.benjamineye.com For the Correction of Nearsightedness, Farsightedness,

More information

QUALITY OF LIFE IN PATIENTS WITH CHOROIDAL MELANOMA

QUALITY OF LIFE IN PATIENTS WITH CHOROIDAL MELANOMA ARCH SOC ESP OFTALMOL 2008; 83: 301-306 ORIGINAL ARTICLE QUALITY OF LIFE IN PATIENTS WITH CHOROIDAL MELANOMA CALIDAD DE VIDA EN PACIENTES CON MELANOMA DE COROIDES BLANCO-RIVERA C 1, CAPEÁNS-TOMÉ C 2, OTERO-CEPEDA

More information

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes Javad Amoozadeh, MD 1 Nima Mirzaee Rad, MD 2 Amir Houshang Beheshtnejad, MD 3 Ahmad Kheirkhah, MD 1 Hassan Hashemi, MD 4,5

More information

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

Journal of Ophthalmology & Clinical Research

Journal of Ophthalmology & Clinical Research Research Article Histopathological Findings in Secondary Refractive Surgery Ectasias ISSN: 2573-9573 Journal of Ophthalmology & Clinical Research Zamora-Ortiz Rocío 1*, Rodríguez-Reyes Abelardo A 2, Hernández-Ayuso

More information

2/7/18. Disclosures: Laser K s: Keratectomy to Keratomileusis with a SMILE. Who Patients Are Listening to

2/7/18. Disclosures: Laser K s: Keratectomy to Keratomileusis with a SMILE. Who Patients Are Listening to Disclosures: Laser K s: Keratectomy to Keratomileusis with a SMILE Glaukos Equinox Alcon Mitch Ibach OD, FAAO Vance Thompson Vision Who Patients Are Listening to Optometrist 36% People who've had surgery

More information

aberration induced by laser

aberration induced by laser How is spherical aberration induced by laser refractive surgery? Geunyoung Yoon, PhD 1 Ian Cox, PhD 2 Scott MacRae,, MD 1 1 Department of Ophthalmology, Center for Visual Science University of Rochester,

More information

The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery

The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery 접수번호 :08-024 Korean Journal of Ophthalmology 2009;23:1-5 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.1.1 The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive

More information

In Practice. Surgical Procedures Diagnosis New Drugs

In Practice. Surgical Procedures Diagnosis New Drugs In Practice Surgical Procedures Diagnosis New Drugs 32 35 Bowman + Bulk = Better Results Mid-stromal lamellar keratoplasty (MSLK) offers a new approach to the management of advanced keratoconus that can

More information

Comparison of Corneal Thickness with Online Optical Coherence Pachymetry and Ultrasound Pachymeter

Comparison of Corneal Thickness with Online Optical Coherence Pachymetry and Ultrasound Pachymeter Comparison of Corneal Thickness with Online Optical Coherence Pachymetry and Ultrasound Pachymeter Rany E. Mitwally 1, MD, M. Tarek El-Naggar 1, MD, FRCS, Mohamed A. Marzouk 1 *, MD. 1 Research Institute

More information

Efficacy of Photorefractive Keratectomy for Military Pilot Recruitment in an Asian Air Force

Efficacy of Photorefractive Keratectomy for Military Pilot Recruitment in an Asian Air Force Efficacy of Photorefractive Keratectomy for Military Pilot Recruitment in an Asian Air Force Brian See, Gerard Nah, Wee Hoe Gan, Robin Low AsMA Annual Scientific Meeting 2013 Chicago, IL, USA Disclosure

More information

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy Original Article Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy Mahmoud Jabbarvand 1, MD; Farshad Askarizadeh 2, PhD; Mohamad Reza Sedaghat 3, MD Hadi Ghadimi

More information

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK?

LASIK. What is LASIK? Eye Words to Know. Who is a good candidate for LASIK? 2014 2015 What is? (laser in situ keratomileusis) is a type of refractive surgery. This kind of surgery uses a laser to treat vision problems caused by refractive errors. You have a refractive error when

More information

Precise measurements of corneal sublayer pachymetry are

Precise measurements of corneal sublayer pachymetry are CLINICAL SCIENCE Semiautomated SD-OCT Measurements of Corneal Sublayer Thickness in Normal and Post-SMILE Eyes Nikolaus Luft, MD,* Michael H. Ring, PhD, Martin Dirisamer, MD, Anna S. Mursch-Edlmayr, MD,*

More information

K eratoconus is a progressive, non-inflammatory condition

K eratoconus is a progressive, non-inflammatory condition 371 EXTENDED REPORT Late onset post-keratoplasty astigmatism in patients with keratoconus L Lim, K Pesudovs, M Goggin, D J Coster... Br J Ophthalmol 2004;88:371 376. doi: 10.1136/bjo.2003.027037 See end

More information

For approximately two decades photorefractive keratectomy. Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK.

For approximately two decades photorefractive keratectomy. Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK. Cornea Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK Anders Ivarsen and Jesper Hjortdal PURPOSE. To examine long-term changes in corneal power and aberrations in myopic patients

More information

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page 7580-7586 Outcome of PRK in Management of Post LISIK Residual Myopia and Myopic Astigmatism Department of Ophthalmology, Faculty

More information

The Visian ICL Advantages

The Visian ICL Advantages The Visian ICL Advantages Many vision correction procedures promise an improved level of vision, but few vision correction alternatives offer the quality and features found with the Visian ICL. These include:

More information

Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery

Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery The Egyptian Journal of Hospital Medicine (April 2018) Vol. 71 (3), Page 2636-2642 Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery Hager

More information

Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia

Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia Issued: September 2013 guidance.nice.org.uk/ipg466 NICE has accredited the process used

More information

US Trends in Refractive Surgery: The 2008 ISRS/AAO Survey

US Trends in Refractive Surgery: The 2008 ISRS/AAO Survey US Trends in Refractive Surgery: The 2008 ISRS/AAO Survey Richard J. Duffey, MD ** David Leaming, MD Refractive Subspecialty Day Atlanta - November 8, 2008 ** TLCV provided partial funding for this survey

More information

NIIOS. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece VASILIS S.

NIIOS. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece VASILIS S. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece Goldman applanation tonometry is still the gold standard for measuring IOP, but its accuracy is

More information

In Practice. Surgical Procedures Diagnosis New Drugs

In Practice. Surgical Procedures Diagnosis New Drugs In Practice Surgical Procedures Diagnosis New Drugs 62 65 A New Carpet Over Broken Tiles Facing an eye with multiple, deep radial keratotomy cuts? Arun Gulani describes how the right mindset and technique

More information

C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY

C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY The characteristics of corneal staining in successful daily and extended disposable contact lens wearers Isabelle Jalbert OD Deborah F Sweeney BOptom

More information

PRELIMINARY RESULTS IN TRANS EPITHELIAL CORNEAL CROSSLINKING

PRELIMINARY RESULTS IN TRANS EPITHELIAL CORNEAL CROSSLINKING PRELIMINARY RESULTS IN TRANS EPITHELIAL CORNEAL CROSSLINKING Authors: Diana Mihu, Adriana Stănilă, Mihaela Florescu, ValericaProştean Ophthalmology Clinic Sibiu Ocular Surface Research Center Sibiu, 2

More information

Central Corneal Thickness and its Association with Ocular Parameters

Central Corneal Thickness and its Association with Ocular Parameters Central Corneal Thickness and its Association with Ocular Parameters Authors: An-Fei Li 1,2 Mei-Ju Chen 1,2 Affiliations: 1 Department of Ophthalmology, Taipei Veterans General Hospital 2 National Yang-Ming

More information