1 Active Articulation E1 Dual Mobility Hip System Design Rationale Knees Hips Extremities Cement and Accessories PMI Trauma Technology
2 Active Articulation E1 Dual Mobility Hip System Dual Mobility: Not a New Concept Dual mobility was introduced in the mid-1970s by Professor G. Bousquet for use in total hip replacement with the goal of preventing chronic dislocations. 1 This goal was successfully achieved 1,2,3 by combining two fundamental principles: the first is the concept of low friction arthroplasty as described by J. Charnley, where by a small head articulating against a polyethylene liner has a low risk of failure due to low wear 4 ; the second is the McKee-Farrar concept, which recommended the use of large diameter bearings to minimize the risk of dislocation. 5 The secondary motion occurs between the E1 bearing and the acetabular cup when a larger range of motion is required (Figure 2). The bearing acts as a large femoral head articulating in the metal cup. Motion 2 Principle 1 (pertaining to the ID) The smaller the head the lower the wear rates. 4 (Charnley) + Principle 2 (pertaining to the OD) The larger the head the greater the joint stability. 5 (McKee-Farrar) = Dual Mobility Biomet has a wealth of experience successfully combining these two principles in dual mobility constructs. The clinically proven AVANTAGE hip system 6 has been a market leader in Europe for over fifteen years. The Active Articulation E1 Hip System merges Biomet s engineering experience in dual mobility with the benefits of Antioxidant Infused Bearing Technology. Large range of motion The dual mobility biomechanical concept is simple. The first motion occurs between the 28 mm femoral head and the inner concave surface of the E1 bearing, until the neck of the femoral stem comes into contact with the E1 bearing (Figure 1). The significant shell-to-neck ratio of the large diameter E1 bearings provide for excellent range of motion (ROM). The ROM for the smallest Active Articulation construct is 161 degrees. 7 Range of Motion (Degrees) Figure 2 Range of Motion Potential for Acetabular Constructs 7 161º 165º AP ROM Motion Max-Rom 36 mm Liner Max-Rom + 40 mm Liner Max-Rom + 44 mm Liner Active Articulation 44 mm Cup Active Articulation 66 mm Cup Figure 3 Figure 1
3 Reduced risk of dislocation The incidence of hip dislocation is gradually increasing, and is currently the fourth largest indication for hip revision at 16%. 8 Large diameter heads have been shown to improve joint stability and increase dislocation resistance. 9,10 The Active Articulation E1 hip bearing ranges from size 38 to 60 mm. Additionally, the retentive properties of the inner-diameter of the E1 bearing can greatly minimize the risk of femoral head dislocation from the E1 bearing. E1 Antioxidant Infused Technology Changing patient demographics show that higher levels of demand are placed on acetabular components related to increased patient activity and longevity. In order to address these evolving patient needs, E1 Antioxidant Infused Technology was developed to surpass the limitations of first generation highly crosslinked polyethylenes and offer an alternative bearing that provides ultra-low wear, maintains mechanical strength and demonstrates remarkable oxidative stability. 7 Revolutionary processing E1 bearings are made from the same isostatically compressed molded polyethylene barstock as Biomet s clinically proven ArCom polyethylene. 11 The bars are then gamma irradiated to induce a high level of crosslinking. Vitamin E is subsequently infused into the highly crosslinked polyethylene to neutralize residual free radicals present after irradiation. Some manufacturers have developed different manufacturing methods in the attempt to reduce the risk of oxidation after crosslinking. One method, remelting, consists of heating the material above its melting temperature. This process reduces the amount of free radicals, which increases the oxidation resistance of the material. 12 However, recent research has demonstrated in vivo oxidation of remelted highly crosslinked bearings. 13,14 Additionally, remelting reduces the strength of the material. 12 Another manufacturing method involves annealing the crosslinked material below the melting temperature. Although annealed materials maintain their strength, the process does not fully eliminate or neutralize all the free radicals which can lead to oxidation. 12 Oxidation in acetabular bearings has been shown to lead to increased wear, decreased strength, material degradation and fracture. 12,13,15,16 Oxidative Stability Cyclic loading, combined with the in vivo environment, may potentially induce cracks in polyethylene. This phenomenon is referred to as environmental stress cracking (ESC). Researchers now believe that oxidation of highly crosslinked UHMWPE may be a result of the absorption of lipids from the synovial fluid and/or cyclic loading, 17 similar to that seen in environmental stress crack testing. 7,18 E1 samples, along with sequentially crosslinked and annealed samples, were tested to determine their resistance to ESC by stressing the samples while subjecting them to adverse aging conditions that lead to oxidation. E1 Antioxidant Infused Technology protects polyethylene from oxidation and cracking during environmental stress crack testing. 19 Oxidation levels were detectable in the sequentially crosslinked and annealed samples and increased with loading. One study demonstrated that vitamin E antioxidant infused material...protect[ed] against lipid-initiated oxidation. 20 Oxidation Index (Absorbance Units) Top Surface Oxidation Profile 7 After ESC testing Center Sequentially Crosslinked and Annealed Loaded Sequentially Crosslinked and Annealed Control E1 Material Loaded E1 Material Control Figure 4 Bottom Surface 1
4 Active Articulation E1 Dual Mobility Hip System Ultra-low wear Laboratory testing was conducted on the Active Articulation E1 Hip System to test wear during suboptimal cup position. Cups were placed at 60 degrees of inclination and the system still yielded ultra-low wear rates after five million cycles Wear Rates for Acetabular Constructs 7 High strength Biomet has carried out extensive testing to compare the mechanical properties of E1 material with both ArCom and other highly crosslinked polyethylenes. The ultimate tensile and yield strengths of the E1 material are similar to those of ArCom polyethylene and greater than those of the remelted polyethylenes (Figure 6). 7,21,22 Tensile Strength Wear Rates 5 Million Cycles % wear reduction over 36 mm ArCom liners 7 62% wear reduction over 44 mm E1 liners 7 Tensile Strength (MPa) mm ArCom Liner (45º Inclination) 44 mm E1 Liner (45º Inclination) Figure 5 60 mm Active Articulation E1 Bearing (60º Inclination) 0 ArCom 7 Polyethylene E1 Unaged Material 7 E1 Aged Material 7 100kGy Remelted 21 50kGy Remelted 22 Yield Strength Yield Strength (MPa) ArCom 7 Polyethylene E1 Unaged Material 7 E1 Aged Material 7 100kGy Remelted 21 50kGy Remelted 22 Figure 6 2
5 Cup Design The Active Articulation Hip System utilizes the clinically proven M 2 a-magnum cup. 23 The M 2 a-magnum cup is an as-cast, high carbon cobalt chrome molybdenum (CoCrMo) alloy, press-fit design, with a titanium PPS Porous Plasma Spray outer surface and a highly polished inner geometry. The exterior geometry contains four pairs of peripheral fins to enhance rotational stability and aid initial fixation (Figure 7). The outer diameter of the cup is fully hemispherical (180 ), with four rim indentations for the stable attachment of the locking impaction device. The inner diameter is designed with up to 165 degrees of acetabular-head coverage to help maximize range-of-motion and minimize subluxation. The interior geometry is honed and buffed to a high polish to achieve a strict spherical tolerance of 200 micro-inches. Proven Fixation PPS Porous Plasma Spray Coating Biomet was the first orthopedic company to introduce a plasma-sprayed prosthesis with the release of the PPS coated Taperloc hip stem in The M 2 a-magnum cup features Biomet s PPS coating, a proprietary process that is instrumental to our clinical success. Biomet s process not only helps guard against osteolysis but allows both immediate and long-term fixation Biomet s proprietary plasma spray application is unique in that only the titanium powder used to create the coating is heated, while the implant s substrate is retained at near ambient temperatures. This unique process enables the implant to maintain its mechanical properties. Figure 7 In order to resist deformation forces, the cup is manufactured with a 6 mm thickness at the dome and an average of 3 mm thickness at the rim (Figure 8). This design is specific to the M 2 a-magnum component and allows for the maximum head-to-cup ratio. Figure 9: Titanium PPS Porous Plasma Spray being applied through a heated plasma arc. The heating effect of the PPS process is transient (lasting only for milliseconds). Therefore, the substrate material remains virtually unaffected, fatigue properties are maintained such that small femoral components are possible with this process and, importantly for M 2 a-magnum articulation, the carbide structure is unaffected. Figure 8 3
6 Active Articulation E1 Dual Mobility Hip System Biomet s PPS coating has irregularly shaped molten titanium particles that splatter upon impaction with the substrate surface. This generates a random distribution of pore size between 100 and 1,000 microns providing a larger contact area between particles and substrate (Figure 10). The larger distribution of pore size in conjunction with the enhanced biocompatibility of titanium, allows immediate fixation via mechanical interlocking and longterm fixation. Straightforward Instrumentation The Active Articulation Hip System utilizes an established and straightforward instrument platform. In fact, it is the same instrumentation used with Biomet s clinically proven, fully-hemispheric M 2 a-magnum cup 23 that has been in use since The system was designed to help facilitate and simplify acetabular preparation, cup and head trialing and cup positioning. The only additional instruments are a bearing press which secures the 28 mm head to the E1 bearing (Figure 11) and a head impactor that is specifically designed to impact the E1 bearing. This instrument platform can be used with all traditional and minimally invasive total hip arthroplasty approaches. See the Active Articulation Dual Mobility Hip System surgical technique for more information. Figure 10: The irregularly shaped titanium particles sprayed onto the substrate result in a wide pore size distribution. Biomet s PPS coating has been clinically proven for over 20 years as seen in a variety of published clinical papers Figure 11 4
7 References 1. Leclercq S, et al. Charnley-Kerboull-Bousquet hybrid THR after 10 years. Total Hip Arthroplasty; 3rd International Symposium, Aubriot P, Lesimple S. Study on Cementless Bousquet Type Cup in One Hundred Hybrid Total Hip Replacements (Femoral Stem Cemented Charnley Type) Average Follow-up 5 Years. Acta Orthopaedica Belgica. 59: Farizon F, de Lavison R et al. Results with a Cementless Alumina Coated Cup with Dual Mobility, a Twelve Years Follow up Study. Int Orthop. 22(4) : , Charnley, J. Long Term Results of Low Friction Arthroplasty of Hip as Primary Intervention, Journal of Bone and Joint Surgery (Br). 54: 61 76, McKee, G., Farrar, J. Replacement of Arthritic Hips by the McKee- Farrar Prosthesis. Journal of Bone and Joint Surgery. 48(2): , Tarasevicius, S. et al. Dual Mobility Cup Reduces Dislocation Rate After Arthroplasty for Femoral Neck Fracture. BMC Musculoskeletal Disorders.11: , Data on File at Biomet. Bench test results not necessarily indicative of clinical performance. 8. Joint Registry for England and Wales 6th Annual Report. P. 66, Beaule, et al. Jumbo Femoral Head for the Treatment of Recurrent Dislocation Following Total Hip Replacement. Journal of Bone and Joint Surgery. 84-A (2): , Burroughs BR, et al. Range of Motion and Stability in Total Hip Arthroplasty with 28, 32, 38, and 44mm Femoral Head Sizes: An in vitro Study. The Journal of Arthroplasty. 20: 11 19, Ritter MA. The Anatomical Graduated Component Total Knee Replacement-A Long Term Evaluation with 20 Year Survival Analysis. Journal of Bone and Joint Surgery (Br) , Kurtz, S. UHMWPE Biomaterials Handbook Second Edition, Currier, et al. In Vivo Oxidation in Remelted Highly Cross-linked Retrievals. JBJS (92) , Rowell, S. et al. Ex vivo stability loss of irradiated and melted UHMWPE. ORS Poster No Bhambri, S. et al. The Effect of Aging on Mechanical Properties of Melt-Annealed Highly Crosslinked UHMWPE. Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements , Wang, A. et al. Wear and Structural Fatigue Simulation of Crosslinked UHMWPE for Hip and Knee Bearing Applications. Crosslinked and Thermally Treated Ultra-High Molecular Weight Polyethylene for Joint Replacements , Multi-Center Study. Data on file at Biomet. 24. Keisu, K. et al. Primary Cementless Total Hip Arthroplasty in Octogenarians: Two to Eleven-Year Follow-Up. Journal of Bone and Joint Surgery. 83: 359, McLaughlin, J. et al. Total Hip Arthroplasty in Young Patients. 8- to 13- Year Results Using an Uncemented Stem. Clinical Orthopaedics and Related Research. 373: , Parvizi, J. et al. Prospective Matched-Pair Analysis of Hydroxyapatite- Coated and Uncoated Femoral Stems in Total Hip Arthroplasty. Journal of Bone and Joint Surgery. 83: 783 6, McLaughlin, J. et al. Total Hip Arthroplasty with an Uncemented Femoral Component. A Long Term Study of the Taperloc Stem. Journal of Arthroplasty. 19(2): 151 6, Meding, K. et al. Minimum Ten-Year Follow-up of a Straight- Stemmed, Plasma-Sprayed, Titanium-Alloy, Uncemented Femoral Component in Primary Total Hip Arthroplasty. Journal of Bone and Joint Surgery. 86: 92 7, McLaughlin, J. et al. Total Hip Arthroplasty with an Uncemented Femoral Component: Excellent Results at Ten Year Follow-Up. Journal of Bone and Joint Surgery. 79B: 900 7, Hozack, W. et al. Primary Cementless Hip Arthroplasty with a Titanium Plasma Sprayed Prosthesis. Clinical Orthopaedics and Related Research. 333: , Head, W. et al. A Titanium Cementless Calcar Replacement Prosthesis in Revision Surgery of the Femur: 13-Year Experience. Journal of Arthroplasty. 16(8): 183 7, Head, W. et al. The Proximal Porous Coating Alternative for Primary Arthroplasty. Orthopedics. 22(9): 813 5, Currier, B.H. et al. Evaluation of Oxidation and Fatigue Damage of Retrieved Crossfire Polyethylene Acetabular Cups. Journal of Bone and Joint Surgery. 89: , Besong, A.A. et al. A study of the combined effects of shelf ageing following irradiation in air and counterface roughness on the wear of UHMWPE. Biomed Mater Eng. 7: 59 65, Muratoglu, O., et al. Ex Vivo Stability Loss of Irradiated and Melted Ultra-High Molecular Weight Polyethylene. The Journal of Bone & Joint Surgery. 92: , Nabar, S. et al. Comparison of Second Generation Highly Crosslinked Polyethylenes Under Adverse Aging Conditions. ORS Poster No FDA Cleared Claim. See biomet.com/e1 for complete claim language. 20. Oral, E., et al. Protection of UHMWPE against lipid-initiated oxidation by antioxidants. 56th Annual Meeting of the Orthopaedic Research Society. Paper No. 234.
8 All trademarks herein are the property of Biomet, Inc. or its subsidiaries unless otherwise indicated. This material is intended for the sole use and benefit of the Biomet sales force and physicians. It is not to be redistributed, duplicated or disclosed without the express written consent of Biomet. For product information, including indications, contraindications, warnings, precautions and potential adverse effects, see the package insert and Biomet s website. P.O. Box 587, Warsaw, IN x Biomet Orthopedics biomet.com Form No. BOI REV043011
Arcos Modular Femoral Revision System
Arcos Modular Femoral Revision System Arcos System Simplify the Complex The Arcos Modular Femoral Revision System meets the demands of complex hip revision surgery by offering surgeons and OR staff the
More informationRegenerex Porous Titanium Construct. Knees Hips Extremities Cement and Accessories PMI Trauma Technology
Regenerex Porous Titanium Construct Knees Hips Extremities Cement and Accessories PMI Trauma Technology One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized
More informationComprehensive Reverse Shoulder System
Comprehensive Reverse Shoulder System Comprehensive Reverse Shoulder System Simple. Versatile. The Comprehensive Reverse Shoulder System is the next generation reverse shoulder prosthesis, offering unmatched
More informationcomprehensive Design Rationale shoulder system Knees Hips Extremities Cement and Accessories PMI TEchnology
comprehensive shoulder system Design Rationale Knees Hips Extremities Cement and Accessories PMI TEchnology . INTRODUCTION The Comprehensive Shoulder System is an evolutionary design based on the successful
More informationCement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter
Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished
More informationExeter. Designed for Anatomic Restoration. Clinical Evidence Education
Exeter Designed for Anatomic Restoration Clinical Evidence Education Anatomic Restoration Exeter re-creating biomechanics In hip replacement, the smallest adjustment can make a big difference. Stryker
More informationVanguard. Total Knee System
Vanguard Total Knee System Why remove healthy ligaments? DISCOVER NATURE AT WORK Bicruciate preserving total knee arthroplasty Thanks to a group of dedicated surgeons and designers, Biomet proudly introduces
More informationG7 ACETABULAR SYSTEM. beautifully. efficient.
G7 ACETABULAR SYSTEM beautifully efficient. HERITAGE MEETS MODERN DESIGN SIMPLICITY, EFFICIENCY AND PERFORMANCE The G7 System is a multi-bearing acetabular platform delivering simplicity, efficiency and
More information*smith&nephew POLARCUP Dual Mobility System. Product Information
Product Information *smith&nephew POLARCUP Dual Mobility System Dual Mobility 30 Years of Clinical Experience Concept combining an extended ROM, a low rate of dislocation and unexpectedly limited wear.
More informationESC. Enhanced Stability Liners. Design Rationale & Surgical Technique
ESC Enhanced Stability Liners Design Rationale & Surgical Technique Choice Without Compromise DePuy Synthes PINNACLE Hip Solutions are designed with a wide range of acetabular cup options, biological and
More informationEnhancing stability and increasing range of motion. Metasul LDH Large Diameter Head
Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that
More informationEnhanced Stability Constrained Liners. Design Rationale Surgical Technique
Enhanced Stability Constrained Liners Design Rationale Surgical Technique The Pinnacle Acetabular Cup System was designed to maximize the number of options available to the surgeon, and provide those options
More informationMARATHON Cross-Linked Polyethylene
MARATHON Cross-Linked Polyethylene Fixation First... 1 Wear Reduction... 2 Oxidative Stability... 3 Mechanical Integrity... 4 Proven Technology...5-6 References... 7 Fixation First FIXATION FIRST Prodigy
More informationReCap Product Rationale
ReCap Product Rationale A Complete resurfacing system designe engineering principles, tried and tested m clinically proven fixation methods Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales
More informationUHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD
UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 66th Annual Meeting February 4-9, 1999 Anaheim, California COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON
More informationOptimum implant geometry
Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available
More informationFIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY
FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY Fixation First FIXATION FIRST 1958 Charnley Hip Over 1,000,000 implanted 1 96.2 percent survivorship at 32 years
More informationOptimum implant geometry
Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available
More informationStability without compromise. Epsilon Durasul Constrained Insert
Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common
More informationACETABULAR CUP SURGICAL TECHNIQUE
ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP DEVICE INDICATIONS FOR USE The ICONACY I-Hip total hip replacement is indicated for the following conditions: 1. A severely painful and/or disabled hip
More informationThe bi-metric XR series of Hip implants blends biomet s proven philosophy and tradition with Advanced technology. porous primary hip series
The bi-metric XR series of Hip implants blends biomet s proven philosophy and tradition with Advanced technology porous primary hip series BI-METRIC XR SERIES implant technology The Bi-Metric XR Hip Series
More informationThe Leader in Orthopaedic Innovation
The Leader in Orthopaedic Innovation Wright is a leading international manufacturer and distributor of superior, easy to use, and innovative orthopaedic implants and instrumentation. For over 50 years,
More informationDePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)
References 1. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital Cost of Dislocation After Primary Total Hip Arthroplasty. J Bone Joint Surg. 2006;88A:290-294. 2. Brodner W, Grübl A, Jankovsky R, Meisinger
More informationInstrumentation Options
Instrumentation Options With the introduction of Microplasty and Premier Instrumentation platforms and advancements such as our patented Slidex Technology, the Vanguard Complete Knee System is the surgeon
More informationFeaturing. Technology. Product Rationale
Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge
More informationA Systematic Review on Performance of the Vanguard Complete Knee System
A Systematic Review on Performance of the Vanguard Complete Knee System Author: Jing Xie, Ph.D. June 30, 2011 Abstract A systematic review was conducted to evaluate the clinical performance of the Vanguard
More informationThe information contained in this document is intended for healthcare professionals only.
The information contained in this document is intended for healthcare professionals only. Trident Hemispherical Acetabular Shell Trident PSL HA Acetabular Shell Featuring: The Trident Acetabular System
More informationTOTAL HIP REPLACEMENT:
THR Prosthesis Design TOTAL HIP REPLACEMENT: PROSTHESIS DESIGN FEATURES JESS JOHNSTON & MELINDA ZIETH History of Hip Prosthesis Joint Replacement Registry Implant Design Technology & Future History and
More informationReaching new heights. Comprehensive. Efficient. Simple.
Reaching new heights Comprehensive. Efficient. Simple. Various acetabular cup choices Compatible with the different head and liner options including VERILAST Technology Reach for proven OR efficient Instrumentation
More informationBIOLOX delta Option Ceramic Femoral Head System
BIOLOX delta Option Ceramic Femoral Head System Product Features and Instructions for Use Knees Hips Extremities Cement and Accessories PMI Technology Product Features BIOLOX delta ceramic heads are composed
More informationCAUTION: Ceramic liners are not approved for use in the United States.
Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of
More informationSURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS
SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments
More informationVanguard Complete Knee System
Complete Knee System Design Rationale Knees Hips Extremities Cement and Accessories PMI Technology 1 One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized
More informationTHE PERSONALIZED KNEE
THE PERSONALIZED KNEE REDEFINING PERSONALIZATION Total knee replacement has long ranked among the most successful procedures in modern medicine. While you can expect excellent implant survivorship with
More informationThe information contained in this document is intended for healthcare professionals only.
The information contained in this document is intended for healthcare professionals only. Trident Poly Acetabular System Trident PSL HA Acetabular Shell with X3 Poly insert Trident Hemispherical Acetabular
More informationHOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX
HOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX HIP IMPLANTS TOTAL HIPS Ultra High Demand (metal on metal) Femur, ingrowth Femoral head, metal Acetabular shell, ingrowth Acetabular liner,
More informationAvenir Femoral Hip System
Avenir Femoral Hip System Avenir Hip System Intuitive and Reproducible During its decade of experience in the bone compaction stem market, the Avenir Hip System has helped surgeons experience flexible,
More informationOver 20 Years of Proven Clinical Success. Zimmer Natural-Knee II System
Over 20 Years of Proven Clinical Success Zimmer Natural-Knee II System CSTi Porous Coating Structurally similar to human bone CSTi porous coating combines the excellent biocompatibility of titanium with
More informationOptimum implant geometry
Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available
More informationZimmer TM. Continuum Acetabular System. Bringing together proven technologies.
Zimmer TM Continuum Acetabular System Bringing together proven technologies. The power to meet individual patient needs. Zimmer Continuum Acetabular System provides highly flexible solutions for orthopaedic
More informationProduct Information. *smith&nephew POLARCUP Dual Mobility System
APPLYING SCIENCE TO MOBILITY Combined with VERILAST Technology for unmatched performance Product Information *smith&nephew POLARCUP Dual Mobility System APPLYING SCIENCE TO MOBILITY Combined with VERILAST
More informationOxford. Partial Knee
Oxford Partial Knee Oxford Partial Knee A Definitive Implant The Oxford Partial Knee is the most widely used 38 and clinically proven 22 partial knee replacement (PKR) system in the world, offering industry
More informationRM Philosophy Innovation based on clinical evidence.
RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher
More informationActive Articulation. Surgical Technique. Dual Mobility Hip System. Knees Hips Extremities Cement and Accessories PMI Trauma Technology
Active Articulation Dual Mobility Hip System Surgical Technique Knees Hips Extremities Cement and Accessories PMI Trauma Technology Active Articulation Dual Mobility Hip System Contents Preoperative Planning,
More informationOSS Modular Arthrodesis System. Assembly Guide
OSS Modular Arthrodesis System Assembly Guide One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art of medical
More informationADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY
E C H E L O N P R I M A R Y H I P S Y S T E M P R O D U C T R A T I O N A L E ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY Echelon Primary Total Hip System HIGH OFFSET STANDARD OFFSET Cementless
More informationUnderstanding Hip Implant Options
Understanding Hip Implant Options Cup (Socket) Bearing (Liner) Head Stem (Femur) Modern Hip Implants (Ball) Metal Femoral Head on Marathon Cross-linked Polyethylene Liner Proven materials Polyethylene
More informationDUAL MOBILITY CUP, CEMENTED OR CEMENTLESS
HIP Documentation DUAL MOBILITY CUP, CEMENTED OR CEMENTLESS >>DUAL MOBILITY ACETABULAR CUPS Dual mobility acetabular cup P r Gilles Bousquet invented the concept of dual mobility more than thirty years
More informationWear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc
Mechanics of Contact and Lubrication, MTM G230 Department of Mechanical & Industrial Enineering Northeastern University Spring 2006 Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc
More informationBIOLOX delta Option Ceramic Femoral Head System. Product Features and Instructions for Use
BIOLOX delta Option Ceramic Femoral Head System Product Features and Instructions for Use Product Features BIOLOX delta ceramic heads are composed of 75% aluminum oxide and approximately 25% zirconia.
More informationCase report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?
Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).
More informationRECLAIM REVISION HIP SYSTEM
RECLAIM REVISION HIP SYSTEM Where Strength and Modularity Connect DESIGN RATIONALE System Overview RECLAIM Modular Revision Hip System WHERE STRENGTH & MODULARITY CONNECT 2 Offset Options Proximal Body
More informationPINNACLE REVISION CUP SYSTEM
PINNACLE REVISION CUP SYSTEM This publication is not intended for distribution in the USA. DESIGN RATIONALE FLEXIBILITY. PRECISION. CONFIDENCE. DePuy Synthes Joint Reconstruction is proud to offer the
More informationTemplating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.
Surgical Technique Contents Templating and Pre Operative Planning 2 Preparation of the Acetabulum 4 Trial Sizing and Impaction of the Shell 5 Cup Positioning 6 Joint Stability 7 Trial sizing and Impaction
More informationArComXL Polyethylene: A New Way to Process Highly Crosslinked Polyethylene
TM ArComXL Polyethylene: A New Way to Process Highly Crosslinked Polyethylene Several years ago, our competitors introduced highly crosslinked polyethylene (HXLP). Biomet chose not to follow that path.
More informationBone Preservation Stem
TRI-LOCK Bone Preservation Stem Featuring GRIPTION Coating Surgical Technique Implant Geometry Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was
More informationWHAT IS THE BEST BONE FIXATION TYPE? 2/11/2011
WHAT IS THE BEST BONE FIXATION TYPE? 2/11/2011 A Comparison of cement vs. bone ingrowth. Thomas P. Gross, M.D. At 2 years of follow-up cemented and uncemented femoral resurfacing is equivalent. Femoral
More informationTaperloc Complete Hip System. Surgical Technique
Taperloc Complete Hip System Surgical Technique One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art of medical
More informationPARTIALS?1 SHOULD BE AT LEAST 20% DID YOU KNOW OF KNEE REPLACEMENTS
PARTIAL KNEE DID YOU KNOW AT LEAST 20% OF KNEE REPLACEMENTS SHOULD BE PARTIALS?1 Research shows that surgeons utilizing Partial Knee Arthroplasty (PKA) for at least 20% of their annual knee arthroplasties
More informationDuraloc CONSTRAINED LINER
SURGICAL TECHNIQUE Duraloc CONSTRAINED LINER A COMPREHENSIVE ACETABULAR REVISION SYSTEM DURALOC CONSTRAINED LINER Introduction Dislocation is the most common postoperative complication in total hip reconstruction.
More informationDesign Rationale and Surgical Technique. *smith&nephew TANDEM. Bipolar and Unipolar Hip System
Design Rationale and Surgical Technique *smith&nephew TANDEM Bipolar and Unipolar Hip System Maximizing performance from every angle Saving time in the operating room is one of the most valuable solutions
More informationEncina Taper Stem. Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA
Stinson Orthopedics Inc. 303 Twin Dolphin Drive, Suite 600 Redwood City, CA 94065 info@stinsonortho.com www.stinsonortho.com Table of Contents Introduction 3 Features 4 Surgical Technique 5 Preoperative
More informationGender Solutions Patello-Femoral Joint System
Zimmer Biomet is the leading company in partial knee arthroplasty (PKA) 1 with over 40 years experience, offering a comprehensive range of anatomic and innovative solutions. Research shows that surgeons
More informationDVR Crosslock Distal Radius Plating System. Product Brochure
DVR Crosslock Distal Radius Plating System Product Brochure One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art
More informationStageOne. Shoulder Cement Spacer Molds. Surgical Technique
StageOne Shoulder Cement Spacer Molds Surgical Technique One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized care to one patient. The science and art
More informationBone Bangalore
Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,
More informationSURGICAL TECHNIQUE. Alpine Cementless Hip Stem
SURGICAL TECHNIQUE Alpine Cementless Hip Stem The following technique is a general guide for the instrumentation of the Alpine Cementless Hip Stem. It is expected that the surgeon is already familiar with
More informationCERAMIC-ON-METAL. References: Cat No: version 2
CERAMIC-ON-METAL References: 1. Beaule, P. E., et al. Jumbo femoral head for the treatment of recurrent dislocation following total hip replacement. J.Bone Joint Surg (2002). 84-A: 256-63. 2. Jolles, B.
More informationLEGION Total Knee System. Exceptional Durability
LEGION Total Knee System Exceptional Durability How many knee systems have been lab-tested to 30 years of simulated wear? Just one: LEGION Primary Knee System with VERILAST Technology The LEGION Primary
More informationCeramic Femoral Heads and Acetabular Cup Liners
Important Information: Please read prior to use in a clinical setting. The Surgeon should be familiar with the operative technique. Caution: Federal (U.S.A) law restricts this device to sale by or on the
More informationScan Bi-Polar 22/28. Operative Technique
Scan Bi-Polar 22/28 Operative Technique Disclaimer This publication and all content, artwork, photographs, names, logos and marks contained in it are protected by copyright, trademarks and other intellectual
More informationBi-Polar 22.2mm & 28mm System - Operative technique
Disclaimer Biomet UK Ltd, as the manufacturer of this device, does not practice medicine and does not recommend any particular surgical technique for use on a specific patient. The surgeon who performs
More informationSidus Stem-Free Shoulder. Secure fixation through a bone-sparing design
Sidus Stem-Free Shoulder Secure fixation through a bone-sparing design An efficient, bone-sparing solution, should revision become necessary Humeral shaft is untouched, minimal bone is removed from the
More informationThis publication is not intended for distribution in the USA. SURGICAL TECHNIQUE
This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE DePuy Synthes DURALOC Surgical Technique CONTENTS Templating and Pre Operative Planning 2 Preparation of the Acetabulum
More informationMicroSeal Total Hip Acetabular System. Like Nothing Else
MicroSeal Total Hip Acetabular System Like Nothing Else Like Nothing Else Unique Patented Sealing and Locking Design The MicroSeal design features: Locking/Sealing mechanisms work together to minimize
More informationPositioning Sleeve Surgical Technique
Positioning Sleeve Surgical Technique Surgical Technique The Comprehensive Fracture System offers a unique tool to evaluate and re-establish humeral height, which can be critical to restoring proper biomechanics
More informationDURALOC. Acetabular Cup System. Surgical Technique
DURALOC Acetabular Cup System Surgical Technique Table of Contents Surgical Technique Templating and Pre Operative Planning 2 Preparation of the Acetabulum 4 Trial Sizing and Impaction of the Shell 5
More informationAOX Antioxidant Polyethylene for SIGMA and LCS Rotating Platform Knee Systems
Value Brief AOX Antioxidant Polyethylene for SIGMA and LCS Rotating Platform Knee Systems Value Summary Advanced Technology: DePuy Orthopaedics is proud to offer AOX Antioxidant Polyethylene, the next
More informationManza Cup HA SURGICAL TECHNIQUE.
1 PRE-OPERATIVE PLANNING. Preoperative assessment of the appropriate size and position of the acetabular component will provide intraoperative guidance for acetabular reaming. To determine the acetabluar
More informationSystematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces
Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole
More informationTri-Lock Bone Preservation Stem
Tri-Lock Bone Preservation Stem Clinical Results of the Tri-lock BPS Femoral Stem at One Year Follow-Up William L. Healy, M.D. Lahey Clinic Medical Center Burlington, Massachusetts Introduction Cementless
More informationPyroCarbon Implants Durability + Biocompatibility
PyroCarbon Implants Durability + Biocompatibility What is PyroCarbon? PyroCarbon is a specific form of carbon that has been tailored for durability and biocompatibility. PyroCarbon exhibits favorable wear
More informationFLH183 04/08. Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom. Tel. +44 (0) Fax: +44 (0)
FLH183 04/08 Biomet UK Ltd Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom Tel. +44 (0)1656 655221 Fax: +44 (0)1656 645454 Exceed ABT Operative Technique The Exceed ABT TM acetabular
More informationI want that next-generation knee
I want that next-generation knee High flex Potentially longer-lasting More normal feeling Gender-optimized Less-invasive procedure Advanced design technology The last ten years have yielded an explosion
More informationDesign Rationale. ECHELON Primary Hip System
Design Rationale ECHELON Primary Hip System ECHELON Primary Total Hip System Addressing clinical issues of cementless hip arthroplasty Cementless total hip arthroplasty has provided a proven method of
More informationPseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008)
Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) 8% at 5yrs FU indicated by preliminary study from same
More informationAllofit /Allofit -S Alloclassic Acetabular Cup System
Allofit /Allofit -S Alloclassic Acetabular Cup System Product Information Convincing Primary Stability The Allofit acetabular cup was developed in 1993. The goal was to create a reliable implant that was
More informationFLH /11
FLH 225 04/11 This publication has been issued by: European Central Marketing Waterton Industrial Estate Bridgend, South Wales CF31 3XA, United Kingdom Tel: +44 (0)1656 655221 Fax: +44 (0)1656 645454 www.biomet.com
More informationApproach Patients with CONFIDENCE
Design Rationale Approach Patients with CONFIDENCE The ACTIS Total Hip System is the first DePuy Synthes stem specifically designed to be utilized with tissue sparing approaches, such as the anterior
More informationACTIS. TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE
ACTIS TOTAL HIP SYSTEM Approach Active Patients with Confidence DESIGN RATIONALE APPROACH ACTIVE PATIENTS WITH CONFIDENCE The ACTIS Total Hip System is the first DePuy Synthes Companies of Johnson & Johnson
More informationSection of Modular Hip Prostheses cemented. TMC-3 Modular Hip Prosthesis, cemented. TMC-3 Modular Hüftprothese, zementiert
Section of Modular Hip Prostheses cemented TMC-3 Modular Hip Prosthesis, cemented TMC-3 Modular Hüftprothese, zementiert Prothèse de hanche modulaire TMC-3, cimentée Indication : The TMC-3 Modular hip
More informationPLR. Proximal Loading Revision Hip System
PLR Proximal Loading Revision Hip System The PLR splined revision stem is designed to recreate the natural stresses in the revised femur, where proximal bone may be compromised. PLR Hip System Design Considerations
More informationMULTIPOLAR BIPOLAR CUP. Surgical Technique
MULTIPOLAR BIPOLAR CUP Surgical Technique MULTIPOLAR BIPOLAR CUP ASSEMBLY 1PREASSEMBLY (BACK TABLE) TECHNIQUE INSERT FEMORAL HEAD Seat the femoral head securely in the polyethylene liner. The inner split
More informationD E S I G N R A T I O N A L E
D E S I G N R A T I O N A L E FREEDOM TO CHOOSE 1 PROVEN MODULAR SOLUTIONS The Pinnacle Acetabular Cup System is uniquely designed with a range of acetabular cup options, biological and mechanical fixation
More informationDual Articular.
Dual Articular 2000 www.biomet.co.uk D e s i g n R a t i o n a l e The Dual Articular Knee has been in use since 1991 and offers a flexible reconstructive solution for total knee surgery; for both revision
More informationThe proven, simple solution. CPT 12/14 Hip System
The proven, simple solution CPT 12/14 Hip System Primary CPT Hip The proven, simple solution 5mm The collarless, polished, doubletaper design concept used in the CPT 12/14 Hip System has proven itself
More information*smith&nephew SL-PLUS Cementless Femoral Hip System. Product Information
Product Information *smith&nephew SL-PLUS Cementless Femoral Hip System First Came the Philosophy to develop a universal hip system that could be used in almost every indication, immaterial to the patient
More informationREFLECTION Constrained Liner Introduction
Surgical Technique REFLECTION Constrained Liner Introduction Surgical treatment of recurrent dislocation should be directed toward correction of any mechanical abnormalities. If dislocation is caused
More informationCase Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases
SAGE-Hindawi Access to Research Advances in Orthopedics Volume 2011, Article ID 519254, 4 pages doi:10.4061/2011/519254 Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs:
More informationGreater Component Accuracy...
Greater Component Accuracy... StageOne cement spacer molds are designed to provide the surgeon with choices never before offered in two-stage revision knee surgery for an infected total joint. StageOne
More information