Comparison of interfragmentary compression in conventional and locked plating of proximal unicondylar tibia fractures : A biomechanical cadaver study

Size: px
Start display at page:

Download "Comparison of interfragmentary compression in conventional and locked plating of proximal unicondylar tibia fractures : A biomechanical cadaver study"

Transcription

1 586 Acta Orthop. s. rahm, Belg., l. 2016, ebneter, 82, b. ellenrieder, c. spross, f. kulling, j. b. erhardt ORIGINAL STUDY Comparison of interfragmentary compression in conventional and locked plating of proximal unicondylar tibia fractures : A biomechanical cadaver study Stefan Rahm, Lukas Ebneter, Birte Schultz, Christian Spross, Fabrice Kulling, Johannes B. Erhardt From the Universitätsklinik Balgrist Zürich, Swiss The extent of interfragmentary compression in intraarticular fractures treated with various fixation methods have not yet been reported. Lateral split fractures were created in six pairs of cadaver tibiae treated using buttress plating with lag screws (group C) or locked buttress plating after clamp compression (group L). Interfragmentary compression and fracture displacement were continuously measured using pressure sensors and a stereoscopic 3-D image correlation system. Significantly larger interfragmentary compression was found initially after clamping the fragment (p < 0.05) in group C (median ± SD ; 45.1 ± 5.0 N/mm 2 ) compared with group L (33.6 ± 3.4 N/mm 2 ), and a statistical trend towards larger compression was also found after cyclic loading (p = 0.05) in group C (45.3 ± 8.6 N/mm 2 ) compared with group L (28.7 ± 5.8 N/mm 2 ). These data indicate that conventional plating with lag screws achieves higher interfragmentary compression in this model compared with external clamp compression and locked plating. Keywords : interfragmentary compression ; tibia plateau fracture ; conventional plating ; locked plating. INTRODUCTION The biomechanics of fracture treatment and healing is of ongoing interest to researchers, surgeons, and patients. A variety of different techniques for fracture treatment can all lead to bone healing. Currently, the two major operative fracture treatments are based on relative and absolute stability, concepts that were introduced by the Swiss study group Arbeitsgemeinschaft für Osteosynthesefragen, now known as the AO Foundation (11). When treating an intra-articular fracture, the concept of absolute stability indicates that the fractured fragments should be anatomically reduced and stabilized to be as rigid as possible. To achieve this, locking plates seem to be advantageous, particularly for use in osteoporotic bone. For simple shaft fractures with good bone quality, however, no advantage in the use of locking plating has been found in biomechanical testing or clinical outcome measures (2,3,6,13). The key for treating osteoporotic intra-articular fractures is to fix the fragments that recreate the articular surface anatomically, and two basic principles are mainly n Stefan Rahm. n Lukas Ebneter. n Birte Ellenrieder. n Christian Spross. n Fabrice Kulling. n Johannes B. Erhardt. Dept. Orthopädie, Universitätsklinik Balgrist Zürich, Swiss. Correspondence : Stefan Rahm, Universitätsklinik Balgrist, Forchstr. 340, 8008 Zürich. orthopaedie@balgrist.ch ch 2016, Acta Orthopædica Belgica. No benefits or funds were received in support of this study. The Acta authors Orthopædica report Belgica, no conflict Vol. of 82 interests rahm-.indd 586 7/10/16 09:11

2 proximal unicondylar tibia fractures 587 used to achieve this. The first is the use of lag screws to achieve maximum interfragmentary compression in combination with a conventional plate, and the second, especially in combination with anatomically pre-shaped locking plates, is the use of a clamp to achieve interfragmentary compression before fixing the fragments to the plate with locking screws. The effect of interfragmentary compression on fracture displacement, as well as the relative amount of interfragmentary forces, has not yet been reported. Furthermore, there is no data available in the literature showing the differences during cyclic loading in interfragmentary compression in various fixation methods. Schatzker I and AO type B1.1 lateral tibia plateau fractures are common and, if displaced more than 2 mm, are usually treated operatively (8,10,12). Although these types of fracture can be treated with a conventional buttress plate, anatomically shaped locking plates have recently become more popular because of their availability, promised increased stability and ease of use (7). Using simple lateral tibia plateau fractures as a clinically relevant test model, the aim of this study was to determine whether there are biomechanical differences in interfragmentary compression and fracture micromotion between traditional buttress plating and the use of a locking plates. The present biomechanical study was designed to show whether there is a difference in the magnitude of interfragmentary compression at different stages of fixation and after cyclic loading between those two fixation techniques. The hypotheses of this study were 1) the use of lag screws in combination with a buttress plate achieves larger magnitudes of interfragmentary compression than clamp fixation and locked buttress plating and 2) the interfragmentary compression will remain higher over time with less interfragmentary micromotion after locked plating compared with the conventional fixation technique. MATERIAL AND METHODS Six matched pairs (6 right and 6 left) of fresh frozen cadaver tibiae were used in this study. The cadaver tibiae were obtained from the anatomical institute of the University of Geneva, where they were managed strictly in Fig. 1. Example of a fixed tibia with a lateral NCB plate showing the four screws and the pressure sensors. line with all regulations as required by law. The local IRB approved this study. At the time of testing, the cadaver tissues were at room temperature. The tibiae were numbered from 1 to 6 with R indicating the right and L indicating the left of the matched pair. The locking group L included the tibia L-1R, L-2L, L-3R, L-4L, L-5R and L-6L and the other 6 tibiae (C-1L, C-2R, C-3L, C-4R, C-5L and C-6R) were assigned to the conventional group C. This design produced two groups with matched pairs that were numbered alternatingly and prepared to avoid the effects of potential variations in the bone mineral density of the tibiae. The tissue setup is shown in Figure 1. rahm-.indd 587 7/10/16 09:11

3 588 s. rahm, l. ebneter, b. ellenrieder, c. spross, f. kulling, j. b. erhardt Implant The implant was a polyaxial nine-hole locking plate for the proximal tibia with three proximal holes (Non- Contact Bridging Plate, NCB ; Zimmer, Warsaw, IN, USA) in all tests. This device was chosen because of the option to use it in either locking or conventional mode. The locking option is activated secondarily with a locking cap. Testing setup Each tibia was distally fixed into the loading device. The NCB plate was secured on the lateral tibia at the desired position for optimal testing. Group conventional (C) Two standard 3.3-mm holes were drilled using a uniaxial drill sleeve, and bicortical 4-mm conventional cortical screws were inserted into the two distal holes (4 and 6) of the plate. Next, the lag screw holes were drilled in two of the three proximal holes, predrilling 5 mm into the lateral fragment and 2.5 mm into the medial fragment, and the depths were verified to be bicortical. Then, the distal screws were loosened and the plate was lifted away approximately 1-2 cm. At this point, the osteotomy was performed using an oscillating saw to simulate a standardized lateral tibial plateau split fracture (Schatzker I), with the goal of passing through 50% of the lateral tibial plateau. The osteotomy gap was opened and three pressure sensors were placed precisely and in the same way in all of the specimens. The fracture was then preliminarily held with a K-wire. Two markers were placed on the medial and two more on the lateral fragments to measure the relative movements of the bone fragments. The distally loosened screws were retightened, but only enough to hold the plate in the desired position. The two measurement systems were calibrated, and then the testing phase was started (Time 0). Group locking plate (L) Standard 3.3-mm holes were drilled using the uniaxial drill sleeve into the two distal spaces (holes 4 and 6 as in group C), and into two of the three proximal holes using the polyaxial drill sleeve. The depths of the holes were measured. At this stage, the osteotomy was performed just as described above for group C, with the osteotomy exactly perpendicular to the drill holes. The three pressure sensors were inserted in the same way as in group C, and the fragments were held together with a K-wire. Next, two markers, as in group C, were placed onto each of the fragments. The two distal screws were tightened enough so that the K-wire could be removed without the fragments coming apart. Then, the calibration was performed and the testing phase was started (Time 0). Testing Phase The first procedure in both groups was to test the buttress effect, and the distal two screws were tightened after removing the K-wire. The next step was to create interfragmentary compression, which was done in group C by tightening the predrilled lag screws and in group L by closing a large reduction clamp. The results of recorded interfragmentary compression and relative fragment micromotions are listed in Table I. Cyclic loading Each specimen was secured in a Zwick universal testing machine (Zwick/Roell, Ulm, Germany). To avoid friction, the load was applied by a ball-and-socket-joint directly perpendicular to the lateral fragment as shown in Figure 1. Each specimen was axially preloaded with 30 N. Then, 1000 cycles of 200 N were applied, similar to the testing protocol used by Blakey et al (1). Compression and Micromotion To measure the interfragmentary compression, three pressure sensors (sensing area : 10 mm 2, SENSOR- TECHNICS, GmbH, Puchheim, Germany) were positioned in each osteotomy gap as described above. One sensor was positioned posterior, one anterior, and one in the middle. The readings from the three pressure sensors were summed and the mean is reported. The resulting voltage values were converted to compressive stress (N/mm 2 ) after calibration. The movement between the bone fragments (the main tibia fragment and the lateral tibia plateau fragment) was measured in 3 dimensions using a stereoscopic, noncontact, full-field digital image correlation system during dynamic loading (Vic-3D System, LIMESS, Correlated Solutions, Columbia, SC, USA). The marker was tracked using pattern recognition performed for the marker placed on each bone fragment. The field of view was mm with an accuracy of ~1 µm in the x- and y-planes and ~2 µm in the z-plane. The sampling rahm-.indd 588 7/10/16 09:11

4 proximal unicondylar tibia fractures 589 Table I. The times are listed from 1 to 4 indicating when the readings for the interfragmentary compression were taken for every tibia tested Measuring the interfragmentary compression : Group C Group L Time 1 : After tightening the two distal bicortical screws After tightening the two distal bicortical screws Time 2 : After tightening the two proximal bicortical lag screws After closing the interfragmentary clamp with reasonable power Time 2a After tightening the two proximal bicortical position screws Time 2b After removing the interfragmentary clamp Time 3 : After leaving the system rest for 60 minutes After leaving the system rest for 60 minutes Time 4 : After cyclic loading After cyclic loading The main finding in this study was that both systems assessed resulted in similar maximum interfragmentary compression magnitudes. However, after relieving the clamp in the locking plate group L, the compression was significantly decreased. Therefore, our first hypothesis was confirmed. The lag screw group C was able to maintain the interfragmentary compression at a higher level over time. Even after cyclic loading, it dropped only 3% of the maximum achieved interfragmentary comfrequency was 10 Hz (4). The 3-dimensional distance and displacment in µm between the two markers was measured during the entire test setup. The displacement between the lateral fragment and the medial main tibia fragment was measured and is given as E in units of μm. Statistics A 2-tailed non-parametric Mann-Whitney U-test and a 2-tailed Pearson Correlation were used to compare the micromotions and compression values between the groups. P-values less than 0.05 were considered statistically significant. RESULTS The results from the first three tibiae (C-4R, L-5R and L-6R) had to be discarded because of a technical error caused by interface failure between the compression measurement, digital image correlation and material testing machine setup. In group C there was a statistical trend towards higher interfragmentary compression compared to group L at all times except at times 1 and 2a, which were immediately after tightening the two proximal bicortical position screws in group L. After removing the interfragmentary clamp from group L, the compression dropped back to levels that were significantly lower than those in group C. Taking the maximum achieved interfragmentary compression to be 100%, the compression after cyclic loading in group L was 15% lower, reaching 85%, but only 3% lower in group C, reaching 97%. The buttress effect alone provided 12% compression in group L and 67% in group C. All of the results are listed in Table II and depicted visually in Figure 2. Micromotion The amount of translational movement of the lateral fragment in group C was not significatnly different compared to group L (refer to Table III). Correlations The 3-dimensional displacements in μm and the mean interfragmentary compression were significantly correlated with a good to very good correlation in group C (p < 0.001, r = 0.686) and a fair to good correlation in group L (p < 0.001, r = 0.433). DISCUSSION rahm-.indd 589 7/10/16 09:11

5 590 s. rahm, l. ebneter, b. ellenrieder, c. spross, f. kulling, j. b. erhardt Table II. The numerical values of the interfragmentary compression are reported in N/mm 2 for each tibia and the statistical significance calculated using a Mann-Whitney U rank test are shown Summary of interfragmentary compression : (N/mm 2 ) Time 1 Time 2 Time 2a Time 2b Time 3 Time 4 Group L L-2L L-1R L-4L L-3L Median Max Min SD Group C C-2R C-1L C-5L C-3R C-6L Median Max Min SD *p = * Mann Whitney-U Test. pression compared with a loss of 15% in group L. Thus, our second hypothesis was rejected. The absolute magnitudes of the interfragmentary compression after the buttress effect at time 1 were to our surprise different between the two groups, but this difference was not statistically significant (Fig. 3). It may be explained by the differences between the anatomic shape of the tibiae and the precontoured NCB plate, which was not placed exactly at the same level on each tibia even though that was our aim. Another point was the relative increase when tightening the locking screw in group L at time 2a, which may be explained by the lag screw effect even though the holes were correctly drilled. Obviously, after the clamp was removed this effect lasted no longer and the compression decreased almost 50% from time 2 to 2a. To our knowledge this is the first study to investigate the magnitude and Fig. 2. Summary of the interfragmentary compression magnitudes at different times. The x-axis shows the different times when a reading was taken. The times from 1 to 4 are explained in Table I. The y-axis shows the median interfragmentary compression in N/mm 2 for each group. The numbers are reported in Table II as the median, Max/Min and SD. rahm-.indd 590 7/10/16 09:11

6 proximal unicondylar tibia fractures 591 Table III. The numerical values of the displacement of the lateral fragment are reported in μm for each tibia and the statistical significance calculated using a Mann-Whitney U rank test are shown, but no significant differences between the groups were found at any time Summary of displacement of the lateral fragment E (μm) Time 1 Time 2 Time 2a Time 2b Time 3 Time 4 Group L L-2L L-1R L-4L L-3L Median Max Min SD Group C C-2R C-1L C-5L C-3R C-6L Median Max Min SD p* = - n.s. n.s. n.s. n.s. n.s. * Mann-Whitney-U Test. mode of interfragmentary compression in a simple fracture model using cadaveric tibiae under cyclic loading. There are a number of limitations to this study, including the small sample size. Despite the limited power, we were able to find statistically significant results. Because no comparable data were available, we did not calculate the power. In addition, we did not measure the bone mineral density of the samples. To minimize variation caused by bone mineral density, we used a matched sample design, including one tibia from each cadaver in each of the two groups. This reduces the influence of varying bone mineral density. Besides the initial failures (exclusion of the first 3 samples), the setup was standardized and not prone to failures, indicating the testing conditions should be reproducible for further studies. In a biomechanical study by Kojima et al. using synthetic bone models in a simple tibia split fracture in osteoporotic bone, the authors showed a significantly higher interfragmentary compression force in fragments fixed with two 6.5-mm lag screws than in those fixed with four 3.5-mm lag screws, whereas fixation with a clamp following a locking plate and screws was no different from the screw results (5). The results of the present study agree with those reported by Kojima et al, except that we additionally showed a significant decrease in compression in the locking plate group after cyclic loading. In other biomechanical studies (7,9), several fixation methods (three cancellous lag screws alone, two cancellous lag screws with an additional antiglide screw, or a buttress plate for fixation) were compared in a simple tibia split fracture, but no significant differences were found between them. However, the authors did not measure the interfragmentary compression. Although we were able to show that the buttress effect itself is already significant, the mode of rahm-.indd 591 7/10/16 09:11

7 592 s. rahm, l. ebneter, b. ellenrieder, c. spross, f. kulling, j. b. erhardt fixation via lag or locking has an additional effect. In addition, we showed that the interfragmentary compression correlated with the micromotion between fragments during cyclic loading, which supports the validity of our test setup. The lag screw group was able to maintain the interfragmentary compression at a higher level over time and during cyclic loading compared with that achieved by the locking group. These results may be of clinical importance when choosing the type of implant and mode of fixation for specific fractures. These biomechanical data support the philosophy of the AO Foundation, especially with respect to the fixation of intra-articular fractures. To compare a next group with a combined lag screw and locking screw fixation with the already two fixation techniques described would be of great interest and should be performed in a following study with maybe even a greater number of species. The use of the combined technique could be of advantage when treating comminuted fractures in middle aged patients. This is the first study in the literature that investigated the interfragmentary compression in a biomechanical test setup in human cadavers. We believe this study provides relevant information for surgeons treating articular fractures using these two techniques. CONCLUSIONS Conventional plating with lag screws achieves higher interfragmentary compression in simple lateral tibial plateau split fractures than external clamp compression and locked plating. REFERENCES 1. Blakey CM, Rennison M, Guy SP et al. A biomechanical study comparing two fixation methods in depression fractures of the lateral tibial plateau in porcine bone. BMC Sports Science, Medicine, and Rehabilitation 2013 ; 5 : Florin M, Arzdorf M, Linke B et al. Assessment of stiffness and strength of 4 different implants available for equine fracture treatment : a study on a 20 degrees oblique long-bone fracture model using a bone substitute. Vet Surg 2005 ;34 : doi : /j X x 3. Gardner MJ, Brophy RH, Campbell D et al. The mechanical behavior of locking compression plates compared with dynamic compression plates in a cadaver radius model. J Orthop Trauma 2005 ; 19 : Giesinger K, Ebneter L, Day RE et al. Can Plate Osteosynthesis of Periprosthethic Femoral Fractures Cause Cement Mantle Failure Around a Stable Hip Stem? A Biomechanical Analysis. The Journal of Arthroplasty 2014 ; 29 : doi : /j.arth Kojima K, Gueorguiev B, Seva G et al. Biomechanical Evaluation of Interfragmentary Compression At Tibia Plateau Fractures In Vitro Using Different Fixation Techniques. Medicine 2015 ; 94 : e 282. doi : / MD Korner J, Lill H, Müller LP et al. The LCP-concept in the operative treatment of distal humerus fractures - biological, biomechanical and surgical aspects. Injury 2003 ; 34 : doi : /j.injury Koval KJ, Polatsch D, Kummer FJ et al. Split fractures of the lateral tibial plateau : evaluation of three fixation methods. J Orthop Trauma 1995 ; 10 : Moore TM, Patzakis MJ, Harvey JP. Tibial plateau fractures : definition, demographics, treatment rationale, and long-term results of closed traction management or operative reduction. J Orthop Trauma 1987 ; 1 : Parker PJ, Tepper KB, Brumback RJ et al. Biomechanical comparison of fixation of type-i fractures of the lateral tibial plateau. Is the antiglide screw effective? J Bone Joint Surg Br 1999 ; 81 : Rasmussen PS. Tibial condylar fractures. Impairment of knee joint stability as an indication for surgical treatment. J Bone Joint Surg Am 1973 ; 55 : Rüedi TP, Murphy WM ; AO Principles of Fracture Management. George Thieme Verlag. 12. Schatzker J, McBroom R, Bruce D. The tibial plateau fracture. The Toronto experience Clin Orthop Relat Res 1979 ; Wenzl ME, Porté T, Fuchs S et al. Delayed and non-union of the humeral diaphysis compression plate or internal plate fixator? Injury 2004 ; 35 : rahm-.indd 592 7/10/16 09:11

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

More information

NCB Proximal Humerus Plating System

NCB Proximal Humerus Plating System NCB Proximal Humerus Plating System Surgical Technique The right locking option for tough fractures Disclaimer This document is intended exclusively for experts in the field, i.e. physicians in particular,

More information

Locking Proximal Humerus Plate. For complex and unstable fractures.

Locking Proximal Humerus Plate. For complex and unstable fractures. Locking Proximal Humerus Plate. For complex and unstable fractures. Features and Benefits Anatomic design & low profile (2.2mm) 95 No bending required Minimised soft tissue irritation Low risk of subacromial

More information

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Technique Guide LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Table of Contents Introduction Features and Benefits 2 AO ASIF Principles 4 Indications 5 Surgical

More information

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis Tibial Plateau Fx. Osteosynthesis Articular Fractures Osteosynthesis Characteristics of the LCP system LCP 3.5 system LCP 4.5 system Many different traditional plates (small, large) Lag screws Rafting

More information

Large Fragment LCP Instrument and Implant Set

Large Fragment LCP Instrument and Implant Set Part of the DePuy Synthes Locking Compression Plate (LCP ) System Large Fragment LCP Instrument and Implant Set Surgical Technique Table of Contents Introduction Large Fragment LCP Instrument and Implant

More information

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. Technique Guide This publication is not intended for distribution in the USA. Instruments

More information

Locked Plating: Clinical Indications

Locked Plating: Clinical Indications Techniques in Orthopaedics 22(3):181 185 2007 Lippincott Williams & Wilkins, Inc. Locked Plating: Clinical Indications Kyle F. Dickson, M.D., M.B.A., John Munz, M.D. Summary: As shown in the previous article,

More information

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia Types of Plates 1. New Dynamic Compression Plate: DCP Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia 1. Undercut adjacent to the holes low contact: less stress shield 2. Undercut at the undersurface

More information

3.5 mm LCP Extra-articular Distal Humerus Plate

3.5 mm LCP Extra-articular Distal Humerus Plate Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Extra-articular Distal Humerus Plate Surgical Technique Table of Contents Introduction 3.5 mm LCP Extra-articular Distal Humerus

More information

Screw Placement and Types Matter

Screw Placement and Types Matter Screw Placement and Types Matter The Blasted Distal Femur Facture: How to Improve Outcomes Abstract ID#: 1506 Friday October 13, 2017 Kyle Jeray University of South Carolina, Greenville Pinpoint open 59

More information

Locked Plate Fixation of Osteoporotic Humeral Shaft Fractures: Are Two Locking Screws Per Segment Enough?

Locked Plate Fixation of Osteoporotic Humeral Shaft Fractures: Are Two Locking Screws Per Segment Enough? Locked Plate Fixation of Osteoporotic Humeral Shaft Fractures: Are Two Locking Screws Per Segment nough? David J. Hak, MD, MBA, Peter Althausen, MD, MBA, and Scott J. Hazelwood, PhD Objective: The purpose

More information

LCP Distal Humerus Plates

LCP Distal Humerus Plates The anatomic fixation system for the distal humerus with angular stability Surgical technique LCP Locking Compression Plate Contents Indications and contraindications 2 Implants 3 Instruments 5 Preparation

More information

LCP Distal Tibia Plate

LCP Distal Tibia Plate Surgical Technique LCP Locking Compression Plate Original Instruments and Implants of the Association for the Study of Internal Fixation AO/ASIF Table of contents Indications 3 Implants/Instruments 5 Surgical

More information

Biomechanical and Clinical Evaluation of a New Operative Technique

Biomechanical and Clinical Evaluation of a New Operative Technique 27 Posterior Olecranon Plating Biomechanical and Clinical Evaluation of a New Operative Technique Nirmal C. Tejwani, M.D., Ian R. Garnham, F.R.C.S., Philip R. Wolinsky, M.D., Frederick J. Kummer, Ph.D.,

More information

Small Fragment Locking Compression Plate (LCP ) System

Small Fragment Locking Compression Plate (LCP ) System Stainless Steel and Titanium Small Fragment Locking Compression Plate (LCP ) System Surgical Technique Table of Contents Introduction Small Fragment Locking Compression Plate (LCP) System 2 AO Principles

More information

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Medial Proximal Tibia Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Medial Proximal Tibia Plates 2 AO

More information

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large). Technique Guide Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large). Image intensifier control Warning This description

More information

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system. LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system. Technique Guide This publication is not intended for distribution in the USA. Instruments and implants approved

More information

Technique Guide. DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP).

Technique Guide. DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP). Technique Guide DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP). Table of Contents Introduction DCP and LC-DCP Systems 2 Indications

More information

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System A locking plate system that expands a surgeon s options in trauma surgery Zimmer NCB Plating System The Power of Choice The power of having true intraoperative options is at your fingertips. Using standard

More information

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Proximal Humerus Plate 2 AO Principles

More information

Biomechanical Performance of Headless Compression Screws: Fixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS

Biomechanical Performance of Headless Compression Screws: Fixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS Biomechanical Performance of Headless Compression Screws: ixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS M. Vallotton, A. Torcasio PhD Stryker Trauma AG, Selzach, Switzerland Abstract Introduction:

More information

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System.

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System. Technique Guide 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System. Table of Contents AO ASIF Principles of Internal Fixation 4 Indications/Contraindications 5 Surgical Technique

More information

PROXIMAL TIBIAL PLATE

PROXIMAL TIBIAL PLATE SURGICAL NÁSTROJE TECHNIQUE PRO ARTROSKOPII PROXIMAL INSTRUMENTS TIBIAL FOR PLATE ARTHROSCOPY Proximal Tibial Plate Description of medical device The Proximal Tibial Plate is used in epyphyseal and metaphyseal

More information

Humerus shaft - Reduction & Fixation - Compression plate - AO Surgery Reference. Compression plating

Humerus shaft - Reduction & Fixation - Compression plate - AO Surgery Reference. Compression plating Humerus shaft 12-A3 ORIF 1. Principles Compression plating Authors Compression plate Compression plating provides fixation with absolute stability for two-part fracture patterns, where the bone fragments

More information

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. Technique Guide LCP Small Fragment System Table of Contents Introduction

More information

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system.

Technique Guide. LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system. Technique Guide LCP Posterior Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment LCP system. Table of Contents Introduction LCP Posterior Medial Proximal Tibial Plate 3.5 2 AO Principles

More information

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium.

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium. Technique Guide Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium. Table of Contents Introduction Small Fragment Locking Compression Plate (LCP) System 2 AO Principles

More information

4.5 mm LCP Medial Proximal Tibia Plates

4.5 mm LCP Medial Proximal Tibia Plates Part of the DePuy Synthes LCP Periarticular Plating System 4.5 mm LCP Medial Proximal Tibia Plates Surgical Technique Table of Contents Introduction 4.5 mm LCP Medial Proximal Tibia Plates 2 AO Principles

More information

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

Technique Guide. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus. Technique Guide PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus. Table of Contents Introduction PHILOS and PHILOS Long 2 AO Principles 4 Indications 5 Surgical Technique

More information

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System A locking plate system that expands a surgeon s options in trauma surgery Zimmer NCB Plating System The Power of Choice The power of having true intraoperative options is at your fingertips. Using standard

More information

Zimmer Small Fragment Universal Locking System. Surgical Technique

Zimmer Small Fragment Universal Locking System. Surgical Technique Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction

More information

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

More information

GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE

GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE GAMMA LOCKING NAIL INSTRUMENTS OPERATIVE TECHNIQUE FEATURES AND BENEFITS The One Shot Device is a new component of the Gamma Locking Nail instrumentation system determining the correct position of the

More information

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique OptiLock Periarticular Plating System For Distal Tibial Fractures Surgical Technique Contents Introduction... Page 1 Indications & Contraindications... Page 6 System Features... Page 7 Surgical Technique...

More information

Distal Femur Fractures in The Elderly The Ideal Construct

Distal Femur Fractures in The Elderly The Ideal Construct Distal Femur Fractures in The Elderly The Ideal Construct Tak-Wing Lau Department of Orthopaedics and Traumatology Queen Mary Hospital The University of Hong Kong Singapore Trauma 2015 Trauma Through the

More information

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation.

Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation. Cannulated Pediatric Osteotomy System (CAPOS). A single system of osteotomy blade plates and cannulated instrumentation. Technique Guide This publication is not intended for distribution in the USA. Instruments

More information

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. Technique Guide LCP Small Fragment System Table of Contents Introduction

More information

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee. Technique Guide TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee. Table of Contents Introduction TomoFix Osteotomy System 2 AO Principles 4 Indications

More information

NCB Distal Femur System. Surgical Technique

NCB Distal Femur System. Surgical Technique NCB Distal Femur System Surgical Technique NCB Distal Femur System Surgical Technique 3 Surgical Technique NCB Distal Femur System Table of Contents Introduction 4 Indications 8 Preoperative Planning

More information

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

More information

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. Technique Guide LCP Small Fragment System Table of Contents Introduction

More information

Plate Fixation Options

Plate Fixation Options Distal tibia extra-articular fractures can be difficult to treat Tenuous soft tissue Complex fracture patterns Plate Fixation Options Medial plating: Minimally invasive approach Technical ease Anterolateral

More information

LCP Extra-articular Distal Humerus Plate.

LCP Extra-articular Distal Humerus Plate. Technique Guide LCP Extra-articular Distal Humerus Plate. The anatomically shaped and angular stable fixation system for extraarticular fractures of the distal humerus. Table of Contents Introduction

More information

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments Part of the 3.5 mm LCP 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments Surgical Technique TABLE OF CONTENTS INTRODUCTION 3.5 mm LCP Low Bend Medial Distal Tibia Plate 2 Aiming Instruments

More information

Open reduction; plate fixation 1 Principles

Open reduction; plate fixation 1 Principles Executive Editor: Peter Trafton Authors: Martin Jaeger, Frankie Leung, Wilson Li Proximal humerus 11-A2 Open reduction, plate fixation Search search... Shortcuts All Preparations All Approaches All Reductions

More information

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Technique Guide LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Table of Contents Introduction LCP Proximal

More information

Technique Guide. 4.5 mm LCP Proximal Tibia Plates. Part of the Synthes LCP Periarticular Plating System.

Technique Guide. 4.5 mm LCP Proximal Tibia Plates. Part of the Synthes LCP Periarticular Plating System. Technique Guide 4.5 mm LCP Proximal Tibia Plates. Part of the Synthes LCP Periarticular Plating System. Table of Contents Introduction 4.5 mm LCP Proximal Tibia Plates 2 AO Principles 4 Indications 5 Surgical

More information

2.7 mm/3.5 mm Variable Angle LCP. Ankle Trauma System

2.7 mm/3.5 mm Variable Angle LCP. Ankle Trauma System Part of the DePuy Synthes Variable Angle Locking Compression Plate (VA LCP ) System 2.7 mm/3.5 mm Variable Angle LCP Ankle Trauma System Surgical Technique Table of Contents Introduction 2.7 mm/3.5 mm

More information

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016

Orthopedics in Motion Tristan Hartzell, MD January 27, 2016 Orthopedics in Motion 2016 Tristan Hartzell, MD January 27, 2016 Humerus fractures Proximal Shaft Distal Objectives 1) Understand the anatomy 2) Epidemiology and mechanisms of injury 3) Types of fractures

More information

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System.

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System. LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction

More information

Biomechanical Comparison of Four Different Lateral Plate Constructs for Distal Fibula Fractures

Biomechanical Comparison of Four Different Lateral Plate Constructs for Distal Fibula Fractures Biomechanical Comparison of Four Different Lateral Plate Constructs for Distal Fibula Fractures Tobin Eckel, MD Richard Glisson, BS Prashanth Anand, MD Selene Parekh, MD Biomechanical Comparison of Four

More information

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon.

Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon. Olecranon Osteotomy Nail. For simple fractures and osteotomies of the olecranon. Technique Guide Discontinued June 2016; AVAILABLE FOR IMPLANT REMOVAL PURPOSES ONLY DSEM/TRM/0517/0843 Table of Contents

More information

Elbow System Anatomy:

Elbow System Anatomy: Elbow System Elbow System Anatomy: Olecranon Fossa Medial Lateral Ulna Olecranon Radius 2 Elbow System AO-classification of distal humerus fractures 3 3 3 Elbow System Case example Transkondylar elbow

More information

MINI TIBIAL PLATEAU LEVELING OSTEOTOMY (TPLO) SYSTEM

MINI TIBIAL PLATEAU LEVELING OSTEOTOMY (TPLO) SYSTEM MINI TIBIAL PLATEAU LEVELING OSTEOTOMY (TPLO) SYSTEM For stabilizing osteotomies of the canine and feline proximal tibia SURGICAL TECHNIQUE TABLE OF CONTENTS INTRODUCTION Mini Tibial Plateau Leveling

More information

MSP TM Metatarsal Shortening System. Surgical Technique

MSP TM Metatarsal Shortening System. Surgical Technique MSP TM Metatarsal Shortening System Surgical Technique Metaphyseal/Diaphyseal osteotomy does not result in plantar displacement of metatarsal head Osteotomy guide integrated into implant allows for controlled

More information

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless Steel and Titanium.

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless Steel and Titanium. Technique Guide Small Fragment Locking Compression Plate (LCP) System. Stainless Steel and Titanium. Table of Contents Introduction Small Fragment Locking Compression Plate (LCP) System 2 AO Principles

More information

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments

LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments LCP Proximal Tibial Plate 4.5/5.0 with Periarticular Aiming Arm Instruments Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO

More information

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. Surgical Technique This publication is not intended for distribution in the USA. Instruments

More information

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

Clinical. Solutions. Synthes Solutions. Foot and Ankle. Clinical Solutions Foot and Ankle. Foot and Ankle. Fractures of the tibial shaft Fractures of the distal fibula Fractures of the distal tibia Fractures and osteotomies of the calcaneus Arthrodesis Fractures,

More information

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System. Technique Guide 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System. Table of Contents Introduction 3.5 mm LCP Low Bend Medial Distal

More information

3.5 mm LCP Distal Humerus Plates

3.5 mm LCP Distal Humerus Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Distal Humerus Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Distal Humerus Plates 2 AO Principles 4 Indications

More information

Surgical Technique. Targeter Systems Overview

Surgical Technique. Targeter Systems Overview Surgical Technique Targeter Systems Overview PERI-LOC Locked Plating System Targeter Systems Overview Table of contents Product overview... 2 Introduction... 2 Indications... 2 Design features and benefits...

More information

For Distal Femur Fractures. 95º Condylar Plate. Quick Reference Chart

For Distal Femur Fractures. 95º Condylar Plate. Quick Reference Chart For Distal Femur Fractures 95º Condylar Plate Quick Reference Chart 95 Condylar Plate. Quick reference chart for distal femur fractures. Insert guide wires Fix condylar fragments with 6.5 mm cancellous

More information

Low Bend Distal Tibia Plates

Low Bend Distal Tibia Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Low Bend Medial Distal Tibia Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Low Bend Medial Distal Tibia

More information

Distal Radius Plate 2.4/2.7 dorsal and volar

Distal Radius Plate 2.4/2.7 dorsal and volar Distal Radius Plate 2.4/2.7 dorsal and volar Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Distal Radius Plate

More information

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures. LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures. Surgical Technique This publication is not intended for distribution in the

More information

Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study

Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study Nailing Stability during Tibia Fracture Early Healing Process: A Biomechanical Study Natacha Rosa, Fernão D. Magalhães, Ricardo Simões and António Torres Marques Enhanced Bone Healing in intramedullary

More information

AxSOS Locking Plate System

AxSOS Locking Plate System AxSOS Locking Plate System Operative Technique Small Fragment Basic Fragment 1 2 Contents Page 1. Introduction 4 2. Features & Benefits 5 4 and 5mm Compression Plates 5 Reconstruction and 1/3 Tubular Locking

More information

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA.

Humerus Block. Discontinued December 2016 DSEM/TRM/0115/0296(1) Surgical Technique. This publication is not intended for distribution in the USA. Humerus Block Surgical Technique Discontinued December 2016 DSEM/TRM/0115/0296(1) This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Contents

More information

ROTATIONAL PILON FRACTURES

ROTATIONAL PILON FRACTURES CHAPTER 31 ROTATIONAL PILON FRACTURES George S. Gumann, DPM The opinions and commentary of the author should not be construed as refl ecting offi cial U.S. Army Medical Department policy. Pilon injuries

More information

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction LCP Distal Fibula Plates 2 AO Principles 4 Indications 5 Surgical Technique

More information

2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1

2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1 2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1 System overview Simply complete: A comprehensive system, consisting of five (5) distal humerus plates and three (3) types of olecranon plates Implant

More information

NCB Large Fragment System. Surgical Technique

NCB Large Fragment System. Surgical Technique NCB Large Fragment System Surgical Technique NCB Large Fragment System Surgical Technique 3 Table of Contents Introduction 4 System Features 7 Indications and Contraindications 10 NCB Large Fragment Plate

More information

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. Surgical Technique LCP Small Fragment System This publication

More information

ANATOMIC LOCKED PLATING SYSTEM

ANATOMIC LOCKED PLATING SYSTEM ANATOMIC LOCKED PLATING SYSTEM There is only one...dvr Anatomic. There is only one... ANATOMIC LOCKED PLATING SYSTEM Distal Tibia TiMAX for strength, biocompatibility and enhanced imaging capabilities

More information

Pre-Operative Planning. Positioning of the Patient

Pre-Operative Planning. Positioning of the Patient Surgical Technique Pre-Operative Planning Decide upon the size and angle of the barrel plate to be used from measuring the x-rays. To maximise the sliding action when using shorter lag screws, the Short

More information

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System. 2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System. Technique Guide Instruments and Implants approved by the AO Foundation Table of Contents Introduction 2.4 mm LCP Radial

More information

2.7 mm/3.5 mm LCP Distal Fibula Plate

2.7 mm/3.5 mm LCP Distal Fibula Plate Part of the DePuy Synthes Locking Compression Plate (LCP ) System 2.7 mm/3.5 mm LCP Distal Fibula Plate Surgical Technique Table of Contents Introduction 2.7 mm/3.5 mm LCP Distal Fibula Plates 2 AO Principles

More information

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Technique Guide 2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Table of Contents Introduction 2.4

More information

Technique Guide. VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology.

Technique Guide. VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology. Technique Guide VA-LCP Distal Humerus Plates 2.7/3.5. The low-profile fixation system with variable angle locking technology. Image intensifier control Warning This description alone does not provide sufficient

More information

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. LCP Extra-articular Distal Humerus Plate. The anatomically shaped and angular stable fixation system for extraarticular fractures of the distal humerus. Surgical Technique This publication is not intended

More information

OPERATIVE TREATMENT OF THE INTERCONDYLAR FRACTURE OF THE FEMUR

OPERATIVE TREATMENT OF THE INTERCONDYLAR FRACTURE OF THE FEMUR OPERATIVE TREATMENT OF THE INTERCONDYLAR FRACTURE OF THE FEMUR S G Chee, K S Lam, B K Tay, N Balachandran SYNOPSIS Operative treatment of 28 intercondylar fractures of femur were done from 98 to 985. The

More information

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 2.7 mm/3.5 mm LCP Distal Fibula Plates 2 AO Principles

More information

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus.

PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus. PHILOS and PHILOS Long. The anatomic fixation system for the proximal humerus. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the

More information

3.5 mm LCP Olecranon Plates

3.5 mm LCP Olecranon Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Olecranon Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Olecranon Plates 2 AO Principles 3 Indications

More information

Instrument and Implant for wrist fracture

Instrument and Implant for wrist fracture Instrument and Implant for wrist fracture Jansri Janpanya Product specialist The Bangkok Unitrade Co,.ltd. Objectives Type of LCP for distal radius Fx. The new LCP design for distal radius Fx. Have knowledge

More information

VariAx Compression Plating System. Operative technique

VariAx Compression Plating System. Operative technique VariAx Compression Plating System Operative technique VariAx Compression Plating System Operative technique VariAx Compression Plating System Contents 1. Introduction... 3 2. Indications, MR safety information

More information

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM 3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM Part of the DePuy Synthes Variable Angle Periarticular Plating System SURGICAL TECHNIQUE TABLE OF CONTENTS INTRODUCTION 3.5 mm VA-LCP Proximal Tibial Plate 2 AO

More information

LCP Locking Compression Plate. Surgical Technique

LCP Locking Compression Plate. Surgical Technique LCP Locking Compression Plate Surgical Technique Image intensifier control This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction by a surgeon

More information

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system.

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system. 2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the locking compression plate (LCP) system. Anatomically contoured Multiple screw options for fixedangle support Coaxial holes minimize screw head

More information

Conventus CAGE PH Surgical Techniques

Conventus CAGE PH Surgical Techniques Conventus CAGE PH Surgical Techniques Conventus Orthopaedics The Conventus CAGE PH (PH Cage) is a permanent implant comprised of an expandable scaffold, made from nitinol and titanium, which is deployed

More information

Osteotomy System. The reference lines on the plate help facilitate the creation of the osteotomy, when a free hand cut is preferred.

Osteotomy System. The reference lines on the plate help facilitate the creation of the osteotomy, when a free hand cut is preferred. Osteotomy System Osteotomy System Since 1988, Acumed has been designing solutions for the demanding situations facing orthopaedic surgeons, hospitals and their patients. Our strategy has been to know the

More information

FLT105 12/02.

FLT105 12/02. FLT105 12/02 www.biometmerck.co.uk Disclaimer Biomet Merck Ltd, as the manufacturer of this device, does not practice medicine and does not recommend this or any other surgical technique for use on a

More information

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual

Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual Orthopedic Bone Nail System - Distal Femoral Nail Surgical Technique Manual Note: The surgical procedures should be performed under the guidance of qualified skilled orthopedic surgeons, and this surgical

More information

3.5 mm Locking Attachment Plate

3.5 mm Locking Attachment Plate For Treatment of Periprosthetic Fractures 3.5 mm Locking Attachment Plate Surgical Technique Table of Contents Introduction 3.5 mm Locking Attachment Plate 2 Indications 4 Surgical Technique Preparation

More information