The Effect of a Recommended Cervical Range. Soft Collar, Used as Normally or Reversed, on Three Planes of of Motion

Size: px
Start display at page:

Download "The Effect of a Recommended Cervical Range. Soft Collar, Used as Normally or Reversed, on Three Planes of of Motion"

Transcription

1 E S E. A R C H S T U D Y The Effect of a Recommended Cervical Range Virginia M. Carter, MPT' )o Ann M. Fasen, MPT, CSCS * )uan M. Roman, )r., MS, MPT~ Karen W. Hayes, PhD, PT4 Cheryl M. Petersen, MS, PT Soft Collar, Used as Normally or Reversed, on Three Planes of of Motion Copyright All rights reserved. ervical orthoses have been prescribed routinely by physicians and physical therapists for the purposes of decreasing inflammation, pain, and spasm; controlling instability; supporting the cervical spine (decreasing compressive loading); and immobilizing the cervical spine following trauma (1,2, 5,6,9,10,13,14). Soft cervical collars are readily available and comfortable for the patient to wear compared with other cervical orthoses, but the degree to which they immobilize the cervical spine is questionable and is the focus of this study. Generally, the effectiveness of a cervical orthosis is determined by its ability to immobilize the affected area. Available studies of subject5 with and without cervical dysfunction disagree whether soft cervical collars can physically limit motion ( 1,2,5,6,9, 10,ll,l.%l5). The factors that affect the effectiveness of the soft cervical collar include height and size. Jones concluded that soft cervical collars with sufficient anterior height were able to restrict motion in the upper cervical region (1 1 ). Other studies have found minimal or no significant restriction in cervical motion while the subject was wearing a soft cervical collar (5,10,15). Hart et al suggested Clinicians recommend soft cervical collars to immobilize the cervical spine following trauma. They apply them either as intended by the manufacturer or reversed (collar rotated 180 ), purportedly to achieve limitation in a specific direction. This study investigated the effectiveness of soh cervical collars in limiting cervical range of motion when worn as recommended or reversed. All planes of cervical range of motion of SO volunteer subjects without current or past cervical dysfunction were measured under three conditions (no collar, recommended use, and reversed) using the Orthopedic Systems Inc. Computerized Anatometry-6000 Spine Motion Analyzer. The instrument has been previously shown to produce measures with high reliability and to correlate strongly with known angular measures. Within subjects analyses of variance indicated significant differences in all six ranges of motions among the three conditions. Post hoc paired t tests showed that wearing a collar either as recommended or reversed decreased motion compared with not wearing a collar, and that the position of the collar affected range of motion in three of the six motions. Differences in range can be attributed to location of the collar closure and initial head posture. Soh cervical collars can physically limit motion when worn either way. Key Words: cervical, orthosis, range of motion ' Staff Physical Therapist, TheraTx, Denver, CO. Address for correspondence: 9601 W. Chatfield Ave., Unit F, Littleton, CO Staff Physical Therapist, Center for Spine, Sport & Occupational Rehabilitation, Chicago, IL ' Staff Physical Therapist, St. Mary of Nazareth, Chicago, II Curriculum Coordinator; Assistant Professor, Programs in Physical Therapy, Northwestern University Medical School, Chicago, IL Instructor, Programs in Physical Therapy, Northwestern University Medical School, Chicago, IL in their review that if the appropriate size was chosen, the soft collar afforded some vertical support of the cervical spine by reducing the amount of weight borne directly by the spine (9). In light of potential sensorimotor-skeletal interaction, Johnson et al suggested that the soft collar is comfortable and may serve as a useful reminder to the patient to restrict his or her own motion (14). In this capacity, a flexible collar should satisfy the goal of providing gentle support (IS). Dudgeon (6) believed, however, that long term use of a cervical orthosis might lead to muscle atrophy, causing further deterioration of a patient's neck. Beavis (2) countered that because immobilization was not complete, the neck muscles were allowed to contract isometrically within the collars, and atrophy and disuse were prevented. Finally, Althoff and Goldie (1) found JOSPT Volume 23 Number J March 1W6

2 Copyright All rights reserved. that cervical collars in general, including soft collars, did not stabilize atlantoaxial instability in patients with rheumatoid arthritis. Studies of subjects with cervical dysfunction that compared wearing a cervical collar and active treatment have also shown mixed results with regard to effectiveness of treatment. Mealy et al assessed residual pain and cervical movement at 4 and 8 weeks following a whiplash accident and found that wearing a soft collar was not as effective as the Maitland technique of neck mobilization in decreasing pain or increasing movement (1 7). In a similar study, Pennie and Agambar (18) compared 2 weeks of rest in soft cervical collars with active treatment of traction and exercises. They found no significant differences between the two treatments (collar vs. active) with regard to improvement in pain scores or cervical movement. These studies used different methods to measure cervical range of motion in assessing the degree of immobility that a soft cervical collar provides. Many studies (5,6,10,11,13, 15) have used radiographs to assess cervical motion, but the risks to sub jects and the cost of radiographs is of concern (20). In addition, studies do not consistently use the same methods for analyzing radiographs and, thus, it is difficult to compare the results of radiographic studies. Other studies (2,17,18) have used goniometry to measure cervical range of motion. The reliability of goniometric measurements varies depending on the joint studied, the movement measured, and the regional anatomy and biomechanics (3.7). Goniometric studies for the cervical spine have indicated intratester intraclass correlation coefficients (ICC 1,l) between 0.78 and 0.90, and interrater ICCs (1,l) between 0.54 and 0.79 (23). Zachman et al found that interexaminer reliability of goniometer measurements, estimated with Pearson product moment correlation coefficients, varied from 0.37 to 0.86 (24). As a result of the substantial measurement error, it is difficult to interpret the results of studies that use the goniometer as their tool. To date, studies examining the degree of immobilization provided by a soft cervical collar have produced conflicting results. It is important to know whether soft cervical collars can restrict motion because they are often prescribed for this goal. If collars are not capable of restricting motion, then they should not be prescribed for that purpose. To determine whether soft cervical collars can restrict cervical range of motion, a reliable method of measuring cervical If collars are not capable of restricting motion, then they should not be prescribed for that purpose. range of motion is necessary. Recent studies have shown that the Orthopedic Systems Inc. Computerized Anatometry-6000 Spine Motion Analyzer (OSI CA-6000 SMA) produces reproducible measurements of spinal range of motion. For cervical range of motion, intrarater ICCs (2,l) were between 0.77 and 0.94 and interrater ICCs (2,4) ranged from 0.87 to 0.93 (12). Angles measured with it are extremely close to known angles (21). In addition to the method of measuring the effectiveness of the collar, the manner in which it is worn must be considered. Depending on the goal of treatment, practitioners recommend that patients wear a collar either as intended by the manufacturer or in the reverse fashion.,wen collars are worn as intended by the manufacturer, they may force the TABLE 1. Means, standard deviations, and ranges of height, weight, and age of 50 subjects (24 men and 26 women). patient into a forward head position, prompting practitioners to suggest turning it around. The effect of the position of the collar on cervical range of motion has not been studied. The purpose of this study was to test the hypothesis that a soft cervical collar when worn as intended by the manufacturer or reversed will limit cervical range of motion. Our null hypotheses were that a soft cervical collar worn either way does not limit cervical range of motion in any direction. METHOD Subjects The study group consisted of 50 adult volunteers (24 men and 26 women). Means, standard deviations, and ranges of subjects' heights, weights, and ages are shown in Table 1. Forty-nine subjects were right-hand dominant and one subject was lefthand dominant. The subjects were recruited through advertisements at Northwestern University's physical therapy, medical, and dental schools and through personal contact by the authors. Subjects had no complaints of local or radicular cervical pain, and no history of cervical injury or disease. Informed consent was obtained from all subjects. Volume 23 Number 3 March 1996 JOSIT

3 Copyright All rights reserved. Examiners Testing was done by three fulltime students in the last year of the entry-level masters degree program at Programs in Physical Therapy, Northwestern University Medical School, Chicago, IL. The examiners reviewed the instruction manual and standardized written protocol provided by the manufacturer of the OSI CA-6000 SMA (4) and were also trained by faculty with previous research experience with this measurement instrument. Before data collection was initiated, all three examiners participated in 6 hours of training in their roles. Instrumentation Subjects were fitted with one of three sizes of the soft cervical collar (small, medium, and extra large). The collars were made of 2.54 cm, medium density foam rubber covered with cotton stockinette. The collars were contoured higher in front with a notch for the chin and a hook and loop fastener in back for adjustable and easy closure (The Saunders Group, Inc., Minneapolis, MN). Cervical range of motion was measured using the OSI CA-6000 SMA (Orthopedic Systems Inc., Hayward, CA). The OSI CA-6000 SMA simultaneously recorded spinal range of motion in three dimensions using six potentiometers mounted on a movable arm that was attached to the subject at two points along the spine. To measure regional motion of the cervical spine between the atlanto-occipital joint and first thoracic vertebra (T,), a skull cap was placed so that the anterior segment of its horizontal band rested slightly above the eyebrows and the posterior segment rested below the curvature of the external occipital protruberance. The thoracic harness, consisting of two rubber straps, was looped around each shoulder and under the axilla to secure the second aluminum mount over the upper thoracic spine. The movable arm linkage was at- FIGURE 1. Posterior view of skull cap and movable arm linkage attached to the thoracic spine. tached between the posterior mount of the skull cap and the second mount over the upper thoracic spine (Figure 1). Examiners observed the subjects prior to and during testing to assure free movement of the linkage. After the movable arm linkage was connected, the subject was instructed to sit squarely against the back of a chair, feet flat on the floor, eyes straight ahead, with their hands on their thighs. Subsequently, the tester operating the OSI CA-6000 A FIGURE 2. Without soit colla. SMA computer electronically set the potentiometers to zero. After each testing condition, the initial starting posture was again assumed and the computer zeroed. Procedure Subjects were seated upright, facing forward in a steel-framed straightbacked chair. To prevent trunk motion, subjects' pelvic and lumbar regions were held squarely against the chair with a standard seat belt, which was buckled at the back of the chair and tightened around subjects' waists. Subjects rested their hands comfortably on their thighs, and they positioned their feet flat on the floor. Practice trials were given before each collar condition to ensure that sub jects understood the commands and to enforce visually that there should be no motion at the thoracic and lumbar regions. Limiting motion at the thoracic and lumbar regions was controlled by observation and verbal instruction; it was not measured. Subjects were fitted with one of three collar sizes so that they perceived the fit as comfortable but snug. The soft cervical collar was fitted to subjects' necks in their initial starting position and fastened using the adjustable closure. Subjects were tested without the soft cervical collar JOSPT Volume 23 Number 3 March 1996

4 Copyright All rights reserved. FIGURE 3. Soh collar worn as intended bv the manufacturer. (Figure 2), with the soft cervical collar worn as intended by the manufacturer (closure posterior) (Figure 3), and with the collar reversed (Figure 4). Three examiners were present. One examiner consistently oriented subjects to the test procedure and fitted the subject. with the measuring device and collar. A second examiner consistently operated the OSI CA SMA computer and instructed subjects when and how to move, while the third examiner recorded data. Subject5 were instructed to perform flexion and extension, right and left lateral flexion, and right and left axial rotation. Subjects were instructed to lead with their eves and move as far as comfortably possible to their cervical end ranges. End range FIGURE 4. Soil collar worn reversed. was defined as the point at which the subjects were unable to move further, with no substitution movements. Specifically, during right and left lateral flexion, subjects were instnlcted not to elevate their shoulders. During right and left rotation, subjects were instructed to lead with their eyes and look over each shoulder. Subject5 were instructed to deform the collar with all movement directions. Before any measurement5 were made, sub jects practiced the movements three times before the no collar condition and one time for each collar condition. Four trials for each plane of movement were recorded for each testing condition (no cervical collar, collar worn as prescribed by the manufacturer, and collar reversed). A reliability analysis of the four trials for each motion among subjects produced intraclass correlation coefficients (2,l) (22) between 0.92 and To examine whether order of collar placement (forward or reversed) affected the results, half of the subjects (even numbered) wore the collar forward first, and half (odd numbered) wore the collar reversed first. Data Analysis Means, standard deviations, and ranges measured from the external occipital protn~berance to T, were used to describe the ranges of motion recorded for each test condition. Within subject analyses of variance were performed to compare the maximum degree of each cervical motion across three conditions (no collar, collar as prescribed by the manufacturer, and collar reversed). Post hoc paired t tests with Bonferroni adjustment (a = 0.017) were used to identify the specific differences. All analyses were performed with the Statistical Package for the Social Sciences (SPSS, Inc., Chicago, IL) and a personal computer (Apple Computer, Inc., Cupertino, CA). RESULTS Muhivariate analysis of variance (ANOVA) indicated no differences in range of motion among subjects with regard to gender (F(df 49) = 1.30, not significant) and order of soft cervical collar placement (F(df49) = 0.95, not significant). Therefore, data for all subject5 were pooled for further analyses. Means and standard deviations of all six ranges of motions within each of the three conditions are listed in Table 2. Significant differences were found in all six ranges of motion among the three conditions (Table 2). Post hoc paired t tests showed that wearing a collar either as recommended or reversed produced significantly less motion than not wearing a collar (Table 3). For the motions of flexion, extension, and right rotation, the two conditions with the collar allowed different amounts of motion. For the motions of right lateral flexion, left lateral flexion, and left rotation, however, range of motion while wearing a collar as recommended did not differ from range of motion while wearing the collar reversed (Table 3). DISCUSSION The results of this study sup ported the hypothesis that healthy subjects wearing the cervical collar in either position produced less motion when compared with not wearing a collar. These results could be attrib uted to the measurement device, the subject tpe, subjects' unwillingness to deform the collar due to the need for more effort or the feeling of restriction, the location of the collar closure, and the starting head posture. The device used to measure cervical range of motion was the OSI CA-6000 SMA, which has been shown to be highly reliable, and it closely reproduces measures of known angles, thereby detecting small degrees of motion with little error (12,21). Volume 23 Number 3. March 19%. JOSFT

5 Copyright All rights reserved. - Motion-Condition X SD F(df 49) P Flexion < No collar Collar forward Collar reversed Extension <O.OOOl No collar Collar forward Collar reversed Right lateral flexion < No collar Collar forward Collar reversed Left lateral flexion < No collar Collar forward Collar reversed Right rotation < No collar Collar forward Collar reversed Left rotation < No collar Collar forward Collar reversed df = Degrees of freedom. Collar forward = Collar as recommended condition. TABLE 2. Means, standard deviations (SD), and within subjects analysis of variance for the six cervical ranges of motion (degrees) for three conditions of collar use. Mean Difference Flexion No vs. For <0.001 No vs. Rev <0.001 For vs. Rev <0.001 Extension No vs. For <0.001 No vs. Rev <0.001 For vs. Rev <0.001 Right lateral flexion No vs. For <0.001 No vs. Rev <0.001 For vs. Rev NS Left lateral flexion No vs. For <0.001 No vs. Rev <0.001 For vs. Rev NS Right rotation No vs. For <0.001 No vs. Rev <0.001 For vs. Rev Left rotation No vs. For <0.001 No vs. Rev <0.001 For vs. Rev NS df = Degrees of freedom. No = No collar condition. For = Collar as recommended condition (forward). Rev = Collar reversed condition. NS = Not significant. TABLE 3. Post hoc paired t tests comparing six ranges of motion (degrees) under three conditions of collar use. Subjects in this study were all healthy volunteers, thus leading to limited variability in relationship with the magnitude of motion that occurred. Small differences in range of motion with and without the collar (.I-1 lo) thus produced effect sizes as large as 2.3 standard deviations (no collar vs. collar forward for left rotation). With effect sizes this large, statistical significance is easy to achieve (19). Differences in range of motion seen with and without the collar may have been due to the increased effort required to deform the collar. In addition, a majority of subjects stated that the testing condition with no cervical collar was most comfortable because they perceived the collar as restrictive. Therefore, subjects may have been unwilling to deform the collar due to its restrictive nature. The location of the collar closure may explain the differential motion restriction in the sagittal plane. The collar closure is a stiffer component of the collar as compared with the soft foam in the rest of the collar. Differences in range can be attrib uted to location of the collar closure depending on whether it was worn as recommended (closure posterior) or reversed (closure anterior). When the collar closure was located posteriorly, extension was more significantly limited and when the closure was located anteriorly, flexion was more significantly limited. Alternatively, differential sagittal motion restriction may be explained by the initial head posture. When the collar is worn as recommended (closure posterior), the collar is higher anteriorly. To clear the anterior height of the collar, the subject may need to a+ sume a forward head posture. In this position, the upper cervical spine a p pears more extended, the middle portion of the cervical spine is relatively extended, and the cervico-thoracic junction is flexed. When the collar is worn reversed, with the opening in front and the height of the collar pos JOSPT Volume 23 Number 3 March 1996

6 Copyright All rights reserved. teriorly, the subject assumes a more neutral position (16). Assuming that the collar limits cervical motion, when subject$ are asked to flex with the collar worn as recommended, they have available the full excursion of upper cervical flexion plus some midcervical flexion. With the collar reversed, they have less upper and midcervical flexion available because they are starting from a more flexed position. Hence, flexion appears to be greater with the collar worn as recommended. Likewise, when subjects are asked to extend with the collar worn as recommended, they have little upper cervical extension and only some midcervical extension avail-.. If the goal of using the collar is to prevent patients from experiencing pain at the extremes of range, the amount of limitation offered by the collar may be clinically meaningful. able. With the collar reversed, they have upper cervical extension as well as midcervical extension available. Thus, extension appears to be greater with the collar worn reversed (8). Regardless of how the collar is worn, it limits motion as little as 3" and as much as 11" compared with not wearing the collar. The total sagittal excursion of the cervical spine is nearly the same, however, in both collar conditions (Table 2). It a p pears, therefore, that the differences in flexion and extension ranges with the different collar positions are more a function of the starting position of the head resulting from the location of the highest portion of the collar than the stiffness of the closure. Right and left lateral flexion and left rotation were limited almost equally regardless of whether the collar was worn as prescribed or reversed. These results would be expected because the height and softness of the collar on both sides are the same. Statistically, right rotation was significantly more limited with the collar worn reversed than when worn as prescribed. However, this difference was less than 2". Left rotation, while not significantly different, had approximately 1" difference in limitation between the prescribed and reversed conditions. The clinical relevance of these results is questionable because the difference in limitation between the two conditions for both motions is so small. The clinical relevance of the limitations in motion produced by the soft cervical collar must be addressed further. The soft cervical collar was able to limit flexion approximately 11.5" when worn reversed, but only about 5.0" when worn as recommended. When the collar was worn as recommended, it limited extension approximately 8.6", and when reversed, it limited extension about 4.7". This finding clearly indicates that the recommended position of the collar should be reversed if the required effect is to limit flexion. Right and left lateral flexion were limited approximately 3.0, whether the collar was worn as recommended or reversed. The soft cervical collar limited right rotation about 5.0" when worn as recommended and about 6.5" when worn reversed. Left rotation was limited about 8.0" with the collar as prescribed and approximately 8.9" when worn reversed. If the goal of using the collar is complete immobilization of the cervical spine, this amount of limitation may not be clinically meaningful. The results of our study are similar to those of Johnson et al ( 13) who used radiographic measurements to document the effectiveness of soft collars in limiting cervical motion. Reporting limitations of motion as percentages of maximum range, they also found that soft cervical collars limit motion in the sagittal plane when compared with not wearing a collar and that flexion was limited more than extension (13). Clearly, the soft cervical collar does not completely eliminate m e tion. On the other hand, if the goal of using the collar is to prevent patients from experiencing pain at the extremes of range, the amount of limitation offered by the collar may be clinically meaningful. Given the results of this study and considering that subjects were instructed to deform the collar, the collar does limit range of motion for all motions. If pain is a factor at the extremes of range, keeping the patient out of extreme ranges of motion addresses the issue of pain relief. This limitation produced by the collar may further serve to relieve minor muscle spasm and offer protection and support from further trauma. The soft cervical collar, however, does not seem to be an appropriate choice if a patient is experiencing serious instability or is in need of complete immobilization because it does not completely limit motion. CONCLUSION The results of this study indicate that soft cervical collars are capable of physically limiting motion in all directions when worn either way, but only up to 11". Motions in the sagittal plane are controlled to a greater degree, depending on collar positioning. The soft cervical collar is inap propriate if complete immobilization of the cervical spine is the goal of treatment. JOSFT ACKNOWLEDGMENTS Our acknowledgments and thanks to all those who volunteered their time to be subjects, particularly Volume 23 Number 3 March 1996 JOSPT

7 Copyright All rights reserved. the Northwestern University Medical School Programs in Physical Therapy classes of 1993 and REFERENCES 1. Althoff B, Goldie IF: Cervical collars in rheumatoid atlanto-axial subluxation. A radiographic comparison. Ann Rheum Dis 39: , Beavis A: Cervical orthoses. Prosthet Orthot Int 13:6-13, Boone DC, Azen SP, Lin C-M, Spence C, Baron C, Lee L: Reliability of goniometric measurements. Phys Ther 58: , CA-6000 Spine Motions Analyzer: User's guide (Version I. I), Hayward, CA: Orthopedic Systems Inc., Colachis SC, Strohm BR, Canter EL: Cervical spine motion in normal women: A radiographic study of effect of cervical collars. Arch Phys Med Rehabil54: , Dudgeon P: The effectiveness of cervical orthoses: The patient's viewpoint. Br J Occup Ther 47: , Gajdosik RL, Bohannon RW: Clinical measurement of range of motion: Review of goniometry emphasizing reliability and validity. Phys Ther 67: , Grieve GP: Modern Manual Therapy of the Vertebral Column, pp Edinburgh: Churchill Livingstone, Hart DL, Johnson RM, Simmons EF, Owen J: Review of cervical orthoses. Phys Ther , Hartman JT, Palumbo F, Hill BJ: Cineradiography of the braced normal cervical spine: A comparative study of five commonly used cervical orthoses. Clin Orthop 1 OW7-102, Jones MD: Cineradiographic studies of the collar-immobilized cervical spine. Neurosurg 17: , Johnson RD, Schuit D, Petersen CM: Reliability of the OSI CA-6000 Spine Motion Analysis System for active cervical range of motion measurement of normal subjects. Ph ys Ther 72:s 108, 1993 (abstract) 13. Johnson RM, Hart DL, Simmons EF, Ramsby GR: Cervical orthoses: A study comparing their effectiveness in restricting cervical motion in normal subjects. J Bone Joint Surg 59A(3): , Johnson RM, Owen JR, Hart DL, Callahan RA: Cervical orthoses: A guide to their selection and use. Clin Orthop 154:34-45, Kaufman BS, Lunsford TR, Lunsford BR, Lance LL: Comparison of three prefabricated cervical collars. Orthot Prosthet 39(4):2 1-28, Kendall FP, McCreary EK: Muscles Testing and Function (4th Ed), pp 66, 91. Baltimore: Williams & Wilkins, 1993 Mealy K, Brennan H, Fenelon GCC: Early mobilisation of acute whiplash injuries. Br Med J 292: , 1986 Pennie BH, Agambar LJ: Whiplash injuries: A trial of early management. J Bone Joint Surg 72B(2): , 1990 Portney LG, Watkins MP: Foundations of Clinical Research: Applications to Practice, pp Norwalk: Appleton & Lange, 1993 Rosen PB, McSwain NE Jr., Arata M, Stahl S, Mercer D: Comparison of two new immobilization collars. Ann EmergMed21: , 1992 Schuit D, Petersen CM, Johnson RD: Reliability and validity of the OSI CA Spine Motion Analyzer for sagittal and frontal plane motions. Phys Ther 72:S62, 1993 (abstract) Shrout PE, Fleiss JL: lntraclass correlation: Uses in assessing rater reliability. Psycho1 Bull 86: , 1979 Youdas JW, Carey JR, Garrett TR: Reliability of measurements of cervical spine range of motion--comparison of three methods. Phys Ther 71 :98-106, 1991 Zachman Z], Traina AD, Keating JC Jr., Bolles ST, Braun-Porter L: Interexaminer reliability and concurrent validity of two instruments for the measurement of cervical ranges of motion. J Manipulative Physiol Ther 72: , 1989 JOSPT Volume 23 Number 3 March 1996

Cervical orthoses A. BEAVIS*

Cervical orthoses A. BEAVIS* Prosthetics and Orthotics International, 1989,13, 6-13 Cervical orthoses A. BEAVIS* Oxford Orthopaedic Engineering Centre, Oxford Abstract A biomechanical study is presented to compare the effectiveness

More information

Effects of Capital Collar Enhanced on Head-Cervical Movements in Comparison with Miami J Advanced and Aspen Vista TX Collars

Effects of Capital Collar Enhanced on Head-Cervical Movements in Comparison with Miami J Advanced and Aspen Vista TX Collars DeRoyal Industries, Inc. 2013 Effects of Capital Collar Enhanced on Head-Cervical Movements in Comparison with Miami J Advanced and Aspen Vista TX Collars Biomechanics/Sports Medicine Laboratory Department

More information

OBJECTIVE AND ACCURATE measurements of lumbar

OBJECTIVE AND ACCURATE measurements of lumbar 99 Strapped Versus Unstrapped Technique of the Prone Press-Up for Measurement of Lumbar Extension Using a Tape Measure: Differences in Magnitude and Reliability of Measurements William D. Bandy, PhD, PT,

More information

subluxation: a radiographic comparison

subluxation: a radiographic comparison Annals of the Rheumatic Diseases, 1980, 39, 485-489 Cervical collars in rheumatoid atlanto-axial subluxation: a radiographic comparison BO ALTHOFF AND IAN F. GOLDIE From the Department of Orthopaedic Surgery

More information

Cervical Spine Exercise and Manual Therapy for the Autonomous Practitioner

Cervical Spine Exercise and Manual Therapy for the Autonomous Practitioner Cervical Spine Exercise and Manual Therapy for the Autonomous Practitioner Eric Chaconas PT, PhD, DPT, FAAOMPT Assistant Professor and Assistant Program Director Doctor of Physical Therapy Program Eric

More information

Bracing for Scoliosis

Bracing for Scoliosis Bracing for Scoliosis The main goal of a brace in scoliosis is to prevent further deformity, as well as to prevent or delay the need for surgery. If surgery is needed, delaying the procedure as long as

More information

The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device

The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device Original Article The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device J. Phys. Ther. Sci. 15: 7 12, 2003 ASGHAR REZASOLTANI, Ph D, PT 1, 2), AMIR AHMADI, B Sc,

More information

Noble Hinge Technology CHANGING THE FOUNDATION OF SPINAL BRACING

Noble Hinge Technology CHANGING THE FOUNDATION OF SPINAL BRACING Noble Hinge Technology CHANGING THE FOUNDATION OF SPINAL RACING DEROYAL SPINAL TECHNOLOGY Noble Hinge Technology: Simplifying the process Noble Hinge Technology (NHT) is a unique and innovative combination

More information

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System Journal of Sport Rehabilitation, 1998, 7, 95-101 0 1998 Human Kinetics Publishers, Inc. lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System Randy Schmitz

More information

Cruciform anterior spinal hyperextension (CASH) brace with round anterior chest pads.

Cruciform anterior spinal hyperextension (CASH) brace with round anterior chest pads. General Description Thoracolumbar orthoses (TLOs) are used mainly to treat fractures between T10 and L2, because their mobility is not restricted by the ribs, unlike fractures between T2 and T9. Immobilization

More information

Chiropractic Glossary

Chiropractic Glossary Chiropractic Glossary Anatomy Articulation: A joint formed where two or more bones in the body meet. Your foot bone, for example, forms an articulation with your leg bone. You call that articulation an

More information

A New Orthosis for Fixation of the Cervical Spine Fronto- Occipito-Zygomatic Orthosis

A New Orthosis for Fixation of the Cervical Spine Fronto- Occipito-Zygomatic Orthosis A New Orthosis for Fixation of the Cervical Spine Fronto- Occipito-Zygomatic Orthosis Toshiro Nakamura, O.A. Mitsuru Oh-Hama, M.D. Hikosuke Shingu, M.D. INTRODUCTION Most of the cervical orthoses for longterm

More information

Spinal Biomechanics & Sitting Posture

Spinal Biomechanics & Sitting Posture Spinal Biomechanics & Sitting Posture Sitting: weight of the body is transferred to a supporting area 1.Main Contact points (seat) Ischial tuberosities Soft tissues 2. Secondary contact points (other)

More information

Reliability of Measuring Trunk Motions in Centimeters

Reliability of Measuring Trunk Motions in Centimeters Reliability of Measuring Trunk Motions in Centimeters MARGARET ROST, SANDRA STUCKEY, LEE ANNE SMALLEY, and GLENDA DORMAN A method of measuring trunk motion and two related motions using a tape measure

More information

Reliability of Lower Extremity Functional Performance Tests

Reliability of Lower Extremity Functional Performance Tests Reliability of Lower Extremity Functional Performance Tests Lori A. Bolgla, PT' Douglas R. Keskula, Ph D, PT, ATC2 ports medicine clinicians commonly see athletes who have sustained a lower extremity injury

More information

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75 The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75 By: Brent L. Arnold and David H. Perrin * Arnold, B.A., & Perrin, D.H. (1993).

More information

Accuracy and validity of Kinetisense joint measures for cardinal movements, compared to current experimental and clinical gold standards.

Accuracy and validity of Kinetisense joint measures for cardinal movements, compared to current experimental and clinical gold standards. Accuracy and validity of Kinetisense joint measures for cardinal movements, compared to current experimental and clinical gold standards. Prepared by Engineering and Human Performance Lab Department of

More information

Goniometric Reliability in a Clinical Setting

Goniometric Reliability in a Clinical Setting Goniometric Reliability in a Clinical Setting Elbow and Knee Measurements JULES M. ROTHSTEIN, PETER J. MILLER, and RICHARD F. ROETTGER Reliability of goniometric measurements has been examined only under

More information

SPINAL IMMOBILIZATION

SPINAL IMMOBILIZATION Spinal Immobilization Decision Assessment Recent studies have shown an increase in mortality for patients with isolated penetrating trauma who are spinally immobilized. Therefore spinal immobilization

More information

FUNDAMENTAL SEATING PRINCIPLES Power Point PDF Bengt Engström Physiotherapist. Concept ENGSTRÖM

FUNDAMENTAL SEATING PRINCIPLES Power Point PDF Bengt Engström Physiotherapist. Concept ENGSTRÖM FUNDAMENTAL SEATING PRINCIPLES Power Point PDF Bengt Engström Physiotherapist Starting with a few questions! How are your clients sitting? What kind of problems do you see? How long time are your clients

More information

The Trunk and Spinal Column Kinesiology Cuneyt Mirzanli Istanbul Gelisim University

The Trunk and Spinal Column Kinesiology Cuneyt Mirzanli Istanbul Gelisim University The Trunk and Spinal Column Kinesiology Cuneyt Mirzanli Istanbul Gelisim University The Trunk and Spinal Column Vertebral column 24 articulating vertebrae 31 pairs of spinal nerves Abdominal muscles some

More information

Journal of Rehabilitation Sciences and Research. Evaluation of the Efficiency of Minerva Collar on Cervical Spine Motions

Journal of Rehabilitation Sciences and Research. Evaluation of the Efficiency of Minerva Collar on Cervical Spine Motions JRSR 2 (2017) 47-52 Journal of Rehabilitation Sciences and Research Journal Home Page: jrsr.sums.ac.ir Original Article Evaluation of the Efficiency of Minerva Collar on Cervical Spine Motions Mohammad

More information

Thoracic Spine Management. Jason Zafereo, PT, OCS, FAAOMPT

Thoracic Spine Management. Jason Zafereo, PT, OCS, FAAOMPT Thoracic Spine Management Jason Zafereo, PT, OCS, FAAOMPT Clinical i l Orthopedic Rehabilitation ti Education Objectives Describe the treatment interventions used for the management of pain from contractile

More information

THE PRO-POSTURE NECK EXERCISER

THE PRO-POSTURE NECK EXERCISER EXERCISE INSTRUCTIONS FOR THE PRO-POSTURE NECK EXERCISER The following are a list of six different exercises that can be performed with the Pro- Posture Neck Exerciser. 1) Basic Pro-Posture Neck Pump Exercise

More information

A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr.

A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr. A Comparison of the Immediate Effects of Eccentric Training vs Static Stretch on Hamstring Flexibility in Basketball Players Dr. Mandeep Thour* *Assistant Professor, Department of Physical Education SGGS

More information

NEO-FLEX POSTURE CORRECTION AND SUPPORT SYSTEM

NEO-FLEX POSTURE CORRECTION AND SUPPORT SYSTEM NEO-FEX POTURE AND UPPORT YTE QUAITY Trunk stabilization EUROPE O p e n i n g u p n e w h o r i z o n s i n m o b i l i t y N.V. Vermeiren N.V. Vermeirenplein 1-15 B-2920 Kalmthout Belgium Tel. : +32 (0)3

More information

ZOROMED TANGENT INFRA LAB TECH. Orthopedic Rehabilitation Aid Products. Tangent Infra Lab Tech 218 B, Ring Road Mall, Sec 3 Rohini, New Delhi

ZOROMED TANGENT INFRA LAB TECH. Orthopedic Rehabilitation Aid Products. Tangent Infra Lab Tech 218 B, Ring Road Mall, Sec 3 Rohini, New Delhi TANGENT INFRA LAB TECH ZOROMED Orthopedic Rehabilitation Aid Products Our range of orthopedic rehabilitation products includes Cervical, Knee, Chest, Back, Ankle and Walking aids. We also have healthcare

More information

NECK & SHOULDER. mediroyal.se

NECK & SHOULDER. mediroyal.se NECK & SHOULDER mediroyal.se 2 Neck & Shoulder MR990 ERIXONE mediroyal.se facebook.com/mediroyalnordic Neck & Shoulder 3 MR990 ErixOne The shoulder joint is one of the most advanced joints in the body.

More information

Gender Based Influences on Seated Postural Responses

Gender Based Influences on Seated Postural Responses Gender Based Influences on Seated Postural Responses Jack P. Callaghan PhD Canada Research Chair in Spine Biomechanics and Injury Prevention Department of Kinesiology Faculty of Applied Health Sciences

More information

Obesity is associated with reduced joint range of motion (Park, 2010), which has been partially

Obesity is associated with reduced joint range of motion (Park, 2010), which has been partially INTRODUCTION Obesity is associated with reduced joint range of motion (Park, 2010), which has been partially attributed to adipose tissues around joints limiting inter-segmental rotations (Gilleard, 2007).

More information

Posture. Kinesiology RHS 341 Lecture 10 Dr. Einas Al-Eisa

Posture. Kinesiology RHS 341 Lecture 10 Dr. Einas Al-Eisa Posture Kinesiology RHS 341 Lecture 10 Dr. Einas Al-Eisa Posture = body alignment = the relative arrangement of parts of the body Changes with the positions and movements of the body throughout the day

More information

Abstract. A novel device to improve sitting posture. Hoda Dalimi (1) Ali Ghorbani (2) Anoushirvan Kezam nejad (3) Mohammad Hossein Alizadeh (4)

Abstract. A novel device to improve sitting posture. Hoda Dalimi (1) Ali Ghorbani (2) Anoushirvan Kezam nejad (3) Mohammad Hossein Alizadeh (4) A novel device to improve sitting posture Hoda Dalimi (1) Ali Ghorbani (2) Anoushirvan Kezam nejad (3) Mohammad Hossein Alizadeh (4) (1) Msc of Sports Science, University of Tehran, Tehran, Iran (2) Msc

More information

Downloaded from umj.umsu.ac.ir at 22: on Friday March 22nd 2019

Downloaded from umj.umsu.ac.ir at 22: on Friday March 22nd 2019 * 91/06/07 : 91/04/02 : :.. -. :. ICC. ICC.. :. // // :. : - - : :.(-) - (SLR) Email: eterafoskouei@tbzmed.ac.ir.(). (ULTT).().. ( ) Straight Leg Raise Upper Limb Tension Tests .().(-) Hough.().() ULTT1.()

More information

CERVICAL STRAIN AND SPRAIN (Whiplash)

CERVICAL STRAIN AND SPRAIN (Whiplash) CERVICAL STRAIN AND SPRAIN (Whiplash) Description time and using proper technique decrease the frequency of Whiplash is an injury to the neck caused when it is forcefully whipped or forced backward or

More information

DEEP SQUAT. Upper torso is parallel with tibia or toward vertical Femur below horizontal Knees are aligned over feet Dowel aligned over feet

DEEP SQUAT. Upper torso is parallel with tibia or toward vertical Femur below horizontal Knees are aligned over feet Dowel aligned over feet APPENDIX 9 SCORING CRITERIA DEEP SQUAT Upper torso is parallel with tibia or toward vertical Femur below horizontal Knees are aligned over feet Dowel aligned over feet Upper torso is parallel with tibia

More information

Scapular Muscle Strengthening

Scapular Muscle Strengthening Original Research Journal of Sport Rehabilitation, 1995, 4, 244-252 O 1995 Human Kinetics Publ~shers, Inc. Scapular Muscle Strengthening Thomas Zmierski, Sam Kegerreis, and James Scarpaci The purposes

More information

Using Biomechanical Principles in the Management of Complex Postural Deviations in Sitting

Using Biomechanical Principles in the Management of Complex Postural Deviations in Sitting Using Biomechanical Principles in the Management of Complex Postural Deviations in Sitting ISS 2016 / Part 2 Managing Complex Postures KELLY WAUGH, PT, MAPT, ATP S E N I O R I N S T R U C TO R / C L I

More information

Active-Assisted Stretches

Active-Assisted Stretches 1 Active-Assisted Stretches Adequate flexibility is fundamental to a functional musculoskeletal system which represents the foundation of movement efficiency. Therefore a commitment toward appropriate

More information

Increase Strength, Flexibility, Range of Motion and Endurance. NeckXsystems Guidelines & Exercise. NeckX is Patent Pending

Increase Strength, Flexibility, Range of Motion and Endurance. NeckXsystems Guidelines & Exercise.   NeckX is Patent Pending Increase Strength, Flexibility, Range of Motion and Endurance NeckXsystems Guidelines & Exercise Instructions www.neckxsystems.com NeckX is Patent Pending Guidelines NeckX is an innovative neck exercise

More information

VIRGINIA ORTHOPEDIC MANUAL PHYSICAL THERAPY INSTITUTE TECHNIQUE MANUAL

VIRGINIA ORTHOPEDIC MANUAL PHYSICAL THERAPY INSTITUTE TECHNIQUE MANUAL VIRGINIA ORTHOPEDIC MANUAL PHYSICAL THERAPY INSTITUTE TECHNIQUE MANUAL Lumbar and Thoracic Spine Lumbar AROM Assessment -Patient Positioning: Standing, appropriately undressed so that the lumbar and thoracic

More information

Improving Thoracic Mobility

Improving Thoracic Mobility Improving Thoracic Mobility By William J. Hanney DPT, PhD, ATC, CSCS Course Description A lack of thoracic mobility can have broad clinical implications and evidence suggests addressing mobility in this

More information

Validity Study of Head and Neck Flexion-Extension Motion Comparing Measurements of a Pendulum Goniometer and Roentgenograms1

Validity Study of Head and Neck Flexion-Extension Motion Comparing Measurements of a Pendulum Goniometer and Roentgenograms1 Validity Study of Head and Neck Flexion-Extension Motion Comparing Measurements of a Pendulum Goniometer and Roentgenograms1 DENYSE BREDENKAMP HERRMANN, MHS, P? Physical therapists measure head and neck

More information

Development of a Novel Biofedelic Skull-Neck- Thorax Model Capable of Quantifying Motions of aged Cervical Spine

Development of a Novel Biofedelic Skull-Neck- Thorax Model Capable of Quantifying Motions of aged Cervical Spine Yale University EliScholar A Digital Platform for Scholarly Publishing at Yale Yale Medicine Thesis Digital Library School of Medicine 11-23-2009 Development of a Novel Biofedelic Skull-Neck- Thorax Model

More information

erhem Thoracal spine section Elastic shoulder brace with support bracing Application:

erhem Thoracal spine section Elastic shoulder brace with support bracing Application: ERH 53/1 Elastic shoulder brace with support bracing Augmentation of proper upright posture Stable fractures of the clavicle Scheuermann s kyphosis Funnel chest Osteoporotic lesions The shoulder brace

More information

IS THERE A LINK BETWEEN SPINE AND HIP MOBILITY?

IS THERE A LINK BETWEEN SPINE AND HIP MOBILITY? EXERCISE AND QUALITY OF LIFE Volume 4, No. 2, 2012, 1-5 UDC 796.012.23 Research article IS THERE A LINK BETWEEN SPINE AND HIP MOBILITY? Miroslav Saviè and S2P, Laboratory for Motor Control and Motor Learning,

More information

Respiration & Trunk control The Great Connection. Brief Review of Normal Development of the Rib Cage

Respiration & Trunk control The Great Connection. Brief Review of Normal Development of the Rib Cage Respiration & Trunk control The Great Connection. are part of a complex combination of interactive systems. Muscles of respiration are part of the musculature of dynamic postural control. First 3 Years

More information

Lumbo TriStep. Three Steps to More Mobility. Information for Practicioners

Lumbo TriStep. Three Steps to More Mobility. Information for Practicioners Lumbo TriStep Three Steps to More Mobility Information for Practicioners 3 Phases for the Rehabilitation Process Lumbo TriStep is a bridging orthosis with a mobilisation function. Its modular design adapts

More information

"Zero-Position" Functional Shoulder Orthosis for Postoperative. management of rotator cuff injuries.

Zero-Position Functional Shoulder Orthosis for Postoperative. management of rotator cuff injuries. "Zero-Position" Functional Shoulder Orthosis for Postoperative Management of Rotator Cuff Injuries Jiro Ozaki, M.D. Ichiro Kawamura INTRODUCTION Many shoulder orthoses such as the airplane splint, the

More information

Reliability of Smartphone Inclinometry to Measure Upper Cervical Range of Motion.

Reliability of Smartphone Inclinometry to Measure Upper Cervical Range of Motion. Sacred Heart University DigitalCommons@SHU Physical Therapy Faculty Publications Physical Therapy & Human Movement Science 7-2018 Reliability of Smartphone Inclinometry to Measure Upper Cervical Range

More information

Reliability of Measurements of Lumbar Spine. Flexible Curve. lames W. Youdas, MS, PT' Vera ). Suman, PhD * Tom R. Garrett, PT3

Reliability of Measurements of Lumbar Spine. Flexible Curve. lames W. Youdas, MS, PT' Vera ). Suman, PhD * Tom R. Garrett, PT3 Reliability of Measurements of Lumbar Spine Sagittal Mobility Obtained With the Flexible Curve lames W. Youdas, MS, PT' Vera ). Suman, PhD * Tom R. Garrett, PT3 hysical therapists are interested in quantifying

More information

Scapulothoracic muscle strength in individuals with neck pain

Scapulothoracic muscle strength in individuals with neck pain Journal of Back and Musculoskeletal Rehabilitation 29 (2016) 549 555 549 DOI 10.3233/BMR-160656 IOS Press Scapulothoracic muscle strength in individuals with neck pain Shannon M. Petersen a,,nathana.domino

More information

CERVICAL STRAIN AND SPRAIN

CERVICAL STRAIN AND SPRAIN CERVICAL STRAIN AND SPRAIN Description Cervical strain/sprain is an injury to the neck caused when it is forcefully whipped or forced backward or forward. The structures involved are the muscles, ligaments,

More information

CORE STABILIZATION EXERCISE. Vincent J. Hudson, PhD, DPT, MBA, ATC Chief Operating Officer OAA Orthopaedic Specialists Allentown, PA

CORE STABILIZATION EXERCISE. Vincent J. Hudson, PhD, DPT, MBA, ATC Chief Operating Officer OAA Orthopaedic Specialists Allentown, PA CORE STABILIZATION EXERCISE Vincent J. Hudson, PhD, DPT, MBA, ATC Chief Operating Officer OAA Orthopaedic Specialists Allentown, PA Stepping Back What are the goals? What is the value? Is it Sport Specific

More information

Life Changing Spinal Innovation. Product Catalog. Upper & Lower Spinal Orthotics

Life Changing Spinal Innovation. Product Catalog. Upper & Lower Spinal Orthotics Life Changing Spinal Innovation Product Catalog Upper & Lower Spinal Orthotics OVER 25 YEARS of Research & Development Aspen Medical Products is a leader in the development of innovative spinal bracing

More information

EFFECIVENESS OF THE WILLIAMS EXCERCISE IN MECHANICAL LOW BACK PAIN

EFFECIVENESS OF THE WILLIAMS EXCERCISE IN MECHANICAL LOW BACK PAIN EFFECIVENESS OF THE WILLIAMS EXCERCISE IN MECHANICAL LOW BACK PAIN Dr.U.Ganapathy Sankar, Ph.D Dean I/C, SRM College of Occupational Therapy, SRMUniversity, Kattankulathur, KancheepuramDistrict, Tamil

More information

UPPER BACK PAIN. Contents What causes Upper Back Pain?... 3

UPPER BACK PAIN. Contents What causes Upper Back Pain?... 3 Contents What causes Upper Back Pain?............. 3 What treatment can I receive?.............. 5 YOUR GUIDE TO UPPER BACK PAIN An IPRS Guide to provide you with exercises and advice to ease your condition

More information

Steeper Essentials New Product Update

Steeper Essentials New Product Update Steeper Essentials New Product Update Steeper Universal PROGLIDE Collars The PROGLIDE collection is a complete solution for individuals with a cervical injury that requires the use of a collar for an extended

More information

Analysis of a Modified Active Knee Extension Test

Analysis of a Modified Active Knee Extension Test Analysis of a Modified Active Knee Extension Test Yvonne Kane, MS, PT' lay Bernasconi, MS, PT2 P hysical therapists frequently assess hamstring muscle length (HML) in patients with hamstring injuries,

More information

Rehabilitation Guidelines for Labral/Bankert Repair

Rehabilitation Guidelines for Labral/Bankert Repair Rehabilitation Guidelines for Labral/Bankert Repair The true shoulder joint is called the glenohumeral joint and consists humeral head and the glenoid. It is a ball and socket joint. Anatomy of the Shoulder

More information

FUSE TECHNICAL REPORT

FUSE TECHNICAL REPORT FUSE TECHNICAL REPORT 1 / 16 Contents Page 3 Page 4 Page 8 Page 10 Page 13 Page 16 Introduction FUSE Accuracy Validation Testing LBD Risk Score Model Details FUSE Risk Score Implementation Details FUSE

More information

The Effects of Posture on Neck Flexion Angle While Using a Smartphone according to Duration

The Effects of Posture on Neck Flexion Angle While Using a Smartphone according to Duration J Korean Soc Phys Med, 2016; 11(3): 35-39 http://dx.doi.org/10.13066/kspm.2016.11.3.35 Online ISSN: 2287-7215 Print ISSN: 1975-311X Research Article Open Access The Effects of Posture on Neck Flexion Angle

More information

Chicago Chiropractic & Sports Injury Centers Dr. Alden Clendenin DC, CCSP

Chicago Chiropractic & Sports Injury Centers Dr. Alden Clendenin DC, CCSP Cervical Lordotic Traction The soft forward [lordotic] curve at the middle of the cervical spine is critical as a shock absorber for the full weight of the head. Loss of cervical lordosis, often called

More information

SpineCor Scoliosis Treatment for Children and Adults

SpineCor Scoliosis Treatment for Children and Adults SpineCor Scoliosis Treatment for Children and Adults SpineCor Treatment Initially a dynamic bracing solution for Adolescent Idiopathic Scoliosis, SpineCor treatment has evolved over the past 20 years and

More information

DUAL POSITION BACK EX/FLEX ATTACHMENT

DUAL POSITION BACK EX/FLEX ATTACHMENT DUAL POSITION BACK EX/FLEX ATTACHMENT OPERATION MANUAL 830-450 FN: 08-261 Rev A 11/15 BIODEX Biodex Medical Systems, Inc. 20 Ramsey Road, Shirley, New York, 11967-4704 Tel: 800-224-6339 (Int l 631-924-9000),

More information

Adolescent Idiopathic Scoliosis

Adolescent Idiopathic Scoliosis Adolescent Idiopathic Scoliosis Adolescent idiopathic scoliosis is characterized by a lateral bending and twisting of the spine. It is the most common spinal deformity affecting adolescents 10 to 16 years

More information

BIOMECHANICAL ANALYSIS OF THE DEADLIFT DURING THE 1999 SPECIAL OLYMPICS WORLD GAMES

BIOMECHANICAL ANALYSIS OF THE DEADLIFT DURING THE 1999 SPECIAL OLYMPICS WORLD GAMES 63 Biomechanics Symposia 2001 / University of San Francisco BIOMECHANICAL ANALYSIS OF THE DEADLIFT DURING THE 1999 SPECIAL OLYMPICS WORLD GAMES Rafael F. Escamilla, Tracy M. Lowry, Daryl C. Osbahr, and

More information

TRAINING THE CORE BEGIN WITH ONE SET OF ALL 17 EXERCISES FOR A TOTAL OF 250 REPS. NEXT, MOVE TO TWO SETS FOR A TOTAL OF 500 REPS.

TRAINING THE CORE BEGIN WITH ONE SET OF ALL 17 EXERCISES FOR A TOTAL OF 250 REPS. NEXT, MOVE TO TWO SETS FOR A TOTAL OF 500 REPS. TRAINING THE CORE 1. LATERAL SIT UPS.X 20 (10 EACH SIDE) 2. HYPEREXTENSIONS.X 10 3. LEG HUGS...X 15 4. RUSSIAN TWIST X 20 (10 EACH SIDE) 5. HIP CURLS..X 14 (7 EACH LEG) 6. JACK KNIFES..X 10 7. REVERSE

More information

Manual Muscle Testing. Yasser Moh. Aneis, PhD, MSc., PT. Lecturer of Physical Therapy Basic Sciences Department

Manual Muscle Testing. Yasser Moh. Aneis, PhD, MSc., PT. Lecturer of Physical Therapy Basic Sciences Department Manual Muscle Testing Yasser Moh. Aneis, PhD, MSc., PT. Lecturer of Physical Therapy Basic Sciences Department Manual Muscle Testing Evaluation of the function and strength of individual muscles and muscles

More information

Core Stabilization for a Pain- Free Posture

Core Stabilization for a Pain- Free Posture PAIN-FREE POSTURE PROGRAM Core Stabilization for a Pain- Free Posture with Mary Ann Foster ABOUT MARY ANN FOSTER Massage therapist since 1981 Somatic educator and movement teacher Author of Somatic Patterning

More information

A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total Daily Stretch Duration

A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total Daily Stretch Duration Journal of Strength and Conditioning Research, 2003, 17(2), 274 278 2003 National Strength & Conditioning Association A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total

More information

OMT Without An OMT Table Workshop. Dennis Dowling, DO FAAO Ann Habenicht, DO FAAO FACOFP

OMT Without An OMT Table Workshop. Dennis Dowling, DO FAAO Ann Habenicht, DO FAAO FACOFP OMT Without An OMT Table Workshop Dennis Dowling, DO FAAO Ann Habenicht, DO FAAO FACOFP Cervical Somatic Dysfunction (C5 SR RR) - Seated 1. Patient position: seated. 2. Physician position: standing facing

More information

2002 Physioball Supplement

2002 Physioball Supplement 2002 Physioball Supplement These exercises are not detailed on the 2002 Off-Ice Training video but will be taught in detail during the 2002 Reach for the Stars Seminar. CORE STRENGTH Physioball/ Sport

More information

Wunzi Strapping Applications

Wunzi Strapping Applications Wunzi Strapping Applications In addition to the basic garment set(s), all Wunzi product configurations now include limb cuff and strapping options that attending therapists can use to achieve higher levels

More information

Spine, October 1, 2003; 28(19): Eythor Kristjansson, Gunnar Leivseth, Paul Brinckmann, Wolfgang Frobin

Spine, October 1, 2003; 28(19): Eythor Kristjansson, Gunnar Leivseth, Paul Brinckmann, Wolfgang Frobin Increased Sagittal Plane Segmental Motion in the Lower Cervical Spine in Women With Chronic Whiplash-Associated Disorders, Grades I-II: A Case-Control Study Using a New Measurement Protocol 1 Spine, October

More information

A Patient s Guide to Dropped Head Syndrome

A Patient s Guide to Dropped Head Syndrome A Patient s Guide to Dropped Head Syndrome 228 West Main, Suite C Missoula, MT 59802 Phone: info@spineuniversity.com DISCLAIMER: The information in this booklet is compiled from a variety of sources. It

More information

Raymond Wiegand, D.C. Spine Rehabilitation Institute of Missouri

Raymond Wiegand, D.C. Spine Rehabilitation Institute of Missouri 2D Pattern matching of frontal plane radiograph to 3D model identifies structural and functional deficiencies of the spinal pelvic system in consideration of mechanical spine pain (AKA Spine distortion

More information

The Bambach Saddle Seat in rehabilitation

The Bambach Saddle Seat in rehabilitation 7 The Bambach Saddle Seat in rehabilitation The Musculo-skeletal System Good design recognises that our body has a centre of gravity (as does each limb) and maintaining posture close to the neutral centre

More information

DIAGNOSTIC VIDEOFLUOROSCOPY IMPRESSIONS and BIOMECHANICS REPORT

DIAGNOSTIC VIDEOFLUOROSCOPY IMPRESSIONS and BIOMECHANICS REPORT P.O. Box 6743 New Albany, IN 47151-6743 (812) 945-5515 (812) 945-5632 Fax WWW.KMX.CC DIAGNOSTIC VIDEOFLUOROSCOPY IMPRESSIONS and BIOMECHANICS REPORT Patient Name: Lubna Ibriham Date of Digitization and

More information

Lumbar Stenosis Rehabilitation Using the Resistance Chair

Lumbar Stenosis Rehabilitation Using the Resistance Chair PRODUCTS HELPING PEOPLE HELP THEMSELVES! Lumbar Stenosis Rehabilitation Using the Resistance Chair a. Description Lumbar spinal stenosis is a term used to describe a narrowing of the spinal canal. The

More information

Treatment of SI dysfunction

Treatment of SI dysfunction + SIJ dysfunction 84 Treatment of SI dysfunction Move innominant into anterior tilt Decreased pain Perform manipulation for correction of a posterior innominant Increased pain/no change Move innominant

More information

Range of global motion of the cervical spine: intraindividual reliability and the influence of measurement device

Range of global motion of the cervical spine: intraindividual reliability and the influence of measurement device Eur Spine J (2000) 9 :379 385 Springer-Verlag 2000 ORIGINAL ARTICLE A. F. Mannion G. N. Klein J. Dvorak C. Lanz Range of global motion of the cervical spine: intraindividual reliability and the influence

More information

Dynamic Splinting of the

Dynamic Splinting of the Dynamic Splinting of the Rheumatoid Hand BY F. RICHARD CONVERY, M.D.,* J. PIERCE CONATY, M.D.** AND VERNON L. NICKEL, M.D.** Rancho Los Amigos Hospital, Downey, California (University of Southern California

More information

Chapter 4B: Reliability Trial 2. Chapter 4B. Reliability. General issues. Inter-rater. Intra - rater. Test - Re-test

Chapter 4B: Reliability Trial 2. Chapter 4B. Reliability. General issues. Inter-rater. Intra - rater. Test - Re-test Chapter 4B: Reliability Trial 2 Chapter 4B Reliability General issues Inter-rater Intra - rater Test - Re-test Trial 2 The second clinical trial was conducted in the spring of 1992 using SPCM Draft 6,

More information

Joint Range of Motion Assessment Techniques. Presentation Created by Ken Baldwin, M.Ed Copyright

Joint Range of Motion Assessment Techniques. Presentation Created by Ken Baldwin, M.Ed Copyright Joint Range of Motion Assessment Techniques Presentation Created by Ken Baldwin, M.Ed Copyright 2001-2006 Objectives Understand how joint range of motion & goniometric assessment is an important component

More information

CLINICAL NOTES. Aust. J. Physiother. 26:5, October,

CLINICAL NOTES. Aust. J. Physiother. 26:5, October, CLINICAL NOTES COMBINED MOVEMENTS IN THE CERVICAL SPINE (C2-7) THEIR VALUE IN EXAMINATION AND TECHNIQUE CHOICE Brian C. Edwards Private Practitioner Lecturer in Manipulative Therapy, Western Australian

More information

Back Safety Healthcare #09-066

Back Safety Healthcare #09-066 Back Safety Healthcare Version #09-066 I. Introduction A. Scope of training This training program applies to healthcare employees whose job requires them to lift patients or other heavy objects. Lifting

More information

Dynamic Flexibility All exercises should be done smoothly while taking care to maintain good posture and good technique.

Dynamic Flexibility All exercises should be done smoothly while taking care to maintain good posture and good technique. Dynamic Flexibility All exercises should be done smoothly while taking care to maintain good posture and good technique. Lying on back: Hip Crossover: Arms out in T position, feet flat on the floor, knees

More information

Cervical Spine: Pearls and Pitfalls

Cervical Spine: Pearls and Pitfalls Cervical Spine: Pearls and Pitfalls Presenters Dr. Rob Donkin Functional Anatomy Current research Cervical Radiculopathy Dr. Gert Ferreira Red flags Case Study Kinesio Taping Chris Neethling Gonstead adjusting

More information

THE VAPC LUMBOSACRAL ORTHOSIS

THE VAPC LUMBOSACRAL ORTHOSIS THE VAPC LUMBOSACRAL ORTHOSIS Gustav Rubin, M.D., F.A.C.S.1, Werner Greenbaum, C.P.O. 2, and Dave Molack 3 Long-established principles and new concepts have been combined in the design of the Veterans

More information

ANTERIOR KNEE PAIN. Explanation. Causes. Symptoms

ANTERIOR KNEE PAIN. Explanation. Causes. Symptoms ANTERIOR KNEE PAIN Explanation Anterior knee pain is most commonly caused by irritation and inflammation of the patellofemoral joint of the knee (where the patella/kneecap connects to the femur/thigh bone).

More information

Shoulder. 36 Shoulder medi orthopaedics

Shoulder. 36 Shoulder medi orthopaedics Shoulder 36 Shoulder medi orthopaedics medi SAS multi Dual purpose 15 abduction / external rotation support post-operative immobilisation following: rotator cuff ruptures humeral head fractures prosthetic

More information

The amount of active range of motion (AROM) for forearm

The amount of active range of motion (AROM) for forearm Reliability of 2 Functional Goniometric Methods for Measuring Forearm Pronation And Supination Active Range of Motion Christos Karagiannopoulos, MPT, MEd, ATC 1 Michael Sitler, EdD, ATC 2 Susan Michlovitz,

More information

DON T JUST PROVIDE A BAND-AID ELIZABETH COLE, MSPT, ATP U.S. Rehab / VGM

DON T JUST PROVIDE A BAND-AID ELIZABETH COLE, MSPT, ATP U.S. Rehab / VGM DON T JUST PROVIDE A BAND-AID ELIZABETH COLE, MSPT, ATP U.S. Rehab / VGM Selecting the most appropriate product solutions for specific postural problems should be a thoughtful exercise in analyzing the

More information

Wheelchair Seating: Are We Speaking the Same Language?

Wheelchair Seating: Are We Speaking the Same Language? Universal Language Wheelchair Seating: Are We Speaking the Same Language? Andrée Gauthier, OT Reg. (Ont.) Seating Clinic, Lyndhurst Centre Toronto Rehab Institute University Health Network andree.gauthier@uhn.ca

More information

The Influence of Spinal Deformities on Acetabular Orientation in Total Hip Arthroplasty

The Influence of Spinal Deformities on Acetabular Orientation in Total Hip Arthroplasty The Influence of Spinal Deformities on Acetabular Orientation in Total Hip Arthroplasty S. SAMUEL BEDERMAN MD PhD FRCSC Scoliosis & Spine Tumor Center S. SAMUEL BEDERMAN MD PhD FRCSC disclosures October

More information

Test instrument for predicting the effect of rigid braces in cases with low back pain

Test instrument for predicting the effect of rigid braces in cases with low back pain Prosthetics and Orthotics International, 1990, 14, 22-26 Test instrument for predicting the effect of rigid braces in cases with low back pain S. W. WILLNER Department of Orthopaedics, Malmö General Hospital,

More information

BACK SPASM. Explanation. Causes. Symptoms

BACK SPASM. Explanation. Causes. Symptoms BACK SPASM Explanation A back spasm occurs when the muscles of the back involuntarily contract due to injury in the musculature of the back or inflammation in the structural spine region within the discs

More information

Spinomed IV. Proven solutions for over 10 million Americans living with osteoporosis 1

Spinomed IV. Proven solutions for over 10 million Americans living with osteoporosis 1 Spinomed IV Proven solutions for over 10 million Americans living with osteoporosis 1 medi. I medi. feel I feel better. Real Problems Osteoporosis Osteoporosis, also known as bone loss, affects 1 in 4

More information

PTA 210 PTA Techniques

PTA 210 PTA Techniques PTA 210 PTA Techniques Patient Positioning and Draping INTRO Patient Positioning Must be considered before, during and after treatment AND when the patient is to be at rest for a prolonged period of time

More information