Rigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device. At-Home Usability and Interpretability Studies for Consumers
|
|
- Laureen Jones
- 5 years ago
- Views:
Transcription
1 Rigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device At-Home Usability and Interpretability Studies for Consumers
2 INTRO Current state of technology indicates that the gold standard mechanisms used by Labcorp, Quest Diagnostics, and Theranos produce results outside clinically defined ranges 8.3%, 7.5%, and 12.2% of the time for LabCorp, Quest, and Theranos, respectively. 1 These data points carry implications on precision and accuracy - namely, it is possible to get gold standard venous CBC results with +/ % differences between the same sample if run twice. As such, a determination was made during the Athelas Clinical Study that the existing gold standard in-lab venous precision was not the most stringent way to characterize accuracy. Additional bench studies were conducted with synthetically concentrated cellular samples of target tranches. Furthermore, consumer-grade usability studies were conducted to highlight the accuracy of the end-to-end device and strip usage in a raw, at-home setting. These studies focused on the accuracy and precision of the model when used by a completely untrained individual with no background. This study found that the Athelas system produced n=0 results outside of clinical range, maintaining reproducibility of results over 20 replicates, and on multiples devices. Further, when users (n=37) attempted to use the Athelas system with only the in-app manual, and paper manual present, all users were able to produce a result that was clinically significant, benchmarked with the Labcorp results generated by a clinician indicating that the Athelas device is understandable and easy to use by an untrained individual. BACKGROUND The purpose of these bench studies is twofold: to exhaustively determine the sensitivity of the Athelas device in various cohorts, and to determine the usability and interpretability of the device when tested in home with only the in-app instructions and the paper manual. Often in diagnostics, accuracies drop when going from clinical studies and in-lab data to the real world with real users. To assess and characterize the performance of the Athelas platform in this end-user setting, Athelas organized a trial focused on patients seeing the device for the first time. 1 J Clin Invest. 2016;126(5): doi: /jci
3 METHODS Bench studies were conducted to test the precision of the test strips, inter device precision, and sensitivity. Additional at home studies were also conducted across a range of demographic groups to characterize the usability of the device and strips and the interpretability of the results. The clinical gold standards2 were in many ways deemed unfit to fully characterize and define the accuracy of the Athelas system due to their own inherent variation. As such the Athelas study made use of synthetic PBMCs3 with tranched concentrations to synthetically create the test samples. Table 1. Collected by B. Kidd in Evaluation of direct to consumer low volume lab tests in healthy adults. J Clin Invest. 2016;126(5): doi: /jci86318 J Clin Invest. 2016;126(5): doi: /jci Kirchner, H., Ch Kleinicke, and W. Digel. "A whole-blood technique for testing production of human interferon by leukocytes." Journal of immunological methods 48.2 (1982):
4 Bench Study - Sensitivity, Precision, Aging The objective of this study was to establish the measurement accuracy stability of the Athelas system as samples aged. A parent vial of 10^8 white blood cells was diluted to create solutions of seven different concentrations (2k, 4k, 5.5k, 9k, 11k, 20k WBC/uL). 4 ul of solution 1 was loaded into 20 test strips. After 5 minutes, each strip was inserted into the machine and a result was obtained and recorded. The same strips were inserted into a second machine and another set of results were obtained and recorded. The procedure was repeated for each solution. At Home Testing - Usability & Interpretability The objective of the study was to demonstrate the accuracy and interpretability of the Athelas System when used by patients. The patients were given the Athelas app (Figure 1) and an instruction manual. Figure 1. Athelas in app instructional gif The patients were asked to perform their own tests from start to completion with no assistance from the Athelas team. The patient's WBC count was compared using the Athelas device versus the results generated by the LabCorp system (Beckman Coulter UniCel DxH 600/800 COulter Cellular Analysis System). Additionally, the patient was asked to report how easy it was to understand the Athelas instructions (Table 2) and the final results (Figure 2). 4
5 A. Did the user understand where the strips were located? B. Did the user understand how to fill the strip? C. Did the user fill the test strip properly? D. Were they able to place it in the correct slot in the device? E. Are they able to perform a successful test? F. If no, was it clear to the user that the test was unsuccessful? G. If no to 1, was it clear that the user should perform another test after their unsuccessful test? H. Did the user perform another test after their unsuccessful test? I. How many times did the user perform a test to ensure a proper reading? Indicate agreement on a 1-5 scale (1=strongly disagree; 5=strongly agree). 1. The instructions were easy to follow. 2. It was easy to apply the sample correctly. 3. It was easy to see and understand the test results (e.g., appearance of the line, change of color). 4. The instructions clearly explain what to do if a test result does not appear or is invalid. 5. I needed help from someone the first time I ran the test. 6. If I did not perform the test properly, I was sure that I had to repeat it. 7. It was clear that this test did not provide a diagnosis. 8. If I performed the test correctly, I knew that I had done it correctly. 1. How difficult is it to read the Labcorp results 1. How difficult is it to read the Athelas results What was the Labcorp Count? What was the Athelas Count? Table 2. Athelas questionnaire to determine patient usability and interpretability. The study was conducted at 275 Moffett Blvd. Mountain View, CA and LabCorp (LabCorp San Jose, 266 N Jackson Ave Ste 4 San Jose, CA 95116). The study plan and proposal was approved by the institutional review board (IRB) at Salus IRB. All participants signed an informed consent form that detailed the objective of the study, the sponsor of the study, the procedures of the study, the right to opt-out, the anonymization of results, and how their anonymized results may be used. 5
6 RESULTS Results from Bench Studies The method used for the bench studies supports that the Athelas system generates accurate results throughout the indicated range (2k-20K WBC/uL) and in the vicinity of cut-off thresholds (Figure 3) with minimal confusion in the border ranges, particularly in the ranges between 4K WBC/uL and 5.5K WBC/uL. This confusion can most likely be attributed to natural (<10%) variation in WBC concentrations. More rigorous testing precision will be done to validate this result. Athelas WBC Consistent Over 20 Trials Across 2K-22K WBC/uL Figure 3. Athelas Bench Study Data conducted at Stanford University, collected on April 3, The methods used has been demonstrated to be linear between 2k/uL-20k/uL within no different in 2k/uL and within no difference in 20k/uL. Setup, takedown, and testing at different times does not affect the results generated by the system. The results also indicate that testing the strip on different devices yields a correlation coefficient > r^2=.8 (Figure 4). 6
7 Inter Device Precision Over 20 Trials Across 2K-22K WBC/uL Figure 4. Athelas Bench Study Data conducted at Stanford University, collected on April 3, Results from At-Home Testing Of the 37 users, aged (51% female, 49% male), three male users (age=17, 69, 71) had to redo the test as they received the N/A screen and 34 users were able to generate their results on their first try. Of the three users who were not able to perform the test on the first attempt, all of them indicated they had done the test incorrectly. Upon second attempt, the three users were able to generate their results. The results generated by the users had an r^2 = with the results obtained by Labcorp for their corresponding sample (Figure 5). 7
8 Labcorp WBC Count vs. User Generated Athelas WBC Count Figure 5. Results of Athelas At-Home Testing conducted by users. In regards to interpretability of results, all 37 individuals in the study were able to correctly identify the different components of their Labcorp blood report accurately upon first attempt, and all 37 individuals in the study were able to correctly identify the different components of their Athelas report on the first attempt. Based on the answers of the users, they believed the interpretability of a Labcorp result to be (4.4)/5 and the interpretability of the Athelas test to be 5/5 with 5 being the easiest to interpret (Table 2). 8
9 Athelas Usability Questionnaire: Answers by Untrained Users 9
10 Athelas Usability Questionnaire: Answers by Untrained Users Table 2. Athelas answers to usability questionnaire. Completed by users 10
11 Conclusion Quality measures of the gold standard mechanism done by Mt. Sinai indicate that the current gold standard technology used by Labcorp generates results for a host of tests, including a WBC count, out of clinical range (0-3.4K WBC/uL, 3.4K-10.8K WBC/uL, >10.8K WBC/uL) in 8.3% of cases. 4 The Athelas system produced n=0 results outside of clinical range, while also maintaining reproducibility of results on 20 replicates of strips, and reproducibility of results on multiple devices. The Athelas system also proves to have developed a substantially equivalent platform to understand erroneous test procedure, and interpretable results as compared to the gold standard. When users attempt the Athelas system with only the in-app, and paper instruction manual, all 37 randomly selected users (age 10-73) were able to generate clinically significant results, benchmarked with the Labcorp results generated by a clinician. 4 J Clin Invest. 2016;126(5): doi: /jci
12 References B. A. Kidd, G. Hoffman, N. Zimmerman, L. Li, J. W. Morgan, P. K. Glowe, G. J. Botwin, S. Parekh, N. Babic, M. W. Doust, G. B. Stock, E. E. Schadt, J. T. Dudley, Evaluation of direct-to-consumer low-volume lab tests in healthy adults. J. Clin. Invest. 126, (2016). Kirchner, H., Ch Kleinicke, and W. Digel. "A whole-blood technique for testing production of human interferon by leukocytes." Journal of immunological methods 48.2 (1982):
1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationGlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.
More informationGlucCell TM SYSTEM USER S GUIDE ver 2.3 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE ver 2.3 Intended Use The GlucCell TM Cell Culture Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx 0244-055-00 OraQuick HCV Rapid Antibody Test Kit with product codes 1001-0270
More informationNICO /14 Nipro Diagnostics, Inc.
Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd CLINICAL
More informationWHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT. Product: SURE CHECK HIV 1/2 Assay Number: PQDx
WHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT Product: SURE CHECK HIV 1/2 Assay Number: PQDx 0054-006-00 Abstract SURE CHECK HIV 1/2 Assay with product code HIV201, manufactured
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx 0159-055-00 OraQuick HIV 1/2 Rapid Antibody Test with product codes 5x4-0010
More informationIDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats
IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HIV 1/2 Rapid Antibody Test WHO reference number: PQDx 0159-055-00 OraQuick HIV 1/2 Rapid Antibody Test with product codes 5x4-0010
More informationCentriVet GK Blood Glucose & Ketone Monitoring System
CentriVet GK Blood Glucose & Ketone Monitoring System FOR ANIMAL USE. NOT FOR HUMAN USE. Welcome and thank you for choosing the CentriVet GK Blood Glucose & Ketone Monitoring System. The CentriVet GK Blood
More informationGetting the Design Right Daniel Luna, Mackenzie Miller, Saloni Parikh, Ben Tebbs
Meet the Team Getting the Design Right Daniel Luna, Mackenzie Miller, Saloni Parikh, Ben Tebbs Mackenzie Miller: Project Manager Daniel Luna: Research Coordinator Saloni Parikh: User Interface Designer
More informationIssue Summary. Topic: Safety and effectiveness of the proposed OraQuick In-Home HIV Test
Blood Products Advisory Committee 102 nd Meeting, May 15, 2012 The Hilton Washington DC North/Gaithersburg 620 Perry Parkway Gaithersburg, Maryland, 20877 Issue Summary Topic: Safety and effectiveness
More informationSound Check: Essentials of a Hearing Screener
Sound Check: Passing the Test? Amanda Wolfe, Au.D., Elizabeth Galster, Au.D., & Beth Thomas, M.S. The importance of early identification and prevention of hearing loss for people of all ages is well established.
More informationEvaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System
Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,
More informationFibrosis stage reference images
Fibrosis stage reference images The following images are provided with the aim of improving inter-observer agreement in the assessment of fibrosis stage in liver biopsies, especially among non-specialist
More informationVQA Control SOP Version 4.0 Roche Amplicor HIV-1 DNA Test, v August 2007
1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationSmart Diagnostics for Point-of-Care!
Smart Diagnostics for Point-of-Care! The one special multi-parameter Point-of-Care (POC) instrument that covers the constantly expanding range of applications with the highest precision. Designed for HbA1c
More informationA Look Into the Determination of Cell Morphology in Hematology in the 21 st Century. Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter
A Look Into the Determination of Cell Morphology in Hematology in the 21 st Century Ramon Simon-Lopez, MD Global Scientific Director Beckman Coulter Is cell morphology important? AML M7 CLL CD5 CD19 NHL
More informationEpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)
EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric) Base Catalog # PLEASE READ THIS ENTIRE USER GUIDE BEFORE USE Uses: The EpiQuik Circulating Acetyl Histone H3K18 ELISA Kit (Colorimetric)
More informationBlood Glucose & Ketone Monitoring System
Blood Glucose & Ketone Monitoring System Self monitoring of blood glucose is an integral part of diabetes care, but the high cost of testing can make this impossible. At ACON, our goal is to provide high
More informationVetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP
VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and
More informationDear Prodigy System Owner:
User's Manual Dear Prodigy System Owner: You have chosen one of the best blood glucose monitoring systems available. This booklet has important information about the system. Please read it carefully. Your
More informationHM5. Hematology Analyzer BETTER. ACTUALLY.
HM5 Hematology Analyzer BETTER. ACTUALLY. Advanced Hematology Five-Part Differential The VetScan HM5 is a fully automated five-part differential hematology analyzer displaying a comprehensive 24-parameter
More informationQMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800
QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole
More informationPsychology, 2010, 1: doi: /psych Published Online August 2010 (
Psychology, 2010, 1: 194-198 doi:10.4236/psych.2010.13026 Published Online August 2010 (http://www.scirp.org/journal/psych) Using Generalizability Theory to Evaluate the Applicability of a Serial Bayes
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: DPP HIV 1/2 Assay WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: DPP HIV 1/2 Assay WHO reference number: PQDx 0053-006-00 DPP HIV 1/2 Assay with product code 65-9506-0, manufactured by Chembio Diagnostic
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationEmpirical Research Methods for Human-Computer Interaction. I. Scott MacKenzie Steven J. Castellucci
Empirical Research Methods for Human-Computer Interaction I. Scott MacKenzie Steven J. Castellucci 1 Topics The what, why, and how of empirical research Group participation in a real experiment Observations
More informationCLINICAL EVIDENCE MADE EASY
CLINICAL EVIDENCE MADE EASY M HARRIS, G TAYLOR & D JACKSON THE BASICS OF EVIDENCE-BASED MEDICINE CLINICAL EVIDENCE MADE EASY CLINICAL EVIDENCE MADE EASY M. Harris General Practitioner and Visiting Senior
More informationRELEASED. Clearing your active insulin
To clear all your settings: 1. Make sure the pump is not connected to your body. 2. Go to the Manage Settings screen. Menu > Utilities > Manage Settings 3. Simultaneously press and hold and until the Manage
More informationQuestionnaire for Using the LeEject Dental Safety Needle & Syringe System
Usability Study The LeEject Dental Safety Needle & Syringe System and instructions for use are surveyed by ten operators that have read the LeEject instructions for use and used the LeEject Dental Safety
More informationAdvanced Hematology Five-Part Differential. Simple Operation
HematologyAnalyzer Advanced Hematology Five-Part Differential analyzer displaying a comprehensive 22-parameter complete blood count (CBC) with cellular histograms on an easy-to-read touch-screen. Its superior
More informationAlign Gluco-Monitoring System (BG1) Quick Start Guide
EN Align Gluco-Monitoring System (BG1) Quick Start Guide WELCOME Welcome to your ihealth Align Gluco-Monitoring System (the ihealth system). The ihealth system is designed to be used with the free ihealth
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationActivity 5.2 Using chemical change to identify an unknown
Activity 5.2 Using chemical change to identify an unknown How can you identify an unknown powder? In this activity, students will develop a method to test five similar-looking powders with four test liquids.
More informationQuestionnaire design. Questionnaire Design: Content. Questionnaire Design. Questionnaire Design: Wording. Questionnaire Design: Wording OUTLINE
Questionnaire design OUTLINE Questionnaire design tests Reliability Validity POINTS TO CONSIDER Identify your research objectives. Identify your population or study sample Decide how to collect the information
More informationA. SAP is the D-Lab's name for a specific set of serum biochemical tests.
Understanding CBC, SAP, UA/Laura J. Steadman, DVM I. CBC - Complete Blood Count A. Three major types of cells are counted 1. Red Blood Cells 2. White Blood Cells 3. Platelets B. Cells are counted at the
More informationStudy on clinical practice guidelines in Estonia
Use of clinical practice guidelines in Estonia, and assessment of barriers to their implementation Pille Taba, University of Tartu April 28, 2011 Study on clinical practice guidelines in Estonia Aims To
More informationClinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013
linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the
More informationWhat I will talk about
EQA of POCT methods: How to interpret results? Anne Stavelin Labquality Days, Helsinki 12 February 2016 www.noklus.no Norwegian quality improvement of primary care laboratories What I will talk about What
More informationHypothesis generation and experimental design. Nick Hunt Bosch Institute School of Medical Sciences
Hypothesis generation and experimental design Nick Hunt Bosch Institute School of Medical Sciences Nicholas.hunt@sydney.edu.au Today s talk Hypothesis generation Hypothesis testing Elements of good experimental
More informationCritical Review Form Meta-analysis
Critical Review Form Meta-analysis Does this Adult Patient Have Septic Arthritis? JAMA 2007; 297: 1497-1488 Objective: To determine the diagnostic value of the history, physical examination, and routine
More informationDetectX. Urinary Creatinine Detection Kit. Catalog Number K002-H1. Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine
World s Only One Component Assay Simple & Easy to Use DetectX Urinary Creatinine Detection Kit Catalog Number K002-H1 Sample Types Validated: Human, Monkey, Dog, Rat and Mouse Urine Please read this insert
More informationAltrix Precision Temperature Management System. Simple. Safe. Efficient.
Altrix Precision Temperature Management System Simple. Safe. Efficient. Altrix Precision Temperature Management System Simple. Safe. Efficient. Patient warming and cooling can be critical in achieving
More informationSQA-V Gold Semiannual Validation/Proficiency/QC Recommendations
MEDICAL ELECTRONIC SYSTEMS 5757 W. Century Blvd. Suite 805 Los Angeles, CA. 90045 Remember, it ALL Started with a Sperm! www.mes global.com SQA-V Gold Semiannual Validation/Proficiency/QC Recommendations
More informationGlucose Meter. User Guide. Veterinary Monitoring System. For dog and cat use only
Glucose Meter User Guide Veterinary Monitoring System For dog and cat use only Gpet instruction Manual 31/5/09 18:06 Page 2 Gpet instruction Manual 31/5/09 18:06 Page 3 TABLE OF CONTENTS Your g-pet system
More informationYour pet s diabetes is manageable
BLOOD GLUCOSE MONITORING SYSTEM Your pet s diabetes is manageable Treatment and Monitoring Plan for: Pet s Name: Clinic Name: Primary Vet: Clinic Phone No.: Clinic Email: Emergency Phone No: Today s Date:
More informationReliability and Validity checks S-005
Reliability and Validity checks S-005 Checking on reliability of the data we collect Compare over time (test-retest) Item analysis Internal consistency Inter-rater agreement Compare over time Test-Retest
More informationVQA HIV DNA Control SOP Version 5.0 HIV DNA Testing 13 March 2012
1. PRINCIPLE 1.1. The Virology Quality Assurance (VQA) Laboratory provides external cell pellet controls for use in the validation of assays that detect HIV proviral DNA. 1.2. HIV seronegative peripheral
More informationShining a Light on MEDs Understanding morphine equivalent dose
Shining a Light on MEDs Understanding morphine equivalent dose In the workers compensation industry, 60.2 percent of claimants utilize opioid analgesics for the treatment of pain caused by a workplace
More informationFUSE TECHNICAL REPORT
FUSE TECHNICAL REPORT 1 / 16 Contents Page 3 Page 4 Page 8 Page 10 Page 13 Page 16 Introduction FUSE Accuracy Validation Testing LBD Risk Score Model Details FUSE Risk Score Implementation Details FUSE
More informationWHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT. Product: First Response HIV Card Test Number: PQDx
WHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT Product: First Response HIV 1-2-0 Card Test Number: PQDx 0018-010-00 Abstract First Response HIV 1-2-0 Card Test with product codes
More informationAn Introduction to Epidemiology
An Introduction to Epidemiology Wei Liu, MPH Biostatistics Core Pennington Biomedical Research Center Baton Rouge, LA Last edited: January, 14 th, 2014 TABLE OF CONTENTS Introduction.................................................................
More informationBlood Glucose Monitoring System
Blood Glucose Monitoring System Contents Chapter 1 OMRON Blood Glucose Monitoring System HEA-232 1.1 Introduction 1.2 List of Kit Contents 1.3 Product and Meter Display Chapter 2 Installing and Setting
More informationarxiv: v2 [cs.ai] 26 Sep 2018
Manipulating and Measuring Model Interpretability arxiv:1802.07810v2 [cs.ai] 26 Sep 2018 Forough Poursabzi-Sangdeh forough.poursabzi@microsoft.com Microsoft Research Jennifer Wortman Vaughan jenn@microsoft.com
More informationTopic: Sanitizing Food Contact Surfaces
Topic: Sanitizing Food Contact Surfaces WHAT IS Back to Basics? Four food safety training activities - to help train your frontline employees. Customize these modules by reviewing your own policies and
More informationDATA is derived either through. Self-Report Observation Measurement
Data Management DATA is derived either through Self-Report Observation Measurement QUESTION ANSWER DATA DATA may be from Structured or Unstructured questions? Quantitative or Qualitative? Numerical or
More informationBiometric Wellness Screening Modality Comparison Venipuncture vs. Fingerstick vs. Qcard Dried Blood Spot Method collection options
Biometric Wellness Screening Modality Comparison Venipuncture vs. Fingerstick vs. Qcard Dried Blood Spot Method options The following is a review of 3 methods of blood : venipuncture, fingerstick, and
More informationBLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S.
BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S. Introducing the DxH 500, an Open-vial Hematology System *. Quality 5-part Differential in a Compact Design The first
More informationCRITICAL EVALUATION OF BIOMEDICAL LITERATURE
Chapter 9 CRITICAL EVALUATION OF BIOMEDICAL LITERATURE M.G.Rajanandh, Department of Pharmacy Practice, SRM College of Pharmacy, SRM University. INTRODUCTION Reviewing the Biomedical Literature poses a
More informationQualitative and Quantitative Approaches Workshop. Comm 151i San Jose State U Dr. T.M. Coopman Okay for non-commercial use with attribution
Qualitative and Quantitative Approaches Workshop Comm 151i San Jose State U Dr. T.M. Coopman Okay for non-commercial use with attribution This Workshop This is a research skill workshop. This workshop
More informationhydrorise implant SIMPLY ACCURATE High-rigidity addition silicone
hydrorise implant SIMPLY ACCURATE High-rigidity addition silicone 2 Hydrorise Implant, simply accurate Dental implants represent a solid solution in prosthetic dentistry and are considered the privileged
More informationEstrone EIA kit. For the quantitative determination of estrone in dried fecal extracts, urine and tissue culture media. Cat. No.
K-ASSAY KAMIYA BIOMEDICAL COMPANY KAMIYA BIOMEDICAL COMPANY Estrone EIA kit For the quantitative determination of estrone in dried fecal extracts, urine and tissue culture media Cat. No. KT-720 For Research
More informationThank you very much for your guidance on this revision. We look forward to hearing from you.
1 PAINTED BUNTING MANUSCRIPT REVIEW HISTORY REVISION NOTES FROM AUTHORS (ROUND 2) We would like to submit a revision of our manuscript. As you requested, we have run new studies that employ additional
More informationEvolution of the Cancer Distress Coach App to Manage Stress in Cancer Survivors. Sophia K. Smith, PhD, MSW Associate Professor Duke School of Nursing
Evolution of the Cancer Distress Coach App to Manage Stress in Cancer Survivors Sophia K. Smith, PhD, MSW Associate Professor Duke School of Nursing Genomic and Precision Medicine Weekly Forum, September
More informationCORESTA RECOMMENDED METHOD N 25
CORESTA RECOMMENDED METHOD N 25 AMBIENT AIR FLOW AROUND CIGARETTES IN ROUTINE ANALYTICAL SMOKING MACHINES : CONTROL AND MONITORING (August 1991) 0. INTRODUCTION The development of smoking machines has
More informationStatistical Analysis August, 2002
Self Discovery Assessment Statistical Analysis August, 2002 An Independent Study By: Jerry MacDaid 2724 SW 14th Drive Gainesville, Florida 32608 Report on the Do What You Are assessment This report summarizes
More informationThe NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16
The NEPHROCHECK Test System Training Program The NEPHROCHECK Training Program US: Astute Medical, Inc. 2014 PN 700001 Rev D 2014/09/16 Learning Objectives At the conclusion of this training, participants
More informationTips for Writing a Research Paper in APA format:
Tips for Writing a Research Paper in APA format: Basics: A research paper (especially one that requires APA style) is different than a term paper, a creative writing paper, a composition-style paper, or
More informationGE Healthcare 9900 Innovation Drive Wauwatosa, WI U.S.A.
2011 General Electric Company All rights reserved. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time
More informationCommonKnowledge. Pacific University. Gina Clark Pacific University. Lauren Murphy Pacific University. Recommended Citation.
Pacific University CommonKnowledge PT Critically Appraised Topics School of Physical Therapy 2012 The diagnostic accuracy of patient subjective history compared to the gold standard of urodynamic testing
More informationRecharge. Relax.Repeat.
Recharge. Relax.Repeat. Recharge. Relax.Repeat. What s Inside Your rechargeable hearing aids will be delivered to you in two boxes. Please note that both ZPower rechargeable batteries and disposable batteries
More informationCritical Review Form Diagnostic Test
Critical Review Form Diagnostic Test The clinical presentation and impact of diagnostic delays on emergency department patients with spinal epidural abscess, J Emerg Med 2004; 26:285-291 Objectives: To
More informationAspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #:
Aspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #: 5605-01 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf
More informationIntro to SPSS. Using SPSS through WebFAS
Intro to SPSS Using SPSS through WebFAS http://www.yorku.ca/computing/students/labs/webfas/ Try it early (make sure it works from your computer) If you need help contact UIT Client Services Voice: 416-736-5800
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationStrength of Vinegar by Acid-Base Titration
Strength of Vinegar by Acid-Base Titration Test Exercise 100 points 1? QUESTIONS? How are acid/base titrations conducted? What is standardization? How do you standardize a solution of a base? How, and
More informationGuidelines for HIV Rapid Test Kits Evaluation
Guidelines for HIV Rapid Test Kits Evaluation George Alemnji, Ph.D. Laboratory Advisor GAP CDC Caribbean Regional Office Email:ikv3@cdc.gov 1 HIV Rapid Test Qualitative assay to detect HIV antibodies Most
More informationAlign Gluco-Monitoring System (BG1) Quick Start Guide
EN Align Gluco-Monitoring System (BG1) Quick Start Guide WELCOME Welcome to your ihealth Align Gluco-Monitoring System (the ihealth system). The ihealth system is designed to be used with the free ihealth
More informationIVD Revised 22 July, 2008
INTENDED USE The DRG One Step HCG Pregnancy Test is an in vitro diagnostic test for the qualitative detection of human chorionic gonadotropin (hcg) in urine. The test is a two-site immunoassay employing
More informationUrinary 8-Isoprostane ELISA kit
Urinary 8-Isoprostane ELISA kit Catalog Number: 8isoU Store at -20 C. FOR RESEARCH USE ONLY V.020420 Introduction This competitive ELISA kit is for determination of free and glucronidated 8-isoprostane
More informationChanges in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature
Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD;
More informationLUNG CANCER CLINICAL TRIALS
UNDERSTANDING SERIES LUNG CANCER CLINICAL TRIALS 1-800-298-2436 LungCancerAlliance.org A GUIDE FOR THE PATIENT THE BASICS ABOUT CLINICAL TRIALS WITH MORE INFORMATION YOU CAN MAKE BETTER CHOICES. The decision
More informationAbbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation
Laboratory Hematology 8:85-90 2002 Carden Jennings Publishing Co.. Ltd. Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation T. SCHISANO, L. VAN HOVE Abbott
More informationExploring Normalization Techniques for Human Judgments of Machine Translation Adequacy Collected Using Amazon Mechanical Turk
Exploring Normalization Techniques for Human Judgments of Machine Translation Adequacy Collected Using Amazon Mechanical Turk Michael Denkowski and Alon Lavie Language Technologies Institute School of
More informationStep-by-Step Instructions For OraQuick HCV Rapid Antibody Test
Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test Complexity: WAIVED for fingerstick whole blood and venipuncture whole blood. A Certificate of CLIA Waiver is required to perform the test
More informationHelicobacter pylori Antigen Test
Helicobacter pylori Antigen Test Instructions For Use Format: Cassette For: Catalog Number: VEL-001-HP(s) Specimen: Fecal Specimen * Please read the instructions carefully before use INTENDED USE Helicobacter
More informationHow can Blood Stabilizers be used for Validation of New Instruments? UK NEQAS for Leucocyte Immunophenotyping
How can Blood Stabilizers be used for Validation of New Instruments? Quality Pathology is essential for about 60% of all diagnoses. Thus, quality of pathology services is critical for good care and management.
More informationREDUCING VARIABILITY IN TEST RESULTS FOR OINDPS WITH AUTOMATED ACTUATION
REDUCING VARIABILITY IN TEST RESULTS FOR OINDPS WITH AUTOMATED ACTUATION In this article, Heli Chauhan, Senior Applications Chemist, and Linda (Lingzhi) Liao, Field Applications Scientist, both of Proveris
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationSingle Cell Quantitative Polymer Chain Reaction (sc-qpcr)
Single Cell Quantitative Polymer Chain Reaction (sc-qpcr) Analyzing gene expression profiles from a bulk population of cells provides an average profile which may obscure important biological differences
More informationEvaluating Quality in Creative Systems. Graeme Ritchie University of Aberdeen
Evaluating Quality in Creative Systems Graeme Ritchie University of Aberdeen Graeme Ritchie {2007} Some Empirical Criteria for Attributing Creativity to a Computer Program. Minds and Machines 17 {1}, pp.67-99.
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationAviva Combo BLOOD GLUCOSE METER. Advanced Owner s Booklet for Self-Testing Only
Aviva Combo BLOOD GLUCOSE METER Advanced Owner s Booklet for Self-Testing Only Before You Start Testing About the meter and test strips Carefully read and follow the instructions in the Getting Started
More informationBIOO RESEARCH PRODUCTS. Aspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #:
BIOO RESEARCH PRODUCTS Aspartate Transaminase (AST) Color Endpoint Assay Kit Manual Catalog #: 5605-01 BIOO Scientific 2010 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure
More information