NICO /14 Nipro Diagnostics, Inc.
|
|
- Cuthbert Randall
- 5 years ago
- Views:
Transcription
1 Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd
2 CLINICAL STUDY SUMMARY EVALUATION OBJECTIVES This clinical evaluation: 1. Evaluates the accuracy of the TRUEresult twist System utilizing international standard ISO 15197:13.. Compares the performance of the TRUEresult twist Blood Glucose Monitoring System utilizing ISO 15197:13 accuracy criteria to the performance of the TRUEresult twist System utilizing the EN ISO 15197:3 international standard. 3. Proves that lay users of the TRUEresult twist Blood Glucose Monitoring System can obtain accurate results consistent with those obtained by trained professionals.. Proves that the TRUEresult twist Blood Glucose Monitoring System is easy to use by first-time users. 5. Proves that the TRUEresult twist Blood Glucose Monitoring System produces precise results when measuring duplicate fingerstick samples. ISO STANDARD 15197:3 1 COMPARED TO ISO 15197:13 An ISO Standard is developed in response to an identified need in the community, and that will eventually be suitable for implementation on as broad a basis as possible. ISO developed the standard for blood glucose monitoring performance based on a consensus of professionals around the world. A clinical accuracy study was conducted according to the methodology requirements of ISO 15197:13. The data were assessed and compared using the accuracy bias limits of both EN ISO 15197:3 and ISO 15197:13. Below is a table outlining the accuracy parameters we evaluate in this study. Accuracy Parameter EN ISO 15197:3 ISO 15197:13 Target blood glucose level from which to base bias or % bias Acceptable bias from reference value for lower target glucose levels Acceptable bias from reference value for higher target glucose levels Acceptable % of all results within bias limits Parkes Error Grid /-.83 +/-.83 +/- % +/- 15% 95% 95% Not required 99% of results within Zones A and B 1 NICO_8_mmol.indd 3
3 CLINICAL STUDY SUMMARY The graph to the right illustrates the comparison of the EN ISO 15197:3 bias limits compared to ISO 15197:13 criteria for accuracy. Bias from YSI () - - Blood Glucose Results ISO 15197:3 ISO 15197:13 SITES AND SUBJECTS YSI Plasma Result () Clinical evaluations of the TRUEresult twist System were conducted at clinical institutions across the U.S. In total, subjects completed the protocol across the evaluation sites. Study demographics shown below. Ethnicity Category White African American Hispanic American Indian Asian Other Not Recorded User Study % 7% % % % % % Accuracy Study 39% 8% % 1% % % % Gender Category Male Female Not Recorded User Study % 53% 1% Accuracy Study 5% 9% 1% Age (Years) Category Average Maximum Minimum User Study Accuracy Study Years of Education Category <1 1 >1 Not Recorded User Study 11% 7% % % Accuracy Study 1% 7% 1% % CLINICAL EVALUATION METHODS Clinical Evaluations performed using the TRUEresult twist Blood Glucose Monitoring System: 1. Accuracy of fingerstick results obtained by patients and healthcare professionals (HCP) compared to the laboratory reference (three test strip lots used).. Accuracy of fingerstick results obtained by healthcare professionals compared to the laboratory reference using Parkes Error Grid analysis (three test strip lots used). 3. User performance evaluation including healthcare professional feedback on first-time user testing technique (one test strip lot used). NICO_8_mmol.indd
4 TRUEresult twist Bias from YSI () TRUEresult twist Clinical Accuracy ISO 15197:13 TOPLINE RESULTS USING ISO CRITERIA - - In this study using fingerstick samples obtained by healthcare professionals, the TRUEresult twist System produced 98.5% of results within the bias limits for EN ISO 15197:3 and 98.8% within the bias limits - for ISO 15197: YSI Plasma () TRUEresult twist Bias from YSI () TRUEresult twist Clinical Accuracy ISO 15197: YSI Plasma () TRUEresult twist Bias from YSI () TRUEresult twist Clinical Accuracy ISO 15197: YSI Plasma () ISO 15197:13 requires 3 lots of test strips to be used in the evaluation TRUEresult twist Clinical Accuracy ISO 15197:3 Summary of TRUEresult twist System accuracy against the YSI reference method using EN ISO 15197:3 and ISO 15197:13. ±.8 EN ISO 15197:3 Samples <. ±.5 ±.83 3/88 7/88 88/88.1% 81.8% 1.% Samples. TRUEresult twist Bias from YSI () - - ±.8 ISO 15197:13 Samples < 5.55 ±.5 ± YSI Plasma () 53/18 1/18 17/18 9.% 7.% 97.8% Samples 5.55 ± 5% ± 1% ± 15% ± % 78/51 8/51 88/51 53/51 5.3% 83.% 95.3% 98.% ± 5% ± 1% ± 15% 55/ 38/ 17/.7% 91.% 99.3% 3 NICO_8_mmol.indd 5
5 TOPLINE RESULTS USING ISO CRITERIA In this study, first-time users of the TRUEresult twist System obtained accurate results compared to the laboratory reference method. TRUEresult twist Bias from YSI () TRUEresult twist User Finger vs. YSI ISO 15197: YSI Plasma () ISO 15197:13 requires 1 data points from user self-testing ±.8 Samples < 5.55 ±.5 ±.83 /3 19/3 3/3.1% 8.% 1.% Samples 5.55 ± 5 % ± 1 % ± 15 % 3/77 9/77 77/77.8% 89.% 1% TRUEresult twist System Accuracy Results for total glucose concentrations. Within +/-.83 and within +/-15% 1/1 (1%) NICO_8_mmol.indd
6 TOPLINE RESULTS USING PARKES ERROR GRID PARKES ERROR GRID 3 Parkes et al. developed an error grid to analyze the clinical significance of the bias between blood glucose system results and lab reference results. For analysis, data points are assigned into one of the five Zones (A-E) on the error grid. Results, or data points, falling into Zones A and B are defined as clinically acceptable, where any observed bias from lab results would not lead to treatment decisions that may put a patient at risk. As the bias, or difference, increases (Zones C, D, and E), there is greater risk of undertreating or overtreating a patient based on the glucose system result. Analysis of data using the Parkes Error Grid is a requirement in ISO 15197:13. The TRUEresult twist System, in the hands of healthcare professionals, is proven to provide accurate fingerstick results that are comparable to the YSI reference method. TRUEresult twist HCP Finger vs. YSI TRUEresult twist HCP Finger Result () E D C B A y = 1.x +.8 R =.99 A B C D ZONE COUNT % Results In Zone Zone # Points % A 1% B % C % D % E % Total 1% YSI Plasma () TRUEresult twist User Finger vs. YSI TRUEresult twist User Finger Result () E D C B y =.99x +.1 R =.98 A 5 A B C D NICO_8_mmol.indd YSI Plasma ()
7 TOPLINE RESULTS USING STANDARD DEVIATION The TRUEresult twist System, in the hands of healthcare professionals, is proven to provide precise results that are comparable to the laboratory reference method. The Standard Deviation for Duplicate Measurements is a statistical calculation that measures how close the individual data points, for each pair of duplicate data points, are to each other. Meter results are more precise with a lower Standard Deviation (S.D.) value. TRUEresult twist S.D. Value.7% S.D. = (Σd /(n)) The Standard Deviation for Duplicate Measurements is calculated by using the above equation where d is the Percent Difference between duplicate data points and n is the number of patient samples. First-time users of the TRUEresult twist System agreed that the system was easy to use and achieved accurate and precise results. Healthcare professionals positively ranked first-time users compliance with the instructions for using the TRUEresult twist System. Was the system easy to use? Was the display easy to read? Were instructions clear on how to read results? Did instructions adequately describe how to apply blood? Were instructions easy to understand? User Evaluation Did patient follow the written instructions? Did patient read result correctly? Did patient apply blood correctly? Did patient insert strip correctly? HCP Evaluation NICO_8_mmol.indd 8
8 CONCLUSION In this study, the TRUEresult twist System demonstrated accurate performance in the hands of both healthcare professionals and first-time users. Clinical data obtained by both first-time users and healthcare professionals exceeded the minimum accuracy criteria for EN ISO 15197:3 and the more rigorous minimum accuracy criteria for ISO 15197:13. First-time users of the TRUEresult twist System agreed that the instructions were easy to understand and the system was easy to use. The TRUEresult twist System produced precise results as evidenced by a Standard Deviation value of.7%. 1. International Organization for Standardization. In vitro diagnostic test systems. Requirements for blood glucose monitoring system for self testing in managing diabetes mellitus. Reference number EN ISO 15197:3 (E). Geneva: International Organization for Standardization; 3.. International Organization for Standardization. In vitro diagnostic test systems. Requirements for blood glucose monitoring system for self testing in managing diabetes mellitus. Reference number ISO 15197:13. Geneva: International Organization for Standardization; Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. ; 3(8): Westgard, James O. Ph.D. Basic Method Validation 3rd Edition. Westgard QC, Inc., 8. FOR ADDITIONAL INFORMATION OR QUESTIONS, ADDRESS WRITTEN REQUESTS TO: FPO Nipro Diagnostics, Inc. NW 55th Ct. Ft. Lauderdale, FL 3339 NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd 1
Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013
linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the
More informationEVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS
EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The
More informationClinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System
Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationGD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips.
GD50 Higher Measuring Accuracy Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Simple. Accurate. Reliable.
More informationBern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.
Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationSmart & Simple Mobile Solutions.
Higher Measuring Accuracy Smart & Simple Mobile Solutions. Test Simply. Live Better. FORA Diamond lets you connect all your Bluetooth devices. *Also available with a Micro USB. ifora Diabetes Manager All
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationTRUE Meter Guide. Accurate and precise results for managing your diabetes effectively and affordably. because it fits your lifestyle!
TRUE Meter Guide Accurate and precise results for managing your diabetes effectively and affordably because it fits your lifestyle! Choosing the right TRUE meter to fit your needs and lifestyle... TRUE
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationReport. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013
Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was
More informationA New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation
691301DSTXXX10.1177/1932296817691301Journal of Diabetes Science and TechnologyChristiansen et al research-article2017 Original Article A New, Wireless-enabled Blood Glucose Monitoring System That Links
More informationClinical Evaluation for. Embrace No Code Blood Glucose Monitoring System
Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic
More informationQUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES
AACC24 QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES Graham Jones Department of Chemical Pathology, St Vincent s Hospital,
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationPerformance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose
ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2
More informationUltra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC
VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More informationA model for managing and monitoring the quality of glucometers used in a high-volume clinical setting
Original papers A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Güzin Aykal*, Ayşenur Yegin, Özgür Tekeli, Necat Yilmaz Central Laboratories, Antalya
More informationIndividual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab
Individual Lab Report Ci-Trol Nov,2016 ST VINCENT MEDICAL CENTER LABORATORY(LAB# 7300 ) 2131 WEST THIRD STREET LOS ANGELES CA USA 90057 Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC SYSMEX
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationGlucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers
Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationPerformance of three portable blood glucose monitoring devices used in a veterinary application
Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,
More informationA validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls
A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls Shyamali Pal R B Diagnostic Private Limited, Lake Town, Kolkata, India INFO ABSTRACT Corresponding
More informationPrecision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System
Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose
More informationPoint-of-Care Blood Glucose Meter Accuracy in the Hospital Setting
In Brief Point-of-care (POC) tests provide analytical information that can be used to make decisions at patients bedside, as opposed to laboratory tests that must be run at a central laboratory. POC testing
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More informationINTERNAL QUALITY CONTROL ( Q I C) QC)
EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)
More informationOwner s Booklet. For assistance, please call UK:
BLOOD GLUCOSE MONITORING SYSTEM BLOOD GLUCOSE MONITORING SYSTEM NO CODING Owner s Booklet For assistance, please call UK: 0800 08 588 08 Distributed by: Nipro Diagnostics (UK) Ltd 25 Barnes Wallis Road
More informationRelationship between glucose meter error and glycemic control efficacy
Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationAccuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial
Journal of Diabetes Science and Technology Volume 7, Issue 5, September 2013 Diabetes Technology Society ORIGINAL ARTICLE Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American
More informationPackage ega. March 21, 2017
Title Error Grid Analysis Version 2.0.0 Package ega March 21, 2017 Maintainer Daniel Schmolze Functions for assigning Clarke or Parkes (Consensus) error grid zones to blood glucose values,
More informationEvaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System
Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais
More informationSystem Accuracy Evaluation LUX Chol HDL test
System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall
More informationOneTouch Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose
Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose Laurence B Katz, Mike Grady, Steven J Setford, and Brian L Levy US Endocrinology SUPPLEMENT
More informationAXIS critical Appraisal of cross sectional Studies
AXIS critical Appraisal of cross sectional Studies Dr. Martin Downes @mjdepi Critical appraisal Systematic evaluation of clinical research to examine Trustworthiness. Valid methods and reporting Clear
More informationEffects of Simulated Altitude on Blood Glucose Meter Performance: Implications for In-Flight Blood Glucose Monitoring
Journal of Diabetes Science and Technology Volume 6, Issue 4, July 2012 Diabetes Technology Society ORIGINAL ARTICLE Effects of Simulated Altitude on Blood Glucose Meter Performance: Tolu, M.B.B.S., M.R.C.P.,
More informationAccuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems
Original Article Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems DOI: 10.7860/JCDR/2018/34055.11671 Internal Medicine Section Anastasios Tentolouris 1,
More informationPRODUCT DATA SHEET DIABETES CARE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE
PRODUCT DATA SHEET DIABETES CARE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE PRODUCT FEATURES Memory
More informationScreening for diabetes
Screening for diabetes Peggy Odegard, Pharm.D, BCPS, CDE What are your risks? 1 Diabetes Mellitus A problem with glucose regulation type 1= pancreas cannot produce insulin so total insulin deficiency,
More informationADMS Sampling Technique and Survey Studies
Principles of Measurement Measurement As a way of understanding, evaluating, and differentiating characteristics Provides a mechanism to achieve precision in this understanding, the extent or quality As
More informationWhy do we need SD LipidoCare?
SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing
More informationThe performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.
Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added
More information1 PROTOCOL. Comparison Study Summary
Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015
More informationValidity and Reliability of a Glucometer Against Industry Reference Standards
514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference
More informationMulti User Model. Product Overview
Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL
More informationCustomer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA).
Customer Bulletin Assignment of Ranges for PTS Panels Multi-Chemistry Controls (Cat. No./REF 721) and PTS Panels Controls (Cat. No./REF 722) Background: PTS Panels Multi-Chemistry Controls and Controls
More informationDose Accuracy of the ClikSTAR, NovoPen 4, and Luxura Insulin Pens: Results of Laboratory and Field Studies
Journal of Diabetes Science and Technology Volume 5, Issue 5, September 2011 Diabetes Technology Society ORIGINAL ARTICLE Arnd, Ph.D., 1 Nils Basso, Ph.D., 2 and Steffen Adler, Ph.D. 1 Abstract Background:
More informationAccuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies
Journal of Diabetes Science and Technology Volume 4, Issue 5, September 2010 Diabetes Technology Society ORIGINAL ARTICLES Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood
More informationPerformance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment
Journal of Diabetes Science and Technology Volume 7, Issue 4, July 2013 Diabetes Technology Society SYMPOSIUM Performance Evaluations of Continuous Glucose Monitoring Systems: Karin, B.S., Günther Schmelzeisen-Redeker,
More informationFreeStyle Libre Report Overview
FreeStyle Libre Report Overview The LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends. Not all meters and meter reports
More informationClinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing
Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Clinical Impact of Prandial State, Exercise, and Preparation on the Equivalence of Alternative- Blood Glucose Testing DAWN M. BINA, CMT ROBYN
More informationQuick Reference Guide
FLASH GLUCOSE MONITORING SYSTEM Quick Reference Guide IMPORTANT USER INFORMATION Before you use your System, review all the product instructions and the Interactive Tutorial. The Quick Reference Guide
More informationComputing the Surveillance Error Grid Analysis: Procedure and Examples
5959DSTXXX.77/99685959Journal of Diabetes Science and TechnologyKovatchev et al research-article Original Article Computing the Surveillance Error Grid Analysis: Procedure and Examples Journal of Diabetes
More informationGB/T Translated English of Chinese Standard: GB/T
Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro
More informationQuick Reference Guide
FLASH GLUCOSE MONITORING SYSTEM Quick Reference Guide IMPORTANT USER INFORMATION Before you use your System, review all the product instructions and the Interactive Tutorial. The Quick Reference Guide
More informationBlood Glucose Monitoring System USER GUIDE. For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips
Blood Glucose Monitoring System USER GUIDE For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips Thank you for choosing the Ascensia CONTOUR Blood Glucose Monitoring System!
More informationPatient survey report Survey of people who use community mental health services 2015 South London and Maudsley NHS Foundation Trust
Patient survey report 2015 Survey of people who use community mental health services 2015 National NHS patient survey programme Survey of people who use community mental health services 2015 The Care
More informationThermometer PRODUCT CATALOGUE
Blood Glucose Monitor Safety Lancet Home Care Blood Pressure Monitor Thermometer PRODUCT CATALOGUE About TaiDoc Since 1998 TaiDoc has been dedicated to manufacturing and marketing premium medical devices
More informationComprehensive Resource Guide for Healthcare Professionals. Professional Monitoring Blood Glucose System
Professional Monitoring Blood Glucose System (using TRUE METRIX PRO Test Strips with GDH-FAD Enzyme) Comprehensive Resource Guide for Healthcare Professionals (Policies, Processes, and Procedures) Information
More informationDetermination of hemoglobin is one of the most commonly
ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning
More informationQuick Reference Guide
FLASH GLUCOSE MONITORING SYSTEM Quick Reference Guide FreeStyle LibreLink app A FreeStyle Libre product IMPORTANT USER INFORMATION Before you use your System, review all the product instructions and the
More informationMQ Comparison of glucometers using whole blood
MQ 2017-1 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices
More informationThe LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends.
Report Overview The LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends. Not all meters and meter reports are available
More informationSummary of Significant Changes at this Revision. Items Required
Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of
More informationMQ Comparison of glucometers using whole blood
MQ 2018-2 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices
More informationSurvey of U.S. Adult Cigarette Smokers
Survey of U.S. Adult Cigarette Smokers Topline Report 13 October 2016 ADULT CIGARETTE SMOKERS (U.S. NATIONWIDE): n=1,000 online surveys S1. First, which of the following ranges includes your age? 0% Under
More informationOmnitest : 2013
ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample
More informationOGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring
More informationMQ Comparison of glucometers using heparin whole blood
MQ 2014-1 Comparison of glucometers using heparin whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing
More informationBlood Glucose Test Strip Guidelines
Berkshire West Integrated Care System Representing Berkshire West Clinical Commisioning Group Royal Berkshire NHS Foundation Trust Berkshire Healthcare NHS Foundation Trust Berkshire West Primary Care
More informationRigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device. At-Home Usability and Interpretability Studies for Consumers
Rigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device At-Home Usability and Interpretability Studies for Consumers INTRO Current state of technology indicates that the gold
More informationAccuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips
Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Sponsored by Arthur Roberts, MD of The Living Heart Foundation The Living Heart
More informationSUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5
USER-EVALUATION ASCENSIA CONTOUR Blood Glucose Meter System Table of contents SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 MATERIALS AND METHODS... 6 ASCENSIA CONTOUR...
More informationHoughton Mifflin Harcourt. Participant s Guide Distractor Rationales Fall 2012 User s Conference By Christina Fritz
Houghton Mifflin Harcourt Participant s Guide Distractor Rationales Fall 2012 User s Conference By Christina Fritz Topics for Discussion High Quality Items Anatomy of a Multiple Choice Item Types of Distractors
More informationA Protocol for Establishing Critical Power in Running
A Protocol for Establishing Critical Power in Running A Working Paper Prepared by The Stryd Team 27 September 201 1 Introduction One of the primary reasons for considering one s power consumption during
More informationEvaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter
Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood
More informationCardiac Assessment Controls
Bio-Rad Laboratories CARDIAC ASSESSMENT CONTROLS Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central
More informationSignificant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters
Journal of Diabetes Science and Technology ORIGINAL ARTICLES Volume 1, Issue 2, March 2007 Diabetes Technology Society Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from
More informationMethod Comparison Report Semi-Annual 1/5/2018
Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community
More informationAlign Gluco-Monitoring System (BG1) Quick Start Guide
EN Align Gluco-Monitoring System (BG1) Quick Start Guide WELCOME Welcome to your ihealth Align Gluco-Monitoring System (the ihealth system). The ihealth system is designed to be used with the free ihealth
More informationSAMPLE. Glucose Monitoring in Settings Without Laboratory Support
POCT13c Glucose Monitoring in Settings Without Laboratory Support This guideline focuses on performance of point-of-care glucose monitoring systems, with an emphasis on safety practices, quality control,
More informationThe Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values
University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary
More informationComparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Emerging Treatments and Technologies O R I G I N A L A R T I C L E Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors BORIS KOVATCHEV, PHD 1 STACEY ANDERSON, MD 1 2 LUTZ
More informationVariable Measurement, Norms & Differences
Variable Measurement, Norms & Differences 1 Expectations Begins with hypothesis (general concept) or question Create specific, testable prediction Prediction can specify relation or group differences Different
More informationComparability Study of Online and Paper and Pencil Tests Using Modified Internally and Externally Matched Criteria
Comparability Study of Online and Paper and Pencil Tests Using Modified Internally and Externally Matched Criteria Thakur Karkee Measurement Incorporated Dong-In Kim CTB/McGraw-Hill Kevin Fatica CTB/McGraw-Hill
More informationAgenda. 15 Dec
Implementing Traceability for Heterogeneous Analytes An Industry View JCTLM Symposium on Reference Measurement Systems for Biologicals, December 15, 2004 R. Miller 15 Dec. 2004 1 Agenda Recent trends in
More informationThe FreeStyle Libre Flash Glucose Monitoring System. Healthcare Professional Training - London
The FreeStyle Libre Flash Glucose Monitoring System Healthcare Professional Training - London The FreeStyle Libre Flash Glucose Monitoring System Indication of Use The FreeStyle Libre flash glucose monitoring
More information