1 Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd

2 CLINICAL STUDY SUMMARY EVALUATION OBJECTIVES This clinical evaluation: 1. Evaluates the accuracy of the TRUEresult twist System utilizing international standard ISO 15197:13.. Compares the performance of the TRUEresult twist Blood Glucose Monitoring System utilizing ISO 15197:13 accuracy criteria to the performance of the TRUEresult twist System utilizing the EN ISO 15197:3 international standard. 3. Proves that lay users of the TRUEresult twist Blood Glucose Monitoring System can obtain accurate results consistent with those obtained by trained professionals.. Proves that the TRUEresult twist Blood Glucose Monitoring System is easy to use by first-time users. 5. Proves that the TRUEresult twist Blood Glucose Monitoring System produces precise results when measuring duplicate fingerstick samples. ISO STANDARD 15197:3 1 COMPARED TO ISO 15197:13 An ISO Standard is developed in response to an identified need in the community, and that will eventually be suitable for implementation on as broad a basis as possible. ISO developed the standard for blood glucose monitoring performance based on a consensus of professionals around the world. A clinical accuracy study was conducted according to the methodology requirements of ISO 15197:13. The data were assessed and compared using the accuracy bias limits of both EN ISO 15197:3 and ISO 15197:13. Below is a table outlining the accuracy parameters we evaluate in this study. Accuracy Parameter EN ISO 15197:3 ISO 15197:13 Target blood glucose level from which to base bias or % bias Acceptable bias from reference value for lower target glucose levels Acceptable bias from reference value for higher target glucose levels Acceptable % of all results within bias limits Parkes Error Grid /-.83 +/-.83 +/- % +/- 15% 95% 95% Not required 99% of results within Zones A and B 1 NICO_8_mmol.indd 3

3 CLINICAL STUDY SUMMARY The graph to the right illustrates the comparison of the EN ISO 15197:3 bias limits compared to ISO 15197:13 criteria for accuracy. Bias from YSI () - - Blood Glucose Results ISO 15197:3 ISO 15197:13 SITES AND SUBJECTS YSI Plasma Result () Clinical evaluations of the TRUEresult twist System were conducted at clinical institutions across the U.S. In total, subjects completed the protocol across the evaluation sites. Study demographics shown below. Ethnicity Category White African American Hispanic American Indian Asian Other Not Recorded User Study % 7% % % % % % Accuracy Study 39% 8% % 1% % % % Gender Category Male Female Not Recorded User Study % 53% 1% Accuracy Study 5% 9% 1% Age (Years) Category Average Maximum Minimum User Study Accuracy Study Years of Education Category <1 1 >1 Not Recorded User Study 11% 7% % % Accuracy Study 1% 7% 1% % CLINICAL EVALUATION METHODS Clinical Evaluations performed using the TRUEresult twist Blood Glucose Monitoring System: 1. Accuracy of fingerstick results obtained by patients and healthcare professionals (HCP) compared to the laboratory reference (three test strip lots used).. Accuracy of fingerstick results obtained by healthcare professionals compared to the laboratory reference using Parkes Error Grid analysis (three test strip lots used). 3. User performance evaluation including healthcare professional feedback on first-time user testing technique (one test strip lot used). NICO_8_mmol.indd

4 TRUEresult twist Bias from YSI () TRUEresult twist Clinical Accuracy ISO 15197:13 TOPLINE RESULTS USING ISO CRITERIA - - In this study using fingerstick samples obtained by healthcare professionals, the TRUEresult twist System produced 98.5% of results within the bias limits for EN ISO 15197:3 and 98.8% within the bias limits - for ISO 15197: YSI Plasma () TRUEresult twist Bias from YSI () TRUEresult twist Clinical Accuracy ISO 15197: YSI Plasma () TRUEresult twist Bias from YSI () TRUEresult twist Clinical Accuracy ISO 15197: YSI Plasma () ISO 15197:13 requires 3 lots of test strips to be used in the evaluation TRUEresult twist Clinical Accuracy ISO 15197:3 Summary of TRUEresult twist System accuracy against the YSI reference method using EN ISO 15197:3 and ISO 15197:13. ±.8 EN ISO 15197:3 Samples <. ±.5 ±.83 3/88 7/88 88/88.1% 81.8% 1.% Samples. TRUEresult twist Bias from YSI () - - ±.8 ISO 15197:13 Samples < 5.55 ±.5 ± YSI Plasma () 53/18 1/18 17/18 9.% 7.% 97.8% Samples 5.55 ± 5% ± 1% ± 15% ± % 78/51 8/51 88/51 53/51 5.3% 83.% 95.3% 98.% ± 5% ± 1% ± 15% 55/ 38/ 17/.7% 91.% 99.3% 3 NICO_8_mmol.indd 5

5 TOPLINE RESULTS USING ISO CRITERIA In this study, first-time users of the TRUEresult twist System obtained accurate results compared to the laboratory reference method. TRUEresult twist Bias from YSI () TRUEresult twist User Finger vs. YSI ISO 15197: YSI Plasma () ISO 15197:13 requires 1 data points from user self-testing ±.8 Samples < 5.55 ±.5 ±.83 /3 19/3 3/3.1% 8.% 1.% Samples 5.55 ± 5 % ± 1 % ± 15 % 3/77 9/77 77/77.8% 89.% 1% TRUEresult twist System Accuracy Results for total glucose concentrations. Within +/-.83 and within +/-15% 1/1 (1%) NICO_8_mmol.indd

6 TOPLINE RESULTS USING PARKES ERROR GRID PARKES ERROR GRID 3 Parkes et al. developed an error grid to analyze the clinical significance of the bias between blood glucose system results and lab reference results. For analysis, data points are assigned into one of the five Zones (A-E) on the error grid. Results, or data points, falling into Zones A and B are defined as clinically acceptable, where any observed bias from lab results would not lead to treatment decisions that may put a patient at risk. As the bias, or difference, increases (Zones C, D, and E), there is greater risk of undertreating or overtreating a patient based on the glucose system result. Analysis of data using the Parkes Error Grid is a requirement in ISO 15197:13. The TRUEresult twist System, in the hands of healthcare professionals, is proven to provide accurate fingerstick results that are comparable to the YSI reference method. TRUEresult twist HCP Finger vs. YSI TRUEresult twist HCP Finger Result () E D C B A y = 1.x +.8 R =.99 A B C D ZONE COUNT % Results In Zone Zone # Points % A 1% B % C % D % E % Total 1% YSI Plasma () TRUEresult twist User Finger vs. YSI TRUEresult twist User Finger Result () E D C B y =.99x +.1 R =.98 A 5 A B C D NICO_8_mmol.indd YSI Plasma ()

7 TOPLINE RESULTS USING STANDARD DEVIATION The TRUEresult twist System, in the hands of healthcare professionals, is proven to provide precise results that are comparable to the laboratory reference method. The Standard Deviation for Duplicate Measurements is a statistical calculation that measures how close the individual data points, for each pair of duplicate data points, are to each other. Meter results are more precise with a lower Standard Deviation (S.D.) value. TRUEresult twist S.D. Value.7% S.D. = (Σd /(n)) The Standard Deviation for Duplicate Measurements is calculated by using the above equation where d is the Percent Difference between duplicate data points and n is the number of patient samples. First-time users of the TRUEresult twist System agreed that the system was easy to use and achieved accurate and precise results. Healthcare professionals positively ranked first-time users compliance with the instructions for using the TRUEresult twist System. Was the system easy to use? Was the display easy to read? Were instructions clear on how to read results? Did instructions adequately describe how to apply blood? Were instructions easy to understand? User Evaluation Did patient follow the written instructions? Did patient read result correctly? Did patient apply blood correctly? Did patient insert strip correctly? HCP Evaluation NICO_8_mmol.indd 8

8 CONCLUSION In this study, the TRUEresult twist System demonstrated accurate performance in the hands of both healthcare professionals and first-time users. Clinical data obtained by both first-time users and healthcare professionals exceeded the minimum accuracy criteria for EN ISO 15197:3 and the more rigorous minimum accuracy criteria for ISO 15197:13. First-time users of the TRUEresult twist System agreed that the instructions were easy to understand and the system was easy to use. The TRUEresult twist System produced precise results as evidenced by a Standard Deviation value of.7%. 1. International Organization for Standardization. In vitro diagnostic test systems. Requirements for blood glucose monitoring system for self testing in managing diabetes mellitus. Reference number EN ISO 15197:3 (E). Geneva: International Organization for Standardization; 3.. International Organization for Standardization. In vitro diagnostic test systems. Requirements for blood glucose monitoring system for self testing in managing diabetes mellitus. Reference number ISO 15197:13. Geneva: International Organization for Standardization; Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. ; 3(8): Westgard, James O. Ph.D. Basic Method Validation 3rd Edition. Westgard QC, Inc., 8. FOR ADDITIONAL INFORMATION OR QUESTIONS, ADDRESS WRITTEN REQUESTS TO: FPO Nipro Diagnostics, Inc. NW 55th Ct. Ft. Lauderdale, FL 3339 NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd 1

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013 linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the

More information

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips.

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Higher Measuring Accuracy Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Simple. Accurate. Reliable.

More information

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D. Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Smart & Simple Mobile Solutions.

Smart & Simple Mobile Solutions. Higher Measuring Accuracy Smart & Simple Mobile Solutions. Test Simply. Live Better. FORA Diamond lets you connect all your Bluetooth devices. *Also available with a Micro USB. ifora Diabetes Manager All

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

TRUE Meter Guide. Accurate and precise results for managing your diabetes effectively and affordably. because it fits your lifestyle!

TRUE Meter Guide. Accurate and precise results for managing your diabetes effectively and affordably. because it fits your lifestyle! TRUE Meter Guide Accurate and precise results for managing your diabetes effectively and affordably because it fits your lifestyle! Choosing the right TRUE meter to fit your needs and lifestyle... TRUE

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was

More information

A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation

A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation 691301DSTXXX10.1177/1932296817691301Journal of Diabetes Science and TechnologyChristiansen et al research-article2017 Original Article A New, Wireless-enabled Blood Glucose Monitoring System That Links

More information

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic

More information

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES AACC24 QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES Graham Jones Department of Chemical Pathology, St Vincent s Hospital,

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch

More information

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2

More information

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values

More information

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the

More information

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Original papers A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Güzin Aykal*, Ayşenur Yegin, Özgür Tekeli, Necat Yilmaz Central Laboratories, Antalya

More information

Individual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab

Individual Lab Report Ci-Trol Nov,2016. Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC Your Lab Individual Lab Report Ci-Trol Nov,2016 ST VINCENT MEDICAL CENTER LABORATORY(LAB# 7300 ) 2131 WEST THIRD STREET LOS ANGELES CA USA 90057 Abnormal Fibrinogen (mg/dl) Abnormal Fbg Control - Lot# LFC SYSMEX

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls Shyamali Pal R B Diagnostic Private Limited, Lake Town, Kolkata, India INFO ABSTRACT Corresponding

More information

Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System

Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose

More information

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting In Brief Point-of-care (POC) tests provide analytical information that can be used to make decisions at patients bedside, as opposed to laboratory tests that must be run at a central laboratory. POC testing

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

INTERNAL QUALITY CONTROL ( Q I C) QC)

INTERNAL QUALITY CONTROL ( Q I C) QC) EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)

More information

Owner s Booklet. For assistance, please call UK:

Owner s Booklet. For assistance, please call UK: BLOOD GLUCOSE MONITORING SYSTEM BLOOD GLUCOSE MONITORING SYSTEM NO CODING Owner s Booklet For assistance, please call UK: 0800 08 588 08 Distributed by: Nipro Diagnostics (UK) Ltd 25 Barnes Wallis Road

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial Journal of Diabetes Science and Technology Volume 7, Issue 5, September 2013 Diabetes Technology Society ORIGINAL ARTICLE Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American

More information

Package ega. March 21, 2017

Package ega. March 21, 2017 Title Error Grid Analysis Version 2.0.0 Package ega March 21, 2017 Maintainer Daniel Schmolze Functions for assigning Clarke or Parkes (Consensus) error grid zones to blood glucose values,

More information

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais

More information

System Accuracy Evaluation LUX Chol HDL test

System Accuracy Evaluation LUX Chol HDL test System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall

More information

OneTouch Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose

OneTouch Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose Laurence B Katz, Mike Grady, Steven J Setford, and Brian L Levy US Endocrinology SUPPLEMENT

More information

AXIS critical Appraisal of cross sectional Studies

AXIS critical Appraisal of cross sectional Studies AXIS critical Appraisal of cross sectional Studies Dr. Martin Downes @mjdepi Critical appraisal Systematic evaluation of clinical research to examine Trustworthiness. Valid methods and reporting Clear

More information

Effects of Simulated Altitude on Blood Glucose Meter Performance: Implications for In-Flight Blood Glucose Monitoring

Effects of Simulated Altitude on Blood Glucose Meter Performance: Implications for In-Flight Blood Glucose Monitoring Journal of Diabetes Science and Technology Volume 6, Issue 4, July 2012 Diabetes Technology Society ORIGINAL ARTICLE Effects of Simulated Altitude on Blood Glucose Meter Performance: Tolu, M.B.B.S., M.R.C.P.,

More information

Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems

Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems Original Article Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems DOI: 10.7860/JCDR/2018/34055.11671 Internal Medicine Section Anastasios Tentolouris 1,

More information

PRODUCT DATA SHEET DIABETES CARE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE

PRODUCT DATA SHEET DIABETES CARE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE PRODUCT DATA SHEET DIABETES CARE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE SMART FOR SELF-TESTING AND POINT-OF-CARE-TESTING OF WHOLE BLOOD GLUCOSE PRODUCT FEATURES Memory

More information

Screening for diabetes

Screening for diabetes Screening for diabetes Peggy Odegard, Pharm.D, BCPS, CDE What are your risks? 1 Diabetes Mellitus A problem with glucose regulation type 1= pancreas cannot produce insulin so total insulin deficiency,

More information

ADMS Sampling Technique and Survey Studies

ADMS Sampling Technique and Survey Studies Principles of Measurement Measurement As a way of understanding, evaluating, and differentiating characteristics Provides a mechanism to achieve precision in this understanding, the extent or quality As

More information

Why do we need SD LipidoCare?

Why do we need SD LipidoCare? SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing

More information

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests. Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added

More information

1 PROTOCOL. Comparison Study Summary

1 PROTOCOL. Comparison Study Summary Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

Multi User Model. Product Overview

Multi User Model. Product Overview Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL

More information

Customer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA).

Customer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA). Customer Bulletin Assignment of Ranges for PTS Panels Multi-Chemistry Controls (Cat. No./REF 721) and PTS Panels Controls (Cat. No./REF 722) Background: PTS Panels Multi-Chemistry Controls and Controls

More information

Dose Accuracy of the ClikSTAR, NovoPen 4, and Luxura Insulin Pens: Results of Laboratory and Field Studies

Dose Accuracy of the ClikSTAR, NovoPen 4, and Luxura Insulin Pens: Results of Laboratory and Field Studies Journal of Diabetes Science and Technology Volume 5, Issue 5, September 2011 Diabetes Technology Society ORIGINAL ARTICLE Arnd, Ph.D., 1 Nils Basso, Ph.D., 2 and Steffen Adler, Ph.D. 1 Abstract Background:

More information

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies Journal of Diabetes Science and Technology Volume 4, Issue 5, September 2010 Diabetes Technology Society ORIGINAL ARTICLES Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood

More information

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment Journal of Diabetes Science and Technology Volume 7, Issue 4, July 2013 Diabetes Technology Society SYMPOSIUM Performance Evaluations of Continuous Glucose Monitoring Systems: Karin, B.S., Günther Schmelzeisen-Redeker,

More information

FreeStyle Libre Report Overview

FreeStyle Libre Report Overview FreeStyle Libre Report Overview The LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends. Not all meters and meter reports

More information

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Clinical Impact of Prandial State, Exercise, and Preparation on the Equivalence of Alternative- Blood Glucose Testing DAWN M. BINA, CMT ROBYN

More information

Quick Reference Guide

Quick Reference Guide FLASH GLUCOSE MONITORING SYSTEM Quick Reference Guide IMPORTANT USER INFORMATION Before you use your System, review all the product instructions and the Interactive Tutorial. The Quick Reference Guide

More information

Computing the Surveillance Error Grid Analysis: Procedure and Examples

Computing the Surveillance Error Grid Analysis: Procedure and Examples 5959DSTXXX.77/99685959Journal of Diabetes Science and TechnologyKovatchev et al research-article Original Article Computing the Surveillance Error Grid Analysis: Procedure and Examples Journal of Diabetes

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro

More information

Quick Reference Guide

Quick Reference Guide FLASH GLUCOSE MONITORING SYSTEM Quick Reference Guide IMPORTANT USER INFORMATION Before you use your System, review all the product instructions and the Interactive Tutorial. The Quick Reference Guide

More information

Blood Glucose Monitoring System USER GUIDE. For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips

Blood Glucose Monitoring System USER GUIDE. For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips Blood Glucose Monitoring System USER GUIDE For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips Thank you for choosing the Ascensia CONTOUR Blood Glucose Monitoring System!

More information

Patient survey report Survey of people who use community mental health services 2015 South London and Maudsley NHS Foundation Trust

Patient survey report Survey of people who use community mental health services 2015 South London and Maudsley NHS Foundation Trust Patient survey report 2015 Survey of people who use community mental health services 2015 National NHS patient survey programme Survey of people who use community mental health services 2015 The Care

More information

Thermometer PRODUCT CATALOGUE

Thermometer PRODUCT CATALOGUE Blood Glucose Monitor Safety Lancet Home Care Blood Pressure Monitor Thermometer PRODUCT CATALOGUE About TaiDoc Since 1998 TaiDoc has been dedicated to manufacturing and marketing premium medical devices

More information

Comprehensive Resource Guide for Healthcare Professionals. Professional Monitoring Blood Glucose System

Comprehensive Resource Guide for Healthcare Professionals. Professional Monitoring Blood Glucose System Professional Monitoring Blood Glucose System (using TRUE METRIX PRO Test Strips with GDH-FAD Enzyme) Comprehensive Resource Guide for Healthcare Professionals (Policies, Processes, and Procedures) Information

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Quick Reference Guide

Quick Reference Guide FLASH GLUCOSE MONITORING SYSTEM Quick Reference Guide FreeStyle LibreLink app A FreeStyle Libre product IMPORTANT USER INFORMATION Before you use your System, review all the product instructions and the

More information

MQ Comparison of glucometers using whole blood

MQ Comparison of glucometers using whole blood MQ 2017-1 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices

More information

The LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends.

The LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends. Report Overview The LibreView system gives you a consistent set of clear, intuitive reports that make it easier and faster to discover patterns and trends. Not all meters and meter reports are available

More information

Summary of Significant Changes at this Revision. Items Required

Summary of Significant Changes at this Revision. Items Required Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of

More information

MQ Comparison of glucometers using whole blood

MQ Comparison of glucometers using whole blood MQ 2018-2 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices

More information

Survey of U.S. Adult Cigarette Smokers

Survey of U.S. Adult Cigarette Smokers Survey of U.S. Adult Cigarette Smokers Topline Report 13 October 2016 ADULT CIGARETTE SMOKERS (U.S. NATIONWIDE): n=1,000 online surveys S1. First, which of the following ranges includes your age? 0% Under

More information

Omnitest : 2013

Omnitest : 2013 ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample

More information

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking. ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring

More information

MQ Comparison of glucometers using heparin whole blood

MQ Comparison of glucometers using heparin whole blood MQ 2014-1 Comparison of glucometers using heparin whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing

More information

Blood Glucose Test Strip Guidelines

Blood Glucose Test Strip Guidelines Berkshire West Integrated Care System Representing Berkshire West Clinical Commisioning Group Royal Berkshire NHS Foundation Trust Berkshire Healthcare NHS Foundation Trust Berkshire West Primary Care

More information

Rigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device. At-Home Usability and Interpretability Studies for Consumers

Rigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device. At-Home Usability and Interpretability Studies for Consumers Rigorous Bench Studies to Determine Sensitivity & Precision of Athelas Strip & Device At-Home Usability and Interpretability Studies for Consumers INTRO Current state of technology indicates that the gold

More information

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Sponsored by Arthur Roberts, MD of The Living Heart Foundation The Living Heart

More information

SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5

SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 USER-EVALUATION ASCENSIA CONTOUR Blood Glucose Meter System Table of contents SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 MATERIALS AND METHODS... 6 ASCENSIA CONTOUR...

More information

Houghton Mifflin Harcourt. Participant s Guide Distractor Rationales Fall 2012 User s Conference By Christina Fritz

Houghton Mifflin Harcourt. Participant s Guide Distractor Rationales Fall 2012 User s Conference By Christina Fritz Houghton Mifflin Harcourt Participant s Guide Distractor Rationales Fall 2012 User s Conference By Christina Fritz Topics for Discussion High Quality Items Anatomy of a Multiple Choice Item Types of Distractors

More information

A Protocol for Establishing Critical Power in Running

A Protocol for Establishing Critical Power in Running A Protocol for Establishing Critical Power in Running A Working Paper Prepared by The Stryd Team 27 September 201 1 Introduction One of the primary reasons for considering one s power consumption during

More information

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood

More information

Cardiac Assessment Controls

Cardiac Assessment Controls Bio-Rad Laboratories CARDIAC ASSESSMENT CONTROLS Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central

More information

Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters

Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from Miscoded Meters Journal of Diabetes Science and Technology ORIGINAL ARTICLES Volume 1, Issue 2, March 2007 Diabetes Technology Society Significant Insulin Dose Errors May Occur if Blood Glucose Results Are Obtained from

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Align Gluco-Monitoring System (BG1) Quick Start Guide

Align Gluco-Monitoring System (BG1) Quick Start Guide EN Align Gluco-Monitoring System (BG1) Quick Start Guide WELCOME Welcome to your ihealth Align Gluco-Monitoring System (the ihealth system). The ihealth system is designed to be used with the free ihealth

More information

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support POCT13c Glucose Monitoring in Settings Without Laboratory Support This guideline focuses on performance of point-of-care glucose monitoring systems, with an emphasis on safety practices, quality control,

More information

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary

More information

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors Emerging Treatments and Technologies O R I G I N A L A R T I C L E Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors BORIS KOVATCHEV, PHD 1 STACEY ANDERSON, MD 1 2 LUTZ

More information

Variable Measurement, Norms & Differences

Variable Measurement, Norms & Differences Variable Measurement, Norms & Differences 1 Expectations Begins with hypothesis (general concept) or question Create specific, testable prediction Prediction can specify relation or group differences Different

More information

Comparability Study of Online and Paper and Pencil Tests Using Modified Internally and Externally Matched Criteria

Comparability Study of Online and Paper and Pencil Tests Using Modified Internally and Externally Matched Criteria Comparability Study of Online and Paper and Pencil Tests Using Modified Internally and Externally Matched Criteria Thakur Karkee Measurement Incorporated Dong-In Kim CTB/McGraw-Hill Kevin Fatica CTB/McGraw-Hill

More information

Agenda. 15 Dec

Agenda. 15 Dec Implementing Traceability for Heterogeneous Analytes An Industry View JCTLM Symposium on Reference Measurement Systems for Biologicals, December 15, 2004 R. Miller 15 Dec. 2004 1 Agenda Recent trends in

More information

The FreeStyle Libre Flash Glucose Monitoring System. Healthcare Professional Training - London

The FreeStyle Libre Flash Glucose Monitoring System. Healthcare Professional Training - London The FreeStyle Libre Flash Glucose Monitoring System Healthcare Professional Training - London The FreeStyle Libre Flash Glucose Monitoring System Indication of Use The FreeStyle Libre flash glucose monitoring

More information