Proportional Assist Ventilation Andreas Schulze, Peter Schaller, Bernd Höhne, Susanne Herber-Jonat

Size: px
Start display at page:

Download "Proportional Assist Ventilation Andreas Schulze, Peter Schaller, Bernd Höhne, Susanne Herber-Jonat"

Transcription

1 1 Proportional Assist Ventilation Andreas Schulze, Peter Schaller, Bernd Höhne, Susanne Herber-Jonat In proportional assist ventilation (PAV, also referred to as elastic and resistive unloading), the patient actually breaths spontaneously. Every breath is supported by the ventilator which increases the airway pressure. The airway pressure increase "follows" the course of the increase in the breath volume and/or breath flow during every inspiration phase. The assist airway pressure is then returned to the adjusted PEEP during the expiration phase. The pressure increase during every inspiration is proportional to the currently inspired volume and/or inspired breath flow. That means that neither the possible inspiration peak pressure nor the inspiration or expiration times are stipulated. These are defined by the patient's breathing attempt and his breathing pattern, i.e. by his biological breathing regulation system. The ventilator only "follows" the patient (the ventilator is "enslaved"). The breathing pressure curve thus varies typically in time with the changes in the spontaneous breath depth and spontaneous breathing frequency (Fig. PAV1.) 9, 13, 14 O) Paw 2 (cm H (l/min) V Time (s) Fig. 1 PAV1: Proportional assist ventilation for a preterm infant. The diagram shows the ventilation pressure signal (Paw) and the breathing gas flow signal ( V & ) over time. The measurement was taken in close proximity

2 2 to the patient at the Y-piece of the ventilation hose system respectively between the endotracheal tube and connector to the ventilation hose system. The inspiration breath flow is shown above the zero line and the expiration breath flow below it. Note the typical fluctuations in breathing frequency, breath volume, inspiration peak pressure and inspiration time. The operator only adjusts the extent of pressure gain per milliliter breath volume and/or per liter/min breathing gas flow. If this is set to "zero", the patient is breathing under CPAP, i.e. he has to do all the breathing work himself. For a positive gain adjustment, the ventilation pressure increases during inspiration so that the ventilator takes on part of the breathing work. The higher the selected gain per volume and/or flow, the lower the breathing work to be performed by the patient for ventilation. In the end, the aim is to reduce the breathing work for a patient with a diseased lung (e.g. stiff lung and/or increased airway resistance) to such an extent that it only corresponds to the breathing work necessary in a normal lung. That means that the patient can breath with a system of apparently normal mechanical breathing parameters, while the ventilator takes on the additional breathing work required because of the illness. Total breathing work for normal spontaneous breathing consists of an elastic and a resistive part. The first is based on the retraction force of the elastic fibers, which increases during inspiration in proportion to the inspired breath volume, and has to be overcome by the breathing muscles. The resistive part is based on the gas flow resistance within the airways: the faster the flow, the greater the pressure required to generate it, i.e. the necessary pressure is proportional to the magnitude of the breathing gas flow. Breathing work unloading can therefore consist of two different components. Elastic unloading. The ventilation pressure is increased in proportion to the inspiration volume. The gain of elastic unloading, i.e. volume-proportional assist, is consequently adjusted in cm H 2 O/ml (pressure per unit of inspired volume) (Fig. PAV2). 6 Resistive unloading. The ventilation pressure is increased in proportion to the breath flow. The gain of resistive unloading, i.e. flow-proportional assist, is consequently adjusted in cm H 2 O/ (l/s) (pressure per unit of inspired flow) (Fig. PAV3). 5

3 3 Breath volume Δ V Δ V Δ V Zeit Ventilation pressure PEEP Gain B Gain A Δ P Δ P Δ P CPAP: gain=0 Zeit Gain = Δ Ventilation pressure (P, cm H 2 O) Δ Breath volume (V, ml) Fig. PAV2: Diagram to show signal progression in breath volume and ventilation pressure over time for three successive spontaneous breaths with differing strength and differing speed during volume proportional assist ventilation (elastic unloading). The ventilation pressure progression follows the same "form" as breath volume progression because it is controlled proportional to the current amount of breath volume, i.e. the ratio of ventilation pressure and breath volume (ΔP/ΔV) is the same at all points in time. The higher this quotient, the greater the support for the patient's spontaneous breathing compared to the elastic retraction tendency of the lung, so that the ventilator takes on all the more elastic breathing work. At the ventilator, the operator adjusts the quotient in cm H 2 O/ml as so-called gain. The illustration shows three different settings. No support corresponds to zero gain with the patient breathing under CPAP, i.e. the ventilation pressure does not change compared to breath volume. Gain "A" is smaller than gain "B".

4 4 Inspiration Expiration 0 Breath flow Zeit PEEP Ventilation pressure Gain B Gain A 0 Gain = CPAP: Gain=0 Δ Ventilation pressure (cm H 2 O) Δ Breath flow (L/s) Zeit Fig. PAV3: Diagram to show signal progression in breath flow and ventilation pressure over time for three successive spontaneous breaths with differing strength and differing speed during flow proportional assist ventilation (resistive unloading). The ventilation pressure progression follows the same "form" as breath flow progression because it is controlled proportional to the current amount of breath flow, i.e. the ratio of ventilation pressure and breath flow (ΔP/ΔV') is the same at all points in time. The higher this quotient, the greater the support for the patient's spontaneous breathing compared to airway resistance, so that the ventilator takes on all the more resistive breathing work. At the ventilator, the operator adjusts the quotient in cm H 2 O/ml as so-called gain. The illustration shows three different settings. No support corresponds to zero gain with the patient breathing under CPAP, i.e. the ventilation pressure does not change compared to breath volume. Gain "A" is smaller than gain "B. The two thin arrows indicate that the changes in ventilation pressure occur "practically at the same time" as the progression in the breath flow signal. Apart from the safety parameters, only three settings are adjusted for proportional assist ventilation: 1. The PEEP level. This influences the degree of functional residual capacity (FRC) The gain of elastic unloading in cm H 2 O/ml depending on the extent of restrictive lung change The gain of resistive unloading in cm H 2 O/l/s depending on the extent of obstructive lung change. 4, 5, 8

5 5 For combined use of elastic and resistive unloading, at every point in time of inspiration, the ventilation pressure curve is the sum of the pressure values generated by the two individual components. Guidelines for individual selection of mechanical breath unloading gain The simplest and most practical approach to using PAV consists of setting the gain to zero to start with, starting CPAP (or low frequency SIMV) and activating the PAV function. The gain of elastic unloading is increased slowly while at the same time, the operator observes the child's breathing pattern, the extent of thoracic contraction and the airway pressure curve on the ventilator monitor. Generally it can be expected that when the PAV has been correctly adjusted, the peak inspiration pressure lies a few cm H 2 O below the peak pressure presumed necessary for conventional ventilation. 12 A suitable unloading gain can also be detected by a reduction in thoracic contractions with regular spontaneous breathing. 2, 11 Consideration must be given to the fact that smaller children will need higher gains than larger children (in absolute terms, i.e. in cm H 2 O/ml), because compliance and breath volume are related to body weight. That means that children with a body weight < 1000g usually need more than 1 cm H 2 O/ml elastic unloading, while smaller settings are often sufficient in larger children with a comparable extent of restrictive disorder. The gain of resistive unloading can initially be set to the estimated resistance value of the endotracheal tube, to compensate at least for the increase in resistive breathing work by the child caused by the endotracheal tube (a 2.5 mm ID endotracheal tube has a resistance of about 25 cm H 2 O/l/s). When the PAV is correctly set, preterm and newborn infants typically breath "fast and flat" with breathing rates frequently reaching 60 to 100/min, and with breath volumes of around or below 5 ml/kg body weight. The inspiration times are short, usually around s. Some preterm infants reveal somewhat higher PaCO 2 volumes under PAV than with conventional ventilation with customary settings. But clinical studies have revealed that this 3, 12 difference in PaCO 2 has no statistical significance. If the gain is set too high, this can result in overcompensation of the elastic retraction force of the lung or of the airway resistance. In this case, an initial airway pressure gain tends to assume a life of its own through the cycle of positive feedback from the airway pressure gain, subsequent increase in lung volume respectively breath flow and resulting further airway pressure gain. Once a breath has begun, under these conditions this causes passive inflation of the lung (even without any further contribution from the patient) until the set upper limits for

6 6 airway pressure or breath volume are reached. Proportional assist ventilation then changes into assist/control. 10 Overcompensation (excessive gain) for elastic unloading is revealed by the airway pressure curve. At the start of inspiration, the pressure increases rapidly in the automatic mode until the upper airway pressure limit setting is reached (or the default limit for breath volume) (Fig. PAV4). In the Stephanie infant ventilator, the airway pressure is then immediately unloaded to the PEEP setting (in the new software version, the airway pressure initially remains on the level of the upper pressure limit and is only reduced when the patient begins the expiration phase or at the end of a maximum time period set to 0.7 s). The greater the overcompensating gain, the faster the increase in insufflatory pressure. O) Paw 2 (cm H Upper airway pressure limit (L/min) V Time (s) Fig. PAV4: Proportional assist ventilation in a preterm infant with well adjusted elastic unloading (left) and elastic overcompensation (right). At the start of a spontaneous breath or when there is an artifact on the flow signal, which the unit interprets as the start of a breath, in overcompensation the airway pressure quickly increases automatically through to the upper pressure limit setting. Once the airway pressure has then been reduced to the PEEP setting, the cycle can be repeated and finally cause excessive ventilation. If overcompensation is detected, the unloading gain should be reduced to avoid excessive ventilation. Overcompensation applies in elastic unloading when the adjusted ventilation pressure gain per breath volume unit (in cm H 2 O/ml) is greater than the simultaneous increase in elastic retraction force of the lung per volume unit. An elastic unloading gain which was adjusted at one point in time to an appropriate setting can therefore become overcompensating

7 7 unloading when there is an improvement in lung compliance. The gain should therefore be reduced appropriately when the lung disease improves. The upper airway pressure limit setting protects the patient from excessive pressure also in the following situations under proportional assist ventilation: Sighing breaths or hiccoughs with high breath volumes which could cause increased pressures under persistent proportional airway pressure gain (Fig. PAV5) Larger leak in the endotracheal tube. The leak flow simulates a larger inspiration volume without this actually reaching the lung. O) Paw (cm 2 H Upper airway pressure limit (L/min) V sigh Time (s) Fig. PAV5: Preterm infant under proportional assist ventilation. The airway pressure also increases proportional to the large volume of a sigh but not beyond the upper airway pressure limit setting. Here it is reduced to the PEEP. Backup ventilation for apnea. During proportional assist ventilation, in the case of apnea, conventional, adequately effective passive ventilation must start in good time (Fig. PAV6). The previous software for the Stephanie infant ventilator presumes apnea when the breath flow signal follows the zero line, but also when all breath volumes ascertained according to the flow signal are so small that they remain under a stipulated minimum. The default for this lower limit for detecting a change in breath volume is set to 1 ml but can be changed (menu "Options" > "Minimum VT" > ). In this case, the breath volume measured during expiration is measured because only this certainly represents the volume which was

8 8 actually in the lung (by contrast, the volume measured during inspiration can contain a leak component which has no effect on the exchange of gases). Passive conventional ventilation then begins after a set delay period (called "apnea time" in the new software). The default delay time which is normally set to 2 s can be changed in the menu "Options" > "Backup delay time" > "Time in seconds; select from 1, 2, 4, 8, 16 seconds". O) Paw 2 (cm H backup backup ventilation (L/min) V Time (s) Fig. PAV6: Passive, conventional mechanical ventilation (*, backup ventilation) is triggered on ascertaining apnea under proportional assist ventilation. The first spontaneous breath to be identified as such during backup ventilation terminates conventional ventilation and returns to proportional assist ventilation (older software version). In the case of apnea under PAV, arterial hemoglobin O 2 saturation falls faster depending on how low the child's functional residual capacity is, i.e. the less oxygen is held in the lung. The setting for the backup delay time (the new software version calls it "apnea time" or "backup begin after apnea time" which can be adjusted in steps of 0.5 seconds) should therefore be adjusted to shorter intervals for more critical pulmonary diseases. Backup ventilation is adjusted by the operator as normal for conventionally controlled ventilation with regard to inspiration peak pressure (Pmax setting adjustment), inspiration and expiration time and the form of inspiration pressure gain (linear, sinusoidal or quasi rectangular gain). The settings should be adjusted so that any possible oxygen de-saturation occurring after the backup delay time (new software: apnea time) can be rectified without resulting in increased hyperventilation of the child. Hyperventilation could result in total suppression of

9 9 spontaneous breathing activity so that it is not possible to return to PAV. The first spontaneous breath to be identified as such by the unit during backup ventilation terminates conventional ventilation and returns to proportional assist ventilation (older software version). The expiration time set for backup ventilation should therefore be long enough (even when using the new software) to allow the patient to take spontaneous breaths between backup insufflation. New software version with improved forms of backup ventilation under PAV. In preterm infants, the first spontaneous breaths after apnea are usually smaller and less regular than during a preceding period with sufficient spontaneous breathing. If backup ventilation is switched off completely during the first breath after apnea, as was the case with the previous software version, the initial spontaneous breathing frequently transforms into apnea again. This can occur particularly for existing or progressive oxygen de-saturation which in turn depress spontaneous breathing in preterm infants (paradox hypoxia in the newborn) (Fig. PAV7). Fig. PAV7. PAV showing the breath flow signal (V ) and ventilation pressure (Paw) in a preterm infant with 800 g birthweight with bouts of apnea. The backup ventilation (previous software version) is triggered after the first apnea phase but is switched off again completely after two ventilation strokes (asterisks) because of small spontaneous breathing attempts which are not sufficient to stop the incipient fall in oxygen saturation (SpO 2 ). They rapidly become apnea again. This repeats several times, with a constant decrease in oxygen saturation. The new software version therefore contains a minimum time period for full backup ventilation ("backup time") together with only gradual reduction of the backup ventilation frequency on the return of spontaneous breathing(fig. PAV8).

10 10 During the return to spontaneous breathing after apnea, it is therefore advisable to warrant additional breathing support for a limited period of time. 1 The new software version contains the following standard functions for backup ventilation: 1. It begins with full backup ventilation as soon as no spontaneous breathing was detected for a stipulated period of time. This period of time stipulated in the unit is called "apnea time". The default setting is 4 seconds and can be modified by the operator in steps of 0.5 seconds in the range from 0.5 to 15 seconds in the menu "Apnea time > backup begins after... seconds". Spontaneous breathing is detected only by the trigger function, i.e. depending on the adjusted trigger limit from the flow signal from inspiration. 2. Backup ventilation is pressure-controlled, time-limited and synchronized if spontaneous breaths occur. The ventilation parameters are adjusted on the front panel of the unit at the knobs for insufflation pressure (P max ), insufflation time and expiration time (frequency) etc. 3. Once backup ventilation has been triggered, it runs initially for a set minimum period of time, i.e. it cannot be interrupted again by one single recurrence of the trigger. This minimum period of time is called "backup time". The default setting is 10 seconds and can be modified by the operator in steps of 5 seconds in the range from 5 to 60 seconds in the "Backup time" menu. 4. The end of backup ventilation is initiated when the trigger is activated by spontaneous inspiration after the end of the backup period. Backup ventilation is brought to a gradual close (Fig. PAV8) by reducing the backup ventilation frequency step by step, in each case after the end of a time period which is the same as the "backup time" setting. The ventilation frequency set on the unit which constitutes the initial backup frequency is reduced in three steps of 50%, 33% and finally 20% of the initial frequency until the backup support ceases completely. The initial form of complete backup ventilation starts up again immediately if "apnea" is ascertained during the weaning phase, i.e. if the trigger is not activated during a period of time corresponding to the stipulated "apnea time".

11 11 Fig. PAV8: The frequency of backup ventilation (asterisks) is gradually reduced as spontaneous breathing returning after apnea gets increasingly stronger. Spontaneous breathing is supported by PAV already in this weaning phase. Preterm infant weighing 840 g under PAV; new software version. The stated standard functions for backup ventilation can be supplemented by the Stefanie unit thanks to the new software which now enables it to accept a pulse oximetry signal (from the NEOSID Nova unit, F. Stephan GmbH, Massimo Technology) to detect oxygen de-saturation and optimize backup ventilation accordingly. Controlling backup ventilation just on the basis of mechanical ventilation criteria is not capable of providing a fully satisfactory solution in certain situations. Oxygen de-saturation resulting from hypoventilation or other causes can precede apnea and even cause or prolong such conditions (paradox hypoxia reaction in the newborn) (Fig. PAV9).

12 12 Fig. PAV9: In spite of regular and strong spontaneous breathing, a fall in oxygen saturation occurs, followed by apnea. This sequence corresponds to so-called paradox hypoxia reaction in the newborn where hypoxia is a secondary cause of breath depression. Here the backup ventilation was controlled purely by mechanical ventilation criteria. When measured values fall below a stipulated pulse oximetry limit, the new software is capable of initiating backup ventilation at this early stage, thus anticipating breath depression in these situations. It is therefore appropriate and, according to clinical tests, effective to start beginning backup ventilation every time the measured values fall below a set lower pulse oximetry limit. 1 The gradual reduction in backup ventilation does not start until this limit value has been exceeded again, at least the set backup period has finished and spontaneous inspiration has been detected. Another new option in the PAV mode for the "Stephanie" unit consists in parallel additional support for spontaneous breathing during periods with low tidal volume. This form of additional support is called "minimum volume guarantee". The user can adjust a minimum target tidal volume for spontaneous breaths using the tidal volume knob (V T ). If the volume falls below this level, the unit gradually adds more ventilation pressure with every following breath in addition to the pressure generated by the PAV function until the target volume is reached. If this target volume is exceeded later by increasingly stronger inspiration efforts, then the additionally supporting inspiration pressure is decreased gradually and then reduced completely so that the PAV function on its own reaches or exceeds the target volume. When the "minimum volume guarantee" function is activated and the tidal volume falls below the set target, the inspiration ventilation pressure no longer progresses just proportional to the inspiration diaphragm muscle effort (i.e. endogenous

13 13 breath drive expressed in the volume and flow signal of spontaneous breathing) but is then out of proportion to the effort. A minimum tidal volume is then continuously enforced unless the spontaneous breathing is so weak that the inspiration flow no longer reaches the adjusted trigger limit line, i.e. spontaneous breaths are no longer detected so that pressurecontrolled backup ventilation starts at the end of the apnea time, as described above. Up to now, the "minimum volume guarantee" function has not been studied in animal experiments or clinical studies. According to theoretical model tests, it effectively prevents longer periods of breaths below the set target level without destabilizing spontaneous breathing activity during PAV. Small preterm infants develop a fast, flat breathing pattern under PAV. It is not known how the "minimum volume guarantee" function modifies this endogenous breathing pattern, particularly depending on the level selected for the minimum volume, or whether for example there is any influence on the arterial CO 2 Wert and its fluctuation. The problem of endotracheal tube leaks under PAV. A leak between the endotracheal tube and the trachea causes breath to flow without contributing to the breath volume. But the "leak flow" passes the flow sensor during inspiration (Fig. PAV10) so that it is initially measured in with the inspiration flow. Leak flow can therefore "simulate" inspiration. Fig. PAV1: "Leak flow" between endotracheal tube and trachea (green, bent arrows) flows initially through the flow sensor and endotracheal tube but then it leaks out without reaching the lung as effective breath volume. The volume of the leak therefore "simulates" inspiration volume at the flow sensor. Under proportional assist ventilation, a leak flow can therefore result in an increase in airway pressure which is out of proportion to the increase in actual lung volume (respectively to the actual breath flow in the lung). A large leak of this kind can hinder proportional assist ventilation, particularly when the leak also varies in size. For an excessive leak, the airway

14 14 pressure changes rapidly up and down between PEEP and the set upper pressure limit. A small leak typically only occurs at the end of the inspiration phase when the airway pressure has increased to such an extent that expansion of the trachea causes a sudden opening at the endotracheal tube, which then closes again during expiration under lower airway pressure. In this case, airway pressure under proportional assist ventilation initially shows typical progression but then suddenly increases (together with the flow signal) at the end of the inspiration phase. The software of the Stefanie ventilator estimates the size of the leak flow using an algorithm and corrects the measured flow signal internally up to a certain degree. This permits PAV to continue in spite of small to medium leaks at the endotracheal tube (Fig. PAV11). O) Paw 2 (cm H V T (ml) Time (s) Fig. PAV11: Proportional assist ventilation in a 12 week old preterm infant with severe chronic lung disease. The diagram (bottom half) shows the volume signal measured between endotracheal tube and ventilation tube system (V T, obtained as integral of the flow signal over time). The volume measured during inspiration is far larger than the expiration volume because of the leak at the endotracheal tube. The volume signal is reset to the zero limit by the unit software at the start of every inspiration phase because otherwise it would increase constantly with every leak. In this illustrated case, the leak is > 50% of the breath volume, but proportional assist ventilation is still possible. The size of the endotracheal tube and the amount of ventilation pressure define the size of the leak. It usually increases clearly, the longer ventilation lasts.

15 15 Supplementary remarks on safety settings. The operator must adjust the following in every case. The upper limit for obtainable airway pressure (at the 'Pmax' know). This pressure limit should be a few cm H 2 O above the inspiration peak pressure resulting during PAV with well adjusted gain. This also defines the inspiration peak pressure used during backup ventilation. The knob for breath volume on the Stefanie respirator acts as an additional safeguard during PAV to prevent excessive airway pressure. The airway pressure (for elastic unloading) is not increased any further on reaching the breath volume adjusted by the VT knob. This setting should therefore be about 5 10 ml higher than the expected "normal" breath volume for PAV (about 4-6 ml/kg). If the VT setting is adjusted too low by mistake, the unit will not build up sufficient inspiration pressure if any, in spite of the presence of spontaneous breathing (knob turned right round to the left). Guaranteed by the unit without additional settings The maximum possible length of the inspiration time is limited by the software. If insufflation lasts longer than 0.7 s, the airway pressure is reduced to the PEEP. LITERATURE 1. Herber-Jonat S, Rieger-Fackeldey E, Hummler H, Schulze A. Adaptive mechanical backup ventilation for preterm infants on respiratory assist modes. Intensive Care Med, eingereicht Musante G, Schulze A, Gerhardt T, Everett R, Claure N, Schaller P, Bancalari E. Respiratory mechanical unloading decreases thoraco-abdominal asynchrony and chest wall distortion in preterm infants. Pediatr Res 2001; 49: Rieger-Fackeldey E, Gerhardt T, Claure N, Everett R, Schulze A, Bancalari E. Crossover comparison of proportional assist ventilation and synchronized intermittent mandatory ventilation in very low birthweight infants with chronic lung disease. Pediatr Res 2001; 49:281A. 4. Schulze A, Schaller P, Dinger J, Winkler U, Gmyrek D. A method of calculating total respiratory system compliance from resonant frequency: validity in a rabbit model. Pediatr Res 1990; 28:

16 16 5. Schulze A, Schaller P, Gehrhardt B, Mädler HJ, Gmyrek D. An infant ventilator technique for resistive unloading during spontaneous breathing. Results in a rabbit model of airway obstruction. Pediatr Res 1990; 28: Schulze A, Schaller P, Jonzon A, Sedin G. Assisted mechanical ventilation using elastic unloading: A study in cats with normal and injured lungs. Pediatr Res 1993; 34: Schulze A, Schaller P, Töpfer A, Kirpalani H. Resistive and elastic unloading to assist spontaneous breathing does not change functional residual capacity. Pediatr Pulmonol 1993; 16: Schulze A, Jonzon A, Schaller P, Sedin G. Effects of ventilator resistance and compliance on phrenic nerve activity in spontaneously breathing cats. Am J Respir Crit Care Med 1996; 153: Schulze A, Schaller P. Assisted mechanical ventilation using resistive and elastic unloading. Semin Neonatol 1997; 2: Schulze A, Rich W, Schellenberg L, Schaller P, Heldt GP. Effects of different gain settings during assisted mechanical ventilation using respiratory unloading in rabbits. Pediatr Res 1998; 44: Schulze A, Suguihara C, Gerhardt T, Schaller P, Claure N, Everett R, Bancalari E. Effects of respiratory mechanical unloading on thoracoabdominal motion in meconium-injured piglets and rabbits. Pediatr Res 1998; 43: Schulze A, Gerhardt T, Musante G, Schaller P, Claure N, Everett R, Gomez-Marin O, Bancalari E. Proportional assist ventilation in low birth weight infants with acute respiratory disease. A comparison to assist/control and conventional mechanical ventilation. J Pediatr 1999; 135: Schulze A, Bancalari E. Proportional assist ventilation in infants. Clin Perinatol 2001; 28: Schulze A. Respiratory mechanical unloading and proportional assist ventilation in infants. Acta Paediatr Suppl 2002; 91:19-22.

Original Paper. Neonatology 2007;92:1 7 DOI: /

Original Paper. Neonatology 2007;92:1 7 DOI: / Original Paper Neonatology 7;9:1 7 DOI: 1.1159/9876 Received: April 4, 6 Accepted after revision: September 11, 6 Published online: January, 7 Randomized Crossover Comparison of Proportional Assist Ventilation

More information

Adaptive mechanical backup ventilation for preterm infants on respiratory assist modes a pilot study

Adaptive mechanical backup ventilation for preterm infants on respiratory assist modes a pilot study Intensive Care Med (2006) 134:302 308 DOI 10.1007/s00134-005-0003-7 PEDIATRIC ORIGINAL Susanne Herber-Jonat Esther Rieger-Fackeldey Helmut Hummler Andreas Schulze Adaptive mechanical backup ventilation

More information

King s Research Portal

King s Research Portal King s Research Portal DOI: 10.1007/s00431-015-2595-4 Document Version Peer reviewed version Link to publication record in King's Research Portal Citation for published version (APA): Shetty, S., Bhat,

More information

King s Research Portal

King s Research Portal King s Research Portal DOI: 10.1007%2Fs00431-015-2595-4 Document Version Peer reviewed version Link to publication record in King's Research Portal Citation for published version (APA): Shetty, S., Bhat,

More information

Dr. AM MAALIM KPA 2018

Dr. AM MAALIM KPA 2018 Dr. AM MAALIM KPA 2018 Journey Towards Lung protection Goals of lung protection Strategies Summary Conclusion Before 1960: Oxygen; impact assessed clinically. The 1960s:President JFK, Ventilators mortality;

More information

Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo

Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo Instant dowload and all chapters Test Bank Pilbeam's Mechanical Ventilation Physiological and Clinical Applications 6th Edition Cairo https://testbanklab.com/download/test-bank-pilbeams-mechanical-ventilation-physiologicalclinical-applications-6th-edition-cairo/

More information

Original Article. Effects of Pressure Support during an Acute Reduction of Synchronized Intermittent Mandatory Ventilation in Preterm Infants

Original Article. Effects of Pressure Support during an Acute Reduction of Synchronized Intermittent Mandatory Ventilation in Preterm Infants Original Article Effects of Pressure Support during an Acute Reduction of Synchronized Intermittent Mandatory Ventilation in Preterm Infants Waldo Osorio, MD Nelson Claure, PhD Carmen D Ugard, RRT Kamlesh

More information

Ventilator Waveforms: Interpretation

Ventilator Waveforms: Interpretation Ventilator Waveforms: Interpretation Albert L. Rafanan, MD, FPCCP Pulmonary, Critical Care and Sleep Medicine Chong Hua Hospital, Cebu City Types of Waveforms Scalars are waveform representations of pressure,

More information

STATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS

STATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS 3K NON-INVASIVE POSITIVE PRESSURE VENTILATION (NIPPV) ADULT EMT EMT-INTERMEDIATE 85 ADVANCED EMT PARAMEDIC Indications: 1. Dyspnea Uncertain Etiology Adult. 2. Dyspnea Asthma Adult. 3. Dyspnea Chronic

More information

GE Healthcare. Non Invasive Ventilation (NIV) For the Engström Ventilator. Relief, Relax, Recovery

GE Healthcare. Non Invasive Ventilation (NIV) For the Engström Ventilator. Relief, Relax, Recovery GE Healthcare Non Invasive Ventilation (NIV) For the Engström Ventilator Relief, Relax, Recovery COPD is currently the fourth leading cause of death in the world, and further increases in the prevalence

More information

TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT:

TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT: fabian HFO Quick guide TO THE OPERATOR AND PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT: This Quick Guide is not a substitute for the Operation Manual. Read the Operation Manual carefully before

More information

Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor

Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor Mechanical Ventilation Prepared by : Bayan Kaddourah RN,MHM. GICU Clinical Instructor 1 Definition Is a supportive therapy to facilitate gas exchange. Most ventilatory support requires an artificial airway.

More information

Comparison of patient spirometry and ventilator spirometry

Comparison of patient spirometry and ventilator spirometry GE Healthcare Comparison of patient spirometry and ventilator spirometry Test results are based on the Master s thesis, Comparison between patient spirometry and ventilator spirometry by Saana Jenu, 2011

More information

Carina. The compact wonder. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Carina. The compact wonder. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care Carina The compact wonder Emergency Care Perioperative Care Critical Care Perinatal Care Home Care The new sub-acute ventilator Carina Sub-acute care is a rapidly growing medical care service for patients

More information

Simulation 3: Post-term Baby in Labor and Delivery

Simulation 3: Post-term Baby in Labor and Delivery Simulation 3: Post-term Baby in Labor and Delivery Opening Scenario (Links to Section 1) You are an evening-shift respiratory therapist in a large hospital with a level III neonatal unit. You are paged

More information

Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014

Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014 Capnography for Pediatric Procedural Sedation Learning Module Last revised: February 18, 2014 Capnography 40 Non-invasive device that continually monitors EtCO 2 While pulse oximetry measures oxygen saturation,

More information

Volume Guarantee Initiation and ongoing clinical management of an infant supported by Volume Guarantee A Case Study

Volume Guarantee Initiation and ongoing clinical management of an infant supported by Volume Guarantee A Case Study D-32084-2011 Volume Guarantee Initiation and ongoing clinical management of an infant supported by Volume Guarantee A Case Study Robert DiBlasi RRT-NPS, FAARC Respiratory Care Manager of Research & Quality

More information

Provide guidelines for the management of mechanical ventilation in infants <34 weeks gestation.

Provide guidelines for the management of mechanical ventilation in infants <34 weeks gestation. Page 1 of 5 PURPOSE: Provide guidelines for the management of mechanical ventilation in infants

More information

NAVA. In Neonates. Howard Stein, M.D. Director Neonatology. Neurally Adjusted Ventilatory Assist. Toledo Children s Hospital Toledo, Ohio

NAVA. In Neonates. Howard Stein, M.D. Director Neonatology. Neurally Adjusted Ventilatory Assist. Toledo Children s Hospital Toledo, Ohio NAVA Neurally Adjusted Ventilatory Assist In Neonates Howard Stein, M.D. Director Neonatology Toledo Children s Hospital Toledo, Ohio Disclaimers Dr Stein: Is discussing products made by Maquet Has no

More information

The Art and Science of Weaning from Mechanical Ventilation

The Art and Science of Weaning from Mechanical Ventilation The Art and Science of Weaning from Mechanical Ventilation Shekhar T. Venkataraman M.D. Professor Departments of Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine Some definitions

More information

Mechanical Ventilation 1. Shari McKeown, RRT Respiratory Services - VGH

Mechanical Ventilation 1. Shari McKeown, RRT Respiratory Services - VGH Mechanical Ventilation 1 Shari McKeown, RRT Respiratory Services - VGH Objectives Describe indications for mcvent Describe types of breaths and modes of ventilation Describe compliance and resistance and

More information

Non Invasive Ventilation In Preterm Infants. Manuel Sanchez Luna Hospital General Universitario Gregorio Marañón Complutense University Madrid

Non Invasive Ventilation In Preterm Infants. Manuel Sanchez Luna Hospital General Universitario Gregorio Marañón Complutense University Madrid Non Invasive Ventilation In Preterm Infants Manuel Sanchez Luna Hospital General Universitario Gregorio Marañón Complutense University Madrid Summary Noninvasive ventilation begings in the delivery room

More information

Competency Title: Continuous Positive Airway Pressure

Competency Title: Continuous Positive Airway Pressure Competency Title: Continuous Positive Airway Pressure Trainee Name: ------------------------------------------------------------- Title: ---------------------------------------------------------------

More information

Potential Conflicts of Interest

Potential Conflicts of Interest Potential Conflicts of Interest Patient Ventilator Synchrony, PAV and NAVA! Bob Kacmarek PhD, RRT Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts 4-27-09 WSRC Received research

More information

NON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV)

NON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV) Table 1. NIV: Mechanisms Of Action Decreases work of breathing Increases functional residual capacity Recruits collapsed alveoli Improves respiratory gas exchange Reverses hypoventilation Maintains upper

More information

Capnography (ILS/ALS)

Capnography (ILS/ALS) Capnography (ILS/ALS) Clinical Indications: 1. Capnography shall be used as soon as possible in conjunction with any airway management adjunct, including endotracheal, Blind Insertion Airway Devices (BIAD)

More information

VENTILATOR GRAPHICS ver.2.0. Charles S. Williams RRT, AE-C

VENTILATOR GRAPHICS ver.2.0. Charles S. Williams RRT, AE-C VENTILATOR GRAPHICS ver.2.0 Charles S. Williams RRT, AE-C Purpose Graphics are waveforms that reflect the patientventilator system and their interaction. Purposes of monitoring graphics: Allow users to

More information

Weaning and extubation in PICU An evidence-based approach

Weaning and extubation in PICU An evidence-based approach Weaning and extubation in PICU An evidence-based approach Suchada Sritippayawan, MD. Div. Pulmonology & Crit Care Dept. Pediatrics Faculty of Medicine Chulalongkorn University Kanokporn Udomittipong, MD.

More information

APRV Ventilation Mode

APRV Ventilation Mode APRV Ventilation Mode Airway Pressure Release Ventilation A Type of CPAP Continuous Positive Airway Pressure (CPAP) with an intermittent release phase. Patient cycles between two levels of CPAP higher

More information

New Modes and New Concepts In Mechanical Ventilation

New Modes and New Concepts In Mechanical Ventilation New Modes and New Concepts In Mechanical Ventilation Prof Department of Anesthesia and Surgical Intensive Care Cairo University 1 2 New Ventilation Modes Dual Control Within-a-breath switches from PC to

More information

1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation.

1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation. Chapter 1: Principles of Mechanical Ventilation TRUE/FALSE 1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation. F

More information

Respiratory Physiology In-Lab Guide

Respiratory Physiology In-Lab Guide Respiratory Physiology In-Lab Guide Read Me Study Guide Check Your Knowledge, before the Practical: 1. Understand the relationship between volume and pressure. Understand the three respiratory pressures

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3

More information

EMS System for Metropolitan Oklahoma City and Tulsa 2017 Medical Control Board Treatment Protocols

EMS System for Metropolitan Oklahoma City and Tulsa 2017 Medical Control Board Treatment Protocols S O EMT EMT-INTERMEDIATE 85 ADVANCED EMT PARAMEDIC 3H WAVEFORM CAPNOGRAPHY ADULT & PEDIATRIC Indications: 1. Medical General Assessment/General Supportive Care. 2. Trauma General Assessment/Trauma & Hypovolemic

More information

17400 Medina Road, Suite 100 Phone: Minneapolis, MN Fax:

17400 Medina Road, Suite 100 Phone: Minneapolis, MN Fax: 17400 Medina Road, Suite 100 Phone: 763-398-8300 Minneapolis, MN 55447-1341 Fax: 763-398-8400 www.pulmonetic.com Clinical Bulletin To: Cc: From: Domestic Sales Representatives and International Distributors

More information

October Paediatric Respiratory Workbook APCP RESPIRATORY COMMITTEE

October Paediatric Respiratory Workbook APCP RESPIRATORY COMMITTEE October 2017 Paediatric Respiratory Workbook APCP RESPIRATORY COMMITTEE This workbook is designed to introduce to you the difference between paediatric and adult anatomy and physiology. It will also give

More information

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients

Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Objectives Describe nocturnal ventilation characteristics that may indicate underlying conditions and benefits of bilevel therapy for specific

More information

The Influence of Altered Pulmonarv

The Influence of Altered Pulmonarv The Influence of Altered Pulmonarv J Mechanics on the Adequacy of Controlled Ventilation Peter Hutchin, M.D., and Richard M. Peters, M.D. W ' hereas during spontaneous respiration the individual determines

More information

Reasons Providers Use Bilevel

Reasons Providers Use Bilevel Reasons Providers Use Bilevel More comfort, improve therapy compliance Noncompliant OSA (NCOSA) 1 Scripts from lab referrals Central/Complex Sleep Apnea 2 For ventilations needs Restrictive Thoracic Disorders/Neuromuscular

More information

Mechanical Ventilation Principles and Practices

Mechanical Ventilation Principles and Practices Mechanical Ventilation Principles and Practices Dr LAU Chun Wing Arthur Department of Intensive Care Pamela Youde Nethersole Eastern Hospital 6 October 2009 In this lecture, you will learn Major concepts

More information

NAVA-korzyści dla noworodka

NAVA-korzyści dla noworodka DISCLOSURE No conflict of interest related to this topic NAVA-korzyści dla noworodka Jan Mazela Poznan University of Medical Sciences Poznan, Poland EUROPE POZNAŃ and WIELKOPOLSKA REGION POLAND WIELKOPOLSKA

More information

Lecture Notes. Chapter 2: Introduction to Respiratory Failure

Lecture Notes. Chapter 2: Introduction to Respiratory Failure Lecture Notes Chapter 2: Introduction to Respiratory Failure Objectives Define respiratory failure, ventilatory failure, and oxygenation failure List the causes of respiratory failure Describe the effects

More information

INTRODUCTION The effect of CPAP works on lung mechanics to improve oxygenation (PaO 2

INTRODUCTION The effect of CPAP works on lung mechanics to improve oxygenation (PaO 2 2 Effects of CPAP INTRODUCTION The effect of CPAP works on lung mechanics to improve oxygenation (PaO 2 ). The effect on CO 2 is only secondary to the primary process of improvement in lung volume and

More information

Recognizing and Correcting Patient-Ventilator Dysynchrony

Recognizing and Correcting Patient-Ventilator Dysynchrony 2019 KRCS Annual State Education Seminar Recognizing and Correcting Patient-Ventilator Dysynchrony Eric Kriner BS,RRT Pulmonary Critical Care Clinical Specialist MedStar Washington Hospital Center Washington,

More information

I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device

I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device II. Policy: Continuous Positive Airway Pressure CPAP by the Down's system will be instituted by Respiratory Therapy personnel

More information

Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis

Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Gal S, Riskin A, Chistyakov I, Shifman N, Srugo I, and Kugelman A Pediatric Department and Pediatric Pulmonary Unit Bnai Zion

More information

Chapter 11 The Respiratory System

Chapter 11 The Respiratory System Biology 12 Name: Respiratory System Per: Date: Chapter 11 The Respiratory System Complete using BC Biology 12, page 342-371 11.1 The Respiratory System pages 346-350 1. Distinguish between A. ventilation:

More information

CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEFINITION

CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEFINITION CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEFINITION Method of maintaining low pressure distension of lungs during inspiration and expiration when infant breathing spontaneously Benefits Improves oxygenation

More information

Table 1: The major changes in AHA / AAP neonatal resuscitation guidelines2010 compared to previous recommendations in 2005

Table 1: The major changes in AHA / AAP neonatal resuscitation guidelines2010 compared to previous recommendations in 2005 Table 1: The major changes in AHA / AAP neonatal guidelines2010 compared to previous recommendations in 2005 Resuscitation step Recommendations (2005) Recommendations (2010) Comments/LOE 1) Assessment

More information

Minimizing Lung Damage During Respiratory Support

Minimizing Lung Damage During Respiratory Support Minimizing Lung Damage During Respiratory Support University of Miami Jackson Memorial Medical Center Care of the Sick Newborn 15 Eduardo Bancalari MD University of Miami Miller School of Medicine Jackson

More information

Objectives. Apnea Definition and Pitfalls. Pathophysiology of Apnea. Apnea of Prematurity and hypoxemia episodes 5/18/2015

Objectives. Apnea Definition and Pitfalls. Pathophysiology of Apnea. Apnea of Prematurity and hypoxemia episodes 5/18/2015 Apnea of Prematurity and hypoxemia episodes Deepak Jain MD Care of Sick Newborn Conference May 2015 Objectives Differentiating between apnea and hypoxemia episodes. Pathophysiology Diagnosis of apnea and

More information

Weaning from Mechanical Ventilation. Dr Azmin Huda Abdul Rahim

Weaning from Mechanical Ventilation. Dr Azmin Huda Abdul Rahim Weaning from Mechanical Ventilation Dr Azmin Huda Abdul Rahim Content Definition Classification Weaning criteria Weaning methods Criteria for extubation Introduction Weaning comprises 40% of the duration

More information

Capnography. Capnography. Oxygenation. Pulmonary Physiology 4/15/2018. non invasive monitor for ventilation. Edward C. Adlesic, DMD.

Capnography. Capnography. Oxygenation. Pulmonary Physiology 4/15/2018. non invasive monitor for ventilation. Edward C. Adlesic, DMD. Capnography Edward C. Adlesic, DMD University of Pittsburgh School of Dental Medicine 2018 North Carolina Program Capnography non invasive monitor for ventilation measures end tidal CO2 early detection

More information

Effect of Body Temperature on the Pattern of Spontaneous Breathing in Extremely Low Birth Weight Infants Supported by Proportional Assist Ventilation

Effect of Body Temperature on the Pattern of Spontaneous Breathing in Extremely Low Birth Weight Infants Supported by Proportional Assist Ventilation 0031-3998/03/5403-0332 PEDIATRIC RESEARCH Vol. 54, No. 3, 2003 Copyright 2003 International Pediatric Research Foundation, Inc. Printed in U.S.A. Effect of Body Temperature on the Pattern of Spontaneous

More information

WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35

WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35 WHAT DO YOU WANT FROM A HOME VENTILATION SYSTEM? 8322_RS_HomeNIV_brochure_v14.ind1 1 4/7/06 12:57:35 D I L E M M A DIFFERENT VENTILATORS DIFFERENT ALGORITHMS TO KNOW YOU VE CHANGED PATIENT LIVES?PATIENT??

More information

Challenging Cases in Pediatric Polysomnography. Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep

Challenging Cases in Pediatric Polysomnography. Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep Challenging Cases in Pediatric Polysomnography Fauziya Hassan, MBBS, MS Assistant Professor Pediatric Pulmonary and Sleep Conflict of Interest None pertaining to this topic Will be using some slides from

More information

Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment

Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment Hypoventilation? Obstructive Sleep Apnea? Different Tests, Different Treatment Judith R. Fischer, MSLS, Editor, Ventilator-Assisted Living (fischer.judith@sbcglobal.net) Thanks to Josh Benditt, MD, University

More information

Hyaline membrane disease. By : Dr. Ch Sarishma Peadiatric Pg

Hyaline membrane disease. By : Dr. Ch Sarishma Peadiatric Pg Hyaline membrane disease By : Dr. Ch Sarishma Peadiatric Pg Also called Respiratory distress syndrome. It occurs primarily in premature infants; its incidence is inversely related to gestational age and

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO 2 ) 3 The Vigileo

More information

Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care

Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care Bergen Community College Division of Health Professions Department of Respiratory Care Fundamentals of Respiratory Critical Care Date Revised: January 2015 Course Description Student Learning Objectives:

More information

RESPIRATORY PHYSIOLOGY Pre-Lab Guide

RESPIRATORY PHYSIOLOGY Pre-Lab Guide RESPIRATORY PHYSIOLOGY Pre-Lab Guide NOTE: A very useful Study Guide! This Pre-lab guide takes you through the important concepts that where discussed in the lab videos. There will be some conceptual questions

More information

Usefulness of DuoPAP in the treatment of very low birth weight preterm infants with neonatal respiratory distress syndrome

Usefulness of DuoPAP in the treatment of very low birth weight preterm infants with neonatal respiratory distress syndrome European Review for Medical and Pharmacological Sciences 2015; 19: 573-577 Usefulness of DuoPAP in the treatment of very low birth weight preterm infants with neonatal respiratory distress syndrome B.

More information

Comparison of automated and static pulse respiratory mechanics during supported ventilation

Comparison of automated and static pulse respiratory mechanics during supported ventilation Comparison of automated and static pulse respiratory mechanics during supported ventilation Alpesh R Patel, Susan Taylor and Andrew D Bersten Respiratory system compliance ( ) and inspiratory resistance

More information

GAS EXCHANGE IB TOPIC 6.4 CARDIOPULMONARY SYSTEM CARDIOPULMONARY SYSTEM. Terminal bronchiole Nasal cavity. Pharynx Left lung Alveoli.

GAS EXCHANGE IB TOPIC 6.4 CARDIOPULMONARY SYSTEM CARDIOPULMONARY SYSTEM. Terminal bronchiole Nasal cavity. Pharynx Left lung Alveoli. IB TOPIC 6.4 GAS EXCHANGE CARDIOPULMONARY SYSTEM CARDIOPULMONARY SYSTEM Branch from the pulmonary artery (oxygen-poor blood) Branch from the pulmonary vein (oxygen-rich blood) Terminal bronchiole Nasal

More information

IB TOPIC 6.4 GAS EXCHANGE

IB TOPIC 6.4 GAS EXCHANGE IB TOPIC 6.4 GAS EXCHANGE CARDIOPULMONARY SYSTEM CARDIOPULMONARY SYSTEM Branch from the pulmonary artery (oxygen-poor blood) Branch from the pulmonary vein (oxygen-rich blood) Terminal bronchiole Nasal

More information

PULMONARY FUNCTION. VOLUMES AND CAPACITIES

PULMONARY FUNCTION. VOLUMES AND CAPACITIES PULMONARY FUNCTION. VOLUMES AND CAPACITIES The volume of air a person inhales (inspires) and exhales (expires) can be measured with a spirometer (spiro = breath, meter = to measure). A bell spirometer

More information

OSA and COPD: What happens when the two OVERLAP?

OSA and COPD: What happens when the two OVERLAP? 2011 ISRC Seminar 1 COPD OSA OSA and COPD: What happens when the two OVERLAP? Overlap Syndrome 1 OSA and COPD: What happens when the two OVERLAP? ResMed 10 JAN Global leaders in sleep and respiratory medicine

More information

Capnography 101. James A Temple BA, NRP, CCP

Capnography 101. James A Temple BA, NRP, CCP Capnography 101 James A Temple BA, NRP, CCP Expected Outcomes 1. Gain a working knowledge of the physiology and science behind End-Tidal CO2. 2.Relate End-Tidal CO2 to ventilation, perfusion, and metabolism.

More information

Lumis Tx: the all-in-one hospital ventilation solution

Lumis Tx: the all-in-one hospital ventilation solution Lumis Tx: the all-in-one hospital ventilation solution The Lumis Tx is a multi-purpose non-invasive ventilator that treats the full range of respiratory conditions and is suitable for a variety of hospital

More information

Weaning: The key questions

Weaning: The key questions Weaning from mechanical ventilation Weaning / Extubation failure: Is it a real problem in the PICU? Reported extubation failure rates in PICUs range from 4.1% to 19% Baisch SD, Wheeler WB, Kurachek SC,

More information

Breathing and pulmonary function

Breathing and pulmonary function EXPERIMENTAL PHYSIOLOGY EXPERIMENT 5 Breathing and pulmonary function Ying-ying Chen, PhD Dept. of Physiology, Zhejiang University School of Medicine bchenyy@zju.edu.cn Breathing Exercise 1: Tests of pulmonary

More information

WILAflow Elite Neonatal Ventilator. Non-invasive treatment for the most delicate patients.

WILAflow Elite Neonatal Ventilator. Non-invasive treatment for the most delicate patients. EN WILAflow Elite Neonatal Ventilator Non-invasive treatment for the most delicate patients. 0197 Infant Ventilation redefined A new generation in Infant Ventilation WILAflow Elite is a microprocessor

More information

Handling Common Problems & Pitfalls During. Oxygen desaturation in patients receiving mechanical ventilation ACUTE SEVERE RESPIRATORY FAILURE

Handling Common Problems & Pitfalls During. Oxygen desaturation in patients receiving mechanical ventilation ACUTE SEVERE RESPIRATORY FAILURE Handling Common Problems & Pitfalls During ACUTE SEVERE RESPIRATORY FAILURE Pravit Jetanachai, MD QSNICH Oxygen desaturation in patients receiving mechanical ventilation Causes of oxygen desaturation 1.

More information

WILAflow Elite Neonatal Ventilator. Non-invasive treatment for the most delicate patients.

WILAflow Elite Neonatal Ventilator. Non-invasive treatment for the most delicate patients. EN WILAflow Elite Neonatal Ventilator Non-invasive treatment for the most delicate patients. 0197 Infant Ventilation redefined A new generation in Infant Ventilation WILAflow Elite is a microprocessor

More information

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation 1 A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation The following 3 minute polysomnogram (PSG) tracing was recorded in a 74-year-old man with severe ischemic cardiomyopathy

More information

Anaesthetic considerations for laparoscopic surgery in canines

Anaesthetic considerations for laparoscopic surgery in canines Vet Times The website for the veterinary profession https://www.vettimes.co.uk Anaesthetic considerations for laparoscopic surgery in canines Author : Chris Miller Categories : Canine, Companion animal,

More information

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy

More information

BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT

BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT BiPAPS/TVAPSCPAPASV???? Lori Davis, B.Sc., R.C.P.T.(P), RPSGT Modes Continuous Positive Airway Pressure (CPAP): One set pressure which is the same on inspiration and expiration Auto-PAP (APAP) - Provides

More information

Stellar 100 Stellar 150

Stellar 100 Stellar 150 Stellar 100 Stellar 150 Invasive and noninvasive ventilator Data Management Guide English The following table shows where data from the Stellar device can be viewed. Data displayed in ResScan can be downloaded

More information

Control of Respiration

Control of Respiration Control of Respiration Graphics are used with permission of: adam.com (http://www.adam.com/) Benjamin Cummings Publishing Co (http://www.awl.com/bc) Page 1. Introduction The basic rhythm of breathing is

More information

BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager

BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager 1 Solutions in Sleep Therapy BPAP 25A Training A.Giudice,RPSGT Clinical Education Manager 2 To access press and hold the On/Off Button and Ramp Button and at the same time connect the power cord into the

More information

Dr. Yasser Fathi M.B.B.S, M.Sc, M.D. Anesthesia Consultant, Head of ICU King Saud Hospital, Unaizah

Dr. Yasser Fathi M.B.B.S, M.Sc, M.D. Anesthesia Consultant, Head of ICU King Saud Hospital, Unaizah BY Dr. Yasser Fathi M.B.B.S, M.Sc, M.D Anesthesia Consultant, Head of ICU King Saud Hospital, Unaizah Objectives For Discussion Respiratory Physiology Pulmonary Graphics BIPAP Graphics Trouble Shootings

More information

9Synchronized and. Mechanical ventilation has improved to the point. Volume-Targeted Ventilation

9Synchronized and. Mechanical ventilation has improved to the point. Volume-Targeted Ventilation Acute Respiratory Care of the Neonate 9Synchronized and Volume-Targeted Ventilation Martin Keszler, MD Mechanical ventilation has improved to the point where few infants now die of acute respiratory failure.

More information

Understanding Breathing Muscle Weakness

Understanding Breathing Muscle Weakness Understanding Breathing Muscle Weakness A N D R E A L. K L E I N P R E S I D E N T / F O U N D E R B R E A T H E W I T H M D w w w.facebook.com/ b r e a t h e w i t h m d h t t p : / / w w w. b r e a t

More information

APPENDIX VI HFOV Quick Guide

APPENDIX VI HFOV Quick Guide APPENDIX VI HFOV Quick Guide Overall goal: Maintain PH in the target range at the minimum tidal volume. This is achieved by favoring higher frequencies over lower P (amplitude). This goal is also promoted

More information

Pedi-Cap CO 2 detector

Pedi-Cap CO 2 detector Pedi-Cap CO 2 detector Presentation redeveloped for this program by Rosemarie Boland from an original presentation by Johnston, Adams & Stewart, (2006) Background Clinical methods of endotracheal tube

More information

Don t let your patients turn blue! Isn t it about time you used etco 2?

Don t let your patients turn blue! Isn t it about time you used etco 2? Don t let your patients turn blue! Isn t it about time you used etco 2? American Association of Critical Care Nurses National Teaching Institute Expo Ed 2013 Susan Thibeault MS, CRNA, APRN, CCRN, EMT-P

More information

GUIDELINE PHYSIOLOGY OF BIRTH ASPHYXIA

GUIDELINE PHYSIOLOGY OF BIRTH ASPHYXIA GUIDELINE PHYSIOLOGY OF BIRTH ASPHYXIA The newborn is not an adult, nor a child. In people of all ages, death can occur from a failure of breathing and / or circulation. The interventions required to aid

More information

Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context

Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context Addendum to the NRP Provider Textbook 6 th Edition Recommendations for specific modifications in the Canadian context A subcommittee of the Canadian Neonatal Resuscitation Program (NRP) Steering Committee

More information

Cardiorespiratory Interactions:

Cardiorespiratory Interactions: Cardiorespiratory Interactions: The Heart - Lung Connection Jon N. Meliones, MD, MS, FCCM Professor of Pediatrics Duke University Medical Director PCVICU Optimizing CRI Cardiorespiratory Economics O2:

More information

Oxygenation. Chapter 45. Re'eda Almashagba 1

Oxygenation. Chapter 45. Re'eda Almashagba 1 Oxygenation Chapter 45 Re'eda Almashagba 1 Respiratory Physiology Structure and function Breathing: inspiration, expiration Lung volumes and capacities Pulmonary circulation Respiratory gas exchange: oxygen,

More information

PRODUCT TRAINING TREND II

PRODUCT TRAINING TREND II Product Training TREND II HOFFRICHTER GmbH 1 General Introduction Complete system spectrum for CPAP therapy TREND II TREND II AUTO TREND II BILEVEL TREND II BILEVEL ST20 AquaTREND uni Therapy Humidifier

More information

Feasibility and physiological effects of noninvasive neurally adjusted ventilatory assist in preterm infants

Feasibility and physiological effects of noninvasive neurally adjusted ventilatory assist in preterm infants Articles Clinical Investigation nature publishing group Feasibility and physiological effects of noninvasive neurally adjusted ventilatory assist in preterm infants Christopher K. Gibu 1,3, Phillip Y.

More information

Lung Wit and Wisdom. Understanding Oxygenation and Ventilation in the Neonate. Jennifer Habert, BHS-RT, RRT-NPS, C-NPT Willow Creek Women s Hospital

Lung Wit and Wisdom. Understanding Oxygenation and Ventilation in the Neonate. Jennifer Habert, BHS-RT, RRT-NPS, C-NPT Willow Creek Women s Hospital Lung Wit and Wisdom Understanding Oxygenation and Ventilation in the Neonate Jennifer Habert, BHS-RT, RRT-NPS, C-NPT Willow Creek Women s Hospital Objectives To review acid base balance and ABG interpretation

More information

By Mark Bachand, RRT-NPS, RPFT. I have no actual or potential conflict of interest in relation to this presentation.

By Mark Bachand, RRT-NPS, RPFT. I have no actual or potential conflict of interest in relation to this presentation. By Mark Bachand, RRT-NPS, RPFT I have no actual or potential conflict of interest in relation to this presentation. Objectives Review state protocols regarding CPAP use. Touch on the different modes that

More information

Apnea Monitors THE FUNDAMENTALS OF...

Apnea Monitors THE FUNDAMENTALS OF... THE FUNDAMENTALS OF... Apnea Monitors Robert M. Dondelinger Apnea is a Greek word meaning without wind. Apnea, in the modern lexicon, refers to the cessation of breathing and is a reversible condition

More information

Monitoring Respiratory Drive and Respiratory Muscle Unloading during Mechanical Ventilation

Monitoring Respiratory Drive and Respiratory Muscle Unloading during Mechanical Ventilation Monitoring Respiratory Drive and Respiratory Muscle Unloading during Mechanical Ventilation J. Beck and C. Sinderby z Introduction Since Galen's description 2000 years ago that the lungs could be inflated

More information

² C Y E N G R E M E ssignac Cardiac Arrest Resuscitation Device uob

² C Y E N G R E M E ssignac Cardiac Arrest Resuscitation Device uob E M E R G E N C Y Boussignac Cardiac Arrest Resuscitation Device ² What is b-card? b-card Boussignac Cardiac Arrest Resuscitation Device has been designed specifically for the treatment of cardiac arrest.

More information

Functional Chest MRI in Children Hyun Woo Goo

Functional Chest MRI in Children Hyun Woo Goo Functional Chest MRI in Children Hyun Woo Goo Department of Radiology and Research Institute of Radiology Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea No ionizing radiation

More information

Proportional Assist Ventilation (PAV) (NAVA) Younes ARRD 1992;145:114. Ventilator output :Triggering, Cycling Control of flow, rise time and pressure

Proportional Assist Ventilation (PAV) (NAVA) Younes ARRD 1992;145:114. Ventilator output :Triggering, Cycling Control of flow, rise time and pressure Conflict of Interest Disclosure Robert M Kacmarek Unconventional Techniques Using Your ICU Ventilator!" 5-5-17 FOCUS Bob Kacmarek PhD, RRT Massachusetts General Hospital, Harvard Medical School, Boston,

More information