1 Alcohol and Alcoholism Advance Access published July 30, 2008 Alcohol & Alcoholism Vol., No. 0, pp. 1 8, 2008 doi: /alcalc/agn058 Automated Measurement of Carbohydrate-Deficient Transferrin Using the Bio-Rad %CDT by the HPLC Test on a Variant TM HPLC System: Evaluation and Comparison with Other Routine Procedures François Schellenberg 1,, Louise Mennetrey 2, Catherine Girre 3, Bertrand Nalpas 4 and Jean Christophe Pagès 1 1 Laboratoire de Biochimie et Biologie Moléculaire, Pôle de Biologie, Hôpital Trousseau, CHRU de Tours, France, 2 CCAA, CHRU, Tours, France, 3 Addictology Unit, Hôpital Fernand Widal, Paris, France and 4 INSERM U567, Hôpital Necker, Paris, France Corresponding author: Laboratoire de Biochimie et Biologie Moléculaire, Pôle de Biologie, Hôpital Trouseau, Tours, France. Tel: ; Fax: ; f.schellenberg@chu-tours.fr (Received 09 November 2007; first review notified 03 January 2008; in revised form 22 May 2008; accepted 24 June 2008) Abstract Aims: In this study, we evaluated the new %CDT by the HPLC method (Bio-Rad, Germany) on a Variant TM HPLC system (Bio-Rad), checked the correlation with well-known methods and calculated the diagnostic value of the test. Methods: Intra-run and day-to-day precision values were calculated for samples with extreme serum transferrin concentrations, high trisialotransferrin and interfering conditions (haemolysed, lactescent and icteric samples). The method was compared with two routine procedures, the %CDT TIA (Bio-Rad, Hercules, CA, USA) and the Capillarys TM CDT (Sebia, France). A total of 350 clinical sera samples were used for a case-control study. Results: Precision values were better in high CDT and medium CDT pools than in low CDT pools. The serum transferrin concentration had no effect on CDT measurement, except in samples with serum transferrin <1 g/l. Haemolysis was the only interfering situation. The method showed high correlation (r 2 > 0.95) with the two other methods (%CDT TIA and CZE %CDT). The global predictive value of the test was >0.90 at 1.9% cut-off. Conclusions: These results demonstrate that the %CDT by the HPLC test is suitable for CDT routine measurement; the results from the high-throughput Variant TM system are well correlated with other methods and are of high diagnostic value. INTRODUCTION Among the biological markers of alcohol abuse investigated over the past decades, CDT (carbohydrate-deficient transferrin) (Stibler, 1991) is considered the most accurate marker for the detection of alcohol abuse and possible relapse (Stibler et al., 1991; Helander 2003; Bortolotti et al., 2006). The main transferrin (Tf) glycoform comprises one polypeptide chain bearing two bi-antennary N-linked oligosaccharide chains, and each antenna is terminated with a sialic acid residue (i.e. four sialic acid residues). The overall number of sialic acid residues (Hatton et al., 1982) is theoretically between 0 (no oligosaccharide chain) and 8 (two tetra-antennae oligosaccharide chains terminated with four sialic acid residues on each). The Tf structure is modified following an alcohol-induced mechanism, resulting in the synthesis of Tf molecules lacking one (disialotransferrin) or both (asialotransferrin) glycans. As disialotransferrin (DiST) is the main modified glycoform (Stibler et al., 1979) and the increase of asialotransferrin (AST) can be observed only in samples with considerable amount of DiST, it was proposed as the target molecule for CDT standardization and calibration (Jeppsson et al., 2007). Over recent years, the use of CDT as a routine test for alcohol abuse has been hampered by laborious and time-consuming methods (Bortolotti et al., 2006). The methods based on microcolumn ion-exchange chromatography allow the global quantification of asialo-, monosialo-, disialo- and a variable amount of trisialotransferrin. The final result is expressed as the ratio between CDT and total Tf. More recently, new methods based on capillary zone electrophoresis (CZE) (Bortolotti et al., 2007; Schellenberg et al., 2007), high-performance liquid chromatography (HPLC) (Jeppsson et al., 1993, Helander and Bergström, 2006) and direct immunonephelometric measurement (Delanghe et al., 2007) have been developed that allow reliable and automated measurement of CDT. HPLC methodology allows the quantitative spectrometric determination of each Tf glycoform and has been proposed as a candidate reference method (Helander et al., 2003). The aim of this study was to evaluate the analytical features and diagnostic value of the commercial kit %CDT by HPLC on the routine high-throughput Variant TM HPLC system (Bio-Rad, Munich, Germany). MATERIALS AND METHODS Analytical procedures %CDT by the HPLC technique. The Bio-Rad %CDT by HPLC was run on a Variant TM HPLC system. This system comprises a two dual-piston pump system, a thermostated (20 C) 100-sample auto sampler and a column compartment (35 C) and a dual wavelength (460 nm and 690 nm) detector. The system is connected to the CDM software (Bio Rad). Before analysis, the samples are pre-treated by 30-min incubation with a ferrous and dextran sulphate solution, followed by 10-min (10,000 g) centrifugation, for both iron saturation and lipid precipitation. The supernatant is then ready for chromatographic analysis. The Tf fractions are separated using an ion-exchange column (600 tests) protected by a guard column (100 tests). The absorbance of the ferrous iron Tf complex is measured at 460 nm; the secondary 690 nm wavelength is used for background noise reduction. The system calculates the area of each Tf glycoform peak following baseline integration from disialo- to pentasialotransferrin. The hexasialotransferrin peak is not integrated; asialotransferrin, when present, is integrated in a separate baseline mode (Fig. 1A and C). The area of each identified peak is expressed as the ratio of its surface to the total under-peak area. The total analysis time is 10 min. %CDT TIA technique. This assay is based on separation of the Tf glycoforms on micro anion exchange chromatography C The Author Published by Oxford University Press on behalf of the Medical Council on Alcohol. All rights reserved

2 2 Schellenberg et al. Fig. 1. Transferrin glycoform pattern obtained using the %CDT by the HPLC test on a Variant TM system. The Y-axis in arbitrary absorbance units. (A) Chromatogram of a homozygote transferrin-c control with a low CDT concentration (1.12%). (B) Chromatogram of a patient with suspected liver cirrhosis, low serum transferrin (1.23 g/l) and DST as a shoulder on the TriST peak, leading to erroneous integration (5.40%). (C) Chromatogram of an alcohol-abusing patient with elevated DST (15.60%) and AST (4.77%) levels. (D) Chromatogram of a serum with high (2 g/l) haemoglobin content. PST: pentasialotransferrin, TeST: tetrasialotransferrin, TriST: trisialotransferrin, DST: disialotransferrin, AST: asialotransferrin. columns followed by nephelometric, or turbidimetric, measurement of the Tf in each eluate, allowing CDT quantification (Schwarz et al., 2003). The Tf glycoforms contained in the CDT eluate are mainly di-, mono- and asialotransferrin, and a minor part of trisialotransferrin (Aldén et al., 2005). In this study, nephelometric determinations were run using an Immage 800 (Beckman Coulter, Brea, CA, USA). The cut-off limit was 2.60% according to the manufacturer s recommendations (Schwan et al., 2004). Capillarys TM CDT technique. This capillary zone electrophoresis (CZE) technique has been developed by SE- BIA (Evry, France) on the routine multicapillary system Capillarys TM (Bortolotti et al., 2007). In this assay, the Tf glycoforms are separated by high voltage electrophoresis (8200 V) in an alkaline buffer (ph 8.8) according to ph and electroosmotic flow. The Tf glycoforms are detected by absorbance measurement at 200 nm wavelength. Samples Six pools of sera with low (1 and 2), medium, critical and high (1 and 2) CDT concentrations were prepared from anonymous routine samples, in which CDT had been previously measured by the routine method %CDT TIA (Bio-Rad). CDT values were <1.9% for the low pools, % for the normal pool, % for the critical pool and >5% and >9% for the high pools 1 and 2, respectively. A panel of 402 samples was analysed in this study: (a) 169 samples from patients with an AUDIT (Alcohol Use Disorders Identification Test) score >10, daily ethanol intake >50 g or

3 Evaluation of the New CDT by the HPLC Method 3 BAC (Blood Alcohol Concentration) >0, which constituted the alcohol abusers group; (b) 112 hospital patients who had no history or suspicion of alcohol abuse and no biological signs (excluding CDT levels) of alcohol abuse, and 69 patients with an AUDIT score <7 and a daily alcohol intake <40 g, constituting a control group (181 patients); (c) 52 samples, surplus of routine examinations, characterized by serum Tf concentration outside the alert values ( g/l) or trisialotransferrin >6% (according to CZE CDT determination). Thirty lipaemic samples were also included upon criteria of high triglyceride concentration, lactescent aspect and high lipemic index determined by a routine chemistry analyser Olympus 2700 (Olympus Corp., Japan). Thirty samples with various CDT concentrations were enriched with nonconjugated bilirubin to a final bilirubin concentration of µmol/l. Thirty samples with various CDT concentrations were enriched with haemoglobin by addition of haemolysed and centrifuged erythrocytes to a final haemoglobin concentration of 1 4 g/l. STUDY DESIGN Precision Intra-run precision was calculated from the repeated measurement of low[1], medium and high[1] pools. These samples were tested 10 times each day for three consecutive days. Intra-day precision was also calculated for each pool. The total precision was determined as described in the CLSI protocol EP5-A2 using low[2], critical and high[2] pools. Each pool was analysed in duplicate 5 days a week by two runs separated by at least 2 h for four consecutive weeks. Interfering situations A. Tf concentration. A panel of samples covering the range of Tf serum concentrations was selected among samples of the (b) and (c) panels to determine if there was a relationship between total Tf concentration and CDT values. The samples were selected according to the following criteria: serum Tf concentrations g/l in 0.1 g/l increments; CDT was assumed as having no reason to be increased, based on a lack of suspected alcohol abuse, according to usual liver parameters (ALT, GGT) and normal CDT values when measured by the routine laboratory method. The relationship between Tf concentration and CDT values was assessed by measuring the slope of the curve CDT = f(tf). B. Serum aspect. Thirty lipaemic samples were analysed in duplicate before and after centrifugation (60,000 g, 30 min). Prior to CDT determination, all samples were submitted to the pre-treatment described by the manufacturer. Thirty samples with various CDT concentrations were enriched with non-conjugated bilirubin to a final bilirubin concentration of µmol/l. CDT levels were measured before and after bilirubin addition using the %CDT by the HPLC technique. Thirty samples with various CDT concentrations were enriched with haemoglobin and the interference with haemolysis was measured by the %CDT by the HPLC technique before and after haemoglobin addition. C. High trisialotransferrin. Thirty samples with a trisialotransferrin (TriST) level >6% of the total Tf, determined by the Capillarys CDT technique, were selected in the (b) and (c) panels to validate the accuracy of the present method in high TriST samples. Correlation study A total of 302 samples from panels (a), (b) and (c) covering the full range of CDT values were analysed in parallel with the %CDT by HPLC, the %CDT TIA and the Capillarys CDT techniques. Highly discrepant results were reanalysed. The methods were compared using the Pearson correlation coefficient and the Passing and Bablock method for regression. The linearity was assessed by the Cusum test for linearity, which tests the random (or not) distribution of the points on each side of the regression line. Case-control study A case-control study was performed to evaluate the diagnostic efficacy of the method and to validate the cut-off proposed by the manufacturer. Samples of panels (a) and (b) were analysed in this purpose. They were 350 participants, aged years, divided into 181 control subjects (115 males, 66 females) and 169 patients with alcohol abuse disorders (AUD) (136 males, 33 females). The participation of these patients and control individuals in the clinical study was approved by the ethical committee of Hôpital Fernand Widal (Paris, France). The %CDT by the HPLC procedure was compared to the %CDT TIA and Capillarys CDT procedures. Statistical calculations were performed using MedCalc 9.2 (MedCalc Software, Mariakerke, Belgium) and Analyse-it (Analyse-it Software, Leeds, UK) software. Pearson s correlation coefficient for non-gaussian distribution was used in the correlation studies. RESULTS Precision Intra-run precision (Table 1) was inversely correlated to the CDT concentrations of the sample. In the low[1] pool, individual measurements varied from simple to double between runs, resulting in daily coefficient of variation (CV) ranging from 9.01% to 19.01%. The precision of CDT determinations in samples with higher CDT concentrations was better, with CVs ranging from 7.84% to 11.74% in the medium pool and 1.43% to 2.90% in the high[1] pool. The high CDT concentration in this pool made possible the integration of the AST peak. The results presented in Table 1 are the sum AST + DiST. Excluding AST did not significantly increase the precision of the technique (data not shown). The EP5-A protocol was applied for day-to-day precision evaluation. In this protocol, within-run, between-run and between-day are not considered as independent factors and consequently a poor within-run precision, as observed in low[2] pool, did not allow the calculation of between-run and betweenday precision (Table 2). We made the same calculations using the EP Evaluator TM software and obtained similar results. Consequently, we calculated separately the within-run precision using the mean of each duplicate as individual value; the CVs ranged from 16.7% (low[2] pool) to 3.60% (high[2] pool excluding AST). Similarly, we used the mean of results

4 4 Schellenberg et al. Table 1. Precision of CDT measurement with the %CDT by the HPLC method intrarun precision Low [1] Medium High [1] day 1 Mean (extreme values) 1.13 ( ) 1.56 ( ) ( ) CV 17.13% 7.84% 1.43% day 2 Mean (extreme values) 1.20 ( ) 1.66 ( ) ( ) CV 19.01% 11.74% 2.80% day 3 Mean (extreme values) 1.12 ( ) 1.65 ( ) ( ) CV 9.01% 8.00% 2.90% Intra-run precision scores for low, medium and high CDT (DiST + AST when detected) values measured by the %CDT by HPLC kit. Intra-run precision was calculated by consecutively analysing 10 aliquots of each pool on three consecutive days. Table 2. Precision of the %CDT by the HPLC test evaluated by the EP5-A2 protocol Within run Between day Total Pool n Mean SD CV(%) SD CV(%) SD CV(%) Low [2] Critical High [2] Disialo High [2] Di- + Asialo Precision scores of the %CDT by the HPLC test according to EP5-5A protocol. Three pools (Low[2], Critical and High[2]) were analysed in duplicate in two separate runs on 5 days per week for 4 weeks. Between-run precision could not be evaluated by the protocol. The missing between-day precision value is due to the high within-run CV for this pool. from each day as individual value to calculate the day-to-day precision. The CVs ranged from 7.18% (low[2] pool) to 2.96% (high[2] pool excluding AST). As observed previously in the intra-run precision study, the coefficients of variation were inversely correlated to CDT concentration in the samples, and the total imprecision according to the EP-5A protocol spread out from 16.7% (low[2] pool) to 5.1% (high[2] pool). In the high[2] pool, AST was not quantifiable in all samples. The exclusion of AST in this pool, as previously recommended (Jeppsson et al., 2007), improved the precision of the technique, resulting in a total imprecision of 4.3%. Interfering situations Peak detection. Technically, the limit of detection (LOD) has been set in the Variant s software at 500 area counts (AC). The limit of quantification (LOQ) results from both the limit of detection and the total Tf concentration. A normal Tf concentration (2.5 g/l) gives a total area of 140, ,000 (AC); consequently, the LOQ is closed to 0.4%. This value is significantly higher than that from the candidate reference method (Helander et al., 2003). This can be attributed to the lower sample volume of the %CDT by the HPLC test, 17 µl versus 28.6 µl of neat serum injected in the column and to a possible superior optical quality of the detection system used for the reference technique (Agilent 1100 LC). Tf concentration. The relationship between serum Tf concentration and CDT measurement was evaluated in 40 samples (Tf range g/l). CDT levels could not be quantified in eight of these samples; in seven samples (Tf range g/l), the DiST peak area was under the integration threshold, and the uneven baseline of one sample (Tf 1.65 g/l) did not allow DiST quantification. The seven samples with no quantifiable CDT had a low Tf concentration and an AC for the sum of all glycoforms below 100,000. Therefore, this value of the total area peak can be considered as a limit for CDT quantification. In the 32 other samples, all CDT values were <1.90%, no significant correlation was observed between Tf and CDT values (r 2 = 0.098), and the slope of the regression line was not significantly different from zero. Lipaemic, icteric and haemolysed samples. Among the 30 lipaemic samples analysed by the %CDT by the HPLC test before and after de-lipidation by ultracentrifugation, the results of untreated and de-lipidated samples were highly correlated (r 2 = 0.971). CDT values showed no difference as attested by the absence of a significant slope (0.944) of the regression equation and by a non-significant paired t-test (P = 0.445). %CDT in the analysed samples was from 0.7% to 11.9%; only two results could be considered as erroneous (difference >10% between both determinations) and were obtained from two samples with high concentrations of triglycerides and chylomicrons, with severely lactescent aspect. Addition of unconjugated bilirubin to non-lipaemic, nonhaemolysed samples had no interfering effect on the %CDT by the HPLC test. The results of supplemented and original samples were highly correlated (r 2 = 0.985), the regression equation showed no significant slope (0.967) and the paired t-test did not reveal any difference (P = 0.439). By contrast, addition of haemoglobin to normal samples resulted in a progressive increase in the baseline of the chromatogram (Fig. 1D), and did not allow a correct integration of the DiST peak for haemoglobin concentrations >2 g/l. In addition, mean values were slightly increased after haemoglobin addition (3.68% versus 3.30%, P = 0.049). High trisialotransferrin. Thirty samples with elevated TriST ( %) were analysed with the %CDT by the HPLC test. In 16 samples, DiST appeared as a shoulder on the TriST peak (Fig. 1B). A low Tf concentration (<1.5 g/l) in 13 of these samples was an additional factor that made quantification of the minor peak difficult, and 9 of these could not be properly

5 Evaluation of the New CDT by the HPLC Method 5 %CDT HPLC (HPLC - CZE) / Average % SD Mean SD AVERAGE of HPLC and CZE HPLC=0.989*CZE %CDT CZE A %CDT HPLC (HPLC - TIA) / Average % SD Mean SD AVERAGE of HPLC and TIA 8 4 HPLC=1.190*TIA B %CDT TIA Fig. 2. Scattergrams of the results with regression lines and 95% confidence intervals for CDT measured by %CDT by the HPLC test and either CZE (A) or TIA (B). The dotted lines correspond to y = x. The insets show the Bland Altman plots of the between-method percentage of the difference. quantified. Six samples were obtained from AUD patients and CDT results were increased in these patients by Capillarys CDT and %CDT by HPLC techniques. Correlation study A total of 302 samples were included in the study. Two were rejected because their chromatogram revealed a BC phenotype for Tf. The CDT concentrations of 17 samples were particularly low and could not be quantified by the %CDT by the HPLC method. The remaining 283 samples covered a wide range of Tf ( g/l) and %CDT ( %) concentrations. We used the paired t-test to compare the overall bias of the results obtained with the different methods. The t values were and 7.822, when the %CDT by the HPLC test was compared with %CDT TIA and Capillarys CDT methods, respectively. These differences were highly significant (P < ), indicating a gap between the results of the compared methods. Fig. 2 (A and B) shows comparisons of the %CDT by the HPLC test with the %CDT TIA and Capillarys CDT methods. The Pearson s correlation coefficients were [95% confidence interval (CI) ] for the %CDT TIA comparison and (95% CI ) for the Capillarys CDT comparison, indicating high concordance between the new %CDT by the HPLC test and the two other methods. Using the Passing and Bablock regression analysis, the slopes of the linear regression lines were calculated as (95% CI ) for the %CDT TIA comparison and (95% CI ) for the Capillarys CDT comparison. This indicated that the difference between %CDT by the HPLC test and %CDT TIA increases with CDT concentration, whereas the difference between %CDT by the HPLC test and the Capillarys CDT method remains approximately constant. The intercepts calculated by the same regression equation were (95% CI to 0.960) for %CDT TIA and (95% CI to 0.503) for Capillarys CDT, both being significantly different from zero. The linearity of the relationship between paired CDT measurements was checked using the Cusum test, which assesses the fit of the data to a linear model. As expected from the Bland and Altman diagrams in Fig. 2 (A and B), the test revealed a non-linear regression between %CDT by the HPLC test and %CDT TIA, and a linear regression between %CDT by the HPLC test and Capillarys CDT. Table 3. Mean CDT values according to sex and alcohol abuse %CDT n mean 95% CI Control subjects Men Women AUD patients Men Women Distribution of %CDT by HPLC values in the study subgroups. A significant difference (t-test P < 0.001) between men and women. All differences between control and AUD groups were significant for CDT (P < 0.001). Case control study Among the 350 subjects studied, 13 had a CDT concentration below the LOQ of the %CDT by the HPLC technique. Two control subjects had a Tf BC phenotype, and their results were not considered in the study. One sample had an uneven baseline that made DiST peak integration not possible. One additional sample from an AUD patient was rejected owing to an abnormally high TriST, resulting in one single peak for TriST and DiST (15.1%). For the remaining 333 subjects included, 165 control subjects (107 males, 58 females) and 168 AUD patients (136 males, 32 females), CDT was between 0.61% and 47.64%. There were no differences in age between the four subgroups of subjects, whereas there was a difference between mean CDT values for male and female subjects in both the control and the AUD patients groups (Table 3). CDT values ranged from 0.61% to 2.66% in the control group and from 0.65% to 47.64% in the AUD group. CDT results in this latter group showed a tailed distribution towards low values, which could be attributed to non-responding patients for CDT, or to an erroneous report of the time since the last alcohol consumption. When calculating the diagnostic values using the cut-off proposed by the manufacturer (1.70%), the sensitivity was 0.74 and the specificity was Another optimized cut-off at 2.0% was proposed by the statistical software from the results of the case-control study reported on the ROC curve (Fig. 3). We next calculated the global predictive value (GPV) of the test, which represents the fraction of subjects correctly classified by

6 6 Schellenberg et al. Sensitivity Specificity Fig. 3. ROC plot (bold line) representing the ability of the %CDT by the HPLC method to distinguish 165 control subjects from 168 AUD patients. The fine lines delimitate the 95% confidence interval of the ROC curve. The arrow indicates the optimum cut-off level calculated by the software (2.0%). Global Predictive Value % Prevalence 1.70% 1.80% 1.90% 2.00% Fig. 4. Global predictive value (GPV ) of the %CDT by HPLC assay in 165 control subjects and 168 AUD patients at 9 degrees of prevalence (10 90%) of AUD for four cut-off values ( %). The global predictive value is the probability for a subject included in the study to be correctly classified by the test as alcohol abuser or not. the test in the studied sample. We used four different cut-off values (1.7%, 1.8%, 1.9% and 2.0%) and nine levels of prevalence (10% to 90%), to check the possible consequences of low or high cut-off levels on the diagnostic value of the test. As indicated in Fig. 4, there is a linear negative relationship between prevalence and GPV, with different slopes according to the cut-off level. DISCUSSION The present study was aimed at evaluating the new %CDT by the HPLC method focusing on the analytical aspects and to compare its diagnostic value to two established methods. Our results suggest that the precision of the test is dependent on the signal, the peak area and thus on CDT concentration. The lack of precision of the %CDT by the HPLC test in low pools does not hamper the clinical interpretation of the results, for no individual result was above the cut-off (1.70%). On the other hand, in the medium pool, CDT was above the cut-off in 7 of the 30 determinations, leading to possible misinterpretation of the result. Optical characteristics of the system probably contribute to this imprecision, since a greater precision was observed (Helander and Bergström, 2006) using the same kit on a high performance HPLC system (Agilent 1100 LC). Our results confirm the better precision of CDT measurement when exclusively quantifying the DiST, the recently proposed (Jeppsson et al., 2007) target analyte. The absence of significant slope of the regression line between Tf and CDT clearly indicates that the total Tf concentration does not by itself introduce a bias in CDT measurement. Nevertheless, this technique does not allow CDT quantification in some samples with Tf <1 g/l while in the 333 samples with Tf >1.5 g/l, only 4 had DiST peak area <500 AC. So, when Tf is in the normal range, the number of samples with no technically relevant result is acceptable ( 1.5%). In 14 patients with high TriST and well-separated TriST and DiST peaks, the %CDT by the HPLC technique gave increased CDT values in 6 AUD patients and below the cut-off in the 8 remaining samples, while the %CDT TIA method gave high CDT in all samples. This discrepancy can probably be attributed to the partial co-elution of TriST in the CDT fraction in the %CDT TIA method (Aldén et al., 2005) that leads to false-positive CDT results in patients with high TriST. In this study, we confirm the findings of a recent study (Arndt et al., 2008) for atypical Tf glycoform distribution. Among the 16 samples with high TriST and poor TriST DiST separation (di-tribridge), 9 of them had low serum Tf (AC< 120,000) and retrospective investigation of their clinical history revealed a liver disease and a suspicion of cirrhosis for 7 out 9 patients. Therefore, high TriST is not by itself a source of inappropriate results, except in patients with low Tf and di-tribridging phenomenon. The high correlation between the %CDT by the HPLC test and the two other techniques is in agreement with previous results (Helander et al., 2001; Schwarz et al., 2003). We observed a linear correlation between the two methods based on Tf glycoforms pattern determination (%CDT by HPLC and Capillarys CDT), and a non-linear correlation of these techniques with the %CDT TIA method. This lack of linearity could be attributed either to a non-linear calibration of the immunoassay or to the incomplete separation between TriST and DiST in the %CDT TIA test. Figure 2 shows a lower dispersion of the results between the HPLC and CZE methods, as opposed to the HPLC and TIA methods. These differences can be attributed to, at least, three factors influencing the TIA method: (i) a poor precision level at low values (CDT < 1.8%), (ii) the partial co-elution of TriST in the CDT fraction and (iii) the cumulative imprecision of the immunoassays used for the separate quantification of Tf and CDT. By contrast, in the HPLC and CZE methods, CDT is a ratio between areas obtained in a unique glycoform separation. Independent of the correlation coefficients, the three compared methods give significantly different CDT results, as

7 Evaluation of the New CDT by the HPLC Method 7 shown by the calculated regression equations. Obviously, a common international standard is needed to obtain a linear correlation between the techniques and to allow comparison of results (Oberrauch et al., 2008). In the control group, we observed a moderate, but significant, difference in the CDT values among genders. This result appears in contradiction with a recent work (Bergström and Helander, 2008) that concluded to an absence of sex difference for CDT values. This might result from differences in the control group definition and from a possible increased alcohol intake in males as compared to women, within the limit of World Health Organization (WHO) recommendations assessed through a questionnaire. Concerning the diagnostic value of CDT measured by HPLC methods, two studies (Simonsson et al., 1996; Werle et al., 1997) reported higher specificity (>0.95), and one study (Werle et al., 1997) reported better sensitivity (0.80). It is, however, hard to determine the respective contributions of the technical features and of the population study selection criteria to explain these differences. To compare the analytical performance of a test, it is crucial to base the population selection on the sole WHO definition of alcohol abuse. The cut-off setting is also a major point as it controls the diagnostic values. The cut-off can be defined from the distribution of values in the control group or as the value that gives the best discrimination between cases and controls in the study population. In this case, the cut-off can be determined using an ROC curve analysis. In our study (Fig. 3), cut-off determination from the ROC curve suggested a value of 2.0%. Selection of this cut-off resulted in a sensitivity of 0.66 with a specificity of Because they are calculated from an established diagnosis, sensitivity and specificity values are nevertheless not highly informative for the physicians. Conversely, the global predictive value (GPV) measures the probability for a subject to be properly classified by the test, independent of his clinical status. With the exception of a test with equal sensitivity and specificity values, GPV fluctuates with the prevalence of the disease (AUD) in the population sample. As indicated in Fig. 4, the higher (2.0% and 1.9%) cut-off values yielded higher GPVs for prevalence values below 50%. Considering a population sample representative of the AUD prevalence in the general population (10%), GPVs were 0.93 for a cut-off at 2.0% or 0.92 for a cut-off at 1.9%. Although the higher cut-off value gives a slightly higher GPV, the heavy social and medical consequences of erroneous interpretations of false-positive CDT value led us to propose 1.9% as cut-off. With the same technique, Helander proposed a cut-off at 1.70% (Helander et al., 2006). Similarly, cut-off values ranging from 1% to 1.75% were selected using other HPLC techniques (Simonsson et al., 1996; Werle et al., 1997; Arndt and Keller, 2004; Helander et al., 2005). These discrepancies might be due to several methodological factors, including the basis for cut-off definition (CDT value distribution in the control group or discrimination between control and AUD groups) and the criteria for inclusion in the control group. The various cut-off values proposed in these studies, as well as in this study, should be considered as interim values in the absence of a value obtained from a study based on the determination of CDT using the candidate reference method (Jeppsson et al., 2007) applied to a population strictly selected upon the WHO definition of alcohol abuse. CONCLUSIONS The data from this study indicate a good validity of results obtained with the %CDT by the HPLC method on the Variant TM system. The precision level is acceptable for routine use and the results are not affected by lipaemic or icteric samples. High trisialotransferrin concentrations are not by themselves a cause of false-positive values. The serum Tf concentration does not affect CDT measurement for concentrations >1 g/l. Furthermore, the results correlate well with those obtained with other routine methods based on other analytical processes. This assay provides accurate clinical information, conjugated to the high throughput allowed by the Variant TM HPLC system. Acknowledgements The authors are grateful to Mrs Annie Détruit, Nadine Puche and Geneviève Wattelet for their technical contribution. They thank Bio-Rad for supporting this study by lending the Variant TM system and providing the reagents. They thank Sebia for providing kits for CZE determinations of CDT. The case-control study was partly supported by INSERM grant ATC Alcool no. ALC0205. REFERENCES Aldén A, Ohlson S, Påhlsson P et al. (2005) HPLC analysis of carbohydrate deficient transferrin isoforms isolated by the Axis-Shield %CDT method. Clin Chim Acta 356: Arndt T, Keller T. (2004) Forensic analysis of carbohydrate-deficient transferrin (CDT): implementation of a screening and confirmatory analysis concept is hampered by the lack of CDT isoform standards. Forensic Sci Int 146:9 16. Arndt T, Van Der Meijden BB, Wielder JPM. (2008) Atypical serum transferrin isoform distribution in liver cirrhosis studied by HPLC, capillary electrophoresis and transferrin genotyping. Clin Chim Acta 394:42 6. Bergström JP, Helander A. (2008) Influence of alcohol use, ethnicity, age, gender, BMI and smoking on the transferrin glycoform pattern: implication for use of carbohydrate-deficient transferrin (CDT) as alcohol biomarker. Clin Chim Acta 388: Bortolotti F, De Paoli G, Pascali JP et al. (2007) Fully automated analysis of carbohydrate-deficient transferrin (CDT) by using a multicapillary electrophoresis system. Clin Chim Acta 380:4 7. Bortolotti F, De Paoli G, Tagliaro F. (2006) Carbohydrate-deficient transferrin (CDT) as a marker of alcohol abuse: a critical review. J Chromatogr B 841: Delanghe JR, Helander A, Wielders JPM et al. (2007) Development and multicenter evaluation of the N Latex CDT direct immunonephelometric assay for serum carbohydrate-deficient transferrin. Clin Chem 53: Hatton MWC, Berry LR, Regoeczi E. (1982) The structural heterogeneity of the carbohydrate moiety of desialylated human transferrin. Can J Biochem 60: Helander A. (2003) Biological markers of alcohol abuse. JNeural Transm 66 (Suppl): Helander A, Bergström JP. (2006) Determination of carbohydratedeficient transferrin in human serum using the Bio-Rad %CDT by HPLC test. Clin Chim Acta 371: Helander A, Fors M, Zakrisson B. (2001) Study of Axis-Shield new %CDT immunoassay for quantification of carbohydrate-deficient transferrin (CDT) in serum. Alcohol Alcohol 36: Helander A, Husa A, Jeppsson JO. (2003) Improved HPLC method for carbohydrate-deficient transferrin in serum. Clin Chem 49: Helander A, Wielders JPM, Te Stroet R et al. (2005) Comparison of HPLC and capillary electrophoresis for confirmatory testing of the alcohol misuse marker carbohydrate-deficient transferrin. Clin Chem 51: Jeppsson JO, Arndt T, Schellenberg F et al. (2007) Toward standardization of carbohydrate-deficient transferrin (CDT) measurements: I. Analyte definition and proposal of a candidate reference method. Clin Chem Lab Med 45:

8 8 Schellenberg et al. Jeppsson JO, Kristensson H, Fimiani C. (1993) Carbohydrate-deficient transferrin quantified by HPLC to determine heavy consumption of alcohol. Clin Chem 39: Oberrauch W, Bergman AC, Helander A. (2008) HPLC and mass spectrometric characterization of a candidate reference material for the alcohol biomarker carbohydrate-deficient transferrin. Clin Chim Acta [Epub ahead of print]. Schellenberg F, Girre C, Nalpas B et al. (2007) Analytical evaluation of a new capillary electrophoresis method for carbohydrate-deficient transferrin. Clin Chim Acta 382: Schwan R, Loiseaux MN, Schellenberg F et al. (2004). Multicenter validation study of the %CDT TIA kit in alcohol abuse and alcohol dependence. Alcohol Clin Exp Res 28: Schwarz MJ, Domke I, Helander A et al. (2003) Multicentre evaluation of a new assay for determination of carbohydrate-deficient transferrin. Alcohol Alcohol 38: Simonsson P, Lindberg S, Alling C. (1996) Carbohydrate-deficient transferrin measured by high-performance liquid chromatography and CDTect TM immunoassay. Alcohol Alcohol 31: Stibler H. (1991) Carbohydrate-deficient transferrin in serum: a new marker of potentially harmful alcohol consumption reviewed. Clin Chem 37: Stibler H, Borg S, Allgulander C. (1979) Clinical significance of abnormal heterogeneity of transferrin in relation to alcohol consumption. Acta Med Scand 206: Stibler H, Borg S, Joustra M. (1991) A modified method for the assay of carbohydrate-deficient transferrin (CDT) in serum. Alcohol Alcohol 1 (Suppl 1): Werle E, Seitz GE, Kohl B et al. (1997) High-performance liquid chromatography improves diagnostic efficiency of carbohydratedeficient transferrin. Alcohol Alcohol 32:71 7.

CE Pharm 2008, October 12-16, 2008, San Francisco, CA, USA

CE Pharm 2008, October 12-16, 2008, San Francisco, CA, USA Clinical Diagnostics by Capillary Electrophoresis: Determination of carbohydrate-deficient transferrin and lamotrigine in human serum under stringent quality control conditions CE Pharm 8, October 1-1,

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

METHOD-DEPENDENT CHARACTERISTICS OF CARBOHYDRATE-DEFICIENT TRANSFERRIN MEASUREMENTS IN THE FOLLOW-UP OF ALCOHOLICS

METHOD-DEPENDENT CHARACTERISTICS OF CARBOHYDRATE-DEFICIENT TRANSFERRIN MEASUREMENTS IN THE FOLLOW-UP OF ALCOHOLICS Alcohol & Alcoholism Vol. 39, No. 1, pp. 59 63, 2004 doi:10.1093/alcalc/agh021, available online at www.alcalc.oupjournals.org METHOD-DEPENDENT CHARACTERISTICS OF CARBOHYDRATE-DEFICIENT TRANSFERRIN MEASUREMENTS

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c Clin Chem Lab Med 215; aop Stéphane Jaisson*, Nathalie Leroy, Emmanuelle Guillard, Aurore Desmons and Philippe Gillery Analytical performances of the D- TM hemoglobin testing system (Bio-Rad) for assay

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Abstract. Author. Biotherapeutics & Biosimilars. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany Sensitive and Reproducible Glycan Analysis of Human Immunoglobulin G The Agilent 1260 Infi nity Bio-inert Quaternary LC System with an Agilent AdvanceBio 2.7 µm Glycan Mapping Column and Fluorescence Detection

More information

retardation in infants. A wide variety of analytical methods for the analysis of

retardation in infants. A wide variety of analytical methods for the analysis of IN THE NAME OF GOD Diagnosis and measurment of Phenylalanine in plasma and dried bilood spot (DBS) by a simple and sensitive HPLC method and compaire with Recipe Reference Health Laboratory Research Center,

More information

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS

DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS FORENSICS AND TOXICOLOGY ANALYSIS DETERMINATION OF CANNABINOIDS, THC AND THC-COOH, IN ORAL FLUID USING AN AGILENT 6490 TRIPLE QUADRUPOLE LC/MS Solutions for Your Analytical Business Markets and Applications

More information

HbA1c enzymatic assay evaluation at AN academic

HbA1c enzymatic assay evaluation at AN academic Peer reviewed ORIGINAL ARTICLE HbA1c enzymatic assay evaluation at AN academic laboratory B Phiri-Ramongane Mmed Chem Path, FCP Path(SA)-Chem Department of Chemical Pathology, Sefako Makgatho Health Sciences

More information

The Adiponectin Turbidimetric Immunoassay Reagent Kit

The Adiponectin Turbidimetric Immunoassay Reagent Kit The Adiponectin Turbidimetric Immunoassay Reagent Kit Catalogue number: 51010 For the quantitative determination of Adiponectin in human serum and plasma This package insert must be read in its entirety

More information

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms

RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms Asian Journal of Chemistry Vol. 22, No. 7 (2010), 5067-5071 RP-HPLC Analysis of Temozolomide in Pharmaceutical Dosage Forms A. LAKSHMANA RAO*, G. TARAKA RAMESH and J.V.L.N.S. RAO Department of Pharmaceutical

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(2):770-775 Validation of Rapid Liquid Chromatographic

More information

Detection and quantitation of hemoglobin (Hb) fractions

Detection and quantitation of hemoglobin (Hb) fractions Evaluation of a Dual Hemoglobin A 2 /A 1c Quantitation Kit on the Bio-Rad Variant II Automated Hemoglobin Analyzer John D. Lafferty, MLT, ART; Andrew G. McFarlane, MLT, ART; David H. K. Chui, MD, FRCP

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer

A Robustness Study for the Agilent 6470 LC-MS/MS Mass Spectrometer A Robustness Study for the Agilent 7 LC-MS/MS Mass Spectrometer Application Note Clinical Research Authors Linda Côté, Siji Joseph, Sreelakshmy Menon, and Kevin McCann Agilent Technologies, Inc. Abstract

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser

More information

The first and only fully-automated, multiplexed solution for Measles, Mumps, Rubella and Varicella-zoster virus antibody testing

The first and only fully-automated, multiplexed solution for Measles, Mumps, Rubella and Varicella-zoster virus antibody testing Bio-Rad Laboratories BioPlex 2200 System BioPlex 2200 MMRV IgG Kit The first and only fully-automated, multiplexed solution for Measles, Mumps, Rubella and Varicella-zoster virus antibody testing Bio-Rad

More information

MEASUREMENT OF CARBOHYDRATE-DEFICIENT TRANSFERRED (CDT) IN A GENERAL MEDICAL CLINIC: IS THIS TEST USEFUL IN ASSESSING ALCOHOL CONSUMPTION?

MEASUREMENT OF CARBOHYDRATE-DEFICIENT TRANSFERRED (CDT) IN A GENERAL MEDICAL CLINIC: IS THIS TEST USEFUL IN ASSESSING ALCOHOL CONSUMPTION? Alcohol & Alcoholism Vol. 33, No. 3, pp. 304-309, 1998 MEASUREMENT OF CARBOHYDRATE-DEFICIENT TRANSFERRED (CDT) IN A GENERAL MEDICAL CLINIC: IS THIS TEST USEFUL IN ASSESSING ALCOHOL CONSUMPTION? GURUPRASAD

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(1):138-144 Simultaneous RP HPLC determination of Latanoprost

More information

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract JJC Jordan Journal of Chemistry Vol. 4 No.4, 2009, pp. 357-365 Development and Validation of Analytical Method for Fluconazole and Fluconazole Related Compounds (A, B, and C) in Capsule Formulations by

More information

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research

Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Analysis of Testosterone, Androstenedione, and Dehydroepiandrosterone Sulfate in Serum for Clinical Research Dominic Foley, Michelle Wills, and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION

More information

Hemoglobinopathy and thalassemia detection

Hemoglobinopathy and thalassemia detection Hemoglobinopathy and thalassemia detection Traditional methods and a novel method capillary electrophoresis technology By Aigars Brants, PhD Published October 2011 In humans, two pairs of unlike globin

More information

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

International Journal of Pharma and Bio Sciences

International Journal of Pharma and Bio Sciences International Journal of Pharma and Bio Sciences RESEARCH ARTICLE PHARMACEUTICAL ANALYSIS DEVELOPMENT AND VALIDATION OF LIQUID CHROMATOGRAPHIC METHOD FOR ESTIMATION OF ESCITALOPRAM OXALATE IN TABLET DOSAGE

More information

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE

METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC FOR ESTIMATION OF ZOLPIDEM TARTARATE WORLD JOURNAL OF PHARMACY AND PHARMACEUTICAL SCIENCES Ramalakshmi et al. SJIF Impact Factor 6.647 Volume 7, Issue 2, 1010-1018 Research Article ISSN 2278 4357 METHOD DEVELOPMENT AND VALIDATION BY RP-HPLC

More information

EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2

EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2 EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2 Dr Barbara abaawild The importance of Hb A 2 measurement Accurate and reliable measurement of Hb A 2 is essential for the diagnosis of beta thalassaemia

More information

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets Journal of PharmaSciTech 0; ():- Research Article Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets * Sayyed Hussain,

More information

LIPASE liquicolor. Design Verification. Multipurpose Reagent

LIPASE liquicolor. Design Verification. Multipurpose Reagent Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of

More information

Instructions for Use. Tg Antibody ELISA

Instructions for Use. Tg Antibody ELISA Instructions for Use Tg Antibody ELISA Enzyme Immunoassay for the Quantitative Determination of Autoantibodies to Thyroglobulin (Tg) in Serum and Plasma I V D REF EA618/96 12 x 8 2 8 C DLD Gesellschaft

More information

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual Page 1 / 20 PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual ZV-4003-0200-10 200 2-8 C Page 2 / 20 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3

More information

liquicolor (AMP Buffer, IFCC) Design Verification

liquicolor (AMP Buffer, IFCC) Design Verification Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Reverse Phase HPLC Analysis of Atomoxetine in Pharmaceutical Dosage Forms

Reverse Phase HPLC Analysis of Atomoxetine in Pharmaceutical Dosage Forms Asian Journal of Chemistry Vol. 21, No. 2 (2009), 829-833 Reverse Phase HPLC Analysis of Atomoxetine in Pharmaceutical Dosage Forms B.V.V.S. JAGADEESH, S. SATYANARAYANA RAJU, V.JAYATHIRTHA RAO and J.V.L.N.

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD

SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD 170 Original Article SIMULTANEOUS ESTIMATION OF VALSARTAN AND HYDROCHLOROTHIAZIDE IN TABLETS BY RP-HPLC METHOD *Lakshmana Rao A, 1 Bhaskara Raju V *V.V. Institute of Pharmaceutical Sciences, Gudlavalleru,

More information

Active-B12 (Holo-TC) ELISA KIT

Active-B12 (Holo-TC) ELISA KIT Active-B12 (Holo-TC) ELISA KIT Cat. No.:DEIA6312 Pkg.Size:96T Intended use The Active-B12 (Holotranscobalamin) assay is an enzyme-immunoassay (ELISA) for the quantitative determination of holotranscobalamin

More information

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method

Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by HPLC Method ISSN: 0973-4945; CODEN ECJAO E- Chemistry http://www.e-journals.net 2012, 9(1), 340-344 Development, Estimation and Validation of Lisinopril in Bulk and its Pharmaceutical Formulation by PLC Method V.

More information

DIRECT ENZYMATIC HbA1c ASSAY

DIRECT ENZYMATIC HbA1c ASSAY DIRECT ENZYMATIC HbA1c ASSAY ACCURATE AND SIMPLE DIAZYME INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses

More information

Pelagia Research Library

Pelagia Research Library Available online at www.pelagiaresearchlibrary.com Der Pharmacia Sinica, 2015, 6(1):6-10 ISSN: 0976-8688 CODEN (USA): PSHIBD Validated RP-HPLC method for simultaneous estimation of metformin hydrochloride

More information

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page: Research Article CODEN: AJPAD7 ISSN: 2321-0923 Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page: www.ajpamc.com ANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF GEFITINIB

More information

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2

Sanjog Ramdharane 1, Dr. Vinay Gaitonde 2 JPSBR: Volume 5, Issue 2: 2015 (151-155) ISS. 2271-3681 A ew Gradient RP- HPLC Method for Quantitative Analysis of : (3-luoro-4- Morpholin-4-yl-Phenyl)-Carbamic Acid Methyl Ester and its Related Substances

More information

ISSN (Print)

ISSN (Print) Scholars Academic Journal of Pharmacy (SAJP) Sch. Acad. J. Pharm., 2014; 3(3): 240-245 Scholars Academic and Scientific Publisher (An International Publisher for Academic and Scientific Resources) www.saspublisher.com

More information

Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum

Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum Screening by immunoassay and confirmation & quantitation by GC-MS of buprenorphine and norbuprenorphine in urine, whole blood and serum NINA KANGAS, SIRPA MYKKÄNEN, SANNA KYLLÖNEN, PÄIVI RAJALA, KARI ARINIEMI

More information

ETHYL GLUCURONIDE DETERMINATION IN HAIR AS AN INDICATOR OF CHRONIC ALCOHOL ABUSE: A FULLY VALIDATED METHOD BY GC-EI-MS/MS

ETHYL GLUCURONIDE DETERMINATION IN HAIR AS AN INDICATOR OF CHRONIC ALCOHOL ABUSE: A FULLY VALIDATED METHOD BY GC-EI-MS/MS FORENSICS AND TOXICOLOGY ANALYSIS ETHYL GLUCURONIDE DETERMINATION IN HAIR AS AN INDICATOR OF CHRONIC ALCOHOL ABUSE: A FULLY VALIDATED METHOD BY GC-EI-MS/MS Solutions for Your Analytical Business Markets

More information

DIHYDROSTREPTOMYCIN and STREPTOMYCIN First draft prepared by Lynn G. Friedlander, Rockville, MD, USA Rainer W. Stephany, Bilthoven, The Netherlands

DIHYDROSTREPTOMYCIN and STREPTOMYCIN First draft prepared by Lynn G. Friedlander, Rockville, MD, USA Rainer W. Stephany, Bilthoven, The Netherlands DIHYDROSTREPTOMYCIN and STREPTOMYCIN First draft prepared by Lynn G. Friedlander, Rockville, MD, USA Rainer W. Stephany, Bilthoven, The Netherlands ADDENDUM To the monograph and addenda prepared by the

More information

Supporting Information

Supporting Information Supporting Information Development of a High Coverage Pseudotargeted Lipidomics Method Based on Ultra-High Performance Liquid Chromatography-Mass Spectrometry Qiuhui Xuan 1,2#, Chunxiu Hu 1#, Di Yu 1,2,

More information

Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin in Pharmaceutical Dosage Form using RP-HPLC Method

Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin in Pharmaceutical Dosage Form using RP-HPLC Method International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.5, No.4, pp 1736-1744, Oct-Dec 2013 Development and Validation for Simultaneous Estimation of Sitagliptin and Metformin

More information

DBC 25-Hydroxyvitamin D

DBC 25-Hydroxyvitamin D D I A G N O S T I C S B I O C H E M C A N A D A DBC 25-Hydroxyvitamin D ELISA OVERVIEW The worldwide aging and chronically ill population is increasing rapidly. It is forecasted that the Global Vitamin

More information

Coping with Analytical Interferences

Coping with Analytical Interferences Coping with Analytical Interferences (Handling Icteric, Hemolytic and Lipaemic Samples) Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Surabaya Indonesia 2016 Acknowledgements

More information

Chapter 2. High-sensitivity C-reactive protein methods examined

Chapter 2. High-sensitivity C-reactive protein methods examined Chapter High-sensitivity C-reactive protein methods examined Snježana Rothkrantz-Kos, Maria PJ Schmitz, Otto Bekers, Paul PCA Menheere, Marja P van Dieijen-Visser Clin Chem ;48:359-6 3 Chapter Abstract

More information

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these

More information

Determination of red blood cell fatty acid profiles in clinical research

Determination of red blood cell fatty acid profiles in clinical research Application Note Clinical Research Determination of red blood cell fatty acid profiles in clinical research Chemical ionization gas chromatography tandem mass spectrometry Authors Yvonne Schober 1, Hans

More information

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form

RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form RP-HPLC Method Development and Validation of Abacavir Sulphate in Bulk and Tablet Dosage Form S. LAVANYA* 1, SK. MANSURA BEGUM 1, K. NAGAMALLESWARA RAO 2, K. GAYATHRI DEVI 3 Department of pharmaceutical

More information

General Chemistry Scheme Guide

General Chemistry Scheme Guide General Chemistry Scheme Guide Copyright WEQAS. All rights reserved. No part of this document may be reproduced or utilised in any form without permission from WEQAS Contents. Scheme details and repertoire.....

More information

DETERMINATION OF LOTEPREDNOL ETABONATE AND TOBRAMYCIN IN COMBINED DOSAGE FORM USING RP-HPLC METHOD

DETERMINATION OF LOTEPREDNOL ETABONATE AND TOBRAMYCIN IN COMBINED DOSAGE FORM USING RP-HPLC METHOD 280 P a g e e-issn: 2248-9126 Vol 5 Issue 4 2015 280-284. Print ISSN: 2248-9118 Indian Journal of Pharmaceutical Science & Research www.ijpsrjournal.com DETERMINATION OF LOTEPREDNOL ETABONATE AND TOBRAMYCIN

More information

New Automated Screening System for the Determination of Basic Compounds in Urine by On-Line Extraction-HPLC- DAD

New Automated Screening System for the Determination of Basic Compounds in Urine by On-Line Extraction-HPLC- DAD New Automated Screening System for the Determination of Basic Compounds in Urine by On-Line Extraction-HPLC- DAD Thomas Grobosch, Lena Schönberg, Dagmar Lampe, Charlotte Kloft Abstract A new automated,

More information

Analytical Method Development and Validation for the Estimation of Guaifenesin and Dextromethorphan by RP-HPLC

Analytical Method Development and Validation for the Estimation of Guaifenesin and Dextromethorphan by RP-HPLC Human Journals Research Article September 2018 Vol.:13, Issue:2 All rights are reserved by Ramya Sree Borra et al. Analytical Method Development and Validation for the Estimation of Guaifenesin and Dextromethorphan

More information

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ

Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Application Note Clinical Research Analysis of anti-epileptic drugs in human serum using an Agilent Ultivo LC/TQ Authors Jennifer Hitchcock 1, Lauren Frick 2, Peter Stone 1, and Vaughn Miller 2 1 Agilent

More information

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products

Application Note. Authors: C. Ledesma, M. Gibert, J.R. Gibert Ingenieria Analitica S.L. Extracts from various food products High-Performance Anion-Exchange Chromatography coupled with Pulsed Electrochemical Detection as a powerful tool to evaluate lactose content in lactose-free labeled products Application Note Authors: C.

More information

36 J App Pharm Vol. 6; Issue 1: 36-42; January, 2014 Rao et al., 2014

36 J App Pharm Vol. 6; Issue 1: 36-42; January, 2014 Rao et al., 2014 36 J App Pharm Vol. 6; Issue 1: 36-42; January, 2014 Rao et al., 2014 Original Research Article SIMPLE AND RAPID LIQUID CHROMATOGRAPHIC METHOD FOR REAL-TIME QUANTIFICATION OF NAPROXEN / ESOMEPRAZOLE MAGNESIUM

More information

Vitamin A-E Serum HPLC Analysis Kit

Vitamin A-E Serum HPLC Analysis Kit Vitamin A-E Serum HPLC Analysis Kit ZV-4013-0200-10 200 2-8 C ZIVAK TEKNOLOJI SANAYI VE TICARET LIMITED SIRKETI Gursel Mh. Eski Besiktas Cd. No:44/1-2 KAGITHANE-ISTANBUL-TURKEY www.zivak.com info@zivak.com

More information

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (http://scholarsresearchlibrary.com/archive.html)

Scholars Research Library. Der Pharmacia Lettre, 2016, 8 (6): (http://scholarsresearchlibrary.com/archive.html) Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2016, 8 (6):217-223 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4

More information

SIMPLE AND VALIDATED RP-HPLC METHOD FOR THE ESTIMATION OF CARBOPLATIN IN BULK AND FORMULATION DOSAGE FORM. Subhashini.Edla* B.

SIMPLE AND VALIDATED RP-HPLC METHOD FOR THE ESTIMATION OF CARBOPLATIN IN BULK AND FORMULATION DOSAGE FORM. Subhashini.Edla* B. SIMPLE AND VALIDATED RP-HPLC METHOD FOR THE ESTIMATION OF CARBOPLATIN IN BULK AND FORMULATION DOSAGE FORM Subhashini.Edla* B.Syama sundhar Abstract Dept of Chemistry, Acharya Nagarjuna University, Nagarjuna

More information

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis

Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis Product Guide for LudgerSep TM R1 HPLC Column for DMB labelled Sialic Acid Analysis Product # LS-R1-4.6x150 Ludger Document # LS-R1-DMB-Guide-v5.1 Ludger Ltd Culham Science Centre Oxford OX14 3EB United

More information

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article

Journal of Chemical and Pharmaceutical Research, 2013, 5(1): Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2013, 5(1):180-184 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 A simple and sensitive RP-HPLC method for estimation

More information

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626 Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi

More information

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual Page 1 / 12 25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual ZV-4027-0500-10 500 2-8 C Page 2 / 12 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. IVD SYMBOLS...

More information

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction

More information

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET

International Journal of Pharma and Bio Sciences DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET International Journal of Pharma and Bio Sciences RESEARCH ARTICLE ANALYTICAL CHEMISTRY DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR THE ESTIMATION OF STRONTIUM RANELATE IN SACHET K.MYTHILI *, S.GAYATRI,

More information

Human Neurology 3-Plex A

Human Neurology 3-Plex A Human Neurology 3-Plex A SUMMARY AND EXPLANATION OF THE TEST The Human N3PA assay is a digital immunoassay for the quantitative determination of total Tau, Aβ42, and Aβ40 in human plasma and CSF. Determination

More information

vii LIST OF TABLES TABLE NO DESCRIPTION PAGE 1.1 System Suitability Parameters and Recommendations Acidic and Alkaline Hydrolysis 15

vii LIST OF TABLES TABLE NO DESCRIPTION PAGE 1.1 System Suitability Parameters and Recommendations Acidic and Alkaline Hydrolysis 15 vii LIST OF TABLES TABLE NO DESCRIPTION PAGE CHAPTER- 1 1.1 System Suitability Parameters and Recommendations 07 1.2 Acidic and Alkaline Hydrolysis 15 1.3 Oxidative Degradation Study 16 1.4 Hydrolysis

More information

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2 Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...

More information

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology Evidence Based HbA1c Accuracy Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology Fresh lycohaemoglobin This program is designed to assist laboratories to assess the accuracy of their hihba1c

More information

Jennifer P. Pascali, PhD. dtolabs, Resana (TV)

Jennifer P. Pascali, PhD. dtolabs, Resana (TV) Jennifer P. Pascali, PhD dtolabs, Resana (TV) Agilent Technologies Users Meeting, 20 May 2014 The project dtolabs dtolabs is an analytical research centre founded by two leading companies in the production

More information

Screening of Antihistamine Agents (Diphenhydramine) with Blood and Urine Samples by REMEDi-HS System

Screening of Antihistamine Agents (Diphenhydramine) with Blood and Urine Samples by REMEDi-HS System Screening of Antihistamine Agents (Diphenhydramine) with Blood and Urine Samples by REMEDi-HS System Ohtsuji M, Ohshima T, Takayasu T, Nishigami J, Kondo T, Lin Z, Minamino T Department of Legal Medicine,

More information

A Validated Chiral Liquid Chromatographic Method for The Enantiomeric Separation of Dapoxetine Hydrochloride

A Validated Chiral Liquid Chromatographic Method for The Enantiomeric Separation of Dapoxetine Hydrochloride Received on 15/05/2012; Revised on 22/05/2012; Accepted on 09/06/2012 A Validated Chiral Liquid Chromatographic thod for The Enantiomeric Separation of Dapoxetine Hydrochloride T.Rohith 1 and S. Ananda

More information

Author. Introduction. Abstract

Author. Introduction. Abstract Simultaneous Measurement of Sirolimus and Tacrolimus Concentrations in Blood by Semi-Automated Extraction and Liquid Chromatography-Electrospray Ionization Mass Spectrometry Application Author Gary L.

More information

RP- HPLC and Visible Spectrophotometric methods for the Estimation of Meropenem in Pure and Pharmaceutical Formulations

RP- HPLC and Visible Spectrophotometric methods for the Estimation of Meropenem in Pure and Pharmaceutical Formulations International Journal of ChemTech Research CODEN( USA): IJCRGG ISSN : 0974-4290 Vol. 3, No.2, pp 605-609, April-June 2011 RP- HPLC and Visible Spectrophotometric methods for the Estimation of Meropenem

More information

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls Shyamali Pal R B Diagnostic Private Limited, Lake Town, Kolkata, India INFO ABSTRACT Corresponding

More information

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2016, 8 (11):236-242 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4

More information

Product Guide for LudgerSep TM C3 anion exchange HPLC Column for Glycan Analysis

Product Guide for LudgerSep TM C3 anion exchange HPLC Column for Glycan Analysis Product Guide for LudgerSep TM C3 anion exchange HPLC Column for Glycan Analysis (Ludger Product Code: LS-C3-7.5x75) Ludger Document # LS-C3-Guide-v3.0 Ludger Ltd Culham Science Centre Oxford OX14 3EB

More information

DEVELOPMENT, VALIDATION AND APPLICATION OF THE HPLC METHOD FOR DETERMINATION OF MIANSERIN IN HUMAN SERUM

DEVELOPMENT, VALIDATION AND APPLICATION OF THE HPLC METHOD FOR DETERMINATION OF MIANSERIN IN HUMAN SERUM Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 64 No. 2 pp. 103ñ107, 2007 ISSN 0001-6837 Polish Pharmaceutical Society ANALYSIS DEVELOPMENT, VALIDATION AND APPLICATION OF THE HPLC METHOD FOR DETERMINATION

More information

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots

More information

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for

More information

LC-MS/MS Method for the Determination of Tenofovir from Plasma

LC-MS/MS Method for the Determination of Tenofovir from Plasma LC-MS/MS Method for the Determination of Tenofovir from Plasma Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 687 Key Words SPE, SOLA CX, Hypersil GOLD, tenofovir Abstract

More information

Available online at Scholars Research Library

Available online at  Scholars Research Library Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2015, 7 (3):157-161 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4

More information

HPLC Analysis of Sugars

HPLC Analysis of Sugars HPLC Analysis of Sugars Pre-Lab Exercise: 1) Read about HPLC, sugars and the experiment and its background. 2) Prepare a flowchart as appropriate for the lab exercise. 3) Note the various sugar concentrations

More information

Direct HbA1c testing capabilities on the RX modena

Direct HbA1c testing capabilities on the RX modena Direct testing capabilities on the RX modena INTRODUCTION What is? The term refers to glycated haemoglobin. It develops when haemoglobin, a protein within red blood cells that carries oxygen throughout

More information

Enzyme Immunoassay for

Enzyme Immunoassay for Enzyme Immunoassay for Prostaglandin E 2 For Research Use Only INTRODUCTION Prostaglandin E 2 EIA Kit Product Number: EA02 Store at 4 C FOR RESEARCH USE ONLY Document Control Number: EA02.120214 Page 1

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), Available online at RESEARCH ARTICLE

Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), Available online at  RESEARCH ARTICLE Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), 26-30 26 Available online at http://jddtonline.info RESEARCH ARTICLE METHOD DEVELOPMENT AND ITS VALIDATION FOR QUANTITATIVE SIMULTANEOUS

More information