AAPM TG178: Towards the Recommendation of a Standard Dosimetry Protocol for Gamma Knife Stereotactic Radiosurgery
|
|
- Domenic Sherman
- 6 years ago
- Views:
Transcription
1 AAPM TG178: Towards the Recommendation of a Standard Dosimetry Protocol for Gamma Knife Stereotactic Radiosurgery Paula L. Petti, PhD Washington Hospital Healthcare System Fremont, California, USA
2 AAPM Task Group 178 Members Paola Alvarez, U. TX, MD Anderson Greg Bednarz, U. Pittsburgh J. Daniel Bourland, Wake Forest U. Larry DeWerd, U. Wisconsin Madison ADCL Robert Drzymala, Washington University, St. Louis, MO Steven Goetsch, San Diego GK Center (Task group chair) David Larson, Washington Hospital, Fremont, CA Lijun Ma, University California, San Francisco Sheridan Meltsner, Duke Univ. Med. Ctr Gennady Neyman, Cleveland Clinic Josef Novotny, Hospital Na Homolce, Prague, Czech Republic Paula Petti, Washington Hospital, Fremont, CA Mark Rivard, Tufts Univ. School of Med. Arjun Sahgal, Sunnybrook Research Institute, Toronto, Ontario, Canada Almon Shiu, Univ. Southern California
3 AAPM Task Group 178 Consultants Jonas Johansson, Elekta Instrument AB, Stockholm, Sweden Keith Kunugi, Univ. Wisconsin, Madison, WI Jan Seuntjens, McGill Univ, Montreal, Quebec, Canada
4 Disclosures Paula Petti, Gennady Neyman, Steven Goetsch and Josef Novotny are consultants for Elekta Instrument AB
5 Importance of a Standardized Dosimetry Protocol Promote consistent dose calibration procedures across all GK institutions Reduce probability of dose delivery error Challenge: Gamma Knife a unique system Calibration protocols specific to linear accelerators not necessarily valid
6 Background: Dose Calibration Protocols Ionization chamber calibrated in air Parameters AAPM TG-21 (1983) Meltsner-DeWerd (2009)* Tabulated based on measurements Monte-Carlo (MC) generated Advantages Not specific to one type of phantom or to one phantom material. Phantom can be water, TE plastic, etc. Not specific to linear accelerator Disadvantages Multiple steps in the calculation Complexity could lead to errors MC parameters need to be calculated for each ion chamber Requires that the user purchase special equipment *Med. Phys. 36, (2009)
7 Background: Dose Calibration Protocols Dosimetry protocols for linear accelerators have evolved away from methods that use an in-air ionization chamber calibration factor and towards those such that the ionization chamber is calibrated in water For in-water-calibrated ionization chambers, only a few simple multiplicative factors are required to convert the ionization chamber reading to dose-to-water.
8 Background: Dose Calibration Protocols Ionization chamber calibrated in water IAEA TRS-398 (2000) AAPM TG-51 (1999) IAEA Working Group (Alfonso 2008)* Parameters Simple measured Simple measured Simple measured + Monte Carlo (MC) generated Advantages Simple to use Machine-specific correction factor Accounts for different ionization chambers and phantom types Disadvantages *Med. Phys. 35, (2008) Designed for use with linear accelerators: 10 x 10 cm2 field Water phantom MC parameters must be generated for each ionization chamber and phantom combination
9 Round-robin dosimetry inter-comparison 9 participating institutions; 10 GK units (3 Model 4C and 7 PFX) Each institution performed at least 8 sets of measurements: 3 different spherical phantoms 1 in-air jig 2 different ionization chambers Compared 4 dosimetry protocols/formalisms TG-21 Meltsner-DeWerd In-air TG-51 IAEA Alfonso formalism Results published: Drzymala, et al., Med. Phys. 42(11), , 2015
10 Equipment and Measurements Elekta ABS Elekta solid water (SW) Phantoms Ionization Chambers PTW TN31010 (PTW) Standard Imaging Exradin A16 (Ex) Electrometer: Collecting Volume cc cc Standard Imaging Max 4000 Phantom laboratories liquid water (LW) Standard Imaging in-air jig In-air measurement
11 Normalization of Data To compare data from each institution, dose rates calculated with each protocol were normalized to the dose rate determined by that institution in the ABS phantom using the TG-21 protocol. This normalization was chosen because historically the ABS has been the most commonly used phantom in GK dosimetry, and TG-21 is currently the only protocol directly applicable to GK for the 3 phantom types considered.
12 Results Averaged over all Institutions Sorted by Dosimetry Protocol/Formalism
13 Results Averaged over all Institutions Sorted by Dosimetry Protocol/Formalism
14 Results Averaged over all Institutions Sorted by Dosimetry Protocol/Formalism
15 Comparing relative advantages of viable protocols TG-21 MDW In-air New IAEA Alfonso Ionization chamber calibration is in terms of absorbed dose to water Can be used for different phantom materials NA User need not purchase new equipment Requires only a few simple calculations Protocol parameters readily available in literature
16 Comparing relative advantages of viable protocols TG-21 MDW In-air New IAEA Alfonso Ionization chamber calibration is in terms of absorbed dose to water Can be used for different phantom materials NA User need not purchase new equipment Requires only a few simple calculations Protocol parameters readily available in literature
17 ff mmmmmm,ff rrrrrr IAEA (Alfonso) Formalism kk QQmmmmmm,QQ factors from Elekta Phys. Rep.* Take a closer look *Johansson and Gorka, Elekta Physics Report: Reference No. SSM 2010/2201, Project nr )
18 ff Alfonso kk mmmmmm,ff rrrrrr factors for PFX from Elekta Phys. Rep. QQmmmmmm,QQ
19 ff Alfonso kk mmmmmm,ff rrrrrr factors for PFX from Elekta Phys. Rep. QQmmmmmm,QQ For IC Vol > 0.05 cm ± ± ±
20 Potential Recommendation for Perfexion 1) Ionization chamber: If possible, select a thimble ionization chamber with collecting volume > 0.05 cc, but small enough to satisfy smallfield conditions (active cylindrical volume < 7 mm length, 5 mm diameter) 2) Phantom: Select 8-cm radius spherical phantom based on personal preference 3) Protocol: New IAEA Alfonso Formalism* with parameters shown on next slide - *TG-21 remains a reasonable option, but not optimal
21 Potential Recommendation for new IAEA Alfonso formalism Perfexion 1) For ionization chambers with collecting volume 0.05 cm 3, * ff mmmmmm,ff rrrrrr kk QQmmmmmm,QQ = for liquid water phantom for Elekta Solid Water phantom for Elekta ABS phantom 2) For ionization chambers with collecting volume < 0.05 cm 3, ff mmmmmm,ff rrrrrr kk QQmmmmmm,QQ = Tabulated value from Elekta Phys. Report, if available Value determined from new M.C. calculations *Contingent upon corroboration of K-factors by other investigations, and upon the publication of these data
22 Thank you for your attention!
Basic dosimetric data for treatment planning system
Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny
More informationDosimetry of the Gamma Knife TM
Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationIAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada
IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans
More informationPhantom Positioning Variation in the Gamma Knife Perfexion Dosimetry
Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View
More informationAbstractID: 8073 Title: Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Seattle, WA May 2008
Knife Radiosurgery S. Goetsch, Ph.D. Page 1 Quality Assurance, Planning and Clinical Results for Gamma Knife Radiosurgery Steven J. Goetsch, Ph.D., FAAPM ACMP 25 th Annual Meeting Seattle, WA May 2008
More informationAAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams
AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams Peter R. Almond, Brown Cancer Center,Louisville, KY 40202 Peter J. Biggs, Department of Radiation Oncology,
More informationCanadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:
Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization
More informationThe RSS and ACRO Program for Distinction in Stereotactic Radiation Therapy
The RSS and ACRO Program for Distinction in Stereotactic Radiation Therapy Practical Medical Physics: Radiation Oncology Accreditation Programs AAPM Annual Meeting, Nashville, TN July 30, 7:30-9:30 Presented
More informationStudy of a spherical phantom for Gamma knife dosimetry
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Study of a spherical phantom for Gamma knife dosimetry Dengsong Zhu, 1a Carlos Austerlitz, 2 Sidi Benhabib, 2 Helvecio Mota,
More informationDose differences between the three dose calculation algorithms in Leksell GammaPlan
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Dose differences between the three dose calculation algorithms in Leksell GammaPlan Andy (Yuanguang) Xu, 1a Jagdish Bhatnagar 1, Greg
More informationRecent advances in dosimetry in reference conditions for proton and light-ion beams
Recent advances in dosimetry in reference conditions for proton and light-ion beams S. Vatnitskiy a), P. Andreo b) and D.T.L. Jones c) a) MedAustron, Wiener Neustadt, Austria b) Medical Radiation Physics,
More informationDosimetric characterization of surface applicators for use with the Xoft ebx system
Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal
More informationBeam Quality Effects in Nonstandard Fields of the Varian TrueBeam
Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center
More informationSmall Field Dosimetry
Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict
More informationIAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy
IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK
More informationImplantable MOSFET dosimeter response to 192 Ir HDR radiation
Implantable MOSFET dosimeter response to 192 Ir HDR radiation Jessica Fagerstrom University of Wisconsin Madison, Department of Medical Physics Medical Radiation Research Center North Central Chapter AAPM
More informationOn the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams
On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More informationSmall Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)
Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationIEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects
IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects 2015 AAPM Annual Meeting Anaheim, CA Wes Culberson, PhD, DABR University of Wisconsin
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationComparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom
Volume 3, No 1/2016 Research Article Malays. j. med. biol. res. ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable
More informationHow accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement
Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement
More informationEric E. Klein, Ph.D. Chair of TG-142
Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing
More informationAssessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
More informationAcknowledgments & Credits 8/2/2012
Dosimetryfor Ir- HDR Brachytherapy Present Status and Future Direction TU-A-213AB L DeWerd 1 *, J Seuntjens 2 *, M Rivard 3 *, M McEwen 4 Approaches for the determination and dissemination of absorbed
More informationIMRT QA: Point Dose Measurements or 2D Array?
QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify
More informationRadiosurgery by Leksell gamma knife. Josef Novotny, Na Homolce Hospital, Prague
Radiosurgery by Leksell gamma knife Josef Novotny, Na Homolce Hospital, Prague Radiosurgery - Definition Professor Lars Leksell The tools used by the surgeon must be adapted to the task and where the human
More informationAAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites
AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites A. Purpose and Rationale Many individual users contact the AAPM Brachytherapy
More informationManik Aima, Larry A. DeWerd, Wesley S. Culberson
Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,
More informationKaren E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria
Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)
More information4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest
SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific
More informationSRS/SBRT Errors and Causes
SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research
More informationTransition to Heterogeneity Corrections. Why have accurate dose algorithms?
Transition to Heterogeneity Corrections Eric E. Klein, M.S., Washington University, St. Louis, MO Craig Stevens, M.D., Ph.D., MD Anderson Cancer Center, Houston, TX Nikos Papinikolou, Ph.D., University
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More information7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.
UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems
More informationMuch of what we will say
New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer
More informationThe need for standardization of dosimetry in experimental radiation biology
The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More information8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit
MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.
More informationIon recombination corrections of ionization chambers in flattening filter-free photon radiation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,
More informationPHYSICS NEEDS IN RADIATION SCIENCES. James Deye, Ph. D. Contractor, NCI Radiation Research Program
PHYSICS NEEDS IN RADIATION SCIENCES James Deye, Ph. D. Contractor, NCI Radiation Research Program 6/21/2016 Reproducibility / Validation / Efficacy Previous Workshops: 2003 recommendations (1) Establish
More informationDisclosure SBRT. SBRT for Spinal Metastases 5/2/2010. No conflicts of interest. Overview
Stereotactic Body Radiotherapy (SBRT) for Recurrent Spine Tumors Arjun Sahgal M.D., F.R.C.P.C. Assistant Professor Princess Margaret Hospital Sunnybrook Health Sciences Center University of Toronto Department
More informationHot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis
Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline
More informationConcepts of Uncertainty and their Application in Gamma Knife SRS. David J. Schlesinger, Ph.D. University of Virginia
Concepts of Uncertainty and their Application in Gamma Knife SRS David J. Schlesinger, Ph.D. University of Virginia (djs9c@virginia.edu) Disclosures Research support: Elekta, AB Learning Objectives Review
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationAlbert Lisbona Medical Physics Department CLCC Nantes Atlantique a
The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department
More informationSmall Field Dosimetry Issues: Theory, Calibration and Treatment Planning
Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Professor, Department of Medical Physics
More informationINTRODUCTION. Material and Methods
Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University
More informationAbsolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro
In Phantoms Vivo treatment for Absolute Dosimetry verification Versatile solid and water phantoms Introduction Introduction Introduction Introtro Intro Intro The rapid development of advanced treatment
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,
More informationPMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.
Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers
More informationFLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system
FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system Collaboration Project: Θ Θ Θ INFN, Milan; Medical Physics Department, Niguarda Ca'Granda Hospital, Milan; Elekta
More informationDosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields
Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements
More informationIMRT point dose measurements with a diamond detector
IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationDetector comparison for small field output factor measurements with flattening filter free photon beams
Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory
More informationPHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)
PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce
More informationCurrent Status of Electronic Brachytherapy Dosimetry
Current Status of Electronic Brachytherapy Dosimetry 2014 NCCAAPM Fall Meeting La Crosse, WI Wes Culberson, PhD, DABR University of Wisconsin Madison University of Wisconsin Medical Radiation Research
More informationMania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane
Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors
More informationLet s speak the same language: Standardization in Terminology
Faculty Disclosure Let s speak the same language: Standardization in Terminology Michael Torrens Gamma Knife Department, Hygeia Hospital, Athens, Greece Why is standardization necessary? Neurosurgical
More informationABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007
ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory
More informationStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry
More informationSmall Field Dosimetry & Radiochromic film
Small Field Dosimetry & Radiochromic film Michael Kissick, Ph.D., Assistant Professor, Medical Physics & Human Oncology & Affiliated with the Morgridge Institute for Research North Central AAPM Chapter
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More informationSmall Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)
Indra J. Das/ Small Fields (Page-1) Small Field Dosimetry: Overview of AAP TG-155 and the IAEA-AAP Code of Practice (Therapy) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana
More informationRotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies
Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies J. Y. C. Cheung and K. N. Yu a Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong,
More informationU.S. Nuclear Regulatory Commission
Page 1 of 8 Index Site Map FAQ Help Glossary Contact Us A U.S. Nuclear Regulatory Commission Home Who We Are What We Do Nuclear Reactors Nuclear Materials Radioactive Waste Facility Info Finder Public
More informationph fax
This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY
More informationThe Journey of Cyberknife Commissioning
The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources
More informationDevelopment and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.
Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationCALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE
CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of
More informationThe determination of timer error and its role in the administration of specified doses
JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School
More informationApplications of Monte Carlo simulations to radiation dosimetry
Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationThree Dimensional Dosimetry. Session Program: Therapy Education Session Organizer: L. John Schreiner
Three Dimensional Dosimetry Session Program: Therapy Education Session Organizer: L. John Schreiner Speakers: LJ Schreiner 1, S Ceberg 2, T Juang 3, G Ibbott 4 1 Cancer Center of Southeastern Ontario,
More informationPhysics. Copyright 2010 Xoft, Inc.
Physics Outline The XOFT System Dosimetry of X-Ray Source: TG43 Balloon Applicators and APBI Endo Rectal Applicators Cervical Applicator TG61 Surface Applicators Radioprotection: Shielding Room exposure
More informationDosimetry and QA of proton and heavier ion beams
Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationSTEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS
PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic
More information8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient
Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,
More informationActive Scanning Beam 3 Checking Delivery/Dosimetry
Active Scanning Beam 3 Checking Delivery/Dosimetry C Algranati, J Salk, A Coray, A Lomax, E Pedroni, T Boeringer, S Lin, E Hug Center for Proton Radiation Therapy Checking Delivery/Dosimetry Contents Absolute
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More informationComparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations
Iran. J. Radiat. Res., 21; 7 (4): 217-222 Comparison of dosimetry parameters of two commercially available Iodine brachytherapy seeds using Monte Carlo calculations Downloaded from ijrr.com at 6:52 +33
More informationComparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry
Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry C. Candela-Juan, J. Vijande, T. García-Martínez, Y. Niatsetski,
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationHow accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues
Accurate Clinical dosimetry Part I: Fundamentals and Issues Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement effect of dose uncertainty
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationSmall Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition
Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of
More informationDosimetry of Small Static Fields Used in External Beam Radiotherapy
Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More information