Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam
|
|
- Vivien Kimberly Kennedy
- 5 years ago
- Views:
Transcription
1 Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center
2 Introduction FFF motivation: FF unnecessary for advanced techniques Faster hypofractionated treatments FFF studies published from Varian TrueBeam TM 6MV and 10MV FFF modes 1. D. M. Galbraith, Med. Phys. 16, (1989) 2. (Image source) Varian Medical Systems, Palo Alto, CA, USA, 2
3 FFF Dosimetry Challenges Profile shape More peaked Less variation with depth Spectrum Softer Less off-axis variation Dose rate Up to 2400 MU/min (4x higher) Volume averaging Energy dependence Beam quality corrections Dose-rate dependence 1 1. (Image source) O. N. Vassiliev, U. Titt, F. Pӧnisch et al., Phys. Med. Biol. 51 (2006) 3
4 FFF Dosimetry TG-51 1,2 Beam quality specified by PDD(10) X FFF accounted for using Pb foil TRS Beam quality specified by TPR 20,10 FFF discrepancies up to 2% Need additional TPR to correct for beam quality Dose relative to 10x Output Factors 6MV 6FFF 10MV 10FFF Side of square field (cm) Alfonso et al. 4 (TG-155) 1. P. R. Almond, P. J. Biggs, B. M. Coursey et al., Med. Phys. 26 (1999) 2. M. McEwen, L. DeWerd, G. Ibbott et al., Med. Phys. 41 (2014) 3. IAEA TRS-398, International Atomic Energy Agency (2004) 4. R. Alfonso, P. Andreo, R. Capote et al., Med. Phys. 35 (2008) 4
5 Beam Quality Correction Factors Corrected for in TG-51:, For FFF 10x10 (MSR) field: For small fields:,,,,,,,,,,,, Ω,,,, Ω,,,, 1. P. R. Almond, P. J. Biggs, B. M. Coursey et al., Med. Phys. 26 (1999) 2. M. McEwen, L. DeWerd, G. Ibbott et al., Med. Phys. 41 (2014) 3. R. Alfonso, P. Andreo, R. Capote et al., Med. Phys. 35 (2008),,,,
6 Measurement Methods Reference detectors L-α-alanine pellets (10x10, 5x5, 2x2 cm 2 ) Exradin W1 PSD (10x10 to 0.6x0.6 cm 2 ) Corrected detectors Ionization chambers Exradin A12 Farmer-type chamber Exradin A1SL scanning chamber Exradin A26 microchamber Diodes Sun Nuclear Edge PTW E diode Measured: D w (alanine, W1) M (chambers, diodes) 6MV, 6FFF, 10MV, 10FFF, 60 Co,,,, (Image source) Far West Technology, Inc., Goleta, CA, USA, 2. Virtual Water TM paddles, manufactured by Med-Cal Inc., Verona, WI 3. (Image source) Standard Imaging, Inc., Middleton, WI 6
7 Monte Carlo Model The TrueBeam TM accelerator was modeled using BEAMnrc/EGSnrc Varian-provided phase spaces Simplified geometries PDDs and profiles were obtained using DOSXYZnrc Compared to commissioning data (measured with CC13 ionization chamber) JAWS: X and Y jaws JAWS: tungsten shield 35 x 35 x 35 cm 3 water phantom CHAMBER: steel baseplate 0.5 x 0.5 x 0.5 cm 3 voxels 7
8 Model Validation: PDDs 8
9 Model Validation: Profiles 9
10 Monte Carlo Methods Chambers modeled in egs_chamber/egsnrc D ch : dose to air in chamber volume D w : dose to water at measurement point,, Point sources at 100 cm SSD 1. B. R. Muir and D. W. O. Rogers, Med Phys 37 (2010) 10
11 Results: MSR Fields,,,, 11
12 Results: A1SL k Q 12
13 Results: A12 k Q 13
14 Results: 6MV Small-Field Corrections Scintillator-based,,,,,,,, 14
15 Results: 6FFF Small-Field Corrections Scintillator-based,,,,,,,, 15
16 Conclusions and Future Work Conclusions Measured,, effects up to 0.7% for the chambers in 10x10 cm 2 fields Agreed with TG-51-predicted values and Monte Carlo results within uncertainties Measured,, effects greater than 6% for chambers and diodes in fields down to 6x6 mm 2 1 Future Work Small-field Monte Carlo Treatment planning study 1. (Image source) Varian Medical Systems, Palo Alto, CA, USA, 16
17 Acknowledgements Dr. Larry DeWerd UW Carbone Cancer Center Dr. Zac Labby Dr. Wesley Culberson Dr. Jessica Miller UWMRRC students and staff Special thanks to: John Micka Mike Lawless Sam Simiele UW Accredited Dosimetry Calibration Lab customers Continued support helps fund student research National Physical Laboratory (UK) Claire Gouldstone Russ Thompson Simon Duane Thank you for your attention! Questions are welcome 17
18 FFF Spectral Effects 18
19 Risks vs. Benefits of FFF: Controversy Benefits Higher dose rate Faster treatment Reduced head scatter Less off-axis spectral variation Easier beam modeling Risks Higher dose rate Interplay effect Target motion errors (coughing) Potentially increased in-patient scatter and surface dose 2 1. C. Liu, M. G. Snyder, and C. G. Orton, Med. Phys. 40 (2013) 2. (Image source) Varian Medical Systems, Palo Alto, CA, USA, 19
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationDetector comparison for small field output factor measurements with flattening filter free photon beams
Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory
More informationIon recombination corrections of ionization chambers in flattening filter-free photon radiation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,
More informationOn the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams
On the impact of ICRU report 90 recommendations on k Q factors for high-energy photon beams Ernesto Mainegra-Hing a) and Bryan R. Muir Measurement Science and Standards, National Research Council of Canada,
More informationDosimetric characterization of surface applicators for use with the Xoft ebx system
Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal
More informationSmall Field Dosimetry
Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict
More informationSmall Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)
Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationCurrent Status of Electronic Brachytherapy Dosimetry
Current Status of Electronic Brachytherapy Dosimetry 2014 NCCAAPM Fall Meeting La Crosse, WI Wes Culberson, PhD, DABR University of Wisconsin Madison University of Wisconsin Medical Radiation Research
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationAn Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center
An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationHot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis
Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline
More informationPMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.
Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers
More informationImplantable MOSFET dosimeter response to 192 Ir HDR radiation
Implantable MOSFET dosimeter response to 192 Ir HDR radiation Jessica Fagerstrom University of Wisconsin Madison, Department of Medical Physics Medical Radiation Research Center North Central Chapter AAPM
More informationThe need for standardization of dosimetry in experimental radiation biology
The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall
More information4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest
SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific
More informationIAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada
IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans
More informationMuch of what we will say
New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer
More informationSmall Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)
Indra J. Das/ Small Fields (Page-1) Small Field Dosimetry: Overview of AAP TG-155 and the IAEA-AAP Code of Practice (Therapy) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationSmall Field Dosimetry Issues: Theory, Calibration and Treatment Planning
Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Professor, Department of Medical Physics
More informationIAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy
IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationBasic dosimetric data for treatment planning system
Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny
More informationReference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor
Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The
More informationAssessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
More informationMEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn
MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the
More informationManik Aima, Larry A. DeWerd, Wesley S. Culberson
Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,
More informationPRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR.
IX Latin American IRPA Regional Congress on Radiation Protection and Safety - IRPA 2013 Rio de Janeiro, RJ, Brazil, April 15-19, 2013 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR PRECISE DOSIMETRY
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationThe Journey of Cyberknife Commissioning
The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources
More informationMania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane
Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors
More informationARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.
ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael
More informationComparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry
Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry C. Candela-Juan, J. Vijande, T. García-Martínez, Y. Niatsetski,
More informationHow accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement
Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement
More information3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe
3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve
More informationDosimetry of the Gamma Knife TM
Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife
More informationThe influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainLAB m3 micro-multileaf collimator
JOURNA OF APPIED CINICA MEDICA PHYSICS, VOUME 13, NUMBER 6, 2012 The influence of field size on stopping-power ratios in- and out-of-field: quantitative data for the BrainAB m3 micro-multileaf collimator
More informationDosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields
Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements
More informationVerification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams
Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health
More informationAAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams
AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams Peter R. Almond, Brown Cancer Center,Louisville, KY 40202 Peter J. Biggs, Department of Radiation Oncology,
More informationA. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.
Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1
More informationDose rate response of Digital Megavolt Imager detector for flattening filter-free beams
Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More information8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit
MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.
More informationKaren E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria
Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)
More informationEXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *
Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West
More informationSmall field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators Madelaine K. Tyler,
More informationIEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects
IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects 2015 AAPM Annual Meeting Anaheim, CA Wes Culberson, PhD, DABR University of Wisconsin
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationSmall Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition
Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of
More informationSRS/SBRT Errors and Causes
SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research
More informationA comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment
Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams
More informationApplications of Monte Carlo simulations to radiation dosimetry
Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationProgress Report on Radiation Dosimetry at NPL
Progress Report on Radiation Dosimetry at NPL Peter Sharpe and Simon Duane National Physical Laboratory Teddington, UK April 2009 1 Introduction This report gives a brief overview of radiation dosimetry
More informationDosimetry of Small Static Fields Used in External Beam Radiotherapy
Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA
More informationAAPM TG178: Towards the Recommendation of a Standard Dosimetry Protocol for Gamma Knife Stereotactic Radiosurgery
AAPM TG178: Towards the Recommendation of a Standard Dosimetry Protocol for Gamma Knife Stereotactic Radiosurgery Paula L. Petti, PhD Washington Hospital Healthcare System Fremont, California, USA AAPM
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationBorges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1
Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationA Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer
IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer
More informationHow accurately can we measure dose clinically? Dosimetric accuracy requirement. Accurate Clinical dosimetry. Part I: Fundamentals and Issues
Accurate Clinical dosimetry Part I: Fundamentals and Issues Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement effect of dose uncertainty
More informationAn Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements
An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements Jean M. Moran, a Donald A. Roberts, Teamour S. Nurushev, Larry E. Antonuk, Youcef El-Mohri, and Benedick A. Fraass Department
More informationtsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM
tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM A Thesis Presented to The Academic Faculty by Ryan J. Underwood In Partial Fulfillment of the Requirements for the Degree Master of Science in the School
More informationInhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system
Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.
More informationSTEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS
PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic
More informationINTRODUCTION. Material and Methods
Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationProtection of the contralateral breast during radiation therapy for breast cancer
Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,
More informationDOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS
Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National
More informationComparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom
Volume 3, No 1/2016 Research Article Malays. j. med. biol. res. ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationCalibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia
Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute
More informationIMRT point dose measurements with a diamond detector
IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationTESTICULAR DOSES IN IMAGE-GUIDED RADIOTHERAPY OF PROSTATE CANCER
doi:10.1016/j.ijrobp.2011.01.071 Int. J. Radiation Oncology Biol. Phys., Vol. 82, No. 1, pp. e39 e47, 2012 Copyright Ó 2012 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ - see front
More informationInfluence of the exposure time in the area monitors at radiodiagnostic
Journal of Physics: Conference Series PAPER OPEN ACCESS Influence of the exposure time in the area monitors at radiodiagnostic To cite this article: F Leyton et al 2016 J. Phys.: Conf. Ser. 733 012087
More informationEnergy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range J. Eduardo Villarreal-Barajas, a and Rao
More informationEvaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf
More informationStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry
More informationNeutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics
Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,
More informationIAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*
THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service
More informationImplementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Implementation of full/half bowtie filter models in a commercial treatment planning system for kilovoltage cone-beam CT dose estimations
More informationElectron Beam ET - Reloaded Therapy - Reloaded
8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles
More informationRecent advances in dosimetry in reference conditions for proton and light-ion beams
Recent advances in dosimetry in reference conditions for proton and light-ion beams S. Vatnitskiy a), P. Andreo b) and D.T.L. Jones c) a) MedAustron, Wiener Neustadt, Austria b) Medical Radiation Physics,
More informationROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning
More informationFlattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group.
Thomas Jefferson University Jefferson Digital Commons Department of Medicine Faculty Papers Department of Medicine 5-8-2015 Flattening filter-free accelerators: a report from the AAPM Therapy Emerging
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationDosimetry and QA of proton and heavier ion beams
Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationNormal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT
Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using
More information