Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Size: px
Start display at page:

Download "Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan"

Transcription

1 Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

2 Oxygen transport in blood Hemoglobin Oxygen O Deoxygen +O O Oximeters

3 Arterial saturation of oxygen SaO SaO is affected by partial pressure of arterial oxygen, PO. At low partial pressures, such as tissue capillary pressure of 40 mmhg, oxygen is released from hemoglobin, leading to a low SaO of about 75%. At high partial pressures, such as pulmonary capillary pressure of 95 mmhg, oxygen binds with hemoglobin, leading to high SaO of about 95 mmhg. Oximeters 3

4 Oxyhemoglobin dissociation curve Oximeters 4

5 Beer-Lambert s law Oximeters 5

6 Light interaction with absorption effect Beer-Lambert s law I x x I x di x ai x dx ax I x I x e 0 ai x x Absorbance of a wavelength of light absorption path length, l concentration, c absorbing species by ελ, molar absorptivity Oximeters 6

7 Molar absorptivities of hemoglobin and oxyhemoglobin Oximeters 7

8 Electromagnetic spectrum Oximeters 8

9 Oxygen saturation SaO, SpO measurement Relative amount of oxygen carried by the hemoglobin Two specific wavelengths : λ 1 : a red wavelength eg. 660 nm λ : a near infrared wavelength eg. 805 nm SaO C C O O C Oximeters 9

10 Light absorption signal produce AC output produce DC output Oximeters 10

11 SaO 0 d d C C IR IR O O IR e I I 0 d d C C R R O O R e I I IR O O IR R O O R DC IR IR DC R R C C C C I I I I IR R ln ln O R R O IR IR R IR O O IR R IR R C C C SaO Oximeters 11

12 Near-infrared spectroscopy NIRS by laser diode Uses the near-infrared region from 800 nm to 500 nm Applications in pharmaceutical and medical diagnostics including blood sugar and oximetry. OxiplexTS, ISS Inc, USA Oximeters 1

13 NIRS sensing by LEDs An sensor with central emitter and eight surrounding, detachable detectors Oximeters 13

14 Evoked cerebral function examination Electroencephalography EEG Near infrared NIR Magnetoencephalography MEG Functional MRI

15 Cerebral functional areas Oximeters 15

16 Neurovascular coupling Neurovascular coupling produces changes in blood flow and oxygen consumption Oximeters 16

17 Continuous-wave NIRS Ratio of reflected to incident light intensity, I/I 0, depends Absorption coefficient Changes of hemoglobin concentration [O ] [H] Path length L and differential path length factor B A I ln I O 0 u L [ O ] H [ H] B L Oximeters 17

18 Continuous-wave NIRS cont. Using two NIRS lights with different wavelengths for computing the change of hemoglobin concentrations ] [ ] [ O H O H H H O H O H O O u u H u u O Oximeters 18

19 Application in brain oxygenation With 0% of oxygen consumption occurring in the human brain, any deficiency in oxygen supply may result in injury OxiplexTS, ISS Inc, USA Biomedical Optic 19

20 Cerebral Near-Infrared spectroscopy monitoring Monitoring of cerebral oxygenation Carotid stenosis or occlusion Brain ischemia Sleep apnea Ischemia of rat model Ischemia Cerebral response to carotid balloon inflation Obstructive sleep apnea

21 Related functional area for NIRS Prefrontal cortices PFC Premotor cortices PMC Supplement motor areas SMA Sensorimotor cortices SMC Oximeters 1

22 Brain oxygenation & reorganization in stroke patients Before rehabilitation oxy After rehabilitation Gait task Total oxy deoxy Lesion Lesion

23 Application in peripheral vascular disease PVD PVD is a narrowing of the vessels carrying blood to the muscles in the legs and arms. Most patients report experiences of pain in the extremities due to inadequate blood flow and oxygen delivery to the exercising muscle. OxiplexTS, ISS Inc, USA Oximeters 3

24 Vessel occulsion test

25 NIRS demonstration Pulse Oximetry NEJM Muscular oxygenation under exercise Functional NIRS

26 Reference G. D. Baura, Medical device technologies, Elsevier Inc., 01. JG. Webster, Medical Instrumentation, application and design, 3rd, Houghton Mifflin, 000. JJ.Carr, JM.Brown, Introduction to Biomedical Equipment Technology, 4nd Edition, Prentice-Hall, 000.

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Pulse oximeter Masimo pulse CO-oximeter http://www.masimo.com/produc ts/continuous/radical-7/

More information

Muscle OxyGen monitor. The Science Behind Moxy

Muscle OxyGen monitor. The Science Behind Moxy Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the

More information

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Takafumi Nagaoka, Kaoru Sakatani, Takayuki Awano, Noriaki Yokose, Tatsuya Hoshino, Yoshihiro

More information

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Research Journal of Pharmaceutical, Biological and Chemical Sciences Research Journal of Pharmaceutical, Biological and Chemical Sciences Oxygen Glucose and Heart rate Monitor: An non Invasive approach Anne Frank Joe A*, Veeramani Saikrupa, and Prajukti Panda Sathyabama

More information

Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University

Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University chanhl@mail.cgu.edu.tw Cerebral function examination Electroencephalography (EEG) Near infrared ray spectroscopy

More information

Disclosures. Set up audience participation. Test question. Outline. Neuromonitoring What and When? IP for monitoring technology licensed to Medtronic

Disclosures. Set up audience participation. Test question. Outline. Neuromonitoring What and When? IP for monitoring technology licensed to Medtronic Neuromonitoring What and When? Disclosures IP for monitoring technology licensed to Medtronic Ken Brady, MD Pediatrics, Anesthesia, Critical Care Texas Children s Hospital Baylor College of Medicine Set

More information

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology Should NIRS be Standard Care for Pediatric CPB Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University Director of Pediatric Cardiac Anesthesiology Stanford Children s Hospital

More information

Optical Sensor System for Hemoglobin Measurement

Optical Sensor System for Hemoglobin Measurement International Journal of Computational Engineering Research Vol, 03 Issue, 7 Optical Sensor System for Hemoglobin Measurement Rajashree Doshi 1, Anagha Panditrao 2 1 Department of Instrumentation and Control,

More information

ALCO Regulations. Protocol pg. 47

ALCO Regulations. Protocol pg. 47 For the EMT-Basic Objectives Understand ALCO regulations relative to monitoring pulse oximetry by the EMT-B Review the signs and symptoms of respiratory compromise. Understand the importance of adequate

More information

Clinically Available Optical Topography System

Clinically Available Optical Topography System Clinically Available Optical Topography System Clinically Available Optical Topography System 18 Fumio Kawaguchi Noriyoshi Ichikawa Noriyuki Fujiwara Yûichi Yamashita Shingo Kawasaki OVERVIEW: Progress

More information

Electroencephalograms and Neuro-Rehabilitation. Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University

Electroencephalograms and Neuro-Rehabilitation. Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University Electroencephalograms and Neuro-Rehabilitation Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University chanhl@mail.cgu.edu.tw Brain Cerebrum ( 大腦 ) Receives and processes conscious sensation

More information

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve SpO2 Monitoring Contents 1 Introduction 1 What is SpO 2? How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve 5 How Do I Use SpO2? Choosing a Sensor

More information

Consumer Near Infrared Spectroscopy as a Training and Recovery Tool. Jon Tarkington

Consumer Near Infrared Spectroscopy as a Training and Recovery Tool. Jon Tarkington Consumer Near Infrared Spectroscopy as a Training and Recovery Tool Jon Tarkington Outline - What is NIRS? - Devices - Data interpretation - Practical Application - Future Impacts - Suffix light Disclaimers

More information

CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018

CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 OBJECTIVES 1. Understand how cerebral oximetry works 2. Understand how cerebral oximetry may guide intervention 3. Learn

More information

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Pulse oximetry revisited Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Topics that will be discussed and dissected: Revisiting physiology

More information

Biomedical Instrumentation D. The Photoplethysmogram

Biomedical Instrumentation D. The Photoplethysmogram Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease

More information

Monitoring Tumor Therapy

Monitoring Tumor Therapy Monitoring Tumor Therapy Zenalux Biomedical, Inc. The Zenascope was used to monitor the physiological response to a vascular disrupting agent in mouse tumors. Changes in hemoglobin concentration, hemoglobin

More information

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices By Zhang Feng, Senior Medical Applications Engineer Marten Smith, Medical Marketing Manager Medical Products Group Microchip

More information

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Nazo Haroon Department of Mechatronics Engineering, CEME National University

More information

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values 1 Life saving technology The availability of the LED provided a convenient light which provides advantages of :- Compact size Modest power requirements

More information

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Takahiro Igarashi, Kaoru Sakatani, Norio Fujiwara, Yoshihiro

More information

Quantitative Performance Assessment Studies in Air Transportation

Quantitative Performance Assessment Studies in Air Transportation Quantitative Performance Assessment Studies in Air Transportation Portable Brain Activity Monitoring Technology Kurtulus Izzetoglu, PhD Drexel University Optical Brain Imaging Group Drexel University,

More information

Estimation of Haemoglobin Using Non-Invasive Technique

Estimation of Haemoglobin Using Non-Invasive Technique Estimation of Haemoglobin Using Non-Invasive Technique Tatiparti Padma #1, Pinjala Jahnavi *2 #Department of ECE, GRIET, Hyderabad, Telangana *Department of ECE, GRIET, Hyderabad, Telangana 1tatipartipadma@gmail.com,

More information

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE PAK BULLET TRAIN (PBT) JFET 22(2) (2015) 39-44 Journal of Faculty of Engineering & Technology journal homepage: www.pu.edu.pk/journals/index.php/jfet/index DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH

More information

HEMOENCEPHALOGRAPHY NEUROFEEDBACK MECHANISM OF ACTION

HEMOENCEPHALOGRAPHY NEUROFEEDBACK MECHANISM OF ACTION Trakia Journal of Sciences, No 4, pp 354-358, 2017 Copyright 2017 Trakia University Available online at: http://www.uni-sz.bg ISSN 1313-7069 (print) ISSN 1313-3551 (online) doi:10.15547/tjs.2017.04.016

More information

JAMES R. KNIGHT

JAMES R. KNIGHT JAMES R. KNIGHT knightjr@ah.org James is currently the supervisor of clinical engineering at Sonora Regional Medical Center. He is the chairman of the CMIA Training & Education Committee. He also develops

More information

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood International Journal of Electronics Engineering Research. ISSN 0975-6450 Volume 9, Number 4 (2017) pp. 483-489 Research India Publications http://www.ripublication.com Development of an Optical Sensor

More information

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood Oxygen Saturation Monitors & Pulse Oximetry D. J. McMahon 141105 rev cewood 2017-11-14 2 3 Key Points Oxygen Saturation Monitors & Pulse Oximetry : Role of hemoglobin in respiration Basic principle of

More information

Appendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires:

Appendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires: Answer Key Appendix D-2 1. An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires: a. oxygen given via nasal cannula b. immediate transport to a medical facility c.

More information

Blood Pressure. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Blood Pressure. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Bood Pressure Hsiao-Lung Chan, Ph.D. Dept Eectrica Engineering Chang Gung University, Taiwan chanh@mai.cgu.edu.tw Bood pressure traves from aorta toward smaer vesses 2 Bood pressure 3 Hagen-Poiseuie s

More information

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method To cite

More information

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Ashita Diwate 1, Yogesh Kumar 2, Dr. Sanjeev Kumar 3, Prof. T.

More information

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers ISSN 1447-4999 ORIGINAL RESEARCH Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers Dr Gyorgyi Kamaras MD, FCEM 1,2, Dr Tamas Geller MD 1, Dr Csaba Dioszeghy MD

More information

Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification

Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification by Larissa Christina Schudlo A thesis submitted in conformity with the requirements for the degree

More information

Saturday Physics Honors presents. Professor Enrico Gratton, Looking Into the Brain with Lasers. (With some demonstrations)

Saturday Physics Honors presents. Professor Enrico Gratton, Looking Into the Brain with Lasers. (With some demonstrations) Saturday Physics Honors presents Professor Enrico Gratton, Looking Into the Brain with Lasers (With some demonstrations) Can we see through the body using light? What we expect to see? Startrek Movie clip

More information

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

Appendix E Choose the sign or symptom that best indicates severe respiratory distress. Appendix E-2 1. In Kansas EMT-B may monitor pulse oximetry: a. after they complete the EMT-B course b. when the service purchases the state approved pulse oximeters c. when the service director receives

More information

Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution

Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution Takahiro Igarashi, Kaoru Sakatani, Tadashi Shibuya, Teruyasu Hirayama,

More information

Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State

Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State Chapter 35 Gender and Age Analyses of NIRS/STAI Pearson Correlation Coefficients at Resting State T. Matsumoto, Y. Fuchita, K. Ichikawa, Y. Fukuda, N. Takemura, and K. Sakatani Abstract According to the

More information

Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain

Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Motor cortex activation fmri Near-infrared imaging Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Sergio Fantini s group, Department of Biomedical

More information

Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs)

Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs) Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs) Nicoladie D. Tam a, and George Zouridakis b a Department of Biological Sciences,

More information

Health Tech Symposium Fall, Dan Sommers P.E. EMT-P

Health Tech Symposium Fall, Dan Sommers P.E. EMT-P Health Tech Symposium Fall, 2009 Dan Sommers P.E. EMT-P Human Physiological Signals Simple Explanations for Complicated Systems Ref: Atlas of Human Anatomy, 4 th Edition Simple Schematic RA LA RV LV PCR

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement

More information

Photoacoustic Imaging and Therapy in Biomedicine. Nicholas Tobey and Grace Yook. Optical Engineering. Dr. Kasra Daneshvar

Photoacoustic Imaging and Therapy in Biomedicine. Nicholas Tobey and Grace Yook. Optical Engineering. Dr. Kasra Daneshvar Photoacoustic Imaging 1 Photoacoustic Imaging and Therapy in Biomedicine Nicholas Tobey and Grace Yook Optical Engineering Dr. Kasra Daneshvar July 16, 2010 Photoacoustic Imaging 2 Abstract When a pulsed

More information

ISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164

ISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164 IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY DESING OF A HYPOXIA DEVICE DETECTOR WITH WIRELESS ALERT Ing. Teo Cortes *1 & Dra. Fatima Moumtadi *2 *Department of Electronics,

More information

Non-invasive Hemoglobin Measurement for Anemia Diagnosis

Non-invasive Hemoglobin Measurement for Anemia Diagnosis Non-invasive Hemoglobin Measurement for Anemia Diagnosis Raditya Artha Rochmanto 1, Hasballah Zakaria 2 Ratih Devi Alviana 3, Nurhalim Shahib 4 Department of Biomedical Engineering Medical Faculty Institut

More information

Is Hypoxia common despite oxygen treatment in the Acute Medical ward?

Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Dr. R Dhar, Dr. RN Harrison and Dr. PD Snashall University Hospital of North Tees, Stockton on Tees, TS198PE, United Kingdom (Received

More information

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion Journal of Physics: Conference Series Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion To cite this article: P A Kyriacou et al 27 J. Phys.: Conf. Ser. 85 1226 Related

More information

Near Infrared Spectroscopy in Medical Diagnostics

Near Infrared Spectroscopy in Medical Diagnostics Near Infrared Spectroscopy in Medical Diagnostics Michael D. Wider, PhD Vice President Of Technology Somanetics Corporation Adjunct Assistant Professor in Physiology Wayne State University School of Medicine

More information

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Maria Angela Franceschini, Sergio Fantini Bioengineering Center, Department of Electrical Engineering and

More information

Introduction to Brain Imaging

Introduction to Brain Imaging Introduction to Brain Imaging Human Brain Imaging NEUR 570 & BIC lecture series September 9, 2013 Petra Schweinhardt, MD PhD Montreal Neurological Institute McGill University Montreal, Canada Various techniques

More information

Intrinsic Signal Optical Imaging

Intrinsic Signal Optical Imaging Intrinsic Signal Optical Imaging Introduction Intrinsic signal optical imaging (ISOI) is a technique used to map dynamics in single cells, brain slices and even and most importantly entire mammalian brains.

More information

How to maintain optimal perfusion during Cardiopulmonary By-pass. Herdono Poernomo, MD

How to maintain optimal perfusion during Cardiopulmonary By-pass. Herdono Poernomo, MD How to maintain optimal perfusion during Cardiopulmonary By-pass Herdono Poernomo, MD Cardiopulmonary By-pass Target Physiologic condition as a healthy person Everything is in Normal Limit How to maintain

More information

Current bedside monitors of brain blood flow and oxygen delivery

Current bedside monitors of brain blood flow and oxygen delivery 24. Brain Chemistry Current bedside monitors of brain blood flow and oxygen delivery Global monitors Cannot detect regional abnormalities Local monitors Sample only a small region of the brain and highly

More information

ESTIMATED CONTRIBUTION OF HEMOGLOBIN AND MYOGLOBIN TO NEAR INFRARED SPECTROSCOPY MICHELLE L DAVIS. B.S., Kansas State University, 2005 A THESIS

ESTIMATED CONTRIBUTION OF HEMOGLOBIN AND MYOGLOBIN TO NEAR INFRARED SPECTROSCOPY MICHELLE L DAVIS. B.S., Kansas State University, 2005 A THESIS ESTIMATED CONTRIBUTION OF HEMOGLOBIN AND MYOGLOBIN TO NEAR INFRARED SPECTROSCOPY by MICHELLE L DAVIS B.S., Kansas State University, 2005 A THESIS submitted in partial fulfillment of the requirements for

More information

Oxygen Level Measurement Techniques: Pulse Oximetry

Oxygen Level Measurement Techniques: Pulse Oximetry Oxygen Level Measurement Techniques: Pulse Oximetry Reem Abedalmoniam Ismail 1 and Sharief F. Babiker 2 1 Computer Engineering Dept., ElNeelain University, Khartoum, Sudan 2 Electrical and Electronics

More information

Neuroimaging. BIE601 Advanced Biological Engineering Dr. Boonserm Kaewkamnerdpong Biological Engineering Program, KMUTT. Human Brain Mapping

Neuroimaging. BIE601 Advanced Biological Engineering Dr. Boonserm Kaewkamnerdpong Biological Engineering Program, KMUTT. Human Brain Mapping 11/8/2013 Neuroimaging N i i BIE601 Advanced Biological Engineering Dr. Boonserm Kaewkamnerdpong Biological Engineering Program, KMUTT 2 Human Brain Mapping H Human m n brain br in m mapping ppin can nb

More information

International Journal of Research and Applications July - Sep 2017 Transactions 4(15):

International Journal of Research and Applications July - Sep 2017 Transactions 4(15): DOI: 10.17812/IJRA.4.15(98)2017 International Journal of Research and Applications July - Sep 2017 Transactions 4(15): 585-591 B I O - M E D I C A L R E V I E W R E P O R T eissn : 2349 0020 pissn : 2394

More information

Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry

Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry Ref: Eye, M.C. Escher, 1946 Ref: Eye, M.C. Escher, 1946

More information

Competitive Intelligence Report Technology & Patents

Competitive Intelligence Report Technology & Patents Competitive Intelligence Report Technology & Patents N o n i n v a s i v e d i a b e t e s m o n i t o r i n g 01 November - 15 November, 2018 7 Published Granted Inactive 5 8 0 25 2 1 Start ups Companies

More information

Supporting Information

Supporting Information This journal is The Royal Society of Chemistry 011 Supporting Information Vertically-Aligned ZnO Nanorods Doped with Lithium for Polymer Solar Cells: Defect Related Photovoltaic Properties Pipat Ruankham,

More information

NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL

NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL Phillip J. Clancy, NIR Technology Systems, 56 Kitchener Pde, Bankstown, NSW, Australia. Near Infrared Transmission (NIT) Spectroscopy

More information

MedStar Health considers Continuous Home Pulse Oximetry medically necessary for the following indications:

MedStar Health considers Continuous Home Pulse Oximetry medically necessary for the following indications: MedStar Health, Inc. POLICY AND PROCEDURE MANUAL MP.006.MH Continuous Home Pulse Oximetry This policy applies to the following lines of business: MedStar Employee (Select) MedStar MA DSNP CSNP MedStar

More information

Near Infrared Spectroscopy in Neonatal Intensive Care

Near Infrared Spectroscopy in Neonatal Intensive Care Near Infrared Spectroscopy in Neonatal Intensive Care Adré J. du Plessis, MBChB Fetal-Neonatal Neurology, Children s Hospital, Boston, MA Children s Hospital and Harvard Medical School Disclosure Statement

More information

CEA and cerebral protection Volodymyr labinskyy, MD

CEA and cerebral protection Volodymyr labinskyy, MD CEA and cerebral protection Volodymyr labinskyy, MD VA Hospital 7/26/2012 63 year old male presents for the vascular evaluation s/p TIA in January 2012 PMH: HTN, long term active smoker, Hep C PSH: None

More information

Studying brain function with near-infrared spectroscopy concurrently with electroencephalography

Studying brain function with near-infrared spectroscopy concurrently with electroencephalography Studying brain function with near-infrared spectroscopy concurrently with electroencephalography Y. Tong* 1, E. J. Rooney 2, P. R. Bergethon 2, J. M. Martin 1, A. Sassaroli 1, B. L. Ehrenberg 1, Vo Van

More information

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM www.osram-os.com Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM Sensor products Exactly what you need: high efficiency and great diversity OSRAM Opto Semiconductors offers a great

More information

Temporal Mapping and Connectivity using NIRS for Language Related Tasks

Temporal Mapping and Connectivity using NIRS for Language Related Tasks Florida International University FIU Digital Commons FIU Electronic Theses and Dissertations University Graduate School 2-9-2012 Temporal Mapping and Connectivity using NIRS for Language Related Tasks

More information

About OMICS International

About OMICS International About OMICS International OMICS International through its Open Access Initiative is committed to make genuine and reliable contributions to the scientific community. OMICS International hosts over 700

More information

The role of pulse oximetry in the accident and emergency department

The role of pulse oximetry in the accident and emergency department Archives of Emergency Medicine, 1989, 6, 211-215 The role of pulse oximetry in the accident and emergency department M. A. LAMBERT & J. CRINNION Accident and Emergency Department, Leeds General Infirmary,

More information

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE 16 CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE OMAR GALAL, MO, PhO; NEIL WILSON, MO Pulse oximetry is a noninvasive

More information

INVOS System Inservice Guide for Pediatric Use. INVOS System Inservice Guide for Pediatric Use

INVOS System Inservice Guide for Pediatric Use. INVOS System Inservice Guide for Pediatric Use INVOS System Inservice Guide for Pediatric Use INVOS System Inservice Guide for Pediatric Use The INVOS System: A Window to Perfusion Adequacy The noninvasive INVOS System reports the venous- weighted

More information

Application of near-infrared spectroscopy for discrimination of mental workloads

Application of near-infrared spectroscopy for discrimination of mental workloads Application of near-infrared spectroscopy for discrimination of mental workloads A. Sassaroli a, F. Zheng a, M. Coutts a L. H. Hirshfield b, A. Girouard b, E. T. Solovey b, R. J. K. Jacob b, Y. Tong c,

More information

International Congress Series 1232 (2002)

International Congress Series 1232 (2002) International Congress Series 1232 (2002) 549-554 Measurement of Cerebral Circulation Using Near-Infrared Spectroscopic Topography during Music-Exercise (trampoline) Therapy as a Method of Brain Rehabilitation

More information

Dr. Puntarica Suwanprathes. Version 2007

Dr. Puntarica Suwanprathes. Version 2007 Dr. Puntarica Suwanprathes Version 2007 O 2 and CO 2 transport in blood Oxyhemoglobin dissociation curve O 2 consumption (VO 2 ) CO 2 production (VCO 2 ) O 2 capacity O 2 content: CaO 2 or CvO 2 %saturation

More information

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR P.Daarani 1 & A.Kavithamani 2 International Journal of Latest Trends in Engineering and Technology pp. 141-147 DOI: http://dx.doi.org/10.21172/1.ires.19

More information

Infinity MCable -Masimo SET

Infinity MCable -Masimo SET Infinity MCable -Masimo SET Bring the advantages of Masimo s Signal Extraction Technology (SET ) to your pulse oximetry monitoring The noninvasive, motion-tolerant Infinity MCable -Masimo SET works with

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Kyriacou, P. A., Shafqat, K. & Pal, S. K. (29). Pilot investigation of photoplethysmographic signals and blood oxygen

More information

Neurovascular Physiology and Pathophysiology

Neurovascular Physiology and Pathophysiology Neurovascular Physiology and Pathophysiology The physiological questions aim at understanding the molecular and biochemical mechanisms, by which the brain adapts local blood flow to neuronal activity and

More information

When Cyanosis is the Norm. Steven M. Schwartz, MD, FRCPC Cardiac Critical Care Medicine The Hospital for Sick Children Toronto

When Cyanosis is the Norm. Steven M. Schwartz, MD, FRCPC Cardiac Critical Care Medicine The Hospital for Sick Children Toronto When Cyanosis is the Norm Steven M. Schwartz, MD, FRCPC Cardiac Critical Care Medicine The Hospital for Sick Children Toronto No Disclosures When Cyanosis is the Norm Physiology of cyanotic congenital

More information

On behalf of the education Committee of the SNACC

On behalf of the education Committee of the SNACC Neuro Quiz 36 Cerebral Oximetry Shobana Rajan M.D, Assistant Professor Anesthesiology Cleveland Clinic Quiz Team; Suneeta Gollapudy M.D, Angele Marie Theard M.D, Verghese Cherian M.D On behalf of the education

More information

Pulse oximetry use in small animals

Pulse oximetry use in small animals Vet Times The website for the veterinary profession https://www.vettimes.co.uk Pulse oximetry use in small animals Author : Clara Rigotti, Marieke de Vries Categories : Vets Date : October 10, 2011 Clara

More information

Penetration Depth of 635 nm Laser Light Into the Biological Tissue

Penetration Depth of 635 nm Laser Light Into the Biological Tissue 2017 Published in 5th International Symposium on Innovative Technologies in Engineering and Science 29-30 September 2017 (ISITES2017 Baku - Azerbaijan) Penetration Depth of 635 nm Laser Light Into the

More information

Near-infrared signals associated with electrical stimulation of peripheral nerves

Near-infrared signals associated with electrical stimulation of peripheral nerves Invited Paper Near-infrared signals associated with electrical stimulation of peripheral nerves Sergio Fantini a, Debbie K. Chen a, Jeffrey M. Martin a,b, Angelo Sassaroli a, and Peter R. Bergethon b a

More information

Noninvasive Detection of Bilirubin in Discrete Vessels

Noninvasive Detection of Bilirubin in Discrete Vessels , July 2-4, 2014, London, U.K. Noninvasive Detection of Bilirubin in Discrete Vessels Mark McEwen, Karen J. Reynolds Abstract Bilirubin, at the levels encountered in hyperbilirubinaemia, makes a similar

More information

April 18, Dear Mr. Lei Chen:

April 18, Dear Mr. Lei Chen: DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Beijing Choice Electronic

More information

Introduction. Device Description. Dual-Band Spectral Output

Introduction. Device Description. Dual-Band Spectral Output Optimum Spectrum and Pulse Shape for Vascular Lesion Treatment: The Science Behind MaxG James Childs, Ph.D.; Andrei Erofeev, Ph.D.; Mikhail Smirnov, Ph.D.; and Gregory Altshuler, Ph.D., Sc.D. Introduction

More information

Pancreas: digestive enzymes Liver: bile-helps to digest fat, stored in gallblader; breaks down toxins, makes some parts of blood

Pancreas: digestive enzymes Liver: bile-helps to digest fat, stored in gallblader; breaks down toxins, makes some parts of blood 1 2 3 Pancreas: digestive enzymes Liver: bile-helps to digest fat, stored in gallblader; breaks down toxins, makes some parts of blood 4 5 6 7 8 9 10 Highlights: - A pulse consists of two phases: contraction

More information

DECODING BRAIN STATES USING FUNCTIONAL BRAIN IMAGING TECHNIQUES

DECODING BRAIN STATES USING FUNCTIONAL BRAIN IMAGING TECHNIQUES DECODING BRAIN STATES USING FUNCTIONAL BRAIN IMAGING TECHNIQUES BY MARIA PEIFER A thesis submitted to the Graduate School New Brunswick Rutgers, The State University of New Jersey in partial fulfillment

More information

Methods of Visualizing the Living Human Brain

Methods of Visualizing the Living Human Brain Methods of Visualizing the Living Human Brain! Contrast X-rays! Computerized Tomography (CT)! Magnetic Resonance Imaging (MRI)! Positron Emission Tomography (PET)! Functional MRI! Magnetoencephalography

More information

Advanced Monitoring of Cardiovascular and Respiratory Systems in Infants Kuwait 2018 Dr. Yasser Elsayed, MD, PhD Director of the Targeted Neonatal

Advanced Monitoring of Cardiovascular and Respiratory Systems in Infants Kuwait 2018 Dr. Yasser Elsayed, MD, PhD Director of the Targeted Neonatal Advanced Monitoring of Cardiovascular and Respiratory Systems in Infants Kuwait 2018 Dr. Yasser Elsayed, MD, PhD Director of the Targeted Neonatal Echocardiography, Point of Care and Hemodynamics Program

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

Effects of positive and negative mood induction on the prefrontal cortex activity measured by near infrared spectroscopy

Effects of positive and negative mood induction on the prefrontal cortex activity measured by near infrared spectroscopy Effects of positive and negative mood induction on the prefrontal cortex activity measured by near infrared spectroscopy A. Compare a, A. Brugnera a, R. Adorni a, and K. Sakatani b a Department of Human

More information

NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed.

NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed. Chapter 5 Monitoring the Emergency Patient NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed. Italicized text is quoted directly from Rosen s. Key Concepts: 1. Monitoring

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO 2 ) 3 The Vigileo

More information

Optical non-invasive methods for characterization of the human health status

Optical non-invasive methods for characterization of the human health status Optical non-invasive methods for characterization of the human health status Abstract Jens Kraitl and Hartmut Ewald Institute of General Electrical Engineering, University of Rostock, Rostock, Germany

More information

SUBJECT INFORMATION AND CONSENT FORM

SUBJECT INFORMATION AND CONSENT FORM Title of Study: Department of Anesthesiology British Columbia s Children Hospital 4480 Oak Street Vancouver V6H 3V4 Tel 604 875 2711 Fax 604 875 3221 SUBJECT INFORMATION AND CONSENT FORM Mobile Health:

More information

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination R. P. Buinevicius 1 J. G. Webster 2 1 PhD Student, Department of Electrical and Computer Engineering University of Wisconsin 1415 Engineering

More information

Anesthesia Monitoring. D. J. McMahon rev cewood

Anesthesia Monitoring. D. J. McMahon rev cewood Anesthesia Monitoring D. J. McMahon 150114 rev cewood 2018-01-19 Key Points Anesthesia Monitoring: - Understand the difference between guidelines & standards - ASA monitoring Standard I states that an

More information

Definition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill

Definition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill By Craig Definition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill pt Can assist diagnosis and decision making

More information