Home Phe Monitor Development Challenge -Advancing the Treatment for PKU- NPKUA SAB Mtg
|
|
- Austin Conrad Lawrence
- 5 years ago
- Views:
Transcription
1 Home Phe Monitor Development Challenge -Advancing the Treatment for PKU- NPKUA SAB Mtg
2 In October 2013 the NPKUA partnered with InnoCentive, Inc. to challenge the scientific and technical community around the world on the development of a Home Phe Monitor NPKUA & InnoCentive worked together to bring funding from The Cundill Foundation in Toronto to help support the project Together the NPKUA and Cundill Foundation have provided $776,000 to support the Home Phe Monitor Challenge Project NPKUA SAB Mtg
3 Phase I Theoretical Challenge opened to the world on how your technology would meet our specifications. 9 awards of $10K each were made to Solvers selected to move forward to Phase II. Phase II A Three Milestone Challenge: (1) to produced a more detailed Project Plan on how the Solver would demonstrate proof of concept for their technology 8 proposals & 5 awards for $10K each (2) provide test results from bench top prototype with phe spiked blood, 3 awards for $50K each & (3) demonstrate technology with whole blood sample from PKU patients 1 award up to $200K to produce a manufacturable & commercializable prototype. Phase III An erequest for Partners that would manufacture and commercialize the product. NPKUA SAB Mtg
4 The 3 Solvers awarded $50K for the Phase II milestone plus 1 other Solver who continued without the $50K award were invited to Emory for testing with real PKU subjects. Dr. Rani Singh and her team facilitated the testing by bringing in patients from their local metabolic clinic. Normal subjects and PKU subjects with low, mid, and high Phe levels were selected by Dr. Singh s team for venous draw to be used by Solvers and referenced by Plasma Amino Acid Analysis and DBS via Tandem Mass Spectrometry. NPKUA SAB Mtg
5 3 Solvers participated in the study: Conductive Technologies, Inc. (Drs. Musho & Szabo) With electrochemical device using sensor strips Dr. Elain Fu from Oregon State University With powerless paper device using microfluidics and enzymatic colorimetric technology Andrew Dennis, sensor researcher/grad. student With sensor strips and colorimetric device A 4 th Solver, BioTem (Dr. Clarence Deffaud), using lateral flow with custom-made antibody was unable to participate due to lack of progress on device development. NPKUA SAB Mtg
6 CTI phenylalanine meter with a test strip inserted. Meter displaying a 53µM PHE level NPKUA SAB Mtg
7 NPKUA SAB Mtg
8 methoxyphenazine methosulfate (mpms), is an electron mediator NPKUA SAB Mtg
9 NPKUA SAB Mtg
10 NPKUA SAB Mtg
11 Lancing device Disposable transfer pipette Running buffer Lateral flow Reader Lateral flow Test General protocol 1-Realizing the blood sample from the fingertip using the lancing device 4- Sample migration for 8 +/- 2 minutes 2-Take 20μL blood from the fingertip using the disposable transfert pipette 5-Read the results of the cassette using the portable Lateral flow reader 3- Add 2 drops of blood sample and running buffer into the lateral flow test NPKUA SAB Mtg
12 Enzyme (phedh & NAD+) instabilities Enzyme lot variations from vendors Calibration concerns due to above issues Hematocrit effect on measurements for some devices (CTI s and Dennis ) Need to rethink specifications for a home phe self monitor vs a true diagnostic device Consequences for FDA approval of a qualitative home self monitoring device vs a quantitative diagnostic device NPKUA SAB Mtg
13 The Final Report from the Emory testing by Dr. Singh s team recommended that none of the 3 devices completely meet the specifications for home monitoring although CTI s device was easiest to use (10-15 sec) and closest to a final product with sensor strip & device fabrications. Solvers were instructed to continue the development of their devices and a follow-up meeting of all Solvers would be held to see how they might collaborate with each other to help improve their devices in order to meet the desired specifications. NPKUA SAB Mtg
14 Solver recommendations going forward: Need further work by enzyme experts to solve the instability and lot variation issues Have renowned metabolic disease clinicians discuss with the FDA the need for a qualitative, self-monitoring device for home use by lay persons vs a quantitative, diagnostic device. Use the $200K to fund the enzyme stability study and to help raise $2-3M to support the product development work that will be necessary to finalize a device for manufacturing that would meet regulatory approval NPKUA SAB Mtg
15 Terence Dunne, Ph.D. at EKF Diagnostics in Dublin, Ireland using colorimetric technology Vamsee Pamula, Ph.D. at Baebies in RTP, NC using microfluidics and detecting phenylpyruvate working with Jerry Vockley at UPMC Rudolf Griss, Ph.D. at Lucentix in Switzerland using luminous technologies Peter Warthoe, Ph.D. at Atonomics in Denmark using trace devices with biomarker modules Marshall Summar, M.D. & Omar Ayyub, Ph.D. at Washington s Children using spectrometric or electrochemical techniques with their own cloned, expressed, and purified GtPheDH from a thermopile bacteria NPKUA SAB Mtg
16 The NPKUA will continue communications with Solvers in the Challenge and those outside the Challenge. Decisions on how to disperse the remaining Challenge funds will be made by the NPKUA Board of Directors in a manner that will hopefully and eventually lead to a successful Home Phe Monitor that meets our specifications.
17 Thomas D. Franklin, Jr., Ph.D. NPKUA Scientific Advisory Board NPKUA SAB Mtg
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationI know my value CoaguChek XS Pro system
I know my value CoaguChek XS Pro system The smart way to manage your anticoagulation clinic CoaguChek XS Pro system A barcode scanner can make a difference There has been an extraordinary increase in the
More informationSP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World
MIT OpenCourseWare http://ocw.mit.edu SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World Spring 2009 For information about citing these materials or our
More informationOGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring
More informationA rapid test for the qualitative detection of Malaria pf and pv antigen in human blood sample
A rapid test for the qualitative detection of Malaria pf and pv antigen in human blood sample For in vitro use only Intended Use For the rapid qualitative determination of Malaria P. falciparum specific
More informationWHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT. Product: Uni-Gold HIV Number: PQDx Abstract
WHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT Product: Uni-Gold HIV Number: PQDx 0149-052-00 Abstract Uni-Gold HIV with product codes 1206502, 1206502N, 1206502E, 1206502N-100 manufactured
More informationImproving the Sensitivity in a Paper-Based Test for the Home Monitoring of Phenylalanine Levels in Phenylketonuria Patients. Liam Wong A THESIS
Improving the Sensitivity in a Paper-Based Test for the Home Monitoring of Phenylalanine Levels in Phenylketonuria Patients By Liam Wong A THESIS submitted to Oregon State University Honors College in
More informationPoint of Care Testing for INR using CoaguChek XS Plus
Point of Care Testing, Pathology Page 1 of 18 Point of Care Testing for INR using CoaguChek XS Plus EDITION No 1.4 DATE OF ISSUE Feb 2014 REVIEW INTERVAL AUTHOR LOCATION OF COPIES 3 YEARS D O Neill 1.
More information2015 PKU Patient Survey Results. National PKU Alliance PO Box 501, Tomahawk, WI
2015 PKU Patient Survey Results National PKU Alliance PO Box 501, Tomahawk, WI 54487 www.npkua.org www.adultswithpku.org The NPKUA Patient Survey demonstrates that strict dietary therapy and treatment
More informationUltra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC
VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values
More informationWHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT. Product: Uni-Gold HIV Number: PQDx Abstract
WHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT Product: Uni-Gold HIV Number: PQDx 0149-052-00 Abstract Uni-Gold HIV with product codes 1206502, 1206502N, 1206502E, 1206502N-100 manufactured
More informationManual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay
TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: DPP HIV 1/2 Assay WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: DPP HIV 1/2 Assay WHO reference number: PQDx 0053-006-00 DPP HIV 1/2 Assay with product code 65-9506-0, manufactured by Chembio Diagnostic
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationProcedure for Blood Analysis. 1.0 Purpose - This procedure specifies the methods for performing blood analysis in forensic casework.
Procedure for Blood Analysis 1.0 Purpose - This procedure specifies the methods for performing blood analysis in forensic casework. 2.0 Scope - This procedure applies to those Forensic Scientists who have
More informationMonitoring of Blood Glucose Level
University of Zagreb Faculty of Electrical Engineering and Computing Biomedical instrumentation Monitoring of Blood Glucose Level Biomedical instrumentation 1/17 Outline Introduction Principles of Blood
More informationMission Hb accurately detects both
Mission Digital Hb-Testing System: Save Cost by testing 2- Parameters with Single Strip real time ie: i.e: Hemoglobin & Hematocrit-[PCV]. Presented By : Remedy Healthcare Private Limited, New Delhi, [INDIA].
More informationCDIA TM Rubella IgG/IgM Rapid Test Kit
CDIA TM Rubella IgG/IgM Rapid Test Kit Cat.No: DTSJZ024 Lot. No. (See product label) Intended Use The CDIA TM Rubella IgG/IgM Rapid Test Kit is a rapid chromatographic immunoassay for the qualitative detection
More informationBlood Glucose Monitoring System. Owner s Booklet
Blood Glucose Monitoring System Owner s Booklet The FreeStyle Freedom Lite Meter and How It Works System Check Screen This display always appears when the meter is turned on. You should check that your
More informationToxicology Blood Lead Comprehensive Testing
Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationComparison of Methods for the Analysis of Lysosomal Enzyme Activities in Quality Control Dried Blood Spot Specimens
Comparison of Methods for the Analysis of Lysosomal Enzyme Activities in Quality Control Dried Blood Spot Specimens Atlanta, GA, 2013 David Millington 1, Qun Shi 1, Gwen Dickerson 1, Hui Zhou 2, Victor
More informationAmerican Association for Clinical Chemistry Emerging Clinical & Laboratory Diagnostics: The Portable Lab April 24-25, 2014
Barry Lutz University of Washington Department of Bioengineering Key collaborators: Paul Yager, UW BioE Elain Fu, Oregon State University American Association for Clinical Chemistry Emerging Clinical &
More informationPepsin Microplate Assay Kit User Manual
Pepsin Microplate Assay Kit User Manual Catalog # CAK1081 Detection and Quantification of Pepsin Activity in Urine, Serum, Plasma, Tissue extracts, Cell lysate, Cell culture media and Other biological
More informationNeovasc Inc. Alexei Marko, CEO. January 2012 OneMedForum
Neovasc Inc. Alexei Marko, CEO January 2012 OneMedForum 2 Forward-Looking Statements Statements contained herein that are not based on historical or current fact, including without limitation statements
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationCoverage and Coding of Continuous Glucose Monitors
Coverage and Coding of Continuous Glucose Monitors Length: 15:43 Date Recorded: 3.15.17 Hello and welcome to another edition of Medicare Minute MD. I m Dr. Robert Hoover, medical director at CGS Administrators,
More informationToxicology Blood Lead Comprehensive Testing
Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where applicable to the scope of services
More informationRayBio Phenylalanine Fluorometric Assay Kit
RayBio Phenylalanine Fluorometric Assay Kit User Manual Version 1.0 June 25, 2014 RayBio Phenylalanine Fluorometric (Cat#: 68-Phenyl-S100) RayBiotech, Inc. We Provide You With Excellent Support And Service
More informationClinical Evaluation for. Embrace No Code Blood Glucose Monitoring System
Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic
More informationNCCLS. Accutest H. pylori WB Test. (Catalog. No. ID-92A225) For Professional and In Vitro Diagnostic Use Only
NCCLS Accutest H. pylori WB Test (Catalog. No. ID-92A225) For Professional and In Vitro Diagnostic Use Only A Rapid Visual Test for the Qualitative Detection of IgG Antibodies Specific to Helicobacter
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationPrerequisites Amino acid synthesis and degradation pathways. Integration of amino acid metabolic pathways with carbohydrate metabolic pathways.
Case 30 Phenylketonuria Focus concept The characteristics of phenylalanine hydroxylase, the enzyme missing in persons afflicted with the genetic disorder phenylketonuria (PKU), are examined. Prerequisites
More informationH.Pylori IgG
DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com
More informationH.Pylori IgG Cat # 1503Z
DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external
More informationRat Hemoglobin A1c (HbA1c) Kit Instructions
V.3 Crystal Chem Rat Hemoglobin A1c (HbA1c) Kit Instructions For the quantitative determination of hemoglobin A1c (HbA1c) in rat whole blood Catalog #80300 96 Assays For research use only. Not for use
More informationMalaria (Pan-LDH) W/B
CORTEZ DIAGNOSTICS, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 USA Tel: (818) 591-3030 Fax: (818) 591-8383 E-mail: onestep@rapidtest.com Web site: www.rapidtest.com See external label
More informationWHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT. Product: SURE CHECK HIV 1/2 Assay Number: PQDx
WHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT Product: SURE CHECK HIV 1/2 Assay Number: PQDx 0054-006-00 Abstract SURE CHECK HIV 1/2 Assay with product code HIV201, manufactured
More informationPhenylalanine. in cases of hyperphenylalaninemia and tetrahydrobiopterin deficiency
Evolving Methods for the Measurement of Phenylalanine A vital diagnostic marker and indicator for follow-up in cases of hyperphenylalaninemia and tetrahydrobiopterin deficiency Dr. Zoltan Lukacs Department
More informationGM700 USER S MANUAL. Blood Glucose Monitoring System GM700
BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan Tel: +886 4 23692388 Fax: +886 4 22617586 http: //www.bionime.com E-mail: info@bionime.com Rev. Date: February/2017
More informationusing the Agilent 7696A Sample Prep
Automated Clean-up for Mineral Oil (Hydrocarbon Oil Index) Analysis using the Agilent 7696A Sample Prep WorkBench Application Note Automated Sample Preparation Authors Frank David, Karine Jacq, and Bart
More informationOmnitest : 2013
ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample
More informationGM700S USER S MANUAL. Blood Glucose Monitoring System
GM700S BIONIME GmbH Tramstrasse 16, 9442 Berneck Switzerland E-mail: info@bionime.ch 0197 101-3GM780-011 EN Rev. Date: September/2014 BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung
More informationToxicology Blood Lead Comprehensive Testing
Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationPKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual
Page 1 / 20 PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual ZV-4003-0200-10 200 2-8 C Page 2 / 20 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3
More informationSummary of Significant Changes at this Revision. Items Required
Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of
More informationab HRV 3C Protease Inhibitor Screening Kit (Colorimetric)
Version 2 Last updated 2 March 2017 ab211089 HRV 3C Protease Inhibitor Screening Kit (Colorimetric) For the rapid, sensitive and accurate screening of potential HRV 3C Protease inhibitors. This product
More informationPerformance of three portable blood glucose monitoring devices used in a veterinary application
Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,
More informationH. pylori IgM. Cat # H. pylori IgM ELISA. ELISA: Enzyme Linked Immunosorbent Assay. ELISA - Indirect; Antigen Coated Plate
DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com H. pylori
More informationWorksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide
LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON
More informationGM300 USER S MANUAL. Blood Glucose Monitoring System
101-3GM300-704 EN BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan Tel: +886 4 23692388 Fax: +886 4 22617586 http: //www.bionime.com E-mail: info@bionime.com Rev.
More informationQUICK REFERENCE INSTRUCTIONS. THYROCHEK TSH Cassette
QUICK REFERENCE INSTRUCTIONS THYROCHEK TSH Cassette A certificate of CLIA waiver is required to perform the testing in a waived setting. If the laboratory does not have a Certificate of Waiver, the Application
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More informationBeneCheck BK6-12M. Plus Multi-Monitoring Meter and Strips
BeneCheck BK6-12M Plus Multi-Monitoring Meter and Strips The BeneCheck BK6-12M multi-monitoring system is an easy to use, handheld device which allows you to check your Total Cholesterol, as well as Blood
More informationPKU (Phenylketonuria) Serum LC-MS/MS Analysis Kit User Manual
Page 1 / 24 PKU (Phenylketonuria) Serum LC-MS/MS Analysis Kit User Manual ZV-3003-0200-10 200 2-8 C Page 2 / 24 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE...
More informationGM260 USER S MANUAL. Blood Glucose Monitoring System
Blood Glucose Monitoring System BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan Tel: +886 4 23692388 Fax: +886 4 22617586 http: //www.bionime.com E-mail: info@bionime.com
More informationC. Bachmann This article was published as. Centre Hospitalier Universitaire Vaudois Number 3 in May 2000 CH 1011 Lausanne, Switzerland
The control of analytical accuracy and day to day precision helps for the follow-up of patients and is essential when using biochemical data from the literature. C. Bachmann This article was published
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationWHO Prequalification of Diagnostics Programme PUBLIC REPORT. Product: HIV 1/2 STAT-PAK Number: PQDx Abstract
WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: HIV 1/2 STAT-PAK Number: PQDx 0007-006-00 Abstract The HIV 1/2 STAT-PAK with product code HIV101, manufactured by Chembio Diagnostic
More informationQiG INS Conference May 22, 2011
Strategy & Vision Growth Electrochem Medical Device Evolution QiG Cardio vascular QiG INS Conference May 22, 2011 Greatbatch Medical Neuro modulation 1 GREATBATCH TODAY 2010 Revenue by Product Line NYSE:
More informationItems in the package:
Intended Use: The EasyLife Hb Monitoring System is designed for in vitro diagnostic use only (external use only), and is suitable for self-testing. The system is for healthcare professionals and persons
More informationH. pylori Antigen ELISA Kit
H. pylori Antigen ELISA Kit Catalog Number KA3142 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of
More informationGlutamine Kit. For the determination of glutamine and glutamate in human EDTA plasma and serum. For research use only K 7732.
Li StarFish S.r.l. Via Cavour, 35-20063 Cernusco S/N (MI), Italy Tel. +39-02-92150794 - Fax. +39-02-92157285 info@listarfish.it -www.listarfish.it Manual Kit For the determination of glutamine and glutamate
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationBern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.
Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,
More informationManage Your Pet s Diabetes at Home
Manage Your Pet s Diabetes at Home Managing your pet s diabetes is easy, accurate and convenient with at-home blood glucose testing with AlphaTRAK 2. If your dog or cat has been diagnosed with diabetes,
More informationEVALUATION OF GLUCOSE MONITORING TECHNOLOGIES FOR COST EFFECTIVE AND QUALITY CONTROL/MANAGEMENT OF DIABETES
EVALUATION OF GLUCOSE MONITORING TECHNOLOGIES FOR COST EFFECTIVE AND QUALITY CONTROL/MANAGEMENT OF DIABETES David P. Paul, III Monmouth University Stacy Ashworth, Leslie Salyers, Sarah Saldanha and Alberto
More informationGM550 USER S MANUAL. Blood Glucose Monitoring System
Blood Glucose Monitoring System BIONIME GmbH Tramstrasse 16, 9442 Berneck Switzerland E-mail: info@bionime.ch 101-3GM550-234 EN BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung City
More information11: Pathways. Reminders:
11: Pathways Reminders: 1. Due in lab NEXT week: a. Pre- Lab 05 b. Amylase lab report 2. Exam 4 Monday (see SPOC for details) Last names A F Ryan Lounge M- 3-721 Last names G - Z here (1 pt for right room)
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationNIH Public Access Author Manuscript Mol Genet Metab. Author manuscript; available in PMC 2014 June 01.
NIH Public Access Author Manuscript Published in final edited form as: Mol Genet Metab. 2013 June ; 109(2): 218 220. doi:10.1016/j.ymgme.2013.03.010. Rapid assays for Gaucher and Hurler diseases in dried
More informationAmal Alamri. Phenylketonuria
Amal Alamri Phenylketonuria Norwegian doctor Asbjørn Følling (left) discovered phenylpyrouvica (later termed phenylketonuria) in 1934 upon discovering that ten mentally retarded patients had phenylpyruvic
More informationHerpes Simplex Virus 2 IgG HSV 2 IgG
DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com
More informationH.pylori IgA Cat #
DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external
More informationWHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT. Product: First Response HIV Card Test Number: PQDx
WHO Prequalification of In Vitro Diagnostics Programme PUBLIC REPORT Product: First Response HIV 1-2-0 Card Test Number: PQDx 0018-010-00 Abstract First Response HIV 1-2-0 Card Test with product codes
More informationVitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid
Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid Application Note Clinical Research Authors Derick Lucas and Limian Zhao Agilent Technologies, Inc. Abstract Lipids from
More informationH.Pylori IgM Cat # 1504Z
DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external
More informationNew Developments and Regulatory Issues SLEEP AND CIRCADIAN RESEARCH. October 11-12, 2018 AGENDA
AGENDA SLEEP AND CIRCADIAN RESEARCH New Developments and Regulatory Issues October 11-12, 2018 Hilton Alexandria Mark Hotel 5000 Seminary Road, Alexandria, VA 22311 The Sleep Research Society would like
More informationPectinase Microplate Assay Kit User Manual
Pectinase Microplate Assay Kit User Manual Catalog # CAK1087 Detection and Quantification of Pectinase Activity in Tissue extracts, Cell lysate, Cell culture media and Other biological fluids Samples.
More informationEXOTESTTM. ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids
DATA SHEET EXOTESTTM ELISA assay for exosome capture, quantification and characterization from cell culture supernatants and biological fluids INTRODUCTION Exosomes are small endosome-derived lipid nanoparticles
More informationMessage from the Toronto HIV/AIDS Community Planning Initiative Co-Champions
Toronto HIV/AIDS Community Planning Initiative (TCPI) Final Report March 2007 Message from the Toronto HIV/AIDS Community Planning Initiative Co-Champions As Co-Champions of the Toronto HIV/AIDS Community
More informationManagement of Central Venous Access Devices. Blood Glucose Monitoring
Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including
More informationD-Mannitol Assay Kit (Colorimetric)
D-Mannitol Assay Kit (Colorimetric) Catalog Number KA3760 100 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information...
More informationMitochondrial Trifunctional Protein (TFP) Protein Quantity Microplate Assay Kit
PROTOCOL Mitochondrial Trifunctional Protein (TFP) Protein Quantity Microplate Assay Kit DESCRIPTION Mitochondrial Trifunctional Protein (TFP) Protein Quantity Microplate Assay Kit Sufficient materials
More informationPhenylalanine Assay Kit
ab83376 Phenylalanine Assay Kit Instructions for Use For the rapid, sensitive and accurate measurement of Phenylalanine levels in various samples This product is for research use only and is not intended
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k063276 B. Purpose for Submission: New device C. Measurand: Urinary Glucose, Blood,
More informationProcedure for Blood Analysis. 1.0 Purpose - This procedure specifies the methods for performing blood analysis in forensic casework.
Procedure for Blood Analysis 1.0 Purpose - This procedure specifies the methods for performing blood analysis in forensic casework. 2.0 Scope - This procedure applies to those Forensic Scientists who have
More informationUpdate on CD4+ T-Cell Counting Technologies
Update on CD4+ T-Cell Counting Technologies Thomas Denny, MSc, M.Phil Professor of Medicine Duke Human Vaccine Institute Duke University Medical School UK NEQAS 24 th June 2013, Sheffield 2 3 4 5 6 Levels
More informationDescription of the technology
Advice Note 2017/ 001 What is the clinical effectiveness, safety and budget impact of the Freestyle Libre System compared with current glucose monitoring methods for people aged 4 years and over with diabetes
More informationGlucCell TM SYSTEM USER S GUIDE Ver 2.1 CELL CULTURE GLUCOSE METER. Important Information. Intended Use. Caution. About the System
GlucCell TM SYSTEM USER S GUIDE Ver 2.1 Intended Use The GlucCell TM Glucose Monitoring System (The GlucCell TM System) is designed to quantitatively measure the concentration of glucose during cell culture.
More informationCancer Awareness & Early Diagnosis Project Examples. Location: Camden (intervention area) and Kensington & Chelsea (control area), London
PROJECT TITLE: Improving breast awareness in women aged 45-54 Location: Camden (intervention area) and Kensington & Chelsea (control area), London PROJECT DETAILS Problem addressed: Breast cancer is now
More informationClinical Chemistry and Clinical Toxicology Devices Panel of the Medical Devices Advisory
This document is scheduled to be published in the Federal Register on 03/01/2018 and available online at https://federalregister.gov/d/2018-04167, and on FDsys.gov 4164-01-P DEPARTMENT OF HEALTH AND HUMAN
More informationRotavirus Test Kit. Instructions For Use. Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA
Rotavirus Test Kit Instructions For Use Format: Cassette Specimen: Fecal Extract Catalog Number: VEL-001-ROTA * Please read the instructions carefully before use INTENDED USE Velotest Rotavirus Test is
More informationLabStyle Innovations Corp. (OTCQB:DRIO)
LabStyle Innovations Corp. (OTCQB:DRIO) End to End Smart Solutions for Diabetes in a Mobile Age Disclaimer Regarding Forward-Looking Statements This presentation of LabStyle Innovations Corp. (the Company
More informationGood to know. Tips and tricks for accurate blood glucose monitoring
Good to know Tips and tricks for accurate blood glucose monitoring Tip 1 Washing your hands No place for dirt Accurate blood glucose readings are essential for your diabetes care. Correct self-testing
More informationEVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS
EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The
More informationDiamine Oxidase Microplate Assay Kit User Manual. Detection and Quantification of Diamine Oxidase (DAO) Activity in
Diamine Oxidase Microplate Assay Kit User Manual Detection and Quantification of Diamine Oxidase (DAO) Activity in Urine, Serum, Plasma, Other biological fluids, Tissue extracts, Cell lysate, Cell culture
More information