System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

Size: px
Start display at page:

Download "System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013"

Transcription

1 System Accuracy Evaluation of Project number: Fora System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th October th October /21

2 System Accuracy Evaluation of Table of contents Abstract... 3 Responsible Parties & Signatures Introduction Material and Methods Recruitment of test subjects Blood glucose monitoring system (BGMS) Reference standard Preparation of test samples Test of blood glucose Statistics Results Ratio of test subjects Test results Accuracy results following ISO 15197:2013 (E) Accuracy results following ISO 15197: Statistical analysis Absolute differences between and YSI 2300 STAT PLUS Conclusion and Discussion References Appendix /21

3 System Accuracy Evaluation of Abstract Objective: The goal of the study was to assess the accuracy and reliability of Fora blood glucose monitoring system (BGMS) in the clinical studies according to the international standard ISO 15197: 2013(E) criteria and more stringent guidelines. Methods: The study sample was collected from 108 subjects comprised 53 patients with types 1 diabetes and 36 patients with types 2 diabetes, and 19 people with type unknown. The capillary blood specimens was collected via fingertip stick from participants at AMCR Institute (700 West El Norte Parkway, Suite 201, Escondido, CA 92026) and tested on three different lots of test strips. Results: The Fora BGMS met the requirements of ISO 15197:2013 from all three lots of test strips tested. Reagent system lot Within ± 5 mg/dl & ± 5 % (Within ± 0.28 mmol/l & ± 5 %) Within ± 10 mg/dl & ± 10 % (Within ± 0.56 mmol/l & ± 10 %) Within ± 15 mg/dl & ± 15 % (Within ± 0.83 mmol/l & ± 15 %) TD14D228-C0E 144 / 200 (72.0 %) 192 / 200 (96.0 %) 197 / 200 (98.5 %) TD14D828-C0E 142 / 200 (71.0 %) 195 / 200 (97.5 %) 197 / 200 (98.5 %) TD14D928-C0E 139 / 200 (69.5 %) 192 / 200 (96.0 %) 198 / 200 (99.0 %) This BGMS is also compliant with the requirements of DIN EN ISO 15197:2003 with all 3 reagent system lot(s). Reagent system lot Within ± 5 mg/dl & ± 5 % (Within ± 0.28 mmol/l & ± 5 %) Within ± 10 mg/dl & ± 10 % (Within ± 0.56 mmol/l & ± 10 %) Within ± 10 mg/dl & ± 15 % (Within ± 0.56 mmol/l & ± 15 %) Within ± 15 mg/dl & ± 20 % (Within ± 0.83 mmol/l & ± 20 %) TD14D228-C0E 117 / 200 (58.5 %) 178 / 200 (89.0 %) 196 / 200 (98.0 %) 199 / 200 (99.5 %) TD14D828-C0E 107 / 200 (53.5 %) 169 / 200 (84.5 %) 197 / 200 (98.5 %) 199 / 200 (99.5 %) TD14D928-C0E 112 / 200 (56.0 %) 171 / 200 (85.5 %) 194 / 200 (97.0 %) 200 / 200 (100 %) Conclusion: The study suggest that the Fora blood glucose monitoring system not only met the current accuracy standard (ISO15197:2003), but also met with the minimum acceptable accuracy standard published in the newer 2013 guideline (ISO15197:2013) when compared the test results with a laboratory reference. 3/21

4

5 System Accuracy Evaluation of 1. Introduction Diabetes is a chronic illness that requires continuing medical care and ongoing patient self-management education and support to prevent acute complications and to reduce the risk of long-term complications. It is predicted that the total number of people with diabetes will rise from 171 million in 2000 to 366 million in Clinical evidence has clearly indicated that improved metabolic control leads to a significant reduction in microvascular complications associated with diabetes. Self-monitoring blood glucose system (SMBG) has become an important and efficacy tool in the daily blood glucose management for people with diabetes. Optimal utilization of SMBG can lead to improvements in glycemic control by providing patients with immediate feedback on the impact of food, exercise, or medication on their blood sugar. Therefore, simple-to-use and accurate performance will be the conditions needed in a good glucose meter. The Fora Blood Glucose monitoring system for this study provides a faster and more effective way of blood glucose measurement. Furthermore, these glucose meters employ good technology which can eliminate miscoding errors that could lead to inaccurate results. Features of self-monitoring blood glucose system (SMBG) include ease-of-use, capillary action, electrodes with carbon, and biosensor technology. This study was to evaluate the system accuracy following ISO 15197: protocol and if possible, the newly published ISO 19197: guideline. EN ISO 15197:2003 is an international standard published by International Standard Origination (ISO) specifying requirements for in vitro glucose monitoring systems on self-testing in managing diabetes mellitus. The minimum acceptable accuracy for results produced by a glucose monitoring system shall be as follows: Ninety-five percent (95 %) of the individual glucose results shall fall within ± 15 mg/dl (± 0.83 mmol/l) of the results of the reference measurement at glucose concentrations < 75 mg/dl (< 4.2 mmol/l) and within ± 20 % at glucose concentrations 75 mg/dl ( 4.2 mmol/l). In 2013, ISO has published a newer standard ISO 15197:2013 (E) which tightens the accuracy acceptance criteria. The minimum acceptable accuracy for results produced by a glucose monitoring system has been changed to the followings: Ninety-five percent (95 %) of the individual glucose results shall fall within ± 15 mg/dl (± 0.83 mmol/l) of the results of the reference measurement at glucose concentrations < 100 mg/dl (< 5.55 mmol/l) and within ± 15 % at glucose concentrations 100 mg/dl ( 5.55 mmol/l). Comparison between 2 ISO standards: EN ISO 15197: 2003 ISO 15197: 2013 (E) Maximum deviation from glucose reference value At glucose levels < 75 mg/dl (< 4.2 mmol/l) At glucose levels 75 mg/dl ( 4.2 mmol/l) ± 15mg/dl (0.83 mmol/l) ± 20% At glucose levels < 100 mg/dl (< 5.5 mmol/l) At glucose levels 100 mg/dl ( 4.2 mmol/l) ± 15mg/dl (0.83 mmol/l) ± 15% 5/21

6 System Accuracy Evaluation of The Clarke error grid (CEG) analysis 4 was part of ISO guideline to quantify the clinical accuracy of blood glucose readings generated by meters as compared to a reference value. According to ISO guideline, the error grids subdivided into five zones. Values in zones A and B represent accurate or acceptable glucose results which have no effect on clinical action. If values fall into in zone C, it may prompt unnecessary corrections that could impact on clinical outcome. Values in zone D could lead to significant medical risk for patients and values in zone E could have dangerous consequence resulted from an inappropriate treatment. To help the treatment for type I diabetes patients, ISO guideline requires 99% of individual glucose measured values shall fall within zones A and B of the CEG. The objective for this study was to access the accuracy of Fora based on ISO protocol and if possible, the new ISO 2013 guideline. 2. Material and Methods 2.1 Recruitment of test subjects: Test subjects were recruited to AMCR Institute based on the inclusion and exclusion criteria set from the study protocol. Inclusion: Male or female subjects with diabetes type 1 or 2 or healthy subjects, potentially including the employees of the sponsor. Minimum age of 18 years. Exclusion: Pregnancy or lactation period Severe acute disease Severe chronic disease with potential risk during the trial Discrepancy between 2 YSI references. According to BS EN ISO 15197: 2013(E), results from the reference method were evaluated to verify sample stability. If the values between first and last reference differ by >4% at glucose 100mg/dl ( 5.55mmol/l) or >4mg/dl (>0.22mmol/l) at glucose <100mg/dl (<5.55mmol/l), indicating glucose instability in the sample, the results for that sample had to be rejected. 2.2 Blood glucose monitoring system (BGMS) Fora glucose monitoring systems are market-ready glucose meters with FDA approval # K Each comes with complete packaging, manual and quick user guide. Twelve meters were used in this study. They were divided into 2 stations with six at each station. Meter: Serial number Station A01-A A07-A12 2 6/21

7 System Accuracy Evaluation of Key features of Fora monitoring system: Blood sample: Capillary and venous blood Sample volume: 0.5 l Measuring range: mg/dl ( mmol/l) Analysis time: 5 seconds Operating temperature: Operating humidity: < 85% R.H. Hematocrit range: % Measurement technology: Glucose dehydrogenase (GDH-FAD) Calibration: Plasma The evaluation was performed with 3 lots of test strips. Reagent system: Blood glucose test strip: Lot Lot number Expiry 1 TD14D228-C0E TD14D828-C0E TD14D928-C0E Control solutions were used to verify the performance and function of the meters each morning during the study. Control solution Lot number Expiry Control Solution Level 2 WA13B Target range Lot 1: 138~186 Lot 2: 139~187 Lot 3: 137~ Reference standard The reference used in this study was a Yellow Springs 2300 (YSI 2300) manufactured by YSI incorporated (Yellow Springs, OH) which detected glucose concentration based on a glucose oxidase method The readings of the YSI analyzer was validated by running the analyzer with the National Institute of Standards and Technology (NIST) secondary reference material SRM 965a which consists of four levels of standard glucose concentrations. Glucose reference was determined as the average of at least duplicated measurements from the YSI. 7/21

8 System Accuracy Evaluation of 2.4 Preparation of Samples Capillary blood samples were collected based on the ISO 15197:2013 guideline to have certain percentage of subject samples in the following concentration categories: Concentration category Percentage of samples [%] Glucose concentration [mg/dl (mmol/l)] Number of test subjects based on ( 2.77) > (> ) > (> ) > (> ) > (> ) > (> ) > 400 (> 22.20) 5 Samples were assigned to the respective glucose concentration category depending on the mean from the YSI measurements. If the total number of test subjects on each concentration category was filled, no additional samples would be added. According to the ISO guideline, the following samples were adjusted by either natural decayed at 37ºC or spiked with additional glucose to fall into specific concentration categories. Concentration category Sample no. 1 85, 96, 97, , 86, 95, 99, 100, 101, 102, 103, 105, 106, , 93, Test of Blood Glucose Sample capillary blood was obtained by a finger stick using a 25G safety lancet. The blood sample was immediately collected onto a Lithium-Heparin Microvette (~ 200 μl) The capillary blood was then placed on a parafilm and tested on Fora blood glucose monitoring systems (6 devices per station with test strip lot 1, 2, and 3). The remaining blood in the lithium-heparin microvette was then immediately placed on a centrifuge to separate plasma. The glucose concentration of the plasma was then determined by the YSI 2300 using one electrode. The measurement was repeated once and the true glucose concentration of the subject glucose was determined by the average of the YSI values. 8/21

9 System Accuracy Evaluation of 2.6 Statistical analysis In this study, the following statistical analyses were performed: Bias analysis according to Bland and Altman 5. Regression analysis according to Passing and Bablok Results 3.1 Test subjects Total of 108 test subjects were enrolled in the study. Among them, 8 test subjects were excluded for the following reasons. Reagent system lot total test subjects Number of subjects rejected Based on ISO 15197: 2013 (E) criteria The change between first (before the measurement procedure) and second (after the measurement procedure) comparison measurement result is > 4 % at glucose concentrations > 100 mg/dl or > 4 mg/dl at glucose concentrations 100 mg/dl (acceptance criterion for sample stability) The concentration category was already filled Additional exclusion criteria based on Fora protocol HCT is high than 50% or below 35% /21

10 System Accuracy Evaluation of The detail test log Reagent system lot Number of included samples Glucose concentration range (YSI 2300 STAT Plus TM glucose analyzer) minimum maximum Hematocrit range [%] Age [years] Type 1 / Type 2 / NA 53 / 36 / / 36 / / 36 / 19 Gender [male / female / unknown] 48 / 50 / / 50 / / 50 / 10 Temperature [ ] (during the test) Humidity [%] (during the test) Test results Accuracy results following ISO 15197:2013 (E) Summary: Glucose concentration range: 37 mg/dl (2.06 mmol/l) to mg/dl (26.4 mmol/l) Test strip lot 1: Lot: TD14D228-C0E Within ± 15 mg/dl & ± 15 % (Within ± 0.83 mmol/l & ± 15 %) 198 / 200 (99.0 %) Test strip lot 2: Lot: TD14D828-C0E Within ± 15 mg/dl & ± 15 % (Within ± 0.83 mmol/l & ± 15 %) 199 / 200 (99.5 %) 10/21

11 System Accuracy Evaluation of Test strip lot 3: Lot: TD14D928-C0E Within ± 15 mg/dl & ± 15 % (Within ± 0.83 mmol/l & ± 15 %) 194 / 200 (97.0 %) All three lots: Within Consensus Error Grid zones A and B 600 / 600 (100 %) Conclusion: The Fora BGMS met the requirements of new ISO standard (ISO 15197:2013). Detail results: System accuracy results for glucose concentrations < 100 mg/dl (5.55 mmol/l) Reagent system lot Within ± 5 mg/dl (Within ± 0.28 mmol/l) Within ± 10 mg/dl (Within ± 0.56 mmol/l) Within ± 15 mg/dl (Within ± 0.83 mmol/l) TD14D228-C0E 47 / 64 (73.4 %) 61 / 64 (95.3 %) 64 / 64 (100 %) TD14D828-C0E 50 / 64 (78.1 %) 61 / 64 (95.3 %) 63 / 64 (98.4 %) TD14D928-C0E 49 / 64 (76.6 %) 64 / 64 (100 %) 64 / 64 (100 %) System accuracy results for glucose concentrations 100 mg/dl (5.55 mmol/l) Reagent system lot Within ± 5 % Within ± 10 % Within ± 15 % TD14D228-C0E 72 / 136 (52.9 %) 118 / 136 (86.8 %) 134 / 136 (98.5 %) TD14D828-C0E 59 / 136 (43.4 %) 109 / 136 (80.2 %) 136 / 136 (100 %) TD14D928-C0E 64 / 136 (47.1 %) 108 / 136 (79.4 %) 130 / 136 (95.6 %) System accuracy results for glucose concentrations between 37 mg/dl (2.06 mmol/l) and mg/dl (26.4 mmol/l) Reagent system lot Within ± 5 mg/dl & ± 5 % (Within ± 0.28 mmol/l & ± 5 %) Within ± 10 mg/dl & ± 10 % (Within ± 0.56 mmol/l & ± 10 %) Within ± 15 mg/dl & ± 15 % (Within ± 0.83 mmol/l & ± 15 %) TD14D228-C0E 144 / 200 (72.0 %) 192 / 200 (96.0 %) 197 / 200 (98.5 %) TD14D828-C0E 142 / 200 (71.0 %) 195 / 200 (97.5 %) 197 / 200 (98.5 %) TD14D928-C0E 139 / 200 (69.5 %) 192 / 200 (96.0 %) 194 / 200 (97.0 %) 11/21

12 System Accuracy Evaluation of Accuracy results following ISO 15197:2003 Summary: Glucose concentration range: 37 mg/dl (2.06 mmol/l) to mg/dl (26.4 mmol/l) Test strip lot 1: Lot: TD14D228-C0E Within ± 15 mg/dl & ± 20 % (Within ± 0.83 mmol/l & ± 20 %) 199 / 200 (99.5 %) Test strip lot 2: Lot: TD14D828-C0E Within ± 15 mg/dl & ± 20 % (Within ± 0.83 mmol/l & ± 20 %) 199 / 200 (99.5 %) Test strip lot 3: Lot: TD14D928-C0E Within ± 15 mg/dl & ± 20 % (Within ± 0.83 mmol/l & ± 20 %) 200 / 200 (100.0 %) All three lots: Within Consensus Error Grid zones A and B 600 / 600 (100 %) Conclusion: Fora BGMS also is in compliance with the requirements of DIN EN ISO 15197:2003 for all 3 reagent system lot(s). 12/21

13 System Accuracy Evaluation of Detail results: System accuracy results for glucose concentrations < 75 mg/dl (4.2 mmol/l) Reagent system lot Within ± 5 mg/dl (Within ± 0.28 mmol/l) Within ± 10 mg/dl (Within ± 0.56 mmol/l) Within ± 15 mg/dl (Within ± 0.83 mmol/l) TD14D228-C0E 29 / 38 (76.3 %) 36 / 38 (94.7 %) 38 / 38 (100 %) TD14D828-C0E 32 / 38 (84.2 %) 35 / 38 (92.1 %) 37 / 38 (97.4 %) TD14D928-C0E 31 / 38 (81.6 %) 38 / 38 (100 %) 38 / 38 (100 %) System accuracy results for glucose concentrations 75 mg/dl (4.2 mmol/l) Reagent system lot Within ± 5 % Within ± 10 % Within ± 15 % Within ± 20 % TD14D228-C0E 88 / 162 (54.3 %) 142 / 162 (87.7 %) 160 / 162 (98.8 %) 161 / 162 (99.4 %) TD14D828-C0E 75 / 162 (46.3 %) 134 / 162 (82.7 %) 162 / 162 (100 %) 162 / 162 (100 %) TD14D928-C0E 81 / 162 (50.0 %) 133 / 162 (82.1 %) 156 / 162 (96.3 %) 162 / 162 (100 %) System accuracy results for glucose concentrations between 37 mg/dl (2.06 mmol/l) and mg/dl (26.4 mmol/l) Reagent system lot Within ± 5 mg/dl & ± 5 % (Within ± 0.28 mmol/l & ± 5 %) Within ± 10 mg/dl & ± 10 % (Within ± 0.56 mmol/l & ± 10 %) Within ± 10 mg/dl & ± 15 % (Within ± 0.56 mmol/l & ± 15 %) Within ± 15 mg/dl & ± 20 % (Within ± 0.83 mmol/l & ± 20 %) TD14D228-C0E 88 / 162 (54.3 %) 142 / 162 (87.7 %) 160 / 162 (98.8 %) 161 / 162 (99.4 %) TD14D828-C0E 75 / 162 (46.3 %) 134 / 162 (82.7 %) 162 / 162 (100 %) 162 / 162 (100 %) TD14D928-C0E 81 / 162 (50.0 %) 133 / 162 (82.1 %) 156 / 162 (96.3 %) 162 / 162 (100 %) 13/21

14 System Accuracy Evaluation of Consensus Error Grid for with three reagent system lots Region Number Percent A % B % C % D % E /21

15 System Accuracy Evaluation of Statistical Analysis Summary of bias analysis (Bland and Altman) n=200 Reagent system lot 1: TD14I615-CDC Reagent system lot 2: TD14I815-CDC Reagent system lot 3: TD14I915-CDC Bias 95% Limits of agreement Bias [%] 95% Limits of agreement [%] to to to to to to /21

16 Summary of regression analysis (Passing and Bablok) Fora System Accuracy Evaluation of n=200 Regression Reagent system lot 1: TD14D228-C0E Reagent system lot 2: TD14 D828-C0E Reagent system lot 3: TD14 D928-C0E Y= X Y= X Y= X 16/21

17 System Accuracy Evaluation of Blood Glucose Monitoring System Absolute differences between and YSI 2300 STAT PLUS 17/21

18 System Accuracy Evaluation of 4. Conclusion and discussion: This study showed that the Fora blood glucose monitoring system not only met the current accuracy standard (ISO15197:2003), but also exceeded the minimum acceptable accuracy standard published in the newer 2013 guideline (ISO15197:2013) when compared the test results with a laboratory reference. In addition, 99% of the data points fall within Zone A & B of Consensus Error Grid, which means that when user obtained a value using Fora monitor, it should not result in a clinical decision that could be deemed inappropriate. There are several glucose monitoring systems out in the market. Since FDA is in preparation of tightening the accuracy standard to ± 15% by adopting the new ISO15197:2013 standard. Based on the results, it will be reasonable assumption that Fora met the new ISO standard, and therefore would be a reasonable selection for users if they like to use meter with a higher accuracy. Reference a International Diabetes Federation, the global burden. b. In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. EN ISO 15197:2003. c. In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013. d. Parkes JL, Slatin SL, Pardo S, Ginsberg BH: A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000; 23: e. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: f. Passing H, Bablok. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 1983; 21: /21

19 System Accuracy Evaluation of Blood Glucose Monitoring System Appendix Raw data with adjusted samples marked Subject- ID Diabetes type [1/2/NA] Sex [f/m] Age [years] Date Temperature [ ] Humidity [%] Haematocrit (meas.) [%] adjusted glucose category sampling time value2 sampling time Lot: TD14D228-C0E Lot: TD14D828-C0E Lot: TD14D928-C0E 1 NA m /10/ :11 07: m /10/ :09 08: m /10/ :28 08: f /10/ :45 08: NA m /10/ :05 09: NA f /10/ :28 09: NA m /10/ :00 10: NA f /10/ :11 10: NA f /10/ :34 10: NA m /10/ :54 10: NA m /10/ :12 11: NA m /10/ :06 13: m /10/ :29 13: f /10/ :51 13: m /10/ :39 14: m /10/ :53 14: f /10/ :32 15: NA /10/ :35 15: f /10/ :56 15: f /10/ :03 16: f /10/ :21 07: m /10/ :37 07: m /10/ :43 07: m /10/ :08 08: f /10/ :44 08: f /10/ :18 09: m /10/ :21 09: f /10/ :37 09: m /10/ :59 10: f /10/ :14 10: f /10/ :30 10: NA f /10/ :50 10: m /10/ :44 11: f /10/ :47 11: NA /10/ :20 13: m /10/ :39 13: f /10/ :45 13: f /10/ :07 14: f /10/ :41 14: f /10/ :07 15: /21

20 System Accuracy Evaluation of Blood Glucose Monitoring System Subject- ID Diabetes type [1/2/NA] Sex [f/m] Age [years] Date Temperature [ ] Humidity [%] Haematocrit (meas.) [%] adjusted glucose category sampling time value2 sampling time Lot: TD14D228-C0E Lot: TD14D828-C0E Lot: TD14D928-C0E 43 2 m /10/ :30 15: m /10/ :43 15: m /10/ :21 07: f /10/ :45 07: f /10/ :02 08: f /10/ :44 08: f /10/ :01 09: f /10/ :27 09: m /10/ :38 09: m /10/ :14 10: m /10/ :35 10: f /10/15 73, :08 11: m /10/ :34 11: m /10/ :58 13: f /10/ :13 13: m /10/ :37 13: m /10/ :01 14: m /10/ :36 14: m /10/ :52 14: f /10/ :09 15: NA /10/ :45 15: m /10/ :36 07: NA NA /10/ :53 07: m /10/ :11 08: m /10/ :33 08: f /10/ :56 08: f /10/ :11 09: NA /10/ :40 09: m /10/ :11 11: m /10/ :47 10: f /10/ :24 11: f /10/ :12 13: f /10/ :39 13: NA /10/ :09 14: NA /10/ :55 14: f /10/ :23 15: f /10/ :20 15: f /10/ :43 15: NA f /10/ :21 16: NA f /10/ :39 16: /21

21 System Accuracy Evaluation of Blood Glucose Monitoring System Subject- ID Diabetes type [1/2/NA] Sex [f/m] Age [years] Date Temperature [ ] Humidity [%] Haematocrit (meas.) [%] adjusted glucose category sampling time value2 sampling time Lot: TD14D228-C0E Lot: TD14D828-C0E Lot: TD14D928-C0E 86 1 m /10/ :03 07: f /10/ :03 08: f /10/ :51 09: m /10/ :07 09: m /10/ :14 10: f /10/ :57 09: NA /10/ :58 12: f /10/ :54 16: m /10/ :32 16: m /10/ :10 16: m /10/ :18 13: m /10/ :23 13: m /10/ :49 16: f /10/ :12 14: NA /10/ :23 15: NA /10/ :30 15: NA f /10/ :04 13: NA m /10/ :18 13: NA m /10/ :12 11: NA f /10/ :11 13: /21

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D. Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips.

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Higher Measuring Accuracy Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Simple. Accurate. Reliable.

More information

Smart & Simple Mobile Solutions.

Smart & Simple Mobile Solutions. Higher Measuring Accuracy Smart & Simple Mobile Solutions. Test Simply. Live Better. FORA Diamond lets you connect all your Bluetooth devices. *Also available with a Micro USB. ifora Diabetes Manager All

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

NICO /14 Nipro Diagnostics, Inc.

NICO /14 Nipro Diagnostics, Inc. Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd CLINICAL

More information

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic

More information

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013 linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation

A New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation 691301DSTXXX10.1177/1932296817691301Journal of Diabetes Science and TechnologyChristiansen et al research-article2017 Original Article A New, Wireless-enabled Blood Glucose Monitoring System That Links

More information

3. System Accuracy Evaluation LUX Hgb test 1. Samples

3. System Accuracy Evaluation LUX Hgb test 1. Samples 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

Multi User Model. Product Overview

Multi User Model. Product Overview Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values

More information

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method

Accuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch

More information

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.

OGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking. ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring

More information

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

A meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson. Report from an evaluation organised by SKUP

A meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson. Report from an evaluation organised by SKUP SKUP Scandinavian evaluation of laboratory equipment for primary health care ONETOUCH Ultra A meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson Report from an evaluation

More information

System Accuracy Evaluation LUX Chol HDL test

System Accuracy Evaluation LUX Chol HDL test System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall

More information

SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5

SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 USER-EVALUATION ASCENSIA CONTOUR Blood Glucose Meter System Table of contents SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 MATERIALS AND METHODS... 6 ASCENSIA CONTOUR...

More information

GB/T Translated English of Chinese Standard: GB/T

GB/T Translated English of Chinese Standard: GB/T Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro

More information

Meter and test strips designed for glucose self-measurement manufactured by Arkray, Inc. Report from an evaluation organised by SKUP

Meter and test strips designed for glucose self-measurement manufactured by Arkray, Inc. Report from an evaluation organised by SKUP SKUP Scandinavian evaluation of laboratory equipment for primary health care GLUCOCARD X-METER GLUCOCARD X-SENSOR Meter and test strips designed for glucose self-measurement manufactured by Arkray, Inc.

More information

MQ Comparison of glucometers using heparin whole blood

MQ Comparison of glucometers using heparin whole blood MQ 2014-1 Comparison of glucometers using heparin whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing

More information

A meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP

A meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Compact Plus A meter designed for glucose self-measurement manufactured by Roche Diagnostics Report from an evaluation

More information

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting

Point-of-Care Blood Glucose Meter Accuracy in the Hospital Setting In Brief Point-of-care (POC) tests provide analytical information that can be used to make decisions at patients bedside, as opposed to laboratory tests that must be run at a central laboratory. POC testing

More information

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support

SAMPLE. Glucose Monitoring in Settings Without Laboratory Support POCT13c Glucose Monitoring in Settings Without Laboratory Support This guideline focuses on performance of point-of-care glucose monitoring systems, with an emphasis on safety practices, quality control,

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

It s Just a Waived Glucose, Isn t It?

It s Just a Waived Glucose, Isn t It? It s Just a Waived Glucose, Isn t It? What Is the Next Step? Becky Damiani, MT (ASCP), Senior Inspection Specialist Laboratory Accreditation Program Objectives Understand the CLIA requirements surrounding

More information

MQ Comparison of glucometers using whole blood

MQ Comparison of glucometers using whole blood MQ 2017-1 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices

More information

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting

A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Original papers A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Güzin Aykal*, Ayşenur Yegin, Özgür Tekeli, Necat Yilmaz Central Laboratories, Antalya

More information

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved: PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:

More information

MQ Comparison of glucometers using whole blood

MQ Comparison of glucometers using whole blood MQ 2018-2 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices

More information

SKUP Scandinavian evaluation of laboratory equipment for primary health care

SKUP Scandinavian evaluation of laboratory equipment for primary health care S K U P i n D e n m a r k, D E K S, R i g s h o s p i t a l e t - G l o s t r u p, DK- 2 6 0 0 G l o s t r u p, P h o n e + 4 5 3 8 6 3 4 4 0 0, w w w. s k u p. n u S K U P i n N o r w a y, N o k l u s,

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

TheraSense FreeStyle blood glucose meter

TheraSense FreeStyle blood glucose meter May 2002 Evaluation Report MDA 02049 TheraSense FreeStyle blood glucose meter Crown Copyright 60 WHAT YOU CAN EXPECT FROM MDA EVALUATION REPORTS The Device Evaluation Service (DES) aims to provide independent

More information

BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY

BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY DAVID C KLONOFF, MD Medical Director Diabetes Research Institute of Mills-Peninsula Health Services San Mateo, California Clinical Professor of Medicine,

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff

More information

Omnitest : 2013

Omnitest : 2013 ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample

More information

ABSTRACT ORIGINAL RESEARCH. Guido Freckmann. Nina Jendrike. Annette Baumstark. Stefan Pleus. Christina Liebing. Cornelia Haug

ABSTRACT ORIGINAL RESEARCH. Guido Freckmann. Nina Jendrike. Annette Baumstark. Stefan Pleus. Christina Liebing. Cornelia Haug Diabetes Ther (2018) 9:683 697 https://doi.org/10.1007/s13300-018-0392-6 ORIGINAL RESEARCH User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria

More information

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.

The performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests. Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added

More information

ISO 15197:2013 / EN ISO 15197:2015 Clinical Validation Study List for Blood Glucose Monitoring System

ISO 15197:2013 / EN ISO 15197:2015 Clinical Validation Study List for Blood Glucose Monitoring System List No. STUDY TITLE DATE / PLACE SCIENTIST BRIEF RESULTS TYPE OF STUDY 01 Intermeiate measurement precision Jan 2018 The FORA 6 Connect fulfille the requirements of evaluation of FORA 6 Connect Multi-functional

More information

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial Journal of Diabetes Science and Technology Volume 7, Issue 5, September 2013 Diabetes Technology Society ORIGINAL ARTICLE Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American

More information

Computing the Surveillance Error Grid Analysis: Procedure and Examples

Computing the Surveillance Error Grid Analysis: Procedure and Examples 5959DSTXXX.77/99685959Journal of Diabetes Science and TechnologyKovatchev et al research-article Original Article Computing the Surveillance Error Grid Analysis: Procedure and Examples Journal of Diabetes

More information

Mission Hb accurately detects both

Mission Hb accurately detects both Mission Digital Hb-Testing System: Save Cost by testing 2- Parameters with Single Strip real time ie: i.e: Hemoglobin & Hematocrit-[PCV]. Presented By : Remedy Healthcare Private Limited, New Delhi, [INDIA].

More information

Analytical error of home glucose monitors: a comparison of 18 systems

Analytical error of home glucose monitors: a comparison of 18 systems Original Article Ann Clin Biochem 1999; 36: 72-79 Analytical error of home glucose monitors: a comparison of 18 systems Roger N Johnson and John R Baker From the Department of Clinical Biochemistry, Green

More information

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values

The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary

More information

Management of Central Venous Access Devices. Blood Glucose Monitoring

Management of Central Venous Access Devices. Blood Glucose Monitoring Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

ISO INTERNATIONAL STANDARD

ISO INTERNATIONAL STANDARD INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais

More information

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON

More information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT

More information

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Clinical Impact of Prandial State, Exercise, and Preparation on the Equivalence of Alternative- Blood Glucose Testing DAWN M. BINA, CMT ROBYN

More information

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies Journal of Diabetes Science and Technology Volume 4, Issue 5, September 2010 Diabetes Technology Society ORIGINAL ARTICLES Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood

More information

Blood Glucose Monitoring System

Blood Glucose Monitoring System Blood Glucose Monitoring System Contents Chapter 1 OMRON Blood Glucose Monitoring System HEA-232 1.1 Introduction 1.2 List of Kit Contents 1.3 Product and Meter Display Chapter 2 Installing and Setting

More information

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood

More information

1 PROTOCOL. Comparison Study Summary

1 PROTOCOL. Comparison Study Summary Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015

More information

Blood Glucose Measurement. Omnitest 5 & Omnican Lance FAQ Frequently Asked Questions

Blood Glucose Measurement. Omnitest 5 & Omnican Lance FAQ Frequently Asked Questions Blood Glucose Measurement Omnitest 5 & Omnican Lance FAQ Frequently Asked Questions FAQ Omnitest 5 & Omnican Lance CONTENT Omnitest 5 Overview 3 Application Range 5 Target Groups 5 Test Strips 5 Settings

More information

Interference and Point-of-Care Testing Devices

Interference and Point-of-Care Testing Devices Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT 1 Learning Objectives

More information

RETAIL BULK DIABETIC SUPPLY TOOL BOX

RETAIL BULK DIABETIC SUPPLY TOOL BOX RETAIL BULK DIABETIC SUPPLY TOOL BOX 1 ADVOCATE Brand Advantage Employers Can Save 52-72% On Diabetic Supplies Free Talking Meter - No Coding Required One of the Most Accurate Blood Glucose Meters on the

More information

SAMPLE. Clinical and Laboratory Standards Institute Advancing Quality in Health Care Testing

SAMPLE. Clinical and Laboratory Standards Institute Advancing Quality in Health Care Testing Vol. 32 No. 12 Replaces EP27-P Vol. 29 No. 16 How to Construct and Interpret an Error Grid for Quantitative Diagnostic Assays; Approved Guideline This guideline describes what an error grid is, why it

More information

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Abbott PXP Glucose Meters 2015 Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Objectives After you complete this Computer-Based Learning (CBL) module, you should be able to: Explain

More information

OneTouch Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose

OneTouch Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose Laurence B Katz, Mike Grady, Steven J Setford, and Brian L Levy US Endocrinology SUPPLEMENT

More information

Summary of Significant Changes at this Revision

Summary of Significant Changes at this Revision COPY Summary of Significant Changes at this Revision CR5890 Removal of Ketone guidelines - refer to Specialist Diabetes nurse (adult wards); Children's ward has their own protocol Remove all references

More information

Thermometer PRODUCT CATALOGUE

Thermometer PRODUCT CATALOGUE Blood Glucose Monitor Safety Lancet Home Care Blood Pressure Monitor Thermometer PRODUCT CATALOGUE About TaiDoc Since 1998 TaiDoc has been dedicated to manufacturing and marketing premium medical devices

More information

Effects of Simulated Altitude on Blood Glucose Meter Performance: Implications for In-Flight Blood Glucose Monitoring

Effects of Simulated Altitude on Blood Glucose Meter Performance: Implications for In-Flight Blood Glucose Monitoring Journal of Diabetes Science and Technology Volume 6, Issue 4, July 2012 Diabetes Technology Society ORIGINAL ARTICLE Effects of Simulated Altitude on Blood Glucose Meter Performance: Tolu, M.B.B.S., M.R.C.P.,

More information

Why do we need SD LipidoCare?

Why do we need SD LipidoCare? SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing

More information

EVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL

EVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL BIOMEDICAL ENGINEERING- APPLICATIONS, BASIS & COMMUNICATIONS EVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL 91 HUNG-CHUN HUANG 1,2, CHIA-HUNG CHIEN 1,2, CHENG-YI WANG 1, FOK-CHING CHONG

More information

FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D.

FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D. FDA Perspective - Public Health Notification: Potentially Fatal Errors with GDH-PQQ Glucose Monitoring Technology Courtney C. Harper, Ph.D., Director, Division of Chemistry and Toxicology Devices Office

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System

Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose

More information

WELL CARD HEALTH TOOL BOX

WELL CARD HEALTH TOOL BOX WELL CARD HEALTH TOOL BOX 1 ADVOCATE Brand Advantage Well Card Health Members Will Save 52-70% On Diabetic Supplies Free Talking Meter Or Dash Meter Programs To Well Card Health Members 3 to 5 day UPS

More information

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues

Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Using Blood Glucose Meters in the Hospital: Defining Critically Ill and Addressing Accreditation Issues Stephen E. Kahn, PhD, DABCC, FACB Professor and Vice Chair, Clinical Services, Pathology Loyola University

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche

COBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where applicable to the scope of services

More information

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are

More information

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,

More information

SKUP Scandinavian evaluation of laboratory equipment for primary health care

SKUP Scandinavian evaluation of laboratory equipment for primary health care SKUP Scandinavian evaluation of laboratory equipment for primary health care Accu-Chek Mobile Meter and test cassettes designed for glucose self-testing manufactured by Roche Diagnostics GmbH Report from

More information

SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World

SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World MIT OpenCourseWare http://ocw.mit.edu SP.718 Special Topics at Edgerton Center: D-Lab Health: Medical Technologies for the Developing World Spring 2009 For information about citing these materials or our

More information

Summary of Significant Changes at this Revision. Items Required

Summary of Significant Changes at this Revision. Items Required Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of

More information

Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers

Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers MD-TIP Workshop at UVA Campus, 2011 Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers Meijuan Li, PhD Acting Team Leader Diagnostic Devices

More information