Laboratory Evaluation Aeroqual S-500 OZU Ozone Sensor

Size: px
Start display at page:

Download "Laboratory Evaluation Aeroqual S-500 OZU Ozone Sensor"

Transcription

1 Laboratory Evaluation Aeroqual S-5 OZU Ozone Sensor

2 Background Three Aeroqual Series S-5 (model OZU, -.15ppm) portable ozone monitors that were previously field-tested at the SCAQMD Rubidoux fixed ambient monitoring station (2/1/215 to 4/14/215) under ambient weather conditions, have now been evaluated in the SCAQMD Chemistry Laboratory under controlled ozone concentration, temperature, and relative humidity. Aeroqual S-5 OZU (3 units tested): Gaseous sensors (metal oxide; non-fem) Each unit measures: Ozone (pphm) Unit cost: ~$5 Time resolution: 1-min Units IDs: AQ #1, AQ #2, AQ #3 FEM instrument: Ozone (Serinus 1, American Ecotech, Providence, RI) Instrument cost: ~$7, Time resolution: 1-min 2

3 Ozone Concentration (ppb) FEM Ozone Conc. (ppb) Aeroqual S-5 Linear Correlation Coefficient with FEM 3 Aeroqual vs FEM (Ozone Conc. Ramping; 2 C, 4% RH) FEM AQ #1 AQ #2 AQ # Ozone Conc. Ramping (-15 ppb; 2 C, 4% RH) y = x R² = Average of Aeroquals (ppb) Three Aeroqual units tracked well with the ozone conc. change as recorded by FEM. At the upper end of Aeroqual s detection limit, the Aeroqual units underestimated the ozone conc. as recorded by FEM. In ozone concentration range of -15 ppb, the three Aeroqual units showed excellent correlation with the corresponding FEM data (R 2 >.99) at 2 C and 4% RH. 3

4 Accuracy (2 C and 4% RH) Aeroqual S-5 Accuracy Steady State (#) Sensor mean (ppb) FEM (ppb) Accuracy (%) The three Aeroqual units showed high accuracy compared to the FEM at 2 C and 4% RH. Accuracy ranges from 83. to 93.1%. Aeroqual Data Recovery & Intra-model Variability Data recovery for ozone from AQ #1, AQ #2, and AQ #3 was 94%, 1%, and 91%, respectively. Low ozone measurement variations were observed among the three Aeroqual units at 2 C and 4% RH. 4

5 Aeroqual S-5 Precision Precision (Effect of ozone conc., temperature and relative humidity) Low Pollutant Level Relative Humidity 15% 4% 65% Medium Pollutant Level Relative Humidity 15% 4% 65% High Pollutant Level Relative Humidity 15% 4% 65% 35 C 35 C 35 C 2 C 2 C 2 C 5 C 5 C 5 C PRECISION (%) PRECISION (%) PRECISION (%) 1% represents high precision. The three Aeroqual units exhibited high precision during almost all tested conditions (ozone concentration, T and RH). FEM s precision was also high across all conditions. 5

6 Ozone Concentration (ppb) Ozone Concentration (ppb) Ozone Concentration (ppb) Ozone Concentration (ppb) Aeroqual S-5 Climate Susceptibility Aeroqual vs FEM (Ozone Conc. Ramping; 5 C, 15% RH) FEM AQ #2 AQ #3 (*AQ #1 failed to record data) Low Temp-Low RH Aeroqual vs FEM (Ozone Conc. Ramping; 35 C, 15% RH) FEM AQ #2 AQ #3 AQ #1 failed to record data. High Temp-Low RH Aeroqual vs FEM (Ozone Conc. Ramping; 5 C, 65% RH) FEM AQ #1 AQ #2 AQ # Low Temp-High RH Aeroqual vs FEM (Ozone Conc. Ramping; 35 C, 65% RH) FEM AQ #1 AQ #2 AQ # High Temp-High RH

7 1-min mean O 3 concentration (ppb) 1-min mean NO 2 concentration (ppb) 5 NO 2 Interferent FEM O3 AQ #1 AQ #3 FRM NO Time (min) -2 In the laboratory, the effect of NO 2 interferent is evaluated by exposing sensors to increasing concentrations of NO 2 at 2 C and 4% RH. As shown in the figure, both the FEM O 3 and Aeroqual sensors maintained their baseline readings throughout the NO 2 concentration ramping from to 7 ppb. 7

8 Discussion Accuracy: The three Aeroqual sensors showed high accuracy compared to the FEM at 2 C and 4% RH. Accuracy ranges from 83. to 93.1%. Precision: The three Aeroqual sensors exhibited high precision during all tested conditions (ozone concentration, T and RH). Intra-model variability: Low ozone measurement variations were observed among the three Aeroqual sensors at 2 C and 4% RH. Data recovery: Data recovery for ozone from AQ#1, AQ#2, and AQ#3 was 94%, 1%, and 91%, respectively. Baseline: At all conditions, except for 35 C and relative humidity higher than 4%, FEM ozone instrument baseline was close to zero, while the sensors baseline was around 5 ppb. At the aforementioned extreme conditions, the FEM ozone baseline was as low as -2 ppb, while the sensors baseline was zero as opposed to 5 ppb difference from the FEM, mainly due to the fact that Aeroqual sensors do not report negative values. Linearity of sensor response: Aeroqual sensors showed excellent linear correlation with the FEM ozone measurement data (R 2 >.99) below 15 ppb at 2 C and 4% RH. Interferent: Sensors were inert to NO 2 at 2 C and 4% RH. When NO 2 was increased from to 7 ppb, the sensors maintained their baseline readings. Drift : Aeroqual sensors had negligible drift. Climate susceptibility: During the lab studies, temperature and relative humidity had little effect on the sensor performance. 8

Brief History of the BAM s for paper thickness s pollution measures spm10fem E-BAM developed PM 2.5 FEM

Brief History of the BAM s for paper thickness s pollution measures spm10fem E-BAM developed PM 2.5 FEM Preliminary PM 2.5 BAM/TEOM/FRM Data Hoke P. Kimball MARAMA Monitoring Meeting Richmond, VA November 17-19, 19, 2008 Brief History of the BAM -1930 s for paper thickness -1970 s pollution measures -1990

More information

US power plant carbon standards and clean air and health co-benefits

US power plant carbon standards and clean air and health co-benefits US power plant carbon standards and clean air and health co-benefits Emissions Modeling For each scenario, we used emissions output from ICF International s Integrated Planning Model (IPM) conducted by

More information

AIR POLLUTION AND EMERGENCY DEPARTMENT VISITS FOR DEPRESSION IN EDMONTON, CANADA

AIR POLLUTION AND EMERGENCY DEPARTMENT VISITS FOR DEPRESSION IN EDMONTON, CANADA International Journal of Occupational Medicine and Environmental Health 2007;20(3):241 245 DOI 10.2478/v10001-007-0024-2 AIR POLLUTION AND EMERGENCY DEPARTMENT VISITS FOR DEPRESSION IN EDMONTON, CANADA

More information

What is the Difference Between the New BAM-1020 PM2.5 FEM and Older BAM-1020 Monitors?

What is the Difference Between the New BAM-1020 PM2.5 FEM and Older BAM-1020 Monitors? Met One Technical Bulletin Revised: April 21, 2009 By: Dennis Hart Met One Instruments, Inc. 1600 N.W. Washington Blvd. Grants Pass, OR 97526 (541) 471-7111 www.metone.com What is the Difference Between

More information

There are many varieties of particulate pollutants in our homes and offices. Ones from outside include industrial

There are many varieties of particulate pollutants in our homes and offices. Ones from outside include industrial Model: IAQ Indoor PM2.5 -PM2.5 Particulate matter (PM) is a particle pollution, which is produced in a great number of ways that can be classified into either mechanical or chemical processes. Traditionally,

More information

LION. Application Note PRECISION. Linear Position Measurement with Eddy-Current Sensors. LA March, Applicable Equipment: Applications:

LION. Application Note PRECISION. Linear Position Measurement with Eddy-Current Sensors. LA March, Applicable Equipment: Applications: LION PRECISION Application Note LA02-0061 March, 2013 Linear Position Measurement with Eddy-Current Sensors Position, Displacement Applicable Equipment: Eddy-Current position measurement systems. Applications:

More information

Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo, Nihon Waters, JAPAN

Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo, Nihon Waters, JAPAN Waters Alliance LC/MS System Analytical Conditions Chromatograms Mass Spectra Calibration Quantitation Analysis of Pesticides (II) Metribuzin & their metabolites in Rice Jun Yonekubo,, JAPAN Key Words

More information

Determination of Copper in Green Olives using ICP-OES

Determination of Copper in Green Olives using ICP-OES Application Note Food and Agriculture Determination of Copper in Green Olives using ICP-OES Intelligent Rinse function reduced analysis time by 60%, saving 191.4 L of argon Authors Ryley Burgess, Agilent

More information

THS74/75 Industrial Temperature & Humidity Transmitter. Operation Manual

THS74/75 Industrial Temperature & Humidity Transmitter. Operation Manual THS74/75 Industrial Temperature & Humidity Transmitter Operation Manual 1. Features 2. Security considerations 3. Installation 4. Connection 5. Display and LED instruction 6. Software setting and calibration

More information

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for

More information

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009 Basic knowledge of POCT instuments The Tide Resort Bangsan Beach 18 June 2009 1 POCT instruments concepts Point of care testing Diagnostic medical devices used at or near the site of patient care Bed site,

More information

David M. Shelow, Restek Corporation, 110 Benner Circle, Bellefonte, PA 16823

David M. Shelow, Restek Corporation, 110 Benner Circle, Bellefonte, PA 16823 USE OF SILCOCAN CANISTERS FOR STORING LOW-LEVEL (1ppb-ppb) REACTIVE SULFURS IN AIR David M. Shelow, Restek Corporation, 11 Benner Circle, Bellefonte, PA 16823 INTRODUCTION Analysis of sulfur-containing

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Spirometry Quality Control Plan

DIAGNOSTIC ACCREDITATION PROGRAM. Spirometry Quality Control Plan DIAGNOSTIC ACCREDITATION PROGRAM Spirometry Quality Control Plan Table of Contents Introduction...1 Spirometry Definitions and Requirements...2 Spirometry Requirements... 2...4 Daily Quality Control (see

More information

Christoph Zuth, Alexander L. Vogel, Sara Ockenfeld, Regina Huesmann and Thorsten Hoffmann * S-1. Supporting Information for:

Christoph Zuth, Alexander L. Vogel, Sara Ockenfeld, Regina Huesmann and Thorsten Hoffmann * S-1. Supporting Information for: Supporting Information for: Ultra-High-Resolution Mass Spectrometry in Real-Time: Atmospheric Pressure Chemical Ionization Orbitrap Mass Spectrometry (APCI- Orbitrap-MS) of Atmospheric Organic Aerosol

More information

Living in the Tail Pipe Pollution Dispersion and Transport David Waugh Air Quality Sciences Environment Canada Dartmouth, NS

Living in the Tail Pipe Pollution Dispersion and Transport David Waugh Air Quality Sciences Environment Canada Dartmouth, NS Living in the Tail Pipe Pollution Dispersion and Transport David Waugh Air Quality Sciences Environment Canada Dartmouth, NS NSEL - AQ Forum - March 2006 1 NSEL AQ Forum March - 2006 Photo Courtesy NOAA/NESDIS

More information

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED from a CURRENT LOOP P18L TYPE USER S MANUAL 1 2 Contents 1. Application... 2 2. Basic requirements, operational safety... 2 3. Installation... 2 3.1. Assembly...

More information

STATE ENVIRONMENTAL HEALTH INDICATORS COLLABORATIVE (SEHIC) CLIMATE AND HEALTH INDICATORS

STATE ENVIRONMENTAL HEALTH INDICATORS COLLABORATIVE (SEHIC) CLIMATE AND HEALTH INDICATORS STATE ENVIRONMENTAL HEALTH INDICATORS COLLABORATIVE (SEHIC) CLIMATE AND HEALTH INDICATORS Category: Indicator: Health Outcome Indicators Allergic Disease MEASURE DESCRIPTION Measure: Time scale: Measurement

More information

TALKING POINTS. Ten Reasons Why the Ozone Air Quality Standard Must Be Strengthened

TALKING POINTS. Ten Reasons Why the Ozone Air Quality Standard Must Be Strengthened October 22, 2014 TALKING POINTS Ten Reasons Why the Ozone Air Quality Standard Must Be Strengthened 1. The current standard is not protective of public health. The Clean Air Act requires EPA to set National

More information

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum DKK-1 ELISA, Cat.No. BI-20413 For the quantitative determination of DKK-1 in human serum ASSAY CHARACTERISTICS Method Standard range Conversion factor Sample volume / well Incubation time, temp. Sensitivity

More information

Agilent 1220 Infinity LC System Performance Specifications

Agilent 1220 Infinity LC System Performance Specifications IET International Equipment Trading, Ltd. www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Agilent 1220 Infinity LC System Performance

More information

Air Quality assessment during indoor use of the Tobacco Heating System 2.2

Air Quality assessment during indoor use of the Tobacco Heating System 2.2 Air Quality assessment during indoor use of the Tobacco Heating System 2.2 THS 2.2. is commercialized under the IQOS brand name Dr Catherine Goujon Ginglinger - Philip Morris International Research & Development,

More information

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer AACC, July 2007 Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer E. Garry, M. Pesta, N. Zampa Advanced Instruments, Inc., Norwood, MA 02062 AbstracT The Advanced Model

More information

Indoor air quality. Save energy by improving indoor air quality with innovative CO 2. sensors

Indoor air quality. Save energy by improving indoor air quality with innovative CO 2. sensors Indoor air quality Save energy by improving indoor air quality with innovative CO 2 sensors (ppm) 2000 CO 2 concentration With windows closed and without forced ventilation Effect on people Lack of concentration

More information

PAH - Sources and Measurement Mike Woodfield - AEAT

PAH - Sources and Measurement Mike Woodfield - AEAT PAH - Sources and Measurement Mike Woodfield - AEAT The Background (1) PAH in Ambient include substances known to produce cancer; World Health Organisation considers the risk to human health from ambient

More information

INSTITUTE OF PHYSICS

INSTITUTE OF PHYSICS FEDERAL UNIVERSITY OF SOUTH MATO GROSSO INSTITUTE OF PHYSICS IMPACTS OF CLIMATE VARIABILITY ON RESPIRATORY MORBIDITY. AUTOR: AMAURY DE SOUZA, FLAVIO ARISTONE, ANA PAULA BERTOSSI OHIO 2014 PROBLEM SITUATION

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Air Pollution and Hospital Admissions for Respiratory Disease in Certain Areas of Korea

Air Pollution and Hospital Admissions for Respiratory Disease in Certain Areas of Korea J Occup Health 2000; 42: 185 191 Journal of Occupational Health Air Pollution and Hospital Admissions for Respiratory Disease in Certain Areas of Korea Belong CHO 1, Jaewook CHOI 2 and Yong-Tae YUM 2 1

More information

What are the Human Health Effects of Air Pollution?

What are the Human Health Effects of Air Pollution? What are the Human Health Effects of Air Pollution? Dr. Judy Guernsey Department of Community Health and Epidemiology Dalhousie University Nova Scotia Department of Environment and Labour Outdoor Air Quality

More information

University of Bristol - Explore Bristol Research. Peer reviewed version. Link to published version (if available): /

University of Bristol - Explore Bristol Research. Peer reviewed version. Link to published version (if available): / Bamashmoos, F., Woznowski, P., & Tryfonas, T. (2018). A review of air quality sensing technologies and their potential interfaces with IoT for asthma management. In Proceedings of the 11th PErvasive Technologies

More information

The Health Effects of Air Pollution. Haim Bibi Carmel Medical Center

The Health Effects of Air Pollution. Haim Bibi Carmel Medical Center The Health Effects of Air Pollution on Children Haim Bibi Carmel Medical Center Haifa Why are children at special risk? Physically, children are more vulnerable to air pollution than adults. Their respiratory

More information

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP

TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED P18L TYPE USER S MANUAL. from a CURRENT LOOP TEMPERATURE and HUMIDITY TRANSDUCER SUPPLIED from a CURRENT LOOP P18L TYPE USER S MANUAL Contents 1. Application... 2 2. Basic requirements, operational safety... 2 3. Installation... 2 3.1. Assembly...

More information

Hotspot Analysis For Examining The Association Between Spatial Air Pollutants And Asthma In New York State, USA Using Kernel Density Estimation (KDE)

Hotspot Analysis For Examining The Association Between Spatial Air Pollutants And Asthma In New York State, USA Using Kernel Density Estimation (KDE) Hotspot Analysis For Examining The Association Between Spatial Air Pollutants And Asthma In New York State, USA Using Kernel Density Estimation (KDE) Francis Tuluri 1, Amit Kr. Gorai 2, Aaron James 1 1

More information

Abstract. Introduction

Abstract. Introduction A Fast Dilute-And-Shoot Method for Simultaneous 5-Hydroxyindoleacetic Acid (5-HIAA), Vanillylmandelic Acid (VMA), and Homovanillic Acid (HVA) LC-MS/MS Analysis in Human Urine By Shun-Hsin Liang and Sue

More information

Rethinking Approaches To Low Dose Extrapolation for Environmental Health Risk Assessment

Rethinking Approaches To Low Dose Extrapolation for Environmental Health Risk Assessment Rethinking Approaches To Low Dose Extrapolation for Environmental Health Risk Assessment Introduction Over the past half-century, methods for quantitatively assessing the human health risks from exposures

More information

Savillex Technical Note

Savillex Technical Note Savillex Technical Note C400d PFA Concentric Nebulizer for ICP-MS Summary The Savillex C-Flow 400d (C400d) PFA concentric nebulizer combines the reliability of a low flow glass concentric nebulizer with

More information

Forensic Analysis of Blood Alcohol Concentration

Forensic Analysis of Blood Alcohol Concentration Application Note Forensics Forensic Analysis of Blood Alcohol Concentration Using the Agilent 886 GC with Agilent J&W DB BAC1 UI and Agilent J&W DB-BAC2 UI columns and the Agilent 7697A headspace sampler

More information

3. System Accuracy Evaluation LUX Hgb test 1. Samples

3. System Accuracy Evaluation LUX Hgb test 1. Samples 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be

More information

BAM Monitor Performance. Seasonal and Geographic Variation in NC

BAM Monitor Performance. Seasonal and Geographic Variation in NC BAM Monitor Performance Seasonal and Geographic Variation in NC 2009-10 2 Presenter Wayne Cornelius 3 Introduction Unsuccessful ARM tests in 2007 and 2009, using a configuration of R&P TEOM monitor. Acquired

More information

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD)

Development and Validation of a Polysorbate 20 Assay in a Therapeutic Antibody Formulation by RP-HPLC and Charged Aerosol Detector (CAD) LIFE SCIENCES AR ENOUGH? HOW DO YOU KNOW RAW MATERIALS ARE PURE? HOW DO YOU EVALUATE PRODUCT PACKAGING? HOW DO YOU DO YOU KNOW WHAT ANALYTICAL TECHNIQUE TO USE? HOW DO YOU SIMULTANEOUSLY TEST FOR TWO BYPRODUCTS?

More information

Development and Validation of Multiresidue Pesticide Methods at FDA/CFSAN

Development and Validation of Multiresidue Pesticide Methods at FDA/CFSAN Development and Validation of Multiresidue Pesticide Methods at FDA/CFSAN Jon W. Wong U.S. Food and Drug Administration Center for Food Safety and Applied Nutrition American Council of Independent Laboratories

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Ammonia Volatilization from Urea, Mechanisms, Magnitude, and Management

Ammonia Volatilization from Urea, Mechanisms, Magnitude, and Management Ammonia Volatilization from Urea, Mechanisms, Magnitude, and Management David E. Kissel University of Georgia Overview Urea Consumption Definitions UAN Solution Reactions Urea Dissolution and Diffusion

More information

Exploration of real-time crash likelihood of Powered-Two Wheelers in Greece. Road safety; real-time data; crash likelihood; powered-two-wheelers

Exploration of real-time crash likelihood of Powered-Two Wheelers in Greece. Road safety; real-time data; crash likelihood; powered-two-wheelers Exploration of real-time crash likelihood of Powered-Two Wheelers in Greece Theofilatos Athanasios 1 *, Yannis George 1, Kopelias Pantelis 2, Papadimitriou Fanis 3 1 National Technical University of Athens,

More information

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms

A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide in Bulk and Pharmaceutical Dosage Forms OPEN ACCESS Eurasian Journal of Analytical Chemistry ISSN: 1306-3057 2017 12(2):31-44 DOI 10.12973/ejac.2017.00152a A New Stability-Indicating and Validated RP-HPLC Method for the Estimation of Liraglutide

More information

Chip Resistors WR 06 X 1000 F T L WW 25 M R002 F T L

Chip Resistors WR 06 X 1000 F T L WW 25 M R002 F T L How to Order R 06 1000 L ype code Size code unctional code Resistance olerance Packaging code ermination code R : iscrete 1~10MR 25 : 2512 (6432) : Normal 24 : 2 significant digits : +/- 1% : 7" reel taping

More information

Ozone and Health: Clinical Studies. Mark W. Frampton MD University of Rochester Medical Center Rochester, NY

Ozone and Health: Clinical Studies. Mark W. Frampton MD University of Rochester Medical Center Rochester, NY Ozone and Health: Clinical Studies Mark W. Frampton MD University of Rochester Medical Center Rochester, NY QUESTIONS What do we know about respiratory effects of ozone? Are there acute cardiovascular

More information

Labeling Summary / Package Insert NIOX MINO

Labeling Summary / Package Insert NIOX MINO Labeling Summary / Package Insert NIOX MINO (According to 21CFR 809.10(b) Labeling for In Vitro Diagnostic Devices) For USA 1 Proprietary and Established Names NIOX MINO Airway Inflammation Monitor measures

More information

Solution range. Superior accuracy in surface interaction analysis

Solution range. Superior accuracy in surface interaction analysis Solution range Superior accuracy in surface interaction analysis [ Q-SENSE ] Explore the nanoscale world with Q-Sense Based on established and powerful quartz crystal microbalance with dissipation (QCM-D)

More information

The setup and validation of an LC-MS/MS assay of Androstenedione & Testosterone a labs experience

The setup and validation of an LC-MS/MS assay of Androstenedione & Testosterone a labs experience The setup and validation of an LC-MS/MS assay of Androstenedione & Testosterone a labs experience Keith Mulready Specialist Senior Medical Scientist Why measure Testosterone? Hypogonadism in Males Cancer

More information

Summary of Analytical Method for Quantitative Estimation of Fingolimod and Fingolimod Phosphate from Human Whole Blood Samples

Summary of Analytical Method for Quantitative Estimation of Fingolimod and Fingolimod Phosphate from Human Whole Blood Samples Fingolimod Whole Blood Analysis Summary of Analytical Method for Quantitative Estimation of Fingolimod and Fingolimod Phosphate from Human Whole Blood Samples Study Detail: - Determination of Fingolimod

More information

Risk Characterization

Risk Characterization Risk Characterization 1 Learning Objectives By the end of this module, participants should have an understanding of: The purpose of risk characterization Integrating the results of hazard identification,

More information

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

MULTI-COMPONENT ANALYSIS OF HEAVY METALS MULTI-COMPONENT ANALYSIS OF HEAVY METALS in Various Foods Using Shimadzu s AA, ICP, ICPMS & EDX HEAVY METALS TESTING HEAVY METALS POSE A SIGNIFICANT THREAT TO HUMAN HEALTH The U.S. Food and Drug Administration

More information

New Materials for Removal of Trace Impurities from Environmental Waters

New Materials for Removal of Trace Impurities from Environmental Waters New Materials for Removal of Trace Impurities from Environmental Waters July, 2010 John Sawyer Life-Changing Research and Development John.Sawyer@matricresearch.com Chemical and Environmental Technologies

More information

DIRECT ENZYMATIC HbA1c ASSAY

DIRECT ENZYMATIC HbA1c ASSAY DIRECT ENZYMATIC HbA1c ASSAY ACCURATE AND SIMPLE DIAZYME INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses

More information

Asthma and air pollution: health effects and prevention

Asthma and air pollution: health effects and prevention Asthma and air pollution: health effects and prevention FuyuenYip PhD, MPH Air Pollution and Respiratory Health Branch Division of Environmental Hazards and Health Effects National Center for Environmental

More information

WNK5 Series High Precision Pressure Transmitter

WNK5 Series High Precision Pressure Transmitter WNK5 Series High Precision Pressure Transmitter WNK5 series pressure transmitter is the high performance pressure transmitters provided by nhui WNK Electronic Engineering Co., Ltd. It transforms the physical

More information

Air quality forecasting and its value as a preventative measure for sensitive people

Air quality forecasting and its value as a preventative measure for sensitive people 2 nd December 2009 Air quality forecasting and its value as a preventative measure for sensitive people www.citeair.eu www.airqualitynow.eu Dr Kirsty Smallbone University of Brighton s European Regional

More information

ACSM ACTRIS intercomparison at the SIRTA French atmospheric supersite in the region of Paris (France)

ACSM ACTRIS intercomparison at the SIRTA French atmospheric supersite in the region of Paris (France) AURORE ACSM ACTRIS intercomparison at the SIRTA French atmospheric supersite in the region of Paris (France) Vincent Crenn 1*, Jean Sciare 1, Olivier Favez, Roman Fröhlich 3, Philip Croteau, Manjula Canagaratna,

More information

Effects of air pollution on the incidence of myocardial infarction

Effects of air pollution on the incidence of myocardial infarction See Featured editorial, page 1721 London School of Hygiene and Tropical Medicine, London, UK Correspondence to: Mr K Bhaskaran, London School of Hygiene and Tropical Medicine, M107, 49 51 Bedford Square,

More information

Climate Change and our Children s Health

Climate Change and our Children s Health Climate Change and our Children s Health Robert J. Laumbach M.D., M.P.H., C.I.H. Department of Environmental and Occupational Medicine Rutgers Robert Wood Johnson Medical School March 1, 2014 Not just

More information

Toxicity of Checkmate LBAM-F and Epiphyas postvittana Pheromone to Ceriodaphnia dubia and Fathead Minnow (Pimephales promelas) Larvae.

Toxicity of Checkmate LBAM-F and Epiphyas postvittana Pheromone to Ceriodaphnia dubia and Fathead Minnow (Pimephales promelas) Larvae. Toxicity of Checkmate LBAM-F and Epiphyas postvittana Pheromone to Ceriodaphnia dubia and Fathead Minnow (Pimephales promelas) Larvae. 28 November, 2007 Aquatic Toxicology Laboratory The University of

More information

Pb-Free Reflow, PCB Degradation, and the Influence of Moisture Absorption

Pb-Free Reflow, PCB Degradation, and the Influence of Moisture Absorption Abstract Pb-Free Reflow, PCB Degradation, and the Influence of Moisture Absorption Kerin O Toole, Bob Esser, Seth Binfield, and Craig Hillman, DfR Solutions, College Park, MD Joe Beers, Gold Circuits,

More information

Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), Available online at RESEARCH ARTICLE

Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), Available online at  RESEARCH ARTICLE Ankit et al Journal of Drug Delivery & Therapeutics; 2013, 3(2), 26-30 26 Available online at http://jddtonline.info RESEARCH ARTICLE METHOD DEVELOPMENT AND ITS VALIDATION FOR QUANTITATIVE SIMULTANEOUS

More information

PtPt-electrode with temperature sensor, Intellical MTC695

PtPt-electrode with temperature sensor, Intellical MTC695 Total Chlorine DOC316.52.93092 Based on 4500-Cl D in Standard Methods for the Examination of Water and Wastewater Amperometric Forward Titration 0.003 5.00 mg/l as Cl 2 1. Introduction This application

More information

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System

Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System Ultrafast Analysis of Benzodiazepines in Urine by the Agilent RapidFire High-Throughput Triple Quadrupole Mass Spectrometry System Application Note Forensic Toxicology Authors Nikunj R. Parikh, Michelle

More information

Updating the Smith Period for Improved Late Blight Management in the UK

Updating the Smith Period for Improved Late Blight Management in the UK Updating the Smith Period for Improved Late Blight Management in the UK Siobhán Dancey Project Supervisors: Dr. Peter Skelsey & Dr. David Cooke DSS in Britain Smith Period SMITH PERIOD (ca. 1950): At least

More information

OralChroma Used worldwide by dentists and researchers

OralChroma Used worldwide by dentists and researchers OralChroma Used worldwide by dentists and researchers Oral Gas Analysis and Oral Health Monitoring OVERVIEW 1. OralChroma & The Dental Practice A pre-examination oral health analysis device Individually

More information

Opportunities to Leverage Aircraft-Derived Atmospheric Observation Data

Opportunities to Leverage Aircraft-Derived Atmospheric Observation Data Opportunities to Leverage Aircraft-Derived Atmospheric Observation Data Michael McPartland 5 December 2017 Sponsor: Gary Pokodner, ANG-C6 DISTRIBUTION STATEMENT A: Approved for public release; distribution

More information

IJRPC 2011, 1(4) Rohan et al. ISSN: INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND CHEMISTRY

IJRPC 2011, 1(4) Rohan et al. ISSN: INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND CHEMISTRY INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND CHEMISTRY Available online at www.ijrpc.com Research Article ABSORPTION CORRECTION METHOD AND SIMULTANEOUS EQUATION METHOD FOR THE SIMULTANEOUS ESTIMATION

More information

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS Int. J. Chem. Sci.: 10(4), 2012, 2039-2047 ISSN 0972-768X www.sadgurupublications.com DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF TRAMADOL IN PURE AND PHARMACEUTICAL FORMULATIONS

More information

Towards High Resolution MS in Regulated Bioanalysis

Towards High Resolution MS in Regulated Bioanalysis Towards High Resolution MS in Regulated Bioanalysis Benno Ingelse MSD 3 rd EB focus meeting June 12 th 13 th Brussels, Belgium Contributors Gary Adamson Ken Anderson Kevin Bateman Cynthia Chavez-Eng Inhou

More information

Title: Pharmacokinetics of daikenchuto, a traditional Japanese medicine (Kampo) after. single oral administration to healthy Japanese volunteers

Title: Pharmacokinetics of daikenchuto, a traditional Japanese medicine (Kampo) after. single oral administration to healthy Japanese volunteers Title: Pharmacokinetics of daikenchuto, a traditional Japanese medicine (Kampo) after single oral administration to healthy Japanese volunteers Authors: Masaya Munekage, Hiroyuki Kitagawa, Kengo Ichikawa,

More information

MAXIMIZING ALMOND QUALITY AND SHELF LIFE ROOM DECEMBER 4, 2018

MAXIMIZING ALMOND QUALITY AND SHELF LIFE ROOM DECEMBER 4, 2018 MAXIMIZING ALMOND QUALITY AND SHELF LIFE ROOM 306-307 DECEMBER 4, 2018 Managing Humidity and Temperature to Preserve Almond Quality 2 Today 1. Almond Physical and Chemical Properties 2. Interactions with

More information

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations

Development and validation of stability indicating RP-LC method for estimation of calcium dobesilate in pharmaceutical formulations Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2016, 8 (11):236-242 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4

More information

Measure breath nitric oxide for airway inflammation with the NObreath FeNO monitor

Measure breath nitric oxide for airway inflammation with the NObreath FeNO monitor Measure breath nitric oxide for airway inflammation with the NObreath FeNO monitor British technology Improving asthma management, one breath at a time. www.bedfont.com w on 01622 851122 or email ask@bedfont.com

More information

Available online Research Article

Available online  Research Article Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2015, 7(9):951-960 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Analytical method development and validation for

More information

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page:

IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page: IJPAR Vol.3 Issue 4 Oct-Dec-2014 Journal Home page: ISSN: 2320-2831 Research article Open Access Method development and validation of tenofovir disoproxil fumerate and emtricitabine in combined tablet

More information

Short guidelines for the use of the MEMORI dosimeter

Short guidelines for the use of the MEMORI dosimeter Short guidelines for the use of the MEMORI dosimeter Introduction This guide provides advice about how the MEMORI dosimeter should be used to obtain the best information about air quality for the safeguarding

More information

Registration Excise and Prohibition Department

Registration Excise and Prohibition Department Registration Excise and Prohibition Department Vikash Bhawan, New Secretariat, Bailey Road, Patna 01 Tender No:Excise/ChemLab/2016/02 Replies to pre bid queries of the bidders as per the pre bid meeting

More information

NOVEL RP-HPLC METHOD. B.Lakshmi et a. concentration range KEY INTRODUCTION. Diltiazem is used to. . It works by of contractionn of the. (dilate).

NOVEL RP-HPLC METHOD. B.Lakshmi et a. concentration range KEY INTRODUCTION. Diltiazem is used to. . It works by of contractionn of the. (dilate). B.Lakshmi et a ISSN-2319-2119 al, The Experiment, January. 2013 Vol..6(4), 365-371 A NOVEL RP-HPLC METHOD FOR THE DETERMINATION OF DILTIAZEM IN PHARMACEUTICAL DRUG PRODUCTS ABSTRACT High resolution RP-HPLC

More information

Dhull Rohit, Kumar Sanjay, Jalwal Pawan Department of Chemistry, OPJS Churu, Rajasthan, India

Dhull Rohit, Kumar Sanjay, Jalwal Pawan Department of Chemistry, OPJS Churu, Rajasthan, India International Journal of Advanced Science and Research ISSN: 2455-4227, Impact Factor: RJIF 5.12 www.allsciencejournal.com Volume 2; Issue 2; March 2017; Page No. 31-40 Validated gradient stability indicating

More information

Quality Assurance Mapping Your QC Program Equipment and Test Quality. Susan Blonshine RRT, RPFT, FAARC, AE-C

Quality Assurance Mapping Your QC Program Equipment and Test Quality. Susan Blonshine RRT, RPFT, FAARC, AE-C Quality Assurance Mapping Your QC Program Equipment and Test Quality Susan Blonshine RRT, RPFT, FAARC, AE-C How to Begin Gather resources Define PF scope of service Procedures performed Equipment Describe

More information

THE ASSOCIATION BETWEEN AIR POLLUTIONS AND RESPIRATORY DISEASES : STUDY IN THAILAND

THE ASSOCIATION BETWEEN AIR POLLUTIONS AND RESPIRATORY DISEASES : STUDY IN THAILAND THE ASSOCIATION BETWEEN AIR POLLUTIONS AND RESPIRATORY DISEASES : STUDY IN THAILAND Anongsin Danpaiboon 1 *, Phongtape Wiwatthadate 2, ChannarongChaisuwan 2,Warungarn Kosawang 3 and Petsawat Kankum 4 1

More information

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these

More information

Novel Activation Technologies for Sodium Persulfate In Situ Chemical Oxidation

Novel Activation Technologies for Sodium Persulfate In Situ Chemical Oxidation Novel Activation Technologies for Sodium Persulfate In Situ Chemical Oxidation Frank Sessa FMC Corporation Philadelphia, PA Dalbir Sethi FMC Corporation Philadelphia, PA Jean Pare ChemCo Quebec, CA Persulfate

More information

PCA Enhanced Kalman Filter for ECG Denoising

PCA Enhanced Kalman Filter for ECG Denoising IOSR Journal of Electronics & Communication Engineering (IOSR-JECE) ISSN(e) : 2278-1684 ISSN(p) : 2320-334X, PP 06-13 www.iosrjournals.org PCA Enhanced Kalman Filter for ECG Denoising Febina Ikbal 1, Prof.M.Mathurakani

More information

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES

A RAPID AND SENSITIVE UPLC/UV/MS METHOD FOR SIMVASTATIN AND LOVA S TAT IN IN SU P P O RT O F C L E A NING VA L I DAT IO N S T U DIES RPID ND SENSITIVE PLC/V/MS METD FR SIMVSTTIN ND LV S TT IN IN S P P RT F C L E NING V L I DT I N S T DIES Erin Chambers, Diane Diehl, and Jeff Mazzeo Waters Corporation, Milford, M, S BCKGRND tilizing

More information

JunkScience.com. June 16, Mr. Ray Wassel National Academy of Sciences 500 Fifth St., N.W. Washington, D.C

JunkScience.com. June 16, Mr. Ray Wassel National Academy of Sciences 500 Fifth St., N.W. Washington, D.C JunkScience.com June 16, 2016 Mr. Ray Wassel National Academy of Sciences 500 Fifth St., N.W. Washington, D.C. 20001 Re: Comment on Assessing Toxicologic Risks to Human Subjects Used in Controlled Exposure

More information

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples ASSAY CHARACTERISTICS Method Competitive Enzyme Immunoassay, HRP/TMB. Microtiter plates are coated with a polyclonal

More information

For more information, please contact: or +1 (302)

For more information, please contact: or +1 (302) Introduction Quantitative Prediction of Tobacco Components using Near-Infrared Diffuse Reflectance Spectroscopy Kristen Frano Katherine Bakeev B&W Tek, Newark, DE Chemical analysis is an extremely important

More information

Wireless Body Analysis Scale

Wireless Body Analysis Scale Wireless Body Analysis Scale FR Balance d'analyse Corporelle Connectée IT Bilancia pesapersone con analisi corporea wireless ES Báscula inalámbrica de análisis corporal DE Drahtlose Körperanalysewaage

More information

Supporting information

Supporting information 1 2 3 Supporting information SO 2 Uptake on Oleic Acid: A New Formation Pathway of Organosulfur Compounds in the Atmosphere 4 5 Jing Shang, Monica Passananti, Yoan Dupart, Raluca Ciuraru, Liselotte Tinel,

More information

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry Application note Environmental Author Terrance Hettipathirana Agilent Technologies Melbourne, Australia

More information

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets

Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets Journal of PharmaSciTech 0; ():- Research Article Development and Validation of a Simultaneous HPLC Method for Quantification of Amlodipine Besylate and Metoprolol Tartrate in Tablets * Sayyed Hussain,

More information

Optimizing Sample. Chromium Analyses in Waters. Jane Timm, James Lovick Jr, Raymond Siery, ato 2011 NEMC, Bellevue, Washington

Optimizing Sample. Chromium Analyses in Waters. Jane Timm, James Lovick Jr, Raymond Siery, ato 2011 NEMC, Bellevue, Washington Optimizing Sample Preservation for Hexavalent Chromium Analyses in Waters Jane Timm, James Lovick Jr, Raymond Siery, and Yongtao Li Underwriters Laboratories ato 2011 NEMC, Bellevue, Washington 2011 Underwriters

More information

Abstract. G. D. Patil 1, P. G. Yeole 1, Manisha Puranik 1 and S. J. Wadher 1 *

Abstract. G. D. Patil 1, P. G. Yeole 1, Manisha Puranik 1 and S. J. Wadher 1 * International Journal of ChemTech Research ISSN : 0974-4290 Vol.1,No.1,pp 16-26, Jan March 2009 A Validated Specific Reverse Phase Liquid Chromatographic Method for the Determination of Valacyclovir in

More information

Solutions for Calibration and Maintenance

Solutions for Calibration and Maintenance SOLUTIONS Solutions for Calibration and Maintenance New Series of Solutions To meet the requirements of various applications, new categories of solutions have been created: HI 50xx series, technical solutions:

More information

FINAL REPORT. Authorship: Professor: Manuel Sillero Quintana 1. Associate Professor: Jakub Grzegorz Adamczyk 2 Student: Sacha Karabás 1

FINAL REPORT. Authorship: Professor: Manuel Sillero Quintana 1. Associate Professor: Jakub Grzegorz Adamczyk 2 Student: Sacha Karabás 1 FINAL REPORT RESEARCH PROJECT: Application of thermography for the establishment of the lower limb thermal profile of Elite Masters Athletes and the influence of athletic competition on their skin temperature.

More information

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum

Rat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum Please read carefully due to Critical Changes, e.g., see Calculation of Results. For Research Use Only. Not For Use In Diagnostic

More information

Impact of Particle Mass Distribution on the Measurement Accuracy of Low-Cost PM-Sensors

Impact of Particle Mass Distribution on the Measurement Accuracy of Low-Cost PM-Sensors Impact of Particle Mass Distribution on the Measurement Accuracy of Low-Cost PM-Sensors Bernd Laquai, 20.10.2017 Commodity PM-measurement devices are used more and more by citizens concerned about air

More information