3. System Accuracy Evaluation LUX Hgb test 1. Samples
|
|
- Morgan Pierce
- 5 years ago
- Views:
Transcription
1 3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be measured by two different meters and by Hemocue. Exclusion criteria, such as packed cell volume (hematocrit), were based on the manufacturer s instructions for use. Capillary blood samples were collected by auto-disabling lancing devices prepared and processed according to the instructions for use, including sample pretreatment if required. Sample containers designed for the collection of capillary blood was used. It is difficult to obtain a sufficient number of fresh capillary blood samples with very low and very high Hemoglobin concentrations. In these cases, modified capillary blood samples in which the Hemoglobin concentration is elevated or lowered was substituted (see below Table 2.) The Hemoglobin concentrations were distributed as specified in Table 1. Once a concentration category is filled, no more samples were added to that category. Table.1 Hemoglobin concentrations of samples for system accuracy evaluation Percentage of samples (%) Hemoglobin concentration g/dl (mmol/l) to 10.9 (0.78 to 1.69) to 14.0 (1.71 to 2.17) to 18.0 (2.19 to 2.79) to 21.0 (2.81 to 3.26) to 25.0 (3.27 to 3.88) Only unaltered capillary blood samples were used for Hemoglobin concentrations of 5 g/dl to 25 g/dl (0.78 mmol/l to 3.88 mmol/l). In order to obtain sufficient samples in the lowest and highest concentration intervals, Hemoglobin concentrations were adjusted as follows. - To obtain additional samples with Hemoglobin < 5 g/dl (< 0.78 mmol/l), capillary blood samples was collected with K 2EDTA anti-coagulant tube and incubated to allow Hemoglobin to hydrolyse. The incubation conditions (Temperature: 23~24, Humidity: 40~50 RH%) to produce sample compatible System Accuracy Evaluation report LUX Hemoglobin Test Page 1 of 9
2 with the system without hemolytic was determined by the manufacturer. - To obtain additional samples with Hemoglobin > 25 g/dl (> 3.88 mmol/dl), capillary blood samples was collected with K 2EDTA anti-coagulant tube and then supplemented with Hemoglobin. Table.2 Number of samples Hemoglobin concentration Number of samples Percentage of samples (%) g/dl (mmol/l) 5.0 to 10.9 (0.78 to 1.69) to 14.0 (1.71 to 2.17) to 18.0 (2.19 to 2.79) to 21.0 (2.81 to 3.26) to 25.0 (3.27 to 3.88) 0 0 Total Strip & Meter 1) Lots of hemoglobin test strips Site # Strip lot number Quantity 1 H strips from 40 vials 2 H strips from 40 vials 2) LUX METER Site #1 Site #2 meter # serial number meter # serial number 1 L L L L ) Instructions for use Two healthcare professionals were given instructions for use of LUX METER for the subject s skin puncture. 3. Test Date and Environmental condition 1) Test period: from 14 th February 2014 to 24 th March ) Temperature Range: 23.0 C to 23.9 C System Accuracy Evaluation report LUX Hemoglobin Test Page 2 of 9
3 4. Evaluation Protocol The following procedure is the minimum experimental design to evaluate system accuracy. It requires at least 100 subjects with Hemoglobin concentrations that span the measuring interval as specified in Table 1. The procedure may be modified to accommodate multiple reagent lots. The reagent system units shall be taken from the same vial/package for each sample. Steps a) to f) shall be performed for each sample. a) Assign numbers to the vials or packages (e.g. 1 to 10). b) Obtain a sample of fresh capillary blood b skin puncture. c) Remove an aliquot of the sample for at least duplicate hemoglobin measurements by the manufacturer s measurement procedure. If the measurement s procedure is designed for plasma samples, the plasma preparation procedure (e.g. centrifugation at 1000g for 10 min to remove cells) shall be performed for each sample. d) Take two reagent system units out of vial/package number one and measure the hemoglobin concentration in the sample using two different meters. Change vials or packages every ten subjects and ensure that reagent system units from all vials are used in the evaluation. Record the results. e) Repeat step c) to obtain a second from the manufacturer s measurement procedure. f) Evaluate the results from the manufacturer s measurement procedure to verify sample stability. If these results demonstrate the effect of drift based on predetermined stability criteria [e.g. change between the first and last results > 4% at hemoglobin > 0.78 mmol/l (5 g/dl), then the results for that subject shall not be included. The rejected sample shall be replaced with another sample in the same concentration interval. g) Obtain a sample from the next subject and repeat steps b) to f). 5. Minimum Acceptable System Accuracy 5.1 System accuracy requirement The minimum acceptable accuracy for results produced by a Hemoglobin monitoring system shall be as follows. Ninety-five percent (95%) of the individual hemoglobin results shall fall within ±10% at hemoglobin concentrations 0.78 mmol/l ( 5 g/dl). System Accuracy Evaluation report LUX Hemoglobin Test Page 3 of 9
4 5.2 Data presentation Results are presented hemoglobin concentration 0.78 mmol/l ( 5 g/dl), results shall be expressed as the percentage of values falling within the following intervals: ±10% 6. Test results System Accuracy Evaluation report LUX Hemoglobin Test Page 4 of 9
5 System Accuracy Evaluation report LUX Hemoglobin Test Page 5 of 9
6 System Accuracy Evaluation report LUX Hemoglobin Test Page 6 of 9
7 System Accuracy Evaluation report LUX Hemoglobin Test Page 7 of 9
8 System Accuracy Evaluation report LUX Hemoglobin Test Page 8 of 9
9 7. Conclusion Ninety-five percent (95%) of the individual hemoglobin results shall fall within ±10% at hemoglobin concentrations 0.78 mmol/l ( 5 g/dl). [Result] A fine and direct correlation between Cobas c111 and LUX hemoglobin strips was confirmed in the 200 blood samples with the correlation coefficient R 2 = The results indicate that the use of the LUX hemoglobin strip generate similar results as the Cobas c111, a laboratory instrument used in hospitals to test for hemoglobin concentration. The data shown indicates that the results are within acceptance criteria. RESULT: Passed System Accuracy Evaluation Regression analysis results for hemoglobin concentration: Linear regression R2 N Sample range [mg/dl] Y=0.9845X to 19.3 System accuracy results hemoglobin concentration Within ± 5% Within ± 10% 200/200 (100%) 200/200 (100%) Arezzo, 24 th March 2014 BIOCHEMICAL SYSTEMS INTERNATIONAL SRL Dr.ssa Annalisa Imbriani QC Manager System Accuracy Evaluation report LUX Hemoglobin Test Page 9 of 9
System Accuracy Evaluation LUX Chol HDL test
System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationClinical Evaluation for. Embrace No Code Blood Glucose Monitoring System
Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic
More informationOGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationEquivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationReport. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013
Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationOmnitest : 2013
ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample
More informationGB/T Translated English of Chinese Standard: GB/T
Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationItems in the package:
Intended Use: The EasyLife Hb Monitoring System is designed for in vitro diagnostic use only (external use only), and is suitable for self-testing. The system is for healthcare professionals and persons
More informationRat cholesterol ELISA Kit
Rat cholesterol ELISA Kit Catalog No. CSB-E11706r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat Cholesterol concentrations in serum, plasma and other biological fluids.
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia
More informationAlere INRatio 2 METHOD AND SAMPLE COLLECTION
Alere INRatio 2 METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing an INR test using the Alere INRatio2 analyser. The INRatio2 meter
More informationWorksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide
LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More informationClinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System
Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More informationUltra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC
VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values
More informationBlood Glucose Monitoring System
Blood Glucose Monitoring System Contents Chapter 1 OMRON Blood Glucose Monitoring System HEA-232 1.1 Introduction 1.2 List of Kit Contents 1.3 Product and Meter Display Chapter 2 Installing and Setting
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More informationACE. Accurate, Compact & Economical Blood Glucose Meter. User Guide
ACE Accurate, Compact & Economical Blood Glucose Meter User Guide NOTICE: The essential features of the type described and illustrations herein remaining unaltered, Pulsatom Healthcare Pvt. Ltd. reserves
More informationHbA1c (Human) ELISA Kit
HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,
More informationRat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit
Rat Mullerian Inhibiting Substance/Anti-Mullerian hormone, MIS/AMH ELISA kit Catalog No. E0228r (96 tests) Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS!
More informationMission Hb accurately detects both
Mission Digital Hb-Testing System: Save Cost by testing 2- Parameters with Single Strip real time ie: i.e: Hemoglobin & Hematocrit-[PCV]. Presented By : Remedy Healthcare Private Limited, New Delhi, [INDIA].
More informationCOBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche
Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma
Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationLIPASE liquicolor. Design Verification. Multipurpose Reagent
Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of
More informationThe performance of Microdot Test Strips has been evaluated both in laboratory and in clinical tests.
Chemical Composition Each Microdot strip contains the enzyme glucose dehydrogenase (Bacillus Sp.) = 1 IU and other ingredients (mediator, NAD, lysing agents etc.) = 200 µg. Test Strips When blood is added
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationForum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007
Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease
More informationINTRO SAFETY INFORMATION GENERAL INFORMATION TESTING PREANALYTICAL PHASE CHECKINGS RUNNING THE TEST ALTERNATIVE SITES MEMORY
04 12 30 60 66 72 INTRO SAFETY INFORMATION GENERAL INFORMATION THE LUX SYSTEM GETTING TO KNOW THE SYSTEM SETTING THE SYSTEM TESTING PREANALYTICAL PHASE CHECKINGS RUNNING THE TEST ALTERNATIVE SITES MEMORY
More informationSummary of Significant Changes at this Revision. Items Required
Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of
More informationBlood Glucose Monitoring System. User Guide
Blood Glucose Monitoring System User Guide Table of Contents Introduction...2 Important Safety Instructions...2 About ipet PRO Blood Glucose Monitoring System...3 About ipet PRO Meter...4 About the ipet
More informationInsulin ELISA. For the quantitative determination of insulin in serum and plasma.
Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United
More informationI know my value CoaguChek XS Pro system
I know my value CoaguChek XS Pro system The smart way to manage your anticoagulation clinic CoaguChek XS Pro system A barcode scanner can make a difference There has been an extraordinary increase in the
More informationEvaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers
Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17
More informationCanine Thyroid Stimulating Hormone, TSH ELISA Kit
Canine Thyroid Stimulating Hormone, TSH ELISA Kit Catalog No: E0463c 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationToxicology Blood Lead Comprehensive Testing
Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where
More informationHuman Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit
Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Catalog No: E0442h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
More informationEVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS
EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The
More informationTitle: Glucose Testing: StatStrip Glucose Meter Page 1 of 6
Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:
More informationTitle: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.
COPY Summary of Significant Changes at this Revision SOP updated Purpose and Scope Items Required 1. The Abbott Optium Xceed meter is a battery-powered device designed for the measurement of blood glucose
More informationEvaluation Report. Eurolyser Haemoglobin Test Kit (ST0190) on the Smart Analyser
Evaluation Report Eurolyser Haemoglobin Test Kit (ST0190) on the Smart Analyser Location Location: Eurolyser Diagnostica GmbH Operators: Michael Gruber Date: 01.08.2009 Specimens Samples taken from the
More informationPrecision ceedpro. Online Training for. Glucometer & Capillary Collection
Online Training for Precision ceedpro Glucometer & Capillary Collection How long will it take? This module takes approximately 30 minutes to complete. What if I have to leave before I finish? Progress
More informationRayBio Acid Phosphatase Activity Colorimetric Assay. Kit. User Manual Version 1.0 May 5, RayBiotech, Inc. Kit Protocol. (Cat#: 68CL-AcPh-S500)
RayBio Acid Phosphatase Activity Colorimetric Assay Kit User Manual Version 1.0 May 5, 2014 RayBio Acid Phosphatase Activity Colorimetric Assay (Cat#: 68CL-AcPh-S500) RayBiotech, Inc. We Provide You With
More informationTable of Contents. Section 1 Introduction... 1 Section 2 Getting Started... 2 Section 3 Components... 5
Table of Contents Section 1 Introduction... 1 Section 2 Getting Started... 2 Section 3 Components... 5 Meter 5 Test Strips 8 Test Devices 8 Control Strips or Control Devices 13 Section 4 Initial Setup...
More informationInsulin Aspart ELISA Kit
Insulin Aspart ELISA Kit Catalogue DEIABL215 For the qualitative determination of antibodies to insulin aspart in human serum and plasma FOR RESEARCH USE ONLY Creative Diagnostics. All rights reserved
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationMouse C3 (Complement Factor 3) ELISA Kit
Mouse C3 (Complement Factor 3) ELISA Kit Cat. No.:DEIA8289 Pkg.Size:96T Intended use The Mouse C3 (Complement Factor 3) ELISA Kit is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring
More informationRat Proinsulin ELISA
Rat Proinsulin ELISA For the quantitative determination of proinsulin in rat serum For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-PINRT-E01 Size: 96 wells Version: June
More informationSerum Amyloid A ELISA
Serum Amyloid A ELISA For the quantitative determination of serum amyloid A (SAA) in serum plasma For Research use Only. Not for Use in Diagnostic Procedures Please see Appendix A for Reference Serum information
More informationBern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.
Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,
More informationToxicology Blood Lead Comprehensive Testing
Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where
More informationRat C-peptide ELISA. For the quantitative determination of C-peptide in rat serum
Rat C-peptide ELISA For the quantitative determination of C-peptide in rat serum Please read carefully due to Critical Changes, e.g., see Calculation of Results. For Research Use Only. Not For Use In Diagnostic
More informationApolipoprotein A-1 ELISA
Apolipoprotein A-1 ELISA For the quantitative determination of apolipoprotein A1 in serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationPLUS One. Blood Glucose Monitoring System. Owner s Manual
PLUS One Blood Glucose Monitoring System Owner s Manual Dear PLUS One System Owner: Thank you for purchasing the PLUS One Blood Glucose Monitoring System. This manual provides important information to
More informationDear Prodigy System Owner:
User's Manual Dear Prodigy System Owner: You have chosen one of the best blood glucose monitoring systems available. This booklet has important information about the system. Please read it carefully. Your
More informationRayBio Phenylalanine Fluorometric Assay Kit
RayBio Phenylalanine Fluorometric Assay Kit User Manual Version 1.0 June 25, 2014 RayBio Phenylalanine Fluorometric (Cat#: 68-Phenyl-S100) RayBiotech, Inc. We Provide You With Excellent Support And Service
More informationSet-up and special features
Set-up and special features Set Time 2 : 45 Set Date 23 Apr 202 Choose to keep your Verio IQ simple or customise it to your desire with its various features Alerts On Of f Please set with care. In the
More informationStandard Operating Procedure
Subject Free Thyroxine (FT4-II) cobas e411 Index Number Lab-1594 Section Regional/Affiliate Subsection Laboratory Category Departmental Contact Goplin, Darcy Last Revised 2/2/2017 References Required document
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationToxicology Blood Lead Comprehensive Testing
Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where applicable to the scope of services
More informationGlucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers
Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,
More informationliquicolor Design Verification
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4
More informationHemoCue Hb METHOD AND SAMPLE COLLECTION
HemoCue Hb 201 + METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing a Haemoglobin (Hb) test using the HemoCue Hb 201 + analyser.
More informationValidity and Reliability of a Glucometer Against Industry Reference Standards
514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference
More informationliquicolor (AMP Buffer, IFCC) Design Verification
Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...
More informationBlood Glucose Monitoring System USER GUIDE. For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips
Blood Glucose Monitoring System USER GUIDE For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips Thank you for choosing the Ascensia CONTOUR Blood Glucose Monitoring System!
More informationMulti User Model. Product Overview
Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL
More informationBlood Glucose & Ketone Monitoring System
Blood Glucose & Ketone Monitoring System Self monitoring of blood glucose is an integral part of diabetes care, but the high cost of testing can make this impossible. At ACON, our goal is to provide high
More informationDetermination of hemoglobin is one of the most commonly
ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning
More informationMouse Ultrasensitive Insulin ELISA
Mouse Ultrasensitive Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationA New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation
691301DSTXXX10.1177/1932296817691301Journal of Diabetes Science and TechnologyChristiansen et al research-article2017 Original Article A New, Wireless-enabled Blood Glucose Monitoring System That Links
More informationGalactose and Lactose Assay Kit
Galactose and Lactose Assay Kit Catalog Number KA0842 100 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 General Information... 4 Materials
More informationRETAIL BULK DIABETIC SUPPLY TOOL BOX
RETAIL BULK DIABETIC SUPPLY TOOL BOX 1 ADVOCATE Brand Advantage Employers Can Save 52-72% On Diabetic Supplies Free Talking Meter - No Coding Required One of the Most Accurate Blood Glucose Meters on the
More informationWELL CARD HEALTH TOOL BOX
WELL CARD HEALTH TOOL BOX 1 ADVOCATE Brand Advantage Well Card Health Members Will Save 52-70% On Diabetic Supplies Free Talking Meter Or Dash Meter Programs To Well Card Health Members 3 to 5 day UPS
More informationSummary of Significant Changes at this Revision
COPY Summary of Significant Changes at this Revision CR5890 Removal of Ketone guidelines - refer to Specialist Diabetes nurse (adult wards); Children's ward has their own protocol Remove all references
More informationBÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:
BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:
More informationHuman IL-2. Pre-Coated ELISA Kit
Human IL-2 (Interleukin 2) Pre-Coated ELISA Kit Catalog No: 90-2083 1 96 well Format (96 tests) Detection Range: 31.2 2000 pg/ml Sensitivity: < 18.75 pg/ml This immunoassay kit allows for the in vitro
More informationHuman HBcAb IgM ELISA kit
Human HBcAb IgM ELISA kit Catalog number: NR-R10163 (96 wells) The kit is designed to qualitatively detect HBcAb IgM in human serum or plasma. FOR RESEARCH USE ONLY. NOT FOR DIAGNOSTIC OR THERAPEUTIC PURPOSES
More informationCoaguChek XS Plus System Whole Blood Protime/INR
1 CoaguChek XS Plus System Whole Blood Protime/INR A. PURPOSE The CoaguChek XS System is intended for use by professional healthcare providers for quantitative protime testing (PT) and calculated INR for
More information8.2 Principles of Quantitative Hematologic Determinations (1)
3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily
More informationBehzad Poopak, DCLS PhD
Behzad Poopak, DCLS PhD Test Report Name Age Critical Low HEMATOLOGY Activated Partial Thromboplastin Time, Plasma Critical High - 150 sec Units Fibrinogen 60 - mg/dl INR (International Normalizing
More information2. Before Testing Monitor Checker Test...10 Inserting Lancets into Lancing Device...11 Quality Control Testing...13
Table of Contents 1. About your HemoSmart Haemoglobin Screening System Contents of Kit...4 HemoSmart Haemoglobin Meter...5 HemoSmart Haemoglobin Test Strip...7 Adjustable Lancing Device and Lancets...8
More informationProfessional Procedure Guide (90821B 04/2008)
Professional Procedure Guide (90821B 04/2008) This test is WAIVED under the Clinical Laboratory Improvement Amendments of 1988 (CLIA). If a laboratory modifies the test instructions, the test will no longer
More informationExendin-4 (Exenatide) ELISA Kit
Exendin-4 (Exenatide) ELISA Kit Catalog: DEIABL227 For the quantitative determination of Exendin-4 in serum or plasma using competitive ELISA method For Research Use Only. Protocol Provided for Informational
More informationPRIMA. Blood Glucose Monitoring System. Owner s Manual
PRIMA Blood Glucose Monitoring System Owner s Manual English Dear FORA Diamond PRIMA System Owner: Thank you for purchasing the FORA Diamond PRIMA Blood Glucose Monitoring System. This manual provides
More information2/2/2011. Blood Components and Transfusions. Why Blood Transfusion?
Blood Components and Transfusions Describe blood components Identify nursing responsibilities r/t blood transfusion Discuss factors r/t blood transfusion including blood typing, Rh factor, and cross matching
More information