Y FILMS DOSIMETR Nederland België / Belgique

Size: px
Start display at page:

Download "Y FILMS DOSIMETR Nederland België / Belgique"

Transcription

1 DOSIMETRY FILMS

2 GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous and caustic chemicals Environmentally friendly with no waste to dispose of Use your dark room space for something more productive For the Characterization, Calibration, Quality Control and Wide-Area Dose Measurement of Therapeutic, Radiological and Industrial Radiation Sources Beam Sources Measure field uniformity Determine field shape Make depth-dose characterizations Map fields in 2- and 3-dimensions Validate IMRT and stereotactic treatment plans Isotope Sources Characterize depth-dose fields of HDR brachytherapy sources Map the activity of LDR brachytherapy seeds Verify the operation of intravascular sources Design and validate treatments with eye plaques Map and measure plaques for implantation Map fields inside irradiators Map and measure hot particles Features Self developing no processing, no chemicals Not light sensitive no darkroom required Unaffected by aqueous fluids can be immersed in water phantoms Energy independent in therapy range Tissue equivalent (except XR Type T) No dose rate or dose fractionation effects Films can be cut to any size and shape Wide dynamic range Almost linear dose-density response Tough, robust, abrasion resistant Quantitative Dosimetry The responses of GAFCHROMIC dosimetry films are dependent upon factors specific to the hospital or laboratory including the energy spectrum of the radiation source, the characteristics of the densitometer, film scanner or other means of measurement, etc. Therefore, the user should determine the dose-density response under the conditions pertaining to their facility. Since the absolute responses of GAFCHROMIC dosimetry films vary from lot-to-lot, the user should re-measure response when changing the lot number.

3 Select from four media types and several formats Custom sizes and shapes available upon request GAFCHROMIC HD-810 and D-200 Single-sided active layer on polyester HD-810 is an 8 x10 sheet film D-200 dosimeters are pre-formatted to fit standard spectrophotometer holders Dose range Gy Very high spatial resolution to 1 micron, or less GAFCHROMIC MD-55 and MD-22 Double active layers laminated between polyester film substrates MD-55 is a sheet film in 5 x5, 10cm x 10cm, 18cm x 18cm and 7cm x 43cm formats MD-22 is a x octagonal film designed for 3-D mapping in anatomical phantoms A,B Dose range 2-Gy Sensitometric Response of GAFCHROMIC HD-810 and D-200 (HD-810 Lot# H1032H810 measured with Radiochromic Densitometer) Sensitometric Response of GAFCHROMIC MD-55 and MD-22 (MD-55 Lot# J1426MD55 measured with Radiochromic Densitometer) A. e.g. Radiosurgery Head Phantom and IMRT Verification System manufactured by CIRS Inc. information available at: and B. Standard Imaging IMRT Dose Verification Phantom available at: %20IMRT%20Brochure%20.pdf GAFCHROMIC HS and HS-14 Single active layer laminated between polyester film substrates HS is a sheet film in 5 x5, 10cm x 10cm, 18cm x 18cm and 7cm x 43cm formats HS-14 are pre-formatted x strips for use with the Galileo intravascular brachytherapy treatment system C Dose range -40Gy Sensitivity of GAFCHROMIC HS and HS-14 (Batch#30263 measured with Radiochromic Densitometer) 6MeVx - rays 18MeVx - rays 6MeVelectrons 18MeVelectrons C. Galileo is a trademark of Guidant Corporation. Information available at GAFCHROMIC XR Type T Single active layer laminated between polyester film substrates Designed primarily for photons with energies <200keV XR Type T is a 5 x5 sheet film Dose range Gy 3.0 Sensitometric Response of GAFCHROMIC XR Type T (Batch # Measured with Radiochromic Densitometer) 120kVp, 2mmAl 80kVp, 2mmAl 60kVp, 2mmAl 40kVp, 1mmAl 20kVp, 0.15mmAl

4 Film Type Configuration Substrate(s) Active layer thickness 1 Response Range Radiation Sources Media Size HD-810 Active layer on substrate 6.5µm 10Gy-400Gy Electrons: >5keV Photons: >1keV UV: 1-300nm 8 x 10 D-200 Active layer on substrate 6.5µm 10Gy-400Gy Electrons: >5keV Photons: >1keV UV: 1-300nm 1.2cm x 6cm formatted for 1cm spectrophotometer cuvette MD-55 MD-22 Multilayer laminate: substrate-active layerlaminating tape-active layer-substrate Multilayer laminate: substrate-active layerlaminating tape-active layer-substrate 2.6mil 2.6mil 2 x 16µm 2 x 16µm 2Gy-Gy 2Gy-Gy 5 x 5, 10cm x 10cm, 18cm x 18cm and 7cm x 43cm x octagonal HS Three-layer laminate: substrate-active layersubstrate 40µm Gy-40Gy 5 x 5, 10cm x 10cm, 18cm x 18cm and 7cm x 43cm HS-14 Three-layer laminate: substrate-active layersubstrate 40µm Gy-40Gy x XR Type T Three-layer laminate: substrate-active layersubstrate Yellow polyester 18µm 0.1Gy-15Gy Photons: keV 5 x 5 1. Nominal thickness. Actual thickness varies slightly from lot-to-lot. Solid Water Eye Phantom and GAFCHROMIC MD-55 film, Image courtesy of Samuel Trichter, NYU Cornell Medical Center, New York

5 Dose Distribution in Blood Irradiator Measured with GAFCHROMIC Dosimetry Film: Field Flatness Expressed % of Central Dose GAFCHROMIC Dosimetry Film Technology GAFCHROMIC Dosimetry Films have been designed for use with gamma, x-rays, electrons, protons, ion beams, α-particles and neutrons and can be used to measure and map sources of all types covering a wide range of energies down to 5keV, or lower. The active component in GAFCHROMIC Dosimetry Films is comprised of sub-micron sized crystals of a radiation-sensitive monomer. When the film is exposed to ionizing radiation, a polymerization reaction is initiated, resulting in the production of a blue-colored dye-polymer. The amount of polymer produced and the depth of the color change is proportional to the dose absorbed in the active layer. GAFCHROMIC Dosimetry Films have inherently high spatial resolution, beyond the capability of conventional silver films. Since the polymerization process does not spread from crystal-to-crystal and there is no post-exposure development process to change the size or shape of the particles, the films are capable of resolution down to the micron range. The spatial resolution is limited principally by the characteristics of the exposure source. The color change in GAFCHROMIC Dosimetry Films may be measured with any densitometer, scanner or spectrophotometer. However, the best response is obtained with instruments that measure at red-light wavelengths in the range of the peak absorbance of the blue dye, i.e. close to 670nm. Examples of such hardware include the Photoelectron Corporation CCD Microdensitometer, the Radlink LaserPro 16TM ( the Tobias TXP Transmission Densitometer, the Nuclear Associates Radiochromic Densitometer Model , or any flatbed color scanner. Spectral Absorbance of Radiochromic Dye in GAFCHROMIC Dosimetry Films Wavelength, nm

6

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Avoid exposure of film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA DESCRIPTION GAFCHROMIC MD-55 radiochromic dosimetry film is designed for the measurement

More information

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad: General Aspects of RCF (TG-235), AAPM July 2015 1 Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad, PhD, FAAPM,

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical

More information

Small Field Dosimetry & Radiochromic film

Small Field Dosimetry & Radiochromic film Small Field Dosimetry & Radiochromic film Michael Kissick, Ph.D., Assistant Professor, Medical Physics & Human Oncology & Affiliated with the Morgridge Institute for Research North Central AAPM Chapter

More information

Applications in brachytherapy

Applications in brachytherapy SAM Practical Medical Physics Session: TU-F-201 Radiochromic Film Dosimetry Update Tuesday, July 14, 2015, 2:45 pm 3:45 pm Applications in brachytherapy Samuel Trichter, M.Sc., Department of Radiation

More information

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients IDEAL DOSIMETRY FOR IMRT/SRS Absolute dose measurement Precise

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions) B3-20 Database System for GEX B3 Dosimetry (gamma and/or electron beam versions) and GEX combine to provide a Genesys 20 version of the flexible 21 CFR Part 11 certified database dosimetry system for use

More information

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau 2018 Council on Ionizing Radiation Measurements and Standards Low Energy (80-300 kev) Electron Beam Calibration Gary Pageau GEX Corporation Centennial, Colorado USA ISO 9001 certified manufacturer of dosimetry

More information

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10 GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10 NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55

Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 Azam Niroomand-Rad, Chair Department of Radiation Medicine, Georgetown University Medical Center, Washington,

More information

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Dosimetry Standards and Dissemination Systems for Radiation Processing in China Dosimetry Standards and Dissemination Systems for Radiation Processing in China Zhang Yanli National Institute of Metrology, China April 2017 in Vienna, Austria Organization Founded in 1955, National Institute

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

Medical Physics 4 I3 Radiation in Medicine

Medical Physics 4 I3 Radiation in Medicine Name: Date: 1. This question is about radiation dosimetry. Medical Physics 4 I3 Radiation in Medicine Define exposure. A patient is injected with a gamma ray emitter. The radiation from the source creates

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Good manufacturing practices (GMP) utilized on human blood irradiation process

Good manufacturing practices (GMP) utilized on human blood irradiation process Artigos Good manufacturing practices (GMP) utilized on human blood irradiation process Cláudio Boghi Pesquisador do IPEN-USP. cboghi@uol.com.br; cboghi@usp.br Dorlivete M. Shitsuka Professora de Organização

More information

Process Control Methods in Radiation Technologies

Process Control Methods in Radiation Technologies Process Control Methods in Radiation Technologies András Kovács Nuclear Security Department Centre for Energy Research Hungarian Academy of Sciences Radiation Processing Applications Radiation sterilization:

More information

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Manik Aima, Larry A. DeWerd, Wesley S. Culberson Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ Eighth Annual Chapman Phytosanitary Irradiation Forum Dosimetry for phytosanitary applications: dos and don ts Florent KUNTZ June 14 th, 2018 What does a dosimeter measure? Question A dosimeter measures

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry Journal of Radiation Research, 2012, 53, 930 935 doi: 10.1093/jrr/rrs030 Advance Access Publication 21 August 2012 Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

More information

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch

More information

3D Dosimetry with Polymer Gel

3D Dosimetry with Polymer Gel 3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer

More information

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range J. Eduardo Villarreal-Barajas, a and Rao

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

3/26/2017. Personal Dosimetry Monitoring and Dose Measurements. Agenda. Dosimetric Terms and Definitions Dose Limits External Dosimetry

3/26/2017. Personal Dosimetry Monitoring and Dose Measurements. Agenda. Dosimetric Terms and Definitions Dose Limits External Dosimetry Speaker David Pellicciarini, CHP, MBA Vice President, Pharmacy Safety, Practice and Technical Operations Cardinal Health Nuclear Pharmacy Services david.pellicciarini@cardinalhealth.com Personal Dosimetry

More information

X-ray Safety Discussion and Tour for Health Physicists

X-ray Safety Discussion and Tour for Health Physicists X-ray Safety Discussion and Tour for Health Physicists April 1, 2016 Health Physics Society Spring Meeting Jerry Bai Environmental Administrator, Field Operations Bureau of Radiation Control Florida Department

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

High-Dose Standardization Study For γ-rays radiation processing at NIM

High-Dose Standardization Study For γ-rays radiation processing at NIM High-Dose Standardization Study For γ-rays radiation processing at NIM ZHANG Yanli, GAO Junchen, YANG Yuandi. Division of Metrology in Ionizing Radiation and medicine, National Institute of Metrology,

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara 1, Z Kuncic 1, J Adamovics 2 and C Baldock 1,3 1 Institute of Medical Physics, School of Physics, University

More information

DOSIMETRY PRODUCTS & SERVICES 2018 PRICE LIST

DOSIMETRY PRODUCTS & SERVICES 2018 PRICE LIST DOSIMETRY PRODUCTS & SERVICES 2018 PRICE LIST Dosimetry Products & Services Dosimetry and Calibration Products and Services All prices are in Canadian Dollars. Delivery for most items is 6 weeks subject

More information

Verification of micro-beam irradiation

Verification of micro-beam irradiation Journal of Physics: Conference Series OPEN ACCESS Verification of micro-beam irradiation To cite this article: Qiongge Li et al 2015 J. Phys.: Conf. Ser. 573 012047 View the article online for updates

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the

Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the Figure 1.1 PHITS geometry for PTB irradiations with: broad beam, upper panel; mono energetic beams, lower panel. Pictures of the setups and of the PMMA ring holder with 7 containers are also shown. Relative

More information

The use of polyvinyl chloride films dyed with methyl red in radiation dosimetry

The use of polyvinyl chloride films dyed with methyl red in radiation dosimetry Volume 14, No 3 International Journal of Radiation Research, July 2016 The use of polyvinyl chloride films dyed with methyl red in radiation dosimetry INTRODUCTION Dosimetry plays an important function

More information

GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7

GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7 GEX Recommended Procedure Eff. Date: 09/21/10 Rev.: D Pg. 1 of 7 NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100

More information

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy Antony L Palmer, 1,2a David Bradley,

More information

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR

Therapeutic Medical Physics. Stephen J. Amadon Jr., Ph.D., DABR Therapeutic Medical Physics Stephen J. Amadon Jr., Ph.D., DABR Outline 1. Why physicists are needed in medicine 2. Branches of medical physics 3. Physics in Radiation Oncology 4. Treatment types and Treatment

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital

More information

Multi-Channel Film Dosimetry and Gamma Map Verification

Multi-Channel Film Dosimetry and Gamma Map Verification Multi-Channel Film Dosimetry and Gamma Map Verification Micke A International Specialty Products Ashland proprietary technology, patents pending Single Channel Film Dosimetry Calibration Curve X=R R ave

More information

Hampton University Proton Therapy Institute

Hampton University Proton Therapy Institute Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Implementation Guide for the DoseControl Dosimetry System

Implementation Guide for the DoseControl Dosimetry System 1.0 SCOPE The GEX DoseControl System is intended for use as a single dosimetry system solution, specifically developed to satisfy the generic dosimetry needs of radiation sterilization application users

More information

Eye plaque dosimetry verification using novel solid state devices

Eye plaque dosimetry verification using novel solid state devices University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2014 Eye plaque dosimetry verification using novel solid state devices Talia

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

CIRMS: A Retrospective & Look Forward

CIRMS: A Retrospective & Look Forward CIRMS: A Retrospective & Look Forward Bert M. Coursey Guest Researcher, Standards Services Group National Institute of Standards and Technology CIRMS 2017 March 27, 2017 CIRMS 2017 Brief history on coordination

More information

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 MODELLING A GAMMA IRRADIATION

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing

Nuclear energy, nuclear technologies, and radiological protection Vocabulary. Part 4: Dosimetry for radiation processing Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 12749-4 First edition 2015-08-15 Nuclear energy, nuclear technologies, and radiological protection Vocabulary Part 4: Dosimetry for radiation processing

More information

Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance

Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance J. Gajewski1, M. Kłosowski1, S. Greilich2, P. Olko1 Institute of Nuclear Physics, Kraków, Poland German Cancer

More information

Radiation Protection Program Update: The Details. July 2010

Radiation Protection Program Update: The Details. July 2010 Radiation Protection Program Update: The Details July 2010 Update Topics 2 Changes mandated by Title 10, Code of Federal Regulations, Part 835, Occupational Radiation Protection (10 CFR 835) How changes

More information

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.

Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation. Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Radiation Dosimeters for Foods. Ashish Anand Department of Biological and Agricultural Engineering Texas A&M University

Radiation Dosimeters for Foods. Ashish Anand Department of Biological and Agricultural Engineering Texas A&M University Radiation Dosimeters for Foods Ashish Anand Department of Biological and Agricultural Engineering Texas A&M University Introduction Interest in food irradiation technology is increasing world-wide. Some

More information

Efficient Dosimetry for Proton Therapy

Efficient Dosimetry for Proton Therapy Efficient Dosimetry for Proton Therapy Why IBA Dosimetry? IBA Dosimetry offers the full product range tailored to fit any Proton Therapy QA needs. PT treatment safety, as well as most efficient dosimetry

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY M. R. NEMTANU, C. OPROIU, M. BRASOVEANU, M. OANE National Institute for Laser, Plasma and Radiation Physics, Electron Accelerator

More information

Option D: Medicinal Chemistry

Option D: Medicinal Chemistry Option D: Medicinal Chemistry Basics - unstable radioactive nuclei emit radiation in the form of smaller particles alpha, beta, positron, proton, neutron, & gamma are all used in nuclear medicine unstable

More information

What is radiation quality?

What is radiation quality? What is radiation quality? Dudley T Goodhead Medical Research Council, UK DoReMi Radiation Quality workshop Brussels. 9-10 July 2013 What is radiation quality? Let s start at the very beginning. A very

More information

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca

More information

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 131 Edition 2001 (E) Polymethylmethacrylate dosimetry systems for ionizing radiation processing of materials and products Systèmes de dosimétrie à polyméthylméthacrylate

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

IBA Dosimetry Company Presentation. Patient Safety & Quality Control

IBA Dosimetry Company Presentation. Patient Safety & Quality Control IBA Dosimetry Company Presentation Patient Safety & Quality Control IBA Mission Protect, enhance and save lives Protect, Enhance, and Save Lives - 2 - The History of IBA Dosimetry A story of innovation

More information

Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy

Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy Varian Acuity BrachyTherapy Suite One Room Integrated Image-Guided Brachytherapy The Acuity BrachyTherapy Suite Integrating Imaging, Planning, and Treatment in a Single Room Each component draws on the

More information

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA

PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA PROCEDURE FOR ABSORBED DOSE TO WATER DETERMINATION IN HIGH ENERGY PHOTON AND ELECTRON BEAMS BY FERROUS SULPHATE DOSIMETER AT INMRI-ENEA M. Pimpinella, A. S. Guerra, S. La Civita and R. F. Laitano Istituto

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

Sang Hun Shin, Wook Jae Yoo, Kyoung Won Jang, Seunghyun Cho, 1 Kum Bae Kim, 2 and Bongsoo Lee 3*

Sang Hun Shin, Wook Jae Yoo, Kyoung Won Jang, Seunghyun Cho, 1 Kum Bae Kim, 2 and Bongsoo Lee 3* Sensors and Materials, Vol. 28, No. 6 (2016) 619 624 MYU Tokyo 619 S & M 1216 Development of an All-in-One Phantom and Scintillator Radiation Sensor for Real-Time Monitoring of Source Position and Dose

More information

Radiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Brachytherapy Radioisotopes in contact with the tumor Overview

More information

B3 RADIOCHROMIC FILM DOSIMETRY

B3 RADIOCHROMIC FILM DOSIMETRY NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100 documents be reviewed periodically to maintain currency and continuity

More information

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Optical Scanning 3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Wuu, C.-S., Hoogcarspel, S.J.,Deh, K., Hsu, W.-Y., Adamovics, J. 2013 Journal of Physics: Conference Series

More information

Amendment No. 2. Item No. 2 (Rfx/ Event number )

Amendment No. 2. Item No. 2 (Rfx/ Event number ) Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have

More information

Verification of Advanced Radiotherapy Techniques

Verification of Advanced Radiotherapy Techniques Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Medical Use of Radioisotopes

Medical Use of Radioisotopes Medical Use of Radioisotopes Therapy Radioisotopes prove to be useful in the application of brachytherapy, the procedure for using temporary irradiation close to the area of disease (i.e. cancer) 10% Medical

More information

Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately.

Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately. TRUS QA Workshop Introduction The goals of using TRUS in prostate brachytherapy Visualize the prostate Need the US to penetrate deeply enough Need sufficient grey scale resolution to be able to visualize

More information

PERSONNEL MONITORING AND DOSIMETRY POLICIES

PERSONNEL MONITORING AND DOSIMETRY POLICIES PERSONNEL MONITORING AND DOSIMETRY POLICIES All individuals who are required to have their exposure to ionizing radiation monitored must be trained prior to using the source(s) of radiation. The radioactive

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information