A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV

Size: px
Start display at page:

Download "A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV"

Transcription

1 A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical Complex of DLNP, Joint Institute for Nuclear Research, , Dubna, Russia 1

2 Aims 1. To compare the response of two types of radiochromic films: EBT and MD55 in a 175 MeV clinical proton beam used at the Joint Institute for Nuclear Research in Dubna. 2. To compare dose distributions calculated by treatment planning system (TPS) and measured with radiochromic films in Alderson phantom. 2

3 Material and methods Radiochromic film : radiation-sensitive organic microcrystal monomers (diacetylene) coated onto polyester film base self-developing (no chemical processing) almost tissue equivalent ( for MD55 Zeff=6.5, for EBT Zeff=6.98) can be immersed in water dose-rate and number of fraction independent not sensitive to room light ( incandescent bulb light) can be used for mapping 2D treatment fields 3

4 Material and methods Differences between two types of GafChromic film (International Specialty Products, Wayne, NJ) MD 55 EBT Dose sensitivity Gy Gy 8 Gy Structure Active layer thickness: 2 x 16 μm Two transparent polyester layers of about 67 μm + 25 μm surface layer and two adhesive layers (45 μm). Active layer thickness:2x17 μm Two transparent polyester layers of about 97 μm + 6 μm surface layer Uniformity Better than ~5% (~6%) Better than 1.5 % Post exposure density growth 12 % over the first 24h, 4 % over the next two weeks ( but 16% over the first 24h Butson et al. 2003) Ready to read < 2h (but ~8% over 6h and ~9% over 24 h Cheung et al )

5 1. Measurements in water Calibration Pieces size of about 2x2 cm 2 were cut and used for calibration. Calibration doses were determined with ionization chamber at the depth of 77 mm of water. Calibration doses for MD55 film were: 10, 20, 30, 40, 50, 60 and 70 Gy. ForEBT:0.5,1,1.5,2,2.5,3,3.5,4, 4.5, 5, 5.5 and 6 Gy All the optical density (OD) readouts were performed with Vidar VXR 16 scanner. 5

6 Measurements in water After calibration the films were irradiated with unmodulated beam in water phantom. Irradiation field 6 cm x 6 cm. The supplied dose at maximum was 60 Gy for MD 55 and 3 Gy for EBT. Film was scanned with resolution of 75 pixel/inch after 48 hours. Calibration curve was used to transform an obtained bitmap into doses. Matrix of doses was drawn in the form of isodoses. For the same, central plane we collected the doses with semiconductor detector (profiles perpendicular to the beam axis - every 3 mm, depth - every 5 mm). Isodoses from film were overlaid on the isodoses obtained with Si-detector. 6

7 Measurements in water -results Isodoses from measurement with MD 55 and EBT (black lines with labels) compared with isodoses measured with Si-detector (colour scale) MD 55 EBT 7

8 Measurements in water -results Differences between measurements with MD 55 film and EBT film comparing with Si-detector on the plane ( Formula: Dose Si -Dose EBT / Dose max) MD 55 EBT % % 8

9 Measurements in water -results Differences between measurements with MD 55 film and EBT film comparing with Si-detector in the form of histograms 88 % MD 55 EBT 98% 9

10 Measurements in water -results Mean difference Max difference MD % (1.95 Gy) 20% (12 Gy) EBT 0.17% (0.005 Gy) 7% (0.02 Gy) 10

11 Measurements in Alderson phantom Clinical case was chosen, target was drawn on the CT scans of phantom and the plans of treatment were prepared. Fixing mask, boluses and collimators were made for phantom irradiation 11

12 Measurements in Alderson phantom The maximum dose was 35 Gy for MD 55 radiochromic film and 3 Gy for EBT respectively. The RCF were scanned after 48 hours after irradiation and recalculated for matrixes of doses (using calibration curve). The main isodoses were drawn on the basis of matrixes and overlaid on the isodoses from treatment plan. 12

13 Measurements in Alderson phantom - results MD55 (black lines) and isodoses from treatment planning system (colour) 13

14 Measurements in Alderson phantom - results EBT (black lines) and isodoses from treatment planning system (colour) 14

15 Measurements in Alderson phantom - results Comparison of dose profiles in the horizontal axis Maximum differences in experimental and calculated isodose as measured from dose profiles Isodose EBT film [cm] MD55 film [cm] 20% % % % % % % % %

16 Measurements in Alderson phantom - results Comparison of dose profiles in the vertical axis Maximum differences in experimental and calculated isodose as measured from dose profiles Isodose EBT film [cm] MD55 film [cm] 20% % % % % % % % %

17 Measurements in Alderson phantom - discussion At the average dose lines from EBT measurement are in better agreement with isodoses calculated by TPS than dose lines from MD55 measurement EBT MD55 Mean of differences between isodoses measured and calculated [cm] Standard deviation of mean Difference in maximum dose 102 % (+0.07 Gy) 101 % (+ 0.3 Gy) 17

18 Discussion Main reason of differences in doses measured with radiochromic film in comparison with semiconductor detector is unhomogeneity of radiochromic emulsion (<6%).. The deviations in the Bragg peak between the MD55 and semiconductor occurred due to LET-dependence of radiochromic films, thus EBT showed much lower LET-dependence. 18

19 Conclusions 1. There is a big accordance between plan and results obtained on the basis of measurements with radiochromic films. 2. Results of measurements showed better accuracy of EBT films than MD55 and its usefulness in proton dosimetry, using much lower doses, which shortens time of exposure. 19

20 References 1. Butson M. J. et al. Radiochromic film for medical radiation dosimetry // Materials Science and Engineering V. 41. p Cheung et al. Post-irradiation colouration of Gafchromic radiochromic film // Phys. Med. Biol. 2005; 50 :N281 N285. ebt 3. Luchin Y. I. et al. The use of radiochromic film in treatment verification of proton radiosurgery // Journal of Radiosurgery V. 3. p Piermattei et al. Radiochromic film dosimetry of a low energy proton beam // Med. Phys V. 27(7). p Vatnitsky S.M. Radiochromic film dosimetry for clinical proton beams // Appl. Radiat. Isot V. 48. p

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

Radiochromic film dosimetry in water phantoms

Radiochromic film dosimetry in water phantoms INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms

More information

Y FILMS DOSIMETR Nederland België / Belgique

Y FILMS DOSIMETR Nederland België / Belgique DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Avoid exposure of film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF

A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital

More information

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film

Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,

More information

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA

GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA GAFCHROMIC MD-55 RADIOCHROMIC DOSIMETRY FILM FOR HIGH-ENERGY PHOTONS CONFIGURATION, SPECIFICATIONS AND PERFORMANCE DATA DESCRIPTION GAFCHROMIC MD-55 radiochromic dosimetry film is designed for the measurement

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad: General Aspects of RCF (TG-235), AAPM July 2015 1 Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad, PhD, FAAPM,

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara 1, Z Kuncic 1, J Adamovics 2 and C Baldock 1,3 1 Institute of Medical Physics, School of Physics, University

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

Experimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants

Experimental measurement with an anthropomorphic phantom of the proton dose distribution in the presence of metal implants E13195: Slightly better series/treatment with PRONE combination with vertical field. - This field is only possible in case the tumor is not too caudal (problem with energy). - It will increase the integral

More information

Verification of micro-beam irradiation

Verification of micro-beam irradiation Journal of Physics: Conference Series OPEN ACCESS Verification of micro-beam irradiation To cite this article: Qiongge Li et al 2015 J. Phys.: Conf. Ser. 573 012047 View the article online for updates

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

Advanced film dosimetry for a UK brachytherapy audit

Advanced film dosimetry for a UK brachytherapy audit Advanced film dosimetry for a UK brachytherapy audit Tony Palmer Lead Scientist Medical Physics, Portsmouth Hospitals NHS Trust, UK University of Surrey, UK Institute of Physics and Engineering in Medicine

More information

Active Scanning Beam 3 Checking Delivery/Dosimetry

Active Scanning Beam 3 Checking Delivery/Dosimetry Active Scanning Beam 3 Checking Delivery/Dosimetry C Algranati, J Salk, A Coray, A Lomax, E Pedroni, T Boeringer, S Lin, E Hug Center for Proton Radiation Therapy Checking Delivery/Dosimetry Contents Absolute

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose

Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Applications in brachytherapy

Applications in brachytherapy SAM Practical Medical Physics Session: TU-F-201 Radiochromic Film Dosimetry Update Tuesday, July 14, 2015, 2:45 pm 3:45 pm Applications in brachytherapy Samuel Trichter, M.Sc., Department of Radiation

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients

NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients IDEAL DOSIMETRY FOR IMRT/SRS Absolute dose measurement Precise

More information

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT

Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT Rajesh

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

Evaluation of triple channel correction acquisition method for radiochromic film dosimetry Journal of Radiation Research, 2012, 53, 930 935 doi: 10.1093/jrr/rrs030 Advance Access Publication 21 August 2012 Evaluation of triple channel correction acquisition method for radiochromic film dosimetry

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy

Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 Evaluation and implementation of triple-channel radiochromic film dosimetry in brachytherapy Antony L Palmer, 1,2a David Bradley,

More information

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau 2018 Council on Ionizing Radiation Measurements and Standards Low Energy (80-300 kev) Electron Beam Calibration Gary Pageau GEX Corporation Centennial, Colorado USA ISO 9001 certified manufacturer of dosimetry

More information

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams.

Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams. IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Verification of relative output factor measurement using Gafchromic films for small-field radiosurgery photon beams. To cite this

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Dosimetric Analysis Report

Dosimetric Analysis Report RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------

More information

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter

Dosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter Applications of Nuclear Techniques (CRETE15) International Journal of Modern Physics: Conference Series Vol. 44 (2016) 1660221 (9 pages) The Author(s) DOI: 10.1142/S2010194516602210 Dosimetric characteristics

More information

Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study

Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study Radiotherapy Quality Assurance of a 50-kV Radiotherapy Unit Using EBT3 GafChromic Film: A Feasibility Study Technology in Cancer Research & Treatment 2016, Vol. 15(1) 163 170 ª The Author(s) 2015 Reprints

More information

In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment Mara Severgnini, 1a

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with

More information

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY

CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY CHARACTERIZATION OF THE ELECTRON BEAM RADIATION FIELD BY CHEMICAL DOSIMETRY M. R. NEMTANU, C. OPROIU, M. BRASOVEANU, M. OANE National Institute for Laser, Plasma and Radiation Physics, Electron Accelerator

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

IORT with mobile linacs: the Italian experience

IORT with mobile linacs: the Italian experience IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams

More information

Multi-Channel Film Dosimetry and Gamma Map Verification

Multi-Channel Film Dosimetry and Gamma Map Verification Multi-Channel Film Dosimetry and Gamma Map Verification Micke A International Specialty Products Ashland proprietary technology, patents pending Single Channel Film Dosimetry Calibration Curve X=R R ave

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar

More information

Electron Beam ET - Reloaded Therapy - Reloaded

Electron Beam ET - Reloaded Therapy - Reloaded 8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles

More information

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range

Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 1, 2014 Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range J. Eduardo Villarreal-Barajas, a and Rao

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ

Eighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ Eighth Annual Chapman Phytosanitary Irradiation Forum Dosimetry for phytosanitary applications: dos and don ts Florent KUNTZ June 14 th, 2018 What does a dosimeter measure? Question A dosimeter measures

More information

Implementation of an Anthropomorphic Phantom for the Evaluation of Proton Therapy Treatment Procedures

Implementation of an Anthropomorphic Phantom for the Evaluation of Proton Therapy Treatment Procedures Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 5-2010 Implementation of an Anthropomorphic Phantom for the Evaluation

More information

Digital film dosimetry in radiotherapy and the development of analytical applications software

Digital film dosimetry in radiotherapy and the development of analytical applications software University of Wollongong Research Online University of Wollongong Thesis Collection 1954-2016 University of Wollongong Thesis Collections 2005 Digital film dosimetry in radiotherapy and the development

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Dosimetry of Rotational Total Skin Electron Irradiation in the Treatment of Mycosis Fungoides

Dosimetry of Rotational Total Skin Electron Irradiation in the Treatment of Mycosis Fungoides International Journal of Advanced Research in Physical Science (IJARPS) Volume 3, Issue 7, 2016, PP 41-47 ISSN 2349-7874 (Print) & ISSN 2349-7882 (Online) www.arcjournals.org Dosimetry of Rotational Total

More information

Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom

Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using a spiral water phantom Journal of Radiation Research, 2013, 54, 1153 1159 doi: 10.1093/jrr/rrt059 Advance Access Publication 17 May 2013 Three-dimensional radiochromic film dosimetry for volumetric modulated arc therapy using

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. YY Translated English of Chinese Standard: YY Translated English of Chinese Standard: YY0831-2011 www.chinesestandard.net Email: Sales@ChineseStandard.net ICS 11.040.50 YY C 43 PHARMACEUTICAL INDUSTRY STANDARD OF THE PEOPLE S REPUBLIC OF CHINA YY

More information

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy

A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy Radiology and Oncology Ljubljana Slovenia www.radioloncol.com research article 307 A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy Jaeman Son 1,2,

More information

Collateral Exposure: The Additional Dose From Radiation Treatment

Collateral Exposure: The Additional Dose From Radiation Treatment Department of Physics and Astronomy, University of Canterbury, Private Bag 4800, Christchurch, New Zealand Collateral Exposure: The Additional Dose From Radiation Treatment Katherine Fricker (BSc) A thesis

More information

3D Dosimetry with Polymer Gel

3D Dosimetry with Polymer Gel 3D Dosimetry with Polymer Gel Yoichi Watanabe, Ph.D. University of Minnesota Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN NCCAAPM Meeting, Minneapolis, MN April 28, 2006 Outline I. Basics of polymer

More information

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE I. Petrovi a, A. Risti -Fira a, L. Kori anac a, J. Požega a, F. Di Rosa b, P. Cirrone b and G. Cuttone

More information

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY doi:10.1016/j.ijrobp.2009.03.029 Int. J. Radiation Oncology Biol. Phys., Vol. 75, No. 1, pp. 253 259, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front

More information

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM A Thesis Presented to The Academic Faculty by Ryan J. Underwood In Partial Fulfillment of the Requirements for the Degree Master of Science in the School

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

Skyscan 1076 in vivo scanning: X-ray dosimetry

Skyscan 1076 in vivo scanning: X-ray dosimetry Skyscan 1076 in vivo scanning: X-ray dosimetry DOSIMETRY OF HIGH RESOLUTION IN VIVO RODENT MICRO-CT IMAGING WITH THE SKYSCAN 1076 An important distinction is drawn between local tissue absorbed dose in

More information

Accurate Accurate boost or Simply Accuboost

Accurate Accurate boost or Simply Accuboost Accurate Accurate boost or Simply Accuboost Zoubir Ouhib MS DABR Lynn Cancer Institute of Boca Raton Community Hospital Disclosure: Advisory board Items to be discussed Big picture on management of Breast

More information

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions) B3-20 Database System for GEX B3 Dosimetry (gamma and/or electron beam versions) and GEX combine to provide a Genesys 20 version of the flexible 21 CFR Part 11 certified database dosimetry system for use

More information

Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55

Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 Azam Niroomand-Rad, Chair Department of Radiation Medicine, Georgetown University Medical Center, Washington,

More information

Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film

Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 2, 2016 Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film Marcus Sonier, 1a Matt Wronski,

More information

Dosimetric Characterization of a Prototype Nanotechnology Microbeam Radiation Therapy Device using Gafchromic EBT2 Film

Dosimetric Characterization of a Prototype Nanotechnology Microbeam Radiation Therapy Device using Gafchromic EBT2 Film Dosimetric Characterization of a Prototype Nanotechnology Microbeam Radiation Therapy Device using Gafchromic EBT2 Film M. Hadsell, R. Ger*, C. Inscoe, E. Schreiber, J.P. Lu, S. Chang, O. Zhou Funding

More information

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report

SOMATOM Drive System Owner Manual Dosimetry and imaging performance report www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

Dosimetry and QA of proton and heavier ion beams

Dosimetry and QA of proton and heavier ion beams Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference

More information

Small Field Dosimetry & Radiochromic film

Small Field Dosimetry & Radiochromic film Small Field Dosimetry & Radiochromic film Michael Kissick, Ph.D., Assistant Professor, Medical Physics & Human Oncology & Affiliated with the Morgridge Institute for Research North Central AAPM Chapter

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

Medical applications round table. Radiotherapy Diagnostic radiology Nuclear medicine Radiation protection

Medical applications round table. Radiotherapy Diagnostic radiology Nuclear medicine Radiation protection Medical applications round table Radiotherapy Diagnostic radiology Nuclear medicine Radiation protection Will be discussing Ion beam therapy/radiotherapy Diagnostic radiology Nuclear medicine Radiation

More information

Medical- and Radiation Physics Research at Surrey

Medical- and Radiation Physics Research at Surrey Medical- and Radiation Physics Research at Surrey D.A. Bradley Department of Physics Centre for Nuclear and Radiation Physics University of Surrey Guildford United Kingdom * d.a.bradley@surrey.ac.uk Topic

More information

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:

MAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by: SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

Discuss the general planning concepts used in proton planning. Review the unique handling of CTV / ITV / PTV when treating with protons

Discuss the general planning concepts used in proton planning. Review the unique handling of CTV / ITV / PTV when treating with protons Mark Pankuch, PhD Discuss the general planning concepts used in proton planning Review the unique handling of CTV / ITV / PTV when treating with protons Pencil Beam distributions and PBS optimization Cover

More information

Good manufacturing practices (GMP) utilized on human blood irradiation process

Good manufacturing practices (GMP) utilized on human blood irradiation process Artigos Good manufacturing practices (GMP) utilized on human blood irradiation process Cláudio Boghi Pesquisador do IPEN-USP. cboghi@uol.com.br; cboghi@usp.br Dorlivete M. Shitsuka Professora de Organização

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Dosimetry Standards and Dissemination Systems for Radiation Processing in China

Dosimetry Standards and Dissemination Systems for Radiation Processing in China Dosimetry Standards and Dissemination Systems for Radiation Processing in China Zhang Yanli National Institute of Metrology, China April 2017 in Vienna, Austria Organization Founded in 1955, National Institute

More information

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution NUKLEONIKA 2007;52(4):153 158 ORIGINAL PAPER Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution Tomasz

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch

More information

PREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY

PREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY PREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY IDENTIFICATION NUMBER: ANS-RT-PROTON-01 BENCHMARK CLASSIFICATION: Radiation Therapy BENCHMARK

More information

Manik Aima, Larry A. DeWerd, Wesley S. Culberson

Manik Aima, Larry A. DeWerd, Wesley S. Culberson Manik Aima, Larry A. DeWerd, Wesley S. Culberson University of Wisconsin Medical Radiation Research Center, Madison, WI 25 th Annual Meeting of the Council of Ionizing Radiation Measurements and Standards,

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry

Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Indigenously developed multipurpose acrylic head phantom for verification of IMRT using film and gel dosimetry N. Gopishankar, 1a

More information

Dosimetric characterization of surface applicators for use with the Xoft ebx system

Dosimetric characterization of surface applicators for use with the Xoft ebx system Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Evaluation of a single-scan protocol for radiochromic film dosimetry

Evaluation of a single-scan protocol for radiochromic film dosimetry JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 2, 2015 Evaluation of a single-scan protocol for radiochromic film dosimetry Yoshinobu Shimohigashi, 1a Fujio Araki, 2 Masato Maruyama, 1

More information