Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance
|
|
- Nathan Strickland
- 5 years ago
- Views:
Transcription
1 Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany
2 Existing methods used in dosimetry 1-dimensional 2-dimensional Films TL pellet diamond OSL ionization chamber semiconductor diode 3-dimensional Gafchromic Radiographic TLD foils matrix of ionisation chambers strip and pixel semiconductor detectors Fricke gel 3-D scintillators TLD personal badge Strip semiconductor detector 3-D Scintillator
3 Presentation plan Principle of operation 2D Thermoluminescence System TL Foils Readers Dosimetric properties 2D TL system for broad beam Quality Assurance TL foils and LET-painting technique Summary and plans for future
4 The principle of 2-D TL Dosimetry Preparation of foil Irradiation Heating Collection of emitted light Data analysis
5 Dosimetric system TLD foils Two types of foils: LiF:Mg,Cu,P (MCP-2D) CaSO :Dy 4 Water resistance and flexibility Up to 20 x 20 cm2 Reusability Zeff MPC-2D CaSO Resolution below 0.1 mm2 Linearity of dose response: Gy
6 Dosimetric system hardware Laboratory Reader: detector size 50 x 50 mm2 resolution 640 x 480 px2 fully adjustable Laboratory reader Clinical Reader: detector size 200 x 200 mm2 resolution 1024 x 1024 px2 easy and safe to use Clinical reader
7 Dosimetric properties uniformity 20 cm After corrections the uniformity of readouts is around 2%
8 Dosimetric properties dose response f(d) f(d) Linearity Index D was constant within ±5% measured with the Laboratory Reader. For the Clinical Reader it is ±2% Laboratory Reader Clinical Reader f(d) D ±2% f(d) D
9 Dosimetric properties Reproducibility More than 20 equivalent irradiations and readouts Repeatability was found below 2% For previous reader and software it was 5% For radiochromic films it is 1-6% [1] McLaughlin et al. Methods Phys. Res. A 302 [2] N. V. Klassen et al. Med. Phys. 24
10 Beam cross-section in PMMA phantom Diode TLD Irradiations have been performed with 60 MeV proton beam at IFJ PAN in Kraków (AIC-144 cyclotron)
11 Profile of proton beam PBP Experimental setup Beam 25 mm collimator PMMA plates TLD foil Pristine Bragg Peak
12 Profile of proton beam SOBP Experimental setup Beam SOBP wheel PMMA plates TLD foil 25 mm collimator Spread Out Bragg Peak
13 Profile of proton beam summary About 100 beam cross-sections have been measured Measured flatness of dose profiles is in 5-12% range (for diode it is 2-3%) Penumbra is broader with deeper location in phantom (difficult to measure with diode) Experiments conducted at IFJ PAN in Kraków show that 2D TLD system is a good tool for proton beam Quality Assurance Dose profiles of beam measured with TL foils and photodiode are comparable 3D visualisation of a cross-section of a proton broad beam
14 MCP-2D relative efficiency η TL response depends not only on dose but also on ionization density.
15 LET-painting with TL foils Biological response depends on ionisation density. Determination of LET is very important for ion radiotherapy. The idea Set of two foils: MCP-2D + CaSO4:Dy Beam MCP-2D CaSO4:Dy Difference in relative efficiency Energy response for CaSO4:Dy foils needs to be verified
16 Plans for future Incoming experiments (HIT-Heidelberg) Energy response for 12 C and p+ ions Tests of QA of scanning beam Calibration curve for dose/let dosimetry Validation of the dose/let response in clinical conditions Other TL foils properties: transparency, angular response, fading, etc. 2D TLD system will be used for QA of therapeutic scanning beam at Bronowice Cyclotron Center in Kraków (C-235 IBA cyclotron)
17 Thank You for Your attention
Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance
Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance J. Gajewski1, M. Kłosowski1, S. Greilich2, P. Olko1 Institute of Nuclear Physics, Kraków, Poland German Cancer
More informationA comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV
A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical
More informationDependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose
Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationResponse evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry
Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba
More informationRadiochromic film dosimetry in water phantoms
INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms
More informationRadiation Monitoring Instruments
Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers
More informationActive Scanning Beam 3 Checking Delivery/Dosimetry
Active Scanning Beam 3 Checking Delivery/Dosimetry C Algranati, J Salk, A Coray, A Lomax, E Pedroni, T Boeringer, S Lin, E Hug Center for Proton Radiation Therapy Checking Delivery/Dosimetry Contents Absolute
More informationTHERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL
THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL G. Gambarini Dipartimento di Fisica dell Università, Milano, Italy e-mail grazia.gambarini http://users.unimi.it/~frixy/
More informationDosimetry and QA of proton and heavier ion beams
Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference
More informationProtons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.
Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic
More informationCHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA
CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy
More informationLearning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan
Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationMAESTRO: Methods and Advanced Equipment for Simulation and Treatment in Radiation Oncology
What is MAESTRO? Sixth Framework Program FP6 (2002-2006) European Commission Integrated Project Duration: 5 years MAESTRO: Methods and Advanced Equipment for Simulation and Treatment in Radiation Oncology
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department
More informationCOMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE
COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE I. Petrovi a, A. Risti -Fira a, L. Kori anac a, J. Požega a, F. Di Rosa b, P. Cirrone b and G. Cuttone
More informationCalibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters
Calibration of 90 Sr+ 90 Y sources used for betatherapy, using a postal kit of thermoluminescent dosimeters Patrícia de Lara Antonio 1*, Rogério M. V. Silva 2, Divanizia do Nascimento Souza 2 and Linda
More informationLuminescent materials for dosimetric applications
Luminescent materials for dosimetric applications Adrie J.J. Bos Delft University of Technology, The Netherlands Delft University of Technology Challenge the future Basic assumption: There is a relation
More informationNUCLEAR PHYSICS FOR MEDICINE - HADRON THERAPY. Pawel Olko Institute of Nuclear Physics Krakow Poland
NUCLEAR PHYSICS FOR MEDICINE - HADRON THERAPY Pawel Olko Institute of Nuclear Physics Krakow Poland The NuPECC report: Nuclear Physics in Medicine http://www.nupecc.org/pub/npmed2014.pdf Outline 1. Why
More informationA preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF
Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital
More informationChapter 4: Radiation Monitoring Instruments
Chapter 4: Radiation Monitoring Instruments Set of 107 slides based on the chapter authored by G. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Review of Radiation Oncology Physics: A
More informationRadiologic Units: What You Need to Know
Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationRADIATION MONITORING DEVICES R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S
RADIATION MONITORING DEVICES 10-526- 1 9 7 R A D I A T I O N P R O T E C T I O N & B I O L O G Y - R H O D E S DETECTION AND MEASUREMENT OF IONIZING RADIATION Dosimeter Dose-measuring device Two classifications:
More informationDosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar
More informationProton Irradiation for Space Research at IFJ PAN Kraków Pawel Olko and Jan Swakoń Institute of Nuclear Physics IFJ PAN
Proton Irradiation for Space Research at IFJ PAN Kraków Pawel Olko and Jan Swakoń Institute of Nuclear Physics IFJ PAN Outline 1. Calibration needs for space dosimetry 2. Photons at IFJ 3. Protons at IFJ
More informationMedical dosimetry using Optically Stimulated Luminescence dots and microstar readers
Medical dosimetry using Optically Stimulated Luminescence dots and microstar readers Christopher A Perks *a, Cliff Yahnke b, Marc Million c a LANDAUER EUROPE, Unit 28, Bankside, Station Approach, Kidlington,
More informationMeasurements of Air Kerma Index in Computed Tomography: A comparison among methodologies
Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationPeak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam
Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch
More informationY FILMS DOSIMETR Nederland België / Belgique
DOSIMETRY FILMS GAFCHROMIC Dosimetry Films The Self-developing Dosimetry Films that Allow You to Go Filmless Do away with the cost and headache of calibrating and maintaining a processor Do away with hazardous
More informationTLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience
IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More information"Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 : Mn DOSIMETER FOR γ- IRRADIATION HIGH-DOSE DOSIMETRY *
Iranian Journal of Science & Technology, Transaction A, Vol. 32, No. A3 Printed in the Islamic Republic of Iran, 2008 Shiraz University "Research Note" CHARACTERIZATION OF A NEW PREPARED Li 2 B 4 O 7 :
More informationAbsorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry
Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,
More informationCharacterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source
Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor
More informationEUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY
: ~ : ; EUROPEAN ORGANZATON FOR NUCLEAR RESEARCH 20 1983 TS-RP/ 113/ CF TS DVSONAL REPORT September 1983 A COMBNATON OF TLD ALBEDO AND SULPHUR ACTVATON TECHNQUES FOR FAST NEUTRON PERSONNEL DOSMETRY J.W.N.
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationNeutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters
133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches
More informationChapter 4 RADIATION MONITORING INSTRUMENTS
Chapter 4 RADIATION MONITORING INSTRUMENTS G. RAJAN Medical Physics and Safety Section, Bhabha Atomic Research Centre, Mumbai, Maharashtra, India J. I Z E W S K A Division of Human Health, International
More informationMEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS
RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training
More informationA comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy
Radiology and Oncology Ljubljana Slovenia www.radioloncol.com research article 307 A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy Jaeman Son 1,2,
More informationPREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY
PREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY IDENTIFICATION NUMBER: ANS-RT-PROTON-01 BENCHMARK CLASSIFICATION: Radiation Therapy BENCHMARK
More informationDosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology
Kantonsspital Luzern Dosimetry Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 Introduction Metrological traceability; the dosimetry chain Absolute / relative
More informationMania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane
Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors
More informationRadPro International GmbH i
RadPro International GmbH i...radiation Protection for the Radiation Professionals... Your provider for complete TLD systems We supply full spectrum of TLD-readers, TLD-materials, dedicated equipment and
More informationVerification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging
Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2
More informationBasic dosimetric data for treatment planning system
Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny
More informationGood manufacturing practices (GMP) utilized on human blood irradiation process
Artigos Good manufacturing practices (GMP) utilized on human blood irradiation process Cláudio Boghi Pesquisador do IPEN-USP. cboghi@uol.com.br; cboghi@usp.br Dorlivete M. Shitsuka Professora de Organização
More informationEye plaque dosimetry verification using novel solid state devices
University of Wollongong Research Online University of Wollongong Thesis Collection University of Wollongong Thesis Collections 2014 Eye plaque dosimetry verification using novel solid state devices Talia
More informationIORT with mobile linacs: the Italian experience
IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams
More informationNEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM. Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients
NEW PERFORMANCE STANDARD: GAFCHROMIC FILM BASED DOSIMETRY SYSTEM Xiang Yu and Andre Micke Advanced Materials Group Ashland Specialty Ingredients IDEAL DOSIMETRY FOR IMRT/SRS Absolute dose measurement Precise
More informationCharacterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning
Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca
More informationDosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor
Dosimetric Characteristics of LiF:Mg,Cu,Na,Si Phosphor Y. M. Nam 1, J. L. Kim 1, S. Y. Chang 1, B. H. Kim 1 and H. J. Kim 2 1 Health Physics Department, Korea Atomic Energy Research Institute, Taejon 305-353
More informationFollow this and additional works at: Part of the Medicine and Health Sciences Commons
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 8-2014 Development and Implementation of an Anthropomorphic Pediatric
More informationCalibration of two 90 Sr+ 90 Y dermatological applicators
Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,
More informationMultilayer Gafchromic film detectors for breast skin dose determination in vivo
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination
More informationRPC s Credentialing Programs for Clinical Trials
RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationImplementation of an Anthropomorphic Phantom for the Evaluation of Proton Therapy Treatment Procedures
Texas Medical Center Library DigitalCommons@TMC UT GSBS Dissertations and Theses (Open Access) Graduate School of Biomedical Sciences 5-2010 Implementation of an Anthropomorphic Phantom for the Evaluation
More informationNeutron dose evaluation in radiotherapy
Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head
More informationDosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for Clinical Electron Beam
Journal of the Korean Physical Society, Vol. 55, No. 6, December 2009, pp. 2566 2570 Dosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for Clinical Electron Beam Jin-Beom
More informationMedical applications round table. Radiotherapy Diagnostic radiology Nuclear medicine Radiation protection
Medical applications round table Radiotherapy Diagnostic radiology Nuclear medicine Radiation protection Will be discussing Ion beam therapy/radiotherapy Diagnostic radiology Nuclear medicine Radiation
More informationAcrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 1, WINTER 2010 Acrylonitrile Butadiene Styrene (ABS) plastic-based low cost tissue equivalent phantom for verification dosimetry in IMRT Rajesh
More informationProton and heavy ion radiotherapy: Effect of LET
Proton and heavy ion radiotherapy: Effect of LET As a low LET particle traverses a DNA molecule, ionizations are far apart and double strand breaks are rare With high LET particles, ionizations are closer
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationNeutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics
Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,
More informationProcess Control Methods in Radiation Technologies
Process Control Methods in Radiation Technologies András Kovács Nuclear Security Department Centre for Energy Research Hungarian Academy of Sciences Radiation Processing Applications Radiation sterilization:
More informationPHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)
PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationThe clinical use of OSLD
The clinical use of OSLD Stephen F. Kry PhD Radiological Physics Center, Imaging and Radiation Oncology Core - Houston IROC Houston The University of Texas MD Anderson Cancer Center, Houston, TX Spring
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationBasic definitions. Dosimetry, radiation protection. Nuclear measurement techniques. Interaction of the nuclear radiation with the matter
Dosimetry, radiation protection. Nuclear measurement techniques. properties measurement dosimetry medical applications of the nuclear radiation Basic definitions Nuclear radiation: Produced in the transition
More informationAdvanced film dosimetry for a UK brachytherapy audit
Advanced film dosimetry for a UK brachytherapy audit Tony Palmer Lead Scientist Medical Physics, Portsmouth Hospitals NHS Trust, UK University of Surrey, UK Institute of Physics and Engineering in Medicine
More informationROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning
More informationEPID dosimetry confi guration and pre-treatment IMRT verifi cation
Rep Pract Oncol Radiother, 2007; 12(6): 307-312 Original Paper Received: 2007.04.26 Accepted: 2007.08.29 Published: 2007.12.27 Authors Contribution: A Study Design B Data Collection C Statistical Analysis
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationPRESCRIBING RECORDING AND REPORTING PROTON BEAM THERAPY ICRU 78. RAJESH THIYAGARAJAN Senior Medical Physicist & RSO Medanta The Medicity
PRESCRIBING RECORDING AND REPORTING PROTON BEAM THERAPY ICRU 78 RAJESH THIYAGARAJAN Senior Medical Physicist & RSO Medanta The Medicity HISTORY OF PROTON THERAPY Robert Wilson proposed the use of proton
More informationOut-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison
MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison
More informationStefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.
Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn
More informationPROGRESS IN HADRONTHERAPY
PROGRESS IN HADRONTHERAPY Saverio Braccini TERA Foundation for Oncological Hadrontherapy IPRD06 - Siena - 01.10.06 - SB 1 Outline Introduction Radiation therapy with X rays and hadrontherapy Hadrontherapy
More informationRADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE
RADIATION MONITORING EXPERIMENT USING THERMOLUMINESCENT DOSIMETER FOR THE TR 19 CYCLOTRON AREA IN NUCLEAR RESEARCH INSTITUTE A. STOCHIOIU, L. CRACIUN, F. MIHAI, I. TUDOR Horia Hulubei National Institute
More informationCalculated LET spectrum from antiproton beams stopping in water
Acta Oncologica, 2009; 48: 223226 ORIGINAL ARTICLE Calculated LET spectrum from antiproton beams stopping in water NIELS BASSLER 1,2 & MICHAEL HOLZSCHEITER 3 1 Department of Experimental Clinical Oncology,
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationIMRT QA: Point Dose Measurements or 2D Array?
QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationJRPR. Quenching Effect in an Optical Fiber Type Small Size Dosimeter Irradiated with 290 MeV u -1 Carbon Ions. Original Research.
Journal of Radiation Protection and Research 26;4(3):222-228 pissn 258-888 eissn 2466-246 Quenching Effect in an Optical Fiber Type Small Size Dosimeter Irradiated with 29 MeV u - Carbon Ions Yuho Hirata,
More informationMeasurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)
Journal of Physics: Conference Series PAPER OPEN ACCESS Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) To cite this
More informationImplementation of personnel dosimetry systems using thermoluminescence dosimetry
Implementation of personnel dosimetry systems using thermoluminescence dosimetry J. C. Rojas 1, S. O. Benavides 1,2, and H. Olaya 3 1 Radiation Safety Committee, Fundación Clínica Valle del Lili, A.A.
More informationEighth Annual Chapman Phytosanitary Irradiation Forum. Dosimetry for phytosanitary applications: dos and don ts. Florent KUNTZ
Eighth Annual Chapman Phytosanitary Irradiation Forum Dosimetry for phytosanitary applications: dos and don ts Florent KUNTZ June 14 th, 2018 What does a dosimeter measure? Question A dosimeter measures
More informationMEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn
MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)
More informationSpecific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)
Azam Niroomand-Rad: General Aspects of RCF (TG-235), AAPM July 2015 1 Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad, PhD, FAAPM,
More informationCanadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Major Dosimetry Equipment. A guidance document on behalf of:
Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Major Dosimetry Equipment A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationCALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE
CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of
More information