Physics update for radiation oncologists (2)

Size: px
Start display at page:

Download "Physics update for radiation oncologists (2)"

Transcription

1 Physics update for radiation oncologists (2) Scott Crowe, Alastair Dick, Alex Livingstone, Craig Lancaster, Darren Cassidy, Kirby Francis, Nancy Yu, Phil Back, Rachael Wilks, Samuel Peet, Steven Sylvander, Tanya Kairn Cancer Care Services, Radiation Oncology Medical Physics

2 Overview 1. Service delivery 2. Staff, interns and students 3. Research output 4. What s on the horizon? 2

3 Service delivery 3

4 Acuros XB Current dose calculation engine in Eclipse is AAA (type-b algorithm) Errors in dose calculated both in and downstream of lung have been observed (and reported by others, such as ACDS) New dose calculation engine in Eclipse is Acuros XB (type-c algorithm) Studies have reported improved accuracy and better agreement with Monte Carlo dose calculations olung VMAT in simple inhomogeneous phantom: AAA up to 2.7% disagreement, Acuros XB within 0.8% (Yan et al.) onasopharyngeal IMRT in anthropomorphic phantom: AAA up to 10% disagreement, Acuros XB within 3% (Kan et al.) Allows reporting of dose as dose-to-water (like earlier TPS algorithms) and dose-to-medium (like Monte Carlo) 4

5 Acuros XB We are commissioning based on international guidelines (IAEA/AAPM): entering beam to allow the system to accurately calculate the dose in water System does this by optimising various source parameters (energy spectrum, penumbra, etc.) Model is being tested across a range of treatment and phantom complexity (square fields > small modulated fields; water tanks > anthropomorphic phantoms) Commissioning is being completed under supervision of Rachael 5

6 Varian Portal Dosimetry Many centres across Australia have switched to EPID-based pretreatment patient-specific quality assurance systems e.g. Varian Portal Dose, Epiqa, Adaptivo, PerFraction Benefits include high resolution, reduction in QA time and improved workflow integration Downsides include lack of gantry angle verification, system potentially not testing the same dose calculation approach used to calculate dose in patient This system is currently being commissioned and tested under supervision of Nancy 6

7 Gel Dosimetry Gel dosimeters change optical properties with ionisation We are able to measure 3D dose distributions, with tremendous potential for verification of, e.g., SABR treatments with high dose gradients, or treatments of multiple mets 7

8 Recent international reports Physics will be reviewing (in conjunction with RTs) the contents of a couple of new reports, with reference to local practice ICRU 91: Prescribing, recording, and reporting of stereotactic treatments with small photon beams Tanya and I have a paper cited in this report! AAPM TG132: Use of image registration and fusion algorithms and techniques in radiotherapy Discusses image registration errors, testing procedures 8

9 Staff, interns and students 9

10 Staffing changes A/Prof Tanya Kairn has joined the physics team, as principal physicist (role previously filled by Bob Fitchew) Tanya previously worked at Genesis Wesley, where she was responsible for stereotactic physics service area and, most recently, the commissioning of a TrueBeam system for use in SRS/SBRT Strong research track record including 78 peer reviewed papers, ARC Linkage grant success and PhD supervision completions Standing includes adjunct A/Prof appointment at QUT, associate editorship at JMIRO & APESM, Boyce Worthley ACPSEM award (2016) Diana Binny has left for ROC Redlands (though her PhD candidature continues with RBWH supervisors) 10

11 2018 internships During 2017 we have had two paid internships in the department: Candice Milewski (Université Paris-Saclay MSc student) and Somayeh Zolfaghari (QUT MSc graduate) Projects included metal artefact reduction techniques for CT systems (which we expect to have access to in the department soon) and dosimetric film preparation (published) Funded in part by RBWH Foundation scholarship for Samuel Peet Samuel Peet and Nancy Yu were both successful in securing scholarships for funding in 2018! These funds will be used to bring in interns once again (for which there are many suitable candidates). 11

12 Students in 2017 (Blue / bold = RBWH) Andre Asena PhD (conferred) Dose distributions in the vicinity of high-density materials in radiotherapy Diana Binny PhD Optimizing patient plan deliveries by characterizing TPSs in radiotherapy Orrie Dancewicz PhD Development and verification of a novel 3D dosimetry technique for tomotherapy Johnny Morales PhD Independent Monte Carlo analysis for stereotactic radiosurgery treatments Samuel Peet PhD Out-of-field dose in contemporary radiation therapy James Rijken PhD Optimisation and uncertainties in SABR spine treatment Shaun Smith PhD (conferred) Development of three-dimensional dosimetry for radiotherapy Nancy Yu PhD Developing new methods of dose evaluation for VMAT patient specific plans Abdullah Alharbi MSc Developing patient specific quality assurance for spine SABR plans David Garrott MSc Monte Carlo modelling of electron beam output for irregular cut-outs Dylan McGrath MSc Commissioning of the Varian Portal Dosimetry (VPD) QA solution Candice Milewski MSc Metal artefact reduction algorithms in CT imaging Emma Whittle Volunteer Commissioning an optical CT solution for gel dosimetry 12

13 Research and publications 13

14 Peer reviewed papers in 2017 (Blue / bold = RBWH; Grey = collaborative) C Chamunyonga, S Crowe, J Burbery, Review of the clinical benefits and implementation of peer review of treatment plans in undergraduate medical dosimetry and radiation therapy training, Journal of Radiotherapy in Practice 16(1): 85-91, 2017 A Asena, ST Smith, T Kairn, SB Crowe, RD Franich, JV Trapp Reduction of artefacts caused by missing ray-sum data in optical-ct imaging of implants in gel dosimeters, Journal of Physics: Conference Series 847: , 2017 T Kairn, A Asena, SB Crowe, A Livingstone, D Papworth, S Smith, B Sutherland, S Sylvander, RD Franich, JV Trapp, Can a commercial gel dosimetry system be used to verify stereotactic spinal radiotherapy treatment dose distributions?, Journal of Physics: Conference Series 847: , 2017 A Asena, ST Smith, T Kairn, SB Crowe, K Hosokawa, SR Sylvander, JV Trapp, Dose distributions in the vicinity of high-density implants using 3D gel dosimeters, Medical Physics 44(4): , 2017 D Binny, E Mezzenga, CM Lancaster, JV Trapp, T Kairn, SB Crowe, Investigating output and energy variations and their relationship to delivery QA results using Statistical Process Control for helical tomotherapy, Physica Medica 38: , 2017 O Dancewicz, SR Sylvander, TS Markwell, SB Crowe, JV Trapp, Radiological properties of 3D printed materials in kilovoltage and megavoltage photon beams, Physica Medica 38: , 2017 S Zolfaghari, KE Francis, T Kairn, SB Crowe, Commissioning a hobby cutting device for radiochromic film preparation, Australasian Physical and Engineering Sciences in Medicine 40(2): , 2017 SK Nabankema, SM Jafari, SC Peet, D Binny, SR Sylvander, SB Crowe, Wearable glass beads for in vivo dosimetry of total skin electron irradiation treatments, Radiation Physics and Chemistry 140: , 2017 D Binny, CM Lancaster, JV Trapp, SB Crowe, Statistical Process Control and verifying positional accuracy of a cobra motion couch using step-wedge quality assurance tool, Journal of Applied Clinical Medical Physics 18(5): 70-79, 2017 B Perrett, P Charles, T Markwell, T Kairn, S Crowe, Feasibility of 3D printed air slab diode caps for small field dosimetry, Australasian Physical and Engineering Sciences in Medicine (in press), 2017 T Aland, E Jhala, T Kairn, J Trapp, Film dosimetry using a smart device camera: a feasibility study for point dose measurements, Physics in Medicine and Biology (in press), 2017 Textbook chapters in 2017 T Kairn. Application of Radiochromic Film to IMRT and VMAT Dosimetry, Radiochromic Film for Radiation Dosimetry. Ed: Indra Das. Abingdon, UK: Taylor and Francis, T Kairn, J Lehmann, S Crowe et al. Postal and clinical trial dosimetry, Radiochromic Film for Radiation Dosimetry. Ed: Indra Das. Abingdon, UK: Taylor and Francis,

15 Output measures Aiming to publish at least 12 papers a year (met in 15 and 16) Other papers in review or being drafted include Use of gel dosimeters for brachytherapy Dose corrections for Fricotan moulds Film dosimetry method comparison Eclipse optimiser behaviour Out of field dosimetry for Tomotherapy History of Australas Phys Eng Sci Med Dose errors in BrainLab SRS calculations Software development and testing in the clinic 15

16 Recent research results Animation, not for PDF Sliding window IMRT treatment being delivered to tonic water: quinine fluoresces in response to ionising radiation Captured with camera: potential uses for high-dose-rate brachy QA, tomographic dose reconstruction in 4D, superficial dose measurement 16

17 What is on the horizon? New systems to be commissioned: superficial, brachytherapy and CT New functionality will be tested on CT system, including: odirect density assignment, in which CT numbers are adjusted for consistency across scanning energies, allowing simpler use of multiple scanning energies odual energy X-ray, allowing improved contrast imaging and more accurate material identification (with potential implications on Monte Carlo dose simulations) ometal artefact reduction algorithms, on which we are currently preparing a scientific manuscript for publication 17

18 What is on the horizon? Nancy and I have organised a school on IGRT systems (including Catalyst, Calypso, ExacTrac, ultrasound) for Queensland physicists tomorrow; at which Sam will be presenting 7 presentations and 2 posters selected for national Engineering and Physical Sciences in Medicine conference in Hobart (29/10 01/11): four resulting from student supervision Presentations at World Congress in Prague next year planned 2 of us are taking part in a Future Medical Physics Rethinking Cancer forum in October NHMRC grant application for small field radiotherapy planned rodash to be presented by Sam 18

Physics update for radiation oncologists

Physics update for radiation oncologists Cancer Care Services, Radiation Oncology Medical Physics Physics update for radiation oncologists Scott Crowe, Alastair Dick, Alex Livingstone, Bob Fitchew, Candice Milewski, Craig Lancaster, Darren Cassidy,

More information

Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST

Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST What is a Medical Physicist? Medical physics application of physics concepts

More information

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

Guideline & Reports 医学物理学会教育委員会資料

Guideline & Reports 医学物理学会教育委員会資料 Guideline & Reports 医学物理学会教育委員会資料 2017/11 更新 report title year keyword 1 keyword 2 AAPM TG211 Classification and evaluation strategies of auto-segmentation approaches for PET 2017 PET autosegmentation

More information

Verification of Advanced Radiotherapy Techniques

Verification of Advanced Radiotherapy Techniques Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December

More information

Future upcoming technologies and what audit needs to address

Future upcoming technologies and what audit needs to address Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital

Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital Image Guided in Radiation Therapy (IGRT) Chumpot Kakanaporn Med Phys Radiation Oncology Siriraj Hospital EBT Process Diagnosis Simulation Tx Planning Tx Verification Tx Delivery X-ray CT MRI NM Patho Process

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Abstract ID 352 Stereotactic radiosurgery for multiple brain metastases: A dose-volume study Tanya Kairn 123, Somayeh Zolfaghari 1, Daniel Papworth

Abstract ID 352 Stereotactic radiosurgery for multiple brain metastases: A dose-volume study Tanya Kairn 123, Somayeh Zolfaghari 1, Daniel Papworth Abstract ID 352 Stereotactic radiosurgery for multiple brain metastases: A dose-volume study Tanya Kairn 123, Somayeh Zolfaghari 1, Daniel Papworth 2, Mark West 2, David Schlect 2, Scott Crowe 13 1 Queensland

More information

Establishing the impact of temporary tissue expanders on

Establishing the impact of temporary tissue expanders on Establishing the impact of temporary tissue expanders on electron and photon beam dose distributions A. Asena 1,*, T. Kairn 1,2, S.B. Crowe 1 and J.V Trapp 1 1 School of Chemistry, Physics and Mechanical

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe 3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

SunCHECK Patient Comprehensive Patient QA

SunCHECK Patient Comprehensive Patient QA The SunCHECK Platform SunCHECK Patient Comprehensive Patient QA SunCHECK Patient // Comprehensive Patient QA Automate Your Patient QA SunCHECK Patient automates all your Patient QA needs from Secondary

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

Unrivaled, End-to-End

Unrivaled, End-to-End PHANTOMS Unrivaled, End-to-End Stereotactic QA Industry-leading 0.1mm accuracy minimizes errors at each link in the stereotactic quality assurance chain. Stereotactic radiosurgery (SRS) is governed by

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of Medicine Indianapolis, IN, USA Indra J. Das,

More information

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose

Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

I. Equipments for external beam radiotherapy

I. Equipments for external beam radiotherapy I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),

More information

ph fax

ph fax This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

SBRT Credentialing: Understanding the Process from Inquiry to Approval

SBRT Credentialing: Understanding the Process from Inquiry to Approval SBRT Credentialing: Understanding the Process from Inquiry to Approval David Followill, Ph.D. (presented by Andrea Molineu, M.S.) IROC Houston QA Center July 15, 2015 What is credentialing? Verification

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques

Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques TROG 08.03 RAVES Quality assurance and credentialing requirements for sites using inverse planned IMRT Techniques Introduction Commissioning and quality assurance of planning systems and treatment delivery

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center

Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center Medical Dosimetry Graduate Certificate Program IU Graduate School & The Department of Radiation Oncology IU Simon Cancer Center All students accepted into the Medical Dosimetry Graduate Certificate Program

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an

More information

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session Stereotactic Body Radiation Therapy Quality Assurance Educational Session J Perks PhD, UC Davis Medical Center, Sacramento CA SBRT fundamentals Extra-cranial treatments Single or small number (2-5) of

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim

More information

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed

More information

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT Purpose The purpose of this curriculum outline is to provide a framework for multidisciplinary training for radiation oncologists, medical

More information

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.

Technical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou. Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution

More information

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System Original Article Research in Oncology 2018; Vol. 14, No. 1: 34-38. DOI: 10.21608/resoncol.2018.3026.1047 Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

More information

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...

More information

Electron Beam ET - Reloaded Therapy - Reloaded

Electron Beam ET - Reloaded Therapy - Reloaded 8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles

More information

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method

A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 4, 2014 A feasibility study of multislice X-ray CT imaging of gel dosimeters using the zero scan method Muhammad B. Kakakhel, 1 Tanya Kairn,

More information

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS ICTP SCHOOL ON MEDICAL PHYSICS FOR RADIATION THERAPY DOSIMETRY AND TREATMENT PLANNING FOR BASIC AND ADVANCED APPLICATIONS March

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy

Introduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction

More information

DELIVERED PATIENT DOSE

DELIVERED PATIENT DOSE SOFTWARE DELIVERED PATIENT DOSE The only software that provides a true picture of patient and dose at each stage of treatment. Adaptivo patient dosimetry software does something other patient dosimetry

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

New Radiation Treatment Modalities in the Treatment of Lung Cancer

New Radiation Treatment Modalities in the Treatment of Lung Cancer New Radiation Treatment Modalities in the Treatment of Lung Cancer David Perry, M.D. Chief, Radiation Oncology Medical Director, CyberKnife Radiosurgery Center Medstar Franklin Square Medical Center Definitions

More information

Patient-Specific QA & QA Process. Sasa Mutic, Ph.D. Washington University School of Medicine

Patient-Specific QA & QA Process. Sasa Mutic, Ph.D. Washington University School of Medicine Patient-Specific QA & QA Process Sasa Mutic, Ph.D. Washington University School of Medicine Outline of Presentation QA and QC in RT Safety vs. quality Patient specific QM program Learning Objectives Describe

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom

The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom The Accuracy of 3-D Inhomogeneity Photon Algorithms in Commercial Treatment Planning Systems using a Heterogeneous Lung Phantom Gary Fisher, B.S. David Followill, Ph.D. Geoffrey Ibbott, Ph.D. This investigation

More information

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar

More information

International Practice Accreditation

International Practice Accreditation International Practice Accreditation Program endorsed and supported by Jean-Xavier Hallet Punta Cana, Nov 2017 Short History Created in 2004 by ESTRO, independent since 2009 Dosimetry Laboratory (IGR):

More information

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a

Albert Lisbona Medical Physics Department CLCC Nantes Atlantique a The Equipment : safety, commissioning, re commissioning calibration : Equipements : sûreté, recette, modification calibration A medical physicist point of view. Albert Lisbona Medical Physics Department

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research

More information

Overview of TG225 - Medical Physics Practice Guideline #1

Overview of TG225 - Medical Physics Practice Guideline #1 Overview of TG225 - Medical Physics Practice Guideline #1 Evaluation and quality assurance of x-ray based image guided radiotherapy systems Jonas Fontenot, Ph.D. MPPG #1 Evaluation and Quality Assurance

More information

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient

8/3/2016. The EPID Strikes Back! - EPID In-Vivo Dosimetry. EPID Research Number of Publications. Why EPID in-vivo? Detectable errors: patient Number of Publications Number of publications 8/3/2016 The Strikes Back! - In-Vivo Dosimetry AAPM, Washington D.C, USA, 2016 Peter Greer 1,2 (1) University of Newcastle, Australia, (2) Calvary Mater Newcastle,

More information

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D. ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS

STEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic

More information

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments

Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor

More information

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times

Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee

More information

Implementation of advanced RT Techniques

Implementation of advanced RT Techniques Implementation of advanced RT Techniques Tibor Major, PhD National Institute of Oncology Budapest, Hungary 2. Kongres radiološke tehnologije, Vukovar, 23-25. September 2016. Current RT equipments at NIO,

More information

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc

Evaluation of Dynamic Delivery Quality Assurance Process for Internal Target Volume Based RapidArc Original Article PMP Progress in Medical Physics 28(4), December 217 https://doi.org/1.14316/pmp.217.28.4.181 pissn 258-4445, eissn 258-4453 Evaluation of Dynamic Delivery Quality Assurance Process for

More information

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases CASE STUDY Institution: Odette Cancer Centre Location: Sunnybrook

More information

Credentialing for the Use of IGRT in Clinical Trials

Credentialing for the Use of IGRT in Clinical Trials Credentialing for the Use of IGRT in Clinical Trials James M. Galvin, DSc Thomas Jefferson University Hospital Jefferson Medical College Philadelphia, PA and The Radiation Therapy Oncology Group RADIATION

More information

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist

The Radixact System Experience at Miami Cancer Institute. Tino Romaguera, D.Sc. Senior Physicist The Radixact System Experience at Miami Cancer Institute Tino Romaguera, D.Sc. Senior Physicist Disclosure An honorarium is provided by Accuray for this presentation The views expressed in this presentation

More information

Origin, funding & collaboration

Origin, funding & collaboration UK Brachytherapy Audit: Origin, funding & collaboration Tony Palmer OVERVIEW Purpose From idea to implementation Methodology, results & reflection Proposal for a Brachytherapy Audit Modern brachytherapy

More information

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this

More information

Measuring radiation dose through the detection of radiation-induced acoustic waves

Measuring radiation dose through the detection of radiation-induced acoustic waves Measuring radiation dose through the detection of radiation-induced acoustic waves Susannah Hickling a, Hao Lei b, Maritza Hobson a, Issam El Naqa a,b a McGill University, Montreal, Canada b University

More information

Hong Kong Association of Medical Physics. Guide to Candidates

Hong Kong Association of Medical Physics. Guide to Candidates Hong Kong Association of Medical Physics Certification of Medical Physicists Professional Assessment Guide to Candidates PA App Form (r2017.1) Contents Chapter 1 p. 3 Professional Assessment System and

More information

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom

Film-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with

More information

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL

Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts New calorimeters Simon Duane, NPL Calorimetry remains the most fundamental method for the measurement of absorbed dose and is the basis of almost

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients 3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients G Asuni *1, T vanbeek 1, E VanUtyven 1, P McCowan 1,2, and B.M.C. McCurdy 1,2,3 1 Medical Physics Department, CancerCare

More information

Image Guided Stereotactic Radiotherapy of the Lung

Image Guided Stereotactic Radiotherapy of the Lung Image Guided Stereotactic Radiotherapy of the Lung Jamie Marie Harris, MS DABR Avera McKennan Radiation Oncology September 25, 2015 Stereotactic Body Radiotherapy - Clinical Dose/Fractionation - Normal

More information

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Optical Scanning 3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Wuu, C.-S., Hoogcarspel, S.J.,Deh, K., Hsu, W.-Y., Adamovics, J. 2013 Journal of Physics: Conference Series

More information