Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Size: px
Start display at page:

Download "Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution"

Transcription

1 Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi István University, Department of Physics and Chemistry, Egyetem tér 1., H- 9026, Gyır, Hungary. b National Centre for Healthcare Audit and Inspection, Váci u. 174., H-1138 Budapest, Hungary. c International Atomic Energy Agency, Division of Human Health, Department of Nuclear Sciences and Applications, Wagramer Strasse 5, A-1400 Vienna, Austria. Abstract. Medical ionization radiation sources give by far the largest contribution to the population dose from man-made sources and most of this contribution comes from diagnostic x-rays. Radiologist constantly face the dilemma of trying to minimize patients exposure whenever possible, while still using exposures that are high enough to produce images of good quality to be able to provide proper diagnosis. Quality assurance provides a framework for achieving this goal. As the part of the QA, there is a need to control patient doses arising from x- ray diagnostic activity. Due to the increased demand for dosimetry measurements in diagnostic and interventional radiology, it has become important to provide traceability of measurements, which can only work satisfactorily if correct calibrations and measurements are performed. In order to advice Secondary Standard Laboratories and end-users in hospitals on how to calibrate diagnostic dosimeters and determine the patient doses resulting from x- ray examinations, the International Atomic Energy Agency initiated the development of a Code of Practice (CoP) for dosimetry in x-ray diagnostic radiology. For testing of various procedures described in the CoP before their broad implementation in practice the IAEA initiated a Coordinated Research Project (CRP) entitled Testing of the Implementation of the Code of Practice on Dosimetry in X-Ray Diagnostic Radiology. Hungarian partner is involved in the following activities: evaluation of measurement procedures general radiography, fluoroscopy, mammography, computed tomography, dental radiology - in hospitals, calibration of KAP meters and TLD dosimetry audit. The Coordinated Research Project is in progress. KEYWORDS: patient dosimetry; diagnostic radiology; medical radiation; x-ray exposure. 1. Introduction Medical ionization radiation sources give by far the largest contribution to the population dose from man-made sources and most of this contribution comes from diagnostic x-rays. Radiologist constantly face the dilemma of trying to minimize patients exposure whenever possible, while still using exposures that are high enough to produce images of good quality to be able to provide proper diagnosis. Quality assurance provides a framework for achieving this goal. As part of the QA, there is a need to control patient doses arising from x-ray diagnostic activity. Due to the increased demand for dosimetry measurements in diagnostic and interventional radiology, it has become important to provide traceability of measurements, which can only work satisfactorily if correct calibrations and measurements are performed. In order to advice Secondary Standard Laboratories and end-users in hospitals on how to calibrate diagnostic dosimeters and determine the patient doses resulting from x-ray examinations, the International Atomic Energy Agency initiated the development of a Code of Practice (CoP) for dosimetry in x-ray diagnostic radiology [1]. For testing of various procedures described in the CoP before their broad implementation in practice the IAEA initiated a Coordinated Research Project (CRP) entitled Testing of the Implementation of the Code of Practice on Dosimetry in X-Ray Diagnostic Radiology. The CoP provides methods for clinical dosimetry in general radiography, fluoroscopy, mammography, CT and dental radiography. For each modality, the procedures for phantom measurements as well as the procedures for measurements on patients are describes. Hungarian partner is involved in the following activities: evaluation of measurement procedures * Presenting author, giczif@sze.hu 1

2 general radiography, fluoroscopy, mammography, computed tomography, dental radiology - in hospitals, calibration of KAP meters and TLD dosimetry audit. The authors illustrate their measurements in chest PA radiography using phantoms as well as patient dosimetry and summarize their preliminary results and conclusions connecting to the other activities in the CoP. 2. Clinical measurements 2.1 General radiography Measurements using phantoms In general radiography the incident air kerma was the dosimetric quantity measured for the chest PA and lumbar spine AP projections. The CDRH chest and CDRH abdomen/lumbar spine phantoms were the standard phantoms used utilizing the x-ray exposure parameters for an average-sized adult patient. In case of chest examination, the technologist was asked to set up the x-ray equipment for a normal adult patient, including the selection of exposure parameters, focus to skin distance, collimation, etc. The phantom was positioned so that it rests directly against the front plate of the vertical Bucky (Figure 1.). The distance between the focus and the surface of the vertical Bucky was measured and recorded, d FTP. The dosimeter was placed in the probe holder, sufficiently above the phantom surface to reduce backscatter and positioned outside the AEC detectors. Distance d m between the reference point of the dosimeter and the Bucky surface was measured and recorded. The ionization chamber of the detector were exposed three times under automatic exposure control or manual control as appropriate. Dosimeter readings and exposure parameters used were recorded. Finally, the ambient temperature and pressure were recorded. Figure 1: Set-up for measurements using a CDRH chest phantom d m t P t FTD The dose calculation was performed in two steps. First the air kerma, K(d), at the measurement point (at a distance, d m, from the Bucky surface) was calculated from the mean value of dosimeter readings using Eq K(d) MNK,Q0 = k k (2.1) Q TP In Eq. 2.1, k TP is the correction factor for temperature and pressure, N K,Q0 is the dosimeter calibration coefficient, k Q is the factor which corrects for differences in the response of the dosimeter at the calibration quality Q 0, and the quality Q of the clinical x-ray beam. In the second step, using the inverse square law (Eq. 2.2), the incident air kerma, K i, to the standard chest was calculated. K i 2 dftd d m K(d) = (2.2) d FTD t p In Eq. 2.2, t P is the thickness of the phantom. Testing of the implementation of CoP methodology were performed on 3 chest radiography x-ray equipment. The main characteristics of the equipment and the measured data can be found in Table 1. 2

3 Table 1: The main characteristics and the measured data for the chest phantom measurement. Identification HU-1 HU-2 HU-3 Generator TOP-X HF EMERIX EDR-750B Vertical Bucky BA-1 BA-1 BA-1 AEC Yes Yes No kv mas HVL (mm Al) d FTD (mm) d m (mm) Dose measurements were performed with a Radcal 9015 dosimeter, calibrated by the Hungarian National Calibration Laboratory. Calibration coefficient N K,Q0 =1.008 mgy/reading. Reference conditions at the calibration were the following: beam quality: BIPM 100, P 0 = kpa, T 0 =20 0 C. It is important to note, that in Hungary, the calibration laboratory specify the calibration coefficient of diagnostic dosimeters only for one beam quality, so no beam quality corrections could be performed. It was a problem that in some cases we could not find the mark of the focus position on the tube housing. In these cases the d FTD was measures from the centreline of the x-ray tube housing. The value of t P is 225 mm for the CDRH standard chest phantom. Dosimetric quantities and correction factors are summarized in Table 2. Table 2: Dosimetric quantities and correction factors for the chest phantom measurement. Identification HU-1 HU-2 HU-3 M (reading) k TP k Q K(d) (mgy) K i (mgy) Uncertainties (mgy) ± ± ± The relative expanded uncertainty (k=2) of the results was 7.2%, applying the scenario 2 in the CoP [1] Patient dosimetry As the first step of patient dosimetry, the x-ray tube output was measured for a representative set of tube voltages and tube loadings which adequately sample the patient exposure parameters used for chest PA projections. Geometry of measurements can be seen in Figure 2. The same geometry was used for the HVL measurements. Figure 2: Set-up for measurement of tube output and HVL 3

4 The detector was positioned to a distance, d, from the focus and sufficiently far from the patient support, avoiding the possible influence of backscatter radiation. In manual exposure control, one of the representative value of tube voltage, tube loading and field size were selected. The ionization chamber was exposed three times, and the readings were recorded. Following the methodology of the CoP, the HVL of the beam was also measured. The measurements were repeated on three tube voltages. Calculation of air kerma, K(d), at the measurement point (at a distance, d, from the x-ray focus) was the same procedure, described earlier, in paragraph Finally, the K(d) values were normalized by the tube loading. Results of the measurements on the chest radiography equipment, identified by HU-2 can be found in Table 3. Table 3: Measurement of tube output and HVL on the chest radiography equipment (HU-2). Tube voltage (kv) HVL (mm Al) M (reading) Y(d) (mgy/mas) Calibration coefficient of the dosimeter was N K,Q0 =1.008 mgy/reading. The correction factor for pressure and temperature was k TP =1.004 and k Q =1, because of the reason mentioned earlier. The measurement point was at a distance of 570 mm from the x-ray focus. Patient and projection data was collected by the technologist performed the chest x-ray examination for 10 patients. The collected data and the calculated incident air kerma as well as entrance surface air kerma values can be found in Table 4. These data are connected to the workplace identified by HU-2. Table 4: Exposure parameters and patient dose values for chest PA projections (HU-2). Patient ID. Tube voltage (kv) Tube loading mas t P (mm) Field size (mm x mm) K i (mgy) K e (mgy) S. J. (male) x ± ± 0.10 T. E. (female) x ± ± 0.10 R. K. (female) x ± ± Sz. B. (female) x ± ± M. J. (male) x ± ± R. R. (male) x ± ± T. T. (male) x ± ± 0.13 D. F. (female) x ± ± Sz. Z. (female) x ± ± 0.05 W. E. (female) x ± ± For each patient, the incident air kerma was calculated from the exposure parameters recorded, using Eq K i 2 d = Y(d)P (2.3) It dftd t p In Eq. 3., the x-ray tube output, Y(d), measured at a distance, d, from the tube focus on the different tube voltages was interpolated from the tube output data in Table 3. P It is the tube loading during the exposure of the patient. The tube focus-to-patient support distance, d FTD, was 1450 mm. The relative expanded uncertainty (k=2) was 7.2%, applying the scenario 2 in the CoP. However, it has to be 4

5 mentioned, that the patient thickness, t P, was measured by the technologist. Consequently the uncertainty of the t P values was not easy to estimate. The entrance surface air kerma was calculated from the incident air kerma values, using the appropriate backscatter factor for water. The backscatter factor was determined by interpolations, based on the tabulated data and the measured HVL and field size used during the exposures. 3. Other activities In Table 5 a summary of the Hungarian contribution to the CoP can be found. Based on our experiences, the following conclusions can be drawn. Generally, the methodologies of the CoP are clear and easy to follow. Datasheets make the data collections in the hospitals easy and fast. There are only some problems can be mentioned. These problems are not serious, but could be appeared in other countries too. Table 5: Hungarian contribution to the CoP. Activity Phantom measurements Patient dosimetry General radiography Chest PA projection 3 measurements on 3 x-ray 30 patient dose measurements on 3 x-ray Lumbar spine AP projection 3 measurements on 3 x-ray 30 patient dose measurements on 3 x-ray Fluoroscopy 3 measurements on 3 interventional cardiology x-ray 14 patient dose measurements on 2 interventional cardiology x-ray Computed tomography Dental radiology 20 free-in-air measurements on 2 CT scanners, 20 dose measurements in CT head phantom, 20 dose measurements in CT body phantom. Bitewing projection - 17 patient dose estimations on 2 CT scanners 6 patient dose estimations on 1 dental x-ray unit Panoramic projection - - Mammography 4 measurements on 4 mammography x-ray Calibration of KAP meters 2 KAP meters 41 patient dose measurements on 2 mammography x-ray There was a general problem, that the calibration laboratory does not specify the calibration coefficient of diagnostic dosimeters for more than one beam quality. Consequently we could not perform the beam quality correction of the dosimeter readings. The other general problem was, that on some of the x-ray tube housing there were not the mark of the focus position. In these cases, we measured the distances from the centre of the housing. In general radiography, the technologist should measure the patient thickness. Our experience is that it can be made only with relatively high uncertainties. For mammography measurements, it s worth mentioning that the calculation of glandular dose is very time consuming. The frequent interpolations of the tabulated data may be the sources of calculation errors. A computer program helping the users should be very useful and is highly suggested. 5

6 4. Conclusion In Hungary the overall patient dose measurements in diagnostic radiology was started in 1989 [2-10]. However, the applied methods were frequently changed, following the methods found in the literature. Consequently, temporal trends in patient doses could hardly be followed. In the framework of the coordinated research project, we could test of various procedures described in the CoP. It can be concluded, that the methods of the CoP are clear and can be followed. Datasheets make easy and fast consistent data collection in the hospitals. Uncertainty estimation promotes the evaluation of the patient doses. The CoP could be the base methodology of nationwide surveys of patient doses arising from diagnostic application of x-rays and determination of diagnostic reference levels. Dissemination of the methods of the CoP can promote the comparability of patient doses. REFERENCES [1] INTERNATIONAL ATOMIC ENERGY AGENCY, Dosimetry in Diagnostic Radiology: An International Code of Practice, Technical Reports Series No. 457, IAEA, Vienna (2007). [2] PERNICKA, F., GICZI. F., et al., Comparison of TLD Air Kerma Measurements in Mammography, International Symposium on Standards and Codes of Practice in Medical Radiation Dosimetry, Vienna, Austria, November 2002., Book of Extended Synopses, IAEA-CN-96-47P, p [3] MACCIA. C., PADOVANI. R., VANO. E., REHANI. MM., et al., Image Quality and Patient Dose Optimization in Mammography in Eastern European Countries, World Congress on Medical Physics and Biomedical Engineering, Sydney, Australia, August [4] Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe), 2005., IAEA-TECDOC-1447 [5] PELLET, S., GICZI, F., et al., A Pilot Study of Radiation Exposures Arising From Interventional Radiology Procedures, Second European IRPA Congress on Radiation Protection, International Radiation Protection Association, Paris, France, Proceeding of Full Papers on CD-Rom. [6] PELLET, S., FAULKNER, K., VANO, E., PADOVANI, R., GICZI, F., et al., Hungarian Contribution to the SENTINEL Project, International Conference on Quality Assurance and New Techniques in Radiation Medicine, International Atomic Energy Agency, Vienna, Austria, November Book of Extended Synopses IAEA-CN-146/261P, p [7] GICZI. F., PELLET. S., et al., Study on the patient dose of photofluorography in Hungary, The Central European Journal of Occupational and Environmental Medicine, (2): [8] PELLET. S., GICZI. F., et al., Hungarian patient dose survey for photofluorography applied in mass chest screening program, Radiation Protection Dosimetry, (Vol. 80, Nos 1-3, pp (1998) [9] GICZI. F., PELLET. S., VANO. E., MACCIA. C., et al. Image quality and patient dose optimization in mammography in Hungary, International Conference on the Radiological Protection of Patients, Málaga, Spain, March Contributed Papers, p [10] PELLET. S., GICZI. F., et al., Patient doses for computed tomography in Hungary, International Conference on the Radiological Protection of Patients, Málaga, Spain, March Contributed Papers, p

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Application of international standards to diagnostic radiology dosimetry

Application of international standards to diagnostic radiology dosimetry Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

Dosimetry in digital mammography

Dosimetry in digital mammography Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3

P T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1, T.Ishigaki 1 and S.Koga 3 Radiation Exposure by Routine Radiographic Examinations: Multicenter Study in Japan with Thermoluminescence Dosimetry and Estimation from the Radiographic Data T.Ishiguchi 1, S.Iwanami 2, S.Kawatsu 1,

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals

Impact of quality control on radiation doses received by patients undergoing abdomen X-ray examination in ten hospitals B.Aghahadi 1,2*,Z.Zhang 1,S.Zareh 3, S. Sarkar 3, P.S.Tayebi 4 1 Department of Physics, Harbin Institute of Technology, Harbin 150080. P.R. China 2 Department of National Radiation Protection, Atomic Energy

More information

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies

Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Measurements of Air Kerma Index in Computed Tomography: A comparison among methodologies Thêssa C. Alonso 1, 2, Arnaldo P. Mourão 1, 3, Teógenes A. Da Silva 1, 2 1 Program of Nuclear Science and Techniques

More information

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology

Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology G K Korir, BSc, MSc (Nuclear Science), PhD (Radiological Science) Department of Physics

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 233-3 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 29 Dosimetry Framework, Formalism including Uncertainties Costas Hourdakis Greek Atomic

More information

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science

Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science PATIENT DOSIMETRY IN DIAGNOSTIC RADIOLOGY MODALITIES Prof. Dr. Doğan BOR Ankara University Institute of Nuclear Science Ankara University Institute of Nuclear Science USE OF RADIATION! INCREASING? Natural

More information

ABSTRACT I. INTRODUCTION

ABSTRACT I. INTRODUCTION 2017 IJSRSET Volume 3 Issue 6 Print ISSN: 2395-1990 Online ISSN : 2394-4099 Themed Section: Engineering and Technology Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography

More information

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration

Establishment of the new IEC mammography qualities in a clinical system used for instruments calibration Establishment of the new IEC 61267 mammography qualities in a clinical system used for instruments calibration 1 Corrêa, E.L.; 2 Vivolo, V.; 3 Potiens, M.P.A. Abstract Instituto de Pesquisas Energéticas

More information

Highlights and Conclusions Thematic Area 4: Radiation of Patients

Highlights and Conclusions Thematic Area 4: Radiation of Patients Highlights and Conclusions Thematic Area 4: Radiation of Patients by Robert H. Corbett and Eliseo Vañó Main topics: Diagnostic Radiology Interventional Radiology Nuclear Medicine Radiotherapy New Techniques

More information

Mammography calibration qualities establishment in a Mo- Mo clinical system

Mammography calibration qualities establishment in a Mo- Mo clinical system Journal of Physics: Conference Series PAPER OPEN ACCESS Mammography calibration qualities establishment in a Mo- Mo clinical system To cite this article: E L Corrêa et al 2016 J. Phys.: Conf. Ser. 733

More information

Dosimetric Consideration in Diagnostic Radiology

Dosimetric Consideration in Diagnostic Radiology Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Patient Dosimetry in Mammography and Tomosynthesis:

Patient Dosimetry in Mammography and Tomosynthesis: 2013 ICTP/IAEA Training Course on Radiation Protection of Patients Trieste Patient Dosimetry in Mammography and Tomosynthesis: What to measure, why and how John M. Boone, Ph.D., FAAPM, FSBI, FACR Professor

More information

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0

BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 BREAST DOSE SURVEYS IN THE NHSBSP: SOFTWARE AND INSTRUCTION MANUAL Version 2.0 Report 04/ July 2004 K C Young National Co-ordinating Centre for the Physics of Mammography Guildford CONTENTS 1. INTRODUCTION...2

More information

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations

Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Caspian Journal of Health Research 2016; 2(2):18-27 Survey Effect of the Corrective Actions and Quality Control in Image Quality and Patient Dose in Radiography Examinations Hossein Rahanjam 1, Hossein

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012

REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012 REPORT OF THE TECHNICAL MEETING: THE NEW DOSE LIMIT FOR THE LENS OF THE EYE IMPLICATIONS AND IMPLEMENTATION VIENNA, 2-4 OCTOBER 2012 Introduction A Technical Meeting on the The new dose limit for the lens

More information

A Standardised Approach to Optimisation

A Standardised Approach to Optimisation The Radiographer vol. 51: 105-110 A Standardised Approach to Optimisation Helen M. Warren-Forward 1 & Ronald Beckhaus 2 ABSTRACT Purpose: Optimisation of radiographic images is said to have been obtained

More information

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,

More information

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY

OPTIMISATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY March 2010 EAST AFRICAN MEDICAL JOURNAL 127 East African Medical Journal Vol. 87 No. 3 March 2010 OPTIMIsATION OF PATIENT PROTECTION AND IMAGE QUALITY IN DIAGNOSTIC RADIOLOGY G. K. Korir, BSc, MSc, PhD

More information

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio

Quality Control in Mammography: Results from a study in Recife, Brazil. H. J. Khoury, V. S. de Barros, M. Sampaio Quality Control in Mammography: Results from a study in Recife, Brazil H. J. Khoury, V. S. de Barros, M. Sampaio Grupo de Dosimetria e Instrumentação, Universidade Federal de Pernambuco, Rua Prof. Luiz

More information

Trends in Occupational Exposure in Malaysia

Trends in Occupational Exposure in Malaysia Trends in Occupational Exposure in Malaysia Noriah Mod Ali Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute for Nuclear Technology Research (MINT) Bangi, 43000 KAJANG, Selangor Darul

More information

Patient dose measurement in common medical X-ray examinations in Iran

Patient dose measurement in common medical X-ray examinations in Iran JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 1, 2016 Patient dose measurement in common medical X-ray examinations in Iran Behrouz Rasuli, 1a Ali Mahmoud-Pashazadeh, 2 Mohammad Ghorbani,

More information

IAEA-TECDOC-796. Radiation doses in diagnostic radiology

IAEA-TECDOC-796. Radiation doses in diagnostic radiology IAEA-TECDOC-796 Radiation doses in diagnostic radiology The IAEA does The originating Section FOREWORD It is well recognized that diagnostic radiology is the largest contributor to the collective dose

More information

IAEA TECDOC 1517 Quality Control in Mammography Software

IAEA TECDOC 1517 Quality Control in Mammography Software IAEA TECDOC 1517 Quality Control in Mammography Software Fredy Somarriba a, Patricia Mora b*, Margarita Chevalier c and Raul Ramirez d a Universidad Nacional Autónoma de Nicaragua, Managua, Nicaragua.

More information

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography

Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Wilson Otto Gomes Batista Instituto Federal da Bahia Rua Emidio dos Santos s/n. Barbalho 40301-015

More information

Trends in Patients Exposure Doses during Radiographic Examination in Japan

Trends in Patients Exposure Doses during Radiographic Examination in Japan Trends in Patients Exposure Doses during Radiographic Examination in Japan S. Suzuki 1*, Y. Asada 1, M. Kobayashi 2, K. Kobayashi 2, Y. Matsunaga 3, T. Otsuka 4 1 School of Health Sciences, Fujita Health

More information

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers

Patient effective dose evaluation for chest X-ray examination in three digital radiography centers Patient effective dose evaluation for chest X-ray examination in three digital radiography centers R. Paydar 1,3*, A. Takavar 1, M.R. Kardan 2,3, A. Babakhani 3,4, M.R. Deevband 3, S. Saber 5 1Medical

More information

Radiation exposure of the Yazd population from medical conventional X-ray examinations

Radiation exposure of the Yazd population from medical conventional X-ray examinations Iran. J. Radiat. Res., 2007; 4 (4): 195-200 Radiation exposure of the Yazd population from medical conventional X-ray examinations F. Bouzarjomehri 1*, M.H. Dashti 2, M.H. Zare 1 1 Department of Medical

More information

Pilot study of patient and phantom breast dose measurements in Bulgaria

Pilot study of patient and phantom breast dose measurements in Bulgaria Pol J Med Phys Eng. 2008;14(1):21-32. PL ISSN 1425-4689 doi: 10.2478/v10013-008-0003-3 website: http://www.pjmpe.waw.pl Simona Avramova-Cholakova, Jenia Vassileva Pilot study of patient and phantom breast

More information

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS

THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS THE STUDY OF PATIENT DOSES IN DIAGNOSTIC RADIOLOGY FOR ESTABLISHMENT OF NATIONAL DIAGNOSTIC REFERENCE LEVELS Larysa Stadnyk, Pylypenko M., Shalepa O., Nosyk O. Grigorev Institute for Medical Radiology,

More information

Computed tomography Acceptance testing and dose measurements

Computed tomography Acceptance testing and dose measurements Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed

More information

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran

Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Patient doses in radiographic examinations in Western and Eastern Azerbyjan provinces of Iran Soraya Emamgholizadeh Minaei 1, Fariba Firouzi 1 2, 3,*, Hamid Reza Khosravi 1 Department of radiology, Shahid

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY

PNRA-RG (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY PNRA-RG-904.05 (Rev. 0) December, 2017 PROTECTION OF PATIENTS IN DIAGNOSTIC RADIOLOGY REGULATORY GUIDE PAKISTAN NUCLEAR REGULATORY AUTHORITY For Further Details Directorate of Policies & Procedures PAKISTAN

More information

CCRI(I)/ International Atomic Energy Agency ACTIVITIES OF THE DOSIMETRY AND MEDICAL RADIATON PHYSICS SECTION Activities in 2011 and 2012

CCRI(I)/ International Atomic Energy Agency ACTIVITIES OF THE DOSIMETRY AND MEDICAL RADIATON PHYSICS SECTION Activities in 2011 and 2012 CCRI(I)/13-22 International Atomic Energy Agency ACTIVITIES OF THE DOSIMETRY AND MEDICAL RADIATON PHYSICS SECTION Activities in 2011 and 2012 1. INTRODUCTION... 1 2. OVERVIEW OF PROJECTS IN DOSIMETRY AND

More information

Patients, Staff, Public, Animals Evaluate RP needs related to emerging technologies. Produce recommendations and user-friendly guidance on RP.

Patients, Staff, Public, Animals Evaluate RP needs related to emerging technologies. Produce recommendations and user-friendly guidance on RP. Committee 3 (Protection in Medicine) is concerned with the protection of persons and unborn children when ionising radiation is used in medical diagnosis, therapy, and biomedical research, as well as protection

More information

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit

Appendix I. List of stakeholders consulted with on the Patient Radiation Protection Manual and members of the Medical Exposure Radiation Unit References References The accuracy, quality and relevance of these works are not guaranteed or uniform and more recent information may have superseded these works. This list is not exhaustive. It does

More information

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use?

Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Doses from pediatric CT examinations in Norway Are pediatric scan protocols developed and in daily use? Eva Godske Friberg * Norwegian Radiation Protection Authority, P.O. Box, Østerås, Norway Abstract.

More information

Patient dose in routine X-ray examinations in Yazd state

Patient dose in routine X-ray examinations in Yazd state Iran. J. Radiat. Res., 2004; 1(4): 199-204 Patient dose in routine X-ray examinations in Yazd state F. Bouzarjomehri Health Physics Dept. Shahid Sadoghi Univ. of Medical Sciences, Yazd, Iran ABSTRACT Background:

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements

More information

Survey of patients CT radiation dose in Jiangsu Province

Survey of patients CT radiation dose in Jiangsu Province Original Article Page 1 of 6 Survey of patients CT radiation dose in Jiangsu Province Yuanyuan Zhou 1, Chunyong Yang 1, Xingjiang Cao 1, Xiang Du 1, Ningle Yu 1, Xianfeng Zhou 2, Baoli Zhu 1, Jin Wang

More information

Introduction and Background

Introduction and Background CT Lung Cancer Screening and the Medical Physicist: Background, Findings and Participant Dosimetry Summary of the National Lung Screening Trial (NLST) Randell Kruger, PhD, DABR Medical Physics Section

More information

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA Full Length Research Article 1 Ibrahim, U, 3 Daniel, I.H., 3 Ayaninola, O., 4 Ibrahim,

More information

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology

Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology Iran. J. Radiat. Res., 2006; 4 (1): 21-27 Application of dose-area product compared with three other dosimetric quantities used to estimate patient effective dose in diagnostic radiology M.T. Bahreyni

More information

Diagnostic reference levels as a quality assurance tool

Diagnostic reference levels as a quality assurance tool Australian Institute of Radiography The Radiographer 2009; 56 (3): 32 37 Literature review Diagnostic reference levels as a quality assurance tool KD Edmonds Medical Physics Section, Medical Radiation

More information

Patient doses in general radiography examinations

Patient doses in general radiography examinations peer reviewed ORIGINAL ARTICLE Patient doses in general radiography examinations Wambani JS 1 MBChB, M. Med (Rad), Paediatric (Rad) Onditi EG 2 MBChB, M.Med (Rad) Korir GK 3 B.Sc, MSc (Nuclear Science),

More information

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS

SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS SPANISH INTERCOMPARISON OF APPROVED PERSONAL DOSIMETRY SERVICES USING PHOTON RADIATION BEAMS I. Villanueva (1), I. Amor (1), A. Brosed (), R. Villanueva () M. Ginjaume (3), X. Ortega (3), M. Roig (3) (1)

More information

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro

PhD Title: Improving QA/QC in mammography screening and breast in Montenegro 3rd Doctoral Coordinated Research Project (CRP) on Advances in medical imaging techniques (E2.40.19) MNE BEL PhD Title: Improving QA/QC in mammography screening and breast in Montenegro MNE BEL PhD Candidate:

More information

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

SSDL Newsletter. Contents. From the editor. No. 51 September IAEA/WHO Network of Secondary Standards Dosimetry Laboratories

SSDL Newsletter. Contents. From the editor. No. 51 September IAEA/WHO Network of Secondary Standards Dosimetry Laboratories SSDL Newsletter No. 51 September 2005 IAEA/WHO Network of Secondary Standards Dosimetry Laboratories Prepared by the Joint IAEA/WHO Secretariat of the SSDL Network http://www-naweb.iaea.org/nahu/external/e3/ssdl.asp

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis 60 Journal of Radiology and Diagnostic Imaging, 2013, 1, 60-64 Quality Assurance and Optimization of Radiation Protection of Patients in X-Ray Radiographic Examinations Using the Rejects Film Analysis

More information

Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology

Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology Horst Aichinger Joachim Dierker Sigrid Joite-Barfuß Manfred Säbel Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology Physical Principles and Clinical Applications Second Edition Springer

More information

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS

PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS Bayero Journal of Pure and Applied Sciences, 2(1): 1-5 Received: October, 2008 Accepted: February, 2009 PATIENT ENTRANCE SKIN DOSES AT MINNA AND IBADAN FOR COMMON DIAGNOSTIC RADIOLOGICAL EXAMINATIONS *Sharifat,

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

The New Jersey Radiographic Quality Assurance Program at 5 Years

The New Jersey Radiographic Quality Assurance Program at 5 Years The New Jersey Radiographic Quality Assurance Program at 5 Years Julie Timins, MD a, Paul Orlando, BS b, Jill Lipoti, PhD b Purpose: Five years ago, the New Jersey Bureau of Radiological Health decided

More information

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME

PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION BASED SCREENING PROGRAMME Number of mammography X-ray sets Radiation Protection Dosimetry (year), Vol. 0, No. 0, pp. 0 0 DOI: 10.1093/rpd/nc0000 SECTION TITLE HERE PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy What is radiation? Radiation is energy emitted from a source, that travels through space and can penetrate matter. Listed below are two types that we are exposed to and contribute to our overall radiation

More information

Influence of the exposure time in the area monitors at radiodiagnostic

Influence of the exposure time in the area monitors at radiodiagnostic Journal of Physics: Conference Series PAPER OPEN ACCESS Influence of the exposure time in the area monitors at radiodiagnostic To cite this article: F Leyton et al 2016 J. Phys.: Conf. Ser. 733 012087

More information

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives

BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation. Introduction. Educational Objectives BICOE Stereotactic Breast Biopsy and Breast Ultrasound Accreditation William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org 1 Introduction Objectives

More information

Bureau of Laboratory Quality Standards

Bureau of Laboratory Quality Standards 1. Condom -Lubricant ISO 4074:2002/Cor.1:2003 Annex C -Thickness ISO 4074:2002/Cor.1:2003 Annex F - Length ISO 4074:2002 / Cor1 : 2003 - Width ISO 4074 : 2002/Cor 1 : 2003 - Bursting pressure and Volume

More information

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED *

ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * ASSESSMENT OF OCCUPATIONAL DOSE IN FLUOROSCOPY PROCEDURES WHEN INDIVIDUAL MONITORING IS NOT UTILIZED * Olivera Ciraj-Bjelac 1**, Danijela Arandjic 1, Predrag Bozovic 1, Sandra Ceklic 2, Jelena Stankovic

More information

Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations

Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 6, NUMBER 1, WINTER 2005 Comparison of six phantoms for entrance skin dose evaluation in 11 standard X-ray examinations Gaetano Compagnone, Laura Pagan,

More information

International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry

International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry IAEA-CN-182 International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry Vienna, Austria 9-12 November 2010 Organized by the International Atomic Energy Agency

More information

International Journal of Radiation Research, January 2017

International Journal of Radiation Research, January 2017 Downloaded from ijrr.com at : +0330 on Tuesday November th Volume 5, No International Journal of Radiation Research, January 7 Radiation dose to the thyroid, eyes and parotid glands of patients undergoing

More information

IAEA SUBPROGRAMME ON DOSIMETRY AND MEDICAL RADIATION PHYSICS (DMRP) REPORT ON ACTIVITIES,

IAEA SUBPROGRAMME ON DOSIMETRY AND MEDICAL RADIATION PHYSICS (DMRP) REPORT ON ACTIVITIES, IAEA SUBPROGRAMME ON DOSIMETRY AND MEDICAL RADIATION PHYSICS (DMRP) REPORT ON ACTIVITIES, 2001 2002 K Shortt, A Meghzifene, J Izewska, F Pernicka, H Tölli, S Vatnitsky, P Bera, L Czap, R Girzikowsky Dosimetry

More information

Radiation Protection Dosimetry Advance Access published April 22, 2008

Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry Advance Access published April 22, 2008 Radiation Protection Dosimetry (2008), pp. 1 5 doi:10.1093/rpd/ncn144 EVALUATION OF EXPOSURE PARAMETERS IN PLAIN RADIOGRAPHY: A COMPARATIVE

More information

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES A THESIS SUBMITTED TO THE GRADUATE FACULTY in partial fulfillment of the requirements

More information

Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application

Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application Strengthening Activity Measurement Quality in Radiation Protection from Metrological Science to Reliable End-user Application Franz Josef Maringer a,b,c*, Robert Brettner-Messler a, Michael Kreuziger a

More information

Doses Received by Patients during Thorax X-Ray Examinations

Doses Received by Patients during Thorax X-Ray Examinations Iranian Journal of Medical Physics Vol. 9, No. 4, Autumn 01, 45-51 Received: September 16, 01; Accepted: February 0, 013 Original Article Doses Received by Patients during Thorax X-Ray Examinations Nsikan

More information

Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand

Evaluation of radiation dose to patients undergoing interventional radiology procedures at Ramathibodi Hospital, Thailand Available online at http://www.biij.org/2011/3/e22 doi: 10.2349/biij.7.3.e22 biij Biomedical Imaging and Intervention Journal ORIGINAL ARTICLE Evaluation of radiation dose to patients undergoing interventional

More information

The relationship of the mean glandular dose with compressed breast thickness in mammography

The relationship of the mean glandular dose with compressed breast thickness in mammography Review Article Page 1 of 8 The relationship of the mean glandular dose with compressed breast thickness in mammography Xiang Du, Ningle Yu, Yimei Zhang, Jin Wang Department of Radiation Protection, Jiangsu

More information

Establishing the Greek National Reference Levels for Interventional Cardiology procedures

Establishing the Greek National Reference Levels for Interventional Cardiology procedures Establishing the Greek National Reference Levels for Interventional Cardiology procedures George Simantirakis*, Christina Koukorava, Maria Kalathaki, Christos Pa

More information

Atoms for Peace and Development

Atoms for Peace and Development Atoms for Peace and Development Vienna International Centre, PO Box 100, 1400 Vienna, Austria Phone: (+43 1) 2600 Fax: (+43 1) 26007 Email: Official.Mail@iaea.org Internet: https://www.iaea.org In reply

More information

Assessing the standard dose to standard patients for x-ray investigations

Assessing the standard dose to standard patients for x-ray investigations Assessing the standard dose to standard patients for x-ray investigations A. Almén and W. Leitz Swedish Radiation Protection Authority, SE-171 16 Stockholm, Sweden E-mail: anja.almen@ssi.se Abstract. The

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Establishment of local diagnostic reference levels for quality control in intraoral radiography

Establishment of local diagnostic reference levels for quality control in intraoral radiography Oral Radiol (27) 33:38 44 DOI.7/s282-6-24-9 ORIGINAL ARTICLE Establishment of local diagnostic reference levels for quality control in intraoral radiography Maki Izawa Yasuo Harata Noriyoshi Shiba Nobuhide

More information

CT Dosimetry in the Clinical Environment: Methods and Analysis

CT Dosimetry in the Clinical Environment: Methods and Analysis CT Dosimetry in the Clinical Environment: Methods and Analysis Manuel Arreola, Ph.D. DABR Associate Chair of Radiology Director, Medical Physics Graduate Program Department of Radiology University of Florida

More information

Why is CT Dose of Interest?

Why is CT Dose of Interest? Why is CT Dose of Interest? CT usage has increased rapidly in the past decade Compared to other medical imaging CT produces a larger radiation dose. There is direct epidemiological evidence for a an increase

More information

Performance of several active personal dosemeters in interventional radiology and cardiology

Performance of several active personal dosemeters in interventional radiology and cardiology Performance of several active personal dosemeters in interventional radiology and cardiology S. Chiriotti 1,*, M. Ginjaume 1, E. Vano 2,3, R. Sanchez 3, J.M. Fernandez 2,3, M.A. Duch 1, J. Sempau 1 1.

More information

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Managing the imaging dose during Image-guided Radiotherapy Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University Radiographic image guidance has emerged as the new paradigm

More information

X-ray Safety Discussion and Tour for Health Physicists

X-ray Safety Discussion and Tour for Health Physicists X-ray Safety Discussion and Tour for Health Physicists April 1, 2016 Health Physics Society Spring Meeting Jerry Bai Environmental Administrator, Field Operations Bureau of Radiation Control Florida Department

More information

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria

A Survey of Radiation Exposure Techniques Factors Used for Common Diagnostics X- Ray Examinations in Akwa Ibom State, Nigeria International Journal of Medical Imaging 2015; 3(4): 69-74 Published online June 6, 2015 (http://www.sciencepublishinggroup.com/j/ijmi) doi: 10.11648/j.ijmi.20150304.11 ISSN: 2330-8303 (Print); ISSN: 2330-832X

More information

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations

Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations Measurement of Kidneys Dose in Three Hospitals during Chest and Abdomen Radiographic Examinations Nadia Alatta 1, Abdelrahaman Elnour 2, Ikhlas Abdelaziz 3, Tugua Tageldeen 4, Duha Abdu 5, Asma Elamin

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

Staff Exposure Monitoring in Interventional Radiology

Staff Exposure Monitoring in Interventional Radiology Conference on Physics in Medicine: From Diagnosis to Treatment "Enhancing safety and quality in radiation medicine". KFMC, Riyadh 7-9 November 2017 Staff Exposure Monitoring in Interventional Radiology

More information

JMSCR Vol 06 Issue 12 Page December 2018

JMSCR Vol 06 Issue 12 Page December 2018 www.jmscr.igmpublication.org Impact Factor (SJIF): 6.379 Index Copernicus Value: 79.54 ISSN (e)-2347-176x ISSN (p) 2455-0450 DOI: https://dx.doi.org/10.18535/jmscr/v6i12.80 Diagnostic Reference Levels:

More information

Improvement of radiation dosimetry: contribution of the IAEA

Improvement of radiation dosimetry: contribution of the IAEA Improvement of radiation dosimetry: contribution of the Ahmed Meghzifene Section Head Dosimetry & Medical Radiation Physics Section Division of Human Health International Atomic Energy Agency Radiation

More information

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433

Mammography. Background and Perspective. Mammography Evolution. Background and Perspective. T.R. Nelson, Ph.D. x41433 - 2015 Background and Perspective 2005 (in US) Women Men Mammography Invasive Breast Cancer Diagnosed 211,240 1,690 Noninvasive Breast Cancer Diagnosed 58,940 Deaths from Breast Cancer 40,410 460 T.R.

More information