JRPR. Prediction of Midline Dose from Entrance and Exit Dose Using OSLD Measurements for Total Body Irradiation. Original Research.
|
|
- Dorcas Randall
- 5 years ago
- Views:
Transcription
1 Journal of Radiation Protection and Research 2017;42(2):77-82 pissn eissn Prediction of Midline Dose from Entrance and Exit Dose Using OSLD Measurements for Total Body Irradiation Chang Heon Choi 1,2,3, Jong Min Park 1,2,3,4, So-Yeon Park 1,2,3, Minsoo Chun 1,2,3,5, Ji Hye Han 1,3, Jin Dong Cho 1,3, Jung-in Kim 1,2,3, * 1 Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea; 2 Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Korea ; 3 Biomedical Research Institute, Seoul National University College of Medicine, Seoul, Korea; 4 Center for Convergence Research on Robotics, Advance Institutes of Convergence Technology, Suwon, Korea; 5 Interdisciplinary Program in Radiation Applied Life Science, Seoul National University College of Medicine, Seoul, Korea Original Research Received March 22, 2017 Revision May 4, 2017 Accepted May 11, 2017 Corresponding author: Jung-in Kim Department of Radiation Oncology, Seoul National University Hospital, 101 Daehak-ro, Jongno-gu, Seoul 03080, Korea Tel: Fax: madangin@gmail.com This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non- Commercial License ( licenses/by-nc/4.0) which permits unrestricted noncommercial use, distribution, and reproduction in any medium, provided the original work is properly cited. Copyright 2017 The Korean Association for Radiation Protection ABSTRACT Background: This study aims to predict the midline dose based on the entrance and exit doses from optically stimulated luminescence detector (OSLD) measurements for total body irradiation (TBI). Materials and Methods: For TBI treatment, beam data sets were measured for 6 MV and 15 MV beams. To evaluate the tissue lateral effect of various thicknesses, the midline dose and peak dose were measured using a solid water phantom (SWP) and ion chamber. The entrance and exit doses were measured using OSLDs. OSLDs were attached onto the central beam axis at the entrance and exit surfaces of the phantom. The predicted midline dose was evaluated as the sum of the entrance and exit doses by OSLD measurement. The ratio of the entrance dose to the exit dose was evaluated at various thicknesses. Results and Discussion: The ratio of the peak dose to the midline dose was 1.12 for a 30 cm thick SWP at both energies. When the patient thickness is greater than 30 cm, the 15 MV should be used to ensure dose homogeneity. The ratio of the entrance dose to the exit dose was less than 1.0 for thicknesses of less than 30 cm and 40 cm at 6 MV and 15 MV, respectively. Therefore, the predicted midline dose can be underestimated for thinner body. At 15 MV, the ratios were approximately 1.06 for a thickness of 50 cm. In cases where adult patients are treated with the 15 MV photon beam, it is possible for the predicted midline dose to be overestimated for parts of the body with a thickness of 50 cm or greater. Conclusion: The predicted midline dose and OSLD-measured midline dose depend on the phantom thickness. For in-vivo dosimetry of TBI, the measurement dose should be corrected in order to accurately predict the midline dose. Keywords: OSLD, In vivo dosimetry, Total body irradiation (TBI), Midline dose Introduction JRPR Total body irradiation (TBI) is a specialized radiotherapy technique that delivers a uniform dose of within ± 10% of the prescribed dose to the entire body of the patient [1]. It is used to prepare a patient for bone marrow or stem cell transplantation by destroying the number of viable cells and suppressing the recipient s immune system 77
2 JRPR Choi CH, et al. with total body exposure [2-6]. TBI usually uses the extended source-to-surface distance (SSD) due to three reasons. The first reason is to be able to deliver the beam with a low dose rate of less than 10 cgymin -1 by reducing the beam efficiency. The second reason is to allow a uniform dose distribution along the beam axis. The final reason is to be able to cover the entire body of the patient with large field sizes (e.g cm 2 ) [7-10]. The parallel-opposed lateral irradiation using extended SSD is the most preferred technique in TBI [11]. This stationary beam technique is very simple and is more comfortable for the patient than non-stationary beam techniques such as the translation beam and sweeping beam [12]. Moreover, the arms of the patient could be used as natural lung compensation for irradiation in the lateral direction in order to lower the risk of radiation-induced lung injury [12]. The tissue lateral effect is observed when a beam of insufficient energy is used for a patient which treated with parallel-opposed beams, and can give rise to an excessively higher dose to the peripheral regions compared to that of the midpoint [9]. The thicker the patient, the higher the beam energy required to produce acceptable dose uniformity for parallelopposed fields [9]. In TBI, the target volume is extended and has an irregular shape; therefore, it is necessary to define one representative point at which the prescribed dose is administered. This point is placed at the patient s midline and close to the abdomen or pelvis [13]. The TBI procedure delivers the prescribed dose to this reference point and should maintain the dose throughout the body to within ± 10% of the prescribed dose. The thickness variation along the medial axis of the patient and the patient diameter along the path of beam also affects dose uniformity. To achieve dose uniformity along the sagittal axis, compensators such as lead or copper plates are designed for the head, neck, shoulder, chest, umbilicus, hip, thigh, knee, calf, and ankle areas, respectively. In vivo dosimetry should be executed with a dosimeter to verify the dose calculation and compensator design [14]. The entrance and exit doses are measured by in vivo dosimetry during treatment and are used to estimate the delivered midline doses [10]. However, it is difficult to predict the midline dose from entrance and exit dose measurements since these surface doses change depending upon the thickness and beam energy [15]. Satory developed an empirical formula to calculate the midline dose from MOSFET (metal-oxidesemiconductor field-effect transistor) measurements of the entrance and exit doses. The dependency of the surface dose on the air gap between the spoiler and the surface was investigated with various phantom thicknesses. The entrance and exit doses were combined using an exponential attenuation formula to give an estimate of the midline dose and this was compared to the midline ion chamber measurement for a range of phantom thicknesses [10]. Although there are many advantages of using MOSFETs, such as its portability, reproducibility, and direct readout capability, it is a quite cumbersome to install and attach the device to the surface of the patient since the MOSFET device requires a cable (to apply a voltage to the detector) and a dedicated reader device. OSLDs have been well established to be suitable dosimeters for in vivo measurements due to their reproducibility, low energy dependence, small size, and ease of use [16]. OS- LDs have high sensitivity and can measure a surface dose of patient with a non-destructive readout (i.e. without removal the cover or packing) and reanalysis using a simple readout system. Moreover, measurement accuracy can be improved using OSLDs with fully filled deep electron/hole traps and optimal bleaching conditions [17, 18]. In this study, the entrance and exit doses were measured using OSLDs at strategic locations. The strategic locations means the point which patient thickness was measured to fabricate a compensator. The midline dose was predicted from these measured doses. We investigated the relationship between the midline dose and the surface dose for a range of patient thicknesses. The relationship was evaluated to predict the midline dose from entrance and exit dose measurements for 6 MV and 15 MV. Materials and Methods 1. OSLD preparation InLight nanodots (Landauer, Inc., Glenwood, IL) were used as the OSLDs. The OSLDs were pre-irradiated at a dose greater than 5 kgy to fill the deep electron and hole traps by means of a 60 Co gamma ray source. The OSLDs were bleached to fall below the residual signal level over 4 h. In our institution, the clinical dosimetry using OSLDs with fully filled deep electron/ hole traps under appropriate bleaching conditions have been shown to be highly stable and accurate, with no change in either the dose sensitivity or linearity. The uncertainty of OSLDs prepared by the process was less than 3%. The R square of calibration curve was 0.99 within calibration range ( cgy). 78
3 Prediction of Midline Dose for TBI In Vivo Dosimetry JRPR A B Fig. 1. Total body irradiation set-up with linac and solid water phantom (A), Measurement setup for measuring entrance, exit and midline dose with solid water phantom (B). 2. TBI measurement Both the 6 MV and 15 MV photon beams were generated with a linear accelerator (Trilogy, Varian Medical Systems, Palo Alto, CA) for the TBI simulation. The midline of the patient was positioned 4 m from the source in consideration of the size of the treatment room, patient setup, and dose rate (Figure 1A). The extended SSD, tissue-maximum ratio (TMR), scatter factor, and transmission factors (block tray, beam spoiler, and compensator) were measured for 6 MV and 15 MV, respectively. The transmission factors was defined as the ratio of the radiation dose with the device to the radiation dose without the device and accounts for the material in the beams path. The beam spoiler, which consisted of a 1 cm-thick acrylic screen, was positioned in the path of the beam to increase the surface dose to the patient. To determine the midline dose (Dm_mid), the dose conversion factor was derived. When one hundred MU (monitor units) was delivered for the reference setup (SSD: 100 cm, field size: 10 cm 10 cm), the integrated coulomb measurement at the depth of the maximum dose (dmax) was measured using a Farmer-type chamber (TN30013, PTW, Freiburg, Germany) and an electrometer (Uniods E, PTW, Freiburg, Germany) in a solid water phantom (SWP, Virtual Watert TM, Radiation Products Design Inc., Albertville, MN) for 6 MV and 15 MV. The dose conversion factor was defined as the measured coulomb (C) at dmax when one hundred monitoring unit was delivered. The unit of dose conversion factor was C Gy -1. The Dm_mid was measured using the Farmer-type chamber that was inserted into a hole in a SWP slab of 2 cm thickness. The chamber was placed at the midline along the central axis of the beam. The coulomb measurements were converted to dose units (Gy) using the dose conversion factor. The solid water phantom slabs were added in order to achieve the desired thickness. In this study, the measurement was performed for a range of phantom thicknesses (6 cm, 10 cm, 16 cm, 20 cm, 22 cm, 26 cm, 30 cm, 36 cm, 40 cm, 46 cm, and 50 cm). To evaluate the dose homogeneity due to the tissue lateral effect for the patient, the dose at a depth of 1 cm (Dpeak) was measured using the Farmer-type chamber. The ratio of Dpeak to Dm_mid was calculated for each thickness. OSLDs were placed on the central axis at the entrance and exit of the phantom to measure the entrance (Den) and exit (Dex) doses (Figure 1B). Three OSLDs were attached on each side to reduce measurement uncertainty. Solid water slabs of different thicknesses were joined in various combinations and attached to the front and back sides of the phantom for each energy. Considering dose homogeneity due to the tissue lateral effect, a 6 MV photon beam was used for phantom thicknesses of 6 cm, 10 cm, 16 cm, 20 cm, 22 cm, 26 cm, 30 cm, and 36 cm, and a 15 MV photon beam was used for phantom thicknesses of 10 cm, 16 cm, 20 cm, 22 cm, 26 cm, 30 cm, 36 cm, 40 cm, 46 cm and 50 cm. In consideration of the fading effect, the OSLDs were read using the InLight MicroStar reader (Landauer, Inc., Glenwood, IL) 15 minutes after irradiation [18]. For a reliable reading of the nanodot OS- LDs, a readout was performed three times for each chip, and the predicted midline dose (Dc_mid) was estimated as the sum of Den and Dex. 79
4 JRPR Choi CH, et al. Results and Discussion Table 1 shows the tray factor, spoiler factor, lead attenuation ratio per 1 mm thickness, and source to axis distance factor for 6 MV and 15 MV. The maximum difference between the measured and calculated TMR data was within 2% and 1% for 6 MV and 15 MV, respectively. Figure 2 shows the plot of the ratio of Dpeak to Dm_mid as a function of the phantom thickness for 6 MV and 15 MV. As the thickness of the phantom was increased, Dpeak also increased rapidly and the lower the energy, the faster the increase. At 6 MV, the ratio increased to 1.32 at a SWP thickness of 36 cm. When the SWP thickness was greater than 30 cm, Dpeak was greater than 10% of Dm_mid at 6 MV. When 15 MV was used at a SWP thickness of 50 cm, Dpeak increased to 125% of the midline dose. At a SWP thickness of 30 cm, the ratio was 1.12 for both energies. Therefore, if the thickness of the patient is greater than 30 cm, 15 MV should be used for treatment in order to achieve dose homogeneity [11]. Table 2 shows the values of Den, Dex, Dc_mid, Dm_mid, and the ratio of Dc_mid to Dm_mid for 6 MV. Den was increased and Dex was decreased as the thickness of the SWP was increased. The ratio of Dc_mid to Dm_mid also increased with increasing SWP thickness. However, when the thickness of the SWP was increased from 10 cm to 16 cm, the ratio decreased slightly because the increase of the entrance dose was smaller than the decrease of the exit dose. This is because the entrance dose depends on the distance between the spoiler and the surface of the entrance. For OSLDs measurements, the measurement point was included in the build-up region. Even though the SSD of the measurement point of the entrance dose was decreased, the dose did not increase considerably [10]. The ratio was approximately 1.0 at a SWP thickness of 30 cm. For a pediatric patient, the thickness of most organs will be less than 30 cm. In the case of the neck or ankle, the thickness will be almost 10 cm. Therefore, we can estimate Dc_mid to be less than Dm_mid. Table 3 shows the results of the 15 MV beam for the same data set as that of the 6 MV beam. The results show a similar tendency to those at 6 MV; at SWP thicknesses between 10 cm and 22 cm, the ratio decreased slightly. In this range, Dc_ mid was approximately 85% of Dm_mid. For most adult patients, the thickness of the hips with both hands placed adjacent to the body is at least 40 cm. Therefore, the 15 MV beam was selected for treatment in consideration of the need for dose homogeneity. The thickness of the head and neck is slightly less than 25 cm. Therefore, Dc_mid at these areas of the body can fall below Dm_mid. When the SWP thickness was greater Dpeak/Dmid Table 1. Basic Parameter for Monitoring Unit Calculation about 6 and 15 MV Energy (MV) Tray Factor Spoiler Factor Lead Attenuation ratio (per mm) Source Axis Distance Factor Solid water phantom thickness (cm) Fig. 2. The ratio of the dose at depth 1cm to that of at the midline as a function of solid water phantom thickness for 6 and 15 MV. Table 2. Comparison of Measured Midline Dose to Predicted Midline Dose Using OSLDs Exit and Entrance Dose Prediction for 6 MV Phantom Thickness Measured Midline dose (Dm_mid) Entrance dose (Den) Exit dose (Dex) Predicted Midline dose (Dc_mid) The ratio to Dc_mid to Dm_mid
5 Prediction of Midline Dose for TBI In Vivo Dosimetry JRPR Table 3. Comparison of Measured Midline Dose to Predicted Midline Dose Using OSLDs Exit and Entrance Dose Prediction for 15 MV Phantom Thickness Measured Midline dose (Dm_mid) Entrance dose (Den) Exit dose (Dex) Predicted Midline dose (Dc_mid) The ratio to Dc_mid to Dm_mid than 20 cm, the ratio increased in keeping with the SWP thickness. If the SWP thickness exceeded 46 cm, the ratio was greater than 1. Most obese patients have organ thicknesses of greater than 50 cm thickness at the shoulder, chest, umbilicus, and hip. Dc_mid of these points can exceed over 10% of the prescription dose. Dc_mid at 15 MV was estimated to be lower than that of 6 MV for same thickness. For 15 MV, a beam spoiler plate of 1 cm thickness is insufficient to produce the initial dose buildup region [10]. Conclusion For TBI treatment, the acceptable range of dose homogeneity is 10% for most protocols. However, the predicted midline dose by entrance and exit dose measurement can be overestimated even though the real delivered dose is within the acceptable range. The midline dose cannot be predicted accurately from the entrance and exit dose measurements with OSLDs due to an insufficiently full build-up region and the lack of backscatter material. In addition, the tissue lateral effect is varies for different thicknesses at the calculation point. Therefore, we need a correction to estimate the midline dose using the entrance and exit dose measurements. However, this correction changes for different thicknesses. The predicted midline dose can be corrected with the proper factor for each calculation point. Consequentially, we can predict a more accurate midline dose using the correction factor and surface dose. In this study, the SWP we have used have discrete thicknesses due to the SWP combination limitations. Therefore, we need to investigate the relationship between the midline dose and the thickness of the SWP with finer resolution. Acknowledgements This study was supported by a grant from the National R&D Program for Cancer Control, Ministry of Health & Welfare, Republic of Korea (No ). References 1. Keane T, Van Dyk JRider W. Idiopathic interstitial pneumonia following bone marrow transplantation: the relationship with total body irradiation. Int. J. Radiat. Oncol. Biol. Phys. 1981;7(10): Kim TH, Khan FMGalvin JM. A report of the work party: Comparison of total body irradiation techniques for bone marrow transplantation. Int. J. Radiat. Oncol. Biol. Phys. 1980;6(6): Appelbaum FR, Sullivan K, Buckner C, Clift R, Deeg HJ, Fefer A, Hill R, Mortimer J, Neiman PE, Sanders JE. Treatment of malignant lymphoma in 100 patients with chemotherapy, total body irradiation, and marrow transplantation. J. Clin. Oncol. 1987; 5(9): Li R, Polishchuk A, DuBois S, Hawkins R, Lee SW, Bagatell R, Shusterman S, Hill-Kayser C. Patterns of relapse in high-risk neuroblastoma patients treated with and without total body irradiation. Int. J. Radiat. Oncol. Biol. Phys. 2017;97(2): Reyzadeoglu M, Oysul K, Dirican B, Arpaci F, Balkan A, Surenkok S, Pak Y. Effect of dose-rate and lung dose in total body irradiation on interstitial pneumonitis after bone marrow transplantation. Tohoku J. Exp. Med. 2004;202(4): Kim MH, Rhu SY, Lim DS, Song JY. Effects of low-dose fractionated total body irradiation on murine immune system. J. Radiat. Prot. Res. 2014;39(3): Andreoli S, Bianchi C, Colleoni P, Fortunato M. Dosimetry in total body irradiation: A measurement set to characterize the treatment technique. Physica Med. 2016;32:
6 JRPR Choi CH, et al. 8. Koch N, Jacqmin D, Mart C, Fugal M, Ashenafi M, Vanek K. Total body irradiation using dynamic arc delivery. Int. J. Radiat. Oncol. Biol. Phys. 2016;96(2):E Khan FM. The physics of radiation therapy. 3rd Ed. Philadelphia, PA. Williams and Wilkins. 2002; Satory P. Calculation of midplane dose for total body irradiation from entrance and exit dose MOSFET measurements. Australas. Phys. Eng. Sci. Med. 2012;35(1): Van Dyk J. Dosimetry for total body irradiation. Radiother. Oncol. 1987;9(2): Quast U. Whole body radiotherapy: A TBI-guideline. J. Med. Phys. 2006;31(1): Hui SK, Das RK, Thomadsen B, Henderson D. CT based analysis of dose homogeneity in total body irradiation using lateral beam. J. Appl. Clin. Med. Phys. 2004;5(4): Akino Y, McMullen KP, Das IJ. Patterns of patient specific dosimetry in total body irradiation. Med. Phys. 2013;40(4): Zhu X. Entrance dose measurements for in-vivo diode dosimetry: Comparison of correction factors for two types of commercial silicon diode detectors. J. Appl. Clin. Med. Phys. 2000;1(3): Mrčela I, Bokulić T, Izewska J, Budanec M, Fröbe AKusić Z. Optically stimulated luminescence in vivo dosimetry for radiotherapy: physical characterization and clinical measurements in 60 Co beams. Phys. Med. Biol. 2011;56(18): Jursinic PA. Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. Med. Phys. 2007;34(12): Jursinic PA. Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose. Med. Phys. 2010;37(1):
Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter
Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter E.K. Chie M.D. 1, S.W. Park M.D. 1, W.S. Kang PhD. 1,2, I.H. Kim M.D. 1,2, S.W. Ha M.D. 1,2,
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationEXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *
Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationMultilayer Gafchromic film detectors for breast skin dose determination in vivo
INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination
More informationPMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.
Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers
More informationThe clinical use of OSLD
The clinical use of OSLD Stephen F. Kry PhD Radiological Physics Center, Imaging and Radiation Oncology Core - Houston IROC Houston The University of Texas MD Anderson Cancer Center, Houston, TX Spring
More informationDr. Rossella Vidimari. Department of Medical Physics Ospedale Maggiore A.S.U.I.T.S Ospedali Riuniti di Trieste
Dr. Rossella Vidimari Department of Medical Physics Ospedale Maggiore A.S.U.I.T.S Ospedali Riuniti di Trieste School on Medical Physics for Radiation Therapy 27 March 7 April 2017 Clinical indications
More informationEffect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution
NUKLEONIKA 2007;52(4):153 158 ORIGINAL PAPER Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution Tomasz
More informationPatient dosimetry for total body irradiation using single-use MOSFET detectors
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,
More informationUse of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments
Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331
More informationPMP. Gamma Evaluation with Portal Dosimetry for Volumetric Modulated Arc Therapy and Intensity-Modulated Radiation Therapy.
Original Article PMP Progress in Medical Physics 28(2), June 2017 https://doi.org/10.14316/pmp.2017.28.2.61 pissn 2508-4445, eissn 2508-4453 Gamma Evaluation with Portal Dosimetry for Volumetric Modulated
More informationAssessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital
Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri
More informationOn the use of bolus for pacemaker dose measurement and reduction in radiation therapy
Received: 8 February 2017 Revised: 16 October 2017 Accepted: 23 October 2017 DOI: 10.1002/acm2.12229 RADIATION ONCOLOGY PHYSICS On the use of bolus for pacemaker dose measurement and reduction in radiation
More informationSpatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.
Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry
More informationMedical dosimetry using Optically Stimulated Luminescence dots and microstar readers
Medical dosimetry using Optically Stimulated Luminescence dots and microstar readers Christopher A Perks *a, Cliff Yahnke b, Marc Million c a LANDAUER EUROPE, Unit 28, Bankside, Station Approach, Kidlington,
More informationClinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations
Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.111 pissn 2508-4445, eissn 2508-4453 Clinical Implications of High Definition Multileaf Collimator
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationDosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for Clinical Electron Beam
Journal of the Korean Physical Society, Vol. 55, No. 6, December 2009, pp. 2566 2570 Dosimetric Characteristics of Standard and Micro MOSFET Dosimeters as In-vivo Dosimeter for Clinical Electron Beam Jin-Beom
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of
More informationNormal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT
Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationEvaluation of Dosimetry Check software for IMRT patient-specific quality assurance
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationTreatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)
Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI) Tagreed AL-ALAWI Medical Physicist King Abdullah Medical City- Jeddah Aim 1. Simplify and standardize
More informationA translational couch technique for total body irradiation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 2, NUMBER 4, FALL 2001 A translational couch technique for total body irradiation Mehrdad Sarfaraz, Cedric Yu, D. J. Chen, and Leon Der Radiation Oncology
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationNeutron Measurements for Intensity Modulated Radiation Therapy
SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,
More information5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany
Evaluation of 3D diamond detectors for application in medical radiation dosimetry K. Kanxheri (1,2), L. Servoli (2), C. Zucchetti (5), A. C. Dipilato (5), M. Iacco (5), S. Lagomarsino (3,4), A. Morozzi
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationInvivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
More informationProtection of the contralateral breast during radiation therapy for breast cancer
Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,
More informationEQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS
PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich
More information3D heterogeneous dose distributions for total body irradiation patients
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 12, NUMBER 3, summer 2011 3D heterogeneous dose distributions for total body irradiation patients Marie-Claude Lavallée, 1,2a Sylviane Aubin, 1 Marie
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationSpinal Cord Doses in Palliative Lung Radiotherapy Schedules
Journal of the Egyptian Nat. Cancer Inst., Vol. 8, No., June: -, 00 Spinal Cord Doses in Palliative Lung Radiotherapy Schedules HODA AL-BOOZ, FRCR FFRRCSI M.D.* and CAROL PARTON, Ph.D.** The Departments
More informationA dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy
A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy Poster No.: R-0008 Congress: 2014 CSM Type: Scientific Exhibit Authors: T. Tran, W. Ding, L. Melvern, M. Chao, B. Subramanian;
More information8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit
MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.
More informationTo Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence
SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of
More informationDiode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment
Journal of Physics: Conference Series PAPER OPEN ACCESS Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment To cite this article: R Marzuki et al 2018 J. Phys.:
More informationDetector comparison for small field output factor measurements with flattening filter free photon beams
Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationOutline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams
Chapter 12 Treatment Planning Combination of Beams Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. http://www.utoledo.edu/med/depts/radther Outline Combination
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department
More informationExtracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas Thomas Samuel Ram, a Paul B. Ravindran,
More informationAn anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery
An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,
More informationDose rate response of Digital Megavolt Imager detector for flattening filter-free beams
Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationExperimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters for Off-Axis Wedged Fields in Radiation Therapy
Iranian Journal of Medical Physics Vol. 12, No. 4, Autumn 201, 262-270 Received: August 1, 201; Accepted: March 12, 2016 Original Article Experimental Evaluation of Depth Dose by Exit Surface Diode Dosimeters
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.
ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael
More informationIMRT QA: Point Dose Measurements or 2D Array?
QA for IMRT IMRT QA: Point Dose Measurements or D Array? General Linac QA for IMRT MLC checks Dose linearity at low MU Patient-specific measurements Performed using various methods Purpose is to verify
More informationAAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction
AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall
More informationDosimetric characteristics of intensity-modulated radiation therapy and RapidArc therapy using a 3D N-isopropylacrylamide gel dosimeter
Applications of Nuclear Techniques (CRETE15) International Journal of Modern Physics: Conference Series Vol. 44 (2016) 1660221 (9 pages) The Author(s) DOI: 10.1142/S2010194516602210 Dosimetric characteristics
More informationDosisverifikation mit Delta 4 Discover während der Behandlung
Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover
More informationEvaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationRADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident
Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate
More informationRadiochromic film dosimetry in water phantoms
INSTITUTE OF PHYSICS PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY Phys. Med. Biol. 46 (2001) N27 N31 www.iop.org/journals/pb PII: S0031-9155(01)16858-2 NOTE Radiochromic film dosimetry in water phantoms
More informationAmendment No. 2. Item No. 2 (Rfx/ Event number )
Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have
More informationMAX-HD SRS PHANTOM THE COMPREHENSIVE END-TO-END SRS PHANTOM SCAN PLAN LOCALIZE TREAT. distributed by:
SRS PHANTOM SCAN PLAN LOCALIZE TREAT THE COMPREHENSIVE END-TO-END SRS PHANTOM distributed by: Tel: +33 (0) 42 88 68 41 info@orion-france.com www.orion-france.com 2, Avenue du General Balfourier 75016 Paris,
More informationTLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience
IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationMEDICAL PHYSICS INTERNATIONAL Journal, vol.3, No.2, 2015 HOW TO
HOW TO QUALITY CONTROL AND PRE-TREATMENT QUALITY ASSURANCE APPLICATION OF EPID (as1000) FOR FF AND FFF BEAM VMAT PLANS Y. Mekuria 1, M. Bjorkqvist 1, J. Kulmala 1 1 Turku University Hospital/Radiotherapy
More informationPhantom Positioning Variation in the Gamma Knife Perfexion Dosimetry
Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View
More informationProtons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.
Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic
More informationFETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS
FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção
More informationTG-191 The clinical use of luminescent dosimeters
TG-191 The clinical use of luminescent dosimeters Task Group Members Stephen F. Kry, chair MD Anderson Cancer Center Paola Alvarez, MD Anderson Cancer Center Joanna Cygler, Ottawa Hospital Cancer Center
More information8/1/2016. TG-191 The clinical use of luminescent dosimeters. Task Group Members. Scope of Report
TG-191 The clinical use of luminescent dosimeters Task Group Members Stephen F. Kry, chair MD Anderson Cancer Center Paola Alvarez, MD Anderson Cancer Center Joanna Cygler, Ottawa Hospital Cancer Center
More informationLong term OSLD reader stability in the ACDS level one audit
Australas Phys Eng Sci Med (2015) 38:151 156 DOI 10.1007/s13246-014-0320-7 TECHNICAL PAPER Long term OSLD reader stability in the ACDS level one audit Andrew D. C. Alves Jessica Lye John Kenny Leon Dunn
More informationThe effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015 The effect of body contouring on the dose distribution delivered with volumetric-modulated arc therapy technique Jaegi Lee, 1,2 Jong
More informationAn Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements
An Active Matrix Flat Panel Dosimeter AMFPD for in-phantom dosimetric measurements Jean M. Moran, a Donald A. Roberts, Teamour S. Nurushev, Larry E. Antonuk, Youcef El-Mohri, and Benedick A. Fraass Department
More informationUse of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments
Use of Electronic Portal Image Detectors for Quality Assurance of Advanced Treatments Jean M. Moran, Ph.D. Department of Radiation Therapy, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationVolumetric Modulated Arc Therapy - Patient Specific QA
Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck
More informationIntroduction. Modalities used in imaging guidance. Flat panel detector. X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy
X-ray Imaging Dose to Patients in the Era of Image-Guided Radiation Therapy George Ding, Ron Price, Charles Coffey Vanderbilt-Ingram Cancer Center Vanderbilt University Medical Center, Nashville, TN Introduction
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationDosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.87 Dosimetric Analysis of Respiratory-Gated RapidArc with Varying Gating Window Times Mee
More informationA Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer
IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationAbstract: A National Project for the in-vivo dosimetry in radiotherapy: first results
Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità
More informationAbsolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro
In Phantoms Vivo treatment for Absolute Dosimetry verification Versatile solid and water phantoms Introduction Introduction Introduction Introtro Intro Intro The rapid development of advanced treatment
More informationIMRT point dose measurements with a diamond detector
IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,
More informationLeila E. A. Nichol Royal Surrey County Hospital
2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationIAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*
THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service
More informationROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationFirst application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities
JBUON 2018; 23(1): 268-272 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com SHORT COMMUNICATION First application of hemi-body electron beam irradiation for Kaposi
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More information3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe
3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve
More informationVerification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging
Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More information