Detector comparison for small field output factor measurements with flattening filter free photon beams

Size: px
Start display at page:

Download "Detector comparison for small field output factor measurements with flattening filter free photon beams"

Transcription

1 Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Austria 2 International Atomic Energy Agency, A-1400 Vienna, Austria

2 Acknowledgements Visit Hugo Palmans, Peter Sharpe, Rebecca Nutbrown, Clare Gouldstone, Russel Thomas, Martin Kelly, Joanna Izewska, Godfrey Azangwe, The financial support by the Federal Ministry of Economy, Family and Youth and the National Foundation for Research, Technology and Development is gratefully acknowledged. 2

3 Motivation Properties of FFF-beams: Higher dose rate, reduced scatter FFF Small fields dosimetric errors are considerably larger Formalism for reference dosimetry of small and nonstandard fields Alfonso et al. (2009) Med Phys 35(11) Which detectors are suitable? Are the detector correction factors of small fields output factor measurements different for FF- and FFF-beams? 3

4 Setup for output factor (OF) measurements Elekta Precise Linac 6MV and 10MV FF- and FFF-beams BrainLab M3 µmlc for field shaping Field sizes: 10x10cm² to 0.6x0.6cm² 5 measurements per field Electrometer: PTW Unidos Webline Bragg peak chamber as reference 6MV FFF 6MV FF 10MV FFF 10MV FF TPR 20/

5 Detectors 14 detectors classified according to their active volume Volume averaging correction factors were calculated for each detector Micro Mini Macro V < 10 mm³ 10 mm³ V 40 mm³ V > 40 mm³ PTW microdiamond IBA CC01 PTW Semiflex IBA SFD PTW PinPoint14 IBA CC13 PTW Diode P PTW PinPoint16 IBA IC10 IBA EFD IBA CC04 NPL 2611 IBA PFD PTW microlion Beam profiles were acquired prior to measurements detector was positioned at the center of the beam 5

6 Detectors Reference dosimeter: Alanine in solid water phantom, same geometry Field width Alanine 5 mm Alanine 2.5 mm 1.8 cm X 1.2 cm X X 0.6 cm X Evaluation: Output factors (OF) were normalized to 3x3 cm² The ratio of the OF measured with detectors to OF measured with alanine was evaluated S S cp cp (detector (alanine ) ) 6

7 Qualitative results Solid state detectors tended to overestimate the small field output factors Ionization chambers tended to underestimate the small field output factors Similar behavior for FF- and FFF-beams 7

8 Results ionization chambers CC01 agreed within2% with alanine; tended to over-respond for larger field sizes PinPoint 16 underestimated OF by up to 9%; agreed well for larger field sizes No substantial difference between FF- and FFF-beams 8

9 Results shielded diodes Substantial difference between 10MV FF- and FFF- beams at 0.6x0.6cm² 9

10 Results shielded diodes Substantial difference between 10MV FF- and FFF- beams at 0.6x0.6cm² Effect not present for 6MV 10

11 Results Volume averaging correction factor No correction was necessary for field widths larger than 1.8cm Field width Micro Mini Macro 0.6 cm Uncertainties (1SD) major component was the positioning uncertainty Field width Micro Mini Macro 0.6 cm 1% 1.2% 2% 11

12 Summary & Conclusion The qualitative behavior of detectors is similar for FF- and FFF-beams Substantial difference between FF- and FFF-beams was only observed for shielded diodes irradiated with 10 MV There might be a difference between correction factors for FF- and FFFbeams for other detectors, but it could not be resolved in a clinical environment Macro detectors should not be used for small field dosimetry Thanks for your attention! 12

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Rhys Carrington, Ysbyty Glan Clwyd, Bodelwyddan. MSc Medical Radiation Physics. Small field dosimetry

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

Basic dosimetric data for treatment planning system

Basic dosimetric data for treatment planning system Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny

More information

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam

Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Beam Quality Effects in Nonstandard Fields of the Varian TrueBeam Megan A. Hyun, M.S. Under the supervision of Larry A. DeWerd, Ph.D. NCCAAPM Fall Meeting October 30, 2015 Medical Radiation Research Center

More information

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn

MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)

More information

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada

IAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy

IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK

More information

Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators

Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 3, 2016 Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators Madelaine K. Tyler,

More information

Small Field Dosimetry

Small Field Dosimetry Small Field Dosimetry Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Emeritus Professor, Department of Medical Physics University of Wisconsin Madison WI Conflict

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane

Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors

More information

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning

Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Small Field Dosimetry Issues: Theory, Calibration and Treatment Planning Thomas Rockwell Mackie Director of Medical Devices Morgridge Institute for Research and Professor, Department of Medical Physics

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe

3D printing of air slab caps for correction free small field dosimetry. Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe 3D printing of air slab caps for correction free small field dosimetry Benjamin Perrett, Paul Charles, Tim Markwell, Tanya Kairn, Scott Crowe Small fields? Defined as fields sizes too small to achieve

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation

Ion recombination corrections of ionization chambers in flattening filter-free photon radiation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy

Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,

More information

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment

Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment Journal of Physics: Conference Series PAPER OPEN ACCESS Diode Dosimetric Characteristics Assessment for In-Vivo Dosimetry in Radiotherapy Treatment To cite this article: R Marzuki et al 2018 J. Phys.:

More information

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR.

PRECISE DOSIMETRY OF SMALL PHOTON BEAMS COLLIMATED BY AN ADD-ON DYNAMIC MICRO-MULTILEAVES COLLIMATOR. IX Latin American IRPA Regional Congress on Radiation Protection and Safety - IRPA 2013 Rio de Janeiro, RJ, Brazil, April 15-19, 2013 SOCIEDADE BRASILEIRA DE PROTEÇÃO RADIOLÓGICA - SBPR PRECISE DOSIMETRY

More information

SRS/SBRT Errors and Causes

SRS/SBRT Errors and Causes SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research

More information

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D.

ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA. By Vibha Chaswal, Ph.D. ARCCHECK: COMPREHENSIVE EVALUATION OF THE DIODE ARRAY PHANTOM - RULES OF THUMB FOR PHANTOM USE FOR QA By Vibha Chaswal, Ph.D. ACKNOWLEDGEMENT Dr. Nilendu Gupta, Dr. Arnab Chakrabarti Dr. Yi Rong, Michael

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition

Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Small Field Dosimetry: An Overview of Reference and Relative Dosimetry in Non-Equilibrium Condition Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of

More information

Protection of the contralateral breast during radiation therapy for breast cancer

Protection of the contralateral breast during radiation therapy for breast cancer Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,

More information

Progress Report on Radiation Dosimetry at NPL

Progress Report on Radiation Dosimetry at NPL Progress Report on Radiation Dosimetry at NPL Peter Sharpe and Simon Duane National Physical Laboratory Teddington, UK April 2009 1 Introduction This report gives a brief overview of radiation dosimetry

More information

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria

Karen E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)

More information

Small-Field Dosimetry: Its Importance In Clinic

Small-Field Dosimetry: Its Importance In Clinic Indra J. Das/ Small Fields (Page-1) Small-Field Dosimetry: Its Importance In Clinic isadministration edia Coverage IJDas (1) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana

More information

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)

Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1) Indra J. Das/ Small Fields (Page-1) Small Field Dosimetry: Overview of AAP TG-155 and the IAEA-AAP Code of Practice (Therapy) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana

More information

Dosimetry of Small Static Fields Used in External Beam Radiotherapy

Dosimetry of Small Static Fields Used in External Beam Radiotherapy Technical Reports SeriEs No.483 Dosimetry of Small Static Fields Used in External Beam Radiotherapy An International Code of Practice for Reference and Relative Dose Determination Sponsored by the IAEA

More information

Detectors assessment for stereotactic radiosurgery with cones

Detectors assessment for stereotactic radiosurgery with cones Received: 27 April 18 Revised: 6 August 18 Accepted: 16 August 18 DOI: 1.2/acm2.12449 RADIATION ONCOLOGY PHYSICS Detectors assessment for stereotactic radiosurgery with cones Nicolas Garnier 1,2 Régis

More information

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)

Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC) Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf

More information

This article was originally published in a journal published by Elsevier, and the attached copy is provided by Elsevier for the author s benefit and for the benefit of the author s institution, for non-commercial

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}

8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS} High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics

More information

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry

Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View

More information

Dosimetry and QA of proton and heavier ion beams

Dosimetry and QA of proton and heavier ion beams Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

The determination of timer error and its role in the administration of specified doses

The determination of timer error and its role in the administration of specified doses JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry

Semiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM

tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM tsmall FIELD DOSE CALIBRATIONS WITH GAFCHROMIC FILM A Thesis Presented to The Academic Faculty by Ryan J. Underwood In Partial Fulfillment of the Requirements for the Degree Master of Science in the School

More information

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields

Dosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy

Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit

8/2/2017. Objectives. Disclosures. Clinical Implementation of an MR-Guided Treatment Unit MR-Linac is a research programme. It is not available for sale and its future availability cannot be guaranteed 8/2/2017 Clinical Implementation of an MR-Guided Treatment Unit Geoffrey S. Ibbott, Ph.D.

More information

Quality Assurance of Helical Tomotherapy Machines

Quality Assurance of Helical Tomotherapy Machines Quality Assurance of Helical Tomotherapy Machines Wolfgang Tomé 1,2, Hazim Jaradat 2, David Westerly 1, John Fenwick 3, Nigel Orton 2, Rock Mackie 1,2, and Bhudatt Paliwal 1,2 1 Human Oncology and 2 Medical

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Future upcoming technologies and what audit needs to address

Future upcoming technologies and what audit needs to address Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple

More information

Progress Report on Radiation Dosimetry at NPL

Progress Report on Radiation Dosimetry at NPL Progress Report on Radiation Dosimetry at NPL Peter Sharpe and Simon Duane National Physical Laboratory Teddington, UK March 2013 1 Introduction This report gives a brief overview of radiation dosimetry

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment

A comparative study between flattening filter-free beams and flattening filter beams in radiotherapy treatment Oncology and Translational Medicine DOI 10.1007/s10330-016-0183-3 December 2017, Vol. 3, No. 6, P260 P266 REVIEW ARTICLE A comparative study between flattening filter-free beams and flattening filter beams

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

A national dosimetric audit of VMAT and Tomotherapy in the UK

A national dosimetric audit of VMAT and Tomotherapy in the UK A national dosimetric audit of VMAT and Tomotherapy in the UK Catharine Clark 1,2,3, M Hussein 1,4, Y Tsang 3,5, D Wilkinson 3,5, R Thomas 2, J Snaith 2, C Gouldstone 2, G Bass 2, S Bolton 6,7, A Nisbet

More information

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2. Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,

More information

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy

Can we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE

CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of

More information

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams

AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams AAPM s TG 51 Protocol for Clinical Reference Dosimetry of High-Energy Photon and Electron Beams Peter R. Almond, Brown Cancer Center,Louisville, KY 40202 Peter J. Biggs, Department of Radiation Oncology,

More information

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Development of Graphite Calorimetry at the NPL for Proton Beam Therapy

Development of Graphite Calorimetry at the NPL for Proton Beam Therapy Development of Graphite Calorimetry at the NPL for Proton Beam Therapy Russell A S Thomas Senior Research & Clinical Scientist Radiation Dosimetry Group 1 Background Current protocols based on chambers

More information

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom

Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable Phantom Volume 3, No 1/2016 Research Article Malays. j. med. biol. res. ISSN 2313-0008 (Print); ISSN 2313-0016 (Online) Comparative Study of Absorbed Doses in Different Phantom Materials and Fabrication of a Suitable

More information

Much of what we will say

Much of what we will say New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA

CHAPTER 5. STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA CHAPTER 5 STUDY OF ANGULAR RESPONSE OF asi 1000 EPID AND IMATRIXX 2-D ARRAY SYSTEM FOR IMRT PATIENT SPECIFIC QA 5.1 Introduction With the advent of new techniques like intensity modulated radiotherapy

More information

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical

More information

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter

Investigation of the clinical performance of a novel solid-state diagnostic dosimeter JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Investigation of the clinical performance of a novel solid-state diagnostic dosimeter Jason Tse, a Donald McLean Medical Physics and

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance

Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training

More information

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations

Clinical Implications of High Definition Multileaf Collimator (HDMLC) Dosimetric Leaf Gap (DLG) Variations Original Article PROGRESS in MEDICAL PHYSICS 27(3), Sept. 2016 http://dx.doi.org/10.14316/pmp.2016.27.3.111 pissn 2508-4445, eissn 2508-4453 Clinical Implications of High Definition Multileaf Collimator

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Zakithi Msimang SAAPMB/SARPA 2011

Zakithi Msimang SAAPMB/SARPA 2011 THE IAEA/WHO TLD POSTAL DOSE AUDIT SERVICE Zakithi Msimang SAAPMB/SARPA 2011 Introduction The idea of organizing a dosimetry audit for radiotherapy centres by the IAEA, was suggested for the first time

More information

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams

Dose rate response of Digital Megavolt Imager detector for flattening filter-free beams Received: 12 May 2017 Revised: 23 October 2017 Accepted: 10 April 2018 DOI: 10.1002/acm2.12358 RADIATION ONCOLOGY PHYSICS Dose rate response of Digital Megavolt Imager detector for flattening filter-free

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D.

7/31/2012. Volumetric modulated arc therapy. UAB Department of Radiation Oncology. Richard Popple, Ph.D. UAB Department of Radiation Oncology Volumetric modulated arc therapy Richard Popple, Ph.D. Disclosures UAB has research agreements with Varian Medical Systems Speaking honoraria from Varian Medical Systems

More information

Evaluation of dosimetric characteristics of diodes and ionization chambers in small megavoltage photon field dosimetry

Evaluation of dosimetric characteristics of diodes and ionization chambers in small megavoltage photon field dosimetry Volume 16, No 3 International Journal of Radiation Research, July 2018 Evaluation of dosimetric characteristics of diodes and ionization chambers in small megavoltage photon field dosimetry H. Keivan1,2,

More information

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry C. Candela-Juan, J. Vijande, T. García-Martínez, Y. Niatsetski,

More information

Radiation treatment planning in lung cancer

Radiation treatment planning in lung cancer Radiation treatment planning in lung cancer Georg Dietmar 1,2 1 Div. Medical Rad. Phys., Dept. of Radiation Oncology / Medical Univ. Vienna & AKH Wien 2 Christian Doppler Laboratory for Medical Radiation

More information

MAY NOT BE SAME AS THE PUBLISHED VERSION

MAY NOT BE SAME AS THE PUBLISHED VERSION MAY NOT BE SAME AS THE PUBLISHED VERSION CCRI supplementary comparison of standards for absorbed dose to water in 60 Co gamma radiation at radiation processing dose levels D.T. Burns a,, P.J. Allisy-Roberts

More information

Cell survival following high dose rate flattening filter free (FFF) and conventional dose rate irradiation

Cell survival following high dose rate flattening filter free (FFF) and conventional dose rate irradiation Cell survival following high dose rate flattening filter free (FFF) and conventional dose rate irradiation Peter Sminia p.sminia@vumc.nl Λαβορατοριυµβεσπρεκινγ 8 νοϖεµβερ 2005 Progress in Radiotherapy:

More information

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa

Analysis of 6 MV Energy Quality File Index using Percentage Depth Dose (PDD) and Tissue Phantom Ratio (TPR) Methods on Linac Siemens and Electa International Journal of Sciences: Basic and Applied Research (IJSBAR) ISSN 237-4531 (Print & Online) http://gssrr.org/index.php?journal=journalofbasicandapplied ---------------------------------------------------------------------------------------------------------------------------

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information