Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
|
|
- Mercy Weaver
- 5 years ago
- Views:
Transcription
1 EURAMED,MELODI and EURADOS Workshop, Helsinki Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical Physicist Kuopio University Hospital Cancer Center Kuopio, Finland
2 Out-of-field doses In radiotherapy high doses of radiation are delivered safely to the tumour. Combining both accurate in- and out-of-field dose calculation is challenging and usually the focus is to determine the in-field doses as accurately as possible. For example the Monte Carlo models used in radiotherapy are always optimized for accurate in-field dosimetry and unfortunately the out-of-field doses are and has been with less concern. However, the out-of-field doses have become more important with the increased survival of the treated cancer patients. The modeling and minimizing the incidence of second cancer has become more and more relevant. The out-of-field doses are the sum of collimator and phantom (patient) scatter and leakage radiation. The smaller the beam size becomes, the higher the demand for the out-of-field dosimetry is, since more radiation will be accumulated from collimator scatter and leakage. 2
3 Background OPERRA project: QA guidance and procedures for biobank of samples of patients exposed to medical radiation in CT examinations and radiotherapy. Validation of several biomarkers with prostate cancer patients treated with stereotactic radiotherapy of 5 x 7.25 Gy with CyberKnife system. Several biomarkers will be analysed from the samples (blood and hair). The biomarkers include free DNA and metabolomics, immunology related markers and mitochondrial DNA analyses of hair samples, mirna, phospho-atm, Raman spectroscopy and microparticles. Classical dicentric assay will be used as a reference for the biomarkers validated. A need to define the integral dose of the treated prostate cancer patients with high accuracy. Both the doses from the planning CT and actual radiotherapy treatments will be investigated. The determination of integral doses of the patients require measurements since treatment planning software are known to underestimate peripheral doses. 3
4 CyberKnife system X-ray tubes Linear accelerator Robot X-ray detectors The treatment accuracy is based on the x-ray imaging system integrated to the robotics with matching coordinates.
5 CyberKnife beam collimation With CyberKnife device the radiation beam is collimated with circular collimators (diameters of 5 60 mm). Fixed collimators Iris-collimator
6 CyberKnife treatment planning CyberKnife has only 6 MV energy Inverse optimization of the dose distributions. The plan optimization defines the optimal beam directions and the user can only define from what directions the beams are not allowed to intersect the patient. The final treatment plan consists of about 200 non-coplanar radiation beams. The treatment time with prostate cancer patients is usually about 30 minutes. 6
7 Aim Determine the integral dose (blood) and dose to organs at risk in the low dose regions in STR of prostate cancer Determine the accuracy of MultiPlan dose calculation algorithms (Monte Carlo and Raytracing) in the out-of-field region Estimate the contribution of out-of-field doses to total integral dose
8 Measurement Set Up IBA blue phantom PTV farmer ionization chamber IBA cc13 ionization chamber NE electrometer EBT3 film Glass dosimetry
9 Measurement Set Up
10 Measurements Measurement of beam profiles (scan width 460 mm) and depth doses 50 mm and 15 mm collimators Iris and Fixed To estimate head scattering/leakage, measurements were repeated with block (field size 0 mm) Point measurements with farmer chamber and NE electrometer-> central axis and 100 mm and 200 mm off axis Three different SSD distance 785 (standard in QA), 650 and 500 mm Farmer and cc13 ionization chambers (volume effect vs. sensitivity) MultiPlan Monte Carlo and Raytracing algorithms
11 Depth dose and Profiles
12 SSD 785 cm, 50 mm Iris collimator Depth 15 mm Depth 100 mm Depth 300 mm
13 SSD 785 cm, 15 mm Iris collimator Depth 15 mm Depth 100 mm Depth 300 mm
14 RT MC Out-of-Field doses Table: Difference in measured and calculated doses in central axis and out-of field region. Measurement Depth 100 mm. 50 mm 15 mm IR DCA (mm) MC 0 0.8% 1.0% 0.4% -0.9% 0.5% -1.1% RT 0 0.0% -0.1% -0.6% 0.1% 1.3% 0.6% % 32.1% 33.8% 65.6% 67.7% 74.0% % 58.1% 60.5% 89.5% 91.0% 94.5% % % % 80.5% 72.7% 69.7% % % % 70.3% 64.0% 61.0%
15 RT MC Out-of-Field doses Table: Difference in measured and calculated doses in central axis and out-of field region. Measurement Depth 100 mm. 50 mm 15 mm IR DCA (mm) MC 0 0.8% 1.0% 0.4% -0.9% 0.5% -1.1% RT 0 0.0% -0.1% -0.6% 0.1% 1.3% 0.6% % 32.1% 33.8% 65.6% 67.7% 74.0% % 58.1% 60.5% 89.5% 91.0% 94.5% % % % 80.5% 72.7% 69.7% % % % 70.3% 64.0% 61.0%
16 RT MC Out-of-Field doses Table: Difference in measured and calculated doses in central axis and out-of field region. Measurement Depth 100 mm. 50 mm 15 mm IR DCA (mm) MC 0 0.8% 1.0% 0.4% -0.9% 0.5% -1.1% RT 0 0.0% -0.1% -0.6% 0.1% 1.3% 0.6% % 32.1% 33.8% 65.6% 67.7% 74.0% % 58.1% 60.5% 89.5% 91.0% 94.5% % % % 80.5% 72.7% 69.7% % % % 70.3% 64.0% 61.0%
17 RT MC Out-of-Field doses Table: Difference in measured and calculated doses in central axis and out-of field region. Measurement Depth 100 mm. 50 mm 15 mm IR DCA (mm) MC 0 0.8% 1.0% 0.4% -0.9% 0.5% -1.1% RT 0 0.0% -0.1% -0.6% 0.1% 1.3% 0.6% % 32.1% 33.8% 65.6% 67.7% 74.0% % 58.1% 60.5% 89.5% 91.0% 94.5% % % % 80.5% 72.7% 69.7% % % % 70.3% 64.0% 61.0%
18 Integral Dose Integral dose was determined as product of dose and volume (Gy*dm^3) Integral dose was determined based on beam profiles measured with Farmer ionization chamber Volume effects were taken into account
19 Integral Dose 50 mm 205 mm - 50 mm IR collimator - SSD 785 In Filed Region Out of Field Region Dose (Gy)
20 Integral Dose Table: Difference (%) between measured and calculated (MC) integral dose 50 mm collimator 15 mm collimator Coll/ SSD (mm) Iris 12.2% 13.0% 18.1% 32.2% 36.0% 50.4% Fixed 13.9% 16.0% 19.7% 49.4% 55.2% 63.5%
21 Out of Field doses and Head Scattering
22 Evaluation of film and glass dosimetry 50 mm Iris Collimator 15 mm 15 mm Iris Collimator
23 Conlcusions Monte Carlo algorithm underestimate the out-of-field dose, and thus, the ID Raytracing more or less useless Contribution of head scattering/ leakage to measured out-of-field doses might be significant Film dosimetry seems to be feasible method for measurement of out-of-field doses
24 Future work Development of robust correction factor for integral and organ dose calculation Validation of correction factor with film (MOSFET and RPL) measurements in anthropomorphic fantom Full Monte Carlo algorithm with more accurate modelling of out-of-field areas Contribution of imaging dose in the total integral dose -> film measurements
25
Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationThe Journey of Cyberknife Commissioning
The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources
More information4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest
SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific
More informationImplementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy
Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out
More informationSRS/SRT Treatment Planning for Skull Base Meningioma
SRS/SRT Treatment Planning for Skull Base Meningioma Janne Heikkilä, Medical Physicist, PhD Centre of Oncology, Kuopio University Hospital, Kuopio, Finland www.cyberknifekuopio.fi cyberknife@kuh.fi 9.11.2018
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More informationVerification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams
Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationFilm-based dose validation of Monte Carlo algorithm for Cyberknife system with a CIRS thorax phantom
Received: 4 August 2017 Revised: 1 January 2018 Accepted: 6 February 2018 DOI: 10.1002/acm2.12314 RADIATION ONCOLOGY PHYSICS Film-based dose validation of Monte Carlo algorithm for Cyberknife system with
More informationBasic dosimetric data for treatment planning system
Measurement of relative output factors for Leksell Gamma Knife Perfexion, BrainLAB m3 multi-leaf collimator and Accuray CyberKnife fixed and Iris collimators by PTW 60019 MicroDiamond detector J. Novotny
More informationPatient-specific quality assurance for intracranial cases in robotic radiosurgery system
JBUON 2018; 23(1): 179-184 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Patient-specific quality assurance for intracranial cases in robotic
More informationLung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014
Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this
More informationProstate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014
Prostate Phantom Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014 GENERAL INFORMATION: Each institution may keep the phantom for a period of time no more than 2 weeks.
More informationIntensity Modulated RadioTherapy
Intensity Modulated RadioTherapy A clinical application of a small accelerator University Medical Center Groningen A.A. van t Veld PhD I. Hoveijn PhD part 1 part2 CERN/KVI accelerator school, Zeegse, June
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationCyberKnife Technology in Ablative Radiation Therapy. Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017
CyberKnife Technology in Ablative Radiation Therapy Jun Yang PhD Cyberknife Center of Philadelphia Drexel University Jan 2017 Objectives Components and work flow of CyberKnife Motion management of CyberKnife
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationAssessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital
Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationIROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014
IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationHampton University Proton Therapy Institute
Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationIROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014
IROC Prostate Phantom Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 The study groups are requesting that each institution keep the phantom for a period of time
More informationA Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer
IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer
More informationSTEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS
PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic
More informationOut-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison
MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison
More informationNormal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT
Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationLevel of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013
Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationA Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer
A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer Tony Kin Ming Lam Radiation Planner Dr Patricia Lindsay, Radiation Physicist Dr John Kim, Radiation Oncologist Dr Kim Ann Ung,
More informationIndependent Dose Verification for IMRT Using Monte Carlo. C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA
Independent Dose Verification for IMRT Using Monte Carlo C-M M Charlie Ma, Ph.D. Department of Radiation Oncology FCCC, Philadelphia, PA 19111, USA Outline Why Monte Carlo for IMRT QA? Experimental verification
More informationProduction and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study
Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad
More informationEXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *
Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West
More informationJOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the
More informationVerification of Advanced Radiotherapy Techniques
Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationQuality Assurance of TPS: comparison of dose calculation for stereotactic patients in Eclipse and iplan RT Dose
Petrovic B Comparison of dose calculation algorithms for stereotaxy Quality Assurance of TPS: comparison of dose calculation for stereotactic patients in and RT Dose Borislava Petrovic 1, Aleksandra Grządziel
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More information5th ADAMAS Workshop at GSI December 15-16, 2016, Darmstadt, Germany
Evaluation of 3D diamond detectors for application in medical radiation dosimetry K. Kanxheri (1,2), L. Servoli (2), C. Zucchetti (5), A. C. Dipilato (5), M. Iacco (5), S. Lagomarsino (3,4), A. Morozzi
More informationPHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)
PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce
More informationStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry
More informationClinical Implementation of SRS/SBRT
Clinical Implementation of SRS/SBRT Anil Sethi, PhD, FAAPM Loyola University Medical Center November 4, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Learning
More informationThe Physics of Oesophageal Cancer Radiotherapy
The Physics of Oesophageal Cancer Radiotherapy Dr. Philip Wai Radiotherapy Physics Royal Marsden Hospital 1 Contents Brief clinical introduction Imaging and Target definition Dose prescription & patient
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationStereotactic Body Radiotherapy for Lung Lesions using the CyberKnife of-the-art and New Innovations
Stereotactic Body Radiotherapy for Lung Lesions using the CyberKnife State-of of-the-art and New Innovations Chad Lee, PhD CK Solutions, Inc. and CyberKnife Centers of San Diego Outline Basic overview
More informationVolumetric Modulated Arc Therapy - Patient Specific QA
Volumetric Modulated Arc Therapy - Patient Specific QA Daliang Cao, PhD, DABR Swedish Cancer Institute, Seattle, WA VMAT plan QA methods Composite dose measurement Film & ion chamber diode array Mapcheck
More informationIROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015
IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During
More informationTechnical Study. Institution University of Texas Health San Antonio. Location San Antonio, Texas. Medical Staff. Daniel Saenz. Niko Papanikolaou.
Technical Study Stereotactic Radiosurgery with Elekta Versa HD and Monaco Accuracy of a single isocenter, multiple brain metastases VMAT plan delivered to a pseudo-patient dosimetric gel phantom Institution
More informationLearning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan
Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams
More informationA preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF
Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital
More informationAssessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 1, 2015 Assessment of a three-dimensional (3D) water scanning system for beam commissioning and measurements on a helical tomotherapy unit
More informationChapters from Clinical Oncology
Chapters from Clinical Oncology Lecture notes University of Szeged Faculty of Medicine Department of Oncotherapy 2012. 1 RADIOTHERAPY Technical aspects Dr. Elemér Szil Introduction There are three possibilities
More informationData Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff
Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment
More informationBorges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1
Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior
More informationDisclosures. Clinical Implementation of SRS/SBRT. Overview. Anil Sethi, PhD. Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch
Clinical Implementation of SRS/SBRT Anil Sethi, PhD Loyola University Medical Center August 3, 2017 Disclosures Speaker: BrainLAB Standard Imaging Research collaboration: RaySearch 2 Overview Physics Considerations
More informationDosimetric characterization of surface applicators for use with the Xoft ebx system
Dosimetric characterization of surface applicators for use with the Xoft ebx system R.M. Kennedy University of Wisconsin Medical Research Center Madison, WI April 23, 2010 Introduction Squamous and basal
More informationMania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane
Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors
More informationEvaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf Collimator (DMLC)
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 4, December, 2015 http://dx.doi.org/10.14316/pmp.2015.26.4.223 Evaluation of Dosimetric Characteristics of a Double-focused Dynamic Micro-Multileaf
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationS Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M
s y n c h r o n y r e s p i r a t o r y t r a c k i n g s y s t e m S Y N C H R O N Y R E S P I R A T O R Y T R A C K I N G S Y S T E M The Synchrony System tracks respiration in real time and automatically
More informationABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007
ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory
More informationIROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015
IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During
More informationIN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS
Romanian Reports in Physics, Vol. 64, No. 3, P. 728 736, 2012 MEDICAL PHYSICS IN VIVO DOSIMETRY MEASUREMENTS FOR BREAST RADIATION TREATMENTS G. VASILE 1,3, MADALINA VASILE 2,3, O.G. DULIU 3 1 Center of
More informationCyberknife Stereotactic Treatment
Cyberknife Stereotactic Treatment Eugene Lief, Ph.D. Christ Hospital Jersey City, New Jersey USA DISCLAIMER: I am not affiliated with any vendor and did not receive any financial support from any vendor.
More informationSRS Uncertainty: Linac and CyberKnife Uncertainties
SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationDosimetry of the Gamma Knife TM
Dosimetry of the Gamma Knife TM Sheridan Griffin Meltsner, Ph.D. Aurora Health Care St. Luke s Medical Center NCCAAPM Spring Meeting Outline Introduction to stereotactic radiosurgery Elekta Gamma Knife
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationMEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS. SB Crowe, E Whittle, C Jones, T Kairn
MEASUREMENT OF TMR AND PDD PROFILES IN SMALL FIELDS SB Crowe, E Whittle, C Jones, T Kairn SMALL FIELD DOSIMETRY The measurement of small field tissue-maximum ratio (TMR) and percentage depth dose (PDD)
More informationCharacterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning
Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca
More informationA TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,
More informationPlan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy
Plan-Specific Correction Factors for Small- Volume Ion Chamber Dosimetry in Modulated Treatments on a Varian Trilogy Vimal K. Desai, M.S. Under the supervision of Dr. Wesley Culberson NCCAAPM 2017 Fall
More informationDOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD
Proceedings of the Tenth EGS4 Users' Meeting in Japan, KEK Proceedings 2002-18, p.65-73 DOSE DISTRIBUTION ANALYZE OF THE BODY STI USED MONTE CARLO METHOD N. Tohyama, H. Saitoh, T. Fujisaki 1, S. Abe 1
More informationSOMATOM Drive System Owner Manual Dosimetry and imaging performance report
www.siemens.com/healthcare SOMATOM Drive System Owner Manual Dosimetry and imaging performance report Table of contents 1 Dosimetry and imaging performance report 5 1.1 Dose information 5 1.1.1 General
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationCT Quality Control Manual FAQs
CT Quality Control Manual FAQs General Question: How often will the QC Manual be updated and how will those updates be communicated? Answer: The ACR CT Physics Subcommittee will review any comments, issues
More informationElectron Beam ET - Reloaded Therapy - Reloaded
8/4/216 Electron Beam ET - Reloaded Therapy - Reloaded Qiuwen Wu, Ph.D. Department of Radiation Oncology Duke University 1 Introduction: e vs. γ vs. p Electron Photon Proton Depth dose and lateral profiles
More informationMonte Carlo for CyberKnife Incise TM MLC
Division of Medical Radiation Physics Monte Carlo for CyberKnife Incise TM MLC Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital, Bern University Hospital, and University
More informationIORT with mobile linacs: the Italian experience
IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams
More informationVerification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film
Verification of treatment planning system parameters in tomotherapy using EBT Radiochromic Film E.B.Rajmohan¹, Pratik Kumar¹, Bhudatt Paliwal,² David Westerly², N.Gopishankar³, R.K.Bisht³, D.Tewatia²,
More informationOnline Care, Online Control, Online Confidence
Online Care, Online Control, Online Confidence * Online Treatment Monitoring and much more! Designed for Efficiency provides you with online control and online confidence of treating your patients exactly
More informationGATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT
GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS
More informationEfficient Dosimetry for Proton Therapy
Efficient Dosimetry for Proton Therapy Why IBA Dosimetry? IBA Dosimetry offers the full product range tailored to fit any Proton Therapy QA needs. PT treatment safety, as well as most efficient dosimetry
More informationDosimetry for Epidemiology Cohorts Who Receive Radiation Therapy
Dosimetry for Epidemiology Cohorts Who Receive Radiation Therapy Wayne Newhauser, PhD Eurados Winter School, Milan, 2016 Introduction About 1 in 2 men and women born today will be diagnosed with some form
More informationIMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)
IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)
More informationAccounting for Imaging Dose
Accounting for Imaging Dose High Profile Over-exposures Lead to Growing Concern FDA issues warning in October 2009-209 patients exposed to 8 times typical dose for CT brain perfusion scan (3-4 Gy) - Some
More informationIMRT point dose measurements with a diamond detector
IMRT point dose measurements with a diamond detector Erin Barnett, Marc MacKenzie, B. Gino Fallone Department of Physics, University of Alberta, and Department of Medical Physics, Cross Cancer Institute,
More informationNIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA MAGELLAN HEALTH RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationTechnology Challenges of Commercial Medical Electron Accelerators
Technology Challenges of Commercial Medical Electron Accelerators John Allen Chief Engineer Elekta Ltd Aims of this talk Medical Radiotherapy is already a large and profitable business Well established
More informationRPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy
RPC Liver Phantom Highly Conformal Stereotactic Body Radiation Therapy Guidelines for Planning and Irradiating the RPC Liver Phantom. Revised Dec 2005 Credentialing for this protocol requires four steps:
More informationCalculation methods in Hermes Medical Solutions dosimetry software
Calculation methods in Hermes Medical Solutions dosimetry software Helena McMeekin MSc. Clinical Applications Scientist, Hermes Medical Solutions MRTDosimetry Scientific Workshop The Principals and Clinical
More information02 CyberKnife: Treatment Delivery
TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options
More informationCredentialing for the Use of IGRT in Clinical Trials
Credentialing for the Use of IGRT in Clinical Trials James M. Galvin, DSc Thomas Jefferson University Hospital Jefferson Medical College Philadelphia, PA and The Radiation Therapy Oncology Group RADIATION
More informationEffective and efficient radiotherapy dosimetry audit: Where to next?
Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More information