1 Hematopathology / Platelet Counting Accuracy in DIC and Acute Leukemia Accuracy of Platelet Counting by Automated Hematologic Analyzers in Acute Leukemia and Disseminated Intravascular Coagulation Potential Effects of Platelet Activation Seon Young Kim, MD, 1 Ji-Eun Kim, MS, 1,2 Hyun Kyung Kim, MD, PhD, 1,2 Kyou-Sup Han, MD,PhD, 1 and Cheng Hock Toh, MD, PhD 3 Key Words: Platelet count; Acute leukemia; Disseminated intravascular coagulation; Platelet activation DOI: 1.139/AJCP88JYLRCSRXPP Abstract Platelet counting in patients with acute leukemia or disseminated intravascular coagulation (DIC) may have a risk for erroneous counts owing to the presence of nonplatelet particles or platelet activation. We evaluated automated platelet counting methods using the Abbott Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA), Sysmex XE-21 (Sysmex, Kobe, Japan), ADVIA 212 (Siemens Diagnostics, Tarrytown, NY), and Beckman Coulter LH 75 (Beckman Coulter, Miami, FL) compared with the international reference method (IRM). Automated platelet counting methods were inaccurate compared with the IRM, without evidence of interfering nonplatelet particles. It is interesting that platelet activation markers were associated with DIC severity and erroneous platelet counting, suggesting that platelet activation is a potential source of inaccuracy. Furthermore, the artifactual in vitro platelet activation induced a high degree of intermethod variation in platelet counts. The inaccuracy of automated platelet counts increased the risk for misdiagnosis of DIC. More attention needs to be given to the accuracy of platelet counts, especially in clinical conditions with florid platelet activation. Measurement of platelet counts using automated hematology analyzers is usually quite precise and accurate. However, the accuracy of automated platelet counts can be compromised when measuring very low platelet counts or in the presence of interference from non-platelet particles or platelet abnormalities. 1 Recent studies, mainly focusing on the counts of low levels of platelets, demonstrated that automated counts were not as accurate in severely thrombocytopenic samples. 2-5 These findings are of concern because current clinical guidelines lowered the prophylactic platelet transfusion threshold to /L for patients without additional risk factors. 6 In addition to this limitation of the technology, automated platelet counts can be inaccurate even at normal or high platelet ranges owing to the characteristics of blood specimens, eg, in specimens with a substantial amount of interfering particles, including WBC fragments, RBC fragments, immune complexes, bacteria, lipid droplets, or protein aggregates. WBC fragments can cause the spurious elevation of platelet counts in patients with acute leukemia at diagnosis and during chemotherapy In 1 study, platelet-like fragments of leukemic blasts were observed in 25% of acute leukemia cases. 11 In addition, granulocyte fragments or microorganisms are possible sources of platelet count overestimation in septic patients RBC fragments, which are often observed in patients with malignancies as part of microangiopathic processes, are also recognized as a cause of erroneous platelet counts. 1 To improve the accuracy of platelet counting and the feasibility of comparing platelet counts between analyzers, the International Council for Standardization in Haematology (ICSH) and the International Society of Laboratory Hematology (ISLH) proposed flow cytometry analysis of monoclonal antibody labeled platelets and the calculation of 634 Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

2 Hematopathology / Original Article platelet counts from the platelet/rbc ratio as a new international reference method (IRM). 15 After the establishment of the IRM, multicenter studies comparing counting methods in platelet concentrates were performed. These studies identified considerable variation between the different counting principles and between different instruments using the same counting principle; this variability was suggested to be related to the increased proportion of activated small platelets When platelets are activated, they become spherical with a hypogranular cytoplasm and release small particles. This may lead to the erroneous detection of platelets when using automated hematology analyzers owing to their deformed morphology. Patients with acute leukemia or disseminated intravascular coagulation (DIC) have an increased risk not only for interference from nonplatelet particles but also for counting errors due to platelet activation because platelet activation is inevitable during the course of a disease in which high levels of thrombin are generated and many inflammatory cytokines induce platelet activation. Recognizing erroneous results of automated platelet counts in these situations is especially critical for a consistent decision in the diagnosis of DIC and for clinical decision making regarding transfusion. The platelet count is an indispensable parameter in the DIC scoring system proposed by the International Society on Thrombosis and Haemostasis Sub- Committee of the Scientific and Standardization Committee on DIC, in which platelet counts of less than /μl (1 1 9 /L) and less than /μl (5 1 9 /L) would score 1 and 2 points, respectively. 19 The aim of this study was to investigate the accuracy of platelet counting in patients with acute leukemia or suspected of having DIC, clinical conditions with a risk for erroneous platelet counts owing to nonplatelet particles or platelet activation. First, we compared 7 different automated platelet counting methods from 4 widely used hematology analyzers, Abbott Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA), Sysmex XE-21 (Sysmex, Kobe, Japan), ADVIA 212 (Siemens Diagnostics, Tarrytown, NY), and Beckman Coulter LH 75 (Beckman Coulter, Miami, FL), with the IRM. Second, to identify the factors responsible for inaccuracies in automated platelet counting, the interference of nonplatelet particles was examined in some specimens. Third, we analyzed the association of intermethod platelet count variation with platelet activation status. Fourth, we explored whether artificial in vitro platelet activation may induce intermethod variation of platelet counts. Finally, we assessed the impact of the inaccuracy in automated platelet counts on the diagnosis of overt DIC. Materials and Methods Study Population EDTA-anticoagulated whole blood samples submitted to the clinical laboratory for CBC counts were analyzed for 2 cohorts of patients. The acute leukemia group consisted of 44 samples from 21 patients with acute leukemia who showed increased peripheral blasts (>2% of WBCs). Besides 21 samples obtained at the initial examination of each patient, 23 samples were repeatedly collected from 1 patients immediately after the initiation of cytotoxic chemotherapy. These samples were included because previous reports suggested an increased incidence of leukemic fragments after chemotherapy. 1 The suspected-dic group consisted of 159 samples from the same number of patients who were clinically suspected of having DIC and had undergone a battery of DIC screening tests. The DIC scores were calculated according to the International Society on Thrombosis and Haemostasis scoring system (scores calculated from platelet count, prothrombin time, fibrinogen, and D-dimer), and overt DIC was defined as a score of 5 or more. 19 The suspected-dic group was divided into 3 subgroups according to an arbitrary DIC score (scores calculated from prothrombin time, fibrinogen, and D-dimer, regardless of platelet count) as follows: patients with an arbitrary score of 5 or more, definite DIC (overt DIC present, regardless of the platelet count); patients with an arbitrary score of 3 or 4, borderline DIC (the presence of overt DIC was dependent on the platelet count); and patients with an arbitrary score of 2 or less, unlikely DIC (no overt DIC, regardless of the platelet count). The characteristics of the patients and details of underlying disorders are shown in Table 1. Whole blood specimens from 15 healthy volunteers (25-4 years old; platelet count, /μl [ /L]) were collected with written informed consent for healthy control samples and an in vitro activation study. This study was reviewed and approved by the Seoul National University College of Medicine Institutional Review Board (Seoul, Republic of Korea). Automated Hematology Analyzers and the IRM Seven different counting methods from 4 automated hematology analyzers were used as follows: immunologic, optical, and impedance methods using the Abbott Cell-Dyn Sapphire (ImmPLT, CD-o, and CD-i, respectively); impedance and optical methods using the Sysmex XE-21 (XE-i and XE-o, respectively); optical methods using the ADVIA 212 (ADVIA); and the impedance method using a Beckman Coulter LH 75 (LH75). The ImmPLT method using the Cell-Dyn Sapphire is a fully automated method using a fluorescein isothiocyanate (FITC)-conjugated CD61 monoclonal antibody, and it generates direct reports of absolute counts. 3 Each analyzer was calibrated according to the manufacturer s guidelines, and quality controls were performed according to the standard procedures. The IRM was performed according to the recommended protocol of the ICSH/ISLH. 15 Two monoclonal antibodies, Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 635

3 Kim et al / Platelet Counting Accuracy in DIC and Acute Leukemia Table 1 Patient Characteristics * Suspected DIC (n = 159) Acute Leukemia Definite DIC Borderline DIC Unlikely DIC Total (n = 18) (n = 21) (n = 41) (n = 64) (n = 54) No. of samples collected Age (y) 59 (2-88) 36 (2-79) 59 (16-79) 62 (2-88) 62 (2-86) Male/female (% male) 113/67 (62.8) 11/1 (52) 24/17 (59) 39/25 (61) 39/15 (72) Platelet count ( 1 9 /L) 87.8 ( ) 72.1 ( ) 77. ( ) 82.7 ( ) ( ) DIC score 4 (-8) 3 (1-8) 6 (5-8) 4 (3-6) 2 (-4) Underlying disorders Hematologic malignancy 3 (16.7) 21 (1) () () 9 (17) Sepsis/infection 55 (3.6) () 13 (32) 26 (41) 16 (3) Solid tumors 42 (23.3) () 11 (27) 19 (3) 12 (22) Hepatic failure 26 (14.4) () 14 (34) 9 (14) 3 (6) Organ destruction 9 (5.) () 1 (2) 3 (5) 4 (7) Trauma 5 (2.8) () 1 (2) 1 (2) 4 (7) Vascular abnormalities 5 (2.8) () 1 (2) 2 (3) 2 (4) Severe immunologic reaction 5 (2.8) () () 2 (3) 3 (6) Obstetric calamities 3 (1.7) () () 2 (3) 1 (2) DIC, disseminated intravascular coagulation. * Data are shown as the median (range) for continuous variables or number (percentage) for categorical variables unless otherwise indicated. P <.5 by Mann-Whitney U test (continuous variable) or χ 2 test (categorical variable) compared with the Unlikely DIC group. Platelet counts and DIC score by the international reference method. Platelet counts are given in Système International units; to convert to conventional units ( 1 3 /μl), divide by 1.. Patients with lymphoma or leukemia who were in remission. anti CD61-FITC (Becton Dickinson, San Jose, CA) and anti CD41-FITC (Becton Dickinson) were used. Flow cytometry analysis was performed using a FACSCalibur with CellQuest software (Becton Dickinson). RBC/platelet ratios were calculated as described elsewhere, 15,2 with the known RBC count determined as the average of each RBC count obtained using the Cell-Dyne Sapphire impedance cell counter, XE-21, ADVIA 212, and LH75. Precision studies were performed on 3 to 1 replicates of 15 specimens with platelet counts ranging from 12 to /μl ( /L) and demonstrated a mean coefficient of variation (CV) of 5.4%. All procedures of platelet counting were performed within 8 hours after venipuncture. Investigation for the Presence of Interfering Nonplatelet Particles To determine whether WBC or RBC fragments or endothelial microparticles were present, we examined peripheral blood smears stained with Wright-Giemsa in all samples. We also performed flow cytometric analysis using the following monoclonal antibodies (all from Becton Dickinson): anti CD45-phycoerythrin (PE), anti CD33-PE, and anti CD54-PE for WBC fragments; anti glycophorin A FITC for RBC fragments; and anti CD62E-PE for endothelial microparticles. Investigation of Platelet Activation Parameters Platelet activation parameters, including mean fluorescence intensity of side scatter (MFI of SSC), mean platelet component concentration (MPC), and percentage of small platelets (pplt-s), were determined in all specimens at the same time platelet counts were measured. MFI of SSC in the gated CD41+/CD61+ platelet events was set out from the scattergram of flow cytometry in the IRM. 21 MPC, the mean refractive index of platelets, was obtained using the ADVIA These parameters reflect the internal complexity and density of platelets and decrease when platelets degranulate on activation. 21 The pplt-s was calculated by dividing the number of small platelets by the total platelet count in the ImmPLT method. 17 The small platelets, which are defined by CD61+ platelets below a fixed lower threshold in 7 light scatter, include platelet microparticle fractions and increase after platelet activation. 17,23 Effect of In Vitro Platelet Activation on Intermethod Variation The 15 EDTA-anticoagulated whole blood specimens from 1 healthy volunteers and 5 patients (platelet counts, /μl [ /L]; DIC scores, 2-4) were divided into aliquots in 3 tubes. Two aliquots from each specimen were incubated with.64 mmol/l fibrin inhibitor (glycine-proline-arginine-proline, Sigma Aldrich, St Louis, MO) for 5 minutes and were then stimulated with 3 or 1 μmol/l adenosine diphosphate (Sigma Aldrich) for 5 minutes at room temperature. The remaining aliquot was kept without manipulation. The platelet counts were obtained using the 7 automated methods and the IRM as described. The degree of platelet activation was estimated by flow cytometric analysis using anti annexin V PE and anti PAC-1-FITC (Becton Dickinson), as well as by the platelet activation parameters obtained from the automated analyzers. 636 Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

4 Hematopathology / Original Article Statistical Analysis Data were compared by using the Mann-Whitney U test and Kruskal-Wallis analysis of variance for continuous variables and the χ 2 test for categorical variables. The comparison between the 2 methods was performed by using Passing-Bablok regression analysis, the Pearson correlation coefficient, and Bland-Altman analysis. Because the difference between the analyzers and the IRM increased as the platelet counts increased, percentage difference plots were used as suggested by Bland and Altman. 24 The 95% limits of agreement (95% LA) were calculated from the 95% range (mean ± 1.96 SD) of the percentage differences to quantify the degree of agreement between the 2 methods. The percentage of erroneous results between each automated method and the IRM was arbitrarily defined by the percentage of specimens that were outside the ± 25% error limits (± 25% difference from zero difference) of the platelet count. This error limit was modified from the Clinical Laboratory Improvement Amendments criteria for the acceptable performance of the quantitative hematology test in proficiency testing programs. 25 The CV across the 7 automated methods and the IRM (intermethod CV) were also calculated for each specimen. The percentage of erroneous results across the quartiles of each platelet parameter were analyzed with adjustments for age, sex, and platelet levels by using linear regression, and the test for linear trend of the percentage was performed on the basis of logistic regression. Statistical analyses were carried out using MedCalc, version 9.6 (MedCalc Software, Mariakerke, Belgium) and SAS, version 9.1 (SAS Institute, Cary, NC). P values less than.5 were considered statistically significant. Results Comparison of Platelet Counts Between Automated Methods and the IRM The Pearson correlation coefficients between each automated method and the IRM ranged from.943 to.976 Figure 1 (upper panels). The correlation was the strongest between ImmPLT and the IRM and weakest between CD-o and the IRM. Despite these good correlations, the slopes from Passing-Bablok regression were less than.9 (from.82 in XE-o to.89 in CD-i), except for ADVIA (.95) (Figure 1, upper panels), suggesting that there were considerable proportional errors in platelet counts in the Figure 1 Comparison of platelet counts between the Cell-Dyn Sapphire immunologic method (ImmPLT) (A), optical method (CD-o) (B), and impedance method (CD-i) (C); XE-21 impedance method (XE-i) (D) and optical method (XE-o) (E); ADVIA 212 (ADVIA) (F) and Coulter LH 75 (LH75) (G) with the international reference method (IRM) in all samples using Passing-Bablok regression (upper panels) and Bland-Altman analysis (lower panels). Shaded areas in the Bland-Altman plots indicate ± 25% error limits (± 25% difference from zero difference) of platelet counts. Upper panels: A, y =.867x 1.187; r =.976. B, y =.844x 1.91; r =.943. C, y =.89x 2.892; r =.971. D, y =.859x 1.587; r =.969. E, y =.821x 2.754; r =.97. F, y =.947x 2.474; r =.968. G, y =.844x 1.144; r =.971. Outside ± 25% error limits, lower panels: A, 19.7% (4/23). B, 39.9% (81/23). C, 14.7% (28/191). * D, 22.7% (46/23). E, 29.% (58/2). F, 13.4% (27/21). G, 28.% (56/2). * For the CD-i, 12 platelet counts were missing because the CD-i did not report platelet counts on samples with a platelet count of <2 1 9 /L. Results were not given for 4 XE-o, 2 ADVIA, and 3 LH75 samples owing to insufficient volume. Platelet counts are given in Système International units; to convert to conventional units ( 1 3 /μl), divide by 1.. IRM, international reference method. (Figure 1 continues on next 2 pages.) A % Difference (ImmPLT IRM) ImmPLT ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD Mean of ImmPLT and IRM ( 1 9 /L) Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 637

5 Kim et al / Platelet Counting Accuracy in DIC and Acute Leukemia B C 3 3 % Difference (CD-o IRM) CD-o ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD % Difference (CD-i IRM) CD-i ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD Mean of CD-o and IRM ( 1 9 /L) Mean of CD-i and IRM ( 1 9 /L) D E 3 3 % Difference (XE-i IRM) XE-i ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD % Difference (XE-o IRM) XE-o ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD Mean of XE-i and IRM ( 1 9 /L) Mean of XE-o and IRM ( 1 9 /L) 638 Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

6 Hematopathology / Original Article automated methods compared with the count in the IRM, except for ADVIA. The 95% LA on the Bland-Altman plots of the ImmPLT, CD-o, CD-i, XE-i, XE-o, ADVIA, and LH75 vs the IRM varied from 46% to +11%, 93% to +37%, 4% to +26%, 54% to +27%, 54% to +24%, 39% to +36%, and 61% to 26%, respectively (Figure 1, solid lines, lower panels). The 95% LA was the widest in the CD-o and narrowest in the ImmPLT. Erroneous platelet count, defined by the percentage of specimens that were outside the ± 25% error limits in the Bland-Altman plot, ranged from 13.4% (ADVIA) to 39.9% (CD-o) (Figure 1, shaded area, lower panels). Investigation of Interfering Nonplatelet Particles The examination of peripheral blood films revealed no definite evidence of interfering nonplatelet particles in any specimen. Flow cytometric analysis using anti-cd45 in 16 specimens from the acute leukemia (n = 31) and suspected-dic (n = 75) groups showed no WBC fragments in the platelet-sized area on the scattergram of the IRM. Analysis using anti-cd33 and anti-cd54 antibodies in 14 samples from the patients with acute myeloid leukemia showed no blast fragments. There were no platelet-sized fragments of endothelial cells labeled with anti-cd62e in 33 samples from the definite-dic group. Flow cytometric analysis with glycophorin A in 22 selected specimens also did not show any platelet-sized fragments of RBCs. Intermethod Variation of Platelet Counts in Terms of Underlying Clinical Condition The intermethod CV (mean ± SD) of platelet counts was significantly higher in the acute leukemia group (2.7% ± 12.2%) than in healthy control subjects (7.6% ± 1.7%; P <.1) Figure 2A. Within the suspected-dic group, the intermethod CV gradually increased as the DIC severity increased (unlikely DIC, 1.6% ± 4.5%; borderline DIC, 13.4% ± 7.4%; definite DIC, 17.4% ± 13.3% Figure 2B, Figure 2C, and Figure 2D. When the acute leukemia specimens were divided into tertiles on the basis of their platelet counts, the 95% LA tended to increase as the platelet levels decreased across all automated methods, and the 95% LA of CD-o was the widest among the automated methods (Figure 2A). Among the suspected-dic groups, the definite-dic group showed the widest 95% F G 3 3 % Difference (ADVIA IRM) ADVIA ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD % Difference (LH75 IRM) LH75 ( 1 9 /L) IRM ( 1 9 /L) Mean SD Mean Mean 1.96 SD Mean of LH75 and IRM ( 1 9 /L) Mean of ADVIA and IRM ( 1 9 /L) Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 639

7 Kim et al / Platelet Counting Accuracy in DIC and Acute Leukemia LA in all methods, and the unlikely-dic group showed the narrowest 95% LA (Figures 2B-2D). The ImmPLT demonstrated the narrowest 95% LA across all methods in the 3 subgroups of suspected DIC. In the definite-dic group, the CD-o, XE-i, and LH75 showed the widest 95% LA (Figure 2B). In the unlikely-dic group, the 95% LA did not show much variability across all methods and platelet tertiles (Figure 2D). A 6 % Difference (Automated Method IRM) Acute leukemia group Total (n = 44) Platelets <6 1 9 /L (n = 17) Platelets /L (n = 14) Platelets > /L (n = 13) Healthy control (n = 15) ImmPLT CD-o CD-i XE-i XE-o ADVIA LH75 B 6 % Difference (Automated Method IRM) ImmPLT CD-o CD-i XE-i Definite DIC group Total (n = 41) Platelets <6 1 9 /L (n = 14) Platelets /L (n = 15) Platelets > /L (n = 12) XE-o ADVIA LH75 Figure 2 Bland-Altman plots showing the mean percentage differences and 95% limits of agreement of platelet counts in different patient groups: acute leukemia (A), definite disseminated intravascular coagulation (DIC) (B), borderline DIC (C), and unlikely DIC (D). The first line represents percentage differences of total specimens in each patient group, and the other lines represent those of tertile subgroups based on platelet levels. Results for 15 healthy subjects are also illustrated in the acute leukemia group (A). 64 Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

8 Hematopathology / Original Article Association of Platelet Activation Markers With Erroneous Platelet Counts and the Underlying Clinical Conditions To investigate whether platelet activation status had an effect on the erroneous measurement of platelet counts, the percentage of specimens with erroneous platelet counts (outside the ±25% error limits on Bland-Altman plots in Figure 1) was shown on the basis of the quartiles of platelet activation parameters in all specimens Figure 3. C 6 % Difference (Automated Method IRM) Borderline DIC group Total (n = 64) Platelets <6 1 9 /L (n = 22) Platelets /L (n = 23) Platelets > /L (n = 19) 16 ImmPLT CD-o CD-i XE-i XE-o ADVIA LH75 D 6 % Difference (Automated Method IRM) Unlikely DIC group Total (n = 54) Platelets <6 1 9 /L (n = 14) Platelets /L (n = 17) Platelets > /L (n = 23) 16 ImmPLT CD-o CD-i XE-i XE-o ADVIA LH75 Intermethod coefficients of variation (ranges) are as follows: A, 2.7% ± 12.2% (7.3%-61.6%); P <.1; B, 17.4% ± 13.3% (4.3%-64.%); P <.1; C, 13.4% ± 7.4% (4.4%-35.1%); P <.2; D, 1.6% ± 4.5% (5.3%-24.4%); P <.3. P by Mann-Whitney U test compared with the intermethod coefficient of variation of healthy control subjects (7.6% ± 1.7%; range, 4.6%-1.9%). ADVIA, ADVIA 212; CD-i, Cell-Dyn Sapphire impedance method; CD-o, Cell-Dyn Sapphire optical method; ImmPLT, Cell-Dyn Sapphire immunologic method; IRM, international reference method; LH75, Coulter LH 75; XE-i, XE-21 impedance method; XE-o, XE-21 optical method. Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 641

9 Kim et al / Platelet Counting Accuracy in DIC and Acute Leukemia Specimens in the lowest quartile (Q1) of the MFI of SSC and MPC, which represent hypogranular activated platelets, showed significantly higher erroneous results than those in the highest quartile (Q4) in the CD-o, XE-i, XE-o, and ADVIA (Figures 3A and 3B). The CD-i and LH75 demonstrated no apparent trends with these parameters. The pplts, the proportion of small and CD61+ platelets representing activated small platelets, showed significant trends of high error rates in the highest quartile (Q4) of pplt-s for most analyzers (Figure 3C). However, the platelet distribution width showed no apparent increasing or decreasing trends across the quartiles, indicating that platelet size variation was not the main factor responsible for the erroneous platelet count (Figure 3D). To investigate the activation status of platelets according to the underlying clinical condition, the mean value of the platelet activation parameter in each patient group was calculated as shown in Table 2. Patients with acute leukemia showed significantly lower levels of the MFI of SSC and MPC and significantly higher levels of pplt-s than patients with unlikely DIC. The definite-dic group showed significantly lower levels of MPC. Effects of In Vitro Platelet Activation on the Agreement of Platelet Counts Fresh whole blood specimens were obtained from 1 healthy volunteers and 5 patients with suspected DIC and were stimulated with a low and a high dose of adenosine diphosphate. Platelet activation status was confirmed with measurements of the percentage of annexin V+ or PAC-1+ platelets, MPC, and pplt-s Figure 4. As expected, the intermethod CV was significantly increased after weak and strong activation (control, 7.2% ± 1.4%; weak activation, 7.7% ± 1.4%; and strong activation, 9.9% ± 5.8%, Figures 4A-4C). Similarly, the 95% LA gradually became wider as the intensity of platelet activation became stronger. A Samples Outside ±25% Error Limits (%) ImmPLT * CD-o CD-i XE-i XE-o * ADVIA P for trend < MFI of SSC Q1 Q2 Q3 Q4 LH B Samples Outside ±25% Error Limits (%) ImmPLT CD-o CD-i XE-i XE-o * ADVIA P for trend < MPC Q1 Q2 Q3 Q4 LH Figure 3 Adjusted percentage of specimens outside the ± 25% error limits in platelet counts by each automated method are presented on the basis of the quartiles (Q1-Q4) of platelet parameters: mean fluorescence intensity of side scatter (MFI of SSC) (A), mean platelet component concentration (MPC) (B), the proportion of small platelets (pplt-s) (C), and platelet distribution width (PDW) (D). 642 Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

10 Hematopathology / Original Article Impacts of Intermethod Variation in Platelet Counts on the Diagnosis of Overt DIC We investigated the clinical impact of the intermethod variation in platelet counts on the diagnosis of overt DIC in the borderline-dic group (n = 64). Platelet counts measured by each automated method were applied to the DIC scoring system and compared with those from the IRM. There was a tendency for all types of analyzers to overdiagnose overt DIC compared with the IRM (false-positive, 3.2%-1.4%; false-negative, %-1.6%) Table 3. The CD-i, XE-i, and ADVIA showed slightly higher κ indices (.9) than the other methods. Discussion This study demonstrated that automated platelet counting methods were inaccurate in patients with acute leukemia or DIC when compared with the IRM and that the intermethod variation of the platelet count correlated with the platelet activation status. Furthermore, in vitro platelet activation induced a higher intermethod variation of platelet counts. The inaccuracies of automated platelet counts have an impact on clinical decision making in the diagnosis of overt DIC. There has been some debate over which counting principle, between the impedance and optical methods, measures platelet counts more accurately. Some studies suggested that the accuracy of the optical methods was superior for thrombocytopenic specimens, 2,26,27 while recent studies demonstrated the impedance method to be more accurate for samples from patients undergoing cytotoxic chemotherapy. 5,28 Our data showed that there was no consistent superiority of platelet measurement in terms of the counting principle. The ImmPLT and ADVIA gave better agreement with the IRM than the other methods, while the CD-o demonstrated substantial inaccuracy in patients with acute leukemia or severe coagulopathy. C Samples Outside ±25% Error Limits (%) ImmPLT * * CD-o CD-i XE-i XE-o ADVIA P for trend < pplt-s Q1 Q2 Q3 Q4 * LH D Samples Outside ±25% Error Limits (%) ImmPLT CD-o CD-i XE-i XE-o ADVIA P for trend PDW Q1 Q2 Q3 Q4 LH All data were adjusted for age, sex, and platelet levels. * P <.5, P <.1, and P <.1 compared with the highest quartile (Q4). ADVIA, ADVIA 212; CD-i, Cell-Dyn Sapphire impedance method; CD-o, Cell-Dyn Sapphire optical method; ImmPLT, Cell-Dyn Sapphire immunologic method; LH75, Coulter LH 75; XE-i, XE-21 impedance method; XE-o, XE-21 optical method. Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 643

11 Kim et al / Platelet Counting Accuracy in DIC and Acute Leukemia Table 2 Levels of Platelet Parameters According to Underlying Clinical Condition in 23 Tested Specimens * Acute Leukemia (n = 44) Definite DIC (n = 41) Borderline DIC (n = 64) Unlikely DIC (n = 54) MFI of SSC (8.47) 7.84 (14.57) (12.39) 73.7 (12.57) pplt-s (%) 2.57 (2.39) 1.71 (2.12) 1.16 (1.36) 1.54 (1.54) MPC (g/dl) 2.75 (1.8) 2.6 (1.5) (1.21) (1.42) PDW (%) (2.12) (2.56) (2.23) (1.64) MFI, mean fluorescence intensity; MPC, mean platelet component; PDW, platelet distribution width; pplt-s, percentage of small platelets; SSC, side scatter. * Data are shown as the mean (SD). P values were calculated by using the Mann-Whitney U test and show comparison with the Unlikely DIC group. P <.1. P <.5. P <.1. A B 1 1 % Difference (Automated Method IRM) % Difference (Automated Method IRM) ImmPLT CD-o CD-i XE-i XE-o ADVIA LH75 4 ImmPLT CD-o CD-i XE-i XE-o ADVIA LH75 C % Difference (Automated Method IRM) ImmPLT CD-o CD-i XE-i XE-o ADVIA LH75 Figure 4 Bland-Altman plots showing the mean percentage differences and 95% limits of agreement of platelet counts in 3 states of platelet activation. Fresh whole blood specimens (n = 15) were stimulated without (control, A) or with 3 μmol/l adenosine diphosphate (ADP) (B) or 1 μmol/l ADP (C). Intermethod coefficients of variation (ranges) are as follows: A, 7.2% ± 1.4% (4.9%-1.4%); B, 7.7% ± 1.4% (4.6%-9.6%); P <.164; C, 9.9% ± 5.8% (4.6%-28.9%); P <.68. P by the paired t test compared with control. A, Annexin V+ (%), 3.9 ±.5; PAC-1+ (%), 2.8 ±.8; MPC (g/ dl), 24.3 ± 1.1; pplt-s (%),.672 ±.52. B, Annexin V+ (%), 12.1 ± 4.2 (P <.1); PAC-1+ (%), 5.3 ± 1.7 (P =.5); MPC (g/dl), 23.8 ± 1. (P =.83); pplt-s (%),.748 ±.577 (P =.48). C, Annexin V+ (%), 14.9 ± 1.6 (P <.1); PAC-1+ (%), 7.8 ± 4.9 (P =.48); MPC (g/dl), 23.6 ± 1.1 (P =.27); pplt-s (%),.786 ±.581 (P =.1). ADVIA, ADVIA 212; CD-i, Cell-Dyn Sapphire impedance method; CD-o, Cell-Dyn Sapphire optical method; ImmPLT, Cell-Dyn Sapphire immunologic method; IRM, international reference method; LH75, Coulter LH 75; MPC, mean platelet component; pplt-s, percentage of small platelets; XE-i, XE-21 impedance method; XE-o, XE-21 optical method. 644 Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

12 Hematopathology / Original Article Table 3 Sensitivity and Specificity of Automated Methods Compared With the IRM for Diagnosis of Overt DIC in the Borderline DIC Group (n = 64) IRM Raw Platelet Count DIC Score * 5 <5 FP (%) FN (%) SN (%) SP (%) κ Index ImmPLT <5 34 CD-o <5 32 CD-i <5 39 XE-i <5 37 XE-o < ADVIA < LH <5 36 ADVIA, ADVIA 212; CD-i, Cell-Dyn Sapphire impedance method; CD-o, Cell-Dyn Sapphire optical method; DIC, disseminated intravascular coagulation; FN, false-negative; FP, false-positive; ImmPLT, Cell-Dyn Sapphire immunologic method; IRM, international reference method; LH75, Coulter LH 75; SN, sensitivity; SP, specificity; XE-i, XE-21 impedance method; XE-o, XE-21 optical method. * Patients with a DIC score of 5 were diagnosed as having overt DIC. The agreement of XE-i and XE-o with the IRM was variable depending on the clinical condition or platelet level. Thus, in our study, the optical counting principles were not superior to the impedance counting principle. The most frequently suggested mechanism of platelet counting error is the inability to discriminate platelet-sized interfering particles from true platelets. 11 Because the specimens included in our study were those having known disorders related to pseudoplatelets, we investigated the possible sources of nonplatelet particles using peripheral blood smears and flow cytometric analysis of CD33, CD45, CD62E, and glycophorin A, but we could not find any nonplatelet particles of platelet sizes in patients with acute leukemia, in patients immediately after chemotherapy, or in patients with DIC. Thus, WBC or RBC fragments are unlikely to be the source of the discrepancies in platelet counts between the counting principles in our study samples. In our results, as the proportion of small and hypogranular platelets increased, the proportion of erroneous platelet counts increased. This association seemed to be stronger for the optical methods than the impedance and immunologic methods. Moreover, these platelet activation parameters, including the MFI of SSC, MPC, and pplt-s, showed a significant trend for high levels in the acute leukemia and definite-dic groups compared with in the unlikely-dic group. These results suggest that platelet activation is a potential candidate for the intermethod variation in platelet counts. We subsequently confirmed that in vitro platelet activation induced the observed intermethod variation of platelet counts. This is the first report describing the association of platelet activation status with intermethod platelet count variation in whole blood specimens. The activated platelets are a minor component of the total platelet count; however, when their proportion is significantly increased in pathologic status, it needs to be considered as a factor causing abnormal platelet counts. Although we could not explore the exact mechanism of the association of platelet activation and the intermethod variation, we could postulate that automated analyzers with different counting principles have different intrinsic detection limits for identifying degranulated small platelets. We finally assessed whether the intermethod variation in automated platelet counts can cause misdiagnosis of overt DIC. There was a potential to overdiagnose overt DIC compared with the IRM when guided by the platelet counts from all types of automated methods because the automated platelet counts tended to underestimate compared with the IRM. One multicenter study using only specimens with a platelet count less than /μl (2 1 9 /L) showed that most automated counting methods had overestimated platelet counts at platelet levels of 5, 1, and /μl (5, 1, and /L). 5 However, other studies reported that automated platelet counts were underestimated at the higher levels of 2 or /μl (2 or /L), similar to the results described herein The inaccuracy of automated platelet counts at higher levels could be a matter of serious concern in clinical practice, not only because of the risk for misdiagnosis but also because patients frequently have risk factors for hemorrhage and higher transfusion thresholds. 6 Despite these inaccuracies of routine automated methods in some abnormal samples, it is not feasible to implement the IRM for routine use as a necessity. Compared with automated analyzers that are much quicker and cost-effective, the IRM is Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 645

13 Kim et al / Platelet Counting Accuracy in DIC and Acute Leukemia time-consuming and requires technical expertise and specific facilities. The ImmPLT, another possible option, is not feasible for routine use either owing to its high cost and the requirement of 1 dedicated automated analyzer (Cell-Dyn). Therefore, to improve platelet count accuracy, further efforts to develop a simple and accurate standard method are necessary. This study has some limitations. First, we performed this accuracy study of platelet counts within 1 clinical laboratory of a tertiary university hospital. Therefore, interinstitutional variation was not investigated. Second, we used routine automated analyzers without advanced calibration of platelet counts just before the study because our study was expected to show intermethod variation only under the current situation of routine automated analyzers that were calibrated on a regular basis according to each of the manufacturer s guidelines. A universal calibrator would be expected to improve the intermethod variation of platelet counts. Third, although we focused on the readily available platelet activation parameters generated by automated analyzers in this study, these parameters are less sensitive than flow cytometric platelet activation markers such as anti-cd62p and PAC-1. Further analysis using the actual counts of activated platelets measured by flow cytometry would be able to give a more quantitative and specific interpretation. Although hematology analyzers usually provide reliable platelet counts, they may be inaccurate at enumerating platelets in patients with acute leukemia or DIC, ultimately resulting in the potential risk to make a wrong decision for DIC. It is interesting that platelet activation markers were associated with the severity of DIC and erroneous platelet counts, suggesting that platelet activation is a potential source for the intermethod variation in platelet counts. More attention needs to be given to improve the accuracy of platelet counts, especially in clinical conditions with high levels of platelet activation. From the 1 Department of Laboratory Medicine and 2 Cancer Research Institute, Seoul National University College of Medicine, Seoul, Korea; and 3 Department of Haematology, University of Liverpool, Liverpool, England. Supported by grant from the National Research Foundation of Korea funded by the Korean Government and by grant from the Priority Research Centers Program through the National Research Foundation of Korea funded by the Ministry of Education, Science and Technology, Daejeon. Address reprint requests to Dr H.K. Kim: Dept of Laboratory Medicine, Seoul National University College of Medicine, 11, Daehang-no Jongno-gu, Seoul , Republic of Korea. References 1. Zandecki M, Genevieve F, Gerard J, et al. Spurious counts and spurious results on haematology analysers: a review, part I: platelets. Int J Lab Hematol. 27;29: Harrison P, Horton A, Grant D, et al. Immunoplatelet counting: a proposed new reference procedure. Br J Haematol. 2;18: Kunz D, Kunz WS, Scott CS, et al. Automated CD61 immunoplatelet analysis of thrombocytopenic samples. Br J Haematol. 21;112: Norris S, Pantelidou D, Smith D, et al. Immunoplatelet counting: potential for reducing the use of platelet transfusions through more accurate platelet counting. Br J Haematol. 23;121: Segal HC, Briggs C, Kunka S, et al. Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusion. Br J Haematol. 25;128: British Committee for Standards in Haematology, Blood Transfusion Task Force. Guidelines for the use of platelet transfusions. Br J Haematol. 23;122: Malcolm ID, Monks P, Katz M. Spurious thrombocytosis in acute myelocytic leukemia [letter]. N Engl J Med. 1978;298: Ballard HS, Sidhu G. Cytoplasmic fragments causing spurious platelet counts in hairy cell leukemia: ultrastructural characterization. Arch Intern Med. 1981;141: Hammerstrom J. Spurious platelet counts in acute leukaemia with DIC due to cell fragmentation. Clin Lab Haematol. 1992;14: Li S, Salhany KE. Spurious elevation of automated platelet counts in secondary acute monocytic leukemia associated with tumor lysis syndrome. Arch Pathol Lab Med. 1999;123: van der Meer W, MacKenzie MA, Dinnissen JW, et al. Pseudoplatelets: a retrospective study of their incidence and interference with platelet counting. J Clin Pathol. 23;56: Krauss JS, Dover RK, Khankhanian NK, et al. Granulocytic fragments in sepsis. Mod Pathol. 1989;2: Hirokawa M, Manabe T, Ishimatsu S, et al. Leukocytic fragments in blood smears. Acta Pathol Jpn. 199;4: Branda JA, Kratz A. Effects of yeast on automated cell counting. Am J Clin Pathol. 26;126: International Council for Standardization in Haematology Expert Panel on Cytometry and International Society of Laboratory Hematology Task Force on Platelet Counting. Platelet counting by the RBC/platelet ratio method: a reference method. Am J Clin Pathol. 21;115: Johannessen B, Haugen T, Scott CS. Standardisation of platelet counting accuracy in blood banks by reference to an automated immunoplatelet procedure: comparative evaluation of Cell-Dyn CD4 impedance and optical platelet counts. Transfus Apher Sci. 21;25: Hervig T, Haugen T, Liseth K, et al. The platelet count accuracy of platelet concentrates obtained by using automated analysis is influenced by instrument bias and activated platelet components. Vox Sang. 24;87: Apelseth TO, Hervig T. In vitro evaluation of platelet concentrates during storage: platelet counts and markers of platelet destruction. Transfus Apher Sci. 27;37: Taylor FB Jr, Toh CH, Hoots WK, et al. Towards definition, clinical and laboratory criteria, and a scoring system for disseminated intravascular coagulation. Thromb Haemost. 21;86: Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP

14 Hematopathology / Original Article 2. Harrison P, Ault KA, Chapman S, et al. An interlaboratory study of a candidate reference method for platelet counting. Am J Clin Pathol. 21;115: Hickerson DH, Bode AP. Flow cytometry of platelets for clinical analysis. Hematol Oncol Clin North Am. 22;16: Macey MG, Carty E, Webb L, et al. Use of mean platelet component to measure platelet activation on the ADVIA 12 haematology system. Cytometry. 1999;38: Springer W, von Ruecker A, Dickerhoff R. Difficulties in determining prophylactic transfusion thresholds of platelets in leukemia patients. Blood. 1998;92: Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 23;22: Department of Health and Human Services. Centers for Medicare and Medicaid Services Centers for Disease Control and Prevention. Medicare, Medicaid, and CLIA programs; laboratory requirements relating to quality systems and certain personnel qualifications: final rule. Fed Regist. 23;68: Briggs C, Harrison P, Grant D, et al. New quantitative parameters on a recently introduced automated blood cell counter: the XE 21. Clin Lab Haematol. 2;22: Chapman DH, Hardin J, Miers M, et al. Reduction of the platelet review rate using the two-dimensional platelet method. Am J Clin Pathol. 21;115: Sandhaus LM, Osei ES, Agrawal NN, et al. Platelet counting by the Coulter LH 75, Sysmex XE 21, and Advia 12: a comparative analysis using the RBC/platelet ratio reference method. Am J Clin Pathol. 22;118: Arroyo JL, Garcia-Marcos MA, Lopez A, et al. Evaluation of a CD61 MoAb method for enumeration of platelets in thrombocytopenic patients and its impact on the transfusion decision-making process. Transfusion. 21;41: Keeney M, Brown W, Yee I. Platelet counts and flagging rates from the LH 75 hematology analyzer compared with the ICSH/ISLH platelet reference method and the Gen S hematology analyzer. Lab Hematol. 21;7: Hong KH, Kim MJ, Lee KW, et al. Platelet count evaluation using three automated haematology analysers compared with the immunoplatelet reference method, and estimation of possible inadequate platelet transfusion. Int J Lab Hematol. 29;31: Am J Clin Pathol 21;134: DOI: 1.139/AJCP88JYLRCSRXPP 647

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer

Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer Science Gene Gulati, PhD, Sean Cote, DO, Eric Behling, MD, William Kocher, MD

More information

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

Changes of hematological references depends on storage period and temperature conditions in rats and dogs

Changes of hematological references depends on storage period and temperature conditions in rats and dogs ISSN 1738-6055 (Print) ISSN 2233-7660 (Online) Lab Anim Res 2016: 32(4), 241-248 https://doi.org/10.5625/lar.2016.32.4.241 Changes of hematological references depends on storage period and temperature

More information

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD

Sue Jung Kim, MD, MS; Yoonjung Kim, MD; Saeam Shin, MD; Jaewoo Song, MD, MS; Jong Rak Choi, MD, PhD Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines Sue

More information

Performance evaluation of Celltac G: a new automated hematology analyzer

Performance evaluation of Celltac G: a new automated hematology analyzer Original Article Performance evaluation of Celltac G: a new automated hematology analyzer Masaaki Sugiyama 1, Tsukasa Kobayashi 1, Yuki Jisyage 2, Shigeko Yamamoto 2, Yutaka Nagai 2,3 and Hiroshi Kondo

More information

Customer Information Literature List White Blood Cells

Customer Information Literature List White Blood Cells Customer Information Literature List White Blood Cells Date: September 2015 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 150901 Note: Whether references are

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Brief Communication Diagnostic Hematology

Brief Communication Diagnostic Hematology Brief Communication Diagnostic Hematology Ann Lab Med 2016;36:244-249 http://dx.doi.org/10.3343/alm.2016.36.3.244 ISSN 2234-3806 eissn 2234-3814 Establishment of Age- and Gender-Specific Reference Ranges

More information

Customer Information Literature List Platelets

Customer Information Literature List Platelets Customer Information Literature List Platelets Date: August 2015 Subject: Literature List Platelets Issued by: Scientific Customer Services Number: 150803 Note: Whether references are given in British

More information

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom

CAROL BRIGGS, DONNA GRANT, SAMUEL J. MACHIN INTRODUCTION. Department of Haematology, University College London Hospital, London, United Kingdom ISLH Laboratory Hematology 7:75-80 2001 Carden Jennings Publishing Co., Ltd. Official Publication Comparison of the Automated Reticulocyte Counts and Immature Reticulocyte Fraction Measurements Obtained

More information

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation

White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation ISLH Laboratory Hematology 7:255 266 2001 Carden Jennings Publishing Co., Ltd. Official Publication White Blood Cell and Platelet Counting Performance by Hematology Analyzers:A Critical Evaluation BERNARD

More information

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature

Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Changes in Automated Complete Blood Cell Count and Differential Leukocyte Count Results Induced by Storage of Blood at Room Temperature Gene L. Gulati, PhD; Lawrence J. Hyland, MD; William Kocher, MD;

More information

Title. Author(s)Hayakawa, Mineji; Gando, Satoshi; Hoshino, Hirokatsu. CitationClinical and Applied Thrombosis/Hemostasis, 13(1): 6. Issue Date

Title. Author(s)Hayakawa, Mineji; Gando, Satoshi; Hoshino, Hirokatsu. CitationClinical and Applied Thrombosis/Hemostasis, 13(1): 6. Issue Date Title A Prospective Comparative Study of Three Sets of Cri vs Japanese Criteria Author(s)Hayakawa, Mineji; Gando, Satoshi; Hoshino, Hirokatsu CitationClinical and Applied Thrombosis/Hemostasis, 13(1):

More information

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation.

Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Getting Beyond the Flags: Quantitative assessment of immature granulocyte (IG) populations may improve the assessment of sepsis and inflammation. Sysmex America White Paper One Nelson C. White Parkway,

More information

EDUCATIONAL COMMENTARY DISSEMINATED INTRAVASCULAR COAGULATION

EDUCATIONAL COMMENTARY DISSEMINATED INTRAVASCULAR COAGULATION EDUCATIONAL COMMENTARY DISSEMINATED INTRAVASCULAR COAGULATION Educational commentary is provided through our affiliation with the American Society for Clinical Pathology (ASCP). To obtain FREE CME/CMLE

More information

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT

Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. ABSTRACT OriginalArticle Comparison of Leukocyte Differential Count by Beckman Coulter Unicel DxH800, Beckman Coulter LH780 and Sysmex XE-5000 Preechaya Wongkrajang, M.D., Chaicharoen Tantanate, M.D. Department

More information

The Evolving Role of Reticulated Platelets

The Evolving Role of Reticulated Platelets The Evolving Role of Kyle Riding, PhD, MLS(ASCP) CM University of Central Florida Session Objectives Platelets Recall the process involved in megakaryopoiesis and thrombocytopoiesis Explain the various

More information

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa

Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa Full Blood Count analysis Is a 3 part-diff good enough? Dr Marion Münster, Sysmex South Africa The Role of the FBC in clinical decision making History Examination Investigations Decision 70% FBC Laboratory

More information

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes

Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes Evaluation of the Diagnostic Performance of the Sysmex XT-2000i Automated Hematology Analyzer in the Detection of Immature Granulocytes A. M. CENCI *1, M. MACONI *2, and B. CASOLARI *1 *1 Laboratory of

More information

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation

Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation Laboratory Hematology 8:85-90 2002 Carden Jennings Publishing Co.. Ltd. Abbott Cell-Dyn Reticulocyte Method Comparison and Reticulocyte Normal Reference Range Evaluation T. SCHISANO, L. VAN HOVE Abbott

More information

IMMATURE PLATELETS CLINICAL USE

IMMATURE PLATELETS CLINICAL USE HAEMATOLOGY FEBRUARY 2017 WHITE PAPER IMMATURE PLATELETS CLINICAL USE Differential diagnosis of thrombocytopenia Thrombocytopenia and automated platelet measurement Thrombocytopenia is a condition characterised

More information

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana

Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana Blutbild 2012: Automatisiert oder manuell? Georg Stüssi Servizio di Ematologia Istituto Oncologico della Svizzera Italiana IS THERE STILL A ROLE FOR THE MANUAL BLOOD COUNT? Indication for examination

More information

Christopher M. Lehman, MD, 1,3 Lori W. Wilson, MT(ASCP), MS, 3 and George M. Rodgers, MD, PhD 1-3. Abstract

Christopher M. Lehman, MD, 1,3 Lori W. Wilson, MT(ASCP), MS, 3 and George M. Rodgers, MD, PhD 1-3. Abstract Coagulation and Transfusion Medicine / D-DIMER FOR THE DIAGNOSIS OF DIC Analytic Validation and Clinical Evaluation of the STA LIATEST Immunoturbidimetric D-Dimer Assay for the Diagnosis of Disseminated

More information

Approccio morfologico alle microangiopatie trombotiche

Approccio morfologico alle microangiopatie trombotiche Approccio morfologico alle microangiopatie trombotiche Gina Zini Polo Oncologia e Ematologia Policlinico A. Gemelli Università Cattolica S. Cuore - Roma 1 Thrombotic microangiopathies Occlusive microangiopathic

More information

8.2 Principles of Quantitative Hematologic Determinations (1)

8.2 Principles of Quantitative Hematologic Determinations (1) 3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily

More information

WHO Prequalification of Diagnostics Programme PUBLIC REPORT

WHO Prequalification of Diagnostics Programme PUBLIC REPORT WHO Prequalification of Diagnostics Programme PUBLIC REPORT Product: BD FACSCount Instrument System with FACSCount Control Kit and BD FACSCount Reagent Kit (Absolute CD4+, CD8+, and CD3+ Counts) Number:

More information

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples

Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Hematopathology / HEMOGLOBIN-BASED BLOOD SUBSTITUTES Automated Quantitation of Hemoglobin-Based Blood Substitutes in Whole Blood Samples Jolanta Kunicka, PhD, 1 Michael Malin, PhD, 1 David Zelmanovic,

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S

First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S 111 Riv Med Lab - JLM, Vol. 4, N. 2, 2003 First comparative evaluation of leucocyte differential and reticulocyte count between Abbott Cell -DYN 3700 and Beckman-Coulter Gen S H. Vandeputte a, D. Darre

More information

Pseudobasophilia in Pediatric Age Group

Pseudobasophilia in Pediatric Age Group Original Article DOI: 10.17354/ijss/2015/259 Pseudobasophilia in Pediatric Age Group Femela Muniraj 1, Vijay Amritraj 2 1 Assistant Professor, Department of Pathology, Chettinad Hospital & Research Institute,

More information

Quantitation of T-lymphocyte Subsets Using the CELL-DYN Sapphire Automated Haematology Analyser

Quantitation of T-lymphocyte Subsets Using the CELL-DYN Sapphire Automated Haematology Analyser ORIGINAL ARTICLE http://dx.doi.org/10.15263/jlmqa.2014.36.1.39 Quantitation of T-lymphocyte Subsets Using the CELL-DYN Sapphire Automated Haematology Analyser Jimyung Kim 1, Seung Ryu 2, Jinsook Lim 1,

More information

Multiple studies are available concerning the use of

Multiple studies are available concerning the use of SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory

More information

M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and Clinical Immunology, MU Sofia

M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and Clinical Immunology, MU Sofia 50, 48, 2012, 3, ADVIA 2120.,.., ANALYSIS OF DIAGNOSTIC RED CELL CYTOGRAMS, GENERATED BY THE ADVIA 2120 HEMATOLOGY ANALYZER M. Velizarova, T. Yacheva and K. Tzatchev Department of Clinical Laboratory and

More information

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer

Evaluation of a Method for Counting Absolute Numbers of Cells with a Flow Cytometer CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, May 1997, p. 309 313 Vol. 4, No. 3 1071-412X/97/$04.00 0 Copyright 1997, American Society for Microbiology Evaluation of a Method for Counting Absolute Numbers

More information

INFECTION/ INFLAMMATION

INFECTION/ INFLAMMATION HAEMATOLOGY OCTOBER 2017* WHITE PAPER INFECTION/ INFLAMMATION Novel haematological parameters for rapidly monitoring the immune system response Patients with inflammatory disease are common on hospital

More information

PointCare NOW TM Technical Note

PointCare NOW TM Technical Note PointCare NOW TM Technical Note Introduction to our Technology And Summary of Evaluation and Performance data 1 of 19 Introduction The PointCare NOW is a portable, and very fast, diagnostic instrument

More information

The Egyptian Journal of Hospital Medicine (July 2018) Vol. 72 (9), Page

The Egyptian Journal of Hospital Medicine (July 2018) Vol. 72 (9), Page The Egyptian Journal of Hospital Medicine (July 2018) Vol. 72 (9), Page 5210-5214 The Utility of Fibrinogen/C-reactive protein Ratio versus D-dimer and Fibrin Degradation Product in Diagnosis of Overt

More information

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags.

WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags. WHITE PAPER IMMATURE GRANULOCYTE (IG) POPULATIONS MAY IMPROVE THE ASSESSMENT OF INFECTION AND SEPSIS. Getting beyond the flags www.sysmex.com/us Introduction For many decades, physicians have relied on

More information

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts

BACKGROUND. Absolute neutrophil counts (ANCs) and absolute phagocyte counts 2681 Utility of Automated Counting to Determine Absolute Neutrophil Counts and Absolute Phagocyte Counts for Pediatric Cancer Treatment Protocols Nobuko Hijiya, M.D. 1,2 Mihaela Onciu, M.D. 3 Scott C.

More information

Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts

Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts Original Article Diagnostic Hematology Ann Lab Med 215;35:28-34 http://dx.doi.org/1.3343/alm.215.35.1.28 ISSN 2234-386 eissn 2234-3814 Flow Cytometric White Blood Cell Differential Using is Excellent for

More information

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential

Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential H e m a t o p a t h o l o g y / REFRIGERATION IMPROVES CBC COUNT STABILITY Refrigerated Storage Improves the Stability of the Complete Blood Cell Count and Automated Differential Brent L. Wood, MD, PhD,

More information

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population

Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population Clinical immunology Determination of monocyte count by hematological analyzers, manual method and flow cytometry in polish population EL BIETA GÓRSKA, URSZULA DEMKOW, ROMAN PIÑKOWSKI, BARBARA JAKUBCZAK,

More information

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2013

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2013 EXPERTISE, SERVICE PROVISION AND CUSTOMER RELATIONS QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY DEFINITIVE ANNUAL REPORT

More information

Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients

Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients Utilizing Sysmex RET He to Evaluate Anemia in Cancer Patients Ellinor I. Peerschke, Ph.D., F.A.H.A. Vice Chair, Laboratory Medicine Chief, Hematology & Coagulation Laboratory Services Memorial Sloan Kettering

More information

Changes in Hematologic Parameters Induced by Thermal Treatment of Human Blood

Changes in Hematologic Parameters Induced by Thermal Treatment of Human Blood Annals of Clinical & Laboratory Science, vol. 32, no. 4, 2002 393 Changes in Hematologic Parameters Induced by Thermal Treatment of Human Blood Jong Weon Choi and Soo Hwan Pai Department of Clinical Pathology,

More information

Literature List Body Fluids

Literature List Body Fluids Literature List Body Fluids Customer Information March 2018 Date: March 2018 Subject: Literature List body fluids Issued by: Scientific Customer Services Number: 180312 Note: Whether references are given

More information

Use of the XE-2100 in a Patient with Cold Auto-immune Hemolytic Anemia

Use of the XE-2100 in a Patient with Cold Auto-immune Hemolytic Anemia Use of the XE-21 in a Patient with Cold Auto-immune Hemolytic Anemia Gudrun STAMMINGER and Lothar BEIER Klinikum Chemnitz ggmbh, Institut für Laboratoriumsmedizin, Flemmingstraße 2, Chemnitz, Germany.

More information

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2014

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2014 EXPERTISE, SERVICE PROVISION AND CUSTOMER RELATIONS QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY DEFINITIVE ANNUAL REPORT

More information

Approach to disseminated intravascular coagulation

Approach to disseminated intravascular coagulation Approach to disseminated intravascular coagulation Khaire Ananta Shankarrao 1, Anil Burley 2, Deshmukh 3 1.MD Scholar, [kayachikitsa] 2.Professor,MD kayachikitsa. 3.Professor and HOD,Kayachikitsa. CSMSS

More information

2008 CAP TODAY Q & A

2008 CAP TODAY Q & A 2008 CAP TODAY Q & A Q. How often should we document hematology competencies to ensure consistency of morphologic observations for manual differentials and fluids? Should we do this every six months or

More information

The measurements of pleural fluid WBC count

The measurements of pleural fluid WBC count Variations in Pleural Fluid WBC Count and Differential Counts With Different Sample Containers and Different Methods* Barrett D. Conner, MD; Y. C. Gary Lee, MD; Paul Branca, MD, FCCP; Jeffrey T. Rogers,

More information

DIC. Bert Vandewiele Fellow Critical Care 23 May 2011

DIC. Bert Vandewiele Fellow Critical Care 23 May 2011 DIC Bert Vandewiele Fellow Critical Care 23 May 2011 Dissiminated Intravascular Coagulopathie 11/3/2011 Dr. Bert Vandewiele 2 Dissiminated Intravascular Coagulopathie = Consumption coagulopathie = Defibrination

More information

Things to never miss in the office. Brett Houston MD FRCPC (PYG-5, hematology) Leonard Minuk MD FRCPC

Things to never miss in the office. Brett Houston MD FRCPC (PYG-5, hematology) Leonard Minuk MD FRCPC Things to never miss in the office Brett Houston MD FRCPC (PYG-5, hematology) Leonard Minuk MD FRCPC Presenter Disclosure Faculty / Speaker s name: Brett Houston / Leonard Minuk Relationships with commercial

More information

Lavender Top Management SUCCESS BEYOND FINANCES

Lavender Top Management SUCCESS BEYOND FINANCES Lavender Top Management Alfonso Ziccardi Ass.Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES 1 . VISION Mission/ Vision To successfully

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2015

DEFINITIVE ANNUAL REPORT FLOW CYTOMETRY: LYMPHOCYTE SUBSET ANALYSIS CD34+ STEM CELL ENUMERATION 2015 EXPERTISE, SERVICE PROVISION AND CUSTOMER RELATIONS QUALITY OF MEDICAL LABORATORIES CLINICAL BIOLOGY COMMISSION COMMITTEE OF EXPERTS EXTERNAL QUALITY ASSESSMENT IN CLINICAL BIOLOGY DEFINITIVE ANNUAL REPORT

More information

Evaluation of Early Post-transplant Leukocyte Recovery Using the Undiluted Erythrocyte Lysing Technique

Evaluation of Early Post-transplant Leukocyte Recovery Using the Undiluted Erythrocyte Lysing Technique Annals of Clinical & Laboratory Science, vol. 32, no. 2, 2002 159 Evaluation of Early Post-transplant Leukocyte Recovery Using the Undiluted Erythrocyte Lysing Technique Myungshin Kim, 1 Ja Young Kim,

More information

Comparative Evaluation of Schistocyte Counting by an Automated Method and by Microscopic Determination

Comparative Evaluation of Schistocyte Counting by an Automated Method and by Microscopic Determination Hematopathology / AUTOMATED AND MICROSCOPIC SCHISTOCYTE COUNTING Comparative Evaluation of Schistocyte Counting by an Automated Method and by Microscopic Determination Jean-François Lesesve, MD, 1 Sylvain

More information

An Approach to the Patient Refractory to Platelets Transfusion. Harold Alvarez, MD

An Approach to the Patient Refractory to Platelets Transfusion. Harold Alvarez, MD Harold Alvarez, MD Objectives Explain the etiology of platelet refractoriness Discuss the different types of platelet refractoriness Describe how platelet refractoriness is diagnosed Discuss different

More information

Effect of the addition of the antioxidant taurine on the complete blood count of whole blood stored at room temperature and at 4 C for up to 7 days

Effect of the addition of the antioxidant taurine on the complete blood count of whole blood stored at room temperature and at 4 C for up to 7 days Original Article Effect of the addition of the antioxidant taurine on the complete blood count of whole blood stored at room temperature and at 4 C for up to 7 days Mahmoud Mohammed Sirdah 1 Abdelnasser

More information

Sysmex Journal International Vol.26 No.1 (2016)

Sysmex Journal International Vol.26 No.1 (2016) Sysmex Journal International Vol.26 No.1 (216) Performance Evaluation of the XN-55 Automated Hematology Analyzer Body Fluid Mode Considerations for Operational Conditions for Cell Counting with Cerebrospinal

More information

The Value-Driven Laboratory. The Role of the Immature Platelet Fraction (IPF) in the Differential Diagnosis of Thrombocytopenia

The Value-Driven Laboratory. The Role of the Immature Platelet Fraction (IPF) in the Differential Diagnosis of Thrombocytopenia The Value-Driven Laboratory The Role of the Immature Platelet Fraction (IPF) in the Differential Diagnosis of Thrombocytopenia Sysmex America White Paper 577 Aptakisic Road Lincolnshire, IL 669 USA www.sysmex.com

More information

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision

Lavender Top Management SUCCESS BEYOND FINANCES. Mission/ Vision Lavender Top Management Alfonso Ziccardi Lab.Operation Manager for AP/CP/Safety Officer New York Hospital Medical Center of Queens SUCCESS BEYOND FINANCES Mission/ Vision. VISION To successfully develop

More information

ADx Bone Marrow Report. Patient Information Referring Physician Specimen Information

ADx Bone Marrow Report. Patient Information Referring Physician Specimen Information ADx Bone Marrow Report Patient Information Referring Physician Specimen Information Patient Name: Specimen: Bone Marrow Site: Left iliac Physician: Accession #: ID#: Reported: 08/19/2014 - CHRONIC MYELOGENOUS

More information

Detection of important abnormalities of the differential

Detection of important abnormalities of the differential J Clin Pathol 1989;42:772-776 Detection of important abnormalities of the differential count using the Coulter STKR blood counter D F BARNARD, S A BARNARD, A B CARTER, S J MACHN, K G PATTERSON, A YARDUMAN

More information

CTAD as a universal anticoagulant

CTAD as a universal anticoagulant Automated Methods & Management in Chemistry Vol. 25, No. 1 (January February 2003) pp. 17 20 CTAD as a universal anticoagulant M. Yokota, N. Tatsumi*, I. Tsuda, T. Nishioka and T. Takubo Department of

More information

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens

Scoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens Original Article Laboratory Informatics Ann Lab Med 2015;35:341-347 http://dx.doi.org/10.3343/alm.2015.35.3.341 ISSN 2234-3806 eissn 2234-3814 Scoring System for Detecting Spurious Hemolysis in Anticoagulated

More information

The importance of thrombocytopenia and its causes

The importance of thrombocytopenia and its causes SYSMEX EDUCATIONAL ENHANCEMENT AND DEVELOPMENT NO 4 2017 SEED HAEMATOLOGY The importance of thrombocytopenia and its causes Key words: Thrombocytopenia, thrombocytopenic, low levels of platelets What is

More information

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine

Do Your Flow Cytometric LDTs. Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine Do Your Flow Cytometric LDTs Conform to the ICSH ICCS Validation Guidelines? Fiona E. Craig, MD University of Pittsburgh School of Medicine How should LDTs be validated? Accuracy Specificity Sensitivity

More information

The Value-Driven Laboratory. The Role of the Immature Platelet Fraction (IPF) in the Differential Diagnosis of Thrombocytopenia

The Value-Driven Laboratory. The Role of the Immature Platelet Fraction (IPF) in the Differential Diagnosis of Thrombocytopenia The Value-Driven Laboratory The Role of the Immature Platelet Fraction (IPF) in the Differential Diagnosis of Thrombocytopenia Sysmex America White Paper 577 Aptakisic Road Lincolnshire, IL 669 USA www.sysmex.com

More information

Research Article Mean Platelet Volume Reflect Hematopoietic Potency and Correlated Blood Group O in Cord Blood from Healthy Newborn

Research Article Mean Platelet Volume Reflect Hematopoietic Potency and Correlated Blood Group O in Cord Blood from Healthy Newborn BioMed Volume 2013, Article ID 754169, 4 pages http://dx.doi.org/10.1155/2013/754169 Research Article Mean Platelet Volume Reflect Hematopoietic Potency and Correlated Blood Group O in Cord Blood from

More information

SIMTI Servizi Srl. Original Article

SIMTI Servizi Srl. Original Article Original Article Evaluation of nucleated red blood cell count by Sysmex XE-2100 in patients with thalassaemia or sickle cell anaemia and in neonates Sabrina Buoro 1, Mauro Vavassori 1, Silvia Pipitone

More information

Evaluation of the Cockroft Gault, Jelliffe and Wright formulae in estimating renal function in elderly cancer patients

Evaluation of the Cockroft Gault, Jelliffe and Wright formulae in estimating renal function in elderly cancer patients Original article Annals of Oncology 15: 291 295, 2004 DOI: 10.1093/annonc/mdh079 Evaluation of the Cockroft Gault, Jelliffe and Wright formulae in estimating renal function in elderly cancer patients G.

More information

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital

Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital HEMATOPATHOLOGY Evaluation of the Abbott Cell-DYN 3500 Hematology Analyzer in a University Hospital JOAN-LLUIS VIVES-CORRONS, MD, ISABEL BESSON, PHD, JOSEP MARIA JOU, MD, AND GABRIELA GUTIERREZ, MB The

More information

Literature List Platelets

Literature List Platelets Literature List Platelets Customer Information February 2018 Date: February 2018 Subject: Literature List Platelets Issued by: Scientific Customer Services Number: 180202 Note: Whether references are given

More information

Automated Blood Cell Counts. State of the Art. Mauro Buttarello, MD, and Mario Plebani, MD. Abstract

Automated Blood Cell Counts. State of the Art. Mauro Buttarello, MD, and Mario Plebani, MD. Abstract Hematopathology / Automated Blood Cell Counts Automated Blood Cell Counts State of the Art Mauro Buttarello, MD, and Mario Plebani, MD Key Words: Blood cell analyzers; CBC count; Leukocyte differential

More information

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN

5/1/2017 DISCUSSION POINTS. Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN DISCUSSION POINTS Importance of hematological immature cell indices Clinical Utility of Immature Cell Indices Beyond the Routine CBC John E. Donnelly BSN, RN Investigate the evidence for clinical utility:

More information

Evaluation of the QBC Star centrifugal three-part differential haematology system

Evaluation of the QBC Star centrifugal three-part differential haematology system ORIGINAL ARTICLE 67 Evaluation of the QBC Star centrifugal three-part differential haematology system O. ERHABOR, G. RICHARDSON, I. MOHAMMED, C. THORNTON, J. BARK, M. HURST, D. HAMER and P. KINSELLA Blood

More information

Cell sorting: A bridge between morphology and flow cytometry. Pr François Mullier, Pr Bernard Chatelain July 19th, 2013

Cell sorting: A bridge between morphology and flow cytometry. Pr François Mullier, Pr Bernard Chatelain July 19th, 2013 Cell sorting: A bridge between morphology and flow cytometry Pr François Mullier, Pr Bernard Chatelain July 19th, 2013 1 CELL SORTING: PRINCIPLES D.Davies CELL SORTING: PRINCIPLES Aims Efficiency Purity

More information

Significant Relationship between Platelet Activation and Apnea-hypopnea Index in Patients with Obstructive Sleep Apnea Syndrome

Significant Relationship between Platelet Activation and Apnea-hypopnea Index in Patients with Obstructive Sleep Apnea Syndrome Tokai J Exp Clin Med., Vol. 36, No. 3, pp. 79-83, 211 Significant Relationship between Platelet Activation and Apnea-hypopnea Index in Patients with Obstructive Sleep Apnea Syndrome Yusuke KONDO *1, Ichiro

More information

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities

Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities Evaluation of Coulter S-Plus Three-part Differential in Population with a High Prevalence of Abnormalities JOANNE CORNBLEET, M.D. AND SALLY KESSINGER, MT(ASCP)SH The authors evaluated the Coulter 15 S-Plus

More information

Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate Specific Antigen

Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate Specific Antigen Letter to the Editor Clinical Chemistry Ann Lab Med 19;39:6-41 https://doi.org/1.3343/alm.19.39.4.6 ISSN 2234-36 eissn 2234-3814 Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate

More information

Palatelet Indices in Ladies With Normal Pregnancy in Red Sea State

Palatelet Indices in Ladies With Normal Pregnancy in Red Sea State Palatelet Indices in Ladies With Normal Pregnancy in Red Sea State 1 Bashir Abdalrahman Bashir,, 2* Mosab Nouraldein Mohammed Hamad, 2 Altag Madani Hassan 1 Asma Osman Ahmed, 1 Roaa Abdalla Hamed, 1 Hawa

More information

Study of correlation of changes in D-DIMER levels and coagulation profile with the development of coagulation disorders

Study of correlation of changes in D-DIMER levels and coagulation profile with the development of coagulation disorders Original article: Study of correlation of changes in D-DIMER levels and coagulation profile with the development of coagulation disorders 1 Dr Gurpreet Singh, 2 Dr Mahima Sharma, 3 Dr R K Chrungoo, 4 Dr.

More information

Autoanalyzer Generated Spurious Basophilia in Adolescents and Adults

Autoanalyzer Generated Spurious Basophilia in Adolescents and Adults Original Article DOI: 10.17354/ijss/2015/294 Autoanalyzer Generated Spurious Basophilia in Adolescents and Adults Femela Muniraj 1, Vijay Amritraj 2, Govindaraju Soundararajan 3 1 Assistant Professor,

More information

CELL-DYN 3700 Strength in Technology, Proven Reliability

CELL-DYN 3700 Strength in Technology, Proven Reliability CELL-DYN 3700 Strength in Technology, Proven Reliability Optical WBC Technology Patented MAPSS Differential Multiple Technologies CELL-DYN 3700 Multiple Technologies One Superior Result Multiple Technologies

More information

Hemostasis Haemostasis means prevention of blood loss from blood vessels.

Hemostasis Haemostasis means prevention of blood loss from blood vessels. 1 Hemostasis Haemostasis means prevention of blood loss from blood vessels. Bleeding is stopped by several mechanisms, which are: 1. Local vasoconstriction 2. Formation of platelet plug 3. Blood coagulation

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228

2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228 2007 Workshop of Society for Hematopathology & European Association for Hematopathology Indianapolis, IN, USA Case # 228 Vishnu V. B Reddy, MD University of Alabama at Birmingham Birmingham, AL USA 11/03/07

More information

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100

Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 Precision and Accuracy of the Leukocyte Differential on the Sysmex XE-2100 R. HERKLOTZ and A. R. HUBER Zentrum für Labormedizin, Kantonsspital Aarau, 5001 Aarau, Switzerland. The automated leukocyte(wbc)

More information

Literature List White Blood Cells

Literature List White Blood Cells Literature List White Blood Cells Customer Information February 2018 Date: February 2018 Subject: Literature List White Blood Cells Issued by: Scientific Customer Services Number: 180202 Note: Whether

More information

Platelet volume indices have low diagnostic efficiency for predicting bone marrow failure in thrombocytopenic patients

Platelet volume indices have low diagnostic efficiency for predicting bone marrow failure in thrombocytopenic patients EXPERIMENTAL AND THERAPEUTIC MEDICINE 5: 209-214, 2013 Platelet volume indices have low diagnostic efficiency for predicting bone marrow failure in thrombocytopenic patients RUI-LONG XU 1,2, ZHAO-JING

More information

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Coagulation and Transfusion Medicine / Discard Tube Unnecessary for Coagulation Assays Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Maarten

More information

Evaluation of a Single-Platform Technology for Lymphocyte Immunophenotyping

Evaluation of a Single-Platform Technology for Lymphocyte Immunophenotyping CLINICAL AND VACCINE IMMUNOLOGY, Oct. 2007, p. 1342 1348 Vol. 14, No. 10 1556-6811/07/$08.00 0 doi:10.1128/cvi.00168-07 Copyright 2007, American Society for Microbiology. All Rights Reserved. Evaluation

More information

XN-SERIES. XN Technology and Case Studies

XN-SERIES. XN Technology and Case Studies XN Technology and Case Studies Karen Hoffman MT(ASCP) Clinical Applications Specialist OBJECTIVES Explain how scattergrams and histogram pictures can provide great insight into abnormal hematology samples

More information

Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay

Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay Phagocytosis of FITC labelled opsonized and non-opsonized E. coli bacteria by monocytes and granulocytes in a whole blood assay APPLICATION NOTE Author: Andreas Spittler, MD, Associate Professor for Pathophysiology

More information

Detailed step-by-step operating procedures for NK cell and CTL degranulation assays

Detailed step-by-step operating procedures for NK cell and CTL degranulation assays Supplemental methods Detailed step-by-step operating procedures for NK cell and CTL degranulation assays Materials PBMC isolated from patients, relatives and healthy donors as control K562 cells (ATCC,

More information

Hemostasis Haemostasis means prevention of blood loss from blood vessels.

Hemostasis Haemostasis means prevention of blood loss from blood vessels. ١ Hemostasis Haemostasis means prevention of blood loss from blood vessels. Bleeding is stopped by several mechanisms, which are: 1. Local vasoconstriction 2. Formation of platelet plug 3. Blood coagulation

More information