Clinical Experience with MRI-Guided Vacuum-Assisted Breast Biopsy

Size: px
Start display at page:

Download "Clinical Experience with MRI-Guided Vacuum-Assisted Breast Biopsy"

Transcription

1 Lehman et al. MRI-Guided Vacuum-Assisted Breast Biopsy Constance D. Lehman 1,2 Elizabeth R. DePeri 3 Sue Peacock 1,2 Michelle D. McDonough 3 Wendy B. DeMartini 1,2 Jennifer Shook 1,2 Lehman CD, DePeri ER, Peacock S, McDonough MD, DeMartini W, Shook J Received June 28, 2004; accepted after revision September 22, Presented at the 2004 annual meeting of the American Roentgen Ray Society, Miami Beach, FL. 1 Department of Radiology, University of Washington Medical Center, 1959 NE Pacific, Seattle, WA Address correspondence to C. D. Lehman (lehman@seattlecca.org). 2 Breast Imaging, Seattle Cancer Care Alliance, 825 Eastlake Ave. E, G4-830, Seattle, WA Mayo Clinic, 4500 San Pablo Rd., Jacksonville, FL AJR 2005;184: X/05/ American Roentgen Ray Society Clinical Experience with MRI-Guided Vacuum-Assisted Breast Biopsy OBJECTIVE. The objective of our study was to evaluate a new commercially available method of MRI-guided vacuum-assisted breast biopsy using an open coil and a closed 1.5-T scanner. MATERIALS AND METHODS. Consecutive MRI-guided vacuum-assisted breast biopsies of 38 lesions in 28 women performed between May and September 2003 at two practice sites in the United States were retrospectively reviewed. Lesion characteristics including size, morphology, and enhancement were recorded. Times to perform each procedure, defined as the time from the start of the first localizing scan to the final scan after biopsy, were recorded. Histologic results for all lesions were obtained, and surgical, imaging, or clinical follow-up was performed. RESULTS. Enhancing masses and foci ranged from 2.5 to 19 mm. Nonmasslike enhancements ranged from 6 to 70 mm. All 38 biopsies (100%) were technically successful, and no complications were associated with any of the biopsy procedures. The average time to perform the 19 single-site MRI-guided procedures was 38 min (range, min). The 11 multiple-site biopsies performed in a single breast averaged 59 min (range, min), and eight bilateral biopsies averaged 64 min (range, min). Histologic results from vacuum-assisted breast biopsy revealed malignancy in 14 lesions (37%), atypical ductal hyperplasia in two lesions (5%), and benign findings in 22 lesions (58%). One of two lesions with atypical ductal hyperplasia was upgraded to ductal carcinoma in situ after surgery, for an overall cancer yield of 40% (15/38). CONCLUSION. This new method of MRI-guided vacuum-assisted breast biopsy is a safe, effective, and time-efficient means of MRI-guided tissue sampling. he use of MRI in the detection and T management of breast cancer is increasing. Prior studies have reported that MRI can detect malignancy that is occult on mammography, sonography, and physical examination [1 7]. However, although MRI shows high sensitivity, it has shown variable specificity. Because of the increasing use of MRI and its high rate of both true- and false-positive lesions occult on other forms of imaging, a method is needed for accurate tissue sampling of lesions seen only on MRI. The most common methods of MRI biopsy currently in use are wire localization and surgical excision or core needle biopsy [8 17]. Under stereotactic mammographic guidance, vacuum-assisted biopsy has proven to be more effective than core needle biopsy and less invasive than surgical excision of breast lesions [18, 19]. The advantages of vacuum-assisted breast biopsy include a single probe insertion with directional sampling and rapid collection of larger samples. MRI-compatible devices have been developed that allow vacuum-assisted needle biopsy to be performed under MR guidance [20, 21]. Clinical studies in Europe have shown that vacuum-assisted breast biopsy was successful in 98% of 341 lesions [22] and have identified important barriers to more widespread use of this technology, including needle artifact, tissue shift during probe insertion, and washout of contrast enhancement during the procedure. The challenges of this early technology have limited application to lesions larger than 10 mm in diameter, and the devices to date are not commercially available in the United States [22, 23]. Recently, a new commercially available method of MRI-guided vacuum-assisted breast biopsy that addresses many of the limitations identified by earlier prototypes has been developed. The purpose of this study was to evaluate this method of vacuum-assisted breast biopsy in women undergoing MRI-guided breast biopsy of lesions identified only on MRI AJR:184, June 2005

2 MRI-Guided Vacuum-Assisted Breast Biopsy Materials and Methods Patients This study was a collaborative study between two breast imaging clinics in the United States. Both sites obtained approval from their respective institutional review offices for this study. Because this study is retrospective, patient informed consent was not required by either institutional review office. Data were collected from our clinical database and medical records of 28 women who had 38 consecutive MRI-guided breast biopsies between May and September All cases selected for MRIguided vacuum-assisted breast biopsy were not palpable and were not clearly visible on mammography or targeted sonography. Lesion Characteristics MRI interpretation criteria, assessments, and recommendations were based on the breast MRI lexicon established by the American College of Radiology [24]. All lesions identified on diagnostic MRI were assessed by the radiologist as category 4 (suspicious abnormality) or category 5 (highly suggestive of malignancy). A targeted second-look sonography examination was performed on all lesions. If the lesion was clearly visible, a sonographically guided core biopsy was performed. If the lesion was not visible on sonography, as was the case in the 38 lesions included in this study, an MRI-guided vacuum-assisted breast biopsy was performed. For all cases, the following lesion characteristics were recorded: type (focus, mass, or nonmasslike enhancement), size, location, and enhancement kinetics (persistent, plateau, or washout). Biopsy Procedure At both institutions, biopsies were performed following standard clinical protocol. The time to perform the procedure, defined as time from the first MRI sequence (localizing) to the last scan sequence (after clip deployment), was recorded for each procedure. Vacuum-Assisted Breast Biopsy Device All biopsies were performed using the ATEC Breast Biopsy and Excision System (Suros Surgical Systems). Components include a breast compression grid that attaches to a breast coil (model OBC- 63 Breast Array Coil, MRI Devices), a single-use MRI-safe introducer kit, and a handheld vacuumassisted biopsy device handpiece (Figs. 1 3). Components of the ATEC Introducer Set (Suros Surgical Systems) include a needle guide, a coaxial introducer sheath, a sharp nonferrous inner stylet, and a plastic localizing obturator (Fig. 2). These components provide access to the target lesion and verification of sheath placement. The combined introducer sheath and stylet allows access to the biopsy site while minimizing fluid egress. The nonartifact-producing localizing obturator appears as a signal void, a black dot, on the image screen in the sagittal view and identifies the position of the sheath in reference to the target lesion. The tip of the obturator corresponds to the center of the sam- Fig. 1. Photograph shows breast coil (model OBC-63 Breast Array Coil, MRI Devices) and components of the ATEC Breast Biopsy and Excision System (Suros Surgical Systems), including compression plate and grid, needle guide, and fiducial marker. Fig. 2. Photograph shows single-use MRI-safe ATEC Introducer Set (Suros Surgical Systems), which includes needle guide, introducer sheath, introducer stylet, and localizing obturator. AJR:184, June

3 Lehman et al. Fig. 3. Photograph shows handpiece of ATEC vacuum-assisted biopsy device (Suros Surgical Systems). Fig. 4. Photograph shows ATEC (Suros Surgical Systems) coaxial introducer sheath (short arrow) placed into breast through localization grid. Vacuum-assisted biopsy device handpiece (long arrow) is placed into introducer sheath to perform biopsy. pling chamber once the obturator is removed and the sampling device is inserted through the sheath. Placement of the biopsy site marker is accomplished with the introducer sheath still in place by guiding the marker through the sheath to the biopsy target site. The ATEC vacuum-assisted breast biopsy handpiece is a pneumatic, MRI-compatible, and fully disposable single-patient-use biopsy device. It is composed of a plastic outer housing attached to a trocar-tipped outer cannula and an inner retractable cutting blade (Fig. 3). Rotating the handpiece allows the 20-mm sampling aperture to be positioned in any 360 position for sampling tissue. A pneumatic drive and vacuum tubing connect the handpiece to the console unit to provide a means by which to continuously sample tissue and aspirate and lavage the sampled tissue through the cutting chamber with lidocaine and A B C D Fig. 5. Sagittal T1-weighted images of 78-year-old woman with recently diagnosed left breast infiltrating ductal carcinoma who presented with suspicious enhancing mass in right breast. A, Image of right breast shows localization grid as low-signal-intensity lines at skin surface and vitamin E marker within grid (arrow). B, Contrast-enhanced image of right breast shows enhancing mass (arrow), which measures 6 mm, in lower outer quadrant. C, Contrast-enhanced image of right breast shows obturator tip (arrow) within mass. D, Contrast-enhanced image of right breast shows clip artifact (arrow) within mass AJR:184, June 2005

4 MRI-Guided Vacuum-Assisted Breast Biopsy TABLE 1 Type and Size of the 38 Lesions in the Study No. (%) Size (mm) Lesion Type of Lesions Range Mean ± SD Masses or foci 24 (63.2) ± 4.4 Nonmasslike enhancement 14 (36.8) ± 18.7 saline. The tissue specimens are automatically delivered to a collection chamber. MRI-Guided Biopsy Procedure After written informed consent was obtained from the patient and IV access was established, the patient was placed prone with the affected breast or breasts (for bilateral biopsies) positioned in the dedicated breast coil (model OBC-63 Breast Array Coil, MRI Devices). The breast was stabilized in a biopsy guidance grid (ATEC, Suros Surgical Systems) by providing moderate compression (Fig. 4). A vitamin E capsule was taped to the skin in the center of the grid for reference. The biopsy guidance grid also allows determination of the x and y coordinates of the target lesion. After localizing sequences had been performed, unenhanced and contrast-enhanced sagittal images were obtained to confirm the persistence of the enhancing lesion noted on the prior diagnostic MRI (Fig. 5). The images were reviewed on the monitor, and the locations of the vitamin E capsule placed on the skin TABLE 2 Time to Perform Biopsies Using MRI-Guided Vacuum-Assisted Breast Biopsy Device Time to Perform Biopsies (min) Biopsy Type No. of Cases Average Range SD Single lesion Multiple lesions in One breast Both breasts TABLE 3 Biopsy Comparison of Histology Results from MRI-Guided Vacuum-Assisted Breast Biopsy and Pathology Results from Surgery Benign Atypical Ductal Hyperplasia Surgical Results Ductal Carcinoma In Situ Invasive Carcinoma Unavailable Benign 1 1 Atypical ductal hyperplasia 1 1 Ductal carcinoma in situ Invasive carcinoma surface and the lesion were marked on a reference diagram. These markings correspond to the x and y coordinates of the lesion and target. In addition, the depths (z coordinates) of the lesion and vitamin E capsule were recorded. The depth (z) difference between the lesion and the vitamin E skin marker was calculated and recorded on the diagram. The MRI table was moved out of the bore of the magnet, the patient s skin was then sterilized with commercially available povidone-iodine and local anesthetic (1% lidocaine mixed with bicarbonate solution [10:1]) was injected. A sterile plastic coaxial sheath with a sharp inner nonferrous metallic stylet was inserted to the calculated depth using the centimeter markers on the exterior of the sheath. Once the coaxial sheath was in place, the inner cutting stylet was removed and replaced with a plastic MRI-visible obturator. With the coaxial sheath and obturator in place, the patient was returned to the bore of the magnet. A sagittal sequence was performed through the region of the obturator and the lesion to confirm accurate targeting (Fig. 5). Once correct targeting was confirmed, the patient was moved out of the bore of the magnet and tissue sampling was performed using the vacuum-assisted breast biopsy device through the coaxial sheath. Sampling was performed preferentially in the direction of the lesion. For example, if the lesion was noted to be inferior in relation to the probe on the sequence performed after obturator insertion, samples were obtained from areas between the 4- and 8-o clock positions. When tissue sampling was complete, the probe was retracted 5 mm and an MRI-safe site marker was placed through the coaxial sheath (ATEC Mark Biopsy Site Identifier, Artemis Medical). A final sagittal sequence through the biopsied region was performed to verify the location of the sampling defect and of the site marker (Fig. 5). At the completion of the biopsy, the coaxial system was removed. Manual compression was applied to achieve hemostasis, and a sterile dressing was applied. Histologic Data Collection Histologic results for all lesions were obtained, and surgical, imaging, or clinical follow-up was recommended. Biopsy histologic results were classified into the following categories: invasive carcinoma, in situ carcinoma, atypical ductal hyperplasia (ADH), or benign. Statistical Analysis The average time to complete an examination and SD were computed. The average lesion size and SD were computed. The positive predictive value was computed by considering ductal carcinoma in situ (DCIS) or invasive carcinoma as positive cases. Results Each of the participating facilities contributed approximately half of the cases included in this study. Patients ranged in age from 36 to 78 years (mean, 54 years). The majority of cases (61%) were referred for MRI evaluation to determine the extent of disease in a patient with a recent diagnosis of breast cancer. Other reasons for MRI evaluation included participation in MRI high-risk breast cancer screening studies (24%) and problem solving (15%). MRI was performed for problem solving in cases of an indeterminate mammographic, sonographic, or clinical abnormality and for follow-up of suspicious MRI examinations performed at an outside institution. Table 1 summarizes the types and sizes of lesions. Targeted masses and enhancing foci ranged from 2.5 to 19 mm (mean, 9.3 mm). Nonmasslike enhancement ranged from 6 to 70 mm (mean, 23.6 mm). Table 2 presents the type of biopsies performed and the average time to complete each type. Half of the study biopsies (n = 19) were on single lesions, 29% (n = 11) of the biopsies involved multiple lesions in a single breast, and the remaining 21% (n = 8) involved multiple lesions in both breasts. The average time to perform the 19 single-biopsy MRI-guided procedures was 38 min (range, min). The 11 multiple biopsies performed in a single breast averaged 59 min (range, min), and the eight bilateral biopsies averaged 64 min (range, min). AJR:184, June

5 Lehman et al. TABLE 4 Authors [reference no.] Comparison of MRI-Guided Biopsy Studies Study Year of Publication Method All 38 biopsies (100%) were technically successful, and no complications were associated with any of the biopsy procedures. Histologic results from vacuum-assisted breast biopsies revealed malignancy in 14 lesions (37%), ADH in two lesions (5%), and benign findings in 22 lesions (58%) (Table 3). Thirteen of the 14 cases of malignancy and the two cases of ADH were followed by surgical excision, lumpectomy, or mastectomy. One patient with infiltrating carcinoma diagnosed by vacuum-assisted breast biopsy elected not to have further surgery when widespread metastases were identified at further imaging evaluation. One of the two lesions diagnosed as ADH by vacuum-assisted breast biopsy was upgraded to DCIS after surgery, and one of four lesions identified as DCIS was upgraded to invasive cancer. One case of invasive carcinoma was downgraded to DCIS on the basis of analysis of the lumpectomy specimen. Two separate lesions originally identified as DCIS or infiltrating ductal carcinoma on vacuum-assisted breast biopsy were downgraded to benign at surgical excision. In both cases, the biopsy cavity from the vacuum-assisted breast biopsy documenting malignancy was identified in the surgical specimen, confirming the malignancies were removed by the initial vacuum-assisted breast biopsy. Two cases with benign histologic findings at vacuumassisted breast biopsy were followed by mastectomy, one in a patient who opted for bilateral mastectomies with known cancer in the opposite breast and one in a woman participating in a high-risk screening trial who opted for prophylactic mastectomy. Surgical pathology from the mastectomy in this second case revealed 5 mm of DCIS near the area of the prior MRI-guided vacuum-assisted breast biopsy. The other case was confirmed to be benign. Discussion We evaluated a new method of MRI-guided vacuum-assisted breast biopsy in 28 patients and found this method to be both safe and accurate with an overall cancer yield of 40% of all biopsies performed. Our results are similar to two prior clinical reports of this method. In their study of 20 cases, Liberman et al. [25] reported that vacuum-assisted breast biopsy could be performed successfully with MRI (96% success rate) and with a small risk of failure (one case Breast Biopsy No. of Lesions (No. of Patients) Average Time (min) to Conduct Biopsy A S M with a minor complication of a hematoma). Orel and colleagues [26] reached a similar conclusion in their analysis of nine MRI-vacuum-assisted breast biopsy biopsies (100% success rate with no complications). The time to perform the procedures and positive biopsy rates in our study are also similar to those reported in these prior studies (Table 4). This method of vacuum-assisted breast biopsy takes less time to acquire tissue compared with prior published reports of MRI-guided core needle biopsy [15, 27, 28]. For example, MRI-guided core needle biopsy of single lesions required 45 and 55 min on average in two prior published studies, compared with 35 and 38 min for single lesions using MRI-guided vacuum-assisted breast biopsy [25, 27, 28]. The time difference appears to be even more dramatic for multiple lesions, with prior reports of 90 min on average to perform multiple core needle biopsies compared with 69 and 61 min on average to perform multiple biopsies with MRIguided vacuum-assisted breast biopsy (Table 4). The data in our study represent our early experience with MRI-guided vacuum-assisted breast biopsy. Of the two facilities in our study, No. (%) of Reported Complications a Technical Success Rate No. (%) of Cases Overall Cancer Yield b Orel et al. [8] 1994 WL 11 (11) 11/11 (100) 5/11 (46) Fischer et al. [10] 1995 WL 28 (26) 60 27/28 (96) 13/28 (46) Kuhl et al. [11] 1997 WL 97 (66) 95/97 (98) 53/97 (55) Daniel et al. [12] 1998 WL 19 (17) 64 1 (5) 19/19 (100) 8/19 (42) Smith et al. [29] 2001 WL 16 (16) 3 (19) 16/16 (100) 5/16 (31) Lampe et al. [30] 2002 WL 132 ( ) 3 (2) 127/132 (96) 70/132 (53) Bedrosian et al. [31] 2002 WL 41 (41) 41/41 (100) 19/41 (46) Morris et al. [14] 2002 WL 101 (69) 31 c 3 (3) 101/101 (100) 31/101 (31) Smith et al. [29] 2001 CN 25 (23) 24/25 (96) 5/25 (20) Daniel et al. [28] 2001 CN 27 (19) /27 (100) 9/27 (33) Kuhl et al. [15] 2001 CN 78 (59) 60 1 (1) 77/78 (99) 27/77 (35) Schneider et al. [27] 2002 CN 21 (21) NA 20/21 (95) 8/21 (38) Chen et al. [17] 2004 CN 35 (29) 34/35 (97) 10/35 (29) Perlet et al. [22] 2002 VA 341 ( ) (5) 334/341 (98) 87/341 (26) Orel et al. [26] 2003 VA 9 (8) 9/9 (100) 4/9 (44) Liberman et al. [25] 2003 VA 27 (20) (4) 26/27 (96) 8/27 (30) This study 2005 VA 38 (28) /38 (100) 15/38 (40) Note. Dash ( ) indicates not reported. A = all biopsies in the study, S = single-lesion biopsy cases, M = multiple-lesion biopsy cases, WL = wire localization, CN = core needle, VA = vacuum-assisted, NA = not applicable. a Complications included hematomas, broken wires, vasovagal reactions, and one ruptured implant. b Cancers detected/lesions biopsied. c This study reported median time rather than average time AJR:184, June 2005

6 MRI-Guided Vacuum-Assisted Breast Biopsy one had no prior experience in MRI-guided vacuum-assisted breast biopsy and the other had only 1 month of experience before the study. The first site did have approximately 1 year of experience with MRI-guided core needle biopsy; the second site had no prior experience with any MRI-guided interventions. It may be that MRI-guided vacuum-assisted breast biopsy can be performed in even less time than reported here once experience is gained. We now find the total table time is approximately 20 min for biopsy of a single lesion. Biopsy of lesions that are far medial or posterior in the breast can prove challenging. Until recently, access for MRI-guided biopsy was limited to the lateral side of the breast. However, coils that allow medial access are now available and new coils that provide a full range of options for access including superior and inferior approaches are being developed. For far posterior lesions, we have found that removing the pad on the coil and rolling the patient to a position similar to the positioning on the stereotactic table for gaining access to far posterior lesions can improve access. This study is the largest to date of MRI-guided vacuum-assisted breast biopsy using the ATEC device. However, there are limitations to this study. Benign biopsies were not followed by surgical excision in 20 of the 22 benign cases. In one case in which the vacuum-assisted breast biopsy sample was benign at histology, DCIS was identified at prophylactic mastectomy. Longer term follow-up of patients with benign histologic results from MRI-guided breast biopsy samples is needed to determine the true rate of false-negative biopsies. This information from long-term follow-up of patients and further study of lesion characteristics associated with various forms of benign abnormalities will help in management decisions regarding suspicious MRI lesions with benign histology. At this time, we determine imaging pathology concordance on a case-by-case basis and recommend surgical excision for all cases of ADH and for any discordant cases. In addition, we recommend 6-month follow-up MRI examinations in all patients with benign histologic results. If the lesion is stable or decreased at the 6-month follow-up examination, we recommend resumption of routine screening. In conclusion, this method of tissue sampling for suspicious MRI lesions is safe and accurate, faster than core needle biopsy, and less invasive than wire localization and surgical excision. As the use of breast MRI grows, this procedure provides an effective and minimally invasive method of sampling lesions that are suspicious for malignancy. Acknowledgments We thank Bonnie Thursten and John McCloskey for their contributions to the success of this project and Tamara Fernando for her invaluable assistance in the preparation of the manuscript. References 1. Kuhl CK, Schmutzler RK, Leutner CC, et al. Breast MR imaging screening in 192 women proved or suspected to be carriers of a breast cancer susceptibility gene: preliminary results. Radiology 2000;215: Warner E, Plewes DB, Shumak RS, et al. Comparison of breast magnetic resonance imaging, mammography, and ultrasound for surveillance of women at high risk for hereditary breast cancer. J Clin Oncol 2001;19: Stoutjesdijk MJ, Boetes C, Jager GJ, et al. Magnetic resonance imaging and mammography in women with a hereditary risk of breast cancer. J Natl Cancer Inst 2001;93: Tilanus-Linthorst MM, Obdeijn IM, Bartels KC, de Koning HJ, Oudkerk M. First experiences in screening women at high risk for breast cancer with MR imaging. Breast Cancer Res Treat 2000;63: Morris EA, Liberman L, Ballon DJ, et al. MRI of occult breast carcinoma in a high-risk population. AJR 2003;181: Liberman L, Morris EA, Dershaw DD, Abramson AF, Tan LK. MR imaging of the ipsilateral breast in women with percutaneously proven breast cancer. AJR 2003;180: Liberman L, Morris EA, Kim CM, et al. MR imaging findings in the contralateral breast of women with recently diagnosed breast cancer. AJR 2003;180: Orel SG, Schnall MD, Newman RW, Powell CM, Torosian MH, Rosato EF. MR imaging-guided localization and biopsy of breast lesions: initial experience. Radiology 1994;193: Fischer U, Vosshenrich R, Keating D, et al. MRguided biopsy of suspect breast lesions with a simple stereotaxic add-on-device for surface coils. Radiology 1994;192: Fischer U, Vosshenrich R, Doler W, Hamadeh A, Oestmann JW, Grabbe E. MR imaging-guided breast intervention: experience with two systems. Radiology 1995;195: Kuhl CK, Elevelt A, Leutner CC, Gieseke J, Pakos E, Schild HH. Interventional breast MR imaging: clinical use of a stereotactic localization and biopsy device. Radiology 1997;204: Daniel BL, Birdwell RL, Ikeda DM, et al. Breast lesion localization: a freehand, interactive MR imaging-guided technique. Radiology 1998;207: Fischer U, Kopka L, Grabbe E. Magnetic resonance guided localization and biopsy of suspicious breast lesions. Top Magn Reson Imaging 1998;9: Morris EA, Liberman L, Dershaw D, et al. Preoperative MR imaging guided needle localization of breast lesions. AJR 2002;178: Kuhl CK, Morakkabati N, Leutner CC, Schmiedel A, Wardelmann E, Schild HH. MR imagingguided large-core (14-gauge) needle biopsy of small lesions visible at breast MR imaging alone. Radiology 2001;220: Lehman CD, Eby PR, Chen X, Dee KE, Thursten B, McCloskey J. MR imaging guided breast biopsy using a coaxial technique with a 14-gauge stainless steel core biopsy needle and a titanium sheath. AJR 2003;181: Chen X, Lehman CD, Dee KE. MRI-guided breast biopsy: clinical experience with 14-gauge stainless steel core biopsy needle. AJR 2004;182: Burbank F. Stereotactic breast biopsy of atypical ductal hyperplasia and ductal carcinoma in situ lesions: improved accuracy with directional, vacuumassisted biopsy. Radiology 1997;202: Philpotts LE, Shaheen NA, Carter D, Lange RC, Lee CH. Comparison of rebiopsy rates after stereotactic core needle biopsy of the breast with 11-gauge vacuum suction probe versus 14-gauge needle and automatic gun. AJR 1999;172: Lehman CD, Aikawo T. MR-guided vacuum-assisted breast biopsy: accuracy of targeting and success in sampling in a phantom model. Radiology 2004;232: Viehweg P, Heinig A, Amaya B, Alberich T, Laniado M, Heywang-Kobrunner SH. MR-guided interventional breast procedures considering vacuum biopsy in particular. Eur J Radiol 2002;42: Perlet C, Heinig A, Prat X, et al. Multicenter study for the evaluation of a dedicated biopsy device for MR-guided vacuum biopsy of the breast. Eur Radiol 2002;12: Heywang-Kobrunner SH, Heinig A, Pickuth D, Alberich T, Spielmann RP. Interventional MRI of the breast: lesion localisation and biopsy. Eur Radiol 2000;10: American College of Radiology. Magnetic resonance imaging. In: American College of Radiology, Breast imaging reporting and data system: breast imaging atlas. Reston, VA: American College of Radiology, Liberman L, Morris EA, Dershaw DD, Thornton CM, Van Zee KJ, Tan LK. Fast MRI-guided vacuum-assisted breast biopsy: initial experience. AJR 2003;181: Orel SG, Schnall MD, Mies CJ. MRI-compatible 12-G vacuum-assisted core biopsy of suspicious enhancing lesions: preliminary experience. (abstr) Radiology 2003;493(P) 27. Schneider JP, Schulz T, Horn LC, Leinung S, Schmidt F, Kahn T. MR-guided percutaneous core biopsy of small breast lesions: first experience with a vertically open 0.5T scanner. J Magn Reson Imaging 2002;15: Daniel BL, Birdwell RL, Butts K, et al. Freehand imri-guided large-gauge core needle biopsy: a new minimally invasive technique for diagnosis of enhancing breast lesions. J Magn Reson Imaging 2001;13: Smith LF, Henry-Tillman R, Mancino AT, et al. Magnetic resonance imaging-guided core needle biopsy and needle localized excision of occult breast lesions. Am J Surg 2001;182: Lampe D, Hefler L, Alberich T, et al. The clinical value of preoperative wire localization of breast lesions by magnetic resonance imaging: a multicenter study. Breast Cancer Res Treat 2002;75: Bedrosian I, Schlencker J, Spitz FR, et al. Magnetic resonance imaging-guided biopsy of mammographically and clinically occult breast lesions. Ann Surg Oncol 2002;9: AJR:184, June

Feasibility of MRI-guided large-core-needle biopsy of suspiscious breast lesions at 3T

Feasibility of MRI-guided large-core-needle biopsy of suspiscious breast lesions at 3T Eur Radiol (2009) 19: 1639 1644 DOI 10.1007/s00330-009-1310-0 BREAST Nicky H. G. M. Peters Carla Meeuwis Chris J. G. Bakker Willem P. Th. M. Mali Arancha M. Fernandez-Gallardo Richard van Hillegersberg

More information

Patient Outcomes in Canceled MRI-Guided Breast Biopsies

Patient Outcomes in Canceled MRI-Guided Breast Biopsies Women s Imaging Original Research Outcomes After Canceled MRI-Guided Breast Biopsies Women s Imaging Original Research Bethany L. Niell 1 Janie M. Lee 1, 2 Christopher Johansen 3 Elkan F. Halpern 4 Elizabeth

More information

Usefulness of magnetic resonance imaging-guided vacuum-assisted breast biopsy in Korean women: a pilot study

Usefulness of magnetic resonance imaging-guided vacuum-assisted breast biopsy in Korean women: a pilot study An et al. World Journal of Surgical Oncology 2013, 11:200 WORLD JOURNAL OF SURGICAL ONCOLOGY RESEARCH Open Access Usefulness of magnetic resonance imaging-guided vacuum-assisted breast biopsy in Korean

More information

National Diagnostic Imaging Symposium 2013 SAM - Breast MRI 1

National Diagnostic Imaging Symposium 2013 SAM - Breast MRI 1 National Diagnostic Imaging Symposium 2013 December 8-12, 2013 Disney s Yacht Club Resort Lake Buena Vista, Florida Self Assessment Module Questions, Answers and References Day SAM Title - Each SAM title

More information

Tips and Tricks to performing Magnetic Resonance Imaging Guided Breast Interventional Procedures Habib Rahbar, MD, FSBI October 23, 2018, 7:00pm ET

Tips and Tricks to performing Magnetic Resonance Imaging Guided Breast Interventional Procedures Habib Rahbar, MD, FSBI October 23, 2018, 7:00pm ET Tips and Tricks to performing Magnetic Resonance Imaging Guided Breast Interventional Procedures Habib Rahbar, MD, FSBI October 23, 2018, 7:00pm ET SAM Questions/Answers/Rationales/References 1. Below

More information

Breast MR biopsy. I Thomassin-Naggara, A.Jalaguier-Coudray, J Chopier

Breast MR biopsy. I Thomassin-Naggara, A.Jalaguier-Coudray, J Chopier Breast MR biopsy I Thomassin-Naggara, A.Jalaguier-Coudray, J Chopier Background EUSOBI When a radiologist perform a MR breast imaging he has to be able to realize or to be apart of a network who is able

More information

MR-guided breast biopsy at 3T: diagnostic yield of large core needle biopsy compared with vacuum-assisted biopsy

MR-guided breast biopsy at 3T: diagnostic yield of large core needle biopsy compared with vacuum-assisted biopsy Eur Radiol (2012) 22:341 349 DOI 10.1007/s00330-011-2272-6 BREAST MR-guided breast biopsy at 3T: diagnostic yield of large core needle biopsy compared with vacuum-assisted biopsy Carla Meeuwis & Jeroen

More information

Pitfalls and Limitations of Breast MRI. Susan Orel Roth, MD Professor of Radiology University of Pennsylvania

Pitfalls and Limitations of Breast MRI. Susan Orel Roth, MD Professor of Radiology University of Pennsylvania Pitfalls and Limitations of Breast MRI Susan Orel Roth, MD Professor of Radiology University of Pennsylvania Objectives Review the etiologies of false negative breast MRI examinations Discuss the limitations

More information

PURPOSE IMAGE-GUIDANCE MODALITIES IMAGE-GUIDED BREAST BIOPSY. US-Techniques. Ultrasound. US guided NLOBB. TH. Helbich

PURPOSE IMAGE-GUIDANCE MODALITIES IMAGE-GUIDED BREAST BIOPSY. US-Techniques. Ultrasound. US guided NLOBB. TH. Helbich IMAGE-GUIDED BREAST BIOPSY PURPOSE TH. Helbich Department of Radiology Division of Molecular & Gender Imaging Medical University of Vienna Imaging techniques Interventional procedures Quality management

More information

Atypical Ductal Hyperplasia and Ductal Carcinoma In Situ as Revealed by Large-Core Needle Breast Biopsy: Results of Surgical Excision

Atypical Ductal Hyperplasia and Ductal Carcinoma In Situ as Revealed by Large-Core Needle Breast Biopsy: Results of Surgical Excision Marla L. Rosenfield Darling 1 Darrell N. Smith 1 Susan C. Lester 2 Carolyn Kaelin 3 Donna-Lee G. Selland 1 Christine M. Denison 1 Pamela J. DiPiro 1 David I. Rose 1 Esther Rhei 3 Jack E. Meyer 1 Received

More information

Diagnostic benefits of ultrasound-guided. CNB) versus mammograph-guided biopsy for suspicious microcalcifications. without definite breast mass

Diagnostic benefits of ultrasound-guided. CNB) versus mammograph-guided biopsy for suspicious microcalcifications. without definite breast mass Volume 118 No. 19 2018, 531-543 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu Diagnostic benefits of ultrasound-guided biopsy versus mammography-guided

More information

Atypical ductal hyperplasia diagnosed at ultrasound guided biopsy of breast mass

Atypical ductal hyperplasia diagnosed at ultrasound guided biopsy of breast mass Atypical ductal hyperplasia diagnosed at ultrasound guided biopsy of breast mass Poster No.: C-1483 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit J. Cho, J. Chung, E. S. Cha, J. E.

More information

Stereotactic 11-Gauge Vacuum- Assisted Breast Biopsy: A Validation Study

Stereotactic 11-Gauge Vacuum- Assisted Breast Biopsy: A Validation Study Georg Pfarl 1 Thomas H. Helbich 1 Christopher C. Riedl 1 Teresa Wagner 2 Michael Gnant 3 Margaretha Rudas 4 Laura Liberman 5 Received March 11, 2002; accepted after revision May 17, 2002. 1 Department

More information

Macrobiopsy under X-Ray Guidance

Macrobiopsy under X-Ray Guidance Macrobiopsy under X-Ray Guidance C. Balleyguier, B. Boyer Radiology Gustave Roussy, Villejuif, France Breast Intervention Imaging Major domain in breast imaging European guidelines recommend a pre surgical

More information

Mammographic features and correlation with biopsy findings using 11-gauge stereotactic vacuum-assisted breast biopsy (SVABB)

Mammographic features and correlation with biopsy findings using 11-gauge stereotactic vacuum-assisted breast biopsy (SVABB) Original article Annals of Oncology 14: 450 454, 2003 DOI: 10.1093/annonc/mdh088 Mammographic features and correlation with biopsy findings using 11-gauge stereotactic vacuum-assisted breast biopsy (SVABB)

More information

Vacuum-assisted breast biopsy using computer-aided 3.0 T- MRI guidance: diagnostic performance in 173 lesions

Vacuum-assisted breast biopsy using computer-aided 3.0 T- MRI guidance: diagnostic performance in 173 lesions Vacuum-assisted breast biopsy using computer-aided 3.0 T- MRI guidance: diagnostic performance in 173 lesions Poster No.: C-2870 Congress: ECR 2017 Type: Scientific Exhibit Authors: A. Pozzetto, L. Camera,

More information

Clinical Practice Guideline for the Indications for Use of Breast Magnetic Resonance Imaging (MRI)

Clinical Practice Guideline for the Indications for Use of Breast Magnetic Resonance Imaging (MRI) CIHRT Exhibit P-2595 Page 1 Question: Clinical Practice Guideline for the Indications for Use of Breast Magnetic Resonance Imaging (MRI) Eastern Health Breast Disease Site Group What are the current indications

More information

High-Risk Lesions at MRI-Guided Breast Biopsy: Frequency and Rate of Underestimation

High-Risk Lesions at MRI-Guided Breast Biopsy: Frequency and Rate of Underestimation Women s Imaging Original Research Lourenco et al. MRI-Guided Biopsy of High-Risk Breast Lesions Women s Imaging Original Research Ana P. Lourenco Hanan Khalil Matthew Sanford,2 Linda Donegan Lourenco AP,

More information

Percutaneous Biopsy of the Breast

Percutaneous Biopsy of the Breast Percutaneous Biopsy of the Breast Expires Tuesday, March 31, 2020 Radiology Brooke A. Caldwell, M.D. Objectives 1. Describe the pros and cons of surgical biopsy and the reasoning behind the development

More information

Underestimation of Atypical Ductal Hyperplasia at Sonographically Guided Core Biopsy of the Breast

Underestimation of Atypical Ductal Hyperplasia at Sonographically Guided Core Biopsy of the Breast Women s Imaging Original Research Jang et al. Sonographic Breast Biopsy Women s Imaging Original Research WOMEN S IMAGING Underestimation of Atypical Ductal Hyperplasia at Sonographically Guided Core Biopsy

More information

Breast Health and Imaging Glossary

Breast Health and Imaging Glossary Contact: Lorna Vaughan HerSpace Breast Imaging & Biopsy Associates 300 State Route 35 South W. Long Branch, NJ 07764 732-571-9100, ext. 104 lorna@breast-imaging.com Breast Health and Imaging Glossary Women

More information

CURRENT METHODS IN IMAGE GUIDED BREAST BIOPSY

CURRENT METHODS IN IMAGE GUIDED BREAST BIOPSY CURRENT METHODS IN IMAGE GUIDED BREAST BIOPSY Stuart Silver April 24, 2004 OBJECTIVES Review development of current techniques Discuss stereotactic breast biopsy Discuss US guided breast biopsy 1 OBJECTIVES

More information

MR Imaging of the Ipsilateral Breast in Women with Percutaneously Proven Breast Cancer

MR Imaging of the Ipsilateral Breast in Women with Percutaneously Proven Breast Cancer Laura Liberman 1 Elizabeth A. Morris 1 D. David Dershaw 1 Andrea F. Abramson 1 Lee K. Tan 2 Received July 23, 2002; accepted after revision September 18, 2002. 1 Breast Imaging Section, Department of Radiology,

More information

Why Are Patients Noncompliant With Follow-Up Recommendations After MRI-Guided Core Needle Biopsy of Suspicious Breast Lesions?

Why Are Patients Noncompliant With Follow-Up Recommendations After MRI-Guided Core Needle Biopsy of Suspicious Breast Lesions? Women s Imaging Original Research Thompson et al. Noncompliance After Biopsy of Breast Lesions Women s Imaging Original Research Matthew O. Thompson 1 Jafi Lipson 2 Bruce Daniel 3 Chivonne Harrigal 2 Paul

More information

Poster No.: C-0466 Congress: ECR 2010 Scientific Exhibit

Poster No.: C-0466 Congress: ECR 2010 Scientific Exhibit Up-right stereotactic vacuum-assisted biopsy (UP-VAB) of non palpable breast lesions: Results and correlations with radiological suspicion (BI-RADS classification) Poster No.: C-0466 Congress: ECR 2010

More information

Consensus Guideline on Image-Guided Percutaneous Biopsy of Palpable and Nonpalpable Breast Lesions

Consensus Guideline on Image-Guided Percutaneous Biopsy of Palpable and Nonpalpable Breast Lesions Consensus Guideline on Image-Guided Percutaneous Biopsy of Palpable and Nonpalpable Breast Lesions Purpose: To outline the use of minimally invasive biopsy techniques (MIBT) for palpable and nonpalpable

More information

Short-term follow-up is the accepted management for mammographic lesions that are probably, but not definitely, benign. For these lesions to be classi

Short-term follow-up is the accepted management for mammographic lesions that are probably, but not definitely, benign. For these lesions to be classi Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Frequency of Malignancy

More information

Is Probably Benign Really Just Benign? Peter R Eby, MD, FSBI Virginia Mason Medical Center Seattle, WA

Is Probably Benign Really Just Benign? Peter R Eby, MD, FSBI Virginia Mason Medical Center Seattle, WA Is Probably Benign Really Just Benign? Peter R Eby, MD, FSBI Virginia Mason Medical Center Seattle, WA Disclosures: CONSULTANT FOR DEVICOR MEDICAL ARS Question 1 Is probably benign really just benign?

More information

Breast calcifications are important mammographic features in the

Breast calcifications are important mammographic features in the Diagn Interv Radiol 2012; 18:354 359 Turkish Society of Radiology 2012 BREAST IMAGING ORIGINAL ARTICLE Causes of failure in removing calcium in microcalcification-only lesions using 11-gauge stereotactic

More information

Sonographically-Guided 14-Gauge Core Needle Biopsy for Papillary Lesions of the Breast

Sonographically-Guided 14-Gauge Core Needle Biopsy for Papillary Lesions of the Breast Sonographically-Guided 14-Gauge Core Needle Biopsy for Papillary Lesions of the Breast Eun Sook Ko, MD Nariya Cho, MD Joo Hee Cha, MD Jeong Seon Park, MD Sun Mi Kim, MD Woo Kyung Moon, MD Index terms:

More information

Management of Palpable Abnormalities in the Breast Katerina Dodelzon, MD July 31, 2018, 7:00pm ET

Management of Palpable Abnormalities in the Breast Katerina Dodelzon, MD July 31, 2018, 7:00pm ET Management of Palpable Abnormalities in the Breast Katerina Dodelzon, MD July 31, 2018, 7:00pm ET SAM Questions 1. 21 year old female presenting with left breast palpable mass, what is the most appropriate

More information

Effective Health Care Program

Effective Health Care Program Comparative Effectiveness Review Number 19 Effective Health Care Program Comparative Effectiveness of Core-Needle and Open Surgical Biopsy for the Diagnosis of Breast Lesions Executive Summary Background

More information

Positive Predictive Value of

Positive Predictive Value of Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Mary C. Mahoney,

More information

Ductal carcinoma in situ, underestimation, ultrasound-guided core needle biopsy

Ductal carcinoma in situ, underestimation, ultrasound-guided core needle biopsy Ductal carcinoma in situ diagnosed after an ultrasoundguided 14-gauge core needle biopsy of breast masses: Can underestimation be predicted preoperatively? Poster No.: C-0442 Congress: ECR 2010 Type: Scientific

More information

Sonographically Guided Core Biopsy of the Breast: Comparison of 14-Gauge Automated Gun and 11-Gauge Directional Vacuum-Assisted Biopsy Methods

Sonographically Guided Core Biopsy of the Breast: Comparison of 14-Gauge Automated Gun and 11-Gauge Directional Vacuum-Assisted Biopsy Methods Sonographically Guided Core Biopsy of the Breast: Comparison of 14-Gauge Automated Gun and 11-Gauge Directional Vacuum-Assisted Biopsy Methods Nariya Cho, MD 1 Woo Kyung Moon, MD 1 Joo Hee Cha, MD 1 Sun

More information

Sonographic Detection and Sonographically Guided Biopsy of Breast Microcalcifications

Sonographic Detection and Sonographically Guided Biopsy of Breast Microcalcifications Sonographic Detection and Sonographically Guided Biopsy of Breast Microcalcifications Mary Scott Soo 1 Jay A. Baker Eric L. Rosen OBJECTIVE. The purpose of this study was to evaluate the ability of sonography

More information

BI-RADS Categorization As a Predictor of Malignancy 1

BI-RADS Categorization As a Predictor of Malignancy 1 Susan G. Orel, MD Nicole Kay, BA Carol Reynolds, MD Daniel C. Sullivan, MD BI-RADS Categorization As a Predictor of Malignancy 1 Index terms: Breast, biopsy, 00.1261 Breast neoplasms, localization, 00.125,

More information

A NEW BIOMARC TISSUE MARKER FOR BREAST BIOPSY: CLINICAL EVALUATION IN ULTRASOUND, MAMMOGRAPHY, CT SCANNING AND BREAST MRI

A NEW BIOMARC TISSUE MARKER FOR BREAST BIOPSY: CLINICAL EVALUATION IN ULTRASOUND, MAMMOGRAPHY, CT SCANNING AND BREAST MRI A NEW BIOMARC TISSUE MARKER FOR BREAST BIOPSY: CLINICAL EVALUATION IN ULTRASOUND, MAMMOGRAPHY, CT SCANNING AND BREAST MRI Michael T. Nelson, M.D. 2 Sina Meisamy, M.D. 1,2 Adeka D. McIntosh, M.D. 1,2 Patrick

More information

Second-look ultrasonography for MRI-detected suspicious breast lesions in patients with breast cancer

Second-look ultrasonography for MRI-detected suspicious breast lesions in patients with breast cancer Second-look ultrasonography for MRI-detected suspicious breast lesions in patients with breast cancer Min Ji Hong 1, Joo Hee Cha 1, Hak Hee Kim 1, Hee Jung Shin 1, Eun Young Chae 1, Ji Eun Shin 1,2, Woo

More information

Breast MRI: Friend or Foe?

Breast MRI: Friend or Foe? Breast MRI: Friend or Foe? UCSF Postgraduate Course May 18, 2013 Cheryl Ewing, MD Clinical Professor of Surgery UCSF Department of Surgery APPLEGATE HAS DOUBLE MASTECTOMY IN CANCER SCARE DIAGNOSED WITH

More information

Current Status of Supplementary Screening With Breast Ultrasound

Current Status of Supplementary Screening With Breast Ultrasound Current Status of Supplementary Screening With Breast Ultrasound Stephen A. Feig, M.D., FACR Fong and Jean Tsai Professor of Women s Imaging Department of Radiologic Sciences University of California,

More information

Spiculated breast masses on MRI: Which category should we choose, 4 or 5?

Spiculated breast masses on MRI: Which category should we choose, 4 or 5? Spiculated breast masses on MRI: Which category should we choose, 4 or 5? Poster No.: C-1394 Congress: ECR 2015 Type: Scientific Exhibit Authors: N. Onishi, S. Kanao, M. Kataoka, M. Kawai, M. Iima, A.

More information

Atypical Ductal Hyperplasia and Papillomas: A Comparison of Ultrasound Guided Breast Biopsy and Stereotactic Guided Breast Biopsy

Atypical Ductal Hyperplasia and Papillomas: A Comparison of Ultrasound Guided Breast Biopsy and Stereotactic Guided Breast Biopsy Atypical Ductal Hyperplasia and Papillomas: A Comparison of Ultrasound Guided Breast Biopsy and Stereotactic Guided Breast Biopsy Breast Cancer is the most common cancer diagnosed in women in the United

More information

Rate of Malignancy in MRI-Detected Probably Benign (BI-RADS 3) Lesions

Rate of Malignancy in MRI-Detected Probably Benign (BI-RADS 3) Lesions Women s Imaging Original Research Spick et al. Malignancy in MRI BI-RADS 3 Lesions Women s Imaging Original Research Claudio Spick 1,2 Dieter H. M. Szolar 1 Pascal A. Baltzer 2 Manfred Tillich 1 Pia Reittner

More information

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School Breast Cancer Screening Early detection of

More information

Breast MRI: Friend or Foe?

Breast MRI: Friend or Foe? Breast : Friend or Foe? APPLEGATE HAS DOUBLE MASTECTOMY IN CANCER SCARE DIAGNOSED WITH CANCER IN ONE BREAST Comments: 0 ASSOCIATED PRESS 8/19/2008 UCSF Postgraduate Course March 19, 2009 E. Shelley Hwang

More information

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School Breast Cancer Screening Early detection of

More information

BI-RADS CATEGORIZATION AND BREAST BIOPSY categorization in the selection of appropriate breast biopsy technique is also discussed. Patients and method

BI-RADS CATEGORIZATION AND BREAST BIOPSY categorization in the selection of appropriate breast biopsy technique is also discussed. Patients and method Original Article Positive Predictive Value of BI-RADS Categorization in an Asian Population Yah-Yuen Tan, Siew-Bock Wee, Mona P.C. Tan and Bee-Kiang Chong, 1 Departments of General Surgery and 1Diagnostic

More information

BREAST MRI. Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School

BREAST MRI. Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School BREAST MRI Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School BREAST MRI Any assessment of the breast parenchyma requires the administration

More information

Angela Gilliam, MD University of Colorado Surgical Grand Rounds November 3, 2008

Angela Gilliam, MD University of Colorado Surgical Grand Rounds November 3, 2008 Angela Gilliam, MD University of Colorado Surgical Grand Rounds November 3, 2008 Breast Cancer Most common cancer in American women 180,000 new cases per year Second most common cause of cancer death 44,000

More information

Pseudoaneurysm of the Breast During Vacuum-Assisted Removal

Pseudoaneurysm of the Breast During Vacuum-Assisted Removal Case Report Pseudoaneurysm of the Breast During Vacuum-Assisted Removal Yu-Mee Sohn, MD, Min Jung Kim, MD, Eun-Kyung Kim, MD, Sang Hoon Chung, MD, Jin Young Kwak, MD, Hee Jung Moon, MD, Soo Jin Kim, MD

More information

Imaging-Guided Core Needle Biopsy of Papillary Lesions of the Breast

Imaging-Guided Core Needle Biopsy of Papillary Lesions of the Breast Eric L. Rosen 1 Rex C. Bentley 2 Jay A. Baker 1 Mary Scott Soo 1 Received January 30, 2002; accepted after revision April 12, 2002. 1 Department of Radiology, Breast Imaging Division, Duke University Medical

More information

Multicenter Evaluation of the Breast Lesion Excision System, a Percutaneous, Vacuum-Assisted, Intact-Specimen Breast Biopsy Device

Multicenter Evaluation of the Breast Lesion Excision System, a Percutaneous, Vacuum-Assisted, Intact-Specimen Breast Biopsy Device 945 Multicenter Evaluation of the Breast Lesion Excision System, a Percutaneous, Vacuum-Assisted, Intact-Specimen Breast Biopsy Device Angela Sie, MD 1 David C. Bryan, MD 2 Victor Gaines, MD 3 Larry K.

More information

DCIS of the Breast--MRI findings with mammographic correlation.

DCIS of the Breast--MRI findings with mammographic correlation. DCIS of the Breast--MRI findings with mammographic correlation. Poster No.: C-1560 Congress: ECR 2013 Type: Educational Exhibit Authors: N. B. Ibrahim, P. Morris, S. ANANDAN; Burlington, MA/US Keywords:

More information

Screening for breast cancer with MRI: recent experience from the Australian Capital Territory

Screening for breast cancer with MRI: recent experience from the Australian Capital Territory Journal of Medical Imaging and Radiation Oncology 53 (2009) 69 80 RADIOLOGY ORIGINAL ARTICLE Screening for breast cancer with MRI: recent experience from the Australian Capital Territory J Price and SW

More information

Using lesion washout volume fraction as a biomarker to improve suspicious breast lesion characterization

Using lesion washout volume fraction as a biomarker to improve suspicious breast lesion characterization JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 5, 2015 Using lesion washout volume fraction as a biomarker to improve suspicious breast lesion characterization Jie Huang, a Sarah M. Schafer,

More information

Excisional biopsy or long term follow-up results in breast high-risk lesions diagnosed at core needle biopsy

Excisional biopsy or long term follow-up results in breast high-risk lesions diagnosed at core needle biopsy Excisional biopsy or long term follow-up results in breast high-risk lesions diagnosed at core needle biopsy Poster No.: C-2515 Congress: ECR 2015 Type: Authors: Scientific Exhibit Ö. S. Okcu 1, A. Oktay

More information

Breast ultrasound appearances after Mammotome vacuumassisted

Breast ultrasound appearances after Mammotome vacuumassisted Breast ultrasound appearances after Mammotome vacuumassisted biopsy. Poster No.: C-1924 Congress: ECR 2011 Type: Educational Exhibit Authors: R. Patel 1, G. R. Kaplan 2 ; 1 London/UK, 2 Herts/UK Keywords:

More information

Stereotactic vacuum-assisted breast biopsy under lateral decubitus position

Stereotactic vacuum-assisted breast biopsy under lateral decubitus position ORIGINAL ARTICLE pissn 2288-6575 eissn 2288-6796 http://dx.doi.org/10.4174/astr.2016.90.1.16 Annals of Surgical Treatment and Research Stereotactic vacuum-assisted breast biopsy under lateral decubitus

More information

Original Article Breast Imaging

Original Article Breast Imaging Original Article Breast Imaging http://dx.doi.org/10.3348/kjr.2014.15.6.697 pissn 1229-6929 eissn 2005-8330 Korean J Radiol 2014;15(6):697-703 Percutaneous Ultrasound-Guided Vacuum-Assisted Removal versus

More information

BREAST MRI. Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School

BREAST MRI. Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School BREAST MRI Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School BREAST MRI Any assessment of the breast parenchyma requires the administration

More information

Stereotactic vacuum-assisted biopsy (SVAB) of Nonpalpable Breast Microcalcifications: Advantage of clip placement (Prospective study)

Stereotactic vacuum-assisted biopsy (SVAB) of Nonpalpable Breast Microcalcifications: Advantage of clip placement (Prospective study) Department of Radiology, University of Ulm Director: Prof. Dr. med. HJ Brambs Stereotactic vacuum-assisted biopsy (SVAB) of Nonpalpable Breast Microcalcifications: Advantage of clip placement (Prospective

More information

Prone table stereotactic breast biopsy

Prone table stereotactic breast biopsy JYH Hui LK Chan RLM Chan AWL Lau J Lo JCS Chan HS Lam MEDICAL PRACTICE Prone table stereotactic breast biopsy!"#$%&'()*+, The prone table machine is a mammographic X-ray system specially designed for use

More information

ANNEX 1 OBJECTIVES. At the completion of the training period, the fellow should be able to:

ANNEX 1 OBJECTIVES. At the completion of the training period, the fellow should be able to: 1 ANNEX 1 OBJECTIVES At the completion of the training period, the fellow should be able to: 1. Breast Surgery Evaluate and manage common benign and malignant breast conditions. Assess the indications

More information

ACR Appropriateness Criteria on Nonpalpable Mammographic Findings (Excluding Calcifications)

ACR Appropriateness Criteria on Nonpalpable Mammographic Findings (Excluding Calcifications) ACR Appropriateness Criteria on Nonpalpable Mammographic Findings (Excluding Calcifications) Mary S. Newell, MD a, Robyn L. Birdwell, MD b, Carl J. D Orsi, MD c, Lawrence W. Bassett, MD d, Mary C. Mahoney,

More information

Breast Cancers Not Detected at MRI: Review of False-Negative Lesions

Breast Cancers Not Detected at MRI: Review of False-Negative Lesions Women s Imaging Original Report Shimauchi et al. MRI-Occult Breast s Women s Imaging Original Report WOMEN S IMAGING Breast s Not Detected at MRI: Review of False-Negative Lesions Akiko Shimauchi 1 Sanaz

More information

Role of Breast MRI in the Preoperative Evaluation of Patients with Newly Diagnosed Breast Cancer

Role of Breast MRI in the Preoperative Evaluation of Patients with Newly Diagnosed Breast Cancer Women s Imaging Original Research Schell et al. Preoperative Breast MRI Women s Imaging Original Research WOMEN S IMAGING Amy M. Schell 1 Kari Rosenkranz 2 Petra J. Lewis 3 Schell AM, Rosenkranz K, Lewis

More information

Learning Curve for Stereotactic Breast Biopsy: How Many Cases Are Enough?

Learning Curve for Stereotactic Breast Biopsy: How Many Cases Are Enough? Laura Liberman 1 Catherine L. Benton D. David Dershaw Andrea F. Abramson Linda R. LaTrenta Elizabeth A. Morris Received July 17, 2000; accepted after revision September 7, 2000. Supported by grant C015709

More information

Original Report. Mucocele-Like Tumors of the Breast: Mammographic and Sonographic Appearances. Katrina Glazebrook 1 Carol Reynolds 2

Original Report. Mucocele-Like Tumors of the Breast: Mammographic and Sonographic Appearances. Katrina Glazebrook 1 Carol Reynolds 2 Katrina Glazebrook 1 Carol Reynolds 2 Received January 2, 2002; accepted after revision August 28, 2002. 1 Department of Radiology, Mayo Clinic, 200 First St. S.W., Rochester, MN 55905. Address correspondence

More information

Six-Month Short-Interval Imaging Follow-Up for Benign Concordant Core Needle Biopsy of the Breast: Outcomes in 1444 Cases With Long-Term Follow-Up

Six-Month Short-Interval Imaging Follow-Up for Benign Concordant Core Needle Biopsy of the Breast: Outcomes in 1444 Cases With Long-Term Follow-Up Women s Imaging Original Research Monticciolo et al. Short-Interval Imaging Follow-Up for Benign Breast Biopsies Women s Imaging Original Research Debra L. Monticciolo 1 Rodney L. Hajdik 1 Mason G. Hicks

More information

Breast Cancers Not Detected at MRI: Review of False-Negative Lesions

Breast Cancers Not Detected at MRI: Review of False-Negative Lesions Women s Imaging Original Report Shimauchi et al. MRI-Occult Breast s Women s Imaging Original Report WOMEN S IMAGING Akiko Shimauchi 1 Sanaz A. Jansen 1 Hiroyuki Abe 1 Nora Jaskowiak 2 Robert A. Schmidt

More information

A Manipulator System for 14-Gauge Large Core Breast Biopsies Inside a High-Field Whole-Body MR Scanner

A Manipulator System for 14-Gauge Large Core Breast Biopsies Inside a High-Field Whole-Body MR Scanner JOURNAL OF MAGNETIC RESONANCE IMAGING 17:493 498 (2003) Technical Note A Manipulator System for 14-Gauge Large Core Breast Biopsies Inside a High-Field Whole-Body MR Scanner Stefan O.R. Pfleiderer, MD,

More information

Magnetic Resonance Imaging Screening of Women at High Risk for Breast Cancer

Magnetic Resonance Imaging Screening of Women at High Risk for Breast Cancer Evidence-Based Series 15-11 Version 2 IN REVIEW A Quality Initiative of the Program in Evidence-Based Care (PEBC), Cancer Care Ontario (CCO) Magnetic Resonance Imaging Screening of Women at High Risk for

More information

CNB vs Surgical Excision

CNB vs Surgical Excision Update on Core Needle Biopsy of Non-palpable Breast Lesions Nour Sneige, M.D. UT MD Anderson Cancer Center Houston, Tx Image-Guided CNB of Breast Lesions An alternative to surgical biospy CNB vs Surgical

More information

Department of Radiology, Research Institute of Radiological Science, Yonsei University, College of Medicine, Seodaemun-gu, Seoul, Republic of Korea

Department of Radiology, Research Institute of Radiological Science, Yonsei University, College of Medicine, Seodaemun-gu, Seoul, Republic of Korea The British Journal of Radiology, 85 (2012), e349 e356 Comparison of the underestimation rate in cases with ductal carcinoma in situ at ultrasound-guided core biopsy: 14-gauge automated core-needle biopsy

More information

Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, cont

Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, cont Note: This copy is for your personal, non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Risk of Upgrade

More information

Aims and objectives. Page 2 of 10

Aims and objectives. Page 2 of 10 Diagnostic performance of automated breast volume scanner (ABVS) versus hand-held ultrasound (HHUS) as second look for breast lesions detected only on magnetic resonance imaging. Poster No.: C-1701 Congress:

More information

Atypical proliferative lesions diagnosed on core biopsy - 6 year review

Atypical proliferative lesions diagnosed on core biopsy - 6 year review Atypical proliferative lesions diagnosed on core biopsy - 6 year review Dr Angela Harris, Dr Julie Weigner & Dr Ricardo Vilain NSW Health Pathology Pathology North, Hunter Anatomical Pathology & Cytology

More information

MRI for breast cancer screening

MRI for breast cancer screening 17 (Supplement 10): x325 x331, 2006 doi:10.1093/annonc/mdl281 MRI for breast cancer screening M. O. Leach Cancer Research UK Clinical Magnetic Resonance Research Group, Institute of Cancer Research, Royal

More information

The Future of Breast MRI Improving Outcomes

The Future of Breast MRI Improving Outcomes The Future of Breast MRI Improving Outcomes Connie Lehman MD PhD Professor of Radiology Harvard Medical School Director of Breast Imaging Massachusetts General Hospital Opportunities New technology provides

More information

Correlation between lesion type and the additional value of digital breast tomosynthesis

Correlation between lesion type and the additional value of digital breast tomosynthesis Correlation between lesion type and the additional value of digital breast tomosynthesis Poster No.: C-1604 Congress: ECR 2011 Type: Scientific Exhibit Authors: C. Van Ongeval, L. Cockmartin, A. Van Steen,

More information

False-Positive Findings at Contrast- Enhanced Breast MRI: A BI-RADS Descriptor Study

False-Positive Findings at Contrast- Enhanced Breast MRI: A BI-RADS Descriptor Study Women s Imaging Original Research Baltzer et al. BI-RADS MRI Women s Imaging Original Research WOMEN S IMAGING Pascal A. T. Baltzer 1 Matthias Benndorf 1 Matthias Dietzel 1 Mieczyslaw Gajda 2 Ingo B. Runnebaum

More information

Here are examples of bilateral analog mammograms from the same patient including CC and MLO projections.

Here are examples of bilateral analog mammograms from the same patient including CC and MLO projections. Good afternoon. It s my pleasure to be discussing Diagnostic Breast Imaging over the next half hour. I m Wei Yang, Professor of Diagnostic Radiology and Chief, the Section of Breast Imaging as well as

More information

Magnetic Resonance Imaging and Breast Ultrasonography as an Adjunct to Mammographic Screening in High-Risk Patients

Magnetic Resonance Imaging and Breast Ultrasonography as an Adjunct to Mammographic Screening in High-Risk Patients Magnetic Resonance Imaging and Breast Ultrasonography as an Adjunct to Mammographic Screening in High-Risk Patients Huong T. Le-Petross, MD,* and Mahesh K. Shetty, MD, FRCR, FACR Screening mammography

More information

Stereotactic Core-Needle Biopsy of Non-Mass Calcifications: Outcome and Accuracy at Long-Term Follow-Up

Stereotactic Core-Needle Biopsy of Non-Mass Calcifications: Outcome and Accuracy at Long-Term Follow-Up Stereotactic Core-Needle Biopsy of Non-Mass Calcifications: Outcome and Accuracy at Long-Term Follow-Up Boo-Kyung Han, MD 1 Yeon Hyeon Choe, MD 1 Young-Hyeh Ko, MD 2 Seok-Jin Nam, MD 3 Jung-Han Kim, MD

More information

Breast Cancer Screening

Breast Cancer Screening Scan for mobile link. Breast Cancer Screening What is breast cancer screening? Screening examinations are tests performed to find disease before symptoms begin. The goal of screening is to detect disease

More information

Computer-Aided Evaluation of Malignancy with Magnetic Resonance Imaging of the Breast. Original Policy Date

Computer-Aided Evaluation of Malignancy with Magnetic Resonance Imaging of the Breast. Original Policy Date MP 6.01.36 Computer-Aided Evaluation of Malignancy with Magnetic Resonance Imaging of the Breast Medical Policy Section Radiology Issue 12:2013 Original Policy Date 12:2013 Last Review Status/Date Reviewed

More information

This is the point of Accuracy

This is the point of Accuracy This is the point of Accuracy Mammotome EX is also available to you, your patients with benign breast disease and the ones they love. Please contact your Breast Care Specialist for more information. This

More information

Computer Aided Detection (CAD) for Breast MRI

Computer Aided Detection (CAD) for Breast MRI Technology in Cancer Research & Treatment ISSN 1533-0346 Volume 4, Number 1, February (2005) Adenine Press (2005) Computer Aided Detection (CAD) for Breast MRI www.tcrt.org Since 1999, there has been a

More information

Electrical impedance scanning of the breast is considered investigational and is not covered.

Electrical impedance scanning of the breast is considered investigational and is not covered. ARBenefits Approval: 09/28/2011 Effective Date: 01/01/2012 Revision Date: Code(s): Medical Policy Title: Electrical Impedance Scanning of the Breast Document: ARB0127 Administered by: Public Statement:

More information

ORIGINAL PAPER. Background parenchymal enhancement in preoperative breast MRI

ORIGINAL PAPER. Background parenchymal enhancement in preoperative breast MRI Nagoya J. Med. Sci. 77. 373 ~ 382, 2015 ORIGINAL PAPER Background parenchymal enhancement in preoperative breast MRI Satoko Kohara 1, Satoko Ishigaki 1, Hiroko Satake 1, Akiko Kawamura 1, Hisashi Kawai

More information

Journal of Breast Cancer

Journal of Breast Cancer Journal of Breast Cancer ORIGINAL ARTICLE J Breast Cancer 2014 September; 17(3): 265-269 Absence of Residual Microcalcifications in Atypical Ductal Hyperplasia Diagnosed via Stereotactic Vacuum-Assisted

More information

Dense Breasts, Get Educated

Dense Breasts, Get Educated Dense Breasts, Get Educated What are Dense Breasts? The normal appearances to breasts, both visually and on mammography, varies greatly. On mammography, one of the important ways breasts differ is breast

More information

ACR PRACTICE GUIDELINE FOR THE PERFORMANCE OF CONTRAST-ENHANCED MAGNETIC RESONANCE IMAGING (MRI) OF THE BREAST

ACR PRACTICE GUIDELINE FOR THE PERFORMANCE OF CONTRAST-ENHANCED MAGNETIC RESONANCE IMAGING (MRI) OF THE BREAST BE IT RESOLVED, Sponsored By: RESOLUTION NO. 12 that the American College of Radiology adopt the ACR Practice Guideline for the Performance of Contrast-Enhanced Magnetic Resonance Imaging (MRI) of the

More information

Breast Lesion Excision System-Intact (BLES): A Stereotactic Method of Biopsy of Suspicius Non-Palpable Mammographic Lesions.

Breast Lesion Excision System-Intact (BLES): A Stereotactic Method of Biopsy of Suspicius Non-Palpable Mammographic Lesions. Breast Lesion Excision System-Intact (BLES): A Stereotactic Method of Biopsy of Suspicius Non-Palpable Mammographic Lesions. Poster No.: C-1595 Congress: ECR 2014 Type: Authors: Scientific Exhibit I. Georgiou

More information

Atypical papillary lesions after core needle biopsy and subsequent breast carcinoma

Atypical papillary lesions after core needle biopsy and subsequent breast carcinoma Asian Biomedicine Vol. 5 No. 2 April 2011; 243-248 DOI: 10.5372/1905-7415.0502.031 Original article Atypical papillary lesions after core needle biopsy and subsequent breast carcinoma Tuenchit Khamapirad

More information

New Palpable Breast Lump With Recent Negative Mammogram: Is Repeat Mammography Necessary?

New Palpable Breast Lump With Recent Negative Mammogram: Is Repeat Mammography Necessary? Women s Imaging Original Research Leung et al. Repeat Mammogram for Breast Lump Found After Negative Mammogram Women s Imaging Original Research Stephanie E. Leung 1 Ilanit Ben-Nachum Anat Kornecki Leung

More information

Medical Audit of Diagnostic Mammography Examinations: Comparison with Screening Outcomes Obtained Concurrently

Medical Audit of Diagnostic Mammography Examinations: Comparison with Screening Outcomes Obtained Concurrently Katherine E. Dee 1,2 Edward A. Sickles 1 Received July 3, 2000; accepted after revision September 12, 2000. Presented in part at the annual meeting of the American Roentgen Ray Society, Washington, DC,

More information