The Effect of Image Processing on the Detection of Cancers in Digital Mammography

Size: px
Start display at page:

Download "The Effect of Image Processing on the Detection of Cancers in Digital Mammography"

Transcription

1 Medical Physics and Informatics Original Research Warren et al. Image Processing in Digital Mammography Medical Physics and Informatics Original Research Lucy M. Warren 1,2 Rosalind M. Given-Wilson 3 Matthew G. Wallis 4,5 Julie Cooke 6 Mark D. Halling-Brown 1 Alistair Mackenzie 1,2 Dev P. Chakraborty 7 Hilde Bosmans 8 David R. Dance 1,2 Kenneth C. Young 1,2 Warren LM, Given-Wilson RM, Wallis MG, et al. Keywords: digital mammography, image processing, observer performance DOI: /AJR Received August 27, 2013; accepted after revision December 3, This work is part of the OPTIMAM project and was supported by Cancer Research UK & Engineering and Physical Sciences Research Council Cancer Imaging Programme in Surrey in association with the Medical Research Council and Department of Health (England). D. P. Chakraborty was supported in part by grants from the U.S. Department of Health & Human Services, National Institutes of Health (ROI-EB and ROI-EB008688). 1 National Coordinating Centre for the Physics of Mammography, Royal Surrey County Hospital, NHS Foundation Trust, Guildford, GU2 7XX, UK. Address correspondence to L. M. Warren (lucy.warren@nhs.net). 2 Department of Physics, Faculty of Engineering and Physical Sciences, University of Surrey, Guildford, UK. 3 Department of Radiology, St. George s Healthcare NHS Trust, Tooting, London, UK. 4 Cambridge Breast Unit, Cambridge University Hospitals, NHS Foundation Trust, Cambridge, UK. 5 NIHR Cambridge Biomedical Research Centre, Cambridge, UK. 6 Jarvis Breast Screening and Diagnostic Centre, Guildford, UK. 7 Department of Radiology, University of Pittsburgh, Pittsburgh, PA. 8 University Hospitals Leuven, Leuven, Belgium. AJR 2014; 203: X/14/ American Roentgen Ray Society The Effect of Image Processing on the Detection of Cancers in Digital Mammography OBJECTIVE. The objective of our study was to investigate the effect of image processing on the detection of cancers in digital mammography images. MATERIALS AND METHODS. Two hundred seventy pairs of breast images (both breasts, one view) were collected from eight systems using Hologic amorphous selenium detectors: 80 image pairs showed breasts containing subtle malignant masses; 30 image pairs, biopsy-proven benign lesions; 80 image pairs, simulated calcification clusters; and 80 image pairs, no cancer (normal). The 270 image pairs were processed with three types of image processing: standard (full enhancement), low contrast (intermediate enhancement), and pseudo film-screen (no enhancement). Seven experienced observers inspected the images, locating and rating regions they suspected to be cancer for likelihood of malignancy. The results were analyzed using a jackknife-alternative free-response receiver operating characteristic (JAFROC) analysis. RESULTS. The detection of calcification clusters was significantly affected by the type of image processing: The JAFROC figure of merit (FOM) decreased from 0.65 with standard image processing to 0.63 with low-contrast image processing (p = 0.04) and from 0.65 with standard image processing to 0.61 with film-screen image processing (p = ). The detection of noncalcification cancers was not significantly different among the image-processing types investigated (p > 0.40). CONCLUSION. These results suggest that image processing has a significant impact on the detection of calcification clusters in digital mammography. For the three image-processing versions and the system investigated, standard image processing was optimal for the detection of calcification clusters. The effect on cancer detection should be considered when selecting the type of image processing in the future. I n digital mammography, the final stage in the image formation process is the application of image processing. Image processing is applied to enhance the diagnostic content within the image for example, by reducing the noise or enhancing the edges and image contrast. Each manufacturer applies its proprietary image-processing algorithm, and the image appearances resulting from this processing can be very different. A number of studies have been performed to investigate the effect of the difference in image appearance as a result of applying different image-processing algorithms: These studies include both preference studies [1, 2] and objective studies [3 9], and the results of these studies are mixed. Some of the objective studies found significant differences among different image-processing algo- rithms [3 5], and others did not find a significant difference [6 9]. However, these studies had the following limitations: Different imageprocessing algorithms were used on systems with different detectors [3], different radiologists inspected different images [8], localization was not included in the task [4], not all cases were from different patients to maintain the independence assumed in the analysis [6], the cancers tended to be obvious and easily detected in all image-processing types investigated [7], and only calcifications were investigated (i.e., noncalcification cancers were not studied) [5, 9]. Therefore, the effect of the difference in image appearance resulting from using different image-processing algorithms on the detection of the various radiologic features of breast cancer is still unclear. The purpose of this work was to perform a retrospective observer study to investigate AJR:203, August

2 Warren et al. the impact of different image-processing algorithms on the detection of cancers in digital mammography. Our approach overcame several of the limitations described previously. Subtle simulated calcifications and subtle noncalcification cancers were included in the study; all the observers inspected all images processed with all image-processing algorithms investigated; the cases were from different patients, thus maintaining independence of cases; the image-processing algorithms investigated were representative of the range currently used in clinical practice; and the study was based on objective measurements rather than observer preference. Materials and Methods The study protocol was approved by the regional research ethics committee. Our ethics committee did not require individual consent because our study used anonymous mammographic images and data. Case Collection An expert breast radiologist (25 years experience and 6000 screening mammography studies read annually) who did not take part in the study inspected the images and the associated clinical information of 238 patients with pathology-proven cancer. One hundred and eighty five of these women presented for screening and 53 were women showing symptoms of cancer. These cases were collected at two sites using eight mammography systems with amorphous selenium detectors (HologicSelenia and Hologic Dimensions, Hologic). In the United Kingdom National Health Service (NHS) Breast Screening Programme, women years old are invited to undergo mammographic screening every 3 years. Women older than 70 years may refer themselves. The examination consists of two views TABLE 1: Malignant Mammographic Abnormalities Used in This Study Abnormality No. (%) of Cases Soft-tissue density (mass) 40 (48.2) Architectural distortion 19 (22.9) Asymmetric density 24 (28.9) of each breast: a mediolateral oblique (MLO) view and a craniocaudal view. The radiologist marked the outline of each cancer in both views (MLO and craniocaudal) on digital images using an in-house web interface [10] and rated lesion conspicuity in each view (very subtle, subtle, or visible). Eighty cases with malignant noncalcification lesions (i.e., mass, architectural distortion, or asymmetric density) were selected from these 238 cases by the expert radiologist. The expert radiologist chose cases in which lesion conspicuity was subtle or very subtle but the lesions were still detectable within the images because obvious cancers would be easily detected under all image-processing options. In total, there were 83 noncalcification cancers in the 80 cases; one case was a patient who had bilateral cancer with noncalcification cancers in both breasts. A more detailed description of these cancers is given in Table 1. In addition, the expert radiologist also inspected the images of 52 consecutive patients with biopsy-proven benign lesions (50 screening-detected cases and two symptomatic cases) and marked the location of the lesions. The images of 30 patients with biopsy-proven benign lesions were randomly selected from these 52 cases for the study. These were a mixture of calcification and noncalcification lesions with a mixture of conspicuities. The images of 160 patients assessed as showing normal findings (i.e., patient not recalled from screening) were also used in the study. Because of the time constraints of the study, the patients had not yet returned for their next screening session, so these normal cases have not been followed up. The interval cancer rate has been shown to be 0.55 cancers per 1000 women screened for the 12-month period after negative findings on a screening study in the United Kingdom [11]. Therefore, of the 270 patients in our study, we could expect 0.15 interval cancers. Although ideally all patients with normal findings would have been followed up, we do not think this lack of follow-up has significantly impacted the results of this study. The number of subtle calcification clusters within the 238 cancer cases collected was smaller than the number of calcification clusters required for sufficient statistical power in this study. Therefore, instead simulated calcifications were inserted into 80 of the 160 images with normal findings. The method used for calcification cluster simulation is described in the next section. At the time of the study, we did not have simulation models of noncalcification cancers that were suitably advanced for use; therefore, images of real noncalcification cancers as described earlier were used. Calcification Cluster Simulation Simulated calcification clusters have been used in observer studies of this type in the past [5, 9, 12]. The main advantage of using simulated cancers is that it is very time-consuming to collect real subtle cancers because of the low prevalence of cancers in breast screening, whereas a database of simulated cancers can be generated in a shorter time period. An additional advantage of using simulated cancers for this study is that cancers detected in the clinic will be found by a radiologist viewing images acquired on a particular digital mammography A Fig. 1 For this study, images of calcification clusters were inserted into breast images that had originally showed normal findings. A C, Three examples of calcification clusters inserted into breast images. Regions shown are mm 2. B C 388 AJR:203, August 2014

3 Image Processing in Digital Mammography system with a particular image-processing algorithm applied. Therefore, if the clinically used image-processing algorithm is one of the arms in the study, because all the cancers have already been detected with this image-processing algorithm, there may be a bias in favor of this clinically used imageprocessing algorithm. When simulated cancers are used, this bias is removed. The simulated calcification cluster images were derived from unprocessed digital images of sliced mastectomy samples imaged on a digital microfocus specimen x-ray cabinet (model MX20 DC2, Faxitron X-Ray, FaxitronBioptics), at five times magnification with an effective pixel size of 10 μm. The clusters within the mastectomy samples were a mixture of ductal carcinoma in situ and invasive carcinoma. From these images, calcification clusters were extracted using a method described elsewhere [9, 13] to generate an image of the calcification cluster with only the surrounding background tissue removed. If inserted into the normal breast images without any other adjustments, the clusters within the image would appear different than if the cluster had been within the patient s breast when imaged. This difference results from differences in magnifications, increased scatter, and decreased subject contrast due to the increased thickness of surrounding breast tissue and the different imaging system and imaging parameters used when acquiring the cluster images and normal breast images. Therefore, before a simulated calcification cluster was inserted, its appearance was adjusted to account for these differences. This process has been described and validated elsewhere [14, 15]. Calcification clusters were inserted into the breast images in a range of locations while avoiding positions very close to the skin edge and nipple. The breasts into which the calcification clusters were inserted were randomly selected from all normal images collected, so the breasts were of all densities. When the location to insert a cluster was being selected, the breast tissue immediately surrounding the cluster was categorized as homogeneously fatty, a mixture of fatty and glandular tissue, or homogeneously glandular. The locations were selected so that there were equal numbers of clusters in the three broad categories of local breast density. In total, 89 calcification clusters were inserted with between one and three clusters inserted per case. Examples of simulated calcification clusters inserted into breast images are shown in Figure 1. It was important that the calcification clusters were realistic in appearance; therefore, once inserted within the breast images, the calcification clusters were inspected by the same expert radiologist who judged them to be realistic in appearance and location and to be subtle in appearance. Image Processing The 80 cases with inserted calcification clusters, 80 normal cases with no cancer present, 80 cases with noncalcification cancers, and 30 cases with biopsy-proven benign lesions made up the set of 270 cases used in the study. These 270 cases were processed using three different types of Hologic image processing (Fig. 2). The first type was the manufacturer s standard image processing and is referred to as the standard image processing (Fig. 2A). The second version has an intermediate amount of enhancement. This type is referred to as the low-contrast image processing (Fig. 2B). The third has no additional enhancement except skin edge enhancement, so it simulates the appearance of a film-screen image. This type is referred to as film-screen image processing (Fig. 2C). The lowcontrast and the film-screen image-processing algorithms were developed by Hologic for investigational purposes only. However, the three types of image processing were selected for this study to be representative of the wide range in image-processing al- A Fig year-old woman with noncalcification cancer. Expert radiologist classified lesion conspicuity in this case, which was used in study, as subtle. A C, Subtle noncalcification cancer is shown in insets with standard image processing (A), low-contrast image processing (B), film-screen image processing (C). B C AJR:203, August

4 Warren et al. gorithms seen on the different digital mammography systems currently used in clinical practice. For all the cases used in the study, images of both breasts were displayed but only one view (MLO or craniocaudal) was shown. Ideally, two views of both breasts would be displayed, as in the clinic. However, only 2D simulated calcification clusters were available and these could not be reoriented for simulation into both views. For normal and benign cases, the view was selected randomly (subject to the constraint that the benign lesions be visible in the randomly selected view); for cases containing noncalcification cancers, the view in which cancer was most subtle while remaining detectable was selected. Of the 270 image pairs used in the study, 142 were MLO views and 128 were craniocaudal views. Observer Study Protocol Seven mammography readers (six radiologists and one radiographer) who had 2 22 years of experience reading mammograms (2 7 years with digital mammography) and who read at least 5000 screening examinations per year took part in the observer study. All observers were certified to interpret mammography images as defined by the United Kingdom NHS Breast Screening Programme [16] and had experience reading similar images (i.e., from this manufacturer) in their clinical practice. Each observer inspected the images in six sessions, and a break was taken halfway through each session. The images were organized so that there was at least 2 weeks between each observer seeing the same image pair with different image processing applied, and the sequence was also randomized so that each observer was shown the images in a different order. Observers were trained before the study in a pilot study (using 180 image pairs) to become familiar with the software, the different image-processing types, and the task they were being asked to perform. Additional training with a smaller set of 10 image pairs was performed at the beginning of the main study. The study was performed using a web-based interface developed in-house [10]. This interface allowed the study to be performed remotely at two sites simultaneously using the clinical workstation in each site. Each workstation contained a pair of 5-megapixel monitors (site 1: model MDMG-5121, Barco; site 2: model GS521-CL, Eizo) calibrated using the vendors software to the DICOM grayscale standard display function [17]. At both sites, room illuminance was maintained at an ambient level during sessions [18]. Observers were told there could be no lesions, one lesion, or multiple lesions present within each image. They were asked to mark the center of any region that they were suspicious of being TABLE 2: Difference in Reader-Averaged Figure of Merit (FOM) When Readers Are Treated as Fixed and Random Factors Using the ANOVA Technique Image-Processing Pair Being Compared Cancer Type cancerous and were asked not to mark benign features. They were asked to identify if they were marking a calcification cluster, noncalcification cancer, or both. The observer was then asked the following question: What is the likelihood that this lesion is malignant? This question was answered on a 5-point scale, with 1 being the lowest likelihood of malignancy and 5 being the highest likelihood of malignancy. Statistical Analysis For data analysis, cases containing biopsyproven malignant noncalcification cancers and simulated calcification clusters were assumed to be positive for disease. Cases containing biopsyproven benign lesions or normal cases were assumed to be negative for disease. A mark made by an observer had to be within the outline of the cancer made by the expert radiologist to be a correctly localized cancer. All other marks were considered false-positives. The ratings given to each lesion were analyzed using jackknife-alternative free-response receiver operating characteristic (JAFROC) analysis [19]. In this analysis, the performance of an observer inspecting images with a particular imageprocessing algorithm applied is expressed as the figure of merit (FOM) and is calculated as the trapezoidal area underneath the alternative free-response receiver operating characteristic (AFROC) curve when extended to (1,1). This straight line extension credits perfect decisions, such as normal images with no marks, and penalizes unmarked lesions [20]. JAFROC analysis was performed using JAFROC software (JAFROC, version 4.0, D. P. Chakraborty). This analysis performs significance testing using the Dorfman-Berbaum-Metz ANOVA technique [21, 22]; the machine code for the requisite ANOVA module was obtained from the authors of that software package. In the primary analysis, only cases were treated as a random Difference in Reader-Averaged FOM (95% CI) Readers as Fixed Factor Readers as Random Factor Standard vs film-screen Mass ( to 0.029) ( to 0.034) Calcification ( ) a ( ) a Standard vs low contrast Mass ( to 0.031) ( to 0.036) Calcification ( ) a ( to 0.060) Film-screen vs low contrast Mass ( to 0.024) ( to 0.029) Calcification ( to 0.006) ( to 0.020) Note When the 95% CIs do not include zero, there is a significant difference between the image-processing pair. a There is a significant difference between the image-processing pair. factor (so that the results apply to the population of cases but only for the readers used in the study). Secondary analysis was also performed with both cases and readers treated as random factors. For both, a p value < 0.05 was required for significance. An additional analysis compared the number of false-positives marked in images for each type of image processing. The number of false-positives was jack-knifed over cases, and Dorfman-Berbaum-Metz ANOVA was performed. This analysis was first performed using false-positives on all cases. The analysis was then repeated for two subgroups of cases: normal cases and cases with biopsy-proven benign lesions. Finally, the number of times a cancer was correctly localized (with any rating) was determined. Results Noncalcification Cancers The reader-averaged AFROC curves for the noncalcification cancers are shown in Figure 3. The reader-averaged FOMs were equal to (95% CI, ), (95% CI, ), and (95% CI, ) for the standard, low-contrast, and filmscreen image-processing types, respectively. The differences between the reader-averaged FOMs for each pair of image-processing types are given in Table 2; the 95% CIs are shown in parentheses. When the 95% CI of the difference does not include zero, this indicates that there is a significant difference between the two image-processing types. No significant difference was found between any of the three image-processing pairs (p > 0.40) when readers were treated as either a fixed or a random factor in the ANOVA analysis. Calcification Cancers The reader-averaged AFROC curves for calcification clusters are shown in Figure 4. The reader-averaged FOMs were equal to 390 AJR:203, August 2014

5 Image Processing in Digital Mammography TABLE 3: Percentage Difference in Number of False-Positives (FPs) Marked in Different Image-Processing Types Percentage Difference in Number of FPs (95% CI) Image-Processing Pair Being Compared Cancer Type All Cases Normal Cases Cases With Biopsy-Proven Benign Lesions Standard vs film-screen Mass 0.66 ( 8.64 to 7.32) 2.81 ( to 10.39) 7.84 ( 7.24 to 22.92) Calcification ( to 0.20) a ( to 0.88) a ( 6.44 to 27.25) Standard vs low contrast Mass 2.19 ( 9.99 to 5.61) 7.38 ( to 4.85) 2.72 ( to 11.95) Calcification 8.08 ( 8.40 to 24.56) 6.78 ( to 30.43) 1.51 ( to 18.61) Film-screen vs low contrast Mass 2.92 ( 5.11 to 10.95) ( 2.49 to 24.69) 4.67 ( to 9.31) Calcification 9.43 ( 8.61 to 27.56) ( 7.12 to 50.48) ( to 4.48) Note When the 95% CIs do not include zero, there is a significant difference between the image-processing pair. a There is a significant difference between the image-processing pair (95% CI, ), (95% CI, ), and (95% CI, ) for the standard, low-contrast, and film-screen image-processing types, respectively. The differences between the readeraveraged FOMs for each pair of image-processing types are given in Table 2. When readers were treated as fixed factors in the ANOVA analysis, the standard image processing was significantly better than both the film-screen image processing (p = ) and the low-contrast image processing (p = 0.04). There was no significant difference between film-screen image processing and low-contrast image processing (p = 0.15). When readers are treated as random factors in the ANOVA analysis, the difference between standard and low-contrast image processing became nonsignificant; however, the difference between standard and film-screen remained significant (p = 0.03). Additional Analysis The number of false-positive noncalcification marks was not significantly different for any of the image-processing pairs investigated (Table 3). The number of false-positive calcification clusters marks increased by 15% for film-screen image processing compared with standard image processing (p = 0.047), as shown in Table 3. When analyzing the subgroup of cases containing biopsy-proven benign lesions, we found that there was no significant difference in the number of false-positive calcification cluster marks for any image-processing pair (p > 0.16). For the subgroup of images with normal findings, the number of false-positive calcification cluster marks was significantly lower for standard image processing than for filmscreen image processing (p = 0.041). The number of times each cancer was correctly localized (mark made within outline of cancer made by expert radiologist) was calculated. The maximum number of times a cancer can be localized is 21 (seven observers reading all three image-processing types). The proportion of times the cancers were localized by all the observers in all the image-processing types and the number of times the cancer was not localized at all are given in Table 4. Discussion We investigated the effect of image processing on the detection of calcification clusters and noncalcification cancers. We found Lesion Localization Fraction Standard IP 0.2 Low-contrast IP Film-screen IP False-Positive Fraction Fig. 3 Reader-averaged alternative free-response receiver operating characteristic curves for noncalcification cancers for three different imageprocessing (IP) types. Empirical curves are shown. Solid lines = measured data, dashed lines = straight line extension to (1,1). that image processing had a significant effect on the detection of calcification clusters but not noncalcification cancers. Several studies have investigated the effect of image processing on breast cancer detection in general, including both calcifications and noncalcification cancers together [4, 6], or on one cancer type only [5, 9]. We analyzed the noncalcification cancers and calcification clusters separately to determine whether image processing had a different effect on the detection of the different radiologic features. Cole et al. [8] and Kamitani et al. [7] also analyzed TABLE 4: Percentage of Cancers Not Correctly Localized by Any of the Seven Observers in Any of the Three Image-Processing Types Investigated and Percentage of Cancers Seen by All Seven Observers in All Three Image-Processing Types Cancers Noncalcification Cancers (%) Calcification Clusters (%) Cancers not localized at all 1 16 Cancers localized every time 26 6 Lesion Localization Fraction Standard IP Low-contrast IP Film-screen IP False-Positive Fraction Fig. 4 Reader-averaged alternative free-response receiver operating characteristic curves for calcification clusters for three different imageprocessing (IP) types. Empirical curves are shown. Solid lines = measured data, dashed lines = straight line extension to (1,1). AJR:203, August

6 Warren et al. calcification and noncalcification cancers separately. However, Kamitani et al. used a smaller number of cancer cases than this study, 11 calcification cases and 34 noncalcification cases. Cole et al. also used a much smaller number of cancer cases acquired on a system with an amorphous selenium detector, eight calcification cases and 13 mass cases, compared with the current study and different observers read images with different image-processing algorithms, whereas in this work the same readers interpreted the same cases with different image-processing algorithms to allow a paired comparison. Our work found that the standard imageprocessing algorithm provided significantly higher detection of calcification clusters than the low-contrast and film-screen image-processing algorithms. The results of this study are in agreement with the work by Zanca et al. [5]; they performed a study with a design similar to the present work and found significant differences in calcification detection between several pairs of image-processing algorithms. However, Warren et al. [9], who also performed a study of similar design, found no significant differences in detection between two different image-processing algorithms. Given that the designs of our study and these two studies were similar, the differences in results may be because of differences in the image-processing algorithms investigated. Our work also found that there were no significant differences in detection of noncalcification cancers between the different types of image processing. This result is in agreement with a study by Cole et al. [8], which found no significant difference in the area under the receiver operating curve for masses. However, that study was limited to small numbers of cases, as described earlier. In this study, we considered marks made by observers on biopsy-proven benign lesions as false-positives. This approach has been used elsewhere [23] and assumes the ground truth is based on biopsy-proven cancers. If, instead, we consider that biopsyproven benign lesions are not false-positives because a radiologist must have recalled this patient originally for a biopsy to be performed (correct recall) and the results are reanalyzed, the same conclusions are reached. Visser et al. [6] found that image processing affected the suspiciousness of normal and malignant features in the breast. In this work we found that there was no significant difference in the number of false-positives with changes in the image processing for noncalcification cancers. However, for calcification clusters there was a significant decrease in the number of false-positives reported on normal images for standard image processing compared with film-screen image processing. Some of these images showed benign features that were not recalled for biopsy in the clinic. The change in the number of false-positives in these images indicates that benign features that did not appear suspicious enough to be recalled in the clinic can appear suspicious enough to be recalled when a different type of image processing has been applied. The biggest limitation to this work is the number of cases and readers. There was large variability among readers, as shown by the wide 95% CI for the reader-averaged FOM for each image-processing type. In addition, a proportion of the noncalcification cancers were seen by all observers in all image-processing algorithms (Table 4). These lesions may have been too obvious to detect possible differences among image-processing algorithms. A proportion of the calcifications were not seen at all (Table 4). These lesions were too subtle to be useful in the study. Therefore, in retrospect this study would have had more power if some more subtle noncalcification cancers and if some more obvious calcifications had been included in the image set and if more observers had been enrolled in the study. There is a limit to the number of readers it is practical to enroll in such studies; therefore, the power limitations of this study emphasize the need for careful case selection when preparing observer studies. In summary, for the particular image-processing algorithms and systems investigated, we found that the standard image processing was superior for the detection of calcifications and that there was no significant difference among image-processing algorithms for noncalcification lesions. The differences measured in the JAFROC FOM for calcification detection were between 4% and 6%. This difference is smaller than the change in calcification detection for other factors such as detector type or dose [9]. Finally, the results in this study are applicable for the particular image-processing algorithms and systems investigated. We recommend that objective measurements, such as the method used in this study, be applied when selecting the optimal image processing to be used by radiologists with other imageprocessing algorithms and systems. Acknowledgments We thank Carole Kliger, Caroline Taylor, Victoria Cooke, Louise Wilkinson, Wilmi Pienaar, Rupika Mehta, and Christine Flis for taking part in the study. We also thank Lindsay Mungutroy and Charul Patel for their assistance in running the observer study. Thanks are also due to Claire Borrelli, Sharon Montaque, and Sue Storr for organizing time on the workstations to complete the study. We thank David Duncan for writing the software used to insert the calcification clusters into breast images. We are grateful to the staff at King s College Hospital in particular, Asif Iqbal and Michael Michell for their assistance collecting breast images. We are also grateful for the help and cooperation of staff from the Jarvis Breast Screening Centre while collecting breast images for the study. Finally, we are grateful to Hologic staff for advice and help in processing the images. References 1. Sivaramakrishna R, Obuchowski NA, Chilcote WA, Cardenosa G, Powell KA. Comparing the performance of mammographic enhancement algorithms: a preference study. AJR 2000; 175: Pisano ED, Cole EB, Major S, et al. Radiologists preferences for digital mammographic display. Radiology 2000; 216: Goldstraw EJ, Castellano I, Ashley S, Allen S. The effect of Premium View post-processing software on digital mammographic reporting. Br J Radiol 2010; 83: Chen B, Wang W, Huang J, et al. Comparison of tissue equalization, and premium view post-processing method in full field digital mammography. Eur J Radiol 2010; 76: Zanca F, Jacobs J, Van Ongeval C, et al. Evaluation of clinical image processing algorithms used in digital mammography. Med Phys 2009; 36: Visser R, Veldkamp WJH, Beijerinck D, et al. Increase in perceived case suspiciousness due to local contrast optimisation in digital screening mammography. Eur Radiol 2012; 22: Kamitani T, Yabuuchi H, Soeda H, et al. Detection of breast cancer by soft-copy reading of digital mammograms: comparison between a routine image-processing parameter and high-contrast parameters. Acta Radiol 2010; 51: Cole EB, Pisano ED, Kistner EO, et al. Diagnostic accuracy of digital mammography in patients with dense breasts who underwent problem-solving mammography: effects of image processing and lesion type. Radiology 2003; 226: Warren LM, Mackenzie A, Cooke J, et al. Effect 392 AJR:203, August 2014

7 Image Processing in Digital Mammography of image quality on calcification detection in digi- calcifications for observer studies in digital mam- cal imaging systems: executive summary of AAPM tal mammography. Med Phys 2012; 39: mography. Phys Med Biol 2013; 58:N217 N228 TG18 report. Med Phys 2005; 32: Looney PT, Halling-Brown M, Given-Wilson R, et 15. Zanca F, Zhang G, Marshall N, et al. Software 19. Chakraborty DP, Berbaum KS. Observer studies in- al. A national digital mammography image re- framework for simulating clusters of microcalcifi- volving detection and localization: modelling, analy- source. (abstract) In: British Institute of Radiology. cations in digital mammography. In: Martí J, Oli- sis and validation. Med Phys 2004; 31: Symposium Mammographicum Conference ver A, Freixenet J, Martí R, eds. Digital mammog- 20. Chakraborty DP. A brief history of FROC para- London, UK: British Institute of Radiology, 2012: Bennett RL, Sellars SJ, Moss SM. Interval cancers in the NHS Breast Cancer Screening Programme in England, Wales and Northern Ireland. Br J Cancer 2011; 104: Ruschin M, Timberg P, Båth M, et al. Dose dependence of mass and microcalcification detection in digital mammography: free response human observer studies. Med Phys 2007; 34: Warren LM, Mackenzie A, Dance DR, Young KC. Comparison of the x-ray attenuation properties of breast calcifications, aluminium, hydroxyapatite and calcium oxalate. Phys Med Biol 2013; 58:N103 N Warren LM, Green FH, Shrestha L, Mackenzie A, Dance DR, Young KC. Validation of simulation of raphy: 10th International Workshop, IWDM 2010, Girona, Catalonia, Spain, June 16 18, 2010 proceedings. Berlin, Germany: Springer- Verlag, 2010; 6136: Liston J, Wilson R, eds. Quality assurance guidelines for breast cancer screening radiology, 2nd ed. Sheffield, England: NHS Breast Screening Programmes, 2011: publication no National Electrical Manufacturers Association. Digital imaging and communications in medicine (DICOM). Part 14. Grayscale standard display function. Rosslyn, VA: National Electrical Manufacturers Association, 2000: publication no. PS Samei E, Banano A, Chakraborty D, et al.; AAPM TG18. Assessment of display performance for medi- digm data analysis. Acad Radiol 2013; 20: Dorfman DD, Berbaum KS, Metz CE. Receiver operating characteristic rating analysis: generalization to the population of readers and patients with jackknife method. Invest Radiol 1992; 27: Hillis SL, Berbaum KS, Metz CE. Recent developments in the Dorfman-Berbaum-Metz procedure for multireader ROC study analysis. Acad Radiol 2008; 15: Rafferty EA, Park JM, Philpotts LE, et al. Assessing radiologist performance using combined digital mammography and breast tomosynthesis compared with digital mammography alone: results of a multicenter, multireader trial. Radiology 2013; 266: AJR:203, August

Since its introduction in 2000, digital mammography has become

Since its introduction in 2000, digital mammography has become Review Article Smith A, PhD email : Andrew.smith@hologic.com Since its introduction in 2000, digital mammography has become an accepted standard of care in breast cancer screening and has paved the way

More information

The value of the craniocaudal mammographic view in breast cancer detection: A preliminary study

The value of the craniocaudal mammographic view in breast cancer detection: A preliminary study The value of the craniocaudal mammographic view in breast cancer detection: A preliminary study P D Trieu 1, Prof. P C Brennan 1, Dr. W Lee 2, Dr. E Ryan 1, Dr. W Reed 1, Dr. M Pietrzyk 1 1. Medical Image

More information

A comparison of the accuracy of film-screen mammography, full-field digital mammography, and digital breast tomosynthesis

A comparison of the accuracy of film-screen mammography, full-field digital mammography, and digital breast tomosynthesis Clinical Radiology xxx (2012) 1e6 Contents lists available at SciVerse ScienceDirect Clinical Radiology journal homepage: www.clinicalradiologyonline.net A comparison of the accuracy of film-screen mammography,

More information

Title: Breast Screen Reader Assessment Strategy (BREAST): UK Collaboration

Title: Breast Screen Reader Assessment Strategy (BREAST): UK Collaboration Susan Williams CoRIPS Research Grant 147 5000 awarded Title: Breast Screen Reader Assessment Strategy (BREAST): UK Collaboration Principle Aim The principal aim of the study is to evaluate the performance

More information

FDA Executive Summary

FDA Executive Summary Meeting of the Radiological Devices Advisory Panel On October 24, 22, the panel will discuss, make recommendations, and vote on a premarket approval application supplement (P83/S) to expand the indications

More information

Improving Reading Time of Digital Breast Tomosynthesis with Concurrent Computer Aided Detection

Improving Reading Time of Digital Breast Tomosynthesis with Concurrent Computer Aided Detection White Paper Improving Reading Time of Digital Breast Tomosynthesis with Concurrent Computer Aided Detection WHITE PAPER 2 3 Abstract PowerLook Tomo Detection, a concurrent computer-aided detection (CAD)

More information

Detection and Classification of Calcifications on Digital Breast Tomosynthesis and 2D Digital Mammography: A Comparison

Detection and Classification of Calcifications on Digital Breast Tomosynthesis and 2D Digital Mammography: A Comparison Women s Imaging Original Research Spangler et al. Digital Breast Tomosynthesis Versus 2D Digital Mammography Women s Imaging Original Research FOCUS ON: M. Lee Spangler 1 Margarita L. Zuley 2 Jules H.

More information

arxiv: v2 [cs.cv] 8 Mar 2018

arxiv: v2 [cs.cv] 8 Mar 2018 Automated soft tissue lesion detection and segmentation in digital mammography using a u-net deep learning network Timothy de Moor a, Alejandro Rodriguez-Ruiz a, Albert Gubern Mérida a, Ritse Mann a, and

More information

Women s Imaging Original Research

Women s Imaging Original Research Women s Imaging Original Research Women s Imaging Original Research David Gur 1 Andriy I. Bandos 2 Howard E. Rockette 2 Margarita L. Zuley 3 Jules H. Sumkin 3 Denise M. Chough 3 Christiane M. Hakim 3 Gur

More information

Financial Disclosures

Financial Disclosures Financial Disclosures 3D Mammography: The Latest Developments in the Breast Imaging Arena I have no financial disclosures Dr. Katharine Lampen-Sachar Breast and Body Radiologist Radiology Associates of

More information

Simulation of spiculated breast lesions

Simulation of spiculated breast lesions Simulation of spiculated breast lesions Premkumar Elangovan* a, Faisal Alrehily b, R Ferrari Pinto a, Alaleh Rashidnasab a,c, David R Dance b,d, Kenneth C Young b,d, Kevin Wells a a Centre for Vision,

More information

Digital Breast Tomosynthesis in the Diagnostic Environment: A Subjective Side-by-Side Review

Digital Breast Tomosynthesis in the Diagnostic Environment: A Subjective Side-by-Side Review Women s Imaging Original Research Hakim et al. Digital Breast Tomosynthesis Women s Imaging Original Research Christiane M. Hakim 1 Denise M. Chough 1 Marie A. Ganott 1 Jules H. Sumkin 1 Margarita L. Zuley

More information

Women s Imaging Original Research

Women s Imaging Original Research Women s Imaging Original Research Brandt et al. DBT for Screening Recalls Without Calcifications Women s Imaging Original Research FOCUS ON: Kathleen R. Brandt 1 Daniel A. Craig 1 Tanya L. Hoskins 2 Tara

More information

Mammography limitations. Clinical performance of digital breast tomosynthesis compared to digital mammography: blinded multi-reader study

Mammography limitations. Clinical performance of digital breast tomosynthesis compared to digital mammography: blinded multi-reader study Clinical performance of digital breast tomosynthesis compared to digital mammography: blinded multi-reader study G. Gennaro (1), A. Toledano (2), E. Baldan (1), E. Bezzon (1), C. di Maggio (1), M. La Grassa

More information

Proven clinical effectiveness at low radiation dose

Proven clinical effectiveness at low radiation dose MicroDose Mammography Solutions Proven clinical effectiveness at low radiation dose Several studies provide evidence that Philips MicroDose Mammography* can provide outstanding image quality at 18% to

More information

Digital Breast Tomosynthesis from a first idea to clinical routine

Digital Breast Tomosynthesis from a first idea to clinical routine International Master Programm Biomedical Engineering Digital Breast Tomosynthesis from a first idea to clinical routine Historical background 2D imaging of 3D objects has important limitations Jörg Barkhausen

More information

Correlation between lesion type and the additional value of digital breast tomosynthesis

Correlation between lesion type and the additional value of digital breast tomosynthesis Correlation between lesion type and the additional value of digital breast tomosynthesis Poster No.: C-1604 Congress: ECR 2011 Type: Scientific Exhibit Authors: C. Van Ongeval, L. Cockmartin, A. Van Steen,

More information

Updates in Mammography. Dr. Yang Faridah A. Aziz Department of Biomedical Imaging University Malaya Medical Centre

Updates in Mammography. Dr. Yang Faridah A. Aziz Department of Biomedical Imaging University Malaya Medical Centre Updates in Mammography Dr. Yang Faridah A. Aziz Department of Biomedical Imaging University Malaya Medical Centre Updates in Mammography Breast Imaging Dr. Yang Faridah A. Aziz Department of Biomedical

More information

Mammography. What is Mammography? What are some common uses of the procedure?

Mammography. What is Mammography? What are some common uses of the procedure? Mammography What is Mammography? Mammography is a specific type of imaging that uses a low-dose x-ray system to examine breasts. A mammography exam, called a mammogram, is used to aid in the early detection

More information

Investigating the performance of a CAD x scheme for mammography in specific BIRADS categories

Investigating the performance of a CAD x scheme for mammography in specific BIRADS categories Investigating the performance of a CAD x scheme for mammography in specific BIRADS categories Andreadis I., Nikita K. Department of Electrical and Computer Engineering National Technical University of

More information

Features of Prospectively Overlooked Computer-Aided Detection Marks on Prior Screening Digital Mammograms in Women With Breast Cancer

Features of Prospectively Overlooked Computer-Aided Detection Marks on Prior Screening Digital Mammograms in Women With Breast Cancer Women s Imaging Original Research Women s Imaging Original Research WOMEN S IMAGING Nariya Cho 1 Seung Ja Kim Hye Young Choi Chae Yeon Lyou Woo Kyung Moon Cho N, Kim SJ, Choi HY, Lyou CY, Moon WK Keywords:

More information

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School Breast Cancer Screening Early detection of

More information

Digital Breast Tomosynthesis

Digital Breast Tomosynthesis Digital Breast Tomosynthesis Policy Number: Original Effective Date: MM.05.012 06/28/2013 Line(s) of Business: Current Effective Date: HMO; PPO; QUEST 06/28/2013 Section: Radiology Place(s) of Service:

More information

Contrast Enhanced Spectral Mammography (CESM) Updates

Contrast Enhanced Spectral Mammography (CESM) Updates Contrast Enhanced Spectral Mammography (CESM) Updates Georgeta Mihai, PhD, DABR Medical Physicist, BIDMC, Boston Assistant Professor, Harvard Medical School, Boston Disclosures None Acknowledgments: Da

More information

The Radiology Aspects

The Radiology Aspects REQUIREMENTS FOR INTERNATIONAL ACCREDITATION OF BREAST CENTERS/UNITS The Radiology Aspects Miri Sklair-Levy, Israel RADIOLOGY GUIDELINES FOR QUALITY ASSURANCE IN BREAST CANCER SCREENING AND DIAGNOSIS Radiologists

More information

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY

EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY EARLY DETECTION: MAMMOGRAPHY AND SONOGRAPHY Elizabeth A. Rafferty, M.D. Avon Comprehensive Breast Center Massachusetts General Hospital Harvard Medical School Breast Cancer Screening Early detection of

More information

The effect of digitising film prior mammograms on radiologists' performance in breast screening: a JAFROC study

The effect of digitising film prior mammograms on radiologists' performance in breast screening: a JAFROC study Loughborough University Institutional Repository The effect of digitising film prior mammograms on radiologists' performance in breast screening: a JAFROC study This item was submitted to Loughborough

More information

Tomosynthesis and breast imaging update. Dr Michael J Michell Consultant Radiologist King's College Hospital NHS Foundation Trust

Tomosynthesis and breast imaging update. Dr Michael J Michell Consultant Radiologist King's College Hospital NHS Foundation Trust Tomosynthesis and breast imaging update Dr Michael J Michell Consultant Radiologist King's College Hospital NHS Foundation Trust Breast imaging new technology BREAST CANCER FLT PET shows different grades

More information

Breast tomosynthesis reduces radiologist performance variability compared to digital mammography

Breast tomosynthesis reduces radiologist performance variability compared to digital mammography Breast tomosynthesis reduces radiologist performance variability compared to digital mammography Andrew Smith 1, Elizabeth Rafferty 2, Loren Niklason 1 1 Hologic, Inc., Bedford MA, USA 2 Massachusetts

More information

CHAPTER 2 MAMMOGRAMS AND COMPUTER AIDED DETECTION

CHAPTER 2 MAMMOGRAMS AND COMPUTER AIDED DETECTION 9 CHAPTER 2 MAMMOGRAMS AND COMPUTER AIDED DETECTION 2.1 INTRODUCTION This chapter provides an introduction to mammogram and a description of the computer aided detection methods of mammography. This discussion

More information

Diagnostic Dilemmas of Breast Imaging

Diagnostic Dilemmas of Breast Imaging Diagnostic Dilemmas of Breast Imaging Common Causes of Error in Breast Cancer Detection By: Jason Cord, M.D. Mammography: Initial Imaging The standard for detection of breast cancer Screening mammography

More information

BI-RADS classification in breast tomosynthesis. Our experience in breast cancer cases categorized as BI-RADS 0 in digital mammography

BI-RADS classification in breast tomosynthesis. Our experience in breast cancer cases categorized as BI-RADS 0 in digital mammography BI-RADS classification in breast tomosynthesis. Our experience in breast cancer cases categorized as BI-RADS 0 in digital mammography Poster No.: C-0562 Congress: ECR 2017 Type: Scientific Exhibit Authors:

More information

Imaging in breast cancer. Mammography and Ultrasound Donya Farrokh.MD Radiologist Mashhad University of Medical Since

Imaging in breast cancer. Mammography and Ultrasound Donya Farrokh.MD Radiologist Mashhad University of Medical Since Imaging in breast cancer Mammography and Ultrasound Donya Farrokh.MD Radiologist Mashhad University of Medical Since A mammogram report is a key component of the breast cancer diagnostic process. A mammogram

More information

Session 4: Test instruments to assess interpretive performance challenges and opportunities Overview of Test Set Design and Use

Session 4: Test instruments to assess interpretive performance challenges and opportunities Overview of Test Set Design and Use Session 4: Test instruments to assess interpretive performance challenges and opportunities Overview of Test Set Design and Use Robert A. Smith, PhD American Cancer Society Test Sets vs. Audits Benefits

More information

Ana Sofia Preto 19/06/2013

Ana Sofia Preto 19/06/2013 Ana Sofia Preto 19/06/2013 Understanding the underlying pathophysiologic processes leading to the various types of calcifications Description and illustration of the several types of calcifications, according

More information

Advice Statement 005/18 April 2018

Advice Statement 005/18 April 2018 Advice Statement 005/18 April 2018 In asymptomatic women attending for breast screening, what is the clinical and cost effectiveness of digital breast tomosynthesis (DBT) in addition to full-field digital

More information

AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS

AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS AN ALGORITHM FOR EARLY BREAST CANCER DETECTION IN MAMMOGRAMS Isaac N. Bankman', William A. Christens-Barryl, Irving N. Weinberg2, Dong W. Kim3, Ralph D. Semmell, and William R. Brody2 The Johns Hopkins

More information

Epworth Healthcare Benign Breast Disease Symposium. Sat Nov 12 th 2016

Epworth Healthcare Benign Breast Disease Symposium. Sat Nov 12 th 2016 Epworth Healthcare Benign Breast Disease Symposium Breast cancer is common Sat Nov 12 th 2016 Benign breast disease is commoner, and anxiety about breast disease commoner still Breast Care Campaign UK

More information

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD

Update of Digital Breast Tomosynthesis. Susan Orel Roth, MD Update of Digital Breast Tomosynthesis Susan Orel Roth, MD NCI estimates that : Why DBT? Approximately 20% of breast cancers are missed at mammography screening Average recall rates approximately 10%

More information

TOMOSYNTHESIS. Daniela Bernardi. U.O. Senologia Clinica e Screening mammografico APSS Trento, Italy

TOMOSYNTHESIS. Daniela Bernardi. U.O. Senologia Clinica e Screening mammografico APSS Trento, Italy TOMOSYNTHESIS Daniela Bernardi U.O. Senologia Clinica e Screening mammografico APSS Trento, Italy BACKGROUND early detection through screening MAMMOGRAPHY is associated with reduced breast cancer morbidity

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy File Name: Origination: Last CAP Review: Next CAP Review: Last Review: digital_breast_tomosynthesis 3/2011 6/2016 6/2017 11/2016 Description of Procedure or Service Conventional

More information

Min Jung Kim Department of Medicine The Graduate School, Yonsei University

Min Jung Kim Department of Medicine The Graduate School, Yonsei University Zoomed image of contact mammography versus magnification mammography in the diagnosis of microcalcifications with soft-copy full field digital mammography Min Jung Kim Department of Medicine The Graduate

More information

Digital Breast Tomosynthesis

Digital Breast Tomosynthesis Applies to all products administered or underwritten by Blue Cross and Blue Shield of Louisiana and its subsidiary, HMO Louisiana, Inc.(collectively referred to as the Company ), unless otherwise provided

More information

Diagnostic benefits of ultrasound-guided. CNB) versus mammograph-guided biopsy for suspicious microcalcifications. without definite breast mass

Diagnostic benefits of ultrasound-guided. CNB) versus mammograph-guided biopsy for suspicious microcalcifications. without definite breast mass Volume 118 No. 19 2018, 531-543 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu Diagnostic benefits of ultrasound-guided biopsy versus mammography-guided

More information

What s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia

What s New in Breast Imaging. Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia What s New in Breast Imaging Jennifer A. Harvey, M.D., FACR Professor of Radiology University of Virginia Disclosure Hologic, Inc. Shareholder and research agreement. Volpara Solutions, Ltd. Shareholder

More information

Name of Policy: Computer-aided Detection (CAD) Mammography

Name of Policy: Computer-aided Detection (CAD) Mammography Name of Policy: Computer-aided Detection (CAD) Mammography Policy #: 112 Latest Review Date: October 2010 Category: Radiology Policy Grade: Active Policy but no longer scheduled for regular literature

More information

Jing Zhang, PhD, Lars J. Grimm, MD, MHS, Joseph Y. Lo, PhD, Karen S. Johnson, MD,

Jing Zhang, PhD, Lars J. Grimm, MD, MHS, Joseph Y. Lo, PhD, Karen S. Johnson, MD, This manuscript has been accepted for publication in Journal of the American College of Radiology. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published

More information

Breast screening: understanding case difficulty and the nature of errors

Breast screening: understanding case difficulty and the nature of errors Loughborough University Institutional Repository Breast screening: understanding case difficulty and the nature of errors This item was submitted to Loughborough University's Institutional Repository by

More information

Computer aided detection of clusters of microcalcifications on full field digital mammograms

Computer aided detection of clusters of microcalcifications on full field digital mammograms Computer aided detection of clusters of microcalcifications on full field digital mammograms Jun Ge, a Berkman Sahiner, Lubomir M. Hadjiiski, Heang-Ping Chan, Jun Wei, Mark A. Helvie, and Chuan Zhou Department

More information

Computer-aided Detection in the United Kingdom National Breast Screening Programme: Prospective Study 1

Computer-aided Detection in the United Kingdom National Breast Screening Programme: Prospective Study 1 Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Lisanne A. L. Khoo,

More information

Fundamentals of Breast Tomosynthesis

Fundamentals of Breast Tomosynthesis Fundamentals of Breast Tomosynthesis Improving the Performance of Mammography Andrew Smith, Ph.D. This white paper is one in a series of research overviws on advanced technologies in women s healthcare.

More information

Amammography report is a key component of the breast

Amammography report is a key component of the breast Review Article Writing a Mammography Report Amammography report is a key component of the breast cancer diagnostic process. Although mammographic findings were not clearly differentiated between benign

More information

Computer Aided Detection of Abnormalities in Mammograms

Computer Aided Detection of Abnormalities in Mammograms Computer Aided Detection of Abnormalities in Mammograms I W Hutt 1, S M Astley 1 & C R M Boggis 2 1: Wolfson Image Analysis Unit, Dept of Medical Biophysics, Manchester University, Oxford. Rd, Manchester,

More information

Dense Breasts, Get Educated

Dense Breasts, Get Educated Dense Breasts, Get Educated What are Dense Breasts? The normal appearances to breasts, both visually and on mammography, varies greatly. On mammography, one of the important ways breasts differ is breast

More information

MEDICAL POLICY SUBJECT: MAMMOGRAPHY: COMPUTER- AIDED DETECTION (CAD) POLICY NUMBER: CATEGORY: Technology Assessment

MEDICAL POLICY SUBJECT: MAMMOGRAPHY: COMPUTER- AIDED DETECTION (CAD) POLICY NUMBER: CATEGORY: Technology Assessment MEDICAL POLICY SUBJECT: MAMMOGRAPHY: COMPUTER- PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including

More information

Detection of suspicious lesion based on Multiresolution Analysis using windowing and adaptive thresholding method.

Detection of suspicious lesion based on Multiresolution Analysis using windowing and adaptive thresholding method. Detection of suspicious lesion based on Multiresolution Analysis using windowing and adaptive thresholding method. Ms. N. S. Pande Assistant Professor, Department of Computer Science and Engineering,MGM

More information

ACCREDITATION DOCUMENT THE RADIOGRAPHER

ACCREDITATION DOCUMENT THE RADIOGRAPHER ACCREDITATION DOCUMENT THE RADIOGRAPHER Nijmegen, October 2012 1. Introduction An optimal quality of mammography is one of the fundamental requirements for successful breast cancer screening programmes.

More information

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis =

#46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? TERMS (AKA) WHAT IS TOMOSYNTHESIS? 3/3/2014. Digital breast tomosynthesis = #46: DIGITAL TOMOSYNTHESIS: What is the Data Really Showing? January K. Lopez, MD Hoag Breast Care Center Newport Beach, CA Disclosures: None TERMS (AKA) Digital breast tomosynthesis = DBT Tomo 3D Full

More information

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program

Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program Impact of digitalization of mammographic units on average glandular doses in the Flemish Breast Cancer Screening Program An De Hauwere, Hubert Thierens Ghent University, Department of Medical Physics,

More information

S. Murgo, MD. Chr St-Joseph, Mons Erasme Hospital, Brussels

S. Murgo, MD. Chr St-Joseph, Mons Erasme Hospital, Brussels S. Murgo, MD Chr St-Joseph, Mons Erasme Hospital, Brussels? Introduction Mammography reports are sometimes ambiguous and indecisive. ACR has developped the BIRADS. BIRADS consists of a lexicon in order

More information

Breast Tomosynthesis

Breast Tomosynthesis Breast Tomosynthesis The Use of Breast Tomosynthesis in a Clinical Setting 2 What s Inside Introduction... 1 Initial Hologic Clinical Trial Purpose and Methodology... 1 Clinical Trial Results... 2 Improved

More information

Breast Tomosynthesis

Breast Tomosynthesis Breast Tomosynthesis Considerations for Routine Clinical Use Breast Tomosynthesis Considerations for routine clinical use Andrew Smith, Ph.D., Vice President - Imaging Sciences, Hologic This white paper

More information

Mammography. What is Mammography?

Mammography. What is Mammography? Scan for mobile link. Mammography Mammography is a specific type of breast imaging that uses low-dose x-rays to detect cancer early before women experience symptoms when it is most treatable. Tell your

More information

Policy Library Clinical Advantages of Digital Breast Tomosynthesis in Symptomatic Patients

Policy Library Clinical Advantages of Digital Breast Tomosynthesis in Symptomatic Patients Policy Library Clinical Advantages of Digital Breast Tomosynthesis in Symptomatic Patients Version: 1 Approved by: Faculty of Clinical Radiology Council Date of approval: Click and type: day month and

More information

Outline. Digital Breast Tomosynthesis: Update and Pearls for Implementation. Tomosynthesis Dataset: 2D/3D (Hologic Combo Acquisition)

Outline. Digital Breast Tomosynthesis: Update and Pearls for Implementation. Tomosynthesis Dataset: 2D/3D (Hologic Combo Acquisition) Outline Digital Breast Tomosynthesis (DBT) the new standard of care Digital Breast Tomosynthesis: Update and Pearls for Implementation Emily F. Conant, M.D. Professor, Chief of Breast Imaging Department

More information

Assessment of extent of disease: digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM)

Assessment of extent of disease: digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM) Assessment of extent of disease: digital breast tomosynthesis (DBT) versus full-field digital mammography (FFDM) Poster No.: C-1237 Congress: ECR 2012 Type: Scientific Paper Authors: N. Seo 1, H. H. Kim

More information

Digital breast tomosynthesis (DBT) occult breast cancers: clinical, radiological and histopathological features.

Digital breast tomosynthesis (DBT) occult breast cancers: clinical, radiological and histopathological features. Digital breast tomosynthesis (DBT) occult breast cancers: clinical, radiological and histopathological features. Poster No.: C-1707 Congress: ECR 2015 Type: Scientific Exhibit Authors: V. Vinci 1, A. Iqbal

More information

Mammographic imaging of nonpalpable breast lesions. Malai Muttarak, MD Department of Radiology Chiang Mai University Chiang Mai, Thailand

Mammographic imaging of nonpalpable breast lesions. Malai Muttarak, MD Department of Radiology Chiang Mai University Chiang Mai, Thailand Mammographic imaging of nonpalpable breast lesions Malai Muttarak, MD Department of Radiology Chiang Mai University Chiang Mai, Thailand Introduction Contents Mammographic signs of nonpalpable breast cancer

More information

One or Two Clusters of Crushed Stone like Calcifications on the Mammogram Produced by Malignancy

One or Two Clusters of Crushed Stone like Calcifications on the Mammogram Produced by Malignancy 66 One or Two Clusters of Crushed Stone like Calcifications on the Mammogram Produced by Malignancy Example 2.13 A 36-year-old woman who recentlyfelt a small hard lump in the upper-outer quadrant of her

More information

Computer-Aided Diagnosis for Microcalcifications in Mammograms

Computer-Aided Diagnosis for Microcalcifications in Mammograms Computer-Aided Diagnosis for Microcalcifications in Mammograms Werapon Chiracharit Department of Electronic and Telecommunication Engineering King Mongkut s University of Technology Thonburi BIE 690, November

More information

Breast Tomosynthesis

Breast Tomosynthesis Breast Tomosynthesis The Use of Breast Tomosynthesis in a Clinical Setting 2 What s Inside Introduction... 1 Initial Hologic Clinical Trial Purpose and Methodology... 1 Clinical Trial Results... 2 Improved

More information

Breast Tomosynthesis An additional screening tool in the fight against breast cancer

Breast Tomosynthesis An additional screening tool in the fight against breast cancer What to Expect Breast Tomosynthesis An additional screening tool in the fight against breast cancer Every woman over 40 should be examined for breast cancer once a year. American Cancer Society What to

More information

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company

WHAT TO EXPECT. Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC. The Women's Health Company WHAT TO EXPECT Breast Tomosynthesis An additional screening tool in the fight against breast cancer HOLOGIC The Women's Health Company ...,. Screening for breast cancer Doctors and scientists agree that

More information

the one name in cancer care.

the one name in cancer care. the one name in cancer care. Landmark study evaluating close to half a million mammography exams published in the Journal of the American Medical Association (JAMA) 1 Hologic 3D Mammography Significantly

More information

3D Mammography. The most exciting advancement in mammography in over 30 years

3D Mammography. The most exciting advancement in mammography in over 30 years What to Expect 3D Mammography The most exciting advancement in mammography in over 30 years Screening for breast cancer Doctors and scientists agree that early detection is the best defense against breast

More information

Mammographic evaluation of palpable breast masses with pathological correlation: a tertiary care centre study in Nepal

Mammographic evaluation of palpable breast masses with pathological correlation: a tertiary care centre study in Nepal Original article 21 Mammographic evaluation of palpable breast masses with pathological correlation: a tertiary care centre study in Nepal G. Gurung, R. K. Ghimire, B. Lohani Department of Radiology and

More information

BI-RADS Categorization As a Predictor of Malignancy 1

BI-RADS Categorization As a Predictor of Malignancy 1 Susan G. Orel, MD Nicole Kay, BA Carol Reynolds, MD Daniel C. Sullivan, MD BI-RADS Categorization As a Predictor of Malignancy 1 Index terms: Breast, biopsy, 00.1261 Breast neoplasms, localization, 00.125,

More information

Clinical Comparison of Full-Field Digital Mammography and Screen- Film Mammography for Detection of Breast Cancer

Clinical Comparison of Full-Field Digital Mammography and Screen- Film Mammography for Detection of Breast Cancer John M. Lewin 1 Carl J. D Orsi 2 R. Edward Hendrick 1,3 Lawrence J. Moss 2 Pamela K. Isaacs 1 ndrew Karellas 2 Gary R. Cutter 4 Received July 6, 2001; accepted after revision February 19, 2002. Supported

More information

Session 2: The Role of Specialist Radiology Technologists

Session 2: The Role of Specialist Radiology Technologists Session 2: The Role of Specialist Radiology Technologists Louise M. Henderson, MSPH PhD Assistant Professor, Department of Radiology The University of North Carolina, Chapel Hill Overview Role of the technologist

More information

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS

CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS ABBY DUROJAYE,M.D CURRENTLY FDA APPROVED ARE FULL FIELD DIGITAL MAMMOGRAPHY SYSTEMS AND FILM SCREEN STILL BEING USED AT SOME INSTITUTIONS BOTH HAVE BEEN SHOWN TO BE EFFECTIVE TOOLS EARLY DETECTION OF BREAST

More information

Department of Radiology, University of Cambridge, Cambridge, UK 2. Aberdeen Biomedical Imaging Centre, University of Aberdeen, Aberdeen, UK 3

Department of Radiology, University of Cambridge, Cambridge, UK 2. Aberdeen Biomedical Imaging Centre, University of Aberdeen, Aberdeen, UK 3 The TOMMY trial: a comparison of TOMosynthesis with digital MammographY in the UK NHS Breast Screening Programme a multicentre retrospective reading study comparing the diagnostic performance of digital

More information

Performance Evaluation of Image Processing Algorithms in Digital Mammography

Performance Evaluation of Image Processing Algorithms in Digital Mammography Performance Evaluation of Image Processing Algorithms in Digital Mammography Federica Zanca* a, Chantal Van Ongeval a, Jurgen Jacobs a, Herman Pauwels a, Guy Marchal a, Hilde Bosmans a a Leuven University

More information

Breast cancer detection: radiologists performance using mammography with and without automated whole-breast ultrasound

Breast cancer detection: radiologists performance using mammography with and without automated whole-breast ultrasound Eur Radiol (2010) 20: 2557 2564 DOI 10.1007/s00330-010-1844-1 BREAST Kevin M. Kelly Judy Dean Sung-Jae Lee W. Scott Comulada Breast cancer detection: radiologists performance using mammography with and

More information

Women s Imaging Original Research

Women s Imaging Original Research Women s Imaging Original Research Waldherr et al. One-View Breast Tomosynthesis Versus Two-View Mammography Women s Imaging Original Research Christian Waldherr 1 Peter Cerny 1 Hans J. Altermatt 2 Gilles

More information

Mammographic Breast Density Classification by a Deep Learning Approach

Mammographic Breast Density Classification by a Deep Learning Approach Mammographic Breast Density Classification by a Deep Learning Approach Aly Mohamed, PhD Robert Nishikawa, PhD Wendie A. Berg, MD, PhD David Gur, ScD Shandong Wu, PhD Department of Radiology, University

More information

Electrical impedance scanning of the breast is considered investigational and is not covered.

Electrical impedance scanning of the breast is considered investigational and is not covered. ARBenefits Approval: 09/28/2011 Effective Date: 01/01/2012 Revision Date: Code(s): Medical Policy Title: Electrical Impedance Scanning of the Breast Document: ARB0127 Administered by: Public Statement:

More information

Digital breast tomosynthesis

Digital breast tomosynthesis GE Healthcare Digital breast tomosynthesis Daniel B. Kopans, M.D., F.A.C.R. Professor of Radiology Harvard Medical School Senior Radiologist - Breast Imaging Division Massachusetts General Hospital Since

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Digital Breast Tomosynthesis Page 1 of 31 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Digital Breast Tomosynthesis Professional Institutional Original Effective

More information

Mammography is a most effective imaging modality in early breast cancer detection. The radiographs are searched for signs of abnormality by expert

Mammography is a most effective imaging modality in early breast cancer detection. The radiographs are searched for signs of abnormality by expert Abstract Methodologies for early detection of breast cancer still remain an open problem in the Research community. Breast cancer continues to be a significant problem in the contemporary world. Nearly

More information

Chief Investigator Adele Francis University of Birmingham UK. Prof MWR Reed (CoI) University of Sheffield

Chief Investigator Adele Francis University of Birmingham UK. Prof MWR Reed (CoI) University of Sheffield The LORIS Trial: A multicentre, randomised phase III trial of standard surgery versus active monitoring in women with newly diagnosed low risk ductal carcinoma in situ. Chief Investigator Adele Francis

More information

Observer Evaluations of Wavelet Methods for the Enhancement and Compression of Digitized Mammograms

Observer Evaluations of Wavelet Methods for the Enhancement and Compression of Digitized Mammograms Observer Evaluations of Wavelet Methods for the Enhancement and Compression of Digitized Mammograms Maria Kallergi 1, John J. Heine 1, and Bradley J. Lucier 2 1 H. Lee Moffitt Cancer Center & Research

More information

Contrast-Enhanced Spectral Mammography

Contrast-Enhanced Spectral Mammography Contrast-Enhanced Spectral Mammography Illuminating Breast Cancer Detection SenoBright HD TM gehealthcare.com/senobright Mammography is the most reliable imaging technique for breasts, but limitations

More information

Retrospective study comparison of Arcadia Lab s Galileo and Parascript AccuDetect Galileo CAD systems in screening environment.

Retrospective study comparison of Arcadia Lab s Galileo and Parascript AccuDetect Galileo CAD systems in screening environment. Retrospective study comparison of Arcadia Lab s Galileo and Parascript AccuDetect Galileo CAD systems in screening environment July 2011 Parascript Page 1 7/8/2011 Investigators Dr. Rossano Girometti (Researcher),

More information

Purpose. [Purpose] [Introduction] Architectural distortion on mammography

Purpose. [Purpose] [Introduction] Architectural distortion on mammography Improved detection of architectural distortion in digital mammography using distortion-weighted image (DiWI): a new mathematical image-filtering technique by the pinwheelframelet processing method Poster

More information

Mammography and Subsequent Whole-Breast Sonography of Nonpalpable Breast Cancers: The Importance of Radiologic Breast Density

Mammography and Subsequent Whole-Breast Sonography of Nonpalpable Breast Cancers: The Importance of Radiologic Breast Density Isabelle Leconte 1 Chantal Feger 1 Christine Galant 2 Martine Berlière 3 Bruno Vande Berg 1 William D Hoore 4 Baudouin Maldague 1 Received July 11, 2002; accepted after revision October 28, 2002. 1 Department

More information

Retrospective Study Comparison of Arcadia Lab s Galileo and Parascript AccuDetect Galileo CAD systems in screening environment.

Retrospective Study Comparison of Arcadia Lab s Galileo and Parascript AccuDetect Galileo CAD systems in screening environment. White Paper Retrospective Study Comparison of Arcadia Lab s Galileo and Parascript AccuDetect Galileo CAD systems in screening environment Takeaway AccuDetect Galileo performs significantly better than

More information

Breast asymmetries in mammography: Management

Breast asymmetries in mammography: Management Breast asymmetries in mammography: Management Poster No.: C-1026 Congress: ECR 2015 Type: Educational Exhibit Authors: V. de Lara Bendahan 1, F. J. Hidalgo Ramos 2, J. L. Ortega Garcia 3, Keywords: DOI:

More information

BreastScreen Victoria tomosynthesis screening trial, Maroondah pilot: Preliminary Outcomes

BreastScreen Victoria tomosynthesis screening trial, Maroondah pilot: Preliminary Outcomes BreastScreen Victoria tomosynthesis screening trial, Maroondah pilot: Preliminary Outcomes Dr Darren Lockie, Director and Chief Radiologist, Maroondah BreastScreen, Eastern Health Ms Michelle Clemson,

More information

Purpose: Materials and Methods: Results: Conclusion: Original Research n Breast Imaging

Purpose: Materials and Methods: Results: Conclusion: Original Research n Breast Imaging Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Original Research

More information

Introduction 1. Executive Summary 5

Introduction 1. Executive Summary 5 Roman_pages 20-09-2005 21:01 Pagina IX Table of contents Introduction 1 Executive Summary 5 1. Epidemiological guidelines for quality assurance in breast cancer screening 15 1.10 Introduction 17 1.20 Local

More information