Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy

Size: px
Start display at page:

Download "Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy"

Transcription

1 Assessment of radiation dose to the chest wall and lung of the patients with breast cancer under electron beam therapy MB. Tavakoli 1, M. Saeb 2, H. Emami 3 1 MB Tavakoli, Department of Medical Physics and Medical Engineering, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran. mbtavakoli@mui.ac.ir 2 M. Saeb, Department of Medical Physics and Medical Engineering, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran 3 H. Emami, Department of Radiation Oncology, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran 0

2 Abstract Radiotherapy treatment for patients who have undergone breast saving operation is basically tangential radiation using cobalt or other high-energy photons. In case of patients who experience modified radical mastectomy (MRM), electron beam therapy is the method of choice. Rapid fall off of the percentage depth dose (DD%), homogeneous dose distribution and limited dose to the sub tumoral tissues are the outstanding features of electron therapy. In the case of MRM, chest wall and auxiliary regions should be treated with electron beams. The treatment field includes chest wall, auxiliary, sub and supra clavicular lymph nodes as well as internal mammary nodes. The aim of this work was to study the depth dose of 10 and 13MeV electron beams in the chest wall, lymph nodes and lung when the chest wall was radiated anteriorly. In this study two high-energy linear accelerators (Neptone 10 and Saturn 20) were used as the sources of electron beams. Measurements of the DD% were performed in a Perspex phantom using TLD (TLD-100) dosimetry method. The phantom was designed based on the chest wall CT images of mastectomized patients. The measured DD% to the internal mammary nodes, auxiliary nodes and chest wall were 97.5%, 96% and 98%, respectively when the electron energy was 13 MeV and they were 90.5%, 77% and 99.7% when 10 MeV electrons beams were used. The anterior lung paranchimal DD% was 78% with 13 MeV electron while the posterior part of the lung received 47% of the maximum dose. With 10 MeV electron beams, the values for anterior and posterior parts of the lung were 83% and 45%, respectively. Using the 13 MeV electron beams, internal mammary and auxiliary lymph nodes as well as chest wall received adequate doses, while lungs were excessively exposed to radiation. With 10 MeV electron radiation therapy, internal mammary nodes and chest wall were well exposed to the radiation, yet auxiliary nodes did not receive enough radiation, hence auxiliary lymph nodes should additionally receive radiation from posterior field as well. Keywords: Electron Beam Therapy; Percentage Depth Dose, Chest Wall, Radiation Dose 1

3 1. Introduction Breast cancer is one of the most common cancers in women. The prevalence age of this cancer is usually between 45 to 75. It has also more incident rate in Jewish and black people. It is less common in Asian and African women. Its treatment of choice is usually surgery or combination of surgery and radiotherapy, chemotherapy and hormone therapy. Radiotherapy is a modality used after surgery to remove probable remained unremoved cancerous cells and also lymph nodes around the excision area such as auxiliary, sternal and sub clavicular nodes. The best radiation therapy modality is assumed to be electron irradiation of the chest wall and also the lymph nodes. Compaired with photon therapy the main advantage of electron therapy is the fast fall off of the depth dose of the electron beam, which results in saving normal underlying tissues especially in lungs. To treat the chest wall and the lymph nodes while saving the lung, it is necessary to have a depth dose of 80% to 90% at the border of the chest wall and the lungs. The choice of electron energy depends on the chest wall thickness and also depth of the lymph nodes. The electron energies of 10 to 13MeV is usually used. The chest wall is usually treated using a single field of electron and in order to compensate the dose to the auxiliary anodes, extra posterior cobalt field is used. The purpose of this work was to compare the measured dose to the target volume and normal tissue with the prescribed dose. 2. Materials and Methods Electron sources used in this research were a Saturn 20 and Nepton 10 Linear accelerators. A phantom of chest wall (without breast) was constructed using Perspex blocks (Perspex is usually easily accessible, hard enough to perform the task, can be cut in different necessary thicknesses and also it is nearly equivalent to the soft tissue). The material used for lung phantom was cork. It is also a material equivalent to lung tissue both in density and atomic number, and hence electron density. The material used for bone phantom was Teflon, which has the properties of bone materials. To obtain the necessary sizes for phantom construction, preliminary measurements of 10 patients referred to the radiotherapy department of Said Al-Shohada hospital were performed. Contour sizes of the patients were obtained for a mean patient size with 95% confidence limits. To measure the percentage depth dose, LiF (Tl) TLD dosimeter was used. In the phantom suitable holes were made at critical locations of the chest wall such as sternal nodes, auxiliary nodes, the border of chest wall and lung and also at different depth in the lung for TLD positioning. To measure the dose at different locations, 47 TLD chips were used. The TLDs were first prepared and calibrated according to the manufacture recommendations. The treatment field was drawn on the phantom chest wall as a rectangle with 26x25cm dimensions by radiotherapist. The irradiation technique was SSD and the applied dose was 200cGy. After positioning the TLDs at the predetermind locations of the phantom and positioning the phantom on the treatment coach of the linacs and adjusting the radiation field, a dose of 200 cgy was applied to the phantom. The above procedure was repeated 6 times for both 10 and 13 MeV electron beams of the two linacs. After each irradiation of the phantom, TLDs were removed and the recorded doses were read using the Solaro 2A TLD reader. 2

4 3. Results 1-The measured percentage depth doses at different locations with 13MeV electron beam are shown in table 1 to 3. The percentage depth dose at different depths in lung is also shown in graph 1 (SSD=100cm, Field size=26x15cm and chest wall thickness was 2.5cm). 2-The measured percentage depth doses at different locations with 10 MeV electron beam are shown in tables 4 to 6. The percentage depth dose at different depths in lung is also shown in graph 2 (SSD=100cm, field size-26x15 and chest wall thickness=2.5cm). 4. Discussion The aim of this work was to compare the prescribed dose with the measured dose in critical points of the chest wall and the surrounding organs. When the 13MeV electron beam from Saturn 20 linear accelerator was used, the results were as follows: Table 1 shows that the dose at the internal mammary lymph nodes were 195cGy while the prescribed dose was 200 cgy. The difference is 2.5%, which is acceptable (the acceptable error is 5%). Table 2 shows that the measured dose in auxiliary lymph nodes was 192 cgy, which is 4% less than the prescribed dose. In this case, the nodes received the acceptable dose. According to table 3, the measured dose at the chest wall was 186cGy compared with the prescribed dose (190 cgy), the difference is 2.1%, which is acceptable. In graph 1, percentage depth doses at different depths using 13MeV electron are shown. It shows that the dose in the lung was relatively high. The dose in the superior level of the lung was 158 cgy, which is 79% of the prescribed dose. This can cause problems for the safety of the lung. It is also be seen that the fall off of the dose to the lung is much more gradual than water or soft tissue, which means entire lung, received a high dose. When 10 MeV electrons from Nepton Linear accelerator was used the results were as follows: Table table 4 shows that, although a dose of 200cGy was prescribed for 3 cm depth but measured dose at this depth was 181cGy, which is 9.5% less than the necessary dose. This is higher than 5% (permitted errors), and hence should be compensated. Table 5 shows that measured the dose at the auxiliary lymph nodes (129cGy) was mush less than the prescribed dose (167cGy), with a difference of about 29%. The necessary treatment dose should be applied to the lymph nodes using an extra boost field. In some centers this is done using a posterior photon field. According to table 6, the difference between measured dose and prescribed dose in the chest wall was 0.2% (209cGy and 208.5cGy), which is in the acceptable error range, and hence the chest wall at the depth of 3 cm receives an acceptable treatment dose. Graph 2 shows the percentage depth dose at different depths of the lung when 10MeV electron was used. The absorbed dose at the border of the lung and chest wall was 162cGy, which was 81% of the maximum dose and decreases to 45cGy at the posterior border of the lung, which is 23% of the tumor dose. Therefore the dose to the lung is sufficiently low to be harmful. From these results it is suggested that when 10MeV electron is used for the treatment of the chest wall, the dose to the sternal nodes and auxiliary lymph nodes should be compensated. 3

5 5. References 1-Carlos A, Perez: Principles and Practice of Radiation Oncology. 3 rd Edition, Williams and Wilkins Vincent T, Devita ET, et al: Practice of Oncology. 5 th Edition (2001). 3-Feletecher GH. Introduction. Tapley N, Ed. Clinical Aplications of the Electron Beam. New York, Wiley, Faiz M Kan, The Physics of Radiation Therapy, 2 nd Ed. Electron Beam Therapy, IAEA. Absorbed Dose determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards dose to Water , (1999) 6-Choi MC, Purdy BJ, Abrat FG, Glasgow GP. Variation in output factor Caused by Secondary for 7-16 MeV Electron Beams. Med. Phys. 37(6).(1979). 4

6 Table 1: Measured doses in the phantom at the sternal nodes position after application of 200 cgy radiation dose from electron beam with 13MeV energy Mean 195 Relative to prescribed 97.5% SD 3.54% Prescribed dose 200 Deference between prescribed and Measured dose 2.5% Table 2: Measured doses in the phantom at the auxiliary nodes position after application of 200 cgy radiation dose from electron beam with 13MeV energy Mean 192 Relative to prescribed 96% SD 4.4% Prescribed dose 200 Deference between prescribed and Measured dose -4% Table 3: Measured doses in the phantom at the chest wall after application of 200 cgy radiation dose from electron beam with 13MeV energy Mean 186 Relative to prescribed 98% SD 4.56% Prescribed dose 190 Deference between prescribed and Measured dose -2.1% 5

7 Table 4: Measured doses in the phantom at the sternal nodes position after application of 200 cgy radiation dose from electron beam with 10MeV energy Mean 181 Relative to prescribed 90.5% SD 4.93% Prescribed dose 200 Deference between prescribed and Measured dose -9.5% Table 5: Measured doses in the phantom at the auxiliary nodes position after application of 200 cgy radiation dose from electron beam with 10MeV energy Mean 129 Relative to prescribed 77% SD 5% Prescribed dose 167 Deference between prescribed and Measured dose -29% Table 6: Measured doses in the phantom at the chest wall after application of 200 cgy radiation dose from electron beam with 10MeV energy Mean Relative to prescribed 99.7% SD 2.66% Prescribed dose 209 Deference between prescribed and Measured dose -0.2% 6

8 %DD Depth from skin (cm) Figure 1: Percentage depth dose in the lungs of the phantom at different depths when irradiation dose at the skin was 200cGy using 13MeV electron beams سوپ حطس زا هير قمع) cm ( ش يريگ هزادنا زد دصرد )cgy( %DD Depth from skin (cm) Figure 2: Percentage depth dose in the lungs of the phantom at different depths when irradiation dose at the skin was 200cGy using 10MeV electron beam. 7

Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas

Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 2, SPRING 2006 Extracranial doses in stereotactic and conventional radiotherapy for pituitary adenomas Thomas Samuel Ram, a Paul B. Ravindran,

More information

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging

Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging Iran. J. Radiat. Res., 2008; 6 (2): 83-87 Verification of the PAGAT polymer gel dosimeter by photon beams using magnetic resonance imaging B. Azadbakht 1, M.H. Zahmatkesh 2 *, k. Hadad 1, S. Bagheri 2

More information

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital

Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital Iran. J. Radiat. Res., 2004; 2 (2): 53-58 Assessment of variation of wedge factor with depth, field size and SSD for Neptun 10PC Linac in Mashhad Imam Reza Hospital M. Hajizadeh Saffar 1*, M.R. Ghavamnasiri

More information

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy

Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy Turkish Journal of Cancer Volume 37, No. 2, 2007 59 Comparison of high and low energy treatment plans by evaluating the dose on the surrounding normal structures in conventional radiotherapy MUHAMMAD BASIM

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

9.5. CONVENTIONAL RADIOTHERAPY TECHNIQUE FOR TREATING THYROID CANCER

9.5. CONVENTIONAL RADIOTHERAPY TECHNIQUE FOR TREATING THYROID CANCER 9.5. CONVENTIONAL RADIOTHERAPY TECHNIQUE FOR TREATING THYROID CANCER ROBERT J. AMDUR, MD, SIYONG KIM, PhD, JONATHAN GANG LI, PhD, CHIRAY LIU, PhD, WILLIAM M. MENDENHALL, MD, AND ERNEST L. MAZZAFERRI, MD,

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom

Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom Iran. J. Radiat. Res., 2003; 1(3): 143-149 Application of MCNP4C Monte Carlo code in radiation dosimetry in heterogeneous phantom A. Mostaar 1, M. Allahverdi 1,2, M. Shahriari 3 1 Medical Physics Department,

More information

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules

Spinal Cord Doses in Palliative Lung Radiotherapy Schedules Journal of the Egyptian Nat. Cancer Inst., Vol. 8, No., June: -, 00 Spinal Cord Doses in Palliative Lung Radiotherapy Schedules HODA AL-BOOZ, FRCR FFRRCSI M.D.* and CAROL PARTON, Ph.D.** The Departments

More information

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS

DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National

More information

Radiation Doses to Contralateral Breast During Irradiation of Breast Cancer

Radiation Doses to Contralateral Breast During Irradiation of Breast Cancer Med. J. Cairo Univ., Vol. 77, No. 2, June: 27-31, 2009 www.medicaljournalofcairouniversity.com Radiation Doses to Contralateral Breast During Irradiation of Breast Cancer ADEL G. WAHBA, M.D.* and REHAM

More information

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department

More information

Estimation of the Thyroid Secondary Cancer Risk on the Patient of Standard Breast External Beam Radiotherapy

Estimation of the Thyroid Secondary Cancer Risk on the Patient of Standard Breast External Beam Radiotherapy Original Article Estimation of the Thyroid Secondary Cancer Risk on the Patient of Standard Breast External Beam Radiotherapy Abstract Background: The purpose of this study was to estimate the secondary

More information

7/16/2009. Cost-benefit, QALYs and the Value of a Statistical Life

7/16/2009. Cost-benefit, QALYs and the Value of a Statistical Life 7/16/29 Cost-benefit, QALYs and the Value of a Statistical Life Alan McKenzie, Bristol The benefits of many developments in radiotherapy may be obvious for instance, replacing hand planning by computer

More information

Multilayer Gafchromic film detectors for breast skin dose determination in vivo

Multilayer Gafchromic film detectors for breast skin dose determination in vivo INSTITUTE OF PHYSICSPUBLISHING Phys. Med. Biol. 47 (2002) N31 N37 PHYSICS INMEDICINE AND BIOLOGY PII: S0031-9155(02)27324-8 NOTE Multilayer Gafchromic film detectors for breast skin dose determination

More information

Protection of the contralateral breast during radiation therapy for breast cancer

Protection of the contralateral breast during radiation therapy for breast cancer Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY

LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY LONG TERM STABILITY OF THE PERFORMANCE OF A CLINICAL LINEAR ACCELERATOR AND Z-SCORE ASSESSEMENT FOR ABSORBED DOSE TO WATER QUANTITY E. BADITA 1, C.VANCEA 1,3, I. CALINA 1,3, D. STROE 2, M. DUMITRACHE 2,3,

More information

Mahdi Aghili*, Parisa Seifi, Farshid Farhan, Ahmad Reza Sebzari, Ehsan Mohamadi, Vahid Vaezzadeh a

Mahdi Aghili*, Parisa Seifi, Farshid Farhan, Ahmad Reza Sebzari, Ehsan Mohamadi, Vahid Vaezzadeh a Original Article Open Access DOI: 10.19187/abc.20163114-18 Assessment of Dose Delivery to Supraclavicular and Axillary Lymph Nodes in Adjuvant Breast Cancer Radiotherapy, with or without Posterior Axillary

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters 133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2.

Stefan Gutiérrez Lores 1, Gonzalo Walwyn Salas 1, Daniel Molina Pérez 1, Raudel Campa Menéndez 2. Experiences of the Secondary Standard Dosimetry Laboratory of the CPHR in the implementation and validation of the Postal Quality Audit in Radiotherapy, using TLDs Stefan Gutiérrez Lores 1, Gonzalo Walwyn

More information

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS

MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging

Comparison of Energy Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams using Magnetic Resonance Imaging Research Journal of Applied Sciences, Engineering and Technology (): 6-, ISSN: -767 Maxwell Scientific Organizational, Submitted: November 9, Accepted: November 9, Published: February, Comparison of Energy

More information

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry

Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara 1, Z Kuncic 1, J Adamovics 2 and C Baldock 1,3 1 Institute of Medical Physics, School of Physics, University

More information

3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast

3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast 1 Angela Kempen February Case Study February 22, 2012 3D Conformal Radiation Therapy for Mucinous Carcinoma of the Breast History of Present Illness: JE is a 45 year-old Caucasian female who underwent

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System Original Article Research in Oncology 2018; Vol. 14, No. 1: 34-38. DOI: 10.21608/resoncol.2018.3026.1047 Verification of performance of Acuros XB Algorithm (AXB) Implemented on Eclipse Planning System

More information

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter E.K. Chie M.D. 1, S.W. Park M.D. 1, W.S. Kang PhD. 1,2, I.H. Kim M.D. 1,2, S.W. Ha M.D. 1,2,

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research)

International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research) International Association of Scientific Innovation and Research (IASIR) (An Association Unifying the Sciences, Engineering, and Applied Research) International Journal of Emerging Technologies in Computational

More information

Electron therapy Class 3: Clinical procedures

Electron therapy Class 3: Clinical procedures Electron therapy Class 3: Clinical procedures Laurence Court lecourt@mdanderson.org Reference: Faiz M. Khan, The Physics of Radiation Therapy Slide acknowledgements: Karl Prado, Rebecca Howell, Kent Gifford,

More information

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Additional Questions for Review 2D & 3D

Additional Questions for Review 2D & 3D Additional Questions for Review 2D & 3D 1. For a 4-field box technique, which of the following will deliver the lowest dose to the femoral heads? a. 100 SSD, equal dmax dose to all fields b. 100 SSD, equal

More information

Mohammad Houshyari 1, Amir ShahramYousefi Kashi 1, Sakineh Soleimani Varaki 2, Afshin Rakhsha 1, Eftekhar Rajab Blookat 3

Mohammad Houshyari 1, Amir ShahramYousefi Kashi 1, Sakineh Soleimani Varaki 2, Afshin Rakhsha 1, Eftekhar Rajab Blookat 3 Electronic Physician (ISSN: 2008-5842) http://www.ephysician.ir Year: 2015, Volume: 7, Issue: 2, Pages: 1032-1038, DOI: 10.14661/2015.1032-1038 Regional lymph node radiotherapy in breast cancer: single

More information

Sarcoma and Radiation Therapy. Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington

Sarcoma and Radiation Therapy. Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington Sarcoma and Radiation Therapy Gabrielle M Kane MB BCh EdD FRCPC Muir Professorship in Radiation Oncology University of Washington Objective: Helping you make informed decisions Introduction Process Radiation

More information

Skin Reaction in Radiation Therapy for Breast Cancer

Skin Reaction in Radiation Therapy for Breast Cancer Iranian Journal of Medical Physics Vol. 11, No. 4, Autumn 2014, 316-321 Received: April 15, 2014; Accepted: July 7, 2014 Original Article Skin Reaction in Radiation Therapy for Breast Cancer Bagher Farhood

More information

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique

Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and Photon Beams Using MRI Technique Australian Journal of Basic and Applied Sciences, 5(11): 879-886, 2011 ISSN 1991-8178 Comparison of Percentage Depth Dose (PDD) and Post Time Dependence of PAGAT Polymer Gel Dosimeter with Electron and

More information

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals

Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals Iran. J. Radiat. Res., 2006; 4 (1): 29-33 Entrance surface dose measurements for routine X-ray examinations in Chaharmahal and Bakhtiari hospitals D. Shahbazi-Gahrouei * Department of Medical Physics and

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Patient dosimetry for total body irradiation using single-use MOSFET detectors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,

More information

Radiation Damage Comparison between Intensity Modulated Radiotherapy (IMRT) and Field-in-field (FIF) Technique In Breast Cancer Treatments

Radiation Damage Comparison between Intensity Modulated Radiotherapy (IMRT) and Field-in-field (FIF) Technique In Breast Cancer Treatments Radiation Damage Comparison between Intensity Modulated Radiotherapy () and Field-in-field (FIF) Technique In Breast Cancer Treatments Huisi Ai 1 and Hualin Zhang 2 1. Department of Radiation Oncology,

More information

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.

Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser. Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic

More information

Left Chest Wall and Supraclavicular Irradiation Using Photon and Electron Energies

Left Chest Wall and Supraclavicular Irradiation Using Photon and Electron Energies 1 Louise Francis March Case Study March 29, 2012 Left Chest Wall and Supraclavicular Irradiation Using Photon and Electron Energies History of Present Illness: RC is 46 year-old Pakistani woman who presented

More information

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS

EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS PAPER EQUIVALENT DOSE FROM SECONDARY NEUTRONS AND SCATTER PHOTONS IN ADVANCE RADIATION THERAPY TECHNIQUES WITH 15 MV PHOTON BEAMS Isra Israngkul Na Ayuthaya *, Sivalee Suriyapee, Phongpheath Pengvanich

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data J. A. Bencomo, N. Hernandez, and W.F. Hanson University of Texas M.D. Anderson Cancer Center Medical physics,

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-861.Volume 6, Issue Ver. II (May-Jun. 201), PP 2- A Dosimetric study of different MLC expansion aperture For the radiotherapy of pancreas cancer

More information

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study

Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Production and dosimetry of simultaneous therapeutic photons and electrons beam by linear accelerator: a monte carlo study Navid Khledi 1, Azim Arbabi 2, Dariush Sardari 1, Mohammad Mohammadi 3, Ahmad

More information

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities JBUON 2018; 23(1): 268-272 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com SHORT COMMUNICATION First application of hemi-body electron beam irradiation for Kaposi

More information

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest

4 Essentials of CK Physics 8/2/2012. SRS using the CyberKnife. Disclaimer/Conflict of Interest SRS using the CyberKnife Sonja Dieterich, PhD, DABR Associate Professor University of California Davis Disclaimer/Conflict of Interest Consulting agreements with Broncus Medical and CyberHeart, Inc. Scientific

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

3D Printed Phantoms for MRT Dosimetry

3D Printed Phantoms for MRT Dosimetry 3D Printed Phantoms for MRT Dosimetry Dr. Andrew Robinson Nuclear Physics Group, The University of Manchester Christie Medical Physics and Engineering, The Christie NHS Foundation Trust BIR/IDUG: The Future

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

IORT with mobile linacs: the Italian experience

IORT with mobile linacs: the Italian experience IORT with mobile linacs: the Italian experience G. Tosi, M. Ciocca lntroduction At the beginning of 1999 a mobile Linac (Novac 7, manufactured by Hitesys, Aprilia, Italy) able to produce electron beams

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

Breast radiotherapy in women with pectus excavatum (funnel chest): is the lateral decubitus technique an answer? A dosimetric study

Breast radiotherapy in women with pectus excavatum (funnel chest): is the lateral decubitus technique an answer? A dosimetric study The British Journal of Radiology, 79 (2006), 785 790 SHORT COMMUNICATION Breast radiotherapy in women with pectus excavatum (funnel chest): is the lateral decubitus technique an answer? A dosimetric study

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of

More information

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY

EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH A COMBINATION OF TLD ALBEDO AND SULPHUR ACTIVATION TECHNIQUES FOR FAST NEUTRON PERSONNEL DOSIMETRY : ~ : ; EUROPEAN ORGANZATON FOR NUCLEAR RESEARCH 20 1983 TS-RP/ 113/ CF TS DVSONAL REPORT September 1983 A COMBNATON OF TLD ALBEDO AND SULPHUR ACTVATON TECHNQUES FOR FAST NEUTRON PERSONNEL DOSMETRY J.W.N.

More information

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT

Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted

More information

International Practice Accreditation

International Practice Accreditation International Practice Accreditation Program endorsed and supported by Jean-Xavier Hallet Punta Cana, Nov 2017 Short History Created in 2004 by ESTRO, independent since 2009 Dosimetry Laboratory (IGR):

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

[Setawati et. al., Vol.4 (Iss10): October, 2017] ISSN: DOI: /zenodo

[Setawati et. al., Vol.4 (Iss10): October, 2017] ISSN: DOI: /zenodo EXTERNAL RADIATION SIMULATION OF LINAC TO DETERMINE EFFECTIVE DOSE IN ORGANS USING MONTE CARLO METHOD Evi Setawati *1, Muchammad Azam 1, Ngurah Ayu Ketut Umiati 1, Hammam Oktajianto 1 *1 Physics Department,

More information

Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy

Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Rasha Makkia Biomedical Physics Ph.D. student East Carolina University March 3 rd, 2016 Outlines The Purpose Wilms Tumor,

More information

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction 1 The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Megan E. Sullivan, B.S., R.T.(T)., Patrick A. Melby, B.S. Ashley Hunzeker, M.S., CMD, Nishele Lenards, M.S., CMD, R.T. (R)(T),

More information

Neutron Measurements for Intensity Modulated Radiation Therapy

Neutron Measurements for Intensity Modulated Radiation Therapy SLAC-PUB-8443 April 2 Neutron Measurements for Intensity Modulated Radiation Therapy N. E. Ipe et al. Presented at Chicago 2 World Congress on Medical Physics and Biomedical Engineering, 7/23/2 7/28/2,

More information

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution

Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution NUKLEONIKA 2007;52(4):153 158 ORIGINAL PAPER Effect of scattered radiation in the total body irradiation technique: evaluation of the spoiler and wall dose component in the depthdose distribution Tomasz

More information

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison

More information

The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI)

The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI) The use of the TLD-100 for quality assurance in Total Body Irradiation (TBI) ARNIE VERDE NOLASCO Dept. de Engenharia Nuclear/UFMG Belo Horizonte - MG BRAZIL arnienolasco@hotmail.com LUIZ OLIVEIRA FARIA

More information

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output

Presented by: Rebecca M. Howell PhD, DABR, FAAPM. Results from Voluntary Independent External Peer Review of Beam Output Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output What do we hope to achieve with independent peer review? Independent peer review

More information

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry

Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Absorbed Dose Response in Water of Kilovoltage X-rays Beams of Radiochromic Film and Thermoluminescent for Brachytherapy Dosimetry Chien-Hau Chu 1, Uei-Tyng Lin 1, Ngot-Swan Chong 2, Wen-Song Hwang 1,

More information

Protocol of Radiotherapy for Breast Cancer

Protocol of Radiotherapy for Breast Cancer 107 年 12 月修訂 Protocol of Radiotherapy for Breast Cancer Indication of radiotherapy Indications for Post-Mastectomy Radiotherapy (1) Axillary lymph node 4 positive (2) Axillary lymph node 1-3 positive:

More information

Contribution of CT to Quantitative Radiation Therapy Planning

Contribution of CT to Quantitative Radiation Therapy Planning 123 Satish C. Prasad1 Miljenko V. Pilepich Carlos A. Perez Received January 1 6. 1 98; accepted after revision August 1 1, 198. All authors: Division of Radiation Oncology, Mallinckrodt Institute of Radiology,

More information

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system

Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system Journal of Physics: Conference Series PAPER OPEN ACCESS Inhomogeneity effect in Varian Trilogy Clinac ix 10 MV photon beam using EGSnrc and Geant4 code system To cite this article: S Yani et al 2016 J.

More information

Anthropomorphic breast phantoms for quality assurance and dose verification

Anthropomorphic breast phantoms for quality assurance and dose verification JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 5, NUMBER 1, WINTER 2004 Anthropomorphic breast phantoms for quality assurance and dose verification 1 2 3 4 José A. Bencomo, 5 Connel Chu, Victor M.

More information

The Effects of DIBH on Liver Dose during Right-Breast Treatments Introduction

The Effects of DIBH on Liver Dose during Right-Breast Treatments Introduction 1 The Effects of DIBH on Liver Dose during Right-Breast Treatments Megan E. Sullivan B.S.R.T.(T)., Patrick A. Melby, B.S. Ashley Hunzeker, M.S., CMD, Nishele Lenards, M.S., CMD Medical Dosimetry Program

More information

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch

More information

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES *

EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * Romanian Reports in Physics, Vol. 66, No. 1, P. 157 165, 2014 EXPERIMENTAL RESULTS IN PERCENTAGE DEPTH DOSE (PDD) DETERMINATION AT THE EXTENDED DISTANCES * M. SPUNEI 1,2, M. MIHAI 3, I. MĂLĂESCU 1 1 West

More information

Optimization of three dimensional planning dosimetric in breast phantom for match region of supraclavicular and tangential fields

Optimization of three dimensional planning dosimetric in breast phantom for match region of supraclavicular and tangential fields Original Article Optimization of three dimensional planning dosimetric in breast phantom for match region of supraclavicular and tangential fields ABSTRACT Aim: Complex geometry of breast tissue causes

More information

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Male Breast Cancer-Dosimetry Evaluation in Teletherapy

Male Breast Cancer-Dosimetry Evaluation in Teletherapy International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2014, 3, 241-251 Published Online November 2014 in SciRes. http://www.scirp.org/journal/ijmpcero http://dx.doi.org/10.4236/ijmpcero.2014.34031

More information

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD)

Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) Journal of Physics: Conference Series PAPER OPEN ACCESS Measurement of computed tomography dose profile with pitch variation using Gafchromic XR-QA2 and thermoluminescence dosimeter (TLD) To cite this

More information

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor

More information

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers

Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Comparison of microlion liquid filled ionisation chamber with Semiflex and Pinpoint air filled chambers Rhys Carrington, Ysbyty Glan Clwyd, Bodelwyddan. MSc Medical Radiation Physics. Small field dosimetry

More information

THERMOLUMINESCENCE RESPONSE OF YTTERBIUM-DOPED AND UNDOPED OF SILICON OPTICAL FIBER SUBJECTED TO PHOTON IRRADIATION

THERMOLUMINESCENCE RESPONSE OF YTTERBIUM-DOPED AND UNDOPED OF SILICON OPTICAL FIBER SUBJECTED TO PHOTON IRRADIATION Digest Journal of Nanomaterials and Biostructures Vol. 10, No. 4, October December 2015, p. 1459-1467 THERMOLUMINESCENCE RESPONSE OF YTTERBIUM-DOPED AND UNDOPED OF SILICON OPTICAL FIBER SUBJECTED TO PHOTON

More information

Zakithi Msimang SAAPMB/SARPA 2011

Zakithi Msimang SAAPMB/SARPA 2011 THE IAEA/WHO TLD POSTAL DOSE AUDIT SERVICE Zakithi Msimang SAAPMB/SARPA 2011 Introduction The idea of organizing a dosimetry audit for radiotherapy centres by the IAEA, was suggested for the first time

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità

More information