1 Presented by: Rebecca M. Howell PhD, DABR, FAAPM Results from Voluntary Independent External Peer Review of Beam Output

2 What do we hope to achieve with independent peer review? Independent peer review is an important tool for improving quality of the clinical physics program. Enhance patient safety Ensure the best possible treatment for our patients Meet regulatory requirements

3 Independent Peer Review Around the World United States MD Anderson RDS IROC-H ADCL MDADL University of Wisconsin ADCL RCL The World IAEA/WHO International Atomic Energy Agency and World Health Organization ACDS - Australian Clinical Dosimetry Service Radiotherapy Trials Quality RTTQA - Radiotherapy Quality Assurance (UK) JCOG Japan Clinical Oncology Group EORTC - European Organization for Research and Treatment of Cancer (audits by IROC-H)

4 MD Anderson Cancer Center Section of Outreach Physics Outreach Physics IROC-H RDS ADCL MDADL

5 Radiation Dosimetry Services RDS Independent Peer Review Services Beam output for external beam RT: megavoltage photon (2-25 MV) electron (4-20 MeV) orthovoltage (1.9 mm Al - 3 mm Cu HVL) Cobalt-60 Dose verification for special procedures: total body dosimetry total skin dosimetry Absorbed Dose in Blood Irradiators Dose Verification Cell Irradiations (in development)

6 Imaging and Radiation Oncology Core Houston IROC-H Independent Peer Review Services (for clinical trial participants) Beam output for external beam RT: megavoltage photon (2-25 MV) electron (4-20 MeV) Cobalt-60 Single beam small field checks Advanced technology end-to-end testing: imaging, planning, delivery Head and neck Prostate Lung (motion phantom) SRS Spine SBRT Liver (motion and multi-target) Also available for fee through MDADL Site Visits on site review of dosimetry TPS versus IROC physicist measurements of basic dosimetry parameters (PDD, OF, WF, OAF) Mechanical evaluations (linac, IGRT, MLC) Programmatic reviews (QA program, staffing, ) Virtual Site Visits (analogous to site visit without measurements)

7 Accredited Dosimetry Calibration Laboratory ADCL AAPM Accredited Calibrations Offered Electrometers Charge and Current Scales External Beam Ionization Chambers: Co-60 Absorbed Dose to Water Co-60 Exposure/Air Kerma X-Ray (Orthovoltage) Well-Ionization Chambers: High Dose Rate (Ir-192) Low Dose Rate (Pd-103, I-125, Cs-131, Cs-137 and Ir-192) Brachytherapy Sources Low Dose Rate (Pd-103, I-125, Cs-131, Cs-137 and Ir-192) See Joint AAPM/IROC Registry for current available source models

8 EBRT Monitoring RDS versus IROC Who, What, How Often Users RDS Any institution that requests service Beams Monitored MV photon and electron, orthovoltage, blood irradiations, specials Frequency of Service Annual, biannual, quarterly, monthly, on request Users IROC Clinical trial participants Beams monitored All photon and 3 electron energies (alternating years), small field Frequency of Service Annual Note: Some institutions use RDS to supplement electron monitoring or to increase frequency of monitoring.

9 History of RDS In 1968, the Texas Regional Medical Physicists (TRMP) was formed by Dr. Shalek, to provide independent peer review. In 1974, the NCI funded the six Centers for Radiological Physics (CRP); TRMP became the Southern CRP to provide QA audit services. In 1985, when the NCI terminated the CRP contracts, Drs. Shalek and Stovall created a new for-fee service, the RDS. Dr. Stovall served as the Director of the RDS for 30 years, (65 years at MDA since 1951) I was appointed Associate Director of RDS in 2015 and promoted to Director in 2016.

10 RDS Users Since 2001 Since 2001, RDS has monitored 21,000 unique EBRT beams at > 2,000 institutions in the United States and 150 institutions

11 Current Users Frequency of Service Frequency of Service Number of Institutions Annual 1005 Quarterly 92 Monthly 12 Bi-annual 19 Semi-annual 345 By-request 799 Other 43 Total 2315 Most institutions monitor beam output on annual basis. Of the 799 institutions set-up for by-request services, 150 placed at least one order in the past year. Note: Total number of distinct institutions is 2014, 10% of institutions use different monitoring frequency for different machines/beam types.

12 Dosimeters TLD-100 (LiF:Mg,Ti) powder from Quantaflux LLC (San Jose, CA) Fully characterize each new batch Batch is named for year in which it is purchased (Currently using B14) Do not reuse Controls are irradiated weekly on 60 Co unit at the MD Anderson ADCL Full review of control irradiations every quarter.

13 Phantoms Acrylic mini-phantoms (3 TLD photon and 6TLD electron) Dimensions based on the type of beam and the energy. Electrons: sufficient for full phantom scatter. Photons: the dimensions are sufficient for electronic scatter equilibrium, but not photon scatter equilibrium. Why mini-phantoms? shipping $$ But. requires BSF to determine output in full phantom.

14 What about the Uncertainty in the TLD Dose? TLD system is calibrated based on the signal-to-dose conversion established with reference TLD irradiated in 60 Co (ADCL), with reference dose of 300 cgy. Uncertainty in the dose determination is 1.3% for TLD. This accuracy allows for a tolerance of ±5% to be used. Alvarez, Kry, Stingo, and Followill. TLD and OSLD dosimetry systems for remote audits of radiotherapy external beam calibration. Rad Meas. In press

15 Quality Assurance TLD Reading Process Measure the mass of every single TLD on high accuracy digital scales. There are 3 TLD for each measurement. SD of reading/mass is reviewed during QC Tolerance is 3%, but is typically < 1.5% Read standards during every session, which includes 24 sets of 3. Read 9 standards per session: 3 at beginning, 3 at middle, 3 at the end.

16 Quality Assurance TLD Dose Calculation Detailed review of results for each session. Any inconsistencies are investigated e.g., Institution X, has historically calibrated in muscle, but in water was reported for the current set of TLD Check the documentation from physicist physicist to confirm Modify calculation as appropriate

17 Quality Assurance TLD Dose Calculation Detailed review of results for each beam.

18 Comparison with historical results. Quality Assurance Review Historical Results

19 Results Confidentiality is a key component of participation. All results are reported in aggregate. Individual examples are anonymized.

20 Summary Statistics All EBRT Beams RDS monitored 155,237 EBRT beams Number of Reports 45,000 40,000 35,000 30,000 25,000 20,000 15,000 10,000 Average 0.999± % 85.5% 64.1% were within ± 3% 2% 1% 11.8% 20.3% 26.0% 17.8% 9.6% < 6% of beams monitored by RDS are outside of ± 3%. 5, % % 0.6% % % % % % 1.05 Dose Ratio (Measured/Expected) 0.2% % %

21 Dose Ratio (Measured/Expected) Summary Statistics All MV Photon Beams RDS monitored 52,257 MV photon beams Number of Reports 16,000 14,000 12,000 10,000 8,000 6,000 4,000 2, Average ± % % % % % % % % % 1.02 Distributed equally between positive and negative ratios 3.7% % % % %

22 Summary Statistics All Electron Beams RDS monitored 101,399 electron beams Number of Reports 25,000 20,000 15,000 10,000 Average 0.999± % 20.2% 24.5% 16.7% 9.4% Distributed somewhat more toward ratios <1.00 5, % 0.2% 0.7% 2.4% 5.5% 4.3% 1.8% 0.7% 0.3% 0.1% 0.1% Dose Ratio (Measured/Expected)

23 Comparison with IROC Results Similar Averages 1.0 with most ± 3% Electron beams Average: ± results ( 22% compare 2014) 480 beams need repeat OSLD Photon beams Average: ± results ( 29% compare 2014) 2000 # results OSLD/Institution

24 Summary Statistics All Orthovoltage Beams RDS monitored 1319 orthovoltage beams Number of Reports % 0.90 Average 0.995± % % % % % % % % % % % % % % % % % % % % 1.10 Dose Ratio (Measured/Expected)

25 Outside of Criteria Setup/documentation Issues SAD vs SSD deliver (and vv) Water vs Muscle (and vv) Wrong field size (20 x 20 vs 10 x 10) Often happens when do electrons 1 st Wrong energy For electrons trouble shoot with PDD 100 MU or 200 MU vs 300 MU And one not to worry about, I left the others in room (out of beam)

26 Beam Type Institution's Frequency of Service Outside of Criteria Real Issues Initial Description Repeated Cobalt-60 Monthly MV photon Monthly MV photon Quarterly 1.22 Physicist found a problem with timer correction factor from most recent calibration. Physicist had ion chamber recalibrated; ADCL determined chamber required repair. Physicist found an 18% error in calibration of a newly commissioned linac MV photon 6-15 MeV electrons Annual Physicist reported incorrect solid water correction factor had been applied, resulting in a 3% calibration error. Error was corrected and 18 MV photons 6-20 MeV electrons Monthly Physicist determined that mercury barometer was malfunctioning, resulting in a 4% error in calibration of a newly commissioned linac MeV electron Quarterly 1.09 Physicist found two problems; depth of dmax was incorrect due to a change in energy, and measured depth dose data did not include a correction for chamber radius. 1.02

27 Outside of Criteria Recent Scenario.. New linac Wanted beam output verification prior to 1 st patient. Rush order Notified by of results. Energy Ratio (MDA/reported) 6 MV MV MV MeV MeV MeV MeV MeV 0.93

28 They did all the right things. Repeated measurements with different Ion chamber Electrometer Physicist Got same result that beams were all properly calibrated and delivering 1cGy/MU. Then..

29 They did all the right things. When taking down equipment, the BNC cable fell apart. Repeated measurements with new cable Results paralleled those reported by RDS. Tuned each of the beams as needed. Repeat TLD Results Energy Ratio (MDA/reported) 6 MV MV MV MeV MeV MeV MeV MeV 1.00

30 Conclusions Independent peer review of beam output is IMPORTANT because effects every patient! Only a small fraction of beam are outside of 3%. Often attributed to set-up errors, but some are actual calibration errors. Don t dismiss it, verify it. Even the best physicists can make mistakes & measurement equipment can malfunction.

31 Thank you Rebecca Howell, PhD, Director, RDS; David Followill, PhD, Director, IROC Houston QA Center; Rita Weathers (Retired 8/2016, Part-time); Debbie Tanner, Office Manager; Loretta Davis, Program Coordinator; Deborah Durdin, Research Technician II; Tera Jones, QA Dosimetrist; Ferrell Munson, Radiological Physics Technician II; Michael Sifuentes, Radiological Physics Technician I; Irene Harris, Research Dosimetrist; Jacob Palmer, Computational Scientist; David Campos, Radiological Physics Technician I; Natosha Ellis, Sr. Data Entry Operator; Ying Qiao, Computational Scientist; Susan Smith, Supervisor QA Dosimetry Services; Sharon McCallister (Retired 12/2016); Samantha Murray, Sr. Coordinator, Research Data; Kimberly Joseph, Sr. Secretary; and TaShun Thomas, Sr. Administrative Assistant. Francisco Aguirre, Medical Physicist. rds.mdanderson.org/

32 MD Anderson Dosimetry Laboratory MDADL Anthropomorphic Phantom Service (for Fee Service) End to end testing: planning and delivery Available services are: IMRT Head and Neck IMRT Prostate IMRT Spine Lung (± Motion Table) SRS Head SRS Single Beam

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

EORTC Member Facility Questionnaire

EORTC Member Facility Questionnaire Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Quality assurance in external radiotherapy

Quality assurance in external radiotherapy Quality assurance in external radiotherapy dr. Marius Laurikaitis medical physicist Oncological Hospital of Kaunas Medical University Hospital Kaunas, 2010-10-14 Acceptance and Commissioning Acceptance

More information

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff

Data Collected During Audits for Clinical Trials. July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Data Collected During Audits for Clinical Trials July 21, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff RPC Programs Assure... Constancy of basic machine calibration (TLD/OSLD Audits) Validity of treatment

More information

The clinical use of OSLD

The clinical use of OSLD The clinical use of OSLD Stephen F. Kry PhD Radiological Physics Center, Imaging and Radiation Oncology Core - Houston IROC Houston The University of Texas MD Anderson Cancer Center, Houston, TX Spring

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned

S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY

More information

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013

Level of Accuracy Practically Achievable in Radiation Therapy. David Followill and RPC staff August 6, 2013 Level of Accuracy Practically Achievable in Radiation Therapy David Followill and RPC staff August 6, 2013 Sources of Treatment Uncertainty Machine functioning Radiation dose determination Patient specific

More information

Zakithi Msimang SAAPMB/SARPA 2011

Zakithi Msimang SAAPMB/SARPA 2011 THE IAEA/WHO TLD POSTAL DOSE AUDIT SERVICE Zakithi Msimang SAAPMB/SARPA 2011 Introduction The idea of organizing a dosimetry audit for radiotherapy centres by the IAEA, was suggested for the first time

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center

An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance. David Followill IROC Houston QA Center An Independent Audit and Analysis of Small Field Dosimetry Quality Assurance David Followill IROC Houston QA Center Outline of Presentation Definition of the problem Results of audits Possible solutions

More information

INTRODUCTION. Material and Methods

INTRODUCTION. Material and Methods Standard Electron Values for Varian, Siemens and Elekta (Philips) Accelerators David S. Followill, Ramesh C. Tailor, Nadia Hernandez and William F. Hanson Department of Radiation Physics The University

More information

IMRT QUESTIONNAIRE. Address: Physicist: Research Associate: Dosimetrist: Responsible Radiation Oncologist(s)

IMRT QUESTIONNAIRE. Address: Physicist:   Research Associate:   Dosimetrist:   Responsible Radiation Oncologist(s) IMRT QUESTIONNAIRE Institution: Date: / / Address: Physicist: e-mail: Telephone: Fax: Research Associate: email: Telephone: Fax: Dosimetrist: email: Telephone: Fax: Responsible Radiation Oncologist(s)

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Activity report from JCOG physics group

Activity report from JCOG physics group 2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer

More information

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015

IROC Liver Phantom. Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 IROC Liver Phantom Guidelines for Planning and Irradiating the IROC Liver Phantom. Revised July 2015 The study groups are requests that each institution keep the phantom for no more than 2 weeks. During

More information

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data

Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data Problems and Shortcomings of the RPC Remote Monitoring Program of Institutions Dosimetry Data J. A. Bencomo, N. Hernandez, and W.F. Hanson University of Texas M.D. Anderson Cancer Center Medical physics,

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies

Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Standardization of dosimetry practices for small and large animal irradiation in radiobiological studies Abdul Kazi, Ph.D., DABR Assistant Professor, Department of Radiation Oncology University of Maryland

More information

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff

The RPC s Evaluation of Advanced Technologies. AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff The RPC s Evaluation of Advanced Technologies AAPM Refresher Course July 29, 2008 Geoffrey S. Ibbott, Ph.D. and RPC Staff 1 http://rpc.mdanderson.org Supported by: NCI grants CA10953 and CA81647, and an

More information

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA

Catharine Clark NPL, Royal Surrey County Hospital and RTTQA Catharine Clark NPL, Royal Surrey County Hospital and RTTQA Background Historically UK has been relatively independent of the rest of the world in audit NPL PSDL International intercomparison for reference

More information

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy

Topics covered 7/21/2014. Radiation Dosimetry for Proton Therapy Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to

More information

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation

8/2/2018. Disclosure. Online MR-IG-ART Dosimetry and Dose Accumulation Online MR-IG-ART Dosimetry and Dose Accumulation Deshan Yang, PhD, Associate Professor Department of Radiation Oncology, School of Medicine Washington University in Saint Louis 1 Disclosure Received research

More information

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014

Prostate Phantom. Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014 Prostate Phantom Guidelines for Planning and Treating the IMRT Prostate Phantom. Revised March 2014 GENERAL INFORMATION: Each institution may keep the phantom for a period of time no more than 2 weeks.

More information

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014

IROC Prostate Phantom. Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 IROC Prostate Phantom Guidelines for Planning and Treating the IROC IMRT Prostate Phantom. Revised March 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised MARCH 2014 The study groups are requesting that each institution keep the phantom for a period of time

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Eric E. Klein, Ph.D. Chair of TG-142

Eric E. Klein, Ph.D. Chair of TG-142 Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing

More information

SBRT Credentialing: Understanding the Process from Inquiry to Approval

SBRT Credentialing: Understanding the Process from Inquiry to Approval SBRT Credentialing: Understanding the Process from Inquiry to Approval David Followill, Ph.D. (presented by Andrea Molineu, M.S.) IROC Houston QA Center July 15, 2015 What is credentialing? Verification

More information

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects

IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects 2015 AAPM Annual Meeting Anaheim, CA Wes Culberson, PhD, DABR University of Wisconsin

More information

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 Lung Spine Phantom Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014 The study groups are requesting that each institution keep the phantom for no more than 2 week. During this

More information

Commissioning and Radiobiology of the INTRABEAM System

Commissioning and Radiobiology of the INTRABEAM System Commissioning and Radiobiology of the INTRABEAM System Susha Pillai and Junan Zhang Scheme INTRABEAM System. Physics Commissioning, QA, and Radiation Protection Radiobiology 1 Disclosure OHSU is an INTRABEAM

More information

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery

An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery An anthropomorphic head phantom with a BANG polymer gel insert for dosimetric evaluation of IMRT treatment delivery G. Ibbott a, M. Beach a, M. Maryanski b a M.D. Anderson Cancer Center, Houston, Texas,

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Who Should Know Radiation Oncology Coding?

Who Should Know Radiation Oncology Coding? Why Should We Learn Radiation Oncology Coding? Terry Wu, Ph.D. Chief Physicist Radiation Oncology Department Willis-Knighton Cancer Center Who Should Know Radiation Oncology Coding? Radiation Oncologist

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.

A. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development. Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1

More information

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY

DEPT OF RADIATION ONCOLOGY DEPT OF RADIATION ONCOLOGY Disclosures Lessons Learned from Misadministrations and Accidents in Radiation Therapy: Teletherapy Ryan Foster, Ph.D. Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX I receive

More information

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w

Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w Brachytherapy Planning and Quality Assurance w Classical implant systems and modern computerized dosimetry w Most common clinical applications w Quality assurance Classical implant systems w Manchester

More information

Brachytherapy Planning and Quality Assurance

Brachytherapy Planning and Quality Assurance Brachytherapy Planning and Quality Assurance Classical implant systems Most common clinical applications and modern dosimetry methods Quality assurance Classical implant systems Manchester (Paterson-Parker)

More information

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce

More information

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy

Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementing New Technologies for Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Implementation of radiosurgery and SBRT requires a fundamentally sound approach Errors don t blur out

More information

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 4, NUMBER 4, FALL 2003 Reference photon dosimetry data: A preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the

More information

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites

AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites AAPM Administrative Policy: Joint AAPM/IROC Houston Registry of Brachytherapy Sources Complying with AAPM Dosimetric Prerequisites A. Purpose and Rationale Many individual users contact the AAPM Brachytherapy

More information

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of:

Canadian Partnership for Quality Radiotherapy. Technical Quality Control Guidelines for Gamma Knife Radiosurgery. A guidance document on behalf of: Canadian Partnership for Quality Radiotherapy Technical Quality Control Guidelines for Gamma Knife Radiosurgery A guidance document on behalf of: Canadian Association of Radiation Oncology Canadian Organization

More information

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System

Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System Calibration of TLD700:LiF for Clinical Radiotherapy Beam Modalities & Verification of a High Dose Rate Brachytherapy Treatment Planning System James D Rijken Thesis submitted for the degree of Master of

More information

Clinical Trial Credentialing:

Clinical Trial Credentialing: IROC Mission Clinical Trial Credentialing: Provide integrated radiation oncology and diagnostic Where imaging to Start quality and control programs in support of the NCI s NCTN Network thereby Resources

More information

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015

IROC Lung Phantom 3D CRT / IMRT. Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 IROC Lung Phantom 3D CRT / IMRT Guidelines for Planning and Irradiating the IROC Lung Phantom. Revised Dec 2015 The IROC requests that each institution keep the phantom for no more than 2 weeks. During

More information

Calibration of dosimeters for small mega voltage photon fields at ARPANSA

Calibration of dosimeters for small mega voltage photon fields at ARPANSA Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection

More information

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT

Normal tissue doses from MV image-guided radiation therapy (IGRT) using orthogonal MV and MV-CBCT Received: 28 September 2017 Revised: 17 November 2017 Accepted: 28 December 2017 DOI: 10.1002/acm2.12276 RADIATION ONCOLOGY PHYSICS Normal tissue doses from MV image-guided radiation therapy (IGRT) using

More information

Effective and efficient radiotherapy dosimetry audit: Where to next?

Effective and efficient radiotherapy dosimetry audit: Where to next? Effective and efficient radiotherapy dosimetry audit: Where to next? Catharine Clark, Eduard Gershkevitsh, Milan Tomsej and Nuria Jornet National Physical Laboratory Royal Surrey County Hospital Radiotherapy

More information

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.

Disclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife

More information

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident

RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Resident s Name: RADIATION ONCOLOGY RESIDENCY PROGRAM Competency Evaluation of Resident Rotation: PHYS 703: Clinical Rotation 2 Inclusive dates of rotation: Feb. 26, 2016 Aug. 25, 2016 Director or Associate

More information

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor

Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Reference Photon Dosimetry Data for the Siemens Primus Linear Accelerator: Preliminary Results for Depth Dose and Output Factor Sang Hyun Cho and Geoffrey S. Ibbott Department of Radiation Physics The

More information

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program?

Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Are Transmission Detectors a Necessary Tool for a Safe Patient Radiation Therapy Program? Can you currently use direct documentation and measurement of delivered dose to ensure you did not underdose or

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction

AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction AAPM Task Group 180 Image Guidance Doses Delivered During Radiotherapy: Quantification, Management, and Reduction Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Fall

More information

TG-191 The clinical use of luminescent dosimeters

TG-191 The clinical use of luminescent dosimeters TG-191 The clinical use of luminescent dosimeters Task Group Members Stephen F. Kry, chair MD Anderson Cancer Center Paola Alvarez, MD Anderson Cancer Center Joanna Cygler, Ottawa Hospital Cancer Center

More information

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia

Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

SRS/SBRT Errors and Causes

SRS/SBRT Errors and Causes SRS/SBRT Errors and Causes Ryan Foster, Ph.D. Assistant Professor Director of Clinical Medical Physics Department of Radiation Oncology UT Southwestern Medical Center Dallas, TX Disclosures I receive research

More information

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998)

Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad: General Aspects of RCF (TG-235), AAPM July 2015 1 Specific Aspects of Radiochromic Film Dosimetry AAPM Task Group 235 An Update to Task Group 55 (1998) Azam Niroomand-Rad, PhD, FAAPM,

More information

International Practice Accreditation

International Practice Accreditation International Practice Accreditation Program endorsed and supported by Jean-Xavier Hallet Punta Cana, Nov 2017 Short History Created in 2004 by ESTRO, independent since 2009 Dosimetry Laboratory (IGR):

More information

8/1/2016. TG-191 The clinical use of luminescent dosimeters. Task Group Members. Scope of Report

8/1/2016. TG-191 The clinical use of luminescent dosimeters. Task Group Members. Scope of Report TG-191 The clinical use of luminescent dosimeters Task Group Members Stephen F. Kry, chair MD Anderson Cancer Center Paola Alvarez, MD Anderson Cancer Center Joanna Cygler, Ottawa Hospital Cancer Center

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

Guideline & Reports 医学物理学会教育委員会資料

Guideline & Reports 医学物理学会教育委員会資料 Guideline & Reports 医学物理学会教育委員会資料 2017/11 更新 report title year keyword 1 keyword 2 AAPM TG211 Classification and evaluation strategies of auto-segmentation approaches for PET 2017 PET autosegmentation

More information

Verification of Advanced Radiotherapy Techniques

Verification of Advanced Radiotherapy Techniques Verification of Advanced Radiotherapy Techniques Catharine Clark, PhD Principal Research Scientist Radiation Dosimetry group and Consultant Clinical Scientist Royal Surrey County Hospital 2 nd December

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET

Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET University of Michigan Department of Radiation Oncology Division of Radiation Physics Special Procedures Rotation I/II SBRT, SRS, TBI, and TSET Resident: Rotation staff mentor/ advisor: _Scott Hadley,

More information

Limits of Precision and Accuracy of Radiation Delivery Systems

Limits of Precision and Accuracy of Radiation Delivery Systems Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann

More information

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011

MVCT Image. Robert Staton, PhD DABR. MD Anderson Cancer Center Orlando. ACMP Annual Meeting 2011 MVCT Image Guidance and QA Robert Staton, PhD DABR MD Anderson Cancer Center Orlando ACMP Annual Meeting 2011 Disclosures MDACCO has received grant funding from TomoTherapy, Inc. Overview TomoTherapy MVCT

More information

Overview IAEA mandate and organization. IAEA activities in medical radiation physics: achievements and challenges. Contents

Overview IAEA mandate and organization. IAEA activities in medical radiation physics: achievements and challenges. Contents activities in medical radiation physics: achievements and challenges Ahmed Meghzifene Head, Dosimetry and Medical Radiation Physics International Atomic Energy Agency Contents : an overview programme development

More information

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Patient dosimetry for total body irradiation using single-use MOSFET detectors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results

Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results Abstract: A National Project for the in-vivo dosimetry in radiotherapy: first results A.Piermattei 1, L. Azario 1, S. Cilla 2, A.Fidanzio 1, F.Greco 1, M.T.Russo 3, S. Zucca 4 1 Istituto di Fisica e Unità

More information

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in:

BICOE Breast Imaging Center of Excellence. What is it? - Requirements. National Mammography Database. What do you get? ACR Accreditation in: BICOE Breast Imaging Center of Excellence What is it? - Requirements William Geiser, MS DABR Senior Medical Physicist MD Anderson Cancer Center Houston, Texas wgeiser@mdanderson.org ACR Accreditation in:

More information

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014

IROC Head and Neck Phantom. Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014 IROC Head and Neck Phantom Guidelines for Planning and Irradiating the IROC IMRT Phantom. Revised April 2014 The IROC Houston is requesting that each institution keep the phantom for two weeks. During

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Quality Assurance in Radiation Therapy

Quality Assurance in Radiation Therapy Quality Assurance in Radiation Therapy Purpose (from George) To make aware of the needs for quality assurance in radiation therapy To identify components of a quality assurance program in radiation therapy

More information

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS

DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course

M. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics

More information

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams

Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Verification of Relative Output Factor (ROF) Measurement for Radiosurgery Small Photon Beams Reduan A a, Mazurawati M b, Nur Iziana M a, Nik Ruzman NI b, Ahmad Z a and Ahmad Lutfi Y b a School of Health

More information

Standards for Radiation Oncology

Standards for Radiation Oncology Standards for Radiation Oncology Radiation Oncology is the independent field of medicine which deals with the therapeutic applications of radiant energy and its modifiers as well as the study and management

More information

Leila E. A. Nichol Royal Surrey County Hospital

Leila E. A. Nichol Royal Surrey County Hospital 2 nd UK and Ireland Dosimetry Check User Meeting Symposium Clatterbridge Cancer Centre, 24 th October 2012 Leila E. A. Nichol Royal Surrey County Hospital Leila.Nichol@nhs.net *My experience with Dosimetry

More information

Hong Kong Association of Medical Physics. Guide to Candidates

Hong Kong Association of Medical Physics. Guide to Candidates Hong Kong Association of Medical Physics Certification of Medical Physicists Professional Assessment Guide to Candidates PA App Form (r2017.1) Contents Chapter 1 p. 3 Professional Assessment System and

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

A new geometric and mechanical verification device for medical LINACs

A new geometric and mechanical verification device for medical LINACs JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A new geometric and mechanical verification device for medical LINACs Keith T. Welsh,* Robert A. Wlodarczyk, and L. E. Reinstein

More information

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D.

Indiana University Health Proton Therapy Center. Chee-Wai Cheng, Ph.D. Indiana University Health Proton Therapy Center Chee-Wai Cheng, Ph.D. Machine configuration and layout Fixed beam line, double scattering and propeller Uniform scanning nozzle and snout Gantry room Range

More information

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy VERSION 2.1 April 2015 Table of Contents Abbreviations & Glossary... 3 Executive Summary...

More information