Gas chromatography. Gas chromatography. Basic principles of GC. Capillary columns for GC. Derivatisation reagents.

Size: px
Start display at page:

Download "Gas chromatography. Gas chromatography. Basic principles of GC. Capillary columns for GC. Derivatisation reagents."

Transcription

1 Gas chromatography Gas chromatography Basic principles of GC Page Introduction to GC Brief selection guide for capillary columns OPTIMA high performance capillary columns for GC 0 6 PERMABOND capillary columns for GC 0 Hints for use and maintenance of capillary columns GC capillary columns for special applications 46 Summary, capillary columns with thick films, fast GC columns 4, amine separations, FAME separations 6, GC analysis of petrochemical compounds 8, environmental analyses 9, triglyceride separation 40, separation of silanes 4, analysis of diethylene glycol 4, enantiomer separation 4 46 Untreated capillaries, retention gaps and deactivated capillaries Fused silica capillary columns not chemically bonded 49 Derivatisation reagents Summary, method development kits, selection guide for derivatisation of important functional groups in GC 0 Acylation, reagents and methods Methylation, reagents and methods Silylation, reagents and methods 8 Standards for GC Environmental standards for GC 9 6 Test mixtures for GC capillary columns 6 6 Accessories for GC 64 7 Accessories for capillary columns 64 6, 68, Valco fittings for capillary GC 66 67, accessories for packed GC columns 69, products for gas flow measurement and in-line gas purification 69 7, ferrules 7, septa 7 0

2 Basic principles of GC Basic principles of GC Introduction to GC Schematic presentation of a GC system gas supply chromatogram sample injection recorder capillary column detector data system In this chapter we want to give a short and general introduction for the beginner in gas chromatography. By no means this is meant as a complete treatise. For further information please refer to the references given below*. The GC system The chromatographic system is composed of the chromatograph and a recorder for plotting chromatograms or a data station for generation and evaluation of chromatograms. The chromatograph consists of the sample injector, gas supplies, oven with temperature control for the chromatographic column and the detector. It would take us too far to describe all injection techniques in detail. We just want to mention the different basic possibilities of direct and indirect sample injection. With direct injection the sample is introduced into the column without contact with other parts from glass or metal (on column injection). With indirect techniques the sample is injected into an evaporator. The vapour then is transferred into the column either completely or partially (split technique). With both techniques the injection can take place at low temperatures, at high temperatures or with temperature programming. The gas supply unit of a chromatograph has to provide all gases needed for a separation: the carrier gas and e.g. the burner gases for the flame ionisation detector (FID). The detector indicates a substance by generation of a more or less intense electrical signal (response). Analogue (recorder) or digital (computer) processing of these signals produces the chromatogram or a numerical report. Some detectors are specific for certain classes of substances or for certain atoms (P, N, etc.). The separation process The chromatographic separation is achieved by repeated distribution of each sample component between two phases. One of these phases is stationary. The second, the mobile phase moves along this stationary phase. In Gas Chromatography (GC) the mobile phase is always a gas (mostly N, H, He). The stationary phase in GSC (Gas Solid Chromatography or adsorption chromatography) is a porous polymer solid, while in GLC (Gas Liquid Chromatography or partition chromatography) the stationary phase is a mostly viscous gumlike liquid. Packed columns are completely filled with a packing, liquid stationary phases being coated onto an inert support. Capillary columns do not require a support, because their inner wall is coated with the stationary phase (WCOT = Wall Coated Open Tubular). Transport of the components is achieved exclusively in the gas phase, separation is accomplished in the stationary phase. The quality of a separation (resolution) depends on how long the components to be separated stay in the stationary phase and on how often they interact with this phase (selectivity). The type of interaction between component and phase (selectivity) is determined by the functional groups. The polarity of the phase is a function of stationary phase substituents. The chromatogram Schematic presentation of a chromatogram Start t 0 t R w / A chromatogram consists of a base line and a number of peaks. The area of a peak allows quantitative determinations. Starting point of a chromatogram is the time of injection of a dissolved sample. The time interval between a peak and the point of injection is called retention time t R. A component can be identified by its retention time (qualitative determination). The retention time is the sum of the residence time of a solute in the mobile phase (t 0 ) and in the stationary phase (t R ' = net retention time); t R i = t 0 + t R i t 0 is also known as dead time. It is the time required by a component to migrate through the chromatographic system without any interaction with the stationary phase (also called air or gas peak). The net retention time t R i is the difference t R 06

3 Basic principles of GC Basic principles of GC Introduction to GC between total retention time t R i and dead time t 0. It indicates how long a substance stays in the stationary phase. = t 0 t R i t R i The capacity factor (k ) is a measure for the position of a sample peak in the chromatogram. The capacity factor is specific for a given compound and constant under constant conditions. t R i t 0 k i = t 0 The relative retention (α), also called separation factor or selectivity coefficient, is defined as α = k k i.e. it is the ratio of two capacity factors, the reference substance always being in the denominator. The relative retention does not provide any information on the quality of a separation, since for equal vaues of α two very broad peaks may overlap, (as shown in trace a), or may be completely resolved (as in trace b), if they are correspondingly narrow. Such a plate is of course a theoretical hypothesis, representing that portion of a column in which the partition equilibrium is established once. The smaller this value the better works the column. The relation between N, R and k' is R = α k i N 4 α k i + th From this formula the number of plates N th required for a given resolution R can be calculated. The Van Deemter equation shows how the plate height h depends on the flow velocity u: HETP h min B u HETP = A + B ū -- + C u C. u u opt. u a b retention time For this reason the peak width at half height (w / ) is taken into account when considering the resolution R according to t R t R R = (w / ) + (w / ) The resolution R is a measure for the quality of a separation. The separation number SN is another value indicating the quality of a separation. It defines how many peaks can be resolved between two members of a homologous series of e.g. n-paraffins. t R t R SN = (w / ) + (w / ) If retention times are not too small, SN can be determined in a temperature programme. The number of theoretical plates (N th ) characterises the quality of a column (it should be determined for k > ): t R N th =.4 ( i ) w / The height equivalent to a theoretical plate (h, HETP) is calculated by dividing the length L of the column by the number of theoretical plates N th : h = HETP = L N th A = Eddy diffusion; A is a function of packing uniformity; for WCOT capillary columns A equals 0 (WCOT = Wall Coated Open Tubular) B = molecular axial diffusion; B is a function of the diffusion coefficient of the component in the respective carrier gas C = resistance to mass transfer In practice often higher velocities than u opt. are chosen, if separation efficiency is sufficient, since higher carrier velocities mean shorter retention times. * References: D. Rood, A practical guide to care, maintainance and troubleshooting of capillary gas chromatographic systems, nd edition, Hüthig Verlag, Heidelberg, 99 W. Blum, R. Aichholz Hochtemperatur-Gas-Chromatographie, Hüthig Verlag, Heidelberg, 99 G. Schomburg, Gas Chromatography, A Practical Course, Verlag Chemie, Weinheim, 990 K. Grob, Classical Split and Splitless Injection Techniques in Capillary GC, Hüthig Verlag, Heidelberg, 986 K. Grob, Making and Manipulation of Capillary Columns for Gas Chromatography, Hüthig Verlag, Heidelberg, 986 P. Sandra, Sample Introduction in Capillary Gas Chromatography, Volume, Hüthig Verlag, Heidelberg, 98 E. Leibnitz, H. G. Struppe, Handbuch der Gas-Chromatographie, Akademische Verlagsgesellschaft Geest u. Portig KG, Leipzig, Germany, 984 W. Jennings, Gas Chromatography with Glass Capillary Columns, Academic Press, New York 978 R. Kaiser, Chromatographie in der Gasphase, BI Hochschultaschenbücher, several volumes, Bibligraphisches Institut, Mannheim, Germany,

4 Brief selection guide for capillary columns The large number of stationary phases and the variability of column parameters have caused a considerable diversification of our column programme. now offers more than 0 different phases with a total of more than 60 Cat. Nos. For this reason especially the beginner may have some problems to select the capillary column best suiting his particular separation needs. This is why we will briefly explain the most important criteria for column selection. A capillary column is characterised by the following parameters: e.g. Cat. No. 76. m x 0. mm ID OPTIMA,.0 µm film A B C D E A = length B = inner diameter C = type of chemical bonding, immobilisation D = polarity of the stationary phase E = film thickness All these parameters have to be optimised for each particular chromatographic task. A. Length The separation efficiency (correctly speaking the number of plates n) of a capillary column is directly proportional to the length. However, the same relation is true for the retention time and thus for the total analysis time, making it desirable to use columns as short as possible. For routine separations in general m or shorter will be sufficient with the advantage of short analysis times. For difficult problems such as the separation of complex mixtures with about 0 components or more 60 m columns may be necessary. Use of longer columns is not recommended. Doubling the column length will only result in a 40% increase in resolution, since the resolution is proportional to the square root of the length. However, the time needed for an isothermal analysis will be doubled. B. Inner diameter The inner diameter also influences the separation efficiency of a capillary column: the lower the inner diameter the higher is the theoretically possible number of plates per meter. This will increase the resolution of the column, which is often required for complex sample mixtures. Due to the low sample capacity one normally has to work with split injection. Increasing the inner diameter while keeping the film thickness constant will increase the capacity. This is especially important for trace analyses and / or for the injection of larger sample volumes. If you halve the diameter of the column while keeping the film thickness constant, the capacity is also reduced by 0%. As a general rule one can say: a column with 0. mm ID and a film thickness of 0. µm has a maximum capacity of 00 ng/component. Overloading the column results in a loss of resolution. a) 0. to 0. mm inner diameter OPTIMA columns with this diameter are especially recommended for GC-MS with the advantage of high resolution and short retention times with low carrier gas flows. b) 0. mm inner diameter Columns with this diameter are used for the analysis of complex mixtures. They are recommended as routine columns for high resolution, especially in research applications. d) 0. mm inner diameter A column with 0. mm ID is the column for routine analyses with short retention times, yet increased capacity. With suitable film thickness it is the column of choice for trace analyses and on-column or splitless injection techniques. If you want to use a mass selective detector (MSD), you are often limited to smaller diameters, because many vacuum pumps cannot handle the high flow rates required for columns with larger diameters. e) 0. mm inner diameter Compared to packed columns these capillaries show more inert surfaces and higher reproducibility with at least equal separation efficiency. They are not only the replacement for packed columns, but the capillary for rapid separations with inert surface and highest capacity. The inner diameter 0. mm is ideal for use with the thermal conductivity detector. C. Immobilisation OPTIMA : the top quality of fused silica (FS) capillaries from MACHEREY-NAGEL with chemically bonded (immobilised) stationary phases. A special treatment after immobilisation of the phase results in lower column bleeding. An immobilised film of stationary phase tolerates injection of large sample volumes in on-column or splitless sample injection techniques. PERMABOND : FS capillary columns, also with immobilised (chemically bonded) stationary phase. FS capillary columns with non-immobilised phases. D. Stationary phase Different substituents in the chemical structures of stationary phases are responsible for the type of interaction (selectivity) between the phase and the solutes to be separated and thus determine the relative retention and the number of theoretical plates required for a given resolution. The stationary phase also limits the temperature range for chromatography. Nonpolar phases separate by efficiency (n) and retention (k') only, while polar phases offer additional interactions, which may influence or improve a separation. Selectivity has to be optimised for the critical pair of components or the main component. You should always select the least polar column which solves your separation problem. About 70% of all separations can be performed on non- to midpolar columns. These columns generally feature high temperature stability. Nonpolar stationary phases (e.g. 00% dimethylpolysiloxane phases) sepa- For applications with GC capillary columns please ask for our catalogue GC Applications 08

5 Brief selection guide for capillary columns rate components mostly according to the height of the boiling point. Typical analytes are linear hydrocarbons (n-alkanes). When increasing the polarity of the phase, e.g. by introduction of phenyl and / or cyanopropyl groups, separation is increasingly influenced by differences in dipole moment and by charge transfer (e.g. for 0% diphenylpolysiloxane phases). Typical analytes are hydrocarbons, which contain oxygen, sulphur, nitrogen, phosphorus or halogen atoms, unsaturated molecules which can be polarised and aromatics. For components featuring different hydrogen bonding capacities and the ability to form strong hydrogen bonds, polyethylene glycol phases (Carbowax) are the best choice for a separation. Typical analytes are alcohols and carboxylic acids. E. Film thickness Our programme of capillary columns comprises films from 0. to.0 µm. The standard film thickness is 0. µm. If you work with thin films (0. 0. µm), in general substances are eluted faster, at lower temperatures and with better resolution than with a thick film (>0. µm). For highboiling compounds, temperature labile or very closely eluting sample substances thin films are very well suited. Increasing the film thickness will increase the capacity and improve inertness. Thick films are ideal for low-boiling compounds, often eliminating the need for cooling during chromatography, since they increase elution temperatures, i.e. the components are retained more strongly. As a general rule one can say: doubling the film thickness results in an increase in elution temperature of about 0 C under isothermal conditions. This is advantageous for aqueous systems as well. Variation of the film thickness is often better than increasing the column length as is shown in the following chromatograms: The upper figure shows a separation with the special column PERMABOND P-00 with 0. µm film thickness and 00 m length, while the lower figure shows the chromatogram on a 0 m column with µm film thickness. The second figure clearly demonstrates the high separation efficiency of a 0 m column with very thick film compared to the 00 m column with thin film. According to the criteria cited above the following capillaries are suited for the larger part of all applications and can be considered as standard equipment for GC:. Cat. No m x 0. mm ID OPTIMA, 0. µm film. Cat. No m x 0. mm ID OPTIMA 70, 0. µm film. Cat. No. 7. m x 0. mm ID PERMABOND CW 0 M, 0. µm film 4. Cat. No m x 0. mm ID OPTIMA δ-, 0. µm film a suitable substitute for. and. Analysis of hydrocarbons C C Capillary column: PERMABOND P-00, 0. µm film, 00 m x 0. mm ID, max. temperature 00/0 C, Cat. No Injection volume: 00 µl Carrier gas: 00 kpa H (. ml/min) Split: 00 ml/min Temperature: C Detector: FID, 0 C, 8. Methane. Ethane. Propane 4. i-butane. Butane 6. Methylbutane 7. Pentane min Separation of hydrocarbons C C Capillary column: OPTIMA,.0 µm film, 0 m x 0. mm ID, max. temperature 80 C, Cat. No Injection volume: 00 µl Carrier gas: 0.08 bar N Split: 60 ml/min Temperature: 6 C Detector: FID, 4. Methane. Ethane. Propane 4. i-butane. n-butane 6. Methylbutane 7. n-pentane min For applications with GC capillary columns visit our web site: 09

6 OPTIMA high performance capillary columns As a result of our efforts in research and development and the continuous improvements in our manufacturing techniques we present OPTIMA a series of high performance capillary columns for gas chromatography. OPTIMA capillary columns provide high thermal stability Improved temperature stability is the reason why OPTIMA capillary columns can be operated at about 40 C higher temperatures compared to standard phases. High-boiling solutes (with very low vapour pressures) "normally" have very long retention times and rather broad peak shapes. OPTIMA columns with their increased operation temperatures elute high-boiling compounds faster and with better peak shapes. Maximum operating temperatures for OPTIMA phases The first temperature is valid for isothermal operation, the second for short isotherms in a temperature programme. Temperature limits for 0. mm ID columns and for columns with film thickness of µm or greater are given with the ordering information on the following pages. Phase max. operating temperatures OPTIMA 40/60 C OPTIMA 40/60 C OPTIMA δ- 40/60 C OPTIMA δ-6 40/60 C OPTIMA 7 0/40 C OPTIMA 0 00/0 C OPTIMA 70 00/0 C OPTIMA 64 80/00 C OPTIMA 0 60/80 C OPTIMA 60/80 C OPTIMA 40 60/80 C OPTIMA CN-00 40/60 C reduced column bleed Improved manufacturing processes offer capillaries with lower bleed levels, which are especially recommended for GC-MS. Less column bleed yields increased sensitivity and accuracy through a better signal to noise ratio for any kind of detector. Reduced column bleed improves detectability of solutes in qualitative and quantitative GC-MS analyses. more inert columns due to optimised deactivation Polar compounds are frequently difficult to analyse, because they often give broad tailing peaks. Advances in deactivation technology for the OPTIMA capillaries result in an excellent chromatographic performance yielding improved peak shapes of polar compounds combined with improved efficiency and sensitivity. In order to guarantee best industry standard capillaries with highest reproducibility from capillary to capillary OPTIMA columns have to meet high specifications. For controlling quality MACHEREY-NAGEL determines the following parameters: efficiency by measuring the separation number (Trennzahl according to Kaiser) in a temperature programme. polarity by measuring retention indices bleeding in a temperature programme with a test mixture including high-boiling hydrocarbons inertness by measuring the peak height ratio for decylamine/c- (for non- to medium polar phases) Capillary column: OPTIMA, 0. µm film, 0 m x 0. mm ID, max. temp. 40/60 C, Cat. No Carrier gas:.0 bar H Temperature: 70 C 0 C (0 min), 0 C/min 0 C ( min) Detector: FID 00 C, 0. C-8. Dibutylamine. C-0 4. Phenylethanol.,6-Dichloroaniline 6.,4-Dichlorophenol 7. C- 8. Decylamine 9. Decanol 0. C-4. C-6. C-8. C-0 4. C-. C-4 6. C-6 7. C-8 8. C-0 9. C min 70 For ordering information of OPTIMA capillary columns please see pages 6 6. For applications with GC capillary columns please ask for our catalogue GC Applications 0

7 OPTIMA high performance capillary columns Comparison of separation properties for selected OPTIMA phases OPTIMA µm film 0 m x 0. mm ID max. temp. 60/80 C Cat. No OPTIMA 0. µm film 0 m x 0. mm ID max. temp. 60/80 C Cat. No OPTIMA 0 0. µm film 0 m x 0. mm ID max. temp. 60/80 C Cat. No OPTIMA µm film 0 m x 0. mm ID max. temp. 0/40 C Cat. No OPTIMA 7 0. µm film 0 m x 0. mm ID max. temp. 0/40 C Cat. No OPTIMA 0. µm film 0 m x 0. mm ID max. temp. 40/60 C Cat. No increasing polarity OPTIMA 0. µm film 0 m x 0. mm ID max. temp. 40/60 C Cat. No time [min] 0 0 min Sample: -OPTIMA test mixture (Cat. No. 76) Injection:.0 µl, split :0 Carrier gas: 80 kpa N Temperature: 80 C T max (isothermal), 8 C/min Detector: FID, C, 6. Undecane. Dodecane. Octanol 4. Dimethylaniline. Decylamine 6. Methyl decanoate 7. Methyl undecanoate 8. Henicosane 9. Docosane 0. Tricosane For applications with GC capillary columns visit our web site:

8 OPTIMA δ the unique family of phases with autoselectivity permanent dipole: constant interactions dipoles of the analytes OV- conventional phase OPTIMA δ- SE-4 OV-7 OV-70 OV-0 OV-40 OPTIMA δ-6 OPTIMA δ permanent dipoles of the stationary phase induced dipoles of OPTIMA δ phases permanent + induced dipole = autoselectivity range of polarities covered by OPTIMA δ phases OPTIMA δ- and OPTIMA δ-6 are stationary phases, which exhibit not only a fixed polarity, like most polysiloxane phases, but additionally a polarity, which is influenced by the sample. The polymers consist of cross-linked methyl-phenylpolysiloxane block polymers with defined composition, and extremely narrow molecular weight distribution, which are exclusively produced for MACHEREY-NAGEL. Especially polar analytes are able to induce a dipole moment in the stationary phase, so that the molecules show stronger interactions with the phase. This enhanced interaction is maintained at higher temperatures, where normally interactions between molecule and phase become reduced due to the Brownian movement. We call this phenomenon autoselectivity, because the stationary phase adjusts itself to the polarity of the analytes. Thus OPTIMA δ phases cover broad ranges of polarities. Compared with conventional phases, OPTIMA δ- polarity ranges from approximately the nonpolar OPTIMA to the midpolar OPTIMA 70, while for OPTIMA δ-6 the polarity covers a range from about the midpolar OPTIMA 7 to the polar OPTIMA 0. Due to this feature, the OPTIMA δ columns show interesting patterns of selectivity. For example, inversions in the sequence of peak elution may occur, which recommends the columns for reference use (e. g. in combination with OPTIMA ). In conventional midpolar phases the polarity is induced by phenyl, but especially by cyano and trifluoromethyl groups. The two latter often cause bleeding, which results in severe problems with some detectors. In contrast, the OPTIMA δ phases show very high temperature limits (40/60 C), as well as the low bleed levels, which makes them ideal for the use with mass selective (MSD) or phosphorus/nitrogen detectors (PND) in the field of environmental trace analysis. Key features of the OPTIMA δ are: Wide range of applications due to autoselectivity Outstanding thermal stability similar to non-polar phases Very low bleed levels Extremely inert Medium polar without CN groups References: W. Röder, D. Lennartz, GIT /99, p. 6 R. Looser, K. Ballschmiter, J. Chromatogr. 86 (999), 7-84 R. Baycan-Keller, M. Oehme, J. Chromatogr. 87 (999), 0 0 For applications with GC capillary columns please ask for our catalogue GC Applications

9 OPTIMA δ the unique family of GC phases with autoselectivity Analysis of isomeric phenols Isomeric phenols, such as chloro- and nitrophenols, are difficult to analyse with standard GC phases (e.g. OPTIMA or OPTIMA 7) because of coelutions. The autoselective OPTIMA δ- is able to separate all phenols due to stronger interactions occurring with more polar molecules, because polar analytes induce a dipole moment in the phase of the OPTIMA δ-. Capillary column: OPTIMA δ-, 0. µm film, 60 m x 0. mm ID, max. temperature 40/60 C, Cat. No Injection:.0 µl, split :80 Carrier gas: He,. bar 8 Temperature: 60 C ( min) 0 C, 6 C/min 6 Detector: MSD HP Phenol 4. -Chlorophenol. -Methylphenol Methylphenol. -Methylphenol 6.,4-Dimethylphenol 7. -Nitrophenol 8.,4-Dichlorophenol 9.,6-Dichlorophenol 0. 4-Chloro--methylphenol ,,-Trichlorophenol 6 0.,4,6-Trichlorophenol.,4,-Trichlorophenol 4.,,4-Trichlorophenol.,,6-Trichlorophenol 6.,,,6-Tetrachlorophenol 9 7.,,4,-Tetrachlorophenol 8.,,4,6-Tetrachlorophenol 9.,4-Dinitrophenol 0.,4,-Trichlorophenol. -Methyl-4,6-dinitrophenol. -Isopropyl-4,6-dinitrophenol 0 0 min Separation of organochlorine pesticides (EPA 808) Capillary column: OPTIMA δ-6, 0. µm film, 0 m x 0. mm ID, max. temperature 40/60 C, Cat. No Sample: EPA 808 organochlorine pesticide calibration mix (Restek), 00 µg/ml each in toluene : hexane ( : ) Injection volume: µl, Split : 0 Carrier gas:.0 bar He 0 Temperature: 80 C 00 C (0 min), 4.0 C/min 4 0 Detector: MSD HP α-bhc 0. γ-bhc (lindane) 4. β-bhc Heptachlor. δ-bhc 6. Aldrin 7. Heptachlor epoxide 8. γ-chlordane 9. α-chlordane Endosulfan I. 4,4'-DDE. Dieldrin. Endrin ,4'-DDD. Endosulfan II 6. 4,4'-DDT 7. Endrin aldehyde 8. Endosulfan sulphate 9. Methoxychlor 0. Endrin ketone 0 0 min Abundance For applications with GC capillary columns visit our web site:

10 OPTIMA high performance capillary columns Unique from, no similar phases Max. temperature for isothermal operation 40 C, max. temperature for short isotherms in a temperature programme 60 C Autoselectivity resulting in a wide range of polarities from approximately the non-polar OPTIMA to the midpolar OPTIMA 70 Analytes determine the polarity of the phase Medium polar without CN groups Ideal for MSD and PND detectors USP 49 Separation of organophosphorus pesticides (EPA 840/84) on OPTIMA δ- Capillary column: OPTIMA δ-, 0. µm film, 0 m x 0. mm ID, max. temperature 40/60 C, Cat. No sample EPA 840 OP pesticide calibration mix (Restek), 00 µg/ml each in hexane : acetone = 9 : Injection: µl, split : 0, carrier gas:.0 bar He, temperature: 0 C 00 C (0 min),. C/min Detector: MSD HP 97. Dichlorvos. Mevinphos. Demeton-s 4. Ethoprop. Naled 6. Phorate 7. Demeton-o 8. Diazinon 9. Disulfoton 0. Parathion-methyl. Ronnel. Fenthion. Chlorpyrifos 4. Trichloronate. Merphos 6. Stirofos 7. Tokuthion 8. Merphos oxidation product 9. Fensulfothion 0. Bolstar. Azinphos-methyl. Coumaphos 4 0 m m m 0 m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0. µm film µm film mm ID (0. mm OD) 0. µm film µm film µm film mm ID (0.8 mm OD).00 µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns please ask for our catalogue GC Applications min

11 OPTIMA high performance capillary columns Unique from, no similar phases Max. temperature for isothermal operation 40 C, max. temperature for short isotherms in a temperature programme 60 C Autoselectivity resulting in a wide range of polarities from approximately the non-polar OPTIMA 7 to the midpolar OPTIMA 0 Analytes determine the polarity of the phase Medium polar without CN groups Ideal for MSD and PND detectors Separation of organophosphorus pesticides (EPA 840/84) on OPTIMA δ-6 Capillary column: OPTIMA δ-6, 0. µm film, 0 m x 0. mm ID, max. temperature 40/60 C, Cat. No Sample: EPA 840 OP pesticide calibration mix (Restek), 00 µg/ml each in hexane : acetone = 9 : Injection volume: µl, split :0 Carrier gas:.0 bar He 8 4 Temperature: 6 9 Detector: MSD HP 97 9, Dichlorvos 0 C 00 C (0 min),. C/min. Mevinphos. Demeton-s 4. Ethoprop. Naled 6. Phorate 7. Demeton-o 8. Diazinon 7 9. Disulfoton 0 0. Parathion-methyl. Ronnel. Fenthion. Chlorpyrifos 4. Trichloronate. Merphos 8 6. Stirofos 7. Tokuthion 8. Merphos oxidation product 9. Fensulfothion 0. Bolstar. Azinphos-methyl. Coumaphos min m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0. µm film mm ID (0. mm OD) 0. µm film µm film mm ID (0.8 mm OD).00 µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns visit our web site:

12 OPTIMA high performance capillary columns 00% dimethylpolysiloxane O CH Si CH n For columns with mm ID and films < µm the max. temperature for isothermal operation is 40 C, the max. temperature for short isotherms in a temperature programme is 60 C for 0. mm ID columns with films < µm the max. temperatures are 0 and 40 C, resp. for thick film columns with films µm the max. temperatures are 00 and 0 C, resp. Nonpolar Separation of components according to boiling points Thick film columns µm film are especially recommended for solvent analyses Similar phases: OV-, DB-, SE-0, HP-, Ultra-, SPB-, CP-SIL CB, Rtx-, 007-, BP, MDN-, AT-, ZB, OV 0 USP G 0 m m m 0 m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film µm film mm ID (0.4 mm OD) 0.0 µm film µm film µm film µm film mm ID (0.4 mm OD) 0.0 µm film µm film µm film µm film mm ID (0. mm OD) 0.0 µm film µm film µm film µm film µm film µm film µm film mm ID (0.8 mm OD) 0.0 µm film µm film µm film µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns please ask for our catalogue GC Applications 6

13 OPTIMA high performance capillary columns Solvent analysis Capillary column: OPTIMA,.0 µm film, 60 m x 0. mm ID, max. temperature 40/60 C, Cat. No Sample: Solvent mixture, courtesy of J. Lutz, Alcan Rorschach, Switzerland Injection volume: 0.4 µl, split : 60 Carrier gas: H, 0 KPa Temperature: 0 C (9 min) 90 C, 4 C/min 80 C ( min), 4 C/min Detector: FID, 00 C, 6 mv min 0. Methanol. Ethanol. Acetone 4. Isopropanol. Methyl acetate 6. n-propanol 7. Methyl ethyl ketone 8. Ethyl acetate 9. Isobutanol 0. n-butanol. -Methoxy--propanol. Isooctane. Ethylglycol 4. Isoheptane. Methyl isobutyl ketone 6. -Ethoxy--propanol 7. Toluene 8. Isobutyl acetate 9. Butyl acetate 0. 4-Hydroxy-4-methyl--pentanone. -Methoxy--propyl acetate. Xylene. Cyclohexanone 4. Ethyl glycol acetate. Butyl glycol 6. Heptanol 7. Ethyldiglycol 8. Butyldiglycol 9. Butyl glycol acetate 0. Butyldiglycol acetate 090 CH Max. temperature for isothermal operation 40 C, max. temperature for short isotherms in a temperature programme 60 C Selectivity identical with OPTIMA Phase with lowest bleeding Ideal for GC/MS and ECD applications and general analyses at trace level 00% dimethylpolysiloxane O Si CH n m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film µm film mm ID (0.4 mm OD) 0. µm film mm ID (0. mm OD) 0. µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns visit our web site: 7

14 OPTIMA high performance capillary columns % phenyl 9% dimethylpolysiloxane CH O Si n CH O Si CH m For columns with mm ID and films < µm the max. temperature for isothermal operation is 40 C, the max. temperature for short isotherms in a temperature programme is 60 C for 0. mm ID columns with films < µm the max. temperatures are 0 and 40 C, resp. for thick film columns with films µm the max. temperatures are 00 and 0 C, resp. Nonpolar Standard phase with large range of application Suitable for GC/MS (columns with small films) Similar phases: SE-4, SE-, DB-, HP-, Ultra-, SPB-, CP-SIL 8, Rtx-, 007-, BP, MDN-, AT-, ZB- USP G 7 0 m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0.0 µm film µm film µm film µm film mm ID (0.4 mm OD) 0.0 µm film µm film µm film µm film µm film mm ID (0. mm OD) 0.0 µm film µm film µm film µm film µm film µm film µm film mm ID (0.8 mm OD) 0.0 µm film µm film µm film µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns please ask for our catalogue GC Applications 8

15 OPTIMA high performance capillary columns % diphenyl 9% dimethylpolysiloxane O Si n O CH Si CH Pyrolysis GC of chloroprene rubber Capillary column: OPTIMA MS, µm film, 60 m x 0. mm ID, max. temperature: 40/60 C, Cat. No Method: g solid are dissolved in 0 ml of hexane or methanol, 00 µl of this solution are pyrolysed in a crucible, headspace Temperature: 60 C ( min) 0 C, 8 C/min Detector: MSD chromatogram courtesy of Mrs. Engel, VW, Wolfsburg, Germany m 0040 Max. temperature for isothermal operation 40 C, max. temperature for short isotherms in a temperature programme 60 C Selectivity identical with OPTIMA NEW! New quality even less bleeding Ideal for GC/MS and ECD applications and general analyses at trace level Analysis of PCB (W congener mix) Capillary column: OPTIMA MS, 0. µm film, 0 m x 0.0 mm ID, max. temperature: 40/60 C, Cat. No Sample: PCB-W congener mix, 0 µl/ml Injection: µl Split: 80 ml/min Carrier gas: 0. bar He Temperature: 0 C 00 C ( min),. C/min Detector: ECD, 00 C 8. PCB 8. PCB. PCB 8 4. PCB 0. PCB PCB PCB 0 8. PCB PCB 8 0. PCB PCB 0. PCB 8. PCB PCB 70. PCB min min 0 m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film µm film mm ID (0.4 mm OD) 0. µm film µm film µm film mm ID (0. mm OD) 0. µm film µm film µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns visit our web site: 9

16 OPTIMA high performance capillary columns phenylmethylpolysiloxane (0% phenyl) O CH Si Max. temperature for isothermal operation 0 C, max. temperature for short isotherms in a temperature programme 40 C for 0. mm ID columns the max. temperatures are 80 and 00 C, resp. Medium polar Suitable for higher temperatures Preferred applications: steroids, pesticides, drug analyses Similar phases: OV-7, DB-7, HP-0+, HP-7, SPB-0, SP-0, Rtx-0, CP-SIL 4 CB, 007-7, ZB-0 USP G n Analysis of aromatic hydrocarbons Capillary column: OPTIMA 7, 0. film, 0 m x 0. mm ID, max. temperature 0/40 C, Cat. No Injection volume: 0. µl Carrier gas: 0 kpa N, split :0 Temperature: 0 C 00 C, C/min Detector: FID, 60 C, 6. Benzene. Toluene 8. Ethylbenzene 4. m-xylene. o-xylene 6. Isopropylbenzene 7. n-propylbenzene p-isopropylbenzene n-butylbenzene 0. n-hexylbenzene. n-octylbenzene min 0 m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0.0 µm film µm film mm ID (0.4 mm OD) 0. µm film µm film µm film mm ID (0. mm OD) 0. µm film µm film µm film µm film mm ID (0.8 mm OD).00 µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns please ask for our catalogue GC Applications 0

17 OPTIMA high performance capillary columns 4% cyanopropyl-phenyl 86% dimethylpolysiloxane CH O Si O Si Analysis of Fusarium mycotoxins F. Walker, B. Meier, Journal of AOAC Int. 8 (998) Capillary column: OPTIMA 70, 0. µm film, m x 0. mm ID, max. temperature 00/0 C, Cat. No Injection: µl, 0 C, splitless Carrier gas: N, 0 cm/s, linearvelocity at 60 C Temperature: 60 C ( min) 40 C, 6 C/min 70 C, 0 C/min ( min) Detector: 6 Ni-ECD, 0 C..9 min Nivalenol (0.007 ppm).. min Deoxynivalenol (0.006 ppm). 7. min -O-Acetyl-4-deoxynivalenol (0.006 ppm) min -Acetyldeoxynivalenol (0.00 ppm) (CH ) CN n CH m 0660 Max. temperature for isothermal operation 00 C, max. temperature for short isotherms in a temperature programme 0 C for 0. mm ID columns the max. temperatures are 80 and 00 C, resp. Medium polar Special selectivity due to high cyanopropyl content Reference column for structure identification, e.g. in combination with OPTIMA Film thickness µm for solvent analyses Similar phases: OV-70, DB-70, CP-SIL 9 CB, HP- 70, Rtx-70, SPB-70, , BP0, ZB m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0.0 µm film mm ID (0.4 mm OD) 0. µm film µm film µm film mm ID (0. mm OD) 0.0 µm film µm film µm film µm film µm film mm ID (0.8 mm OD).00 µm film µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns visit our web site:

18 OPTIMA high performance capillary columns 6% cyanopropyl-phenyl 94% dimethylpolysiloxane CH O Si O Si Max. temperature for isothermal operation 00 C, max. temperature for short isotherms in a temperature programme 0 C Medium polar Ideal for pesticide analyses Similar phases: HP-0, DB-0, SPB-0, Rtx-0, CP-0, (CH ) CN n CH m 0 m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0. µm film mm ID (0. mm OD) 0. µm film µm film mm ID (0.8 mm OD).00 µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. Analysis of a pesticide mixture Capillary column: OPTIMA 0, 0. µm film, 60 m x 0. mm ID, max. temperature 00/0 C, Cat. No Injection: µl, 80 C ( min) 0 C ( min), pulsed splitless Carrier gas: He, 4 ml/min Temperature: 80 C ( min) 90 C, 0 C/min ( min) 40 C, C/min ( min) 60 C, 0 C/min (0 min) Detector: ECD Peaks (0, ng/µl):. Propyzamide. Vinclozolin. Bromophos-ethyl 4 4.,4-DDT. Brompropylate Analysis of a PCB mixture Capillary column: OPTIMA 0, 0. µm film, 60 m x 0. mm ID, max. temperature 00/0 C, Cat. No Injection: µl, 80 C ( min) 0 C ( min), pulsed splitless Carrier gas: He, 4 ml/min Temperature: 80 C ( min) 90 C, 0 C/min ( min) 40 C, C/min ( min) 60 C, 0 C/min (0 min) Detector: ECD Peaks (0, ng/µl):. PCB 8. PCB. PCB 8 4. PCB. PCB PCB min min For applications with GC capillary columns please ask for our catalogue GC Applications

19 OPTIMA high performance capillary columns cyanopropyl-phenyl-dimethylpolysiloxane Max. temperature for isothermal operation 80 C, max. temperature for short isotherms in a temperature programme 00 C Medium polar Recommended for environmental analyses Similar phases: HP-64, HP-VOC, DB-64, DB-VRX, SPB-64, CP-64, RTX-64, RTX-Volatiles, , BP64, VOCOL USP G 4 Separation of volatile hydrocarbons by Headspace GC Column: OPTIMA 64,.8 µm film, 0 m x 0. mm ID, max. temperature: 80/00 C, Cat. No , Injection: HS cryofocussion, splitless, 0 C Carrier gas: N,. ml/min Temperature: 6 C (4 min), 00 C, 0 C/min, ( min) Detector: ECD, 60 C. Dichloromethane (6,69 min). trans-,-dichloroethene (7,7 min). cis-,-dichloroethene (8,96 min) 4. Trichloromethane (9,46 min).,,-trichloroethane (9,8 min) 6. Tetrachloromethane (0,7 min) 7. Trichloroethene (,44 min) 8.,,-Trichloroethane (,94 min) 9. Tetrachloroethene (4,6 min) min Chromatogram courtesy of R. Mueller, G&P Torsten Plaar Umweltanalytik, Oldenburg, Germany 0 m m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD).0 µm film mm ID (0.4 mm OD).40 µm film mm ID (0. mm OD).80 µm film mm ID (0.8 mm OD).00 µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns visit our web site:

20 OPTIMA high performance capillary columns trifluoropropyl-methylpolysiloxane (0% trifluoropropyl) CH O Si (CH ) CF n Aromatic hydrocarbons (BTX) Capillary column: OPTIMA 0, 0. µm film, 0 m x 0. mm ID, max. temperature 40/60 C, Cat. No Injection volume: 0. µl Carrier gas: 0 kpa N (. ml/min) Split: 0 ml/min Temperature: 0 C Detector: FID, 0 C, 6. Benzene. Toluene. Ethylbenzene 4. p-xylene. m-xylene 6. o-xylene 000 Max. temperature for isothermal operation 60 C, max. temperature for short isotherms in a temperature programme 80 C Polar Recommended for environmental analyses Similar phases: OV-0, DB-0, Rtx-00, close equivalent to USP G min 0 0 m m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0. µm film µm film mm ID (0. mm OD) 0. µm film µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns please ask for our catalogue GC Applications 4

21 OPTIMA high performance capillary columns 0% cyanopropyl-methyl 0% phenylmethylpolysiloxane CH CH Max. temperature for isothermal operation 60 C, max. temperature for short isotherms in a temperature programme 80 C Polar Recommended for fatty acid analyses Similar phases: DB-, HP-, SP-0, OV-, Rtx-, CP-SIL 4, 007-, BP close equivalent to USP G 7 O Si (CH ) CN O Si n Analysis of FAME in porcine fat Column: OPTIMA, 0. µm film, m x 0. mm ID, max. temperature 60/80 C, Cat. No. 76. Injection volume: µl, split :40 Carrier gas: 60 kpa H Temperature: 0 C ( min) C, 0 C/min 60 C, C/min 80 C, 0 C/min 00 C, C/min 0 C, 0 C/min (0 min) Detector: FID 60 C. C 4:0 8. C 8:0. C :0 9. C 8:. C 6:0 0. C 8: 4. C 8:0. C 8:. C 0:0. C 9:0 6. C :0. C 0:0 7. C :0 4. C 0: 8. C :0. C 0: 9. C : 6. C 0:4 0. C 4:0 7. C 0:. C 4: 8. C 0:. C :0 9. C :0. C : 0. C : 4. C 6:0. C :. C 6:. C :6 6. C 7:0. C 4:0 7. C 7: 4. C 4: mv FAME standard 0 4 mv FAME in porcine fat Chromatograms courtesy of Dr. Bantleon, Mr. Leusche, Mr. Hagemann, VFG-Labor, Versmold, Germany min min m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0. µm film mm ID (0. mm OD) 0. µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns visit our web site:

22 OPTIMA high performance capillary columns % cyanopropyl-methyl 67% dimethylpolysiloxane CH CH No similar phases O Si O Si Max. temperature for isothermal operation 60 C, max. temperature for short isotherms in a temperature programme 80 C Highly polar Recommended for FAMEs, dioxins (CH ) CN n CH m Fatty acid methyl esters cis/trans C 8: (FAME) Capillary column: OPTIMA 40, 0. film, 60 m x 0. mm ID, max. temperature 60/80 C, Cat. No Sample: FAME mixture Injection volume:.0 µl, split : Carrier gas: 0 kpa H Temperature: 80 C 0 C, 0 C/min 60 C (0 min), C/min Detector: FID, 80 C 4. C 4:0 7. trans-c 8: mv. C :0 8. cis-c 8: C 8:0 9. C 8: 0 4. C 0:0 0. C 8: 8. C :0. C 8: 6. C :0. C 0:0 7. C :0. C 0: C 4:0 4. C 0: 9. C 4:. C 0: 0. C :0 6. C 0:4 6. C : 7. C 0:. C 6:0 8. C :0. C 6: 9. C : 4 4. C 7:0 0. C :. C 7:. C 4: 6. C 8: min 0 m m 0 m 0 m 60 m 0. mm ID (0.4 mm OD) 0. µm film µm film mm ID (0. mm OD) 0. µm film µm film µm film Each column is individually tested and supplied with test certificate and test chromatogram, but without fittings or ferrules. Column ends are closed with septa, and thus protected from atmospheric oxygen. Additionally, we supply the corresponding test mixture with each column. In addition to this standard programme we will be happy to supply columns custom-made to your specifications. For applications with GC capillary columns please ask for our catalogue GC Applications 6

Gas chromatography. Gas chromatography

Gas chromatography. Gas chromatography Gas chromatography Gas chromatography Basic principles of GC Page Introduction to gas chromatography 4 Brief selection guide for capillary columns 6 7 OPTIMA high performance capillary columns for GC 8

More information

Capillary columns for GC

Capillary columns for GC GC Contents Basic principles of GC 98 99 Summary of MN standard phases for GC 00 0 OPTIMA high performance capillary columns 0 0 PERMABOND capillary columns Capillary columns for special applications 4

More information

Gas chromatography. 298

Gas chromatography. 298 Gas chromatography 98 Basics The GC system 6 5 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 4 Transport of the components occurs exclusively in the mobile phase, while separation only takes place in the

More information

Table 2 Summary of groundwater field paramaters - September 2018

Table 2 Summary of groundwater field paramaters - September 2018 Table 2 Summary of groundwater field paramaters - September 2018 Well ID Sampling Event Date Measured SWL (mbtoc) BOC (mbtoc) Dissolved Oxygen (ppm) Electrical Conductivity (us/cm) ph Ox Red Potential

More information

Colonne Gascromatografiche

Colonne Gascromatografiche SGE dispone di un ampia gamma di colonne capillari per GC utilizzabili praticamente in qualunque applicazione cromatografica. Di seguito sono riportate in modo sintetico le principali caratteristiche dei

More information

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column Application Note Food Testing & Agriculture Authors Pat Sasso and Ken Lynam Agilent Technologies, Inc. Abstract

More information

HT5 - High Temperature Stationary Phase for Capillary Gas Chromatography

HT5 - High Temperature Stationary Phase for Capillary Gas Chromatography HT5 - High Temperature Stationary Phase for Capillary Gas Chromatography INTRODUCTION HT5 was the first carborane modified siloxane phase to be commercially available on fused silica capillary columns.

More information

Improving the Analysis of 37 Fatty Acid Methyl Esters

Improving the Analysis of 37 Fatty Acid Methyl Esters Application Note Food Testing Improving the Analysis of 37 Fatty Acid Methyl Esters Using Three Types of Capillary GC columns Authors Yun Zou Agilent Technologies (Shanghai) Co.Ltd, Shanghai 200131 P.R.China

More information

e-brochure 2009 CAPILLARY COLUMNS LABORATORY dex FAST-GC MegaHT High Temperature Columns Standard conventional GC columns

e-brochure 2009 CAPILLARY COLUMNS LABORATORY  dex FAST-GC MegaHT High Temperature Columns Standard conventional GC columns CAPILLAY COLUMNS Our Experience in Gas Chromatography at Your Service -1-5 -WAX -1701-17 -SE54 Staard conventional GC columns CUSTOM DEDICATED COLUMNS dex chiral columns FAST-GC MegaHT High Temperature

More information

Gas chromatography columns and accessories

Gas chromatography columns and accessories Capillary gas chromatography columns Application specific columns GC-MS columns Chiral columns High temperature columns Fast gc columns Retention gaps, press-fit connectors www.dot-red.com GC columns:

More information

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph Application Note Food Rapid Analysis of 37 FAMEs with the Agilent 88 Gas Chromatograph Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd., Shanghai 131 P. R. China Abstract An Agilent 88 GC

More information

CAPILLARY COLUMNS: 007-1

CAPILLARY COLUMNS: 007-1 CAPILLARY COLUMNS: 007-1 007-1 DIMETHYLPOLYSILOXANE Non polar Excellent thermal stability Comparable to: SE-30, OV-1,OV-101 Similar bonded phases: DB-1, DB-2887, DB-Petro, Rtx-1, Rtx-1PONA, Rtx-2887, HP-1,

More information

Evaluation of a New Ionic Liquid Stationary Phase with PEG Like Selectivity

Evaluation of a New Ionic Liquid Stationary Phase with PEG Like Selectivity Evaluation of a New Ionic Liquid Stationary Phase with PEG Like Selectivity L.M.Sidisky, G. A. Baney, J.L. Desorcie, D.L. Shollenberger, G. Serrano, and K.K. Stenerson Supelco, Div. of Sigma-Aldrich Bellefonte,

More information

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column Application Note Food Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column Authors Ingrid Van Der Meer, Yun Zou, and Gustavo Serrano Agilent

More information

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases APPLICATION NOTE Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases No. 21557 Aaron L. Lamb Thermo Fisher Scientific,

More information

3-Acetyldeoxynivalenol. 15-Acetyldeoxynivalenol

3-Acetyldeoxynivalenol. 15-Acetyldeoxynivalenol 3-Acetyldeoxynivalenol 15-Acetyldeoxynivalenol [Methods listed in the Feed Analysis Standards] 1 Simultaneous analysis of trichothecene mycotoxin by gas chromatography [Feed Analysis Standards, Chapter

More information

A comparison study of the analysis of volatile organic acids and fatty acids

A comparison study of the analysis of volatile organic acids and fatty acids Application Note Food Testing A comparison study of the analysis of volatile organic acids and fatty acids Using DB-FATWAX Ultra Inert and other WAX GC columns Author Yun Zou Agilent Technologies (Shanghai)

More information

Rapid Separation of Fatty Acid Methyl Esters

Rapid Separation of Fatty Acid Methyl Esters Application Note Food Testing & Agriculture Rapid Separation of Fatty Acid Methyl Esters Using DB-FastFAME Intuvo GC columns Author Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Shanghai 200131 P. R.

More information

Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -1301Sil MS Column

Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -1301Sil MS Column Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -101Sil MS Column By Rebecca Stevens and Chris English Abstract Chromatographic conditions were developed for a fast GC-MS glycol

More information

Highly Reproducible Detailed cis/trans FAMEs Analysis Ensured by New Optimized Rt-2560 Column Manufacturing and Application-Specific QC Test

Highly Reproducible Detailed cis/trans FAMEs Analysis Ensured by New Optimized Rt-2560 Column Manufacturing and Application-Specific QC Test Highly Reproducible Detailed cis/trans FAMEs Analysis Ensured by New Optimized Rt-2560 Manufacturing and Application-Specific QC Test By Kristi Sellers and Rebecca Stevens Restek s Rt-2560 GC column is

More information

Forensic Analysis of Blood Alcohol Concentration

Forensic Analysis of Blood Alcohol Concentration Application Note Forensics Forensic Analysis of Blood Alcohol Concentration Using the Agilent 886 GC with Agilent J&W DB BAC1 UI and Agilent J&W DB-BAC2 UI columns and the Agilent 7697A headspace sampler

More information

Column Selection for the Analysis of Fatty Acid Methyl Esters Application

Column Selection for the Analysis of Fatty Acid Methyl Esters Application Column Selection for the Analysis of Fatty Acid Methyl Esters Application Food Analysis Authors Frank David Research Institute for Chromatography President Kennedy Park B- Kortrijk, Belgium Pat Sandra

More information

INTERNATIONAL OLIVE COUNCIL

INTERNATIONAL OLIVE COUNCIL INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 33/Rev.1 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

IDENTIFICATION AND CONTROLOFRESIDUALSOLVENTS Identification and control of residual solvents EUROPEAN PHARMACOPOEIA 6.

IDENTIFICATION AND CONTROLOFRESIDUALSOLVENTS Identification and control of residual solvents EUROPEAN PHARMACOPOEIA 6. EUROPEAN PHARMACOPOEIA 6.0 2.4.24. Identification and control of residual solvents paper and wash each filter with 3 quantities, each of 15 ml, of methylenechlorider.placethecombinedfiltrateand washings

More information

DETERMINATION OF COMPOSITION OF TRIACYLGLYCEROLS AND COMPOSITION AND CONTENT OF DI-ACYLGLYCEROLS BY CAPILLARY GAS CHROMATOGRAPHY, IN VEGETABLE OILS

DETERMINATION OF COMPOSITION OF TRIACYLGLYCEROLS AND COMPOSITION AND CONTENT OF DI-ACYLGLYCEROLS BY CAPILLARY GAS CHROMATOGRAPHY, IN VEGETABLE OILS INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 32 November 2013 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS Application Note Clinical Research Authors Frank David and Bart Tienpont, Research Institute for Chromatography,

More information

VOLUNTARY MONOGRAPH. Council for Responsible Nutrition March 2006

VOLUNTARY MONOGRAPH. Council for Responsible Nutrition March 2006 VOLUNTARY MONOGRAPH Council for Responsible Nutrition March 2006 Omega-3 DHA Omega-3 EPA Omega-3 DHA & EPA DEFINITION Omega-3 fatty acids, EPA and DHA, consist of the all cis forms of 5, 8, 11, 14, 17-eicosapentaenoic

More information

REPORT of ANALYSIS OF PESTICIDE RESIDUES. in SOFT DRINK SAMPLES sent by DIRECTORATE GENERAL OF HEALTH SERVICES, NEW DELHI to CFL, Mysore

REPORT of ANALYSIS OF PESTICIDE RESIDUES. in SOFT DRINK SAMPLES sent by DIRECTORATE GENERAL OF HEALTH SERVICES, NEW DELHI to CFL, Mysore REPORT of ANALYSIS OF PESTICIDE RESIDUES in SOFT DRINK SAMPLES sent by DIRECTORATE GENERAL OF HEALTH SERVICES, NEW DELHI to CFL, Mysore Dated AUGUST 14, 2003 FROM CENTRAL FOOD LABORATORY AT CENTRAL FOOD

More information

CERTIFICATE OF ANALYSIS

CERTIFICATE OF ANALYSIS EM1008926 False CERTIFICATE OF ANALYSIS : 1 of 8 Laboratory : Environmental Division Melbourne Contact : MR JUAN CASTILLO Contact : Steven McGrath Address : 34-36 Lakeside Ave. Address : 4 Westall Rd Springvale

More information

Investigating GCxGC separations using selective column chemistry and compound derivatization pairings for common metabolomics chemical compounds

Investigating GCxGC separations using selective column chemistry and compound derivatization pairings for common metabolomics chemical compounds Investigating GCxGC separations using selective column chemistry and compound derivatization pairings for common metabolomics chemical compounds Julie Kowalski, Michelle Misselwitz and Jack Cochran Restek

More information

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds

Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds Independent Column Temperature Control Using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds Application Note Food and Flavors Author Frank David Research Institute for

More information

dex ~ r~,~ c hiral columns Mega High Temperature columns PRESS-AT connectors since

dex ~ r~,~ c hiral columns Mega High Temperature columns PRESS-AT connectors since dex ~ r~,~ c hiral columns Mega High Temperature columns since 1980 PRESS-AT connectors GC Column Dimensions Available &Jfl~,ru 17 IllbUII" CAPlL...l...AAY COLUIVINS L...AE30A.D>J"'"ClAY 10 0.05 mm 0.10

More information

Chiral Gas Chromatography

Chiral Gas Chromatography GC Chiral Gas Chromatography Chiral phases overview poster included Stop scrying start planning LIPDEX and HYDRDEX cyclodextrin phases for high enantiomeric recognition MACHEREY-NAGEL A pioneer in chiral

More information

Tenofovir disoproxil fumarate (Tenofoviri disoproxili fumaras)

Tenofovir disoproxil fumarate (Tenofoviri disoproxili fumaras) C 19 H 30 N 5 O 10 P. C 4 H 4 O 4 Relative molecular mass. 635.5. Chemical names. bis(1-methylethyl) 5-{[(1R)-2-(6-amino-9H-purin-9-yl)-1-methylethoxy]methyl}-5-oxo-2,4,6,8-tetraoxa-5-λ 5 - phosphanonanedioate

More information

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants

Selectivity Comparison of Agilent Poroshell 120 Phases in the Separation of Butter Antioxidants Selectivity Comparison of Agilent Poroshell 1 Phases in the Separation of Butter Antioxidants Application Note Food Testing & Agriculture Author Rongjie Fu Agilent Technologies (Shanghai) Co. Ltd. Abstract

More information

Rapid and Robust Detection of THC and Its Metabolites in Blood

Rapid and Robust Detection of THC and Its Metabolites in Blood Rapid and Robust Detection of THC and Its Metabolites in Blood Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. Santa Clara CA 95051 USA Abstract A robust method

More information

Secrets to Successful GC-MS/MS Pesticide Method Development

Secrets to Successful GC-MS/MS Pesticide Method Development Secrets to Successful GC-MS/MS Pesticide Method Development Katerina Mastovska Nutritional Chemistry and Food Safety Covance Laboratories QuEChERS vs. Traditional Methods Quick, Easy, Cheap, Effective,

More information

Lutein Esters from Tagetes Erecta

Lutein Esters from Tagetes Erecta Residue Monograph prepared by the meeting of the Joint FAO/WHO Expert Committee on Food Additives (JECFA), 82 nd meeting 2016 Lutein Esters from Tagetes Erecta This monograph was also published in: Compendium

More information

Columns for HPLC. Columns for HPLC

Columns for HPLC. Columns for HPLC Basic principles of preparative HPLC Basically, preparative HPLC follows the same rules as analytical scale chromatography. However, there are important differences in the aims of the two techniques. In

More information

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases

Phenyl-Hexyl. UHPLC Columns. lternate, complementary selectivity to C18 and C8 bonded phases Phenyl-Hexyl UHPLC Columns lternate, complementary selectivity to A C8 and C8 bonded phases articularly recommended for P compounds containing aromatic groups xcellent bonded phase stability for E durable,

More information

using the Agilent 7696A Sample Prep

using the Agilent 7696A Sample Prep Automated Clean-up for Mineral Oil (Hydrocarbon Oil Index) Analysis using the Agilent 7696A Sample Prep WorkBench Application Note Automated Sample Preparation Authors Frank David, Karine Jacq, and Bart

More information

Off-line supercritical fluid extraction/gas chromatography-mass spectrometry analysis of pesticides in fish

Off-line supercritical fluid extraction/gas chromatography-mass spectrometry analysis of pesticides in fish PO-CON1815E Off-line supercritical fluid extraction/gas chromatography-mass spectrometry analysis of pesticides in fish ASMS 2018 William Hedgepeth 1, Tairo Ogura 1, Riki Kitano 1, June Black 2 1 Shimadzu

More information

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path

Analyze Drugs of Abuse with Agilent J&W Ultimate Plus Tubing in an Inert Flow Path Analyze Drugs of Abuse with J&W Ultimate Plus Tubing in an Inert Flow Path Application Note Forensic Toxicology Author Ngoc A Dang Technologies, Inc. Abstract J&W Ultimate Plus deactivated fused silica

More information

Roc On with These Dependable LC Columns

Roc On with These Dependable LC Columns Roc LC s Conventional column built to be the cornerstone for your lab; pressure rated for any 400 bar system. Solid and reliable delivers the peak shape, reproducibility, ruggedness, and performance you

More information

AppNote 4/2002. Stir Bar Sorptive Extraction: Capacity and Competition Effects KEYWORDS ABSTRACT. SBSE, SPME, capacity

AppNote 4/2002. Stir Bar Sorptive Extraction: Capacity and Competition Effects KEYWORDS ABSTRACT. SBSE, SPME, capacity AppNote 4/22 Stir Bar Sorptive Extraction: Capacity and Competition Effects Edward Pfannkoch, Jacqueline Whitecavage Gerstel, Inc., 71 Digital Drive, Suite J, Linthicum, MD 219, USA Andreas Hoffmann Gerstel

More information

GC Capillary Columns. SGE GC Capillary Column innovation, manufacture and selection % Dimethyl Polysiloxane BP1 83 BP1 PONA 84 BPX1 84

GC Capillary Columns. SGE GC Capillary Column innovation, manufacture and selection % Dimethyl Polysiloxane BP1 83 BP1 PONA 84 BPX1 84 SGE GC Capillary Column innovation, manufacture and selection 74-82 100% Dimethyl Polysiloxane BP1 83 BP1 PONA 84 BPX1 84 GC Capillary Columns Tel: +98 (21) 8891 6706, Fax: +98 (21) 8889 7273, E-mail:

More information

CORESTA RECOMMENDED METHOD NÄ 9

CORESTA RECOMMENDED METHOD NÄ 9 CORESTA RECOMMENDED METHOD NÄ 9 DETERMINATION OF NICOTINE IN CIGARETTE FILTERS BY GAS CHROMATOGRAPHIC ANALYSIS (April 2009) 0. INTRODUCTION In 2001 the CORESTA Routine Analytical Chemistry Sub-Group was

More information

Organic Chemistry. Chapter 23. Hill, Petrucci, McCreary & Perry 4 th. Ed. Alkane to Substituent Group methane CH 4 methyl CH 3

Organic Chemistry. Chapter 23. Hill, Petrucci, McCreary & Perry 4 th. Ed. Alkane to Substituent Group methane CH 4 methyl CH 3 hapter 23 rganic hemistry ill, Petrucci, Mcreary & Perry 4 th Ed. Alkane to Substituent Group methane 4 methyl 3 ethane 3 3 ethyl 3 2 propane 3 2 3 propyl 3 2 2 isopropyl ( 3 ) 2 or 3 3 butyl 3 2 2 2 butane

More information

E17 ETHYLCELLULOSE. Revision 3 Stage 4

E17 ETHYLCELLULOSE. Revision 3 Stage 4 00-205PDG.pdf 2 E7 ETHYLCELLULOSE Revision 3 Stage 4 3 4 5 6 7 8 9 0 2 3 4 5 6 7 8 9 20 2 22 23 24 25 26 27 28 29 30 3 32 33 34 35 36 37 DEFINITION Ethylcellulose is a partly O-ethylated cellulose. It

More information

Ultra Columns. also available. ordering note

Ultra Columns. also available. ordering note : 3μm or 5μm particles; 100Å pore size Our broadest selection of stationary phases, including unique phases. High density bondings, for maximum retention. High-purity, Type B silica gives excellent peak

More information

Supplement of Evaluation of NO + reagent ion chemistry for online measurements of atmospheric volatile organic compounds

Supplement of Evaluation of NO + reagent ion chemistry for online measurements of atmospheric volatile organic compounds Supplement of Atmos. Meas. Tech., 9, 2909 2925, 2016 http://www.atmos-meas-tech.net/9/2909/2016/ doi:10.5194/amt-9-2909-2016-supplement Author(s) 2016. CC Attribution 3.0 License. Supplement of Evaluation

More information

CAPILLARY COLUMNS. Cap. simply Quality

CAPILLARY COLUMNS. Cap. simply Quality CAPILLARY COLUMNS Cap simply Quality CAPILLARY COLUMNS GREYHOUND - setting the standard Range As a trusted name in the supply of chromatography consumables and certified reference standards, Greyhound

More information

Development of non-aqueous single stage derivatisation method for the determination of putrescine and cadaverine using GC-MS

Development of non-aqueous single stage derivatisation method for the determination of putrescine and cadaverine using GC-MS Cent. Eur. J. Chem. 6(2) 2008 229 236 DOI: 10.2478/s11532-008-0007-6 Central European Journal of Chemistry Development of non-aqueous single stage derivatisation method for the determination of putrescine

More information

NEW! 200 m GC Columns for Detailed Analysis of cis/trans FAME Isomers

NEW! 200 m GC Columns for Detailed Analysis of cis/trans FAME Isomers Order: 00--00 (U.S.) -- (Global) NEW! 00 m GC Columns for Detailed Analysis of cis/trans FAME Isomers Leonard M. Sidisky, R&D Manager; and Michael D. Buchanan, Product Manager mike.buchanan@sial.com Over

More information

ANALYSIS OF -HYDROXYBUTYRATE (GHB) AND -BUTYROLACTONE (GBL) IN LIQUIDS PERFORMED AT NATIONAL LABORATORY OF FORENSIC SCIENCE (SKL), SWEDEN

ANALYSIS OF -HYDROXYBUTYRATE (GHB) AND -BUTYROLACTONE (GBL) IN LIQUIDS PERFORMED AT NATIONAL LABORATORY OF FORENSIC SCIENCE (SKL), SWEDEN ANALYSIS OF -HYDROXYBUTYRATE (GHB) AND -BUTYROLACTONE (GBL) IN LIQUIDS PERFORMED AT NATIONAL LABORATORY OF FORENSIC SCIENCE (SKL), SWEDEN Per LUNDQUIST National Laboratory of Forensic Science, Linköping,

More information

Idaho Bureau of Laboratories 2220 Old Penitentiary Road Boise, Idaho Phone: (208) Fax: (208)

Idaho Bureau of Laboratories 2220 Old Penitentiary Road Boise, Idaho Phone: (208) Fax: (208) October 13, 2010 Report To Todd Vitek 507 W Bird Ave Nampa, ID 83686 Sample Submitted by Walk-In Client - Private RE: Workorder E100700201 072710 Profile WA-General Environmental Enclosed are the analytical

More information

Comparison of PAL SPME Fibers with established Fibers

Comparison of PAL SPME Fibers with established Fibers Comparison of PAL SPME Fibers with established Fibers Abstract The performance of SPME fibers is an essential aspect of an analytical method. This study shows the comparison of three different SPME fibers

More information

ANALYSIS OF CARNAUBA WAX ON BPX5

ANALYSIS OF CARNAUBA WAX ON BPX5 FOO 01 - FOOD ANALYSIS OF CARNAUBA WAX ON BPX5 CARNAUBA WAX Column Part No.: 054118 Phase: BPX5, 0.25 µm film 12 m x 0.32 mm ID Initial Temp: 100 C, 2 min Rate: 10 C/min Final Temp: Detector: Sensitivity:

More information

Type Analysis of Citrus Essential Oils by Multidimensional Supercritical Fluid Chromatography/ Gas Chromatography

Type Analysis of Citrus Essential Oils by Multidimensional Supercritical Fluid Chromatography/ Gas Chromatography 25 Type Analysis of Citrus Essential Oils by Multidimensional Supercritical Fluid Chromatography/ Gas Chromatography Takashi YARITA, Akira NoMURA and Yoshiyuki HoRIMOTo Department of Analytical Chemistry,

More information

Ultra Columns HPLC COLUMNS

Ultra Columns HPLC COLUMNS Acetaminophen, Narcotic Analgesics.............519 Aldehydes, Ketones...........................500 Beclomethasone.............................523 Corticosteroids..............................529 Explosives..................................501

More information

Eurospher II the logical choice. Physical properties of Eurospher II silica gel: Silica gel: ultra pure, > % Metal content: < 10 ppm

Eurospher II the logical choice. Physical properties of Eurospher II silica gel: Silica gel: ultra pure, > % Metal content: < 10 ppm HPLC SMB smometry Euroline Eurospher II Eurospher II the logical choice After more than years on the market, our Eurospher stationary phase has established a reputation for being a first-class packing

More information

FATTY ACID PROFILING BY GAS CHROMATOGRAPHY FOR THE SHERLOCK MIS

FATTY ACID PROFILING BY GAS CHROMATOGRAPHY FOR THE SHERLOCK MIS FATTY ACID PROFILING BY GAS CHROMATOGRAPHY FOR THE SHERLOCK MIS Traditional gas chromatography of complex mixtures of compounds requires precision on the part of the chromatography equipment and considerable

More information

Catalog No. : Lot No. : Container Size : 300 µl Autosampler Vial Pkg Amt : 150 µl

Catalog No. : Lot No. : Container Size : 300 µl Autosampler Vial Pkg Amt : 150 µl Catalog No. : 30731 Lot No. : 052764 Description : DHA PONA Standard Container Size : 300 µl Autosampler Vial Pkg Amt : 150 µl Expiration Date : 12 months unopened* Storage : 4 or colder *From date of

More information

DEVELOPMENT AND VALIDATION OF THE GC METHOD FOR THE QUANTITATIVE DETERMINATION OF SEMI-VOLATILE SOLVENTS IN THE PHARMACEUTICAL SUBSTANCE BOSENTAN

DEVELOPMENT AND VALIDATION OF THE GC METHOD FOR THE QUANTITATIVE DETERMINATION OF SEMI-VOLATILE SOLVENTS IN THE PHARMACEUTICAL SUBSTANCE BOSENTAN Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 71 No. 6 pp. 1107ñ1113, 2014 ISSN 0001-6837 Polish Pharmaceutical Society DEVELOPMENT AND VALIDATION OF THE GC METHOD FOR THE QUANTITATIVE DETERMINATION

More information

Application Note. Authors. Abstract. Petrochemical

Application Note. Authors. Abstract. Petrochemical Fast screening of impurities in biodiesel using the Agilent 160 Infinity Analytical SFC System in combination with evaporative light scattering detection Application Note Petrochemical Authors Maria Rambla-Alegre,

More information

GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil

GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil GC-µECD Analysis and Confirmation of Contract Laboratory Protocol Pesticides in Olive Oil Application Note Food Safety and Environmental Markets Authors Jennifer McCulley and Ken Lynam Agilent Technologies,

More information

CORESTA Recommended Method No. 84

CORESTA Recommended Method No. 84 Cooperation Centre for Scientific Research Relative to Tobacco E-Vapour Sub-Group CORESTA Recommended Method No. 84 DETERMINATION OF GLYCERIN, PROPYLENE GLYCOL, WATER, AND NICOTINE IN THE AEROSOL OF E-CIGARETTES

More information

A Rapid Method for Trans-Fatty Acid Determination Using a Single Capillary GC Seiichi Shirasawa, Akiko Sasaki, Yasue Saida and Chiemi Satoh

A Rapid Method for Trans-Fatty Acid Determination Using a Single Capillary GC Seiichi Shirasawa, Akiko Sasaki, Yasue Saida and Chiemi Satoh Journal of Oleo Science Copyright 2007 by Japan Oil Chemists Society A Rapid Method for Trans-Fatty Acid Determination Using a Single Capillary GC Seiichi Shirasawa, Akiko Sasaki, Yasue Saida and Chiemi

More information

CORESTA RECOMMENDED METHOD N 8

CORESTA RECOMMENDED METHOD N 8 CORESTA RECOMMENDED METHOD N 8 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARETTES BY GAS CHROMATOGRAPHIC ANALYSIS (August 1991) 1. FIELD OF APPLICATION The method is applicable to the particulate

More information

Sepax Technologies, Inc.

Sepax Technologies, Inc. Sepax Technologies, Inc. Sepax Technologies, Inc. develops and manufactures products in the area of chemical and biological separations, biosurfaces and proteomics. Sepax product portfolio includes ) liquid

More information

Pesticide and Chemical Residues. Nonalcoholic Beverages & Concentrates. Environmental & Water Standards. Page 1 of 10. August 2014 John Barron

Pesticide and Chemical Residues. Nonalcoholic Beverages & Concentrates. Environmental & Water Standards. Page 1 of 10. August 2014 John Barron Pesticide and Chemical Residues Nonalcoholic Beverages & Concentrates Environmental & Water Standards Page 1 of 10 Reagents and Standards for Pesticide and Chemical Residues As per Official Methods of

More information

Volatile Organic Reference Materials for U.S. EPA Method 8260

Volatile Organic Reference Materials for U.S. EPA Method 8260 Volatile Organic Reference Materials for U.S. EPA Method 8260 Prepared in the fewest solutions compatible with maximum shelf life; fewer ampuls to open means fewer errors in preparation. Ready-to-use mixes

More information

AS Application Note 1602

AS Application Note 1602 Determination of the fatty acid composition in refined oils and fats by alkaline transesterification by the ASAN 1602 Status: February 2018 Page 1 / 12 Introduction Animal and vegetable fats play a key

More information

TO-3 BTEX and TPH. Table 13.1a Summary of QC Criteria for TO-3 BTEX Oxygenates

TO-3 BTEX and TPH. Table 13.1a Summary of QC Criteria for TO-3 BTEX Oxygenates TO-3 BTEX and TPH Method TO-3 uses cryogenic trapping and a gas chromatograph with a flame ionization detector (FID) to measure volatile organic compounds collected in Summa Canisters or Tedlar Bags. This

More information

DETERMINATION OF FATTY ACIDS IN EDIBLE OILS BY CAPILARY GC

DETERMINATION OF FATTY ACIDS IN EDIBLE OILS BY CAPILARY GC DETERMINATION OF FATTY ACIDS IN EDIBLE OILS BY CAPILARY GC Vesna Kostik 1 University Goce Delcev Stip Faculty of Medicine Department of Pharmacy 1 WHY FATTY ACID (FA) ANALYSIS IN EDIBLE OILS The content

More information

Chromatography Vacuum Ultraviolet Spectroscopy

Chromatography Vacuum Ultraviolet Spectroscopy Application Note Differentiation and Determination Differentiation and Determination of Fatty Acid Methyl of Fatty Esters Acid by Gas Methyl Chromatography Esters by Vacuum Gas Ultraviolet Spectroscopy

More information

SeQuant ZIC -HILIC For all who expect more...

SeQuant ZIC -HILIC For all who expect more... SeQuant ZIC -ILIC For all who expect more... The better choice for PLC and LC/MS of all types of polar and hydrophilic compounds EMD Millipore Corp. is a subsidiary of Merck KGaA, Darmstadt, Germany Poor

More information

HYDROXYPROPYLCELLULOSE, LOW SUBSTITUTED Stage 4, Revision 1 CP: USP BRIEFING NOTE

HYDROXYPROPYLCELLULOSE, LOW SUBSTITUTED Stage 4, Revision 1 CP: USP BRIEFING NOTE 002-0901PDG.pdf HYDROXYPROPYLCELLULOSE, LOW SUBSTITUTED Stage 4, Revision 1 CP: USP BRIEFING NOTE Compared to the Stage 4, document the following changes are proposed: 1. Assay: a determination of the

More information

The Raptor HILIC-Si Column

The Raptor HILIC-Si Column The Raptor HILIC-Si Column With Raptor LC columns, Restek chemists became the first to combine the speed of superficially porous particles (also known as SPP or core-shell particles) with the resolution

More information

Chromatogram Search Tool Search by compound name, synonym, CAS # or keyword Pinnacle II Columns

Chromatogram Search Tool Search by compound name, synonym, CAS # or keyword   Pinnacle II Columns : 3μm or 5μm particles; 110Å pore size Silica at, for total control of quality and reproducibility. Excellent replacement for the original Hypersil material. Physical and chromatographic properties similar

More information

Pesticides. Summary of Features & Benefits: Commercial Benefits. Technical Benefits

Pesticides. Summary of Features & Benefits: Commercial Benefits. Technical Benefits Pesticides Pesticides Summary of Features & Benefits: Commercial Benefits Ready to use (dilute for use as calibration and/or quality control standards) Extensive range of organic compound mixes and single

More information

DEVELOPMENT AND VALIDATION OF GAS CHROMATOGRAPHY METHODS FOR THE CONTROL OF VOLATILE IMPURITIES IN THE PHARMACEUTICAL SUBSTANCE DUTASTERIDE

DEVELOPMENT AND VALIDATION OF GAS CHROMATOGRAPHY METHODS FOR THE CONTROL OF VOLATILE IMPURITIES IN THE PHARMACEUTICAL SUBSTANCE DUTASTERIDE Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 74 No. 5 pp. 1343ñ1351, 2017 ISSN 0001-6837 Polish Pharmaceutical Society ANALYSIS DEVELOPMENT AND VALIDATION OF GAS CHROMATOGRAPHY METHODS FOR THE CONTROL

More information

Analytical and Preparative SFC Columns

Analytical and Preparative SFC Columns Analytical and Preparative SFC Columns Best Performance for Supercritical Fluid Chromatography Sepax SFC-Cyano... Sepax SFC-Amino... Sepax SFC-Pyridine... Sepax SFC-SCX... 5 Sepax SFC-Diol... 6 Sepax SFC-Silica...

More information

SUPPLEMENTARY INFORMATION. Materials and Methods

SUPPLEMENTARY INFORMATION. Materials and Methods GSA DATA REPOSITORY 2013030 I. Melendez et al. SUPPLEMENTARY INFORMATION Materials and Methods Sample Description A calcareous concretion containing a well preserved invertebrate fossil was collected from

More information

CHROMATOGRAPHIC ANALYSIS OF ORGANIC COMPOUNDS ON IMPREGNATED CHEMICALLY BONDED STATIONARY PHASES. PART 1

CHROMATOGRAPHIC ANALYSIS OF ORGANIC COMPOUNDS ON IMPREGNATED CHEMICALLY BONDED STATIONARY PHASES. PART 1 ACTA CHROMATOGRAPHICA, NO. 17, 2006 CHROMATOGRAPHIC ANALYSIS OF ORGANIC COMPOUNDS ON IMPREGNATED CHEMICALLY BONDED STATIONARY PHASES. PART 1 G. Grygierczyk Institute of Chemistry, Silesian University,

More information

Hyper-fast & Super-rugged

Hyper-fast & Super-rugged Hyper-fast & Super-rugged UHPLC & HPLC columns FALL 2013 PRODUCT CATALOG Hyper-fast & Super-rugged UHPLC & HPLC columns he introduction of HALO columns, developed using innovative Fused-Core particle

More information

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS

A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS A Novel Solution for Vitamin K₁ and K₂ Analysis in Human Plasma by LC-MS/MS By Shun-Hsin Liang and Frances Carroll Abstract Vitamin K₁ and K₂ analysis is typically complex and time-consuming because these

More information

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract

F. Al-Rimawi* Faculty of Science and Technology, Al-Quds University, P.O. Box 20002, East Jerusalem. Abstract JJC Jordan Journal of Chemistry Vol. 4 No.4, 2009, pp. 357-365 Development and Validation of Analytical Method for Fluconazole and Fluconazole Related Compounds (A, B, and C) in Capsule Formulations by

More information

FLUNITRAZEPAM Latest Revision: January 24, 2006

FLUNITRAZEPAM Latest Revision: January 24, 2006 FLUNITRAZEPAM Latest Revision: January 24, 2006 1. SYNONYMS CFR: Flunitrazepam CAS #: 1622-62-4 Other Names: 5-(2-Fluorophenyl)-1,3-dihydro-1-methyl-7-nitro-2H- 1,4-benzodiazepin-2-one Flunitrax Hipnosedon

More information

METHOD 3620B FLORISIL CLEANUP

METHOD 3620B FLORISIL CLEANUP METHOD 3620B FLORISIL CLEANUP 1.0 SCOPE AND APPLICATION 1.1 Florisil, a registered trade name of U. S. Silica Co., is a magnesium silicate with basic properties. It is used to separate analytes from interfering

More information

Analysis of FAMEs Using Cold EI GC/MS for Enhanced Molecular Ion Selectivity

Analysis of FAMEs Using Cold EI GC/MS for Enhanced Molecular Ion Selectivity APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Adam Patkin PerkinElmer, Inc. Shelton, CT Analysis of FAMEs Using GC/MS for Enhanced Molecular Ion Selectivity Introduction Characterization

More information

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns

Maximizing Efficiency Using Agilent InfinityLab Poroshell 120 Columns Maximizing Efficiency Using Agilent InfinityLab Poroshell 12 Columns 1, Plates in Less Than Minutes Using Coupled Column Technology Application Note Food, Environmental, Chemical, Pharmaceutical Authors

More information

AppNote 3/2001. Elimination of Polar Matrix Components Prior to GC Analysis using Stir Bar Sorptive Extraction (SBSE)

AppNote 3/2001. Elimination of Polar Matrix Components Prior to GC Analysis using Stir Bar Sorptive Extraction (SBSE) AppNote 3/2001 Elimination of Polar Matrix Components Prior to GC Analysis using Stir Bar Sorptive Extraction (SBSE) Edward Pfannkoch, Jacqueline Whitecavage Gerstel Inc., 1510 Caton Center Drive, Suite

More information

Koji URAKAMI,*,a Chiaki KOBAYASHI, a Yoshihiko MIYAZAKI, a Koji NISHIJIMA, a Yoshinobu YOSHIMURA, a and Keiji HASHIMOTO b

Koji URAKAMI,*,a Chiaki KOBAYASHI, a Yoshihiko MIYAZAKI, a Koji NISHIJIMA, a Yoshinobu YOSHIMURA, a and Keiji HASHIMOTO b September 2000 Chem. Pharm. Bull. 48(9) 1299 1303 (2000) 1299 Degradation Products Generated by Sonication of Benzyl Alcohol, a Sample Preparation Solvent for the Determination of Residual Solvents in

More information

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Dr. Syed Tufail Hussain Sherazi E-mail: tufail_sherazi@yahoo.com National Center of Excellence in

More information

Organic Chemistry Laboratory Fall Lecture 3 Gas Chromatography and Mass Spectrometry June

Organic Chemistry Laboratory Fall Lecture 3 Gas Chromatography and Mass Spectrometry June 344 Organic Chemistry Laboratory Fall 2013 Lecture 3 Gas Chromatography and Mass Spectrometry June 19 2013 Chromatography Chromatography separation of a mixture into individual components Paper, Column,

More information

176 YMC Chiral Columns

176 YMC Chiral Columns 176 YMC Chiral Columns YMC Chiral Columns 177 Chiral Columns Content YMC Chiral NEA(R)(S)... 178-181 YMC Chiral CD BR... 182-185 Ordering Information... 186 HPLC Columns for Optical Isomer Separation Introduction

More information

DET REPORT NO. 69 JUNE 2015

DET REPORT NO. 69 JUNE 2015 1.) THINKING BEYOND THE NPD BOX - INEXPENSIVE CONVERSION OF NPD EQUIPMENT TO MULTIPLE MODES OF SELECTIVE GC DETECTION. 2.) GC-CCID DIFFERENTIATION BETWEEN SATURATE VS. UNSATURATE AND MONO-UNSATURATE VS.

More information

Organic Chemistry Diversity of Carbon Compounds

Organic Chemistry Diversity of Carbon Compounds Organic Chemistry Diversity of Carbon Compounds Hydrocarbons The Alkanes The Alkenes The Alkynes Naming Hydrocarbons Cyclic Hydrocarbons Alkyl Groups Aromatic Hydrocarbons Naming Complex Hydrocarbons Chemical

More information

An Introduction and Overview on Comprehensive Two-Dimensional Gas Chromatography (GCxGC): New Opportunities for Unresolved Complex Mixtures

An Introduction and Overview on Comprehensive Two-Dimensional Gas Chromatography (GCxGC): New Opportunities for Unresolved Complex Mixtures An Introduction and Overview on Comprehensive Two-Dimensional Gas Chromatography (GCxGC): New Opportunities for Unresolved Complex Mixtures Daniela Cavagnino GC Product Manager Thermo Fisher Scientific,

More information