In Vitro Magnetic Resonance Imaging Evaluation of Ossicular Implants at 3 T
|
|
- Felix Caldwell
- 5 years ago
- Views:
Transcription
1 Otology & Neurotology 33:871Y877 Ó 2012, Otology & Neurotology, Inc. In Vitro Magnetic Resonance Imaging Evaluation of Ossicular Implants at 3 T *Frank G. Shellock, Lauren N. Meepos, Matthew R. Stapleton, and Sam Valencerina *Keck School of Medicine, University of Southern California and Institute for Magnetic Resonance Safety, Education, and Research, Los Angeles, California; ÞUniversity of Pennsylvania, Philadelphia, Pennsylvania; þcreighton University, School of Medicine, Omaha, Nebraska; and University of Southern California, University Hospital, Los Angeles, California, U.S.A. Hypothesis: Ossicular implants made from metallic materials may be acceptable or pose hazards for patients referred for magnetic resonance imaging (MRI) examinations, depending on the outcome of proper MRI testing procedures. Background: Using a 3-T MR system, 2 ossicular implants were tested for magnetic field interactions, heating, and artifacts. Materials and Methods: Two different ossicular implants (Stainless Steel/Fluoroplastic Sanna-Type Piston [6 mm in length] and the Offset ALTO Total Prosthesis [15 mm in length, titanium/ silicone]; Grace Medical, Memphis, TN, USA) were selected for testing, which represented the largest metallic mass and materials with the highest magnetic susceptibilities, with the intent of applying the MRI findings to other ossicular implants. The implants were evaluated at 3-T for magnetic field interactions, heating, and artifacts using standard previously described techniques. Results: Each ossicular implant showed relatively minor magnetic field interactions that will not be associated with movement in situ. Heating was not excessive (highest temperature change, e1.6-c; background temperature change, 1.5-C). Artifacts, although relatively small, may create issues for diagnostic imaging if the area of interest is in the same area or close to these ossicular implants. Conclusion: The results of this investigation demonstrated that it would be acceptable (i.e., MR conditional using current terminology) for patients with these ossicular implants to undergo MRI examinations at 3 T or less. In consideration of the materials and dimensions of the implants that underwent testing, these findings pertain to many other similar ossicular implants from the same manufacturer. Key Words: ArtifactsVImplantsV Magnetic resonance imagingvsafetyvspecific absorption rate. Otol Neurotol 33:871Y877, A variety of prostheses may be used in ossicular reconstruction procedures with the goal of improving or restoring hearing. Ossicular or middle ear implants are made from either nonmetallic or metallic materials. Notably, the presence of a metallic ossicular implant in an individual referred for a magnetic resonance imaging (MRI) procedure must be given careful consideration owing to the possible issues that exist, including those related to magnetic field interactions, MRI-related heating, and artifacts (1,2). This is even more important given the increased use of 3-T MR systems, which is the highest-field-strength magnet type used in the clinical setting. Accordingly, there have been many reports Address correspondence and reprint requests to Frank G. Shellock, Ph.D., Institute for Magnetic Resonance Safety, Education, and Research, 7511 McConnell Avenue, Los Angeles, CA 90045; frank.shellock@gte.net Supported by an unrestricted educational grant from Grace Medical, Memphis, Tennessee, U.S.A. directed toward addressing these issues for ossicular implants (3Y23). To date, several MR-unsafe ossicular implants exist, necessitating the careful pre-mri screening of patients to identify these contraindicated devices (2,21,22,24). With the continued development of ossicular implants and because many have yet to undergo proper MRI testing, it is necessary to conduct in vitro testing, particularly at 3 T (i.e., worst-case conditions for clinical MRI) to characterize MRI issues for these devices. Importantly, in consideration of the close anatomic relationship for these metallic implants in the temporal bone, it is necessary to know whether it is hazardous to patients with metallic middle ear implants regarding displacement and a rise in temperature. The presence of artifacts may also be problematic and compromise the diagnostic use of MRI. Therefore, this investigation evaluated the MRI issues (i.e., magnetic field interactions, MRI-related heating, 871
2 872 F. G. SHELLOCK ET AL. and artifacts) at 3 T for 2 different metallic ossicular implants. In consideration of the materials and dimensions of the implants that underwent testing, these findings pertain to many other similar ossicular implants from the same manufacturer that have smaller sizes and are made from materials with lower magnetic susceptibilities (25), thus expanding the MRI information for many other devices. This same strategy of MRI testing for implants has been successfully applied to aneurysm clips and hemostatic clips (26,27). METHODS Ossicular Implants Two different ossicular implants (Stainless Steel/Fluoroplastic Sanna-Type Piston [6 mm in length] and Offset ALTO Total Prosthesis [15 mm in length, titanium/silicone]; Grace Medical, Memphis, TN, USA) (Figs. 1 and 2) were selected for testing because they represented the largest metallic masses and sizes and they had the highest magnetic susceptibility values (i.e., based on material information for all items) (25) among many other similar ossicular implants from the same manufacturer listed in the Appendix. Accordingly, for the other ossicular FIG. 2. The Offset ALTO Total Prosthesis, 15 mm in length, titanium/silicone that underwent MRI testing at 3 T. implants from the same manufacturer, these items were made from materials with lower magnetic susceptibilities than stainless steel or titanium and/or had smaller dimensions. Magnetic Field Interactions Each ossicular implant was evaluated for translational attraction and torque in association with a 3-T MR system (Excite, Software G B; General Electric Healthcare, Milwaukee, WI, USA; active-shielded, horizontal field scanner). FIG. 1. The Stainless Steel/Fluoroplastic Sanna-Type Piston, 6 mm in length, which underwent MRI testing at 3 T. Translational Attraction To evaluate translational attraction for the ossicular implants, the deflection angle technique was used according to previously described methods (26Y28). Each ossicular implant was connected to a test fixture to determine the deflection angle in the 3-T MR system. The test fixture incorporated a protractor with 1-degree graduated markings (26Y28). The test sample was suspended on the apparatus by a lightweight string (20 cm in length; weight, G1% of the weight of each implant) that was fixed at the 0-degree indicator of the protractor. Deflection angles for each ossicular implant were assessed at the point of the highest spatial magnetic gradient for the 3-T MR system (26Y28). The highest patient accessible spatial gradient magnetic field for the 3-T MR scanner used in this investigation is 720 G/cm and occurs at an off-axis position that is 74 cm from isocenter of the MR system (26,27,29). The maximum deflection angle from the vertical direction to the nearest 1 degree was measured 3 times for each ossicular implant, and an average value was calculated (26Y28).
3 OSSICULAR IMPLANTS AND 3-T MRI 873 Torque Torque for each ossicular implant in association with exposure to the 3-T MR system was determined using a previously described, qualitative assessment technique (26,27). This involved the use of a flat plastic device with a millimeter grid (26,27). Each ossicular implant was placed on the test apparatus in an orientation that was 45 degrees relative to the static magnetic field of the 3-T MR system. The test apparatus was then positioned in the center of the scanner, where the effect of torque is the greatest (26,27). Each implant was carefully observed for possible alignment or rotation relative to the 3- T static magnetic field. The ossicular implant was then moved 45 degrees relative to its previous position and observed for alignment or rotation. This process was repeated to encompass a full 360-degree rotation of positions for each ossicular implant. The following previously described, qualitative scale was applied to the results (26,27): 0, no torque; +1, mild or low torque, the ossicular implant slightly changed orientation but did not align to the magnetic field; +2, moderate torque, the ossicular implant aligned gradually to the magnetic field; +3, strong torque, the ossicular implant showed rapid and forceful alignment to the magnetic field; +4, very strong torque, the ossicular implant showed very rapid and very forceful alignment to the magnetic field. MRI-Related Heating Magnetic resonance imagingyrelated heating at 3 T/128 MHz was assessed for each ossicular implant. This procedure used a plastic, ASTM phantom (i.e., which has dimensions that simulate the human head and torso) filled to a depth of 10 cm with gelled saline (i.e., 1.32 g/l NaCl plus 10 g/l polyacrylic acid in distilled water), with each implant placed in a position in the phantom where there was a highly uniform electric field tangential to the implant, ensuring extreme radiofrequency (RF) heating conditions for this experimental setup (i.e., based on an analysis of the ASTM phantom and the MRI conditions used for this assessment) (26,27,30). A relatively high level of RF energy was applied during the MRI-related heating experiment, as previously described (26,27). Because this experimental setup lacks blood flow, it simulates an extreme condition used to assess MRI-related heating for the ossicular implants. Temperature Recording System and Placement of Thermometry Probes Temperature measurements were obtained using a fluoroptic thermometry system (Model 3100; LumaSense Technologies, Santa Clara, CA, USA). The fluoroptic thermometry probes (0.5 mm in diameter) were positioned on each ossicular implant to record representative temperatures, as follows: Probe no. 1, sensor portion of the probe placed in contact with one end of the implant; Probe no. 2, sensor portion of the probe placed in contact with opposite end of the implant; Probe no. 3, sensor portion of the probe placed in contact with middle portion of the implant. The positions of the thermometry probes were inspected and verified immediately before and after each MRIrelated heating experiment. MRI Conditions Magnetic resonance imaging was performed at 3 T/128 MHz (Excite, Software G B; General Electric Healthcare). The body RF coil was used to transmit RF energy. Magnetic resonance imaging parameters were selected to generate a relatively high level of RF energy (26,27), producing an MR system reported, whole-body-averaged specific absorption rate (SAR) of 2.9 W/kg for 15 minutes. The landmark position (i.e., the center position or anatomic region for the MRI procedure) and section locations were selected to encompass the entire area of each ossicular implant. Experimental Protocol Each ossicular implant was placed in the ASTM head/torso phantom at a position mid line on the left side, slightly (5 mm) below the mid depth (vertical orientation) of the gelled saline. For this particular 3-T/128-MHz MR system and experimental setup, the left side of the ASTM head/torso phantom was found to be associated with a greater temperature rise than the right side of the head/torso phantom for a given implant or device (i.e., based on pilot experiments). Therefore, each ossicular implant was placed on the left side of the ASTM head/torso phantom to yield the worst-case temperature rise for the described measurement conditions, based on previous analyses of device heating for this particular MR system (i.e., due to asymmetry in heating patterns for this phantom and MR system) (26,27). Each ossicular implant was positioned in the plastic phantom using a grid and small plastic after set up as previously described (26,27). The fluoroptic thermometry system was calibrated, and the fluoroptic thermometry probes were applied. The phantom was filled with the gelled saline and allowed to equilibrate to the environmental temperature for more than 24 hours. The MR system fan was not on during the MRI-related heating investigations. The room and MR system bore temperatures were at constant levels throughout each experimental session. After recording baseline temperatures (5 min), MRI was performed for 15 minutes with temperatures recorded at 5-second intervals. This procedure was repeated for the next ossicular implant after the gelled saline returned to thermoequilibrium facilitated by manual mixing and verified by recording temperatures at multiple positions in the phantom. The highest temperature changes recorded by the fluoroptic thermometry probes are reported for each ossicular implant. The background temperature was also recorded in the gelled salineyfilled ASTM phantom. Accordingly, the temperature change was recorded at the same position, middle temperature probe position (i.e., corresponding to the position for Probe no. 3 for the MRI-related heating test with the implant present) in the phantom in association with MRI-related heating of the gelled salineyfilled phantom without the implant present (26,27,30). To record the background temperature, a fluoroptic thermometry probe was placed in the ASTM head/torso phantom at a position mid line on the left side, slightly (5 mm) below the mid depth (vertical orientation) of the gelled saline, and recordings were obtained as described previously. Artifacts Magnetic resonance imaging artifacts were assessed at 3 T for each ossicular implant. This test was conducted with each ossicular implant attached to a plastic frame and then placed in a gadolinium-doped, saline-filled plastic phantom (26,27). Magnetic resonance imaging was conducted using a 3-T MR system (Excite, Software G B; General Electric Healthcare), a transmit/receive RF head coil, and the following pulse sequences (26,27): 1. T1-weighted, spin echo pulse sequence: repetition time, 500 milliseconds; echo time, 20 milliseconds; matrix size, ; section thickness, 10 mm; field of view, 24 cm; number of excitations, 2; bandwidth; 16 khz.
4 874 F. G. SHELLOCK ET AL. 2. Gradient echo pulse sequence: repetition time, 100 milliseconds; echo time, 15 milliseconds; flip angle, 30 degrees; matrix size, ; section thickness, 10 mm; field of view, 24 cm; number of excitations, 2; bandwidth, 16 khz. The imaging planes were oriented to encompass the long axis and short axis of each ossicular implant. The frequencyencoding direction was parallel to the plane of imaging. The image locations obtained through each ossicular implant represented the largest or worst-case artifacts (i.e., based on reviewing multiple section locations in each imaging plane for each implant), and these were selected for evaluation. Planimetry software was used to measure (accuracy and resolution + 10%) the cross-sectional area of the largest artifact size for each ossicular implant, for each pulse sequence, and for each orientation of the section location (26,27). The image display parameters (i.e., window and level settings, magnification, etc.) were carefully selected and used in a consistent manner to provide valid measurements of sizes for the artifacts (26,27). This methodology has been used in many previous reported involving the characterization of artifacts for metallic implants and permits comparison to many other implants that have undergone similar evaluations of artifacts (2,26,27). RESULTS The average deflection angle was 1 degree in each case for the each different ossicular implant and the qualitatively measured torque was 0, no torque. Findings for the MRI-related heating assessments for the ossicular implants indicated that the highest temperature changes were equal to 1.6-C or less (range, 1.5Y1.6-C). The background temperature change was 1.5-C in each case. Artifact test results are presented in Table 1. In general, the artifacts associated with the 2 different ossicular implants were observed as signal losses or voids that were small in relation to the size and shape of each implant, with the gradient echo pulse sequence producing larger artifacts than the T1-weighted, spin echo pulse sequence. Figure 3 shows examples of artifacts for the ossicular implants, as observed on the gradient pulse sequence in the view oriented to the long axis of the respective implant. DISCUSSION Magnetic Field Interactions The average deflection angle was 1 degree, in each case, for the selected ossicular implants that underwent testing at 3 T. These translational attraction findings FIG. 3. Magnetic resonance imaging artifacts associated with the 2 different ossicular implants: (A) Stainless Steel/Fluoroplastic Sanna-Type Piston and (B) Offset ALTO Total Prosthesis (gradient echo pulse sequence; repetition time/echo time, 100/15 ms; flip angle, 30 degrees; long-axis imaging plane). are considered with regard to the document from the American Society for Testing and Materials International (30), which states If the implant deflects less than 45-, then the magnetically induced deflection force is less than the force on the implant due to gravity (its weight). For this condition, it is assumed that any risk imposed by the application of the magnetically induced force is no greater than any risk imposed by normal daily activity in the Earth s gravitational field. Therefore, the 2 different ossicular implants passed this acceptance criterion relative to the 3-T MR system that was used in this investigation. The qualitatively measured torque was 0, no torque, in each case. Accordingly, these ossicular implants displayed very low magnetic field interactions and, thus, will not present a hazard to a patient in a 3-T or less MRI environment. For a metallic implant or device, the associated magnetic field interactions are dependent on the strength of the static magnetic field, the maximum spatial gradient magnetic field, the mass of the object, the shape of the object, and the magnetic susceptibility of the material(s) (1,2,22,25Y29). The 2 different ossicular implants were specifically selected in this study for MRI testing because they represented the largest metallic masses and sizes with the highest magnetic susceptibility values compared with many other similar ossicular implants from the same manufacturer (Appendix). Therefore, because these additional ossicular implants have lower masses and dimensions, along with being made from materials with lower TABLE 1. Summary of artifact sizes for the ossicular implants evaluated at 3 T Pulse sequence T1-SE T1-SE GRE GRE Stainless Steel/Fluoroplastic Sanna-Type Piston Signal void size (mm 2 ) Imaging plane Parallel (long axis) Perpendicular (short axis) Parallel (long axis) Perpendicular (short axis) Offset ALTO Total Prosthesis Signal void size (mm 2 ) Imaging plane Parallel (long axis) Perpendicular (short axis) Parallel (long axis) Perpendicular (short axis) Imaging plane is relative to each ossicular implant. GRE indicates gradient echo; T1-SE, T1-weighted, spin echo.
5 OSSICULAR IMPLANTS AND 3-T MRI 875 magnetic susceptibilities, the magnetic qualities will be inherently less relative to the 3-T MRI environment. Interestingly, some reports and documents have presented information for magnetic field interactions relative to ossicular implants using MR systems operating at 7 T and even 9.4 T (19,24,31). However, this information may not be helpful insofar as the MRI evaluations were not conducted using an MRI-relating heating assessment, or the heating test was improperly performed (i.e., not in adherence with the ASTM International guidance information) (30), which is a critical part of the proper characterization of MRI issues for implants and devices according the U.S. Food and Drug Administration (32). Furthermore, rarely is a patient with a metal implant scanned using these experimental, very-high-fieldstrength scanners. At the present time, 3 T is the highestfield-strength MR system currently used worldwide in the clinical setting. MRI-Related Heating With regard to MRI-related heating, extreme conditions were applied to determine the temperature rises for the 2 different ossicular implants. Thus, using a relatively high level of RF energy at 3 T (i.e., the MR system reported, whole-body-averaged SAR, 2.9 W/kg) with each ossicular implant placed in a worst-case position in the gelled salineyfilled phantom, the highest temperature changes ranged from 1.5 to 1.6-C. Under the same MRI experimental conditions, the background temperature change was 1.5-C. Importantly, the recorded temperature elevations for the ossicular implants are not considered to be physiologically consequential for a human subject. Potentially injurious, MRI-related heating may be generated in an implant or device depending on the dimensions of the object with regard to its length and/or if it is in the shape of a closed loop with a relatively large diameter (2,33Y35). For an ossicular implant, the length and closed-loop aspects for these various devices, when considered relative to MRI-related heating, do not involve dimensional aspects that will be associated with substantial temperature rises insofar as they are relatively short and have very small, closed loops (Fig. 2). Thus, it is not surprising that the measured temperature increases (i.e., 1.5 and 1.6-C, respectively), even during extreme experimental conditions at 3 T, were not substantial compared with the background temperature rise of 1.5-C. These temperature rises will not impose thermophysiological stress to a human subject. Because the 2 ossicular implants tested had the largest dimensions (albeit relatively short) compared with those listed in the Appendix, the findings of the MRI-related heating tests can be applied to these other implants from the same manufacturer, with a similarly presumed lack of excessive temperature rises. Comparable findings have been reported in the evaluation of intracranial aneurysm clips (26) and hemostatic clips (27) from investigations that followed a similar MRI testing strategy as that applied to this assessment of ossicular implants. Artifacts The presence of a metallic implant during an MRI examination can cause substantial image artifacts, including signal loss, failure of fat suppression, geometric distortion, and other issues (36). Although many factors are known to affect the size of an artifact observed with a metallic implant, it is well known that, for ossicular implants, the extent of the size is predominantly dependent on the magnetic susceptibility of the material (3,25Y27,36). The associated artifacts may affect the diagnostic use of the MRI examination if the area of interest is the same as or close to the site of the ossicular implant. Careful parameter and pulse sequence selections can significantly reduce artifacts from metallic devices (36). Overall, the artifact areas (i.e., in relation to the size of the implant) observed on MRI using T1-weighted, spin echo and gradient echo pulse sequences for the Stainless Steel/Fluoroplastic Sanna-Type Piston were smaller than those associated with the Offset ALTO Total Prosthesis, although stainless steel has a higher magnetic susceptibility than titanium. Obviously, the larger artifact for the titanium implant was because it was larger (15 mm) than the stainless steel one (6 mm). Fortunately, because of the small dimensions and the materials used for the ossicular implants that underwent testing, the artifacts were correspondingly small and should not present substantial problems. However, even a small artifact could limit MRI evaluation of the middle ear. The artifact findings are vital for interpreting radiologists to know to avoid diagnostic dilemmas, including misconstruing an artifact as an abnormality on MRI. MRI Recommendations The current standard of care when managing patients in the MRI setting is for health care professionals to conduct a pre-mri screening procedure that involves identifying implants in the patients and determining whether these objects are MR safe, MR conditional, or MR unsafe (1,2,37,38). On the basis of the methodology used for testing in this study, MRI-specific labeling for each ossicular implant includes the following information (37,38): Nonclinical testing demonstrated that patients with these specific ossicular implants can undergo MRI safely, immediately after implantation under the following conditions: Y Static magnetic field of 3 T or less. Y Maximum spatial gradient magnetic field of 720 G/cm or less. Y MR system reported whole-body-averaged SAR of 2.9 W/kg for 15 minutes of scanning (i.e., per pulse sequence).
6 876 F. G. SHELLOCK ET AL. Implications for Other Metallic and Nonmetallic Ossicular Implants The MR-conditional findings for the 2 metallic ossicular implants can be applied to many other ossicular implants from the same company that have the same or smaller dimensions and made from materials with lower magnetic susceptibilities (Appendix). In addition, many ossicular implants are made from nonmetallic, nonconducting materials, including hydroxyapatite, fluoroplastic, and silicone, which are also included in the Appendix. Magnetic resonance imaging health care professionals may not know that these particular ossicular implants exist and, more importantly, may not be aware that these are, in fact, MR safe. The term MR safe according to the current criteria and labeling terminology is any item that poses no known hazards in all MRI environments because it is made from nonconducting, nonmetallic material(s) (37,38). CONCLUSION In consideration of the minor magnetic field interactions, relative little heating (i.e., above the background temperature when using extreme experimental conditions), and the characterization of artifacts, the results of this investigation demonstrated that it would be acceptable, or MR conditional (i.e., using current MR labeling terminology) (37,38), for patients with the Stainless Steel/Fluoroplastic Sanna-Type Piston and Offset ALTO Total Prosthesis to undergo MRI procedures at 3 T or less. In consideration of the materials and dimensions of the implants that underwent testing, these findings pertain to many other similar ossicular implants from the same manufacturer (Appendix). REFERENCES 1. Shellock FG, Spinazzi A. MRI safety update 2008: Part 2. Screening patients for MRI. AJR Am J Roentgenol 2008;191:12Y Shellock FG. Reference Manual for Magnetic Resonance Safety, Implants and Devices: 2012 Edition. Los Angeles, CA: Biomedical Research Publishing Group, Applebaum EL, Valvassori GE. Effects of magnetic resonance imaging fields on stapedectomy prostheses. Arch Otolaryngol 1985;11: 820Y1. 4. Mattucci KF, Setzen M, Hyman R, Chaturvedi G. The effect of nuclear magnetic resonance imaging on metallic middle ear prostheses. Otolaryngol Head Neck Surg 1986;94:441Y3. 5. Leon JA, Gabriele OF. Middle ear prosthesis: significance in magnetic resonance imaging. Magn Reson Imaging 1987;5:405Y6. 6. Huttenbrink KB. Metallic middle-ear implants in magnetic resonance tomography: hearing impairment and other hazards. Radiologe 1987;27:377Y9. 7. Applebaum EL, Valvassori GE. Further studies on the effects of magnetic resonance fields on middle ear implants. Ann Otol Rhinol Laryngol 1990;99:801Y4. 8. Garaventa G, Satragno L, Vellucci F, Pagano M, Pallestrini EA. The interactions between metal stapes prostheses and high-intensity magnetic fields during magnetic resonance tomography. Acta Otorhinolaryngol Ital 1991;11:455Y Shellock FG, Schatz CJ. Metallic otologic implants: in vitro assessment of ferromagnetism at 1.5 T. AJNR Am J Neuroradiol 1991; 12:279Y Nogueira M, Shellock FG. Otologic bioimplants: ex vivo assessment of ferromagnetism and artifacts at 1.5-tesla. AJR Am J Roentgenol 1995;163:1472Y Syms AJ, Petermann GW. Magnetic resonance imaging of stapes prostheses. Am J Otol 2000;21:494Y Williams MD, Antonelli PJ, Williams LS, Moorhead JE. Middle ear prosthesis displacement in high-strength magnetic fields. Otol Neurotol 2001;22:158Y Syms MJ, Peterman DW. Vibratory sample magnetometry of middle ear prostheses and manufacturing materials. Otol Neurotol 2001;22:487Y Syms MJ. Safety of magnetic resonance imaging of stapes prostheses. Laryngoscope 2005;115:381Y Tohme SM, Karkas AA, Romanos BH. Safety of MRI with metallic middle ear implants. J Med Liban 2003;51:127Y Kwok P, Waldeck A, Strutz J. How do metallic middle ear implants behave in the MRI? Laryngorhinootologie 2003;82:13Y Fritsch MH, Gutt JJ. Ferromagnetic movements of middle ear implants and stapes prostheses in a 3-T magnetic resonance field. Otol Neurotol 2005;26:225Y Martin AD, Driscoll CL, Wood CP, Felmlee JP. Safety evaluation of titanium middle ear prostheses at 3.0 tesla. Otolaryngol Head Neck Surg 2005;132:537Y Thelen A, Bauknecht HC, Asbach P, Schrom T. Behavior of metal implants used in ENT surgery in 7 tesla magnetic resonance imaging. Eur Arch Otorhinolaryngol 2006;263:900Y Wild DC, Head K, Hall DA. Safe magnetic resonance scanning of patients with metallic middle ear implants. Clin Otolaryngol 2006; 31:508Y Fritsch MH. MRI scanners and the stapes prosthesis. Otol Neurotol 2007;28:733Y Fritsch MH, Mosier KM. MRI compatibility issues in otology. Curr Opin Otolaryngol Head Neck Surg 2007;15:335Y Bauknecht HC, Jach C, Krug L, Schrom T. Behaviour of titanium middle ear implants at 1.5 and 3 tesla field strength in magnetic resonance imaging. Laryngorhinootologie 2009;88: 236Y Irving, TX: Kurz Medical, Inc. Available at: en/products/; english.pdf. Accessed December 29, Schenck JF. The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds. Med Phys 1996;23:815Y Shellock FG, Valencerina S. In vitro evaluation of MR imaging issues at 3-T for aneurysm clips made from MP35N: findings and information applied to 155 additional aneurysm clips. AJNR Am J Neuroradiol 2010;31:615Y Gill A, Shellock FG. Assessment of MRI issues at 3-tesla for metallic surgical implants: findings applied to 61 additional skin closure staples and vessel ligation clips. J Cardiovasc Magn Reson 2012;14: American Society for Testing and Materials (ASTM) Designation: F e1 Standard test method for measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment. In: Annual Book of ASTM Standards. Section 13: Medical Devices and Services. Volume Medical and Surgical Materials and Devices. West Conshohocken, PA: ASTM International; 2006:1576Y Shellock FG, Kanal E, Gilk T. Confusion regarding the value reported for the term spatial gradient magnetic field and how this information is applied to labeling of medical implants and devices. AJR Am J Roentgenol 2011;196:142Y American Society for Testing and Materials International Designation: F Standard test method for measurement of radio frequency induced heating near passive implants during magnetic resonance imaging. In: Annual Book of ASTM Standards. Section 13: Medical Devices and Services. Volume Medical and Surgical Materials and Devices. West Conshohocken, PA: ASTM International, 2011.
7 OSSICULAR IMPLANTS AND 3-T MRI Fritsch MH, Gutt JJ, Naumann IC. Magnetic properties of middle ear and stapes implants in a 9.4-T magnetic resonance field. Otol Neurotol 2006;27:1064Y Woods TO. Standards for medical devices in MRI: present and future. J Magn Reson Imaging 2007;26:1186Y Dempsey MF, Condon B, Hadley DM. Investigation of the factors responsible for burns during MRI. J Magn Reson Imaging 2001;13: 627Y Kainz W. MR heating tests of MR critical implants. J Magn Reson Imaging 2007;26:450Y Shellock FG. Comments on MR heating tests of critical implants. J Magn Reson Imaging 2007;26:1182Y Hargreaves BA, Worters PW, Pauly KB, Pauly JM, Koch KM, Gold GE. Metal-induced artifacts in MRI. AJR Am J Roentgenol 2011;197:547Y American Society for Testing and Materials (ASTM) International, Designation: F Standard practice for marking medical devices and other items for safety in the magnetic resonance environment. In: Annual Book of ASTM Standards. Section 13: Medical Devices and Services. Volume Medical and Surgical Materials and Devices. West Conshohocken, PA: ASTM International; Shellock FG, Woods TO, Crues JV. MRI labeling information for implants and devices: Explanation of terminology. Radiology 2009; 253:26Y30. APPENDIX In consideration of the materials and dimensions of the implants that underwent testing (Stainless Steel/Fluoroplastic Sanna-Type Piston [6 mm in length] and the Offset ALTO Total Prosthesis [15 mm in length, titanium/silicone]), these findings pertain to many other similar ossicular implants from the same manufacturer that have smaller sizes and are made from materials with lower magnetic susceptibilities, thus expanding the MRI information for many other devices (all ossicular implants are from Grace Medical, Memphis, TN, USA; Device family Family product number(s) Device material(s) MR-conditional ossicular implants ALTO Partial and Total Prostheses 6XX Titanium, HA, silicone K-Helix Prostheses 756-XXX, 757-XXX Titanium Strasnick Prostheses 220-XXX, 270-XXX Titanium, silicone Precise Prostheses 700-XXX, 705-XXX, 720-XXX, 750-XXX, 765-XXX, 770-XXX Titanium, HA Stapes Prostheses Y Piston Stapes Prostheses Y Buckets MR-safe ossicular implants Stapes Prostheses Y Pistons 409-XXX, 410-XXX, 411-XXX, 412-XXX, 413-XXX, 415-XXX, 417-XXX, 418-XXX, 419-XXX, 452-XXX, 453-XXX, 465-XXX, 466-XXX, 467-XXX, 468-XXX, 469-XXX 420-XXX, 421-XXX, 422-XXX, 423-XXX, 424-XXX, 425-XXX, 426-XXX, 427-XXX, 428-XXX, 429-XXX, 430-XXX Titanium, platinum, stainless steel, nitinol, fluoroplastic Titanium, nitinol 401-XXX, 402-XXX, 403-XXX, 404-XXX, 405-XXX, Fluoroplastic 406-XXX, 408-XXX Partial and Total Prostheses 1XX, 190-XXX HA Partial and Total Prostheses 2XX, 193 HA, silicone HA indicates hydroxyapatite.
Aneurysm Clips: Evaluation of Magnetic Field Interactions With an 8.0 T MR System
JOURNAL OF MAGNETIC RESONANCE IMAGING 12:107 111 (2000) Original Research Aneurysm Clips: Evaluation of Magnetic Field Interactions With an 8.0 T MR System Allahyar Kangarlu, PhD, 1* and Frank G. Shellock,
More information*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants
Knee Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Knee Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All knee implants of Smith
More information*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants
Shoulder Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Shoulder Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All shoulder implants
More information*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants
Hip Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Hip Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All hip implants of Smith &
More informationProsthetic Heart Valves and Annuloplasty Rings: Assessment of Magnetic Field Interactions, Heating, and Artifacts at 1.5 Tesla
Journal of Cardiovascular Magnetic Resonance, 3(4), 317 324 (2001) Prosthetic Heart Valves and Annuloplasty Rings: Assessment of Magnetic Field Interactions, Heating, and Artifacts at 1.5 Tesla Frank G.
More informationMRI Information for Medtronic Spinal and Biologics Devices
M708348B263E Rev C Release Date: 2013-02 MRI Information for Medtronic Spinal and Biologics Devices Not for Distribution in the U.S or its territories. 1 P age Table of Contents MRI Information for Medtronic
More informationLow-Profile Wrist Fixator. For stabilization of fractures of the distal radius.
Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius. Technique Guide Part of the External Fixation System Low-Profile Wrist Fixator Indications Intended for stabilization of
More informationMRI Guidelines for the Axonics Sacral Neuromodulation System
MRI Guidelines for the Axonics Sacral Neuromodulation System For Healthcare Professionals Rx only Note: This document is a supplement to the physician manuals provided with the Axonics SNM System. This
More informationVNS Therapy System Physician s Manual
VNS Therapy System Physician s Manual Pulse Generator Model 102 Pulse Duo Generator Model 102R Demipulse Generator Model 103 Demipulse Duo Generator Model 104 AspireHC Generator Model 105 AspireSR Generator
More informationMedium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps.
Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps. Technique Guide Part of the Medium External Fixation System MRI Information Synthes Medium External Fixation devices
More information1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)
1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500). ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro
More informationCochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)
Cochlear Implants Medical Procedures for MED EL Implant Systems AW33290_2.0 (English US) This manual provides important instructions and safety information for MED EL Implant System users who have to undergo
More informationExternal Distal Radius Fixator. Supplement to the 8 mm rod fixator system
External Distal Radius Fixator. Supplement to the 8 mm rod fixator system Surgical technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation
More informationMedium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use.
Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use. Technique Guide Part of the Medium External Fixation System MRI Information Synthes Medium External Fixation devices
More information1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)
1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro
More informationSafety Issue of Static Magnetic Field (1)
MRI Safety Safety Issue of Static Magnetic Field (1) Currently, typical MRI utilizes static field with strengths of 0.5 to 3 Tesla (FDA guideline is max 4 T for infants < 1 month and 8 T for others) 1
More informationReplacement Heart Valve Product Description (Stented Tissue) Models Reference
Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC
More informationWallFlex Biliary RX Partially Covered Stent System Prescriptive Information
Caution/Rx Only: Federal Law (USA) restricts this device to sale by or on the order of a physician. Intended Use/Indications for Use The WallFlex Biliary RX is indicated for use in the palliative treatment
More informationAlgovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x
Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 201x Algovita
More informationImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System
ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System CAUTION: Federal law restricts this device to sale, distribution and use by or on the order of a physician. 91035972-02
More informationMemo on The Safety of MRI Scanning of Coronary Artery Stents
Memo on The Safety of MRI Scanning of Coronary Artery Stents Developed by Tobias Gilk, Gilk Radiology Consultants for The problem: Many coronary artery stents have (a) no manufacturer guidance, or (b)
More informationRead this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System.
Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 0300-000175-001
More informationReplacement Heart Valve Product Description (Stented Tissue) Models Reference
Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC
More informationACR MRI Accreditation Program. ACR MRI Accreditation Program Update. Educational Objectives. ACR accreditation. History. New Modular Program
ACR MRI Accreditation Program Update Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives Present requirements of the new
More informationCochlear Implants. Medical Procedures. for MED EL CI Systems. AW33290_5.0 (English US)
Cochlear Implants Medical Procedures for MED EL CI Systems AW33290_5.0 (English US) This manual provides important instructions and safety information for MED EL CI System users who have to undergo a medical
More information1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System
1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System ONLY 11096 Rev B 1 NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro Corp.
More informationJOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp , 2003
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 5, No. 2, pp. 387 397, 2003 SAFETY Cardiac s, ICDs, and Loop Recorder: Evaluation of Translational Attraction Using Conventional ( Long-Bore ) and Short-Bore
More informationACR MRI Accreditation: Medical Physicist Role in the Application Process
ACR MRI Accreditation: Medical Physicist Role in the Application Process Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics University of Texas M.D. Anderson Cancer Center Educational Objectives
More informationMRI GUIDELINES FOR INSPIRE THERAPY
MRI GUIDELINES FOR INSPIRE THERAPY Clinician s Manual ONLY The following is a trademark of Inspire Medical Systems, Inc.: Inspire Introduction Read the information in this manual prior to conducting an
More informationImageReady MRI Head Only Guidelines for Spectra WaveWriter Spinal Cord Stimulator System
ImageReady MRI Head Only Guidelines for Spectra WaveWriter Spinal Cord Stimulator System CAUTION: Federal law restricts this device to sale, distribution and use by or on the order of a physician. 91171762-01
More information2014 AAPM 56 th Annual Meeting
2014 AAPM 56 th Annual Meeting SAM Diagnostic Radiology MR Safety - Deep Brain Stimulator and Other Neurostimulators Yunhong Shu, Ph.D. Mayo Clinic, Rochester, MN Outline Background of Neurostimulator
More informationDISTAL RADIUS. Instructions for Use
DISTAL RADIUS Instructions for Use CAUTION: FEDERAL LAW RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN. Table of Contents 1. INTENDED USE... 3 2. DEVICE DESCRIPTION... 3 3. METHOD OF
More informationMicra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR)
Micra MC1VR01 MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR) MRI Technical Manual The following list includes trademarks or registered
More informationMRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH
MRI for Patients with MRI-Conditional Pacing System: Radiographers Role Lawrance Yip DM, DR, QMH MRI & Pacemaker MRI is a crucial and growing imaging modality Plays an important role in clinical management
More informationMRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers
MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers Frank G. Shellock, Ph.D., FACC Adjunct Clinical Professor of Radiology and Medicine Keck School of Medicine,
More informationPATIENT CARE AND MRI SAFETY Module 7
PATIENT CARE AND MRI SAFETY Module 7 1 Biological Considerations There are no reported adverse biological effects of extended exposure to MRI. However, several inconsequential and reversible effects of
More informationMRI Safety Information for the HiRes Ultra 3D Cochlear Implant
Advanced Bionics AG Laubisrütistrasse 28 8712 Stäfa, Switzerland +41 58 928 78 00 Manufactured by: Advanced Bionics LLC California, U.S.A. +1 661 362 1400 MRI Safety Information for the HiRes Ultra 3D
More informationWarning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed.
Prometra and Prometra II Programmable Pumps Magnetic Resonance Imaging (MRI) Safety Information GENERAL MR Conditional WARNING: FAILURE TO EMPTY THE PUMP PRIOR TO EXPOSURE TO MRI ENVIRONMENT COULD RESULT
More informationA quality control program for MR-guided focused ultrasound ablation therapy
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 2, SPRING 2002 A quality control program for MR-guided focused ultrasound ablation therapy Tao Wu* and Joel P. Felmlee Department of Radiology,
More informationMR QA/QC for MRgRT. Rick Layman, PhD, DABR Department of Radiology July 13, 2015
MR QA/QC for MRgRT Rick Layman, PhD, DABR Department of Radiology July 13, 2015 The Ohio State University Comprehensive Cancer Center Arthur G. James Cancer Hospital and Richard J. Solove Research Institute
More informationOrthopedic Hardware Imaging Part II: MRI v. Metal
Orthopedic Hardware Imaging Trent Roth, MD And Lauren Ladd, MD Indiana University School of Medicine IU Health Physicians-Radiology Recap: Imaging Techniques Radiography Standard for initial and surveillance
More informationLarge External Fixator Modular Knee Bridge. Using multi-pin clamps.
Large External Fixator Modular Knee Bridge. Using multi-pin clamps. Technique Guide Part of the Large External Fixation System Large External Fixator Modular Knee Bridge Technique Overview Insert Schanz
More informationThe Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.
The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.
More informationMRI GUIDELINES FOR INSPIRE THERAPY
MRI GUIDELINES FOR INSPIRE THERAPY Clinician s Manual The following is a trademark of Inspire Medical Systems, Inc.: Inspire This product and/or the use of this product in a method may be covered by one
More informationComparative MRI Compatibility of 316L Stainless Steel Alloy and Nickel Titanium Alloy Stents
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE w Vol. 4, No. 4, pp. 423 430, 2002 ORIGINAL ARTICLE Technical Comparative MRI Compatibility of 316L Stainless Steel Alloy and Nickel Titanium Alloy Stents Andrea
More informationCARDIAC MRI. Cardiovascular Disease. Cardiovascular Disease. Cardiovascular Disease. Overview
CARDIAC MRI Dr Yang Faridah A. Aziz Department of Biomedical Imaging University of Malaya Medical Centre Cardiovascular Disease Diseases of the circulatory system, also called cardiovascular disease (CVD),
More informationSafety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator
european urology 50 (2006) 311 316 available at www.sciencedirect.com journal homepage: www.europeanurology.com Neuro-urology Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator
More informationREFILL KIT For use with Prometra Programmable Infusion Systems
REFILL KIT Caution: Federal (USA) Law restricts this device to sale by or on the order of a physician. Table of Contents Contents... 3 Description... 3 Indications... 3 Contraindications... 3 Warnings...
More informationOrthopedic Metal Artifact Reduction Distortion correction in the presence of a orthopedic implant
O-MAR MR software Orthopedic Metal Artifact Reduction Distortion correction in the presence of a orthopedic implant By S. Hey, D. Hoogenraad, V. Elanchezhian, M. van Meel and F. Admiraal-Behloul In recent
More informationOTOLOGICAL MIDDLE EAR PROSTHESES
MIDDLE EAR IMPLANTS OTOLOGICAL MIDDLE EAR PROSTHESES Tita-prosthesis type Partial Vario Campana (OSSEOUS: PORP) Tita-prosthesis type Vario Cask (OSSEOUS: TORP) PTFE Prosthesis (TORP) PTFE Partial Prosthesis
More informationDistal Radius Plate 2.4/2.7 dorsal and volar
Distal Radius Plate 2.4/2.7 dorsal and volar Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Distal Radius Plate
More informationOBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.
LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. Surgical Technique LCP Small Fragment System This publication
More informationDescriptions of NDT Projects Fall 2004 October 31, 2004
Descriptions of NDT Projects Fall 2004 October 31, 2004 Introduction There are two separate NDT labs in Magister: ULTRA for ultrasound and EDDY for eddy current. Both labs are equipped with mechanical
More informationBuilding physician referrals and improve patient care for an aging population with
MR Conditional implants Building physician referrals and improve patient care for an aging population with MR Conditional implants 12.5 million people in the USA are presently carrying an orthopedic or
More informationRECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE
In Practice RECENT ADVANCES IN CLINICAL MR OF ARTICULAR CARTILAGE By Atsuya Watanabe, MD, PhD, Director, Advanced Diagnostic Imaging Center and Associate Professor, Department of Orthopedic Surgery, Teikyo
More informationQuality Assurance of Ultrasound Imaging in Radiation Therapy. Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St.
Quality Assurance of Ultrasound Imaging in Radiation Therapy Zuofeng Li, D.Sc. Murty S. Goddu, Ph.D. Washington University St. Louis, Missouri Typical Applications of Ultrasound Imaging in Radiation Therapy
More informationComparison of Fluoroplastic Causse Loop Piston and Titanium Soft-Clip in Stapedotomy
Original Article Iranian Journal of Otorhinolaryngology, Vol.29(1), Serial No.90, Jan 2017 Comparison of Fluoroplastic Causse Loop Piston and Titanium Soft-Clip in Stapedotomy Mohammad Faramarzi 1, Nafiseh
More informationMagnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T
International Congress Series 1273 (2004) 394 398 www.ics-elsevier.com Magnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T F. Risi*, A. Saldanha, R. Leigh, P. Gibson Cochlear Limited,
More informationRevision A Date:
Biomet Trauma 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581 USA 01-50-4053 Revision A Date: 2016-06 Biomet Unite3D Bridge with OsseoTi Technology ATTENTION OPERATING SURGEON DESCRIPTION Biomet
More informationMagnetic Resonance Imaging on Soft Tissue. Jiten K. Mistry Calvin Gan
Magnetic Resonance Imaging on Soft Tissue 1 Jiten K. Mistry Calvin Gan Outline Background of Medical Imaging Introduction to MRI How MRI works MRI of Soft Tissue Benefits & Risks Recent Advances 2 The
More informationPersonal use only. MRI Metal Artifact Reduction: Shoulder Implants and Arthroplasty. Reto Sutter, MD
MRI Metal Artifact Reduction: Shoulder Implants and Arthroplasty Reto Sutter, MD University Hospital Balgrist Zurich University of Zurich Cor PD fat sat 56-year old male patient with positive lift-off
More informationLCP Metaphyseal Plates. For extra-articular fractures.
LCP Metaphyseal Plates. For extra-articular fractures. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier
More informationUser Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017
User Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017 Ives EEG Solutions; 25 Storey Ave., #118, Newburyport, MA, 01950; ph: 978-358-8006; fax: 978-358-7825; www.iveseegsolutions.com
More informationFor Small-Statured Adults and Pediatric Patients. Medium External Fixator Basic Modular Frame
For Small-Statured Adults and Pediatric Patients Medium External Fixator Basic Modular Frame Surgical Technique MRI Information DePuy Synthes Medium External Fixation devices are labeled MR Conditional
More informationLead Management for MRI
Lead Management for MRI Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services The Ohio State University Medical Center Columbus, OH USA President, International
More informationMR Evaluation Ex Vivo and In Vivo of a Covered Stent-Graft for Abdominal Aortic Aneurysms: Ferromagnetism, Heating, Artifacts, and Velocity Mapping
JOURNAL OF MAGNETIC RESONANCE IMAGING 12:112 121 (2000) Original Research MR Evaluation Ex Vivo and In Vivo of a Covered Stent-Graft for Abdominal Aortic Aneurysms: Ferromagnetism, Heating, Artifacts,
More informationLCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.
LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Surgical Technique This publication is not intended for
More informationAPPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS
APPLICATION AND DEPLOYMENT OF ADVANCED NDE TECHNIQUES IN HIGH PRESSURE VESSELS Jeffrey P. Milligan, Daniel T. Peters, Structural Integrity Associates, Inc., USA Many advances in Non-Destructive Examination
More informationMagnetic Resonance Angiography
Magnetic Resonance Angiography 1 Magnetic Resonance Angiography exploits flow enhancement of GR sequences saturation of venous flow allows arterial visualization saturation of arterial flow allows venous
More informationSterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation.
Sterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation. Large External Fixator Ankle Frame Kit Large External Fixator Trauma Kit Large
More informationGUIDELINE 2-G-004 (formerly APPENDIX 3) HSREB (March 2005) Page 1
HSREB (March 2005) Page 1 Guidelines for Ethics Approval of Research Protocols Involving Human Exposure to Magnetic Resonance Imaging Required Wording in Informed Consent Documentation; The Food & Drug
More informationSyllabus References. Resources. Video: MRI Introduction
MRI Lesson Outline Syllabus References 9.6.4.2.5 Define precessing and relate the frequency of the precessing to the composition of the nuclei and the strength of the applied external magnetic field 9.6.4.2.6
More informationLarge External Fixator Modular Knee Bridge
Using Multi-pin Clamps Large External Fixator Modular Knee Bridge Surgical Technique Large External Fixator Modular Knee Bridge DePuy Synthes Large External Fixation devices are labeled MR Conditional
More informationNon Contrast MRA. Mayil Krishnam. Director, Cardiovascular and Thoracic Imaging University of California, Irvine
Non Contrast MRA Mayil Krishnam Director, Cardiovascular and Thoracic Imaging University of California, Irvine No disclosures Non contrast MRA-Why? Limitations of CTA Radiation exposure Iodinated contrast
More informationAnnual Associate Safety Module Magnetic Resonance Imaging (MRI)
Annual Associate Safety Module Magnetic Resonance Imaging (MRI) MRI Safety and YOU! Rev 12/16 Magnetic Resonance Imaging (MRI) MRI uses a large magnet with a powerful magnetic field, radiofrequencies (RF
More informationACR Accreditation Update in MRI
ACR Accreditation Update in MRI Whole Body Systems Extremity (MSK) Ron Price Vanderbilt University Medical Center Nashville, TN Dedicated Breast MRI Accreditation Update 1. ACR MRI Accreditation: Overview,
More informationFOR CMS (MEDICARE) MEMBERS ONLY NATIONAL COVERAGE DETERMINATION (NCD) FOR MAGNETIC RESONANCE IMAGING:
National Imaging Associates, Inc. Clinical guidelines BONE MARROW MRI Original Date: July 2008 Page 1 of 5 CPT Codes: 77084 Last Review Date: September 2014 NCD 220.2 MRI Last Effective Date: July 2011
More informationANJON BREMER LIGHT VEST
PATENTED TECHNOLOGY, TARGETED TO YOUR NEEDS ANJON BREMER APPLICATION, REMOVAL, AND CPR INSTRUCTIONS WWW.GENERIXMEDICAL.COM INTRODUCTORY INFORMATION FOR THE ANJON BREMER MRI Safety Information: Non-clinical
More informationButton Plate. Reinforcement for transosseous fixations.
Button Plate. Reinforcement for transosseous fixations. Product Information This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier
More informationCT Scanning Protocol For V2R Guided Surgery Solutions
CT Scanning Protocol For V2R Guided Surgery Solutions 2 V2R CT Scanning Protocol \\ Contents Contents General requirements... 3 V2R Dual Scan Protocol... 5 V2R Single Scan Protocol... 8 Overview... 10
More informationHigh Field MR of the Spine
Department of Radiology University of California San Diego 3T for MR Applications Advantages High Field MR of the Spine Increased signal-to-noise Better fat suppression Increased enhancement with gadolinium
More informationP2 Visual - Perception
P2 Visual - Perception 2014 SOSE Neuroimaging of high-level visual functions gyula.kovacs@uni-jena.de 11/09/06 Functional magnetic resonance imaging (fmri) The very basics What is fmri? What is MRI? The
More information2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.
2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Surgical Technique This publication is not intended
More informationMedium External Fixator Delta Frame Ankle Bridge
For Small-Statured Adults Medium External Fixator Delta Frame Ankle Bridge Surgical Technique MRI Information DePuy Synthes Medium External Fixation devices are labeled MR Conditional according to the
More informationLarge External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia.
Large External Fixator Delta Frame Ankle Bridge. For staged fixation of the distal tibia. Technique Guide Part of the Large External Fixation System Large External Fixator Delta Frame Ankle Bridge Technique
More informationMR Advance Techniques. Vascular Imaging. Class II
MR Advance Techniques Vascular Imaging Class II 1 Vascular Imaging There are several methods that can be used to evaluate the cardiovascular systems with the use of MRI. MRI will aloud to evaluate morphology
More informationMRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System
MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. WARNING: This
More informationSensitivity and Specificity in Detection of Labral Tears with 3.0-T MRI of the Shoulder
Magee and Williams MRI for Detection of Labral Tears Musculoskeletal Imaging Clinical Observations C M E D E N T U R I C L I M G I N G JR 2006; 187:1448 1452 0361 803X/06/1876 1448 merican Roentgen Ray
More informationCarry this card with you at all times. Show this card to any medical professional treating you. Patient Implant Card. Option ELITE Vena Cava Filter.
Patient Guide A Safe Option for a Healthier You! P/N: P-2017-0175-00 Rev B 1. Static magnetic field of 3 Tesla or less. 2. Spatial gradient magnetic field of 720 Gauss/cm or less. 3. Maximum whole body
More informationLOW PROFILE NEURO. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE
LOW PROFILE NEURO This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE TABLE OF CONTENTS INTRODUCTION Low Profile Neuro Plating System 2 Intended Use, Indications, Contraindications
More informationw. D. Jolly, F. A. Bruton, and C. Fedor
ULTRASONIC TRANSDUCER CHARACTERIZATION STATION w. D. Jolly, F. A. Bruton, and C. Fedor Southwest Research Institute 6220 Culebra Road San Antonio, Texas 78284 INTRODUCTION The portable ultrasonic transducer
More informationEssentials of Clinical MR, 2 nd edition. 99. MRA Principles and Carotid MRA
99. MRA Principles and Carotid MRA As described in Chapter 12, time of flight (TOF) magnetic resonance angiography (MRA) is commonly utilized in the evaluation of the circle of Willis. TOF MRA allows depiction
More informationMeniere s disease and Sudden Sensorineural Hearing Loss
Meniere s disease and Sudden Sensorineural Hearing Loss Tsutomu Nakashima 1,2 1 Ichinomiya Medical Treatment & Habilitation Center, Ichinomiya, Japan 2 Department of Otorhinolaryngology, Nagoya University,
More informationDevelopment of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion
Development of Ultrasound Based Techniques for Measuring Skeletal Muscle Motion Jason Silver August 26, 2009 Presentation Outline Introduction Thesis Objectives Mathematical Model and Principles Methods
More informationLCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.
LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Surgical Technique This publication is not intended for
More informationSTANDARD OPERATING PROCEDURE (SOP)
STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0
More informationUC Davis Imaging Research Center MRI Safety Manual
UC Davis Imaging Research Center MRI Safety Manual 1 Rev. 3, 1/5/18 Table of Contents Section 1... 4 Introduction... 4 Components of the MRI scanner system... 4 The Static or main Magnetic field... 4 The
More informationCardiac Pacemakers and Implantable Cardioverter Defibrillators: In Vitro Magnetic Resonance Imaging Evaluation at 1.5-Tesla
Journal of Cardiovascular Magnetic Resonance (2007) 9, 21 31 Copyright c 2007 Informa Healthcare ISSN: 1097-6647 print / 1532-429X online DOI: 10.1080/10976640600897237 Cardiac Pacemakers and Implantable
More informationLarge External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.
Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts. Technique Guide Part of the Large External Fixation System Large External Fixator Delta Frame Ankle Bridge Technique
More informationNew Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor. Cardiothoracic Radiology
New Cardiovascular Devices and Interventions: Non-Contrast MRI for TAVR Abhishek Chaturvedi Assistant Professor Cardiothoracic Radiology Disclosure I have no disclosure pertinent to this presentation.
More information