TIM Status and Contributions to the Total Solar Irradiance Record

Size: px
Start display at page:

Download "TIM Status and Contributions to the Total Solar Irradiance Record"

Transcription

1 TIM Status and Contributions to the Total Solar Irradiance Record Greg Kopp Laboratory for Atmospheric and Space Physics University of Colorado 1234 Innovation Dr., Boulder, CO 80303, USA TIM Status Greg Kopp, Page 1

2 Basis of TSI Proxies Is Recent Measurements Sunspot and cosmogenic isotope records give longterm TSI proxies to compare with climate. TSI proxies are extrapolations from recent observations. TSI is compared to sunspot record for last 27 years. Sunspot record is compared to cosmogenic isotopes for last 400 yrs. Ilya Usoskin, Physical Review Letters, (Groveman and Landsburg; 1979) (Lean et al.; 1995). Courtesy of J. Lean, NRL. TSI measurements need to be accurate and well-connected to existing 27-year record. TIM Status Greg Kopp, Page 2

3 TIM TSI Record TIM Status Greg Kopp, Page 3

4 TSI Record updated regularly at TIM Status Greg Kopp, Page 4

5 TSI Accuracy Workshop Held at NIST, Gaithersburg in July 2005 Representatives of 7 of the space-borne TSI instruments: ACRIM I, II, and III; ERBE; VIRGO/PMO; VIRGO/DIARAD, SOLCON, SOVA; TIM 3-day agenda Review the absolute accuracies of each space-borne TSI instrument Review degradation corrections and the ability to track long-term solar variability Discuss possible future improvements to calibration accuracy TIM Status Greg Kopp, Page 5

6 Agreement Between TIM Cones Remains ~270 ppm TIM cone agreement ~270 ppm 1- standard deviation Variations likely caused by uncertainty in applying linearity corrections from ground TIMs to each cone of flight unit TIM Status Greg Kopp, Page 6

7 Comparison to Trapped Diode Compare TIM to transfer standard 6- element BoulderMetrics trapped diode Power comparison (not irradiance) Optical setup Power levels 1-10 mw Intended TIM power 68 mw Laser stability 10-3 Spatial filter improves uniformity Expanded (~6 mm diam) beam gives appropriate cone illumination Still have small calibration details to include for the TIM Witness Unit Will be doing similar test at NIST for all current TSI instruments (ACRIM, VIRGO/PMO, VIRGO/DIARAD, TIM) Setup: David Harber, Karl Heuerman TIM Status Greg Kopp, Page 7

8 Space-Borne TSI Record Relies on Continuity Glory TIM Status Greg Kopp, Page 8

9 10 ppm/yr Stability Facilitates Trend Detection? TIM s 4 duty-cycled cavities and photodiodes monitor degradation. SORCE/TIM has measured degradation of <50 ppm/yr. Corrected with an uncertainty <10 ppm/yr TIM Status Greg Kopp, Page 9

10 Cone Intercomparisons Show <100 ppm Degradation in Primary Cone (B) after First Two Years in Orbit Weekly simultaneous solar observations with Primary Cone (B) and Secondary Cone (A) track changes in Primary Cone Photodiodes monitoring reflectance from each cone show similar changes Degradation is very small and is corrected in data processing TIM Status Greg Kopp, Page 10

11 Cumulative Exposure of Backup Cones Is Low At current exposure rates, Secondary Cone will not start to degrade significantly during 5-year SORCE mission x20 x50 x50 TIM Status Greg Kopp, Page 11

12 Low Noise Detects Dynamic Solar Effects 2003 Drives instrument requirements on thermal stability thermistor noise shutter cycle period amplifier & AtoD noise SORCE/TIM was the first TSI instrument with sufficient sensitivity to detect a solar flare. TIM Status Greg Kopp, Page 12

13 TSI Flare Observations Measure Total Flare Energy See presentation by T. Woods on flares TIM Status Greg Kopp, Page 13

14 Data Processing Has TSI Data Available In 1 Week TIM TSI measurements from 1 March 2003 to the present are available via Daily data updates follow data acquisition by roughly 7 days Evaluating making processed data available within a day of telemetry of raw data for quicker comparisons with other solar activity indicators TIM Status Greg Kopp, Page 14

15 TIM Will Next Be Flying on Glory... S/C TSI Observations Glory TIM Instrument on Glory TIM Instrument TPS TIM Status Greg Kopp, Page 15

16 ...and then on NPOESS TIM S/C TSI Observations SIM Glory TIM Status Greg Kopp, Page 16

17 Summary TIM TSI values remain ~4.5 W/m 2 lower than other S/C TSI instruments. The NIST TSI Accuracy Workshop in July 2005 reaffirms our confidence in this value The TIM shows very little optical degradation from solar exposure, and the electronics show no signs of change since launch. TIM degradation is low at <50 ppm/yr This is tracked and corrected in data analysis Results from 1 March 2003 to the present are released regularly and are available 7 days after acquisition. TIM Status Greg Kopp, Page 17

SOLSTICE I and II: Calibrations and Comparisons

SOLSTICE I and II: Calibrations and Comparisons SOLSTICE I and II: Calibrations and Comparisons Marty Snow, Bill McClintock, Tom Woods, and Gary Rottman Laboratory for Atmospheric and Space Physics Outline SOLSTICE I and II SOLSTICE I Validation review

More information

NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment

NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment NMIJ measurement service on ultrasonic field parameters available to demonstrate performance and safety of ultrasonic medical equipment Masahiro Yoshioka National Metrology Institute of Japan (NMIJ) National

More information

The need for standardization of dosimetry in experimental radiation biology

The need for standardization of dosimetry in experimental radiation biology The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall

More information

Clinical Implementation of a New Ultrasound Guidance System. Vikren Sarkar Bill Salter Martin Szegedi

Clinical Implementation of a New Ultrasound Guidance System. Vikren Sarkar Bill Salter Martin Szegedi Clinical Implementation of a New Ultrasound Guidance System Vikren Sarkar Bill Salter Martin Szegedi Disclosure The University of Utah has research agreements with Elekta Agenda Historical Review Trans-Abdominal

More information

INTRODUCTION TO MALDI IMAGING

INTRODUCTION TO MALDI IMAGING INTRODUCTION TO MALDI IMAGING Marten F. Snel, Emmanuelle Claude, Thérèse McKenna, and James I. Langridge Waters Corporation, Manchester, UK INT RODUCTION The last few years have seen a rapid increase in

More information

Importance of Radiation Dosimetry standards in preclinical radiobiology studies

Importance of Radiation Dosimetry standards in preclinical radiobiology studies Importance of Radiation Dosimetry standards in preclinical radiobiology studies Ceferino Obcemea Radiation Research Program National Cancer Institute, Bethesda, MD, USA CIRMS 2018 No financial conflict

More information

PHYSICS NEEDS IN RADIATION SCIENCES. James Deye, Ph. D. Contractor, NCI Radiation Research Program

PHYSICS NEEDS IN RADIATION SCIENCES. James Deye, Ph. D. Contractor, NCI Radiation Research Program PHYSICS NEEDS IN RADIATION SCIENCES James Deye, Ph. D. Contractor, NCI Radiation Research Program 6/21/2016 Reproducibility / Validation / Efficacy Previous Workshops: 2003 recommendations (1) Establish

More information

Followup on Recent Debates with Dan Murphy on AMS Quantification

Followup on Recent Debates with Dan Murphy on AMS Quantification Followup on Recent Debates with Dan Murphy on AMS Quantification Jose-Luis Jimenez et al. University of Colorado-Boulder AMS Users Meeting Portland, OR, 21-Oct-2016 http://www.tandfonline.com/doi/full/10.1080/02786826.2015.1136403

More information

EOS Validation Investigation Annual Report

EOS Validation Investigation Annual Report EOS Validation Investigation Annual Report Title: Validation and Correction for the Terra MODIS Spatial Response NASA Grant No: NAG5-6339 Period: June 1, 2000 September 30, 2001 Principal Investigator:

More information

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007

ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007 ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory

More information

WHITE PAPER. Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions for Optical Design and Simulation

WHITE PAPER. Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions for Optical Design and Simulation Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions for Optical Design and Simulation Efficient Measurement of Large Light Source Near-Field Color and Luminance Distributions

More information

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4

Practical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4 Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability

More information

icat Precise Policy and Procedure Manual

icat Precise Policy and Procedure Manual ORAL SURGERY GROUP P.A. PRACTICE LIMITED TO ORAL AND MAXILLOFACIAL SURGERY DENTAL IMPLANTS, WISDOM TEETH, ORAL PATHOLOGY, AND ORAL MEDICINE Dr. David M. Rauch Dr. Philip S. Engel Dr. Richard K. Stern Dr.

More information

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM

THE TUFFEST STUFF CT REGISTRY REVIEW Live Lecture Seminar SATURDAY CURRICULUM 1. The CT Imaging Chain-10 major components & their functions a. The x-ray tube b. Generator c. Filter d. Pre-patient collimator e. Pre-detector collimator f. Detector system g. Analog to digital converter

More information

Calibration Standards

Calibration Standards T H I N F I L M Calibration Standards psi Standards Calibration Standards for the Semiconductor Industry S I M P L Y T H E B E S T PSI Standards Move into the future of calibration and ISO compliance with

More information

GEX Technical Information Report

GEX Technical Information Report Transfer Standard Alanine Study September 2017 Background Ionizing radiation is used to produce various desired effects in products. Examples of applications include; the sterilization of medical products,

More information

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties

Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry

More information

WFC3-IR Thermal Vacuum Testing: IR Channel Dark Current

WFC3-IR Thermal Vacuum Testing: IR Channel Dark Current WFC3-IR Thermal Vacuum Testing: IR Channel Dark Current B. Hilbert and M. Robberto September 9, 2005 ABSTRACT During the 2004 thermal vacuum testing campaign, a series of tests were performed in order

More information

Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance

Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance Bruker Daltonics Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance The new ultraflextreme exceeds all current expectations of MALDI-TOF/TOF technology: A proprietary khz smartbeam-ii TM MALDI

More information

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy

Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16

More information

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy?

Stereotaxy. Outlines. Establishing SBRT Program: Physics & Dosimetry. SBRT - Simulation. Body Localizer. Sim. Sim. Sim. Stereotaxy? Establishing SBRT Program: Physics & Dosimetry Lu Wang, Ph.D. Radiation Oncology Department Fox Chase Cancer Center Outlines Illustrate the difference between SBRT vs. CRT Introduce the major procedures

More information

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant:

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant: DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 General Requirements Y, N, N/A Demonstration of Conformance G.1 Latest version of protocols or procedures G.2 Latest version of dosimeter specifications

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Efficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation

Efficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation Efficient Measurement of Large Light Source Near-field Color and Luminance Distributions for Optical Design and Simulation Hubert Kostal*, Douglas Kreysar, Ronald Rykowski Radiant Imaging, Inc., 22908

More information

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom

Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson

More information

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis

Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline

More information

The Journey of Cyberknife Commissioning

The Journey of Cyberknife Commissioning The Journey of Cyberknife Commissioning Jun Yang Ph.D 1, Alan Cohen M.S. 2 1) Adjunct Associate Professor Drexel University Alliance Oncology 2) Chief Medical Physicist Accuray Incorporated X-ray Sources

More information

Time-of-Flight Aerosol Mass Spectrometer Instrument & Winter Data

Time-of-Flight Aerosol Mass Spectrometer Instrument & Winter Data Time-of-Flight Aerosol Mass Spectrometer Instrument & Winter Data First Field Deployment during PMTACS-NY Frank Drewnick +, Silke Hings +, Silke Weimer, Marc Gonin *, Jose Jimenez &, Peter DeCarlo &, Doug

More information

Supporting Information

Supporting Information Supporting Information Identification of Proprietary Amino Ethoxylates in ydraulic Fracturing Wastewater Using Liquid Chromatography/Time-of-Flight Mass Spectrometry with Solid Phase Extraction Kurban

More information

Exposure to Background Radiation In Australia

Exposure to Background Radiation In Australia AU9816851 Exposure to Background Radiation In Australia S.B.SOLOMON, Australian Radiation Laboratory, Lower Plenty Road, Yallambie, Victoria, 3085 SUMMARY The average effective dose received by the Australian

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

PGY-1. Resident Review Session Schedule

PGY-1. Resident Review Session Schedule 1. August Simulation & Treatment 1.1. Sim Setup 1.2. Sim Techniques 1.3. 4DCT 1.4. Breath Hold / Gating 1.5. Treatment Setup 1.6. Treatment Delivery 1.7. Filming 1.7.1. Port film 1.7.2. kv 1.7.3. CBCT

More information

Potential conflicts-of-interest. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy. Educational objectives. Overview.

Potential conflicts-of-interest. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy. Educational objectives. Overview. Respiratory Gated and Four-Dimensional Tumor Tracking Radiotherapy Potential conflicts-of-interest I am PI of a sponsored research agreement between Stanford University and Varian Medical Systems P Keall

More information

Dosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology

Dosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology Kantonsspital Luzern Dosimetry Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 Introduction Metrological traceability; the dosimetry chain Absolute / relative

More information

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS

ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS ACQUITY UPLC WITH PDA DETECTION: DETERMINING THE SENSITIVITY LIMITS OF OXYBUTYNIN AND RELATED COMPOUNDS Tanya Jenkins Waters Corporation, Milford, MA, USA INTRODUCTION Some of the most challenging methods

More information

Policy Setting Considerations

Policy Setting Considerations Radon Review and Update: How Good is the Science Mar 7, 2012 EPA Office of Radiation and Indoor Air (ORIA) Indoor Environments Division (IED) CAPT Susan Conrath, PhD, MPH, US Public Health Service conrath.susan@epa.gov

More information

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter

3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Optical Scanning 3-D dose verification by cone-beam optical CT scanning of PRESAGE dosimeter Wuu, C.-S., Hoogcarspel, S.J.,Deh, K., Hsu, W.-Y., Adamovics, J. 2013 Journal of Physics: Conference Series

More information

ISO Dentistry Digitizing devices for CAD/CAM systems for indirect dental restorations Test methods for assessing accuracy

ISO Dentistry Digitizing devices for CAD/CAM systems for indirect dental restorations Test methods for assessing accuracy INTERNATIONAL STANDARD ISO 12836 First edition 2012-09-15 Dentistry Digitizing devices for CAD/CAM systems for indirect dental restorations Test methods for assessing accuracy Médecine bucco-dentaire Dispositifs

More information

Small field diode dosimetry

Small field diode dosimetry Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

Analysis of drugs of abuse in biological matrix using Time of Flight technology

Analysis of drugs of abuse in biological matrix using Time of Flight technology Analysis of drugs of abuse in biological matrix using Time of Flight technology Niclas Stephanson, PhD Department of Clinical Pharmacology Karolinska University Hospital Karolinska Institutet Stockholm,

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,

More information

A Novel Dosimetry System for Military Use in Response of Nuclear Emergencies

A Novel Dosimetry System for Military Use in Response of Nuclear Emergencies A Novel Dosimetry System for Military Use in Response of Nuclear Emergencies Mirela Kirr 1, Craig Yoder, PhD 1, Chris Passmore, CHP 1 1 Landauer Inc. 2 Science Road, Glenwood IL 60425-1586 USA Abstract.

More information

Application(s) of Alanine

Application(s) of Alanine Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for

More information

EUEC 2015 Experience using Cavity Enhanced Absorption Spectroscopy (CEAS) for HCl measurement on a coal-fired power plant

EUEC 2015 Experience using Cavity Enhanced Absorption Spectroscopy (CEAS) for HCl measurement on a coal-fired power plant Steve Gibbons, Business Development - CEMS EUEC 2015 Experience using Cavity Enhanced Absorption Spectroscopy (CEAS) for HCl measurement on a coal-fired power plant Introduction Steve Gibbons Business

More information

Image Guided Proton Therapy and Treatment Adaptation

Image Guided Proton Therapy and Treatment Adaptation Image Guided Proton Therapy and Treatment Adaptation www.hollandptc.nl d.r.schaart@tudelft.nl Cancer in The Netherlands About 1 in 3 people get cancer in some stage of their life 86.800 new cancer patients

More information

Recent Advances in Breast Radiotherapy

Recent Advances in Breast Radiotherapy Recent Advances in Breast Radiotherapy Dr Anna Kirby Consultant Clinical Oncologist 2 The Royal Marsden Overview: Key innovations 1. 2D to 3D planning 2. Hypofractionation 3. Intensity modulated radiotherapy

More information

S1. Derivatization efficiency tests

S1. Derivatization efficiency tests Supplement of Atmos. Meas. Tech., 7, 4417 4430, 2014 http://www.atmos-meas-tech.net/7/4417/2014/ doi:10.5194/amt-7-4417-2014-supplement Author(s) 2014. CC Attribution 3.0 License. Supplement of Online

More information

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland

IGRT1 technologies. Paweł Kukołowicz Warsaw, Poland IGRT1 technologies Paweł Kukołowicz Warsaw, Poland Minimal prerequisite for good, efficient radiotherapy ICTP 2015 Paweł Kukołowicz 2/29 Minimal prerequisite for good, efficient radiotherapy Well trained

More information

Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract

Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract Progress Report No. 2-3, March 31, 1999 The Home Automation and Healthcare Consortium Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy Prof. Kamal Youcef-Toumi Principal Investigator

More information

Collapsed Cone Convolution 2D illustration

Collapsed Cone Convolution 2D illustration Collapsed Cone Convolution 2D illustration 8 cones Energy desposition decreases very quickly with distance Energy is absorber in blue pixels only. IGRT1 technologies Paweł Kukołowicz Warsaw, Poland IGRT

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Fundamentals of Soft Ionization and MS Instrumentation

Fundamentals of Soft Ionization and MS Instrumentation Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:

More information

IES LM Test Report

IES LM Test Report IES LM 79 08 Test Report Test results reported for: B 3 DL report: LEAG002 010 Original issue date: 01 May 2012 Prepared for: Testing performed by: The Leader Leader Mfg., Inc 1003 7th Ave. 1540 Leader

More information

Remote Healing Measured by the Biofield Meter

Remote Healing Measured by the Biofield Meter Remote Healing Measured by the Biofield Meter Buryl Payne, Ph. D. Psychology, Psychophysics Laboratories, P.O. Box 514, Soquel, CA. 95073 Abstract A force around the human body different from electric,

More information

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies

Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation

More information

Zakithi Msimang SAAPMB/SARPA 2011

Zakithi Msimang SAAPMB/SARPA 2011 THE IAEA/WHO TLD POSTAL DOSE AUDIT SERVICE Zakithi Msimang SAAPMB/SARPA 2011 Introduction The idea of organizing a dosimetry audit for radiotherapy centres by the IAEA, was suggested for the first time

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement

More information

The merits of mental age as an additional measure of intellectual ability in the low ability range. Simon Whitaker

The merits of mental age as an additional measure of intellectual ability in the low ability range. Simon Whitaker The merits of mental age as an additional measure of intellectual ability in the low ability range By Simon Whitaker It is argued that mental age may have some merit as a measure of intellectual ability,

More information

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry

Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative

More information

SRS Uncertainty: Linac and CyberKnife Uncertainties

SRS Uncertainty: Linac and CyberKnife Uncertainties SRS Uncertainty: Linac and CyberKnife Uncertainties Sonja Dieterich, PhD Linac/CyberKnife Technological Uncertainties 1 Linac Mechanical/Radiation Isocenters Depuydt, Tom, et al. "Computer aided analysis

More information

EMDOLA Education Program: First Year Details. History of lasers: Optics data : The quantum nature of the light:

EMDOLA Education Program: First Year Details. History of lasers: Optics data : The quantum nature of the light: EMDOLA Education Program Module 1: Optics Module 2: Physics of lasers Module 3:Interaction laser-tissues Module 4: Laser Safety Properties of lasers Module 5:Lasers conservative dentistry Laser in caries

More information

Why Measure Aerosol Fluxes?

Why Measure Aerosol Fluxes? Application of aerosol mass spectroscopy for eddycovariance measurements of speciated particle fluxes Eiko Nemitz, Gavin Phillips, Rick Thomas: Centre for Ecology and Hydrology (CEH), Edinburgh, U.K. JoseLuis

More information

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10

GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10 GEX Recommended Procedure Eff. Date: 07/27/07 Rev.: C Pg. 1 of 10 NOTICE: This document is version controlled and was produced as a part of the GEX Information Program which requires that all Series 100

More information

Internal Validation Guide of Y-STR Systems for Forensic Laboratories Printed in USA. 11/12 Part# GE713

Internal Validation Guide of Y-STR Systems for Forensic Laboratories Printed in USA. 11/12 Part# GE713 REFERENCE MANUAL Internal Validation Guide of Y-STR Systems for Forensic Laboratories 11/12 Internal Validation Guide of Y-STR Systems for Forensic Laboratories PLEASE DISCARD PREVIOUS VERSIONS. All technical

More information

Routine Quality Assurance Cookbook

Routine Quality Assurance Cookbook This Cookbook is a companion guide to the AIUM Routine Quality Assurance (QA) for Diagnostic Ultrasound Equipment document, which outlines the basic QA requirements for AIUM-accredited practices. The Guide

More information

Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately.

Introduction. The goal of TRUS QA is to ensure your system can do all of this accurately. TRUS QA Workshop Introduction The goals of using TRUS in prostate brachytherapy Visualize the prostate Need the US to penetrate deeply enough Need sufficient grey scale resolution to be able to visualize

More information

General Single Ion Calibration. Pete 14-May-09

General Single Ion Calibration. Pete 14-May-09 General Single Ion Calibration Pete 14-May-09 Purpose of SI calibration Measure the instrument response of a single ion. Necessary for understanding of error in instrument (counting statistics) Calculation

More information

Multi-Channel Film Dosimetry and Gamma Map Verification

Multi-Channel Film Dosimetry and Gamma Map Verification Multi-Channel Film Dosimetry and Gamma Map Verification Micke A International Specialty Products Ashland proprietary technology, patents pending Single Channel Film Dosimetry Calibration Curve X=R R ave

More information

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood

Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system. Dr Craig Moore & Dr Tim Wood Cone Beam CT Protocol Optimisation for Prostate Imaging with the Varian Radiotherapy OBI imaging system Dr Craig Moore & Dr Tim Wood Background With the increasing use of CBCT imaging alongside complex

More information

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv

More information

EXPOSURE SCHEDULES FOR SUNTANNING PRODUCTS

EXPOSURE SCHEDULES FOR SUNTANNING PRODUCTS EXPOSURE SCHEDULES FOR SUNTANNING PRODUCTS It is generally accepted that proper use of indoor tanning methods may be safer for the tanner than the uncontrolled environment of outdoor tanning. However,

More information

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL

Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Implementation of the 2012 ACR CT QC Manual in a Community Hospital Setting BRUCE E. HASSELQUIST, PH.D., DABR, DABSNM ASPIRUS WAUSAU HOSPITAL Conflict of Interest Disclaimer Employee of Aspirus Wausau

More information

OIML R 132 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 132 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 132 Edition 2001 (E) Alanine EPR dosimetry systems for ionizing radiation processing of materials and products Systèmes de dosimétrie à alanine pour le traitement par

More information

DOSE MEASUREMENTS ON BOARD THE ISS WITH THE PILLE TL SYSTEM

DOSE MEASUREMENTS ON BOARD THE ISS WITH THE PILLE TL SYSTEM DOSE MEASUREMENTS ON BOARD THE ISS WITH THE PILLE TL SYSTEM I. Apáthy 1,S. Deme 1,P. Szántó 1, L. Bodnár 2, Y. A. Akatov 3, V. V. Arkhangelsky 3 I. Nikolaev 4 1 HAS KFKI Atomic Energy Research Institute,

More information

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19,

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19, MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH December 19, 2014 1 The principle of MALDI imaging Spatially resolved mass spectra are recorded Each mass signal represents

More information

An innovative method, the dust flux one allows an evaluation of the short ragweed fight of a municipality (or other administrative structures)

An innovative method, the dust flux one allows an evaluation of the short ragweed fight of a municipality (or other administrative structures) An innovative method, the dust flux one allows an evaluation of the short ragweed fight of a municipality (or other administrative structures) Chantal Déchamp 1, Henriette Méon 1,2 1 French foundation

More information

Persistence in the WFC3 IR Detector: Intrinsic Variability

Persistence in the WFC3 IR Detector: Intrinsic Variability Persistence in the WFC3 IR Detector: Intrinsic Variability Knox S. Long, & Sylvia M. Baggett March 29, 2018 ABSTRACT When the WFC3 IR detector is exposed to a bright source or sources, the sources can

More information

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG

CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG CHAPTER 2 SIMULTANEOUS DETRMINATION OF ANASTROZOLE AND TEMOZOLOMIDE IN TEMOZOLOMIDE CAPSULES 20 MG AND ANASTROZOLE TABLETS 1 MG ANALYTICAL METHOD VALIDATION REPORT FOR ASSAY 43 2.1 Introduction Analytical

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.

Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,

More information

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance

Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 3, 2015 Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance Ganesh Narayanasamy, Travis Zalman, Chul S. Ha,

More information

Chandra Users Committee. CXC Manager s Status Report

Chandra Users Committee. CXC Manager s Status Report Chandra Users Committee CXC Manager s Status Report For the period October 2015 September 2016 Roger Brissenden CXC Deputy Manager 27 September 2016 Topics Program Management Mission Operations Spacecraft

More information

Luminosity Monitor and Test Run Results. Presented by: Luis Mercado UMASS Amherst 7/23/08

Luminosity Monitor and Test Run Results. Presented by: Luis Mercado UMASS Amherst 7/23/08 Luminosity Monitor and Test Run Results Presented by: Luis Mercado UMASS Amherst 7/23/08 Overview Luminosity Monitors Lumi Setup Components PMT Linearity Motivation/Formalism Experimental Test Setup Acceptable

More information

SOLASE DENTAL DIODE LASER

SOLASE DENTAL DIODE LASER TECHNICAI SPECIFICATIONS Wavelength Power Output Pulse Duration Pulse Repetition Rate Tip Diameter Power Supply Input Battery Pack Size / Weight Two models, 808nm or 976nm 976nm: 0.-6W@pulse mode, 0.-0W@CW

More information

MEGGITT SENSING SYSTEMS Endevco short course guide

MEGGITT SENSING SYSTEMS Endevco short course guide MEGGITT SENSING SYSTEMS Endevco short course guide Shock and vibration measurement technology I Shock and vibration measurement technology II Instructors: Jon Wilson, Jon S. Wilson Consulting Michael Phan,

More information

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau

2018 Council on Ionizing Radiation Measurements and Standards. Low Energy ( kev) Electron Beam Calibration. Gary Pageau 2018 Council on Ionizing Radiation Measurements and Standards Low Energy (80-300 kev) Electron Beam Calibration Gary Pageau GEX Corporation Centennial, Colorado USA ISO 9001 certified manufacturer of dosimetry

More information

Sam Howard PhD Tests & measurements, Onda Corporation. Timothy Demchek PhD - Human tissue heating, Indian State University

Sam Howard PhD Tests & measurements, Onda Corporation. Timothy Demchek PhD - Human tissue heating, Indian State University Should you be using Therapeutic Ultrasound and how can we improve our research Stephen J. Straub PhD, ATC Thanks Lenn Johns PhD - cell biologist, QU Sam Howard PhD Tests & measurements, Onda Corporation

More information

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry

Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Measurement Guided Dose Reconstruction (MGDR) Transitioning VMAT QA from phantom to patient geometry Raj Varadhan, PhD, DABMP Minneapolis Radiation Oncology Conflict of interest None Acknowledgement: Jim

More information

Dosisverifikation mit Delta 4 Discover während der Behandlung

Dosisverifikation mit Delta 4 Discover während der Behandlung Dosisverifikation mit Delta 4 Discover während der Behandlung Anders Adolfson, Regional Sales Manager at Scandidos AK IMRT 2015 in Erlangen, March 19 th 2015 Agenda Which problem does Delta 4 Discover

More information

Experimental Objectives

Experimental Objectives opyright Waters orp. 2 onen 1_1519 by apillary HPL Steven A. ohen and ozette M. uppett Waters orporation, 34 Maple Street, Milford, MA, 1757 Experimental bjectives Study sensitivity limits in capillary

More information

MEDIUM-FLOW PNEUMOTACH TRANSDUCER

MEDIUM-FLOW PNEUMOTACH TRANSDUCER MEDIUM-FLOW PNEUMOTACH TRANSDUCER SS11LA for MP3x and MP45 System TSD117 & TSD117-MRI for MP150/MP100 System RX117 Replacement Airflow Head See also: AFT series of accessories for airflow and gas analysis

More information

Predicting Sleep/Wake Behaviour in Operational Settings

Predicting Sleep/Wake Behaviour in Operational Settings Predicting Sleep/Wake Behaviour in Operational Settings Peter Page March 2017 1 InterDynamics Background: Not Researchers We specialise in Decision Support Solutions in particular we provide a full suite

More information

Feng Xiujuan National Institute of Metrology (NIM),China

Feng Xiujuan National Institute of Metrology (NIM),China The acoustic calibration service in transportation at NIM Feng Xiujuan National Institute of Metrology (NIM),China 1. Calibration requirements 2. Calibration service at NIM 2.1 Microphone 2.2 Ultrasonic

More information

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

OIML R 131 RECOMMENDATION. Edition 2001 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION INTERNATIONAL RECOMMENDATION OIML R 131 Edition 2001 (E) Polymethylmethacrylate dosimetry systems for ionizing radiation processing of materials and products Systèmes de dosimétrie à polyméthylméthacrylate

More information

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique

More information

Fundus Autofluorescence

Fundus Autofluorescence Brittany Bateman, BS Fundus autofluorescence imaging is used to record fluorescence that may occur naturally in ocular structures or as a byproduct of a disease process. This technique allows the topographic

More information

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions)

Aer ODE B3-20. Database System for GEX B3 Dosimetry. Aérial. Optical Dosimetry Equipment. (gamma and/or electron beam versions) B3-20 Database System for GEX B3 Dosimetry (gamma and/or electron beam versions) and GEX combine to provide a Genesys 20 version of the flexible 21 CFR Part 11 certified database dosimetry system for use

More information

HCS 7367 Speech Perception

HCS 7367 Speech Perception Long-term spectrum of speech HCS 7367 Speech Perception Connected speech Absolute threshold Males Dr. Peter Assmann Fall 212 Females Long-term spectrum of speech Vowels Males Females 2) Absolute threshold

More information