Beijing, China, East, Dongcheng District, North Third Ring Road. Beijing Global Trade Center #36

Size: px
Start display at page:

Download "Beijing, China, East, Dongcheng District, North Third Ring Road. Beijing Global Trade Center #36"

Transcription

1 Ref: ES B 2 EN A France Technopôle de l Aube BP Troyes Cedex 9 France +33 (0) China Unit 08, Level 16, Building A, Beijing Global Trade Center #36 North Third Ring Road East, Dongcheng District, Beijing, China, Brazil Rua Bela Vista, No 77, Centro Sao Bernardo do Campo SP, CEP Brazil United States 4030 West Braker Lane, Suite 360 Austin, Texas LDR, LDR Spine, LDR Médical, BF+, BF+(ph), Easyspine, Laminotome, MC+, Mobi, Mobi-C, Mobi-L Mobidisc, ROI, ROI-A, ROI-MC+ and ROI-T are trademarks or registered trademarks of LDR Holding Corporation or its affiliates in France, the United States and other countries. P E D I C L E S C R E W S Y S T E M

2 Designed by leading spine surgeons, the Easyspine Pedicle Screw System features a simplified surgical technique and adaptable implants to accommodate varied anatomies Unique self-contained multi-axial connection Multi-axial joint is integrated into the pedicle screw Locking component is pre-assembled A new standard: the LDR flattened rod Variable stiffness with a constant diameter Choice of stiffness Flat-on-flat connection Protection safety stop Facilitates restoration of lordosis Mechanical optimization Reliable locking (flat-on-flat) Multi-axial freedom of movement Multi-axial locking Excellent physiological mechanical properties Sterility and traceability Sterile packaging Absolute traceability Optimized management Table of Contents Static Tests - Slippage Static Tests - Multi-axial Resistance Rod Stiffness Rod Deflection Static Testing Static and Dynamic Construct Testing Competitive Comparison A Summary and Comparison of Results for the Mechanical Testing Performed on the Easyspine System

3 Static Tests - Slippage To determine the slippage resistance of the Easyspine Rod-Screw Connection, the product was tested according To ASTM The worst case size was determined to be the Rigid (R) rod, since this size has the least surface contact. The construct was assembled with a torque of 8.5 N.m (75 in-lbs) according to the surgical technique. R Rod 14.0mm 2 M Rod 18.8mm 2 S Rod 19.6mm 2 Contact Surface The Easyspine multi-axial connection was able to withstand an axial load of 1400 N (314 lbs). Static Tests - Multi-axial Resistance F The Multi-axial resistance of the Easyspine rod/screw connection was tested and compared to published results of competitive systems*. The system was tested in a worse case scenario by measuring the force required to move the Rod/Screw angle from 80 to 90. It is important to note that the diameter of the screws for the Easyspine system was less than competitive systems tested (6.0mm vs for Stryker/DePuy). In addition the locking torque for Easyspine was 8.5 N.m (75 in-lbs), which is 30-50% less than that of competitive systems (13 N.m for Stryker and 15 N.m for DePuy). 80 Screw System Load to Failure (N) Screw Diameter (mm) Silhouette (Zimmer) BMI (Blackstone Medical) Moss Miami (DePuy) M8 (Medtronic) Click X (Synthes) SD 90 (Surgical Dynamics) Easyspine Alpha Screw (LDR) Xia (Stryker) Monarch (DePuy) Magnum (DePuy) *Guy R. Fogel, MD, Charles A. Reitman, MD, Weiqiang Liu, PhD, and Stephen I. Esses, MD. Physical Characteristics of Polyaxial-headed Pedicle Screws and Biomechanical Comparison of Load with their Failure. SPINE Volume 28, Number 5, pp (2003) 30 mm The Maximum Load for the Easyspine Alpha Screw without failure was determined to be 375 N (85 lbs). 2-3

4 Rod Stiffness LDR Rods Conventional Rods Comparison of Stiffness Between LDR Rods and Conventional Round Rods h Ø R Rod: M Rod: S Rod: h = 5.6mm h = 5.0mm h = 4.0mm Ø 5.9mm Ø 5.5mm Ø 4.6mm Elastic Limit for LDR Rods (Newton) R Rod M Rod Stiffness100% Stiffness 80% S Rod Stiffness 50% The Easyspine rods (R, M, S) are roughly equivalent to traditional round rods that are 5.9mm, 5.5mm, and 4.6mm. Rod Deflection Static Testing 5 million dynamic compression cycles were performed with a cyclic load ranging from 28 to 280 N according to ASTM F These tests were performed on the R Rods. Note that up to 280 N, the M and S rods are still within the elastic deformation zone. 35 mm F Blocking system Displacement After 5 million dynamic compression cycles there were no failures. 4-5

5 Static and Dynamic Construct Testing The Easyspine system construct was tested in accordance with Vertebrectomy Model defined in ASTM F Static Test To determine the elastic limit of a standard assembly Dynamic Test To determine a maximum load applied for a fixed number of cycle (5 Million Cycles) Locking torque 8.5 N.m With a loading cycle of 20 N to 200 N, the Easyspine Alpha Screw withstood 5 million cycles without breakage. Competitive Comparison It is important to note that the Easyspine system uses (as per the surgical technique) a lower locking torque than competitive systems (13 N.m for Stryker Screws vs. 8.5 N.m for LDR Screws). In addition, the diameter of the screws is less in the Easyspine testing than competitive systems tested (6.5mm for Stryker Screws vs. 6.0mm for LDR Screws). Static Testing Results Chart Screw System Elastic Limit (N) Screw Diameter Locking Torque Xia (Stryker Spine) mm 13 N.m Moss Miami Ti (DePuy) mm 10/15 N.m Tenor (Medtronic) mm 10 N.m Moss Miami SS (DePuy) mm 10/15 N.m Synergy VLS closed mm 14 N.m (Biomet) Synergy VLS open mm 11 N.m (Biomet) Easyspine Alpha Screw (LDR) mm 8.5 N.m Dynamic Testing Results Chart Screw System Load Screw diameter Xia (Stryker Spine) 220 N: Failure 6.5mm Moss Miami Ti (DePuy) 208 N: Failure 6.0mm Synergy VLS Closed (Biomet) 199 N: Failure 6.5mm Synergy VLS Open (Biomet) 219 N: Failure 6.5mm Easyspine Standard Screw (LDR) 280 N: No Failure 6.0mm Easyspine Alpha Screw (LDR) 200 N: No Failure 6.0mm Ralph Edward Stanford, FRACS, PhD, Andreas Herman Loefler, FRACS, Philip Mark Stanford, DipAppSci, and William R. Walsh, PhD, Multiaxial Pedicle Screw Designs: Static and Dynamic Mechanical Testing, SPINE Volume 29, Number 4, pp (2004) The Easyspine Standard Screw sustained a dynamic testing load of 280 N (63 lbs) without failure, outlasting competitive systems that failed under lower loads, all while using 30-50% less locking torque. 6-7

6 A Summary and Comparison of Results for the Mechanical Testing Performed on the Easyspine System Presented by: LDR Executive Summary The LDR Easyspine Pedicle Screw System was subjected to testing that confirmed its equivalence to other systems in the market. In the US, the Easyspine System is cleared for use as a posterior pedicle screw system intended for immobilization and stability of spinal segments when used as an adjunct to fusion in the thoracic, lumbar and sacral spine. There are specific tests required by the FDA and there are other industry-recognized tests that are performed to further distinguish a particular system over the competition. Testing of the LDR Easyspine Pedicle Screw System clearly demonstrated that it meets the standards for static and dynamic compression, as well as torsional stability characteristics. When compared to peerreviewed, published test results for other popular systems, Easyspine compares very favorably. When evaluated and compared for more specific attributes, the LDR Easyspine system demonstrates superior results. For example, load-to-failure testing clearly demonstrated that the LDR polyaxial screw has the highest resistance to shear failure of any published polyaxial screw in the market, having comparable thread diameters. In other words, the Easyspine screw is less likely to shear at the bone interface. Due to the unique rod design with flat locking surfaces, the LDR Easyspine system provides three rod stiffness options, with 1/3rd the inventory. Every rod in the system is compatible with every available diameter of Easyspine screw. In addition, the flat locking surfaces and fine thread in the integrated locking mechanism provide higher assurance of a secure lock with the polyaxial tightening screw at significantly lower torque values than any other system in the market. The patented Easyspine system provides all of the basic mechanical characteristics offered by competitive systems, yet clearly goes beyond those systems to provide much more in the form of assurance of system flexibility and relative mechanical properties. Test Methods To date, the Easyspine System has been subjected to all required tests and clearly demonstrated both its equivalence and superiority to other polyaxial pedicle screw fixation systems currently available in the market. In addition, the system has been subjected to additional testing and evaluation techniques to further distinguish itself as a better choice for a fusion system. Below is a description of the testing: ASTM F Standard Test Methods for Spinal Constructs in a Vertebrectomy Model: a) These test methods cover the materials and methods for the static and fatigue testing of spinal implant assemblies in a vertebrectomy model. b) These test methods are intended to provide a basis for the mechanical comparison among past, present, and future spinal implant assemblies. They allow comparison of spinal implant constructs with different intended spinal locations and methods of application to the spine. These test methods are not intended to define levels of performance. c) These test methods set out guidelines for load types and methods of applying loads. Methods for three static load types and one fatigue test are defined for the comparative evaluation of spinal implant assemblies. d) These test methods establish guidelines for measuring displacements, determining the yield load, and evaluating the stiffness and strength of the spinal implant assembly. There are essentially four tests defined by this ASTM standard which have been prescribed to evaluate the spinal implant assemblies. These are the same tests used to determine substantial equivalence to previously cleared spinal construct systems: three static mechanical tests and one dynamic test. The three static mechanical tests are 1) compression bending, 2) tensile bending, and 3) torsion. The dynamic test is 4) compression bending fatigue. Due to variability in designs and intended application of components, the ASTM test standards have been constructed with the intent to minimize variations when systems are compared to one another for potential in-vivo performance. They are only intended for comparison of relative mechanical parameters. Static Compression Bending Test: Run with a transverse bar to simulate the potential for additional fatigue generated at the site of the cross-connector clamp. (It is more critical to demonstrate this requirement in the compression fatigue test, but is usually utilized in the static test for consistency of test set-up). Static Tension Bending Test: Used to further differentiate fatigue properties of the polyaxial head, when needed. In general, this test is specific to cervical devices (typically plates) and not normally utilized for evaluation of thoraco-lumbar fixation systems. Static Torsional Test: Run without a transverse bar to simulate the weakest configuration - without additional torsional stability provided by a cross-connection (transverse bar). Compression Bending Fatigue Testing: Run with a transverse bar to simulate the potential for additional fatigue generated at the site of the cross-connector clamp. (As always, all testing should be done with the weakest and/or worst case configuration). The effect of environment may be significant. In order to provide consistent, relevant comparisons, these tests should be initially performed dry (ambient room conditions). The maximum recommended frequency for this type of cyclic testing should be 5 Hz. All testing should have a minimum of five (5) samples. The fatigue testing should result in at least two samples surviving 5,000,000 (minimum) cycles at a load with a difference of no more than 10% of the (static) compression bending ultimate load. Designation: F (Reapproved 2003): Standard Guide for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants a) This standard covers the measurement of uniaxial static and fatigue strength, and resistance to loosening of the component interconnection mechanisms of spinal arthrodesis implants. b) The purpose of this standard is to provide a means of mechanically characterizing different designs of spinal implant interconnections. Ultimately, the various components and interconnections should be combined for static and fatigue testing of the spinal implant construct, (ASTM F ). Polyaxial Locking - Load to Failure: - (Variant of ASTM F : Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants): A test of the physical characteristics of the polyaxial locking mechanism and resistance to failure. In this case, the objective for Easyspine was to determine the (worst case scenario) load at which the recommended locking torque for the polyaxial screw would fail to hold the associate spinal rod when the screw was subjected to a cantilever load. Static Resistance to Sliding (Static Axial Grip Capacity Test) - (Variant of ASTM F : Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants): This test was performed to determine the amount of axial load (applied along the longitudinal axis of rod) that would be required to cause the rod to slip in the screw, after the polyaxial screw was tightened to the manufacturer s recommended torque. Dynamic Resistance to Sliding (Dynamic Axial Grip Capacity Test) - (Variant of ASTM F : Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants): This test was performed to determine the amount of dynamic axial load (applied along the longitudinal axis of rod) that would be required to cause the rod to slip in the screw, after the polyaxial screw was tightened to the recommended torque. Loads were determined based on a percentage of maximum static load achieved in the previous test. Significance and Use of ASTM standards Spinal implants are generally composed of several components (screws, hooks, rods, etc.) which, when connected together, form a spinal implant assembly. Spinal implant assemblies are designed to provide some stability to the spine while arthrodesis takes place. The ASTM test methods outline standard materials and methods for the evaluation of different spinal implant assemblies so that comparison between different designs may be facilitated. These test methods are used to quantify the static and dynamic mechanical characteristics of different designs assemblies. The mechanical tests are conducted in vitro using simplified load schemes and do not attempt to mimic the complex loads of the spine. The results obtained cannot be used directly to predict in vivo performance. However, the results can be used to further distinguish a system when compared to other system designs in terms of their relative mechanical potential. Finite Element Analysis: This is a computer simulation to demonstrate the stress magnitude and pattern that develops in each (screw) design when tested and validated per the ASTM 1717 benchmark for design performance. Rod Flexion / Stiffness Analysis: Is a measure of resistance to bending. The Elastic Limit is the greatest stress that can be applied to a material without causing permanent (bending) deformation. In the case of Easyspine, the rod stiffness (elastic limit) was measured to allow comparison to conventional (round) rods made from the same material (Ti6Al4V). 8-9

7 ASTM : Standard Test Methods for Spinal Constructs in a Vertebrectomy Model Static Compression Bending (Std. Screws) I MTS Scale Used 0-25kN Ø6 x L40mm Std. Screw Control: Displacement Speed: 10mm / min Screw Tightening Torque: 8.5 N.m Additional Static Compression Bending (Alpha Screws) II MTS 810 Scale Used 0-25kN Ø6 x L40mm Alpha Screw Control: Displacement Speed: 10mm / min Screw Tightening Torque: 9.5 N.m Specimens Test # Maximum Displacement Displacement Stiffness Load (mm) 2% of to 2% of Elastic (N) 1 Elastic Limit (mm) 2% offset Limit (N) yield Ø6 x L40mm mm R-rod Ø6 x L40mm mm R-rod Ø6 x L40mm mm R-rod Mean Specimens # of Test Mean Value of Yield Bending Mean Value of Maximum Samples Load on the Polyaxial Recorded Stress Connection (N) (N) Ø6 x L40mm Alpha Screw w/ mm R-rod Static Torsion Test (Std. Screws) III MTS Mini Bionix II Ø6 x L40mm Std. Screw Control: Angulation Speed: 10º / min Screw Tightening Torque: 8.5 N.m Specimens Test # Maximum Maximum Torque Angulation Stiffness Torque Torque to 2% to 2% of Elastic (N/º) (Nm) Angulation of Elastic Limit (mm) ( ) Limit (Nm) Ø6 x L40mm mm R-rod Ø6 x L40mm mm R-rod Mean (Dynamic) Compression Bending Fatigue Test (Std. Screws) IV ESM with a capacity of 10kN Ø6 x L40mm Std. Screw Control: Load Load Applied: 50% 25% % % of max static compression bending load (1118 N) Ratio: 0.1 Frequency: 5 Hz Duration: 5,000,000 cycles or failure of implant assembly Screw Tightening Torque: 8.5 N.m Specimens Test # % of Load Amplitude Numbers Type of Maximum Cycles of of Cycles Failure Static Load Applied Displacement Achieved Observed Obtained (N) (mm) Ø6 x L40mm to ,486 Broken Screw 100mm R-rod to ,000,000 No Failure to ,759,153 Broken Screw to ,000,000 No Failure to ,785,253 Broken Rod to ,000,000 No Failure Additional (Dynamic) Compression Fatigue Testing (Alpha Screws) V ESM with a capacity of 10kN Ø6 x L40mm ALPHA Screw Control: Load Load Applied: 50% - 25% % % of max static compression bending load (1118 N) Ratio: 0.1 Frequency: 5 Hz Duration: 5,000,000 cycles or failure of implant assembly Screw Tightening Torque: 8.5 N.m Specimens Test # % of Load Numbers Type of Maximum Cycles of Cycles Failure Static Load Applied Achieved Observed Obtained (N) Ø6 x L40mm to 100 5,000,000 No Failure Alpha Screw w/ 90mm R-rod to 150 5,000,000 No Failure to 200 5,000,000 No Failure to , to ,829 Broken Screw at PE Block Interface 2 Broken Screws at PE Block Interface N = Newton: A metric measurement of Force. (1 N lbs.; or, 1 lb N)

8 Comparison of Easyspine Test Results to Literature : For direct comparison of results between the Easyspine System and other commercially available polyaxial pedicle screw systems, we turn to literature. The most relevant recent reference is by Stanford, et al., published in Spine in February, 2004 vi. Table 1 - Specifications and Construction Torques for Screw Assemblies v * Tenor design locked by failure of a shear nut. # Root diameter of screw measured immediately below rod screw link, this either tapered to a smaller diameter at the tip or was constant to the tip (straight). Torque applied to locking screw for the Alpha Screw design only. Torque applied to inner and outer locking nuts, respectively. Screw Design Manufacturer Material Outer Screw Root Screw) Rod Locking Diameter (mm) Diameter (mm) Diameter (mm) Torque (Nm) Easyspine Std. & LDR Titanium alloy tapered / 9.5 Alpha Screws Tenor Medtronic Titanium alloy tapered * Moss Miami (SS) DePuy Stainless Steel straight /15 Moss Miami (Ti) DePuy Titanium alloy straight /15 Xia Stryker Spine Titanium alloy tapered Synergy VLS Biomet Titanium alloy tapered (Open) Synergy VLS Biomet Titanium alloy tapered (Closed) Table 2 - Static and Dynamic Compression Structural Properties of Assemblies by Screw Design Static Data Dynamic Cycles to Failure (at described loads x 10 3 cycles) Screw Screw 2% Ultimate 50% Load 75% Load Design Diameter (mm) 2% Offset of Elastic Bending Load Actual Load Actual Load Yield (N/mm) Limit (N) (N) Easyspine N 420 N Std. Screw (5,000) (3759) Tenor N (5,000) 498 N (4,719) Moss Miami (SS) Moss Miami (Ti) N (4,280) 312 N (42) Xia N (4,394) 337 N (106) Synergy VLS (Open) N (5,000) 219 N (987) Synergy VLS (Closed) N (532) 298 N (110) Alpha Screw Screw Yield Ultimate 50% Load * 79% Load * Design Diameter 2% Offset Bending Bending Actual Load Actual Load (mm) Yield (N/mm) Load (N) Load (N) Easyspine Alpha 6.0 Not Recorded N 250 N Screw Design (5,000) (643) Table 3 - Number and Mode of Failure of Screw Designs in Static and Compressive Loading to Failure Static Compression Failure Dynamic Compression Failure Screw Number Tested Mode Number Tested Mode Design out of 5 out of 5 Easyspine (α) 5 No Rotation of the Rod 25% Fracture of screw at the PE test block Alpha 6.0mm VII Longitudinal Slip of 50% interface, no slip of link or fretting. Rod/screw Link Interface 63% Rods tipped in α-screw cradle at 10º 75% in relation to screw axis. Tenor 1 Rotational Slip of Rod/screw Link 25% Fracture of rod adjacent to 50% rod / screw link, 75% no slip of link or fretting. Moss Miami (SS) 5 Rotational Slip of Rod/screw Link Moss Miami (Ti) 5 Rotational Slip of Rod/screw Link 25% Fracture of neck of screw, 50% just below rod / screw link, 75% no slippage or fretting. Xia 5 Rotational Slip of Rod/screw Link 25% Rotational slip of rod / screw link, 50% with cracking of neck of 75% screw adjacent to link. Synergy VLS 5 Rotational Slip of Rod/screw Link 25% Rotational slip of rod / screw link with (Open) 50% cracking of link casing and fretting 75% of serrated components. Synergy VLS 5 Rotational Slip of Rod/screw Link 25% Rotational slip of rod / screw link with (Closed) 50% cracking of link casing and fretting 75% of serrated components. Discussion of Results from Typical Static and Dynamic Testing : By itself, the raw data is not really that telling, however, when one compares the results of the Easyspine system with the data from various well known systems available in the market, one begins to note that the Easyspine screws provide significant advantages in terms of relative mechanical parameters. For example, both the Alpha and Standard (6.0mm) Easyspine screw designs have a higher static yield bending load than any other 6.0mm screw, while the Standard design has significantly better properties in nearly every category. When compared to conventional rod and screw systems tested by independent evaluators and published in peer-reviewed journals, the Easyspine System is the only system that had no rotational slippage between the rod and locking screw during static or dynamic testing. This could imply that the surgeon should be less concerned about the potential for de-rotation of the spinal segments when using the Easyspine System. As noted previously, these test methods are used to further quantify the static and dynamic mechanical characteristics of design assemblies. These mechanical tests are also conducted in vitro using simplified load schemes and cannot mimic the complex loads of the spine. The results obtained cannot be used directly to predict in vivo performance. However, the results can be used to further distinguish a system when compared to other system designs in terms of their relative mechanical potential. Slippage Tests ASTM F : Standard for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants A slippage test demonstrates the ability of the system to resist movement between the rod and the locking screw once the system is assembled with the appropriate tightening torque. Assembly was tested according to ASTM ALL tests were conducted made with the R Rod *** and a locking torque of only 8.5 N.m. Additionally, one should note that these results were obtained using the Easyspine polyaxial screw that has a self-contained locking screw that requires a lower tightening torque than most (if not all) other screws in the market

9 Illustrations of the three types of rods are provided below. *** The R rod has the smallest contact surface area (and highest stiffness) for the rods in the Easyspine system. R Rod 14.0mm 2 (Rigid) M Rod 18.8mm 2 (Medium) S Rod 19.6mm 2 (Slender) Contact Surface Static Resistance to Sliding Test - (Rod / Std. Screw Axial Gripping Capacity) MTS Scale Used 0-25kN Control: Displacement Speed: 2mm/min Screw Tightening Torque : 8.5 N.m Specimens Test # Maximum Load Displacement of Obtained (N) Maximum Load (mm)** Ø6 x L40mm mm R-rod Mean ** Displacement of Maximum Load: Failure of grip strength was deemed to have occurred when the rod was able to slip under the polyaxial head a distance of approx. 0.2mm (or the approx. equivalent to 2.5 x the thickness of a human hair). (Dynamic) Resistance to Sliding Test - (Rod / Std. Screw Axial Gripping Capacity) ESM Scale Used 0-25kN Control: Load Load Applied: 75% - 80% - 85% of max static load (1399 N) Ratio: 0.1 Frequency: 5 Hz Duration: 2,500,000 cycles or failure of implant / connection Screw Tightening Torque: 8.5 N.m Specimens Test # % of Load Numbers Type of Maximum Cycles of Cycles Failure Static Load Applied Achieved Observed Obtained (N) Ø6 x L40mm 1 75% 100 to ,500,000 No Break or Sliding 60mm R-rod 2 80% 110 to ,500,000 No Break or Sliding 3 85% 120 to ,500,000 No Breakage - Sliding between the Screw and the Screw and Rod occurred at the very end of the test Polyaxial Locking - Load to Failure: for the Alpha (α) Screw These diagrams represent a variation of the screw angle from 80 to 90 respectively. In order to test a worst case loading scenario, it was necessary to determine the load needed to deflect or break the connection between the multi-axial Alpha (α) screw head and rod, from an extreme, locked position (80 ) to a point approaching 90, (the theoretically strongest position). A1 B1 X2 A1 B1 X2 For 5 alpha screws (Ø 6mm), the results of the maximum force required to shift the rod / screw axis angle from 80 to 90 were 481N, 439N, 419N, 187N, and 344N. The average of these 5 results is 375 N. It should be noted that there was no breakage of the screw or polyaxial mechanism, only the locking torque resistance (8.5 Nm) was overcome. Comparison of Easyspine Results to Peer-Reviewed, Published Data Additional peer-reviewed published literature which provides further direct comparison of load-to failure results between commercially available polyaxial pedicle screw systems can also be found in Spine. The most relevant reference for this comparison is by Fogel, et. al, published in Spine in February, 2003 VIII. Utilizing a modified version of ASTM F , the author compared the loads to failure for nine different polyaxial pedicle screws. For comparison, the test results for the Easyspine Alpha screw is shown in shaded format to the right. Screw System Load to Failure (N) Screw Diameter (mm) Silhouette (Zimmer) BMI (Blackstone Medical Inc) Moss Miami (DePuy) M8 (Medtronic) Click X (Synthes) SD 90 (Surgical Dynamics) Alpha (α) Screw (LDR) Xia (Stryker) Monarch (DePuy) Magnum (DePuy) Analysis and Conclusions from Slippage and Locking Mechanism Evaluations : One should note that the static load required to overcome the resistance to slippage between the rod and the locking screw (1399 N) for the Std. Easyspine screw, was 25% greater than the static compression bending load (1118 N) noted previously for the Easyspine System. When comparable published data could be found, the Easyspine system demonstrated clearly superior results. The results from the load-to-failure testing of the polyaxial pedicle screws, presented by Fogel et. al,, demonstrates that the weakest point of the construct is the head-to-screw coupling. The author suggests that this failure of the polyaxial head may be a protective factor for the pedicle screw shaft, preventing early breakage. However, the author fails to note that the load-to-failure for most designs of polyaxial screws is in direct correlation to the cross-section of the screw at the polyaxial swivel joint. This cross-sectional diameter is controlled by a function of the minor diameter of the screw thread and the major diameter of the polyaxial head that must swivel over the ball-joint at the end of the screw head. As demonstrated above, all screws tested (with the exception of Moss Miami and Easyspine), were between 6.5mm and 7.5mm in diameter). The Easyspine Screw provides a uniquely different (patented) solution to this limitation by incorporating the polyaxial swivel within the self-contained locking screw in the head of each screw. Subsequently, there is no built-in stress-riser or weak point at the bone interface, (polyaxial ball-joint). The only strength limitation of the screw is the minor diameter of the thread itself, which is stronger than any polyaxial ball-joint found in typical polyaxial screws, similar to those described by the Interpore-Cross / Biomet 555 patent. Additional System-Specific Tests : In support of the analysis and conclusion drawn above, the Easyspine screw design was subjected to a detailed Finite Element Analysis (FEA) to determine the magnitude and pattern of stresses which could develop when subjected to the simulated loading scenario required by the various ASTM testing protocols. Additionally, the Easyspine rods were evaluated for stiffness and compared to conventional round titanium rods. Finite Element Analysis of the Easyspine Polyaxial Screw IX The finite element method was used to evaluate three different pedicle screw design configurations: Alpha size 5, Standard size 5, and Standard size 6. The stress magnitude and pattern that develops in each design when tested per ASTM 1717 was used to benchmark the design performance. Methods: Finite Element models of each design were constructed using LDR CAD files in STEP format. The files were imported into the FEA environment (FEMAP), where the meshing and boundary conditions were applied to simulate ASTM 1717 testing. Because only the screw component of the spinal fixation assembly was simulated, an equivalent load was applied to the screw as determined by rigid body mechanics. All designs were simulated using the same boundary conditions (fixation point and equivalent load). Each FEA model was tested against a rigorous validation and verification protocol. All models demonstrated appropriate accuracy for the required application, with differences from closed form approximations of less than 10%. For the available configurations (size 5 Alpha and size 6 Standard), the FEA correctly predicted the failure location as observed in real-world tests. Complete model details, validation and verification, and results are documented in ProbaSci reports: Finite Element Analysis simulating ASTM 1717 testing on LDR's Pedicle Screw Design Alpha, size 5; Finite Element Analysis simulating ASTM 1717 testing on LDR's Pedicle Screw Design Standard, size 5; Finite Element Analysis simulating ASTM 1717 testing on LDR's Pedicle Screw Design Standard, size 6. Results: Figure 1 A-C shows the FEA predicted stresses for each design. The stress magnitudes in the expected failure zone were: Alpha size MPa, Standard size MPa, and Standard size MPa. All three screw design configurations exhibited the same pattern of stress. The highest stresses were observed in the threaded region, near to the constraint location, with the head regions experiencing much lower stress. When subjected to this loading, all three designs are expected to fail in the same location: in the threaded region, near the constraint point. The difference in the Alpha size 5 and the Standard size 5 is 17 MPa, which is 4% of the predicted stress. This difference is within the estimated model errors and is therefore too small to be practically significant. In other words, the Alpha size 5 and Standard size 5 designs perform equally, within the error of the model. While the Standard size 6 demonstrated similar stress patterns as the Standard 5 and Alpha 5, the magnitude of the stress was about 1/2 of the size 5 designs. It should be noted that these FEA findings represent a very powerful method of analysis. Due to the highly variable nature of real world fatigue tests, an equivalent real world test protocol would require a very large amount of samples in order to detect a 4% difference in design performance

10 Comparison Alpha 5, Standard 5, and Standard 6 Fig. 1- A. (Alpha 5) Fig. 1- B. (Standard 5) Fig. 1 C. (Standard 6) Figure 1: FEA predicted stresses during ASTM 1717 for screw design configurations: (A) Alpha size 5, (B) Standard size 5, (C) Standard size 6. Conclusion: The Alpha size 5, Standard size 5, and Standard size 6 all demonstrate similar FEA predicted stress patterns when simulating ASTM 1717 loading. The Alpha size 5, Standard size 5, and Standard size 6 all demonstrated much lower stresses in the head section than in the threaded section. The Alpha size 5, Standard size 5, and Standard size 6 could all be expected to fail in a similar location that is in the threaded region, near the constraint point. The Alpha size 5 and Standard size 5 demonstrated equivalent stresses. The Alpha size 5 stress was 419 MPa and the Standard size 5 was 402 MPa. The Standard size 6 had approximately 1/2 the stress level as the Standard size 5 and the Alpha size 5 under identical loading. The Standard size 6 stress was 219 MPa. In short: By relocation the polyaxial mechanism to the inside of the screw head, only the minor diameter of the thread affects the strength of the screw construct, not the polyaxial connection, as with all other conventional designs. Finally: In addition to the FEA performed on the screws, the mechanical characteristics of the rods were determined using both mechanical and theoretical analysis methods. LDR Rods - Flexion Test and Comparison Flexion test protocol Flexion Tested Rods Rod's length F 35 mm Blocking system R Rod M Rod S Rod Displacement Flexion test results on LDR Rods (Elastic limit for each type of LDR rods) Load (Newton) R Rod M Rod S Rod - 5 million dynamic compression cycles were performed with a cyclic load ranging from 28 to 280 N according to ASTM F These tests were performed on the "R" Rods. - All LDR rods have a constant Ø6.0mm, with a variable (milled cross-section) height, roughly corresponding to industry accepted diameters for Ø4.0mm, Ø5.0mm, and Ø6.0mm rods - One should note that up to 280N, the "M" and "S" Rods are still within the elastic deformation zone. Comparison of Stiffness Between LDR Rods and Conventional Round Rods In bending, the deflection is determined by the following equation: F: applied force f = Fl 3 / 3EI : f: deflection l: rod's length E: Young modulus I: moment of inertia If one compares the deflection calculation with an LDR Rod and with a round Rod, the only variation is the moment of inertia. The calculation of the moment inertia implies the equivalence of the following stiffness: LDR Rods Conventional Rods Analysis / Comparison of Results for the Easyspine Rods: As noted in the in-vitro results presented above the unique design for the Easyspine system rods will afford essentially equal performance compared to any standard 6.0mm round titanium rod offered by any other system and superior results for stiffness when compared to any other 5.0mm or 4.0mm round rod. Again, one should note that any size rod offered in the Easyspine System can be used interchangeably with any size screw in the system. The net result of this means that the surgeon need never worry about the ability to match the appropriate rod stiffness and fusion stability of a construct to match the need of the patient, regardless of screw size. The results presented cannot be used directly to predict in vivo performance. However, the results can be used to further distinguish a system when compared to other system designs in terms of their relative mechanical potential. Summary: R Rod: M Rod: S Rod: h h = 5.6mm h = 5.0mm h = 4.0mm Ø 5.9mm Ø 5.5mm Ø 4.6mm Ø Testing of the LDR Easyspine System clearly demonstrated that it meets all safety and efficacy standards for static and dynamic compression, as well as torsional stability characteristics. When compared to peerreviewed, published test results for other popular systems, Easyspine compares very favorably. When evaluated and compared for more specific attributes, the LDR Easyspine system demonstrates superior results. For example, load-to-failure testing clearly demonstrated that the LDR polyaxial screw has the highest resistance to shear failure of any published polyaxial screw in the market, having comparable thread diameters. This is further confirmed and validated by FEA. In other words, the Easyspine screw is less likely to shear at the bone interface. Further, testing has demonstrated the ability of the alpha screw to withstand higher failure loads (locking torque resistance) than any other polyaxial screw of comparable diameter. Due to the unique rod design with flat locking surfaces, the LDR Easyspine system provides three rod stiffness options, with 1/3rd the inventory. Every rod in the system is compatible with every available diameter of Easyspine screw. In addition, the flat locking surfaces and fine thread in the integrated locking mechanism provide higher assurance of a secure lock with the polyaxial tightening screw at significantly lower torque values than any other system in the market. The patented Easyspine system provides all of the basic mechanical characteristics offered by competitive systems, yet clearly goes beyond those systems to provide much more in the form of assurance of system flexibility and relative mechanical parameters. Bibliography I CRITT Report # B : Static and Dynamic Tests of Compression and Torsion on Easyspine Implants According to ASTM F 1717 and ASTM II LNE - Laboratoire National D Essais - Test Report: Dossier D Document CQPE/2, April 10, 2003; Mechanical Characteristics of the Easyspine (Ø6.0mm) α-screw by Means of Static and Dynamic Tests. III IV CRITT Report # B : Static and Dynamic Tests of Compression and Torsion on Easyspine Implants According to ASTM F 1717 and ASTM CRITT Report # B : Static and Dynamic Tests of Compression and Torsion on Easyspine Implants According to ASTM F 1717 and ASTM V LNE - Laboratoire National D Essais - Test Report: Dossier D Document CQPE/2, April 10, 2003; Mechanical Characteristics of the Easyspine (Ø6.0mm) α-screw by Means of Static and Dynamic Tests. VI Multiaxial Pedicle Screw Designs: Static and Dynamic Mechanical Testing; Ralph E. Stanford, FRACS, PhD, Andreas H. Loefler, FRACS,, Philip M. Standford, DipAppSci, and William R. Walsh, PhD; SPINE Volume 29, Number 4, pp (2004) VII LNE - Laboratoire National D Essais - Test Report: Dossier D Document CQPE/2, April 10, 2003; Mechanical Characteristics of the Easyspine (Ø6.0mm) α-screw by Means of Static and Dynamic Tests. VIII IX Physical Characteristics of Polyaxial-headed Pedicle Screws and Biomechanical Comparison of Load with their Failure: Guy R. Fogel, MD, Charles A. Reitman, MD, Weiqiang Liu, PhD, and Stephen I. Esses, MD SPINE Volume 28, Number 5, pp (2003) ProbaSci Final Report - June 7, 2005: Comparison of FEA Results for the Alpha Size 5, Standard Size 5, and Standard Size 6 Easyspine Pedicle Screw Designs

CERVICAL MODULAR CAGE

CERVICAL MODULAR CAGE CERVICAL MODULAR CAGE CERVICAL MODULAR CAGE Modularity and exclusive technical characteristics Anchoring clip Open cage Closed cage Bone substitute (made to measure) Stand-alone and versatile Stability

More information

SURGICAL TECHNIQUE ROI-T TM TLIF INTERSOMATIC IMPLANT TRANS-FORAMINAL APPROACH

SURGICAL TECHNIQUE ROI-T TM TLIF INTERSOMATIC IMPLANT TRANS-FORAMINAL APPROACH TLIF INTERSOMATIC IMPLANT TRANS-FORAMINAL APPROACH Table of Contents page Step 1 - Articular resection.................................................................................. 3 Step 2 - Pedicle

More information

SURGICAL TECHNIQUE. Easyspine PEDICLE SCREW SYSTEM

SURGICAL TECHNIQUE. Easyspine PEDICLE SCREW SYSTEM SURGICAL TECHNIQUE Easyspine PEDICLE SCREW SYSTEM Easyspine PEDICLE SCREW SYSTEM Designed by leading spine surgeons, Easyspine features a simplified surgical technique and adaptable implants to accommodate

More information

SURGICAL TECHNIQUE ROI-C TM ANTERIOR CERVICAL CAGE

SURGICAL TECHNIQUE ROI-C TM ANTERIOR CERVICAL CAGE SURGICAL TECHNIQUE ROI-C TM ANTERIOR CERVICAL CAGE SURGICAL TECHNIQUE ROI-C TM Table of Contents page 1 - Disc location............................................................................................

More information

SURGICAL TECHNIQUE MODULAR CERVICAL CAGE

SURGICAL TECHNIQUE MODULAR CERVICAL CAGE MODULAR CERVICAL CAGE Table of Contents Surgical approach to the disc................................................................. 3 1 to 2 Sizing of the implant..............................................................................

More information

Mobi-C CERVICAL DISC TWO-LEVEL APPLICATIONS

Mobi-C CERVICAL DISC TWO-LEVEL APPLICATIONS Mobi-C CERVICAL DISC Product Brochure Features TWO-LEVEL APPLICATIONS INDICATIONS mobi-c Cervical disc bone sparing two-level indications Ease of insertion mobile bearing Over 17,000 Mobi-C discs have

More information

A U X I L I A R Y C O N N E C T O R S Surgical Technique

A U X I L I A R Y C O N N E C T O R S Surgical Technique A U X I L I A R Y C O N N E C T O R S Surgical Technique AUXILIARY CONNECTORS ISSYS LP Auxiliary Connectors The ISSYS LP auxiliary connectors were designed to provide medial-lateral variability for the

More information

A Fusion of Versatility, Performance and Ease.

A Fusion of Versatility, Performance and Ease. MATRIX Spine System. Degenerative and Deformity. System Guide A Fusion of Versatility, Performance and Ease. Instruments and implants approved by the AO Foundation The MATRIX Spine System is a universal

More information

Lumbar Dynamic. Reinhold Knebel Product Manager

Lumbar Dynamic. Reinhold Knebel Product Manager Lumbar Dynamic Reinhold Knebel Product Manager The DSS system provides options for dynamic and rigid stabilization to treat hypermobile segments and segments that require fusion. Allows pedicle displacement

More information

EXCELLA ll. Spinal System

EXCELLA ll. Spinal System EXCELLA ll Spinal System Excella II Spinal System INDICATIONS FOR USE The Innovasis Excella II Spinal System is intended for use in the non-cervical area of the spine. WARNING: The safety and effectiveness

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that JTL America 3205 Clairmont Ct., Ste. B Fort

More information

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants INTERNATIONAL STANDARD ISO 14801 Second edition 2007-11-15 Dentistry Implants Dynamic fatigue test for endosseous dental implants Art dentaire Implants Essai de fatigue dynamique pour implants dentaires

More information

USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine.

USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine. USS Variable Axis Screw (VAS) System. For posterior fixation of the lumbar spine. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction USS Variable Axis

More information

CERTIFICATE OF ACCREDITATION

CERTIFICATE OF ACCREDITATION CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that JTL America 3205 Clairmont Ct., Ste. B Fort

More information

USS Variable Axis Screw

USS Variable Axis Screw USS Variable Axis Screw Polyaxial side-opening pedicle screw Surgical technique Original Instruments and Implants of the Association for the Study of Internal Fixation AO/ASIF USS Variable Axis Screw

More information

Biomechanics of Fractures and Fixation

Biomechanics of Fractures and Fixation Biomechanics of Fractures and Fixation Theodore Toan Le, MD Original Author: Gary E. Benedetti, MD; March 2004 New Author: Theodore Toan Le, MD; Revised October 09 Basic Biomechanics Material Properties

More information

4.5 System. Surgical Technique. This publication is not intended for distribution in the USA.

4.5 System. Surgical Technique. This publication is not intended for distribution in the USA. 4.5 System Surgical Technique This publication is not intended for distribution in the USA. Contents EXPEDIUM 4.5 Spine System 2 Features and Benefits 3 Surgical Technique Extended Tandem Connector 4 Placement

More information

Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine.

Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine. Thoracolumbar Anterior Compression (TAC) System. Allows distraction, compression, and lateral fixation of the lower thoracic and lumbar spine. Technique Guide Instruments and implants approved by the AO

More information

operative technique Universal Application

operative technique Universal Application operative technique Universal Application Introduction Introduction Building upon the design rationale of the Xia Spinal System, the new Xia Spinal System represents the latest advancement in spinal implant

More information

surgical technique ROI-T

surgical technique ROI-T surgical technique ROI-T TLIF CAGE Surgical Technique ROI-T TLIF cage Table of Contents Approach and facet/pedicle preparation... 3 Controlled distraction... 4 Discectomy and endplate preparation... 6

More information

The Versatile Polyaxial Solution for the Universal Spine Systems. USS II Polyaxial. Surgical Technique

The Versatile Polyaxial Solution for the Universal Spine Systems. USS II Polyaxial. Surgical Technique The Versatile Polyaxial Solution for the Universal Spine Systems USS II Polyaxial Surgical Technique Image intensifier control This description alone does not provide sufficient background for direct use

More information

VLIFT System Overview. Vertebral Body Replacement System

VLIFT System Overview. Vertebral Body Replacement System VLIFT System Overview Vertebral Body Replacement System VLIFT System System Description The VLIFT Vertebral Body Replacement System consists of a Distractible In Situ (DIS) implant, which enables the surgeon

More information

SerratoTM. Surgical technique

SerratoTM. Surgical technique TM Surgical technique Table of contents Key design features.... 3 System overview.... 4 Surgical technique Patient positioning... 7 Preparing the pedicle.... 8 Screw insertion... 11 Rod contouring... 12

More information

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 EMPIRICAL TESTING CORP. 4628 Northpark Drive Colorado Springs, CO 80918 Dawn Lissy Phone: 719 264 9937 dlissy@empiricaltesting.com MECHANICAL Valid To: January

More information

The CerviFix System. Including the StarLock Components. CerviFix Clamp and Screw. StarLock Clamp and Screw

The CerviFix System. Including the StarLock Components. CerviFix Clamp and Screw. StarLock Clamp and Screw The CerviFix System Including the StarLock Components CerviFix Clamp and Screw StarLock Clamp and Screw The CerviFix System The CerviFix System is a comprehensive set of implants, including clamps, screws,

More information

Specializing in Mechanical Testing for Medical Devices. Office:

Specializing in Mechanical Testing for Medical Devices. Office: Specializing in Mechanical Testing for Medical Devices Office: 260.489.1444 Email: sales@jtlamerica.com www.jtlamerica.com Spine Spinal implant test standards can t keep up with the rapid pace of device

More information

Valencia Pedicle Screw Surgical Technique

Valencia Pedicle Screw Surgical Technique Valencia Pedicle Screw Surgical Technique VALENCIA CIRCUIT TABLE OF CONTENTS Design Rationale Indications for Use Surgical Technique 1. Pedicle Preparation 2. Screw Insertion 3. Rod Placement 4. Locking

More information

Spinal System. Aesculap Posterior Thoracolumbar Stabilization System S 4. Aesculap Spine

Spinal System. Aesculap Posterior Thoracolumbar Stabilization System S 4. Aesculap Spine Aesculap Posterior Thoracolumbar Stabilization System S 4 Spinal System Aesculap Spine S 4 Spinal System Small The S 4 Spinal System features a revolutionary pressure vessel design capable of delivering

More information

Freedom. Lumbar Disc Polymer-Metal Bond Integrity. CAUTION: Investigational device. Limited by Federal law to investigational use.

Freedom. Lumbar Disc Polymer-Metal Bond Integrity. CAUTION: Investigational device. Limited by Federal law to investigational use. Freedom Lumbar Disc Polymer-Metal Bond Integrity CAUTION: Investigational device. Limited by Federal law to investigational use. 1 White Paper Freedom Lumbar Disc Polymer-Metal Bond Integrity Abstract

More information

SFX Cross Connector System. Surgical Technique Guide

SFX Cross Connector System. Surgical Technique Guide SFX Cross Connector System Surgical Technique Guide Introduction The EXPEDIUM SFX Cross Connector System redefines ease-of-use, implant versatility, and construct security with an advanced top-loading

More information

Y o u r Id e a s En g i n e e r e d t o Li f e

Y o u r Id e a s En g i n e e r e d t o Li f e ISSYS LP Spinal Fixation System Surgical Guide Y o u r Id e a s En g i n e e r e d t o Li f e In t r o d u c t i o n ISSYS LP Sp i n a l Fixation System The foundation of the ISSYS LP Spinal Fixation System

More information

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine.

Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine. Thoracolumbar Spine Locking Plate (TSLP) System. A low-profile plating system for anterior stabilization of the thoracic and lumbar spine. Technique Guide Instruments and implants approved by the AO Foundation

More information

Simply Complete. VA-LCP Elbow Plating System.

Simply Complete. VA-LCP Elbow Plating System. Simply Complete. VA-LCP Elbow Plating System. Building on success Founded upon the established principles of Synthes LCP technology and the first generation elbow plates, the Variable Angle LCP Elbow Plating

More information

RECLAIM REVISION HIP SYSTEM

RECLAIM REVISION HIP SYSTEM RECLAIM REVISION HIP SYSTEM Where Strength and Modularity Connect DESIGN RATIONALE System Overview RECLAIM Modular Revision Hip System WHERE STRENGTH & MODULARITY CONNECT 2 Offset Options Proximal Body

More information

EXTERNAL FIXATION SYSTEM

EXTERNAL FIXATION SYSTEM EXTERNAL FIXATION SYSTEM The Apex Pin Fixation System is a one step procedure reducing insertion time and reducing insertion temperature. There are three types of fixation pins: Self-drilling / Self-tapping

More information

Surgical Technique. Occipito-Cervico-Thoracic System

Surgical Technique. Occipito-Cervico-Thoracic System OASYS Surgical Technique Occipito-Cervico-Thoracic System System Overview The OASYS Occipito-Cervico-Thoracic System was developed to provide the surgeon with versatility for the treatment of pathologies

More information

L8 Spine System SURGICAL TECHNIQUE. Add: No.1-8, Tianshan Road, Xinbei District, Changzhou, Jiangsu, China

L8 Spine System SURGICAL TECHNIQUE. Add: No.1-8, Tianshan Road, Xinbei District, Changzhou, Jiangsu, China Add: No.-8, Tianshan Road, Xinbei District, Changzhou, Jiangsu, China 23022 Tel: 0086 59 8595556 Fax: 0086 59 859555 Http://www.kanghui.com Add: F25, Shanghai International Pharmaceutical Trad & Exhibition

More information

Graft and fusion. The specifications of the roi implant aim at satisfying the requirements of lumbar intersomatic arthrodesis.

Graft and fusion. The specifications of the roi implant aim at satisfying the requirements of lumbar intersomatic arthrodesis. Patented system New exclusive LDR Medical concept Graft and fusion The specifications of the roi implant aim at satisfying the requirements of lumbar intersomatic arthrodesis. Roi intersomatic cages were

More information

AcUMEDr. Anatomic Midshaft Forearm Plates

AcUMEDr. Anatomic Midshaft Forearm Plates AcUMEDr Anatomic Midshaft Forearm Plates Anatomic Midshaft Forearm Plates Since 1988, Acumed has been designing solutions to the demanding situations facing orthopaedic surgeons, hospitals and their patients.

More information

3.5 mm Locking Attachment Plate

3.5 mm Locking Attachment Plate For Treatment of Periprosthetic Fractures 3.5 mm Locking Attachment Plate Surgical Technique Table of Contents Introduction 3.5 mm Locking Attachment Plate 2 Indications 4 Surgical Technique Preparation

More information

SURGICAL TECHNIQUE GUIDE SOLANAS. Posterior Cervico-Thoracic Fixation System Adjustable Bridge System

SURGICAL TECHNIQUE GUIDE SOLANAS. Posterior Cervico-Thoracic Fixation System Adjustable Bridge System SURGICAL TECHNIQUE GUIDE SOLANAS Posterior Cervico-Thoracic Fixation System Adjustable Bridge System 2 SURGICAL TECHNIQUE GUIDE Preface SOLANAS Posterior Cervico-Thoracic Fixation System is designed to

More information

Ref: Q400-09T1 EBI Spine. September 05/VS02. c/o BIOMET Spain Orthopaedics, S.L.

Ref: Q400-09T1 EBI Spine. September 05/VS02. c/o BIOMET Spain Orthopaedics, S.L. Ref: Q400-09T1 EBI Spine. September 05/VS02 c/o BIOMET Spain Orthopaedics, S.L. www.ebimedical.com EBI Omega 21 TM LP Since its introduction in 1996, and with thousands of patients treated so far, the

More information

GIZA Surgical Technique

GIZA Surgical Technique GIZA Surgical Technique Vertebral Body Replacement System Manufactured by Titanium alloy material provides mechanical integrity during insertion and distraction, x-ray visibility, and biocompatibility*

More information

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM

For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM For the Attention of the Operating Surgeon: IMPORTANT INFORMATION ON THE MATRIXRIB FIXATION SYSTEM 10/16 GP2685-E-CAN DESCRIPTION The MatrixRIB Fixation System consists of locking plates, locking screws,

More information

AxSOS Locking Plate System

AxSOS Locking Plate System AxSOS Locking Plate System Operative Technique Small Fragment Basic Fragment 1 2 Contents Page 1. Introduction 4 2. Features & Benefits 5 4 and 5mm Compression Plates 5 Reconstruction and 1/3 Tubular Locking

More information

HydraLok. Operative Technique. Polyaxial Pedicle Screw System

HydraLok. Operative Technique. Polyaxial Pedicle Screw System HydraLok Operative Technique Polyaxial Pedicle Screw System Table of Contents Introduction...1 OPERATIVE TECHNIQUE OVERVIEW...2 DETAILED OPERATIVE TECHNIQUE...4 LOCATE AND PREPARE THE PEDICLE...4 PROBE

More information

DISTRACTION PRODUCT OVERVIEW. For a wide variety of facial applications

DISTRACTION PRODUCT OVERVIEW. For a wide variety of facial applications DISTRACTION PRODUCT OVERVIEW For a wide variety of facial applications DISTRACTION PRODUCT OVERVIEW. STRONG, MODULAR, VERSATILE CRANIOFACIAL DISTRACTION External Midface Distractor Distraction of the maxilla,

More information

3.5 mm LCP Extra-articular Distal Humerus Plate

3.5 mm LCP Extra-articular Distal Humerus Plate Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Extra-articular Distal Humerus Plate Surgical Technique Table of Contents Introduction 3.5 mm LCP Extra-articular Distal Humerus

More information

The Most Reliable Spinal Solution

The Most Reliable Spinal Solution MFG : 09.02.20 REV : 1 4CIS VANE introduces you the reliable pedicle screw The Most Reliable Spinal Solution Non-slip threaded joint Ergonomic design Wedged trapezoid thread Superior locking mechanism

More information

Technique Guide. Universal Spinal System (USS). Designed to achieve the goals of scoliosis surgery.

Technique Guide. Universal Spinal System (USS). Designed to achieve the goals of scoliosis surgery. Technique Guide Universal Spinal System (USS). Designed to achieve the goals of scoliosis surgery. Table of Contents Introduction Universal Spinal System 2 AO ASIF Principles of Internal Fixation 3 Indications

More information

SURGICAL TECHNIQUE GUIDE TRESTLE. Anterior Cervical Plating System

SURGICAL TECHNIQUE GUIDE TRESTLE. Anterior Cervical Plating System SURGICAL TECHNIQUE GUIDE TRESTLE Anterior Cervical Plating System 2 SURGICAL TECHNIQUE GUIDE SURGICAL TECHNIQUE GUIDE System Features Large window enables visualization of graft site and end plates Screw

More information

POLYAXIAL SPINE SYSTEM D E G E N E R A T I V E

POLYAXIAL SPINE SYSTEM D E G E N E R A T I V E S U R G I C A L T E C H N I Q U E En POLYAXIAL SPINE SYSTEM D E G E N E R A T I V E LOW PROFILE POLYAXIAL SPINE SYSTEM A01110050 Straight spatula A01110060 Curved spatula A01110070 Bone probe A01114500

More information

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique. Hip Knee Spine Navigation

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique. Hip Knee Spine Navigation ONE SYSTEM, MULTIPLE OPTIONS Surgical Technique Hip Knee Spine Navigation MUST MINI Surgical Technique Hip Knee Spine Navigation INTRODUCTION The M.U.S.T. Mini posterior cervical screw system is a modular

More information

Occipital Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput.

Occipital Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput. Occipital Cervical Fusion System. Implants and instruments designed to optimize fixation to the occiput. Technique Guide and implants approved by the AO Foundation Table of Contents Introduction Occipital

More information

Surgical Technique FOR U.S. DOMESTIC USE ONLY

Surgical Technique FOR U.S. DOMESTIC USE ONLY Surgical Technique FOR U.S. DOMESTIC USE ONLY Polyaxial Screws HA Coated Polyaxial Screws Uniplanar Screws Reduction Screws Reduction Uniplanar Screws Modular Screws TABLE OF CONTENTS Reform Pedicle Screw

More information

Polaris Deformity System Trivium Derotation System

Polaris Deformity System Trivium Derotation System Polaris Deformity System Trivium Derotation System A Three Dimensional Approach to Deformity Correction The Polaris Deformity System in combination with the Trivium Derotation System incorporates Enbloc

More information

Hoffmann II MRI External Fixation Systems. Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis

Hoffmann II MRI External Fixation Systems. Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis Hoffmann II MRI External Fixation Systems Hoffmann II MRI Hoffmann II Compact MRI Modular Systems for Upper Extremity Lower Extremity Pelvis Introduction In the ever changing world of medicine, new technology

More information

EXPEDIUM 5.5 Spine System Family Product Catalogue

EXPEDIUM 5.5 Spine System Family Product Catalogue EXPEDIUM 5.5 Spine System Family Product Catalogue Including: EXPEDIUM EXPEDIUM Vertebral Body Derotation Universal Connector Set Cobalt Chromium Rods INTRODUCTION Contents EXPEDIUM Spine System is a comprehensive

More information

Asnis. Micro Cannulated screw system. Xpress operative technique

Asnis. Micro Cannulated screw system. Xpress operative technique Asnis Micro Cannulated screw system Xpress operative technique Asnis Micro Cannulated screw system Table of contents Indications, precautions & contraindications 3 Operative technique 4 This publication

More information

VERTEX SELECT. surgical technique. adjustability. Flexibility. adaptability. Reconstruction System

VERTEX SELECT. surgical technique. adjustability. Flexibility. adaptability. Reconstruction System VERTEX SELECT Reconstruction System surgical technique adjustability. Flexibility. adaptability. adjustability. Flexibility. adaptability. The VERTEX SELECT Reconstruction System is a comprehensive set

More information

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments

Torsion Property and Cyclic Fatigue Fracture Behavior of. Nickel-Titanium Endodontic Instruments Key Engineering Materials Online: 2005-06-15 ISSN: 1662-9795, Vols. 288-289, pp 603-606 doi:10.4028/www.scientific.net/kem.288-289.603 2005 Trans Tech Publications, Switzerland Torsion Property and Cyclic

More information

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area.

Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area. Mandible External Fixator II. Provides treatment for fractures of the maxillofacial area. Technique Guide This publication is not intended for distribution in the USA. Instruments and implants approved

More information

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique

ONE SYSTEM, MULTIPLE OPTIONS. Surgical Technique ONE SYSTEM, MULTIPLE OPTIONS Surgical Technique Joint Spine Sports Med M.U.S.T. Mini Surgical Technique 2 INDEX 1. INTRODUCTION 4 1.1 Indications 4 1.2 Contraindications 4 1.3 Pre-Operative Planning 5

More information

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Olecranon Plates 2 AO Principles 3 Indications 3 Clinical

More information

I. II. SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE

I. II. SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE I. GS1 II. SPINAL SYSTEM SURGICAL TECHNIQUE GUIDE Introduction The Gold Standard Orthopaedics, LLC GS1 Spinal System designed in conjunction with Richard Holt, M.D. incorporates both strength and function

More information

Low Profile Neuro Plating System. Surgical Technique

Low Profile Neuro Plating System. Surgical Technique Low Profile Neuro Plating System Surgical Technique TABLE OF CONTENTS INTRODUCTION Low Profile Neuro Plating System 2 SURGICAL TECHNIQUE Technique 5 PRODUCT INFORMATION Low Profile Neuro Plates 10 Low

More information

Royal Oak Cervical Plate System

Royal Oak Cervical Plate System Royal Oak Cervical Plate System Manufactured by Nexxt Spine, Inc. Royal Oak Cervical Plate System INTRODUCTION FEATURES AND BENEFITS Table of Contents SURGICAL TECHNIQUE Step 1. Patient Positioning Step

More information

Biomet Large Cannulated Screw System

Biomet Large Cannulated Screw System Biomet Large Cannulated Screw System s u r g i c a l t e c h n i q u e A Complete System for Simplified Fracture Fixation 6.5mm & 7.3mm The Titanium, Self-drilling, Self-tapping Large Cannulated Screw

More information

Biomechanics of Interspinous Process Fixation and Lateral Modular Plate Fixation to Support Lateral Lumbar Interbody Fusion (LLIF)

Biomechanics of Interspinous Process Fixation and Lateral Modular Plate Fixation to Support Lateral Lumbar Interbody Fusion (LLIF) Biomechanics of Interspinous Process Fixation and Lateral Modular Plate Fixation to Support Lateral Lumbar Interbody Fusion (LLIF) Calusa Ambulatory Spine Conference 2016 Jason Inzana, PhD 1 ; Anup Gandhi,

More information

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013

Finite Element Analysis of Radius and Ulna. Eli Pavlatos April 24, 2013 Finite Element Analysis of Radius and Ulna Eli Pavlatos April 24, 2013 Outline Review bone biomechanics bone behavior during fracture Stress analysis concepts properties of bone for models Fracture simulations

More information

Thoracic Lumbar Pelvic Posterior Anterior. Universal Spine System (USS). A versatile side-loading system portfolio.

Thoracic Lumbar Pelvic Posterior Anterior. Universal Spine System (USS). A versatile side-loading system portfolio. Thoracic Lumbar Pelvic Posterior Anterior Universal Spine System (USS). A versatile side-loading system portfolio. Universal Spine System (USS) The Synthes USS product portfolio is based on more than

More information

Surgical Technique & Product Catalogue. Guide for Open & MIS Procedures

Surgical Technique & Product Catalogue. Guide for Open & MIS Procedures Surgical Technique & Product Catalogue Guide for Open & MIS Procedures INTRODUCTION The VIPER Cortical Fix Fenestrated Screw System is the first pedicle screw implant to offer enhanced fixation in both

More information

Visit our website on www.biotech-medical.com The DLP - Dorso-Lumbar Polyaxial Screw System has been designed to address the pathologies of the thoracolumbar spine. The DLP System contains a wide range

More information

mykres spinal system SHOWA IKA small implants with a big difference

mykres spinal system SHOWA IKA small implants with a big difference mykres spinal system small implants with a big difference SHOWA IKA 8 years of innovative development 6000 succesful cases 1 innovative spinal system Introduction Mykres is a unique system with a unique

More information

Xiapedia Xiapedia_TLXIABR12100_US.indd 1 3/5/13 8:58 AM

Xiapedia Xiapedia_TLXIABR12100_US.indd 1 3/5/13 8:58 AM Xiapedia 2.0 Xiapedia Introduction The Stryker Spine Thoracolumbar Brand Marketing Team is proud to introduce Xiapedia 2.0 the latest one stop resource for comprehensive information on the Xia, Xia 4.5,

More information

External Fixator Brochure

External Fixator Brochure External Fixator Brochure Response Ortho is a global orthopaedic trauma solutions manufacturer offering premium products created under its founding principles of innovation, excellence by design and functional

More information

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Low Bend Medial Distal Tibia Plates

More information

Thunderbolt. surgical technique. MIS Pedicle Screw System. Where Nimble and Secure Intersect

Thunderbolt. surgical technique. MIS Pedicle Screw System. Where Nimble and Secure Intersect Thunderbolt TM MIS Pedicle Screw System Where Nimble and Secure Intersect surgical technique i www.choicespine.com System Features Dovetail set screw: Minimizes head splay and cross-threading Secure connection

More information

July 10, X-spine Systems, Incorporated David Kirschman, MD Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342

July 10, X-spine Systems, Incorporated David Kirschman, MD Chief Medical Officer 452 Alexandersville Road Miamisburg, Ohio 45342 DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service July 10, 2015 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002 X-spine

More information

3.5 mm LCP Olecranon Plates

3.5 mm LCP Olecranon Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Olecranon Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Olecranon Plates 2 AO Principles 3 Indications

More information

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10

TABLE OF CONTENTS. Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview. Implants 3. Instruments 5. Surgical Technique 10 Surgical Technique TABLE OF CONTENTS Vault C Anterior Cervical Discectomy 2 and Fusion (ACDF) System Overview Indications 2 Implants 3 Instruments 5 Surgical Technique 10 1. Preoperative planning 10 2.

More information

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates

Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based screw insertion into locking plates Ioannou et al. Journal of Orthopaedic Surgery and Research (2016) 11:118 DOI 10.1186/s13018-016-0458-y RESEARCH ARTICLE Effectiveness of the surgical torque limiter: a model comparing drill- and hand-based

More information

XRL A modular expandable radiolucent vertebral body replacement system

XRL A modular expandable radiolucent vertebral body replacement system XRL A modular expandable radiolucent vertebral body replacement system This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE Table of Contents Introduction XRL 2 AO Spine Principles

More information

low ProfIle neuro PlaTIng system

low ProfIle neuro PlaTIng system low ProfIle neuro PlaTIng system surgical TeChnIque Table of Contents Introduction Low Profile Neuro Cranial Plating System 2 Surgical Technique Technique 5 Product Information Low Profile Neuro Plates

More information

SYNEX The vertebral body replacement with ratchet mechanism

SYNEX The vertebral body replacement with ratchet mechanism SYNEX The vertebral body replacement with ratchet mechanism Instruments and implants approved by the AO Foundation. This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE Image

More information

Introduction. Our hope is that this Surgical Technique Guide enhances your knowledge and contributes to clinical success for your patients.

Introduction. Our hope is that this Surgical Technique Guide enhances your knowledge and contributes to clinical success for your patients. Surgical Technique Introduction DePuy Spine continues to support the goal of expanding the spine surgeon s options for the treatment of spinal disorders. Collaborating with renowned spine specialists,

More information

AcUMEDr. Olecranon Threaded Compression Rod

AcUMEDr. Olecranon Threaded Compression Rod AcUMEDr Olecranon Threaded Compression Rod Olecranon Threaded Compression Rod Since 1988, Acumed has been designing solutions to the demanding situations facing orthopaedic surgeons, hospitals and their

More information

CSS Combined Spine System

CSS Combined Spine System CSS Combined Spine System H- CSS Combined Spine System H- CSS Combined Spinal System : 800000 H-3 CSS Combined Spinal System : 800000 Ref.Number Description Pieces Ref.Number Description Pieces 00000 Open

More information

Hoffmann II Micro External Fixation System

Hoffmann II Micro External Fixation System Trauma Hoffmann II Micro External Fixation System Indications for the Hand Components 9 3 4 5 1 2 7 8 6 1 Hoffmann II Micro Multi-Pin Clamp 2 Hoffmann II Micro 90 Multi-Pin Clamp 3 Hoffmann II Micro Rod

More information

Dual-Opening USS. For deformity correction.

Dual-Opening USS. For deformity correction. Dual-Opening USS. For deformity correction. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction Dual-Opening USS 2 AO Principles 4 Indications 5 Preoperative

More information

TM TM Surgical Technique

TM TM Surgical Technique TM TM Surgical Technique TABLE OF CONTENTS Reli SP Spinous Plating System Overview Device Description Implant Features Indications Instruments Access Instruments Preparation Instruments Insertion Instruments

More information

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE

ACP. Anterior Cervical Plate System SURGICAL TECHNIQUE ACP Anterior Cervical Plate System SURGICAL TECHNIQUE ACP TABLE OF CONTENTS INTRODUCTION 4 INDICATIONS AND CONTRAINDICATIONS 5 WARNINGS AND PRECAUTIONS 6 IMPLANT DESCRIPTION 7 INSTRUMENTS 10 SURGICAL

More information

Zimmer Small Fragment Universal Locking System. Surgical Technique

Zimmer Small Fragment Universal Locking System. Surgical Technique Zimmer Small Fragment Universal Locking System Surgical Technique Zimmer Small Fragment Universal Locking System 1 Zimmer Small Fragment Universal Locking System Surgical Technique Table of Contents Introduction

More information

Virage OCT Spinal Fixation System Surgical Technique

Virage OCT Spinal Fixation System Surgical Technique Virage OCT Spinal Fixation System Surgical Technique 2 Virage OCT Spinal Fixation System Surgical Technique Virage OCT Spinal Fixation System Surgical Technique Description, Indications & Contraindications...3

More information

Interbody fusion cage for the transforaminal approach. Travios. Surgical Technique

Interbody fusion cage for the transforaminal approach. Travios. Surgical Technique Interbody fusion cage for the transforaminal approach Travios Surgical Technique Image intensifier control This description alone does not provide sufficient background for direct use of DePuy Synthes

More information

X-spine Surgical Technique

X-spine Surgical Technique X-spine Surgical Technique The X90 Pedicle Screw System Revolutionary Design and Function This document is intended exclusively for experts in the field, particularly physicians, and is not intended for

More information

designed to advance the treatment of hip fractures.

designed to advance the treatment of hip fractures. designed to advance the treatment of hip fractures. introducing the tfn-advanced proximal femoral nailing system (tfna). the tfna system is a new system designed to solve a wide range of unmet needs for

More information

MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES

MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES MODELING AND ANALYSIS OF SOME STABILIZATION SYSTEMS OF RADIUS FRACTURES Prof.dr.eng. Mirela TOTH-TAŞCĂU, Lecturer.dr.eng. Lucian RUSU, PhD.eng. Flavia BĂLĂNEAN, PhD.eng. Cristian TOADER-PASTI, Politehnica

More information

Technique Guide. Epoca Revision Instrument Set. For extraction and revision of Epoca shoulder implants.

Technique Guide. Epoca Revision Instrument Set. For extraction and revision of Epoca shoulder implants. Technique Guide Epoca Revision Instrument Set. For extraction and revision of Epoca shoulder implants. Table of Contents Introduction Epoca Revision Instrument Set 2 Implantation of a Long or Extra-long

More information

OPTIONAL/ADDITIONAL INSTRUMENTS

OPTIONAL/ADDITIONAL INSTRUMENTS 20 Elevated 6320-36-22 20 Elevated Rim Liner, 36mm OD x 22mm ID 6320-80-22 20 Elevated Rim Liner, 80mm OD x 22mm ID 6320-42-26 20 Elevated Rim Liner, 42mm OD x 26mm ID 6320-80-26 20 Elevated Rim Liner,

More information