Nonlinear Stress Analysis of Titanium Implants by Finite Element Method

Size: px
Start display at page:

Download "Nonlinear Stress Analysis of Titanium Implants by Finite Element Method"

Transcription

1 Dental Materials Journal 2008; 27(4): Original Paper Nonlinear Stress Analysis of Titanium Implants by Finite Element Method Sakae NAGASAWA 1,2, Keigo HAYANO 1, Tooru NIINO 1, Kazunori YAMAKURA 1, Takamitsu YOSHIDA 1,3, Toshihide MIZOGUCHI 3, Nobuyosi TERASHIMA 1, Kaoru TAMURA 2, Michio ITO 1,2, Hiroshi YAGASAKI 2,3, Osamu KUBOTA 4 and Masayuki YOSHIMURA 4 1 Department of Dental Materials, Matsumoto Dental University, 1780 Gobara, Hirooka Shiojiri, Nagano , Japan 2 Division of Biomaterials, Department of Hard Tissue Research, Graduate School of Oral Medicine, Matsumoto Dental University, 1780 Gobara, Hirooka Shiojiri, Nagano , Japan 3 Division of Implantology, Institute for Oral Science, Matsumoto Dental University, 1780 Gobara, Hirooka Shiojiri, Nagano , Japan 4 Yoshioka Co. Ltd., Shimohira, Komagane, Nagano , Japan Corresponding author, Sakae NAGASAWA; snagasawa@po.mdu.ac.jp With use of dental implants on the rise, there is also a tandem increase in the number of implant fracture reports. To the end of investigating the stress occurring in implants, elasticity and plasticity analyses were performed using the finite element method. The following results were obtained: (1) With one-piece type of implants of 3.3 mm diameter, elasticity analysis showed that after applying 500 N in a 45-degree direction, stress exceeding 500 MPa which is the proof stress of grade 4 pure titanium occurred. This suggested the possibility of fatigue destruction due to abnormal occlusal force, such as during bruxism. (2) With two-piece type of implants that can tolerate vertical loading of 5,000 N, plasticity analysis suggested the possibility of screw area fracture after applying 500 N in a 45-degree direction. (3) On the combined use of an abutment and a fixture from different manufacturers, fracture destruction of even Ti-6Al- 4V, which has a high degree of strength, was predicted. Key words: Finite element method, Titanium implant, Nonlinear stress analysis Received Apr 26, 2006: Accepted May 21, 2008 INTRODUCTION With use of dental implants unabatedly on the rise, there is also a tandem increase in the number of implant fracture reports 1,2). Perennially, implant success is of utmost concern and importance to all patients at all times. Against this backdrop, strength analyses and fatigue tests have been performed to predict and determine the clinical performance of dental implants 3-7). In particular, for restoration with ITI implants (Φ: 3.3 mm, L: 12 mm, pure titanium: grade 4), Merz et al. of the Institute Straumann performed fatigue tests to analyze the fatigue resistance of implantabutment connectors 8). They reported that fatigue destruction occurred when force larger than 380 N was applied in a 30-degree direction. They also reported that fatigue strength differed among original titanium lots. Differences in fatigue strength among original material lots are due to differences in the residual degree of process distortion, as we have previously reported 9). Therefore, many cases of implant fracture, as shown in Fig. 1, were due to insufficient understanding of the non-uniform mechanical strengths of implants caused by non-uniformity of original materials, as well as due to limitation on mechanical strength as imposed by load application angle. In this study, nonlinear stress analysis using the finite element method was performed with twofold purposes: (1) to investigate the stress occurring in implants caused by occlusal forces; and (2) to assess the limitation on mechanical strength as imposed by load application angle on implants. MATERIALS AND METHODS Analytical methods Analysis was performed using a finite element method program (ANSYS Inc., Canonsburg, PA, USA). Since the elastic modulus and Poisson s ratio of pure titanium are almost the same as those of titanium alloy 10), a model with unified abutment and implant areas (one-piece type) was produced for the purpose of performing elasticity analysis. In addition, a model in which the abutment and implant areas were separated into two pieces (two-piece type) was produced to perform nonlinear plasticity analysis. Analytical model Grade 2 (G2) pure titanium implant (Φ: 3.3 mm, 4 mm) (Yoshioka, Yokohama, Japan) was selected as

2 634 Dent Mater J 2008; 27(4): Fig. 1 A case of fractured titanium implant. Fig. 2 A figure of the division of elements. Table 1 Physical properties of materials used Material Elastic modulus (GPa) Poisson s ratio Friction coefficient (static, dynamic) Ref No. C.P. titanium grade , , 12 C.P. titanium grade , , 12 Ti-6Al-4V alloy , , 12 Compact bone the analytical model. In addition, G2 pure titanium implant and Ti-6Al-4V alloy abutment (Nitto Kogyo, Nagoya, Japan) were also combined as an analytical model. Each analytical model was divided into approximately 30,000 three-dimensional elements as shown in Fig. 2. Since human maximum occlusal force is almost the same as human body weight (approximately kgf), 500 N was applied to the abutment head area vertically and in the directions of 12.5, 30, and 45 degrees. Assuming that abutment and implants are connected by friction, the compact bone s basal area was fixed. Physical properties Table 1 shows the values of the physical properties of materials used for elasticity analysis 11-13). Grade 2 (G2) and grade 4 (G4) pure titanium and Ti-6Al-4V alloy were used as original implant materials. Figure 3 shows a true stress-true strain curve of the titanium material used for plasticity analysis. The true stress-true strain curve was obtained from a stress-strain curve and the diameter of fractured area after a tension test of a titanium round rod (G2 Ti-6Al-4V alloy: Kobe Steel Ltd., Kobe, Japan; G4: L. Klein SA, Bienne, Switzerland). The latter was used for implant production by Yoshioka (Yoshioka, Yokohama, Japan), and tension test was carried out using a test machine (Model 5882, Instron, Canton, Fig. 3 True stress-true strain curve of pure titanium round rod (Φ: 5 mm). G2: grade 2 pure titanium, G4: grade 4 pure titanium, Ti-6Al-4V: Ti-6Al-4V (wt %) alloy. USA) 14). As the main purpose of this study was to analyze the strength of implants, the jawbone was regarded as the compact bone and being elastic in property. Evaluation of the validity of analyses To evaluate the validity of analyses, an implant manufactured by Yoshioka (Yoshioka, Yokohama, Japan) was fixed using a jig, and load test was

3 Dent Mater J 2008; 27(4): performed by applying a prescribed load using a test machine (AG-500D, Shimadzu, Kyoto, Japan). After load application, the implant was longitudinally sectioned using a polishing machine (ECOMET3, Buehler, Lake Bluff, USA) and the sectioned area was compared with analysis results. RESULTS Elasticity analysis Figure 4 shows the deformation and distribution of first main stress in a general one-piece type of implant of 3.3 mm diameter after a vertical application of 500 N. The amount of deformation is indicated by actual length, and the amount of stress is indicated by contours. With a vertical load application, stress applied to one-piece type of implant was only approximately 100 MPa or smaller. Figure 5 shows the deformation and distribution of first main stress in a one-piece type of implant of 3.3 mm diameter after applying 500 N in a 45-degree direction. Stress exceeding 500 MPa, which is the proof stress of G4 pure titanium, occurred in this case. Figure 6 shows the deformation and distribution of first main stress in a one-piece type of implant of 4.0 mm diameter after applying 500 N in a 45-degree direction. In this case, stress decreased to approximately 320 MPa. Plasticity analysis Figure 7 shows the deformation and distribution of Fig. 4 Results of elastic analysis after vertical application of 500 N to a one-piece type of implant (diameter: 3.3 mm) (actual deformation, first main stress distribution). Fig. 6 Results of elastic analysis after application of 500 N to a one-piece type of implant (diameter: 4.0 mm) in a 45-degree direction (actual deformation, first main Fig. 5 Results of elastic analysis after application of 500 N to a one-piece type of implant (diameter: 3.3 mm) in a 45-degree direction (actual deformation, first main Fig. 7 Results of plastic analysis after vertical application of 500 N to a G2 implant (diameter: 3.3 mm) (actual deformation, von Mises

4 636 Dent Mater J 2008; 27(4): Fig. 8 Results of plastic analysis after application of 500 N to a G2 implant (diameter: 3.3 mm) in a 45-degree direction (actual deformation, von Mises Fig. 10 Results of plastic analysis after application of 500 N to a G2 implant (diameter: 3.3 mm) in a 12.5-degree direction (actual deformation, von Mises Fig. 9 Results of plastic analysis after application of 500 N to a G2 implant (diameter: 3.3 mm) in a 30-degree direction (actual deformation, von Mises Fig. 11 Results of plastic analysis after application of 500 N to a G2 implant (diameter: 4.0 mm) in a 45-degree direction (actual deformation, von Mises von Mises stress 15) in a G2 pure titanium, two-piece type of implant of 3.3 mm diameter after a vertical application of 500 N. With two-piece type of implant, stress of approximately 150 MPa was concentrated in the abutment screw area, showing a higher stress concentration compared with one-piece type of implant. Figure 8 shows the deformation and distribution of von Mises stress in a G2 pure titanium, two-piece type of implant of 3.3 mm diameter after applying 500 N in a 45-degree direction. Concentration of stress exceeding 550 MPa was noted in the abutment neck area, superior margin of the bearing area, and superior margin of the compact bone, with evident indication of large deformation. Conversely, in Fig. 5, the amount of elastic deformation was very little as most of the deformation was plastic deformation. Figures 9 and 10 show the deformation and distribution of von Mises stress in a G2 pure titanium, two-piece type of implant of 3.3 mm diameter after applying 500 N in (Fig. 10) and 30-degree (Fig. 9) directions. Maximum stress applied to the implant decreased to 400 MPa at the angle of 30 degrees, and 300 MPa at the angle of 12.5 degrees. A slight deformation occurred in the screw area at the angle of 12.5 degrees. Figure 11 shows the deformation and distribution of von Mises stress in a G2 pure titanium, two-piece

5 637 Dent Mater J 2008; 27(4): Fig. 12 Results of plastic analysis after application of 500 N to a G4 implant (diameter: 3.3 mm) in a 45-degree direction (actual deformation, von Mises Fig. 14 Comparison of experimental results and analytical results (A: no load applied, B: 5,000 N vertical load, 45 degrees 500 N load, -1: experiment, -2: analysis). diameter and an abutment (Ti-6Al-4V alloy) manufactured by Nitto Kogyo, after applying 500 N in a 45-degree direction. Stress exceeding 1 GPa, which is larger than the proof stress of 900 MPa of Ti-6Al-4V alloy, was concentrated in the abutment neck area. Fig. 13 Results of plastic analysis after application of 500 N to an implant (diameter: 3.3 mm) of the combination of an implant and an abutment of different manufacturers in a 45-degree direction (actual deformation, von Mises type of implant of 4.0 mm diameter after applying 500 N in a 45-degree direction. Maximum stress decreased to 390 MPa or lower, and deformation was slight. Figure 12 shows the deformation and distribution of von Mises stress in a G4 pure titanium, two-piece type of implant of 3.3 mm diameter after applying 500 N in a 45-degree direction. Although deformation was smaller than in G2, stress exceeding 550 MPa was concentrated in the abutment neck and bearing areas. Figure 13 shows the deformation and distribution of von Mises stress in an implant, which was a combination of a G2 pure titanium implant of 3.3 mm Evaluation of the validity of analyses Experimental versus analysis results are compared in Fig. 14. Figure 14A shows the longitudinal section of an implant manufactured by Yoshioka, to which no load was applied. Figure 14B-1 shows the longitudinal section of the implant after a vertical application of 5,000 N. Figure 14B-2 shows the analysis results in G2 implant after a vertical application of 5,000 N. Figure 14C-1 shows the longitudinal section of the G2 implant after applying 500 N in a 45-degree direction. Figure 14C-2 shows the analysis results in G2 implant after applying 500 N in a 45-degree direction. Although the implant head area was greatly crushed after a vertical application of 5,000 N according to the analysis results, it was because friction by the pressure plate of the autograph was not considered, and analysis results were markedly in accordance with experimental results. DISCUSSION As shown in Fig. 14, the analysis results in this study were markedly in accordance with the experimental results. On this basis, our analysis results were sufficiently reliable to discuss the limitation imposed on implants mechanical strength due to load application angle. As shown in Figs. 4 and 7, current implants have

6 638 Dent Mater J 2008; 27(4): enough strength to bear the vertical loading of occlusal forces. However, as shown in Fig. 5, even in a one-piece type of implant that is the strongest structurally, an application of 500 N in a 45-degree direction to the implant of 3.3 mm diameter resulted in stress exceeding 500 MPa which is the proof stress of G4 pure titanium. This result suggested that since the fatigue strengths of titanium and titanium alloy are almost the same as the proof stress 16,17), titanium implants of 3.3 mm diameter in any structural morphology may cause fatigue destruction. On the other hand, when the diameter was 4.0 mm, as shown in Fig. 6, maximum stress decreased to 320 MPa. This finding thus suggested that whereas it is impossible for G2 pure titanium with a proof stress of approximately 250 MPa to bear the maximum occlusal force of the molars, G4 pure titanium can do so adequately. Plasticity analysis results showed that whereas G2 pure titanium could bear the load of 500 N when load application direction was within 12.5 degrees (Fig. 10), G4 pure titanium could up the ante by bearing the same load even when the direction was 30 degrees (Fig. 9). Nonetheless, fatigue destruction resulted for G4 pure titanium when load application direction was 45 degrees (Fig. 12). When the diameter was 4.0 mm (Fig. 11), maximum stress was greatly decreased but the possibility of fatigue destruction remained high for G2 pure titanium. When G2 pure titanium implant and Ti-6Al-4V abutment, which are products of different manufacturers, were combined (Fig. 13), markedly large stress concentration occurred in the abutment neck area thereby indicating a higher possibility of fatigue destruction. One example of implant fracture is shown in Fig. 1, and this fracture occurred due to this combination. Therefore, although it is mechanically possible to combine implants and abutments of different manufacturers, fit of the junction area cannot be obtained and stress is concentrated in the screw area. With explicit suggestion of a high fracture risk, combined use of implants and abutments of different manufacturers is therefore not recommended. CONCLUSIONS Elasticity and nonlinear stress analyses were performed using the finite element method for twofold purposes: (1) to investigate the stress occurring in implants caused by occlusal forces; and (2) to assess the limitation on mechanical strength as imposed by load application angle on implants. The following conclusions were obtained: (1) When an occlusal force of 500 N was applied to implants of 3.3 mm diameter in a 45-degree direction, stress exceeding 500 MPa which is the proof stress of G4 pure titanium occurred even in one-piece type of implants which have no combined structures. (2) Since all grades of pure titanium have almost the same elastic modulus, possibility of fatigue destruction is present in all types of titanium implants such as Bränemark and ITI. Therefore, use of these implants warrants attention and vigilant carefulness. (3) Since structural design of titanium implants involves limitation of strength, careless use may result in easy fracture. (4) Combined use of implants and abutments of different manufacturers is contraindicated, since the possibility of fatigue destruction is markedly high. REFERENCES 1) Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JYK. Clinical complaints and implant prostheses. J Prosthetic Dent 2003; 90: ) Wakimoto M, Yamamoto K, Nagashima Y, Kido H, Matsuura M. Fracture cases of fixtures with internal locking system. J. Jpn Soc Oral Implant 2005; 18: ) Iplikcioglu H, Akca K, Cehreli CM, Sahin S. Comparison of non-linear finite element stress analysis with in vitro strain gauge measurements on a morse taper implant. Int J Oral Maxillofac Implants 2003; 18: ) Lang AL, Kang B, Wang R, Lang RB. Finite element analysis to determine implant preload. J Prosthet Dent 2003; 90: ) Watanabe F, Hata Y, Komatsu S, Ramos TC, Fukuda H. Finite element analysis of the influence of implant inclination, loading position, and load direction on stress distribution. N Z Dent J 2003; 91: ) Kikuchi M, Takahashi M, Okuno O. Mechanical properties and grindability of dental cast Ti-Nb alloys. Dent Mater J 2003; 22: ) Kitagawa T, Tanimoto Y, Nemoto K, Aida M. Influence of cortical bone quality on stress distribution in bone around dental implant. Dent Mater J 2005; 24: ) Merz RB, Hunenbart S, Belser CU. Mechanics of the implant-abutment connection: An 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000; 15: ) Ono H, Hayano K, Taira K, Nagasawa S, Ito M. The mechanical properties of the titanium for implant materials An influence by the purity. Matsumoto Shigaku 2005; 31: ) Ito Y. The metal material utilization encyclopedia, 1st ed, Sangyo Chosakai, Tokyo, 2000, p ) Kobayashi T, Hirose T. Physics vade mecum, 1st ed, Maruzen, Tokyo, 1989, p ) Kobayashi T, Hirose T. Physics vade mecum, 1st ed, Maruzen, Tokyo, 1989, p.53.

7 Dent Mater J 2008; 27(4): ) Nakamura T, Ohyama T, Waki T, Kinuta S, Wakabayashi K, Takano N, Yatani H. Finite element analysis of fiber-reinforced fixed partial dentures. Dent Mater J 2005; 24: ) The Japan Society for Technology of Plasticity. Nonlinear finite element method for metal forming analysis, 1st ed, Corona Publishing Co. Ltd., Tokyo, 1997, p ) The Japan Society for Technology of Plasticity. Nonlinear finite element method for metal forming analysis, 1st ed, Corona Publishing Co. Ltd., Tokyo, 1997, p ) National Institute for Materials Science, Japan. NIMS Fatigue data sheet No. 95, National Institute for Materials Science, Japan, Tokyo, 2004, pp ) National Institute for Materials Science, Japan. NIMS Fatigue data sheet No. 98, National Institute for Materials Science, Japan, Tokyo, 2005, pp.1-10.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics.

In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. In-Silico approach on Offset placement of implant-supported bridges placed in bone of different density in Orthodontics. Chandrasenan.P 1, Vishnu.G 2, Akshay K Nair 3 1M Tech student, Department of Mechanical

More information

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant International Journal of Mechanical & Mechatronics Engineering IJMME-IJENS Vol:15 No:06 84 Experimental and Numerical Comparison of Stresses Level between Titanium and Novel Composite Single Tooth Implant

More information

Anuradha Borse 1* Rubina Tabassum 2 Omkar Shetty 3

Anuradha Borse 1* Rubina Tabassum 2 Omkar Shetty 3 Research AHB Article Adv Hum Biol 2014; 4(2):22-31. A Comparative Evaluation of Stress Distribution at the Implant Abutment Interface between Platforms Switched Morse Taper and Pseudo Platform Switched

More information

Finite element modeling technique for predicting mechanical behaviors on mandible bone during mastication

Finite element modeling technique for predicting mechanical behaviors on mandible bone during mastication ORIGINAL ARTICLE http://dx.doi.org/10.4047/jap.01.4.4.18 Finite element modeling technique for predicting mechanical behaviors on mandible bone during mastication Hee-Sun Kim 1 *, PhD, Jae-Yong Park 1,

More information

Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw Fracture in Dental Implant System

Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw Fracture in Dental Implant System Hindawi Journal of Healthcare Engineering Volume 2017, Article ID 4842072, 6 pages https://doi.org/10.1155/2017/4842072 Research Article Hollow Abutment Screw Design for Easy Retrieval in Case of Screw

More information

Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone

Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone Original Article Three Dimensional Finite Element Analysis to Detect Stress Distribution in Spiral Implants and Surrounding Bone Matteo Danza*, Annalisa Palmieri**, Francesca Farinella***, Giorgio Brunelli****,

More information

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis

Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A Three-Dimensional Finite Element Analysis J Indian Prosthodont Soc (Oct-Dec 2013) 13(4):555 559 DOI 10.1007/s13191-012-0189-7 ORIGINAL ARTICLE Evaluation of Stress Distribution in Bone of Different Densities Using Different Implant Designs: A

More information

INFLUENCE OF CANTILEVER ON STRESS DISTRIBUITION IN BONE AROUND DENTAL IMPLANTS: A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS

INFLUENCE OF CANTILEVER ON STRESS DISTRIBUITION IN BONE AROUND DENTAL IMPLANTS: A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS INFLUENCE OF CANTILEVER ON STRESS DISTRIBUITION IN BONE AROUND DENTAL IMPLANTS: A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS Roberto Brunow Lehmann Universidade Federal Fluminense Av. dos Trabalhadores,

More information

STUDYING THE EFFECT OF THE PRELOAD INDUCED BY SCREW TIGHTENING ON STRESS DISTRIBUTION OF A DENTAL IMPLANT

STUDYING THE EFFECT OF THE PRELOAD INDUCED BY SCREW TIGHTENING ON STRESS DISTRIBUTION OF A DENTAL IMPLANT Belagh & Kashani, 2017 Volume 3 Issue 3, pp.195-211 Date of Publication: 19 th December, 2017 DOI-https://dx.doi.org/10.20319/mijst.2017.33.195211 This paper can be cited as: Belagh, A.B.F & Kashani, H.M.

More information

Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA

Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA ISSN 2395-1621 Effect of Screw Profile on Stress Distribution Pattern of Dental Implants by FEA #1 RaiphaleGovind, #2 Shinde B.M. 1 govindaraiphale@gmail.com*1 2bharat.shinde@zealeducation.com 1 (M.E Student

More information

While the high rate of success of dental

While the high rate of success of dental Mechanics of the Implant-Abutment Connection: An 8-Degree Taper Compared to a Butt Joint Connection Beat R. Merz, Dr sc techn, MBA 1 /Stephan Hunenbart, Dipl Eng TU 2 / Urs C. Belser, DMD, Prof Dr med

More information

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint

International Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses

More information

Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect

Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect Journal of Oral Rehabilitation 2007 Three-dimensional finite element analysis of the application of attachment for obturator framework in unilateral maxillary defect S. JIAN*, J. TING*, T. YING &W.DONG-MEI

More information

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study

Influence of Different Angulations of Force in Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study World Journal of Medical Sciences 10 (2): 217-221, 2014 ISSN 1817-3055 IDOSI Publications, 2014 DOI: 10.5829/idosi.wjms.2014.10.2.82275 Influence of Different Angulations of Force in Stress Distribution

More information

Original Article Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results

Original Article Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results Ming-Su Lee, DDS, MS Attending Doctor, Department of Dentistry, National

More information

Abstract. Keywords: Fatigue Failure, Surface Roughness, Fatigue Strength Factor, Bio-Compatible Dental Implant, Electron Beam Melting (EBM)

Abstract. Keywords: Fatigue Failure, Surface Roughness, Fatigue Strength Factor, Bio-Compatible Dental Implant, Electron Beam Melting (EBM) THE NUMERICAL MODELING OF FATIGUE PROPERTIES OF A BIO- COMPATIBLE DENTAL IMPLANT PRODUCED BY ELECTRON BEAM MELTING (EBM) M. Jamshidinia 1, F. Kong 1, R. Kovacevic 1, * Research Center for Advanced Manufacturing

More information

BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA

BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2008, 152(2):1. T. Goldmann, S. Ihde, J. Kuzelka, L. Himmlova 1 BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL

More information

NobelActive. The NobelActive technical story

NobelActive. The NobelActive technical story NobelActive The NobelActive technical story The NobelActive technical story NobelActive origins The life cycle of NobelActive began following years of research and development on a self-drilling and bone-condensing

More information

Journal of Dental Research, Dental Clinics, Dental Prospects. Original Article

Journal of Dental Research, Dental Clinics, Dental Prospects. Original Article Journal of Dental Research, Dental Clinics, Dental Prospects Original Article Comparison of the Effect of Three Abutment-implant Connections on Stress Distribution at the Internal Surface of Dental Implants:

More information

SCREW LOOSENING has been reported as

SCREW LOOSENING has been reported as BASIC SCIENCE RESEARCH Effect of Casting Procedures on Screw Loosening in UCLA-Type Abutments Stefania Kano, DDS, MS, PhD; 1 Donald A. Curtis, DMD; 2 Paul Binon, DDS, MSA; 3 and Gerson Bonfante, DDS, MSc,

More information

The development of modern Oral

The development of modern Oral RESEARCH In Vitro Evaluation of the Removal Force of Abutments in Frictional Dental Implants João César Zielak, DDS, MSc, PhD* Murilo Rorbacker, DDS Rodrigo Gomes, DDS Celso Yamashita, DDS, MSc, PhD Carla

More information

Design and Analysis of Dental Surgical Fixture

Design and Analysis of Dental Surgical Fixture Volume 119 No. 10 2018, 1951-1956 ISSN: 1311-8080 (printed version); ISSN: 1314-3395 (on-line version) url: http://www.ijpam.eu ijpam.eu Design and Analysis of Dental Surgical Fixture Suhas H.G 1,*, Rajkumar

More information

NARROW DIAMETER implant

NARROW DIAMETER implant ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball

More information

Department of Prosthodontics and Periodontics, Piracicaba Dental School, Campinas State University, Piracicaba, SP, Brazil

Department of Prosthodontics and Periodontics, Piracicaba Dental School, Campinas State University, Piracicaba, SP, Brazil 239 Journal of Oral Science, Vol. 55, No. 3, 239-244, 2013 Original Stress distribution in fixed-partial prosthesis and peri-implant bone tissue with different framework materials and vertical misfit levels:

More information

ABSTRACT INTRODUCTION

ABSTRACT INTRODUCTION ABSTRACT Purpose: Evaluation of stress distribution maxilla under parafunctional loading on short dental implant using 3-dimensional (3D) finite element models. Material and method: 3D maxillary and mandibular

More information

Monday Morning Pearls of Practice by Bobby Baig

Monday Morning Pearls of Practice by Bobby Baig Oct 17, 2016 Monday Morning Pearls of Practice by Bobby Baig baig@buildyoursmile.com Prosthodontic Associates 2300 Yonge St, suite 905 Toronto, M4P1E4 www.buildyoursmile.com Mechanical Integrity of Retrieved

More information

Effect of surface treatment with commercial primers on tensile bond strength of auto-polymerizing resin to magnetic stainless steel

Effect of surface treatment with commercial primers on tensile bond strength of auto-polymerizing resin to magnetic stainless steel Effect of surface treatment with commercial primers on tensile bond strength of auto-polymerizing resin to magnetic stainless steel Jian-rong Chen, DDS, a,b Kenji Oka, DDS, PhD, c Wei Hua, DDS, a and Tetsuo

More information

Straumann Screw-retained Prosthetics

Straumann Screw-retained Prosthetics Straumann Screw-retained Prosthetics Straumann Screw-retained Abutments Straumann Screw-retained abutments Indication and product features Abutments for fixed screw-retained restorations; for multi-unit

More information

Dental Research Journal

Dental Research Journal Dental Research Journal Original Article The influence of thread geometry on biomechanical load transfer to bone: A finite element analysis comparing two implant thread designs Manisha G Herekar, Viraj

More information

Double-casting method for fixed prosthodontics with functionally generated path

Double-casting method for fixed prosthodontics with functionally generated path Double-casting method for fixed prosthodontics with functionally generated path Shogo Minagi, DDS, PhD, a Toshiyuki Tanaka, DDS, b Takashi Sato, DDS, PhD, c and Tadashi Matsunaga, DDS, PhD d Department

More information

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY A PATH FOR HORIZING YOUR INNOVATIVE WORK STUDY AND ANALYSIS OF KNEE IMPLANT IN HUMAN BODY GAJANAN D. MANDAVGADE Assistant

More information

Bending strengths and hardness of autopolymerized acrylic resin

Bending strengths and hardness of autopolymerized acrylic resin Bending strengths and hardness of autopolymerized acrylic resin Tomoko Osada, DMD, PhD, Takako Ishimoto, DMD, PhD, Takayuki Aoki, DMD, PhD, Yasunori Suzuki, DMD, PhD, Norio Takishin, DDS, PhD, Chikahiro

More information

The fundamental biomechanics of oral implants. Dr. Szűcs Attila

The fundamental biomechanics of oral implants. Dr. Szűcs Attila The fundamental biomechanics of oral implants Dr. Szűcs Attila Physical elements I. Mechanics solving of technical problems by physical laws»» investigation of prosthesis and the connection to abutments

More information

FINITE ELEMENT ANALYSIS OF PERIODONTAL STRESSES IN FIXED PROSTHODONTICS

FINITE ELEMENT ANALYSIS OF PERIODONTAL STRESSES IN FIXED PROSTHODONTICS FINITE ELEMENT ANALYSIS OF PERIODONTAL STRESSES IN FIXED PROSTHODONTICS Valeria Pendefunda, Arina Ciocan Pendefunda, Nicoleta Ioanid, Alina Apostu, Oana Ţănculescu Odontology, Periodontology and Fixed

More information

CHECK-UP Determination of Impression type & Method after implant placement

CHECK-UP Determination of Impression type & Method after implant placement CHECK-UP Determination of Impression type & Method after implant placement Fixture level impression Abutment level impression Transfer impression Conventional method Pick-up impression Plastic impression

More information

Finite Element Analysis of Dental Implant as Orthodontic Anchorage

Finite Element Analysis of Dental Implant as Orthodontic Anchorage JCDP 10.5005/jp-journals-10024-1044 ORIGINAL RESEARCH Finite Element Analysis of Dental Implant as Orthodontic Anchorage Finite Element Analysis of Dental Implant as Orthodontic Anchorage Anirban Sarmah,

More information

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis

Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis 1 Research Article Stress and Displacement Analysis of Dental Implant Threads Using Three-Dimensional Finite Element Analysis Aswin Yodrux* Department of Materials Handling and Logistics for Engineering,

More information

IJPCDR INTRODUCTION ABSTRACT

IJPCDR INTRODUCTION ABSTRACT Original Research Comparative Evaluation of Stress Distribution 10.5005/jp-journals-10052-0001 within Computerized Bone Model Comparative Evaluation of Stress Distribution within Computerized Bone Model

More information

Mechanical and technical risks in implant therapy.

Mechanical and technical risks in implant therapy. Mechanical and technical risks in implant therapy. Salvi GE, Brägger U. Int J Oral Maxillofac Implants. 2009;24 Suppl:69-85. To systematically appraise the impact of mechanical/technical risk factors on

More information

International Journal of Engineering Research and General Science Volume 3, Issue 1, January-February, 2015 ISSN

International Journal of Engineering Research and General Science Volume 3, Issue 1, January-February, 2015 ISSN Study of Factors Affecting Success Rate of Immediate Implant Loading Archana N Mahajan 1, Dr Ms K N Kadam 2 1 Assistant Professor, Vishwatmak Om Gurudev College of Engineering, Atgaon, Mumbai University,

More information

Comparative Study of Fixation Devices for Intertrochanteric Fractures

Comparative Study of Fixation Devices for Intertrochanteric Fractures Comparative Study of Fixation Devices for Intertrochanteric Fractures C. Sticlaru * A. Davidescu Politehnica University of Timişoara Politehnica University of Timişoara Timişoara, România Timişoara, România

More information

Summer Workshop of Applied Mechanics. Theoretical Analysis of Total Hip Joint Replacement of Sandwich design by FEM

Summer Workshop of Applied Mechanics. Theoretical Analysis of Total Hip Joint Replacement of Sandwich design by FEM Summer Workshop of Applied Mechanics June 2002 Department of Mechanics Faculty of Mechanical Engineering Czech Technical University in Prague Theoretical Analysis of Total Hip Joint Replacement of Sandwich

More information

Stress distribution within circumferential clasp arms

Stress distribution within circumferential clasp arms Journal of Oral Rehabilitation, 1976, Volume 3, pages 387-394 Stress distribution within circumferential clasp arms H. MORRIS, J. W. FARAH, R. G. CRAIG and J. A. A. HOOD Veterans Administration Hospital,

More information

Student Presentation # 3 Biomechanics

Student Presentation # 3 Biomechanics Student Presentation # 3 Biomechanics Anu Phogat Peter Ho 1 BIOMECHANICS OF DENTAL IMPLANTS 2 STRUCTURE OF TOOTH 3 INTERESTING FACTS According to the National Institutes of Health,by age 50, the average

More information

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS

FINITE ELEMENT ANALYSIS OF KNEE IMPLANT AND PROTOTYPE FABRICATION USING RPT PROCESS International Journal of Mechanical Engineering and Technology (IJMET) Volume 8, Issue 12, December 2017, pp. 197 205, Article ID: IJMET_08_12_020 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=8&itype=12

More information

Inlay Fixed Partial Denture Framework 3-D Structural Integrity Validation Using COMSOL Multiphysics 3.5a

Inlay Fixed Partial Denture Framework 3-D Structural Integrity Validation Using COMSOL Multiphysics 3.5a Excerpt from the Proceedings of the COMSOL Conference 2009 Milan Inlay Fixed Partial Denture Framework 3-D Structural Integrity Validation Using COMSOL Multiphysics 3.5a Timo M. R. Alho * Department of

More information

Since the introduction of the Branemark

Since the introduction of the Branemark RESEARCH Influence of the Implant Diameter With Different Sizes of Hexagon: Analysis by 3-Dimensional Finite Element Method Eduardo Piza Pellizzer, DDS, PhD, MSc 1 * Fellippo Ramos Verri, DDS, PhD, MSc

More information

Dynamic retentive force of a mandibular unilateral removable partial denture framework with a back-action clasp

Dynamic retentive force of a mandibular unilateral removable partial denture framework with a back-action clasp J Med Dent Sci 2001; 48: 105 111 Original Article Dynamic retentive force of a mandibular unilateral removable partial denture framework with a back-action clasp Innim Park, DDS*, Miho Eto, DDS*, Noriyuki

More information

The Use of Alpha-Bio Tec's Narrow NeO Implants with Cone Connection for Restoration of Limited Width Ridges

The Use of Alpha-Bio Tec's Narrow NeO Implants with Cone Connection for Restoration of Limited Width Ridges Case Study 48 The Use of Alpha-Bio Tec's Narrow NeO Implants with Cone Connection for Restoration of Limited Width Ridges Dr. Amir Gazmawe DMD, Specialist in Prosthodontics, Israel Dr. Amir Gazmawe graduated

More information

Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver

Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver Original Research Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver Jalil Ghanbarzadeh 1, Mahboobeh Rajabzadeh 2, Mohammadreza Nakhaei 3, Hamidreza

More information

Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element analysis

Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element analysis ISSN 1 746-7233, England, UK World Journal of Modelling and Simulation Vol. 2 (2006) No. 2, pp. 134-140 Comparative study of 12 thread shapes of dental implant designs: a three-dimensional finite element

More information

Australian Dental Journal

Australian Dental Journal Australian Dental Journal The official journal of the Australian Dental Association SCIENTIFIC ARTICLE Australian Dental Journal 2012; 57: 440 445 doi: 10.1111/adj.12002 Load transfer in tilted implants

More information

Characteristics of the Tooth in the Initial Movement: The Influence of the Restraint Site to the Periodontal Ligament and the Alveolar Bone

Characteristics of the Tooth in the Initial Movement: The Influence of the Restraint Site to the Periodontal Ligament and the Alveolar Bone The Open Dentistry Journal, 2009, 3, 85-91 85 Open Access Characteristics of the Tooth in the Initial Movement: The Influence of the Restraint Site to the Periodontal Ligament and the Alveolar Bone Kyoko

More information

Dental Implant Treatment with Diffe Title for Sinus Floor Elevation-A Case Re. Sekine, H; Taguchi, T; Seta, S; Tak Author(s) T; Kakizawa, T

Dental Implant Treatment with Diffe Title for Sinus Floor Elevation-A Case Re. Sekine, H; Taguchi, T; Seta, S; Tak Author(s) T; Kakizawa, T Dental Implant Treatment with Diffe Title for Sinus Floor Elevation-A Case Re Sekine, H; Taguchi, T; Seta, S; Tak Author(s) T; Kakizawa, T Journal Bulletin of Tokyo Dental College, 4 URL http://hdl.handle.net/10130/200

More information

Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis.

Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis. Platform switching for marginal bone preservation around dental implants: a systematic review and meta-analysis. Atieh MA, Ibrahim HM, Atieh AH. J Periodontol. 2010 Oct;81(10):1350-66. BACKGROUND: Platform

More information

SHAPE OPTIMIZATION AND TOLERANCE ANALYSIS OF DENTAL IMPLANTS BY MEANS OF VIRTUAL MODELS

SHAPE OPTIMIZATION AND TOLERANCE ANALYSIS OF DENTAL IMPLANTS BY MEANS OF VIRTUAL MODELS SHAPE OPTIMIZATION AND TOLERANCE ANALYSIS OF DENTAL IMPLANTS BY MEANS OF VIRTUAL MODELS PEZZUTI, Eugenio; PISCOPO, Giampietro; UBERTINI, Alessio; VALENTINI, Pier Paolo University of Rome Tor Vergata Dept

More information

Straumann Original Clinical Review. General Information

Straumann Original Clinical Review. General Information Straumann Original Clinical Review General Information Perfect, precise components match Proven mechanical performance Long-term clinical effect PROVEN MECHANICAL PERFORMANCE A number of studies have investigated

More information

Clinical evaluation of jacket crowns made of the Estenia indirect composite

Clinical evaluation of jacket crowns made of the Estenia indirect composite Clinical evaluation of jacket crowns made of the Estenia indirect composite Hideaki Shiono, DDS, PhD, a,b Hiroyasu Koizumi, DDS, PhD, b,c Mika Nemoto, DDS, PhD, b,c Yumi Ishikawa, DDS, b Hideo Matsumura,

More information

EFFECT OF CASTING PROCEDURE ON SCREW LOOSENING OF UCLA ABUTMENT IN TWO IMPLANT-ABUTMENT CONNECTION SYSTEMS

EFFECT OF CASTING PROCEDURE ON SCREW LOOSENING OF UCLA ABUTMENT IN TWO IMPLANT-ABUTMENT CONNECTION SYSTEMS ORIGINAL ARTICLE EFFECT OF CASTING PROCEDURE ON SCREW LOOSENING OF UCLA ABUTMENT IN TWO IMPLANT-ABUTMENT CONNECTION SYSTEMS Chun-Yeo Ha 1, DDS, MSD, Chang-Whe Kim 2, DDS, MSD, PhD, Young-Jun Lim 3, DDS,

More information

The Sheffield Fit Test - Revisited. Academic Unit of Restorative Dentistry, School of Clinical Dentistry, University of

The Sheffield Fit Test - Revisited. Academic Unit of Restorative Dentistry, School of Clinical Dentistry, University of The Sheffield Fit Test - Revisited Dr A. Johnson 1 MMedSci; PhD; LCGI; MCGI; FHEA, Dr C.A. Miller 2 BEng (hons), PhD. Dr G.E. White 1 MMedSci; PhD; CGIA. 1 Academic Unit of Restorative Dentistry, School

More information

An Innovative Technique to select Angled. Abutment using Inclination Gauge ORIGINAL ARTICLE ABSTRACT MATERIALS AND METHODS INTRODUCTION

An Innovative Technique to select Angled. Abutment using Inclination Gauge ORIGINAL ARTICLE ABSTRACT MATERIALS AND METHODS INTRODUCTION ORIGINAL ARTICLE An Innovative Technique to select Angled 10.5005/jp-journals-10012-1149 Abutment using Inclination Gauge An Innovative Technique to select Angled Abutment using Inclination Gauge 1 Anshul

More information

Since Branemark introduced the osseointegration

Since Branemark introduced the osseointegration RESEARCH Influence of Abutment Screw Design and Surface Coating on the Bending Flexural Strength of the Implant Set Célio Jesus do Prado 1 * Flávio Domingues das Neves 1 Carlos José Soares 2 Kelly Abadia

More information

IQ IMPLANTS Smart Choice PRODUCT C ATA L O G

IQ IMPLANTS Smart Choice PRODUCT C ATA L O G IQ IMPLANTS Smart Choice PRODUCT C ATA L O G www.iqimplants.com info@iqimplants.com 2 TABLE OF CONTENTS GENERAL Company Profile 3 IMPLANTS 5 PROSTHETICS 13 TOOLS 27 3 COMPANY PROFILE IQ Implants Ltd. reflects

More information

Available online at ScienceDirect

Available online at   ScienceDirect Available online at www.sciencedirect.com ScienceDirect Procedia Engineering 69 ( 2014 ) 1251 1257 24th DAAAM International Symposium on Intelligent Manufacturing and Automation, 2013 Comparative Study

More information

Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis

Stress distribution of two commercial dental implant systems: A three-dimensional finite element analysis Journal of Dental Sciences (2013) 8, 261e271 Available online at www.sciencedirect.com journal homepage: www.e-jds.com ORIGINAL ARTICLE Stress distribution of two commercial dental implant systems: A three-dimensional

More information

ß For cement-retained crowns and bridges. Note: synocta cementable abutments. ß Can be especially helpful with large multi-unit cases Note:

ß For cement-retained crowns and bridges. Note: synocta cementable abutments. ß Can be especially helpful with large multi-unit cases Note: Choosing the correct Straumann abutment for crowns, bridges, and bars Straumann abutments are torqued to 35 Ncm by using the ratchet with torque control device and the appropriate abutment driver. Abutments

More information

More than pure esthetics. The natural and strong solution.

More than pure esthetics. The natural and strong solution. More than pure esthetics. The natural and strong solution. Expand your patient pool with an innovative solution. With the Straumann PURE Ceramic Implant, clinicians have an additional option to convince

More information

The Screw Loosening and Fatigue Analyses of Three Dimensional Dental Implant Model

The Screw Loosening and Fatigue Analyses of Three Dimensional Dental Implant Model The Screw Loosening and Fatigue Analyses of Three Dimensional Dental Implant Model M. Wierszycki 1,2, W. Kąkol 2, T. Łodygowski 1 1 Institute of Structural Engineering, Poznań University of Technology,

More information

In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis

In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis ORIGINAL RESEARCH Cobalt-Chromium 10.5005/jp-journals-10024-1836 and Nickel-titanium Clasps In vitro Assessment of Clasps of Cobalt-Chromium and Nickel-titanium Alloys in Removable Prosthesis 1 Mohammed

More information

Students Opinion of a Predoctoral Implant Training Program

Students Opinion of a Predoctoral Implant Training Program Students Opinion of a Predoctoral Implant Training Program Hirofumi Kido, D.D.S., Ph.D.; Katsuki Yamamoto, D.D.S., Ph.D.; Kae Kakura, D.D.S., Ph.D.; Kenzo Morinaga, D.D.S., Ph.D.; Takashi Matsuura, D.D.S.,

More information

fracture resistance of abutment retention screws made of titanium, polyetheretherketone and carbon fiber-reinforced polyetheretherketone

fracture resistance of abutment retention screws made of titanium, polyetheretherketone and carbon fiber-reinforced polyetheretherketone Original Research Implantology Fracture resistance of abutment screws made of titanium, polyetheretherketone, and carbon fiber-reinforced polyetheretherketone Eduardo Aloisio Fleck NEUMANN (a) Cristina

More information

NUMERICAL STUDY OF STRESSES IN DENTAL MATERIALS BY USING FINITE ELEMENT ANALYSIS

NUMERICAL STUDY OF STRESSES IN DENTAL MATERIALS BY USING FINITE ELEMENT ANALYSIS NUMERICAL STUDY OF STRESSES IN DENTAL MATERIALS BY USING FINITE ELEMENT ANALYSIS Gunjal Vinayak B. PG Student, Department of Mechanical Engineering, Rajarambapu Institute of Technology, Sakharale, India

More information

(c) B. Ravi, IIT Bombay 1

(c) B. Ravi, IIT Bombay 1 Collaborative Engineering Evaluation, Testing TKP Safety, TKP System Mechanism FEA Physical testing Fatigue Wear What How Evaluate Testing Judgement Criteria's Testing Conditions Parameters Measured OrthoCAD

More information

A Three-Dimensional Finite Element Study to Characterize the Influence of Load Direction on Stress Distribution in Bone Around Dental Implant

A Three-Dimensional Finite Element Study to Characterize the Influence of Load Direction on Stress Distribution in Bone Around Dental Implant Copyright 2014 American Scientific Publishers All rights reserved Printed in the United States of America Journal of Biomaterials and Tissue Engineering Vol. 4, 1 7, 2014 A Three-Dimensional Finite Element

More information

Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring

Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring Original Research Three dimensional finite element analysis of immediate loading mini over denture implants with and without acrylonitrile O ring Srinivasan Jayaraman, Sreekanth Mallan 1, Babu Rajan, Murugan

More information

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants

ISO INTERNATIONAL STANDARD. Dentistry Implants Dynamic fatigue test for endosseous dental implants INTERNATIONAL STANDARD ISO 14801 Second edition 2007-11-15 Dentistry Implants Dynamic fatigue test for endosseous dental implants Art dentaire Implants Essai de fatigue dynamique pour implants dentaires

More information

Implant Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis

Implant Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis Implant Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis Xi Ding, MDS, 1 Xing-Hao Zhu, MDS, 2 Sheng-Hui Liao, PhD, 3

More information

MICRODENT EXTERNAL CONNECTION PRODUCTS GUIDE EXTERNAL CONNECTION IMPLANTS IMPLANTOLOGY SYSTEMS. Product presentation. Digital information

MICRODENT EXTERNAL CONNECTION PRODUCTS GUIDE EXTERNAL CONNECTION IMPLANTS IMPLANTOLOGY SYSTEMS. Product presentation. Digital information Product presentation Digital information IMPLANTOLOGY SYSTEMS EXTERNAL CONNECTION PRODUCTS GUIDE www.microdentsystem.com Implant catalogues SYSTEM UNIVERSAL Implantes de conexión externa Other catalogues

More information

ORIGINAL ARTICLE. Key words: Osseointegration, Mini-implant, Finite element analysis, Von-Mises stress INTRODUCTION

ORIGINAL ARTICLE. Key words: Osseointegration, Mini-implant, Finite element analysis, Von-Mises stress INTRODUCTION ORIGINAL ARTICLE Three-dimensional finite element analysis for determining the stress distribution after loading the bone surface with two-component mini-implants of varying length Bohm Choi, DDS, MSD,

More information

Mechanical and technical risks in implant therapy.

Mechanical and technical risks in implant therapy. Mechanical and technical risks in implant therapy. Salvi GE, Brägger U. Int J Oral Maxillofac Implants. 2009;24 Suppl:69-85. Department of Periodontology, School of Dental Medicine, University of Bern,

More information

Effects of Mesiodistal Inclination of Implants on Stress Distribution in Implant-Supported Fixed Prostheses

Effects of Mesiodistal Inclination of Implants on Stress Distribution in Implant-Supported Fixed Prostheses Effects of Mesiodistal Inclination of Implants on Stress Distribution in Implant-Supported Fixed Prostheses Alper Çağlar, PhD 1 /Cemal Aydin, PhD 2 /Jülide Özen, PhD 3 /Caner Yilmaz, PhD 2 /Turan Korkmaz,

More information

STRESS ANALYSIS OF DENTAL IMPLANT INSERTED IN THE MANDIBLE

STRESS ANALYSIS OF DENTAL IMPLANT INSERTED IN THE MANDIBLE Journal of MECHANICAL ENGINEERING Strojnícky časopis, VOL 68 (2018), NO 1, 25-32 STRESS ANALYSIS OF DENTAL IMPLANT INSERTED IN THE MANDIBLE ŁAGODA Agnieszka 1, NIESŁONY Adam 2 1 Faculty of Mechanical Engineering,

More information

Root fracture is a common and serious cause of

Root fracture is a common and serious cause of Root Fracture of Abutment Teeth for Partial Removable Dental Prostheses Yoko Mizuno, DDS, PhD 1 /Tomoya Gonda, DDS, PhD 2 /Toshihito Takahashi, DDS, PhD 3 / Akiko Tomita, DDS, PhD 4 /Yoshinobu Maeda, DDS,

More information

More than an implant. A sense of trust. Straumann Dental Implant System

More than an implant. A sense of trust. Straumann Dental Implant System More than an implant. A sense of trust. Straumann Dental Implant System More than an implant. A sense of trust. When patients need a tooth replacement, you want to provide them with the best possible solution

More information

At present, there are several

At present, there are several CLINICAL USE OF NEW MAGNETIC ATTACHMENTS FOR IMPLANT-SUPPORTED OVERDENTURES Takehiro Fujimoto, DDS, PhD Atsushi Niimi, DDS, DMSc Itsuki Murakami, DDS, MS, PhD Minoru Ueda, DDS, PhD KEY WORDS Brånemark

More information

SPECIALIST IN CUSTOMIZED PROSTHETICS

SPECIALIST IN CUSTOMIZED PROSTHETICS STRAUMANN CARES digital solutions SPECIALIST IN CUSTOMIZED PROSTHETICS Excellent solutions with Straumann Specialist in customized prosthetics Digitalizing your workflow Attaining an accurate restoration

More information

COMPARISON OF THE COMPRESSIVE STRENGTH

COMPARISON OF THE COMPRESSIVE STRENGTH RESEARCH COMPARISON OF THE COMPRESSIVE STRENGTH OF 3DIFFERENT IMPLANT DESIGN SYSTEMS Jose E. Pedroza, DMD, MS; Ysidora Torrealba, DDS, MSD; Augusto Elias, DMD, MSD; Walter Psoter, DDS, PhD The aims of

More information

Novel Removable Keeper System for Magnetic Attachments on Overdenture Abutments

Novel Removable Keeper System for Magnetic Attachments on Overdenture Abutments Showa Univ J Med Sci 27 2, 137 142, June 2015 Technical Note Novel Removable Keeper System for Magnetic Attachments on Overdenture Abutments Sawako TAKEUCHI 1, Yuji SATO 1, Dan NATHANSON 2 Noboru KITAGAWA

More information

More than an implant. A sense of trust.

More than an implant. A sense of trust. Straumann Dental Implant System More than an implant. A sense of trust. 45X.002.indd 1 20.02.14 14:36 More than an implant. A sense of trust. When patients need a tooth replacement, you want to provide

More information

Accuracy of Manual Torque-limiting Devices for Use in Oral Implantology

Accuracy of Manual Torque-limiting Devices for Use in Oral Implantology 48 EDI BDIZ EDI Q&R Committee Accuracy of Manual Torque-limiting Devices for Use in Oral Implantology Dr Jörg Neugebauer, Dr Thea Lingohr, Dr Viktor E. Karapetian, Raid Shihadeh, Prof Joachim E. Zöller

More information

In Vitro Study of Reduction of Stress Transferred onto Tissues around Implants Using a Resilient Material in Maxillary Implant Overdentures

In Vitro Study of Reduction of Stress Transferred onto Tissues around Implants Using a Resilient Material in Maxillary Implant Overdentures J Med Dent Sci 2007; 54: 17 23 Original Article In Vitro Study of Reduction of Stress Transferred onto Tissues around Implants Using a Resilient Material in Maxillary Implant Overdentures Manabu Kanazawa,

More information

EFFECTIVE DATE: 04/24/14 REVISED DATE: 04/23/15, 04/28/16, 06/22/17, 06/28/18 POLICY NUMBER: CATEGORY: Dental

EFFECTIVE DATE: 04/24/14 REVISED DATE: 04/23/15, 04/28/16, 06/22/17, 06/28/18 POLICY NUMBER: CATEGORY: Dental MEDICAL POLICY SUBJECT: DENTAL IMPLANTS PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product (including an Essential

More information

Surveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1

Surveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1 د. فائزة Lec.3 Prosthodontics Surveying The ideal requirements for successful removable partial denture are: 1. Be easily inserted and removed by the patient. 2. Resist dislodging forces. 3. It should

More information

A biomechanical investigation of mandibular molar implants: reproducibility and validity of a finite element analysis model

A biomechanical investigation of mandibular molar implants: reproducibility and validity of a finite element analysis model Omori et al. International Journal of Implant Dentistry (2015) 1:10 DOI 10.1186/s40729-015-0011-5 RESEARCH Open Access A biomechanical investigation of mandibular molar implants: reproducibility and validity

More information

Stress analysis of different configurations of 3 implants to support a fixed prosthesis in an edentulous jaw

Stress analysis of different configurations of 3 implants to support a fixed prosthesis in an edentulous jaw Prosthodontics Stress analysis of different configurations of 3 implants to support a fixed prosthesis in an edentulous jaw João Paulo da Silva-Neto (a) Marcele Jardim Pimentel (a) Flávio Domingues das

More information

US Implant System 2013 PROSTHETIC PROCEDURE

US Implant System 2013 PROSTHETIC PROCEDURE US Implant System 2013 PROSTHETIC PROCEDURE Contents TS Implant System Cement retained restoration 06 Cement retained bridges with the Cement abutment system 18 Cement retained crown with the Angled abutment

More information

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282

Anchorage system. tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282 tomas / Sets Page 270 tomas / Pins Page 271 Instruments and accessories Page 273 Patient consultation material Page 282 266 . innovative comprehensive efficient. Dentaurum Online Shop shop.dentaurum.com

More information

For the oral rehabilitation

For the oral rehabilitation For the oral rehabilitation Contents For superstructures Chapter 1 Overview of prosthetic modes 1. Prosthetic modes of the PLATON system 2. Factors involved in prosthetic design Chapter 2 Cement-retained

More information

An In Vitro Load Evaluation of a Conical Implant System with 2 Abutment Designs and 3 Different Retaining-Screw Alloys

An In Vitro Load Evaluation of a Conical Implant System with 2 Abutment Designs and 3 Different Retaining-Screw Alloys An In Vitro Load Evaluation of a Conical Implant System with 2 Abutment Designs and 3 Different Retaining-Screw Alloys Christian Erneklint, DT 1 /Per Ödman, LDS, Odont Dr 2 /Ulf Örtengren, LDS, Odont Dr

More information