Evaluation of some factors that may affect the accuracy of implant
|
|
- Lily Glenn
- 5 years ago
- Views:
Transcription
1 Cairo Dental Journal (25) Number (2), 219:226 May, 2009 Evaluation of some factors that may affect the accuracy of implant transfer impression Ashraf A. Amer 1 and Dina H. Mostafa 2 1. Assistant Professor, Prosthodontic Department, Faculty of Oral and Dental Medicine, Cairo University. 2. Assistant Professor, Biomaterials Department, Faculty of Oral and Dental Medicine, Cairo University. Abstract Three implants were vertically arranged in a straight line parallel to one another in an acrylic block. Three impression transfers were manually tightened to the implants by 10 different operators and by a 30 N/Cm torque as a control. The distance between the transfers was measured under microscope from top and side views. Three abutments were connected to the implants and an electroformed bridge was constructed to fit them. Four groups of impressions were made for the implants using abutments as an impression transfers and the ordinary transfers; both were tightened manually and by 30 N/Cm torque. Impressions were poured into stone using new implant analogues. Another two groups of impressions were made using abutments that were tightened manually and by torque, but the stone models were made using old implant analogues. The abutments were connected to each stone model and the electroformed bridge was placed over them. The bridge seating was evaluated using a travelling microscope. Conclusions: Within the limitation of this study: 1- On a microscopic scale, the tightening force of the transfer screw does affect its relation to the implant. 2- There is no practical difference between using the transfers or the actual abutments while making impression for implants, whether they were manually tightened or tightened with a controlled torque ratchet. 3- The repeated use of implant analogues for several times may affect the clinical accuracy of the transfer impression.
2 (220) Ashraf A. Amer and Dina H. Mostafa C.D.J. Vol. 25. No. (2) INTRODUCTION Implants are more conservative than fixed bridges as it does not destroy healthy tooth structure. 1 Since implants are indirectly restored, the relationship between the implants and the surrounding hard and soft tissues should be accurately transferred to the working models. Errors in transfer during the impression procedure reduce the possibility of the prosthesis passive fit. In addition, problems with articulation of the working cast, axial contour and interproximal contacts, open margins, lack of retention and resistance to displacement, and errors in occlusal relationships of the frameworks may occur. 2 Several factors affect the accuracy of the transmission of this relationship: accuracy of impression and cast materials 3,4 as well as the implant impression techniques. 5,6 As a result of inaccurate transfer, two main clinical problems may develop, the poor marginal fit of the impassive framework and the abutment screw loosening. 7,8 In the majority of the implant systems, the abutment is fastened to the fixture by an abutment fixation screw. The recommended torque values are based on each manufacture system. The optimum preload is the tightening torque above the level of the initial contact force and falls within the material elastic range of the abutment screw. 9 The screw is either tightened manually or mechanically. The amount of torque placed on the dental implant screw manually is operator dependent, under-torque and over-torque may occur. Moreover, experienced operators deliver varying amounts of torque to a dental implant screw. Under-torquing leads to screw loosening and over-torquing have been recommended to maintain better retention of the abutment screw. 10 However, this might lead to fracture of the joint. 11 Screw loosening in implant dentistry is a problem facing many operators experienced or inexperienced. During tightening of an implant screw, a force is exerted causing a preload stress in the abutment screw within its elastic range. During function the abutment is subjected to external forces and stresses. It is at this point that the screw joint is said to be protected against external load applications as long as these external loads do not exceed the preload. When the preload is achieved, the screw will transmit the entire external load applied during function to the clamped parts. This protection is beneficial to the fatigue performance of the screw. However, when the total external load experienced by the screw is greater than the yield strength of the screw, then the protection afforded by the preload is lost and the joint becomes unstable, the potential for screw loosening and/or fracture exists. 12,26 Many factors affect the precision of fit between the framework and the implant structure as: the structure and properties of the different types of the implant materials, the accuracy during the transfer of the relationship of the implants to the working cast, the techniques and types of framework constructed, the dentist s skill, and the torquing between the implant and the abutment that may affect the screw joint itself The exact transfer of the implants position to the working cast may be considered of the most important factors. Any minute variation in this relationship, or any defect or deformation of the used materials, affects the framework s accuracy of fit. Errors that result from the transfer of implant position during the impression procedures often make it necessary to section and solder metal frameworks repeatedly. 17 Some operators modified the copings during recording the impression to decrease the possibility of inaccuracy The precision of fit of the framework is more important when fixed prostheses are connected to implants than to natural teeth as natural teeth have the ability to adjust to the misfit because of the mobility of the periodontal ligament. 22 To insure long-term service precise fit must be present. Poor fit of frameworks leads to failure of the restoration 23 as well as the supporting structures as a deformation of the surrounding bone in the form of compression between the implants. 24 This study was designed to 1. Evaluate the effect of the tightening force of the implant-transfer screw on the accuracy of recording the relationship between implants in the working cast.
3 Evaluation of some factors (221) 2. Estimate the clinical acceptance of the fitness of a bridge framework to the implant abutments with the following variables: a. Manual torque versus controlled torque ratchet to fasten the transfers. b. Using the definitive abutment as transfers versus the regular transfers while making the impression. c. The repeated use of the implant analogues. MATERIALS and METHODS 1. Effect of tightening force of the impression abutment on their position in relation to implants Three implants 1 were vertically arranged in a straight line parallel to one another and secured in this position. Self cure Acrylic resin was poured on the assembly. This resulted in an acrylic block with 3 implants fixed in position (Figure 1). Three impression transfers were tightened to the implants. Using a sharp lancet, the transfers were marked with a fine scratch at the middle of the side corresponding to the buccal surface. The mark was extended to the top surface of the transfer (Figure 2). Marks on the first, second and third transfer were referred to as a,b, and c respectively. They were confirmed to be visible under a stereomicroscope 2 with an electronic camera, specially designed for digital photomicrography, which was connected to a computer (Figure 3). On the computer s monitor, three measurements were made to measure the distances between the transfers from both the lateral and the top views (Figure 4): The experiment measurements were done when a group of operators tightened the transfers manually. Five males (group B) and five females (group C) operators, with more than 3 years of clinical experience in implant dentistry, were asked to hand-torque the three transfers into the implants. The operators were seated in their normal operating position, wearing examination latex gloves and each operator was asked to tighten the transfers to the implant independently 5 separate times with the force normally used in clinical situations. The transfers were tightened ten separate times using a torque controlled ratchet at 30N/cm force to act as the control group (group A). Each time, the distance between the three transfers was measured from the lateral and top views. Fig. (1) The implants within the acrylic block. Measurement 1, from a to b (distance between first and second implant) Measurement 2, from b to c (distance between second and third implant) Measurement 3, from a to c (distance between first and third implant) Fig. (2) Marked transfer (Top view) 1. Tapered Screw-Vent Implant System, Zimmer Dental Inc, USA. 2. Olympus Stereomicroscope with Olympus DP-10 Digital Camera, Olympus America, Inc., USA
4 (222) Ashraf A. Amer and Dina H. Mostafa C.D.J. Vol. 25. No. (2) restoration was made and tried on the abutments (Fig. 5); this bridge was considered as a master bridge. Seven groups of transfer impressions were made, by different methods, for the implants using mechanically mixed polyether impression material in an indexed custom acrylic tray to control the impression variables every time the impression was taken. Impression was then removed and transfers were unscrewed from the implants. Implant analogue was attached to each impression coping, snapped into its place in the impression, boxed and poured in extra-hard stone. Fig. (3) The stereomicroscope The transfers were removed and replaced by the abutments on the stone model. Then the bridge was seated on the abutments by applying gentle finger pressure. The accuracy of seating of this bridge was measured in 16 marked locations: 6 marks bucally (2 for each abutment), 6 marks lingually (2 for each abutment), and 2 marks at the side of each peripheral abutment. The accuracy of seating was represented by the vertical distance between the bridge margin and a groove on the abutment at definite marked points. This was measured using a travelling microscope 1 with 1µ accuracy. Each measure was repeated five times for each model obtained from the following groups of impression methods: Group 1: using manually tightened transfer. Fig. (4) Distance measurements (lateral view) 2. Effect of the impression method on the prosthesis fitness The same acrylic model was used in this part of the study. Three crown and bridge abutments were fastened to the implants using the 1.25mm hex driver and the 30 N/Cm torque controlled ratchet. An impression was made for the abutments and poured directly into extra hard stone. On that model, an electroformed gold bridge Fig. (5) The electroformed bridge on the master model Nikon SC-102
5 Evaluation of some factors (223) Group 2: using transfer tightened with controlled torque. Group 3: using an abutment as a transfer that was manually tightened. Group 4: using an abutment as a transfer that was tightened with controlled torque. Group 5: using an abutment as a transfer manually tightened to a used implant analogue. Group 6: using an abutment as a transfer tightened with controlled torque to a used implant analogue. Group 7: (control group) bridge on the acrylic master model tightened with torque. To properly function as impression transfer in groups 3, 4, 5 and 6, the abutment were modified by cementing custom made metal caps on them (Fig 6). These caps provided flat sides and undercuts to resemble transfers. Statistical analysis was carried out using SAS program. 26 Student t test (Procedure T-TEST of SAS) was run to compare the effect of new and old analogues. One way analysis of variance (Procedure ANOVA of SAS) followed by Duncan s multiple Range Test were used to test the effect of techniques on bridge fitness and distance between analogues. Data were presented as mean and standard deviation values in tables 1 and 2. Fig. (6) Modified abutments on stone model Results 1- The effect of tightening force on the accuracy of recording the relationship of the implants to the working cast Table 1 shows the means and standard deviation values, ANOVA and Duncans test results for the effect of different operators forces on the distance between the transfers from the top and lateral views. Statistically significant difference was found between each group, with the males mean tightening forces value being closer to the mean control group value. Table (2) The means, standard deviation values of the effect of modified impression techniques on the fitness of the frameworks to the abutment/implant. Group Analogue Mean (µ) S.D. dt Transfer - manual (group1) new a Transfer with controlled torque (group2) new a Abutment - manual (group3) new a Abutment with controlled torque (group4) new a Abutment controlled torque (group5) old b Abutment- manual (group 6) old c Control (group7) new a S.D. = Standard deviation. dt = Duncans Multiple Range Test. Means with the same letter within each column are not significantly different at p 0.05.
6 (224) Ashraf A. Amer and Dina H. Mostafa C.D.J. Vol. 25. No. (2) Table (1) The effect of the tightening forces on the distance between the transfers. Force No. of Trials lateral view top view Mean (Pixels) S.D. dt Mean (Pixels) S.D. dt Group A ±0.6 a ±0.8 a Group B ±10.8 b ±17.2 b Group C ±32.2 c ±29.4 c S.D. = Standard deviation. dt = Duncans Multiple Range Test. Means with the same letter within each column are not significantly different at p The effect of using abutments as an impression transfers on the fitness of the bridge framework when using used and unused analogues: Table 2 shows the means and standard deviation values of ANOVA and Duncans test of the effect of using modified abutments as impression transfers on the framework seating on the abutments. It also compares the framework seating in case of manual and torque controlled tightening of both transfers and abutments while making the impression. Both impression techniques gave insignificant results between the four groups. Table 2 also compares the means, standard deviation values of ANOVA and Duncans test of the effect of using new and old analogs on the fitness of the framework to the abutments. A remarkable misfit of the framework to abutments was found when old analogs were used. Discussion Passive seating of the suprastructure whatever fixed or removable is of paramount importance for the long serviceability of the implants. Clinically the suprastructure in many cases with natural teeth can be seated while applying some pressure. Sometimes the patient may tolerate this situation and adapt to it. But this is not applied to implant supported prosthesis where undue forces will be applied continuously especially in fixed restoration cases. On the long run this may not be tolerated by the supporting bone and may affect the implant prognosis. However, with removable restorations, the problem is less but still present as the restoration is not continuously seated in the patient mouth Errors in transfer during the impression procedure reduce the possibility of the prosthesis passive fit. Tightening the screw attaching the transfer to the implant is an important point to be considered. The mean habitual closing torque of different operators is not constant. It was found that individual finger forces differ by gender, and not by hand dimension and age. 28 It is also reported that some male operators do not normally apply maximum possible tightening force clinically for fear of endangering the osseointegration. 18, 27 Thus, to obtain optimal preload in the screw with least transmission of force to the bone, it is recommended to use an appropriate torque driver. 12,28 In this study, tightening the screw manually by different individuals was different than that done with the controlled torque ratchet. This difference was clearly reflected in the measurements done under the microscope which indicates that different screw torquing does affect the position of the transfer in relation to the implant. However it seems that the effect of this microscopic difference does not clinically exceed the acceptance level of accuracy as manifested by the insignificant difference in bridge seating in manual and definite torque groups. The results of this study showed no difference between manual torquing and controlled torquing during
7 Evaluation of some factors (225) impression making, this may be attributed to using the same method of torquing during impression making, model pouring and abutment tightening for testing the bridge seating on that cast. However it looks that the force used during screw tightening is important as its variation during different steps of construction from one subject to another, for example the dentist and the laboratory technicians, may be reflected on the passive seating of the final prosthesis. This may be avoided by using a fixed torque during clinical and laboratory steps through the use of controlled torque ratchets. This point needs further investigation using definite variable torques during different steps. It was clear from this study that using old implant analogue does not allow the transfer or the abutment to be precisely positioned in place, so the resulting model will not accurately represent the real situation and the restoration may apply undue forces to the implants. Thus the implant analogue should be used for only one time or be fabricated from a rigid material that does not deform easily under tightening force. Many implant systems in the market now use a single component to act as an implant insertion mount, to do the function of an impression transfer, and used as final abutment. Although the results of this study showed no clinically significant difference between using the traditional transfer and the use of abutment as a transfer during impression making, a potential disadvantage of this technique may occur in the multiple screwing and unscrewing of the abutment many times that may deteriorate the hex angels of the implant-abutment joint, leading to less accurate fitting between them However, repeated torquing does not affect the ultimate strength of the screw itself. 32,33 Conclusions Within the limitation of this study the following may be concluded: 1. On a microscopic scale, the tightening force of the transfer screw does affect its relation to the implant. 2. There is no practical difference between using the transfers or the actual abutments while making impressions for implants, whether they were manually tightened or tightened with a controlled torque ratchet. 3. The repeated use of implant analogues for several times may affect the clinical accuracy of the transfer impression. Acknowledgment The authors would like to express their deepest gratitude to Professor Dr. Inas Samy for her generous help and extreme support. References 1. Hebel KS, Galindo D, Gajjar RC. Implant position record and implant position cast: Minimizing errors, procedures and patient visits in the fabrication of the milled-bar prosthesis. J Prosthet Dent 2000; 83 (1): Azizogli MA, Catania EM and Weiner S. Comparison of the accuracy of working casts made by the direct and transfer coping procedures. J Prothet Dent 1999; 84(4): Barrett MG, de Rijk WG, Burgess JO. The accuracy of six impression techniques for osseointegrated implants. J Prosthodont. 1993; 2: Hsu CC, Millstein PL, Stein RS. A comparative analysis of the accuracy of implant transfer techniques. J Prosthet Dent. 1993;69: Herbst D, Nel JC, Driessen CH, Becker PJ. Evaluation of impression accuracy for osseointegrated implant supported superstructures. J Prosthet Dent 2000; 83(5): Vigolo P, Fonzi F, Majzoub Z, Cordioli G. An evaluation of impression techniques for multiple internal connection implant prostheses. J Prosthet Dent 2004; 92(5): Jemt T, Laney WR, and Harris D. Osseeintegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study. Int J Oral Maxillofac Implant 1991; 6:29-35.
8 (226) Ashraf A. Amer and Dina H. Mostafa C.D.J. Vol. 25. No. (2) 8. Jemt T, Pettersson P. A 3-year fellow-up study on single implant treatment. J Dent 1993;21: Brunski JB. Biomaterials and biomechanics in dental implant design. Int J Oral Maxillofac Implants. 1988;3: Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants. 1995;10: Pesun IJ, Brosky ME, Korioth TWP, Hodges J, Devoe BJ. Operator-induced compressive axial forces during implant gold screw fastening. J Prosthet Dent 2001; 86 (1): Lang LA. May KB. and Wang RF. The effect of the use of a counter-torque device on the abutment-implant complex. J Prosthet Dent 1999; 81: Jemt T, Linden B. Fixed implant supported prosthesis with welded titanium frameworks. Int J Periodontics Restorative Dent. 1992;12: Linehan AD, Windeler AS. Passive fit of implant-retained prosthetic framework improved by electric discharge machining. J Prosthodont. 1994;3: Shiau JC, Chen LL, Wu CT. An accurate impression method for implant prosthesis fabrication. J Prosthet Dent. 1994;72: May KB, Lang BR, Lang BE, Wang RF. Periotest method: Implant-supported framework fit evaluation in vivo. J Prosthet Dent 1998; 79 (6): Wee AG, Schneider RL, Aquilino SA, Huff TL, Lindquist TJ, Williamson DL. Evaluation of the accuracy of solid implant casts. J Prosthodont. 1998;7: Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J Prosthet Dent 2002; 88 (2): Assif D, Marshak B and Schmidt A. Accuracy of implant impression techniques. Int. J Oral Maxillofac. Implants. 1996; 11: Assif D, Nissan J, Varsano I, et al: Accuracy of implant impression splinted techniques: Effect of splinting material. Int J Oral Maxillofac Implants 1999; 14: Vigolo P, Majzoub Z, Cordioli G. In vitro comparison of master cast accuracy for single-tooth implant replacement. J Prosthet Dent 2000; 83 (5): Takahashi T and Johan G. Fit of implant frameworks: An invitro comparison between two fabrication techniques. J Prothet Dent2003; 89: Carlsson B, Carlsson GE. Prosthodontic complications in osseointegrated dental implant treatment. Int J Oral Maxillofac Implants. 1994;9: Jemt T, Lekholm U. Measurements of bone and frame-work deformations induced by misfit of implant superstructures: A pilot study in rabbits. Clin Oral Implants Res 1998;9: Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001; 86 (1): SAS (1988): Statistical Analysis Systems, STAT/User s Guide, Release 6.03 ed., SAS Institute, Cary NC, USA. 27. Bickford JH. An introduction to the design and behavior of bolted joints. New York: Marcel Dekker, Inc; Gross M. Kozak D. Laufer BZ and Weiss EI. Manual closing torque in five implant abutment systems: An invitro comparative study. J Prothet Dent 1999; 8(5): Pylatiuk, C. Kargov, A. Schulz, S. And Döderlein L. Distribution of grip force in three different functional prehension patterns. J Med. Eng. & technol. 2006; 30(3): Dixon D.L, Breeding L.C, Sadler J.P and Mckay M.L. Comparison of screw loosening, rotation, and deflection among three implant designs. J Prosthet Dent 1995; 74 (3): Khraisat A, Stegaroiu R, Nomura S, Miyakawa O. Fatigue resistance of two implant/abutment joint designs. J Prosthet Dent. 2002;88: Khraisat A, Hashimoto A, Nomura S, Miyakawa O Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system Journal of Prosthetic Dentistry April 2004 (Vol. 91, Issue 4, Pages ) 33. Al Rafee MA, Nagy WW, Fournelle RA, Dhuru VB, Tzenakis GK, Pechous CE. The effect of repeated torque on the ultimate tensile strength of slotted gold prosthetic screws. J. Prosthet. Dent (2):
Objectives: Previous studies that evaluated the torque needed for removing dental
www.scielo.br/jaos A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants Felipe Miguel SALIBA 1, Mayra CARDOSO 2, Marcelo Ferreira TORRES 2, Alexandre Carvalho
More informationAccuracy of Electronic Implant Torque Controllers Following Time in Clinical Service
Accuracy of Electronic Implant Torque Controllers Following Time in Clinical Service Ricardo Mitrani, DDS, MSD 1 /Jack I. Nicholls, PhD 2 /Keith M. Phillips, DMS, MSD 3 /Tsun Ma, DMD, MS, MDS 4 Tightening
More informationLegacy Prosthetic Catalog
Legacy Prosthetic Catalog Straight Snap-On Straight Contoured Angled Contoured Ball Attachment Gold/Plastic Castable Locator Attachment Plastic Temporary Screw Receiving Plastic Non-Engaging Castable 15
More informationStraight Octagon Abutment for Tissue Level Implant
Cover screw in place The presentation that follows lists only one combination of parts. Obviously the clinical situation may call for substitution of another part on this slide Prosthetic instruments needed
More informationThe Sheffield Fit Test - Revisited. Academic Unit of Restorative Dentistry, School of Clinical Dentistry, University of
The Sheffield Fit Test - Revisited Dr A. Johnson 1 MMedSci; PhD; LCGI; MCGI; FHEA, Dr C.A. Miller 2 BEng (hons), PhD. Dr G.E. White 1 MMedSci; PhD; CGIA. 1 Academic Unit of Restorative Dentistry, School
More informationSCREW LOOSENING has been reported as
BASIC SCIENCE RESEARCH Effect of Casting Procedures on Screw Loosening in UCLA-Type Abutments Stefania Kano, DDS, MS, PhD; 1 Donald A. Curtis, DMD; 2 Paul Binon, DDS, MSA; 3 and Gerson Bonfante, DDS, MSc,
More informationOT EQUATOR CASTABLE Single Attachment for Overdentures
OT EQUATOR CASTABLE Single Attachment for Overdentures pick-up impression OT EQUATOR PROFILE CASTABLE MALE ATTACHMENT individual tray RETENTIVE CAPS OT EQUATOR HOUSING VIOLET CAP RIGID RETENTION (2.7Kg)
More informationSS Implant System 2013 PROSTHETIC PROCEDURE
SS Implant System 2013 PROSTHETIC PROCEDURE Contents TS Implant System Cement retained restoration 06 Cement-retained bridges with the Solid abutment system (non-modified abutment) 16 Cement-retained bridges
More information9. Literaturübersicht
9. Literaturübersicht 1. Adell R, Lekholm U, Rockler B, Branemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 10 (6): 387-416, 1981. 2. Assif
More informationThe LOCATOR concept. Simplicity and versatility for prosthesis fixation
The concept Simplicity and versatility for prosthesis fixation The concept Experience the freedom in prosthesis fixation Simple and secure fixation of implant-supported prostheses is essential for successful
More informationVery small abutment head easy and secure handling. Ankylos. The SmartFix concept. Prosthetic solution on angled implants
Very small abutment head easy and secure handling Ankylos The SmartFix concept Prosthetic solution on angled implants Stable prosthetic fit The area supporting the prosthesis is extended distally by the
More informationAbutment level impression Ⅰ. Combi abutment/dual abutment Multi units
Pr osthetic par t Abutment level impression Ⅰ Combi abutment/dual abutment Multi units Prosthetic part Remove Cover screw Prosthetic part Remove Healing abutment Prosthetic part Select suitable Combi abutment,
More informationTreatment Options for Restoring Edentulous Jaws using One- and Two-Piece Implants from Implant Direct Int l
Treatment Options for Restoring Edentulous Jaws using One- and Two-Piece Implants from Implant Direct Int l Two-Piece ReActive Tri-Lobe Implants with Multi-Unit Abutments One-Piece ScrewIndirect Implants
More informationsynocta Meso abutment for cement-retained restorations PROSTHETICS Step-by-step instructions
synocta Meso abutment for cement-retained restorations PROSTHETICS Step-by-step instructions INTRODUCTION synocta Meso abutment for cement-retained restorations 1 The synocta Meso abutments, available
More informationINSTRUCTIONS FOR THE DENTIST. synocta impression procedure, Screw-retained and Snap-on. Screw-retained (Open tray) Snap-on (Closed tray)
INSTRUCTIONS FOR THE DENTIST synocta impression procedure, Screw-retained and Snap-on Screw-retained (Open tray) Snap-on (Closed tray) click 1. Attach synocta Impression caps to implants. Hand-tighten
More informationImplant Restorations: A Step-By-Step Guide
Implant Restorations: A Step-By-Step Guide Drago, Carl DDS, MS ISBN-13: 9780813828831 Table of Contents Contributors. Foreword. Acknowledgments. Chapter 1. Introduction To Implant Dentistry. 1. Introduction.
More informationBlue Sky Bio Multi unit Abutments BIO Multi-Units are available for all our Bone Level Implants
BIO Multi-Units are available for all our Bone Level Implants Multi-Units abutments are ideal for following restorative plans Blue Sky Bio offers a variety of components that can be used with BIO Multi-units.
More informationCASE REPORT ABSTRACT INTRODUCTION. AIMS AND OBJECTIVES Aims
An in vitro Study to Compare the Accuracy of Master Cast Fabricated by Three Different 10.5005/jp-journals-10012-1143 Transfer Techniques for Single Tooth CASE REPORT An in vitro Study to Compare the Accuracy
More informationIMPRESSION PROCEDURES TRAINING MANUAL
IMPRESSION PROCEDURES TRAINING MANUAL 01 01 A.B. DENTAL A.B. Dental is proud to present this impression taking procedure protocol. This manual explains, step by step, the procedure while using A.B. Dental
More informationHex-Lock Abutment System. Restorative Manual
System Restorative Manual 32 Restorative options with s s are manufactured from titanium alloy and used as the support foundation for single- or multiple-unit cement-retained, partially edentulous fixed
More informationNARROW DIAMETER implant
ND NARROW DIAMETER implant TABLE OF CONTENTS ND - NARROW DIAMETER implant Implant characteristics page 04 Dental implant page 05 Open Tray Impression Transfer page 06 Titanium Abutments page 07 O-Ball
More informationcement-retained single crowns using cementable abutments
cement-retained single crowns using cementable abutments cement-retained restorations cement-retained single crowns using cementable abutments Cement-retained implant restorations are similar to conventional
More informationSimpleLine II Surgical / Prosthesis Manual
SimpleLine II Surgical / Prosthesis Manual SimpleLine II SURGICAL MANUAL Surgical Drill Sequence 04 Fixture Installation 05 Fixture Connection 06 Surgical Kit Maintenance 08 Warnings 09 04 SimpleLine ll
More information4766 Research Dr. San Antonio, TX insightdentalsystems.com
OVERVIEW OF THE INSIGHT DENTAL IMPLANT DELIVERY SYSTEM The IDS system comes in a unit dose implant system where its advantage provides sterile instrumentation in one single-use kit. It is organized to
More informationRestoring Straumann implants. with Locator abutments.
Restoring Straumann implants with Locator abutments. 1 1 Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education. Contents.
More informationAdvanced restorative techniques and the full mouth reconstruction: the use of gold copings in bridgework
Clinical Advanced restorative techniques and the full mouth reconstruction: the use of gold copings in bridgework Paul Tipton 1 Introduction From the studies produced by Lindhe and Nyman, described earlier
More informationTHE EFFECT OF ABUTMENT HEIGHT ON SCREW LOOSENING IN SINGLE IMPLANT-SUPPORTED PROSTHESES AFTER DYNAMIC CYCLIC LOADING
J Korean Acad Prosthodont : Volume 42, Number 6, 2004 THE EFFECT OF ABUTMENT HEIGHT ON SCREW LOOSENING IN SINGLE IMPLANT-SUPPORTED PROSTHESES AFTER DYNAMIC CYCLIC LOADING Nam-Gun Kim, D.D.S., Yung-Soo
More informationIMPRESSION-TAKING, BITE REGISTRATION, AND TEMPORARY RESTORATION ON CAMLOG IMPLANTS. a perfect fit
a perfect fit IMPRESSION-TAKING, BITE REGISTRATION, AND TEMPORARY RESTORATION ON CAMLOG IMPLANTS Open and closed impression-taking Impression-taking for option platform switching Bite registration Temporary
More informationUDELL DENTAL LABORATORY Instructions for Use PREAT Precision Attachments
Indications Instructions The Locator Root Attachment is designed for use with overdentures or partial dentures, retained in whole or in part by endodontically treated roots in the mandibular or maxilla.
More informationA New Procedure Reduces Laboratory Time to 6 Hours for the Elaboration of Immediate Loading Prostheses with a Titanium Frame Following Implant Placement Huard C, Bessadet M, Nicolas E, Veyrune JL * International
More informationSCD Case Study. Attachments in Prosthodontics
SCD Case Study Attachments in Prosthodontics Precision attachments are retention units milled out of alloy. They have male and female parts that fit together with tolerances of 10 microns. It is important
More informationB U J O D. Invited Review IMPLANT IMPRESSIONS SIMPLIFIED. Reddy et al. Vol. 2 Issue-3 Sept
Invited Review IMPLANT IMPRESSIONS SIMPLIFIED AUTHORS: Vahini Reddy*, Venkateswara Allu Reddy**, Sadhvi K.V*** Introduction It is said that you never get a second chance to make a first impression but
More informationOriginal Article Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results
Effect of Anti-rotational Design and Countertorque Device on Peri-implant Bone Stress During and After Preload A Preliminary Results Ming-Su Lee, DDS, MS Attending Doctor, Department of Dentistry, National
More informationCAUSE OF TECHNICAL FAILURES OF CONICAL CROWN-RETAINED DENTURE (CCRD): A CLINICAL REPORT
J Korean Acad Prosthodont : Volume 41, Number 6, 2003 CAUSE OF TECHNICAL FAILURES OF CONICAL CROWN-RETAINED DENTURE (CCRD): A CLINICAL REPORT Yang-Jin Yi, D.D.S.,M.S.D.,Ph.D., Lee-Ra Cho, D.D.S.,M.S.D.,Ph.D.,
More informationImplant Phase Flow Chart
Implant Phase Flow Chart 1. Surgical Phase Immediate Load 1. Place implant 1-2 mm below bone. 2. Attach crown posts at time of surgery using S17000 series abutments and (AS7001) Alan Key, and (AS7033)
More informationProsthetic Options in Implant Dentistry. Hakimeh Siadat, DDS, MSc Associate Professor
Prosthetic Options in Dentistry Hakimeh Siadat, DDS, MSc Associate Professor Dental Research Center, Department of Prosthodontics & Dental s Faculty of Dentistry, Tehran University of Medical Sciences
More informationLect. 14 Prosthodontics Dr. Osama
Lect. 14 Prosthodontics Dr. Osama Principles of Removable Partial Denture Design Difference in Prosthesis Support and Influence on Design: For a tooth-supported prosthesis, the movement potential is less
More informationSwissPlant Prosthetic Catalog. SwissPlant
SwissPlant Prosthetic Catalog Contoured Snap-On Angled Contoured Ball Attachment Gold/Plastic Castable Locator Attachment Plastic Temporary Screw Receiving Plastic Non-Engaging Castable Narrow Titanium
More informationCHECK-UP Determination of Impression type & Method after implant placement
CHECK-UP Determination of Impression type & Method after implant placement Fixture level impression Abutment level impression Transfer impression Conventional method Pick-up impression Plastic impression
More informationRock-solid fit. A connection you can depend on. Internal Hex with Friction-Fit
Rock-solid fit. A connection you can depend on. Internal Hex with Friction-Fit Zimmer Dental s patented friction-fit abutment connection. Drive a piton deep into rock, and it locks firm, giving the climber
More informationTelescopic Retainers: An Old or New Solution? A Second Chance to Have Normal Dental Function
Telescopic Retainers: An Old or New Solution? A Second Chance to Have Normal Dental Function Joseph B. Breitman, DMD, FACP, 1,2 Scott Nakamura, DMD, 3 Arnold L. Freedman, DDS, 4 & Irving L. Yalisove, DDS
More informationINTRODUCTION. Why Choose Cortex Prosthetics?
Prosthetic Manual 4 INDEX Introduction 6-14 Impression Techniques 15-29 Abutments Selection Handling 30-33 Temporary Restorations 34-38 Screw Reiained Restorations 39-52 Removable Restorations 53-63 5
More informationINSTRUCTIONS FOR USE STRAUMANN VARIOBASE ABUTMENT. Straumann Variobase Abutment
INSTRUCTIONS FOR USE STRAUMANN VARIOBASE ABUTMENT Straumann Variobase Abutment The ITI (International Team for Implantology) is the academic partner of Institut Straumann in the areas of research and education.
More informationTorque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver
Original Research Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver Jalil Ghanbarzadeh 1, Mahboobeh Rajabzadeh 2, Mohammadreza Nakhaei 3, Hamidreza
More informationFor the oral rehabilitation
For the oral rehabilitation Contents For superstructures Chapter 1 Overview of prosthetic modes 1. Prosthetic modes of the PLATON system 2. Factors involved in prosthetic design Chapter 2 Cement-retained
More informationß For cement-retained crowns and bridges. Note: synocta cementable abutments. ß Can be especially helpful with large multi-unit cases Note:
Choosing the correct Straumann abutment for crowns, bridges, and bars Straumann abutments are torqued to 35 Ncm by using the ratchet with torque control device and the appropriate abutment driver. Abutments
More informationSCD Case Study. Implant-supported overdentures
SCD Case Study Implant-supported overdentures An implant-retained overdenture may be indicated in patients with changed anatomy, neuromuscular disorders, significant gag reflex or considerable ridge resorption
More informationEffect of impression coping and implant angulation on the accuracy of implant impressions: an in vitro study
ORIGINAL ARTICLE DOI:10.4047/jap.2010.2.4.128 Effect of impression coping and implant angulation on the accuracy of implant impressions: an in vitro study Si-Hoon Jo, DDS, Kyoung-Il Kim, DDS, Jae-Min Seo,
More informationTrabecular Metal and Tapered Screw-Vent Implant Systems. Restorative Manual
Trabecular Metal and Tapered Screw-Vent Implant Systems Restorative Manual Table of Contents Overview 3 General Information 3 Platform Plus Connection 3 Restorative Options 4 Partially Edentulous 4 Fully
More informationFixed Partial Dentures /FPDs/, Implant Supported. in implant prosthodontics
Fixed Partial Dentures /FPDs/, Implant Supported Prosthesis/ISP/ in implant prosthodontics Prof.dr.Tamas Divinyi Semmelweis University, Faculty of Dentistry Department of Oral and Maxillofacial Surgery
More informationverification jig fabrication
verification jig fabrication full-arch restorations verification jig fabrication Use this technique to verify and achieve passive fitting metal framework for a bridge, hybrid prosthesis or for an overdenture
More informationUDELL DENTAL LABORATORY Instructions for Use
DALLA BONA Summary Cylindrical Dalla Bona Rigid precision attachment. Radicular telescopic stud. Adjustable frictional retention. Gold alloy male (Elitor) and female (OSV). Fixation: Male - soldered to
More informationQuick start Prosthetics
EN Quick start Prosthetics Easy and effective YELLOW PROSTHETIC PLATFORM QN CEMENTED RESTORATIONS ESTHETIC AUTMENTS OPEN TRAY POST IMPLANT ANALOG Indexed components with internal conical connection. 0.7
More informationNEW. The SmartFix Concept Prosthetic solution on tilted implants. SmartFix
NEW The SmartFix Concept Prosthetic solution on tilted implants SmartFix The SmartFix Concept Prosthetic solution on tilted implants The SmartFix Concept is an effective and time-saving technique for immediate
More informationprosthetic technique manual
prosthetic technique manual table of contents table of contents introduction modules why choose BioHorizons prosthetics? surgical and prosthetic options & impression technique overview impression technique
More informationUS Implant System 2013 PROSTHETIC PROCEDURE
US Implant System 2013 PROSTHETIC PROCEDURE Contents TS Implant System Cement retained restoration 06 Cement retained bridges with the Cement abutment system 18 Cement retained crown with the Angled abutment
More informationRemovable Partial Dentures
Removable Partial Dentures Replacement of missing teeth Fixed partial denture Removable partial denture Complete removable Transitional denture Partial removable Implant retained prosthodontics No prosthetic
More informationProsthodontic Procedure
Prosthodontic Procedure TSH BNT Regeneration Implant Systems CAD-CAM Digital Solutions Services Important: before using Phibo The innovative and patented design of the Phibo implant systems incorporates
More informationMaking an implant level cast with the angled abutment and acrylic resin cap : A clinical report
Medical Communication Biosci. Biotech. Res. Comm. 10(2): 319-323 (2017) Making an implant level cast with the angled abutment and acrylic resin cap : A clinical report Hamid Neshandar Asli 1 and Bardia
More informationStep-by-Step Cemented Bridge Using Using CPK Abutments. Internal Hex. Implant System
4. 6. Step-by-Step Cemented Bridge Using Using CPK Abutments Internal Hex. Implant System MIS Corporation. All rights reserved. Published by MIS, which reserves the right to ameliorate the products described
More informationRESTORATION OF A FULLY EDENTULOUS PATIENT UTILIZING SIMPLE TECHNIQUES FOR IMPRESSION AND FABRICATION OF A HYBRID BRIDGE
Case Report International Journal of Dental and Health Sciences Volume 02,Issue 01 RESTORATION OF A FULLY EDENTULOUS PATIENT UTILIZING SIMPLE TECHNIQUES FOR IMPRESSION AND FABRICATION OF A HYBRID BRIDGE
More informationFabrication of a Hybrid Screw-Retained and Cement-Retained Implant Prosthesis: A Case Report
Journal Dental School 2012; 30(2):198-202 Case Report Fabrication of a Hybrid Screw-Retained and Cement-Retained Implant Prosthesis: A Case Report 1 Abolfazl Saboury * 2 Hamid Neshandar Asli 1 Associate
More informationMulti-Unit Abutment System SIC Safe on Four. Optimum use of available bone by angled placement of implants
Multi-Unit Abutment System SIC Safe on Four Optimum use of available bone by angled placement of implants Multi-Unit Abutment System SIC Safe on Four Safe on Four The SIC Safe on Four system is a further
More informationAnkylos. Product Catalog
Product Catalog General prosthetic advice for In order to obtain utmost benefit from the unique restorative features of the implant system, please observe the following guidelines: Avoid occlusal overload,
More informationMEDICAL UNIVERSITY OF VARNA FACULTY OF DENTAL MEDICINE DEPARTMENT OF PROSTHETIC DENTAL MEDICINE GOVERNMENT EXAMINATION SYLLABUS
MEDICAL UNIVERSITY OF VARNA FACULTY OF DENTAL MEDICINE DEPARTMENT OF PROSTHETIC DENTAL MEDICINE GOVERNMENT EXAMINATION SYLLABUS OF PROSTHETIC DENTAL MEDICINE ACADEMIC YEAR 2016/2017 1 1. Biomechanics of
More informationCOMPARISON OF THE COMPRESSIVE STRENGTH
RESEARCH COMPARISON OF THE COMPRESSIVE STRENGTH OF 3DIFFERENT IMPLANT DESIGN SYSTEMS Jose E. Pedroza, DMD, MS; Ysidora Torrealba, DDS, MSD; Augusto Elias, DMD, MSD; Walter Psoter, DDS, PhD The aims of
More informationObjectives: The purpose of this in vitro
www.scielo.br/jaos and cast implant single Rudys Rodolfo DE JESUS TAVAREZ 1, Wellington Cardoso BONACHELA 2, Anuar Antônio XIBLE 3 1- DDS, MSc, PhD, Professor, Maranhão University Center (UNICEUMA), São
More informationStudent Presentation # 3 Biomechanics
Student Presentation # 3 Biomechanics Anu Phogat Peter Ho 1 BIOMECHANICS OF DENTAL IMPLANTS 2 STRUCTURE OF TOOTH 3 INTERESTING FACTS According to the National Institutes of Health,by age 50, the average
More informationDisplacement of Implant Components from Impressions to Definitive Casts
Displacement of Implant Components from Impressions to Definitive Casts Sunjai Kim, DDS, PhD 1 /Jack I. Nicholls, PhD 2 /Chong-Hyun Han, DDS, PhD 3 /Keun-Woo Lee, DDS, PhD 4 Purpose: Four possible displacements
More informationTS Implant System 2013 PROSTHETIC PROCEDURE
TS Implant System 2013 PROSTHETIC PROCEDURE Contents TS Implant System Cement retained restoration Screw retained restoration 08 When abutment reduction is unnecessary 66 Screw retained restoration 21
More informationRestorative Driven Implant Solutions Utilizing the Latest Technology
Restorative Driven Implant Solutions Utilizing the Latest Technology Go online for in-depth content by Timothy F. Kosinski, DDS, MAGD As a general dentist who has placed nearly 7,000 dental implants, I
More informationPreci Clix Instructions for Ball Retained Overdenture Attachments
www.preat.com 800-232-7732 Preci Clix Instructions for Ball Retained Overdenture Attachments Dentist Direct Placement Technique #1293 / 1297 In one appointment, an existing removable partial denture may
More informationSmarter Thinking. Simpler Design. TiLobe Prosthetic Manual for PrimaConnex & Genesis implants
Smarter Thinking. Simpler Design. TiLobe Prosthetic Manual for PrimaConnex & Genesis implants Dear Customer, Thank you for allowing Keystone Dental Inc. the opportunity to provide solutions which we hope
More informationKAREN O. From Failure to Fantastic
KAREN O From Failure to Fantastic THE Pi TEAM THOMAS J. BALSHI, DDS, PhD, FACP GLENN J. WOLFINGER, DMD, FACP JAMES R. BOWERS, DDS IAN SMITH, DMD KAREN O Karen suffered from painful and swollen gum tissue
More informationSwishPlant. SwishPlant System Catalog. 4.8mmD mmD. 4.8mmD mmD. 3.7mmD mmD. 6.5mmD
SwishPlant System Catalog ------------ 4.1mmD ------------ 3.7mmD ------------ 3.3mmD SwishPlant 6.5mmD ------------ 6.5mmD ------------ 5.7mmD Table of Contents SwishPlant Fixture Mount Packaging Page
More informationPreci Clix Instructions for Ball Retained Overdenture Attachments
www.preat.com 800-232-7732 Preci Clix Instructions for Ball Retained Overdenture Attachments Dentist Direct Placement Technique In one appointment, an existing removable partial denture may be converted
More informationContinuing Education. Implant Prosthetics for the General Practice. Samuel M. Strong, DDS
Implant Prosthetics for the General Practice Samuel M. Strong, DDS We are pleased to offer you free CE in your Dentaltown Magazine! You can now read a CE article in the magazine and go online to www. to
More informationInfluence of the implant abutment types and the dynamic loading on initial screw loosening
http://jap.or.kr J Adv Prosthodont 2013;5:21-8 Despite the high clinical success of the implant, Laney et al. 3 reported several complications after 3-year follow up research of a single-implant prostheses.
More informationStraumann synocta Prosthetic System Basic Information. Technical Information
Straumann synocta Prosthetic System Basic Information Technical Information About this guide The Basic Information on the Straumann synocta Prosthetic System provide dental practitioners and related specialists
More informationBridging the Gap Between Original and Multi-Platform Systems.
Product Information Straumann CARES Screw-Retained Bars & Bridges Bridging the Gap Between Original and Multi-Platform Systems. CARES SRBB solutions are compatible with the following implant systems: Neodent
More informationSurveying. 3rd year / College of Dentistry/University of Baghdad ( ) Page 1
د. فائزة Lec.3 Prosthodontics Surveying The ideal requirements for successful removable partial denture are: 1. Be easily inserted and removed by the patient. 2. Resist dislodging forces. 3. It should
More informationABSTRACT. The Effect of Impression Technique, Connection Type and Implant Angulation on Impression Accuracy Joanna Kempler, DDS
ABSTRACT Title of Thesis: Thesis Directed by: The Effect of Impression Technique, Connection Type and Implant Angulation on Impression Accuracy Joanna Kempler, DDS Dr. Radi Masri Assistant Professor Department
More informationEvaluation of the Accuracy of Implant-Level Impression Techniques for Internal-Connection Implant Prostheses in Parallel and Divergent Models
Evaluation of the Accuracy of Implant-Level Impression Techniques for Internal-Connection Implant Prostheses in Parallel and Divergent Models Jung-Han Choi, DDS, MSD 1 /Young-Jun Lim, DDS, MSD, PhD 2 /Soon-Ho
More informationSuperLine & Implantium Surgical / Prosthesis Manual
SuperLine & Implantium Surgical / Prosthesis Manual SuperLine & Implantium SURGICAL MANUAL Surgical Drill Sequence I 04 Surgical Drill Sequence II 06 Drilling Depth Guide 08 Fixture Connection 10 Installation
More informationUser Manual Complete Prosthetic Kit (CPK)
User Manual Complete Prosthetic Kit (CPK) The Complete Prosthetic Kit is designed to enable easy impression and transfer technique and to prevent the need to remove and to re-tighten the abutments. Advantages
More informationOD Secure chairside pick-up using existing denture
OD Secure chairside pick-up using existing denture full-arch restorations OD Secure chairside pick-up using existing denture Use this technique for chairside pick-up of the OD Secure housing into an existing
More informationResearch Article Comparative Analysis of Screw Loosening in Titanium and Zirconia Implant-Supported Abutments-An In vitro Study
Cronicon OPEN ACCESS DENTAL SCIENCE Research Article Comparative Analysis of Screw Loosening in Titanium and Zirconia Implant-Supported Abutments-An In vitro Rafaella Monteiro Oliveira Cid 1,3 *, Gustavo
More informationTITANIUM BASES CAD/CAM FOR CROWN AND BRIDGE RESTORATIONS CAD/CAM PROSTHETICS ON CAMLOG AND CONELOG IMPLANTS
TITANIUM BASES CAD/CAM FOR CROWN AND BRIDGE RESTORATIONS CAD/CAM PROSTHETICS ON CAMLOG AND CONELOG IMPLANTS TABLE OF CONTENTS GENERAL SYSTEM INFORMATION 2 CAMLOG AND CONELOG TITANIUM BASES CAD/CAM PRODUCT
More informationProsthodontic Procedure
Prosthodontic Procedure Aurea Regeneration Implant Systems CAD-CAM Digital Solutions Services AUREA IMPLANT SYSTEM PROSTHODONTIC PROCEDURE IMPORTANT BEFORE USING AUREA IMPLANT SYSTEM The innovative and
More informationConus Concept: A Rewarding Complete Denture Treatment
Conus Concept: A Rewarding Complete Denture Treatment Complete dentures have largely become the domain of the denturist due to the dissatisfaction general dentists feel with this treatment. Multiple visits,
More informationImplant-based fixed rehabilitation of
CLINICAL Treatment of a Maxillary Dento-Alveolar Defect Using an Immediately Loaded Definitive Zygoma Implant-Retained Prosthesis With 11-Month Follow-Up: A Clinical Report Mirza Rustum Baig, MDS, MRD
More informationIndian Journal of Medical Research and Pharmaceutical Sciences August 2016;3(8) ISSN: ISSN: DOI: /zenodo Impact Factor: 3.
HYBRID VS CERAMOMETAL IMPLANT PROSTHESIS Dr.Rahul Patil* *Mds Prosthodontics, Senior Lecturer Sddc Dental College, Parbhani Keywords: Abstract A completely edentulous patient has few treatment options
More informationHeight* FC width Prep depth RC width 4.0mm+ 4.3mm N/A 6.3mm
STERN ERA IMPLANT ABUTMENT Summary Resilient precision overdenture attachment. Universal hinge with vertical movement. Titanium abutment, nylon male. Manufactured for most popular screw and cylinder implants.
More informationReal World Implant Prosthetics: Fixed and Removable Samuel M. Strong, DDS
Real World Implant Prosthetics: Fixed and Removable Samuel M. Strong, DDS Presurgical planning Health history-systemic conditions Case presentation Financial agreement Radiographs- PA s, FMX, Panoramic,
More informationBasic information on the. Straumann Pro Arch TL. Straumann Pro Arch TL
Basic information on the Straumann Pro Arch TL Straumann Pro Arch TL Contents 1. Introduction 2 1.1 Discover more treatment options with the 4 mm Short Implant 2 2. Technical information 3 3. Step-by-step
More informationNEW NEW. An Extended Solution for Narrow Ridges
NEW mm NEW An Extended Solution for Narrow Ridges NICE System: An Extended Solution for Narrow Ridges Quality Alpha-Bio Tec presents an advanced new NARROW implant solution for narrow alveolar ridges and
More informationCROWN AND BRIDGE RESTORATIONS WITH THE SYNOCTA PROSTHETIC SYSTEM
CROWN AND BRIDGE RESTORATIONS WITH THE SYNOCTA PROSTHETIC SYSTEM Straumann is the industrial partner of the ITI (International Team for Implantology) in the areas of research, development, and education.
More informationA Clinical Study of the Efficacy of Gold-Tite Square Abutment Screws in Cement-Retained Implant Restorations. Carl J.
A Clinical Study of the Efficacy of Gold-Tite Square Abutment Screws in Cement-Retained Implant Restorations Carl J. Drago, DDS, MS 1 Purpose: The purpose of this study was to record the effectiveness
More informationControlling Tissue Contours with a Prosthetically Driven Approach to Implant Dentistry
Controlling Tissue Contours with a Prosthetically Driven Approach to Implant Dentistry Go online for in-depth content by Timothy F. Kosinski, DDS, MAGD With continual improvements in the design and production
More informationIntended and Achieved Torque of Implant Abutment s Screw using Manual Wrenches /jp-journals
Intended and Achieved Torque of Implant Abutment s Screw using Manual Wrenches 10.5005/jp-journals-10024-1950 in Simulated Clinical Setting ORIGINAL RESEARCH Intended and Achieved Torque of Implant Abutment
More information