1 European Journal of Ophthalmology / Vol. 16 no. 0, 2006 / pp Glaucoma detection using the GDx nerve fiber analyzer and the retinal thickness analyzer E.M. HOFFMANN 1, 2, C. BOWD 2, N. KLEIN 1, K. STEMBERGER 1, F.H. GRUS 1, N. PFEIFFER 1 1 Universitäts-Augenklinik, Mainz - Germany 2 Hamilton Glaucoma Center, University of California, San Diego - USA PU R P O S E. To compare the ability of the nerve fiber analyzer (GDx) and the retinal thickness analyzer (RTA) to discriminate between glaucomatous and healthy eyes. ME T H O D S. Thirty-seven glaucoma patients (early to moderate severity) and 34 healthy controls were included. Glaucoma patients were defined as those with two repeatable abnormal visual fields by automated perimetry within 1 year. All subjects were examined with a GDx scanning laser polarimeter and RTA. Twelve GDx retinal nerve fiber layer parameters and 12 RTA optic disk topography parameters were obtained. GDx and RTA measure m e n t s w e re compared between both experimental groups using t-tests. Areas under the re c e i v e r operating characteristic curves (AUROC) for discriminating between healthy and glaucomatous eyes using GDx and RTA parameters were calculated and compared, and sensitivities at 80% and 95% specificity were re p o r t e d. RE S U LT S. Statistically significant differences between glaucoma and healthy eyes were found for most GDx and RTA parameters. For GDx, the parameter with the largest AUROC for discriminating between healthy and glaucomatous eyes was the number (AUROC = 0.91, sensitivity = 85% at specificity = 84%, sensitivity = 73% at specificity = 95%). For RTA, the parameter with the largest AUROC was mean cup depth (AUROC = 0.79, sensitivity = 61% at specificity = 82%, sensitivity = 33% at specificity = 95%). The AUROC for the GDx number was significantly larger than the AUROC for RTA mean cup depth (p<0.05). CO N C L U S I O N S. GDx showed better discrimination and better sensitivities at fixed specificities than RTA. The currently available RTA optic disk analysis software likely cannot replace GDx RNFL analysis software for successful glaucoma diagnosis. (Eur J Ophthalmol 2006; 16: 2 1) KE Y WO R D S. Glaucoma, Nerve fiber analyzer, Optic nerve imaging, Retinal thickness analyzer Accepted: November 15, 2005 INTRODUCTION The importance of early and objective diagnosis of glaucoma is evident. In the last decade, optical imaging devices have been developed to quantify structural characteristics of the optic nerve head, retinal nerve fiber layer (RNFL), and retina in an attempt to objectively identify glaucoma (1-8). The recently introduced retinal thickness analyzer (RTA) (6-8) was developed as a competitor to other alre a d y well-established imaging devices, such as the GDx nerve fiber analyzer (9-11), to aid in the diagnosis of glaucoma (12-14). To date, no published study has evaluated glaucoma detection using RTA optic disc measurements or compared RTA optic disc measurements and RNFL measurements using GDx for discriminating between healthy and glaucomatous eyes. The purpose of this study was to evaluate and determine the diagnostic accuracy of the RTA compared to the GDx in glaucomatous and healthy eyes. Wichtig Editore, /21-00$15.00/0

2 PLEASE ADD SHORT TITLE METHODS Subjects Seventy-one randomly chosen eyes from 71 subjects recruited consecutively from the Glaucoma Unit of the Department of Ophthalmology, University of Mainz, Germany, were included. Thirty-seven patients had open angle glaucoma (OAG, defined below) and 34 participants served as healthy cont rols. Subjects included in the study had best-corre c t e d visual acuity (BCVA) of 20/40 or better, a spherical refraction within ±5.0 D and cylinder ±3.0 D, clear media, and were at least 20 years of age. All patients underwent a complete ophthalmologic assessment including determination of BCVA, slit-lamp biom i c ro s c o p y, gonioscopy, stereophotography of the optic disk, automated visual field testing, intraocular pre s s u re (IOP) measurement with Goldmann applanation tonometry in a repeated 24-hour IOP measurement design, pachymetry (DGH 500 Pachette, DGH Technology Inc., Exton), and dilated fundus examination. The optic nerve head was assessed by a 90- or 78- diopter (D) lens. Visual field testing was performed using the Octopus 101, V 3.16e (Haag-Streit, Interzeag AG, Switzerland). Patients were only included if they had reliable visual fields with less than 25% false-positive errors, and less than 25% false-negative errors. A dynamic testing strategy was performed with white Goldmann size III stimuli presented on a apostilb background. Visual field indices were used to assess the severity of glaucoma according to Flammer et al (15) and Brusini using the Brusini Glaucoma Staging System (GSS) (16). This system is based on the indices mean deviation (MD) and corrected loss variance (CLV) and classifies visual field defects into five stages. Normal Octopus perimetry results according to the GSS are ±2 db MD and <4 db 2 for CLV (17). GSS Stage 1 represents early, subtle glaucomatous visual field loss; stages 2 and 3 represent moderate visual field loss; and stages 4 and 5 represent advanced visual field loss (18). No patients with advanced glaucoma were included in the current study. Inclusion and exclusion criteria for the experimental groups are summarized below. This study adhered to the Declaration of Helsinki, and informed written consent was obtained from all participants. Fig. 1 - A rea under the receiver operating characteristic curves for the best parameter of GDx nerve fiber analyzer (the number) and the retinal thickness analyzer (mean cup depth). Criteria for open angle glaucoma Patients with OAG had open anterior chamber angles and no goniodysgenesis confirmed by gonioscopy, in addition to glaucomatous optic neuropathy (GON) based on both slit-lamp funduscopy and evaluation of stere o p h o- tographs of the optic nerve head. GON was defined based on the presence of neuro retinal rim thinning, notching, excavation, nerve fiber layer defects, and/or localized pallor indicative of glaucoma. Patients with type beta peripapillary a t rophy were excluded. At least two consecutive abnormal visual fields also were re q u i red (Brusini Stage 1 3). Criteria for healthy subjects Healthy subjects had no family history of glaucoma, no history of diabetes or other systemic diseases, normal appearing disks on examination, and two normal visual field test results (see criteria above). Instrumentation Scanning laser polarimetry. RNFL thickness was measured by scanning laser polarimetry (GDx Nerve Fiber Analyzer, Laser Diagnostic Technologies, San Diego, CA) using a confocal scanning laser opht halmoscope (wavelength 780 nm) with an integrated polarimeter and fixed corneal polarization compensator. Briefly, the shift (i.e., retardation), caused by the birefringent nature of RN- FL microtubules, of polarized light reflected from the reti- 22

3 Hoffman et al na is assumed to be linearly related to RNFL thickness. The light beam is directed sequentially over each of 256 x 256 peripapillary locations to obtain a thickness map based on the corresponding retardation value at each location. The time to acquire these 65,536 pixel measurements is about 0.7 seconds. At least three images of high quality (passed by software-based image quality assessment and visual assessment by an experienced glaucoma specialist) were acquired using a field of view of 15x15 degrees, and a baseline retardation map was constructed for analysis by creating a mean image of these three measurements. High quality images required good focus and centering with even and appropriate illumination (i.e., not under- or overexposed). Images with 90 shifted RNFL retardation pattern (suggesting inappropriate corneal compensation) were excluded. The GDx Nerve Fiber Analyzer used in this study (software version 3.1) uses a fixed anterior segment compensator to neutralize the birefringence of the cornea and the lens (assumed slow axis of corneal polarization of 15 degrees nasally downwards with a magnitude of 60 nm). Neither pupil dilatation nor corneal contact is required. The GDx parameters investigated are listed in Table I. Retinal thickness analysis. Optic disc topography meas u red with the Retinal Thickness Analyzer (RTA; Talia Visionary Diagnostics, Neve-Ilan, Israel) is a digitized laser slit lamp that uses a helium-neon laser (wavelength 543 nm) as a light source. The RTA used in this study projected a 2 mm laser slit beam obliquely onto the retina in one brief scan. Each scan obtained 10 optical cross-sections, 100 µm TABLE I - DESCRIPTION OF GDX PARAMETERS Number:...Neural network number that indicates the likelihood that glaucoma is present (0 30 normal, borderline, glaucoma) Symmetry:...Ratio of the average of the 1500 thickest pixels in the superior region over the average of the 1500 thickest pixels in the inferior region Superior ratio:...ratio of the average of the 1500 thickest pixels in the superior region over the average of the 1500 median pixels in the temporal region Inferior ratio:...ratio of the average of the 1500 thickest pixels in the inferior region over the average of the 1500 median pixels in the temporal region Superior/nasal:...Ratio of the average of the 1500 thickest pixels in the superior region over the average of the 1500 median pixels in the nasal region Max. modulation:...the difference between the thickest and thinnest measurements within the image Ellipse modulation:...indication of the difference between the thickest parts of the nerve fiber layer and the thinnest parts. Uses the pixels covered by surrounding the optic nerve Average thickness (µm):..average thickness for all usable pixels outside the user-defined ellipse and divided by the number of pixels used Ellipse average (µm):...average thickness for the pixels along the measuring ellipse Superior average (µm):...average thickness for the pixels in the superior quadrant of the measuring ellipse Inferior average (µm):...average thickness for the pixels in the inferior quadrant of the measuring ellipse Superior integral (mm 2 ):..The total area under the curve (or total volume) of the nerve fiber layer beneath the superior portion of the ellipse surrounding the optic nerve TABLE II - DESCRIPTION OF RTA PARAMETERS Disk area (mm 2 ):...Total area of the disk delimited by the operator-defined contour line Cup area (mm 2 ):...Total cup area within the contour line Cup/disk area ratio:...the ratio between the cup area and the disk area, presented as coefficient Rim area (mm 2 ):...Total rim area above the reference plane and within the contour line Cup volume (mm 2 ):...The volume of the optic disk underneath the reference plane Rim volume (mm 2 ):...The volume of the rim area above the reference plane enclosed by the contour line Mean cup depth (mm):...the mean value of all calculated depths inside the cup area Maximum cup depth (mm):...the maximum calculated cup depth Cup shape measure:...the disk s slope distribution measurement (negative values represent a mild slope; positive values represent steep slopes) Height variation contour (mm):.the distance between the lowest and the highest point of the contour line Mean RNFL thickness (mm):...the mean height of the retinal surface along the contour line and above the reference plane RNFL cross section area (mm 2 ): The mean retinal nerve fiber layer thickness multiplied by the length of the contour line 23

4 ADD SHORT TITLE wide and 200 µm apart. The area of one scan is 2x2 and the measurement is obtained in 0.3 seconds. Interpolation is used to depict a continuous surface. Recently developed s o f t w a re enables the creation of a three-dimensional topographic map of the optic nerve head for glaucoma diagnosis (software version 3.0). We chose to investigate optic nerve topography using the RTA because this software was specifically designed for glaucoma diagnosis. To define optic disk topography, four overlapping images in the superio r, inferotemporal, inferonasal, and temporal areas were performed for each measurement and an operator drawn contour line was placed, outlining the optic disk marg i n. S t e reometric optic disk parameters and a classification analysis provided by the RTA, some of which rely on the contour line, are similar to those provided by the Heidelb e rg Retina Tomograph (HRT). RTA images were obtained by a glaucoma specialist. Twenty minutes before the meas u rements were taken, the pupil was dilated with 0.5% t ropicamide and 5% Neo-Synephrine eye drops. Prior to the topographic disk analysis, a contour line around the optic nerve head was placed using the inner edge of E l s c h n i g s scleral ring while viewing stereophotographs of the optic disk. All images were reviewed for quality by evaluating clarity and even illumination and optic nerve head centering. The software provides the analysis of 12 disk parameters, presented in color-coded, two- and three- dimensional topographic maps. The parameters and definitions are shown in Table II. Statistical analyses t-tests were used to evaluate the diff e rences in GDx and RTA measurements between experimental gro u p s. After Bonferroni correction for multiple comparisons, a p value of no more than was considered statistically significant. The areas under receiver operator characteristics curves (AUROC) were used to describe the ability of each parameter from both instruments to differentiate between glaucoma and healthy eyes. The AUROC shows the trade-off between sensitivity and 1 specificity. An area under the curve of 1.0 indicates perfect discrimination, whereas an area under the curve of 0.5 re p re s e n t s chance discrimination. The method of DeLong et al (19) was used to determine statistically significant differences among AUROCs. Statistical analyses were performed using commercially available software (JMP Version 5.11, SAS Institute, Cary, NC; and SPSS , Chicago, IL). RESULTS The age, sex, average IOP, BCVA, central corneal thickness (CCT), mean deviation (MD), and corrected loss variance (CLV) of the visual field test nearest to the imaging date are presented in Table III. There was a statistically significant diff e rence in age between glaucoma patients and healthy subjects (t-test, p<0.001). Sex, BCVA, and CCT were not significantly different between the two experimental groups (t-tests, p=0.61, p=0.08, and p=0.18, respectively). Average IOP (SD) of glaucoma eyes at the time of imaging was 20.1 (6.6) mm Hg. Average maximum IOP was 28.2 (8.7) mm Hg. Average IOP (SD) of the healthy eyes at the time of imaging was 13.8 (3.1) mm Hg. The stage of visual field defects for each patient was determined according to the Brusini classification system (16, 18). Among OAG eyes, 19 (51%) patients were classified as having early glaucomatous loss (Stage 1 of the GSS) and 18 (49%) subjects had moderate glaucomatous loss (Stage 2 and 3). Average visual field MD (SD) for OAG and healthy eyes were 3.04 (5.02) db and 0.19 (2.54) db, respectively (t-test, p<0.001). Average CLV (SD) for OAG and healthy eyes were (8.09) db 2 and 1.11 (0.87) db 2 (t-test, p<0.001). Statistically significant differences between experimental groups were found for all parameters except symmetry (t-test, p=0.17) and superior integral (t-test, p=0.02). Table IV presents the mean RNFL values of the GDx parameters for OAG and healthy controls. Statistically significant differences between groups were found for 5 of 12 parameters. Parameters that were not significantly diff e rent between groups (by t-test) were mean RNFL thickness (p=0.38), disk area (p=0.48), cup area (p=0.01), rim area (p=0.02), height variation contour (p=0.02), rim volume (p=0.39), and RNFL cross sectional area (p=0.90). Table V presents the mean disk topography values of the RTA parameters for OAG and healthy eyes. The AUROCs and sensitivities at fixed specificities for discriminating between glaucomatous and healthy eyes for GDx and RTA parameters are shown in Table VI. For GDx, ROC curve areas ranged from 0.58 (symmetry) to 0.91 (the number). The AUROC for the number was significantly greater than for all other GDx parameters (method of DeLong, p<0.05). For RTA, the AUROC ranged from 0.45 (RNFL cross sectional area) to 0.79 (mean cup depth). The AUROC for mean cup depth (0.79) was significantly g reater than for disk area (0.53), rim volume (0.61), and RNFL cross sectional area (0.45) (method of DeLong, 24

5 Hoffman et al p<0.05). The AUROC for the GDx number was significantly greater than the AUROC for RTA measured mean cup depth and all other RTA parameters (method of DeLong, p<0.05). The AUROC for GDx number was significantly larg e r (method of DeLong, p<0.05) than the AUROC from RTA mean cup depth. Figure 1 compares AUROCs for GDx number and RTA mean cup depth. DISCUSSION Based on AUROC curves analyses, our results suggest that GDx can more successfully discriminate between glaucomatous eyes and healthy eyes than RTA. The largest AUROCs for GDx and RTA were 0.91 and 0.79, res p e c t i v e l y. In addition, sensitivity at 80% and 95% s p e c i f i c i t y, for the parameter with the highest sensitivity for each instrument, was considerably higher for GDx than RTA (Tab. VI). Previously reported AUROCs for GDx range from 0.79 to 0.94 (20-22). In studies that included early to moderate glaucoma, sensitivities ranged from 33 to 81% at high fixed specificities of 90% (22, 23). These results are comparable to our results using GDx with fixed corneal compensation. Although a newer generation GDx currently is available, the GDx VCC (with variable corneal compensation) that incorporates a subject-specific variable corneal polarization axis and magnitude compens a t o r, the earlier version of the GDx currently is used in many clinics. In addition, much longitudinal data has been collected that likely will appear in future articles to assess the ability of scanning laser polarimetry in general to pre- TABLE III - CHARACTERISTICS OF THE STUDY POPULATION OAG (n=37) Normal (n=34) p Value* Age, y 63.5 (10.0) 52.2 (16.1) <0.001 F/M 24/13 24/ Average IOP, mm Hg 20.1 (6.6) 13.8 (3.1) < BCVA 20/25 20/ CCT, µm (33.2) (28.5) 0.18 MD, db 3.04 (5.02) 0.19 (2.54) < Data are given as mean (SD) * t - Test, unless otherwise indicated OAG = Open angle glaucoma; IOP = Intraocular pre s s u re; BCVA = Best-corrected visual acuity; CCT = Central corneal thickness; MD = Mean deviation TABLE IV - RETINAL NERVE FIBER LAYER (RNFL) MEASUREMENTS IN GLAUCOMA PATIENTS AND NORMAL CONTROLS USING GDX NERVE FIBER ANALYZER WITH FIXED CORNEAL COMPENSATION Parameter OAG (n=37) Normal (n=34) p Value* Number 62.5 (22.7) 24.6 (15.4) < Symmetry 0.09 (0.05) 0.08 (0.06) 0.17 Superior ratio 1.66 (0.29) 2.15 (0.42) < Inferior ratio 1.69 (0.33) 2.19 (0.40) < Superior/nasal ratio 1.55 (0.20) 1.88 (0.35) < Maximal modulation 0.82 (0.31) 1.30 (0.40) < Ellipse modulation 1.55 (0.60) 2.21 (0.77) Average thickness 55.6 (12.2) 65.3 (13.7) Ellipse average 56.7 (12.7) 68.2 (13.9) < Superior average 61.2 (13.8) 75.2 (15.0) < Inferior average 63.4 (14.7) 80.1 (16.7) < Superior integral 0.18 (0.05) 0.21 (0.04) 0.02 Data are given as mean (SD) * t - Test; after Bonferroni correction a p value of no more than was considered statistically significant. Because symmetry is calculated as the superior to inferior RNFL thickness ratio, the analysis was conducted on the absolute value of symmetry 1, to re p resent the distance from perfect symmetry monotonically OAG = Open angle glaucoma 25

6 ADD SHORT TITLE TABLE V - OPTIC DISK TOPOGRAPHY OF GLAUCOMA PATIENTS AND NORMAL CONTROLS USING THE RETINAL THICK- NESS ANALYZER (RTA) Parameter OAG (n=37) Normal (n=34) p Value* Mean RNFL thickness 0.14 (0.14) 0.17 (0.09) 0.38 Disk area 2.79 (0.69) 2.69 (0.41) 0.48 Cup area 1.57 (0.71) 1.12 (0.72) 0.01 Cup/disk area ratio 0.61 (0.05) 0.41 (0.04) <0.001 Rim area 1.10 (0.88) 1.57 (0.76) 0.02 Cup volume 0.49 (0.37) 0.24 (0.26) <0.001 Rim volume 0.23 (0.29) 0.29 (0.26) 0.39 Mean cup depth 0.27 (0.10) 0.17 (0.08) < Maximum cup depth 0.67 (0.18) 0.52 (0.17) <0.001 Cup shape measure (0.09) (0.09) <0.001 Height variation contour 0.53 (0.25) 0.39 (0.21) 0.02 RNFL cross section area 0.85 (0.98) 0.82 (0.66) 0.90 Data are given as mean (SD) * t - Test; after Bonferroni correction a p value of no more than was considered statistically significant OAG = Open angle glaucoma; RNFL = Retinal nerve fiber layer TABLE VI - AREAS UNDER THE RECEIVER OPERATING CHARACTERISTICS (ROC) CURVES AND SENSITIVITIES AT FIXED SPECIFICITIES GDx parameter ROC (SE) Sensitivity at 80% specificity Sensitivity at 95% specificity Number 0.91 (0.03) 85/84 73/95 Symmetry* 0.58 (0.07) 29/81 14/95 Superior ratio 0.83 (0.07) 68/84 53/95 Inferior ratio 0.84 (0.06) 68/81 53/95 Superior/nasal ratio 0.79 (0.06) 74/81 47/95 Maximal modulation 0.82 (0.05) 68/81 53/95 Ellipse modulation 0.74 (0.06) 61/81 41/95 Average thickness 0.72 (0.02) 41/81 18/95 Ellipse average 0.73 (0.02) 47/84 12/95 Superior average 0.76 (0.02) 50/81 21/95 Inferior average 0.79 (0.02) 53/84 27/95 Superior integral 0.68 (0.08) 47/81 5/95 GDx parameter ROC (SE) Sensitivity at 80% specificity Sensitivity at 95% specificity Mean RNFL thickness 0.62 (0.08) 29/81 2/95 Disk area 0.53 (0.05) 8/81 8/95 Cup area 0.68 (0.08) 38/81 14/96 Cup/disk area ratio 0.70 (0.09) 47/81 15/95 Rim area 0.68 (0.08) 38/81 5/96 Cup volume 0.74 (0.06) 70/81 12/97 Rim volume 0.61 (0.07) 24/81 5/96 Mean cup depth 0.79 (0.08) 61/82 33/95 Maximum cup depth 0.72 (0.08) 55/80 21/97 Cup shape measure 0.77 (0.07) 55/84 36/97 Height variation contour 0.70 (0.08) 55/83 27/95 RNFL cross section area 0.45 (0.09) 9/81 3/95 *Because symmetry is calculated as the superior to inferior retinal nerve fiber layer (RNFL) thickness ratio, the analysis was conducted on the absolute value of symmetry 1, to re p resent the distance from perfect symmetry monotonically 26

7 Hoffman et al dict glaucomatous conversion and pro g ression (24, 25). Our results, although not based on GDx VCC technology, provide encouraging results for the use of scanning laser polarimetry for the diagnosis of glaucoma in general. Add i t i o n a l l y, our results provide new information about the ability of RTA technology in the diagnosis of glaucoma and suggest that RTA optic disk topography measurement is not yet acceptable as a new alternative for scanning laser polarimetry RNFL measurements. We agree with previous reports that VCC technology can improve the ability of SLP to discriminate between healthy and glaucomatous eyes (5, 26, 27). However, several of the reported improvements using GDx VCC are slight (26, 28), and it has been shown that the ratio and modulation parameters obtained using GDx VCC are similar to those obtained using GDx with fixed corneal compensation (26 27) (although thickness parameters provide better glaucoma versus healthy discrimination using VCC). In addition, the current GDx VCC retains some limitations. One observed concern is the presence of atypical scan patterns that have been identified in some subjects that suggest areas of artificially high retardation (i.e., thick RNFL thickness) (5). The poor classification results associated with most RTA optic disk topography measurements in the curre n t study might be the result of several factors. First, the RTA uses slit lamp technology to obtain cross sections of the respective tissue by reflection, which may provide nonoptimal image quality, and thus inaccurate measurements in some eyes (29). In addition, non-optimal image quality could lead to difficulty in subjectively defining the contour line used to calculate most RTA topographic parameters, thus compromising measurement accuracy and therefore diagnostic precision. The Heidelberg Retina To m o g r a p h, which uses confocal scanning technology to obtain retinal height measurements used to depict optic disk topography and which also uses a subjectively defined reference plane, has been shown to have similar glaucoma detection ability as the GDx (20, 26, 30). Therefore, it is possible that RTA technology is not yet well developed enough for subjective re f e rence plane-based optic disk topography measurement. One further concern might be the required interpolation used by the RTA to produce a surface map of the optic disk. It is possible that this strategy could overlook defects or changes in disk topography. Only one study (12) reported AUROCs using RTA posterior pole measurements in the diagnosis of glaucoma. AUROCs for posterior pole indices ranged from 0.82 to The sensitivity at fixed specificity was highest to differentiate between normal and glaucoma by using the parameter perifoveal minimum thickness (sensitivity = 95% at specificity = 90%). One limitation of the current study is limited sample size that could decrease the generalization of the results and mask small but significant diff e rences between gro u p s observable in large population studies. In addition, healthy participants and patients with glaucoma diff e re d significantly in age. Although age might have aff e c t e d quality of images, each image from both instruments was reviewed for good quality by an experienced examiner. T h e re f o re, influence on the results of the study by poor quality images is unlikely. It is furthermore not likely that age affected discrimination for either instrument because we examined the association between age and each parameter from both instruments in the healthy eyes using linear re g ression. Only the association between age and GDx number was significant (r 2 = 0.13, p=0.04), indicating that age had little effect on the measurements from these instruments (in our population with a limited age range). Moreover, the purpose of this study was to compare the discriminative ability of GDx and RTA in the same population. Therefore, any difference in age likely affected measurements for each instrument similarly. Because age was significantly associated with GDx number, we recalculated the AUROC for this parameter while including age as a discriminant factor. The AUROC changed minimally from 0.91 to Overall, topographic RTA parameters were less able to discriminate between healthy and glaucomatous eyes than GDx RNFL parameters. Despite the use of GDx with fixed corneal compensation, discrimination was significantly better using GDx compared to RTA, suggesting that currently available RTA optic disk analysis softw a re likely cannot replace GDx RNFL analysis software for glaucoma diagnosis. The authors have no financial interest in any of the devices used in this article. No financial support was provided. The study was performed at the Department of Ophthalmology, Mainz, Germany. [QUERY: Please update address] Reprint requests to: Esther Hoffmann, MD University of California of San Diego Dept. of Ophthalmology Hamilton Glaucoma Center 9500 Gilman Drive La Jolla, CA ehoffmann@glaucoma.ucsd.edu ehoffman@mail.uni-mainz.de 27

8 ADD SHORT TITLE REFERENCES 1. Wilson MR, Khanna S. The value of diff e rent screening techniques for glaucoma. Curr Opin Ophthalmol 1994; 5: Weinreb RN. Laser scanning tomography to diagnose and monitor glaucoma. Curr Opin Ophthalmol 1993; 4: We i n reb RN, Shakiba S, Zangwill L. Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol 1995; 119: Weinreb RN, Zangwill LM. Imaging technologies for assessing neuroprotection in glaucomatous optic neuropathy. Eur J Ophthalmol 1999; 9 Suppl 1: S Medeiros FA, Zangwill LM, Bowd C, Weinreb RN. Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and Stratus OCT optical coherence tomograph for the detection of glaucoma. Arch Ophthalmol 2004; 122: Asrani S, Zeimer R, Goldberg MF, Zou S. Application of rapid scanning retinal thickness analysis in retinal diseases. Ophthalmology 1997; 104: Asrani S, Zou S, d Anna S, Vitale S, Zeimer R. Noninvasive mapping of the normal retinal thickness at the posterior pole. Ophthalmology 1999; 106: Zeimer R, Shahidi M, Mori M, Zou S, Asrani S. A new method for rapid mapping of the retinal thickness at the posterior pole. Invest Ophthalmol Vis Sci 1996; 37: Weinreb RN. Evaluating the retinal nerve fiber layer in glaucoma with scanning laser polarimetry. Arch Ophthalmol 1999; 117: Hoh ST, Greenfield DS, Liebmann JM, et al. Factors affecting image acquisition during scanning laser polarimetry. Ophthalmic Surg Lasers 1998; 29: Choplin NT, Lundy DC. The sensitivity and specificity of scanning laser polarimetry in the detection of glaucoma in a clinical setting. Ophthalmology 2001; 108: Tanito M, Itai N, Ohira A, Chihara E. Reduction of posterior pole retinal thickness in glaucoma detected using the Retinal Thickness Analyzer. Ophthalmology 2004; 111: Martinez de la Casa JM, Garcia Feijoo J, Castillo Gomez A, G a rcia Sanchez J. [Correlations between retinal thickness analyzer (RTA) and confocal scanning laser tomography (HRT) in optic disc analysis.] Arch Soc Esp Oftalmol 2004; 79: Itai N, Tanito M, Chihara E. Comparison of optic disc topography measured by Retinal Thickness Analyzer with m e a s u rement by Heidelberg Retina Tomograph II. Jpn J Ophthalmol 2003; 47: Flammer J, Drance SM, Augustiny L, Funkhouser A. Quantification of glaucomatous visual field defects with automated perimetry. Invest Ophthalmol Vis Sci 1985; 26: Brusini P. Clinical use of a new method for visual field damage classification in glaucoma. Eur J Ophthalmol 1996; 6: Zulauf M, LeBlanc RP, Flammer J. Normal visual fields m e as u red with Octopus-Program G1. II. Global visual field indices. Graefes Arch Clin Exp Ophthalmol 1994; 232: Kocak I, Zulauf M, Bergamin O. Evaluation of the Brusini glaucoma staging system for typing and staging of perimetric results. Ophthalmologica 1998; 212: DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 1988; 44: G reaney MJ, Hoffman DC, Garway-Heath DF, Nakla M, Coleman AL, Caprioli J. Comparison of optic nerve imaging methods to distinguish normal eyes from those with glaucoma. Invest Ophthalmol Vis Sci 2002; 43: Bowd C, Zangwill LM, Berry CC, et al. Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci 2001; 42: We i n reb RN, Zangwill L, Berry CC, Bathija R, Sample PA. Detection of glaucoma with scanning laser polarimetry. Arch Ophthalmol 1998; 116: Sinai MJ, Essock EA, Fechtner RD, Srinivasan N. Diff u s e and localized nerve fiber layer loss measured with a scanning laser polarimeter: sensitivity and specificity of detecting glaucoma. J Glaucoma 2000; 9: Mohammadi K, Bowd C, We i n reb RN, Medeiros FA, Sample PA, Zangwill LM. Retinal nerve fiber layer thickness measurements with scanning laser polarimetry predict glaucomatous visual field loss. Am J Ophthalmol 2004; 138: Hollo G, Szabo A, Vargha P. Scanning laser polarimetry versus frequency-doubling perimetry and conventional threshold perimetry: changes during a 12-month follow-up in p reperimetric glaucoma. A pilot study. Acta Ophthalmol Scand 2001; 79: We i n reb RN, Bowd C, Zangwill LM. Glaucoma detection using scanning laser polarimetry with variable corneal polarization compensation. Arch Ophthalmol 2003; 121: Choplin NT, Zhou Q, Knighton RW. Effect of individualized compensation for anterior segment birefringence on retinal nerve fiber layer assessments as determined by scanning laser polarimetry. Ophthalmology 2003; 110: G reenfield DS, Knighton RW, Feuer WJ, Schiffman JC, Zangwill L, Weinreb RN. Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry. Am J Ophthalmol 2002; 134: Polito A, Shah SM, Haller JA, et al. Comparison between retinal thickness analyzer and optical coherence tomography for assessment of foveal thickness in eyes with macular disease. Am J Ophthalmol 2002; 134: Zangwill LM, Bowd C, Berry CC, et al. Discriminating between normal and glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical C o h e rence Tomograph. Arch Ophthalmol 2001; 119:

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

Comparison of Optic Disc Topography Measured by Retinal Thickness Analyzer with Measurement by Heidelberg Retina Tomograph II

Comparison of Optic Disc Topography Measured by Retinal Thickness Analyzer with Measurement by Heidelberg Retina Tomograph II Comparison of Optic Disc Topography Measured by Retinal Analyzer with Measurement by Heidelberg Retina Tomograph II Noriko Itai*, Masaki Tanito*, and Etsuo Chihara* *Senshokai Eye Institute, Uji, Kyoto,

More information

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD CLINICAL SCIENCES Comparison of the GDx VCC Scanning Laser Polarimeter, HRT II Confocal Scanning Laser Ophthalmoscope, and Stratus OCT Optical Coherence Tomograph for the Detection of Glaucoma Felipe A.

More information

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module

More information

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:

More information

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Korean J Ophthalmol Vol. 18:1-8, 2004 Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Sun Young Lee, MD,

More information

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 61 Abstract Purpose The GDx is a scanning laser polarimeter that assesses

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Reza Zarei, MD 1 Mohammad Soleimani, MD 2 Sasan Moghimi, MD 3 Mohammad Yaser Kiarudi, MD 2 Mahmoud Jabbarvand, MD 1 Yadollah

More information

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Felipe A. Medeiros, Christopher Bowd, Linda M. Zangwill, Chirag Patel, and Robert N. Weinreb From the Hamilton

More information

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation Diagnostic Accuracy of Scanning Laser olarimetry with Enhanced versus Variable Corneal T. A. Mai, MD, N. J. Reus, MD, hd, H. G. Lemij, MD, hd urpose: To compare the diagnostic accuracy of scanning laser

More information

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan Correspondence to: Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute,

More information

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting (2007) 21, 1050 1055 & 2007 Nature Publishing Group All rights reserved 0950-222X/07 $30.00 www.nature.com/eye CLINICAL STUDY A comparison of HRT II and GDx imaging for glaucoma detection in a primary

More information

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous

More information

Reliability analyses of the GDx nerve-fiber analyzer

Reliability analyses of the GDx nerve-fiber analyzer VOL. 29 NO. 2 PHILIPPINE JOURNAL OF Ophthalmology APRIL ORIGINAL ARTICLE - JUNE 2004 1, 2, 3 Patricia M. Khu, MD, MS Edgardo U. Dorotheo, MD 1 Lawrence Tinio, MD 2 Cynthia P. Cordero, MS 4 1, 2, 3 Manuel

More information

Scanning laser polarimetry (SLP) provides real-time, objective

Scanning laser polarimetry (SLP) provides real-time, objective The Effect of Atypical Birefringence Patterns on Glaucoma Detection Using Scanning Laser Polarimetry with Variable Corneal Compensation Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda

More information

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of

More information

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry European Journal of Ophthalmology / Vol. 13 no. 4, 2003 / pp. 377-382 The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

More information

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Jimmy S. M. Lai*, Clement C. Y. Tham, Jonathan C. H. Chan*, Nelson K. F. Yip, Wilson W. T. Tang, Patrick S. H. Li*, Jane

More information

OtticaFisiopatologica

OtticaFisiopatologica Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza

More information

Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer

Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer European Journal of Ophthalmology / Vol. 15 no. 2, 2005 / pp. 246-254 Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer M. IESTER 1,2, A. MERMOUD 1 1 Hopital Ophtalmique

More information

Correlating Structure With Function in End-Stage Glaucoma

Correlating Structure With Function in End-Stage Glaucoma C L I N I C A L S C I E N C E Correlating Structure With Function in End-Stage Glaucoma Eytan Z. Blumenthal, MD; Amjad Horani, MD; Rajesh Sasikumar, MD; Chandrasekhar Garudadri, MD; Addepalli Udaykumar,

More information

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T et al Original article Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T1, Aggarwal A1, Bali SJ2, Sharma A1, Shah

More information

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma (2009) 23, 2182 2186 & 2009 Macmillan Publishers Limited All rights reserved 09-222X/09 $32.00 www.nature.com/eye CLINICAL STUDY Relationship between GDx VCC and Stratus OCT in juvenile glaucoma R Zareii,

More information

Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma

Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma Felipe A. Medeiros, Linda M. Zangwill, Christopher Bowd, Antje S. Bernd, and Robert N. Weinreb

More information

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro

More information

Clinical decision making based on data from GDx: One year observations

Clinical decision making based on data from GDx: One year observations Washington University School of Medicine Digital Commons@Becker Open Access Publications 2002 Clinical decision making based on data from GDx: One year observations James C. Bobrow Washington University

More information

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements 586 Glaucoma Service, Department of Ophthalmology, University of Campinas, Campinas, Brazil R Lauande-Pimentel R A Carvalho H C Oliveira D C Gonçalves L M Silva V P Costa Glaucoma Service, Department of

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

Scanning laser polarimetry (SLP) incorporates a confocal

Scanning laser polarimetry (SLP) incorporates a confocal Scanning Laser Polarimetry with Enhanced Corneal Compensation and Optical Coherence Tomography in Normal and Glaucomatous Eyes Mitra Sehi, 1 Stephen Ume, 1 David S. Greenfield, 1 and Advanced Imaging in

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December-2013 108 Name of Chief and Corresponding Author : Dr Chandrima Paul TITLE : Comparison of glaucoma diagnostic ability

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

Relationship Between Structure

Relationship Between Structure Original Article Relationship Between Structure and Function of the Optic Nerve Head-Glaucoma versus Normal Dr Savita Bhat, Dr Anna Elias, Dr Siddharth Pawar, Dr S.J. Saikumar, Dr Alpesh Rajput, superior,

More information

Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender

Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Bowd et al. Vol. 19, No. 1/January 2002/J. Opt. Soc. Am. A 197 Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Christopher Bowd,

More information

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: OPHTHALMOLOGIC TECHNIQUES PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical

More information

In some patients with glaucoma, standard (achromatic) automated

In some patients with glaucoma, standard (achromatic) automated Detecting Early Glaucoma by Assessment of Retinal Nerve Fiber Layer Thickness and Visual Function Christopher Bowd, 1 Linda M. Zangwill, 1 Charles C. Berry, 2 Eytan Z. Blumenthal, 1 Cristiana Vasile, 1

More information

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS Nerve fiber layer thickness in normals and glaucoma patients 403 NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS HIROTAKA SUZUMURA, KAYOKO HARASAWA, AKIKO KOBAYASHI and NARIYOSHI ENDO Department

More information

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Felipe A. Medeiros, Luciana M. Alencar, Linda M. Zangwill, Christopher

More information

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects 접수번호 : 2008-105 Korean Journal of Ophthalmology 2009;23:169-175 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.3.169 Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve

More information

MEDICAL POLICY. Proprietary Information of YourCare Healthcare

MEDICAL POLICY. Proprietary Information of YourCare Healthcare MEDICAL POLICY PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy criteria are not applied.

More information

The Role of the RNFL in the Diagnosis of Glaucoma

The Role of the RNFL in the Diagnosis of Glaucoma Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer

More information

CLINICAL SCIENCES. Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma

CLINICAL SCIENCES. Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma CLINICAL SCIENCES Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma The Confocal Scanning Laser Ophthalmoscopy Ancillary Study to the Ocular

More information

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Available online at   Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6): Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve

More information

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography 194 Review Article Evaluating the Optic Nerve and Retinal Nerve Fibre Layer: The Roles of Heidelberg Retina Tomography, Scanning Laser Polarimetry and Optical Coherence Tomography Sek-Tien Hoh, 1 MBBS,

More information

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS The Relationship between Standard Automated Perimetry and GDx VCC Measurements Nicolaas J. Reus and Hans G. Lemij PURPOSE. To investigate the relationship between retinal light sensitivity measured with

More information

Assessment of retinal ganglion cell loss is critical for the

Assessment of retinal ganglion cell loss is critical for the Scanning Laser Polarimetry with Variable Corneal Compensation: Identification and Correction for Corneal Birefringence in Eyes with Macular Disease Harmohina Bagga, David S. Greenfield, and Robert W. Knighton

More information

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Hiroshi Nakamura,* Toshine Maeda,* Yasuyuki Suzuki and Yoichi Inoue* *Eye Division of Olympia Medical Clinic, Tokyo, Japan; Department of

More information

Discriminating between Normal and Glaucoma-Damaged Eyes with the Heidelberg Retina Tomograph 3

Discriminating between Normal and Glaucoma-Damaged Eyes with the Heidelberg Retina Tomograph 3 Discriminating between Normal and Glaucoma-Damaged Eyes with the Heidelberg Retina Tomograph 3 Antonio Ferreras, MD, PhD, 1 Luís E. Pablo, MD, PhD, 1 José M. Larrosa, MD, PhD, 1 Vicente Polo, MD, PhD,

More information

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2

More information

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC Department of Ophthalmology, Medical University of Vienna, Austria Correspondence to Clemens Vass, Department of Ophthalmology and Optometry, Medical University of Vienna, General Hospital, Währinger Gürtel

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

Glaucoma: Diagnostic Modalities

Glaucoma: Diagnostic Modalities Glaucoma: Diagnostic Modalities - Dr. Barun Kumar Nayak, Dr. Sarika Ramugade Glaucoma is a leading cause of blindness in the world, especially in older people. Early detection and treatment by ophthalmologist

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013 268 Comparison of optic nerve morphology in eyes with glaucoma and eyes with non-arteritic anterior ischemic optic neuropathy by Fourier domain optical coherence tomography YUXIN YANG 1, HAITAO ZHANG 1,

More information

A vailability of the scanning laser polarimeter now permits

A vailability of the scanning laser polarimeter now permits 70 CLINICAL SCIENCE Specificity and sensitivity of glaucoma detection in the Japanese population using scanning laser polarimetry Shigeo Funaki, Motohiro Shirakashi, Kiyoshi Yaoeda, Haruki Abe, Shiho Kunimatsu,

More information

Intro to Glaucoma/2006

Intro to Glaucoma/2006 Intro to Glaucoma/2006 Managing Patients with Glaucoma is Exciting Interesting Challenging But can often be frustrating! Clinical Challenges To identify patients with risk factors for possible glaucoma.

More information

G laucoma is a progressive optic neuropathy in which

G laucoma is a progressive optic neuropathy in which 1135 EXTENDED REPORT The correlation between optic nerve head topographic measurements, peripapillary nerve fibre layer thickness, and visual field indices in glaucoma Y-W Lan, D B Henson, A J Kwartz...

More information

Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation

Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation Korean Journal of Ophthalmology 22(1):18-25, 2008 DOI : 10.3341/kjo.2008.22.1.18 Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation

More information

Optic nerve head morphometry in healthy adults using confocal laser scanning tomography

Optic nerve head morphometry in healthy adults using confocal laser scanning tomography 761 SCIENTIFIC REPORT Optic nerve head morphometry in healthy adults using confocal laser scanning tomography M M Hermann, I Theofylaktopoulos, N Bangard, C Jonescu-Cuypers, S Coburger, M Diestelhorst...

More information

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Acta Medica Anatolia Volume 2 Issue 1 2014 Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Remzi Mısır 1, Sinan Sarıcaoğlu

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

Assessment of the Retinal Nerve Fiber Layer in Clinical Trials of Glaucoma Neuroprotection

Assessment of the Retinal Nerve Fiber Layer in Clinical Trials of Glaucoma Neuroprotection SURVEY OF OPHTHALMOLOGY VOLUME 45 SUPPLEMENT 3 MAY 2001 Assessment of the Retinal Nerve Fiber Layer in Clinical Trials of Glaucoma Neuroprotection Eytan Z. Blumenthal, MD, and Robert N. Weinreb, MD Glaucoma

More information

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e The GDxVCC Early answers and ongoing assessment for glaucoma The quantifiable approach to quality care Only Humphrey GPA software Early insight

More information

Although measurements of the optic disc and retinal nerve

Although measurements of the optic disc and retinal nerve Longitudinal Variability of Optic Disc and Retinal Nerve Fiber Layer Measurements Christopher Kai-shun Leung, 1,2 Carol Yim-lui Cheung, 1 Dusheng Lin, 1,3 Chi Pui Pang, 1 Dennis S. C. Lam, 1 and Robert

More information

Structural examina.on: Imaging

Structural examina.on: Imaging ManaMa: Glaucoma Structural examina.on: Imaging Luís Abegão Pinto, MD, PhD Department of Ophthalmology CHLC Lisbon Faculty of Medicine, Lisbon University 1 11-10- 2013 Structural changes Qualitative changes

More information

Assessing the Relationship between Central Corneal Thickness and Retinal Nerve Fiber Layer Thickness in Healthy Subjects

Assessing the Relationship between Central Corneal Thickness and Retinal Nerve Fiber Layer Thickness in Healthy Subjects 1 1 1 1 1 1 Assessing the Relationship between Central Corneal Thickness and Retinal Nerve Fiber Layer Thickness in Healthy Subjects Authors: Tarkan Mumcuoglu, 1* Kelly A Townsend, 1* Gadi Wollstein, 1

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Glaucoma, Evaluation by Ophthalmologic Techniques File Name: Origination: Last CAP Review: Next CAP Review: Last Review: glaucoma_evaluation_by_ophthalmologic_techniques 3/2001

More information

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Nerve Fiber Layer

More information

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,

More information

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Case 1 Normal Tension Glaucoma 53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Table 1. Right Eye Left Eye Visual acuity 6/6 6/6 Ishihara

More information

A Comparative Analysis of Ganglion Cell Complex Parameters in Nigerian Negroes with Glaucoma and Macular Disease

A Comparative Analysis of Ganglion Cell Complex Parameters in Nigerian Negroes with Glaucoma and Macular Disease Nigerian Journal of Ophthalmology 2012; 20(1): 14-20 A Comparative Analysis of Ganglion Cell Complex Parameters in Nigerian Negroes with Glaucoma and Macular Disease O Oderinlo, AO Ogunro FWACS, AO Hassan

More information

Correspondence should be addressed to Verena Prokosch;

Correspondence should be addressed to Verena Prokosch; Hindawi Ophthalmology Volume 2017, Article ID 8014294, 6 pages https://doi.org/10.1155/2017/8014294 Research Article Comparison between the Correlations of Retinal Nerve Fiber Layer Thickness Measured

More information

OCT in the Diagnosis and Follow-up of Glaucoma

OCT in the Diagnosis and Follow-up of Glaucoma OCT in the Diagnosis and Follow-up of Glaucoma Karim A Raafat MD. Professor Of Ophthalmology Cairo University Hmmmm! Do I have Glaucoma or not?! 1 Visual Function 100% - N Gl Structure : - 5000 axon /

More information

The only measure of RNFL Integrity: GDxPRO

The only measure of RNFL Integrity: GDxPRO The only measure of RNFL Integrity: GDxPRO Look beyond thickness. Underlying structural organization is the key to RNFL At times, stepping onto ice is a risky proposition. Just knowing its thickness might

More information

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthaology Scanning Laser Polarimetry, but Not Optical Coherence Tomography Predicts Permanent Visual Field Loss in Acute Nonarteritic Anterior Ischemic

More information

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD Journal of Glaucoma 10:163 169 2001 Lippincott Williams & Wilkins, Inc. Influence of Scan Radius Correction for Ocular Magnification and Relationship Between Scan Radius With Retinal Nerve Fiber Layer

More information

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS Darshana Soma A research report submitted to the Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, in partial

More information

Peripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept.

Peripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept. Peripapillary Retinal Thickness Maps in the Evaluation of Glaucoma Patients: A Novel Concept The Harvard community has made this article openly available. Please share how this access benefits you. Your

More information

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients Yoshio Yamazaki, Kenji Mizuki, Fukuko Hayamizu and Chizuru Tanaka Department of Ophthalmology, Nihon

More information

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

STANDARD AUTOMATED PERIMETRY IS A GENERALLY Comparison of Long-term Variability for Standard and Short-wavelength Automated Perimetry in Stable Glaucoma Patients EYTAN Z. BLUMENTHAL, MD, PAMELA A. SAMPLE, PHD, LINDA ZANGWILL, PHD, ALEXANDER C. LEE,

More information

RETINAL NERVE FIBER LAYER

RETINAL NERVE FIBER LAYER CLINICAL SCIENCES The Effect of Scan Diameter on Retinal Nerve Fiber Layer Thickness Measurement Using Stratus Optic Coherence Tomography Giacomo Savini, MD; Piero Barboni, MD; Michele Carbonelli, MD;

More information

Optic disc damage has been shown to precede visual field

Optic disc damage has been shown to precede visual field Confocal Scanning Laser Ophthalmoscopy Classifiers and Stereophotograph Evaluation for rediction of Visual Field Abnormalities in Glaucoma-Suspect Eyes Christopher Bowd, 1 Linda M. Zangwill, 1 Felipe A.

More information

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography CLINICAL SCIENCES Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography Seong Bae Park, MD; Kyung Rim Sung, MD, PhD; Sung Yong Kang, MD; Kyung Ri Kim, BS;

More information

The only measure of RNFL Integrity : GDxPRO

The only measure of RNFL Integrity : GDxPRO The ideal glaucoma assessment tool for your practice The only measure of RNFL Integrity : GDxPRO Proven Clinical Performance Numerous studies have validated the diagnostic accuracy and performance of the

More information

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2015;29(6):418-423 http://dx.doi.org/10.3341/kjo.2015.29.6.418 Original Article Differences between Non-arteritic Anterior Ischemic Optic Neuropathy

More information

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence

More information

Relation of optic disc topography and age to thickness of retinal nerve fibre layer as measured using scanning laser polarimetry, in normal subjects

Relation of optic disc topography and age to thickness of retinal nerve fibre layer as measured using scanning laser polarimetry, in normal subjects Br J Ophthalmol 2000;84:473 478 473 Department of Ophthalmology, Celal Bayar University School of Medicine, Manisa, Turkey A B Toprak Department of Ophthalmology, Beyoglu Hospital, Taksim, Istanbul, Turkey

More information

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

More information

Because technological advances have provided the ability. Slope of the Peripapillary Nerve Fiber Layer Surface in Glaucoma

Because technological advances have provided the ability. Slope of the Peripapillary Nerve Fiber Layer Surface in Glaucoma Slope of the Peripapillary Nerve Fiber Layer Surface in Glaucoma Joseph Caprioli, Hyun Joon Park, Seyda Ugurlu, and Douglas Hoffman PURPOSE. TO develop structural markers of early glaucomatous optic nerve

More information

Early Detection Of Glaucoma Clinical Clues. Points To Live By. Glaucoma Risk Factors 10/3/2014

Early Detection Of Glaucoma Clinical Clues. Points To Live By. Glaucoma Risk Factors 10/3/2014 Early Detection Of Glaucoma Clinical Clues Eric E. Schmidt, O.D. Omni Eye Specialists Wilmington, NC schmidtyvision@msn.com Points To Live By 25% of G pxs NEVER have IOP >21mm 50% of G pxs have trough

More information

Introduction. Hemma Resch, Gabor Deak, Ivania Pereira and Clemens Vass. e225. Acta Ophthalmologica 2012

Introduction. Hemma Resch, Gabor Deak, Ivania Pereira and Clemens Vass. e225. Acta Ophthalmologica 2012 Comparison of optic disc parameters using spectral domain cirrus high-definition optical coherence tomography and confocal scanning laser ophthalmoscopy in normal eyes Hemma Resch, Gabor Deak, Ivania Pereira

More information

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(4):330-336 http://dx.doi.org/10.3341/kjo.2014.28.4.330 Original Article Patterns of Subsequent Progression of Localized Retinal Nerve Fiber

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2012 September ; 130(9): 1107 1116. doi:10.1001/archophthalmol.2012.827. A Combined Index of Structure and Function

More information

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach ORIGINAL STUDY Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach Joong Won Shin, MD, Kyung Rim Sung, MD, PhD, Jiyun

More information

Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph

Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph Yoshio Yamazaki,* Keiji Yoshikawa, Shiho Kunimatsu, Nobuyuki Koseki, Yasuyuki Suzuki, Shun Matsumoto and Makoto

More information

New Normative Database of Inner Macular Layer Thickness Measured by Spectralis OCT Used as Reference Standard for Glaucoma Detection

New Normative Database of Inner Macular Layer Thickness Measured by Spectralis OCT Used as Reference Standard for Glaucoma Detection Article https://doi.org/10.1167/tvst.7.1.20 New Normative Database of Inner Macular Layer Thickness Measured by Spectralis OCT Used as Reference Standard for Glaucoma Detection María Nieves-Moreno 1,2,

More information

Linking structure and function in glaucoma

Linking structure and function in glaucoma CET CONTINUING Sponsored by 1 CET POINT Linking structure and function in glaucoma 50 Dr Samantha McGinnigle PhD, BSc (Hons), MCOptom, AHEA This article will give an overview of the latest imaging technology

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,900 116,000 120M Open access books available International authors and editors Downloads Our

More information

A Review Of Risk Factors. Early Detection Of Glaucoma Clinical Clues. A risk factor analysis is critical. Points To Live By

A Review Of Risk Factors. Early Detection Of Glaucoma Clinical Clues. A risk factor analysis is critical. Points To Live By A Review Of Risk Factors Early Detection Of Glaucoma Clinical Clues Eric E. Schmidt, O.D. Omni Eye Specialists Wilmington, NC schmidtyvision@msn.com FINDACAR Family history IOP Nearsightedness Diabetes/Vascular

More information