Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma

Size: px
Start display at page:

Download "Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma"

Transcription

1 Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma Felipe A. Medeiros, Linda M. Zangwill, Christopher Bowd, Antje S. Bernd, and Robert N. Weinreb From the Hamilton Glaucoma Center and Diagnostic Imaging Laboratory, Department of Ophthalmology, University of California, San Diego, California. Supported in part by the Foundation for Eye Research (FAM), and by National Eye Institute Grant EY11008 (LMZ). Submitted for publication August 12, 2002; revised November 1, 2002; accepted January 30, Disclosure: F.A. Medeiros, None; L.M. Zangwill, None; C. Bowd, None; A.S. Bernd, None; R.N. Weinreb, None The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked advertisement in accordance with 18 U.S.C solely to indicate this fact. Corresponding author: Robert N. Weinreb, Hamilton Glaucoma Center, University of California, San Diego, 9500 Gilman Drive, La Jolla, CA ; PURPOSE. To apply Fourier analysis to the retinal nerve fiber layer (RNFL) thickness measurements obtained with scanning laser polarimetry (SLP), by using variable corneal compensation, and to evaluate the ability of this method to discriminate glaucomatous from normal eyes. METHODS. The study included one eye each of 55 patients with glaucoma and 52 healthy subjects. RNFL thickness measurements were obtained with a modified commercial scanning laser polarimeter (GDx Nerve Fiber Analyzer; Laser Diagnostic Technologies, Inc., San Diego, CA) so that corneal birefringence could be corrected on a subject-specific variable basis. The shape of the RNFL thickness double-hump pattern was analyzed by Fourier analysis of polarimetry data. Fourier coefficients and GDx parameters were compared between the two groups. A linear discriminant function was developed to identify and combine the most useful Fourier coefficients to separate the two groups. Receiver operating characteristic (ROC) curves were obtained for each measurement, and sensitivity values (at fixed specificities) were calculated. RESULTS. The Fourier-based linear discriminant function (LDF Fourier) resulted in a sensitivity of 84% for a specificity set at 92%. For similar specificity, the GDx software provided parameters had sensitivities ranging from 24% to 69%. The area under ROC curve for the LDF Fourier was 0.949, significantly larger than the ROC curve area for the single best GDx software provided parameter, superior average (0.870). CONCLUSIONS. The combination of Fourier RNFL thickness measures in an LDF, obtained using SLP with variable corneal compensation, improved the ability to discriminate glaucomatous from healthy eyes, compared with the GDx software provided parameters. (Invest Ophthalmol Vis Sci. 2003;44: ) DOI: /iovs Scanning laser polarimetry (SLP) quantitatively and objectively measures parapapillary retinal nerve fiber layer (RNFL) thickness. It is based on the principle that polarized light passing through a birefringent medium undergoes a measurable phase shift, known as retardation, that is directly proportional to the thickness of the medium. 1 4 The measure of RNFL thickness is based on the linear relationship between the retardation of reflected light and histologically measured RNFL tissue thickness. 1 Differences in retardation between normal and glaucomatous eyes, as well as good correlation between retardation measurements and visual field loss, have been described In normal subjects, the distribution of RNFL thickness around the optic disc has a characteristic double-hump pattern, due to a thicker RNFL in the superior and inferior regions, compared with temporal and nasal regions. 4,14 RNFL loss in glaucoma leads to a change in the appearance of this pattern, either by reducing its amplitude or changing its shape. However, as RNFL thickness is known to vary widely among healthy eyes, the usefulness of absolute thickness values to separate glaucomatous from normal eyes is limited Therefore, it is possible that an analysis of the global shape of the distribution of RNFL thickness around the optic disc may be more effective in detecting RNFL loss in glaucoma than an assessment of particular thicknesses. This analysis can be accomplished by using the mathematical procedure, Fourier analysis. 15,16 Using Fourier analysis, a complex waveform, such as the double-hump curve, can be resolved into a set of sinusoidal related components that yield the original pattern, when added point by point. Fourier analysis provides the amplitudes and phases of the sinusoidal components that can then be used to study the original waveform quantitatively. Essock et al. 16 applied Fourier analysis to the RNFL measurements of scanning laser polarimetry (SLP) and found an improved detection of glaucoma. However, a source of error in these studies was most likely introduced by the erroneous compensation of anterior segment birefringence The RNFL is not the only birefringent structure in the eye. The cornea and Henle fiber layer of the macula, and to a lesser extent the lens, also are birefringent. 21,22 To address anterior segment birefringence, the first commercial SLP had a fixed corneal compensator. The compensator was calibrated based on the assumption that all individuals had a slow axis of corneal birefringence 15 nasally downward with a magnitude of retardance of 60 nm. However, there is a wide variation, both in the axis and in the magnitude of corneal birefringence in normal and glaucomatous eyes An improvement of the SLP technology consisting of the variable compensation of anterior segment birefringence has been recently described. 26 Weinreb et al. 27 have shown that SLP with variable corneal compensation results in improvement of the sensitivity and specificity of several parameters to discriminate between healthy and glaucomatous eyes. In the present study, Fourier analysis was applied to the RNFL thickness measurements obtained with SLP by using variable corneal compensation, and the ability of this method to discriminate between glaucomatous and normal eyes was evaluated. Investigative Ophthalmology & Visual Science, June 2003, Vol. 44, No Copyright Association for Research in Vision and Ophthalmology

2 IOVS, June 2003, Vol. 44, No. 6 Measurement of Variable Corneal Compensation in Glaucoma 2607 SUBJECTS AND METHODS Fifty-two healthy subjects and 55 patients with glaucoma who met entry criteria were enrolled in this study. All patients were evaluated at the Hamilton Glaucoma Center (University of California, San Diego, CA), and informed consent was obtained from all participants. The University of California San Diego Human Subjects Committee approved all protocols and the methods described adhered to the tenets of the Declaration of Helsinki. Mean age SD of normal subjects and patients with glaucoma was years (range, 49 86) and years (range, 50 89), respectively (P 0.004; Student s t-test). Normal subjects were chosen so that the age range was similar to that of the patients with glaucoma. Among the healthy subjects, 33 (63%) were women and 19 (37%) were men. Among the patients with glaucoma, 29 (53%) were women and 26 (47%) were men. Ninety individuals were white, five were Hispanic, five were of African American descent, four were Asian American, and three were self-reported as unknown. One randomly chosen eye per subject was included in the study. Each subject underwent a comprehensive ophthalmic examination, including review of medical history, best corrected visual acuity, slit-lamp biomicroscopy, intraocular pressure (IOP) measurement with Goldmann applanation tonometry, gonioscopy, dilated fundoscopic examination with a 78-D lens, stereoscopic optic disc photography, and automated perimetry with 24-2 full-threshold standard automated perimetry (SAP) or Swedish interactive threshold algorithm (SITA; Humphrey Field Analyzer; Zeiss-Humphrey Systems, Dublin, CA). To be included, each subject had to have best corrected visual acuity of 20/40 or better. Refractive error ranged from 9.00 to 4.75 D. Eyes with coexisting retinal disease, uveitis, or nonglaucomatous optic neuropathy were excluded from the investigation. Normal control eyes had an IOP of 22 mm Hg or less, with no history of increased IOP, and a normal visual field result. Normal visual field was defined as a mean deviation (MD) and pattern standard deviation (PSD) within 95% confidence limits, and a Glaucoma Hemifield Test result (GHT; Zeiss-Humphrey Systems) within normal limits. Normal control eyes also had a healthy appearance of the optic disc and RNFL (no diffuse or focal rim thinning, cupping, optic disc hemorrhage, or RNFL defects), as evaluated by optic disc photographs. Eyes were classified as glaucomatous if they had repeatable (two consecutive) abnormal visual field test results, defined as a PSD outside the 95% normal confidence limits, or a GHT result outside 99% normal confidence limits, regardless of the appearance of the optic disc. Because we were evaluating SLP-measured RNFL thickness, we chose the best indicator of glaucoma that is not dependent of RNFL or optic disc appearance, to avoid bias in the evaluation of the technique. 28 Average MD of the glaucomatous eyes on the visual field test nearest the imaging date was 5.91 db (range, to 0.26 db). According to the Hodapp-Parrish-Anderson 29 grading scale of severity of visual field defects, 21 (38%) patients were classified as having early visual field defects, 30 (55%) as having moderate defects, and 4 (7%) as having severe defects. SLP Measurements In this study, all patients were imaged using a prototype version of the scanning laser polarimeter with variable corneal polarization compensation. This prototype consisted of a modification of the previously commercially available scanning laser polarimeter (GDx Nerve Fiber Analyzer, Laser Diagnostic Technologies, Inc., San Diego, CA) so that the original fixed corneal compensator was replaced with a variable custom corneal compensator (VCC) as described by Zhou and Weinreb. 26 This method of variable corneal polarization compensation has been incorporated into the new commercially available version of the scanning laser polarimeter (GDx VCC, software version 5.0.1; Laser Diagnostic Technologies, Inc.) Briefly, the VCC scanning laser polarimeter consists of a set of four linear retarders in the path of the measurement laser beam. The first two adjustable retarders are optical lenses that have equal retardance and form a variable corneal and lens compensator. The third retarder is composed of the cornea and lens, and the fourth retarder is the retinal birefringent structure (RNFL or macular Henle fibers). To determine the eye-specific corneal polarization axis (CPA) and magnitude (CPM), the compensating retarders were adjusted to 0 nm, and the macula was imaged. The retardation at a locus of points along a circle centered on the fovea was calculated. This measurement circle was 80 pixels in diameter, which corresponds to 1.41 mm in an emmetropic eye. The resultant retardation profile represented the additive effects of cornea, lens, and macular Henle fiber birefringence. The axis and magnitude of anterior segment birefringence were determined from the measured macular retardation profile. The macula was imaged three times in each subject, and the average CPMs and CPAs from the three macular scans that resulted in adequate compensation were recorded. The compensating retarders were then adjusted to minimize the effects of anterior segment birefringence. Once the VCC was correctly adjusted, the total retardation measured with the scanning laser polarimeter was the birefringence of the RNFL or that of the Henle fiber layer. As a result, an SLP image of the macula should show a radially symmetric uniform pattern that is, a flat profile. To confirm the effectiveness of the VCC, the macula was imaged again with a subject-specific CPM and CPA, to assure that the macular retardation profile was flat when these values were used. Next, three corneal birefringence-compensated parapapillary SLP images from each eye were obtained with the appropriate eye-specific VCC CPM and CPA. The three images were automatically combined by GDx software version 3.0 to create a composite mean image that was used for analysis of RNFL thickness. The disc margin was established by an experienced operator who used an ellipse to outline the inner margin of the peripapillary scleral ring, based on the information from the SLP reflectance image. A 10-pixel-wide elliptical band was automatically positioned concentric with the disc margin outline and at 1.7 disc diameters from the center of the optic disc. This elliptical band was divided into 32 sectors (11.25 each), and the average retardation was recorded in each of the 32 sectors. These measurements were used in the Fourier analysis, to be described later. The GDx software estimates RNFL thickness (in micrometers) by correcting retardation measurements (in degrees) with a factor determined in a previous histologic comparison in monkey eyes. 1 The GDx software also calculates summary parameters based on quadrants, which are defined as temporal ( unit circle), superior ( ), nasal ( ), and inferior ( ). The GDx software provided parameters investigated in this study were symmetry (superior quadrant thickness/inferior quadrant thickness), superior ratio (superior quadrant thickness/temporal quadrant thickness), inferior ratio, superior/nasal ratio, maximum modulation (thickest quadrant/thinnest quadrant)/thinnest quadrant, average thickness, superior maximum (average of the 1500 thickest points in the superior quadrant), inferior maximum, ellipse average, ellipse modulation, superior average, inferior average, and superior integral. The latter five parameters are measured in relation to the 10-pixel-wide elliptical band. Each of these parameters have been described in detail elsewhere. 11 Fourier Analysis Fourier analysis is a mathematical procedure in which a complex waveform pattern (such as the double-hump pattern of the RNFL) is resolved into a series of harmonically related sinusoids of specific frequencies, amplitudes, and phases. 30 These sinusoids, when summed point by point, reproduce the original pattern. The fast Fourier transform (FFT) was used to determine the coefficients (i.e., the amplitude and phase of each sinusoid). Mathematical software (Mathematica ver. 4; Wolfram Research, Inc., Champaign, IL) was used to perform the FFT algorithm on the data from SLP. Details from the FFT procedure have been published elsewhere In brief, when Fourier analysis is performed on a set of discrete data that describe a waveform, the

3 2608 Medeiros et al. IOVS, June 2003, Vol. 44, No. 6 number of sine waves generated equals half the number of data points. The lowest frequency component, also called the fundamental (F1), corresponds to a sine wave pattern with one cycle. The other components are sine waves with frequencies that are integer multiples of the fundamental. For example, the second harmonic has a frequency that is two times that of the fundamental. The pattern of thickness measurements from SLP in the 32 sectors around the optic disc was analyzed to obtain the Fourier coefficients. Because the Fourier analysis was performed on 32 data points, 16 sine waves were obtained. The amplitude of each sine wave was the peak-to-trough distance, and the phase value described where the waveform begins in relation to some reference. The amplitude of each component indicated its relative contribution to the shape of the composite curve fitting the RNFL thickness measurements. Also produced by Fourier analysis was the component DC, which represented the overall level of all amplitude measures and corresponded to the mean RNFL thickness. The use of more than 32 sectors (data points) resulted in higher-frequency components with very low amplitudes that did not contribute significantly to the discrimination of glaucomatous from normal eyes, even when combined with other lower-frequency components. Statistical Analysis Student s t-tests were used to evaluate differences in GDx software provided parameters and Fourier coefficients between glaucomatous and healthy eyes. The Hotelling T-squared generalized means test was used to compare the set of means of Fourier measures between the two groups. This test controls the inflation of type I error that arises when making a series of t-tests to compare group means of several dependent variables. Results of Bonferroni adjustments based on the number of comparisons were also provided. The Fourier components (DC component and all amplitude and phase values) from all subjects were entered into a stepwise discriminant analysis to develop a classification function (linear discriminant function; LDF) designed to identify and combine the best measures to differentiate glaucomatous from normal eyes. The relative importance of each variable in the discriminant analysis was assessed by using the ratio of within-group variance to total-group variance (Wilks ). The absolute phase coefficients were used for this procedure to emphasize deviations in shape, regardless of direction. Receiver operating characteristic (ROC) curves were used to describe the ability to differentiate glaucomatous from normal eyes of each GDx software provided parameter and also of the LDF obtained from Fourier measures. The ROC curve shows the tradeoff between sensitivity and 1 specificity. An area under the ROC curve of 1.0 represents perfect discrimination whereas an area of 0.5 represents chance discrimination. The method of DeLong et al. 34 was used to compare areas under the ROC curve. Minimum specificity cutoffs of 80% and 90% were used for comparing the sensitivity of the GDx parameters and the Fourier-based LDF by the McNemar s test for paired proportions. The use of the same population to identify and test a discriminant function can lead to a significant bias or optimism with respect to the predictive ability of the model when applied to an independent data set. To provide an internal validation estimate, we used a computerintensive bootstrap approach. 35,36 The bootstrap procedure provides stable and nearly unbiased estimates of performance, and is said to be the most efficient validation method available, comparing favorably with standard cross-validation techniques The superiority of bootstrap estimates over traditional methods seems to be particularly evident for error rate estimation in two-class discrimination problems when samples sizes are relatively small. 39 Bootstrapping is a resampling method that allows one to make inferences about the population from which the sample originated by drawing B samples (B 1000 in the present study), with replacement from the original data set, of the same size as the original data set. In the bootstrap method, a model is built for each bootstrap sample and evaluated only in those FIGURE 1. Mean RNFL thickness distribution around the optic disc in healthy and glaucomatous eyes. subjects not sampled. The prediction errors are then averaged over all bootstrap samples (test performance). Because the evaluation is based on an independent data set, this method can be seen as a direct extension or as a smoothed version of cross-validation. Because the data set is sampled with replacement, on average 63.2% of the subjects are included at least once in a bootstrap sample, giving the method its name. 35 The estimated performance is a weighted combination of the apparent performance (resubstitution error estimate on the full data set) and test performance. The bootstrap is an extension of the method applying a regularizing coefficient based on the amount of overfitting and has the advantage of performing well even when there is severe overfitting. 36 In our study, the area under the ROC curve was used as a measure of predictive performance of the LDF, and the bootstrap method was applied to estimate the optimism or bias of this measure. The application of the method in a similar situation has been reported elsewhere. 37 Bootstrap methods were also used to calculate bootstrap bias-corrected and accelerated (BCa) 95% confidence intervals for the area under ROC curves. 40,41 Statistical analyses were performed on computer (S-Plus 2000 software; MathSoft, Inc., Seattle, WA). RESULTS The mean RNFL thickness estimates for the 32 sectors around the optic disc in normal and glaucomatous eyes are represented in Figure 1. The general double-hump pattern is evident in both groups, but the glaucomatous mean RNFL profile is clearly blunted in relation to the mean profile of normal subjects. Fourier analysis of the thickness measurements in the 32 sectors for each patient resulted in 16 components. The mean amplitudes and phases of the Fourier components in glaucomatous and normal subjects are shown in Table 1. The Hotelling T-squared test, when applied to the set of Fourier measures, showed a statistically significant difference between the two groups (P 0.001). Several Fourier measures showed statistically significant differences between glaucomatous and healthy eyes at P 0.05 (Table 1). After Bonferroni correction, significant differences were observed for the amplitudes of the F2, F5, and F14 components and also for the DC component and phase of the second component. The stepwise discriminant analysis applied to the Fourier measures resulted in the following LDF (LDF Fourier): LDF (0.170 F2 ampli-

4 IOVS, June 2003, Vol. 44, No. 6 Measurement of Variable Corneal Compensation in Glaucoma 2609 TABLE 1. Mean Amplitudes and Phases of the Fourier Components in Healthy and Glaucomatous Eyes Fourier Component Amplitude Phase Healthy Glaucomatous P Healthy Glaucomatous P DC* F F2* F F F5* F F F F F F F F F14* F F Data are expressed as the mean SD of results in 52 healthy and 55 glaucomatous eyes. * Statistically significant difference between amplitudes after Bonferroni correction ( 0.001). Statistically significant differences between phases after Bonferroni correction ( 0.001). tude) (0.744 F2 phase) (0.481 F5 amplitude) (1.280 F14 amplitude) (0.057 DC). This discriminant function had an area under the ROC curve of (bias-corrected 95% confidence interval: ). The bootstrap estimate of bias under the ROC curve was With a cutoff of as normal, the LDF correctly labeled 48 of 52 healthy eyes (specificity 92%) and 46 of the 55 glaucomatous eyes (sensitivity 84%). The most important variable in the LDF (in terms of discriminative power) was the amplitude of the second Fourier component. Figure 2 shows the graph resulting from inverse Fourier transform using only the mean values from the F2 component and the overall mean RNFL thickness (DC component) of glaucomatous and normal eyes. The F2 component represents a sine wave with a frequency that is two times that of the fundamental (sine wave with one cycle). Thus, the F2 component has two cycles, or humps. The curve profile resulting from the F2 component roughly corresponds to the doublehump pattern of RNFL thickness shown in Figure 1. Table 2 shows the measures for the GDx software provided parameters in normal and glaucomatous eyes. A Hotelling T- squared test of the set of GDx parameters showed a statistically significant difference between the two groups (P 0.001). After Bonferroni correction, significant differences were observed for all evaluated parameters except symmetry. Table 3 shows ROC curve areas for all parameters and also for the LDF Fourier. ROC curve areas for the GDx parameters ranged from to The GDx software provided parameter with largest area under the ROC curve was the superior average. The ROC curve area for the LDF Fourier was significantly larger than the area for the superior average (P 0.01; Fig. 3). Our study had a power ranging from 100% to 78% to detect statistically significant differences (at 0.05) between an area under the ROC curve of 0.95 and areas under the ROC curve ranging from 0.61 to Table 3 also shows the sensitivities at fixed specificities ( 80% and 90%) for the GDx parameters and LDF Fourier. The LDF Fourier had significantly higher sensitivities (P 0.05, McNemar test) than did the single best GDx software provided parameter (superior average) at both specificities. TABLE 2. GDx Software Provided Parameters in Healthy and Glaucomatous Eyes Parameter Healthy Glaucoma P* Symmetry Superior ratio Inferior ratio Superior/nasal ratio Maximum modulation Ellipse modulation Average thickness ( m) Ellipse average ( m) Superior average ( m) Inferior average ( m) Superior integral (mm 2 ) Superior maximum ( m) Inferior maximum ( m) FIGURE 2. Inverse Fourier transform obtained from mean values of the second and DC Fourier components in healthy and glaucomatous eyes. Data are expressed as the mean SD of results in 52 healthy and 55 glaucomatous eyes. * Student s t-test.

5 2610 Medeiros et al. IOVS, June 2003, Vol. 44, No. 6 TABLE 3. Areas under the ROC Curves for Discriminating between Healthy and Glaucomatous Eyes Area under ROC Curve (95% CI)* Sensitivity/Specificity (Specificity >90%) Sensitivity/Specificity (Specificity >80%) LDF Fourier ( ) 84/92 93/81 Superior average ( ) 69/90 82/81 Ellipse average ( ) 69/90 80/81 Inferior average ( ) 66/90 78/81 Superior maximum ( ) 55/90 72/81 Superior integral ( ) 66/90 67/81 Ellipse modulation ( ) 49/90 69/81 Superior ratio ( ) 35/90 62/81 Average thickness ( ) 53/90 64/88 Superior/nasal ( ) 49/90 60/81 Inferior maximum ( ) 53/90 64/81 Maximum modulation ( ) 36/90 56/81 Inferior ratio ( ) 40/90 49/81 Symmetry ( ) 24/90 40/81 Sensitivities for diagnosing glaucoma at fixed specificities. * 95% Confidence interval calculated using bootstrap bias-corrected and accelerated method. DISCUSSION The present study shows that the Fourier analysis of the RNFL thickness measurements obtained with SLP using variable corneal compensation improves the discrimination of glaucomatous from normal eyes, compared with the parameters provided by the GDx software. The area under the ROC curve for the LDF combining Fourier measures (LDF Fourier) was greater than the ROC area of any of the GDx software provided parameters investigated. In this study, we also calculated sensitivities at high ( 90%) and moderate ( 80%) specificities. Depending on the patient population and diagnostic criteria, each of these conditions may be more desirable. At a fixed specificity of at least 90%, the LDF Fourier had a sensitivity of 84% to detect glaucoma, whereas the GDx parameters had sensitivities ranging from 24% to 69%. At a fixed specificity of at least 80%, the sensitivities of GDx parameters ranged from 40% to 82%, whereas the LDF Fourier had sensitivity of 93%. Previous reports have also shown that GDx software provided parameters have a limited ability to discriminate between normal and glaucomatous eyes The lower diagnostic ability of the GDx parameters may be related to the wide variability of absolute RNFL thickness measurements in healthy subjects. FIGURE 3. ROC curves for the LDF obtained from Fourier measures and for the single parameter with the largest ROC curve area (superior average). The RNFL thickness has been found to vary widely in the normal population. 4,14 This may limit the identification of glaucomatous eyes with loss of nerve fibers, but with absolute RNFL thickness still within normal limits. Rather than emphasizing absolute thickness itself, the Fourier analysis of RNFL thickness measurements provides a global measure that takes into account the whole shape of thickness distribution around the optic disc, emphasizing relative differences between local areas. In addition, by comparing relative differences in the shape of RNFL thickness distribution curve, the Fourier method may better identify patients with glaucoma with localized RNFL defects compared with the GDx parameters, which are usually calculated based on RNFL thicknesses averaged over a large region. Although different methods of analyzing thickness measurements from SLP may result in an improved detection of eyes with glaucomatous RNFL loss compared with the GDx parameters, a persistent source of error may be related to the erroneous compensation for anterior segment birefringence The erroneous compensation for anterior segment birefringence produces a wider range of retardation measurements in normal eyes, which may complicate the identification of abnormalities. The effects of the axis and magnitude of corneal birefringence on RNFL retardation measurements have been described, 18,23 25 and algorithms designed to correct for this have been reported. 19,20,26 Garway-Heath et al. 20 proposed a correction of RNFL retardation measurements obtained using the GDx with fixed corneal compensation using perifoveal or peripapillary temporal retardation values. This method resulted in a narrower normal range of retardation measurements and improvement in the discrimination between normal and glaucomatous eyes. In another approach, Greenfield et al. 19 showed that the incorporation of CPMs improved the discriminatory ability of some GDx parameters. We used SLP data obtained using variable corneal compensation according to the method described by Zhou and Weinreb. 26 This method of anterior segment polarization compensation has been incorporated into the new commercially available scanning laser polarimeter (GDx VCC; Laser Diagnostic Technologies, Inc.) and is based on the determination of the magnitude and axis of anterior segment birefringence by polarimetry imaging of the Henle fiber layer. Individualized anterior segment compensation can be achieved with this method so that the measured retardation largely reflects the RNFL retardance. In a recent work, Weinreb et al. 27 showed that the diagnostic ability of several GDx parameters to classify eyes as glaucomatous or

6 IOVS, June 2003, Vol. 44, No. 6 Measurement of Variable Corneal Compensation in Glaucoma 2611 normal is improved considerably with SLP, using variable corneal compensation compared with SLP using fixed corneal compensation. This improvement was stronger for thickness parameters than for ratio modulation parameters, probably because the latter may already compensate for some of the changes in retardation measurements caused by an inadequate corneal compensation in some patients. In our study, the areas under the ROC curve for the thickness parameters were generally greater than the ROC areas for the ratio/modulation parameters. The three summary parameters superior average, ellipse average and inferior average performed comparably. However, the area under the ROC curve for the best GDx parameter, superior average, was still significantly inferior to the area under the ROC curve for the LDF Fourier. The GDx parameters evaluated in our study were the ones shown in the standard GDx printout and previously identified as having the best performance to discriminate glaucomatous from normal eyes. Thus, it is not surprising that the means of all GDx parameters (except symmetry) showed significant differences between glaucomatous and normal eyes. In contrast, among all the Fourier components, just a few seem to be important for discriminating glaucomatous from normal eyes. This is also not a surprising result, because the RNFL thickness profile has certain characteristics (like its double-hump shape) which are better described by some Fourier components than by others. When the Fourier components are combined, as in the LDF developed in our study, they become a powerful tool for analysis of polarimetry data, showing higher sensitivity and specificity than GDx parameters. In the present study, we did not examine the diagnostic ability of the GDx parameter, the number, because this parameter has been developed as a best means to interpret data obtained from SLP, using fixed corneal compensation. This parameter is calculated with a neural network approach and has been reported to be the best of the GDx software provided parameters in the discrimination of glaucomatous from normal eyes. 9 11,42 44 It is possible that a new neural network based parameter using SLP-VCC data will have better diagnostic precision than the GDx parameters reported in our study, and the comparison of this method to the Fourier analysis of RNFL measurements should be investigated. The application of Fourier analysis to the RNFL polarimetry data was initially reported by Essock et al. 16 Using SLP with fixed corneal compensation, they found a sensitivity and specificity of 96% and 90%, respectively, in the differentiation of glaucomatous from normal eyes. In our study, for a similar level of specificity, we found a lower level of sensitivity (84%). Different methods to evaluate the Fourier measurements and different population characteristics may be related to the different results. In the study by Essock et al., the average visual field MD of patients with glaucoma was 8.9 db, considerably higher than the average MD of the patients included in our study ( 5.9dB). Therefore, one of the reasons for the discrepancies in the results of the two studies may be related to different severity of glaucoma in the patient population. In another study using RNFL data obtained from SLP with fixed corneal compensation, Sinai et al. (Sinai MJ, Bowd C, Essock EA, Zangwill LM, Weinreb RN, ARVO Abstract 717, 2001) found an area under the ROC curve of for discrimination between glaucoma and healthy eyes, using a Fourier-based LDF. Although this LDF significantly outperformed GDx parameters in that study, the overall unbiased assessment of its performance by split-half analysis resulted in a sensitivity of 73% with specificity of 73%. Fourier measures included in their LDF were the amplitudes of the 2nd, 12th, and 13th components, and also the phases of the 12th and 14th components. In the present study, we found an area under the ROC curve of for the LDF, using RNFL data obtained from SLP using variable corneal compensation. The bootstrap estimate of bias of the area under the ROC curve was small and the application of a standard split-half analysis as an internal validation procedure for our LDF resulted in a sensitivity and specificity of 86% and 89%, respectively. This is an improvement over the previously reported Fourier-based LDF obtained with fixed corneal compensation data. Furthermore, the internal validation analysis of our LDF indicates that its robustness seems to be superior to the discriminant function developed using SLP with fixed corneal compensation data. The lower degree of robustness of the Fourier LDF developed with SLP FCC may be related to the wider variability of RNFL retardance measurements introduced by the erroneous compensation of anterior segment birefringence. In our LDF, the second Fourier component was the most important (in terms of discriminative power) in the LDF equation. Both the amplitude and phase of this component were entered in the LDF after stepwise discriminant analysis. This is an expected result, because the second component has two humps and hence contributes the most to the shape of the RNFL double-hump curve. The other components serve to shape the pattern provided by the second component, so that the composite curve matches the original RNFL pattern. It is likely that the incorporation of higher-frequency components in the LDF may also be involved in the identification of glaucomatous eyes with localized RNFL defects, although our study did not directly address this question. Also, the pattern of the RNFL thickness profile of normal and glaucomatous eyes shows an apparent indentation of the superior hump (Fig. 1). This indentation may be related to the split of nerve fiber layer bundles, as described by Colen and Lemij. 45 Higher-frequency Fourier components may be also necessary to better describe this characteristic of the RNFL thickness profile. The mean overall RNFL thickness, or DC component, was also incorporated in our LDF obtained from SLP-VCC data, as opposed to the previously reported LDF obtained from SLP-FCC data. In that the absolute thickness parameters have improved diagnostic precision with the VCC compared with the FCC, 27 this is not a surprising finding. One limitation of our study was that we did not test our discriminant function on another cohort of subjects to estimate its sensitivity and specificity independently. However, the provided bootstrap estimate of bias under the area of the ROC curve was relatively small, resulting in a reduction of the area under the ROC curve of 1.7%. Therefore, we expect that our results of sensitivity and specificity may be overstated by approximately the same amount. A recent analysis of our LDF with an independent sample resulted in an area under the ROC curve of with sensitivity of 80% and specificity of 90% to discriminate glaucomatous from normal eyes (Michael Sinai, Ph.D., Laser Diagnostic Technologies, unpublished data, September 2002). In conclusion, we showed that the combination of Fourier RNFL measures in an LDF obtained using SLP with variable corneal compensation improved the ability to differentiate between healthy eyes and eyes with glaucomatous visual field loss, compared with the GDx software provided parameters. Further studies using larger samples and different populations are needed to develop and validate standard Fourier-based measures to be incorporated and tested in clinical practice. Acknowledgments The authors thank Michael Sinai, Ph.D., for valuable suggestions provided during a critical review of the manuscript.

7 2612 Medeiros et al. IOVS, June 2003, Vol. 44, No. 6 References 1. Weinreb RN, Dreher AW, Coleman A, Quigley H, Shaw B, Reiter K. Histopathologic validation of Fourier-ellipsometry measurements of retinal nerve fiber layer thickness. Arch Ophthalmol. 1990;108: Dreher AW, Reiter K, Weinreb RN. Spatially resolved birefringence of the retinal nerve fiber layer assessed with a retinal laser ellipsometer. Appl Opt. 1992;31: Zhou Q, Knighton RW. Light scattering and form birefringence of parallel cylindrical arrays that represent cellular organelles of the retinal nerve fiber layer. Appl Opt. 1997;36: Weinreb RN, Shakiba S, Zangwill L. Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol. 1995;119: Tjon-Fo-Sang MJ, Lemij HG. The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. Am J Ophthalmol. 1997;123: Choplin NT, Lundy DC, Dreher AW. Differentiating patients with glaucoma from glaucoma suspects and normal subjects by nerve fiber layer assessment with scanning laser polarimetry. Ophthalmology. 1998;105: Sinai MJ, Essock EA, Fechtner RD, Srinivasan N. Diffuse and localized nerve fiber layer loss measured with a scanning laser polarimeter: sensitivity and specificity of detecting glaucoma. J Glaucoma. 2000;9: Hoh ST, Greenfield DS, Mistlberger A, et al. Optical coherence tomography and scanning laser polarimetry in normal, ocular hypertensive, and glaucomatous eyes. Am J Ophthalmol. 2000;129: Zangwill LM, Bowd C, Berry CC, et al. Discriminating between normal and glaucomatous eyes using the Heidelberg Retina Tomograph, GDx Nerve Fiber Analyzer, and Optical Coherence Tomograph. Arch Ophthalmol. 2001;119: Bowd C, Zangwill LM, Berry CC, et al. Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci. 2001;42: Weinreb RN, Zangwill LM, Berry CC, Bathija R, Sample PA. Detection of glaucoma with scanning laser polarimetry. Arch Ophthalmol. 1998;116: Kwon YH, Hong S, Honkanen RA, Alward WL. Correlation of automated visual field parameters and peripapillary nerve fiber layer thickness as measured by scanning laser polarimetry. J Glaucoma. 2000;9: Chen YY, Chen PP, Xu L, et al. Correlation of peripapillary nerve fiber layer thickness by scanning laser polarimetry with visual field defects in patients with glaucoma. J Glaucoma. 1998;7: Dichtl A, Jonas J, Naumann GOH. Retinal nerve fiber layer thickness in human eyes. Graefes Arch Clin Exp Ophthalmol. 1999; 273: Jindra LF, van Saarloos PP, Eikelbloom RH, Cooper RL. Fourier analysis of the nerve fiber layer by digital imaging techniques. Ophthalmic Surg. 1989;20: Essock EA, Sinai MJ, Fechtner RD, Srinivasan N, Bryant ND. Fourier analysis of nerve fiber layer measurements from scanning laser polarimetry in glaucoma: emphasizing shape characteristics of the double hump pattern. J Glaucoma. 2000;9: Weinreb RN. Evaluating the retinal nerve fiber layer in glaucoma using scanning laser polarimetry. Arch Ophthalmol. 1999;117: Knighton RW, Huang XR, Greenfield DS. Analytical model of scanning laser polarimetry for retinal nerve fiber layer assessment. Invest Ophthalmol Vis Sci. 2002;43: Greenfield DS, Knighton RW, Feuer WJ, Schiffman JC, Zangwill L, Weinreb RN. Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry. Am J Ophthalmol. 2002;134: Garway-Heath DF, Greaney MJ, Caprioli J. Correction for the erroneous compensation of anterior segment birefringence with the scanning laser polarimeter for glaucoma diagnosis. Invest Ophthalmol Vis Sci. 2002;43: klein Brink HB, van Blokland GJ. Birefringence of the human foveal area assessed in vivo with Mueller-matrix ellipsometry. J Opt Soc Am A. 1988;5: Van Blokland GJ, Verhelst SC. Corneal polarization in the living human eye explained with a biaxial model. J Opt Soc Am A. 1987;4: Greenfield DS, Knighton RW, Huang X. Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry. Am J Ophthalmol. 2000;129: Knighton RW, Huang XR. Linear birefringence of the central human cornea. Invest Ophthalmol Vis Sci. 2002;43: Weinreb RN, Bowd C, Greenfield DS, Zangwill LM. Measurement of the magnitude and axis of corneal polarization with scanning laser polarimetry. Arch Ophthalmol. 2002;120: Zhou Q, Weinreb RN. Individualized compensation of anterior segment birefringence during scanning laser polarimetry. Invest Ophthalmol Vis Sci. 2002;43: Weinreb RN, Bowd C, Zangwill LM. Glaucoma detection using scanning laser polarimetry with variable corneal compensation. Arch Ophthalmol. 2003;121: Garway-Heath D, Hitchings RA. Sources of bias in studies of optic disc and retinal nerve fibre layer morphology (Commentary). Br J Ophthalmol. 1998;82: Hodapp E, Parrish RK II, Anderson DR. Clinical Decisions in Glaucoma. St. Louis: Mosby-Year Book; 1993: Bloomfield P. Fourier analysis of time series: an introduction. Wiley Series in Probability and Statistics. New York: John Wiley & Sons, Inc.; 2000: Bracewell RN. The Fourier Transform and its Applications. New York: McGraw-Hill; Cooley JW, Tukey JW. An algorithm for the machine computation of complex Fourier series. Math Comput. 1965;19: Zonst E. Understanding the FFT. A tutorial on the Algorithm and Software for Laymen, Students, Technicians and Working Engineers. Titusville, FL: Citrus Press; DeLong ER, DeLong DM, Clarke-Pearson D. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44: Efron B, Tibshirani R. An introduction to the bootstrap. Monogr Stat Appl Probability. New York: Chapman & Hall; Efron B, Tibshirani R. Improvements on cross-validation: the.632 bootstrap method. J Am Stat Assoc. 1997;92: Steyerberg EW, Harrel FE, Borsboom GJJM, et al. Internal validation of predictive models: efficiency of some procedures for logistic regression analysis. J Clin Epidemiol. 2001;54: Harrel FE, Lee KL, Mark DB. Multivariable prognostic models: issues in developing models, evaluating assumptions and adequacy, and measuring and reducing errors. Stat Med. 1996;15: Chernick MR. Bootstrap methods: a practitioner s guide. Wiley Series in Probability and Statistics. New York: John Wiley & Sons, Inc.; 1999: Obuchowski NA, Lieber ML. Confidence intervals for the receiver operating characteristic area in studies with small samples. Acad Radiol. 1998;5: Margolis DJ, Bilker W, Boston R, et al. Statistical characteristics of area under the receiver operating characteristic curve for a simple prognostic model using traditional and bootstrapped approaches. J Clin Epidemiol. 2002;55: Horn FK, Jonas JB, Martus P, Mardin CY, Burde WM. Polarimetric measurement of retinal nerve fiber layer thickness in glaucoma diagnosis. J Glaucoma. 1999;8: Trible JR, Schultz RD, Robinson JC, et al. Accuracy of scanning laser polarimetry in the diagnosis of glaucoma. Arch Ophthalmol. 1999;117: Paczka JA, Friedman DS, Quigley HA, Barron Y, Vitale S. Diagnostic capabilities of frequency-doubling technology, scanning laser polarimetry, and nerve fiber layer photographs to distinguish glaucomatous damage. Am J Ophthalmol. 2001;131: Colen TP, Lemij HG. Prevalence of split nerve fiber layer bundles in healthy eyes imaged with scanning laser polarimetry. Ophthalmology. 2001;108:

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 61 Abstract Purpose The GDx is a scanning laser polarimeter that assesses

More information

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of

More information

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Korean J Ophthalmol Vol. 18:1-8, 2004 Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Sun Young Lee, MD,

More information

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Felipe A. Medeiros, Christopher Bowd, Linda M. Zangwill, Chirag Patel, and Robert N. Weinreb From the Hamilton

More information

Scanning laser polarimetry (SLP) provides real-time, objective

Scanning laser polarimetry (SLP) provides real-time, objective The Effect of Atypical Birefringence Patterns on Glaucoma Detection Using Scanning Laser Polarimetry with Variable Corneal Compensation Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda

More information

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD CLINICAL SCIENCES Comparison of the GDx VCC Scanning Laser Polarimeter, HRT II Confocal Scanning Laser Ophthalmoscope, and Stratus OCT Optical Coherence Tomograph for the Detection of Glaucoma Felipe A.

More information

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements 586 Glaucoma Service, Department of Ophthalmology, University of Campinas, Campinas, Brazil R Lauande-Pimentel R A Carvalho H C Oliveira D C Gonçalves L M Silva V P Costa Glaucoma Service, Department of

More information

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module

More information

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting (2007) 21, 1050 1055 & 2007 Nature Publishing Group All rights reserved 0950-222X/07 $30.00 www.nature.com/eye CLINICAL STUDY A comparison of HRT II and GDx imaging for glaucoma detection in a primary

More information

Assessment of retinal ganglion cell loss is critical for the

Assessment of retinal ganglion cell loss is critical for the Scanning Laser Polarimetry with Variable Corneal Compensation: Identification and Correction for Corneal Birefringence in Eyes with Macular Disease Harmohina Bagga, David S. Greenfield, and Robert W. Knighton

More information

Clinical decision making based on data from GDx: One year observations

Clinical decision making based on data from GDx: One year observations Washington University School of Medicine Digital Commons@Becker Open Access Publications 2002 Clinical decision making based on data from GDx: One year observations James C. Bobrow Washington University

More information

OtticaFisiopatologica

OtticaFisiopatologica Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza

More information

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation Diagnostic Accuracy of Scanning Laser olarimetry with Enhanced versus Variable Corneal T. A. Mai, MD, N. J. Reus, MD, hd, H. G. Lemij, MD, hd urpose: To compare the diagnostic accuracy of scanning laser

More information

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:

More information

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan Correspondence to: Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute,

More information

Correlating Structure With Function in End-Stage Glaucoma

Correlating Structure With Function in End-Stage Glaucoma C L I N I C A L S C I E N C E Correlating Structure With Function in End-Stage Glaucoma Eytan Z. Blumenthal, MD; Amjad Horani, MD; Rajesh Sasikumar, MD; Chandrasekhar Garudadri, MD; Addepalli Udaykumar,

More information

Reliability analyses of the GDx nerve-fiber analyzer

Reliability analyses of the GDx nerve-fiber analyzer VOL. 29 NO. 2 PHILIPPINE JOURNAL OF Ophthalmology APRIL ORIGINAL ARTICLE - JUNE 2004 1, 2, 3 Patricia M. Khu, MD, MS Edgardo U. Dorotheo, MD 1 Lawrence Tinio, MD 2 Cynthia P. Cordero, MS 4 1, 2, 3 Manuel

More information

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Reza Zarei, MD 1 Mohammad Soleimani, MD 2 Sasan Moghimi, MD 3 Mohammad Yaser Kiarudi, MD 2 Mahmoud Jabbarvand, MD 1 Yadollah

More information

In some patients with glaucoma, standard (achromatic) automated

In some patients with glaucoma, standard (achromatic) automated Detecting Early Glaucoma by Assessment of Retinal Nerve Fiber Layer Thickness and Visual Function Christopher Bowd, 1 Linda M. Zangwill, 1 Charles C. Berry, 2 Eytan Z. Blumenthal, 1 Cristiana Vasile, 1

More information

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS The Relationship between Standard Automated Perimetry and GDx VCC Measurements Nicolaas J. Reus and Hans G. Lemij PURPOSE. To investigate the relationship between retinal light sensitivity measured with

More information

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro

More information

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Felipe A. Medeiros, Luciana M. Alencar, Linda M. Zangwill, Christopher

More information

Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer

Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer European Journal of Ophthalmology / Vol. 15 no. 2, 2005 / pp. 246-254 Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer M. IESTER 1,2, A. MERMOUD 1 1 Hopital Ophtalmique

More information

Analysis of GDx-VCC Polarimetry Data by Wavelet-Fourier Analysis across Glaucoma Stages

Analysis of GDx-VCC Polarimetry Data by Wavelet-Fourier Analysis across Glaucoma Stages Analysis of GDx-VCC Polarimetry Data by Wavelet-Fourier Analysis across Glaucoma Stages Edward A. Essock, 1,2 Yufeng Zheng, 1 and Pinakin Gunvant 1 PURPOSE. The purpose of this study was to apply shape-based

More information

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T et al Original article Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T1, Aggarwal A1, Bali SJ2, Sharma A1, Shah

More information

Relationship Between Structure

Relationship Between Structure Original Article Relationship Between Structure and Function of the Optic Nerve Head-Glaucoma versus Normal Dr Savita Bhat, Dr Anna Elias, Dr Siddharth Pawar, Dr S.J. Saikumar, Dr Alpesh Rajput, superior,

More information

The Role of the RNFL in the Diagnosis of Glaucoma

The Role of the RNFL in the Diagnosis of Glaucoma Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

Scanning laser polarimetry (SLP) incorporates a confocal

Scanning laser polarimetry (SLP) incorporates a confocal Scanning Laser Polarimetry with Enhanced Corneal Compensation and Optical Coherence Tomography in Normal and Glaucomatous Eyes Mitra Sehi, 1 Stephen Ume, 1 David S. Greenfield, 1 and Advanced Imaging in

More information

A vailability of the scanning laser polarimeter now permits

A vailability of the scanning laser polarimeter now permits 70 CLINICAL SCIENCE Specificity and sensitivity of glaucoma detection in the Japanese population using scanning laser polarimetry Shigeo Funaki, Motohiro Shirakashi, Kiyoshi Yaoeda, Haruki Abe, Shiho Kunimatsu,

More information

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry European Journal of Ophthalmology / Vol. 13 no. 4, 2003 / pp. 377-382 The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

More information

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma (2009) 23, 2182 2186 & 2009 Macmillan Publishers Limited All rights reserved 09-222X/09 $32.00 www.nature.com/eye CLINICAL STUDY Relationship between GDx VCC and Stratus OCT in juvenile glaucoma R Zareii,

More information

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology

Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthaology Scanning Laser Polarimetry, but Not Optical Coherence Tomography Predicts Permanent Visual Field Loss in Acute Nonarteritic Anterior Ischemic

More information

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC Department of Ophthalmology, Medical University of Vienna, Austria Correspondence to Clemens Vass, Department of Ophthalmology and Optometry, Medical University of Vienna, General Hospital, Währinger Gürtel

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

Eye (2005), 1 9 & 2005 Nature Publishing Group All rights reserved X/05 $

Eye (2005), 1 9 & 2005 Nature Publishing Group All rights reserved X/05 $ (2005), 1 9 & 2005 Nature Publishing Group All rights reserved 0950-222X/05 $30.00 www.nature.com/eye Automated detection of wedgeshaped defects in polarimetric images of the retinal nerve fibre layer

More information

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects 접수번호 : 2008-105 Korean Journal of Ophthalmology 2009;23:169-175 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.3.169 Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

G laucoma is a progressive optic neuropathy in which

G laucoma is a progressive optic neuropathy in which 1135 EXTENDED REPORT The correlation between optic nerve head topographic measurements, peripapillary nerve fibre layer thickness, and visual field indices in glaucoma Y-W Lan, D B Henson, A J Kwartz...

More information

Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender

Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Bowd et al. Vol. 19, No. 1/January 2002/J. Opt. Soc. Am. A 197 Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Christopher Bowd,

More information

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS Nerve fiber layer thickness in normals and glaucoma patients 403 NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS HIROTAKA SUZUMURA, KAYOKO HARASAWA, AKIKO KOBAYASHI and NARIYOSHI ENDO Department

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography 194 Review Article Evaluating the Optic Nerve and Retinal Nerve Fibre Layer: The Roles of Heidelberg Retina Tomography, Scanning Laser Polarimetry and Optical Coherence Tomography Sek-Tien Hoh, 1 MBBS,

More information

Correlation Between Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry in Glaucoma Suspects and Glaucomatous Eyes

Correlation Between Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry in Glaucoma Suspects and Glaucomatous Eyes Korean J Ophthalmol Vol. 18:89-99, 2004 Correlation Between Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry in Glaucoma Suspects and Glaucomatous Eyes Su Hyun Kim, MD, Hunei Hong,

More information

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous

More information

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,

More information

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD Journal of Glaucoma 10:163 169 2001 Lippincott Williams & Wilkins, Inc. Influence of Scan Radius Correction for Ocular Magnification and Relationship Between Scan Radius With Retinal Nerve Fiber Layer

More information

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Jimmy S. M. Lai*, Clement C. Y. Tham, Jonathan C. H. Chan*, Nelson K. F. Yip, Wilson W. T. Tang, Patrick S. H. Li*, Jane

More information

The only measure of RNFL Integrity : GDxPRO

The only measure of RNFL Integrity : GDxPRO The ideal glaucoma assessment tool for your practice The only measure of RNFL Integrity : GDxPRO Proven Clinical Performance Numerous studies have validated the diagnostic accuracy and performance of the

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2012 September ; 130(9): 1107 1116. doi:10.1001/archophthalmol.2012.827. A Combined Index of Structure and Function

More information

Glaucoma detection using the GDx nerve fiber analyzer and the retinal thickness analyzer

Glaucoma detection using the GDx nerve fiber analyzer and the retinal thickness analyzer European Journal of Ophthalmology / Vol. 16 no. 0, 2006 / pp. 21-000 Glaucoma detection using the GDx nerve fiber analyzer and the retinal thickness analyzer E.M. HOFFMANN 1, 2, C. BOWD 2, N. KLEIN 1,

More information

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography CLINICAL SCIENCES Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography Seong Bae Park, MD; Kyung Rim Sung, MD, PhD; Sung Yong Kang, MD; Kyung Ri Kim, BS;

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: OPHTHALMOLOGIC TECHNIQUES PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical

More information

Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation

Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation Korean Journal of Ophthalmology 22(1):18-25, 2008 DOI : 10.3341/kjo.2008.22.1.18 Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation

More information

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of

More information

Although measurements of the optic disc and retinal nerve

Although measurements of the optic disc and retinal nerve Longitudinal Variability of Optic Disc and Retinal Nerve Fiber Layer Measurements Christopher Kai-shun Leung, 1,2 Carol Yim-lui Cheung, 1 Dusheng Lin, 1,3 Chi Pui Pang, 1 Dennis S. C. Lam, 1 and Robert

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December-2013 108 Name of Chief and Corresponding Author : Dr Chandrima Paul TITLE : Comparison of glaucoma diagnostic ability

More information

The only measure of RNFL Integrity: GDxPRO

The only measure of RNFL Integrity: GDxPRO The only measure of RNFL Integrity: GDxPRO Look beyond thickness. Underlying structural organization is the key to RNFL At times, stepping onto ice is a risky proposition. Just knowing its thickness might

More information

Optical coherence tomography (OCT) is a noninvasive,

Optical coherence tomography (OCT) is a noninvasive, Ability of Stratus OCT to Detect Progressive Retinal Nerve Fiber Layer Atrophy in Glaucoma Eun Ji Lee, 1,2 Tae-Woo Kim, 1,2 Ki Ho Park, 2 Mincheol Seong, 3 Hyunjoong Kim, 4 and Dong Myung Kim 2 PURPOSE.

More information

Relation of optic disc topography and age to thickness of retinal nerve fibre layer as measured using scanning laser polarimetry, in normal subjects

Relation of optic disc topography and age to thickness of retinal nerve fibre layer as measured using scanning laser polarimetry, in normal subjects Br J Ophthalmol 2000;84:473 478 473 Department of Ophthalmology, Celal Bayar University School of Medicine, Manisa, Turkey A B Toprak Department of Ophthalmology, Beyoglu Hospital, Taksim, Istanbul, Turkey

More information

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2

More information

RETINAL NERVE FIBER LAYER

RETINAL NERVE FIBER LAYER CLINICAL SCIENCES The Effect of Scan Diameter on Retinal Nerve Fiber Layer Thickness Measurement Using Stratus Optic Coherence Tomography Giacomo Savini, MD; Piero Barboni, MD; Michele Carbonelli, MD;

More information

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Available online at   Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6): Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve

More information

Optic disc damage has been shown to precede visual field

Optic disc damage has been shown to precede visual field Confocal Scanning Laser Ophthalmoscopy Classifiers and Stereophotograph Evaluation for rediction of Visual Field Abnormalities in Glaucoma-Suspect Eyes Christopher Bowd, 1 Linda M. Zangwill, 1 Felipe A.

More information

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

More information

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(3):166-173 DOI: 10.3341/kjo.2011.25.3.166 Original Article Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis

More information

Glaucoma is an optic neuropathy characterized by a gradual

Glaucoma is an optic neuropathy characterized by a gradual Optical Coherence Tomography Machine Learning Classifiers for Glaucoma Detection: A Preliminary Study Zvia Burgansky-Eliash, 1,2 Gadi Wollstein, 1,2 Tianjiao Chu, 3 Joseph D. Ramsey, 4 Clark Glymour, 4

More information

MEDICAL POLICY. Proprietary Information of YourCare Healthcare

MEDICAL POLICY. Proprietary Information of YourCare Healthcare MEDICAL POLICY PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy criteria are not applied.

More information

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Hiroshi Nakamura,* Toshine Maeda,* Yasuyuki Suzuki and Yoichi Inoue* *Eye Division of Olympia Medical Clinic, Tokyo, Japan; Department of

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013 268 Comparison of optic nerve morphology in eyes with glaucoma and eyes with non-arteritic anterior ischemic optic neuropathy by Fourier domain optical coherence tomography YUXIN YANG 1, HAITAO ZHANG 1,

More information

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Acta Medica Anatolia Volume 2 Issue 1 2014 Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Remzi Mısır 1, Sinan Sarıcaoğlu

More information

The influence of depolarization and corneal birefringence on ocular polarization

The influence of depolarization and corneal birefringence on ocular polarization INSTITUTE OF PHYSICS PUBLISHING JOURNAL OF OPTICS A: PURE AND APPLIED OPTICS J. Opt. A: Pure Appl. Opt. 6 (24) S91 S99 PII: S1464-4258(4)67964-5 The influence of depolarization and corneal birefringence

More information

CLINICAL SCIENCES. optic neuropathy characterized

CLINICAL SCIENCES. optic neuropathy characterized CLINICAL SCIENCES Spectral-Domain Optical Coherence Tomography for Detection of Localized Retinal Nerve Fiber Layer Defects in Patients With Open-Angle Glaucoma Na Rae Kim, MD; Eun Suk Lee, MD, PhD; Gong

More information

Discriminating between Normal and Glaucoma-Damaged Eyes with the Heidelberg Retina Tomograph 3

Discriminating between Normal and Glaucoma-Damaged Eyes with the Heidelberg Retina Tomograph 3 Discriminating between Normal and Glaucoma-Damaged Eyes with the Heidelberg Retina Tomograph 3 Antonio Ferreras, MD, PhD, 1 Luís E. Pablo, MD, PhD, 1 José M. Larrosa, MD, PhD, 1 Vicente Polo, MD, PhD,

More information

Receiver operating characteristic (ROC) curves are a wellaccepted

Receiver operating characteristic (ROC) curves are a wellaccepted A Statistical Approach to the Evaluation of Covariate Effects on the Receiver Operating Characteristic Curves of Diagnostic Tests in Glaucoma Felipe A. Medeiros, 1 Pamela A. Sample, 1 Linda M. Zangwill,

More information

NIH Public Access Author Manuscript Ophthalmology. Author manuscript; available in PMC 2009 October 4.

NIH Public Access Author Manuscript Ophthalmology. Author manuscript; available in PMC 2009 October 4. NIH Public Access Author Manuscript Published in final edited form as: Ophthalmology. 2008 August ; 115(8): 1352 1357.e2. doi:10.1016/j.ophtha.2008.01.011. Combining Nerve Fiber Layer to Optimize Glaucoma

More information

Glaucoma: Diagnostic Modalities

Glaucoma: Diagnostic Modalities Glaucoma: Diagnostic Modalities - Dr. Barun Kumar Nayak, Dr. Sarika Ramugade Glaucoma is a leading cause of blindness in the world, especially in older people. Early detection and treatment by ophthalmologist

More information

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e The GDxVCC Early answers and ongoing assessment for glaucoma The quantifiable approach to quality care Only Humphrey GPA software Early insight

More information

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Nerve Fiber Layer

More information

RETINAL SCANNING LASER POLARIMETRY AND

RETINAL SCANNING LASER POLARIMETRY AND RETINAL SCANNING LASER POLARIMETRY AND METHODS TO COMPENSATE FOR CORNEAL BIREFRINGENCE ZHOU Q. 1,2 ABSTRACT Scanning laser polarimetry (SLP) was developed to provide objective assessment of the retinal

More information

Comparative study of new imaging technologies for the diagnosis of glaucoma: Protocol Approved by the Ethics Committee

Comparative study of new imaging technologies for the diagnosis of glaucoma: Protocol Approved by the Ethics Committee Comparative study of new imaging technologies for the diagnosis of glaucoma: Protocol Approved by the Ethics Committee HTA 09/22/111. Applicants: Augusto Azuara-Blanco (CI), Jennifer Burr,, Rodolfo Hernández,

More information

The thickness of the retinal nerve fiber layer (RNFL) is an

The thickness of the retinal nerve fiber layer (RNFL) is an Multidisciplinary Ophthalmic Imaging Atypical Retardation Patterns in Scanning Laser Polarimetry Are Associated with Low Peripapillary Choroidal Thickness Ralf P. Tornow, 1 Wolfgang A. Schrems, 1 Delia

More information

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma Hindawi Ophthalmology Volume 2017, Article ID 78365, 8 pages https://doi.org/10.1155/2017/78365 Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer

More information

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(5):369-377 http://dx.doi.org/10.3341/kjo.2012.26.5.369 Original Article A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT)

More information

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma

LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS. Darshana Soma LARGE DISCS WITH LARGE CUPS A DIAGNOSTIC CHALLENGE IN AFRICAN PATIENTS Darshana Soma A research report submitted to the Faculty of Health Sciences, University of the Witwatersrand, Johannesburg, in partial

More information

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Donald C. Hood 1,2,* and Ali S. Raza 1 1 Department of Psychology, Columbia

More information

Advances in OCT Murray Fingeret, OD

Advances in OCT Murray Fingeret, OD Disclosures Advances in OCT Murray Fingeret, OD Consultant Alcon, Allergan, Bausch & Lomb, Carl Zeiss Meditec, Diopsys, Heidelberg Engineering, Reichert, Topcon Currently Approved OCT Devices OCT Devices

More information

Targeting Intraocular Pressure in Glaucoma: a Teaching Case Report 1

Targeting Intraocular Pressure in Glaucoma: a Teaching Case Report 1 Targeting Intraocular Pressure in Glaucoma: a Teaching Case Report 1 By: Andrew Kemp, OD, Marcus Gonzales, OD, FAAO, Joe DeLoach, OD, FAAO, and Zanna Kruoch, OD FAAO Background Glaucoma is a range of conditions

More information

Science & Technologies

Science & Technologies STANDARD COMPUTERIZED PERIMETRY IN FUNCTION OF DIAGNOSTIC GLAUCOMA Iljaz Ismaili, 1 Gazepov Strahil, 2, Goshevska Dashtevska Emilija 1 1 University Eye Clinic,Skopje 2 Clinical Hospital, Shtip Abstract

More information

Because of the progressive nature of the disease, early

Because of the progressive nature of the disease, early Glaucoma Improving Glaucoma Detection Using Spatially Correspondent Clusters of Damage and by Combining Standard Automated Perimetry and Optical Coherence Tomography Ali S. Raza, 1,2 Xian Zhang, 1 Carlos

More information

CLINICAL SCIENCES. Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD

CLINICAL SCIENCES. Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD CLINICAL SCIENCES Achromatic and Short-Wavelength Automated Perimetry in Patients With Glaucomatous Large Cups Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD

More information

Ophthalmology. Proofs of ophthal and PTS# 4157 ===== Dear Author:

Ophthalmology. Proofs of ophthal and PTS# 4157 ===== Dear Author: Ophthalmology Copy of e-mail Notification zf1 Proofs of ophthal - 00 and PTS# 1 ===== Dear Author: A RESPONSE TO THIS EMAIL IS IMPERATIVE WITHIN HOURS, EITHER WITH CORRECTIONS OR A BRIEF STATEMENT OF ACCEPTANCE.

More information

Peripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept.

Peripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept. Peripapillary Retinal Thickness Maps in the Evaluation of Glaucoma Patients: A Novel Concept The Harvard community has made this article openly available. Please share how this access benefits you. Your

More information

CLINICAL SCIENCES. Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma

CLINICAL SCIENCES. Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma CLINICAL SCIENCES Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma The Confocal Scanning Laser Ophthalmoscopy Ancillary Study to the Ocular

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2010 November 18.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2010 November 18. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2009 July ; 127(7): 875 881. doi:10.1001/archophthalmol.2009.145. Measurement of Local Retinal Ganglion Cell Layer

More information

Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes

Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes Mohammad-Mehdi Parvaresh, MD 1 Marjan Imani, MD 2 Mohsen Bahmani-Kashkouli,

More information