Although measurements of the optic disc and retinal nerve
|
|
- Anne Fowler
- 5 years ago
- Views:
Transcription
1 Longitudinal Variability of Optic Disc and Retinal Nerve Fiber Layer Measurements Christopher Kai-shun Leung, 1,2 Carol Yim-lui Cheung, 1 Dusheng Lin, 1,3 Chi Pui Pang, 1 Dennis S. C. Lam, 1 and Robert N. Weinreb 2 PURPOSE. To evaluate the longitudinal variability of optic disc and retinal nerve fiber layer (RNFL) measurements obtained from optical coherence tomography (OCT), scanning laser polarimetry (SLP), and confocal scanning laser ophthalmoscopy (CSLO). METHODS. Forty-five normal and 43 glaucomatous eyes of 88 subjects were analyzed in this longitudinal study. RNFL thickness was measured by OCT (StatusOCT; Carl Zeiss Meditec, Dublin, CA) and SLP (GDx VCC; Carl Zeiss Meditec). Neuroretinal rim area was measured by CSLO (HRT 3; Heidelberg Engineering, GmbH, Dossenheim, Germany). Three separate measurements taken over an average period of months were used to evaluate measurement variability. Reproducibility coefficient, coefficient of variation, intraclass correlation coefficient (ICC), and sensitivity to change [(97.5 percentile value 2.5 percentile value)/2 within-subject standard deviation (Sw)] of the global measures were calculated. The association between RNFL and rim area measurement variability and visual field MD (mean deviation) was evaluated with regression analysis. RESULTS. Low variability was found for global and sectorial rim area and RNFL measurements. The reproducibility coefficient, ICC, and sensitivity to change for OCT average RNFL thickness, GDx VCC TSNIT average, and HRT global rim area were 11.7 m (95% confidence interval [CI]: m), 0.97 ( ), 10.2 ( ); 4.7 m ( m), 0.98 ( ), 11.3 ( ); and 0.22 mm 2 ( mm 2 ), 0.97 ( ), 9.3 ( ), respectively. No association was found between OCT (r [95% CI: ], P 0.924) and SLP (r [95% CI: ], P 0.756) RNFL thickness variances and visual field MD. The association between CSLO rim area variance and visual field MD became insignificant after adjustment for reference height variance. CONCLUSIONS. Longitudinal RNFL and neuroretinal rim measurements obtained with OCT, SLP, and CSLO have low variability. As the measurement variability does not change with the severity of glaucoma, these parameters are useful for assessment of glaucoma progression. (Invest Ophthalmol Vis Sci. 2008;49: ) DOI: /iovs Although measurements of the optic disc and retinal nerve fiber layer (RNFL) with optical coherence tomography (OCT), scanning laser polarimetry (SLP), and confocal scanning laser ophthalmoscope (CSLO) have been shown to have high diagnostic performance for detection of glaucomatous damage, 1 5 their roles in analyzing glaucoma progression are still evolving. Distinguishing genuine change from measurement variability is a challenge in studying glaucoma progression with these imaging instruments. Repeatability and reproducibility are two commonly used indices to indicate intra- and intervisit measurement variability, respectively. 6,7 Intervisit variability is generally higher than intravisit variability, probably because of the potential differences in the scanning techniques among different operators and the different scanning angles and locations during different visits. 8,9 Most studies on RNFL imaging have been focused on repeatability with measurements taken in a single visit, or reproducibility over a few days to a few weeks However, in the usual clinical setting, imaging is seldom repeated within a day or a month. Determining reproducibility over a longer period with multiple measurements would be clinically more useful and relevant to reflect the long-term measurement variability of these imaging instruments. The StratusOCT (Carl Zeiss Meditec, Dublin, CA), HRT 3 (Heidelberg Engineering, GmbH, Dossenheim, Germany), and GDx VCC (Carl Zeiss Meditec) are the latest commercially available versions of OCT, CSLO, and SLP, respectively. Although no difference has been found in their diagnostic sensitivity for glaucoma detection, 4,17 it is uncertain whether the results that they obtain are equally reproducible during longitudinal assessment. The purpose of this study was to evaluate the longitudinal variability of optic disc and RNFL measurements in normal subjects and in patients with glaucoma, with the StratusOCT, HRT 3, and GDx VCC. From the 1 Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong, Peoples Republic of China; the 2 Department of Ophthalmology, Hamilton Glaucoma Center, University of California, San Diego, California; and 3 The STU/ CUHK Joint Shantou International Eye Centre (JSIEC), Shantou University Medical College, Peoples Republic of China. Submitted for publication September 11, 2007; revised November 23, 2007, and January 14, February 19, April 5, and May 8, 2008; accepted September 5, Disclosure: C.K. Leung, Carl Zeiss Meditec (F); C.Y. Cheung, None; D. Lin, None; C.P. Pang, None; D.S.C. Lam, None; R.N. Weinreb, Carl Zeiss Meditec (C, F) The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked advertisement in accordance with 18 U.S.C solely to indicate this fact. Corresponding author: Christopher Kai-shun Leung, Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, 147K Argyle Street, Kowloon, Hong Kong, Peoples Republic of China; tlims00@hotmail.com. METHODS Subjects Forty-five normal subjects and 43 patients with glaucoma followed up during the period from February 2006 to October 2007 at the University Eye Center, The Chinese University of Hong Kong, were included in the analysis. These subjects were enrolled in a glaucoma imaging study designed to evaluate test retest variability longitudinally and glaucoma progression and had had at least three serial measurements obtained before October All subjects underwent a full ophthalmic examination including visual acuity, refraction, intraocular pressure measurement with Goldmann tonometry, gonioscopy, and fundus examination. The inclusion criteria were best corrected visual acuity of not worse than 20/40, spherical refractive error within the range of 6.0 to 6.0 D, and 6.0 D of cylinder. Individuals were excluded if they had a history of any retinal diseases, surgery or laser procedures, diabetes mellitus, or neurologic diseases. Normal subjects were indi- Investigative Ophthalmology & Visual Science, November 2008, Vol. 49, No Copyright Association for Research in Vision and Ophthalmology
2 IOVS, November 2008, Vol. 49, No. 11 Longitudinal Variability of Optic Disc and RNFL Measurements 4887 viduals with no ocular diseases. In particular, they had no visual field defects, no structural optic disc abnormalities, no history of intraocular pressure higher than 21 mm Hg, and no history of ocular or neurologic diseases. The patients with glaucoma were defined based on the presence of visual field defects (described later). Only one eye was selected randomly in each subject for analysis. Each included eye underwent OCT, CSLO, and SLP imaging in three separate visits over an average period of months, with a between-visit duration of at least 3 months. This study was conducted in accordance with the standards stated in the Declaration of Helsinki and approved by Clinical Research Ethics Committee of Hong Kong Hospital Authority Kowloon West Cluster, with informed consent obtained. Visual Field Testing Standard visual field testing was performed using static automated white-on-white threshold perimetry (SITA Standard 24-2, Humphrey Field Analyzer II; Carl Zeiss Meditec). Visual field testing was repeated at each follow-up visit in patients with glaucoma. A visual field was defined as reliable when fixation loss was 20% and false-positive and -negative rates were 25%. Visual field sensitivity was expressed as MD (mean deviation) and PSD (pattern standard deviation), as calculated by the perimetry software. None of the normal subjects had a visual field defect. A field defect was defined as having three or more significant (P 0.05) non edge-contiguous points with at least one at the P 0.01 level on the same side of the horizontal meridian in the PSD plot and as having results outside normal limits in the glaucoma hemifield test, confirmed with at least two visual field tests. Visual field progression was evaluated with the Collaborative Initial Glaucoma Treatment Study (CIGTS) visual field scoring system. 18 Visual field testing was repeated when there was evidence suggestive of progression. One subject was not included in the analysis because of progressive visual field change. Optic Disc and RNFL Imaging The pupils were not routinely dilated with a pharmacologic agent during optic disc and RNFL imaging. However, dilation with tropicamide and phenylephrine (0.5% each) was performed when the pupil size was not large enough to obtain an image with the required quality (described later). Images with poor centration, poor focus, or missing data were detected by the operator at the time of imaging, with rescanning performed in the same session. Subjects were excluded if good-quality images could not be obtained after three consecutive scans in the same visit. Optical Coherence Tomography. OCT was performed with the StratusOCT and analyzed with software version 4.0.1(Carl Zeiss Meditec Inc.). Each subject underwent RNFL scanning with the fast RNFL (3.4 mm; 256 A-scans) protocol. The nerve fiber layer, with its high reflectivity signal, can be visualized as the first layer in red on the scan. Its thickness is determined by the difference in distance between the vitreoretinal interface and a posterior border based on a predefined reflectivity signal level. The RNFL thickness was reported in the analysis printout after averaging the results of three sequential circular scans. All the images obtained had a signal strength of at least 7. Three subjects were excluded because of suboptimal signal strength in one of the three visits. Confocal Scanning Laser Ophthalmoscopy. CSLO was performed with the HRT 3 (Heidelberg Engineering) and analyzed with Heidelberg Eye Explorer version In brief, a three-dimensional topographic image consisting of up to pixels is constructed from multiple focal planes axially along the optic nerve head. An average of three consecutive scans is obtained and aligned to compose a single mean topography for analysis. An experienced examiner outlined the optic disc margin on the mean topographic image. Images with an image quality SD greater than 30 m were excluded. Once the optic disc margin was drawn, the software automatically transferred the disc margin to subsequent visits and calculated all the optic disc measurements. The reference plane was defined at 50 m posterior to the mean retinal height between 350 and 356 along the contour line. The area above the reference plane was defined as the rim and that below as the cup. In this study, we focused only on the neuroretinal rim area measurements. Two subjects with image quality SD more than 30 m in one of the three visits were excluded. Scanning Laser Polarimetry. SLP was performed with the GDx VCC and analyzed with software version (Carl Zeiss Meditec). The GDx VCC quantifies the RNFL by first measuring the eyespecific corneal birefringence, consisting of the corneal polarization axis and magnitude. It is determined with a macular image acquired with the retardance of VCC set to 0. The Henle fiber layer and corneal retardation can then be measured from the macular retardation profile. Subsequent imaging sessions for the same individual used the previously obtained values of corneal polarization magnitude and axis. To ensure accurate measurement, the software provides an image quality check score (1 10) based on the correct alignment, fixation, and refraction of the scan. A minimum score of 8 was set to be the minimum standard for a good-quality scan. Ten subjects were excluded because of a poor image quality score in one or more visits. One normal subject and one with glaucoma were also excluded because of images with atypical birefringence patterns (defined as having variable and splotchy areas of high and low retardation not conforming to the normal anatomic profile of RNFL in the retardation map and with a typical scan score less than 60). 19 The superior average, inferior average, and TSNIT average reported in the standard printout were analyzed in this study. The engineers from Carl Zeiss Meditec and Heidelberg Engineering check the calibration of our OCT and CSLO on a regular basis, as recommended by the manufacturers (the calibration of both OCT and CSLO was checked during the study period). For SLP, there is a daily system self-test that verifies the scan field centration and retardation measurement reproducibility. If the results exceed the preset acceptance limits, an error message is displayed, and the instrument is sent to the company for calibration. Statistical Analysis Statistical analyses were performed with commercial software (SPSS ver. 15.0; SPSS Inc, Chicago, IL). Differences in age, refraction, visual field MD between the diagnostics groups were evaluated by independent t-test. By setting the confidence interval as 20% on either side of the estimate of within-subject standard deviation (Sw): n /[ (m 1)], where n is the number of subjects and m is the number of observations, 20 it could be estimated that a minimum of 25 subjects in each group would be necessary. The Sw was calculated as the square root of the within-subject mean square of error (the unbiased estimator of the component of variance due to random error) in a one-way random-effects model. 20 Intersession sectorial and global measurements were compared with repeated-measures analysis of variance. The intersession within-subject standard deviation (Sw), coefficient of variation, reproducibility coefficient (2.77 Sw), and intraclass correlation coefficient (ICC) were computed. Statistical comparison of ICC was performed based on work described by Fisher. 21 A sample size calculation estimated that 87 subjects would be necessary for 80% power at a 5% two-sided significance level for the statistical test to detect a mean difference in ICC of 0.10 between 0.80 and The relationships between visual field MD and each of the global measures variances (average RNFL thickness for OCT; TSNIT average for GDx VCC; and global rim area for HRT) and image quality score variances (signal strength for OCT, image quality SD for HRT, and quality score for GDx VCC) were evaluated with multiple linear regression models with visual field MD as a dependent variable, global measures variances, and image quality score variances as predictors. The association between the SD and the mean of the repeated measurements was evaluated with correlation analysis (Kendall s correlation coefficient). Sensitivity to change 23 was calculated as (97.5 percentile value 2.5 percentile value)/2 Sw. This parameter refers
3 4888 Leung et al. IOVS, November 2008, Vol. 49, No. 11 TABLE 1. Demographics of Normal Subjects and Patients with Glaucoma to the number of progression events that can be detected in a given cohort of patients. A higher number signifies a higher ability to detect change. In all the analyses, P 0.05 was considered statistically significant. RESULTS Normal (n 45) Glaucoma (n 43) P Male/female 18/27 27/ * Age (SD) (y) 47.2 (12.9) 52.0 (13.6) Range 3.0 to to 73.0 Refraction (SD) (D) 0.69 (2.31) 1.04 (3.07) Range 4.75 to to 4.25 Visual field MD (SD) (db) 0.27 (1.29) 9.24 (7.32) Range 3.35 to to 0.53 * 2 test. Independent samples t-test. Forty-five normal and 43 glaucomatous eyes of 88 subjects were included in the analysis after excluding 17 subjects with inadequate image quality. There was no difference in age and refraction between the normal and glaucoma groups (Table 1). The average period ( SD) between the first and second visits was months, and that between the second and third visits was months. No significant changes in RNFL thicknesses and rim area measures (all with P 0.239, repeatedmeasures analysis of variance) in the glaucoma group were detected within the study period ( months). The overall image quality indices for OCT, HRT, and GDx VCC were (signal strength), (image quality SD) and (quality score), respectively. The HRT global rim area variance was significantly correlated with the variance of image quality SD (r 0.284, P 0.007), whereas no association was found between the variances of OCT average RNFL thickness (P 0.837)/GDx VCC TSNIT average (P 0.756) and the variances of their respective image-quality indices. The mean, Sw, and reproducibility coefficient of each measured parameter are shown in Tables 2 (normal group) and 3 (glaucoma group). Low measurement variability (low values of reproducibility coefficient) was evident in most of the parameters in normal subjects and in those with glaucoma. In the normal group, the reproducibility coefficient of RNFL measurements ranged from to m for OCT and from 3.60 to 6.41 m for GDx VCC. The reproducibility coefficient of HRT 3 rim area ranged from to mm 2. In the glaucoma group, the reproducibility coefficient of OCT and GDx VCC measurements were generally comparable with those in the normal group, whereas the reproducibility coefficients of HRT measurements were generally higher in the glaucoma group. To investigate the effect of glaucomatous damage on the measurements reproducibility, the association between within-subject variance of the global measures (average RNFL thickness for OCT, TSNIT average for GDx VCC, and global rim area for HRT) and visual field MD were examined with regression analyses. Although no association was found for OCT (r [95% confidence interval (CI): ], P 0.924) and GDx VCC (r [95% CI: ], P 0.756) average RNFL thicknesses, there was a significant association between within-subject variance of HRT global rim area and visual field MD (r [95% CI: to 0.273], P 0.001). Nevertheless, this association became insignificant (standardized coefficient [95% CI: ], P 0.079) when the variances of reference height (mean variance m; [95% CI: ], P 0.001) and image quality SD (mean variance m; [95% CI: ], P 0.211) were included in the multivariate regression model (R , P 0.001). The coefficient of variation, ICC, and sensitivity to change of the global measures are shown in Table 4. No significant difference in ICC was found (P 0.119). There was no association between the SD and the mean of the intersession measurements in each of the global parameters (Fig. 1). TABLE 2. The Overall Mean, Sw, and Reproducibility Coefficient of StratusOCT, HRT 3, and GDx VCC Measurements in the Normal Group Overall Mean (SD) Sw (95% CI) Reproducibility Coefficient (95% CI) StratusOCT RNFL Thickness Measurement ( m) Superior (16.23) 6.77 ( ) ( ) Nasal (14.61) 6.55 ( ) ( ) Inferior (18.15) 8.02 ( ) ( ) Temporal (15.10) 6.22 ( ) ( ) Average (11.46) 4.53 ( ) ( ) HRT 3 Rim Area Measurement (mm 2 ) Temporal (0.1099) ( ) ( ) Superotemporal (0.0445) ( ) ( ) Inferotemporal (0.0495) ( ) ( ) Nasal (0.1107) ( ) ( ) Superonasal (0.0544) ( ) ( ) Inferonasal (0.0528) ( ) ( ) Global (0.3561) ( ) ( ) GDx VCC RNFL Measurement ( m) Inferior average (8.19) 2.31 ( ) 6.41 ( ) Superior average (8.10) 2.24 ( ) 6.21 ( ) TSNIT average (5.44) 1.30 ( ) 3.60 ( ) n 45.
4 IOVS, November 2008, Vol. 49, No. 11 Longitudinal Variability of Optic Disc and RNFL Measurements 4889 TABLE 3. The Overall Mean, Sw, and Reproducibility Coefficient of StratusOCT, HRT 3 and GDx VCC Measurements in the Glaucoma Group Overall Mean (SD) Sw (95% CI) Reproducibility Coefficient (95% CI) StratusOCT RNFL Measurement ( m) Superior (23.16) 6.03 ( ) ( ) Nasal (15.90) 5.74 ( ) ( ) Inferior (27.62) 7.26 ( ) ( ) Temporal (16.13) 4.47 ( ) ( ) Average (16.44) 3.85 ( ) ( ) HRT 3 Rim Area Measurement (mm 2 ) Temporal (0.0497) ( ) ( ) Superotemporal (0.0552) ( ) ( ) Inferotemporal (0.0503) ( ) ( ) Nasal (0.1338) ( ) ( ) Superonasal (0.0581) ( ) ( ) Inferonasal (0.0566) ( ) ( ) Global (0.2866) ( ) ( ) GDx VCC RNFL Measurement ( m) Inferior average (10.92) 2.64 ( ) 7.31 ( ) Superior average (12.86) 2.36 ( ) 6.54 ( ) TSNIT average (8.25) 2.00 ( ) 5.54 ( ) n 43. DISCUSSION High reproducibility was observed in most of the parameters measured by OCT, GDx VCC, and HRT, as indicated by their relatively low reproducibility coefficients. For example, the difference in average RNFL thickness obtained in two separate visits would be expected to be less than 11.7 m in OCT and 4.7 m in GDx VCC (Table 4) for 95% of pairs of observations. Although the reproducibility coefficient is lower in GDx VCC, it does not suggest that GDx VCC measurement is more reproducible. It is notable that although both instruments express RNFL thickness in micrometers, they are in different scales. RNFL thickness measured by OCT was higher than that measured by GDx VCC 24 and thus resulted in the higher reproducibility coefficients. In fact, all the global measurements demonstrated high ICCs (all 0.97), indicating that they are reproducible. In studying progression, a parameter should be both reproducible and sensitive to change. Reproducibility refers to the ability of an instrument to provide consistent measurements. It is expressed as Sw. Jampel et al. 23 introduced sensitivity to change, defined as [(97.5 percentile value 2.5 percentile value)/2 Sw], to measure the ability of a parameter to detect change. This parameter indicates the number of progression events that can be detected in a given cohort of patients (i.e., from completely healthy to end-stage glaucoma). In other words, this parameter reflects the ability of an instrument to be clinically useful for monitoring glaucoma progression. In this regard, there may be no significant difference among the three imaging instruments. We compared only the global parameters. As the proportion of sectorial distribution of the structural measures in each imaging test is different, sectorial measurements cannot be directly compared. Conflicting results have been reported regarding the effect of glaucomatous damage on optic disc/rnfl measurement repeatability. Jampel et al. 23 found no association between the degree of glaucomatous damage (both in visual field MD and HRT rim area) and the SE of measurement of rim area, cup shape, and retinal thickness. On the contrary, DeLeon Ortega et al. 4 showed significant correlation between visual field MD and variances of OCT average RNFL thickness, GDx VCC TSNIT average, and HRT global rim area. In the present study, there was no significant association between the degree of glaucomatous damage and the variances of global RNFL/rim area measurements suggesting that longitudinal variability does not change with the severity of glaucoma. This conclusion is also supported by the finding that the SD was unrelated to the magnitude of the measurements (Fig. 1). In fact, the Sw s of the OCT and GDx VCC measurements were very similar between the normal and glaucoma groups (Tables 2, 3). Of note, the apparent association between HRT global rim area variance and visual field MD became insignificant when it was adjusted for the variances of reference height and image quality SD. These two predictors were selected in the multivariate analysis, because they were found to be the most influential for rim area variability. 8,25 In a study by Strouthidis et al., 8 50% of intravisit rim area difference was explained by reference height difference. Reference height and its relationship to reference plane are crucial in defining the measurement of the rim and TABLE 4. Comparison of Reproducibility, Coefficient of Variation, and Sensitivity to Change of the Global Measures of StratusOCT, HRT 3, and GDx VCC Reproducibility Coefficient (99% CI) Coefficient of Variation (%) (95% CI) Intraclass Correlation Coefficient (95% CI) Sensitivity to Change (95% CI) StratusOCT average RNFL thickness ( m) ( ) 4.79 ( ) 0.97 ( ) 10.2 ( ) HRT 3 global rim area (mm 2 ) ( ) 6.22 ( ) 0.97 ( ) 9.3 ( ) GDx VCC TSNIT average ( m) 4.65 ( ) 3.52 ( ) 0.98 ( ) 11.3 ( ) n 88.
5 4890 Leung et al. IOVS, November 2008, Vol. 49, No. 11 FIGURE 1. Scatterplots of SD versus the (A) mean of OCT average RNFL thickness, (B) GDx VCC TSNIT average, and (C) HRT global rim area. the cup. Change in the orientation or tilting of the optic disc surface during repeated imaging may result in a shift in the level of reference plane, causing measurement variability of the rim. This variability may be more substantial in glaucomatous disc in which the surface of the disc could be lower than the parapapillary retinal surface, resulting in a more superficially located reference plane. Even slight shifts or tilts in the images could lead to significant changes in rim area measurement. Although rim area variability is unlikely to be related to the severity of glaucomatous damage, glaucomatous optic disc per se may result in higher reference height variation, accounting for the higher rim area variability.
6 IOVS, November 2008, Vol. 49, No. 11 Longitudinal Variability of Optic Disc and RNFL Measurements 4891 Determining the reproducibility of optic disc and RNFL measurements is fundamental in differentiating measurement variability from pathologic changes in the detection of glaucoma progression. Measurement variability could be intrinsic to the instrument, or it could be due to physiological change in the cornea, lens, or retina over time. Investigating longitudinal variability is thus clinically important in deciding whether the change observed is within the limits of physiological variation. In this study, the average follow-up time for imaging was 4.4 months, which is closely matched with the average follow-up schedule for patients with glaucoma in a glaucoma clinic, and a total of three visits, spanning a period of 8.8 months, were used in the calculation of reproducibility. The study was designed in a way to resemble clinical practice in which imaging tests are usually repeated over several months (and not in a day or a few weeks), and yet in a relatively short period (8.8 months), to minimize potential pathologic changes. The observations that the mean measurements remained unchanged (all with P 0.239, repeated-measures ANOVA) and the Sw s of RNFL thicknesses measured by OCT and GDx VCC were comparable between normal subjects and those with glaucoma support the assumption that no progression occurred over the study period. HRT defines rim area by the area bounded by the disc margin and the cup margin. Although the disc margin was automatically exported between visits, thus eliminating the variability of disc area measurement, cup area measurement varied between visits because of the variation of the HRT reference height. Given that the cup area is measured where the reference plane intersects the optic disc on the z-axis, any axial shifts between the optic disc and reference plane due to variations of optic disc orientation, scanning alignment, and centering of optic disc could lead to variability of cup and rim area measurements. Because each imaging device has its unique features and limitations, the current findings could provide a useful reference for clinicians to compare the measurement variability of OCT, CSLO, and SLP when the scan quality has been optimized. It is notable that although ICC may be useful to reflect the relative measurement reliability among different imaging instruments in the same study population, it cannot be compared in different studies because of population heterogeneity. Our study is limited in that we analyzed only quadrant and average RNFL thicknesses from the OCT, TSNIT and superior/ inferior RNFL average from the GDx VCC, and sectorial and global rim area measurements from the HRT. These parameters were selected because they represent the most commonly used indicators in the evaluation of glaucoma. The measurement of neuroretinal rim area to differentiate normal subjects from those with glaucoma has been advocated The data from the nerve fiber indicator (NFI) of the GDx VCC were not analyzed, because this parameter was calculated based on different variables and not designed for serial monitoring. All the images taken in this study were obtained by highly trained and experienced technicians in an imaging research unit with high image-quality scores and consistent image centration. Results obtained in this study may not be directly translated to other populations in different clinical settings. Although intraocular pressure reduction could affect topographical measurements of the optic disc, 30 it is controversial whether there is a significant change in RNFL thickness with the lowering of intraocular pressure. 31,32 In summary, longitudinal RNFL and rim measurements obtained from the StratusOCT, GDx VCC, and HRT 3 have high reproducibility. The observation that the measurement variability was independent of the severity of glaucoma suggests that serial monitoring of these measurements is useful to detect glaucoma progression. A longitudinal study with longer follow-up time is needed to compare the performance of these imaging instruments in detection of glaucoma progression. References 1. Schuman JS, Hee MR, Puliafito CA, et al. Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography. Arch Ophthalmol. 1995;113: Budenz DL, Michael A, Chang RT, et al. Sensitivity and specificity of the StratusOCT for perimetric glaucoma. Ophthalmology. 2005; 112: Reus NJ, de Graaf M, Lemij HG. Accuracy of GDx VCC, HRT I, and clinical assessment of stereoscopic optic nerve head photographs for diagnosing glaucoma. Br J Ophthalmol. 2007;91: Deleon-Ortega JE, Arthur SN, McGwin G Jr, et al. Discrimination between glaucomatous and nonglaucomatous eyes using quantitative imaging devices and subjective optic nerve head assessment. Invest Ophthalmol Vis Sci. 2006;47: Leung CK, Medeiros FA, Zangwill LM, et al. American Chinese glaucoma imaging study: a comparison of the optic disc and retinal nerve fiber layer in detecting glaucomatous damage. Invest Ophthalmol Vis Sci. 2007;48: British Standards Institution. Accuracy (Trueness and Precision) of Measurement Methods and Results: General Principles and Definitions. London: HMO; BS ISO 5725 part 1; British Standards Institution. Accuracy (Trueness and Precision) of Measurement Methods and Results: Basic Methods for the Determination of Repeatability and Reproducibility of a Standard Measurement Method. London: HMO; BS ISO 5725 part 2; Strouthidis NG, White ET, Owen VM, et al. Factors affecting the test-retest variability of Heidelberg retina tomograph and Heidelberg retina tomograph II measurements. Br J Ophthalmol. 2005; 89: Paunescu LA, Schuman JS, Price LL, et al. Reproducibility of nerve fiber thickness, macular thickness, and optic nerve head measurements using StratusOCT. Invest Ophthalmol Vis Sci. 2004;45: Budenz DL, Chang RT, Huang X, et al. Reproducibility of retinal nerve fiber thickness measurements using the stratus OCT in normal and glaucomatous eyes. Invest Ophthalmol Vis Sci. 2005; 46: Wu Z, Vazeen M, Varma R, et al. Factors associated with variability in retinal nerve fiber layer thickness measurements obtained by optical coherence tomography. Ophthalmology. 2007;114(8): DeLeon Ortega JE, Sakata LM, Kakati B, et al. Effect of glaucomatous damage on repeatability of confocal scanning laser ophthalmoscope, scanning laser polarimetry, and optical coherence tomography. Invest Ophthalmol Vis Sci. 2007;48: Frenkel S, Slonim E, Horani A, et al. Operator learning effect and interoperator reproducibility of the scanning laser polarimeter with variable corneal compensation. Ophthalmology. 2005;112: Gurses-Ozden R, Teng C, Vessani R, et al. Macular and retinal nerve fiber layer thickness measurement reproducibility using optical coherence tomography (OCT-3). J Glaucoma. 2004;13: Hoh ST, Ishikawa H, Greenfield DS, et al. Peripapillary nerve fiber layer thickness measurement reproducibility using scanning laser polarimetry. J Glaucoma. 1998;7: Iacono P, Da Pozzo S, Fuser M, et al. Intersession reproducibility of retinal nerve fiber layer thickness measurements by GDx-VCC in healthy and glaucomatous eyes. Ophthalmologica. 2006;220(4): Medeiros FA, Zangwill LM, Bowd C, et al. Comparison of scanning laser polarimetry using variable corneal compensation and retinal nerve fiber layer photography for detection of glaucoma. Arch Ophthalmol. 2004;122: Musch DC, Lichter PR, Guire KE, et al. The Collaborative Initial Glaucoma Treatment Study: study design, methods, and baseline characteristics of enrolled patients. Ophthalmology. 1999;106: Bagga H, Greenfield DS, Feuer WJ. Quantitative assessment of atypical birefringence images using scanning laser polarimetry with variable corneal compensation. Am J Ophthalmol. 2005;139: Bland JM, Altman DG. Measurement error. BMJ. 1996;313:744.
7 4892 Leung et al. IOVS, November 2008, Vol. 49, No Fisher RA. Statistical Methods for Research Workers. Darien, CT: Hafner Publishing Company; Stratford PW, Spadoni GF. Sample size estimation for the comparison of competing measures reliability coefficients. Physiother Can. 2003;55: Jampel HD, Vitale S, Ding Y, et al. Test-retest variability in structural and functional parameters of glaucoma damage in the glaucoma imaging longitudinal study. J Glaucoma. 2006;15: Leung CK, Chan WM, Chong KL, et al. Comparative study of retinal nerve fiber layer measurement by StratusOCT and GDx VCC: (1) Correlation analysis in glaucoma. Invest Ophthalmol Vis Sci. 2005; 46: Tan JC, Garway-Heath DF, Fitzke FW, et al. Reasons for rim area variability in scanning laser tomography. Invest Ophthalmol Vis Sci. 2003;44: Britton RJ, Drance SM, Schulzer M, et al. The area of the neuroretinal rim of the optic nerve in normal eyes. Am J Ophthalmol. 1987;103: Caprioli J, Miller JM. Optic disc rim area is related to disc size in normal subjects. Arch Ophthalmol. 1987;105: Montgomery DM. Measurement of optic disc and neuroretinal rim areas in normal and glaucomatous eyes: a new clinical method. Ophthalmology. 1991;98: Wollstein G, Garway-Heath DF, Hitchings RA. Identification of early glaucoma cases with the scanning laser ophthalmoscope. Ophthalmology. 1998;105: Tan JC, Hitchings RA. Reversal of disc cupping after intraocular pressure reduction in topographic image series. J Glaucoma. 2004;13: Chang PT, Sekhon N, Budenz DL, et al. Effect of lowering intraocular pressure on optical coherence tomography measurement of peripapillary retinal nerve fiber layer thickness. Ophthalmology. 2007;114(12): Yamada N, Tomita G, Yamamoto T, et al. Changes in the nerve fiber layer thickness following a reduction of intraocular pressure after trabeculectomy. J Glaucoma. 2000;9:
The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation
C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND
More informationCLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD
CLINICAL SCIENCES Comparison of the GDx VCC Scanning Laser Polarimeter, HRT II Confocal Scanning Laser Ophthalmoscope, and Stratus OCT Optical Coherence Tomograph for the Detection of Glaucoma Felipe A.
More informationDetection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation
Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Felipe A. Medeiros, Christopher Bowd, Linda M. Zangwill, Chirag Patel, and Robert N. Weinreb From the Hamilton
More informationOtticaFisiopatologica
Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza
More informationRelationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma
Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Reza Zarei, MD 1 Mohammad Soleimani, MD 2 Sasan Moghimi, MD 3 Mohammad Yaser Kiarudi, MD 2 Mahmoud Jabbarvand, MD 1 Yadollah
More informationRetinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography
Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo
More informationSTRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx
STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module
More informationScanning laser polarimetry (SLP) provides real-time, objective
The Effect of Atypical Birefringence Patterns on Glaucoma Detection Using Scanning Laser Polarimetry with Variable Corneal Compensation Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda
More informationSeiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods
Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous
More informationTHE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS
Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:
More informationS Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science
Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan Correspondence to: Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute,
More informationScanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects
접수번호 : 2008-105 Korean Journal of Ophthalmology 2009;23:169-175 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.3.169 Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve
More informationRelationship between GDx VCC and Stratus OCT in juvenile glaucoma
(2009) 23, 2182 2186 & 2009 Macmillan Publishers Limited All rights reserved 09-222X/09 $32.00 www.nature.com/eye CLINICAL STUDY Relationship between GDx VCC and Stratus OCT in juvenile glaucoma R Zareii,
More informationRetinal nerve fiber layer thickness in Indian eyes with optical coherence tomography
Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,
More informationDetection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation
Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Felipe A. Medeiros, Luciana M. Alencar, Linda M. Zangwill, Christopher
More informationOptical coherence tomography (OCT) is a noninvasive,
Ability of Stratus OCT to Detect Progressive Retinal Nerve Fiber Layer Atrophy in Glaucoma Eun Ji Lee, 1,2 Tae-Woo Kim, 1,2 Ki Ho Park, 2 Mincheol Seong, 3 Hyunjoong Kim, 4 and Dong Myung Kim 2 PURPOSE.
More informationReproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography
Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,
More informationRelationship Between Structure
Original Article Relationship Between Structure and Function of the Optic Nerve Head-Glaucoma versus Normal Dr Savita Bhat, Dr Anna Elias, Dr Siddharth Pawar, Dr S.J. Saikumar, Dr Alpesh Rajput, superior,
More informationDiagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS
Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2
More informationPosition of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC
Department of Ophthalmology, Medical University of Vienna, Austria Correspondence to Clemens Vass, Department of Ophthalmology and Optometry, Medical University of Vienna, General Hospital, Währinger Gürtel
More informationComparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements
Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,
More informationThe Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS
The Relationship between Standard Automated Perimetry and GDx VCC Measurements Nicolaas J. Reus and Hans G. Lemij PURPOSE. To investigate the relationship between retinal light sensitivity measured with
More informationCorrelating Structure With Function in End-Stage Glaucoma
C L I N I C A L S C I E N C E Correlating Structure With Function in End-Stage Glaucoma Eytan Z. Blumenthal, MD; Amjad Horani, MD; Rajesh Sasikumar, MD; Chandrasekhar Garudadri, MD; Addepalli Udaykumar,
More informationDiagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation
Diagnostic Accuracy of Scanning Laser olarimetry with Enhanced versus Variable Corneal T. A. Mai, MD, N. J. Reus, MD, hd, H. G. Lemij, MD, hd urpose: To compare the diagnostic accuracy of scanning laser
More informationRetinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence
More informationRelationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation
Korean Journal of Ophthalmology 22(1):18-25, 2008 DOI : 10.3341/kjo.2008.22.1.18 Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation
More informationIntroduction. Hemma Resch, Gabor Deak, Ivania Pereira and Clemens Vass. e225. Acta Ophthalmologica 2012
Comparison of optic disc parameters using spectral domain cirrus high-definition optical coherence tomography and confocal scanning laser ophthalmoscopy in normal eyes Hemma Resch, Gabor Deak, Ivania Pereira
More informationStructural examina.on: Imaging
ManaMa: Glaucoma Structural examina.on: Imaging Luís Abegão Pinto, MD, PhD Department of Ophthalmology CHLC Lisbon Faculty of Medicine, Lisbon University 1 11-10- 2013 Structural changes Qualitative changes
More informationParafoveal Scanning Laser Polarimetry for Early Glaucoma Detection
Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of
More informationStudy of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma
Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro
More informationEvaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)
Dada T et al Original article Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T1, Aggarwal A1, Bali SJ2, Sharma A1, Shah
More informationScanning laser polarimetry (SLP) incorporates a confocal
Scanning Laser Polarimetry with Enhanced Corneal Compensation and Optical Coherence Tomography in Normal and Glaucomatous Eyes Mitra Sehi, 1 Stephen Ume, 1 David S. Greenfield, 1 and Advanced Imaging in
More informationMEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association
MEDICAL POLICY SUBJECT: OPHTHALMOLOGIC TECHNIQUES PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical
More informationCLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography
CLINICAL SCIENCES Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography Seong Bae Park, MD; Kyung Rim Sung, MD, PhD; Sung Yong Kang, MD; Kyung Ri Kim, BS;
More informationThe only measure of RNFL Integrity: GDxPRO
The only measure of RNFL Integrity: GDxPRO Look beyond thickness. Underlying structural organization is the key to RNFL At times, stepping onto ice is a risky proposition. Just knowing its thickness might
More informationGanglion cell complex scan in the early prediction of glaucoma
Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell
More informationC a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma
C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e The GDxVCC Early answers and ongoing assessment for glaucoma The quantifiable approach to quality care Only Humphrey GPA software Early insight
More informationKey words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography
194 Review Article Evaluating the Optic Nerve and Retinal Nerve Fibre Layer: The Roles of Heidelberg Retina Tomography, Scanning Laser Polarimetry and Optical Coherence Tomography Sek-Tien Hoh, 1 MBBS,
More informationSensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7
Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 61 Abstract Purpose The GDx is a scanning laser polarimeter that assesses
More informationA comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting
(2007) 21, 1050 1055 & 2007 Nature Publishing Group All rights reserved 0950-222X/07 $30.00 www.nature.com/eye CLINICAL STUDY A comparison of HRT II and GDx imaging for glaucoma detection in a primary
More informationThe only measure of RNFL Integrity : GDxPRO
The ideal glaucoma assessment tool for your practice The only measure of RNFL Integrity : GDxPRO Proven Clinical Performance Numerous studies have validated the diagnostic accuracy and performance of the
More informationNERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS
Nerve fiber layer thickness in normals and glaucoma patients 403 NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS HIROTAKA SUZUMURA, KAYOKO HARASAWA, AKIKO KOBAYASHI and NARIYOSHI ENDO Department
More informationAdvances in OCT Murray Fingeret, OD
Disclosures Advances in OCT Murray Fingeret, OD Consultant Alcon, Allergan, Bausch & Lomb, Carl Zeiss Meditec, Diopsys, Heidelberg Engineering, Reichert, Topcon Currently Approved OCT Devices OCT Devices
More informationReliability analyses of the GDx nerve-fiber analyzer
VOL. 29 NO. 2 PHILIPPINE JOURNAL OF Ophthalmology APRIL ORIGINAL ARTICLE - JUNE 2004 1, 2, 3 Patricia M. Khu, MD, MS Edgardo U. Dorotheo, MD 1 Lawrence Tinio, MD 2 Cynthia P. Cordero, MS 4 1, 2, 3 Manuel
More informationOptic disc damage has been shown to precede visual field
Confocal Scanning Laser Ophthalmoscopy Classifiers and Stereophotograph Evaluation for rediction of Visual Field Abnormalities in Glaucoma-Suspect Eyes Christopher Bowd, 1 Linda M. Zangwill, 1 Felipe A.
More informationGlaucoma is a chronic progressive optic neuropathy,
ORIGINAL STUDY Comparison of Neuroretinal Rim Area Measurements Made by the Heidelberg Retina Tomograph I and the Heidelberg Retina Tomograph II Ya Xing Wang, MD,*w Neil O Leary, MSc,zy Nicholas G. Strouthidis,
More informationScanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure
Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Jimmy S. M. Lai*, Clement C. Y. Tham, Jonathan C. H. Chan*, Nelson K. F. Yip, Wilson W. T. Tang, Patrick S. H. Li*, Jane
More informationMEDICAL POLICY. Proprietary Information of YourCare Healthcare
MEDICAL POLICY PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy criteria are not applied.
More informationFactors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach
ORIGINAL STUDY Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach Joong Won Shin, MD, Kyung Rim Sung, MD, PhD, Jiyun
More informationInternational Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN
International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December-2013 108 Name of Chief and Corresponding Author : Dr Chandrima Paul TITLE : Comparison of glaucoma diagnostic ability
More informationReproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography
ORIGINAL STUDY Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography Keunheung Park, MD,* Jinmi Kim, PhD, and Jiwoong Lee, MD, PhD*
More informationTo assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography
American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of
More informationClinical Use of OCT in Assessing Glaucoma Progression
r e v i e w Clinical Use of OCT in Assessing Glaucoma Progression Jacek Kotowski, MD; Gadi Wollstein, MD; Lindsey S. Folio, BS; Hiroshi Ishikawa, MD; Joel S. Schuman, MD ABSTRACT Detection of disease progression
More informationLinking structure and function in glaucoma
CET CONTINUING Sponsored by 1 CET POINT Linking structure and function in glaucoma 50 Dr Samantha McGinnigle PhD, BSc (Hons), MCOptom, AHEA This article will give an overview of the latest imaging technology
More informationAssessing the Relationship between Central Corneal Thickness and Retinal Nerve Fiber Layer Thickness in Healthy Subjects
1 1 1 1 1 1 Assessing the Relationship between Central Corneal Thickness and Retinal Nerve Fiber Layer Thickness in Healthy Subjects Authors: Tarkan Mumcuoglu, 1* Kelly A Townsend, 1* Gadi Wollstein, 1
More informationEXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013
268 Comparison of optic nerve morphology in eyes with glaucoma and eyes with non-arteritic anterior ischemic optic neuropathy by Fourier domain optical coherence tomography YUXIN YANG 1, HAITAO ZHANG 1,
More informationImaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender
Bowd et al. Vol. 19, No. 1/January 2002/J. Opt. Soc. Am. A 197 Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Christopher Bowd,
More informationThe Role of the RNFL in the Diagnosis of Glaucoma
Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer
More informationRETINAL NERVE FIBER LAYER
CLINICAL SCIENCES The Effect of Scan Diameter on Retinal Nerve Fiber Layer Thickness Measurement Using Stratus Optic Coherence Tomography Giacomo Savini, MD; Piero Barboni, MD; Michele Carbonelli, MD;
More informationAbility of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population
Korean J Ophthalmol Vol. 18:1-8, 2004 Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Sun Young Lee, MD,
More informationCLINICAL SCIENCES. Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma
CLINICAL SCIENCES Baseline Topographic Optic Disc Measurements Are Associated With the Development of Primary Open-Angle Glaucoma The Confocal Scanning Laser Ophthalmoscopy Ancillary Study to the Ocular
More informationG laucoma is a progressive optic neuropathy in which
1135 EXTENDED REPORT The correlation between optic nerve head topographic measurements, peripapillary nerve fibre layer thickness, and visual field indices in glaucoma Y-W Lan, D B Henson, A J Kwartz...
More informationEye Movements, Strabismus, Amblyopia, and Neuro-Ophthalmology
Eye Movements, Strabismus, Amblyopia, and Neuro-Ophthaology Scanning Laser Polarimetry, but Not Optical Coherence Tomography Predicts Permanent Visual Field Loss in Acute Nonarteritic Anterior Ischemic
More informationPeripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept.
Peripapillary Retinal Thickness Maps in the Evaluation of Glaucoma Patients: A Novel Concept The Harvard community has made this article openly available. Please share how this access benefits you. Your
More informationNoel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD
Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database
More informationA Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(5):369-377 http://dx.doi.org/10.3341/kjo.2012.26.5.369 Original Article A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT)
More informationAvailable online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):
Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve
More informationReproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography
ORIGINAL STUDY Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography Huijuan Wu, MD, PhD,*w Johannes F. de Boer, PhD,z and Teresa C. Chen,
More informationMethod for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.
Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Donald C. Hood 1,2,* and Ali S. Raza 1 1 Department of Psychology, Columbia
More informationNIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01.
NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2012 September ; 130(9): 1107 1116. doi:10.1001/archophthalmol.2012.827. A Combined Index of Structure and Function
More informationComparison of Optic Disc Topography Measured by Retinal Thickness Analyzer with Measurement by Heidelberg Retina Tomograph II
Comparison of Optic Disc Topography Measured by Retinal Analyzer with Measurement by Heidelberg Retina Tomograph II Noriko Itai*, Masaki Tanito*, and Etsuo Chihara* *Senshokai Eye Institute, Uji, Kyoto,
More informationTranslating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects
Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects
More informationAdvances in the Structural Evaluation of Glaucoma with Optical Coherence Tomography
Curr Ophthalmol Rep (2013) 1:98 105 DOI 10.1007/s40135-013-0014-4 DIAGNOSIS AND MONITORING OF GLAUCOMA (S SMITH, SECTION EDITOR) Advances in the Structural Evaluation of Glaucoma with Optical Coherence
More informationClinical decision making based on data from GDx: One year observations
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2002 Clinical decision making based on data from GDx: One year observations James C. Bobrow Washington University
More informationCLINICAL SCIENCES. optic neuropathy characterized
CLINICAL SCIENCES Spectral-Domain Optical Coherence Tomography for Detection of Localized Retinal Nerve Fiber Layer Defects in Patients With Open-Angle Glaucoma Na Rae Kim, MD; Eun Suk Lee, MD, PhD; Gong
More information* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD
Journal of Glaucoma 10:163 169 2001 Lippincott Williams & Wilkins, Inc. Influence of Scan Radius Correction for Ocular Magnification and Relationship Between Scan Radius With Retinal Nerve Fiber Layer
More informationNIH Public Access Author Manuscript Ophthalmology. Author manuscript; available in PMC 2009 October 4.
NIH Public Access Author Manuscript Published in final edited form as: Ophthalmology. 2008 August ; 115(8): 1352 1357.e2. doi:10.1016/j.ophtha.2008.01.011. Combining Nerve Fiber Layer to Optimize Glaucoma
More informationCLINICAL SCIENCES. Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography
CLINICAL SCIENCES Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography Antonio Polito, MD; Michele Del Borrello, MD; Miriam Isola, MHS; Nicola Zemella,
More informationRetinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer
European Journal of Ophthalmology / Vol. 15 no. 2, 2005 / pp. 246-254 Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer M. IESTER 1,2, A. MERMOUD 1 1 Hopital Ophtalmique
More informationMacular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma
Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma
More informationCirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis
Glaucoma Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis Javier Moreno-Montañés, Natalia Olmo, Aurora Alvarez, Noelia García,
More informationScanning Laser Tomography to Evaluate Optic Discs of Normal Eyes
Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Hiroshi Nakamura,* Toshine Maeda,* Yasuyuki Suzuki and Yoichi Inoue* *Eye Division of Olympia Medical Clinic, Tokyo, Japan; Department of
More informationReproducibility of measurements and variability of the classification algorithm of Stratus OCT in normal, hypertensive, and glaucomatous patients
ORIGINAL RESEARCH Reproducibility of measurements and variability of the classification algorithm of Stratus OCT in normal, hypertensive, and glaucomatous patients Alfonso Antón 1,2,3 Marta Castany 1,2
More informationComparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(3):166-173 DOI: 10.3341/kjo.2011.25.3.166 Original Article Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis
More informationEarly detection and control of disease progression or
Glaucoma Agreement Among Spectral-Domain Optical Coherence Tomography, Standard Automated Perimetry, and Stereophotography in the Detection of Glaucoma Progression Sebastián A. Banegas, 1 Alfonso Antón,
More informationGlaucoma Diagnosis & Tracking with Optical Coherence Tomography
Glaucoma Diagnosis & Tracking with Optical Coherence Tomography David Huang, MD, PhD Charles C. Manger III, MD Chair of Corneal Laser Surgery Assoc. Prof. of Ophthalmology & Biomedical Engineering Doheny
More informationThe Measure of Confidence
Heidelberg_936357.qxd:Layout 1 5/9/08 12:01 PM 12:02 Page 1 (Cyan (Magenta (Yellow (Black (UV Five Powerful Solutions to Fit Your Practice PowerCheck Glaucoma FastCheck+ GPS Software and Retina Edema Index
More informationCorrespondence should be addressed to Verena Prokosch;
Hindawi Ophthalmology Volume 2017, Article ID 8014294, 6 pages https://doi.org/10.1155/2017/8014294 Research Article Comparison between the Correlations of Retinal Nerve Fiber Layer Thickness Measured
More informationInvestigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension
Acta Medica Anatolia Volume 2 Issue 1 2014 Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Remzi Mısır 1, Sinan Sarıcaoğlu
More informationDifferences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect
pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2015;29(6):418-423 http://dx.doi.org/10.3341/kjo.2015.29.6.418 Original Article Differences between Non-arteritic Anterior Ischemic Optic Neuropathy
More informationScreening for Glaucoma in High-Risk Populations Using Optical Coherence Tomography
Screening for Glaucoma in High-Risk Populations Using Optical Coherence Tomography Gisèle Li, MD, 1,2 Alvine Kamdeu Fansi, MD, PhD, 2 Jean-François Boivin, MD, ScD, 1 Lawrence Joseph, PhD, 1 Paul Harasymowycz,
More informationPerformance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening
Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening Boel Bengtsson, Sabina Andersson and Anders Heijl Department of Clinical Sciences, Ophthalmology in Malmo,
More informationMATERIALS AND METHODS
Glaucoma Analysis of Peripapillary Retinal Nerve Fiber Distribution in Normal Young Adults Seung Woo Hong, 1,2 Myung Douk Ahn, 2 Shin Hee Kang, 1,3 and Seong Kyu Im 1,4 PURPOSE. To determine the anatomic
More informationOPTOMETRY RESEARCH PAPER
C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY RESEARCH PAPER Interocular symmetry of retinal nerve fibre layer thickness in healthy eyes: a spectral-domain optical coherence tomographic study
More informationGlaucoma is an optic neuropathy characterized by a gradual
Optical Coherence Tomography Machine Learning Classifiers for Glaucoma Detection: A Preliminary Study Zvia Burgansky-Eliash, 1,2 Gadi Wollstein, 1,2 Tianjiao Chu, 3 Joseph D. Ramsey, 4 Clark Glymour, 4
More informationRepeatability and Reproducibility of Macular Thickness Measurements Using Fourier Domain Optical Coherence Tomography
10 The Open Ophthalmology Journal, 009, 3, 10-14 Open Access Repeatability and Reproducibility of Macular Thickness Measurements Using Fourier Domain Optical Coherence Tomography Alison Bruce 1, Ian E.
More information21st Century Visual Field Testing
Supplement to Supported by an educational grant from Carl Zeiss Meditec, Inc. Winter 2011 21st Century Visual Field Testing the Evolution Continues 21st Century Visual Field Testing 21st Century Visual
More informationFourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma
Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma Felipe A. Medeiros, Linda M. Zangwill, Christopher Bowd, Antje S. Bernd, and Robert N. Weinreb
More informationThe evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry
European Journal of Ophthalmology / Vol. 13 no. 4, 2003 / pp. 377-382 The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry
More informationResearch Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography
Ophthalmology Volume 2015, Article ID 605940, 5 pages http://dx.doi.org/10.1155/2015/605940 Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence
More information