Optical coherence tomography (OCT) is a noninvasive,

Size: px
Start display at page:

Download "Optical coherence tomography (OCT) is a noninvasive,"

Transcription

1 Ability of Stratus OCT to Detect Progressive Retinal Nerve Fiber Layer Atrophy in Glaucoma Eun Ji Lee, 1,2 Tae-Woo Kim, 1,2 Ki Ho Park, 2 Mincheol Seong, 3 Hyunjoong Kim, 4 and Dong Myung Kim 2 PURPOSE. To evaluate the ability of Stratus optical coherence tomography (OCT version 4.0; Carl Zeiss Meditec, Inc., Dublin, CA) to detect progressive glaucomatous retinal nerve fiber layer (RNFL) atrophy observed by red-free RNFL photography. METHODS. Intersession test retest variability of each clock hour, quadrant, and average RNFL thickness was determined in 53 control subjects. The sensitivity and specificity of OCT for identification of progressive RNFL atrophy were tested on subjects in whom this condition was clearly observed in redfree RNFL photographs (n 27) and in another control group (n 62), according to criteria derived from test retest variability. RESULTS. The sensitivity of Stratus OCT RNFL measurement ranged from 14.8% (for average RNFL thickness) to 85.2% (for clock hour thickness) when tested at the 95% confidence level. The specificity of Stratus OCT RNFL measurement was approximately 95% for average RNFL thickness, but decreased considerably with clock hour (59.7%) and quadrant thickness (77.4%). This is presumably because multiple testing was used for multiple clock hours and quadrants. When calculated based on two consecutive follow-up examinations, the specificity for the clock hour measurements increased to 86.6% and that for quadrant thickness increased to 92.5%. The OCT-measured RNFL thickness change showed excellent topographic agreement with the progressive RNFL atrophy observed using RNFL photography. CONCLUSIONS. Within the limits of retest variability, Stratus OCT detects progressive RNFL atrophy with high sensitivity and moderate specificity in cases showing localized progressive loss of retinal nerve fibers in red-free photographs. The specificity can be improved by use of multiple measurements. Stratus OCT is a potentially useful technique for detection of glaucoma progression. (Invest Ophthalmol Vis Sci. 2009;50: ) DOI: /iovs From the 1 Department of Ophthalmology, Seoul National University Bundang Hospital, Seongnam, Korea; the 2 Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Korea; the 3 Department of Ophthalmology, Hanyang University College of Medicine, Guri Hospital, Guri, Korea; and the 4 Department of Applied Statistics, Yonsei University, Seoul, Korea Supported by a grant from the Seoul National University Bundang Hospital Research Fund. Submitted for publication January 4, 2008; revised June 14 and August 11, 2008; accepted December 8, Disclosure: E.J. Lee, None; T.-W. Kim, None; K.H. Park, None; M. Seong, None; H. Kim, None; D.M. Kim, None The publication costs of this article were defrayed in part by page charge payment. This article must therefore be marked advertisement in accordance with 18 U.S.C solely to indicate this fact. Corresponding author: Tae-Woo Kim, Department of Ophthalmology, Seoul National University Bundang Hospital, Seoul National University College of Medicine, 300 Gumi-dong, Bundang-gu, Seongnam , Korea; twkim7@snu.ac.kr. Optical coherence tomography (OCT) is a noninvasive, high-resolution, cross-sectional imaging technique that allows in vivo measurement of tissue thickness. 1 The Stratus OCT (Carl Zeiss Meditec, Inc., Dublin, CA) is a third-generation machine that has a resolution of 8 to 10 m and is capable of differentiating between healthy eyes and eyes with glaucoma. 2 6 Several recent studies have demonstrated the high sensitivity and specificity of the Stratus OCT with its internal normative database for diagnosing glaucoma. 7,8 New-generation imaging devices such as the Stratus OCT must be able to detect progressive changes if it is to assist in better management of patients with glaucoma. Currently, limited information is available regarding the ability of OCT to detect progressive glaucomatous change. Wollstein et al. 9 compared OCT and automated perimetry for detection of glaucoma progression. They suggested that OCT is more sensitive but could not exclude the possibility of a higher number of OCTrelated type I errors (false positives). Sehi et al. 10 reported that both OCT and scanning laser polarimetry were capable of documenting and measuring progressive glaucomatous RNFL atrophy. However, that study involved only one case. Moreover, that patient had not been compliant with antiglaucoma therapy for 4 years, resulting in an intraocular pressure 30 mm Hg, which led to marked visual field defect progression. Therefore, although that study demonstrated that OCT has the potential to detect progressive glaucomatous RNFL atrophy, it remains unclear whether this new imaging device can accurately detect small progressive changes. This is a critical issue because early detection of progressive change and proper management provide the best prospect for preventing significant visual function loss since glaucomatous damage is irreversible. In the present study, we investigated the ability of Stratus OCT to detect small progressive glaucomatous changes. METHODS The study was approved by the Seoul National University Bundang Hospital institutional review board, and conformed to the Declaration of Helsinki. Informed consent was obtained from all subjects. For the purpose of the present study, three subject groups meeting defined eligibility requirements were used: (1) a stable glaucoma group to establish test retest limits, (2) a stable glaucoma group to test OCT specificity at baseline and at one or two follow-ups, and (3) a progressive glaucoma group to test OCT sensitivity at baseline and at one or two follow-ups. Patients with Stable Glaucoma Eligible patients with glaucoma were consecutively enrolled in the study during regular follow-up visits and were divided into two groups: control group 1, whose data were used to measure OCT reproducibility, and control group 2, whose data were used to evaluate OCT specificity in the detection of progressive glaucomatous RNFL atrophy. Only patients with IOP 18 mm Hg on the last three consecutive visits were included. Exclusion criteria were best corrected visual acuity worse than 20/40; spherical refraction greater than 5.0 D; cylinder correction greater than 3.0 D; history of ocular surgery (including Investigative Ophthalmology & Visual Science, February 2009, Vol. 50, No Copyright Association for Research in Vision and Ophthalmology

2 IOVS, February 2009, Vol. 50, No. 2 Stratus OCT in Progressive Glaucoma 663 FIGURE 1. Determination of the clock hour location of the expanded defect on red-free fundus photographs. A clock face circle was placed around the optic nerve head, the location and the diameter of which corresponded as closely as possible to the circle displayed in the video mode of the OCT RNFL thickness analysis report (inset). The clock hour location of the expanded portion of the defect was then determined. Note the progression in the 7 o clock sector in this patient. To determine the angular width of the expansion, the two points where the borders of the defect met the circle (small arrows) were then identified, and the directional angle of the border points with respect to the center of the circle was then measured. The angular width of the expansion was calculated based on that measurement. Expansion of the defect was more obvious in the more peripheral area (large arrows). cataract extraction between examinations); or presence of a disease that could affect the peripapillary area (where OCT measurements are obtained). All subjects were assessed with the peripapillary Fast RNFL program of the Stratus OCT, either twice for control group 1 or three times for control group 2; all measurements were 3 to 30 days apart. The same Stratus OCT instrument was used by the same operator for all testing sessions. Patients with Progressive Glaucomatous Change Eyes showing progressive glaucomatous change that met the eligibility criteria were consecutively enrolled from a database of patients examined for glaucoma between February 2006 and April 2008 in the Department of Ophthalmology, Seoul National University Bundang Hospital. Before the study, all patients underwent a complete ophthalmic examination including visual acuity, refraction, slit-lamp biomicroscopy, gonioscopy, Goldmann applanation tonometry, and dilated stereoscopic examination of the optic disc. They also underwent red-free fundus photography, OCT and standard automated perimetry (SAP), each of which was performed by the same operator. Patients having a localized RNFL defect and showing definite progressive RNFL atrophy between visits according to red-free RNFL photography were enrolled. To be included, patients should have OCT results obtained taken within 1 month of red-free photography on each occasion. Exclusion criteria were a best corrected visual acuity worse than 20/40, a spherical refraction greater than 5.0 D, cylinder correction greater than 3.0 D, a history of ocular surgery including cataract extraction between examinations, and the presence of disease that may affect the peripapillary area (where OCT measurements are obtained). The sensitivity of OCT for detection of progressive glaucomatous atrophy was tested by single and repeated testing. When a patient underwent only one OCT examination after the first detection of a progressive change in the RNFL photography, the patient was invited into the study and underwent repeated OCT examinations. Defining Progression Progression was primarily determined based on red-free RNFL photographs, which were acquired with a digital camera (EOS D60; Canon, Utsunomiyashi, Tochigiken, Japan) after maximum pupil dilation. Sixty degree, wide-angle views of the optic disc focusing on the retina with the built-in split-lines focusing device and centered between the fovea and the optic disc 11 were obtained and reviewed on an LCD monitor. Red-free RNFL photographs were independently evaluated by two authors (EJL and TWK) who were masked to the patient s clinical information including OCT results. Each examiner classified eyes into one of the following three categories: no change, progressed, or ambiguous (which may have been due to poor image quality or a change too subtle to be considered progression). Progression was defined as clearly visible widening of the preexisting localized defect borders according to red-free photography and the development of new localized defects. The quadrant where the progression was detected was recorded. Subjects were enrolled when both observers classified the defect as progression, and identified it in the same quadrant. Any discrepancy between the observers was resolved by consensus. After patients were enrolled, the two observers together reviewed the RNFL photographs to agree on a clock location of the progression, which enabled a topographic comparison with the OCT-measured RNFL thickness change. For this assessment, the directional angle of the expanded RNFL defect was measured in a way that allowed alignment of the OCT images with the RNFL photographs. This method has been described in detail elsewhere. 8,12,13 Briefly, a clock-face circle was drawn around the optic nerve head on the red-free photograph, the diameter and location of which corresponded as closely as possible to the circle displayed in the video mode of the RNFL thickness analysis report (Fig. 1). The expanded portion of the defect was identified and its clock hour location was determined. To determine the angular width of the expansion, the two points where the borders of the defect met the circle (small arrows in Fig. 1) were then identified, and the directional angle of the border points with respect to the center of the circle was then measured. The angular width of the expansion was calculated based on that measurement. The clock hour was assessed in a clockwise direction for right eyes and a

3 664 Lee et al. IOVS, February 2009, Vol. 50, No. 2 TABLE 1. Demographic Characteristics of Study Subjects Subjects with Progression (n 27) Control Group 1 (n 53) Control Group 2 (n 62) Age (y), mean SD Male/female 13/14 26/7 31/31 MD (db), mean SD PSD (db), mean SD counterclockwise direction for left eyes, with the temporal sector at 9 o clock to correspond with the OCT. Optical Coherence Tomography Subjects were assessed using the peripapillary Fast RNFL program of the Stratus OCT (Carl-Zeiss Meditec, Inc.) after a pupillary dilation to a minimum diameter of 5 mm. The imaging lens was positioned 1 cm from the eye to be examined and adjusted independently until the retina was in focus. The internal fixation target was used owing to its higher reproducibility. 2 Satisfactory quality was defined as: good centration on the optic disc and a signal strength 6 (10 maximum). Data were analyzed (Stratus version 4.0 software; Carl-Zeiss Meditec, Inc.). With the Fast RNFL program, RNFL thickness was determined at 256 points at a set diameter (3.4 mm) around the center of the optic disc three times during a single scan. These values were averaged to yield 12-clock-hour thicknesses (represented by clock hours, with the 9-o clock (3-o clock for the left eyes) value representing the segment between 345 and 15 for the right eyes), four-quadrant thicknesses, and a global average RNFL thickness measurement (360 measure). Data Analysis For control group 1, we obtained data on intraclass correlation coefficients (ICCs), coefficients of variation (COVs), and inter-session test retest variability for clock hour thickness, quadrant thickness, and 360 average thickness. The ICC was defined as the ratio of the intersubject component of variance to the total variance. For the purpose of the present study, intersession test retest variability was defined at the 95% and 80% confidence level. It was calculated as 1.96-fold greater than the intervisit SD for the 95% level 14 and was calculated as fold greater than the intervisit SD for the 80% level. For control group 2, we measured OCT specificity. The sensitivity and specificity of the Stratus OCT were tested using the following parameters (each of which had an upper 95% confidence limit of test retest variability defined at the 95% and 80% level): (1) RNFL thickness decrease beyond the limit in more than 1 clock hour; (2) RNFL thickness decrease beyond the limit in more than one quadrant; and (3) 360 average RNFL thickness decrease beyond the limit. The number of subjects and measurements allowed us to establish the lower 95% confidence limit for an ICC of 0.85 as not lower than 0.75, which is considered an acceptable lower cutoff level for good reproducibility. 15 Statistical analyses were performed with commercial software (SPSS, ver. 15.0; SPSS Inc, Chicago, IL) and R package ( P 0.05 was considered to indicate a significant difference. RESULTS Subject Characteristics We evaluated 692 eyes of 380 subjects by using red-free photography, stereo disc photography (SDP), SAP-SITA, and Stratus OCT at least twice. All eyes had localized RNFL defects in the baseline red-free photographs and met the other inclusion criteria. Based on RNFL photographs, we identified progressive change in 29 eyes of 29 patients and no progression in 571 eyes of 323 subjects. In 92 eyes of 61 subjects, the diagnosis was ambiguous owing to poor-quality photography (89 eyes of 59 subjects) or to very subtle progression (3 eyes of 2 subjects). After the first grading, there was disagreement between the two observers on approximately 21 eyes. Disagreements regarding no change versus ambiguity were the most common (n 19), followed by progression versus ambiguity (n 2). Cases in which there was disagreement regarding no change versus ambiguity were finally classified as either no change or ambiguity and excluded from further analysis. Cases in which there was disagreement regarding progression versus ambiguity were classified as ambiguous and also excluded from further analysis. In cases initially classified as progression by both graders, both agreed on the location of the defect expansion or of the new defect. Of 29 eyes classified as progressive, either the baseline or follow-up OCT scans from 2 (6.9%) subjects were deemed unacceptable owing to an inadequate signal strength ( 6 db) or to the presence of an uncentered circular ring around the optic disc. These measurements were excluded from further analyses. The final sample consisted of 27 eyes of 27 patients. The mean age of the 27 patients who showed progressive change in RNFL photographs was years, and 14 were women. The visual field mean deviation and pattern standard deviation at baseline were db and db, respectively (Table 1). Of the 27 eyes, 26 showed expansion of one preexisting RNFL defect, and 1 eye showed expansion of two preexisting RNFL defects. The mean angular width of the expanded defect was (range, ) on red-free photography. The mean interval between the baseline red-free photography and the follow-up red-free photography where the progressive RNFL atrophy was noted was months. Although progression was determined primarily using RNFL photographs, progression was also observed in 19 (70.4%) eyes by using SDP, and in 14 (51.9%) eyes with SAP (according to criteria suggested by the European Glaucoma Society, 16 which was modified from the Hoddap-Parrish-Anderson criteria 17 ). Sixteen eyes showed progression at 7 o clock, four eyes at 11 o clock, two eyes each at 6 and 8 o clock, and one eye each at 5 and 10 o clock. One eye showed progression at both 7 and 11 o clock (Fig. 2). We enrolled 115 patients, the first 53 consecutive patients in control group 1 and the next 62 patients in control group 2. The mean age of control group 1 was years and that of control group 2 was years. The visual field mean deviation and pattern standard deviation at baseline were and , respectively, for control group 1, and and , respectively, for control group 2 (Table 1). Reproducibility of OCT RNFL Thickness Measurements Table 2 shows the intersession ICCs, COVs, and test retest variability for RNFL tests performed on two separate days during a 3- to 30-day period. The test retest variability defined at the 95% level for average thickness was 6.4 m. The quad-

4 IOVS, February 2009, Vol. 50, No. 2 Stratus OCT in Progressive Glaucoma 665 FIGURE 2. Frequency distribution of progressive glaucomatous change in 27 eyes according to RNFL photography. rant and clock-hour measurements had higher test retest variability. The test retest variabilities of the temporal and inferior quadrant seemed to be less than that of the nasal and superior quadrant. Sensitivity and Specificity of OCT RNFL Thickness Parameters for Detection of Glaucoma Progression The OCT global average thickness after progression detected in photographs was m, which was significantly less than the baseline measurement of m (P ). Table 3 shows the sensitivity and specificity of OCT, which were tested using criteria corresponding to the upper 95% limit of test retest variability defined at the 95% level. The sensitivity was highest for the criterion based on clock-hour thickness (85.2% [CI, %]) and lowest (14.8% [CI, %]) for the criterion based on average RNFL thickness. The specificity was highest (98.4% [CI, %]) for the criterion based on average RNFL thickness, followed by quadrant thickness (77.4% [CI, %]) and clock-hour thickness (59.7% [CI, %]). The specificities for the criteria based on clock hour and quadrant measurement increased to 86.6% (CI, %) and 92.5% (CI, %), respectively, when calculations were based on the two consecutive follow-up examinations. Table 4 shows the sensitivity and specificity of OCT using criteria corresponding to the upper 95% limit of test retest variability defined at the 80% confidence level. Again, the sensitivity was highest for the clock hour criterion (100.0% [CI, %]), followed by quadrant (92.6% [CI, %]) and average RNFL thickness criteria (40.7% [CI, %]). The specificity was highest for the average RNFL thickness (85.5% [ %]), followed by quadrant (41.9% [CI, %]) and clock hour (24.2% [CI, %]). The sensitivity of OCT RNFL thickness measurement showed a marginally significant difference depending on the degree of RNFL thickness expansion. For 12 subjects with expansion less than 10, the sensitivity was 75.0%. For 15 subjects with expansion greater than or equal to 10, the sensitivity was 93.3% (P 0.096; Table 5). Topographic Relationship between OCT-Measured Clock-Hour RNFL Thickness Change and Progressive RNFL Atrophy, According to RNFL Photography Of the 23 subjects whose clock hour thickness progressed, we noted progression in the OCT in the same clock hour as in the RNFL photography for 17 patients. Six patients exhibited progressive changes in clock hour that was not observed in the red-free photographs; in one of whom the clock hour determined as having progressed in the OCT examination was just next to the clock hour noted as having expanded in the red-free photography (Table 6). DISCUSSION It has been reported that there is a greater likelihood of detecting glaucomatous progression by OCT than by automated perimetry. 9 However, there is limited information regarding the ability of OCT to detect progressive glaucomatous change. TABLE 2. Intersession Correlations and Test Retest Variability of Fast RNFL Measurements Location ICC* COV Test-Retest Variability 95 ( m) Test-Retest Variability 80 ( m) Global average 0.88 (0.80) ( ) 4.2 ( ) Temporal quadrant 0.86 (0.76) ( ) 5.7 ( ) Superior quadrant 0.91 (0.85) ( ) 6.5 ( ) Nasal quadrant 0.83 (0.73) ( ) 6.9 ( ) Inferior quadrant 0.95 (0.92) ( ) 5.2 ( ) 9 o clock 0.86 (0.76) ( ) 5.4 ( ) 10 o clock 0.91 (0.85) ( ) 7.4 ( ) 11 o clock 0.94 (0.89) ( ) 8.6 ( ) 12 o clock 0.89 (0.82) ( ) 8.1 ( ) 1 o clock 0.80 (0.68) ( ) 9.6 ( ) 2 o clock 0.80 (0.68) ( ) 9.3 ( ) 3 o clock 0.79 (0.66) ( ) 7.6 ( ) 4 o clock 0.83 (0.72) ( ) 8.3 ( ) 5 o clock 0.91 (0.85) ( ) 6.9 ( ) 6 o clock 0.95 (0.92) ( ) 6.5 ( ) 7 o clock 0.97 (0.94) ( ) 6.6 ( ) 8 o clock 0.91 (0.84) ( ) 5.6 ( ) n 53. * Shown with lower 95% confidence interval (CI) in parentheses. Test retest variability defined at the 95% level. Shown with 95% CI in parentheses. Test retest variability defined at the 80% level. Shown with 95% CI in parentheses.

5 666 Lee et al. IOVS, February 2009, Vol. 50, No. 2 TABLE 3. Sensitivity and Specificity of OCT RNFL Thickness Parameters for Detection of Progressive Glaucoma Identified on RNFL Photographs Based on the Upper 95% Limit of Test Retest Variability Defined at the 95% Confidence Level OCT Parameter Sensitivity * Sensitivity Based Specificity * Specificity Based In 1 clock hour ( ) ( ) ( ) ( ) In 1 quadrant ( ) ( ) ( ) ( ) In 360 average thickness ( ) (0 17.3) ( ) ( ) * 95% CI is in parentheses. The present study found that using specific criteria, Stratus OCT detected progressive RNFL atrophy with high sensitivity and moderate specificity. This appears to be the first relatively large study to investigate the use of Stratus OCT for detecting glaucoma progression. The reproducibility of Status OCT and several other findings of the present study are similar to those reported recently by Budenz et al. 18 They reported a test retest variability of 6.7 m for average RNFL thickness, as measured by the Fast RNFL program. The reproducibility was variable depending on the size and location of the area that was measured. Specifically, we found less test retest variability for average thickness than for quadrants and less test retest variability in quadrants than in their respective clock-hour sectors. In addition, for both studies, the measurement was least reproducible (lowest ICCs and highest COVs) in the nasal quadrant and in the respective clock-hour sectors. We tested the sensitivity and specificity of Stratus OCT for detection of glaucoma progression using criteria corresponding to the upper 95% limit of test retest variability defined at the 95% and 80% confidence level. These criteria used different thickness values for individual clock hours or quadrants. It seemed unreasonable to use identical criteria for all clock hours (e.g., 15 m decrease from the baseline measurement) or quadrants owing to the considerable variation between clock hour measurements and quadrant measurements. This was also the case when the test retest variability was converted to percentages (data not presented). For the criteria we tested, the clock-hour criterion appeared to be the most sensitive. This is somewhat contrary to our expectations based on measurements of test retest variability. The ability to detect changes due to disease depends largely on the test retest variability of measurements. When variability is low, small changes can be detected with confidence. Our test retest variability was lowest for average RNFL thickness followed by quadrant thickness, and so we expected that the criteria for average RNFL thickness would have the highest sensitivity, followed by quadrant criterion. However, this was the opposite of our observations. To explain this, we propose that glaucoma progression occurred focally, at least in subjects who had expansion of a localized RNFL defect, and this had little affect on average RNFL thickness and quadrant thickness. Indeed, the difference in average RNFL thickness between the baseline and follow-up examinations in patients with progressive RNFL atrophy ( m) was far less than the 95% upper limit of test retest variability defined at the 95% confidence level (7.9 m). Even when tested with the test retest variability defined at the 80% confidence level, the sensitivity was still approximately 40%. We expected that the RNFL thickness would change by no more than the upper 95% limit of our test retest variability defined at 95% confidence level 95% of the time (assuming there was no change in patient age or disease progression). This means that, with the criteria corresponding to the upper 95% limit of test retest variability, there should be a specificity of 95%. However, this expectation is applicable only to cases where the specificity is tested once. In the present study, the specificity for criterion based on average RNFL thickness (which needs a single test) was 98.4%. However, for the criteria based on clock-hour and quadrant thickness, the specificities were 59.7% and 77.4%, respectively. This decrease in specificity is in agreement with our statistical considerations. If parameters for each sector are set at the 95% confidence level, 12 simultaneous tests for 12 different clock-hour measurements and four simultaneous tests for four quadrant measurements may have the confidence level of 54.0% ( ) and 81.5% ( ), respectively. In clinical practice, the inherent limitations of simultaneous multiple testing can be overcome by repeat OCT testing. Given that the specificity of a diagnostic device is 60% for a single test, the specificity should increase to 84% ( ) when the test is performed twice and to 93.6% ( ) when it is performed three times. The specificity calculated from two consecutive examinations in the present study agrees with this expectation. Meanwhile, the sensitivity of OCT did not change considerably when calculated from two consecutive examina- TABLE 4. Sensitivity and Specificity of OCT RNFL Thickness Parameters for Detection of Progressive Glaucomatous Change Identified on RNFL Photography Based on the Upper 95% Limit of Test Retest Variability Defined at the 80% Confidence Level OCT Parameter Sensitivity * Sensitivity Based Specificity * Specificity Based In 1 clock hour ( ) ( ) ( ) ( ) In 1 quadrant ( ) ( ) ( ) ( ) In 360 average thickness ( ) ( ) ( ) ( ) * 95% CI is in parentheses.

6 IOVS, February 2009, Vol. 50, No. 2 Stratus OCT in Progressive Glaucoma 667 tions and maintained acceptable sensitivity for the clock-hour criterion. In the present study, test retest variability as well as sensitivity and specificity were measured in patients with early-tomoderate glaucoma. It is possible that the test retest variability determined in the present study would not be valid for evaluating the ability of OCT to detect progressive change in patients with advanced glaucoma who have a much thinner RNFL at baseline. Moreover, the test retest variability in such patients may significantly differ from that observed in the present patients. Thus, further studies may be required to evaluate the ability of OCT to detect progressive glaucomatous change in patients with advanced damage. In the present study, progressive change in glaucoma was defined primarily based on changes identified in RNFL photographs. It is common to define progression based on VF changes when studying glaucoma progression. However, VF data may fluctuate, leading to incorrect conclusions. In addition, detectable functional change does not occur simultaneously with structural changes in glaucoma Thus, it is inappropriate to consider VF change as the gold standard when investigating the ability of a diagnostic instrument to detect progressive structural change. In contrast, both OCT RNFL thickness analysis and RNFL photography evaluate the same anatomic structure, the RNFL. As such, it is reasonable to assume that any RNFL change should be detected by OCT if OCT can accurately measure RNFL thickness. Furthermore, by defining progression based on RNFL photography, we were able to evaluate the topographic relationship between the OCT-measured RNFL thickness change location and the progressive RNFL atrophy location based on red-free RNFL photography. We should note that the video image acquisition of the Stratus OCT is recorded after the actual scan has been taken and may not represent the exact location of the acquired scan. If this disparity is substantial, it may lead to a weak topographical relationship between the RNFL photography and OCT. In the present study, we noted an excellent topographical correlation in subjects who showed progressive atrophy in the photography. This observation suggests that the disparity between the actual scan and the true location of acquired scan in OCT is likely to be minimal. The present study included only eyes with localized defects and with an obvious border. Although localized RNFL defects are easily identified with red-free photography, it is sometimes difficult to define clearly the diffuse atrophy borders using RNFL photography, 19 which may lead to problems in measuring any defect expansion. As such, the present study excluded eyes with diffuse atrophy. Owing to the design of our study, our results provide information regarding the ability of OCT to detect progressive RNFL atrophy only of patients who have localized RNFL defects. The performance of OCT for the detection of glaucoma TABLE 5. Angular Width of RNFL Defect Expansion and Sensitivity of OCT Angular Width of RNFL Defect Expansion (deg)* Eyes (n) Eyes Showing RNFL Thickness Decrease (n) Sensitivity of OCT The criterion was the RNFL thickness decrease beyond the upper 95% limit of test retest variability defined at the 95% confidence level in more than 1 clock hour. * Measurement based on RNFL photography. TABLE 6. The Relationship of Clock Hour Location of OCT- Determined Progression and RNFL Expansion According to Photography When the OCT Progression Was Determined Clock Hour Location of OCT-Determined Progression progression of patients with diffuse RNFL damage remains to be determined. Progression of the RNFL defect was defined as the widening of the preexisting defect or development of a new defect. However, it is possible that progressive changes occur in a different fashion. Remnant nerve fibers in a region previously defined as defective may be further damaged without any change in the defect border (i.e., deepening of the defect). It is difficult to accurately detect this type of progression, and the inclusion of such change in the definition of progression may provide a source of bias. Thus, the present study considered only defect widening or a new defect as the criteria for progression. In eyes in which we detected progressive OCT change, we often observed the progressive change even in clock hour sectors where progression was not noted in the photography. It is possible that OCT-detected changes represent true RNFL loss that can only be detected using OCT. At present, it is impossible to test this, as there is no reference standard that can be used to definitively indicate progressive structural changes in glaucoma. A prospective longitudinal study could investigate this question. In the present study, we included only the subjects who had high-quality OCT measurements (that is, good centration on the optic disc of the scan and signal strength 6). These inclusion criteria may have biased our results toward a better performance of OCT for detection of glaucoma progression than typically occurs in clinical practice. In summary, the present study demonstrated that within the limits of retest variability, Stratus OCT can detect progressive RNFL atrophy with high sensitivity and moderate specificity in cases showing localized progressive loss of retinal nerve fibers in red-free photographs. The OCT-measured RNFL thickness change showed excellent topographic agreement with the progressive RNFL atrophy observed using RNFL photography. We also demonstrated that the specificity of Stratus OCT can be improved by repeated testing. Stratus OCT is a potentially useful method for detection of glaucoma progression. References Eyes (n) Only in the same clock hour as in the RNFL photography 4 In the same clock hour as in the RNFL photography, but also in the clock hour where progression was not noted in the RNFL photography 13 Only in the clock hour where progression was not noted in the RNFL photography 6* The criterion was set at the upper 95% limit of test retest variability defined at the 95% confidence level. * In one of the six subjects, the OCT clock hour that was determined as having progressed was just next to the clock hour noted as having expanded in the red-free photography. 1. Huang D, Swanson EA, Lin CP, et al. Optical coherence tomography. Science. 1991;254(5035): Schuman JS, Pedut-Kloizman T, Hertzmark E, et al. Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography. Ophthalmology. 1996;103(11): Schuman JS, Hee MR, Puliafito CA, et al. Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography. Arch Ophthalmol. 1995;113(5): Bowd C, Weinreb RN, Williams JM, Zangwill LM. The retinal nerve fiber layer thickness in ocular hypertensive, normal, and glauco-

7 668 Lee et al. IOVS, February 2009, Vol. 50, No. 2 matous eyes with optical coherence tomography. Arch Ophthalmol. 2000;118(1): Bowd C, Zangwill LM, Berry CC, et al. Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function. Invest Ophthalmol Vis Sci. 2001;42(9): Williams ZY, Schuman JS, Gamell L, et al. Optical coherence tomography measurement of nerve fiber layer thickness and the likelihood of a visual field defect. Am J Ophthalmol. 2002;134(4): Budenz DL, Michael A, Chang RT, et al. Sensitivity and specificity of the StratusOCT for perimetric glaucoma. Ophthalmology. 2005; 112(1): Jeoung JW, Park KH, Kim TW, et al. Diagnostic ability of optical coherence tomography with a normative database to detect localized retinal nerve fiber layer defects. Ophthalmology. 2005; 112(12): Wollstein G, Schuman JS, Price LL, et al. Optical coherence tomography longitudinal evaluation of retinal nerve fiber layer thickness in glaucoma. Arch Ophthalmol. 2005;123(4): Sehi M, Greenfield DS. Assessment of retinal nerve fiber layer using optical coherence tomography and scanning laser polarimetry in progressive glaucomatous optic neuropathy. Am J Ophthalmol. 2006;142(6): Hoyt WF, Frisen L, Newman NM. Fundoscopy of nerve fiber layer defects in glaucoma. Invest Ophthalmol. 1973;12(11): Hwang JM, Kim TW, Park KH, et al. Correlation between topographic profiles of localized retinal nerve fiber layer defects as determined by optical coherence tomography and red-free fundus photography. J Glaucoma. 2006;15(3): Kim TW, Park UC, Park KH, Kim DM. Ability of Stratus OCT to identify localized retinal nerve fiber layer defects in patients with normal standard automated perimetry results. Invest Ophthalmol Vis Sci. 2007;48(4): Medeiros FA, Doshi R, Zanwill LM, et al. Long-term variability of GDx VCC retinal nerve fiber layer thickness measurements. J Glaucoma. 2007;16(3): Fleiss JL, Levin B, Paik MC. Statistical Methods for Rates and Proportions. Hoboken, NJ: John Wiley & Sons, Inc.; European Glaucoma Society. Terminology and guidelines for glaucoma 2nd ed. Sanova, Italy: Dogma Srl; 2003: Hodapp E, Parrish R, Anderson S. Clinical Decisions in Glaucoma. St. Louis: Mosby; Budenz DL, Fredette MJ, Feuer WJ, Anderson DR. Reproducibility of peripapillary retinal nerve fiber thickness measurements with stratus OCT in glaucomatous eyes. Ophthalmology. 2008;115(4): Airaksinen PJ, Tuulonen A, Werner EB. Clinical evaluation of the optic disc and retinal nerve fiber layer. In: Ritch R, Shields MB, Krupin T, eds. The Glaucomas. Vol. 1. St. Louis: Mosby; 1996:

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2

More information

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

CLINICAL SCIENCES. optic neuropathy characterized

CLINICAL SCIENCES. optic neuropathy characterized CLINICAL SCIENCES Spectral-Domain Optical Coherence Tomography for Detection of Localized Retinal Nerve Fiber Layer Defects in Patients With Open-Angle Glaucoma Na Rae Kim, MD; Eun Suk Lee, MD, PhD; Gong

More information

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects 접수번호 : 2008-105 Korean Journal of Ophthalmology 2009;23:169-175 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.3.169 Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(4):330-336 http://dx.doi.org/10.3341/kjo.2014.28.4.330 Original Article Patterns of Subsequent Progression of Localized Retinal Nerve Fiber

More information

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography CLINICAL SCIENCES Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography Seong Bae Park, MD; Kyung Rim Sung, MD, PhD; Sung Yong Kang, MD; Kyung Ri Kim, BS;

More information

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence

More information

Relationship Between Structure

Relationship Between Structure Original Article Relationship Between Structure and Function of the Optic Nerve Head-Glaucoma versus Normal Dr Savita Bhat, Dr Anna Elias, Dr Siddharth Pawar, Dr S.J. Saikumar, Dr Alpesh Rajput, superior,

More information

-Zone Parapapillary Atrophy and the Rate of Retinal Nerve Fiber Layer Thinning in Glaucoma

-Zone Parapapillary Atrophy and the Rate of Retinal Nerve Fiber Layer Thinning in Glaucoma Glaucoma -Zone Parapapillary Atrophy and the Rate of Retinal Nerve Fiber Layer Thinning in Glaucoma Eun Ji Lee, 1,2 Tae-Woo Kim, 1,3 Robert N. Weinreb, 4 Ki Ho Park, 1 Seok Hwan Kim, 1,5 and Dong Myung

More information

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,

More information

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Reza Zarei, MD 1 Mohammad Soleimani, MD 2 Sasan Moghimi, MD 3 Mohammad Yaser Kiarudi, MD 2 Mahmoud Jabbarvand, MD 1 Yadollah

More information

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module

More information

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(3):166-173 DOI: 10.3341/kjo.2011.25.3.166 Original Article Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis

More information

OtticaFisiopatologica

OtticaFisiopatologica Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza

More information

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:

More information

Scanning laser polarimetry (SLP) provides real-time, objective

Scanning laser polarimetry (SLP) provides real-time, objective The Effect of Atypical Birefringence Patterns on Glaucoma Detection Using Scanning Laser Polarimetry with Variable Corneal Compensation Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda

More information

RETINAL NERVE FIBER LAYER

RETINAL NERVE FIBER LAYER CLINICAL SCIENCES The Effect of Scan Diameter on Retinal Nerve Fiber Layer Thickness Measurement Using Stratus Optic Coherence Tomography Giacomo Savini, MD; Piero Barboni, MD; Michele Carbonelli, MD;

More information

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan Correspondence to: Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute,

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension

Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Acta Medica Anatolia Volume 2 Issue 1 2014 Investigation of the relationship between central corneal thickness and retinal nerve fiber layer thickness in ocular hypertension Remzi Mısır 1, Sinan Sarıcaoğlu

More information

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography

Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography ORIGINAL STUDY Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using Spectral Domain Optical Coherence Tomography Huijuan Wu, MD, PhD,*w Johannes F. de Boer, PhD,z and Teresa C. Chen,

More information

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma.

Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Method for comparing visual field defects to local RNFL and RGC damage seen on frequency domain OCT in patients with glaucoma. Donald C. Hood 1,2,* and Ali S. Raza 1 1 Department of Psychology, Columbia

More information

Reproducibility of measurements and variability of the classification algorithm of Stratus OCT in normal, hypertensive, and glaucomatous patients

Reproducibility of measurements and variability of the classification algorithm of Stratus OCT in normal, hypertensive, and glaucomatous patients ORIGINAL RESEARCH Reproducibility of measurements and variability of the classification algorithm of Stratus OCT in normal, hypertensive, and glaucomatous patients Alfonso Antón 1,2,3 Marta Castany 1,2

More information

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(5):369-377 http://dx.doi.org/10.3341/kjo.2012.26.5.369 Original Article A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT)

More information

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of

More information

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD CLINICAL SCIENCES Comparison of the GDx VCC Scanning Laser Polarimeter, HRT II Confocal Scanning Laser Ophthalmoscope, and Stratus OCT Optical Coherence Tomograph for the Detection of Glaucoma Felipe A.

More information

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Available online at   Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6): Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve

More information

Comparison of macular GCIPL and peripapillary RNFL deviation maps for detection of glaucomatous eye with localized RNFL defect

Comparison of macular GCIPL and peripapillary RNFL deviation maps for detection of glaucomatous eye with localized RNFL defect Comparison of macular GCIPL and peripapillary RNFL deviation maps for detection of glaucomatous eye with localized RNFL defect Mi Jeung Kim, 1,2 Ki Ho Park, 1,2 Beong Wook Yoo, 3 Jin Wook Jeoung, 1,2 Hee

More information

MATERIALS AND METHODS

MATERIALS AND METHODS Glaucoma Analysis of Peripapillary Retinal Nerve Fiber Distribution in Normal Young Adults Seung Woo Hong, 1,2 Myung Douk Ahn, 2 Shin Hee Kang, 1,3 and Seong Kyu Im 1,4 PURPOSE. To determine the anatomic

More information

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma (2009) 23, 2182 2186 & 2009 Macmillan Publishers Limited All rights reserved 09-222X/09 $32.00 www.nature.com/eye CLINICAL STUDY Relationship between GDx VCC and Stratus OCT in juvenile glaucoma R Zareii,

More information

Clinical Use of OCT in Assessing Glaucoma Progression

Clinical Use of OCT in Assessing Glaucoma Progression r e v i e w Clinical Use of OCT in Assessing Glaucoma Progression Jacek Kotowski, MD; Gadi Wollstein, MD; Lindsey S. Folio, BS; Hiroshi Ishikawa, MD; Joel S. Schuman, MD ABSTRACT Detection of disease progression

More information

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro

More information

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach ORIGINAL STUDY Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach Joong Won Shin, MD, Kyung Rim Sung, MD, PhD, Jiyun

More information

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Felipe A. Medeiros, Luciana M. Alencar, Linda M. Zangwill, Christopher

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

Although measurements of the optic disc and retinal nerve

Although measurements of the optic disc and retinal nerve Longitudinal Variability of Optic Disc and Retinal Nerve Fiber Layer Measurements Christopher Kai-shun Leung, 1,2 Carol Yim-lui Cheung, 1 Dusheng Lin, 1,3 Chi Pui Pang, 1 Dennis S. C. Lam, 1 and Robert

More information

The Role of the RNFL in the Diagnosis of Glaucoma

The Role of the RNFL in the Diagnosis of Glaucoma Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December-2013 108 Name of Chief and Corresponding Author : Dr Chandrima Paul TITLE : Comparison of glaucoma diagnostic ability

More information

Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening

Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening Performance of time-domain and spectral-domain Optical Coherence Tomography for glaucoma screening Boel Bengtsson, Sabina Andersson and Anders Heijl Department of Clinical Sciences, Ophthalmology in Malmo,

More information

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous

More information

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Felipe A. Medeiros, Christopher Bowd, Linda M. Zangwill, Chirag Patel, and Robert N. Weinreb From the Hamilton

More information

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD

* Şükrü Bayraktar, MD, Zerrin Bayraktar, MD, and Ömer Faruk Yilmaz, MD Journal of Glaucoma 10:163 169 2001 Lippincott Williams & Wilkins, Inc. Influence of Scan Radius Correction for Ocular Magnification and Relationship Between Scan Radius With Retinal Nerve Fiber Layer

More information

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of

More information

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2015;29(6):418-423 http://dx.doi.org/10.3341/kjo.2015.29.6.418 Original Article Differences between Non-arteritic Anterior Ischemic Optic Neuropathy

More information

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T et al Original article Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T1, Aggarwal A1, Bali SJ2, Sharma A1, Shah

More information

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma

Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer Thickness Changes in Glaucoma Hindawi Ophthalmology Volume 2017, Article ID 78365, 8 pages https://doi.org/10.1155/2017/78365 Research Article The Pattern of Retinal Nerve Fiber Layer and Macular Ganglion Cell-Inner Plexiform Layer

More information

Clinical decision making based on data from GDx: One year observations

Clinical decision making based on data from GDx: One year observations Washington University School of Medicine Digital Commons@Becker Open Access Publications 2002 Clinical decision making based on data from GDx: One year observations James C. Bobrow Washington University

More information

OPTOMETRY RESEARCH PAPER

OPTOMETRY RESEARCH PAPER C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY RESEARCH PAPER Interocular symmetry of retinal nerve fibre layer thickness in healthy eyes: a spectral-domain optical coherence tomographic study

More information

Ophthalmology Department, Lozano Blesa University Hospital, c/ San Juan Bosco 15, Zaragoza, Spain 2

Ophthalmology Department, Lozano Blesa University Hospital, c/ San Juan Bosco 15, Zaragoza, Spain 2 Ophthalmology Volume 2012, Article ID 107053, 6 pages doi:10.1155/2012/107053 Clinical Study Comparison of Retinal Nerve Fiber Layer Thickness Measurements in Healthy Subjects Using Fourier and Time Domain

More information

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Jimmy S. M. Lai*, Clement C. Y. Tham, Jonathan C. H. Chan*, Nelson K. F. Yip, Wilson W. T. Tang, Patrick S. H. Li*, Jane

More information

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 61 Abstract Purpose The GDx is a scanning laser polarimeter that assesses

More information

CLINICAL SCIENCES. Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography CLINICAL SCIENCES Repeatability and Reproducibility of Fast Macular Thickness Mapping With Stratus Optical Coherence Tomography Antonio Polito, MD; Michele Del Borrello, MD; Miriam Isola, MHS; Nicola Zemella,

More information

Scanning laser polarimetry (SLP) incorporates a confocal

Scanning laser polarimetry (SLP) incorporates a confocal Scanning Laser Polarimetry with Enhanced Corneal Compensation and Optical Coherence Tomography in Normal and Glaucomatous Eyes Mitra Sehi, 1 Stephen Ume, 1 David S. Greenfield, 1 and Advanced Imaging in

More information

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal

Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 2: Optic Nerve Head and Retinal Nerve Fiber Layer

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013 268 Comparison of optic nerve morphology in eyes with glaucoma and eyes with non-arteritic anterior ischemic optic neuropathy by Fourier domain optical coherence tomography YUXIN YANG 1, HAITAO ZHANG 1,

More information

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation Diagnostic Accuracy of Scanning Laser olarimetry with Enhanced versus Variable Corneal T. A. Mai, MD, N. J. Reus, MD, hd, H. G. Lemij, MD, hd urpose: To compare the diagnostic accuracy of scanning laser

More information

OCT in the Diagnosis and Follow-up of Glaucoma

OCT in the Diagnosis and Follow-up of Glaucoma OCT in the Diagnosis and Follow-up of Glaucoma Karim A Raafat MD. Professor Of Ophthalmology Cairo University Hmmmm! Do I have Glaucoma or not?! 1 Visual Function 100% - N Gl Structure : - 5000 axon /

More information

Reliability analyses of the GDx nerve-fiber analyzer

Reliability analyses of the GDx nerve-fiber analyzer VOL. 29 NO. 2 PHILIPPINE JOURNAL OF Ophthalmology APRIL ORIGINAL ARTICLE - JUNE 2004 1, 2, 3 Patricia M. Khu, MD, MS Edgardo U. Dorotheo, MD 1 Lawrence Tinio, MD 2 Cynthia P. Cordero, MS 4 1, 2, 3 Manuel

More information

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC Department of Ophthalmology, Medical University of Vienna, Austria Correspondence to Clemens Vass, Department of Ophthalmology and Optometry, Medical University of Vienna, General Hospital, Währinger Gürtel

More information

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Korean J Ophthalmol Vol. 18:1-8, 2004 Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Sun Young Lee, MD,

More information

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 April 29.

NIH Public Access Author Manuscript Br J Ophthalmol. Author manuscript; available in PMC 2010 April 29. NIH Public Access Author Manuscript Published in final edited form as: Br J Ophthalmol. 2009 August ; 93(8): 1057 1063. doi:10.1136/bjo.2009.157875. Retinal nerve fibre layer thickness measurement reproducibility

More information

Optical coherence tomography of the retinal nerve fibre layer in mild papilloedema and pseudopapilloedema

Optical coherence tomography of the retinal nerve fibre layer in mild papilloedema and pseudopapilloedema 294 SCIENTIFIC REPORT Optical coherence tomography of the retinal nerve fibre layer in mild papilloedema and pseudopapilloedema E Z Karam, T R Hedges... Aims: To determine the degree to which optical coherence

More information

Jong Chul Han, Da Ye Choi, and Changwon Kee. 1. Introduction

Jong Chul Han, Da Ye Choi, and Changwon Kee. 1. Introduction Ophthalmology Volume 2015, Article ID 641204, 7 pages http://dx.doi.org/10.1155/2015/641204 Clinical Study The Different Characteristics of Cirrus Optical Coherence Tomography between Superior Segmental

More information

Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia

Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia Divya Singh et al Original REASEARCH 10.5005/jp-journals-10028-1223 Assessment of Retinal Nerve Fiber Layer Changes by Cirrus High-definition Optical Coherence Tomography in Myopia 1 Divya Singh, 2 Sanjay

More information

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Case 1 Normal Tension Glaucoma 53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Table 1. Right Eye Left Eye Visual acuity 6/6 6/6 Ishihara

More information

Early detection and control of disease progression or

Early detection and control of disease progression or Glaucoma Agreement Among Spectral-Domain Optical Coherence Tomography, Standard Automated Perimetry, and Stereophotography in the Detection of Glaucoma Progression Sebastián A. Banegas, 1 Alfonso Antón,

More information

Changes in Peripapillary Retinal Nerve Fiber Layer Thickness after Pattern Scanning Laser Photocoagulation in Patients with Diabetic Retinopathy

Changes in Peripapillary Retinal Nerve Fiber Layer Thickness after Pattern Scanning Laser Photocoagulation in Patients with Diabetic Retinopathy pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(3):220-225 http://dx.doi.org/10.3341/kjo.2014.28.3.220 Original Article Changes in Peripapillary Retinal Nerve Fiber Layer Thickness after

More information

Advances in OCT Murray Fingeret, OD

Advances in OCT Murray Fingeret, OD Disclosures Advances in OCT Murray Fingeret, OD Consultant Alcon, Allergan, Bausch & Lomb, Carl Zeiss Meditec, Diopsys, Heidelberg Engineering, Reichert, Topcon Currently Approved OCT Devices OCT Devices

More information

Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes

Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes Optical Coherence Tomography-Measured Nerve Fiber Layer and Macular Thickness in Emmetropic, High-Myopic and High-Hyperopic Eyes Mohammad-Mehdi Parvaresh, MD 1 Marjan Imani, MD 2 Mohsen Bahmani-Kashkouli,

More information

CLINICAL SCIENCES. (FDP) was designed to emphasize the response characteristics of the parasol

CLINICAL SCIENCES. (FDP) was designed to emphasize the response characteristics of the parasol CLINICAL SCIENCES Detecting Visual Function Abnormalities Using the Swedish Interactive Threshold Algorithm and Matrix Perimetry in Eyes With Glaucomatous Appearance of the Optic Disc Lisandro M. Sakata,

More information

Evaluation of Retinal nerve fiber layer thickness, mean deviation and visual field

Evaluation of Retinal nerve fiber layer thickness, mean deviation and visual field Evaluation of Retinal nerve fiber layer thickness, mean deviation and visual field index in progressive glaucoma. Sebastián A. Banegas,¹ Alfonso Antón, ² Antonio Morilla,² Marco Bogado, ² Eleonora M.Ayala²,

More information

Glaucoma Diagnosis. Definition of Glaucoma. Diagnosing Glaucoma. Vision Institute Annual Fall Conference

Glaucoma Diagnosis. Definition of Glaucoma. Diagnosing Glaucoma. Vision Institute Annual Fall Conference Glaucoma Diagnosis Vision Institute Annual Fall Conference Mitchell W. Dul, OD, MS, FAAO mdul@sunyopt.edu Richard J. Madonna, MA, OD, FAAO rmadonna@sunyopt.edu Definition of Glaucoma Glaucoma can be regarded

More information

Peripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept.

Peripapillary Retinal Thickness. Maps in the Evaluation of Glaucoma Patients: A Novel Concept. Peripapillary Retinal Thickness Maps in the Evaluation of Glaucoma Patients: A Novel Concept The Harvard community has made this article openly available. Please share how this access benefits you. Your

More information

Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia

Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia Thickness Changes in the Fovea and Peripapillary Retinal Nerve Fiber Layer Depend on the Degree of Myopia Sung-Won Choi, MD, Seok-Joon Lee, MD Department of Ophthalmology, Wonju Christian Hospital. Yonsei

More information

Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis

Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis Glaucoma Cirrus High-Definition Optical Coherence Tomography Compared with Stratus Optical Coherence Tomography in Glaucoma Diagnosis Javier Moreno-Montañés, Natalia Olmo, Aurora Alvarez, Noelia García,

More information

Diabetic retinopathy accounts for much of the visual impairment

Diabetic retinopathy accounts for much of the visual impairment Diabetic Macular Edema Assessed with Optical Coherence Tomography and Stereo Fundus Photography Charlotte Strøm, 1 Birgit Sander, 1 Nicolai Larsen, 1,2 Michael Larsen, 1 and Henrik Lund-Andersen 1 PURPOSE.

More information

Glaucoma: Diagnostic Modalities

Glaucoma: Diagnostic Modalities Glaucoma: Diagnostic Modalities - Dr. Barun Kumar Nayak, Dr. Sarika Ramugade Glaucoma is a leading cause of blindness in the world, especially in older people. Early detection and treatment by ophthalmologist

More information

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS The Relationship between Standard Automated Perimetry and GDx VCC Measurements Nicolaas J. Reus and Hans G. Lemij PURPOSE. To investigate the relationship between retinal light sensitivity measured with

More information

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements 586 Glaucoma Service, Department of Ophthalmology, University of Campinas, Campinas, Brazil R Lauande-Pimentel R A Carvalho H C Oliveira D C Gonçalves L M Silva V P Costa Glaucoma Service, Department of

More information

Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography

Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography ORIGINAL STUDY Reproducibility of Bruch Membrane Opening-Minimum Rim Width Measurements With Spectral Domain Optical Coherence Tomography Keunheung Park, MD,* Jinmi Kim, PhD, and Jiwoong Lee, MD, PhD*

More information

Optical coherence tomography (OCT) is a new noninvasive

Optical coherence tomography (OCT) is a new noninvasive Use of Optical Coherence Tomography to Assess Variations in Macular Retinal Thickness in Myopia Marcus C. C. Lim, 1 Sek-Tien Hoh, 1 Paul J. Foster, 1,2 Tock-Han Lim, 3,4 Sek-Jin Chew, 1 Steve K. L. Seah,

More information

The improvement in perimetric results occurring over

The improvement in perimetric results occurring over ORIGINAL STUDY Learning Effect of Humphrey Matrix Frequency Doubling Technology Perimetry in Patients With Ocular Hypertension Marco Centofanti, MD, PhD,*w Paolo Fogagnolo, MD,w Francesco Oddone, MD,w

More information

NIH Public Access Author Manuscript Ophthalmology. Author manuscript; available in PMC 2009 October 4.

NIH Public Access Author Manuscript Ophthalmology. Author manuscript; available in PMC 2009 October 4. NIH Public Access Author Manuscript Published in final edited form as: Ophthalmology. 2008 August ; 115(8): 1352 1357.e2. doi:10.1016/j.ophtha.2008.01.011. Combining Nerve Fiber Layer to Optimize Glaucoma

More information

Characterization of Peripapillary Atrophy Using Spectral Domain Optical Coherence Tomography

Characterization of Peripapillary Atrophy Using Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2010;24(6):353-359 DOI: 10.3341/kjo.2010.24.6.353 Characterization of Peripapillary trophy Using Spectral Domain Optical Coherence Tomography Original

More information

Correlating Structure With Function in End-Stage Glaucoma

Correlating Structure With Function in End-Stage Glaucoma C L I N I C A L S C I E N C E Correlating Structure With Function in End-Stage Glaucoma Eytan Z. Blumenthal, MD; Amjad Horani, MD; Rajesh Sasikumar, MD; Chandrasekhar Garudadri, MD; Addepalli Udaykumar,

More information

Joon Mo Kim 1, Ki Ho Park 2*, Seon Jeong Kim 2, Hyo Ju Jang 1, Eun Noh 3, Mi Jeung Kim 2, Tae Woo Kim 4, Dong Myung Kim 2 and Joseph Caprioli 5

Joon Mo Kim 1, Ki Ho Park 2*, Seon Jeong Kim 2, Hyo Ju Jang 1, Eun Noh 3, Mi Jeung Kim 2, Tae Woo Kim 4, Dong Myung Kim 2 and Joseph Caprioli 5 Kim et al. BMC Ophthalmology 2013, 13:67 RESEARCH ARTICLE Open Access Comparison of localized retinal nerve fiber layer defects in highly myopic, myopic, and non-myopic patients with normal-tension glaucoma:

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2013 October 01. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2012 September ; 130(9): 1107 1116. doi:10.1001/archophthalmol.2012.827. A Combined Index of Structure and Function

More information

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry European Journal of Ophthalmology / Vol. 13 no. 4, 2003 / pp. 377-382 The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: OPHTHALMOLOGIC TECHNIQUES PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical

More information

Progressive glaucomatous optic disc atrophy is characterized

Progressive glaucomatous optic disc atrophy is characterized Glaucoma Longitudinal Analysis of Progression in Glaucoma Using Spectral-Domain Optical Coherence Tomography Julia M. Wessel, 1 Folkert K. Horn, 1 Ralf P. Tornow, 1 Matthias Schmid, 2 Christian Y. Mardin,

More information

Diabetic macular edema (DME) is the most common cause. Effect of Diabetic Macular Edema on Peripapillary Retinal Nerve Fiber Layer Thickness Profiles

Diabetic macular edema (DME) is the most common cause. Effect of Diabetic Macular Edema on Peripapillary Retinal Nerve Fiber Layer Thickness Profiles Retina Effect of Diabetic Macular Edema on Peripapillary Retinal Nerve Fiber Layer Thickness Profiles Duck Jin Hwang, 1,2 Eun Ji Lee, 1 Sang Yoon Lee, 3 Kyu Hyung Park, 1 and Se Joon Woo 1 1 Department

More information

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2010 November 18.

NIH Public Access Author Manuscript Arch Ophthalmol. Author manuscript; available in PMC 2010 November 18. NIH Public Access Author Manuscript Published in final edited form as: Arch Ophthalmol. 2009 July ; 127(7): 875 881. doi:10.1001/archophthalmol.2009.145. Measurement of Local Retinal Ganglion Cell Layer

More information

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS Nerve fiber layer thickness in normals and glaucoma patients 403 NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS HIROTAKA SUZUMURA, KAYOKO HARASAWA, AKIKO KOBAYASHI and NARIYOSHI ENDO Department

More information

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

STANDARD AUTOMATED PERIMETRY IS A GENERALLY Comparison of Long-term Variability for Standard and Short-wavelength Automated Perimetry in Stable Glaucoma Patients EYTAN Z. BLUMENTHAL, MD, PAMELA A. SAMPLE, PHD, LINDA ZANGWILL, PHD, ALEXANDER C. LEE,

More information

Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography

Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence Tomography Ophthalmology Volume 2015, Article ID 605940, 5 pages http://dx.doi.org/10.1155/2015/605940 Research Article Repeatability of Perimacular Ganglion Cell Complex Analysis with Spectral-Domain Optical Coherence

More information