Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator

Size: px
Start display at page:

Download "Urine/CSF Albumin Reagent. Urine/CSF Albumin Calibrator"

Transcription

1 December 2014 PRODUCT ANNOUNCEMENT AU Clinical Chemistry Systems AU400/400 e, AU480, AU640/640 e, AU680, AU2700, AU5400, and AU5800 Product Urine/CSF Albumin Reagent Urine/CSF Albumin Reagent Urine/CSF Albumin Calibrator Kit REF B38858 B46435 B38859 Dear Beckman Coulter AU Customer: Beckman Coulter, Inc. is pleased to announce the release of a new Urine/CSF Albumin reagent in addition to a new Urine/CSF Albumin calibrator for use on the Beckman Coulter AU clinical chemistry analyzers. The new Beckman Coulter Urine/CSF Albumin reagent and calibrator will replace the current Microalbumin reagent, REF OSR6x67, and calibrator, REF ODR3024. The reason for this transition is to harmonize reagents between Beckman Coulter AU and Synchron platforms plus the added features of extended upper analytical range and increased prozone tolerance. As recommended by the National Kidney Disease Education Program - International Federation of Clinical Chemistry (IFCC) Working Group on the standardization of albumin in urine, the name of the assay will now change from Microalbumin to Urine/CSF Albumin in order to reflect current practice. The new Urine/CSF Albumin reagent is a turbidimetric method for the quantitative determination of albumin concentration in human urine and cerebrospinal fluid (CSF). The new reagent kit consists of a 2-component Urine/CSF Albumin reagent and is available in 2 kit sizes. The tests per vial for each kit size on each AU analyser are outlined below. New Reagent REF Package Configuration Analyser Tests/ Vial Test/Kit Urine/CSF Albumin B38858 R1: 4x32.6 ml R2: 4x4.4 ml AU400/400 e /480/640/ 640 e / AU2700/5400/ B58824-AA

2 New Reagent REF Package Configuration Analyser Tests/ Vial Test/Kit Urine/CSF Albumin B46435 R1: 4x54.8 ml R2: 4x7.3 ml AU400/400 e /480/640/ 640 e / AU2700/5400/ A new multi-level ready to use liquid calibrator is also available; New Calibrator REF Package Configuration Urine/CSF Albumin Calibrator Level 1, 2, 3, 4, 5 B38859 Levels 1-5: 2 ml Features of the new Urine/CSF Albumin assay include: 2-part ready to use liquid reagent for human Urine and CSF Albumin 60 day reagent on-board stability 60 day calibration stability Analytical range: Urine mg/l, CSF mg/l On-board dilution for samples exceeding the analytical range (1 in 10 dilution with deionized water or 0.9% saline, using automatic dilution function) Limit of Quantitation (LOQ): Urine 7 mg/l ( 0.7 mg/dl), CSF 10 mg/l ( 1.0 mg/dl) Prozone tolerance up to 20,000 mg/l using Data Check Parameters Reagent probe Contamination Avoidance Parameters have been developed for use with the new reagent. New sample probe Contamination Avoidance Parameters have also been developed for use when changing from Serum to Urine samples on the AU2700, AU5400, AU5800, AU480 and AU680. Recommended quality control materials for the new AU Urine/CSF Albumin assay include the Bio-Rad Liquichek Microalbumin and Spinal Fluid controls. The following marketed lots contain value assignment for the new AU Urine/CSF Albumin assay. The target values and ranges are available on the Bio-Rad website; Control REF Lots Bio-Rad Liquichek Microalbumin Controls Level 1 and Level 2 Bio-Rad Liquichek Spinal Fluid Controls Level 1 and Level 2 378, , , 96670, 96680, 96690, , 55590, 55600, 55610, The Urine and CSF Method Comparison data for Urine/CSF Albumin reagent B46435/B38858 versus Microalbumin reagent OSR6x67 on an AU5800 analyzer is attached. The new Urine/CSF Albumin assay recovers higher than the current Microalbumin assay due to the difference in B58824-AA Page 2 of 5

3 calibrator traceability. The new assay is traceable to IFCC standard CRM 470 while the current assay is traceable to a purified Albumin Standard quantified using a theoretical extinction coefficient. The Urine and CSF Method comparison data for the Beckman Coulter Urine/CSF Albumin reagent B46435/B38858 versus an Albumin test on a Siemens BN ProSpec Nephelometer is also attached. For both Urine and CSF, the new Urine/CSF Albumin reagent on the AU5800 correlates extremely well with the Siemens nephelometry method. The Urine/CSF Albumin Instructions for Use (BLB38858) will be available in multiple languages on the Technical Documents page of the Beckman Coulter website through the following link: Instrument Settings/ Parameters (BSB38858), Calibrator Value Assignment Sheets, and (Material) Safety Data Sheets (M)SDSs as well as Contamination Avoidance Parameters will also be available at the above link. Please share this information with your laboratory staff and retain this notification as part of your laboratory documentation. If you have any questions, please contact Beckman Coulter Technical Support at from the United States and Canada or from Europe or your local Beckman Coulter Representative. Thank you for your continued support of Beckman Coulter products and services. Sincerely, Kerri Breasley Senior Marketing Manager Chemistry Systems Enclosures: Method Comparison graphs for OSR6167 versus B46435/B38858 on AU5800 for CSF and Urine Method Comparison graphs for B46435/B38858 on AU5800 versus an Albumin test on a Siemens BN ProSpec Nephelometer for CSF and Urine 2014 Beckman Coulter, Inc. All rights reserved. Beckman Coulter, the Beckman Coulter logo, and AU are trademarks of Beckman Coulter, Inc. and are registered in the USPTO. Bio-Rad and Liquichek are trademarks of Bio-Rad Laboratories Ltd. Siemens and ProSpec are trademarks of Siemens. All other trademarks are the property of their respective owners. B58824-AA Page 3 of 5

4 Method Comparison of B46435/B38858 versus OSR6167 on an AU5800 analyser A. Urine samples B. CSF samples n: 117 Slope: Intercept: mg/l r: n: 49 Slope: Intercept: mg/l r: Figure 1. Method comparison graphs for OSR6167 vs. B46435 using A. Urine samples and B. CSF samples including Deming regression summary details. B58824-AA Page 4 of 5

5 Method Comparison of B46435/B38858 on an AU5800 versus an Albumin Urine/CSF test on a Siemens BN ProSpec Nephelometer A. Urine samples B. CSF samples n:121 Slope: Intercept: mg/l r: n: 51 Slope: Intercept: mg/L r: Figure 2. Method comparison graphs for B46435 vs. an Albumin Urine/CSF test on a Siemens BN ProSpec Nephelometer method using A. Urine samples and B. CSF samples including Deming regression summary details. B58824-AA Page 5 of 5

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

Cardiac Assessment Controls

Cardiac Assessment Controls Bio-Rad Laboratories CARDIAC ASSESSMENT CONTROLS Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central

More information

Third Party Controls for Your Dade Behring Systems

Third Party Controls for Your Dade Behring Systems Bio-Rad Laboratories Q U A L I T Y C O N T R O L Third Party Controls for Your Dade Behring Systems The Complete QC Solution from Bio-Rad Laboratories Bio-Rad has the control you need for your Dade Behring

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,

More information

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems Bio-Rad Laboratories Cardiac Assessment Controls Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central

More information

liquicolor (AMP Buffer, IFCC) Design Verification

liquicolor (AMP Buffer, IFCC) Design Verification Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...

More information

LIPASE liquicolor. Design Verification. Multipurpose Reagent

LIPASE liquicolor. Design Verification. Multipurpose Reagent Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of

More information

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for

More information

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

Cardiac Assessment Controls A critical element of reliable cardiac testing

Cardiac Assessment Controls A critical element of reliable cardiac testing Cardiac Assessment Controls Cardiac Assessment Controls A critical element of reliable cardiac testing Choose the products best suited for your cardiac assessment needs Each year, 17.7 million people on

More information

Quick Reference Guide. on Turbodyne SC

Quick Reference Guide. on Turbodyne SC RF Turbodyne RF Quantitative turbidimetric immunoassay for determination of Rheumatoid Factors Turbodyne RF is used for the detection of Rheumatoid Arthritis (RA), Differential diagnosis of RA from Rheumatic

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by

More information

1 PROTOCOL. Comparison Study Summary

1 PROTOCOL. Comparison Study Summary Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015

More information

The Adiponectin Turbidimetric Immunoassay Reagent Kit

The Adiponectin Turbidimetric Immunoassay Reagent Kit The Adiponectin Turbidimetric Immunoassay Reagent Kit Catalogue number: 51010 For the quantitative determination of Adiponectin in human serum and plasma This package insert must be read in its entirety

More information

ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System

ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System C ARK Methotrexate Assay Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 4600 Chemistry System Reference No. 5026000100 Intended for the Quantitative Determination of Methotrexate

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

Clinical Chemistry Specific Proteins Presentation

Clinical Chemistry Specific Proteins Presentation Clinical Chemistry Specific Proteins Presentation November 2004 AEROSET and c8000 are registered trademarks of Abbott Laboratories. All other trademarks, brands, trade names and product names are the property

More information

Beckman Coulter introduces Sentinel Pancreatic Amylase reagent and calibrator on AU Clinical Chemistry Analyzers

Beckman Coulter introduces Sentinel Pancreatic Amylase reagent and calibrator on AU Clinical Chemistry Analyzers PREVIOUSLY RELEASED BREA, CALIF. (June 23, 2016) Beckman Coulter Diagnostics is pleased to announce the release of a new pancreas specific amylase assay. The Sentinel Pancreatic Amylase reagent and calibrator

More information

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

DRI METHAQUALONE APPLICATION BECKMAN COULTER AU5800

DRI METHAQUALONE APPLICATION BECKMAN COULTER AU5800 DRI METHAQUALONE APPLICATION BECKMAN COULTER AU5800 Catalog No. 0514, 0515 Homogeneous Enzyme Immunoassay for the Qualitative and Semiquantitative Determination of Methaqualone Levels in Urine For In Vitro

More information

CEDIA OPIATE APPLICATION BECKMAN COULTER AU5800

CEDIA OPIATE APPLICATION BECKMAN COULTER AU5800 CEDIA OPIATE APPLICATION BECKMAN COULTER AU5800 Catalog No. 100089, 100098, 1661248 Intended for the qualitative and semiquantitative determination of opiates in human urine For In Vitro Diagnostic and

More information

HbA1c3 PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

HbA1c3 PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. HbA1c3 Hemoglobin A1c3 REF B36415 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

Product Correction Urgent Immediate Action Required

Product Correction Urgent Immediate Action Required Abbott Ireland Diagnostics Division Lisnamuck, Longford Co. Longford, Ireland DIAGNOSTICS Urgent Field Safety Notice Product Correction Urgent Immediate Action Required Date Issued October 12, 2018 Product

More information

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS DIAZYME PRODUCT LIST LIST PRODUCT 2 15 INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses its proprietary

More information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

human Total Cathepsin B Catalog Number: DY2176

human Total Cathepsin B Catalog Number: DY2176 human Total Cathepsin B Catalog Number: DY2176 This DuoSet ELISA Development kit contains the basic components required for the development of sandwich ELISAs to measure natural and recombinant human Total

More information

RealLine HIV quantitative Str-Format

RealLine HIV quantitative Str-Format Instructions for use DETECTION AND QUANTIFICATION OF THE HUMAN IMMUNODEFICIENCY VIRUS RNA BY REAL TIME PCR Research Use Only (RUO) Attention! Please read the information about quantification process carefully!

More information

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS DIAZYME PRODUCT LIST LIST PRODUCT 2 14 INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses its proprietary

More information

C-Peptide I and II (Rat) ELISA

C-Peptide I and II (Rat) ELISA C-Peptide I and II (Rat) ELISA For the quantitative determination of rat C-peptide in plasma, serum, urine, and cell culture supernatant Please read carefully due to Critical Changes, e.g., blot plate

More information

Urgent Field Corrective Action

Urgent Field Corrective Action Siemens Healthcare Diagnostics Products GmbH N Latex CDT Urgent Field Corrective Action PP-18-002.A.OUS May 2018 Negative bias of %CDT values Our records indicate that your facility may have received the

More information

HumaTex CRP. Design Verification. Contents

HumaTex CRP. Design Verification. Contents Design Verification HumaTex CRP Contents 1 Function... 2 2 Reproducibility... 2 3 Sensitivity and dynamic range... 2 Preparation of serum control panel... 2 Sensitivity test results... 2 Prozone check...

More information

Technical Bulletin. Performance Characteristics of the BN ProSpecä System. By Liana Frances Romero, BS, MT(ASCP)

Technical Bulletin. Performance Characteristics of the BN ProSpecä System. By Liana Frances Romero, BS, MT(ASCP) Technical Bulletin Performance Characteristics of the BN ProSpecä System By Liana Frances Romero, BS, MT(ASCP) Introduction The BN ProSpecä System is an automated nephelometry analyzer for in vitro diagnostic

More information

READ HIGHLIGHTED CHANGES

READ HIGHLIGHTED CHANGES BIO-FLASH anti-hbs 3000-8581 100 tests The BIO-FLASH anti-hbs is a fully automated chemiluminescent simultaneous immunoassay for quantitative measurement of antibodies to Hepatitis B surface antigen (anti-hbs)

More information

Smart Diagnostics for Point-of-Care!

Smart Diagnostics for Point-of-Care! Smart Diagnostics for Point-of-Care! The one special multi-parameter Point-of-Care (POC) instrument that covers the constantly expanding range of applications with the highest precision. Designed for HbA1c

More information

BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S.

BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S. BLOOD IS COMPLEX ANALYZING IT SHOULDN T BE DxH 500 * *Not available for sale in the U.S. Introducing the DxH 500, an Open-vial Hematology System *. Quality 5-part Differential in a Compact Design The first

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and

More information

Rat C-Peptide EIA. Cat. No. YII-YK010-EX FOR LABORATORY USE ONLY

Rat C-Peptide EIA. Cat. No. YII-YK010-EX FOR LABORATORY USE ONLY Rat C-Peptide EIA Cat. No. YII-YK010-EX FOR LABORATORY USE ONLY TOYO 2CHOME, KOTO-KU, TOKYO, 135-0016, JAPAN http://www.cosmobio.co.jp e-mail : export@cosmobio.co.jp 1 Phone : +81-3-5632-9617 FAX : +81-3-5632-9618

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

ab ORAC Assay Kit

ab ORAC Assay Kit Version 1 Last updated 10 April 2018 ab233473 ORAC Assay Kit For the measurement of ORAC activity in cell lysate, plasma, serum, tissue homogenates and food extracts. This product is for research use only

More information

NEPHROCHECK Liquid Control Kit Package Insert

NEPHROCHECK Liquid Control Kit Package Insert NEPHROCHECK Liquid Control Kit Package Insert Manufactured for Astute Medical, Inc. 3550 General Atomics Ct. Building 2 San Diego, CA 92121 USA Intended Use The NEPHROCHECK Liquid Controls are used for

More information

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS

DIAZYME PRODUCT LIST LIST INNOVATIONS IN CLINICAL DIAGNOSTICS DIAZYME PRODUCT LIST LIST PRODUCT 2 13 INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses its proprietary

More information

ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2)

ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2) ASSAYED URINE CONTROL - LEVEL 2 (URN ASY CONTROL 2) CAT. NO. AU 2352 LOT NO. 718UC SIZE: 12 x 10ml EXPIRY: 2017-10 INTENDED USE This product is intended for in vitro diagnostic use in the quality control

More information

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date AU US Instructions For Use 2018 Beckman Coulter, Inc. All rights reserved. TRIG TRIGLYCERIDE OSR60118 4 x 20 ml R1, 4 x 5 ml R2 OSR61118 4 x 50 ml R1, 4 x 12.5 ml R2 OSR66118 4 x 173 ml R1, 4 x 48 ml R2

More information

GC & LC Application Note

GC & LC Application Note GC & LC Application Note www.palsystem.com Comparison of the manual and automated generation of calibration standards Comparison of the manual and automated generation of calibration standards Thomas Funke,

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Date. Measurements of cocaine and cocaine metabolites are used in the diagnosis and treatment of cocaine use or overdose.

Date. Measurements of cocaine and cocaine metabolites are used in the diagnosis and treatment of cocaine use or overdose. SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Cocaine Metabolite REF 475003 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE INSTRUCTIONS FOR USE VITROS Chemistry Products malb Reagent malb 680 1740 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products malb Reagent is used on the VITROS 5,1 FS Chemistry

More information

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet Interleukin5 (IL5) is a secreted glycoprotein that belongs to the α-helical group of cytokines (1, 2, 3). Unlike other family members, it is present as a covalently linked antiparallel dimer (4, 5). IL-5

More information

91154 PRIME-XV T Cell CDM Liquid 1 L Additional package sizes are available at request

91154 PRIME-XV T Cell CDM Liquid 1 L Additional package sizes are available at request PRIME-XV T CELL CDM PRIME-XV T Cell CDM is a ready-to-use chemically-defined, animal component-free medium. It is optimized and designed for the culture of T cells of human origin and recommended for use

More information

Original Article. Evaluation of two new high-sensitivity methods for C-reactive protein. Introduction. Abstract

Original Article. Evaluation of two new high-sensitivity methods for C-reactive protein. Introduction. Abstract Original Article Evaluation of two new high-sensitivity methods for C-reactive protein SnjezÆana Rothkrantz-Kos 1, Otto Bekers 1, Armand Gubbels 1, Marjolein Drent 2, Maria PJ Schmitz 1 and Marja P van

More information

HbA1c PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

HbA1c PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Hemoglobin A1c REF 650262 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED

More information

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. 2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846

More information

Rat Glicentin EIA FOR RESEARCH USE ONLY. <Distributed by> DF Kasumigaseki Place, 3-6-7, Kasumigaseki, Chiyoda-ku Tokyo Japan

Rat Glicentin EIA FOR RESEARCH USE ONLY. <Distributed by> DF Kasumigaseki Place, 3-6-7, Kasumigaseki, Chiyoda-ku Tokyo Japan YK111 Rat Glicentin EIA FOR RESEARCH USE ONLY DF Kasumigaseki Place, 3-6-7, Kasumigaseki, Chiyoda-ku Tokyo 100-0013 Japan URL: http://www.sceti.co.jp/export/ e-mail: exp-pet@sceti.co.jp

More information

Daily Quality Control Calibration Verification / Linearity. Quality control made easy. Product Information

Daily Quality Control Calibration Verification / Linearity. Quality control made easy. Product Information Daily Quality Control Calibration Verification / Linearity Quality control made easy. Product Information AUDIT MicroControls Making life in the lab easier. AUDIT provides value to customers through high-quality

More information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT

More information

ASSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3)

ASSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3) SSAYED URINE CONTROL - LEVEL 3 (URN ASY CONTROL 3) CAT. NO. AU 2353 LOT NO. 797UC SIZE: 12 x 10 ml EXPIRY: 2019-03 INTENDED USE This product is intended for in vitro diagnostic use, in the quality control

More information

New York State Department of Health. Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine

New York State Department of Health. Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine New York State Department of Health Notice of Voluntary Recall of Influenza A (H1N1) 2009 Monovalent Vaccine The U.S. Centers for Disease Control and Prevention have announced a Non safety Related Voluntary

More information

AmpliX HBV Quantitative

AmpliX HBV Quantitative Instructions for use REAL TIME PCR DETECTION AND QUANTITATION KIT OF HEPATITIS B VIRUS DNA Research Use Only (RUO) (Lyo-format) VBD0595 96 rcs valid from May 2013 Explanation of symbols used in labeling

More information

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153

S150 KEEP Analytical Methods. American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153 S150 KEEP 2009 Analytical Methods American Journal of Kidney Diseases, Vol 55, No 3, Suppl 2, 2010:pp S150-S153 S151 The Kidney Early Evaluation program (KEEP) is a free, communitybased health screening

More information

Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing. Vincent Delatour, PhD

Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing. Vincent Delatour, PhD Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing Vincent Delatour, PhD 1 Traceability in laboratory medicine : regulatory drivers Reform of medical biology in France

More information

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA

Evidence Based Commutability: Bias 2 Study. Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA Evidence Based Commutability: Bias 2 Study Janice Gill Manager RCPAQAP Chemical Pathology Adelaide SA Australian Bias Studies conducted by Gus Koerbin, ACT Pathology on behalf of AACB Harmonisation Committee

More information

Nori Rabbit IL-2 ELISA Kit DataSheet

Nori Rabbit IL-2 ELISA Kit DataSheet Nori Rabbit IL-2 ELISA Kit DataSheet IL-2 is an interleukin, a type of cytokine immune system signaling molecule. IL-2 is a T cell stimulatory cytokine best known for inducing T cell proliferation, the

More information

Chapter 2. High-sensitivity C-reactive protein methods examined

Chapter 2. High-sensitivity C-reactive protein methods examined Chapter High-sensitivity C-reactive protein methods examined Snježana Rothkrantz-Kos, Maria PJ Schmitz, Otto Bekers, Paul PCA Menheere, Marja P van Dieijen-Visser Clin Chem ;48:359-6 3 Chapter Abstract

More information

Performance Guide. icheck Iodine The test kit to measure iodine in salt

Performance Guide. icheck Iodine The test kit to measure iodine in salt Performance Guide icheck Iodine The test kit to measure iodine in salt Performance explained This Performance Guide is intended to explain the validity of our test kit icheck Iodine in a nutshell to facilitate

More information

Rat Leptin ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN

Rat Leptin ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN YK050 Rat Leptin ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC. 2480-1 AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN 418-0011 Contents Introduction 2 Characteristics 3 Composition 4 Method 5-6 Notes

More information

The superior solution for disease state diagnosis. IMMAGE 800 Special Chemistry System

The superior solution for disease state diagnosis. IMMAGE 800 Special Chemistry System The superior solution for disease state diagnosis. IMMAGE 800 Special Chemistry System Blood Banking Centrifugation Chemistry Flow Cytometry Hematology Hemostasis Immunoassay Information Systems Lab Automation

More information

YK052 Mouse Leptin ELISA

YK052 Mouse Leptin ELISA YK052 Mouse Leptin ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC. 2480-1 AWAKURA, FUJINOMIYA-SHI SHIZUOKA, JAPAN 418-0011 Contents Ⅰ. Introduction 2 Ⅱ. Characteristics 3 Ⅲ. Composition 4 Ⅳ. Method

More information

INSULIN MOUSE SERUM KITS

INSULIN MOUSE SERUM KITS INSULIN MOUSE SERUM KITS PROTOCOL Part # 62IN3PEF & 62IN3PEB Test size#: 200 tests (62IN3PEF), 5 x 200 tests (62IN3PEB) - assay volume: 20 µl Revision: 03-Jan.2018 Store at: 2-8 C (62IN3PEF); 2-8 C (62IN3PEB)

More information

Mouse GLP-2 EIA FOR LABORATORY USE ONLY

Mouse GLP-2 EIA FOR LABORATORY USE ONLY YK142 Mouse GLP-2 EIA FOR LABORATORY USE ONLY Kasumigaseki place, 3-6-7, Kasumigaseki, Chiyoda-ku, Tokyo 100-0013 Japan http://www.sceti.co.jp/english/export e-mail exp-pet@sceti.co.jp

More information

1. Calibra - H - Store at 2-8 ºC.

1. Calibra - H - Store at 2-8 ºC. CALIBRA H Insert Ref.:80 Lot Expiration Calibrator Attention. It is suggested to verify carefully if the lot number printed in this insert corresponds to the lot on the bottle label. Intended use. is a

More information

Thyroxine (T4) Human ELISA Kit

Thyroxine (T4) Human ELISA Kit ab108662 Thyroxine (T4) Human ELISA Kit Instructions for Use For the quantitative measurement of Human Thyroxine (T4) concentrations in serum. This product is for research use only and is not intended

More information

Product information: Human Apolipoprotein B kit 10,000 tests 1. ASSAY DESCRIPTION AND INTENDED USE 2. PROTOCOL AT GLANCE

Product information: Human Apolipoprotein B kit 10,000 tests 1. ASSAY DESCRIPTION AND INTENDED USE 2. PROTOCOL AT GLANCE Headquarters & Europe Office Cisbio Bioassays Phone: +33 ()4 66 79 67 5 Fax: +33 ()4 66 79 19 2 bioassays@cisbio.com USA Office Cisbio US, Inc. Phone: +1 888 963 4567 Fax: +1 781 687 15 htrfinfo@cisbio.us

More information

Evaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser

Evaluation Report. Eurolyser Phenobarbital test kit (VT0211) on. solo Analyser Evaluation Report Eurolyser Phenobarbital test kit (VT0211) on solo Analyser Location Location: Eurolyser Diagnostica GmbH Operator: Bettina Schönauer Date: February 2016 - March 2016 Specimens The specimens

More information

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2 Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...

More information

GLP-2 (Rat) ELISA. For the quantitative determination of glucagon-like peptide 2 (GLP-2) in rat serum and plasma

GLP-2 (Rat) ELISA. For the quantitative determination of glucagon-like peptide 2 (GLP-2) in rat serum and plasma GLP-2 (Rat) ELISA For the quantitative determination of glucagon-like peptide 2 (GLP-2) in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 48-GP2RT-E01

More information

PRODUCTS FOR EXPORT 2018

PRODUCTS FOR EXPORT 2018 PRODUCTS FOR EXPORT 2018 MEDICON HELLAS SA 5-7 Melitona st., 153 44 Gerakas, Attiki, Greece - Tel.:+30 210 6606000 FAX:+30 210 6612666 e-mail:diag@mediconsa.com, http://www.mediconsa.com General Electronic

More information

SERUM GLUCAGON KITS PROTOCOL

SERUM GLUCAGON KITS PROTOCOL SERUM GLUCAGON KITS PROTOCOL Part # 62SGLPEB & 62SGLPEF Test size#: 5 x 200 tests (62SGLPEB), 200 tests (62SGLPEF) - assay volume: 20 µl Revision: 02-Jan.2018 Store at: -60 C or below (62SGLPEB); -60 C

More information

Total Cost of Ownership (TCO): An evidence-based approach to compare laboratory equipment

Total Cost of Ownership (TCO): An evidence-based approach to compare laboratory equipment Total Cost of Ownership (TCO): An evidence-based approach to compare laboratory equipment P.C.G. Gontard 1, L.I. Stankevich 1, B.G. Gorodetsky 1 SUMMARY Clinical laboratories across the globe operate in

More information

RealLine HCV quantitative Str-Format

RealLine HCV quantitative Str-Format Instructions for use QUANTITATIVE KIT FOR REAL TIME PCR DETECTION FOR RNA OF HEPATITIS C VIRUS Attention! Please read the information about quantification process carefully! Research Use Only (RUO) (Str

More information

Urgent field safety notice

Urgent field safety notice Accu-Chek Aviva Urgent field safety notice May 2018 Important information on selected lots of Accu-Chek Aviva (50s, 10s) potentially showing an increased number of strip errors prior to dosing or biased

More information

RealLine Mycoplasma genitalium Str-Format

RealLine Mycoplasma genitalium Str-Format Instructions for use ASSAY KIT FOR THE QUALITATIVE DETECTION OF MYCOPLASMA GENITALIUM DNA BY REAL-TIME PCR METHOD In vitro Diagnostics () VBD4396 96 Tests valid from December 2018 Rev06_1218_EN Page 1

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Urinary Albumin and Urinary Creatinine Laboratory Survey Years: 2005 to 2006 SAS Transport File: ALB_CR_D.XPT November 2007 NHANES 2005 2006 Data Documentation

More information

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold

Comparison of Biotage Extrahera vs. Manual Sample Processing Using a Vacuum Manifold Technical Note PPS366 Extraction of 25-OH Vitamin D from Plasma Using ISOLUTE SLE+ Page 1 Comparison of Biotage vs. Sample Processing Using a Vacuum Manifold Extraction of 25-OH Vitamin D from Plasma Using

More information

Rat Leptin-HS ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN

Rat Leptin-HS ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN YK051 Rat Leptin-HS ELISA FOR LABORATORY USE ONLY YANAIHARA INSTITUTE INC. 2480-1 AWAKURA, FUJINOMIYA - SHI SHIZUOKA, JAPAN 418 0011 Contents Introduction 2 Characteristics 3 Composition 4 Method 5-6 Notes

More information

NEPHROCHECK Calibration Verification Kit Package Insert

NEPHROCHECK Calibration Verification Kit Package Insert NEPHROCHECK Calibration Verification Kit Package Insert Manufactured for Astute Medical, Inc. 3550 General Atomics Ct. Building 2 San Diego, CA 92121 USA Intended Use The NEPHROCHECK Calibration Verification

More information

Abstract. Introduction

Abstract. Introduction A Fast Dilute-And-Shoot Method for Simultaneous 5-Hydroxyindoleacetic Acid (5-HIAA), Vanillylmandelic Acid (VMA), and Homovanillic Acid (HVA) LC-MS/MS Analysis in Human Urine By Shun-Hsin Liang and Sue

More information

I. Introduction. II. Characteristics

I. Introduction. II. Characteristics YK050 Rat Leptin ELISA I. Introduction Leptin, which is a product of ob gene, is a protein consisting of 167 amino acids and it is secreted from white adipose tissue. It is known that leptin acts on hypothalamus

More information

ab Lipoxygenase Inhibitor Screening Assay Kit

ab Lipoxygenase Inhibitor Screening Assay Kit ab133087 Lipoxygenase Inhibitor Screening Assay Kit Instructions for Use For the detection of hydroperoxides produced in the lipoxygenation reaction using a purified Lipoxygenases. This product is for

More information

FOR LABORATORY USE ONLY

FOR LABORATORY USE ONLY YK060 Insulin ELISA FOR LABORATORY USE ONLY Kasumigaseki Place, 3-6-7, Kasumigaseki, Chiyoda-ku Tokyo 100-0013 Japan URL: http://www.sceti.co.jp/export/ e-mail: medical@sceti.co.jp Contents

More information

The NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16

The NEPHROCHECK Test System Training Program. The NEPHROCHECK Training Program US: Astute Medical, Inc PN Rev D 2014/09/16 The NEPHROCHECK Test System Training Program The NEPHROCHECK Training Program US: Astute Medical, Inc. 2014 PN 700001 Rev D 2014/09/16 Learning Objectives At the conclusion of this training, participants

More information