1 Changes in Intraocular Pressure After Laser in situ Keratomileusis Khaled M. Rashad, MD; Ahmed A. Bahnassy, MSc, PhD ABSTRACT PURPOSE: To evaluate changes in intraocular pressure (IOP) measurements by Goldmann applanation tonometry after laser in situ keratomileusis (LASIK) for myopia and myopic astigmatism, and to assess the accuracy of Goldmann applanation tonometry measurements after LASIK in these eyes. METHODS: LASIK was performed on 166 eyes of 93 patients for correction of myopia and myopic astigmatism. Intraocular pressure was measured by Goldmann applanation tonometry at the central and temporal parts of the cornea before and at 1, 3, 6, and 12 months after LASIK. The amount of change in IOP was computed and its relation to different variables was evaluated by regression analysis. RESULTS: Intraocular pressure measured at the center of the cornea was reduced by a mean of 3.69 ± 1.63 mmhg after LASIK. Multiple regression analysis showed that the decrease in IOP was related to the preoperative IOP and the change in central corneal thickness after LASIK. Measurements of IOP at the temporal part of the cornea were also reduced by a mean of 2.39 ± 1.71 mmhg. There was wide variability in the amount of difference between the temporal and central measurements after LASIK (temporal measurements were higher than central by 0 to +4 mmhg). CONCLUSION: LASIK for myopia produced underestimation of IOP measured by Goldmann applanation tonometry at the central part of the cornea by a mean of 3.69 ± 1.63 mmhg. The decrease of IOP was related to preoperative IOP and the change in central corneal thickness after LASIK. Temporal, although decreased after LASIK, were less reliable. [J Refract Surg 2001;17: ] From the Ophthalmology Department, Faculty of Medicine, Alexandria University, Egypt and Almana General Hospital, Dammam, Saudi Arabia (Rashad) and Department of Family and Community Medicine, College of Medicine, King Faisal University, Saudi Arabia (Bahnassy). The authors have no proprietary interest in the materials presented herein. Correspondence: Khaled M. Rashad, MD, AL Mana General Hospital, P. O. Box 2366, Dammam Saudi Arabia. Tel: ext.#1007; Fax: Received: April 3, 2000 Accepted: February 16, 2001 Laser in situ keratomileusis (LASIK) has become an accepted method to treat many refractive errors. Recent studies have shown that it is a safe, effective, and predictable procedure for the correction of myopia and myopic astigmatism. 1,2 As the number of patients undergoing LASIK increases, there is concern about the accuracy of intraocular pressure (IOP) measurements after excimer laser refractive surgery. Goldmann applanation tonometry is considered the standard method for assessment of IOP, and is used routinely in ophthalmic practice. 3,4 Measurements of IOP by Goldmann applanation tonometry are influenced by corneal thickness and to a lesser extent by corneal curvature. 5 In a normal population, thin corneas tend to give lower readings (underestimation) by Goldmann applanation tonometry, whereas thick corneas are associated with elevated readings (overestimation). 5,6 Previous studies reported decreased corneal thickness in many cases of low-tension glaucoma, and increased corneal thickness in cases of ocular hypertension. 7,8 Studies also showed a positive correlation between IOP measurements by Goldmann applanation tonometry and increasing corneal curvature. 9 Thus, flat corneas can cause the IOP measurement to be underestimated by Goldmann applanation tonometry. 10 LASIK for myopia (and myopic astigmatism) alters corneal parameters including central corneal thickness and corneal curvature, hence LASIK can change the accuracy of IOP measurements by Goldmann applanation tonometry. 11,12 This study prospectively evaluates the changes that occur in IOP measurements by Goldmann applanation tonometry after LASIK for myopia and myopic astigmatism. PATIENTS AND METHODS LASIK was performed on 166 eyes of 93 patients for correction of myopia and myopic astigmatism. Fifty-two patients (55.9%) were males, and 41 (44.1%) were females. The age of the patients 420 Journal of Refractive Surgery Volume 17 July/August 2001

2 ranged from 20 to 64 years with a mean age of 30.6 ± 8.0 years. The spherical equivalent manifest refraction ranged from to diopters (D) with a mean of ± 3.75 D. The amount of refractive astigmatism (as measured by manifest refraction) ranged from 0 to 6.50 D with a mean of 0.81 ± 0.99 D. Preoperative examination included manifest refraction, uncorrected and spectacle-corrected visual acuity, slit-lamp biomicroscopy, and fundus examination. Corneal topography (videokeratography) was examined using the Computed Anatomy Topographic Modeling System TMS-1 (Computed Anatomy Inc., New York, NY). The average keratometric power was calculated from the Sim K (simulated keratometry) readings of the corneal topography. 12 The intraocular pressure (IOP) was measured using a Goldmann applanation tonometer (Haag- Streit). The calibration of the tonometer was checked periodically. To measure IOP, corneal anesthesia was achieved with oxybuprocaine hydrochloride 0.4% (Novesin) eye drops. The tear film was stained by touching the lower tarsal conjunctiva with a moistened strip of fluorescein impregnated paper. The patient was asked to look straight ahead and to avoid blinking, and IOP was measured at the center of the cornea. Three measurements were done in each eye, and IOP was recorded as the mean of the three readings. Then, the patient was asked to look slightly nasally, and IOP was measured at the temporal part of the cornea close to, but still inside, the limbus. Again, three measurements were done, and the mean was recorded. The central corneal thickness was measured by ultrasonic pachymetry using a Teknar Ophthasonic pachymeter (Teknar Corporation, St. Louis, MO). The calibration of the pachymeter was checked according to the manufacturer s instructions. After topical anesthesia, the ultrasonic probe was held perpendicular to the corneal surface opposite the center of the pupil. Three measurements were recorded, and the mean was calculated. Selection criteria included an age of 20 years or older, with stable refraction over the past year, and a spectacle-corrected visual acuity of 20/50 or better. Patients with keratometric power more than D, those with keratoconus or suspected keratoconus, or those with IOP greater than 22 mmhg by Goldmann applanation tonometry were excluded. Patients had no corneal pathology or ocular disease apart from myopia (and astigmatism). Patients using soft contact lenses were asked to discontinue contact lens use for 1 week (4 weeks for gaspermeable hard lenses) prior to preoperative evaluation. Patients with diabetes or collagen diseases that are likely to affect corneal healing were also excluded. All LASIK procedures were performed under topical anesthesia by the same surgeon (KMR) at Almana General Hospital in Dammam between March and November The refractive goal was emmetropia in all eyes. The surgical technique has been described in detail. 1,2 In the first 80 eyes, the Automated Corneal Shaper (Chiron Vision, Claremont, CA) was used to create a 160-µm, nasally hinged corneal flap, about 8.5 mm in diameter. Laser ablation was performed using the Chiron- Technolas Keracor 116 excimer laser (Chiron Technolas GMBH, Munich, Germany), with multizone ablation (usually ablation zones of 5.0 and 6.0 mm were used with an outer blend zone of 7.0 mm). In the remaining 86 eyes, the Carriazo- Barraquer microkeratome (Moria, Antony, France) was used to create a 160-µm corneal flap with a superior hinge, about 9 to 9.5 mm in diameter. Laser ablation was performed using the Keracor 117C PlanoScan excimer laser (Technolas Bausch and Lomb Surgical), with an ablation zone size of 5.5 to 6.0 mm and a transition zone of 8.1 to 9.0 mm. Postoperatively, spersadexoline eye drops (0.1% dexamethasone, 0.5% chloramphenicol, 0.025% tetrahydrozoline hydrochloride, Ciba Vision, Hettlingen, Switzerland) were prescribed four times daily for 1 week, and artificial tears were used for 6 to 8 weeks. Follow-up examinations were scheduled at 1 day, 1 week, 1, 3, 6, and 12 months after LASIK for all eyes. Postoperative examination included manifest refraction, uncorrected and spectacle-corrected visual acuity, and slit-lamp biomicroscopy. From the second postoperative examination (1 week after LASIK), the examination included corneal topography, measurement of IOP by Goldmann applanation tonometry in the central and temporal parts of the cornea, and ultrasonic pachymetry to measure central corneal thickness. The preoperative and postoperative were performed at the same time interval (between 5 and 8 PM) to minimize the possible effects of diurnal IOP variations. Statistical Analysis The data were analyzed using SPSS Statistical Package for windows, Version 7.5 (SPSS Inc., Redmond, WA). Descriptive statistics were analyzed for all variables. Paired t-test was used to compare Journal of Refractive Surgery Volume 17 July/August

3 Table 1 Paired t-test Comparing Data for 166 Eyes Before and 1 Year After LASIK Preoperative Postoperative Change Paired t-test Mean ± SD (Range) Mean ± SD (Range) Mean ± SD (Range) Significance (P-value) Spherical equivalent manifest refraction (D) ± ± ± 3.52 <.001* (-1.13 to ) (+0.75 to -3.00) (1.13 to 18.00) Refractive astigmatism (D) 0.81 ± ± ± 0.79 <.001* (0.00 to 6.50) (0.00 to 2.75) (0.00 to 6.25) Average keratometric power (D) ± ± ± 2.04 <.001* (40.40 to 47.00) (31.40 to 45.00) (-0.8 to -10.8) Central corneal thickness (µm) ± ± ± <.001* (405.0 to 634.0) (380.0 to 614.0) (-10 to -136) Central Goldmann applanation tonometry (mmhg) ± ± ± 1.63 <.001* (7.0 to 20.0) (4.0 to 14.0) (0.0 to -9.0) Temporal Goldmann applanation tonometry (mmhg) ± ± ± 1.71 <.001* (7.0 to 20.0) (5.0 to 16.0) (0.0 to -7.0) * Statistically significant (P<.001) preoperative and postoperative data. Pearson correlation coefficient was computed to find the linear relationship between IOP and different variables. Stepwise multiple linear regression analysis was used to predict the change in IOP after LASIK from the related variables. P values less than.05 were considered statistically significant. RESULTS Table 1 shows the preoperative and 1-year postoperative spherical equivalent manifest refraction, refractive astigmatism, average keratometric power, central corneal thickness, central and temporal of intraocular pressure (IOP), and the amount of change that occurred after LASIK in these parameters. The mean ablation depth, as estimated by the software of the excimer laser, was ± µm (range, 22 to 182 µm). The actual change in central corneal thickness after LASIK, as measured by ultrasonic pachymetry, was less than the estimated ablation depth in all eyes (Table 1). Intraocular Pressure Before surgery, there were no significant differences between IOP measured by Goldmann applanation tonometry at the central and temporal parts of the cornea (Table 2). Measurements of IOP by Goldmann applanation tonometry at the central part of the cornea ranged from 7 to 20 mmhg with a mean of ± 2.69 mmhg. At 1 year after LASIK, IOP measured by Goldmann applanation tonometry at the center of the cornea ranged from 4 to 14 mmhg (mean, 8.94 ± 2.13 mmhg). Comparing the IOP measurements at the center of the cornea among the different postoperative examinations, there were no significant changes during the 12-month period of follow-up (P >.05). None of the operated eyes had a postoperative IOP higher than the preoperative measurements. After LASIK, IOP measured by Goldmann applanation tonometry at the center of the cornea was reduced by a mean of 3.69 ± 1.63 mmhg (range, 0 to -9 mmhg), compared to preoperative measurements. The decrease of IOP after LASIK was statistically significant (Table 1). After LASIK, measurements of IOP by Goldmann applanation tonometry at the temporal part of the cornea were generally higher than those obtained from the center of the cornea. At 1 year after LASIK, the temporal measurements of IOP by Goldmann applanation tonometry were higher than the central measurements by a mean of 1.38 ± 1.11 mmhg. The difference between the temporal and central measurements of IOP by Goldmann applanation tonometry after LASIK was statistically significant at all postoperative examinations (Table 2). However, there was wide variability in the amount of difference between the temporal and central measurements of IOP after LASIK. In 48 of the operated eyes (28.9%), the two measurements were identical; 422 Journal of Refractive Surgery Volume 17 July/August 2001

4 Table 2 Paired t-test Comparing IOP Measured by Goldmann Applanation Tonometry at the Central and Temporal Parts of the Cornea Before and After LASIK (166 Eyes) Central Temporal Difference Paired t-test Mean ± SD (Range) Mean ± SD (Range) Mean ± SD (Range) Significance (P-value) Preoperative ± ± ± (7.0 to 20.0) (7.0 to 20.0) (-1.0 to +1.0) Postoperative 1 month 8.77 ± ± ± 1.09 <.001* (5.0 to 14.0) (5.0 to 17.0) (0.0 to +4.0) 3 months 8.86 ± ± ± 1.17 <.001* (4.0 to 15.0) (5.0 to 16.0) (0.0 to +4.0) 6 months 8.99 ± ± ± 1.23 <.001* (5.0 to 14.0) (5.0 to 16.0) (0.0 to + 3.0) 12 months 8.94 ± ± ± 1.11 <.001* (4.0 to 14.0) (5.0 to 16.0) (0.0 to +4.0) * Statistically significant (P<.001) Table 3 Correlation Coefficient of Preoperative Central Goldmann Applanation Tonometry Measurements of IOP With Different Variables (166 Eyes) Variable Pearson Statistical Correlation Significance (r) With Pre- (P-value) operative IOP Age 0.139* Sex Preoperative central corneal thickness Preoperative average keratometric power Preoperative spherical equivalent refraction * * Correlation is statistically significant at the.05 level in three eyes (1.8%), the temporal measurements were higher than the central measurements by 4 mmhg. In the remaining 115 eyes (69.3%), the temporal measurements were higher than the central measurements by 1 to 3 mmhg. There were no significant changes in the postoperative temporally measured IOP during the follow-up period. At 1 year after LASIK, the IOP measured by Goldmann applanation tonometry at the temporal part of the cornea was reduced by a mean of 2.39 ± 1.71 mmhg (range, 0 to -7 mmhg), compared to preoperative measurements (Table 1). Correlation of IOP With Different Parameters Before surgery, there was a significant correlation between the preoperative IOP (measured by Goldmann applanation tonometry at the center of the cornea) and the age of the patients (Table 3). The preoperative IOP also had a significant negative correlation with the spherical equivalent refraction (ie, patients with higher amounts of myopia had higher IOP). The correlation between preoperative IOP and preoperative central corneal thickness almost reached statistical significance (P =.055). After LASIK, the postoperative IOP (measured by Goldmann applanation tonometry at the center of the cornea) maintained a significant positive correlation with age. The postoperative IOP also had a statistically significant correlation with the preoperative IOP, and with the postoperative central corneal thickness (Table 4). A significant correlation was found between the amount of change in IOP after LASIK (postoperative minus preoperative IOP measured by Goldmann applanation tonometry at the center of the cornea) and the preoperative IOP. This means that eyes with higher preoperative IOP had a greater decrease in IOP after LASIK. The change in IOP also had a statistically significant correlation with the change in central corneal thickness, change in average keratometric power, and the change in spherical equivalent refraction (Table 5). Journal of Refractive Surgery Volume 17 July/August

5 Table 4 Correlation Coefficient of 1-year Postoperative Central Goldmann Applanation Tonometry Measurements of IOP With Different Variables (166 Eyes) Variable Pearson Statistical Correlation Significance (r) With Pre- (P-value) operative IOP Age 0.190*.007 Sex Preoperative IOP 0.794* <.001 Change in central corneal thickness Change in spherical equivalent refraction Change in average keratometric power Postoperative central corneal thickness 0.182*.010 Postoperative average keratometric power * Correlation is statistically significant at the.01 level Table 5 Correlation Coefficient of Change in Central Goldmann Applanation Tonometry Measurements of IOP After LASIK With Different Variables (166 Eyes) Variable Pearson Statistical Correlation Significance (r) With Change (P-value) in IOP Age Sex Preoperative IOP * <.001 Change in central corneal thickness 0.266* <.001 Change in average keratometric power 0.245*.001 Change in spherical equivalent refraction *.003 Postoperative central corneal thickness Postoperative average keratometric power 0.212*.003 * Correlation is statistically significant at the.01 level Table 6 Model Derived by Stepwise Multiple Linear Regression Analysis to Predict IOP After LASIK From Related Independent Variables Variable B t Statistical Significance (P-value) Preoperative IOP <.001 Change in central corneal thickness <.001 Age (Constant) R 2 = The change in average keratometric power and the change in the spherical equivalent refraction were excluded because they did not affect the predictability of postoperative IOP. Predictability of Change in IOP After LASIK The preoperative IOP measured by Goldmann applanation tonometry at the center of the cornea, the change in central corneal thickness after LASIK, and the patient s age, were significant predictors for postoperative IOP (R 2 = 0.674). This means that these variables explained 67.4% of the variability in the postoperative IOP (Table 6). The equation derived by stepwise multiple linear regression analysis to predict the postoperative IOP was: Table 7 Model Derived by Stepwise Multiple Linear Regression Analysis to Predict Change in IOP After LASIK From Related Independent Variables Variable B t Statistical Significance (P-value) Preoperative IOP <.001 Change in central corneal thickness <.001 Age (Constant) R 2 = The change in average keratometric power and the change in the spherical equivalent refraction were excluded because they did not affect the predictability of the change in IOP after LASIK. [Postoperative IOP (mmhg) = (constant) x preoperative IOP (mmhg) x change in central corneal thickness (µm) x patient s age (yr)]. Preoperative IOP, change in central corneal thickness after LASIK, and the patient s age were also significant predictors for the amount of change in IOP after LASIK (R 2 = 0.443). These variables explained 44.3% of the variability in the amount of decrease of IOP after LASIK (Table 7). The equation 424 Journal of Refractive Surgery Volume 17 July/August 2001

6 derived by stepwise multiple linear regression analysis to predict the change in IOP after LASIK was: [Change in IOP (mmhg) = (constant) x preoperative IOP (mmhg) x change in central corneal thickness (µm) x patient s age (yr)]. DISCUSSION Evaluation of IOP is a fundamental part of ocular examination in both health and disease. IOP measurement is not only important in the diagnosis and management of glaucoma, but its assessment is important in the management of corneal, lenticular, uveal, and retinal diseases. 5 Currently, the most common way to assess IOP is by Goldmann applanation tonometry, which is considered the standard method for measurement of IOP. 3,4 Goldmann applanation tonometry gives accurate measurements of IOP in eyes with normal corneal curvature, and with normal (average) corneal thickness of 520 µm. 5,14-17 In this study, the centrally measured IOP by Goldmann applanation tonometry was reduced by a mean of 3.69 ± 1.63 mmhg after LASIK for myopia. The reduction of IOP by Goldmann applanation tonometry after LASIK for myopia was consistent throughout the 12-month follow-up period. During the study, all preoperative and postoperative were performed at the same time interval (between 5 and 8 PM) to minimize the possible effects of diurnal IOP variations. Also, corticosteroid response could not affect the measurements because topical corticosteroids were used for 1 week only postoperatively. The cause of the decrease or underestimation of the centrally measured IOP after LASIK for myopia must be related to the change in the corneal parameters. A reduction of IOP measured by Goldmann applanation tonometry has been reported in different studies after excimer laser photorefractive keratectomy (PRK) and LASIK for myopia. 1,2,6,10-13,17-22 Rashad and Kalaway were the first to report the decrease of IOP measurement by Goldmann applanation tonometry after PRK for myopia. In that study, PRK was performed on 307 eyes with mean myopia of D, and IOP decreased from a mean of 12.9 ± 3.1 mmhg preoperatively to a mean of 11.0 ± 2.7 mmhg after PRK. 23 Schipper and colleagues performed PRK on 35 eyes with mean myopia of D and found a 2- to 3-mmHg decrease in centrally measured IOP after PRK; temporal measurements remained unchanged. 24 The reduction of IOP by Goldmann applanation tonometry after PRK for myopia in other reported studies varied between 2.0 and 3.1 mmhg. 10,17,20,21 Studies of LASIK for myopia also demonstrated underestimation of the IOP by Goldmann applanation tonometry after surgery. In our previous study on LASIK performed on 157 eyes with mean myopia of ± 3.74 D, mean IOP was reduced from 11.9 ± 2.7 mmhg preoperatively to 9.0 ± 2.4 mmhg after LASIK. 1 Emara and coauthors reported that IOP decreased from 16.1 ± 2.9 mmhg before surgery to 13.6 ± 3.3 mmhg after LASIK for myopia. 13 Underestimation of the IOP measurements by Goldmann applanation tonometry after LASIK for myopia in other reported studies varied between 1.9 and 3.8 mmhg. 11,18,19 In this study, there were no significant differences in IOP measured from the central or temporal parts of the cornea before surgery. Similar observations have been reported in previous studies of PRK. 17,24 At 1 year after LASIK, the measurements of IOP by Goldmann applanation tonometry at the temporal part of the cornea were higher than the central measurements by a mean of 1.38 ± 1.11 mmhg. In LASIK for myopia (or myopic astigmatism), the deepest ablation is in the center of the cornea, with less ablation toward the edge of the ablation zone. 13,18,25 Higher temporal measurements of IOP are explained by the relatively increased thickness of the peripheral cornea. However, our results showed that there was variability in the amount of difference between temporal and central IOP measurements by Goldmann applanation tonometry after LASIK. In some eyes the two measurements were identical; in other eyes the temporal readings were higher than the central measurements by as much as 4.0 mmhg. Measurements of IOP at the temporal part of the cornea after LASIK were still lower than preoperative measurements by a mean of 2.39 ± 1.71 mmhg. Our results differ from the studies of Schipper and Abbasoglu where central measurements were lower than peripheral measurements by 2 to 3 mmhg after PRK. 17,24 In these studies, ablation of the cornea was performed only in the central 6.0-mm zone, and the peripheral portion of the cornea was not ablated. In the present study, the diameter of the ablation zone was usually 6.0 mm with a large transition zone extending to 9.0 mm in some eyes, ie, the peripheral portion of the cornea was also partially ablated, although to a lesser extent than the central zone. Therefore, in our study, the peripheral corneal thickness was also reduced after LASIK. This may explain the variability in the amount of Journal of Refractive Surgery Volume 17 July/August

7 difference between the temporal and central after LASIK in our study. Before surgery, the correlation between IOP and central corneal thickness in this study almost reached statistical significance. After LASIK, central corneal thickness was reduced by a mean of ± µm. Regression analysis showed that there was a strong correlation between the drop in IOP and the change in central corneal thickness after LASIK. Several clinical and manometric studies have demonstrated a significant relationship between central corneal thickness and the error in IOP measurements by Goldmann applanation tonometry. 5,6,12,13,20,26-29 The Goldmann tonometer was found to give accurate readings when the central corneal thickness was 520 µm. 28 Previous studies showed that thin corneas produced underestimation of IOP by as much as 5 mmhg. 5,6,26,28 Many eyes diagnosed as having low-tension glaucoma were found to have thin corneas. 6,7,27 It is possible that central corneal thickness is related to the resistance of the cornea to applanation (corneal rigidity). 6,26 A thin cornea will be less rigid than a thick cornea and thus the force required to overcome corneal rigidity is reduced. When corneal rigidity is decreased, as a result of thinning of the central cornea after LASIK, the cornea requires less pressure to be applanated, resulting in low tonometer readings and underestimation of IOP. 5,6,12,13,20,26 The assumed thickness of corneal tissue removed, as estimated by the excimer laser software (estimated ablation depth), was always greater than the actual decrease in central corneal thickness observed after LASIK (as measured by ultrasonic pachymetry). Schipper 24 and Faucher 10 reported similar observations after PRK. Our results showed that patients with higher preoperative IOP had a larger decrease in IOP measurements after LASIK. Fournier and colleagues also reported that the decrease in IOP after myopic LASIK was greater if preoperative IOP was high. 11 The reduction of IOP in this study had a positive correlation with the reduction of average keratometric power after LASIK. Previous studies reported that Goldmann applanation tonometry readings are influenced by corneal curvature. 5,9,30,31 It is speculated that the steeper the corneal curvature, the more the cornea must be indented to produce the same area of applanation. 14 Flatter corneas are more readily applanated, and this may result in underestimation of IOP measurements by Goldmann applanation tonometry. 17,30 The flatter cornea after LASIK for myopia will require less force by the tonometer to reach the desired applanation diameter. 5,20 Also, flattening of the corneal curvature may cause an increased force of capillary attraction of the precorneal tear film, resulting in lower tonometer readings. 5 Results of multiple regression analysis in this study showed that preoperative IOP and change in central corneal thickness were important factors in determining the amount of reduction of IOP after LASIK for myopia. Fournier and colleagues also showed that reduction in IOP after LASIK was related to preoperative IOP. 11 Chatterjee and colleagues correlated the amount of IOP reduction to the amount of myopia corrected by PRK. 21 Other studies of PRK or LASIK could not demonstrate a statistically significant relationship between the amount of reduction of IOP after surgery and the changes in central corneal thickness, keratometric power, spherical equivalent refraction, or estimated ablation depth. 10,11,13,17,18,20,22,24 Myopic patients have an increased risk of developing glaucoma. 5,15,24,32 Assessment of visual fields and cup-disc ratios in these patients is more difficult than in patients without myopia. 19 Monitoring of IOP in these patients is important for detecting ocular hypertension, steroid-induced glaucoma, or primary glaucoma. Currently, Goldmann applanation tonometry is considered the standard method for measurement of IOP. 3,4 However, Goldmann applanation tonometry measurements were found to be inaccurate after LASIK. Ophthalmologists should keep in mind the underestimation of IOP by Goldmann applanation tonometry when treating myopic patients who have undergone LASIK. Postoperative of IOP at the temporal part of the cornea, which usually gave higher readings than the central measurements, were still lower than preoperative measurements. Measuring IOP by Goldmann applanation tonometry at the temporal part of the cornea did not give reliable information about the true postoperative IOP. A more practical approach to assess IOP after LASIK is to take Goldmann applanation tonometry measurements at the center of the cornea in the usual technique and then add 3 to 4 mmhg to the reading, or use a prediction formula (similar to that in this study) to calculate the approximate value of the true IOP. It is advisable that each patient who undergoes corneal refractive surgery should be given a record of his Goldmann applanation tonometry measurements before and after surgery. Such 426 Journal of Refractive Surgery Volume 17 July/August 2001

8 a record will be useful to compare future measurements. The results of this study indicate that Goldmann applanation tonometry measurements, obtained in the standard technique from the center of the cornea, will underestimate IOP after LASIK for myopia (by a mean of 3.69 ± 1.63 mmhg in this study). This underestimation of IOP persisted throughout the 12-month duration of this study, and was correlated with preoperative IOP and decrease in the central corneal thickness after LASIK. This may delay the diagnosis and proper management of future glaucoma in these myopic patients. With increasing numbers of refractive surgery procedures for myopia, there is an increased need for awareness of underestimation of IOP by Goldmann applanation tonometry after LASIK. Although Goldmann applanation tonometry measurements from the temporal part of the cornea usually gave higher readings than central measurements after LASIK, these temporal measurements were not reliable and also underestimated postoperative IOP. There is a need for a more accurate tonometer to assess true IOP after LASIK. Patients with ocular hypertension or glaucoma should not have LASIK until a valid method for evaluating IOP after corneal refractive surgery is available. REFERENCES 1. Rashad KM. Laser assisted in situ keratomileusis (LASIK) for correction of myopia. Egypt J Cataract Refract Surg 1996;2: Rashad KM. Laser in situ keratomileusis for myopic astigmatism. J Refract Surg 1999;15: Midelfart A, Wigers A. Clinical comparison of the Pro Ton and Tono-Pen tonometers with the Goldmann applanation tonometer. Br J Ophthalmol 1994;78: Holladay JT, Allison ME, Prager TC. Goldmann applanation tonometry in patients with regular corneal astigmatism. Am J Ophthalmol 1983;96: Kohlhaas M, Lerche RC, Draeger J, Klemm M, Ehlers N, Hjortdal J, Olsen H, Barraquer C, Barraquer JI, Flicker D, Rivera F, Carriazo C. The influence of corneal thickness and corneal curvature on tonometry readings after corneal refractive surgery. Eur J Implant Ref Surg 1995;7: Shah S, Chatterjee A, Mathai M, Kelly SP, Kwartz J, Henson D, McLeod D. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology 1999;106: Ehlers N, Hansen FK. Central corneal thickness in low-tension glaucoma. Acta Ophthalmol 1974;52: Johnson M, Kass MA, Moses RA, Grodzki WJ. Increased corneal thickness simulating elevated intraocular pressure. Arch Ophthalmol 1978;96: Mark HH. Corneal curvature in applanation tonometry. Am J Ophthalmol 1973;76: Faucher A, Gregoire J, Blondeau P. Accuracy of Goldmann tonometry after refractive surgery. J Cataract Refract Surg 1997;23: Fournier AV, Podtetenev M, Lemire J, Thompson P, Duchesne R, Perreault C, Chehade N, Blondeau P. Intraocular pressure change measured by Goldmann tonometry after laser in situ keratomileusis. J Cataract Refract Surg 1998;24: Price FW, Koller DL, Price MO. Central corneal pachymetry in patients undergoing laser in situ keratomileusis. Ophthalmology 1999;106: Emara B, Probst LE, Tingey DP, Kennedy DW, Williams LJ, Machat J. Correlation of intraocular pressure and corneal thickness in normal myopic eyes and after laser in situ keratomileusis. J Cataract Refract Surg 1998;24: Whitacre MM, Stein R. Sources of error with the use of Goldmann-type tonometers. Surv Ophthalmol 1993;38: Phelan PS, McGhee CN, Bryce IG. Excimer laser PRK and corticosteroid induced IOP elevation: the tip of an emerging iceberg. Br J Ophthalmol 1994;78: Brubaker RF. Tonometry. In: Duane's Ophthalmology, CD- Rom edition. Philadelphia, PA: Lippincott-Raven Publishers;1997:Record Abbasoglu OE, Bowman RW, Cavanagh HD, McCulley JP. Reliability of intraocular pressure measurements after myopic excimer photorefractive keratectomy. Ophthalmology 1998;105: Zadok D, Tran DB, Twa M, Carpenter M, Schanzlin DJ. Pneumotonometry versus Goldmann tonometry after laser in situ keratomileusis for myopia. J Cataract Refract Surg 1999;25: Perez-Santonja JJ, Bellot J, Claramonte P, Ismail MM, Alio JL. Laser in situ keratomileusis to correct high myopia. J Cataract Refract Surg 1997;23: Levy Y, Zadok D, Glovinsky Y, Krakowski D, Nemet P. Tonopen versus Goldmann tonometry after excimer laser photorefractive keratectomy. J Cataract Refract Surg 1999;25: Chatterjee A, Shah S, Bessant DA, Naroo SA, Doyle SJ. Reduction in intraocular pressure after excimer laser photorefractive keratectomy. Correlation with pretreatment myopia. Ophthalmology 1997;104: Mardelli PG, Piebenga LW, Whitacre MM, Siegmund KD. The effect of excimer laser photorefractive keratectomy on intraocular pressure measurements using the Goldmann applanation tonometer. Ophthalmology 1997;104: Rashad K, Kalaway H. Photorefractive keratectomy (PRK) for myopia using a wide ablation zone. Bull Ophthalmol Soc Egypt 1994;87: Schipper I, Senn P, Thomann U, Suppiger M. Intraocular pressure after excimer laser photorefractive keratectomy for myopia. J Refract Surg 1995;11: Probst LE, Machat JJ. Mathematics of laser in situ keratomileusis for high myopia. J Cataract Refract Surg 1998;24: Whitacre MM, Stein R, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993;115: Herndon LW, Choudhri SA, Cox T, Damji KF, Shields MB, Allingham RR. Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes. Arch Ophthalmol 1997;115: Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol 1975;53: Argus WA. Ocular hypertension and central corneal thickness. Ophthalmology 1995;102: Tomlinson A, Leighton DA. Ocular dimensions in low tension glaucoma compared with open angle glaucoma and the normal. Br J Ophthalmol 1972;56: Argento C, Cosentino MJ, Moussalli MA. Intraocular pressure measurement following hyperopic LASIK. J Cataract Refract Surg 1998;24: Dougherty PJ, Wellish KL, Maloney RK. Excimer laser ablation rate and corneal hydration. Am J Ophthalmol 1994;118: Journal of Refractive Surgery Volume 17 July/August

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany

INTRODUCTION S. HERDENER, D. HAFIZOVIC, M. PACHE, S. LAUTEBACH, J. FUNK. University Eye Hospital, Freiburg - Germany European Journal of Ophthalmology / Vol. 18 no. 1, 2008 / pp. 39-43 Is the PASCAL -Tonometer suitable for measuring intraocular pressure in clinical routine? Long- and short-term reproducibility of dynamic

More information

Laser in situ keratomileusis (LASIK) has proven to be

Laser in situ keratomileusis (LASIK) has proven to be Autorefractometry after laser in situ keratomileusis Dimitrios S. Siganos, MD, PhD, Corina Popescu, MD, Nikolaos Bessis, DOpt, Georgios Papastergiou, MD Purpose: To correlate cycloplegic subjective refraction

More information

CLINICAL SCIENCES. Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry

CLINICAL SCIENCES. Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry CLINICAL SCIENCES Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry Markus Kohlhaas, MD; Andreas G. Boehm, MD; Eberhard Spoerl, PhD; Antje Pürsten, Dipl-Ing

More information

Central corneal thickness determined with optical coherence tomography in various types of glaucoma

Central corneal thickness determined with optical coherence tomography in various types of glaucoma Br J Ophthalmol 2000;84:1233 1237 1233 Department of Ophthalmology, Ludwig-Maximilians- University, Munich, Germany M Bechmann M J Thiel B Roesen S Ullrich M W Ulbig K Ludwig Correspondence to: Martin

More information

The two currently accepted methods for correcting

The two currently accepted methods for correcting New Technique Therapeutic Alloplastic Laser in situ Keratomileusis for Myopia Arturo Maldonado-Bas, MD; Ruben Pulido-Garcia, MD ABSTRACT BACKGROUND: A new technique, therapeutic alloplastic laser in situ

More information

Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia

Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia Seyed Mohammad Reza Taheri, MD 1 Azita Kheiltash, MD, MPH 2 Hassan Hashemi, MD 1,3 Abstract Purpose: To

More information

Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer

Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer 518 EXTENDED REPORT Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer P Gunvant, M Baskaran, L Vijaya, I S Joseph, R J Watkins,

More information

Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, and Julian Garcia-Sanchez METHODS

Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, and Julian Garcia-Sanchez METHODS Ocular Response Analyzer versus Goldmann Applanation Tonometry for Intraocular Pressure Measurements Jose M. Martinez-de-la-Casa, Julian Garcia-Feijoo, Ana Fernandez-Vidal, Carmen Mendez-Hernandez, and

More information

Recent evidence has stressed the clinical role of the assessment

Recent evidence has stressed the clinical role of the assessment Underestimate of Tonometric Readings after Photorefractive Keratectomy Increases at Higher Intraocular Pressure Levels Ciro Tamburrelli, Andrea Giudiceandrea, Agostino Salvatore Vaiano, Carmela Grazia

More information

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma

CLINICAL SCIENCES. Clinical Significance of Central Corneal Thickness in the Management of Glaucoma CLINICAL SCIENCES Clinical Significance of Central Corneal Thickness in the Management of Glaucoma Carolyn Y. Shih, MD; Joshua S. Graff Zivin, PhD; Stephen L. Trokel, MD; James C. Tsai, MD Objective: To

More information

Central corneal thickness among glaucoma patients attending Menelik II Hospital, Addis Ababa, Ethiopia

Central corneal thickness among glaucoma patients attending Menelik II Hospital, Addis Ababa, Ethiopia Original article Central corneal thickness among glaucoma patients attending Menelik II Hospital, Addis Ababa, Ethiopia Zelalem Gizachew 1, Abiy Mulugeta 1* Abstract Background: Glaucoma is one of the

More information

PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET

PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET 616.365.5775 www.keillasik.com PHOTOREFRACTIVE KERATECTOMY (PRK) PATIENT INFORMATION BOOKLET Please read this entire booklet. Discuss its contents with your doctor so that questions are answered to your

More information

Nature and Science 2017;15(11) Mohamed Elmoddather. MD

Nature and Science 2017;15(11)   Mohamed Elmoddather. MD Outcome of PRK in Management of Post LISIK Residual Myopia and Myopic Astigmatism Mohamed Elmoddather. MD Ophthalmology Faculty of Medicine, Al-Azhar University, Assuit, Egypt shahdmsaleh@hotmail.com Abstract:

More information

Cataract Surgery in the Patient with a History of LASIK or PRK

Cataract Surgery in the Patient with a History of LASIK or PRK Cataract Surgery in the Patient with a History of LASIK or PRK #56996-RS April 2018 Sebastian Lesniak, MD Matossian Eye Associates None Disclosures Bio Matossian Eye Associates, Hopewell NJ, 7/2015 Present

More information

Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery

Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery DOI 10.1007/s00417-010-1431-9 REFRACTIVE SURGERY Comparison between Pascal dynamic contour tonometer and Goldmann applanation tonometer after different types of refractive surgery Antonios P. Aristeidou

More information

Recent concerns regarding the depth of tissue ablation with

Recent concerns regarding the depth of tissue ablation with Volume Estimation of Excimer Laser Tissue Ablation for Correction of Spherical Myopia and Hyperopia Damien Gatinel, 1 Thanh Hoang-Xuan, 1 and Dimitri T. Azar 1,2 PURPOSE. To determine the theoretical volumes

More information

Summary Recommendations for Keratorefractive Laser Surgery June 2013

Summary Recommendations for Keratorefractive Laser Surgery June 2013 Summary Recommendations for Keratorefractive Laser Surgery June 2013 Background Laser assisted in-situ keratomileusis (LASIK) surgery is the most commonly performed keratorefractive surgery; altering the

More information

Four-year Postoperative Results of the US ALLEGRETTO WAVE Clinical Trial for the Treatment of Hyperopia

Four-year Postoperative Results of the US ALLEGRETTO WAVE Clinical Trial for the Treatment of Hyperopia Four-year Postoperative Results of the US ALLEGRETTO WAVE Clinical Trial for the Treatment of Hyperopia Guy M. Kezirian, MD, FACS; Charles R. Moore, MD, FICS; Karl G. Stonecipher, MD; SurgiVision Consultants

More information

펨토초레이저와미세각막절삭기를이용한근시교정수술에서임상성적비교

펨토초레이저와미세각막절삭기를이용한근시교정수술에서임상성적비교 펨토초레이저와미세각막절삭기를이용한근시교정수술에서임상성적비교 4 Table. Summary of data about patient who had LASIK operation using second laser and keratome Age (years) Male:Female (eyes) Pre-op. IOP (mmhg) Pre-op SE * (D) Pre-op

More information

The Egyptian Journal of Hospital Medicine (January 2018) Vol. 70 (8), Page

The Egyptian Journal of Hospital Medicine (January 2018) Vol. 70 (8), Page The Egyptian Journal of Hospital Medicine (January 2018) Vol. 70 (8), Page 1371-1376 Relationship Between Corneal Astigmatism and Intraocular Pressure Abdulrahman E. Algarni 1, Mohammad Abahussin 2, Nuha

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Pachymetry Page 1 of 8 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Pachymetry Professional Institutional Original Effective Date: March 11, 2004 Original Effective

More information

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation

Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation Management of Unpredictable Post-PRK Corneal Ectasia with Intacs Implantation Mohammad Naser Hashemian, MD 1 Mahdi AliZadeh, MD 2 Hassan Hashemi, MD 1,3 Firoozeh Rahimi, MD 4 Abstract Purpose: To present

More information

The pinnacle of refractive performance.

The pinnacle of refractive performance. The pinnacle of refractive performance. WaveLight REFRACTIVE PORTFOLIO Advancing REFRACTIVE SURGERY Contoura Vision sets a new standard in LASIK outcomes More than 98% of patients would choose it again.

More information

Photorefractive Keratectomy as A Retreatment of Residual Myopia after Previous Laser in Situ Keratomileusis

Photorefractive Keratectomy as A Retreatment of Residual Myopia after Previous Laser in Situ Keratomileusis Photorefractive Keratectomy as A Retreatment of Residual Myopia after Previous Laser in Situ Keratomileusis Alireza Ashtari, MD 1 Hassan Razmju, MD 2 Amin Masjedi, MD 3 Alireza Zandi, MD 1 Mehdi Tavakoli,

More information

Wavefront-optimized Versus Wavefrontguided LASIK for Myopic Astigmatism With the ALLEGRETTO WAVE: Three-month Results of a Prospective FDA Trial

Wavefront-optimized Versus Wavefrontguided LASIK for Myopic Astigmatism With the ALLEGRETTO WAVE: Three-month Results of a Prospective FDA Trial Wavefront-optimized Versus Wavefrontguided LASIK for Myopic Astigmatism With the ALLEGRETTO WAVE: Three-month Results of a Prospective FDA Trial Karl G. Stonecipher, MD; Guy M. Kezirian, MD, FACS ABSTRACT

More information

Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes

Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes (2009) 23, 262 269 & 2009 Macmillan Publishers Limited All rights reserved 0950-222X/09 $32.00 www.nature.com/eye CLINICAL STUDY Comparison of Goldmann applanation tonometry and dynamic contour tonometry

More information

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract

Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes. Abstract Effects of Intracorneal Ring Segment on Corneal Biomechanics in Keratoconic Eyes Javad Amoozadeh, MD 1 Nima Mirzaee Rad, MD 2 Amir Houshang Beheshtnejad, MD 3 Ahmad Kheirkhah, MD 1 Hassan Hashemi, MD 4,5

More information

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by Pentacam as A New Imaging Technique Mohammad Ali Zare, MD 1 Hassan

More information

Dina H. Erickson, O.D., a Denise Goodwin, O.D., a Michael Rollins, O.D., b Amber Belaustegui, O.D., c and Chad Anderson a

Dina H. Erickson, O.D., a Denise Goodwin, O.D., a Michael Rollins, O.D., b Amber Belaustegui, O.D., c and Chad Anderson a Optometry (2009) 80, 169-174 Comparison of dynamic contour tonometry and Goldmann applanation tonometry and their relationship to corneal properties, refractive error, and ocular pulse amplitude Dina H.

More information

Flap characteristics, predictability, and safety of the Ziemer FEMTO LDV femtosecond laser with the disposable suction ring for LASIK

Flap characteristics, predictability, and safety of the Ziemer FEMTO LDV femtosecond laser with the disposable suction ring for LASIK (2014) 28, 66 71 & 2014 Macmillan Publishers Limited All rights reserved 0950-222X/14 www.nature.com/eye CLINICAL STUDY Flap characteristics, predictability, and safety of the Ziemer FEMTO LDV femtosecond

More information

Nature and Science 2016;14(9)

Nature and Science 2016;14(9) Comparison between Implantable Contact Lens (ICL) versus Acry Sof Cachet Phakic Intraocular Lenses in Correction of Moderate to High Myopia Khaled Nada, M. Sc.; Mohammed Attia, M. D and Ashraf El Habbak,

More information

December 2016, Volume: 8, Issue: 12, Pages: , DOI:

December 2016, Volume: 8, Issue: 12, Pages: , DOI: Electronic Physician (ISSN: 2008-5842) http://www.ephysician.ir December 2016, Volume: 8, Issue: 12, Pages: 3429-3433, DOI: http://dx.doi.org/10.19082/3429 Refinement of Intraocular Pressure Measurements

More information

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND,, PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including

More information

Description of iatrogenic corneal ectasia in patients without traditional risk factors

Description of iatrogenic corneal ectasia in patients without traditional risk factors ARTICLE Description of iatrogenic corneal ectasia in patients without traditional risk factors Julio Ortega-Usobiaga, MD, PhD 1 ; Rosario Cobo-Soriano, MD, PhD 1 ; Fernando Llovet-Osuna, MD, PhD 1 ; Stephan

More information

Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery

Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery The Egyptian Journal of Hospital Medicine (April 2018) Vol. 71 (3), Page 2636-2642 Evaluation of The Accuracy of Sub Bowman s Keratomileusis (SBK) Microkeratome in Flap Creation during Lasik Surgery Hager

More information

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page 7580-7586 Outcome of PRK in Management of Post LISIK Residual Myopia and Myopic Astigmatism Department of Ophthalmology, Faculty

More information

CATARACT SURGERY AFTER RADIAL KERATOTOMY

CATARACT SURGERY AFTER RADIAL KERATOTOMY AFTER RADIAL KERATOTOMY How to avoid disappointment. BY BERNARD MATHYS, MD CATARACT SURGERY Radial keratotomy (RK; Figure 1) was a popular refractive surgical procedure to correct myopia in the 1970s and

More information

Ocular Rigidity Evaluation After Photorefractive Keratectomy: An Experimental Study

Ocular Rigidity Evaluation After Photorefractive Keratectomy: An Experimental Study BIOMECHANICS Ocular Rigidity Evaluation After Photorefractive Keratectomy: An Experimental Study George D. Kymionis, MD, PhD; Vasilios F. Diakonis, MD; George Kounis, BSc; Spyridon Charisis, MD; Dimitrios

More information

For approximately two decades photorefractive keratectomy. Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK.

For approximately two decades photorefractive keratectomy. Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK. Cornea Seven-Year Changes in Corneal Power and Aberrations after PRK or LASIK Anders Ivarsen and Jesper Hjortdal PURPOSE. To examine long-term changes in corneal power and aberrations in myopic patients

More information

Comparison of Intraocular Pressure Measurements by the Ocular Response Analyzer and Goldmann Applanation Tonometer after Penetrating Keratoplasty

Comparison of Intraocular Pressure Measurements by the Ocular Response Analyzer and Goldmann Applanation Tonometer after Penetrating Keratoplasty Comparison of Intraocular Pressure Measurements by the Ocular Response Analyzer and Goldmann Applanation Tonometer after Penetrating Keratoplasty in Keratoconic Patients Sepehr Feizi, MD 1 Mohammad Pakravan,

More information

Correlating central corneal thickness and intraocular pressure in ocular hypertension and glaucoma

Correlating central corneal thickness and intraocular pressure in ocular hypertension and glaucoma VOL. 3 NO. 1 PHILIPPINE JOURNAL OF Ophthalmology JANUARY ORIGINAL ARTICLE JUNE 07 Jonathan G. Soriano, MD 1 Ma. Margarita L. Lat-Luna, MD 1, 3 Patricia M. Khu, MD 1, 1 Department of Ophthalmology and Visual

More information

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis Han-Yin Sun 1, 2, 3, Hsiu-Wan Yang 4, I-Tsung Wu 4, Jung-Kai Tseng 2, 3 1, 5* and Shun-Fa

More information

Author s Affiliation. Original Article. Visual outcomes after LASIK (laser-assisted in-situ keratomileusis) for various refractive errors.

Author s Affiliation. Original Article. Visual outcomes after LASIK (laser-assisted in-situ keratomileusis) for various refractive errors. Original Article Visual outcomes after LASIK (laser-assisted in-situ keratomileusis) for various refractive errors. Author s Affiliation Sobia Tufail Imran Ahmad Asad Aslam Khan Correspondence Author:

More information

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Jack T. Holladay, MD, MSEE, FACS; Warren E. Hill, MD, FACS; Andreas Steinmueller, MSc ABSTRACT PURPOSE:

More information

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy Original Article Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy Mahmoud Jabbarvand 1, MD; Farshad Askarizadeh 2, PhD; Mohamad Reza Sedaghat 3, MD Hadi Ghadimi

More information

L. Spadea, R. Ferrante, F. Romani, A. Di Gregorio

L. Spadea, R. Ferrante, F. Romani, A. Di Gregorio University of L Aquila Eye Clinic Head: Prof. Leopoldo Spadea ULTRAFAST EXCIMER LASER FOR TRANS-EPITHELIAL CUSTOMIZED PHOTOREFRACTIVE SURGERIES: CLINICAL RESULTS WITH 6 MONTHS FOLLOW UP L. Spadea, R. Ferrante,

More information

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus 273 This is an Open Access article licensed under the terms of the Creative Commons Attribution- NonCommercial-NoDerivs 3.0 License (www.karger.com/oa-license), applicable to the online version of the

More information

POST-LASIK ECTASIA MANAGEMENT

POST-LASIK ECTASIA MANAGEMENT POST-LASIK ECTASIA MANAGEMENT A. John Kanellopoulos MD 1,2 1: Laservision.gr Clinical & Research Eye Institute, Athens, Greece 2: NYU Medical School Department of Ophthalmology, NY, NY Financial interests:

More information

Central corneal thickness and vascular risk factors in normal tension glaucoma

Central corneal thickness and vascular risk factors in normal tension glaucoma Central corneal thickness and vascular risk factors in normal tension glaucoma Aoife Doyle, Ahmed Bensaid and Yves Lachkar L Institut du Glaucome, Fondation Hoˆpital St. Joseph, Paris, France ABSTRACT.

More information

Contoura TM Vision Correction

Contoura TM Vision Correction Contoura TM Vision Correction Fernando Faria Correia, Financial Disclosures: Alcon/Wavelight Cairo (Egypt) 26/01/2018 Topography-guided ablations Topography guided ablations Evolution from complicated

More information

Intro to Glaucoma/2006

Intro to Glaucoma/2006 Intro to Glaucoma/2006 Managing Patients with Glaucoma is Exciting Interesting Challenging But can often be frustrating! Clinical Challenges To identify patients with risk factors for possible glaucoma.

More information

The Association of Central corneal thickness with Intra-ocular Pressure and Refractive Error in a Nigerian Population

The Association of Central corneal thickness with Intra-ocular Pressure and Refractive Error in a Nigerian Population Peer Reviewed, Open Access, Free Published Quarterly Mangalore, South India ISSN 0972-5997 Volume 6, Issue 3; Jul-Sep 2007 Original Article The Association of Central corneal thickness with Intra-ocular

More information

Three-year Follow-up of the Artisan Phakic Intraocular Lens for Hypermetropia

Three-year Follow-up of the Artisan Phakic Intraocular Lens for Hypermetropia Three-year Follow-up of the Artisan Phakic Intraocular Lens for Hypermetropia Ruchi Saxena, MS, 1 Monika Landesz, MD, PhD, 2 Bastiaantje Noordzij, 1 Gregorius P. M. Luyten, MD, PhD 1 Purpose: We report

More information

The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery

The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery 접수번호 :08-024 Korean Journal of Ophthalmology 2009;23:1-5 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.1.1 The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive

More information

Comparison of Intraocular Lens Power Calculation Using the Binkhorst and SRK Formulae: A Clinical Study

Comparison of Intraocular Lens Power Calculation Using the Binkhorst and SRK Formulae: A Clinical Study Comparison of Intraocular Lens Power Calculation Using the Binkhorst and SRK Formulae: A Clinical Study Pages with reference to book, From 38 To 40 Manzoor Hussain, Jehangir Durrani ( Department of Ophthalmology,

More information

Intraocular pressure changes and relationship. Corneal Biomechanics After SMILE and FS-LASIK

Intraocular pressure changes and relationship. Corneal Biomechanics After SMILE and FS-LASIK Cornea Intraocular Pressure Changes and Relationship With Corneal Biomechanics After SMILE and FS-LASIK Hua Li, Yan Wang, Rui Dou, Pinghui Wei, Jiamei Zhang, Wei Zhao, and Liuyang Li Tianjin Eye Hospital

More information

CLINICAL SCIENCES. Management of Post-LASIK Corneal Ectasia With Intacs Inserts

CLINICAL SCIENCES. Management of Post-LASIK Corneal Ectasia With Intacs Inserts Management of Post-LASIK Corneal Ectasia With Intacs Inserts One-Year Results CLINICAL SCIENCES George D. Kymionis, MD, PhD; Charalambos S. Siganos, MD, PhD; George Kounis, BSc; Nikolaos Astyrakakis, OD;

More information

Simultaneous Topography-guided Surface Ablation with Collagen Cross-linking for Keratoconus

Simultaneous Topography-guided Surface Ablation with Collagen Cross-linking for Keratoconus IJKECD Case series Simultaneous Topography-guided Surface Ablation with Collagen 10.5005/jp-journals-10025-1124 Cross-linking for Keratoconus Simultaneous Topography-guided Surface Ablation with Collagen

More information

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (10), Page

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (10), Page The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (10), Page 7707-7712 LASIK surgery in pediatric anisometropia Hany M. Bayoumy, Ihab A. El-shaikh, Nour-eldeen A. Salah * Department of Ophthalmology,

More information

Wavefront-Optimized Technology in Hyperopic Correction Stability Using Different Optical Zones

Wavefront-Optimized Technology in Hyperopic Correction Stability Using Different Optical Zones Med. J. Cairo Univ., Vol. 78, No. 1, December 559-563, 2010 www.medicaljournalofcairouniversity.com Wavefront-Optimized Technology in Hyperopic Correction Stability Using Different Optical Zones ADNAN

More information

KNOW THE OPTIONS. Discover how the latest advances in vision correction can improve your sight.

KNOW THE OPTIONS. Discover how the latest advances in vision correction can improve your sight. KNOW THE OPTIONS. LASIK VISIAN ICL PRK Discover how the latest advances in vision correction can improve your sight. Today, you can choose from several choices of permanent vision correction procedures

More information

Comparison of Corneal Shape Changes and Aberrations Induced By FS-LASIK and SMILE for Myopia

Comparison of Corneal Shape Changes and Aberrations Induced By FS-LASIK and SMILE for Myopia ORIGINAL ARTICLE Comparison of Corneal Shape Changes and Aberrations Induced By FS-LASIK and SMILE for Myopia Anders Gyldenkerne, MS; Anders Ivarsen, MD, PhD; Jesper Ø. Hjortdal, MD, PhD ABSTRACT PURPOSE:

More information

Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma

Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma Original Article DOI 10.3349/ymj.2009.50.2.206 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 50(2):206-210, 2009 Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure

More information

Our experience with Athens protocol - simultaneous topo-guided photorefractive keratectomy followed by corneal collagen cross linking for keratoconus

Our experience with Athens protocol - simultaneous topo-guided photorefractive keratectomy followed by corneal collagen cross linking for keratoconus International Journal of Research in Medical Sciences Shah S et al. Int J Res Med Sci. 2016 Jul;4(7):2639-2644 www.msjonline.org pissn 2320-6071 eissn 2320-6012 Research Article DOI: http://dx.doi.org/10.18203/2320-6012.ijrms20161924

More information

Circadian variations in central corneal thickness and intraocular pressure in patients with glaucoma

Circadian variations in central corneal thickness and intraocular pressure in patients with glaucoma SCIENTIFIC REPORT Circadian variations in central corneal thickness and intraocular pressure in patients with glaucoma P Fogagnolo, L Rossetti, F Mazzolani, N Orzalesi... Br J Ophthalmol ;9:. doi: 1.113/bjo.5.795

More information

IOL Master. Photorefractive keratectomy

IOL Master. Photorefractive keratectomy 89/11/3 : 89/9/3 : 1389 /119 / IOL Master * Photorefractive keratectomy 3 2 1 2 4 Photorefractive keratectomy(prk) (IOL) :. IoL-Master.. (clinical trial) :. 30 20 35.. PRK. IOL Master ( GM300,china) IOL.

More information

Predictability and accuracy of IOL formulas in high myopia

Predictability and accuracy of IOL formulas in high myopia ARTICLE Predictability and accuracy of IOL formulas in high myopia Mohamed Yasser Sayed Saif, MD 1 ; Mohamed Othman Abdel Khalek, MD 1 ; Ahmed Tamer Sayed Saif, MD 2 ; Passant Sayed Saif, MD 3 ; Sherif

More information

Refractive corneal surgery (RCS) using a laser, such as. Glaucoma Progression in Eyes with a History of Refractive Corneal Surgery.

Refractive corneal surgery (RCS) using a laser, such as. Glaucoma Progression in Eyes with a History of Refractive Corneal Surgery. Glaucoma Glaucoma Progression in Eyes with a History of Refractive Corneal Surgery Yoon Jeon Kim, 1 Sung-Cheol Yun, 2 Jung Hwa Na, 3 Hung Won Tchah, 1 Jong Jin Jung, 3 and Kyung Rim Sung 1 PURPOSE. To

More information

The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology

The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology 1395 SCIENTIFIC REPORT The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology A C Browning, A Bhan, A P Rotchford, S Shah, H S Dua... Background/aim: To

More information

Role of central corneal thickness measurement in management of open angle glaucoma and glaucoma suspects in Calabar, Nigeria

Role of central corneal thickness measurement in management of open angle glaucoma and glaucoma suspects in Calabar, Nigeria Original Research Article Role of central corneal thickness measurement in management of open angle glaucoma and glaucoma suspects in Calabar, Nigeria Nkanga DG 1,2, Ibanga AA 1,2, Nkanga ED 2, Etim BA

More information

aberration induced by laser

aberration induced by laser How is spherical aberration induced by laser refractive surgery? Geunyoung Yoon, PhD 1 Ian Cox, PhD 2 Scott MacRae,, MD 1 1 Department of Ophthalmology, Center for Visual Science University of Rochester,

More information

LASIK for 6.00 to D of Myopia With up to 3.00 D of Cylinder Using the ALLEGRETTO WAVE: 3- and 6-month Results With the 200- and 400-Hz Platforms

LASIK for 6.00 to D of Myopia With up to 3.00 D of Cylinder Using the ALLEGRETTO WAVE: 3- and 6-month Results With the 200- and 400-Hz Platforms LASIK for 6.00 to 12.00 D of Myopia With up to 3.00 D of Cylinder Using the ALLEGRETTO WAVE: 3- and 6-month Results With the 200- and 400-Hz Platforms Karl G. Stonecipher, MD; Guy M. Kezirian, MD; Megan

More information

ALTERNATIVES TO PHAKIC IMPLANT SURGERY

ALTERNATIVES TO PHAKIC IMPLANT SURGERY Visian ICL Consent INTRODUCTION This information is being provided to you so that you can make an informed decision about having eye surgery to reduce or eliminate your nearsightedness. Only you and your

More information

INFORMED CONSENT FOR PHOTOREFRACTIVE KERATECTOMY (PRK) AND ADVANCE SURFACE ABLATION (ASA)

INFORMED CONSENT FOR PHOTOREFRACTIVE KERATECTOMY (PRK) AND ADVANCE SURFACE ABLATION (ASA) INFORMED CONSENT FOR PHOTOREFRACTIVE KERATECTOMY (PRK) AND ADVANCE SURFACE ABLATION (ASA) This information and the Patient Information booklet must be reviewed so you can make an informed decision regarding

More information

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page

The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page The Egyptian Journal of Hospital Medicine (October 2018) Vol. 73 (9), Page 7574-7579 Biomechanical Corneal Changes Post LASIK with Mechanical Microkeratome Flap versus Femtosecond Flap Department of Ophthalmology,

More information

Keratoconus Clinic. Optometric Co-management Opportunities

Keratoconus Clinic. Optometric Co-management Opportunities Keratoconus Clinic Optometric Co-management Opportunities The Bochner Eye Institute established the first Keratoconus Clinic in Canada in 2008. The consultation and advanced imaging are OHIP covered. All

More information

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Learn Connect Succeed. JCAHPO Regional Meetings 2017 Learn Connect Succeed JCAHPO Regional Meetings 2017 Faculty Biometry and IOL Calculations ASCRS and ASOA Symposium and Congress Los Angeles, CA Daniel H. Chang, M.D. - Empire Eye and Laser Center Bakersfield,

More information

NIIOS. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece VASILIS S.

NIIOS. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece VASILIS S. Cornea Specialist, Melles Cornea Clinic, NIIOS, Rotterdam, the Netherlands Naval Hospital, Athens, Greece Goldman applanation tonometry is still the gold standard for measuring IOP, but its accuracy is

More information

Laser-Assisted In Situ Keratomileusis for Patients With Dry Eye

Laser-Assisted In Situ Keratomileusis for Patients With Dry Eye CLINICAL SCIENCES Laser-Assisted In Situ Keratomileusis for Patients With Dry Eye Ikuko Toda, MD; Naoko Asano-Kato, MD; Yoshiko Hori-Komai, MD; Kazuo Tsubota, MD Objective: To evaluate the efficacy and

More information

Laser in Situ Keratomileusis versus Laser Assisted Subepithelial Keratectomy for the Correction of Low to Moderate Myopia and Astigmatism

Laser in Situ Keratomileusis versus Laser Assisted Subepithelial Keratectomy for the Correction of Low to Moderate Myopia and Astigmatism Laser in Situ Keratomileusis versus Laser Assisted Subepithelial Keratectomy for the Correction of Low to Moderate Myopia and Astigmatism Seyed Javad Hashemian, MD 1 Hossein Aghaei, MD 2 Alireza Foroutan,

More information

Trabeculectomy is an effective method for lowering

Trabeculectomy is an effective method for lowering ORIGINAL STUDY Refractive Outcome of Cataract Surgery in Eyes With Prior Trabeculectomy: Risk Factors for Postoperative Myopia Oliver L. Yeh, MD, Karine D. Bojikian, MD, Mark A. Slabaugh, MD, and Philip

More information

Changes in the anterior cornea during the early stages of severe myopia prior to and following LASIK, as detected by confocal microscopy

Changes in the anterior cornea during the early stages of severe myopia prior to and following LASIK, as detected by confocal microscopy EXPERIMENTAL AND THERAPEUTIC MEDICINE 14: 2869-2874, 2017 Changes in the anterior cornea during the early stages of severe myopia prior to and following LASIK, as detected by confocal microscopy JINRONG

More information

Outcomes of NIDEK Optical Path Difference Custom Ablation Treatments (OPDCAT) for Myopia With or Without Astigmatism

Outcomes of NIDEK Optical Path Difference Custom Ablation Treatments (OPDCAT) for Myopia With or Without Astigmatism Outcomes of NIDEK Optical Path Difference Custom Ablation Treatments (OPDCAT) for Myopia With or Without Astigmatism Shihao Chen, MD, OD, MS; Yibo Wang, MD; Qinmei Wang, MD ABSTRACT PURPOSE: To report

More information

Number 80. Laser Eye Surgery in Myopia. Date of decision October 2017 Date of review October 2020 GUIDANCE

Number 80. Laser Eye Surgery in Myopia. Date of decision October 2017 Date of review October 2020 GUIDANCE Excluded: Procedure not routinely funded Bedfordshire, Hertfordshire, West Essex, Luton and Milton Keynes Priorities Forum statement - adapted for Bedfordshire CCG Number 80 Subject Date of decision October

More information

Retrospective Testing of the Score for the Detection of Ectasia Susceptibility: A Case Report of Ectasia 7 Years after LASIK

Retrospective Testing of the Score for the Detection of Ectasia Susceptibility: A Case Report of Ectasia 7 Years after LASIK IJKECD 10.5005/jp-journals-10025-1055 Case Retrospective Report Testing of the Score for the Detection of Ectasia Susceptibility: Case Report of Ectasia 7 Years after LSIK Retrospective Testing of the

More information

Refractive Dilemma. Challenging Case

Refractive Dilemma. Challenging Case Challenging Case Refractive Dilemma Section Editor: Alireza Baradaran-Rafii, MD Case presentation A 21-year old man was referred to an ophthalmology clinic insisting on getting rid of his glasses which

More information

LASER IN SITU KERATOMILEUSIS (LASIK) HAS BECOME

LASER IN SITU KERATOMILEUSIS (LASIK) HAS BECOME A Predictive Model for Postoperative Intraocular Pressure Among Patients Undergoing Laser in Situ Keratomileusis (LASIK) CHIA-CHING YANG, MD, I-JONG WANG, MD, YUE-CUNE CHANG, PHD, LUKE LONG-KUANG LIN,

More information

Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia

Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia Issued: September 2013 guidance.nice.org.uk/ipg466 NICE has accredited the process used

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

PresbyMax Outcomes in Myopia, Hyperopia, Emmetropia and Patients post Lasik

PresbyMax Outcomes in Myopia, Hyperopia, Emmetropia and Patients post Lasik PresbyMax Outcomes in Myopia, Hyperopia, Emmetropia and Patients post Lasik SCHWIND eye-tech-solutions Lunch Symposium Prof. Jorge L. Alió MD, PhD. UNIVERSIDAD MIGUEL HERNÁNDEZ VISSUM INSTITUTO OFTALMOLÓGICO

More information

Original Article High myopia as a risk factor for post-lasik ectasia: a case report

Original Article High myopia as a risk factor for post-lasik ectasia: a case report Original Article High myopia as a risk factor for post-lasik ectasia: a case report Mona Harissi-Dagher, MD, a,b Sonja A. F. Frimmel, c and Samir Melki, MD, PhD a,d Author affiliations: a Massachusetts

More information

Comparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera

Comparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(6):497-505 https://doi.org/10.3341/kjo.2018.0008 Original Article Comparison of Intraocular Lens Power Calculation Methods Following Myopic

More information

Introduction. Anders Vestergaard, 1,2 Anders Ivarsen, 1 Sven Asp 1 and Jesper Ø. Hjortdal 1. Acta Ophthalmologica 2013

Introduction. Anders Vestergaard, 1,2 Anders Ivarsen, 1 Sven Asp 1 and Jesper Ø. Hjortdal 1. Acta Ophthalmologica 2013 Femtosecond (FS) laser vision correction procedure for moderate to high myopia: a prospective study of ReLEx Ò flex and comparison with a retrospective study of FS-laser in situ keratomileusis Anders Vestergaard,

More information

Research Article Effect of Flat Cornea on Visual Outcome after LASIK

Research Article Effect of Flat Cornea on Visual Outcome after LASIK Hindawi Publishing Corporation Journal of Ophthalmology Volume 2015, Article ID 794854, 7 pages http://dx.doi.org/10.1155/2015/794854 Research Article Effect of Flat Cornea on Visual Outcome after LASIK

More information

Informed Consent For Refractive Lens Exchange (RLE) For the Correction of Hyperopia (Farsightedness) Or Myopia (Nearsightedness)

Informed Consent For Refractive Lens Exchange (RLE) For the Correction of Hyperopia (Farsightedness) Or Myopia (Nearsightedness) Informed Consent For Refractive Lens Exchange (RLE) For the Correction of Hyperopia (Farsightedness) Or Myopia (Nearsightedness) INTRODUCTION This surgery, called a refractive lens exchange or RLE, involves

More information

Evolution in Visual Freedom.

Evolution in Visual Freedom. Evolution in Visual Freedom. The EVO Visian ICL Advantages Many vision correction procedures promise an improved level of vision, but few vision correction alternatives offer the quality and features

More information

LASIK Flap Thickness Accuracy after Using Mechanical Microkeratome

LASIK Flap Thickness Accuracy after Using Mechanical Microkeratome Med. J. Cairo Univ., Vol. 84, No. 1, June: 733-737, 2016 www.medicaljournalofcairouniversity.net LASIK Flap Thickness Accuracy after Using Mechanical Microkeratome AYA H. FARAG, M.Sc.; RANIA M. SOBHY,

More information

Conductive keratoplasty (CK) (Refractec, Inc.) is a

Conductive keratoplasty (CK) (Refractec, Inc.) is a Conductive keratoplasty to correct residual hyperopia after corneal surgery Ian F. Comaish, FRCOphth, Michael A. Lawless, FRACO, FRCOphth Conductive keratoplasty (CK) is an electrical-current-based technique

More information

Interventional procedures guidance Published: 25 September 2013 nice.org.uk/guidance/ipg466

Interventional procedures guidance Published: 25 September 2013 nice.org.uk/guidance/ipg466 Photochemical corneal collagen cross-linkage using riboflavin and ultraviolet A for keratoconus and keratectasia Interventional procedures guidance Published: 25 September 2013 nice.org.uk/guidance/ipg466

More information