Correlation Between Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry in Glaucoma Suspects and Glaucomatous Eyes

Size: px
Start display at page:

Download "Correlation Between Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry in Glaucoma Suspects and Glaucomatous Eyes"

Transcription

1 Korean J Ophthalmol Vol. 18:89-99, 2004 Correlation Between Frequency Doubling Technology Perimetry and Scanning Laser Polarimetry in Glaucoma Suspects and Glaucomatous Eyes Su Hyun Kim, MD, Hunei Hong, BS*, Hee Jo Koo, BS*, Sung-jae Yang, MD**, Hungwon Tchah, MD**, Michael S Kook, MD** Department of Ophthalmology, University of Soonchunhyang, College of Medicine *Unit for Consulting Biostatistics, Asan Medical Center **Department of Ophthalmology, University of Ulsan, College of Medicine, Asan Medical Center, Seoul, Korea The aim of this study was to determine the relationship between the frequency doubling technology (FDT) screening algorithm and parapapillary retinal nerve fiber layer (RNFL) thickness in the eyes of glaucoma suspects and patients with open angle glaucoma. FDT C20-1 screening program and a scanning laser polarimetry (SLP) system (GDx-NFA) was used to assess 53 glaucomatous eyes, 53 glaucoma suspects and 36 normal control eyes. In glaucomatous eyes, there were correlations between the FDT the screening algorithm and RNFL retardation values in several polarimetric indices, most significantly inferior thickness (r = , P = 0.029). In the eyes of glaucoma suspects, however, we observed no correlation between the FDT results and RNFL retardation values (r = 0.080, P > 0.05, inferior thickness ). In glaucomatous eyes, the abnormal scores obtained with FDT screening program correlated negatively with RNFL retardation values, as measured by SLP. Despite poor correlation between the FDT abnormal score and RNFL retardation value in glaucoma suspects, detection of abnormality using the FDT screening protocol may aid in the assessment of early glaucomatous structural damage. Key words: frequency doubling technology perimetry (FDT), glaucoma, glaucoma suspects, retinal nerve fiber layer (RNFL), scanning laser polarimetry (SLP) INTRODUCTION In patients with glaucoma, detectable structural loss of the retinal nerve fiber layer (RNFL) has been Reprint requests to Michael S Kook, MD, Department of Ophthalmology, University of Ulsan, College of Medicine, Asan Medical Center, pungnap-dong, songpa-gu, Seoul, , Korea. This study was presented in part at the 4 th International Glaucoma Symposium, Barcelona, Spain, 19-22, March, found to precede apparent visual field dysfunction. 1-3 Recent studies have shown that various scanning laser devices, such as the GDx-Nerve Fiber Analyzer (NFA), provide reproducible and objective evaluation of RNFL, and these devices have been used to evaluate differences in retardation between normal and glaucomatous eyes and between normal subjects and patients with ocular hypertension (OHTN). 4-9 In addition, polarimetric measurements of the RNFL thickness may detect preperimetric glaucomatous optic nerve damage. 10

2 90 SH Kim, et al Since the description of frequency doubling illusion in 1996, threshold testing has been shown to detect glaucomatous damage earlier than white-onwhite conventional perimetry FDT screening mode has also shown promise for the detection of glaucoma. This latter technique has a number of additional advantages as a diagnostic device, including rapid assessment, relatively low cost, and easy operation by trained nonphysicians. Using a Humphrey C 24-2 full-threshold algorithm as the definition of glaucoma, the screening mode of FDT has been reported to have a sensitivity of 91% and a specificity of 94% with an abnormal GHT, 14 although it is not yet known if this method can detect very early forms of glaucoma including preperimetric glaucoma. Although an association between functional properties and structural alterations of the nerve fiber layer in glaucoma has been described, attempts to correlate structural changes, as measured by SLP, with functional alterations, as measured by FDT, have been limited in eyes with early glaucomatous damage. We therefore performed a cross-sectional study to evaluate the potential of the FDT screening mode in the early detection of glaucoma by examining RNFL parameters obtained with GDx-NFA in normal controls, glaucoma suspects, and glaucomatous eyes. We also determined the correlation between RNFL retardation, as measured by scanning laser polarimetry(slp), and FDT findings in our series of glaucoma suspects and glaucomatous eyes. MATERIALS AND METHODS All study patients and control subjects were consecutively enrolled and examined from June 2002 to May 2003 at the Asan Medical Center, College of Medicine, University of Ulsan, Seoul, Korea. All subjects had best corrected visual acuity of 20/40 or better. If both eyes of a patient met the inclusion criteria, we randomly selected for 1 eye for inclusion in the study. The study followed the tenets of the Helsinki declaration for research involving human subjects, and informed consent was obtained from all participants. We excluded patients with a history of laser surgery, intraocular surgery in either eye, intracranial abnormalities, established diagnosis of neurological disease, or a lesion upon neurological examination. To reduce the influence of age and high prevalence of glaucoma over the age of 40 years, study subjects were restricted to those between 50 and 75 years. Fifty-three eyes of 53 patients were enrolled as glaucoma suspects, defined as having one or more abnormal locations on FDT C20-1 screening program. They also had suspicious changes of optic nerve head, including cup/disc ratio greater than 0.6, cup/disc ratios higher vertically than horizontally, an asymmetry in cup/disc ratio greater than 0.2 between eyes, or optic disc hemorrhage. The appearance of the RNFL was not taken into account. Each glaucoma suspect had normal open angles on gonioscopy without any evidence of pigmentation, pseudoexfoliation, or other secondary causes of glaucoma on anterior segment examination. Each had normal findings with a Humphrey field analyzer (HFA; Humphrey Zeiss, Dublin, CA, USA) based on our defined criteria. None of the glaucoma suspects had typical visual field defects of neurologic origin including quadrantanopia, hemianopia, altitudinal defects, and central scotomas based on the HFA. Eyes with optic disc abnormalities suggestive of nonglaucomatous origins including congenital anomalies of the optic disk, various types of optic atrophy, and inflammatory and ischemic disorders of the optic nerve head, were excluded from the study. Of the 53 eyes in the glaucoma suspect group, 12 (23%) had IOP greater than or equal to 22mmHg on Goldman applanation tonometry (GAT) at multiple times during the evaluations. The glaucoma group consisted of 53 eyes of 53 patients with open-angle glaucoma with or without IOP elevation over 21 mm Hg. None had secondary open-angle glaucoma due to pigment dispersion syndrome, pseudoexfoliation, traumatic anterior chamber angle recession, or other secondary causes. Patients with glaucoma had to also fulfill the following criteria: presence of glaucomatous optic neuropathy and associated glaucomatous field based on HFA. Glaucomatous optic neuropathy was defined as either cup/disk asymmetry between fellow eyes of greater than 0.2, rim thinning, notching, excavation, or RNFL defect. Of the 53 eyes in the glaucoma group, 13 (24.5%) had normal-tension glaucoma (NTG) based on the following criteria: an untreated

3 CORRELATION BETWEEN FDT AND SLP 91 intraocular pressure less than or equal to 21 mmhg at all times over a 24-hour period, measured every 2 hours on admission, a normal open angle of the anterior chamber, the presence of optic nerve as defined above and glaucomatous visual field changes, and no ocular or systemic disorders that could be responsible for optic nerve damage or visual field changes. Thirty-six healthy volunteers were included in this study as a control group. They were hospital administrative staff members and their relatives. The normal appearance of the optic nerve head in both eyes was confirmed by ophthalmoscopy and stereoscopic photography, and no other significant ocular diseases except for mild age-related cataracts were noted by slit lamp biomicroscopy. None of the control group had a history of IOP elevation greater than 21 mm Hg or ocular trauma. The control subjects also did not have a glaucomatous visual field as defined below. All subjects underwent an HFA C30-2 or 24-2, full-threshold, Glaucoma Hemifield test (GHT) program. All subjects, including controls, had undergone prior visual field examinations more than once, and we used the last visual field test for analysis. All subjects had to meet a less than 25% rate of false-positive, false-negative responses as well as a 25% fixation loss. None of the glaucoma suspects or normal controls exhibited any specific glaucomatous visual field defect, defined as: (1) three adjacent locations decreased by a sensitivity of 5 db each or greater with one decreased by at least 10, (2) two adjacent locations decreased by at least 10 db each, (3) a 10 db difference across the nasal horizontal meridian at two adjacent locations, or (4) GHT outside normal limits. For the FDT testing, all subjects, including normal controls, were tested with distance or bifocal corrective lenses in place. All patients had a refractive error of less than +/- 7 diopters in the sphere or less than +3 diopters in the cylinder. One technician (HWH), trained for the instrument, tested all the subjects, who were given a full explanation together with a chart showing pictures of stimuli. Before the recorded test, the subjects practiced several times. The C-20-1 screening program was used according to the fixed protocol. Test results with good reliability, defined as fixation losses, and false positives of 33% or less were entered for analysis in all subjects. A plot of 17 visual field locations was produced for each eye. If any defect was noted in 17 locations per eye in the first trial after a few practice sessions, the test was repeated 30 minutes later to determine whether there was a learning effect. In these cases, the results of the second test were used as the final FDT data. Three levels of shading indicated mild, moderate, and severe relative loss based on the expected contrast values from a database at multiple levels that would be detected in normal persons of comparable age (99%, 99.5%, and worse than 99.5% level). For quantification of the defect score, we used the FDT results according to the criteria devised by Quigley, in which the total defect score was assessed by the number of abnormal points in 17 zones, including the central fixation zone times the severity of the measured defect on a scale from 1 (mild) to 2 (moderate) to 3 (severe). 14 Regional defect score was calculated according to the same method in each hemifield of eight zones, excluding the central fixation zone. Each patient in the glaucoma suspect group had to have one or more abnormal points somewhere in the field, including the central point, depressed to P value less than 1% with a minimal defect score of 1 and at least one abnormal block. All subjects were examined with a SLP device(gdx-nfa, Laser Diagnostic Technologies Inc, San Diego, CA, USA) by the same experienced operator (HWH). For each eye, 65,536 retinal locations were measured to create a retardation map corresponding to RNFL measurements over a retinal area in 0.7 second. While undergoing scanning, the pupils of the subjects were undilated. We used the mean of three good images for analysis, with an average standard deviation of less than 8 um. As defined by the manufacturer, a good image had even illumination and sharp and well-defined edges of blood vessels, as well as lacking red saturation. The optic disc margin was approximated by a circle or ellipse placed around the inner margin of the peripapillary scleral ring by the same operator. The instrument was then placed a measuring circle or ellipse (10 pixels width) 1.75 disc diameters away and concentric with the margin of the optic disc. Defaulted quadrant positions (supplied by the

4 92 SH Kim, et al Table 1. Patient demographics Characteristics Normal Glaucoma suspects Glaucoma P value controls (N = 36) (N = 53) (N = 53) Age(yr) ± ± ± Refractive error(d) 0.20 ± ± ± Intraocular pressure(mmhg) ± ± ± Cup/Disc ratio 0.28 ± ± ± * HFA MD ± ± ± * HFA PSD 2.00 ± ± ± * HFA CPSD 1.01 ± ± ± * HFA: Humphrey field analyzer, MD: mean deviation, PSD: pattern standard deviation, CPSD: corrected pattern standard deviation, Data are given as mean ± standard deviation, *: statistically significant manufacturer) were applied. The peripapillary band was divided into superior and inferior segments of 120 each, a temporal segment of 50, and a nasal segment of 70. The GDx Analysis program automatically calculated for 16 RNFL parameters symmetry, superior ratio, inferior ratio, superior/ nasal, maximum modulation, ellipse modulation, the number, average thickness, ellipse average, superior average, inferior average, superior integral, and deviation from normal- for printout in each of the four quadrants. All participants underwent FDT testing, HFA examination, and GDx-NFA examination within 1 month of initial clinical evaluation. We compared the FDT defect score, number of abnormal blocks in the total region, FDT test duration, and GDx parameters among the 3 different groups. Since both the defect score and the number of abnormal blocks in the FDT screening testing represented deviations from an age-matched normal database, we correlated each with the average of superior and inferior deviations from normal values (GDx mean deviation ) in the corresponding GDx-NFA printout. Similar correlation analysis was performed between the FDT hemifield defect score and the corresponding superior or inferior quadrant deviation from normal values in the GDx printout. Finally, we performed correlation analysis of the total defect score and the number of abnormal blocks and GDx-NFA indices. Data were statistically analyzed using software package SPSS (Statistical Package for Social Sciences Inc., version 10.0, Chicago, IL, USA). The Kruskall-Wallis test was used to evaluate differences in FDT and GDx-NFA measurement values among the 3 subject groups. Sensitivity and specificity were obtained based on the FDT abnormality criteria (one or more abnormal locations out of 17 locations). Significance of pairwise comparisons was determined using the Mann-Whitney test. Bonferroni s correction was applied to compare multiple parameters when appropriate. Correlation between FDT data and different parameters of GDx- NFA was assessed with the Pearson correlation coefficient. The multiple regression model was used to assess the GDx-NFA individual parameter with dependent variables. Statistical significance was set at P < All numbers are reported as mean +/- standard deviation. RESULTS The number, age, refractive error, last IOP, and C/D ratio of all subjects are presented in Table 1. There were no statistically significant differences in mean age, refractive error, or IOP among the three groups. The total mean deviation (MD), pattern standard deviation (PSD), and corrected pattern standard deviation (CPSD) of the 30-2 or 24-2 HFA program, however, were significantly higher in the glaucomatous eyes than in either of the other two groups (P < for each; Table 1). We found significant differences in mean total defect score and number of abnormal blocks on FDT testing among the 3 groups (Table 2). One or more abnormal locations had a sensitivity of 84.9% (45/53) and a specificity of 100% (36/36). In addition, statistically significant differences in the mean

5 CORRELATION BETWEEN FDT AND SLP 93 Table 2. FDT parameters FDT Normal controls Glaucoma suspects Glaucoma (N = 36) (N = 53) (N = 53) P value Total defect score 0.00 ± ± ± Total number of abnormal blocks 0.00 ± ± ± Test time (sec) ± ± ± FDT: frequency doubling technology perimetry, Data are given as mean ± standard deviation, *statistically significant Table 3. RNFL Measures by GDx NFA Parameters Normal controls Glaucoma suspects Glaucoma P value Mean deviation 2.47 ± ± ± * (sup. + inf. Deviation)/2 Symmetry 0.96 ± ± ± Superior ratio 2.04 ± ± ± * Inferior ratio 2.17 ± ± ± * Superior-nasal ratio 1.77 ± ± ± * Maximum modulation 1.25 ± ± ± * Ellipse modulation 2.29 ± ± ± * GDx number ± ± ± * Average thickness (µm) ± ± ± * Ellipse average (µm) ± ± ± * Superior average (µm) ± ± ± * Inferior average (µm) ± ± ± * Superior integral (micron squared) 0.21 ± ± ± * RNFL: retinal nerve fiber layer, Data are given as mean ± standard deviation, *: statistically significant FDT testing time were found among the 3 groups (Table 2). We also observed significant differences in GDx mean deviation values from the age-matched normal database, as well as 12 RNFL thickness measurements by GDx-NFA, between the normal controls and each of the other two groups in all pairwise comparisons except symmetry (P < 0.001) (Tables 3 and 4). In contrast, no differences in RNFL parameters were found between glaucoma suspects (normal SAP and abnormal FDT) and glaucomatous eyes (abnormal SAP) except inferior ratio. Except for symmetry, all measured values for the glaucoma suspects were intermediate between the measurements for the normal controls and those for glaucomatous eyes. Although the means of the measured variables differed significantly between the normal controls and each of the other two groups, there were noticeable overlaps between controls and glaucoma suspects, and between glaucoma suspects and glaucomatous eyes, for all GDx parameters (Table 3). In the glaucomatous eyes, we observed negative linear correlations between total FDT defect score and GDx mean deviation, and between total number of abnormal blocks and GDx mean deviation (P < 0.05, Fig. 1). Similar to the full field, FDT defect score in the superior hemifield was negatively correlated with inferior GDx deviation value in glaucomatous eyes (P = 0.01, Fig. 2). In contrast, we found no significant association in the glaucoma suspects between GDx mean deviation and either FDT total defect score or the number of abnormal blocks (Fig. 1). Furthermore, we observed poor regional correlation between FDT hemifield abnormal score and corresponding GDx deviation value in these eyes (Fig. 2). When we analyzed FDT abnormal data and the 12 GDx-NFA parameters (Tables 5, 6), we found

6 94 SH Kim, et al Fig. 1. Correlation between FDT abnormality and GDx Mean deviation. Fig. 2. Correlation between FDT hemifield defect score and corresponding quadrant GDx deviation. Table 4. P value for pairwise comparison Parameters Normals controls vs. GS vs. GS GL GL Mean deviation (sup. + inf. Deviation)/ * 0.000* Symmetry Superior ratio 0.000* 0.000* Inferior ratio 0.000* 0.000* Superior-nasal ratio * Maximum modulation (ratio) 0.000* 0.000* Ellipse modulation (ratio) 0.000* 0.000* GDx number** 0.000* 0.000* Average thickness (µm) 0.000* 0.002* Ellipse average (µm) 0.000* 0.001* Superior average (µm) 0.000* 0.001* Inferior average (µm) 0.000* 0.000* Superior integral (micron squared) 0.001* 0.001* GS: glaucoma suspects, GL: glaucoma, *: Statistically significant with Mann-Whitney U-test after Bonferroni correction (P < ), **: Determined using GDx software (version ): Laser Diagnostic Technologies, San Diego, CA. that total FDT defect score increased with decreasing inferior average in the glaucomatous eyes (r = , P < 0.05). In addition, the total number of abnormal blocks in FDT increased with increasing GDx Number (r = 0.296, P < 0.05), and with decreasing inferior average, ellipse average, and

7 CORRELATION BETWEEN FDT AND SLP 95 Table 5. Pearson correlation coefficient (r) between 12 GDx parameters and FDT total defect score FDT total defect score vs. Variables Glaucoma suspects Glaucoma r P value r P value Symmetry Superior ratio Inferior ratio Superior-nasal ratio Maximum modulation (ratio) Ellipse modulation (ratio) GDx number Average thickness (µm) Ellipse average (µm) Superior average (µm) Inferior average (µm) * Superior integral (micron squared) *: statistically significant (P < 0.05) Table 6. Pearson correlation coefficient (r) between 12 GDx parameters and FDT total number of abnormal blocks FDT total number of abnormal blocks vs. Variables Glaucoma suspects Glaucoma r P value r P value Symmetry Symmetry Superior ratio Inferior ratio Superior-nasal ratio Maximum modulation (ratio) Ellipse modulation(ratio) GDx number * Average thickness ((m) Ellipse average ((m) * Superior average ((m) * Inferior average ((m) * Superior integral (micron squared) *: statistically significant (P < 0.05) superior average (P < 0.05 for each), in the glaucomatous eyes. In the eyes of glaucoma suspects, however, we observed no significant correlation between FDT and SLP data. Since there was some correlation between number of abnormal blocks in FDT and various GDx-NFA parameters for the glaucomatous eyes, we used a multiple linear regression model to determine the independent contributions of these parameters. Of the GDx parameters, we found that the most important predictor of FDT total number of abnormal blocks was inferior average in the glaucomatous eyes (b = , SE = 0.104, t = ; P < 0.01).

8 96 SH Kim, et al DISCUSSION In glaucoma, it is generally accepted that a significant loss of retinal ganglion cells occurs before functional visual loss appears on conventional achromatic visual field testing. This has led to a search for more sensitive ways of detecting retinal ganglion cell damage at a relatively early stage. Psychophysical examination, such as FDT testing, has been reported to be more sensitive to early visual field loss in eyes with glaucomatous optic neuropathy than standard automated perimetry (SAP). This is accomplished by evaluating the functional properties of a subset of larger retinal ganglion cells (My ganglion cells) which may be more susceptible to glaucomatous optic neuropathy than other ganglion cells. 13,22-28 When tested on a group of patients with glaucoma and OHTN, FDT was shown to be superior to SAP for identifying eyes with glaucomatous optic neuropathy. In addition, FDT identified a larger percentage of the abnormal results in the eyes of patients with OHTN than SAP or other psychophysical tests including short-wavelength automated perimetry (SWAP) and motion automated perimetry. 22 In patients with open angle glaucoma and visual field defects in only one hemifield, FDT with threshold N-30 program detected visual field changes in approximately 40% of the hemifields that were intact on HFA testing. The Heidelberg Retinal Tomograph (HRT) was shown to detect a significantly smaller rim volume in the corresponding region of the optic disc in eyes with hemifield defects than in control eyes with normal FDT results indicating that FDT may be able to predict abnormal structures in the optic disc or RNFL at an early stage of glaucoma, before a specific conventional visual field defect occurs. 13 A suprathreshold screening protocol using FDT has shown reasonably good sensitivity and specificity for the detection of glaucomatous visual field loss. 14,23,29,30 In one study, 91% of eyes with abnormal Humphrey perimetric results had two or more abnormal locations in 17 sectors in the FDT C-20-1 screening program, 14 while a second study showed that one or more missed locations on the FDT C-20-1 screening program had a sensitivity of 83.5% and specificity of 100% in glaucomatous eyes. 23 The program s testing is relatively quick, and test-retest reliability has been shown to be good. To our knowledge, however, the usefulness of the FDT screening protocol had not been evaluated in the early detection of glaucoma, i.e., before visual field defects are present in conventional achromatic perimetry. We therefore tested the ability of this program to predict early glaucomatous structural damage in glaucoma suspects, and the possible functional correlates of structural change in RNFL by using GDx-NFA with FDT screening algorithms (defect score and number of abnormal blocks). Our glaucoma suspects were enrolled on the basis of an abnormality of reliable FDT screening result and normal achromatic visual field test, while the examiner (MSK) was masked with respect to the status of the optic nerve or RNFL. After enrollment, those eyes with any suspicious nonglaucomatous optic neuropathy were excluded from our study. We hypothesized that these eyes may have sustained prefield glaucomatous damage. When these eyes were evaluated by another diagnostic modality that delineated structural changes in the RNFL, we found that the mean values of all polarimetric parameters, except symmetry, were intermediate between those of normal controls and glaucomatous eyes. While the means of measured GDx parameters differed significantly between the glaucoma suspects and control group, there was a marked overlap between glaucoma suspects and glaucomatous eyes. These findings are consistent with an earlier result showing statistical differences in all GDx-NFA values, except for the symmetry variable, between a preperimetric glaucoma group and a control group. 10 Our findings also suggest that the use of the FDT screening algorithm may enhance the detection of prefield glaucomatous damage in glaucoma suspects by unmasking morphological changes of the optic disc or RNFL damage before abnormalities appear on achromatic visual field examinations. Despite its high specificity, the FDT screening protocol using the abnormal criteria of two or more abnormal blocks might be insensitive to early preperimetric glaucomatous damage in suspects. 14 One study found that only 9% of glaucoma suspects

9 CORRELATION BETWEEN FDT AND SLP 97 had abnormal FDT with one or more defective blocks, suggesting that the sensitivity and specificity of the FDT screening protocol in detecting early structural damage in preperimetric glaucoma suspects require further confirmation. 23 On the other hand, detection of an abnormality using the FDT screening protocol, together with suspicious signs of glaucoma in the optic disc despite normal conventional perimetry, suggests that the high specificity associated with the FDT C-20-1 program may strengthen its ability to unmask patients at early stages of glaucoma. We also investigated whether the FDT screening algorithm could be correlated with RNFL retardation parameters in glaucoma suspects and glaucomatous eyes. Comparisons of achromatic conventional perimetry with SLP have demonstrated correlations between quantitative measurements of the RNFL and visual field loss, as measured both globally and regionally in glaucoma patients. 17,31 In our series of glaucoma suspects, however, neither the defect score nor the total number of abnormal blocks could be strongly correlated with the average of inferior and superior GDx deviations from normal values ( mean deviation ). We also found poor correlations between the FDT abnormality score in each hemifield and GDx-NFA measures in the corresponding quadrant. These findings may result from each diagnostic mode targeting on different visual properties of glaucoma. Alternatively, damage to the magnocellular optic nerve fibers, as measured by FDT testing, may not be reflected in the early structural loss as detected by polarimetry, although some histologic evidence suggests that such damage to the magnocellular optic nerve fibers occurs first in the course of glaucoma. 32 As observed in glaucomatous eyes, more advanced disease leads to more profound structural loss of RNFL, such that the visual field defect, as detected by FDT, may be better correlated with the structural parameters. Lack of correlation at an early stage of glaucoma may also reflect high inter-individual variability of RNFL thickness. The wide distribution of normative data as employed by GDx-NFA results in part from the presence of corneal birefringence artifacts which are poorly neutralized by the fixed compensator. This can limit the correlating power between the visual function obtained using FDT and RNFL measurements by GDx-NFA among our glaucoma suspects. This device requires further refinements regarding the influence of anterior segment polarization, individual disc size, split nerve fiber bundle effect, and myopia. Due to the lack of an accepted standard, the diagnostic value of the FDT screening algorithms we utilized may also require further confirmation, as well as a novel classification method based on their efficacy. While previous studies showed that the highest correlation with Humphrey mean deviation among GDx parameters in perimetric glaucomatous eyes was The Number, 10,31,33 our findings showed that inferior average had the highest correlation with the number of abnormal FDT screening blocks, followed by The Number, ellipse average, and superior average. Since The Number devised by the neural network is derived from computations based on all available RNFL data, both in normal subjects and glaucomatous eyes with abnormal conventional visual field defects, it may not correlate as well as inferior average in the eyes of subjects with early glaucomatous damage. Since our group of glaucomatous eyes included a significant number of eyes with NTG (13/53, or 25%), our finding that The Number did not have the highest correlation with the number of abnormal FDT screening blocks may be due to variations in glaucoma severity in our subjects. Other possible explanations include the pattern of glaucoma damage, age, race, study design, and use of different functional tests (i.e. FDT vs. HVF). Despite the statistical significance of our findings, several factors limited our results. A number of patients in this study were referred for possible glaucoma by outside ophthalmologists as we are engaged in a university-based practice. Although the appearance of the optic nerve head or RNFL was not part of the inclusion criteria, there might have been artificially increased sensitivity and specificity of glaucomatous RNFL change among our patients with defective FDT outcome. In addition, in this study we used FDT as our standard to quantitatively evaluate the functional glaucomatous damage in the group of glaucoma suspect and glaucomatous eyes. This predefined criterion using FDT results may have artificially increased

10 98 SH Kim, et al the classification and inclusion selection bias among study groups. It must also be considered that our results are derived from cross-sectional, as opposed to longitudinal, data, suggesting the need for a longitudinal study to confirm whether glaucoma suspects are more prone to develop conventional glaucomatous visual field defects. Another limitation of this study is the technique we used for measuring the polarimetry of RNFL with respect to the variable axis and magnitude of corneal polarization. The distribution of the variable corneal polarization axis and magnitude among individuals has been reported to lead to suboptimal RNFL images with elevated retardation or rotated pattern, primarily because, to neutralize corneal birefringence, the GDx-NFA uses a fixed compensator with a magnitude of 60nm and a fast axis oriented at 15 degrees nasally downward. 34 Therefore, measurement errors inherent to incomplete compensation of corneal birefringence were not accounted for. However, our effort was to compare the retardation values of the studied eyes with those of normal controls and provide the relative degree of retardation measures among the three groups, as they were equally affected by this inherent error. At the time of this study, we only had access to GDx with fixed corneal compensation (GDx-NFA). The validity of the RNFL measurement would have been enhanced if the inter-individual variability of the corneal polarization could be taken into account by using a variable corneal compensator. In conclusion, our results suggest that a FDT screening program may aid in the detection of prefield glaucomatous damage, and that our schematic scoring system may predict the relative degree of structural damage in glaucomatous eyes. The lack of correlation between the results of the two diagnostic devices in glaucoma suspects necessitates determination of the role of magnocellular ganglion cells (My cells) in the early glaucomatous disease process. Use of a more refined polarimeter in glaucoma suspects may better elucidate the relationship between the structural alterations observed with SLP and the functional losses detected by the FDT screening protocol. Furthermore, longitudinal studies are needed to determine whether the FDT screening protocol would be useful for predicting future glaucomatous visual field defects. REFERENCES 1. Sommer A, Katz J, Quigley HA, Miller NR, Robin AL, Richter RC, Witt KA. Clinically detectable nerve fiber atrophy precedes the onset of glaucomatous field loss. Arch Ophthalmol. 1991;109: Quigley HA, Reacher M, Katz J, Strahlman E, Gilbert D, Scott R. Quantitative grading of nerve fiber layer photographs. Ophthalmology. 1993;100: Quigley HA. Examination of the retinal nerve fiber layer in the recognition of early glaucoma damage. Trans Am Ophthalmol Soc. 1986;84: Weinreb RN, Shakiba S, Zangwill L. Scanning laser polarimetry to measure the nerve fiber layer of normal and glaucomatous eyes. Am J Ophthalmol. 1995; 119: Zangwill L, Berry CA, Garden VS, Weinreb RN. Reproducibility of retardation measurements with the nerve fiber analyzer II. J Glaucoma. 1997;6: Anton A, Zangwill L, Emdadi A, Weinreb RN. Nerve fiber layer measurements with scanning laser polarimetry in ocular hypertension. Arch Ophthalmol. 1997;115: Tjon-Fo-Sang MJ, de Vries J, Lemij HG. Measurement by nerve fiber analyzer of retinal nerve fiber layer thickness in normal subjects and patients with ocular hypertension. Am J Ophthalmol. 1996;122: Weinreb RN, Zangwill L, Berry CC, Bathija R, Sample PA. Detection of glaucoma with scanning laser polarimetry. Arch Ophthalmol. 1998;116: Choplin NT, Landy DC, Dreher AW. Differentiating patients with glaucoma suspects and normal subjects by nerve fiber layer assessment with scanning laser polarimetry. Ophthalmology. 1998;105: Horn FK, Jonas JB, Martus P, Mardin CY, Budde WM. Polarimetric measurement of retinal nerve fiber layer thickness in glaucoma diagnosis. J Glaucoma. 1999;8: Kondo Y, Yamamoto T, Sato Y, Matsubara M, Kitazawa Y. A frequency-doubling perimetry study in normal-tension glaucoma with hemifield defect. J Glaucoma. 1998;7: Reyes RD, Tomita G, Kitazawa Y. Retinal nerve fiber layer thickness within the area of apparently normal visual field in normal tension glaucoma with hemifield defect. J Glaucoma. 1998;7: Wu LL, Suzuki Y, Kunimatsu S, Araie M, Iwase A, Tomita G. Frequency doubling technology and con-

11 CORRELATION BETWEEN FDT AND SLP 99 focal scanning ophthalmoscopic optic disc analysis in open-angle glaucoma with hemifield defects. J Glaucoma. 2001;10: Quigley HA. Identification of glaucoma-related visual field abnormality with the screening protocol of frequency doubling technology. Am J Ophthalmol. 1998;125: Lachenmayr B, Airaksinen PJ, Drance SM, Wijsman K. Correlation of retinal nerve-fiber-layer loss, changes at the optic nerve head and various psychophysical criteria in glaucoma. Graefe s Arch Clin Exp Ophthalmol. 1991;229: Horn FK, Jonas JB, Junemann A, Korth M, Junemann A, Grundler A. The full field flicker test in early diagnosis of chronic open-angle glaucoma. Am J Ophthalmol. 1997;123: Weinreb RN, Shakiba S, Sample PA, Shahrokni S, van Horn S, Garden VS, Asawaphureekorn S, Zangwill L. Association between quantitative nerve fiber layer measurement and visual field loss in glaucoma. Am J Ophthalmol. 1995;120: Eid TM, Spaeth GL, Katz LJ, Azura-Blanco A, Agusberger J, Nicholl J. Quantitative estimation of retinal nerve fiber layer height in glaucoma and the relationship with optic nerve head topography and visual field. J Glaucoma. 1997;6: Tjon-Fo-Sang MJ, Lemij HG. The sensitivity and specificity of nerve fiber layer measurements in glaucoma as determined with scanning laser polarimetry. Am J Ophthalmol. 1997;123: Teesalu P, Airaksinen PJ, Tuulonen A. Blue-on-yellow visual field and retinal nerve fiber layer in ocular hypertension and glaucoma. Ophthalmology. 1998; 105: Abecia E, Honrubia FM. Retinal nerve fiber defects and automated perimetry evaluation in ocular hypertensives. Int Ophthalmol. 1992;16: Sample PA, Bosworth CF, Blumenthal EZ, Girkin C, Weinreb RN. Visual function-specific perimetry for indirect comparison of different ganglion cell populations in glaucoma. Invest Ophthalmol Vis Sci. 2000; 41: Paczka JA, Friedman DS, Quigley HA, Barron Y, Vitale S. Diagnostic capabilities of frequency-doubling technology, scanning laser polarimetry, and nerve fiber layer photographs to distinguish glaucomatous damage. Am J Ophthalmol. 2001;131: Kelly DH. Frequency doubling in visual responses. J Opt Soc Am. 1996;110: Johnson CA, Samuels SJ. Screening for glaucomatous visual field loss with frequency-doubling perimetry. Invest Ophthalmol Vis Sci. 1997;38; Dandona L, Hendrickson A, Quigley HA. Selective effects of experimental glaucoma on axonal transport by retinal ganglion cells to the dorsal lateral geniculate nucleus. Invest Ophthalmol Vis Sci. 1991;32: Chaturvedi N, Hedley-Whyte ET, Dreyer EB. Lateral geniculate nucleus in glaucoma. Am J Ophthalmol. 1993;116: Glovinsky Y, Quigley HA, Pease ME. Foveal ganglion cell loss is size dependent in experimental glaucoma. Invest Ophthalmol Vis Sci. 1993;34: Burnstein Y, Ellish NJ, Magbalon M, Higginbotham EJ. Comparison of frequency doubling perimetry with humphrey visual field analysis in a glaucoma practice. Am J Ophthalmol. 2000;129: Sponsel WE, Arango S, Trigo Y, Mensah J. Clinical classification of glaucomatous visual field loss by frequency doubling perimetry. Am J Ophthalmol. 1998;125: Kwon YH, Hong SP, Honkanen RA, Alward WL. Correlation of automated visual field parameters and peripapillary nerve fiber layer thickness as measured by scanning laser polarimetry. J Glaucoma. 2000;9: Quigley HA, Sanchez RM, Dunkelberger GR, L Hernault NL, Baginski TA. Chronic glaucoma selectively damages large optic nerve fibers. Invest Ophthalmol Vis Sci. 1987;28: Lauande-Pimentel R, Carvalho RA, Oliveira HC, Goncalves DC, Silva LM, Costa VP. Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements. Br J Ophthalmol. 2001;85: Greenfield DS, Knighton RW, Huang SR. Effect of corneal polarization axis on assessment of retinal laser polarimetry. Am J Ophthalmol. 2000;129;

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population

Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Korean J Ophthalmol Vol. 18:1-8, 2004 Ability of Scanning Laser Polarimetry (GDx) to Discriminate among Early Glaucomatous, Ocular Hypertensive and Normal Eyes in the Korean Population Sun Young Lee, MD,

More information

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma

Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Study of Retinal Nerve Fiber Layer Thickness Within Normal Hemivisual Field in Primary Open-Angle Glaucoma and Normal-Tension Glaucoma Chiharu Matsumoto, Shiroaki Shirato, Mai Haneda, Hiroko Yamashiro

More information

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements

Discrimination between normal and glaucomatous eyes with visual field and scanning laser polarimetry measurements 586 Glaucoma Service, Department of Ophthalmology, University of Campinas, Campinas, Brazil R Lauande-Pimentel R A Carvalho H C Oliveira D C Gonçalves L M Silva V P Costa Glaucoma Service, Department of

More information

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7

Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 Sensitivity and specificity of new GDx parameters Colen TP, Tang NEML, Mulder PGH and Lemij HG Submitted for publication CHAPTER 7 61 Abstract Purpose The GDx is a scanning laser polarimeter that assesses

More information

Correlating Structure With Function in End-Stage Glaucoma

Correlating Structure With Function in End-Stage Glaucoma C L I N I C A L S C I E N C E Correlating Structure With Function in End-Stage Glaucoma Eytan Z. Blumenthal, MD; Amjad Horani, MD; Rajesh Sasikumar, MD; Chandrasekhar Garudadri, MD; Addepalli Udaykumar,

More information

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection

Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Yamanashi Med. J. 18(1), 15~ 20, 2003 Original Article Parafoveal Scanning Laser Polarimetry for Early Glaucoma Detection Satoshi KOGURE, Yoshiki TODA, Hiroyuki IIJIMA and Shigeo TSUKAHARA Department of

More information

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS

NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS Nerve fiber layer thickness in normals and glaucoma patients 403 NERVE FIBER LAYER THICKNESS IN NORMALS AND GLAUCOMA PATIENTS HIROTAKA SUZUMURA, KAYOKO HARASAWA, AKIKO KOBAYASHI and NARIYOSHI ENDO Department

More information

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation

The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation C L I N I C A L S C I E N C E The Effect of Pupil Dilation on Scanning Laser Polarimetry With Variable Corneal Compensation Amjad Horani, MD; Shahar Frenkel, MD, PhD; Eytan Z. Blumenthal, MD BACKGROUND

More information

Clinical decision making based on data from GDx: One year observations

Clinical decision making based on data from GDx: One year observations Washington University School of Medicine Digital Commons@Becker Open Access Publications 2002 Clinical decision making based on data from GDx: One year observations James C. Bobrow Washington University

More information

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx

STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma. Module 3a GDx STRUCTURE & FUNCTION An Integrated Approach for the Detection and Follow-up of Glaucoma Module 3a GDx Educational Slide Deck Carl Zeiss Meditec, Inc. November 2005 1 Structure & Function Modules Module

More information

T he retinal ganglion cells of different sizes have distinct

T he retinal ganglion cells of different sizes have distinct 604 CLINICAL SCIENCE Agreement between frequency doubling perimetry and static perimetry in eyes with high tension glaucoma and normal tension glaucoma S Kogure, Y Toda, D Crabb, K Kashiwagi, F W Fitzke,

More information

A vailability of the scanning laser polarimeter now permits

A vailability of the scanning laser polarimeter now permits 70 CLINICAL SCIENCE Specificity and sensitivity of glaucoma detection in the Japanese population using scanning laser polarimetry Shigeo Funaki, Motohiro Shirakashi, Kiyoshi Yaoeda, Haruki Abe, Shiho Kunimatsu,

More information

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting

A comparison of HRT II and GDx imaging for glaucoma detection in a primary care eye clinic setting (2007) 21, 1050 1055 & 2007 Nature Publishing Group All rights reserved 0950-222X/07 $30.00 www.nature.com/eye CLINICAL STUDY A comparison of HRT II and GDx imaging for glaucoma detection in a primary

More information

G laucoma is a progressive optic neuropathy in which

G laucoma is a progressive optic neuropathy in which 1135 EXTENDED REPORT The correlation between optic nerve head topographic measurements, peripapillary nerve fibre layer thickness, and visual field indices in glaucoma Y-W Lan, D B Henson, A J Kwartz...

More information

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma

Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Relationship between the GDx VCC and Stratus OCT in Primary Open Angle Glaucoma Reza Zarei, MD 1 Mohammad Soleimani, MD 2 Sasan Moghimi, MD 3 Mohammad Yaser Kiarudi, MD 2 Mahmoud Jabbarvand, MD 1 Yadollah

More information

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science

S Morishita, T Tanabe, S Yu, M Hangai, T Ojima, H Aikawa, N Yoshimura. Clinical science Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan Correspondence to: Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute,

More information

The Role of the RNFL in the Diagnosis of Glaucoma

The Role of the RNFL in the Diagnosis of Glaucoma Chapter 1. The Role of the RNFL in the Diagnosis of Glaucoma Introduction Glaucoma is an optic neuropathy characterized by a loss of of retinal ganglion cells and their axons, the Retinal Nerve Fiber Layer

More information

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography

Retinal nerve fiber layer thickness in Indian eyes with optical coherence tomography Original articles in Indian eyes with optical coherence tomography Malik A, Singh M, Arya SK, Sood S, Ichhpujani P Department of Ophthalmology Government Medical College and Hospital, Sector 32, Chandigarh,

More information

Relationship Between Structure

Relationship Between Structure Original Article Relationship Between Structure and Function of the Optic Nerve Head-Glaucoma versus Normal Dr Savita Bhat, Dr Anna Elias, Dr Siddharth Pawar, Dr S.J. Saikumar, Dr Alpesh Rajput, superior,

More information

Reliability analyses of the GDx nerve-fiber analyzer

Reliability analyses of the GDx nerve-fiber analyzer VOL. 29 NO. 2 PHILIPPINE JOURNAL OF Ophthalmology APRIL ORIGINAL ARTICLE - JUNE 2004 1, 2, 3 Patricia M. Khu, MD, MS Edgardo U. Dorotheo, MD 1 Lawrence Tinio, MD 2 Cynthia P. Cordero, MS 4 1, 2, 3 Manuel

More information

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma Med. J. Cairo Univ., Vol. 83, No. 2, September: 67-72, 2015 www.medicaljournalofcairouniversity.net Macular Ganglion Cell Complex Measurement Using Spectral Domain Optical Coherence Tomography in Glaucoma

More information

OtticaFisiopatologica

OtticaFisiopatologica Anno quindicesimo dicembre 2010 How to assess the retinal nerve fiber layer thickness Antonio Ferreras Miguel Servet University Hospital, Zaragoza. Aragón Health Sciences Institute University of Zaragoza

More information

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects

Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve Fiber Layer Defects 접수번호 : 2008-105 Korean Journal of Ophthalmology 2009;23:169-175 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.3.169 Scanning Laser Polarimetry and Optical Coherence Tomography for Detection of Retinal Nerve

More information

In some patients with glaucoma, standard (achromatic) automated

In some patients with glaucoma, standard (achromatic) automated Detecting Early Glaucoma by Assessment of Retinal Nerve Fiber Layer Thickness and Visual Function Christopher Bowd, 1 Linda M. Zangwill, 1 Charles C. Berry, 2 Eytan Z. Blumenthal, 1 Cristiana Vasile, 1

More information

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS

THE BASIC PATHOLOGIC CHANGE IN GLAUCOMA IS Quantitative Assessment of Atypical Birefringence Images Using Scanning Laser Polarimetry With Variable Corneal Compensation HARMOHINA BAGGA, MD, DAVID S. GREENFIELD, MD, AND WILLIAM J. FEUER, MS PURPOSE:

More information

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Original Article Philippine Journal of OPHTHALMOLOGY Retinal Nerve Fiber Layer Measurements in Myopia Using Optical Coherence Tomography Dennis L. del Rosario, MD and Mario M. Yatco, MD University of Santo

More information

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods

Seiji T. Takagi, Yoshiyuki Kita, Asuka Takeyama, and Goji Tomita. 1. Introduction. 2. Subjects and Methods Ophthalmology Volume 2011, Article ID 914250, 5 pages doi:10.1155/2011/914250 Clinical Study Macular Retinal Ganglion Cell Complex Thickness and Its Relationship to the Optic Nerve Head Topography in Glaucomatous

More information

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects Romanian Journal of Ophthalmology, Volume 60, Issue 3, July-September 2016. pp:158-164 GENERAL ARTICLE Translating data and measurements from stratus to cirrus OCT in glaucoma patients and healthy subjects

More information

Ganglion cell complex scan in the early prediction of glaucoma

Ganglion cell complex scan in the early prediction of glaucoma Original article in the early prediction of glaucoma Ganekal S Nayana Super Specialty Eye Hospital and Research Center, Davangere, Karnataka, India Abstract Objective: To compare the macular ganglion cell

More information

Test-retest variability in visual field testing using frequency doubling technology

Test-retest variability in visual field testing using frequency doubling technology European Journal of Ophthalmology / Vol. 17 no. 2, 2007 / pp. 203-207 Test-retest variability in visual field testing using frequency doubling technology A. HORANI, S. FRENKEL, E.Z. BLUMENTHAL Department

More information

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS

The Relationship between Standard Automated Perimetry and GDx VCC Measurements METHODS The Relationship between Standard Automated Perimetry and GDx VCC Measurements Nicolaas J. Reus and Hans G. Lemij PURPOSE. To investigate the relationship between retinal light sensitivity measured with

More information

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry European Journal of Ophthalmology / Vol. 13 no. 4, 2003 / pp. 377-382 The evaluation of retinal nerve fiber layer in pigment dispersion syndrome and pigmentary glaucoma using scanning laser polarimetry

More information

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD

CLINICAL SCIENCES. Felipe A. Medeiros, MD; Linda M. Zangwill, PhD; Christopher Bowd, PhD; Robert N. Weinreb, MD CLINICAL SCIENCES Comparison of the GDx VCC Scanning Laser Polarimeter, HRT II Confocal Scanning Laser Ophthalmoscope, and Stratus OCT Optical Coherence Tomograph for the Detection of Glaucoma Felipe A.

More information

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation

Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Detection of Glaucoma Using Scanning Laser Polarimetry with Enhanced Corneal Compensation Felipe A. Medeiros, Christopher Bowd, Linda M. Zangwill, Chirag Patel, and Robert N. Weinreb From the Hamilton

More information

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma

Relationship between GDx VCC and Stratus OCT in juvenile glaucoma (2009) 23, 2182 2186 & 2009 Macmillan Publishers Limited All rights reserved 09-222X/09 $32.00 www.nature.com/eye CLINICAL STUDY Relationship between GDx VCC and Stratus OCT in juvenile glaucoma R Zareii,

More information

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients

Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients Correlation of Blue Chromatic Macular Sensitivity with Optic Disc Change in Early Glaucoma Patients Yoshio Yamazaki, Kenji Mizuki, Fukuko Hayamizu and Chizuru Tanaka Department of Ophthalmology, Nihon

More information

Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer

Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer European Journal of Ophthalmology / Vol. 15 no. 2, 2005 / pp. 246-254 Retinal nerve fiber layer measured by Heidelberg retina tomograph and nerve fiber analyzer M. IESTER 1,2, A. MERMOUD 1 1 Hopital Ophtalmique

More information

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma

Patterns of Subsequent Progression of Localized Retinal Nerve Fiber Layer Defects on Red-free Fundus Photographs in Normal-tension Glaucoma pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(4):330-336 http://dx.doi.org/10.3341/kjo.2014.28.4.330 Original Article Patterns of Subsequent Progression of Localized Retinal Nerve Fiber

More information

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation

Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Detection of Progressive Retinal Nerve Fiber Layer Loss in Glaucoma Using Scanning Laser Polarimetry with Variable Corneal Compensation Felipe A. Medeiros, Luciana M. Alencar, Linda M. Zangwill, Christopher

More information

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS

Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study METHODS Glaucoma Diagnostic Accuracy of OCT with a Normative Database to Detect Diffuse Retinal Nerve Fiber Layer Atrophy: Diffuse Atrophy Imaging Study Jin Wook Jeoung, 1,2 Seok Hwan Kim, 1,3 Ki Ho Park, 1,2

More information

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect

Differences between Non-arteritic Anterior Ischemic Optic Neuropathy and Open Angle Glaucoma with Altitudinal Visual Field Defect pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2015;29(6):418-423 http://dx.doi.org/10.3341/kjo.2015.29.6.418 Original Article Differences between Non-arteritic Anterior Ischemic Optic Neuropathy

More information

Intro to Glaucoma/2006

Intro to Glaucoma/2006 Intro to Glaucoma/2006 Managing Patients with Glaucoma is Exciting Interesting Challenging But can often be frustrating! Clinical Challenges To identify patients with risk factors for possible glaucoma.

More information

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure

Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Scanning Laser Polarimetry in Patients with Acute Attack of Primary Angle Closure Jimmy S. M. Lai*, Clement C. Y. Tham, Jonathan C. H. Chan*, Nelson K. F. Yip, Wilson W. T. Tang, Patrick S. H. Li*, Jane

More information

Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender

Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Bowd et al. Vol. 19, No. 1/January 2002/J. Opt. Soc. Am. A 197 Imaging of the optic disc and retinal nerve fiber layer: the effects of age, optic disc area, refractive error, and gender Christopher Bowd,

More information

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation

Diagnostic Accuracy of Scanning Laser Polarimetry with Enhanced versus Variable Corneal Compensation Diagnostic Accuracy of Scanning Laser olarimetry with Enhanced versus Variable Corneal T. A. Mai, MD, N. J. Reus, MD, hd, H. G. Lemij, MD, hd urpose: To compare the diagnostic accuracy of scanning laser

More information

CLINICAL SCIENCES. Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD

CLINICAL SCIENCES. Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD CLINICAL SCIENCES Achromatic and Short-Wavelength Automated Perimetry in Patients With Glaucomatous Large Cups Steven L. Mansberger, MD; Pamela A. Sample, PhD; Linda Zangwill, PhD; Robert N. Weinreb, MD

More information

OCT in the Diagnosis and Follow-up of Glaucoma

OCT in the Diagnosis and Follow-up of Glaucoma OCT in the Diagnosis and Follow-up of Glaucoma Karim A Raafat MD. Professor Of Ophthalmology Cairo University Hmmmm! Do I have Glaucoma or not?! 1 Visual Function 100% - N Gl Structure : - 5000 axon /

More information

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC)

Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T et al Original article Evaluation of retinal nerve fiber layer thickness parameters in myopic population using scanning laser polarimetry (GDxVCC) Dada T1, Aggarwal A1, Bali SJ2, Sharma A1, Shah

More information

Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma

Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma Fourier Analysis of Scanning Laser Polarimetry Measurements with Variable Corneal Compensation in Glaucoma Felipe A. Medeiros, Linda M. Zangwill, Christopher Bowd, Antje S. Bernd, and Robert N. Weinreb

More information

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC

Position of retinal blood vessels correlates with retinal nerve fibre layer thickness profiles as measured with GDx VCC and ECC Department of Ophthalmology, Medical University of Vienna, Austria Correspondence to Clemens Vass, Department of Ophthalmology and Optometry, Medical University of Vienna, General Hospital, Währinger Gürtel

More information

STANDARD AUTOMATED PERIMETRY IS A GENERALLY

STANDARD AUTOMATED PERIMETRY IS A GENERALLY Comparison of Long-term Variability for Standard and Short-wavelength Automated Perimetry in Stable Glaucoma Patients EYTAN Z. BLUMENTHAL, MD, PAMELA A. SAMPLE, PHD, LINDA ZANGWILL, PHD, ALEXANDER C. LEE,

More information

CLINICAL SCIENCES. optic neuropathy characterized

CLINICAL SCIENCES. optic neuropathy characterized CLINICAL SCIENCES Spectral-Domain Optical Coherence Tomography for Detection of Localized Retinal Nerve Fiber Layer Defects in Patients With Open-Angle Glaucoma Na Rae Kim, MD; Eun Suk Lee, MD, PhD; Gong

More information

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note.

53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Case 1 Normal Tension Glaucoma 53 year old woman attends your practice for routine exam. She has no past medical history or family history of note. Table 1. Right Eye Left Eye Visual acuity 6/6 6/6 Ishihara

More information

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography

CLINICAL SCIENCES. Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography CLINICAL SCIENCES Comparison of Glaucoma Diagnostic Capabilities of Cirrus HD and Stratus Optical Coherence Tomography Seong Bae Park, MD; Kyung Rim Sung, MD, PhD; Sung Yong Kang, MD; Kyung Ri Kim, BS;

More information

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD

Noel de Jesus Atienza, MD, MSc and Joseph Anthony Tumbocon, MD Original Article Philippine Journal of OPHTHALMOLOGY Diagnostic Accuracy of the Optical Coherence Tomography in Assessing Glaucoma Among Filipinos. Part 1: Categorical Outcomes Based on a Normative Database

More information

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography

To assess the glaucoma diagnostic ability of Fourier Domain Optical Coherence Tomography American Journal of Engineering Research (AJER) e-issn : 2320-0847 p-issn : 2320-0936 Volume-02, Issue-11, pp-104-110 www.ajer.org Research Paper Open Access To assess the glaucoma diagnostic ability of

More information

Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation

Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation Korean Journal of Ophthalmology 22(1):18-25, 2008 DOI : 10.3341/kjo.2008.22.1.18 Relationship between Scanning Laser Polarimetry with Enhanced Corneal Compensation and with Variable Corneal Compensation

More information

Linking structure and function in glaucoma

Linking structure and function in glaucoma CET CONTINUING Sponsored by 1 CET POINT Linking structure and function in glaucoma 50 Dr Samantha McGinnigle PhD, BSc (Hons), MCOptom, AHEA This article will give an overview of the latest imaging technology

More information

Early Detection Of Glaucoma Clinical Clues. Points To Live By. Glaucoma Risk Factors 10/3/2014

Early Detection Of Glaucoma Clinical Clues. Points To Live By. Glaucoma Risk Factors 10/3/2014 Early Detection Of Glaucoma Clinical Clues Eric E. Schmidt, O.D. Omni Eye Specialists Wilmington, NC schmidtyvision@msn.com Points To Live By 25% of G pxs NEVER have IOP >21mm 50% of G pxs have trough

More information

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT

Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis OCT pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(3):166-173 DOI: 10.3341/kjo.2011.25.3.166 Original Article Comparison of Retinal Nerve Fiber Layer Thickness between Stratus and Spectralis

More information

Preperimetric glaucoma diagnosis by confocal scanning laser tomography of the optic disc

Preperimetric glaucoma diagnosis by confocal scanning laser tomography of the optic disc Br J Ophthalmol 1999;83:299 304 299 Department of Ophthalmology and Eye Hospital, Friedrich-Alexander- University Erlangen-Nürnberg, Erlangen, Germany C Y Mardin F K Horn J B Jonas W M Budde Correspondence

More information

Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph

Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph Influence of Myopic Disc Shape on the Diagnostic Precision of the Heidelberg Retina Tomograph Yoshio Yamazaki,* Keiji Yoshikawa, Shiho Kunimatsu, Nobuyuki Koseki, Yasuyuki Suzuki, Shun Matsumoto and Makoto

More information

Characterization of Peripapillary Atrophy Using Spectral Domain Optical Coherence Tomography

Characterization of Peripapillary Atrophy Using Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2010;24(6):353-359 DOI: 10.3341/kjo.2010.24.6.353 Characterization of Peripapillary trophy Using Spectral Domain Optical Coherence Tomography Original

More information

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography

Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence Tomography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(1):32-38 http://dx.doi.org/10.3341/kjo.2012.26.1.32 Retinal Nerve Fiber Layer Measurement Variability with Spectral Domain Optical Coherence

More information

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association

MEDICAL POLICY. Proprietary Information of Excellus Health Plan, Inc. A nonprofit independent licensee of the BlueCross BlueShield Association MEDICAL POLICY SUBJECT: OPHTHALMOLOGIC TECHNIQUES PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical

More information

Assessment of retinal ganglion cell loss is critical for the

Assessment of retinal ganglion cell loss is critical for the Scanning Laser Polarimetry with Variable Corneal Compensation: Identification and Correction for Corneal Birefringence in Eyes with Macular Disease Harmohina Bagga, David S. Greenfield, and Robert W. Knighton

More information

Available online at Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6):

Available online at   Pelagia Research Library. Advances in Applied Science Research, 2013, 4(6): Available online at www.pelagiaresearchlibrary.com Advances in Applied Science Research, 2013, 4(6):201-206 ISSN: 0976-8610 CODEN (USA): AASRFC Comparison of glaucoma diagnostic ability of retinal nerve

More information

Eye (2005), 1 9 & 2005 Nature Publishing Group All rights reserved X/05 $

Eye (2005), 1 9 & 2005 Nature Publishing Group All rights reserved X/05 $ (2005), 1 9 & 2005 Nature Publishing Group All rights reserved 0950-222X/05 $30.00 www.nature.com/eye Automated detection of wedgeshaped defects in polarimetric images of the retinal nerve fibre layer

More information

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma

C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e. The GDxVCC Early answers and ongoing assessment for glaucoma C a t a r a c t G l a u c o m a R e t i n a R e f r a c t i v e The GDxVCC Early answers and ongoing assessment for glaucoma The quantifiable approach to quality care Only Humphrey GPA software Early insight

More information

A Review Of Risk Factors. Early Detection Of Glaucoma Clinical Clues. A risk factor analysis is critical. Points To Live By

A Review Of Risk Factors. Early Detection Of Glaucoma Clinical Clues. A risk factor analysis is critical. Points To Live By A Review Of Risk Factors Early Detection Of Glaucoma Clinical Clues Eric E. Schmidt, O.D. Omni Eye Specialists Wilmington, NC schmidtyvision@msn.com FINDACAR Family history IOP Nearsightedness Diabetes/Vascular

More information

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes

Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Scanning Laser Tomography to Evaluate Optic Discs of Normal Eyes Hiroshi Nakamura,* Toshine Maeda,* Yasuyuki Suzuki and Yoichi Inoue* *Eye Division of Olympia Medical Clinic, Tokyo, Japan; Department of

More information

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN

International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December ISSN International Journal of Scientific & Engineering Research, Volume 4, Issue 12, December-2013 108 Name of Chief and Corresponding Author : Dr Chandrima Paul TITLE : Comparison of glaucoma diagnostic ability

More information

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography

Key words: Glaucoma, Imaging, Ophthalmoscopy, Optic neuropathy, Topography 194 Review Article Evaluating the Optic Nerve and Retinal Nerve Fibre Layer: The Roles of Heidelberg Retina Tomography, Scanning Laser Polarimetry and Optical Coherence Tomography Sek-Tien Hoh, 1 MBBS,

More information

Scanning laser polarimetry (SLP) provides real-time, objective

Scanning laser polarimetry (SLP) provides real-time, objective The Effect of Atypical Birefringence Patterns on Glaucoma Detection Using Scanning Laser Polarimetry with Variable Corneal Compensation Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda

More information

Clinical Study Visual Field Loss Morphology in High- and Normal-Tension Glaucoma

Clinical Study Visual Field Loss Morphology in High- and Normal-Tension Glaucoma Journal of Ophthalmology Volume 2012, Article ID 327326, 8 pages doi:10.1155/2012/327326 Clinical Study Visual Field Loss Morphology in High- and Normal-Tension Glaucoma Michele Iester, 1, 2 Fabio De Feo,

More information

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013

EXPERIMENTAL AND THERAPEUTIC MEDICINE 6: , 2013 268 Comparison of optic nerve morphology in eyes with glaucoma and eyes with non-arteritic anterior ischemic optic neuropathy by Fourier domain optical coherence tomography YUXIN YANG 1, HAITAO ZHANG 1,

More information

MEDICAL POLICY. Proprietary Information of YourCare Healthcare

MEDICAL POLICY. Proprietary Information of YourCare Healthcare MEDICAL POLICY PAGE: 1 OF: 7 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy criteria are not applied.

More information

The frequency-doubling illusion occurs when a low-spatialfrequency

The frequency-doubling illusion occurs when a low-spatialfrequency Sensitivity and Specificity of Frequency Doubling Perimetry in Neuro-ophthalmic Disorders: A Comparison with Conventional Automated Perimetry Michael Wall, 1,2 Richard K. Neahring, 1 and Kimberly R. Woodward

More information

T he extent of damage to the optic nerve caused by nonarteritic

T he extent of damage to the optic nerve caused by nonarteritic 1274 EXTENDED REPORT Frequency doubling technology perimetry in non-arteritic ischaemic optic neuropathy with altitudinal defects C A Girkin, G McGwin Jr, J DeLeon-Ortega... See end of article for authors

More information

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements

Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Original article Comparative evaluation of time domain and spectral domain optical coherence tomography in retinal nerve fiber layer thickness measurements Dewang Angmo, 1 Shibal Bhartiya, 1 Sanjay K Mishra,

More information

RETINAL NERVE FIBER LAYER

RETINAL NERVE FIBER LAYER CLINICAL SCIENCES The Effect of Scan Diameter on Retinal Nerve Fiber Layer Thickness Measurement Using Stratus Optic Coherence Tomography Giacomo Savini, MD; Piero Barboni, MD; Michele Carbonelli, MD;

More information

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography

Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Reproducibility of Nerve Fiber Layer Thickness Measurements by Use of Optical Coherence Tomography Eytan Z. Blumenthal, MD, 1 Julia M. Williams, BS, 1 Robert N. Weinreb, MD, 1 Christopher A. Girkin, MD,

More information

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach

Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach ORIGINAL STUDY Factors Associated With Visual Field Progression in Cirrus Optical Coherence Tomography-guided Progression Analysis: A Topographic Approach Joong Won Shin, MD, Kyung Rim Sung, MD, PhD, Jiyun

More information

The only measure of RNFL Integrity : GDxPRO

The only measure of RNFL Integrity : GDxPRO The ideal glaucoma assessment tool for your practice The only measure of RNFL Integrity : GDxPRO Proven Clinical Performance Numerous studies have validated the diagnostic accuracy and performance of the

More information

Diagnosing open-angle glaucoma may be particularly

Diagnosing open-angle glaucoma may be particularly A New Look at Selective Perimetry What is its role in clinical practice? BY MURRAY FINGERET, OD Diagnosing open-angle glaucoma may be particularly challenging when damage is mild or early. Although most

More information

Glaucoma: Diagnostic Modalities

Glaucoma: Diagnostic Modalities Glaucoma: Diagnostic Modalities - Dr. Barun Kumar Nayak, Dr. Sarika Ramugade Glaucoma is a leading cause of blindness in the world, especially in older people. Early detection and treatment by ophthalmologist

More information

3/16/2018. Perimetry

3/16/2018. Perimetry Perimetry The normal visual field extends further away from fixation temporally and inferiorly than superiorly and nasally. From the center of the retina this sensitivity decreases towards the periphery,

More information

Structural examina.on: Imaging

Structural examina.on: Imaging ManaMa: Glaucoma Structural examina.on: Imaging Luís Abegão Pinto, MD, PhD Department of Ophthalmology CHLC Lisbon Faculty of Medicine, Lisbon University 1 11-10- 2013 Structural changes Qualitative changes

More information

Summary HTA HTA-Report Summary Validity and cost-effectiveness of methods for screening of primary open angle glau- coma

Summary HTA HTA-Report Summary Validity and cost-effectiveness of methods for screening of primary open angle glau- coma Summary HTA HTA-Report Summary Validity and cost-effectiveness of methods for screening of primary open angle glaucoma Antony K, Genser D, Fröschl B DAHTA@DIMDI Waisenhausgasse 36-38a D-50676 Köln Tel.:

More information

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements

A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT) Retinal Nerve Fiber Layer Measurements Based on Time Domain OCT Measurements pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2012;26(5):369-377 http://dx.doi.org/10.3341/kjo.2012.26.5.369 Original Article A Formula to Predict Spectral Domain Optical Coherence Tomography (OCT)

More information

Optic Nerve: Clinical Examination

Optic Nerve: Clinical Examination Optic Nerve: Clinical Examination Marcelo T. Nicolela 2 Core Messages Optic disc evaluation is of fundamental importance in the management of glaucoma. Clinical examination of the optic disc is best performed

More information

The only measure of RNFL Integrity: GDxPRO

The only measure of RNFL Integrity: GDxPRO The only measure of RNFL Integrity: GDxPRO Look beyond thickness. Underlying structural organization is the key to RNFL At times, stepping onto ice is a risky proposition. Just knowing its thickness might

More information

Optical coherence tomography (OCT) is a noninvasive,

Optical coherence tomography (OCT) is a noninvasive, Ability of Stratus OCT to Detect Progressive Retinal Nerve Fiber Layer Atrophy in Glaucoma Eun Ji Lee, 1,2 Tae-Woo Kim, 1,2 Ki Ho Park, 2 Mincheol Seong, 3 Hyunjoong Kim, 4 and Dong Myung Kim 2 PURPOSE.

More information

The evaluation of the optic nerve and retinal nerve

The evaluation of the optic nerve and retinal nerve Five rules to evaluate the optic disc and retinal nerve fiber layer for glaucoma Murray Fingeret, O.D., a,b Felipe A. Medeiros, M.D., c Remo Susanna, Jr, M.D., d and Robert N. Weinreb, M.D. c a Department

More information

CLINICAL SCIENCES. Differences in Visual Function and Optic Nerve Structure Between Healthy Eyes of Blacks and Whites

CLINICAL SCIENCES. Differences in Visual Function and Optic Nerve Structure Between Healthy Eyes of Blacks and Whites CLINICAL SCIENCES Differences in Visual Function and Optic Nerve Structure Between Healthy Eyes of and Lyne Racette, PhD; Catherine Boden, PhD; Shannon L. Kleinhandler, BSc; Christopher A. Girkin, MD;

More information

Access to the published version may require journal subscription. Published with permission from: Elsevier

Access to the published version may require journal subscription. Published with permission from: Elsevier This is an author produced version of a paper published in Ophthalmology. This paper has been peer-reviewed but does not include the final publisher proof-corrections or journal pagination. Citation for

More information

21st Century Visual Field Testing

21st Century Visual Field Testing Supplement to Supported by an educational grant from Carl Zeiss Meditec, Inc. Winter 2011 21st Century Visual Field Testing the Evolution Continues 21st Century Visual Field Testing 21st Century Visual

More information

Relation of optic disc topography and age to thickness of retinal nerve fibre layer as measured using scanning laser polarimetry, in normal subjects

Relation of optic disc topography and age to thickness of retinal nerve fibre layer as measured using scanning laser polarimetry, in normal subjects Br J Ophthalmol 2000;84:473 478 473 Department of Ophthalmology, Celal Bayar University School of Medicine, Manisa, Turkey A B Toprak Department of Ophthalmology, Beyoglu Hospital, Taksim, Istanbul, Turkey

More information

Jong Chul Han, Da Ye Choi, and Changwon Kee. 1. Introduction

Jong Chul Han, Da Ye Choi, and Changwon Kee. 1. Introduction Ophthalmology Volume 2015, Article ID 641204, 7 pages http://dx.doi.org/10.1155/2015/641204 Clinical Study The Different Characteristics of Cirrus Optical Coherence Tomography between Superior Segmental

More information

F requency doubling technology (FDT) perimetry was

F requency doubling technology (FDT) perimetry was 131 EXTENDED REPORT Clinical evaluation of frequency doubling technology perimetry using the Humphrey Matrix 24-2 threshold strategy P G D Spry, H M Hussin, J M Sparrow... See end of article for authors

More information