Blood flow and structure interactions in a stented abdominal aortic aneurysm model

Size: px
Start display at page:

Download "Blood flow and structure interactions in a stented abdominal aortic aneurysm model"

Transcription

1 Medical Engineering & Physics 27 (2005) Blood flow and structure interactions in a stented abdominal aortic aneurysm model Zhonghua Li a, Clement Kleinstreuer b, a Department of Mechanical and Aerospace Engineering, North Carolina State University, Campus Box 7910, 3198 Broughton Hall, Raleigh, NC , USA b Department of Mechanical and Aerospace Engineering and Department of Biomedical Engineering, North Carolina State University, Campus Box 7910, 3198 Broughton Hall, Raleigh, NC , USA Received 10 September 2004; received in revised form 7 December 2004; accepted 12 December 2004 Abstract Since the introduction of endovascular techniques in the early 1990s for the treatment of abdominal aortic aneurysms (AAAs), the insertion of an endovascular graft (EVG) into the affected artery segment has been greatly successful for a certain group of AAA patients and is continuously evolving. However, although minimally invasive endovascular aneurysm repair (EVAR) is very attractive, post-operative complications may occur. Typically, they are the result of excessive fluid structure interaction dynamics, possibly leading to EVG migration. Considering a 3D stented AAA, a coupled fluid flow and solid mechanics solver was employed to simulate and analyze the interactive dynamics, i.e., pulsatile blood flow in the EVG lumen, pressure levels in the stagnant blood filling the AAA cavity, as well as stresses and displacements in the EVG and AAA walls. The validated numerical results show that a securely placed EVG shields the diseased AAA wall from the pulsatile blood pressure and hence keeps the maximum wall stress 20 times below the wall stress value in the non-stented AAA. The sac pressure is reduced significantly but remains non-zero and transient, caused by the complex fluid structure interactions between luminal blood flow, EVG wall, stagnant sac blood, and aneurysm wall. The time-varying drag force on the EVG exerted by physiological blood flow is unavoidable, where for patients with severe hypertension the risk of EVG migration is very high IPEM. Published by Elsevier Ltd. All rights reserved. Keywords: Fluid structure interaction; Stented abdominal aortic aneurysm; Endovascular graft; Sac pressure; Wall stress; Drag force; EVG migration 1. Introduction Aneurysms, an irreversible ballooning of weakened artery segments, occur most frequently in the abdominal aorta. As the aneurysm expands, it may eventually rupture, making it the 13th leading cause of mortality in the US. Alternative to open surgery, where the diseased aorta segment is replaced with a synthetic graft, a minimally invasive technique has evolved over the last decade called endovascular aneurysm repair (EVAR). In EVAR, an endovascular graft (EVG) is guided from the iliac to the affected area where the Corresponding author. Tel.: ; fax: address: ck@eos.ncsu.edu (C. Kleinstreuer). URL: CFPDlab/index.html (C. Kleinstreuer). EVG expands and forms a new artificial blood vessel, shielding the aneurysm from the pulsatile blood flow. The EVG or stent-graft is basically a cylindrical wire mesh embedded in synthetic graft material. While EVAR has shown outstanding success, especially for non-distorted abdominal aortic aneurysms (AAAs), EVG failure may occur due to blood leakage into the aneurysm cavity, which elevates the sac pressure and may cause rupture. This may also be caused by EVG migration when the drag force exerted on the EVG exceeds the fixation force, exposing the aneurysm sac again to the pulsatile blood flow. So far, experimental studies and computational work have focused on the analysis of blood flow induced wall stresses of either the AAA or the EVG [1 12]. For example, Fillinger et al. [1,2] computed the peak wall stresses in realistic AAA /$ see front matter 2005 IPEM. Published by Elsevier Ltd. All rights reserved. doi: /j.medengphy

2 370 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) configurations and declared the maximum stress to be the most important indicator of AAA rupture. Di Martino et al. [3,4] and Finol et al. [9] simulated the interactions between blood flow and aneurysm wall to analyze system parameters which are of concern in AAA-rupture risk assessment. Raghavan et al. [5] and Thubrikar et al. [7] investigated wall stress distribution on three-dimensionally reconstructed models of human abdominal aortic aneurysms. Chong and How [11] measured the flow patterns in an endovascular EVG for abdominal aortic aneurysm repair. Liffman et al. [12], Morris et al. [13], and Mohan et al. [14] investigated numerically or clinically the forces on a bifurcated EVG, assuming a rigid EVG wall. Typically, the EVG is anchored to the aortic neck due to frictional forces, where arterial ingrowth and/or stent barbs and hooks may play supportive roles. Clearly, a long cylindrical AAA neck with healthy tissue is most desirable for secure EVG placement [15,16]. Nevertheless, Resch et al. [17] reported that 45% of patients after EVAR showed migration of their EVGs. Other than problematic AAA necks, Volodos et al. [18] added systemic hypertension as a major factor in EVG migration. So far, most publications focused on AAA-wall stress or EVG-lumen flow separately. However, a stented AAA is a complex and strongly coupled system between blood flow and EVG/AAA wall. Thus, in order to evaluate blood flow patterns, wall stress distributions, sac pressure and EVG drag forces, fluid structure interaction (FSI) dynamics have to be considered. Presently, no publication dealt with fully coupled FSI dynamics of stented AAAs. The purpose of this paper was to analyze the pulsatile 3D hemodynamics and its impact on EVG placement, EVG/AAA wall stress distributions, sac pressure generation, and EVG drag force in a representative stented AAA. 2. Methods 2.1. System geometry The representative 3D asymmetric stented AAA model is shown in Fig. 1. The interacting materials include the luminal blood, EVG wall, stagnant blood in the AAA cavity, and AAA wall. The EVG is assumed to be a uniform 3D bifurcating shell attached to the proximal neck and iliac artery wall. The cavity between EVG and aneurysm wall is filled with stagnant blood, experiencing a time-dependent pressure as a result of the dynamic fluid structure interactions. The drag force exerted on the EVG, due to fluid friction and net momentum change, was computed under physiologically realistic conditions to assess incipient EVG migration Flow equations For transient three-dimensional incompressible fluid flow, the governing equations in tensor (or comma) notation, following Einstein s repeated index convention, are: continuity : u i,i = 0 (1) momentum : in F Ω(t) ρ u i t +ρ(u j û j )u i,j = p i,j + τ ij,j + ρf i (2a) stress tensor : τ ij = η γ ij (2b) non-newtonian fluid model : η p η = [ ( k 0 +k γ 1/2 r 1+ γ 1/2 r ) ] 2 (2c) Ht where u i is the velocity vector, p i the pressure scalar, ρ the fluid density, f i the body force at time t per unit mass, û i the wall displacement velocity at time t, F Ω(t) the moving spatial domain upon which the fluid is described, γ ij the shear rate tensor, η p the plasma viscosity, γ r = γ/ γ c a relative shear rate, γ c is defined by a phenomenological kinetic model [19], k 0 the lower limit Quemada viscosity constant, k the upper limit Quemada viscosity constant, and Ht is the hemotocrit. The parameter values for the blood and the structures are given in Section Structure equations The general governing equations for structure dynamics are: momentum : ρa i = σ ij,j + ρf i in S Ω(t) (3a) a i = dû i dt where (3b) equilibrium of condition : σ ij n i = T i on S Γ (t) (4) and constitutive : σ ij = D ijkl ε kl in S Ω(t) (5) Here, S Ω(t) is the structure domain at time t, n i the outward pointing normal on the wall surface S Γ (t), T i the surface traction vector at time t, S Γ (t) the boundary of the structure domain, σ ij the mechanical stress tensor, D ijkl the Lagrangian elasticity tensor, and ε kl is the strain tensor. In order to analyze the stress distributions in the aneurysm wall, the Von Mises stress, used as a material fracture criterion in complicated geometries, is employed. Especially, 2 σ Von Mises = (σ 1 σ 2 ) 2 + (σ 2 σ 3 ) 2 + (σ 3 σ 1 ) 2 2 (6) where σ 1, σ 2, and σ 3 are the three principal stresses.

3 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 1. (a) Representative AAA with EVG, (b) system inlet and outlet profiles, and (c) actual EVG photograph (Zenith, Cook Incorporated 2003C).

4 372 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Table 1 Assumptions for blood flow and structure characteristics Blood flow Incompressible Non-Newtonian fluid (Quemada model) Laminar blood flow No slip at the wall Blood particle effects not considered No endoleaks in stented aneurysm models Stagnant blood in cavity Structure characteristics (artery wall, aneurysm wall, and EVG) Isotropic and elastic Incompressible Non-linear (large deformations) No tissue growth on walls No residual stresses EVG consists of equivalently uniform material No EVG migration 2.4. Numerical method The underlying assumptions for simulating the coupled fluid structure interactions are listed in Table 1. The blood parameter values in Eq. (2c) are ρ = g/cm 3, k 0 = , k = , Ht = 40%, and η p = dyn/cm 2 [19]. Table 2 lists the structure parameter values used in the present simulations. With respect to Young s moduli for the arterial wall and aneurysm, experimental data indicate that Young s modulus of an aneurysm is much higher than for a normal artery [7]. In this paper, the Young s modulus is assumed to be 4.66 MPa. The healthy artery section (neck) is incompressible with a Poisson ratio of 0.49, and the aneurysm wall is nearly incompressible with a Poisson ratio of 0.45 [3]. For a bifurcated NiTi-stent interwoven with graft material (EVG), no direct experimental data are available; thus, an equivalent Young s modulus for the uniform EVG configuration was assumed to be10 MPa [20]. The physiologically representative inflow velocity waveform is shown in Fig. 1 with a maximum Reynolds number (Re max ) of 1950 and average Reynolds number (Re average )of 330. For the outlet pressure (see Fig. 1), the peak and average pressures are 122 and 98.7 mmhg, respectively [21]. The pulse period is chosen as T = 1.2 s. The inlet velocity profile is assumed to be parabolic. For the composite structures, the boundary conditions are a fixed degrees-of-freedom (DOF) at the inlet and the exit, and a free DOF on the wall. The fluid structure interactions occur on the interfaces between luminal blood flow and EVG wall, as well as the sac s stagnant blood and aneurysm/evg wall. The finite element software package (ANSYS7.1) for linear and non-linear multi-physics analysis in Arbitrary Fig. 2. Flow chart of fluid structure interaction solution algorithm. Lagrangian Eulerian (ALE) formulation has been employed to solve this fluid structure interaction problem. Specifically, it uses separately ANSYS FLOTRAN for the fluid domain and ANSYS structural solver for the solid parts. It transfers fluid forces, solid displacements, and velocities across the fluid solid interface (see Fig. 2). A total of 76,730 8-node fluid elements were needed for meshing the luminal blood flow region, 66,820 structure elements for the EVG and AAA Table 2 Parameters required in the simulation Parameters Normal artery Aneurysm EVG Wall thickness 1.5 mm 1.0 mm Equivalent: 0.2 mm Diameter Neck aorta (inner): 17 mm 60 mm Main body diameter: 17 mm Iliac artery (inner): 11 mm Iliac leg: 11 mm Length Neck aorta:30 mm 80 mm Main body: 60 mm Iliac artery:30 mm Iliac leg: 70 mm Young s modulus 1.2 MPa 4.66 MPa Equivalent: 10 MPa Poisson ratio Equivalent: 0.27 Density 1.12 g/cm g/cm 3 Equivalent: 6.0 g/cm 3

5 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Table 3a Comparison of results between simulations and theoretical analyses ANSYS FSI simulation results Theoretical results [22] Error (%) Straight artery Radial deformation (internal wall) cm cm 4.1 Circumferential stress (internal wall) MPa MPa 5.0 Radial strain (internal wall) Radial strain (external wall) Aneurysm Maximum circumferential stress MPa 0.16 MPa (Laplace s equation) 8.6% wall, and 19,250 non-net-flow fluid elements for the stagnant cavity. The number of elements on the FSI interface was The algorithm continues to loop through the solid and fluid analyses until convergence is reached for each time step (see Fig. 2). Convergence in the stagger loop is based on the quantities being transferred at the fluid solid interface. A variable time step was employed, where t min = s with 60 total time steps per cycle. Six cycles were required to achieve convergence for the transient analysis. Using a single processor, the total CPU time was about 50 h on an IBM p690 workstation Model validations Comparison to theoretical results and clinical observations In order to test the accuracy of the ANSYS-ALE-FSI solver for simulating interacting flow and wall phenomena, several computer model validation studies were performed. The comparison between our present simulations and theoretical analyses as well as clinical observations is listed in Tables 3a and 3b Comparison with Womersley s wave propagation theory The wave propagation equation for viscous blood flow in a longitudinally tethered elastic vessel is 1 c F10 = c 0 1 σ 2 (7) where c 0 is the Moens Korteweg wave speed, which depends on the system s geometric and material properties; F 10 a function of the Womersley number, Bessel functions, and Young s modulus; and σ is Poisson s ratio of the elastic tube. Using a Womersley number of 10 and Young s modulus of 1.0 MPa, compared with the wave propagation Table 3c Experimental parameters for stented aneurysm Aneurysm model type Latex fusiform EVG type Knitted polyethylene Perfusate Saline solution Aneurysm length 50 mm Maximum aneurysm diameter 60 mm Aneurysm volume 80 ml Neck diameter 18 mm Wall compliance (%/100 mmhg) 6 coats: 3.45 ± 0.5; 12 coats: 1.4 ± 0.2 Pressure mmhg Flow rate l/min value of c * = cm/s from Eq. (7) for a longitudinally tethered elastic tube, ANSYS-ALE-FSI simulation results yielded c * = 880 cm/s, i.e., the error is 5.9% Comparison with experimental data Gawenda et al. [26] measured the sac pressure with in vitro stented aneurysm models consisting of 6 or 12 layers of latex. The model parameters are listed in Table 3c. Now, the sac pressures were computed for similar stented aneurysm models and flow conditions employing the ANSYS-ALE-FSI-solver. Fig. 3 presents the comparison between the experimental and numerical results. It can be seen that the simulation results are in good agreement with the experimental data. 3. Results The results are divided into three groups, i.e., the beneficial impact of EVG insertion (Figs. 4 and 5), the luminal blood velocity fields, wall stress distributions and sac pressure levels at three selected time levels during the cardiac cycle (Figs. 6 8), and the effect of blood pressure waveforms on the transient EVG drag force, calculated from the wall Table 3b Comparison of results between simulations and clinical observations ANSYS FSI simulation results Clinical observations Von Mises stress drop 82% 75% [25] Diameter decrease 12.0% 10% [23] Maximum wall displacement 1.52 mm in non-stented aneurysm; mm in stented aneurysm. 1.0 mm in non-stented aneurysm and 0.2 mm in stented aneurysm [24]

6 374 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 3. Comparison between simulation results and experimental data. shear stress distribution and the net momentum change (see Fig. 9) EVG impact A securely placed EVG forms a new smooth conduit for blood flow (Fig. 1) and hence protects the weakened aneurysm wall from high pressure and stress levels (Fig. 4). In fact, in case of zero leakage into the cavity, the sac pressure is reduced by a factor of 10 and the maximum wall stress decreases by a factor of 20 throughout the cardiac cycle when an EVG is inserted (see Fig. 4a and b). As a result, the largest change in maximum AAA diameter drops during systole from 2 to 0.17 mm (Fig. 4c). Indeed, as reported by Malina et al. [24], EVG placement can reduce the maximum wall deformation to 0.2 mm. Clearly, these dynamic system parameter variations due to cyclic fluid structure interactions are powered by the given blood pressure waveform (cf. Figs. 1 and 4, keeping in mind that the inlet pressure differs only by 5% from the outlet pressure). A snapshot during the critical deceleration phase (t/t = 0.27, peak blood pressure) shows the velocity fields as well as the pressure and wall stress distributions in both the open AAA and the stented AAA (Fig. 5). Referring to Fig. 5a, the angulated AAA neck guides the jet-like blood stream into the large cavity, forming immediately two strong vortices due to the sudden area expansion. The greatest impact of the angled blood stream occurs, as can be expected, not where the an eurysm has its maximum diameter, but well below near the iliac bifurcation, i.e., the impact area of highest net momentum change. Interestingly, the maximum wall stress and wall deflection in non-stented AAA have their highest magnitudes in different locations, i.e., the maximum wall stress (0.59 MPa) occurs on the right side above the bifurcation; while the maximum wall deformation (2.3 mm) occurs on the left side near the angled neck. Now, with an EVG in place, the blood flow is tubular and rather uniform, except for a few areas of secondary flows caused by neck and iliac angles (Fig. 5b). As a result of complex fluid structure interactions between lumen blood, EVG wall, stagnant sac, and AAA wall, the sac pressure still remains at mmhg, i.e., 11.8% of the lumen pressure. Triggered by the EVG wall shear stress and net momentum change, the drag force acting on the EVG is at that moment almost 2 N. While the stress in the AAA wall and the pressure in the cavity are very low, the EVG carries the blood flow impact as numerically indicated with a peak EVG wall stress of 1.7 MPa at the stagnation (or bifurcation) point (see Fig. 5b and c) Transient fluid structure interactions In order to illustrate the dynamics of pulsatile blood flow influencing the stented AAA parameters (see Figs. 6 8), three representative time levels were selected, i.e., reverse flow at t/t = 0.1 (Re = 70), peak systole at t/t = 0.2 (Re = 1950), and flow deceleration at t/t = 0.27 (Re = 1200). In general, the sac pressure stays very low throughout the cycle; however, it increases slightly as the flow input waveform progresses (and time elapses) because of the blood pressure transmission from the EVG lumen via the distensible EVG wall into the aneurysm cavity. The same can be observed for the peak σ max in the AAA wall, occurring always in the same location, which is basically an inflection point in the geometric midplane wall function. Naturally, the stress in the (healthy) neck tissue is the highest because of the necessary oversizing to keep the EVG anchored. The luminal blood flow meanders from the AAA neck part through the aneurysm with typically four small recirculation zones until it bifurcates into the EVG daughter tubes, i.e., the iliac legs. As expected, the maximum EVG wall stress can be found at the EVG bifurcation, nonlinearly increasing during the observation time. Although the present AAA geometry is anterior posterior symmetric, the flow field and wall stress distributions are asymmetric, i.e., the results are fully three-dimensional (see Figs. 6 8). Comparing slices C C in Figs. 6 8 more closely, reveals that the location of the maximum EVG stress switches in the daughter tubes between up-flow (Fig. 6) and peak downflow (Fig. 7), and back again during decelerating down-flow (Fig. 8). Strong secondary flows appear before the EVG flow

7 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 4. EVG impacts on AAA: (a) EVG impact on sac pressure, (b) EVG impact on maximum wall stress, and (c) EVG impact on d AAA,max changes. division (see slices B B in Figs. 6 8), an area which also experiences a relatively high wall stress EVG migration If the actual EVG drag force starts to exceed the fixation force, the EVG will migrate or dislodge. As mentioned, the drag force is composed of the integral over the surface shear plus the net pressure, where the shear stress contribution is typically only 3% of the total. The fixation force for an EVG without barbs and hooks is the friction between the proximal EVG segment and the AAA neck which is usually oversized to supply, at least initially, solid anchoring. In addition, the EVG ends may be secured via frictional effects

8 376 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 5. Comparison between non-stented AAA and stented AAA (t/t = 0.27): (a) non-stented AAA, (b) stented AAA, and (c) wall stress distribution in EVG.

9 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 6. Fluid structure interactions of stented AAA (t/t = 0.1).

10 378 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 7. Fluid structure interactions of stented AAA (t/t = 0.2).

11 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 8. Fluid structure interactions of stented AAA (t/t = 0.27).

12 380 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) Fig. 9. Effect of blood pressure waveform on EVG drag force. to the iliac lumen. In case the EVG migrates, blood can leak into the AAA cavity, leading to a dangerous pressure buildup and subsequently aneurysm rupture. As alluded to in Section 1, ideally the AAA neck should be long, cylindrical, and of healthy tissue. Obviously, AAA patients with severe hypertension (see waveform IV in Fig. 9) are especially at risk for EVG failure because in that case the drag force has a maximum and the relatively large pressure difference during the cardiac cycle may reduce the EVG fixation to a calcified, i.e., hardened, neck tissue. Specifically, the pressure descriptions of waveforms I IV range from normal to severe, where p max,normal = 40 mmhg and p max,severe = 70 mmhg. For example, Mohan et al. [14] declared that high blood pressure is one important factor to cause EVG migration, while Morris et al. [13] found that the drag force may vary over the cardiac cycle between 3.9 and 5.5 N. Thus, for an EVAR patient with severe hypertension, extra fixation should be considered. Because of the incompressible-fluid condition, the transient drag force exhibits, with a very small time lag, basically the same trend as the given inlet pressure waveform (see Fig. 9). 4. Discussion Although minimally invasive endovascular aneurysm repair is very attractive, post-operative complications may occur, which are the result of excessive fluid structure interaction dynamics. Thus, the fluid structure interactions for a representative AAA model with and without a realistic EVG have been investigated in terms of pulsatile blood flow influencing EVG movement, which is transmitted via the stagnant blood in the cavity to the aneurysm wall. Of interest are the beneficial impact of an EVG, the blood velocity field, the highest pressure level in the aneurysm sac, the stress distributions, and displacements of both of the EVG wall and the AAA wall, as well as the maximum drag force exerted on the EVG. It was readily demonstrated that a securely placed EVG shields the diseased AAA wall from the pulsatile blood pressure and hence keeps the maximum wall stress 20 times below the wall stress value in the non-stented AAA. An interesting finding is that the sac pressure is reduced significantly but is not zero after the sac is completely excluded by the EVG, i.e., no blood leakage exists. Thus, our simulation shows that even in the absence of endoleaks, the sac pressure can be generated by the complex fluid structure interactions between luminal blood flow, EVG wall, stagnant sac blood, and aneurysm wall. The EVG/AAA wall compliance plays an important role in sac pressure generation. As indicated in Section 2.5, our simulation results are in good agreement with the experimental data of Gawenda et al. [26]. As shown in Figs. 7 9, the intra-sac pressure varies from 9.8 to 14.4 mmhg (4.6 mmhg pulsatility) during one cycle, while the average EVG lumen blood pressure (p lumen ) ranges from 82.3 to mmhg (38 mmhg pulsatility). Sonesson et al.[25] found that the sac pressure pulsatility varied from 0 to 6 mmhg clinically. Dias et al. [27] indicated that the mean sac pressure pulsatility ranged from 2 to 10 mmhg for their 30 patients. Compared to the lumen pressure pulse, the pressure pulsatility in the sac is not substantial. While the AAA volume should typically shrink after successful EVAR, aneurysm enlargement might still occur because an elevated sac pressure could cause a delayed aortic aneurysm enlargement, even after successful EVAR [28]. EVG wall deformations are insignificant due to the high material stiffness; indeed, our results show that the

13 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) maximum graft deformation is less than 1 mm. However, EVG wall deformation is a main factor when determining the sac pressure [26,32]. Furthermore, even relatively small, repetitive EVG-wall deformations may result in interactions between the metallic wire and the interwoven graft material, leading to fabric abrasion and holes [30]. In addition, transient solid fluid interactions can lead to material fatigue and hence device failure. Clinically, that includes metallic fracture (metal wire fracture, stent-strut, and barb/hook breakage), graft fabric holes and suture breakage, and/or separation. Additionally, device corrosion has also been observed clinically. So far, there is no model which can predict such device deteriorations [33]. The reason is the serious lack of time-dependent clinical data due to short follow-ups, especially for the second generation of stentgrafts. Another reason is that most of the patients have been asymptomatic and have not as of yet needed interventions for device fatigue [34]. In this study, we did not consider device failure cases. However, it is possible to simulate device fatigue employing a fluid structure interaction solver for a realistic stented AAA model in order to assess such problems. Endoleaks were not considered in this paper. However, in cases of a loose neck attachment, graft defect and/or minor branch backflows, blood may leak into the AAA sac after EVAR. For example, Zarins et al. [37] indicated that endoleaks were diagnosed with CT scans in 38% of 398 patients. Endoleaks may cause an increase in sac pressure and hence higher stresses in the AAA wall. Serious endoleaks can result in EVR failure, AAA rupture, and the need for second procedures. In a preliminary study, we found that if an endoleak volume of 3% is added to the sac volume, the sac pressure and AAA wall stress can increase by 60% [32]. Even though EVG placement may reduce significantly the sac pressure and wall stress, the hemodynamics incurs a drag force which may trigger EVG migration. It is shown that the risk of migration can be high for patients with severe hypertension (see Fig. 9). EVG migration is a common problem for EVAR patients. For example, Zarins et al. [29] reported that the migration rate was up to 8.4% for their 1119 patients within 12 months after EVAR. Serious EVG migration can cause endoleaks, EVG twisting or kinking and hence EVG failure. In this paper, the maximum EVG drag force is about 2 N because the selected EVG size is relative small (17 mm). However, our preliminary research results indicated that the EVG drag force can exceed 5 N for AAAs with a large neck angle, iliac angle, large EVG size, and aorto-uni-iliac EVG [32]. Thus, the fixation of self-expandable or balloonexpandable EVG contact may be inadequate to withstand the pulling forces caused by net momentum changes inside the endovascular graft. Means of extra fixation should be taken into account when possible. Concerning the supportive literature, Liffman et al. [12] and Morris et al. [13] confirmed that the drag force increases with EVG size. Mohan et al. reported that the drag force is non-linearly increasing with iliac angle [14]. Sternbergh et al.[35] declared that the migration rate increases by 30% for AAAs with neck angles greater than 40. Interestingly, compared with open surgery, the mortality of EVAR is almost the same. However, the key advantages of EVAR are reduced morbidity, shorter hospitalisation, and quicker recovery; but, these benefits may be offset by the cost of the device, the need for continuous control measurements, and eventually the need for late intervention or conversion to open repair [31]. It should be noted that in this study the AAA wall was assumed to be smooth. Intraluminal thrombus and small branches were not considered. In patient-specific stented AAAs, intraluminal thrombi and small braches may affect the fluid structure interactions, sac pressure and hence possible EVG migration. In the range of pressure loads from 80 to 120 mmhg, linearly elastic property values are generally used [3,6,7]. However, non-linear wall properties may provide even more realistic results [5,9,36]. Considering the difficulty to model an actual EVG, i.e., a NiTi-wire mesh interwoven with synthetic graft material, presently the EVG is assumed to be a uniform shell made of an equivalent composite material. Furthermore, the blood particle effects on the wall were not considered. In a future study, patient-specific CT scan models, non-linear material properties, impact of an intraluminal thrombus, blood particle effects, and additional EVG configurations will be considered. In summary, this study provides a new technique, i.e., fully coupled fluid structure interaction, to evaluate the impact of EVG placement, determine wall stresses and sac pressure values, calculate EVG migration forces, and provide physical insight into the biomechanics of stented AAAs. To the authors knowledge, this is the first computational FSI study of a stented AAA. Additional clinical applications, including optimal EVG designs, minimizing sac pressure magnitude, preventing endoleaks, and EVG migration, will be performed as future work. References [1] Fillinger MF, Raghavan ML, Marra P, Cronenwett L, Kennedy E. In vivo analysis of mechanical wall stress and abdominal aortic aneurysm rupture risk. J Vasc Surg 2002;36: [2] Fillinger MF, Marra PS, Raghavan ML, Kennedy EF. Prediction of rupture in abdominal aortic aneurysm during observation: wall stress versus diameter. J Vasc Surg 2003;37: [3] Di Martino ES, Guadagni G, Fumero A, Ballerini G, Spirito R, Biglioli P, et al. Fluid structure interaction within realistic threedimensional models of the aneurysmatic aorta as a guidance to assess the risk of rupture of the aneurysm. Med Eng Phys 2001;23: [4] Di Martino ES, Guadagni G, Fumero A, Spirito R, Redaelli A. In: Middleton J, Jones M, Shrive N, Pande G, editors. A computational study of the fluid structure interaction within a realistic aneurysmatic vessel model obtained from CT scans image processing, in computer methods in biomechanics and biomedical engineering. 3rd ed. Newark, NJ: Gordon and Breach Science Publishers; p

14 382 Z. Li, C. Kleinstreuer / Medical Engineering & Physics 27 (2005) [5] Raghavan M, Vorp D, Federle M, Makaroun M, Webster M. Wall stress distribution on three-dimensionally reconstructed models of human abdominal aortic aneurysm. J Vasc Surg 2000;31: [6] Vorp DA, Raghavan M, Webster M. Mechanical wall stress in abdominal aortic aneurysm: influence of diameter and asymmetry. J Vasc Surg 1998;27: [7] Thubrikar M, Al-Soudi J, Robicsek F. Wall stress studies of abdominal aortic aneurysm in a clinical model. Ann Vasc Surg 2001;3: [8] Finol EA, Keyhani K, Amon CH. The effect of asymmetry in abdominal aortic aneurysms under physiologically realistic pulsatile flow conditions. J Biomech Eng-Trans ASME 2003;125: [9] Finol EA, Di Martino ES, Vorp DA, Amon CH. Fluid structure interaction and structural analyses of an aneurysm model. In: Proceedings of the ASME 2003 summer bioengineering conference, p [10] Finol EA, Di Martino ES, Vorp DA, Amon CH. Biomechanics of patient specific abdominal aortic aneurysms: computational analysis of fluid flow. In: Proceedings of the 28th Annual IEEE northeast bioengineering conference, p [11] Chong CK, How TV. Flow patterns in an endovascular EVG for abdominal aortic aneurysm repair. J Biomech 2004;37: [12] Liffman K, Lawrence M, Semmens B, Bui A, Rudman M, Hartley D. Analytical modeling and numerical simulation of forces in an endoluminal graft. J Endovasc Ther 2001;8: [13] Morris L, Delassus P, Walsh M, McGloughlin T. A mathematical model to predict the in vivo pulsatile drag forces acting on bifurcated stent grafts used in endovascular treatment of abdominal aortic aneurysms (AAA). J Biomech 2004;37: [14] Mohan IV, Harris PL, van Marrewijk CJ, Laheij RJ, How TV. Factors and forces influencing EVG migration after endovascular aortic aneurysm repair. J Endovasc Ther 2002;9: [15] Lambert AW, Williams DJ, Budd JS, Horrocks M. Experimental assessment of proximal EVG (IntervascularTM) fixation in human cadaveric infrarenal aortas. Eur J Endovasc Surg 1998;17: [16] Resch T, Malina M, Lindblad B, Malina J, Brunkwall J, Ivancev K. The impact of stent design on proximal EVG fixation in the abdominal aorta: an experimental study. Eur J Endovasc Surg 2000;20: [17] Resch T, Ivancev K, Brunkwall J, Nyman U, Malina M, Lindblad B. Distal migration of EVGs after endovascular repair of abdominal aortic aneurysms. J Vasc Interv Radiol 1999;10: [18] Volodos SM, Sayers RD, Gostelow JP, Bell P. Factors affecting the displacement force exerted on a stent graft after AAA repair an in vitro study. Eur J Vasc Endovasc Surg 2003;26: [19] Buchanan JR, Kleinstreuer C, Comer JK. Rheological effects on pulsatile hemodynamics in a stenosed tube. Comput Fluids 2001;29: [20] Suzuki K, Ishiguchi T, Kawatsu S, Iwai H, Maruyama K, Ishigaki T. Dilatation of EVGs by luminal pressures: experimental evaluation of polytetrafluorothylene (PTFE) and Woven polyester grafts. Cardiovasc Interv Radiol 2001;24:94 8. [21] Meter O. Numerical simulation and experimental validation of blood flow in arteries with structured-tree outflow conditions. Ann Biomed Eng 2000;28: [22] Nichols WW, O Rourke MF. McDonald s blood flow in arteries. Philadelphia, London, UK: LEA & FEBIGER; p [23] Blankensteijn JD, Prinssen M. Does fresh clot shrink faster than preexistent mural thrombus after endovascular AAA repair? J Endovasc Ther 2002;9: [24] Malina M, Länne T, Ivancev K, Lindblad B, Brunkwall J. Reduced pulsatile wall motion of abdominal aortic aneurysms after endovascular repair. J Vasc Surg 1998;27: [25] Sonesson B, Dias N, Malina M, Olofsson P, Griffin D, Lindblad B, et al. Intra-aneurysm pressure measurement in successfully excluded abdominal aortic aneurysm after endovascular repair. J Vasc Surg 2003;37: [26] Gawenda M, Knez P, Winter S, Jaschke G, Wassmer G, Schmitz- Rixen T, et al. Endotension is influenced by wall compliance in a latex aneurysm model. Eur J Vasc Endovasc Surg 2004;27: [27] Dias N, Ivancev K, Malina M, Resch T, Lindblad B, Sonesson B. Intra-aneurysm sac pressure measurements after endovascular aneurysm repair: differences between shrinking, unchanged, and expanding aneurysms with and without endoleaks. J Vasc Surg 2004;39: [28] Lin PH, Bush RL, Katzman JB, Zemel G, Puente OA, Katzen BT, et al. Delayed aortic aneurysm enlargement due to endotension after endovascular abdominal aortic aneurysm repair. J Vasc Surg 2003;38: [29] Zarins CK, Bloch DA, Crabtree T, Matsumoto AH, White RA, Fogarty TJ. Stent graft migration after endovascular aneurysm repair: importance of proximal fixation. J Vasc Surg 2003;38: [30] Chakfe N, Dieval F, Riepe G, Mathieu D, Zbali I, Thaveau F, et al. Influence of the textile structure on the degradation of explanted aortic endoprostheses. Eur J Vasc Endovasc Surg 2004;27: [31] Lester JS, Bosch JL, Kaufman JA, Halpern EF, Gazelle GS. Inpatient cost of routine endovascular repair of abdominal aortic aneurysm. Acad Radiol 2001;7: [32] Li Z. Computational analysis and simulations of fluid structure interactions for stented abdominal aortic aneurysms. Ph.D. thesis. North Carolina State University; [33] Najibi S, Steinberg J, Katzen BT, Zemel G, Lin PH, Weiss VJ, et al. Detection of isolated hook fractures 36 months after implantation of the Ancure endograft: a cautionary note. J Vasc Surg 2001;34: [34] Jacobs TS, Won J, Gravereaux EC, Faries PL, Morrissey N, Teodorescu VJ, et al. Mechanical failure of prosthetic human implants: a 10-year experience with aortic stent graft devices. J Vasc Surg 2003;37: [35] Sternbergh WC, Carter G, York JW, Yoselevitz M, Money SR. Aortic neck angulation predicts adverse outcome with endovascular abdominal aortic aneurysm repair. J Vasc Surg 2002;35: [36] Wang D, Makaroun M, Webster M, Vorp DA. Effect of intraluminal thrombus on wall stress in patient specific model of abdominal aortic aneurysm. J Vasc Surg 2002;3: [37] Zarins CK, White RA, Hodgson KJ, Schwarten D, Fogarty TJ. Endoleak as a predictor of outcome after endovascular aneurysm repair: aneurx multicenter clinical trial. J Vasc Surg 2000;32:

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms

Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms Chapter 8 Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms Efstratios Georgakarakos, Antonios Xenakis, George S. Georgiadis, Konstantinos C. Kapoulas, Evagelos Nikolopoulos and

More information

A Computational Study of the Magnitude and Direction of Migration Forces in Patient-specific Abdominal Aortic Aneurysm Stent-Grafts

A Computational Study of the Magnitude and Direction of Migration Forces in Patient-specific Abdominal Aortic Aneurysm Stent-Grafts Eur J Vasc Endovasc Surg (2010) 40, 332e339 A Computational Study of the Magnitude and Direction of Migration Forces in Patient-specific Abdominal Aortic Aneurysm Stent-Grafts D.S. Molony a, E.G. Kavanagh

More information

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD International Journal of Mechanical Engineering and Technology (IJMET) Volume 9, Issue 11, November 2018, pp. 684 688, Article ID: IJMET_09_11_068 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=9&itype=11

More information

From 1996 to 1999, a total of 1,193 patients with

From 1996 to 1999, a total of 1,193 patients with THE ANEURX CLINICAL TRIAL AT 8 YEARS Lessons learned following the US AneuRx clinical trial from 1996 to 2004. BY CHRISTOPHER K. ZARINS, MD From 1996 to 1999, a total of 1,193 patients with infrarenal

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta No Author Given No Institute Given Abstract. Abdominal Aortic Aneurysm is a potentially life-threatening disease if not treated adequately.

More information

aneurysm initiation and fully developed aneurysms

aneurysm initiation and fully developed aneurysms Calhoun: The NPS Institutional Archive Faculty and Researcher Publications Faculty and Researcher Publications Collection 2015 Investigation of material modeling in fluid structure interaction analysis

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 75-83 Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Taedong Kim, Taewon Seo* 1,2 and Abdul.I.

More information

TriVascular Ovation Prime Abdominal Stent Graft System

TriVascular Ovation Prime Abdominal Stent Graft System TriVascular Ovation Prime Abdominal Stent Graft System Science of the Seal O-Ring Sealing Technology O-Ring Sealing in Proven Engineering Solutions O-rings are designed to seal by blocking the flow of

More information

Mathematical Model Analysis of Wallstent and AneuRx

Mathematical Model Analysis of Wallstent and AneuRx Computational Investigation Suncica Canic, PhD K. Ravi-Chandar, PhD Zvonimir Krajcer, MD, FACC Dragan Mirkovic, PhD Serguei Lapin, PhD Key words: Aortic aneurysm, abdominal; blood flow velocity; blood

More information

Eccentric stent graft compression: An indicator of insecure proximal fixation of aortic stent graft

Eccentric stent graft compression: An indicator of insecure proximal fixation of aortic stent graft Eccentric stent graft compression: An indicator of insecure proximal fixation of aortic stent graft Yehuda G. Wolf, MD, Bradley B. Hill, MD, W. Anthony Lee, MD, Christine M. Corcoran, RN, MS, Thomas J.

More information

A Comparison Between the Dynamic Responses of Bare-Metal Wallstent Endoprosthesis and AneuRx Stent-Graft: A Mathematical Model Analysis

A Comparison Between the Dynamic Responses of Bare-Metal Wallstent Endoprosthesis and AneuRx Stent-Graft: A Mathematical Model Analysis 1 A Comparison Between the Dynamic Responses of Bare-Metal Wallstent Endoprosthesis and AneuRx Stent-Graft: A Mathematical Model Analysis Sunčica Čanić, Ph.D.(corresponding author) Department of Mathematics,

More information

Fluid-Structure Interaction Analyses of Stented Abdominal Aortic Aneurysms

Fluid-Structure Interaction Analyses of Stented Abdominal Aortic Aneurysms Annu. Rev. Biomed. Eng. 2007. 9:169 204 First published online as a Review in Advance on March 15, 2007 The Annual Review of Biomedical Engineering is online at bioeng.annualreviews.org This article s

More information

ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE

ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE Badreddin Giuma s.k\ Kahar Osman 1 and Mohamed Rafiq Abdul Kadir 1,2 1Faculty of Mechanical Engineering,

More information

Iliac fixation inhibits migration of both suprarenal and infrarenal aortic endografts

Iliac fixation inhibits migration of both suprarenal and infrarenal aortic endografts From the Society for Vascular Surgery Iliac fixation inhibits migration of both suprarenal and infrarenal aortic endografts Peyman Benharash, MD, Jason T. Lee, MD, Oscar J. Abilez, MD, Tami Crabtree, MS,

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

The effect of angulation in abdominal aortic aneurysms: fluid structure interaction simulations of idealized geometries

The effect of angulation in abdominal aortic aneurysms: fluid structure interaction simulations of idealized geometries DOI 10.1007/s11517-010-0714-y ORIGINAL ARTICLE The effect of angulation in abdominal aortic aneurysms: fluid structure interaction simulations of idealized geometries Michalis Xenos Yared Alemu Dan Zamfir

More information

Computational mechanics of Nitinol stent grafts

Computational mechanics of Nitinol stent grafts Journal of Biomechanics 41 (2008) 2370 2378 www.elsevier.com/locate/jbiomech www.jbiomech.com Computational mechanics of Nitinol stent grafts C. Kleinstreuer a,b,d,,z.li c, C.A. Basciano a, S. Seelecke

More information

Access More Patients. Customize Each Seal.

Access More Patients. Customize Each Seal. Access More. Customize Each Seal. The Least Invasive Path Towards Proven Patency ULTRA LOW PROFILE TO EASE ADVANCEMENT The flexible, ultra-low 12F ID Ovation ix delivery system enables you to navigate

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Degeneration of the Neck Post Implementation - a New Era of AAA Stent

Degeneration of the Neck Post Implementation - a New Era of AAA Stent Degeneration of the Neck Post Implementation - a New Era of AAA Stent New Mexico Heart Institute Albuquerque, New Mexico USA - Gore Current FDA-Approved EVAR Devices Sealing Mechanism Endologix Lombard

More information

Nellix Endovascular System: Clinical Outcomes and Device Overview

Nellix Endovascular System: Clinical Outcomes and Device Overview Nellix Endovascular System: Clinical Outcomes and Device Overview Jeffrey P. Carpenter, MD Professor and Chief, Department of Surgery CAUTION: Investigational device. This product is under clinical investigation

More information

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully

Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully Physician Training Bifurcated system Proximal suprarenal stent Modular (aortic main body and two iliac legs) Full thickness woven polyester graft material Fully supported by self-expanding z-stents H&L-B

More information

Management of Endoleaks

Management of Endoleaks Management of Endoleaks Sarah Ikponmwosa, MD Brooklyn VA 6/20/08 Questions Advantages of endovascular repair Definition of an endoleak Types of endoleaks Management of type lll endoleak Diagnosis of type

More information

Description. Section: Surgery Effective Date: April 15, Subsection: Surgery Original Policy Date: December 6, 2012 Subject:

Description. Section: Surgery Effective Date: April 15, Subsection: Surgery Original Policy Date: December 6, 2012 Subject: Last Review Status/Date: March 2015 Page: 1 of 6 Description Wireless sensors implanted in an aortic aneurysm sac after endovascular repair are being investigated to measure post procedural pressure. It

More information

Periprosthetic leak and rupture after endovascular repair of abdominal aortic aneurysm: The significance of device design for long-term results

Periprosthetic leak and rupture after endovascular repair of abdominal aortic aneurysm: The significance of device design for long-term results CASE REPORTS Periprosthetic leak and rupture after endovascular repair of abdominal aortic aneurysm: The significance of device design for long-term results Kirsten Krohg-Sørensen, MD, PhD, Magne Brekke,

More information

MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE

MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE MODERN METHODS FOR TREATING ABDOMINAL ANEURYSMS AND THORACIC AORTIC DISEASE AAA FACTS 200,000 New Cases Each Year Ruptured AAA = 15,000 Deaths per Year in U.S. 13th Leading Cause of Death 80% Chance of

More information

Increased Flexibility of AneuRx Stent-Graft Reduces Need for Secondary Intervention Following Endovascular Aneurysm Repair

Increased Flexibility of AneuRx Stent-Graft Reduces Need for Secondary Intervention Following Endovascular Aneurysm Repair 583 Increased Flexibility of AneuRx Stent-Graft Reduces Need for Secondary Intervention Following Endovascular Aneurysm Repair Frank R. Arko, MD; W. Anthony Lee, MD; Bradley B. Hill, MD; Paul Cipriano,

More information

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta Robert H.P. McGregor 1, Dominik Szczerba 1,andGábor Székely 1 Computer Vision Laboratory, Sternwartstr. 7, 8092 Zürich, Switzerland

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

In Vitro Evaluation of Endovascular Stents to Assess Suitability for Endovascular Graft Fixation*

In Vitro Evaluation of Endovascular Stents to Assess Suitability for Endovascular Graft Fixation* Eur J Vasc Endovasc Surg 9, 403-407 (1995) In Vitro Evaluation of Endovascular Stents to Assess Suitability for Endovascular Graft Fixation* S. M. Andrews ~, A. W. Anson 2, R. M. Greenhalgh ~ and D. M.

More information

Abdominal and thoracic aneurysm repair

Abdominal and thoracic aneurysm repair Abdominal and thoracic aneurysm repair William A. Gray MD Director, Endovascular Intervention Cardiovascular Research Foundation Columbia University Medical Center Abdominal Aortic Aneurysm Endografts

More information

Length Measurements of the Aorta After Endovascular Abdominal Aortic Aneurysm Repair

Length Measurements of the Aorta After Endovascular Abdominal Aortic Aneurysm Repair Eur J Vasc Endovasc Surg 18, 481 486 (1999) Article No. ejvs.1999.0882 Length Measurements of the Aorta After Endovascular Abdominal Aortic Aneurysm Repair J. J. Wever, J. D. Blankensteijn, I. A. M. J.

More information

Improved Assessment and Treatment of Abdominal Aortic Aneurysms: The Use of 3D Reconstructions as a Surgical Guidance Tool in Endovascular Repair

Improved Assessment and Treatment of Abdominal Aortic Aneurysms: The Use of 3D Reconstructions as a Surgical Guidance Tool in Endovascular Repair Improved Assessment and Treatment of Abdominal Aortic Aneurysms: The Use of 3D Reconstructions as a Surgical Guidance Tool in Endovascular Repair Barry J. Doyle, 1 Pierce A. Grace, 1,2 Eamon G. Kavanagh,

More information

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation EasyChair Preprint 259 Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation Tianyi Xia, Matthew Doyle, Thomas Forbes and Cristina H. Amon EasyChair preprints

More information

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction The Study o Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction Taewon Seo*, Guek-don Shin* and Do-Il Kim** *Department o Mechanical and Automotive Engineering, Andong National University,

More information

An Overview of Post-EVAR Endoleaks: Imaging Findings and Management. Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC

An Overview of Post-EVAR Endoleaks: Imaging Findings and Management. Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC An Overview of Post-EVAR Endoleaks: Imaging Findings and Management Ravi Shergill BSc Sean A. Kennedy MD Mark O. Baerlocher MD FRCPC Disclosure Slide Mark O. Baerlocher: Current: Consultant for Boston

More information

Hostile Proximal Neck: A New Conformable EVAR Device

Hostile Proximal Neck: A New Conformable EVAR Device Hostile Proximal Neck: A New Conformable EVAR Device Young-Guk Ko, M.D., Seoul, Korea Currently Available Devices for EVAR in Korea, 2018 Zenith Flex, Cook Endurant IIs, Medtronic INCRAFT, Cordis AFX2,

More information

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis Biomedical & Pharmacology Journal Vol. 8(1), 123-131 (2015) Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis M. JAHANGIRI 1 *, M.

More information

FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall

FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall Engineering Letters, 13:2, EL_13_2_15 (Advance online publication: 4 August 2006) FSI within Aortic Arch Model over Cardiac Cycle and Influence of Wall Stiffness on Wall Stress in Layered Wall F. Gao and

More information

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca,

More information

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

Technique and Tips for Complicated AAA Cases with Stent Graft

Technique and Tips for Complicated AAA Cases with Stent Graft Technique and Tips for Complicated AAA Cases with Stent Graft Seung-Woon Rha, MD, PhD FACC, FAHA, FESC, FSCAI, FAPSIC Cardiovascular Center, Korea University Guro Hospital Mar 15, 2018 LINC AP 2018 Endoleak;

More information

REGIONAL MECHANICAL PROPERTIES OF AAA TISSUE AND FINITE ELEMENT ANALYSIS OF RUPTURE

REGIONAL MECHANICAL PROPERTIES OF AAA TISSUE AND FINITE ELEMENT ANALYSIS OF RUPTURE USE OF REGIONAL MECHANICAL PROPERTIES OF ABDOMINAL AORTIC ANEURYSMS TO ADVANCE FINITE REGIONAL MECHANICAL PROPERTIES OF AAA ELEMENT MODELING OF RUPTURE RISK TISSUE AND FINITE ELEMENT ANALYSIS OF RUPTURE

More information

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Eann A. Patterson Department of Mechanical Engineering, The University of Sheffield Mappin Street, Sheffield, S1 3JD

More information

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents.

Keywords: Angioplasty, Explicit finite elements method, Tube hidroforming, Stents. Blucher Mechanical Engineering Proceedings May 2014, vol. 1, num. 1 www.proceedings.blucher.com.br/evento/10wccm AN ANALYSIS OF THE CONTACT BETWEEN THE STENT AND THE ARTERY USING TUBE HIDROFORMING SIMULATION

More information

Zenith Renu AAA Converter Graft. Device Description Planning and Sizing Deployment Sequence Patient Follow-Up

Zenith Renu AAA Converter Graft. Device Description Planning and Sizing Deployment Sequence Patient Follow-Up Zenith Renu AAA Converter Graft Device Description Planning and Sizing Deployment Sequence Patient Follow-Up Device description: Device indications The Zenith Renu AAA Converter Graft with Z-Trak Introduction

More information

Quantitative Indicator of Abdominal Aortic Aneurysm Rupture Risk Based on its Geometric Parameters

Quantitative Indicator of Abdominal Aortic Aneurysm Rupture Risk Based on its Geometric Parameters Quantitative Indicator of Abdominal Aortic Aneurysm Rupture Risk Based on its Geometric Parameters Guillermo Vilalta, Félix Nieto, Carlos Vaquero, José A. Vilalta Abstract Abdominal aortic aneurysms rupture

More information

Aortic stents, types, selection, tricks in deployment.

Aortic stents, types, selection, tricks in deployment. Aortic stents, types, selection, tricks in deployment. Hamdy Soliman.M.D,FSCAI Consultant of Cardiology&Head of Endovascular Unit National Heart Institute Endovascular Treatment of Thoracic Aortic Aneurysms

More information

Endologix PowerWeb System EPW?

Endologix PowerWeb System EPW? 13 579 583 2004 Endologix PowerWeb System EPW? Endologix PowerWeb System EPW (AAA) 1993 7 2003 11 AAA 176 155 21 52 897240 120mm 53.5mm EPWEPW 1 2 proximal neck PN 15mm 3 PN 23mm 4 distal neck DN 15mm

More information

Talent Abdominal Stent Graft

Talent Abdominal Stent Graft Talent Abdominal with THE Xcelerant Hydro Delivery System Expanding the Indications for EVAR Treat More Patients Short Necks The Talent Abdominal is the only FDA-approved device for proximal aortic neck

More information

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ

History of the Powerlink System Design and Clinical Results. Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ History of the Powerlink System Design and Clinical Results Edward B. Diethrich Arizona Heart Hospital Phoenix, AZ Powerlink System: Unibody-Bifurcated Design Long Main Body Low-Porosity Proprietary eptfe

More information

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim

Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University Hospital Sangmin Kim Endovascular Procedures for Isolated Common Iliac and Internal Iliac Aneurysm Chungbuk Regional Cardiovascular Center, Division of Cardiology, Departments of Internal Medicine, Chungbuk National University

More information

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Saccular Aneurysm Formation in Curved and Bifurcating Arteries AJNR Am J Neuroradiol 20:1309 1317, August 1999 Saccular Aneurysm Formation in Curved and Bifurcating Arteries George N. Foutrakis, Howard Yonas, and Robert J. Sclabassi BACKGROUND AND PURPOSE: Distinguishing

More information

Abdominal Aortic Aneurysm (AAA)

Abdominal Aortic Aneurysm (AAA) Abdominal Aortic Aneurysm (AAA) Vascular Workshop: Objectives Anatomy Keith VanHaltren Indications Technique Cases Abdominal Aorta: Normal Size Abdominal aortic aneurysm: Definition Normal diameter of

More information

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging Original Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging P. Khamdaengyodtai 1, T. Khamdaeng 1, P. Sakulchangsatjatai 1, N. Kammuang-lue

More information

Abdominal Aortic Aneurysms. A Surgeons Perspective Dr. Derek D. Muehrcke

Abdominal Aortic Aneurysms. A Surgeons Perspective Dr. Derek D. Muehrcke Abdominal Aortic Aneurysms A Surgeons Perspective Dr. Derek D. Muehrcke Aneurysm Definition The abnormal enlargement or bulging of an artery caused by an injury or weakness in the blood vessel wall A localized

More information

Considerations for a Durable Repair

Considerations for a Durable Repair Considerations for a Durable Repair Eric Verhoeven, MD, PhD, A. Katsargyris, MD Department of Vascular and Endovascular Surgery, Paracelsus Medical University, Nuremberg, Germany Disclosures William Cook

More information

Reliable World Class Insights Your Silicon Valley Partner in Simulation ANSYS Sales, Consulting, Training & Support

Reliable World Class Insights Your Silicon Valley Partner in Simulation ANSYS Sales, Consulting, Training & Support www.ozeninc.com info@ozeninc.com (408) 732 4665 1210 E Arques Ave St 207 Sunnyvale, CA 94085 Reliable World Class Insights Your Silicon Valley Partner in Simulation ANSYS Sales, Consulting, Training &

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

BioMedical Engineering OnLine. Open Access RESEARCH

BioMedical Engineering OnLine. Open Access RESEARCH Li et al. BioMed Eng OnLine 2016, 15(Suppl 2):135 DOI 10.1186/s12938-016-0249-z BioMedical Engineering OnLine RESEARCH Open Access Influence of endoleak positions on the pressure shielding ability of stent

More information

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA)

Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA) Computer Aided Surgery (CAS) for Abdominal Aortic Aneurysm (AAA) Aurélien Dumenil Pierre Louat Thierry Marchal Michel Rochette 2009 ANSYS, Inc. All rights reserved. 1 ANSYS, Inc. Proprietary Outline Clinical

More information

Anatomical challenges in EVAR

Anatomical challenges in EVAR Anatomical challenges in EVAR M.H. EL DESSOKI, MD,FRCS PROFESSOR OF VASCULAR SURGERY CAIRO UNIVERSITY Disclosure Speaker name:... I have the following potential conflicts of interest to report: Consulting

More information

From the Western Vascular Society

From the Western Vascular Society From the Western Vascular Society The role of aortic neck dilation and elongation in the etiology of stent graft migration after endovascular abdominal aortic aneurysm repair with a passive fixation device

More information

Post-conditioning. P a g e 1. To my Thesis Committee,

Post-conditioning. P a g e 1. To my Thesis Committee, P a g e 1 To my Thesis Committee, This document seeks to clarify my research project. After describing what post-conditioning (PC) is, I will explain differences between my research and the recent peristaltic

More information

Endograft Technology: Highlights of the Past 10 Years

Endograft Technology: Highlights of the Past 10 Years II-192 J ENDOVASC THER REVIEW Endograft Technology: Highlights of the Past 10 Years Thomas J. Fogarty, MD; Frank R. Arko, MD; and Christopher K. Zarins, MD Division of Vascular Surgery, Stanford University

More information

Blood flow in vessels with artificial or pathological geometrical changes

Blood flow in vessels with artificial or pathological geometrical changes Blood flow in vessels with artificial or pathological geometrical changes P. Tibaut 1, B. Wiesler 1, M. Mayer 2 & R. Wegenkittel 3 1 AVL LIST GmbH, Graz, Austria 2 VRVIs, Vienna, Austria 3 Tiani Medgraph

More information

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA

Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type 1b Endoleak following an EVAR for Ruptured AAA Hindawi Case Reports in Vascular Medicine Volume 2017, Article ID 4931282, 5 pages https://doi.org/10.1155/2017/4931282 Case Report Early and Late Endograft Limb Proximal Migration with Resulting Type

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

Management of Endoleaks. Michael Meuse, M.D Vascular and Interventional Radiology 12/14/09

Management of Endoleaks. Michael Meuse, M.D Vascular and Interventional Radiology 12/14/09 Management of Endoleaks Michael Meuse, M.D Vascular and Interventional Radiology 12/14/09 Endoleak Failure to totally exclude the abdominal aortic aneurysm (AAA) from continued perfusion and pressurization

More information

Haemodynamics and Stresses in Abdominal Aortic Aneurysms: A Fluid-Structure Interaction Study into the Effect of Neck and. Iliac Bifurcation Angle

Haemodynamics and Stresses in Abdominal Aortic Aneurysms: A Fluid-Structure Interaction Study into the Effect of Neck and. Iliac Bifurcation Angle Haemodynamics and Stresses in Abdominal Aortic Aneurysms: A Fluid-Structure Interaction Study into the Effect of Neck and Iliac Bifurcation Angle Corey J. Drewe, 1,2 Louis Parker, 1,2 Lachlan J. Kelsey,

More information

The evolution. AORFIX AAA Stent Graft now with the new AORFLEX Delivery System

The evolution. AORFIX AAA Stent Graft now with the new AORFLEX Delivery System The evolution of AORFIX AORFIX AAA Stent Graft now with the new AORFLEX Delivery System The confident choice for the right angles AORFIX the only AAA stent graft approved for all angulations from 0 90,

More information

A study of computational modeling of abdominal aortic aneurysm

A study of computational modeling of abdominal aortic aneurysm A study of computational modeling of abdominal aortic aneurysm S.M. Kim (a), M.P Lee (b), H.C Kang (c), J.H Lee (d), J.H Jeong (e), S.G Roh (f) (a) Department of Medical Bio Engineering, College of Biosystem,

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties

Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Numerical Analysis of the Influence of Stent Parameters on the Fatigue Properties Lin Chen, Shen Jingfeng & Chen Bing School of mechanical engineering, University of Shanghai for Science & Technology,

More information

Aneurysm sac pressure monitoring: Does the direction of pressure measurement matter in fibrinous thrombus?

Aneurysm sac pressure monitoring: Does the direction of pressure measurement matter in fibrinous thrombus? Aneurysm sac pressure monitoring: Does the direction of pressure measurement matter in fibrinous thrombus? Jan-Willem Hinnen, MD, a Daniel J. Rixen, PhD, b Olivier H. Koning, MD, a Hajo J. Van Bockel,

More information

Tips and techniques for optimal stent graft placement in angulated aneurysm necks

Tips and techniques for optimal stent graft placement in angulated aneurysm necks VASCULAR AND ENDOVASCULAR TECHNIQUES Thomas L. Forbes, MD, Section Editor Tips and techniques for optimal stent graft placement in angulated aneurysm necks Jasper W. van Keulen, MD, Frans L. Moll, MD,

More information

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System,

Young-Guk Ko, M.D. Severance Cardiovascular Hospital, Yonsei University Health System, Young-Guk Ko, M.D., Dangas G, J Am Coll Cardiol Intv 2012;5:1071 All-cause Mortality Dangas Severance G, J Am Coll Cardiovascular Cardiol Intv Hospital, 2012;5:1071 Yonsei University Health System Aneurysm-related

More information

Right Choice for Right Angles

Right Choice for Right Angles Right Choice for Right Angles The Anatomy of Technology Aorfix gives you technology that conforms to patient anatomy, optimising both procedure and post-operative performance. Fishmouth for optimum neck

More information

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Flow Analysis with Stent Placement in the Cerebral Aneurysm Proceedings of the ECCOMAS Thematic International Conference on Simulation and Modeling of Biological Flows (SIMBIO 2011) September 21-23, 2011, VUB, Brussels, Belgium Flow Analysis with Stent Placement

More information

PAPER. Morphologic Changes and Outcome Following Endovascular Abdominal Aortic Aneurysm Repair as a Function of Aneurysm Size

PAPER. Morphologic Changes and Outcome Following Endovascular Abdominal Aortic Aneurysm Repair as a Function of Aneurysm Size PAPER Morphologic Changes and Outcome Following Endovascular Abdominal Aortic Aneurysm Repair as a Function of Aneurysm Size Frank R. Arko, MD; Konstantinos A. Filis, MD; Bradley B. Hill, MD; Thomas J.

More information

The diagnostic and treatment challenge of type IIIb endoleaks

The diagnostic and treatment challenge of type IIIb endoleaks The diagnostic and treatment challenge of type IIIb endoleaks Rodolfo Pini, MD, Gianluca Faggioli, MD, Chiara Mascoli, MD, Antonio Freyrie, MD, Mauro Gargiulo, MD, and Andrea Stella, MD, Bologna, Italy

More information

A New EVAR Device for Infrarenal AAAs

A New EVAR Device for Infrarenal AAAs A New EVAR Device for Infrarenal AAAs Peter Nelson, MD, MS Assistant Professor of Surgery MM0203 Rev. 01 Current U.S. EVAR Devices Anatomical Fixation Proximal Fixation Powerlink - Endologix Excluder WL

More information

Changes in aneurysm volume after endovascular repair of abdominal aortic aneurysm

Changes in aneurysm volume after endovascular repair of abdominal aortic aneurysm Changes in aneurysm volume after endovascular repair of abdominal aortic aneurysm Yehuda G. Wolf, MD, a Manfred Tillich, MD, b W. Anthony Lee, MD, a Thomas J. Fogarty, MD, a Christopher K. Zarins, MD,

More information

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC

Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC Anatomy-Driven Endograft Selection for Abdominal Aortic Aneurysm Repair S. Jay Mathews, MD, MS, FACC Interventional Cardiologist/Endovascular Specialist Bradenton Cardiology Center Bradenton, FL, USA Disclosures

More information

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Simulations of the blood flow in the arterio-venous fistula for haemodialysis Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI,

More information

Durability of The Endurant Stent-Graft through 5 Years

Durability of The Endurant Stent-Graft through 5 Years Durability of The Endurant Stent-Graft through 5 Years Michel S. Makaroun MD Co-Director, UPMC Heart and Vascular Institute Professor and Chair, Division of Vascular Surgery University of Pittsburgh School

More information

A Finite Element Analysis Rupture Index (FEARI) as an. Additional Tool for Abdominal Aortic Aneurysm Rupture. Prediction

A Finite Element Analysis Rupture Index (FEARI) as an. Additional Tool for Abdominal Aortic Aneurysm Rupture. Prediction A Finite Element Analysis Rupture Index (FEARI) as an Additional Tool for Abdominal Aortic Aneurysm Rupture Prediction Barry J. Doyle, 1 Anthony Callanan, 1 Michael T. Walsh, 1 Pierce A. Grace, 1,2 and

More information

IDEALIZED ABDOMINAL AORTIC ANEURYSM (AAA) GEOMETRY AS PREDICTOR OF HEMODYNAMICS STRESSES

IDEALIZED ABDOMINAL AORTIC ANEURYSM (AAA) GEOMETRY AS PREDICTOR OF HEMODYNAMICS STRESSES European Congress on Computational Methods in Applied Sciences and Engineering (ECCOMAS 2012) J. Eberhardsteiner et.al. (eds.) Vienna, Austria, September 10-14, 2012 IDEALIZED ABDOMINAL AORTIC ANEURYSM

More information

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Arteriovenous Graft Modeling and Hemodynamic Interpretation Open Journal of Fluid Dynamics, 2012, 2, 324-330 http://dx.doi.org/10.4236/ojfd.2012.24a040 Published Online December 2012 (http://www.scirp.org/journal/ojfd) Arteriovenous Graft Modeling and Hemodynamic

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

Abdominal Aortic Aneurysm Risk of Rupture: Patient-Specific FSI Simulations Using Anisotropic Model

Abdominal Aortic Aneurysm Risk of Rupture: Patient-Specific FSI Simulations Using Anisotropic Model Peter Rissland Yared Alemu Shmuel Einav Department of Biomedical Engineering, Stony Brook University, Stony Brook, NY 11794-8181 John Ricotta Department of Surgery, Stony Brook University Hospital, 101

More information

Endovascular aneurysm repair alters renal artery movement: a preliminary evaluation using dynamic CTA.

Endovascular aneurysm repair alters renal artery movement: a preliminary evaluation using dynamic CTA. 3/1/2007 o/06-3928 muss diss pag 77 Endovascular aneurysm repair alters renal artery movement: a preliminary evaluation using dynamic CTA. 6 Bart E. Muhs MD 1, Arno Teutelink MD 1, Matthias Prokop MD,

More information

Type IIIb endoleak and relining: a mathematical model on distraction forces

Type IIIb endoleak and relining: a mathematical model on distraction forces Type IIIb endoleak and relining: a mathematical model on distraction forces Abstract Purpose : Stent graft degradation with type IIIb endoleak represents a serious complication of endovascular aneurysm

More information

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents Augusto F. Sanches and Eva Gutheil Interdisciplinary Center for Scientific Computing, Heidelberg University Flow diverters

More information

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement Interactive CardioVascular and Thoracic Surgery 19 (2014) 16 20 doi:10.1093/icvts/ivu034 Advance Access publication 22 April 2014 ORIGINAL ARTICLE ADULTCARDIAC A computational fluid dynamics simulation

More information

Type-II Endoleaks Following Endovascular AAA Repair: Preoperative Predictors and Long-term Effects

Type-II Endoleaks Following Endovascular AAA Repair: Preoperative Predictors and Long-term Effects 503 VASCULAR FELLOWS FORUM 2001, FIRST PLACE Type-II Endoleaks Following Endovascular AAA Repair: Preoperative Predictors and Long-term Effects Frank R. Arko, MD; Geoffrey D. Rubin, MD; Bonnie L. Johnson,

More information