Blood flow in vessels with artificial or pathological geometrical changes

Size: px
Start display at page:

Download "Blood flow in vessels with artificial or pathological geometrical changes"

Transcription

1 Blood flow in vessels with artificial or pathological geometrical changes P. Tibaut 1, B. Wiesler 1, M. Mayer 2 & R. Wegenkittel 3 1 AVL LIST GmbH, Graz, Austria 2 VRVIs, Vienna, Austria 3 Tiani Medgraph AG, Vienna, Austria Abstract For more than a decade numerical investigations of blood flow inside vessels have been performed. Due to an increase in software user-friendliness and computer performance, these simulations are gaining more and more practical importance. The influence of geometrical changes of the vessels, e.g. by stenotic lesions, aneurysm, or by-pass operations, are subject of these investigations. A number of special features have to be considered to accurately calculate blood flow. Mainly the transient flow behavior due to the heartbeat and the non-newtonian flow effects, especially the dependency of the blood viscosity on the local shear rate, has to be mentioned in that respect. The blood flow in a healthy natural junction, arteria vertebralis dextra and sinistra with ateria basilaris, is compared with the flow in an artificial junction, resulting from a by-pass operation. The areas for preferred formation of stenotic lesions are identified. The shape of the vessels is significantly different in many areas from one human being to another. A detailed geometrical representation of the particular situation is required in such cases to achieve useful results, i.e. the vessel geometry of a particular person has to be known. This is valid for pathological geometrical changes as well. The entire process and the subsequent flow calculation and discussion of results is demonstrated on the basis of a bifurcation area, arteria carotis communis with arteria carotis interna and externa, with a pathological dilatation. Although numerous studies of blood flow have been performed, only a few have considered realistic anatomies under correct flow conditions.

2 462 Design and Nature II 1 Introduction Vascular diseases such as arteriosclerosis and aneurysms are becoming frequent disorders in the industrialized world due to excess sedentary and rich food. Causing more deaths then cancer, cardiovascular diseases are the leading cause of death in the western world. Aneurysms are caused by weakness of arterial wall resulting in a bulge in a shape of a small balloon. The greatest danger is that the aneurysm could break, particularly if the patient has elevated blood pressure, and produce bleeding within the brain. On histological level, aneurysms are caused by damage to intima cells in the arterial wall. Damage is believed to be caused by shear stress due to blood flow. Shear stress generates heat that breaks down these cells. Once aneurysm is formed, fluctuations in blood flow within aneurysm are of critical importance because they can induce vibration of the aneurysm wall that contributes to progression and eventual rupture. Arterosclerotic disease of carotid artery is a leading cause of stroke. Arterosclerotic plaque in the carotid artery obstructs blood flow to a brain and stimulates the formation of thromoembli that occlude downstream vessels. Unusual shear stress patterns and disturbed flows are related to plaque rupture, plaque erosion, and the formation of thromboembli. The risk of stroke from carotid artery stenosis progressively increases with increasing degree of stenosis, but the degree of stenosis does not completely explain stroke risk; moderate carotid artery stenosis is associated with stroke, trough not quite as strongly as high-grade stroke. Given the well-recognized predisposition to arterosclerotic plaque formation of specific arterial regions with curvature, such as the carotid bifurcation, hemodynamics and arterial geometry may both offer useful information for carotid artery stenosis evaluation. To date there are no reliable methods to determine wall shear stress in the recirculation zones downstream of the stenosis in clinical evaluation [1]. Thus, patient-specific information on hemodynamics generated from computational fluid dynamics (CFD) simulations may complement the evaluation of carotid artery stenosis. Knowledge of the carotid hemodynamics could clarify the relationship between carotid artery stenosis and symptoms, and, ultimately, the risk of stroke. 2 Mathematical model Blood flow can be mathematically modeled using the time-dependent Navier- Stokes equations for an incompressible fluid. Assuming the flow is laminar, the governing equations are written as: u = 0 (1) u (2) ρ + u u = p + 2 u t µ

3 Design and Nature II 463 where u is the fluid velocity, ρ the density, µ the viscosity and p the pressure. Solution of these partial differential equations requires the geometric boundaries and Dirichlet and/or Neumann conditions to be specified, and the choice of boundary conditions is very important since the predicted velocity field and shear stress can be quite sensitive to the flow conditions imposed at the boundaries. The majority of previous numerical studies of arterial flows assumed fully developed flows at the inlet and outlets of the model concerned. However, fully developed flows rarely occur in human arteries due to their relatively short lengths and their non-planar curvature and branching. Solution of the governing equations is impossible by the traditional analytical techniques for such a complicated system involving 3D irregular geometry, pulsatile flow and non-newtonian viscosity. Numerical techniques are required, hence the need for CFD. By now CFD has matured in its original engineering applications, and the application of CFD has become more and more important in cardiovascular fluid mechanics. Figure 1: Typical time course of the flow velocity during the cardiac cycle. Values on x-axis denote the time (T) normalised to a period of cardiac cycle (TP). A number of special features have to be considered to calculate blood flow accurately. Mainly the transient flow behavior due to the heartbeat and the non- Newtonian flow effects, especially the dependency of the blood viscosity on the local shear rate, has to be mentioned in that respect. For the pulsatile flow the diagram on Fig. 1 has been assumed. Assuming that µ in Eq.2 is a constant is equivalent to adopting a Newtonian model for the blood rheology. However, a better approximation to the fluid behavior of blood is given by the Cross model (ref. Lit.). The difference in axial velocity profiles between Newtonian and non-newtonian fluid is shown on Fig. 2.

4 464 Design and Nature II Figure 2: Comparison of axial velocity profiles of both: Newtonian and non- Newtonian case. Figure 3: Comparison between measured (a) and calculated (b) functional dependency based on Cross model. Dependency of blood viscosity on the local shear rate is described with equation of Cross, (Eq. (3)) where Rouleaux-effect is taken into account. For given formulation the comparison between calculated and measured functional dependency is shown on Fig. 3. The formulation of Cross describes well measured data.

5 Design and Nature II 465 η = η η η o ηo η + 1+ Aγ Ns = m Ns = m, A = 8 (3) Above expression was implemented in commercial CFD code AVL SWIFT. 3 Workflow A successful analysis incorporates a description of the process from (nonintrusive) computer tomography to the generation of surface data of the vessels in STL format, the smoothing of the surface and the generation of the computational meshes. The workflow consists of the following steps: Computer tomography Automatic generation of CAD surface data Smoothing of the surface using Surface Laplace smoothing algorithm, which does not conserve the volume of original model, but very small number of iterations (e.g. 2) does not change geometry significantly. Automatic generation of the computational meshes. Non-structured grids automate the process of fitting elements to the complex geometries of the human anatomy. One of the most important steps in any CFD simulations is to generate a good quality mesh, which represents accurately the fluid domain of interest. Here the local mesh refinement, number of boundary layer cells, proper smoothed and corrected cells play an important role. For the problem concerned here, the mesh has been generated with automatic meshing tool AVL FAME. Transient flow simulation considering pulsatile flow and physical properties of blood Result assessment, e.g. the influence of stenotic lesions, aneurysm, by-pass operations. 4 Results The procedures described above were applied to two cases in order to illustrate complete process of pre-processing, solving and post-processing the medical phenomena. A more detailed discussion of the flow results is beyond the scope of this paper.

6 466 Design and Nature II Arteria Vertebralis Dextra Arteria Vertebralis Sinistra Arteria Basilaris Natural carotid bifurcation Artificial bifurcation after by-pass i Figure 4: Natural carotid bifurcation and artificial bifurcation after by-pass operation. (a) (b) Figure 5: a) Carotid bifurcation in ascending and descending pulsatile cycle - distribution of velocity vectors, b) natural and artificial bifurcation distribution of wall shear rates during systolic acceleration.

7 Design and Nature II 467 The first case shows a healthy natural junction, arteria vertebralis dextra and sinistra with ateria basilaris compared with the flow in an artificial junction, resulted from a by-pass operation in temple area. Visualisation of the results of CFD simulation of the pulsating flow using the grids shown on Fig. 4 are presented in Fig. 5, where the distribution of velocity vectors and wall shear stresses are shown. From the results, it is evident, that the pulsatile flow doesn t change the nature of the flow behavior significantly, but the level of the flow quantities (e.g. velocity vectors, Fig. 5a). Bigger is the influence of the junction angle, where local wall shear rates (Fig. 5b) are significantly higher by increasing the bifurcation angle (artificial bifurcation after by-pass operation). The second case shows arteria carotis bifurcation geometry extracted from circulus arteriosus (willis). Pathological geometry deformation of the vessels is evident. Figure 6: MR image, generated CAD surface (circulus arteriosus and arteria carotis) and computational grid geometry of the aneurysm and parent vessel. Surface velocity values are mainly within 0.25m/s. Streamlines show uniform flow in the arteria communis as well as in arteria interna and externa. The flow in aneurysm has mixing and tumbling tendency. Kinetic energy dissipates there and induces higher pressure drop.

8 468 Design and Nature II Figure 7: Surface maps of the velocity magnitude, streamlines and total pressure. General characteristic of blood flow in the aneurysm are illustrated in Fig. 8. Streamlines spread and began to mix as they entered central part of the neck. Streamlines mixed as they impacted the posterior wall of the aneurysm and swirled in the left, right and inferior direction. Figure 8: Overview of the complex aneurysnm hemodynamics - velocity vectors and streamlines distribution in the center and on the surface of aneurysm. Flow in the upper part of the bifurcation is direct, whereas the lower part represents area of recirculation and stagnation. 5 Conclusion Realistic modeling can yield new insights into the arterial blood flow. Furthermore, for the purposes of diagnosis and surgical planning, realistic patient specific modeling is needed because of the large anatomic and physiologic variability among individuals. These examples demonstrate how CFD tools can

9 Design and Nature II 469 be used to assess the distributions of blood velocity and biochemical forces imposed on the arterial wall by the blood. This may provide better understanding of the relationship between hemodynamics and vascular disease, and may one day lead to a non-invasive tool for early perdition of carotid disease or surgical planning. References [1] Cebral J.R., Yim P.J., Loehner R., Soto O., Choyke P.L., Blood Flow Modeling in Carotid Arteries with Computational Fluid Dynamics and MR Imaging, Academic Radiology, Vol 9, No 11, pp [2] Cebral J.R., Loehner R., Burgess J.E., Computer Simulation of Cerebral Artery Clipping: Relevance to Aneurysm Neuro-Surgery Planning, European Congress on Computational Methods in Applied Sciences and Engineering, ECCOMAS 2000, Barcelona. [3] Xu X.Y, Zhao S.Z., Ariff B., Long Q., Hughes A.D., Thom S.A., CFD and FE Stress Analysis of the Carotid Bifurcation Based on Individual Anatomical and Flow Data Acquired Using MRI and Ultrasound, BED- Vol. 50, 2001 Bioengineering Conference ASME 2001, pp [4] SWIFT Manuals, AVL LIST GmbH, Graz.

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD Roopa.V.Chanashetty 1, Dr.Channappa Bhyri 2 and Vijaykumar Chanashetty 3 1 Department of Electronics and Communication Engineering,

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Arteriovenous Graft Modeling and Hemodynamic Interpretation Open Journal of Fluid Dynamics, 2012, 2, 324-330 http://dx.doi.org/10.4236/ojfd.2012.24a040 Published Online December 2012 (http://www.scirp.org/journal/ojfd) Arteriovenous Graft Modeling and Hemodynamic

More information

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Simulations of the blood flow in the arterio-venous fistula for haemodialysis Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI,

More information

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 75-83 Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Taedong Kim, Taewon Seo* 1,2 and Abdul.I.

More information

Numerical analysis of blood flow in human abdominal aorta

Numerical analysis of blood flow in human abdominal aorta Advances in Fluid Mechanics VI 603 Numerical analysis of blood flow in human abdominal aorta M. Podyma 1, I. Zbicinski 1, J. Walecki 2, M. L. Nowicki 2, P. Andziak 3, P. Makowski 4 & L. Stefanczyk 5 1

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION Jayme Pinto Ortiz University of São Paulo - Avenida Prof. Luciano Gualberto, travessa3 nº 380 - CEP - 05508-900

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

FOR many decades, cardiovascular disease has been one of

FOR many decades, cardiovascular disease has been one of Vol:1, No:2, 27 Effect of Non-Newtonian Behaviour of Blood on Pulsatile Flows in Stenotic Arteries Somkid Amornsamankul, Benchawan Wiwatanapataphee, Yong Hong Wu, Yongwimon Lenbury International Science

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Flow Analysis with Stent Placement in the Cerebral Aneurysm Proceedings of the ECCOMAS Thematic International Conference on Simulation and Modeling of Biological Flows (SIMBIO 2011) September 21-23, 2011, VUB, Brussels, Belgium Flow Analysis with Stent Placement

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

Post-conditioning. P a g e 1. To my Thesis Committee,

Post-conditioning. P a g e 1. To my Thesis Committee, P a g e 1 To my Thesis Committee, This document seeks to clarify my research project. After describing what post-conditioning (PC) is, I will explain differences between my research and the recent peristaltic

More information

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery 1 Sajid S. Mulani, 2 P. I. Jagad 1,2 Department of Mechanical Engineering, SCoE, Pune 411041, India Email: 1 sajidsmulani21@gmail.com,

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement Interactive CardioVascular and Thoracic Surgery 19 (2014) 16 20 doi:10.1093/icvts/ivu034 Advance Access publication 22 April 2014 ORIGINAL ARTICLE ADULTCARDIAC A computational fluid dynamics simulation

More information

Deposited on: 21 January 2009

Deposited on: 21 January 2009 Paul, M.C. and Molla, M.M. and Roditi, G. (2009) Large-Eddy simulation of pulsatile blood flow. Medical Engineering and Physics, 31 (1). pp. 153-159. ISSN 1350-4533 http://eprints.gla.ac.uk/4891/ Deposited

More information

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Applied Mathematics Volume, Article ID 77, pages http://dx.doi.org/.//77 Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Mahsa Dabagh,, Wakako

More information

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents Augusto F. Sanches and Eva Gutheil Interdisciplinary Center for Scientific Computing, Heidelberg University Flow diverters

More information

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング Hao LIU Advanced Computer and Information Division, RIKEN 2-1, Hirosawa, Wako-shi, Saitama 351-0198 JAPAN e-mail:

More information

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION Maurizio Bordone *, Eugenio Oñate *, Paula Rudenick, Bart Bijnens and Eduardo Soudah * * International Centre for Numerical Methods in Engineering (CIMNE) Campus

More information

Computational design of Intracranial Stent using 3D visualization system

Computational design of Intracranial Stent using 3D visualization system Computational design of Intracranial Stent using 3D visualization system Institute of Fluid Science, Tohoku University Makoto OHTA Graduate school of Engineering Hitomi Anzai Graduate school of Biomedical

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT

NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT Journal of Pure and Applied Mathematics: Advances and Applications Volume 9, Number 2, 2013, Pages 81-106 NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT BURASKORN

More information

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca,

More information

A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling

A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling International Mathematical Forum, Vol. 7, 2012, no. 54, 2687-2693 A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling Duangkamol Poltem Department of Mathematics, Faculty of Science

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta No Author Given No Institute Given Abstract. Abdominal Aortic Aneurysm is a potentially life-threatening disease if not treated adequately.

More information

Predicting Aneurysm Rupture: Computer Modeling of Geometry and Hemodynamics

Predicting Aneurysm Rupture: Computer Modeling of Geometry and Hemodynamics Predicting Aneurysm Rupture: Computer Modeling of Geometry and Hemodynamics Robert E. Harbaugh, MD, FACS, FAHA Director, Penn State Institute of the Neurosciences University Distinguished Professor & Chair,

More information

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2009 Conference Title 18th IMACS World Congress MODSIM09

More information

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch?

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch? Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch? J. Renner 1, D. Loyd 1, T. Länne 2, M. Karlsson 1 1 Department of Management and Engineering 2 Department of Medical and Health

More information

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Saccular Aneurysm Formation in Curved and Bifurcating Arteries AJNR Am J Neuroradiol 20:1309 1317, August 1999 Saccular Aneurysm Formation in Curved and Bifurcating Arteries George N. Foutrakis, Howard Yonas, and Robert J. Sclabassi BACKGROUND AND PURPOSE: Distinguishing

More information

Sensitivity Analysis of Geometry Changes in the Simulation of Basilar Aneurysms

Sensitivity Analysis of Geometry Changes in the Simulation of Basilar Aneurysms University of Wisconsin Milwaukee UWM Digital Commons Theses and Dissertations December 2016 Sensitivity Analysis of Geometry Changes in the Simulation of Basilar Aneurysms Paul Eserkaln University of

More information

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY Proceedings of the International Conference on Mechanical Engineering and Renewable Energy 2017 (ICMERE2017) 18 20 December, 2017, Chittagong, Bangladesh ICMERE2017-PI-325 NUMERICAL SIMULATION OF EFFECTS

More information

FSI Analysis of Diseased Coronary using Patient Specific Data

FSI Analysis of Diseased Coronary using Patient Specific Data FSI Analysis of Diseased Coronary using Patient Specific Data Mingchao Cai, * Haofei Liu, a Chun Yang, a,c Jie Zheng, b Richard Bach, b Mehmet H. Kural, d Kristen L. Billiar, d David Muccigrosso, b Dongsi

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0110_0000 May 27, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved. 1. Clinical

More information

Subject-Specific Modeling of Intracranial Aneurysms

Subject-Specific Modeling of Intracranial Aneurysms Subject-Specific Modeling of Intracranial Aneurysms Juan R. Cebral* a, Monica Henandez b, Alejandro Frangi b, Christopher Putman c, Richard Pergolesi c, James Burgess c a School of Computational Sciences,

More information

A 1D lumped-parameter/3d CFD approach for pressure drop in the Aortic Coarctation

A 1D lumped-parameter/3d CFD approach for pressure drop in the Aortic Coarctation A 1D lumped-parameter/3d CFD approach for pressure drop in the Aortic Coarctation Eduardo Soudah, Maurizio Bordone, Pooyan Davdan and Riccardo Rossi CIMNE Biomedical Engineering Department. Centre Internacional

More information

Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures

Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures DOI 10.1186/s12938-017-0326-y BioMedical Engineering OnLine REVIEW Open Access Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures

More information

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 101-108 Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Hun Jung*, Jong Wook Choi 1 and Chan Guk Park 2 Graduate School,

More information

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics 54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics Andžela Šešok 1, Donatas Lukšys 2 Vilnius Gediminas Technical University,

More information

Edinburgh Imaging Academy online distance learning courses

Edinburgh Imaging Academy online distance learning courses Course: Biomechanics Semester 1 / Autumn 10 Credits Each Course is composed of Modules & Activities. Modules: Biomechanics basics Ultrasound advanced Cardiovascular IMSc IMSc IMSc Each Module is composed

More information

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis Biomedical & Pharmacology Journal Vol. 8(1), 123-131 (2015) Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis M. JAHANGIRI 1 *, M.

More information

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta Robert H.P. McGregor 1, Dominik Szczerba 1,andGábor Székely 1 Computer Vision Laboratory, Sternwartstr. 7, 8092 Zürich, Switzerland

More information

BLOOD FLOW ANALYSIS OF STA-MCA ANASTOMOSIS USING CFD

BLOOD FLOW ANALYSIS OF STA-MCA ANASTOMOSIS USING CFD VI International Conference on Computational Bioengineering ICCB 2015 M. Cerrolaza and S.Oller (Eds) BLOOD FLOW ANALYSIS OF - ANASTOMOSIS USING CFD SHO TAKAYAMA *1, MITSUYOSHI WATANABE 2, HIROYUKI TAKAO

More information

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique International Journal of Engineering Research and Technology. ISSN 0974-3154 Volume 6, Number 3 (2013), pp. 399-406 International Research Publication House http://www.irphouse.com Study of Newtonian and

More information

Analysis of the GPATD : Geometrical Influence on Blood Clot Extraction using CFD Simulation

Analysis of the GPATD : Geometrical Influence on Blood Clot Extraction using CFD Simulation 2014 UKSim-AMSS 16th International Conference on Computer Modelling and Simulation Analysis of the GPATD : Geometrical Influence on Blood Clot Extraction using CFD Simulation Gregorio Romero, M.Luisa Martínez

More information

The anterior communicating artery (AcomA) is a recognized

The anterior communicating artery (AcomA) is a recognized ORIGINAL RESEARCH M.A. Castro C.M. Putman M.J. Sheridan J.R. Cebral Hemodynamic Patterns of Anterior Communicating Artery Aneurysms: A Possible Association with Rupture BACKGROUND AND PURPOSE: The anterior

More information

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model Background Coarctation of the aorta (CoA) accounts for 8%-11% of congenital heart defects, affecting tens of thousands

More information

EXPLORING ULTRASOUND IMAGES OF THE CAROTID ARTERIES USING NEURAL NETWORK TOOLS

EXPLORING ULTRASOUND IMAGES OF THE CAROTID ARTERIES USING NEURAL NETWORK TOOLS Proceedings of the 6th International Conference on Mechanics and Materials in Design, Editors: J.F. Silva Gomes & S.A. Meguid, P.Delgada/Azores, 26-30 July 2015 PAPER REF: 5430 EXPLORING ULTRASOUND IMAGES

More information

Senior Honors Thesis Prospectus

Senior Honors Thesis Prospectus Senior Honors Thesis Prospectus [Author s Name] [Author s Contact Info] [Thesis Director s Name] University of Houston Biomedical Engineering Background Coronary Atherosclerosis -Stents -Oscillatory Wall

More information

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation EasyChair Preprint 259 Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation Tianyi Xia, Matthew Doyle, Thomas Forbes and Cristina H. Amon EasyChair preprints

More information

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass BioImpacts, 2011, 1(2), 99-104 http://bi.tbzmed.ac.ir/ Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass Seyed Esmail Razavi *, Ramin Zanbouri, Omid Arjmandi-Tash

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis PEARANAT CHUCHARD 1, BENCHAWAN WIWATANAPATAPHEE 1, THITIKOM PUAPANSAWAT 1, THANONGCHAI SIRIAPISITH 2 1 Department

More information

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation American Journal of Applied Sciences 7 (9): 1237-1242, 2010 ISSN 1546-9239 2010 Science Publications Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation 1 Vahab Dehlaghi and

More information

CPM Specifications Document Aortic Coarctation: Exercise

CPM Specifications Document Aortic Coarctation: Exercise CPM Specifications Document Aortic Coarctation: Exercise OSMSC 0091_2000 0102_2000 0107_0000 0111_0000 May 29, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software

More information

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Sinjae Hyun, PhD, a Clement Kleinstreuer, PhD, b and Joseph P. Archie, Jr, PhD,

More information

A Two-layered Model for the Analysis of Arterial Rheology

A Two-layered Model for the Analysis of Arterial Rheology IJCSIT International Journal of Computer Science and Information Technology, Vol., o., June, pp. 37- A Two-layered Model for the Analysis of Arterial Rheology Sapna Singh Department of Mathematics, Harcourt

More information

SHREYAS S. HEGDE et al: COMPUTATIONAL FLUID DYNAMIC APPROACH TO UNDERSTAND THE EFFECT

SHREYAS S. HEGDE et al: COMPUTATIONAL FLUID DYNAMIC APPROACH TO UNDERSTAND THE EFFECT Computational Fluid Dynamic Approach to Understand the Effect of Increasing Blockage on Wall Shear Stress and Region of Rupture in Arteries Blocked by Arthesclerotic Plaque Shreyas S. Hegde 1, Anindya

More information

Computational Fluid Dynamics Analysis of Blood Flow in Human Aorta

Computational Fluid Dynamics Analysis of Blood Flow in Human Aorta Computational Fluid Dynamics Analysis of Blood Flow in Human Aorta Yogesh V. Borse 1, Prof. S.A. Giri 2 M. Tech Scholar, Dept of Mechanical Engg, Ramdeobaba College of Engineering and Management, Nagpur,

More information

Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology

Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology Poster No.: B-044 Congress: ECR 2011 Type: Scientific Paper Topic:

More information

Implications for the treatment of incidentally found IAs have

Implications for the treatment of incidentally found IAs have Published September 8, 2011 as 10.3174/ajnr.A2655 ORIGINAL RESEARCH Y. Qian H. Takao M. Umezu Y. Murayama Risk Analysis of Unruptured Aneurysms Using Computational Fluid Dynamics Technology: Preliminary

More information

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries World Applied Sciences Journal 27 (10): 1336-1344, 2013 ISSN 1818-4952 IDOSI Publications, 2013 DOI: 10.5829/idosi.wasj.2013.27.10.2908 Estimation and Comparison of T Graft Versus Conventional Graft for

More information

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt Washington University in St. Louis Washington University Open Scholarship Mechanical Engineering and Materials Science Independent Study Mechanical Engineering & Materials Science 12-23-2017 Computational

More information

Blood flow in S-shaped in-plane and out-of-plane coronary arteries

Blood flow in S-shaped in-plane and out-of-plane coronary arteries Blood flow in S-shaped in-plane and out-of-plane coronary arteries Author Johnston, Peter, Johnston, Barbara Published 2008 Journal Title The A N Z I A M Journal DOI https://doi.org/10.21914/anziamj.v49i0.330

More information

A study of computational modeling of abdominal aortic aneurysm

A study of computational modeling of abdominal aortic aneurysm A study of computational modeling of abdominal aortic aneurysm S.M. Kim (a), M.P Lee (b), H.C Kang (c), J.H Lee (d), J.H Jeong (e), S.G Roh (f) (a) Department of Medical Bio Engineering, College of Biosystem,

More information

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/323417296 Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid

More information

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Korea-Australia Rheology Journal Vol. 21, No. 2, June 2009 pp. 119-126 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Kun Hyuk Sung, Kyoung

More information

Flow in re-stenosed artery after angioplasty

Flow in re-stenosed artery after angioplasty Data Management and Security 209 Flow in re-stenosed artery after angioplasty S. I. Bernad 1, A. Totorean 2, E. S. Bernad 3 & R. Susan-Resiga 2 1 Romanian Academy, Timisoara Branch, Romania 2 Politehnica

More information

Index. average stress 146. see ACIS

Index. average stress 146. see ACIS Index ACIS (autonomous catheter insertion system) 156, 237 39, 241 49 acute stroke treatment 59, 69, 71 anatomical model 88 aneurismal clipping treatment 106, 110 aneurysm 2 3, 26, 47 50, 52 55, 67 68,

More information

Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms

Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms Posted: December 19, 2017 Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms Kirsten B. Giesbrecht *1 Advisor: Dr. Ellen R. Swanson 1 1 Centre College, Department of Mathematics;

More information

There has been extensive speculation in the literature about

There has been extensive speculation in the literature about TECHNICAL NOTE A. Mantha C. Karmonik G. Benndorf C. Strother R. Metcalfe Hemodynamics in a Cerebral Artery before and after the Formation of an Aneurysm SUMMARY: Using data obtained from 3D digital subtraction

More information

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD International Journal of Mechanical Engineering and Technology (IJMET) Volume 9, Issue 11, November 2018, pp. 684 688, Article ID: IJMET_09_11_068 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=9&itype=11

More information

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery Xi Gu, Guan Heng Yeoh, Victoria Timchenko Abstract In the current work, a three-dimensional geometry

More information

CVS Hemodynamics. Change in blood pressure:

CVS Hemodynamics. Change in blood pressure: CVS Hemodynamics -The distribution of blood inside the circulation: The major part of blood volume is found in the venous system 60% (2/3), that s why veins are called the capacitance vessels. -Arteries

More information

Extraction and FSI modeling of Left coronary artery structure from patient s CTA images

Extraction and FSI modeling of Left coronary artery structure from patient s CTA images Extraction and FSI modeling of Left coronary artery structure from patient s CTA images Safia Salim Dept. of Electronics & Communication College of Engineering, Kerala University Trivandrum. 695016 Email:

More information

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Andrzej Polańczyk, MSc Ireneusz Zbiciński, PhD, DSc Abstract The aim of this study was to estimate whether

More information

Controversy exists regarding the mechanisms responsible

Controversy exists regarding the mechanisms responsible ORIGINAL RESEARCH J.R. Cebral M. Sheridan C.M. Putman Hemodynamics and Bleb Formation in Intracranial Aneurysms BACKGROUND AND PURPOSE: Intracranial aneurysms with irregular shapes and blebs or secondary

More information

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow?

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow? CASE 8 A 65-year-old man with a history of hypertension and coronary artery disease presents to the emergency center with complaints of left-sided facial numbness and weakness. His blood pressure is normal,

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0003_0000 0006_0000 May 24, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved.

More information

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft Korea-Australia Rheology Journal, Vol.26, No.3, pp.327-334 (August 2014) DOI: 10.1007/s13367-014-0037-x www.springer.com/13367 A numerical study on the effect of hematocrit on hemodynamic characteristics

More information

Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques

Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques BEE 4530 Spring 2013 Group 8 Alexandra Braun Stellie Justin Ford Marina Shumakovich Alden Sonnenfeldt

More information

Closed-loop CFD Model of the Self-Powered Fontan Circulation for the Hypoplastic Left Heart Syndrome

Closed-loop CFD Model of the Self-Powered Fontan Circulation for the Hypoplastic Left Heart Syndrome McNair Scholars Research Journal Volume 3 Article 4 016 Closed-loop CFD Model of the Self-Powered Fontan Circulation for the Hypoplastic Left Heart Syndrome Nathalie E. Quintero Embry-Riddle Aeronautical

More information

A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt Flow with an Additional Pulsatile Flow Through a modified Blalock-Taussig Shunt

A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt Flow with an Additional Pulsatile Flow Through a modified Blalock-Taussig Shunt University of New Orleans ScholarWorks@UNO University of New Orleans Theses and Dissertations Dissertations and Theses Spring 5-19-2017 A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt

More information

Rheology update in Cardiovascular disease

Rheology update in Cardiovascular disease Rheology update in Cardiovascular disease Byoung Kwon Lee, MD,PhD Deptment of Medicine/Cardiology, Sanggye Paik Hospital Inje University School of Medicine, Seoul, Korea Atherosclerosis is a result product

More information

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Titus PETRILA 1, Balazs ALBERT,2 Corresponding author 1 Vasile Goldis Western University,

More information

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging Original Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging P. Khamdaengyodtai 1, T. Khamdaeng 1, P. Sakulchangsatjatai 1, N. Kammuang-lue

More information

David Rutkowski Department of Mechanical Engineering Valparaiso University 1900 Chapel Drive Valparaiso, IN 46383

David Rutkowski Department of Mechanical Engineering Valparaiso University 1900 Chapel Drive Valparaiso, IN 46383 Proceedings of The National Conference On Undergraduate Research (NCUR) 2014 University of Kentucky Lexington, KY April 3-5, 2014 Computational Fluid Dynamics of the Pediatric Trachea David Rutkowski Department

More information

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction The Study o Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction Taewon Seo*, Guek-don Shin* and Do-Il Kim** *Department o Mechanical and Automotive Engineering, Andong National University,

More information

VFI Technology to Change the Way You Work

VFI Technology to Change the Way You Work Analogic Ultrasound VFI Technology to Change the Way You Work Vascular Ultrasound Made Easier Vector Flow Imaging VFI VFI is a ground-breaking technology that can revolutionize the workflow for many Doppler

More information

Human Heart Blood Flow Simulations Based on CFD

Human Heart Blood Flow Simulations Based on CFD Human Heart Blood Flow Simulations Based on CFD Dean Vucinic 1, Tao Yang 1,2, Andrey A. Aksenov 3 1 Vrije Universiteit Brussel, Belgium 12 Xi an Modern Control Technology Institute, China 3 Capvidia/TESIS,

More information

Vascular Hemodynamics CFD Modeling

Vascular Hemodynamics CFD Modeling University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange Masters Theses Graduate School 12-2007 Vascular Hemodynamics CFD Modeling Shashank Kiran Karra University of Tennessee

More information

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2008-09-01 Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational

More information