Computational design of Intracranial Stent using 3D visualization system

Size: px
Start display at page:

Download "Computational design of Intracranial Stent using 3D visualization system"

Transcription

1 Computational design of Intracranial Stent using 3D visualization system Institute of Fluid Science, Tohoku University Makoto OHTA Graduate school of Engineering Hitomi Anzai Graduate school of Biomedical Engineering Toshio Nakayama 1

2 Background Causes of death Cause-specific death rate in Japan (H19) (Population Survey Report ) other Malignant neoplasm Pneumonia 35.8% 11.7% Stent placement Endovascular treatment Sole stenting Minimal invasiveness Good prognosis of pationts Reduction of flow in aneurysm aneurysm M. Aenis et al. (1997) Cerebral vascular disease Heart disease 13.5% 18.7% Cardiovascular disease 32.3% stent strut 2

3 Integration of realistic stent data to realistic patient data Our team firstly succeeded to develop this method in the world And provided this techniques to VISC (2006). 3

4 Method Aneurysm and stent geometry was reconstructedcted merging geometries using Digital Subtraction Angiography and Micro CT using 3D CAD technique Inlet Numerical simulation Inlet Finite Volume Method Velocity : [m/s] Incompressible-Newtonian Density : 1050 [kg/m 3 ] fluid Viscosity : [Pa s] Equation of continuity Reynolds number : about 200 Navier-Stokes equations (assuming Basilar Artery) Steady flow Outlet Fluent 6.3 (Fluent, Inc.) Pressure : 0 [Pa] Outlet Boundary condition on the wall Non-Slip Mesh The number : about 1,200,000 (in all cases) Type : Tetrahedron Using Size Function 6

5 Figure: Stent image 5

6 Silk stent Still rupture... Risk of embolization Still underfinished.. Less materials is better

7 We need to reduce the material volume Look for the most effective (reduce-able) point 7

8 Realization Workspace (CAVE) in IFS 3D realization may help a lot for understanding the flow 8

9 Specification of inflow zone 9

10 Background and Purpose Purpose To verify the validity of 3D visualization to understand the flow on stent as a new way of stent development Flowchart Numerical Simulation 3D Visualization Good stent t? Good knowledge about effective stenting? Discussions Improvement of stent

11 Specification of inflow zone cc Numerical simulation (before strut addition) & 3D visualization Inflow from Inlet Inflow zone Addition of strut to cover inflow zone Outflow to Outlet Flow Inflow zone Add a strut at this position 10

12 Addition of strut Strut diameter : 0.15 [mm] Curvature : 667 [1/m] Change strut position x=0.50[mm] x=1.30[mm] x=2.00[mm] 12

13 The role of stent strut on the flow

14 Addition of strut Flow Velocity WSS x=0.50[mm] x=1.00[mm] x=1.30[mm] x=1.65 [mm] x=2.00[mm] 15

15 The most effective stenting position? Flow is very complex around aneurysm! We cannot understand what factor can reduce the flow without visual information! Purpose To verify the validity of 3D visualization to understand the flow on stent as a new way of stent development 15

16 Method Models We used 3types of real arterial geometry Flowchart Numerical Simulation Conditions Inlet uniform velocity profile steady flow rate: [m 3 /s] 3D visualization Outlet Pressure: 0 [Pa] Viscosity: [Pa s] Density: [kg/m 3 ] 3D Visualization Discussions We visualized numerical results in Realization Workspace (AFI-IFS) Software : EnSight Gold 8.2 (CEI) Improvement of stent 16

17 Results1 Specification of inflow zone 1 inlet & 1 outlet 2 inlets & 1 outlet 1 inlet & 3 outlets Inflow zone Inflow zone Inflow zone Proceedings of ASME rd Joint US-European Fluids Engineering Summer Meeting and 8th International Conference on Nanochannels, Microchannels, and Minichannels FEDSM2010-ICNMM2010 August 1-5, 2010, Montreal, Canada FEDSM-ICNMM THE EFFECT OF 3D VISUALIZATION ON OPTIMAL DESIGN FOR STRUT POSITION OF INTRACRANIAL STENT Hitomi Anzai, Toshio Nakayama, Yuriko Takeshima, Makoto Ohta 17

18 Results2 Strut placement Strut position Strut position was decided to disturb inflow from parent artery. Strut was placed using 3D CAD techniques. 1 inlet & 1 outlet 2 inlet s& 1 outlet Without strut With one strut After placement The flow direction is changed. WSS distribution ti is changed. Strut 1 inlet & 3 outlets Strut Without strut With one strut 18

19 Parameter Study of Hemodynamics Simulation at Internal Carotid Stenosis Toshio Nakayama Hitoshi Hayase Koji Tokunaga Makoto Ohta Institute of Fluid Science, Tohoku University Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences Neurological Surgery, Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences Institute of Fluid Science, Tohoku University

20 Nakayama, Hayase, Tokunaga, Ohta, J. Flu. Sci. Tech. 3, 544, 08 The effect of outlet pressure on the flow distribution 0Pa 280Pa 420Pa

21 Simulation Results Flow rate proportion & maximum flow velocity of ICA ECA Pressure to internal carotid artery [Pa] Max.Velocity (ICA) [m/s] Max.Velocity(ECA) [m/s] Flow rate proportion Output of ultrasound ICA: 1.124[m/s], 124[m/s] ECA: [m/s] Flow rate proportion:

22 The proper flow rate portion Flow ra ate to pro The proper of flow rate: Flow rate of simulation = Flow rate of ultrasound Intersection ti in two lines Form figure, Internal carotid artery: [Pa] Pressure [Pa] :Flow rate of ultrasound :Flow rate of simulation

23 Results The Proper Case Results Outlet Condition ICA: [Pa] ECA:0 [Pa] Velocity (Measurement Position) Internal carotid artery: [m/s] External carotid artery: [m/s]

24 Optimisation of Stents Makoto OHTA, K Srinivas Institute of Flow Science, Tohoku University Sydney University

25 Optimisation Latin Hypercube sample Variables FLUENT Evaluate Fitness 60 samples Krigging The best candidate Two-dimensional optimisation of a stent for cerebral aneurysm K. Srinivas, M. Ohta, T. Nakayama, S. Obayashi, T. Yamaguchi Journal of Medical Devices (accepted) 25

26 Objective Function Maximise Velocity Reduction in Cavity Shear Stress Reduction in Cavity Vorticity Reduction in Cavity 26

27

28 Computational Details 5mm 5mm 4mm 10mm 40mm 2D Computation with 75,000 nodes Reynolds Number 300, Inlet velocity 0.3 m/s. Standard Boundary Conditions Each calculation about 2 mins on Super Computer

29 Non Dominated Solutions (Aneurysm) 29

30 Optimal Strut-Gap Arrangements V -max -max compromise V -max -max compromise 30

31 Proceedings of ASME rd Joint US-European Fluids Engineering Summer Meeting and 8th International Conference on Nanochannels, Microchannels, and Minichannels FEDSM2010-ICNMM2010 August 2-4, 2010, Montreal, Canada FEDSM-ICNMM DEVELOPMENT OF STENT STRUT PATTERN FOR CEREBRAL ANEURYSM Toshio Nakayama, Shinkyu Jeong, Srinivas karkenahalli, Makoto 31 Ohta

32 Conclusion WSS distribution and flow pattern were changed by one strut. 3D visualization system may be useful to observe flow pattern around strut. These 3D information may be helpful to design stent strut pattern. 32

33 Thank you for your attention Acknowledgement Global COE in IFS, Tohoku University, JSPS Core to core in IFS, Tohoku University, JSPS Grant-in-aid, id JNIH Grant-in-aid, JSPS 33

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Flow Analysis with Stent Placement in the Cerebral Aneurysm Proceedings of the ECCOMAS Thematic International Conference on Simulation and Modeling of Biological Flows (SIMBIO 2011) September 21-23, 2011, VUB, Brussels, Belgium Flow Analysis with Stent Placement

More information

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 75-83 Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Taedong Kim, Taewon Seo* 1,2 and Abdul.I.

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Arteriovenous Graft Modeling and Hemodynamic Interpretation Open Journal of Fluid Dynamics, 2012, 2, 324-330 http://dx.doi.org/10.4236/ojfd.2012.24a040 Published Online December 2012 (http://www.scirp.org/journal/ojfd) Arteriovenous Graft Modeling and Hemodynamic

More information

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents Augusto F. Sanches and Eva Gutheil Interdisciplinary Center for Scientific Computing, Heidelberg University Flow diverters

More information

Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology

Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology Poster No.: B-044 Congress: ECR 2011 Type: Scientific Paper Topic:

More information

Blood flow in vessels with artificial or pathological geometrical changes

Blood flow in vessels with artificial or pathological geometrical changes Blood flow in vessels with artificial or pathological geometrical changes P. Tibaut 1, B. Wiesler 1, M. Mayer 2 & R. Wegenkittel 3 1 AVL LIST GmbH, Graz, Austria 2 VRVIs, Vienna, Austria 3 Tiani Medgraph

More information

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca,

More information

E. Schuit July 2007 BMTE 08.15

E. Schuit July 2007 BMTE 08.15 Internship Pulsatile flow in a stented and non-stented 2D cerebral aneurysm model E. Schuit July 2007 BMTE 08.15 ir. G. Mulder prof. dr. ir. F.N. v.d. Vosse Eindhoven University of Technology Department

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

There has been extensive speculation in the literature about

There has been extensive speculation in the literature about TECHNICAL NOTE A. Mantha C. Karmonik G. Benndorf C. Strother R. Metcalfe Hemodynamics in a Cerebral Artery before and after the Formation of an Aneurysm SUMMARY: Using data obtained from 3D digital subtraction

More information

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique International Journal of Engineering Research and Technology. ISSN 0974-3154 Volume 6, Number 3 (2013), pp. 399-406 International Research Publication House http://www.irphouse.com Study of Newtonian and

More information

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt Washington University in St. Louis Washington University Open Scholarship Mechanical Engineering and Materials Science Independent Study Mechanical Engineering & Materials Science 12-23-2017 Computational

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Sinjae Hyun, PhD, a Clement Kleinstreuer, PhD, b and Joseph P. Archie, Jr, PhD,

More information

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Andrzej Polańczyk, MSc Ireneusz Zbiciński, PhD, DSc Abstract The aim of this study was to estimate whether

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

Medical device design using Computational Fluid Dynamics (CFD)

Medical device design using Computational Fluid Dynamics (CFD) Medical device design using Computational Fluid Dynamics (CFD) Session: Winter 2016 IMPORTANT NOTE: This project has 8 deliverables, for each one timely work is expected. 1. General Design Specifications

More information

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD

EVALUATION OF ABDOMINAL AORTIC ANEURYSM WALL STESS BASED ON FLOW INDUCED LOAD International Journal of Mechanical Engineering and Technology (IJMET) Volume 9, Issue 11, November 2018, pp. 684 688, Article ID: IJMET_09_11_068 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=9&itype=11

More information

BLOOD FLOW ANALYSIS OF STA-MCA ANASTOMOSIS USING CFD

BLOOD FLOW ANALYSIS OF STA-MCA ANASTOMOSIS USING CFD VI International Conference on Computational Bioengineering ICCB 2015 M. Cerrolaza and S.Oller (Eds) BLOOD FLOW ANALYSIS OF - ANASTOMOSIS USING CFD SHO TAKAYAMA *1, MITSUYOSHI WATANABE 2, HIROYUKI TAKAO

More information

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Applied Mathematics Volume, Article ID 77, pages http://dx.doi.org/.//77 Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Mahsa Dabagh,, Wakako

More information

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch?

Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch? Is a Flat Inlet Profile Sufficient for WSS Estimation in the Aortic Arch? J. Renner 1, D. Loyd 1, T. Länne 2, M. Karlsson 1 1 Department of Management and Engineering 2 Department of Medical and Health

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Simulations of the blood flow in the arterio-venous fistula for haemodialysis Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI,

More information

Numerical analysis of blood flow in human abdominal aorta

Numerical analysis of blood flow in human abdominal aorta Advances in Fluid Mechanics VI 603 Numerical analysis of blood flow in human abdominal aorta M. Podyma 1, I. Zbicinski 1, J. Walecki 2, M. L. Nowicki 2, P. Andziak 3, P. Makowski 4 & L. Stefanczyk 5 1

More information

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2009 Conference Title 18th IMACS World Congress MODSIM09

More information

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation

EasyChair Preprint. Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation EasyChair Preprint 259 Computational Fluid Dynamics Simulations of Flow in the Renal Arteries after Stent Graft Implantation Tianyi Xia, Matthew Doyle, Thomas Forbes and Cristina H. Amon EasyChair preprints

More information

Senior Honors Thesis Prospectus

Senior Honors Thesis Prospectus Senior Honors Thesis Prospectus [Author s Name] [Author s Contact Info] [Thesis Director s Name] University of Houston Biomedical Engineering Background Coronary Atherosclerosis -Stents -Oscillatory Wall

More information

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics 54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics Andžela Šešok 1, Donatas Lukšys 2 Vilnius Gediminas Technical University,

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD

BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD BLOOD FLOW VISUALISATION THROUGH CAROTID BIFURCATION USING ANSYS CFD Roopa.V.Chanashetty 1, Dr.Channappa Bhyri 2 and Vijaykumar Chanashetty 3 1 Department of Electronics and Communication Engineering,

More information

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery 1 Sajid S. Mulani, 2 P. I. Jagad 1,2 Department of Mechanical Engineering, SCoE, Pune 411041, India Email: 1 sajidsmulani21@gmail.com,

More information

Index. average stress 146. see ACIS

Index. average stress 146. see ACIS Index ACIS (autonomous catheter insertion system) 156, 237 39, 241 49 acute stroke treatment 59, 69, 71 anatomical model 88 aneurismal clipping treatment 106, 110 aneurysm 2 3, 26, 47 50, 52 55, 67 68,

More information

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries World Applied Sciences Journal 27 (10): 1336-1344, 2013 ISSN 1818-4952 IDOSI Publications, 2013 DOI: 10.5829/idosi.wasj.2013.27.10.2908 Estimation and Comparison of T Graft Versus Conventional Graft for

More information

A study of computational modeling of abdominal aortic aneurysm

A study of computational modeling of abdominal aortic aneurysm A study of computational modeling of abdominal aortic aneurysm S.M. Kim (a), M.P Lee (b), H.C Kang (c), J.H Lee (d), J.H Jeong (e), S.G Roh (f) (a) Department of Medical Bio Engineering, College of Biosystem,

More information

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent

Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent Development of Computational Models for Evaluation of Mechanical and Hemodynamic Behavior of an Intravascular Stent Kuang-Huei Lee, Hao-Ming Hsiao, Ying-Chih Liao, Yi-Hsiang Chiu, Yu-Seng Tee Dept. of

More information

Computational Analysis on Commercially Available Stent Designs

Computational Analysis on Commercially Available Stent Designs Computational Analysis on Commercially Available Stent Designs Abhijit Chanda 1, Shuvrangsu Das 2, Sounak Bhattacharjee 2, Pranab Ghosh 2, K. Basu 3 1 School of Bio Science and Engineering, Jadavpur University,

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION Jayme Pinto Ortiz University of São Paulo - Avenida Prof. Luciano Gualberto, travessa3 nº 380 - CEP - 05508-900

More information

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Titus PETRILA 1, Balazs ALBERT,2 Corresponding author 1 Vasile Goldis Western University,

More information

Introducing a New Treatment Method for Brain Aneurysms

Introducing a New Treatment Method for Brain Aneurysms Pipeline Embolization Device Introducing a New Treatment Method for Brain Aneurysms UNDERSTANDING ANEURYSMS What is a brain aneurysm? An aneurysm is an outpouching in an artery caused by weakness in the

More information

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement Interactive CardioVascular and Thoracic Surgery 19 (2014) 16 20 doi:10.1093/icvts/ivu034 Advance Access publication 22 April 2014 ORIGINAL ARTICLE ADULTCARDIAC A computational fluid dynamics simulation

More information

Thermal Fluids Modeling of Localized Hypothermia in a Canine Brain

Thermal Fluids Modeling of Localized Hypothermia in a Canine Brain www.heartandstroke.com Thermal Fluids Modeling of Localized Hypothermia in a Canine Brain Graduate Student: Ryan A. Sikorski Thesis Advisor: Dr. Thomas L. Merrill Rowan University Department of Mechanical

More information

Journal of Biomechanics

Journal of Biomechanics Journal of Biomechanics ] (]]]]) ]]] ]]] Contents lists available at SciVerse ScienceDirect Journal of Biomechanics journal homepage: www.elsevier.com/locate/jbiomech www.jbiomech.com Flow resistance analysis

More information

Blood flow in S-shaped in-plane and out-of-plane coronary arteries

Blood flow in S-shaped in-plane and out-of-plane coronary arteries Blood flow in S-shaped in-plane and out-of-plane coronary arteries Author Johnston, Peter, Johnston, Barbara Published 2008 Journal Title The A N Z I A M Journal DOI https://doi.org/10.21914/anziamj.v49i0.330

More information

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2008-09-01 Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational

More information

Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures

Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures DOI 10.1186/s12938-017-0326-y BioMedical Engineering OnLine REVIEW Open Access Computational medical imaging and hemodynamics framework for functional analysis and assessment of cardiovascular structures

More information

Research Article Propose a Wall Shear Stress Divergence to Estimate the Risks of Intracranial Aneurysm Rupture

Research Article Propose a Wall Shear Stress Divergence to Estimate the Risks of Intracranial Aneurysm Rupture The Scientific World Journal, Article ID 508131, 8 pages http://dx.doi.org/10.1155/2013/508131 Research Article Propose a Wall Shear Stress Divergence to Estimate the Risks of Intracranial Aneurysm Rupture

More information

Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms

Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms Posted: December 19, 2017 Exploring Blood Flow Dynamics Underlying Pathogenesis of Splenic Artery Aneurysms Kirsten B. Giesbrecht *1 Advisor: Dr. Ellen R. Swanson 1 1 Centre College, Department of Mathematics;

More information

The anterior communicating artery (AcomA) is a recognized

The anterior communicating artery (AcomA) is a recognized ORIGINAL RESEARCH M.A. Castro C.M. Putman M.J. Sheridan J.R. Cebral Hemodynamic Patterns of Anterior Communicating Artery Aneurysms: A Possible Association with Rupture BACKGROUND AND PURPOSE: The anterior

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0003_0000 0006_0000 May 24, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved.

More information

Flow Analysis on Fusiform Aneurysm during Exercise Condition

Flow Analysis on Fusiform Aneurysm during Exercise Condition Flow Analysis on Fusiform Aneurysm during Exercise Condition Ishkrizat Taib 1, Kahar osman 2, Shahrin Hisham Amirnordin 3, Hamidon Salleh 4 1 Faculty of Mechanical and Manufacturing Engineering, Universiti

More information

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング

ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング ISSUES ON COMPUTATIONAL MODELING FOR COMPUTATION-AIDED DIAGNOSIS 臨床診断支援ツールのための計算力学モデリング Hao LIU Advanced Computer and Information Division, RIKEN 2-1, Hirosawa, Wako-shi, Saitama 351-0198 JAPAN e-mail:

More information

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass BioImpacts, 2011, 1(2), 99-104 http://bi.tbzmed.ac.ir/ Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass Seyed Esmail Razavi *, Ramin Zanbouri, Omid Arjmandi-Tash

More information

CPM Specifications Document Aortic Coarctation: Exercise

CPM Specifications Document Aortic Coarctation: Exercise CPM Specifications Document Aortic Coarctation: Exercise OSMSC 0091_2000 0102_2000 0107_0000 0111_0000 May 29, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software

More information

Evaluation of the carotid artery stenosis based on minimization of mechanical energy loss of the blood flow

Evaluation of the carotid artery stenosis based on minimization of mechanical energy loss of the blood flow Original Article Evaluation of the carotid artery stenosis based on minimization of mechanical energy loss of the blood flow Proc IMechE Part H: J Engineering in Medicine 2016, Vol. 230(11) 1051 1058 Ó

More information

Developing Pulsatile Flow in a Deployed Coronary Stent

Developing Pulsatile Flow in a Deployed Coronary Stent Divakar Rajamohan Department of Mechanical Engineering, University of Cincinnati, Cincinnati, OH 45221 Rupak K. Banerjee 1 Department of Mechanical Engineering, and Department of Biomedical Engineering,

More information

SHREYAS S. HEGDE et al: COMPUTATIONAL FLUID DYNAMIC APPROACH TO UNDERSTAND THE EFFECT

SHREYAS S. HEGDE et al: COMPUTATIONAL FLUID DYNAMIC APPROACH TO UNDERSTAND THE EFFECT Computational Fluid Dynamic Approach to Understand the Effect of Increasing Blockage on Wall Shear Stress and Region of Rupture in Arteries Blocked by Arthesclerotic Plaque Shreyas S. Hegde 1, Anindya

More information

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model

CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model CFD Challenge: Simulation of Hemodynamics in a Patient-Specific Aortic Coarctation Model Background Coarctation of the aorta (CoA) accounts for 8%-11% of congenital heart defects, affecting tens of thousands

More information

UNIVERSITY OF CINCINNATI

UNIVERSITY OF CINCINNATI UNIVERSITY OF CINCINNATI Date: 5-Nov-2010 I, Justin Niehaus, hereby submit this original work as part of the requirements for the degree of: Master of Science in Aerospace Engineering It is entitled: Use

More information

Saccular intracranial aneurysms generally occur at arterial

Saccular intracranial aneurysms generally occur at arterial TECHNICAL NOTE H. Meng Z. Wang M. Kim R.D. Ecker L.N. Hopkins Saccular Aneurysms on Straight and Curved Vessels Are Subject to Different Hemodynamics: Implications of Intravascular Stenting SUMMARY: Our

More information

Closed-loop CFD Model of the Self-Powered Fontan Circulation for the Hypoplastic Left Heart Syndrome

Closed-loop CFD Model of the Self-Powered Fontan Circulation for the Hypoplastic Left Heart Syndrome McNair Scholars Research Journal Volume 3 Article 4 016 Closed-loop CFD Model of the Self-Powered Fontan Circulation for the Hypoplastic Left Heart Syndrome Nathalie E. Quintero Embry-Riddle Aeronautical

More information

Predicting Aneurysm Rupture: Computer Modeling of Geometry and Hemodynamics

Predicting Aneurysm Rupture: Computer Modeling of Geometry and Hemodynamics Predicting Aneurysm Rupture: Computer Modeling of Geometry and Hemodynamics Robert E. Harbaugh, MD, FACS, FAHA Director, Penn State Institute of the Neurosciences University Distinguished Professor & Chair,

More information

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation

Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation American Journal of Applied Sciences 7 (9): 1237-1242, 2010 ISSN 1546-9239 2010 Science Publications Wall Shear Stress Distribution in Patient Specific Coronary Artery Bifurcation 1 Vahab Dehlaghi and

More information

Lattice Boltzmann Analysis of the Flow Reduction Mechanism

Lattice Boltzmann Analysis of the Flow Reduction Mechanism Lattice Boltzmann Analysis of the Flow Reduction Mechanism in Stented Cerebral Aneurysms for the Endovascular Treatment Miki Hirabayashi 1,Makoto Ohta 2,DanielA.Rüfenacht 2, and Bastien Chopard 1 1 University

More information

FOR many decades, cardiovascular disease has been one of

FOR many decades, cardiovascular disease has been one of Vol:1, No:2, 27 Effect of Non-Newtonian Behaviour of Blood on Pulsatile Flows in Stenotic Arteries Somkid Amornsamankul, Benchawan Wiwatanapataphee, Yong Hong Wu, Yongwimon Lenbury International Science

More information

An Evaluation of Proposed and Current Cardiovascular Stent Design Approaches to Aortoiliac Occlusive Disease

An Evaluation of Proposed and Current Cardiovascular Stent Design Approaches to Aortoiliac Occlusive Disease Washington University in St. Louis Washington University Open Scholarship Engineering and Applied Science Theses & Dissertations Engineering and Applied Science Spring 5-18-2018 An Evaluation of Proposed

More information

Advances in Engineering Mechanics and Materials. Novel Aortic Cannula with Spiral Flow Inducing Design

Advances in Engineering Mechanics and Materials. Novel Aortic Cannula with Spiral Flow Inducing Design Novel Aortic Cannula with Spiral Flow Inducing Design Nadia S. Shafii, N. Darlis, Jeswant Dillon, Kahar Osman,Ahmad Zahran Md Khudzari, and E. Supriyanto Abstract Aortic cannula is one of the medical parts

More information

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery

Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Korea-Australia Rheology Journal Vol. 21, No. 2, June 2009 pp. 119-126 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Kun Hyuk Sung, Kyoung

More information

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION

STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION STUDY OF FLOW PHENOMENA IN AORTIC DISSECTION Maurizio Bordone *, Eugenio Oñate *, Paula Rudenick, Bart Bijnens and Eduardo Soudah * * International Centre for Numerical Methods in Engineering (CIMNE) Campus

More information

3D Modeling of Plaque Progression in the Human Coronary Artery

3D Modeling of Plaque Progression in the Human Coronary Artery Proceedings 3D Modeling of Plaque Progression in the Human Coronary Artery Igor Saveljic 1,2, *, Dalibor Nikolic 1,2, Zarko Milosevic 1,2, Velibor Isailovic 1,2, Milica Nikolic 1,2, Oberdan Parodi 3 and

More information

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Saccular Aneurysm Formation in Curved and Bifurcating Arteries AJNR Am J Neuroradiol 20:1309 1317, August 1999 Saccular Aneurysm Formation in Curved and Bifurcating Arteries George N. Foutrakis, Howard Yonas, and Robert J. Sclabassi BACKGROUND AND PURPOSE: Distinguishing

More information

Endovascular treatment for pseudoocclusion of the internal carotid artery

Endovascular treatment for pseudoocclusion of the internal carotid artery Endovascular treatment for pseudoocclusion of the internal carotid artery Daqiao Guo, Xiao Tang, Weiguo Fu Institute of Vascular Surgery, Fudan University, Department of Vascular Surgery, Zhongshan Hospital

More information

Imaging and Hemodynamics in Aneurysm Evolution

Imaging and Hemodynamics in Aneurysm Evolution Disclosures Imaging and Hemodynamics in Aneurysm Evolution David Saloner, PhD There are no financial conflicts of interest to report Department of Radiology and Biomedical Imaging VA Medical Center San

More information

Article. Reference. Towards the patient-specific design of flow diverters made from helix-like wires: an optimization study. ZHANG, Mingzi, et al.

Article. Reference. Towards the patient-specific design of flow diverters made from helix-like wires: an optimization study. ZHANG, Mingzi, et al. Article Towards the patient-specific design of flow diverters made from helix-like wires: an optimization study ZHANG, Mingzi, et al. Abstract Background Flow diverter (FD) intervention is an emerging

More information

Edinburgh Imaging Academy online distance learning courses

Edinburgh Imaging Academy online distance learning courses Course: Biomechanics Semester 1 / Autumn 10 Credits Each Course is composed of Modules & Activities. Modules: Biomechanics basics Ultrasound advanced Cardiovascular IMSc IMSc IMSc Each Module is composed

More information

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta

A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta A Multiphysics Simulation of a Healthy and a Diseased Abdominal Aorta No Author Given No Institute Given Abstract. Abdominal Aortic Aneurysm is a potentially life-threatening disease if not treated adequately.

More information

FFR Fundamentals and Measurements

FFR Fundamentals and Measurements FFR Fundamentals and Measurements Ghassan S. Kassab Thomas Linnemeier Chair Professor Biomedical Engineering, Indiana University Purdue University Indianapolis Principle of FFR Q S ( P P ) / R P max d

More information

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 101-108 Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Hun Jung*, Jong Wook Choi 1 and Chan Guk Park 2 Graduate School,

More information

ON FLOW DYNAMICS IN TYPE B AORTIC DISSECTION (TBAD)

ON FLOW DYNAMICS IN TYPE B AORTIC DISSECTION (TBAD) ON FLOW DYNAMICS IN TYPE B AORTIC DISSECTION (TBAD) A. Rabin 1, R. Karmeli 1, P. Bar-Yoseph 2 1- Vascular & Endovascular surgery department Hadassah university hospital, Jerusalem, Israel 2- Mechanical

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0110_0000 May 27, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved. 1. Clinical

More information

Translating In-vitro and Computational Multi-scale Models for the Vascular Surgeon

Translating In-vitro and Computational Multi-scale Models for the Vascular Surgeon Translating In-vitro and Computational Multi-scale Models for the Vascular Surgeon Benjamin A. Eslahpazir MS, BSME 1, Kyle W. Beggs 2, Zoe M. Lauters 2 Alain J. Kassab PhD 2 William M. DeCampli MD, PhD

More information

Microrheology P38. Laminar blood flow in stenotic microchannels

Microrheology P38. Laminar blood flow in stenotic microchannels Microrheology P38 Laminar blood flow in stenotic microchannels Joana A. C. Calejo, Valdemar Garcia, Carla S. Fernandes School of Technology and Management, Polytechnic Institute of Bragança, Campus de

More information

Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms

Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms Chapter 8 Studying the Flow Dynamics Within Endografts in Abdominal Aortic Aneurysms Efstratios Georgakarakos, Antonios Xenakis, George S. Georgiadis, Konstantinos C. Kapoulas, Evagelos Nikolopoulos and

More information

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction The Study o Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction Taewon Seo*, Guek-don Shin* and Do-Il Kim** *Department o Mechanical and Automotive Engineering, Andong National University,

More information

Numerical Analysis of Blood Flow in Human Arteries and Medical Devices

Numerical Analysis of Blood Flow in Human Arteries and Medical Devices Numerical Analysis of Blood Flow in Human Arteries and Medical Devices Prof. Dr.-Ing. Mehdi Behbahani Faculty of Medical Engineering and Technomathematics FH Aachen, Campus Jülich, Institute of Bioengineering,

More information

Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques

Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques Modeling Flow Characteristics in Carotid Artery Bifurcation Afflicted with Atherosclerotic Plaques BEE 4530 Spring 2013 Group 8 Alexandra Braun Stellie Justin Ford Marina Shumakovich Alden Sonnenfeldt

More information

Vascular Modeling & Hemodynamics Research OUTLINE. CLINICAL MOTIVATION: Stroke. BACKGROUND: Diagnosis

Vascular Modeling & Hemodynamics Research OUTLINE. CLINICAL MOTIVATION: Stroke. BACKGROUND: Diagnosis Biomedical Ultrasonics Research Lab Vascular Modeling & Hemodynamics Research Tamie Poepping Biomedical Ultrasonics Research Laboratory Dept. of Physics & Astronomy The University of Western Ontario UWO:

More information

The Effects of Aortic Stenosis on the Hemodynamic Flow Properties using Computational Fluid Dynamics

The Effects of Aortic Stenosis on the Hemodynamic Flow Properties using Computational Fluid Dynamics The Effects of Aortic Stenosis on the Hemodynamic Flow Properties using Computational Fluid Dynamics Adi Azriff Basri 1, Muhammad Zuber 1, Mohamad Shukri Zakaria 1,2, Ernnie Illyani Basri 1, Ahmad Fazli

More information

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/323417296 Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid

More information

Making the difference with Live Image Guidance

Making the difference with Live Image Guidance AneurysmFlow Interventional X-ray Making the difference with Live Image Guidance Enhance insight into cerebral aneurysm flow Key benefits Visualizes blood flow patterns in the parent vessel and aneurysm

More information

Predicting Aneurysm Rupture

Predicting Aneurysm Rupture Predicting Aneurysm Rupture Robert E. Harbaugh, MD, FAANS, FACS, FAHA Director, Penn State Institute of the Neurosciences University Distinguished Professor & Chair, Department of Neurosurgery Professor,

More information

Correlation between Reynolds number and stenosis morphology in eccentric and concentric artery models

Correlation between Reynolds number and stenosis morphology in eccentric and concentric artery models 10TH INTERNATIONAL SYMPOSIUM ON PARTICLE IMAGE VELOCIMETRY PIV13 Delft, The Netherlands, July 2-4, 2013 Correlation between Reynolds number and stenosis morphology in eccentric and concentric artery models

More information

Numerical investigation of the blood flow through the middle cerebral artery

Numerical investigation of the blood flow through the middle cerebral artery BioImpacts Razavi SE., et al., BioImpacts, 18, 8(3), 195- doi: 1.15171/bi.18. http://bi.tbzmed.ac.ir/ TUOMS Publishing Group Numerical investigation of the blood flow through the middle cerebral artery

More information