Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery

Size: px
Start display at page:

Download "Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery"

Transcription

1 Korea-Australia Rheology Journal Vol. 21, No. 2, June 2009 pp Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou* School of Mechanical Engineering, Chung-Ang University, 221 HeukSuk-Dong, Dongjak-Gu, Seoul , Korea (Received March 13, 2009; final revision received May 4, 2009) Abstract A numerical analysis is performed to investigate the effect of rotation on the blood flow characteristics with four different angular velocities. The artery has a cylindrical shape with 50% stenosis rate symmetrically distributed at the middle. Blood flow is considered a non-newtonian fluid. Using the Carreau model, we apply the pulsatile velocity profile at the inlet boundary. The period of the heart beat is one second. In comparison with no-rotation case, the flow recirculation zone (FRZ) contracts and its duration is reduced in axially rotating artery. Also wall shear stress is larger after the FRZ disappears. Although the geometry of artery is axisymmetry, the spiral wave and asymmetric flow occur clearly at the small rotation rate. It is caused that the flow is influenced by the effects of the rotation and the stenosis at same time. keywords : blood flow, axially rotating velocity, stenosis, pulsatile flow, non-newtoninan 1. Introduction Arteriosclerosis is one of the most widespread diseases in human beings. It is a significant factor in the death rate because it affects hemodynamics and reduces the flow rate of blood to the heart and the brain. In particular, sudden body movement causes the blood pressure fluctuations of a person with an arterial diseases and an ischemic poverty of blood to increase more than those of a healthy person. In a serious case, when the blood flow rate decreases, the person may fall because of vertigo and experience temporary eyesight and hearing trouble. Thus, human body movements can affect the blood flow characteristics of persons with arterial diseases more than those of healthy persons. In research on human acceleration, Burton et al. (1974) studied eyesight trouble caused by a change of blood flow in extreme gravity circumstances. Hooks et al. (1972) conducted a clinical study of the side effects of body acceleration. In research on blood flow and acceleration, Misra and Sahu (1988) developed a mathematical model to study the blood flow through large arteries under the action of periodic body acceleration. Belardinelli et al. (1989) performed an experimental study of the effect of blood pressure by shock acceleration, and Mandal et al. (2007) performed a numerical study on the blood characteristics of a cylindrical blood vessel with periodic accelerations. Nakamura et al. (1988) and Luo et al. (1992) carried out an investigation *Corresponding author: cfdmec@cau.ac.kr 2009 by The Korean Society of Rheology of the blood flow characteristics of stenosed and bifurcated blood vessels. Ro et al. (2008) performed a numerical study on the blood characteristic of the carotid bifurcation artery with periodic accelerations. However, none of these researches focused on the effect of body acceleration on the characteristics of blood flow without rotation. From the medical side, studies on the effects of cervical or spinal rotation on hemodynamic in arteries have been executed, but they concentrated on the velocity distribution caused by the change in artery volume. There have been only a few studies on how the rotational movement of the human body affects blood flow characteristics. In fluid dynamics, the flow characteristics of an axially rotating pipe without stenosis have been studied. For example, Imao et al. (1992) showed the flow instability problem in axially rotating pipes at the critical ratio of the circumferential velocity to the mean axial velocity. Kikuyama et al. (1983) showed that the transition of flow state from the laminar to the turbulent in axially rotating pipe can occur at a low Reynolds number. Both of these experiments showed that rotation caused destabilization of the flow in an axially rotating pipe. However, these experiments used water, so they did not address the characteristics of blood, which has non-newtonian viscosity. Generally, the assumption of Newtonian behavior of blood is acceptable for high shear-rate flow, but it is not valid when the shear rate is low (0.1 s -1 ), as it is in small arteries or on the downstream side of stenosis (Chien et al., 1982). It has also been pointed out that, in some diseased conditions e.g. patients with severe myocardial infarction, cerebrovascular diseases Korea-Australia Rheology Journal June 2009 Vol. 21, No

2 Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou and hypertension blood exhibits remarkable non-newtonian behavior. Thus, non-newtonian viscosity must be considered when analyzing the characteristics of flow in stenosed arteries. Young (1979) showed that head loss is a nonlinear function of stenosis and that pressure losses become significant only for stenoses greater than 50 ~70%. Thus, the stenosis affects significantly the pressure distribution in the artery and puts the person with an arterial disease in jeopardy. The experiment showed the effect of inlet velocity on the flow characteristics at the downstream of stenosis in the artery (Deplano et al., 1999). Extremely, the angular velocity is about 8 revolutions a second when the figure skater performs a standing spin. Although that case is rare to the common people, they undergo rotations of body through daily exercise. So rotation of body can affect the characteristic of blood flow such as pressure drop, wall shear stress, and flow recirculation zone. However, in spite of the importance of rotation effect, there is no research on the effect of rotation on the flow characteristics considered pulsatile velocity profile in stenosed arteries. Hence, for the basic study on the rotation effect to the blood flow, we select the common carotid artery because the artery rotates axially when people do a standing spin. Therefore the purpose of this paper is a numerical analysis of the effect of rotation and an unsteady pulsatile velocity profile in a stenosed artery. 2. Numerical details 2.1. Governing equations In order to simulate the blood flow characteristics, mass and momentum conservation equations are required and the non-newtonian viscosity and pulsatile flow must be considered. In addition, the axially rotating, centrifugal force must be added to the momentum equation as a source term. ρ ( ρν) = 0 t --- ( ρν) + ( ρν ν) = t ( pδ + µ ( ν+ ( ν)t) ) ρω ( ω r) (2) To simulate a non-newtonian fluid problem, a constitution equation is required for blood rheology characteristics described by the second invariant of shear rate tensor: τ= ηγ (3) where η and γ are apparent viscosity and shear rate. Shear rate γ is represented as: γ 1 = -- γ ijγji 2 (4) i j (1) Fig. 1. The schematic view and grid generation of stenosed blood vessel. We use the Carreau viscosity model because it is more suitable for representing blood rheology characteristics (Cho, 1985): η = η + ( η 0 η )[ 1 + ( λγ ) 2 ] ( n 1) 2 (5) where η 0 is the zero shear viscosity (0.056 Pa s), η is the infinite shear viscosity ( Pa s), λ is the time constant (3.313 s) and n is the power law index (0.356) Modeling of an artery with stenosis Fig. 1 shows a schematic view and grid generation of a stenosed blood vessel. The stenosis, where is located between 1D upstream and 1D downstream from the center of stenosis, is modeled by Young s model (Young, 1968), as shown in equation (6). The diameter of the blood vessel is 8 mm, and the minimum diameter of the stenosis is half the size of the blood vessel. The stenosis rate is 50% with no eccentricity. Rz () = [ R a R[ 1 + cosπ( z z 1 ) z 0 ]] (6) where, a is the stenosis rate, z 0 is half the length of the stenosis,1d, and z 1 is the axial position from the starting point of the stenosis where is 1D upstream from the center of stenosis. The grid independent test is performed with four different number of grid cells which are 43,587, 130,720, 434,808 and 580,320 when the angular velocity is 6 rev/s. In Fig. 2(a) and (b), the averaged wall shear stress (WSS) with 434,808 grid cells follows that with 580,320 grid cells at each 3D and 5D downstream from the center of stenosis. The difference of averaged WSS is within 5% for a period of pulsatile in both cases. Thus, 434,808 hexahedral grid cells is selected for numerical analysis. The computing time for each case was about 8 hours with 8 nodes, 2.0 GHz CPU Boundary and initial condition For the numerical simulation, the 3-D time-dependent Navier-Stokes equations were solved by the ANSYS CFX V11.0 based on the finite volume method with the pressure-based coupled solver. Fixed time step, s, was used with the Second Order Backward Euler scheme for transient term. The flow is assumed to be a laminar flow, incompressible, non-newtonian, and the wall of artery is 120 Korea-Australia Rheology Journal

3 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Fig. 3. The pulsatile inlet velocity profile. rigid with no slip conditions. For the study of the effect of the pulsatile flow on the blood flow characteristics in an artery, the idealized pulsatile velocity profile of the common carotid artery is used as the inlet boundary condition (Gijsen et al., 1999). Fig. 3 shows the pulsatile velocity profile which is dimensionless by the period of pulsatile cycle, t p, which is 1 s. The timeaveraged dynamic viscosity for a period of pulsatile is Pa s. The peak of velocity is 0.20 m/s at inlet in pulsatile and the density of blood is 1100 kg/m 3. The pressure boundary condition is used in the outlet of the artery. The initial velocity through whole domain is equal to the start of systole, t/t p =0, in pulsatile. For the study on the effect of rotation on the blood flow characteristics, we use a MRF (Multiple Reference Frame) method for application to the rotating effect of blood vessel (Luo et al., 1994). Through the MRF method, the only wall is rotated. 3. Results and discussion Fig. 2. Averaged axial WSS on circumferential lines at two different locations from the stenosis in pulsatile for various number of grid cells. ((a) At 3D downstream from the stenosis, (b) At 5D downstream from the stenosis). For the validation of our numerical method, the numerical results are compared with the experiment in axially rotating pipe (Imao et al., 1992). The angular velocity of an axially rotating artery is varied as 1, 2, 4 and 6 revolutions per a second (rev/s), and the results are compared with those of the no-rotation case. We compare the axial velocity profiles of Newtonian fluid flow to that of non-newtonian fluid flow for 6 rev/s. Consequently, the Newtonian fluid flow is more unstable than another due to the magnitude of viscosity. Overall, the viscosity of non-newtonian fluid is larger than that of Newtonian fluid. For unsteady flow, the simulations are executed over at least three cycles to achieve a periodic solution. The velocity variation after two cycles is less than 1% at test points behind the stenosis and results are saved for the final cycle. In this section, the blood flow characteristics such as axial velocity profiles, pressure distribution, flow recirculation zone (FRZ) and wall shear stress (WSS) distribution are presented. The blood flow characteristics are obtained for the entire flow domain at four different instants (t / t p = 0.14, 0.16, 0.40, 0.78) in pulsatile Validations Due to the instability problem of flow in the experiment Korea-Australia Rheology Journal June 2009 Vol. 21, No

4 Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou Fig. 5. Axial velocity profiles and flow recirculation zone at different time phase in pulsatile. The position of profile is presented as the multiple of diameter distal to stenosis. ((a) No rotation model, t/t p =0.14, (b) No rotation model, t/t p =0.16, (c) No rotation model, t/t p =0.40, (d) No rotation model, t/t p =0.78, (e) 6 rev/s model, t/t p =0.40, (f) 6 rev/s model, t/t p =0.78). The numerical results follow the experimental results except the axial velocity profile at Z = 120. The profile of measured axial velocity tends to become the turbulent at Z = 120. This reason is that the flow state changes from laminar to turbulent by the rotation effect in experiment. However, the circumferential velocity profile approaches the solid-body rotation gradually as the numerical simulation predicts. Fig. 4. Dimensionless velocity profile ((a) The dimesionless circumferential velocity, (b) The dimensionless axial velocity). (Imao et al., 1992), the spiral wave appears as the ratio of circumferential velocity to the axial velocity increases and it is clear when the rotation rate is 3 in axially rotating pipe. Thus, we have been performed the numerical analysis when the rotation rate is 3. In the experiment, the Reynolds number is 500 and the ratio of circumferential velocity to the axial velocity is 3. The circumferential velocity of a pipe and the flow rate of water are given as constant. Fig. 4(a) shows the circumferential velocity which is dimensionless by the wall velocity and Fig. 4(b) shows the axial velocity which is dimensionless by the inlet velocity in axially rotating pipe Results of numerical analysis The axial velocity profiles are presented in a plane containing the axis of artery because the geometry is axisymmetric. Fig. 5(a), (b), (c) and (d) show axial velocity profiles and FRZs when the artery is not rotating at different time phases in pulsatile. The contour of zero velocity indicates the boundary of FRZ, in which the flow is either stagnant or reversed. The jet velocity profile like a piston shape appears behind the stenosis in flow acceleration phase. Due to the decrease of the area at the stenosis and reverse flow near the vessel wall, axial velocity of the center line of artery increases in order to satisfy the flow rate conservation law at 1D downstream from the stenosis. The FRZ appears at all time in pulsatile because the stenosis causes disturbance to the flow and its size varies due to the change 122 Korea-Australia Rheology Journal

5 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery Fig. 7. The Axial velocity contours ((a) Axial velocity contours when the angular velocity is 6 rev/s at four different time phase in pulsatile, (b) Axial velocity contours as variation on the angular velocity at t/tp=0.40). Fig. 6. Pressure distribution at t/tp=0.40 ((a) Pressure distribution along the axial line at t/tp=0.40, (b) Pressure contour at yzplane as x=0). of magnitude of velocity in pulsatile. The FRZ grows and expands to the downstream of stenosis as flow accelerates and extends up to the whole downstream at t/tp =0.4. After that time, FRZ contracts rapidly and its size becomes constant because the gradient of velocity is zero in pulsatile. Due to the centrifugal force by rotating of vessel, the pressure along axial line decreases but the pressure near wall increases at the same time as shown in Fig. 6(a). Consequently, the iso-pressure lines become more convex as the angular velocity increases like Fig. 6(b). This causes the friction increases and this phenomenon suppresses the heart due to the increase of pressure drop at the rotating artery. Fig. 5(e) and (f) show axial velocity profiles and FRZs when the angular velocity is 6 rev/s. In comparison with the no-rotation case, the FRZs are reduced remarkably due to the increase of radial velocity from the centrifugal force at t/tp = 0.4. Also, axial velocity profiles and FRZs are asymmetry although the geometry of artery is axisymmetry. Fig. 7(a) shows the axial velocity contours when the angular velocity is 6 rev/s at four different time phase in pulsatile. Asymmetric contour with three or four protrusions appears at t/tp =0.4 and possesses the eccentricity. In Fig. 6, the protrusions constitute the spiral wave as the angular velocity is 4 rev/s. Fig. 7(b) shows axial velocity Korea-Australia Rheology Journal contours as variation on the angular velocity at t/tp =0.40. As the fluid flows to the downsteram, the four protrusions of contour become dim at 6 rev/s. On the other hand, the shape of four protrusions is maintained at 4 rev/s. The asymmetric coutours appear after 3D downstream from the stenosis at the whole rotating cases and after that location, 3D, the shape of protrusions rotates slowly. Imao et al. (1992) investigated the structure and characteristic of a spiral wave with the flow visualization technique in axially rotating pipe. In experimental study, the spiral wave occurs as the variation of rotation rate which is the ratio of circumferential velocity to the axial velocity. As the rotation rate is 3, the spiral wave is most amplified and is reduced with a greater rotation rate. Fig. 8 shows the Iso-contour of the axial velocity at four different instants in pulsatile. As the angular velocity is 4 rev/s, the spiral wave occurs at t/tp =0.4. The spiral wave appears and disappears as the variation of rotation rate due to the pulsatile inlet velocity. The stenosis affects the stability of flow in the artery. Tang et al. (1999) showed the asymmetric flow patterns occur and become unstable in an axisymmetric geometry with the stenosis. Buchanan et al. (1998) showed in their numerical study that two co-rotating vortices occurred in a 75% (area reduction) axisymmetrical stenosed model when the flow started to decelerate in pulsatile. In our numerical analysis, the spiral wave and asymmetric flow occur clearly although the rotation rate is smaller than 3. It is caused by the flow is influenced by June 2009 Vol. 21, No

6 Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou Fig. 8. ISO-contour of the axial velocity which is m/s when the angular velocity is 4 rev/s. the effects of the rotation and the stenosis at the same time. Variations of averaged axial WSS on circumferential lines are shown in Fig. 9 at four different downstream locations through the whole pulsatile. Considering for the asymmetric flow due to the rotation and stenosis of artery, WSS are averaged on circumferential line of each location. The intersections of WSS curves and the horizontal axis are two points at which WSS changes signs and hence correspond to the flow separation point (WSS changes from positive to negative) and reattachment point (WSS changes from negative to positive). At 1D downstream from the stenosis, there is no intersection point and the sign of WSS is negative all the time. Those indicate FRZ occurs through the whole cycle. After 3D downstream from the stenosis, it is notable that the difference of negative WSS caused by the reverse flow. The WSS of no-rotation case is about 40% and nearly twice smaller than that of rotating case due to the stronger reverse flow at 5D and 7D downstream from the center of stenosis, respectively. The centrifugal force caused by rotation effect suppresses the blood flow toward wall and this pheonomenon decreases the reverse Fig. 9. Averaged axial WSS on circumferential lines at four different downstream locations in pulsatile ((a) 1D downstream from the stenosis, (b) 3D downstream from the stenosis, (c) 5D downstream from the stenosis, (d) 7D downstream from the stenosis). 124 Korea-Australia Rheology Journal

7 Numerical investigation on the blood flow characteristics considering the axial rotation in stenosed artery flow and the FRZs. But, the WSS of rotation is larger after the FRZs disappear because the circumferential gradient of velocity is producted by the rotation of artery. This causes the new vascular disorder by injuring endothelium of artery such as arteroscelrosis at the downstream of stenosis (Fry, 1972). Also, when the artery is rotating, the existence time of FRZs shortens further from the stenosis. In comparison with no-rotation case, the existence time is smaller about 12.5% at 7D downstream from the center of stenosis. 4. Conclusion In this paper, the rotation effects has been studied numerically on a stenosed blood vessel with the pulsatile inlet velocity profile. The FRZs occur through the whole pulsatile because the stenosis causes disturbance to the flow and its size varies in pulsatile. In comparison with the no-rotation case, the size and existence time of FRZs are reduced remarkably due to the increase of radial velocity from the centrifugal force in axially rotating artery. But, the friction increases due to the increase of pressure by the centrifugal force. This phenomenon suppresses the heart due to the increase of pressure drop at the rotating artery. Also, the WSS of no-rotation case is about 40% and nearly twice smaller than that of rotating case due to the stronger reverse flow at 5D and 7D downstream from the stenosis, respectively. But the WSS of rotation is larger after the FRZs disappear because the circumferential gradient of velocity is producted by the rotation of artery. This causes the new vascular disorder by injuring endothelium of artery such as arteroscelrosis at the downstream of stenosis. Although the geometry of artery is axisymmetry, the spiral wave and asymmetric flow occur clearly in spite of the small rotation rate. The contour of axial velocity is asymmetric with four or three protrusions at t/t p =0.4 and possesses the eccentricity after 3D downstream from the stenosis at the whole rotating cases. And it is caused that the flow is influenced by the effects of the rotation and the stenosis at the same time. Acknowledgement This research was partially supported by the Chung-Ang University Grant in List of symbols R [m] radius of blood vessel D [m] the maximum artery diameter t [s] simulation time t p [s] a period of pulsatile z 0 [m] half the length of the entire stenosis region, 1D z 1 [m] the axial position from the starting point of the stenosis region a [-] the rate of stenosis v [-] dimensionless circumferential velocity, v /R ω r [-] dimensionless radial distance, r /R w [-] dimensionless axial velocity ρ [kg/m 3 ] density ν [m/s] blood velocity tensor η [Pa s] apparent viscosity γ [1/s] shear rate ω [rev/s] angular velocity Subscripts 0 zero shear, amplitude infinite shear Superscripts ( ) Dimension value References Belardinelli, E., M. Ursino and E. Iemmi, 1989, A preliminary theoretical study of arterial pressure perturbations under shock acceleration, J. Biomech. Eng.-Trans. ASME 111, Buchanan Jr, J. and C. Kleinstreuer, 1998, Simulation of particlehemodynamics in a partially occluded artery segment with implications to the initiation of microemboli and secondary stenoses, J. Biomech. Eng.-Trans. ASME 120, Burton, R. R., S. D. Leverett Jr and E. D. Michaelson, 1974, Man at high sustained +G(z) acceleration: a review, AEROSPACE MED. 45, Chien, S., 1982, Hemorheology in clinical medicine, Clin. Hemorheol. 2, Cho, Y. I., L. H. Back and D. W. Crawford, 1985, Experimental investigation of branch flow ratio, angle, and Reynolds number effects on the pressure and flow fields in arterial branch models, J. Biomech. Eng.-Trans. ASME 107, Deplano, V. and M. Siouffi, 1999, Experimental and numerical study of pulsatile flows through stenosis: Wall shear stress analysis, J. Biomech. 32, Fry, D., 1972, Response of the arterial wall to certain physical factors. Atherogenesis: Initiating factors, A Ciba Foundation Symp., Gijsen, F. J. H., E. Allanic, F. N. Van De Vosse and J. D. Janssen, 1999, The influence of the non-newtonian properties of blood on the flow in large arteries: Unsteady flow in a 90 curved tube, J. Biomech. 32, Hooks, L., R. Nerem and T. Benson, 1972, A momentum integral solution for pulsatile flow in a rigid tube with and without longitudinal vibration, Int. J. Eng. Sci. 10, Imao, S., M. Itoh, Y. Yamada and Q. Zhang, 1992, The characteristics of spiral waves in an axially rotating pipe, Exp. Fluids Korea-Australia Rheology Journal June 2009 Vol. 21, No

8 Kun Hyuk Sung, Kyoung Chul Ro and Hong Sun Ryou 12, Kikuyama, K., M. Murakami, K. Nishibori and K. Maeda, 1983, Flow in an Axially Rotating Pipe: A calculation of flow in the saturated region, B. JSME. 26, Luo, J. Y., R. I. Issa and A. D. Gosman, 1994, Prediction of impeller-induced flows in mixing vessels using multiple frames of reference, IChemE Symp. Ser. 136, Luo, X. and Z. Kuang, 1992, Non-Newtonian flow patterns associated with an arterial stenosis, J. Biomech. Eng.-Trans. ASME 114, Mandal, P. K., S. Chakravarty, A. Mandal and N. Amin, 2007, Effect of body acceleration on unsteady pulsatile flow of nonnewtonian fluid through a stenosed artery, Appl. Math. Comput. 189, Misra, J. C. and B. K. Sahu, 1988, Flow through blood vessels under the action of a periodic acceleration field. A mathematical analysis, J. Comput. Appl. Math. 16, Nakamura, M. and T. Sawada, 1988, Numerical study on the flow of a non-newtonian fluid through an axisymmetric stenosis, J. Biomech. Eng.-Trans. ASME 110, Ro, K. C., S. H. Lee, S. W. Cho and H. S. Ryou, 2008, Numerical Study on Blood Flow Characteristics of the Stenosed Blood Vessel with Periodic Acceleration and Rotating Effect, Springer Proceedings in Physics Series 124, Tang, D., C. Yang and D. N. Ku, 1999, A 3-D thin-wall model with fluid-structure interactions for blood flow in carotid arteries with symmetric and asymmetric stenoses, Comput. Struct. 72, Young, D. F., 1968, Effect of a time dependent stenosis on flow through a tube, J. Engng. Ind. Trans. ASME 90, Young, D. F., 1979, Fluid mechanics of arterial stenoses, J. Biomech. Eng.-Trans. ASME 101, Korea-Australia Rheology Journal

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery

Numerical Simulation of Blood Flow through Asymmetric and Symmetric Occlusion in Carotid Artery Proceedings of the 3 rd International Conference on Fluid Flow, Heat and Mass Transfer (FFHMT 16) Ottawa, Canada May 2 3, 2016 Paper No. 170 Numerical Simulation of Blood Flow through Asymmetric and Symmetric

More information

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm

Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm Non-Newtonian pulsatile blood flow in a modeled artery with a stenosis and an aneurysm I. Husain, C. Langdon and J. Schwark Department of Mathematics Luther College University of Regina Regina, Saskatchewan

More information

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries

CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries Copyright 2015 Tech Science Press MCB, vol.12, no.1, pp.37-47, 2015 CFD Analysis of Pulsatile Flow and Non-Newtonian Behavior of Blood in Arteries P. Jhunjhunwala,, P.M. Padole, and S.B. Thombre, Abstract:

More information

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1

JADAVPUR UNIVERSITY & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 Presented at the COMSOL Conference 2010 India ABHIRUP ROY CHOUDHURY 1, KRITTIKA DASGUPTA 2, ABHIJIT CHANDA 1,2, DEBABRATA NAG 1 1 DEPARTMENT OF MECHANICAL ENGINEERING & 2 SCHOOL OF BIOSCIENCE AND ENGINEERING

More information

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft

A numerical study on the effect of hematocrit on hemodynamic characteristics in arteriovenous graft Korea-Australia Rheology Journal, Vol.26, No.3, pp.327-334 (August 2014) DOI: 10.1007/s13367-014-0037-x www.springer.com/13367 A numerical study on the effect of hematocrit on hemodynamic characteristics

More information

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches

Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 75-83 Numerical simulations of fluid mechanical interactions between two abdominal aortic branches Taedong Kim, Taewon Seo* 1,2 and Abdul.I.

More information

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery

Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Korea-Australia Rheology Journal Vol. 16, No. 2, June 2004 pp. 101-108 Asymmetric flows of non-newtonian fluids in symmetric stenosed artery Hun Jung*, Jong Wook Choi 1 and Chan Guk Park 2 Graduate School,

More information

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery

Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Excerpt from the Proceedings of the COMSOL Conference 2010 Paris (COMSOL Conference) Using Computational Fluid Dynamics Model to Predict Changes in Velocity properties in Stented Carotid Artery Vaidehi

More information

FOR many decades, cardiovascular disease has been one of

FOR many decades, cardiovascular disease has been one of Vol:1, No:2, 27 Effect of Non-Newtonian Behaviour of Blood on Pulsatile Flows in Stenotic Arteries Somkid Amornsamankul, Benchawan Wiwatanapataphee, Yong Hong Wu, Yongwimon Lenbury International Science

More information

Analysis of the effects of plaque deposits on the blood flow through human artery

Analysis of the effects of plaque deposits on the blood flow through human artery ISSN 2395-1621 Analysis of the effects of plaque deposits on the blood flow through human artery #1 Sajid S. Mulani, #2 Pankaj I. Jagad 1 sajidsmulani21@gmail.com 2 pjjagad.scoe@sinhgad.edu #12 Department

More information

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION

PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION PHYSIOLOGICAL PULSATILE WAVEFORM THROUGH AXISYMMETRIC STENOSED ARTERIES: NUMERICAL SIMULATION Jayme Pinto Ortiz University of São Paulo - Avenida Prof. Luciano Gualberto, travessa3 nº 380 - CEP - 05508-900

More information

Arteriovenous Graft Modeling and Hemodynamic Interpretation

Arteriovenous Graft Modeling and Hemodynamic Interpretation Open Journal of Fluid Dynamics, 2012, 2, 324-330 http://dx.doi.org/10.4236/ojfd.2012.24a040 Published Online December 2012 (http://www.scirp.org/journal/ojfd) Arteriovenous Graft Modeling and Hemodynamic

More information

A Two-layered Model for the Analysis of Arterial Rheology

A Two-layered Model for the Analysis of Arterial Rheology IJCSIT International Journal of Computer Science and Information Technology, Vol., o., June, pp. 37- A Two-layered Model for the Analysis of Arterial Rheology Sapna Singh Department of Mathematics, Harcourt

More information

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study

Hematocrit Level on Blood flow through a Stenosed Artery with Permeable Wall: A Theoretical Study Available at http://pvamu.edu/aam Appl. Appl. Math. ISSN: 1932-9466 Vol. 12, Issue 1 (June 2017), pp. 291-304 Applications and Applied Mathematics: An International Journal (AAM) Hematocrit Level on Blood

More information

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis

Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis Biomedical & Pharmacology Journal Vol. 8(1), 123-131 (2015) Effects of Non-Newtonian Behavior of Blood on Wall Shear Stress in an Elastic Vessel with Simple and Consecutive Stenosis M. JAHANGIRI 1 *, M.

More information

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement

A computational fluid dynamics simulation study of coronary blood flow affected by graft placement Interactive CardioVascular and Thoracic Surgery 19 (2014) 16 20 doi:10.1093/icvts/ivu034 Advance Access publication 22 April 2014 ORIGINAL ARTICLE ADULTCARDIAC A computational fluid dynamics simulation

More information

Blood Flow Simulation toward Actual Application at Hospital

Blood Flow Simulation toward Actual Application at Hospital THE 5 TH ASIAN COMPUTAITIONAL FLUID DYNAMICS BUSAN, KOREA, OCTOBER 27 ~ OCTOBER 30, 2003 Blood Flow Simulation toward Actual Application at Hospital Abstract R. Himeno 1 1. Advanced Center for Computing

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with and without Bypass Graft BURASKORN NUNTADILOK 1, BENCHAWAN WIWATANAPATAPHEE 1 MEECHOKE CHUEDOUNG 1, THANONGCHAI SIRIAPISITH

More information

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis

Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid Artery with Stenosis See discussions, stats, and author profiles for this publication at: https://www.researchgate.net/publication/323417296 Simulation of Physiological Non-Newtonian Blood Flow through 3-D Geometry of Carotid

More information

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture

Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture Refinements in Mathematical Models to Predict Aneurysm Growth and Rupture RAMON BERGUER, a,b JOSEPH L. BULL, a,b AND KHALIL KHANAFER a a Vascular Mechanics Laboratory, Department of Biomedical Engineering,

More information

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY

NUMERICAL SIMULATION OF EFFECTS OF REYNOLDS NUMBER ON NON-NEWTONIAN BLOOD FLOW WITH SPIRAL COMPONENT THROUGH A REGULAR STENOSED ARTERY Proceedings of the International Conference on Mechanical Engineering and Renewable Energy 2017 (ICMERE2017) 18 20 December, 2017, Chittagong, Bangladesh ICMERE2017-PI-325 NUMERICAL SIMULATION OF EFFECTS

More information

Simulations of the blood flow in the arterio-venous fistula for haemodialysis

Simulations of the blood flow in the arterio-venous fistula for haemodialysis Acta of Bioengineering and Biomechanics Vol. 16, No. 1, 2014 Original paper DOI: 10.5277/abb140109 Simulations of the blood flow in the arterio-venous fistula for haemodialysis DANIEL JODKO*, DAMIAN OBIDOWSKI,

More information

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM

COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM COMPUTER SIMULATION OF BLOOD FLOW IN ARTERIES AFFECTED BY MULTIPLE ANEURYSM H. GIRIJA BAI 1 and K.B. NAIDU 2 Department of Mathematics, Sathyabama University, Chennai-600 119, Tamil Nadu, India 1 girijanameprakash@gmail.com

More information

Blood flow in vessels with artificial or pathological geometrical changes

Blood flow in vessels with artificial or pathological geometrical changes Blood flow in vessels with artificial or pathological geometrical changes P. Tibaut 1, B. Wiesler 1, M. Mayer 2 & R. Wegenkittel 3 1 AVL LIST GmbH, Graz, Austria 2 VRVIs, Vienna, Austria 3 Tiani Medgraph

More information

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery

A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery A Review of Study of the Effects of Plaque Deposits on the Blood Flow through Human Artery 1 Sajid S. Mulani, 2 P. I. Jagad 1,2 Department of Mechanical Engineering, SCoE, Pune 411041, India Email: 1 sajidsmulani21@gmail.com,

More information

Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel

Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel RESEARCH ARTICLE OPEN ACCESS Mathematical Modeling of Bingham Plastic Model of Blood Flow Through Stenotic Vessel S.R. Verma Department of Mathematics, D.A-V. (P.G.) College, Kanpur-208001, India Abstract

More information

Study of Cavitation Instabilities in Double- Suction Centrifugal Pump

Study of Cavitation Instabilities in Double- Suction Centrifugal Pump nternational Journal of Fluid Machinery and Systems DO: http://dx.doi.org/10.5293/jfms.2014.7.3.094 Vol. 7, No. 3, July-September 2014 SSN (Online): 1882-9554 Original Paper (nvited) Study of Cavitation

More information

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents

Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational Fluid Dynamics Analyses of Implanted Coronary Stents Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2008-09-01 Assessment of the Effects of Increasing Levels of Physiological Realism in the Computational

More information

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics

54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics 54. Simulation and research on the influence of the shape and the geometrical parameters of a blood vessel bypass graft upon hemodynamics Andžela Šešok 1, Donatas Lukšys 2 Vilnius Gediminas Technical University,

More information

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass

Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass BioImpacts, 2011, 1(2), 99-104 http://bi.tbzmed.ac.ir/ Simulation of Blood Flow Coronary Artery with Consecutive Stenosis and Coronary-Coronary Bypass Seyed Esmail Razavi *, Ramin Zanbouri, Omid Arjmandi-Tash

More information

Effects of surface geometry and non-newtonian viscosity on the flow field in arterial stenoses

Effects of surface geometry and non-newtonian viscosity on the flow field in arterial stenoses Journal of Mechanical Science and Technology 23 (2009) 2424~2433 Journal of Mechanical Science and Technology www.springerlink.com/content/1738-494x DOI 10.1007/s12206-009-0627-6 Effects of surface geometry

More information

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics

Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Computational analysis of effects of external carotid artery flow and occlusion on adverse carotid bifurcation hemodynamics Sinjae Hyun, PhD, a Clement Kleinstreuer, PhD, b and Joseph P. Archie, Jr, PhD,

More information

Flow Analysis with Stent Placement in the Cerebral Aneurysm

Flow Analysis with Stent Placement in the Cerebral Aneurysm Proceedings of the ECCOMAS Thematic International Conference on Simulation and Modeling of Biological Flows (SIMBIO 2011) September 21-23, 2011, VUB, Brussels, Belgium Flow Analysis with Stent Placement

More information

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery

Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery Three Dimensional Large Eddy Simulation of Blood Flow and Deformation in an Elastic Constricted Artery Xi Gu, Guan Heng Yeoh, Victoria Timchenko Abstract In the current work, a three-dimensional geometry

More information

Mathematical and Computational study of blood flow through diseased artery

Mathematical and Computational study of blood flow through diseased artery Mathematical and Computational study of blood flow through diseased artery Abstract This paper presents the study of blood flow through a tapered stenosed artery. The fluid (blood) medium is assumed to

More information

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases Proceedings of the 215 2nd International Symposium on Physics and Technology of Sensors, 8-1th March, 215, Pune, India Design and Simulation of Blocked Blood Vessel for Early Detection of Heart Diseases

More information

Medical device design using Computational Fluid Dynamics (CFD)

Medical device design using Computational Fluid Dynamics (CFD) Medical device design using Computational Fluid Dynamics (CFD) Session: Winter 2016 IMPORTANT NOTE: This project has 8 deliverables, for each one timely work is expected. 1. General Design Specifications

More information

Deposited on: 21 January 2009

Deposited on: 21 January 2009 Paul, M.C. and Molla, M.M. and Roditi, G. (2009) Large-Eddy simulation of pulsatile blood flow. Medical Engineering and Physics, 31 (1). pp. 153-159. ISSN 1350-4533 http://eprints.gla.ac.uk/4891/ Deposited

More information

A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling

A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling International Mathematical Forum, Vol. 7, 2012, no. 54, 2687-2693 A Review: Hemodynamics of Cerebral Aneurysm with Mathematical Modeling Duangkamol Poltem Department of Mathematics, Faculty of Science

More information

NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT

NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT Journal of Pure and Applied Mathematics: Advances and Applications Volume 9, Number 2, 2013, Pages 81-106 NUMERICAL STUDY OF PULSATILE BLOOD FLOW IN THE CORONARY SYSTEM WITH THE RCA BYPASS GRAFT BURASKORN

More information

Comparison of Stent Designs using Computational Fluid Dynamics

Comparison of Stent Designs using Computational Fluid Dynamics Dublin Institute of Technology ARROW@DIT Conference Papers School of Mechanical and Design Engineering 2007-03-28 Comparison of Stent Designs using Computational Fluid Dynamics Jonathan Murphy Dublin Institute

More information

4D model of hemodynamics in the abdominal aorta

4D model of hemodynamics in the abdominal aorta Bio-Medical Materials and Engineering 26 (2015) S257 S264 DOI 10.3233/BME-151312 IOS Press S257 4D model of hemodynamics in the abdominal aorta Ireneusz Zbicinski a,*, Natalia Veshkina a and Ludomir Stefa

More information

Non-axisymmetric Flow Development in Pulsatile Blood Flow through an Aneurysm

Non-axisymmetric Flow Development in Pulsatile Blood Flow through an Aneurysm 16 th Australasian Fluid Mechanics Conference Crown Plaza, Gold Coast, Australia 2-7 December 2007 Non-axisymmetric Flow Development in Pulsatile Blood Flow through an Aneurysm R.A. Jamison 1,2, G.J. Sheard

More information

Physiological flow analysis in significant human coronary artery stenoses

Physiological flow analysis in significant human coronary artery stenoses Biorheology 40 (2003) 451 476 451 IOS Press Physiological flow analysis in significant human coronary artery stenoses Rupak K. Banerjee a,, Lloyd H. Back b, Martin R. Back c and Young I. Cho d a Department

More information

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY

FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY LABORATORY OF BIOLOGICAL STRUCTURE MECHANICS www.labsmech.polimi.it FLUID MECHANICAL PERTURBATIONS INDUCED BY STENT IMPLANTATION: A NUMERICAL STUDY Rossella Balossino, Francesca Gervaso, Francesco Migliavacca,

More information

Flow in re-stenosed artery after angioplasty

Flow in re-stenosed artery after angioplasty Data Management and Security 209 Flow in re-stenosed artery after angioplasty S. I. Bernad 1, A. Totorean 2, E. S. Bernad 3 & R. Susan-Resiga 2 1 Romanian Academy, Timisoara Branch, Romania 2 Politehnica

More information

BLOOD FLOW THROUGH A COMPOSITE STENOSIS IN CATHETERIZED ARTERIES INTRODUCTION

BLOOD FLOW THROUGH A COMPOSITE STENOSIS IN CATHETERIZED ARTERIES INTRODUCTION 55 BLOOD FLOW THROUGH A COMPOSITE STENOSIS IN CATHETERIZED ARTERIES V P Srivastava, Rochana Vishnoi, Shailesh Mishra, Poonam Sinha * Department of Mathematics, Krishna Girls Engineering College, Kanpur-917,

More information

Post-conditioning. P a g e 1. To my Thesis Committee,

Post-conditioning. P a g e 1. To my Thesis Committee, P a g e 1 To my Thesis Committee, This document seeks to clarify my research project. After describing what post-conditioning (PC) is, I will explain differences between my research and the recent peristaltic

More information

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries

Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Simulations of pulsatile blood flow in tapered S-shaped inplane and out-of-plane coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2009 Conference Title 18th IMACS World Congress MODSIM09

More information

Non-Newtonian blood flow in human right coronary arteries: Transient simulations

Non-Newtonian blood flow in human right coronary arteries: Transient simulations Non-Newtonian blood flow in human right coronary arteries: Transient simulations Author Johnston, Barbara, Johnston, Peter, Corney, Stuart, Kilpatrick, David Published 2006 Journal Title Journal of Biomechanics

More information

Developing Pulsatile Flow in a Deployed Coronary Stent

Developing Pulsatile Flow in a Deployed Coronary Stent Divakar Rajamohan Department of Mechanical Engineering, University of Cincinnati, Cincinnati, OH 45221 Rupak K. Banerjee 1 Department of Mechanical Engineering, and Department of Biomedical Engineering,

More information

Particle Hemodynamics Analysis after Coronary Angioplasty

Particle Hemodynamics Analysis after Coronary Angioplasty , June 30 - July 2, 2010, London, U.K. Particle Hemodynamics Analysis after Coronary Angioplasty S.I. Bernad, A.F. Totorean, V.F. Vinatu, R.F. Susan-Resiga Abstract The resistance to flow through a stenosis

More information

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt

Computational Fluid Dynamics Analysis of Blalock-Taussig Shunt Washington University in St. Louis Washington University Open Scholarship Mechanical Engineering and Materials Science Independent Study Mechanical Engineering & Materials Science 12-23-2017 Computational

More information

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System

Contents 1 Computational Haemodynamics An Introduction 2 The Human Cardiovascular System Contents 1 Computational Haemodynamics An Introduction... 1 1.1 What is Computational Haemodynamics (CHD)... 1 1.2 Advantages of CHD... 3 1.3 Applications in the Cardiovascular System... 4 1.3.1 CHD as

More information

Numerical analysis of blood flow in human abdominal aorta

Numerical analysis of blood flow in human abdominal aorta Advances in Fluid Mechanics VI 603 Numerical analysis of blood flow in human abdominal aorta M. Podyma 1, I. Zbicinski 1, J. Walecki 2, M. L. Nowicki 2, P. Andziak 3, P. Makowski 4 & L. Stefanczyk 5 1

More information

Vascular Stenosis Asymmetry Influences Considerably Pressure Gradient and Flow Volume

Vascular Stenosis Asymmetry Influences Considerably Pressure Gradient and Flow Volume Physiol. Res. 65: 63-69, 2016 Vascular Stenosis Asymmetry Influences Considerably Pressure Gradient and Flow Volume L. NOVAKOVA 1, J. KOLINSKY 1, J. ADAMEC 1, J. KUDLICKA 2, J. MALIK 2 1 Department of

More information

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction

The Study of Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction The Study o Abdominal Aortic Aneurysm Rupture Risk using Fluid-Structure Interaction Taewon Seo*, Guek-don Shin* and Do-Il Kim** *Department o Mechanical and Automotive Engineering, Andong National University,

More information

Research Article Effect of Pulsatile Flow Waveform and Womersley Number on theflowinstenosedarterialgeometry

Research Article Effect of Pulsatile Flow Waveform and Womersley Number on theflowinstenosedarterialgeometry International Scholarly Research Network ISRN Biomathematics Volume 212, Article ID 85356, 17 pages doi:1.542/212/85356 Research Article Effect of Pulsatile Flow Waveform and Womersley Number on theflowinstenosedarterialgeometry

More information

Quantifying turbulent wall shear stress in a subject specific human aorta using large eddy simulation

Quantifying turbulent wall shear stress in a subject specific human aorta using large eddy simulation Quantifying turbulent wall shear stress in a subject specific human aorta using large eddy simulation Jonas Lantz, Roland Gårdhagen and Matts Karlsson Linköping University Post Print N.B.: When citing

More information

SUPPRESSION OF CAVITATION SURGE OF A HELICAL INDUCER OCCURRING IN PARTIAL FLOW CONDITIONS

SUPPRESSION OF CAVITATION SURGE OF A HELICAL INDUCER OCCURRING IN PARTIAL FLOW CONDITIONS OS-4-3 Fifth International Symposium on Cavitation (CAV23) Osaka, Japan, November -4, 23 SUPPRESSION OF CAVITATION SURGE OF A HELICAL INDUCER OCCURRING IN PARTIAL FLOW CONDITIONS Naoki ENOMOTO Graduate

More information

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images

Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Applied Mathematics Volume, Article ID 77, pages http://dx.doi.org/.//77 Research Article Hemodynamic Features in Stenosed Coronary Arteries: CFD Analysis Based on Histological Images Mahsa Dabagh,, Wakako

More information

Research Article Numerical Study of Turbulent Pulsatile Blood Flow through Stenosed Artery Using Fluid-Solid Interaction

Research Article Numerical Study of Turbulent Pulsatile Blood Flow through Stenosed Artery Using Fluid-Solid Interaction Computational and Mathematical Methods in Medicine Volume 215, Article ID 515613, 1 pages http://dx.doi.org/1.1155/215/515613 Research Article Numerical Study of Turbulent Pulsatile Blood Flow through

More information

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents

CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents CFD Study of the Blood Flow in Cerebral Aneurysms Treated with Flow Diverter Stents Augusto F. Sanches and Eva Gutheil Interdisciplinary Center for Scientific Computing, Heidelberg University Flow diverters

More information

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes

Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Calculation of the Wall Shear Stress in the case of an Internal Carotid Artery with stenoses of different sizes Titus PETRILA 1, Balazs ALBERT,2 Corresponding author 1 Vasile Goldis Western University,

More information

Rheology update in Cardiovascular disease

Rheology update in Cardiovascular disease Rheology update in Cardiovascular disease Byoung Kwon Lee, MD,PhD Deptment of Medicine/Cardiology, Sanggye Paik Hospital Inje University School of Medicine, Seoul, Korea Atherosclerosis is a result product

More information

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis

Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis Numerical Simulation of Blood Flow in the System of Human Coronary Arteries with Stenosis PEARANAT CHUCHARD 1, BENCHAWAN WIWATANAPATAPHEE 1, THITIKOM PUAPANSAWAT 1, THANONGCHAI SIRIAPISITH 2 1 Department

More information

Blood flow in S-shaped in-plane and out-of-plane coronary arteries

Blood flow in S-shaped in-plane and out-of-plane coronary arteries Blood flow in S-shaped in-plane and out-of-plane coronary arteries Author Johnston, Peter, Johnston, Barbara Published 2008 Journal Title The A N Z I A M Journal DOI https://doi.org/10.21914/anziamj.v49i0.330

More information

A Mathematical Model for Flow and Diffusion through Stenotic Capillary-Tissue Exchange System

A Mathematical Model for Flow and Diffusion through Stenotic Capillary-Tissue Exchange System 1 A Mathematical Model for Flow and Diffusion through Stenotic Capillary-Tissue Exchange System Shailesh Mishra *, S.U.Siddiqui Department of Mathematics Harcourt Butler Technological Institute Kanpur-282,

More information

Investigation of blood flow rheology using second-grade viscoelastic model (Phan-Thien Tanner) within carotid artery

Investigation of blood flow rheology using second-grade viscoelastic model (Phan-Thien Tanner) within carotid artery Acta of Bioengineering and Biomechanics Vol. 19, No. 3, 2017 Original paper DOI: 10.5277//ABB-00775-2016-05 Investigation of blood flow rheology using second-grade viscoelastic model (Phan-Thien Tanner)

More information

¼ density (kg/m 3 ) ij ¼ Reynolds stress tensor (kg/m-s 2 ) ! ¼ specific dissipation rate ("=k)

¼ density (kg/m 3 ) ij ¼ Reynolds stress tensor (kg/m-s 2 ) ! ¼ specific dissipation rate (=k) The current issue and full text archive of this journal is available at www.emeraldinsight.com/0961-5539.htm Numerical simulation of pulsatile turbulent flow in tapering stenosed arteries Bin Xiao and

More information

Developed pulsatile flow in a deployed coronary stent

Developed pulsatile flow in a deployed coronary stent Biorheology 44 (2007) 91 102 91 IOS Press Developed pulsatile flow in a deployed coronary stent Rupak K. Banerjee a,b,, Surendra B. Devarakonda a, Divakar Rajamohan a and Lloyd H. Back c a Mechanical Engineering

More information

UNIVERSITY OF CINCINNATI

UNIVERSITY OF CINCINNATI UNIVERSITY OF CINCINNATI Date: 5-Nov-2010 I, Justin Niehaus, hereby submit this original work as part of the requirements for the degree of: Master of Science in Aerospace Engineering It is entitled: Use

More information

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries

The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries The relative effects of arterial curvature and lumen diameter on wall shear stress distributions in human right coronary arteries Author Johnston, Barbara, Johnston, Peter Published 2007 Journal Title

More information

Experimental Flow Studies on Carotid Artery Models with and without Stents

Experimental Flow Studies on Carotid Artery Models with and without Stents Experimental Flow Studies on Carotid Artery Models with and without Stents Liepsch, D 1, Schmid, Th. 1, Klügel, G. 1, Sakurai, A. 3, Berger, H. 2, Greil, O. 2 1 FB 05, Fachhochschule München, Lothstr.

More information

Vascular reconstruction: CFD predictions of bypass graft haemodynamics

Vascular reconstruction: CFD predictions of bypass graft haemodynamics Vascular reconstruction: CFD predictions of bypass graft haemodynamics J.S. Cole 1, J.K. Watterson 1 & M.J.G. O Reilly 2 1 School of Mechanical and Aerospace Engineering, The Queen s University of Belfast,

More information

Computational design of Intracranial Stent using 3D visualization system

Computational design of Intracranial Stent using 3D visualization system Computational design of Intracranial Stent using 3D visualization system Institute of Fluid Science, Tohoku University Makoto OHTA Graduate school of Engineering Hitomi Anzai Graduate school of Biomedical

More information

Saccular Aneurysm Formation in Curved and Bifurcating Arteries

Saccular Aneurysm Formation in Curved and Bifurcating Arteries AJNR Am J Neuroradiol 20:1309 1317, August 1999 Saccular Aneurysm Formation in Curved and Bifurcating Arteries George N. Foutrakis, Howard Yonas, and Robert J. Sclabassi BACKGROUND AND PURPOSE: Distinguishing

More information

FFR Fundamentals and Measurements

FFR Fundamentals and Measurements FFR Fundamentals and Measurements Ghassan S. Kassab Thomas Linnemeier Chair Professor Biomedical Engineering, Indiana University Purdue University Indianapolis Principle of FFR Q S ( P P ) / R P max d

More information

ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE

ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE ANALYSIS OF FLUID-STRUCTURE INTERACTION IN ABDOMINAL AORTIC ANEURYSM WITH HIGH AND NORMAL BLOOD PRESSURE Badreddin Giuma s.k\ Kahar Osman 1 and Mohamed Rafiq Abdul Kadir 1,2 1Faculty of Mechanical Engineering,

More information

FSI Analysis of Diseased Coronary using Patient Specific Data

FSI Analysis of Diseased Coronary using Patient Specific Data FSI Analysis of Diseased Coronary using Patient Specific Data Mingchao Cai, * Haofei Liu, a Chun Yang, a,c Jie Zheng, b Richard Bach, b Mehmet H. Kural, d Kristen L. Billiar, d David Muccigrosso, b Dongsi

More information

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging

Original. Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging Original Stresses and Strains Distributions in Three-Dimension Three-Layer Abdominal Aortic Wall Based on in vivo Ultrasound Imaging P. Khamdaengyodtai 1, T. Khamdaeng 1, P. Sakulchangsatjatai 1, N. Kammuang-lue

More information

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction

Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Simulating the Motion of Heart Valves Under Fluid Flows Induced by Cardiac Contraction Eann A. Patterson Department of Mechanical Engineering, The University of Sheffield Mappin Street, Sheffield, S1 3JD

More information

The Pressure Losses in the Model of Human Lungs Michaela Chovancova, Pavel Niedoba

The Pressure Losses in the Model of Human Lungs Michaela Chovancova, Pavel Niedoba The Pressure Losses in the Model of Human Lungs Michaela Chovancova, Pavel Niedoba Abstract For the treatment of acute and chronic lung diseases it is preferred to deliver medicaments by inhalation. The

More information

IDEALIZED ABDOMINAL AORTIC ANEURYSM (AAA) GEOMETRY AS PREDICTOR OF HEMODYNAMICS STRESSES

IDEALIZED ABDOMINAL AORTIC ANEURYSM (AAA) GEOMETRY AS PREDICTOR OF HEMODYNAMICS STRESSES European Congress on Computational Methods in Applied Sciences and Engineering (ECCOMAS 2012) J. Eberhardsteiner et.al. (eds.) Vienna, Austria, September 10-14, 2012 IDEALIZED ABDOMINAL AORTIC ANEURYSM

More information

There has been extensive speculation in the literature about

There has been extensive speculation in the literature about TECHNICAL NOTE A. Mantha C. Karmonik G. Benndorf C. Strother R. Metcalfe Hemodynamics in a Cerebral Artery before and after the Formation of an Aneurysm SUMMARY: Using data obtained from 3D digital subtraction

More information

A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt Flow with an Additional Pulsatile Flow Through a modified Blalock-Taussig Shunt

A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt Flow with an Additional Pulsatile Flow Through a modified Blalock-Taussig Shunt University of New Orleans ScholarWorks@UNO University of New Orleans Theses and Dissertations Dissertations and Theses Spring 5-19-2017 A Computational Fluid Dynamics Study on Bidirectional Glenn Shunt

More information

Blood Flow At Arterial Branches: Complexities To Resolve For The Angioplasty Suite

Blood Flow At Arterial Branches: Complexities To Resolve For The Angioplasty Suite Blood Flow At Arterial Branches: Complexities To Resolve For The Angioplasty Suite P.D. Richardson 1, I.V. Pivkin 2, G.E. Karniadakis 2, D.H. Laidlaw 3 1 Division of Engineering 2 Division of Applied Mathematics

More information

CPM Specifications Document Aortofemoral Normal:

CPM Specifications Document Aortofemoral Normal: CPM Specifications Document Aortofemoral Normal: OSMSC 0110_0000 May 27, 2013 Version 1 Open Source Medical Software Corporation 2013 Open Source Medical Software Corporation. All Rights Reserved. 1. Clinical

More information

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique

Study of Newtonian and Non-Newtonian Effect of Blood Flow in Portal Vein in Normal and Hypertension Conditions using CFD Technique International Journal of Engineering Research and Technology. ISSN 0974-3154 Volume 6, Number 3 (2013), pp. 399-406 International Research Publication House http://www.irphouse.com Study of Newtonian and

More information

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries

Estimation and Comparison of T Graft Versus Conventional Graft for Coronary Arteries World Applied Sciences Journal 27 (10): 1336-1344, 2013 ISSN 1818-4952 IDOSI Publications, 2013 DOI: 10.5829/idosi.wasj.2013.27.10.2908 Estimation and Comparison of T Graft Versus Conventional Graft for

More information

Numerical analysis of the effect of turbulence transition on the hemodynamic parameters in human coronary arteries

Numerical analysis of the effect of turbulence transition on the hemodynamic parameters in human coronary arteries Original Article Numerical analysis of the effect of turbulence transition on the hemodynamic parameters in human coronary arteries Arun Mahalingam 1, Udhav Ulhas Gawandalkar 1, Girish Kini 1, Abdulrajak

More information

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm

Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Simulation of blood flow through endovascular prosthesis in patients with Abdominal Aortic Aneurysm Andrzej Polańczyk, MSc Ireneusz Zbiciński, PhD, DSc Abstract The aim of this study was to estimate whether

More information

Physiological Flow Simulation in Residual Human Stenoses After Coronary Angioplasty

Physiological Flow Simulation in Residual Human Stenoses After Coronary Angioplasty Rupak K. Banerjee Staff Scientist, Bioengineering and Physical Science Program, Bldg. 13, Rm. 3N17, National Institute of Health (NIH), Bethesda, MD 20892 Lloyd H. Back Jet Propulsion Laboratory, California

More information

The anterior communicating artery (AcomA) is a recognized

The anterior communicating artery (AcomA) is a recognized ORIGINAL RESEARCH M.A. Castro C.M. Putman M.J. Sheridan J.R. Cebral Hemodynamic Patterns of Anterior Communicating Artery Aneurysms: A Possible Association with Rupture BACKGROUND AND PURPOSE: The anterior

More information

Dynamic Role of the Cardiac Jelly

Dynamic Role of the Cardiac Jelly 61 Chapter 6 Dynamic Role of the Cardiac Jelly Before looping, when the embryonic heart is still a straight tube, the cardiac jelly occupies the bulk of the heart tube walls. Despite its preeminence in

More information

Microrheology P38. Laminar blood flow in stenotic microchannels

Microrheology P38. Laminar blood flow in stenotic microchannels Microrheology P38 Laminar blood flow in stenotic microchannels Joana A. C. Calejo, Valdemar Garcia, Carla S. Fernandes School of Technology and Management, Polytechnic Institute of Bragança, Campus de

More information

Flow Analysis on Fusiform Aneurysm during Exercise Condition

Flow Analysis on Fusiform Aneurysm during Exercise Condition Flow Analysis on Fusiform Aneurysm during Exercise Condition Ishkrizat Taib 1, Kahar osman 2, Shahrin Hisham Amirnordin 3, Hamidon Salleh 4 1 Faculty of Mechanical and Manufacturing Engineering, Universiti

More information

Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation

Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation Identification of Atherosclerotic Lesion-Prone Sites through Patient-Specific Simulation of Low-Density Lipoprotein Accumulation Ufuk Olgac 1, Vartan Kurtcuoglu 1, Stefan C. Saur 2, and Dimos Poulikakos

More information

Numerical analysis of urine flow through the side holes of a double J stent in a ureteral stenosis

Numerical analysis of urine flow through the side holes of a double J stent in a ureteral stenosis Technology and Health Care 25 (2017) S63 S72 DOI 10.3233/THC-171307 IOS Press S63 Numerical analysis of urine flow through the side holes of a double J stent in a ureteral stenosis Hyoung-Ho Kim a, Young

More information

Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration

Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration Journal of Medical and Biological Engineering, 34(5): 487-494 487 Numerical Study on Effects of Drug-coating Position of Drug-eluting Stents on Drug Concentration Yu Chen 1 Fei Yan 1 Wen-Tao Jiang 1,*

More information