Inflation and deflation timing of the AutoCAT 2 WAVE intra-aortic balloon pump using the autopilot mode in a clinical setting

Size: px
Start display at page:

Download "Inflation and deflation timing of the AutoCAT 2 WAVE intra-aortic balloon pump using the autopilot mode in a clinical setting"

Transcription

1 450060PRF / Bakker EWM et al.perfusion 2012 Inflation and deflation timing of the AutoCAT 2 WAVE intra-aortic balloon pump using the autopilot mode in a clinical setting Perfusion 0(0) 1 6 The Author(s) 2012 Reprints and permission: sagepub. co.uk/journalspermissions.nav DOI: / prf.sagepub.com EWM Bakker, 1 K Visser, 1 A van der Wal, 1 MA Kuiper, 2 M Koopmans 2 and R Breedveld 3 Abstract The primary goal of this observational clinical study was to register the occurrence of incorrect inflation and deflation timing of an intra-aortic balloon pump in autopilot mode. The secondary goal was to identify possible causes of incorrect timing. During IABP assistance of 60 patients, every four hours a strip was printed with the IABP frequency set to 1:2. Strips were examined for timing discrepancies beyond 40 ms from the dicrotic notch (inflation) and the end of the diastolic phase (deflation). In this way, 320 printed strips were examined. A total of 52 strips (16%) showed incorrect timing. On 24 of these strips, the incorrect timing was called incidental, as it showed on only one or a few beats. The other 28 cases of erroneous timing were called consistent, as more than 50% of the beats on the strip showed incorrect timing. We observed arrhythmia in 69% of all cases of incorrect timing. When timing was correct, arrhythmia was found on 13 (5%) of 268 strips. A poor quality electrocardiograph (ECG) signal showed on 37% of all strips with incorrect timing and 11% of all strips with proper timing. We conclude that inflation and deflation timing of the IABP is not always correct when using the autopilot mode. The quality of the ECG input signal and the occurrence of arrhythmia appear to be related to erroneous timing. Switching from autopilot mode to operator mode may not always prevent incorrect timing. Keywords intra-aortic balloon pump; IABP; timing; deflation; inflation; AutoCAT 2 WAVE; autopilot Introduction Since the introduction of the intra-aortic balloon pump (IABP) in cardiac care, it has been applied as an effective method to improve left ventricular performance. 1,2 Currently, it is the most frequently employed type of cardiac assist device, used to counter a period of ischemic left ventricular failure. Effective use of the IABP depends on accurate inflation and deflation timing. 3 Late inflation and early deflation lead to suboptimal assistance, where early inflation and late deflation may augment the afterload, thereby, causing an increase in cardiac oxygen consumption and impaired left ventricular performance. 4 Some new methods have been introduced in the last decade for determining the points of inflation and deflation. The fibre-optic sensor provides a reliable pressure signal. 5 From this signal the aortic flow is calculated on a beat-to-beat basis, providing the aortic valve closure point for that beat. With the use of this sensor, new algorithms have been applied to focus on accurate detection of the dicrotic notch 6,7 and better handling of arrhythmias. 8 Furthermore, trigger timing and trigger choice may change without human intervention as the autopilot accurately monitors the effects of the IABP on patient parameters. The autopilot ensures a timely change to the arterial pressure trigger 1 Departement of Perfusion, Medisch Centrum Leeuwarden, 2 Departement of Intensive Care, Medisch Centrum Leeuwarden, 3 Departement of Cardiac Care, Medisch Centrum Leeuwarden, Corresponding author: Edwin Bakker Dept. of Extracorporeal Circulation Hart Centrum, Medisch Centrum Leeuwarden H. Dunantweg 2, 8901 BR Leeuwarden edwin.bakker@znb.nl

2 2 Perfusion 0(0) when the ECG signal is lost, and back when it is regained. It may also change the algorithm used, for example, when arrhythmias are detected. With these changes the need for human intervention decreases. But this advantage may have its drawbacks. When the IABP function is entrusted to the autopilot, timing errors due to haemodynamic changes, poor quality trigger signals or incorrect position of the catheter could be missed by the nursing staff. The new algorithms have proven to perform well, 6,8,9 but the reliability of the autopilot, in all of its facets, has not been extensively examined in a clinical setting. Osentowski 10 describes the performance of the autopilot when using the Datascope CS100 IABP (Maquet, Hirrlingen, Germany). Incorrect timing could be demonstrated in 16.7% and 55.6% of inflations and deflations, respectively. We hypothesized that inflation and deflation timing of the IABP may not always be correct using the autopilot mode of the IABP in a clinical setting. Therefore, we investigated the timing of the Arrow AutoCAT 2 WAVE (Windkessel Aortic Valve Equation) IABP with 60 patients requiring IABP assistance. The input signals of ECG and blood pressure may present with arrhythmias or artefacts, possibly caused by other clinical equipment or movement of the catheter within the aorta. 14 Therefore, the second aim of this study was the identification of the causes of erroneous timing, wherever possible. Methods During a period of 29 months ( ), the inflation and deflation timing of the AutoCAT 2 WAVE IABP (Teleflex Incorporated, Hilversum, ) was tested while using the autopilot mode. The IABP timing was observed during IABP assistance of 60 patients on the cardiac care unit or the intensive care unit of the Medical Centre Leeuwarden. In 11 of these cases, observation started with a patient in the operating theatre. The IABP was supplied with software update 2.23 (SPSS Inc., Chicago, IL, USA), ensuring optimal handling of arrhythmias, breathing artefacts and trigger changes. This update also changed the timing of inflation from 20% to 15% of maximal (calculated) flow, which would decrease the occurrence of (borderline) early inflation. A 40-cc, 7.5 Fr. UltraFlex TM IAB-Catheter (Teleflex Incorporated) with fibre-optic sensor was used in 39 patients. This type of catheter was recalled in The reason for the recall was that some catheters were not recognized by the IABP console, resulting in an inflation volume of 5 cc instead of 40-cc. As a result of this recall, a 40-cc, 8 Fr. Fidelity TM IAB-catheter (Maquet, Getinge, Sweden), without a fibre-optic sensor, was used in 19 patients. A 40-cc, 8 Fr. NarrowFlex IAB- catheter (Teleflex Incorporated), without a fibre-optic sensor, was used in two patients. Catheters were inserted through the femoral artery. The IABP was used in the autopilot mode. Every four hours of assistance, the IABP frequency was temporarily set to 1:2 and a strip was printed from the IABP console, presenting the ECG and the aortic blood pressure waveform of at least five heart beats. ECG leads from the patient led to the IABP console directly. Where possible, blood pressure was measured by the fibre-optic sensor. In the operating theatre, the ECG signal was slaved from the monitor. The central lumen of the catheter was used as the blood pressure signal source when a fibre-optic sensor was not present. Factors that might affect the inflation or deflation timing were written down by the nursing staff. The strips were independently examined by two perfusionists and an intensivist for incorrect timing of the inflation or deflation, according to the following guidelines: 1) inflation should be less than 40 ms (1 mm) from the dicrotic notch, 2) deflation should be less than 40 ms from the start of the next systole and should visibly reduce end diastolic pressure. These guidelines were chosen after consulting the company (Teleflex) about the deviation justifying a switch from autopilot mode to operator mode. The number of early or late inflations and early or late deflations were noted. Incorrect timing was called consistent if it showed on >50% of all beats of a printed strip. Only one or a few incorrectly timed beats (with a maximum of 50% of all beats) on a print was called incidental. Results We included 320 from 339 printed strips. Eighteen strips were excluded, as the IABP was changed from autopilot mode to operator mode. The reasons for changing the mode were late deflations (four times), early deflations (three times) and early inflations (once). In 10 cases, the reason for conversion to operator mode was not identified. One strip showing a very poor quality blood pressure curve was excluded. Consistent incorrect timing was seen on 28 strips (9%) from 15 patients. As seen in Table 1, incorrect deflation appeared on 21 of these strips where, in seven cases, inflation was not correct. Examining possible causes for erroneous timing, we found incorrect timing to occur when arrhythmia and/ or distorted ECG signals were observed (Table 2). Incorrect positioning of the catheter on X-thorax was seen in combination with distorted ECG signals in three cases. On three strips, we found late inflations that may relate to haemodynamics (short systole). The quality of the pressure curve was poor on one strip

3 Bakker EWM et al. 3 Table 1. Incidence of consistent incorrect timing of the IABP Type of incorrect timing Early inflation 4 (1%) Late inflation 3 (1%) Incorrect inflation 7 (2%) Early deflation 11 (3%) Late deflation 20 (6%) Incorrect deflation 21 (7%) * Total 28 (9%) number of strips (n=320) *The total number of strips with incorrect deflation timing was lower than the sum of early and late deflations because both types of incorrect deflation were seen on 10 strips. IABP: intra-aortic balloon pump. with incorrect timing. Incorrect timing appeared not to be related to the type of catheter used, although damped pressure curves were not seen when using a FOS signal. On 24 strips (7%), we found incidental incorrect timing, 21 concerning deflation, three with erroneous inflation timing. During all cases of incidental incorrect timing, arrhythmia or incidental irregular beats or artefacts of the ECG signal were observed, as seen in Table 3. In one case of early inflation, we found two augmentations during systole, as shown in Figure 1. Arrhythmia was found in 36 cases of the 52 strips that showed consistent or incidental incorrect timing (69%). In 13 cases (5%) of all 268 strips with correct timing, an irregular rhythm was found (Table 4). In 19 of the 52 strips with incorrect timing (37%), ECG quality was poor. Of all strips with correct timing, 30 out of 268 strips (11%) presented with a poor quality ECG, as shown in Table 4. The quality of the ECG was determined in two ways. The AutoCAT 2 IABP has a quantitative expression of the ECG quality printed on the strip ( ECG score in Table 2 and Table 3). In 16 cases, the quality score of the chosen lead was four or higher, which means that the ECG quality is poor. In 33 cases, the ECG was significantly disturbed, in spite of a good ECG quality score (an example is seen in Figure 1). Table 2. Consistent incorrect timing and observations Strip nr Inflation Deflation Observations Cath Trigger (algorithm) ECG score 34 late Short systole NF ECG, skin (pattern) 0 35 late Short systole NF ECG, skin (pattern) 0 36 late Short systole NF ECG, skin (pattern) 0 40 late - NF ECG, skin (pattern) 0 48 late APqual NF ECG, skin (pattern) 0 63 late+early ECGqual FOS ECG, skin (pattern) 0 71 early - FOS ECG, skin (pattern) 0 73 early ECGqual FOS ECG, skin (pattern) late+early Arrhythmia FOS ECG, skin (peak) late Arrhythmia FOS ECG, skin (pattern) late+early Arrhythmia/ECGqual NF ECG, skin (pattern) late+early Arrhythmia NF ECG, skin (peak) late+early Arrhythmia/ECGqual NF ECG, skin (peak) early Arrhythmia NF AP late Arrhythmia NF ECG, skin (pattern) late+early Arrhythmia/ECGqual NF ECG, skin (peak) late POS/ECGqual NF ECG, MON (peak) late POS/ECGqual NF ECG, MON (pattern) late POS/ECGqual NF ECG, skin (pattern) late ECGqual NF ECG, skin (pattern) early FOS ECG, MON (pattern) early FOS ECG, MON (pattern) late Arrhythmia FOS ECG, skin (peak) late+early Arrhythmia FOS ECG, skin (peak) late Arrhythmia/ECGqual FOS ECG, skin (peak) late+early Arrhythmia/ECGqual FOS ECG, skin (pattern) late+early Arrhythmia FOS ECG, skin (peak) late+early Arrhythmia FOS ECG, skin (peak) 0 APqual: poor quality arterial pressure signal, ECGqual: poor quality ECG signal, POS: position of the IAB catheter too low, Cath: catheter type, FOS: catheter with fibre-optic sensor, NF: catheter without fibre-optic sensor, ECG skin: ECG signal directly from patient skin, ECG MON: ECG signal slaved from monitor, AP: arterial pressure signal.

4 4 Perfusion 0(0) Table 3. Incidental incorrect timing and observations Strip nr Inflation Deflation Observation Cath Trigger (algorithm) ECG score 143 late Arrhythmia FOS ECG, skin (pattern) late Arrhythmia FOS ECG, skin (pattern) late+early Arrhythmia FOS ECG, skin (pattern) late Arrhythmia/ECGqual NF AP early Arrhythmia/ECGqual NF ECG, skin (peak) late Arrhythmia/ECGqual NF ECG, skin (pattern) early Arrhythmia/APqual NF ECG, skin (pattern) early Arrhythmia/APqual NF ECG, skin (pattern) late Arrhythmia/ECGqual NF ECG, skin (pattern) early Arrhythmia/ECGqual NF ECG, skin (pattern) late Arrhythmia NF ECG, skin (peak) early ECG qual NF ECG, skin (pattern) late Arrhythmia NF ECG, skin (pattern) early ECG qual NF ECG, skin (pattern) late Arrhythmia FOS ECG, skin (pattern) late Arrhythmia FOS ECG, skin (pattern) late+early Arrhythmia/ECGqual NF ECG, skin (peak) late+early Arrhythmia FOS ECG, skin (peak) early Arrhythmia FOS ECG, skin (pattern) early Arrhythmia FOS ECG, skin (pattern) late Arrhythmia FOS ECG, skin (pattern) late Arrhythmia FOS ECG, skin (pattern) late Arrhythmia/ECGqual FOS ECG, skin (pattern) late Arrhythmia FOS ECG, skin (pattern) 0 APqual: poor quality arterial pressure signal, ECGqual: poor quality ECG signal, POS: position of the IAB catheter too low, Cath: catheter type, FOS: catheter with fibre-optic sensor, NF: catheter without fibre-optic sensor, ECG skin: ECG signal directly from patient skin, ECG MON: ECG signal slaved from monitor, AP: arterial pressure signal. Figure 1. Incidental incorrect timing with two inflations during systole (arrows) on strip nr 191, most likely caused by ECG artefacts. Table 4. Prevalence of poor quality ECG and arrhythmia during correct and incorrect timing Correct timing Poor quality ECG 11% 37% Arrhythmia 5% 69% Incorrect timing In our observations, we found poor quality pressure curves in 14 cases of all strips. Two of these cases presented with incorrect timing, but, in these cases, pattern trigger (using the ECG signal) was used. From our Figure 2. Arterial pressure signal with artefacts during diastolic augmentation. results, we do not conclude a poor quality blood pressure signal to be a cause of incorrect timing. With the use of a fibre-optic sensor, we did not find pressure curves of poor quality, although we did see artefacts during diastolic augmentation (Figure 2). These artefacts may result from whipping catheters or movement against the aortic wall. 14 Interpretation of inflation and deflation timing may be hampered by artefacts or poor quality pressure curves.

5 Bakker EWM et al. 5 Figure 3. Two cases of consistent early inflation (A=strip 71, B= strip 263). In five cases, we found a catheter presenting too low in the descending aorta on X-thorax. In three of these cases, timing was incorrect (consistent late deflation). After repositioning the catheter, timing was correct. Discussion Adding our observations of consistent and incidental incorrect timing, we found a total of 3% and 13% of inflations and deflations, respectively, to be incorrect. In 18 cases (5%), IABP assistance was switched to operator mode. We found consistently incorrect timing at some point of IABP assistance in 15 of 60 patients (25%). Our results should not be compared with the results of Osentowski, 10 using Datascope IABPs. We allowed a maximum of 40 ms deviation from the dicrotic notch or the end of diastole. Osentowski did not mention allowing any deviation from the dicrotic notch, and accepted no less than 5 to 10 mmhg reduction of end diastolic pressure. Furthermore, Osentowski only made a single evaluation per patient. According to Schreuder et al., 6 the fibre-optic sensor of the ultraflex catheter would effectively result in correct inflation timing. Inflation timing deviated only 0.6 ms from the dicrotic notch on average, with a range of 11 to +12 ms in 27 patients. We found four cases (with two patients) of consistent early inflation when using the fibre-optic sensor of the IABP (Figure 3). In these cases, switching to operator mode to adjust inflation timing would be justified. When discussing the accuracy of the autopilot during arrhythmias, early deflations could probably not be prevented if timing was assessed manually. The irregular rhythm would still cause the predictive nature of the algorithm to be disturbed, leading to timing discrepancies. Early deflations may be considered a safe way to handle irregular rhythms. The peak trigger, often used during arrhythmias, does not use R-wave deflation. It has a complex algorithm that continuously checks whether deflation empties the balloon for more than 40 percent before the start of the next systole. On the other hand, it determines its deflation point according to the average of a number of beats, so incidental late and early deflations will occur. It continuously adapts its deflation point to make it the safest, but still effective, point. Switching to the afib algorithm, which does use R-wave deflation, may prevent incidental timing errors during arrhythmias. The disadvantage of this trigger type is that all deflations may be relatively late. The cause of these late deflations could be the volumeregulated nature of the pneumatic system of the Autocat IABP, which may empty the balloon somewhat slower than a pressure-regulated pump. With the 2.23 software update, R-wave deflation can be provided without switching from autopilot mode to operator mode by turning arrhythmia timing on. In general, incorrect timing during arrhythmias may not always be prevented by switching to operator mode. Optimization of clinical care should, therefore, be focussed on resolving the irregular rhythm. One case of incidental incorrect timing is remarkable. We found two beats with inflations during systole (Figure 1). The landmarks show that the IABP misinterpreted the heavily distorted ECG signal. These incorrect inflations may have been prevented with a pacemaker trigger. The delay of the ECG signal when slaved from the monitor will influence timing. 15 We found four cases of monitor use in combination with incorrect timing, compared to seven cases with correct timing. The incidence is too low for conclusions from our observations. In one patient, we found late inflations on three strips (nr 34, 35 and 36 in Table 2) with very small systolic pressure curves followed by a strong augmentation curve. The duration of the systole was short (160 ms), as seen in hypovolemia. Later, during IABP assistance, the systolic time period increases to 240 ms, with a higher systolic pressure and correct IABP timing. In these cases of late inflations, the IABP used the pattern trigger mode, without fibre-optic sensor, resulting in a predictivepredictive timing method. If a fibre-optic sensor had been used, the wave-predictive algorithm might have prevented the late inflations. From our results, we recommend frequent checks of inflation and deflation timing. If incorrect timing is observed, switching to operator mode should be considered, particularly when timing is consistently incorrect. In a case of a poor ECG signal, a pacemaker trigger or

6 6 Perfusion 0(0) arterial pressure trigger may be considered if the signal remains poor. Timing errors during arrhythmias cannot always be prevented. The use of R-wave deflation (arrhythmia timing or afib trigger) to overrule the peak trigger may be considered if end diastolic pressure remains acceptable. Furthermore, a catheter with a fibre-optic sensor is recommended for its robust pressure signal and the application of the wave-predictive algorithm that may prevent late inflations. The clinical relevance of incorrect timing as seen in our observations has been a point of discussion during the last two decades. 16 Using high fidelity pressure measurements at the aortic root 17,18 or a conductance catheter to measure volume and pressure inside the left ventricle 9 may open the debate. We are aware of the fact that interpreting a print from a blood pressure measurement at the descending aorta is a method with low accuracy. Still, it is the clinical way of interpreting inflation and deflation timing in our daily clinical practice. Conclusions Inflation and deflation timing of the Arrow AutoCAT 2 WAVE, while using the autopilot mode, is not always correct. The quality of the ECG input signal and the occurrence of arrhythmia appear to be related to erroneous timing. Switching from autopilot mode to operator mode may not always prevent incorrect timing. Funding This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors. Conflict of Interest Statement None declared. References 1. Trost JC, Hillis D. Intra-aortic balloon counterpulsation. Am J Cardiol 2006; 97: Kantrowitz A, Kantrowitz A. Experimental augmentation of coronary flow by retardation of the arterial pressure pulse. Surgery 1953; 34: Zelano JA, LI JKJ, Welkowitz W. A closed loop control scheme for intraaortic balloon pumping. IE EE Trans Biomed Eng 1990; 37: Schreuder JJ, Maisano F, Donelli A, et al. Beat-to-beat effects of intraaortic balloon pump timing on left ventricular performance in patients with low ejection fraction. Ann Thorac Surg 2005; 79: Reesink KD, Nagel T, Bovelander J, et al. Feasibility study of a fiber-optic system for invasive blood pressure measurements. Cathet Cardiovasc Intervent 2002; 57: Schreuder JJ, Castiglioni A, Donelli A, et al. Automatic intraaortic balloon pump timing using an intrabeat dicrotic notch prediction algorithm. Ann Thorac Surg 2005; 79: Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993; 74: Donelli A, Jansen JRC, Hoeksel B, et al. Performance of a real-time dicrotic notch detection and prediction algorithm in arrhythmic human aortic pressure signals. J Clin Monit 2002; 17: Hoeksel SAAP, Jansen JRC, Blom JA, Schreuder JJ. Detection of dicrotic notch in arterial pressure. J Clin Monit 1997; 13: Osentowski MK, Holt DW. Evaluating the efficacy of intra-aortic balloon pump timing using the auto-timing mode of operation with the Datascope CS100. J Extra Corpor Technol 2007; 39: Patel SI, Souter MJ. Equipment-related electrocardiographic artifacts: causes, characteristics, consequences, and correction. Anesthesiology 2008; 108: Sakiewicz PG, Wright E, Robinson O, et al. Abnormal electrical stimulus of an intra-aortic balloon pump with concurrent support with continuous veno-venous hemodialysis. ASAIO J 2000; 46: Bernstein AD, Parsonnet V. R-wave triggering of external instrumentation during cardiac pacing. PACE 1983; 6: Parissis H, Leotsinidis M, Dougenis D, Richens D. The way the intra-aortic balloon catheter moves within the aorta as a possible mechanism of balloon associated morbidity. Interact CardioVasc Thorac Surg 2007; 6: Arrow International. Operatoring Manual, Arrow Autocat tm 2 Series Intra-Aortic Balloon Pump (IABP) System, IAM-9005, Rev. 1, 5.1 ECG connections, Kern MJ, Aquirre FV, Caracciolo EA, et al. Hemodynamic effects of new intra-aortic balloon counterpulsation timing methods in patients: a multicenter evaluation. Am Heart J 1999; 137: Pantalos GM, Koenig SC, Gillars KJ, Haugh GS, Dowling RD, Gray LA Jr. Intraaortic balloon pump timing discrepancies in adult patients. Artif Organs 2011; 35: Pantalos GM, Minich LL, Tani LY, McGough EC, Hawkins JA. Estimation of timing errors for the intraaortic balloon pump use in pediatric patients. ASAIO J 1999; 45:

Arrow AC3 Optimus Intra-Aortic Balloon Pump Greater precision with increased simplicity; an evolution in IABP performance

Arrow AC3 Optimus Intra-Aortic Balloon Pump Greater precision with increased simplicity; an evolution in IABP performance Arrow AC3 Optimus Intra-Aortic Balloon Pump Greater precision with increased simplicity; an evolution in IABP performance Unmatched simplicity, right from the start The AC3 Optimus Intra-Aortic Balloon

More information

Arrow AC3 Optimus Intra-Aortic Balloon Pump Greater precision with increased simplicity; an evolution in IABP performance

Arrow AC3 Optimus Intra-Aortic Balloon Pump Greater precision with increased simplicity; an evolution in IABP performance Arrow AC3 Optimus Intra-Aortic Balloon Pump Greater precision with increased simplicity; an evolution in IABP performance Optimized therapy even in the most challenging patient conditions When already-compromised

More information

IABP Timing & Fidelity. Pocket Reference Guide

IABP Timing & Fidelity. Pocket Reference Guide IABP Timing & Fidelity Pocket Reference Guide Correct IABP Timing A = One complete cardiac cycle R B = Unassisted aortic end diastolic pressure P T C = Unassisted systolic pressure D = Diastolic augmentation

More information

Section 6 Intra Aortic Balloon Pump

Section 6 Intra Aortic Balloon Pump Section 6 Intra Aortic Balloon Pump The Intra Aortic Balloon Pump (IABP) The balloon is synthetic and is made for single use only. It is threaded into the aorta, usually via a femoral approach. The balloon

More information

CSC Review Course: Intra Aortic Balloon Pump (IABP) Paula M Davis ARNP, MSN, CCRN-CSC

CSC Review Course: Intra Aortic Balloon Pump (IABP) Paula M Davis ARNP, MSN, CCRN-CSC CSC Review Course: Intra Aortic Balloon Pump (IABP) Paula M Davis ARNP, MSN, CCRN-CSC What is an IABP? The Intra-Aortic Balloon Counterpulsation system is a volume displacement device. It should be positioned

More information

SENSATION and CS300 Smaller meets faster. IABP System

SENSATION and CS300 Smaller meets faster. IABP System SENSATION and CS300 Smaller meets faster IABP System Cardiac Assist SENSATION and CS300 3 Cardiac Assist SENSATION and CS300 5 Smaller The SENSATION 7Fr. Fiber-Optic IAB Catheter Faster The CS300 with

More information

Mechanics of Cath Lab Support Devices

Mechanics of Cath Lab Support Devices Mechanics of Cath Lab Support Devices Issam D. Moussa, MD Chief Medical Officer First Coast Cardiovascular Institute, Jacksonville, FL Professor of Medicine, UCF, Orlando, FL None DISCLOSURE Percutaneous

More information

Principles of intra-aortic balloon pump counterpulsation

Principles of intra-aortic balloon pump counterpulsation Principles of intra-aortic balloon pump counterpulsation Murli Krishna MBBS FRCA FFPMRCA Kai Zacharowski MD PhD FRCA Key points The primary goal of intraaortic balloon pump (IABP) treatment is to increase

More information

Mechanics of Cath Lab Support Devices

Mechanics of Cath Lab Support Devices Mechanics of Cath Lab Support Devices Issam D. Moussa, MD Professor of Medicine Mayo Clinic College of Medicine Chair, Division of Cardiovascular Diseases Mayo Clinic Jacksonville, Florida DISCLOSURE Presenter:

More information

Impedance Cardiography (ICG) Method, Technology and Validity

Impedance Cardiography (ICG) Method, Technology and Validity Method, Technology and Validity Hemodynamic Basics Cardiovascular System Cardiac Output (CO) Mean arterial pressure (MAP) Variable resistance (SVR) Aortic valve Left ventricle Elastic arteries / Aorta

More information

THE journal OF EXTRA-CORPOREAL TECHNOLOGY ABSTRACT

THE journal OF EXTRA-CORPOREAL TECHNOLOGY ABSTRACT Original Article Innovations in Intra-Aortic Balloon Pump Management: Computer Modem Technology John M. Ingram, BS, CCP Cedars Medical Center Miami, Florida Keywords: intra-aortic balloon pump; computer;

More information

PROBLEM SET 2. Assigned: February 10, 2004 Due: February 19, 2004

PROBLEM SET 2. Assigned: February 10, 2004 Due: February 19, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

Beat-to-Beat Effects of Intraaortic Balloon Pump Timing on Left Ventricular Performance in Patients With Low Ejection Fraction

Beat-to-Beat Effects of Intraaortic Balloon Pump Timing on Left Ventricular Performance in Patients With Low Ejection Fraction Beat-to-Beat Effects of Intraaortic Balloon Pump Timing on Left Ventricular Performance in Patients With Low Ejection Fraction Jan J. Schreuder, MD, PhD, Francesco Maisano, MD, Andrea Donelli, MS, Jos

More information

Counterpulsation. John N. Nanas, MD, PhD. Professor and Head, 3 rd Cardiology Dept, University of Athens, Athens, Greece

Counterpulsation. John N. Nanas, MD, PhD. Professor and Head, 3 rd Cardiology Dept, University of Athens, Athens, Greece John N. Nanas, MD, PhD Professor and Head, 3 rd Cardiology Dept, University of Athens, Athens, Greece History of counterpulsation 1952 Augmentation of CBF Adrian and Arthur Kantrowitz, Surgery 1952;14:678-87

More information

CLINICAL SUPPORT SERVICES DEVELOPING AN IABP PRECEPTOR STRATEGY

CLINICAL SUPPORT SERVICES DEVELOPING AN IABP PRECEPTOR STRATEGY CLINICAL SUPPORT SERVICES DEVELOPING AN IABP PRECEPTOR STRATEGY DATASCOPE IS NOW MAQUET CARDIOVASCULAR Datascope is now MAQUET Cardiovascular In early 2009, the purchase agreement between Datascope and

More information

Hemodynamic Monitoring and Circulatory Assist Devices

Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic Monitoring and Circulatory Assist Devices Speaker: Jana Ogden Learning Unit 2: Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic monitoring refers to the measurement of pressure,

More information

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018 WAVES CATCH A WAVE.. W I S C O N S I N P A R A M E D I C S E M I N A R A P R I L 2 0 1 8 K E R I W Y D N E R K R A U S E R N, C C R N, E M T - P Have you considered that if you don't make waves, nobody

More information

Ventricular Assisting Devices in the Cathlab. Unrestricted

Ventricular Assisting Devices in the Cathlab. Unrestricted Ventricular Assisting Devices in the Cathlab Unrestricted What is a VAD? A single system device that is surgically attached to the left ventricle of the heart and to the aorta for left ventricular support

More information

Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock

Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock Nickolas Trubov, B.Sc., C.C.P. and Steven J. Phillips, M.D. Mercy Hospital, Des Moines, Iowa 50314 ABSTRACT An 80 cc

More information

Selected age-associated changes in the cardiovascular system

Selected age-associated changes in the cardiovascular system Selected age-associated changes in the cardiovascular system Tamara Harris, M.D., M.S. Chief, Interdisciplinary Studies of Aging Acting Co-Chief, Laboratory of Epidemiology and Population Sciences Intramural

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

HOW TO PERFORM LEFT VENTRICULAR ASSISTANCE IN THE CATHLAB. Andreas Baumbach, MD FESC FRCP Bristol Heart Institute University Hospitals Bristol UK

HOW TO PERFORM LEFT VENTRICULAR ASSISTANCE IN THE CATHLAB. Andreas Baumbach, MD FESC FRCP Bristol Heart Institute University Hospitals Bristol UK HOW TO PERFORM LEFT VENTRICULAR ASSISTANCE IN THE CATHLAB Andreas Baumbach, MD FESC FRCP Bristol Heart Institute University Hospitals Bristol UK Disclosure I have no conflicts of interest regarding this

More information

Impedance Cardiography (ICG) Application of ICG for Hypertension Management

Impedance Cardiography (ICG) Application of ICG for Hypertension Management Application of ICG for Hypertension Management 1mA @ 100 khz Impedance Cardiography (ICG) Non-invasive Beat-to-beat Hemodynamic Monitoring Diastole Systole Aortic valve is closed No blood flow in the aorta

More information

THE HEART THE CIRCULATORY SYSTEM

THE HEART THE CIRCULATORY SYSTEM THE HEART THE CIRCULATORY SYSTEM There are three primary closed cycles: 1) Cardiac circulation pathway of blood within the heart 2) Pulmonary circulation blood from the heart to lungs and back 3) Systemic

More information

Competency Title: Caring for a patient with an arterial line

Competency Title: Caring for a patient with an arterial line Appendix 2 Competency Title: Caring for a patient with an arterial line Aims and objectives Following completion of this competency document the practitioner will be able to: Discuss the indications for

More information

Epidemiology of Heart Failure in Adults

Epidemiology of Heart Failure in Adults Cardiac Critical Care : Focused on IABP & PCPS Epidemiology of Heart Failure in Adults Prevalence Incidence Mortality 2004 Hospital Cost 2007 2004 Age 20+ (New Cases) All Ages Discharges Age 35+ 2004 All

More information

CVICU EXAM. Mrs. Jennings is a 71-year-old post-op CABG x5 with an IABP in her left femoral artery

CVICU EXAM. Mrs. Jennings is a 71-year-old post-op CABG x5 with an IABP in her left femoral artery CVICU EXAM 1111 North 3rd Street Mrs. Jennings is a 71-year-old post-op CABG x5 with an IABP in her left femoral artery 1. Nursing standards for a patient on an IABP device include: a. Know results of

More information

Adult Echocardiography Examination Content Outline

Adult Echocardiography Examination Content Outline Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,

More information

LINEAR. 7.5Fr. IAB Catheter. This document is intended to provide information to an international audience outside of the US.

LINEAR. 7.5Fr. IAB Catheter. This document is intended to provide information to an international audience outside of the US. LINEAR 7.5Fr. IAB Catheter This document is intended to provide information to an international audience outside of the US. Maquet offers a range of convenient and innovative clinical solutions for counterpulsation

More information

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah CARDIOVASCULAR MONITORING Prof. Yasser Mostafa Kadah Introduction Cardiovascular monitoring covers monitoring of heart and circulatory functions It makes it possible to commence interventions quickly in

More information

IN VIVO ASSESSMENT OF CENTRAL AND PERIPHERAL HEMODYNAMIC IMPACT OF THE C-PULSE SYSTEM

IN VIVO ASSESSMENT OF CENTRAL AND PERIPHERAL HEMODYNAMIC IMPACT OF THE C-PULSE SYSTEM IN VIVO ASSESSMENT OF CENTRAL AND PERIPHERAL HEMODYNAMIC IMPACT OF THE C-PULSE SYSTEM Francisco Javier Londono Hoyos, PhD; Dimitrios Georgakopoulos, PhD; Oliver Fey; Dori Jones, MS; Christian Schlensak,

More information

Bridging With Percutaneous Devices: Tandem Heart and Impella

Bridging With Percutaneous Devices: Tandem Heart and Impella Bridging With Percutaneous Devices: Tandem Heart and Impella DAVID A. BARAN, MD, FACC, FSCAI SYSTEM DIRECTOR, ADVANCED HEART FAILURE, TX AND MCS SENTARA HEART HOSPITAL NORFOLK, VA PROFESSOR OF MEDICINE

More information

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time.

Cardiac Cycle. Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. The Heartbeat Cardiac Cycle Each heartbeat is called a cardiac cycle. First the two atria contract at the same time. Next the two ventricles contract at the same time. Then all the chambers relax. http://www.youtube.com/watch?v=frd3k6lkhws

More information

Guideline compliance, utilization trends

Guideline compliance, utilization trends Guideline compliance, utilization trends and device selection Tilmann Schwab Cardiology / Intensive care Cardiac support IABP LVAD Transluminal l LVAD Cardiac support Emergency cardiac life support (ECLS)

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

CHRONIC HEART FAILURE : WHAT ELSE COULD WE OFFER TO OUR PATIENTS? Cardiac Rehabilitation Society of Thailand

CHRONIC HEART FAILURE : WHAT ELSE COULD WE OFFER TO OUR PATIENTS? Cardiac Rehabilitation Society of Thailand CHRONIC HEART FAILURE : WHAT ELSE COULD WE OFFER TO OUR PATIENTS? Cardiac Rehabilitation Society of Thailand ENHANCED EXTERNAL COUNTER PULSATION Piyanuj Ruckpanich, MD. Cardiac Rehabilitation Center Perfect

More information

LINEAR 7.5Fr. IAB Catheter

LINEAR 7.5Fr. IAB Catheter LINEAR 7.5Fr. IAB Catheter The LINEAR 7.5Fr. Catheter Easier to insert* Proprietary IAB Membrane provides better performance* Abrasion resistance improved 43%* Improved fatigue resistance* Immediate inflation

More information

Forward Looking Statement

Forward Looking Statement Forward Looking Statement This presentation contains forward-looking statements. All forward looking statements are management s (Dave Rosa) present expectations of future events and are subject to a number

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information

The Counter HF Clinical Study for Heart Failure

The Counter HF Clinical Study for Heart Failure The Counter HF Clinical Study for Heart Failure CAUTION: C-Pulse is an investigational device. It is limited by Federal (or United States) Law to investigational use only. 13-111-B Agenda Heart Failure

More information

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Oliver Gödje, MD, Christian Thiel, MS, Peter Lamm, MD, Hermann Reichenspurner, MD, PhD, Christof Schmitz, MD,

More information

Technique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial)

Technique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial) Critical Care Monitoring Hemodynamic Monitoring Arterial Blood Pressure Cannulate artery Uses 2 Technique Sites Locate artery, prep 3 1 Technique Local anesthetic? Aseptic technique Hyper-extend (if radial)

More information

Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication vs Benefit? Mortality? Morbidity?

Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication vs Benefit? Mortality? Morbidity? Preoperative intraaortic balloon counterpulsation in high-risk CABG Stefan Klotz, M.D. Preoperative IABP in high-risk CABG Questions?? Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication

More information

Citation Acta medica Nagasakiensia. 1984, 29

Citation Acta medica Nagasakiensia. 1984, 29 NAOSITE: Nagasaki University's Ac Title Author(s) Efficacy of Coenzyme Q10 Administra Aortic Stenosis and Pacemaker Induc Igarashi, Katsuro Citation Acta medica Nagasakiensia. 1984, 29 Issue Date 1984-10-25

More information

12/19/2017. Learning Objectives. Mechanical Circulatory Support. Mechanical Aids to External Massage. Noninvasive Mechanical Support Devices

12/19/2017. Learning Objectives. Mechanical Circulatory Support. Mechanical Aids to External Massage. Noninvasive Mechanical Support Devices Learning Objectives Mechanical Circulatory Support Explain the indications, function, & complications for selected mechanical circulatory support devices Describe examples of ventricular assist devices

More information

QUIZ 1. Tuesday, March 2, 2004

QUIZ 1. Tuesday, March 2, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition Table of Contents Volume 1 Chapter 1: Cardiovascular Anatomy and Physiology Basic Cardiac

More information

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity

Clinical application of Arterial stiffness. pulse wave analysis pulse wave velocity Clinical application of Arterial stiffness pulse wave analysis pulse wave velocity Arterial system 1. Large arteries: elastic arteries Aorta, carotid, iliac, Buffering reserve: store blood during systole

More information

THE CARDIOVASCULAR SYSTEM

THE CARDIOVASCULAR SYSTEM THE CARDIOVASCULAR SYSTEM AND RESPONSES TO EXERCISE Mr. S. Kelly PSK 4U North Grenville DHS THE HEART: A REVIEW Cardiac muscle = myocardium Heart divided into two sides, 4 chambers (L & R) RS: pulmonary

More information

Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU

Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU Ailin Barseghian El-Farra, MD, FACC Assistant Professor, Interventional Cardiology University of California, Irvine Department of Cardiology

More information

Warm Up- Monday -AND- Setup Cornell Notes.

Warm Up- Monday -AND- Setup Cornell Notes. Warm Up- Monday Brainstorm in your notebook: If the heart sends blood to all organs, how and where does the heart get blood to provide oxygen for its muscles? -AND- Setup Cornell Notes. Announcements Unit

More information

Relax and Learn At the Farm 2012

Relax and Learn At the Farm 2012 Relax and Learn At the Farm Session 9: Invasive Hemodynamic Assessment and What to Do with the Data Carol Jacobson RN, MN Cardiovascular Nursing Education Associates Function of CV system is to deliver

More information

Experience with Intraaortic Balloon Counterpulsation

Experience with Intraaortic Balloon Counterpulsation Experience with Intraaortic Balloon Counterpulsation Peter M. Sanfelippo, M.D., Norman H. Baker, M.D., H. Gene Ewy, M.D., Patrick J. Moore, M.D., John W. Thomas, M.D., George J. Brahos, M.D., and Robert

More information

Panorama. Arrhythmia Analysis Frequently Asked Questions

Panorama. Arrhythmia Analysis Frequently Asked Questions Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead

More information

Rhondalyn C. McLean. 2 ND YEAR RESEARCH ELECTIVE RESIDENT S JOURNAL Volume VII, A. Study Purpose and Rationale

Rhondalyn C. McLean. 2 ND YEAR RESEARCH ELECTIVE RESIDENT S JOURNAL Volume VII, A. Study Purpose and Rationale A Randomized Clinical Study To Compare The Intra-Aortic Balloon Pump To A Percutaneous Left Atrial-To-Femoral Arterial Bypass Device For Treatment Of Cardiogenic Shock Following Acute Myocardial Infarction.

More information

Cath Lab Essentials : LV Assist Devices for Hemodynamic Support (IABP, Impella, Tandem Heart, ECMO)

Cath Lab Essentials : LV Assist Devices for Hemodynamic Support (IABP, Impella, Tandem Heart, ECMO) Cath Lab Essentials : LV Assist Devices for Hemodynamic Support (IABP, Impella, Tandem Heart, ECMO) Michael A. Gibson, MD Assistant Professor of Medicine University of California, Irvine Division of Cardiology

More information

Closed Loop Stimulation vs. Conventional DDDR Pacing: Benefits of Hemodynamic Pacing

Closed Loop Stimulation vs. Conventional DDDR Pacing: Benefits of Hemodynamic Pacing 292 June 2000 Closed Loop Stimulation vs. Conventional DDDR Pacing: Benefits of Hemodynamic Pacing P. ZECCHI, F. BELLOCCI, T. SANNA, G. DI MARTINO Catholic University, Policlinico Gemelli, Institute of

More information

NURSING PRACTICE & SKILL

NURSING PRACTICE & SKILL NURSING PRACTICE & SKILL Authors Tanja Schub, BS Mary Woten, RN, BSN Reviewers Eliza Schub, RN, BSN Teresa-Lynn Spears, RN, MSN Nursing Practice Council Glendale Adventist Medical Center, Glendale, CA

More information

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Heart Pump and Cardiac Cycle. Faisal I. Mohammed, MD, PhD

Heart Pump and Cardiac Cycle. Faisal I. Mohammed, MD, PhD Heart Pump and Cardiac Cycle Faisal I. Mohammed, MD, PhD 1 Objectives To understand the volume, mechanical, pressure and electrical changes during the cardiac cycle To understand the inter-relationship

More information

Topic 6: Human Physiology

Topic 6: Human Physiology Topic 6: Human Physiology 6.2 The Blood System D.4 The Heart Essential Questions: 6.2 The blood system continuously transports substances to cells and simultaneously collects waste products. D.3 The chemical

More information

MODULE 2 THE CLINICAL ENIGMA: RANDOMIZED TRIALS vs CLINICAL PRACTICE. Nico H. J. Pijls, MD, PhD Catharina Hospital Eindhoven The Netherlands

MODULE 2 THE CLINICAL ENIGMA: RANDOMIZED TRIALS vs CLINICAL PRACTICE. Nico H. J. Pijls, MD, PhD Catharina Hospital Eindhoven The Netherlands MODULE 2 THE CLINICAL ENIGMA: RANDOMIZED TRIALS vs CLINICAL PRACTICE Nico H. J. Pijls, MD, PhD Catharina Hospital Eindhoven The Netherlands Disclosure All presenters have a speaker agreement with Maquet

More information

Impella Ins & Outs. CarVasz November :45 12:15

Impella Ins & Outs. CarVasz November :45 12:15 Impella Ins & Outs CarVasz November 21 2014 10:45 12:15 Nicolas M. Van Mieghem, MD, PhD, FESC Clinical Director of Interventional Cardiology Thoraxcenter, Erasmus MC Rotterdam Background IABP is widely

More information

Rationale for Prophylactic Support During Percutaneous Coronary Intervention

Rationale for Prophylactic Support During Percutaneous Coronary Intervention Rationale for Prophylactic Support During Percutaneous Coronary Intervention Navin K. Kapur, MD, FACC, FSCAI Assistant Director, Interventional Cardiology Director, Interventional Research Laboratories

More information

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS CARDIAC MECHANICAL SUPPORT PROGRAM GUIDELINES CARDIAC MECHANICAL SUPPORT: LVAD BASICS FREQUENT SCENARIOS AND TROUBLESHOOTING Review Date: July 2011

More information

Cardiovascular Physiology

Cardiovascular Physiology Cardiovascular Physiology Introduction The cardiovascular system consists of the heart and two vascular systems, the systemic and pulmonary circulations. The heart pumps blood through two vascular systems

More information

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The

More information

HEARTWARE HVAD WAVEFORM APP INSTRUCTIONS

HEARTWARE HVAD WAVEFORM APP INSTRUCTIONS HEARTWARE HVAD WAVEFORM APP INSTRUCTIONS TABLE OF CONTENTS Welcome... 3 HVAD Waveforms 1. Characteristics... 4 2. Theory of Operation... 5 3. Ao & LV Pressure... 6 4. HQ Curve... 7 5. PV Loops... 8 Home

More information

Principles of Biomedical Systems & Devices. Lecture 8: Cardiovascular Dynamics Dr. Maria Tahamont

Principles of Biomedical Systems & Devices. Lecture 8: Cardiovascular Dynamics Dr. Maria Tahamont Principles of Biomedical Systems & Devices Lecture 8: Cardiovascular Dynamics Dr. Maria Tahamont Review of Cardiac Anatomy Four chambers Two atria-receive blood from the vena cave and pulmonary veins Two

More information

Feasibility of Leadless Cardiac Pacing Using Injectable. Magnetic Microparticles

Feasibility of Leadless Cardiac Pacing Using Injectable. Magnetic Microparticles Supplementary Information for Feasibility of Leadless Cardiac Pacing Using Injectable Magnetic Microparticles Menahem Y. Rotenberg, Hovav Gabay, Yoram Etzion and Smadar Cohen. Correspondence to: scohen@bgu.ac.il,

More information

Mechanical Cardiac Support in Acute Heart Failure. Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research

Mechanical Cardiac Support in Acute Heart Failure. Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research Mechanical Cardiac Support in Acute Heart Failure Michael Felker, MD, MHS Associate Professor of Medicine Director of Heart Failure Research Disclosures Research Support and/or Consulting NHLBI Amgen Cytokinetics

More information

Index of subjects. effect on ventricular tachycardia 30 treatment with 101, 116 boosterpump 80 Brockenbrough phenomenon 55, 125

Index of subjects. effect on ventricular tachycardia 30 treatment with 101, 116 boosterpump 80 Brockenbrough phenomenon 55, 125 145 Index of subjects A accessory pathways 3 amiodarone 4, 5, 6, 23, 30, 97, 102 angina pectoris 4, 24, 1l0, 137, 139, 140 angulation, of cavity 73, 74 aorta aortic flow velocity 2 aortic insufficiency

More information

LINEAR. 7.5Fr. IAB CATHETER INSTRUCTIONS FOR USE. INSTRUCTIONS FOR USE LINEAR 7.5Fr. Intra-Aortic Balloon Catheter 1

LINEAR. 7.5Fr. IAB CATHETER INSTRUCTIONS FOR USE. INSTRUCTIONS FOR USE LINEAR 7.5Fr. Intra-Aortic Balloon Catheter 1 LINEAR 7.5Fr. IAB CATHETER INSTRUCTIONS FOR USE Read this entire booklet before opening inner pack. INSTRUCTIONS FOR USE LINEAR 7.5Fr. Intra-Aortic Balloon Catheter 1 INTENDED USE The intra-aortic balloon

More information

ISPUB.COM. Intra Aortic Balloon Pump (IABP) Counterpulsation. P Oberwalder HISTORY PHYSIOLOGIC EFFECTS OF IABP THERAPY

ISPUB.COM. Intra Aortic Balloon Pump (IABP) Counterpulsation. P Oberwalder HISTORY PHYSIOLOGIC EFFECTS OF IABP THERAPY ISPUB.COM The Internet Journal of Thoracic and Cardiovascular Surgery Volume 2 Number 2 Intra Aortic Balloon Pump (IABP) Counterpulsation P Oberwalder Citation P Oberwalder. Intra Aortic Balloon Pump (IABP)

More information

Modern Left Ventricular Assist Devices (LVAD) : An Intro, Complications, and Emergencies

Modern Left Ventricular Assist Devices (LVAD) : An Intro, Complications, and Emergencies Modern Left Ventricular Assist Devices (LVAD) : An Intro, Complications, and Emergencies ERIC T. ROME D.O. HEART FAILURE, MECHANICAL ASSISTANCE AND TRANSPLANTATION CVI Left Ventricular Assist Device An

More information

Chapter 13 The Cardiovascular System: Cardiac Function

Chapter 13 The Cardiovascular System: Cardiac Function Chapter 13 The Cardiovascular System: Cardiac Function Overview of the Cardiovascular System The Path of Blood Flow through the Heart and Vasculature Anatomy of the Heart Electrical Activity of the Heart

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval BPU22 with the AViTA BPM63S Devices Item 9 HARTMANN Veroval BPU22 AViTA BPM63S Pictures Display Image Validation Equivalence ESH 2010 ESH 2002 BHS AAMI Category Arm Type

More information

C postoperative care of patients undergoing cardiac. Evaluation of an Extraaortic Counterpulsation Device in Severe Cardiac Failure

C postoperative care of patients undergoing cardiac. Evaluation of an Extraaortic Counterpulsation Device in Severe Cardiac Failure Evaluation of an Extraaortic Counterpulsation Device in Severe Cardiac Failure John A. Zelano, PhD, Wilson KO, MD, Richard Lazzaro, MD, W. Douglas Lazenby, MD, Stephen Topaz, MS, Willem J. Kolff, MD, 0.

More information

EACTS Adult Cardiac Database

EACTS Adult Cardiac Database EACTS Adult Cardiac Database Quality Improvement Programme List of changes to Version 2.0, 13 th Dec 2018, compared to version 1.0, 1 st May 2014. INTRODUCTORY NOTES This document s purpose is to list

More information

*Generating blood pressure *Routing blood: separates. *Ensuring one-way blood. *Regulating blood supply *Changes in contraction

*Generating blood pressure *Routing blood: separates. *Ensuring one-way blood. *Regulating blood supply *Changes in contraction *Generating blood pressure *Routing blood: separates pulmonary and systemic circulations *Ensuring one-way blood flow: valves *Regulating blood supply *Changes in contraction rate and force match blood

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium

More information

Effect of physiological heart rate changes on left ventricular dimensions and mitral blood flow velocities in the normal fetus

Effect of physiological heart rate changes on left ventricular dimensions and mitral blood flow velocities in the normal fetus ELSEVIER Early Human Development 40 (1995) 109-114 Effect of physiological heart rate changes on left ventricular dimensions and mitral blood flow velocities in the normal fetus P.B. Tsyvian a, K.V. Malkin

More information

BACHELOR OF SCIENCE IN CARDIO VASCULAR TECHNOLOGY

BACHELOR OF SCIENCE IN CARDIO VASCULAR TECHNOLOGY BACHELOR OF SCIENCE IN CARDIO VASCULAR TECHNOLOGY BCVT 101 Eco Cardiography BCVT 102 ECG-Stream-Holter BCVT 103 Cat lab BCVT 104 Anatomy 1 st YEAR BCVT 105 Physiology BCVT 106 Pathology & Pathophysiology

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval wrist blood pressure monitor with the AViTA BPM15S Devices HARTMANN Veroval wrist blood pressure monitor AViTA BPM15S Pictures Display

More information

Device Equivalence Evaluation Form

Device Equivalence Evaluation Form Device Equivalence Evaluation Form Comparison of the HARTMANN Veroval upper arm blood pressure monitor with the AViTA BPM63S Devices Pictures HARTMANN Veroval upper arm blood pressure monitor AViTA BPM63S

More information

RV dysfunction and failure PATHOPHYSIOLOGY. Adam Torbicki MD, Dept Chest Medicine Institute of Tuberculosis and Lung Diseases Warszawa, Poland

RV dysfunction and failure PATHOPHYSIOLOGY. Adam Torbicki MD, Dept Chest Medicine Institute of Tuberculosis and Lung Diseases Warszawa, Poland RV dysfunction and failure PATHOPHYSIOLOGY Adam Torbicki MD, Dept Chest Medicine Institute of Tuberculosis and Lung Diseases Warszawa, Poland Normal Right Ventricle (RV) Thinner wall Weaker myocytes Differences

More information

Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring

Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring Critical Care Nursing Program August to November, 2015 Full-time Lesson A13 Pumping and Perfusion III Basic Hemodynamic Monitoring August 2015 RN Professional Development Centre Page 1 Lesson Thirteen

More information

Percutaneous Mechanical Circulatory Support Devices

Percutaneous Mechanical Circulatory Support Devices Percutaneous Mechanical Circulatory Support Devices Daniel Vazquez RN, RCIS Miami Cardiac & Vascular Institute FINANCIAL DISCLOSURES none CASE STUDY CASE STUDY 52 year old gentlemen Complaining of dyspnea

More information

Smith, Miller and Grab er(4) state that the maintenance of an efficient

Smith, Miller and Grab er(4) state that the maintenance of an efficient THE SIGNIFICANCE OF THE DIASTOLIC AND SYSTOLIC BLOOD-PRESSURES FOR THE MAINTENANCE OF THE CORONARY CIRCULATION. BY G. V. ANREP AND B. KING. (From the Physiological Laboratory, Cambridge.) IT is generally

More information

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart Cardiovascular Physiology Heart Physiology Introduction The cardiovascular system consists of the heart and two vascular systems, the systemic and pulmonary circulations. The heart pumps blood through

More information

Acute heart failure: ECMO Cardiology & Vascular Medicine 2012

Acute heart failure: ECMO Cardiology & Vascular Medicine 2012 Acute heart failure: ECMO Cardiology & Vascular Medicine 2012 Lucia Jewbali cardiologist-intensivist 14 beds/8 ICU beds Acute coronary syndromes Heart failure/ Cardiogenic shock Post cardiotomy Heart

More information

Management of Cardiogenic Shock. Dr Stephen Pettit, Consultant Cardiologist

Management of Cardiogenic Shock. Dr Stephen Pettit, Consultant Cardiologist Dr Stephen Pettit, Consultant Cardiologist Cardiogenic shock Management of Cardiogenic Shock Outline Definition, INTERMACS classification Medical management of cardiogenic shock PA catheters and haemodynamic

More information

Coronary Anomalies & Hemodynamic Identification

Coronary Anomalies & Hemodynamic Identification Coronary Anomalies & Hemodynamic Identification David Stultz, MD Cardiology Fellow, PGY 6 May 2, 2006 Anomaly #1 Anomaly #2 Anomaly #3 Figure 18-27 Anomalous origin of the left circumflex artery.

More information

Cite this article as:

Cite this article as: doi: 10.21037/acs.2018.08.06 Cite this article as: Loforte A, Baiocchi M, Gliozzi G, Coppola G, Di Bartolomeo R, Lorusso R. Percutaneous pulmonary artery venting via jugular vein while on peripheral extracorporeal

More information

SECTION I FUNDAMENTALS AND CLINICAL APPLICATIONS OF HEMODYNAMICS: COPYRIGHTED MATERIAL

SECTION I FUNDAMENTALS AND CLINICAL APPLICATIONS OF HEMODYNAMICS: COPYRIGHTED MATERIAL SECTION I FUNDAMENTALS AND CLINICAL APPLICATIONS OF HEMODYNAMICS: COPYRIGHTED MATERIAL UNDERSTANDING THE PRESSURE WAVES IN THE HEART: THE WIGGER S DIAGRAM Everything you want to know about hemodynamics

More information