Europace (2001) 3, doi: /eupc , available online at on

Size: px
Start display at page:

Download "Europace (2001) 3, doi: /eupc , available online at on"

Transcription

1 Europace (21) 3, doi:1.153/eupc , available online at on ICDs Clinical evaluation of defibrillation efficacy with a new single-capacitor biphasic in patients undergoing implantation of an implantable cardioverter defibrillator J. Brugada 1, B. Herse 2, B. Sandsted 3, U. Michel 4, B. D. Schubert 4 and S. J. Hahn 5 1 Arrhythmia Unit, Cardiovascular Institute, Hospital Clinic, University of Barcelona, Spain; 2 University Clinic, Goettingen, Germany; 3 University Clinic, Gothenburg, Sweden; 4 Guidant Research, Brussels, Belgium; 5 Guidant Tachyarrhythmia Research, St Paul, U.S.A. Aims Improvements in the size and shape of implantable cardioverter defibrillators (ICDs) might be obtained by using one capacitor instead of the series connection of two capacitors traditionally used in ICDs. The aim of this study was to determine whether a biphasic delivered from a single 336 μf capacitor had the same defibrillation efficacy as a standard biphasic. Methods and Results Randomized, paired defibrillation threshold testing was acutely performed in 54 patients undergoing ICD implantation. A standard 14 μf 8% tilt biphasic (two 28 μf capacitors connected in series) was compared with an experimental biphasic delivered from a single 336 μf capacitor at either 6% tilt (33 patients) or 8% tilt (21 patients). All s had a 6/4 phase1/phase2 duration ratio. Compared with the standard, the 6% tilt experimental had a lower delivered energy ( vs joules, P< 2), lower peak voltage ( vs V, P< 1), and a slightly longer pulse duration ( vs ms, P< 1). Conversely, the 8% tilt experimental had a higher delivered energy ( vs joules, P< 1), a lower peak voltage ( vs V, P< 1) and a much longer pulse duration ( vs ms,P< 1). Conclusion Waveforms delivered from a large capacitance are feasible but require a lower tilt. This technique may allow smaller, thinner ICDs without jeopardizing defibrillation success. (Europace 21; 3: ) 21 The European Society of Cardiology Key Words: Human, defibrillation, pulse-, capacitor, pulse- tilt. Introduction Over the past several years, research and technological advances have allowed important improvements in Manuscript submitted 13 November 2, revised 18 June 21, and accepted 2 July 21. This study was supported in part by Guidant Research, St Paul, Minnesota, U.S.A. and Brussels, Belgium. Correspondence and/or reprint requests: Dr Josep Brugada, Director of the Arrhythmia Unit, Cardiovascular Institute, Hospital Clinic, c/ Villarroel 17, 836 Barcelona, Spain. jepbrugada@grn.es implantable cardioverter defibrillators (ICD). New and highly effective algorithms for detection, treatment and follow-up have been developed [1,2,3] along with significant reductions in the size of the device [4,5]. However, current device sizes and shapes are still not ideal and further research and technological advances are needed in this area. The size and shape of an implantable pulse generator is partially dependent on the physical size of the capacitors that are used to generate the defibrillating shock since they consume approximately 1/3 of the total volume of the ICD. The energy stored on a capacitor is proportional to its capacitance and to the square of its voltage (E= 1 2 CV2 ) but its size is usually primarily /1/ $35./ 21 The European Society of Cardiology

2 Single-capacitor defibrillation 279 dictated by its capacitance. Thus, lower capacitance, higher voltage capacitors would appear to be desirable, but traditional capacitor technologies suitable for ICDs are limited to maximum voltages between 35 and 39 Volts. Thus, ICDs have used a combination of capacitors connected in series to provide adequate voltage (7 8 V) for effective defibrillation. Defibrillating s from ICDs are commonly referred to as single-capacitor s, even though two or more physical capacitors are actually used. The term single capacitance would be more correct. Recently there has been a great deal of interest in improving the biphasic by searching for more optimum capacitance values. The majority of this research has focused on trying to reduce the energy requirements for defibrillation by shortening the pulse duration. This is most easily accomplished by reducing capacitance but at the expense of increasing the voltage and current requirements [6 11]. However, present and evolving capacitor technologies suggest a search for s that defibrillate with lower voltages and currents may be a more effective strategy for improving device size and shape. Lengthening the pulse duration, and reducing the tilt of the defibrillation are two ways that defibrillation voltages can be reduced [12 17]. One means to lengthen pulse duration is to increase the capacitance value used for defibrillation. Block et al. [18] showed a single 45 μf capacitor reduced voltage requirements by 45% with no change in energy requirements. However, 45 μf has a longer than is desired (>2 ms) and may provide more energy than necessary for future ICDs. In this study we investigated the use of an intermediate capacitance value (336 μf) and determined the importance of tilt on the successful application of this larger capacitance. Specifically, we compared the defibrillation efficacy of a standard 14 μf, 8% tilt biphasic to a new experimental biphasic generated from a single 336 μf capacitor at 6% tilt (1st series) and at 8% tilt (2nd series) in patients undergoing implantation of an implantable cardioverter-defibrillator. Methods Patients Fifty-four patients undergoing implantation of a new transvenous cardioverter-defibrillator system were included in the study (33 patients in the 1st series and 21 patients in the 2nd series). Written informed consent was obtained for all patients under a protocol approved by the Ethics Committee of each participating institution. Test procedure Under general anaesthesia, each patient underwent routine transvenous implantation of a single endocardial Volts Volts Volts ms ms (a) (b) (c) 14 µf, 8% tilt standard 6% tilt 1st series 8% tilt 2nd series ms Figure 1 Waveforms tested in this study. Each is shown with a peak voltage appropriate for a 1 Joule shock (stored energy). (a) Standard 14 μf, 8% tilt biphasic tested in all patients, (b) 6% tilt, 336 μf experimental tested in the first series of 33 patients, (c) 336 μf experimental tested in the second series of 21 patients. defibrillation lead (Endotak Model 95 or 1, Guidant Corporation, St Paul, MN, U.S.A.) using fluoroscopic guidance. A hot can emulator (Model 6967, Guidant Corporation) was placed in a subcutaneous pocket in the left pectoral region. Defibrillation efficacy was tested using a right ventricle (leading edge of 1st phase cathodic) to superior vena cava and hot can (leading edge of 1st phase anodic) lead configuration (RV SVC+Can). Two specifically modified external research defibrillators were used for delivery of the two biphasic s (ERD, Guidant Corporation). The standard 14 μf had an 8% overall tilt and a 6/4 phase1/phase2 duration ratio (Fig. 1a). In the 1st series

3 28 J. Brugada et al. of patients, the 336 μf experimental had an overall tilt of 6% with a 6/4 phase duration ratio (Fig. 1b). In the 2nd series of patients the 336 μf experimental had an overall tilt of 8% with a 6/4 phase duration ratio (Fig. 1c). In each patient, defibrillation efficacy using the standard was tested in parallel with the experimental using a randomized, interleaved, step-down defibrillation threshold protocol. Fibrillation was induced and allowed to continue for 1 s before application of the test shock. Only one test shock during each fibrillation episode was used for data analysis. Upon failure of any test shock, rescue shocks were delivered using a standard external cardioverter defibrillator. All testing began with 15 Joules. Testing continued by stepping down to successively lower energy levels (1, 5, 12, 1, 8, 5 J) until one failure was observed with each device or the lowest energy level of 5 Joules was attained. At least 5 min was allowed between each fibrillation episode. The defibrillation threshold (DFT), was defined as either the lowest energy that succeeded or 5 Joules. Upon completion of the defibrillation test protocol the hot can emulator was removed and the ICD implant was completed in a normal way. Data analysis and statistics All data were expressed as mean standard deviation. When quoting pairs of means, the data for the standard were always given first. Within each series of patients, paired Student s t-tests were used to compare the stored energy, delivered energy, peak voltage, peak current, system impedance and pulse duration at the defibrillation threshold between the standard and the experimental. Differences were considered statistically significant at the P< 5 level. Only patients who completed defibrillation threshold testing with both s were included in the analysis. Statistical analysis between 1st and 2nd series of patients was performed using a two-sample t-test (unpaired) on the corresponding data assuming equal variance in the two samples. Sample size was determined prospectively by assuming an α= 5 and a power (1 β) of 8% to detect a 2 Joule difference in energy requirements between the standard and experimental s. Assuming an expected mean energy of 9 5 Joules for the standard yielded a sample size of 41 patients for each series of patients. Results Patient demographics Fifty-six patients were enrolled in this study. The actual mean and standard deviation of the standard turned out to be significantly lower than assumed in the Table 1 Patient characteristics in 1st and 2nd series 1st series 2nd series n Male/female 27/6 2/1 Age (years) LVEF (%) Weight (kg) CAD 46% 48% Primary VF 37% 19% Amiodarone 18% 19% Sotalol 15% % CAD=coronary artery disease; LVEF=left ventricular ejection fraction; VF=ventricular fibrillation. sample size calculations so the first series of patients was stopped after 33 patients. These 33 patients had an actual power of 98% to resolve a 2 Joule difference in energy requirements. The second series of patients was stopped after 21 patients since a statistically significant result had been reached. The 21 patients in the second series had an actual power of 89% to resolve a 2 Joule difference in energy requirements. The clinical characteristics of each series of patients are shown in Table 1. No complications were observed during defibrillation threshold testing in any of the patients in the 1st or 2nd series. All patients were successfully implanted with an ICD system. 1st series of patients (standard vs 6% tilt experimental s) In the first series of 33 patients the single capacitor 336 μf experimental had a 6% tilt. The mean impedance of the lead system was 43 5 Ω with both standard and experimental s. At that impedance the experimental had a % longer pulse duration than the standard ( vs ms, P< 1). Mean stored energy for defibrillation was similar between the standard and the experimental s (8 3 3 vs Joules, P= 71). However, the mean delivered energy requirements were 15% lower with the experimental ( vs Joules, P= 27). Peak voltages were 35% lower with the experimental ( vs Volts, P< 1) and peak currents were 36% lower (8 1 6 vs Amps, P< 1). 2nd series of patients (standard vs 8% tilt experimental) In the second series of 21 patients the 336 μf single capacitor experimental had an 8% tilt. Mean impedance of the lead system was 46 5 Ω with both s. At this impedance, the experimental now had a much longer pulse duration than

4 Single-capacitor defibrillation 281 Series 1 33 patients Series 2 21 patients ± ± 2 8 P = ± ± 3 5 P = 1 Energy (J) µf, 6% tilt, 14 µf, Figure 2 Comparison of mean delivered energy defibrillation thresholds (DFTs) for the standard ( 14 μf; ) and experimental (336 μf) s (). The experimental had a 6% tilt in the first series of 33 patients, and an 8% tilt in the second series of 21 patients. the standard ( vs ms, P< 1). The experimental with 8% tilt required 38% more stored energy ( vs9 3 3 Joules, P= 2) and 44% more delivered energy ( vs Joules, P= 1) than the standard. While mean energy requirements increased, peak voltage (32 51 vs Volts, P< 1) and peak current ( vs Amps, P< 1) requirements were 23% lower using the experimental. second group of patients, it is likely that the true difference between 6% tilt and 8% tilt s was underestimated. To obtain a lesser assessment of the true differences between s the 336 μf experimental data was normalized by the control data within each patient. An unpaired t-test of the normalized data showed the 8% tilt had higher stored energy (41%, P< 1), higher delivered energy (62%, P< 1), and higher peak voltage (12%, P= 1). Cross-comparison of data from series 1 and 2 Unpaired comparisons between data from the two series of patients showed the standard had a 7% higher lead system impedance (P= 13), a 7% longer pulse duration and a 15% lower peak current (P= 12) in the second series of patients compared with the first series. The standard also showed a trend towards lower stored energy (P= 66), delivered energy (P= 58) (Fig. 2) and peak voltage (P= 8) in the second series of patients. Turning to the 336 μf experimental s, a 6% tilt was tested in series 1 and a 8% tilt was tested in series 2. The second series of patients had a 7% higher lead system impedance (P= 13) but a 92% longer (P< 1) (Fig. 3), due to the increase in tilt. Delivered energy at DFT was 35% higher with the 8% tilt (P= 7) (Fig. 2). Stored energy and peak voltage also trended higher but were not statistically significant. However, since the control s mean defibrillation requirements were lower in the Discussion In this study we tested patients undergoing defibrillator implantation to determine whether the use of a shock generated from single 336 μf capacitor would result in the same defibrillating efficacy as a standard 14 μf. Our data showed that a 6% tilt biphasic from a single 336 μf capacitor provided the same defibrillation efficacy as the standard with the additional benefit of significantly reduced peak voltage and current. However, an 8% tilt was significantly less effective than the 6% tilt when 336 μf was used. Biphasic strength-duration relationships: impact on device design It is well known that the shape of a defibrillating has a significant impact on defibrillation strength requirements. Strength-duration relationships for defibrillation with fixed tilt monophasic s

5 282 J. Brugada et al. 3 Series 1 33 patients 11 ± ± 1 4 P < 1 Series 2 21 patients 11 ± 1 7 ± 2 5 P <.1 Pulse duration (ms) µf, 6% tilt, 14 µf, Figure 3 Comparison of mean pulse durations at DFT for the standard ( 14 μf) and experimental (336 μf) s. The experimental had a 6% tilt in the first series of 33 patients, and an 8% tilt in the second series of 21 patients. Abbreviations as in Figure 2. are similar in shape to those known for pacing [12,14]. Defibrillation energies increase monotonically between 2 and 2 ms while voltage and current decrease to a stable level beyond 1 ms [14]. Therefore, it should be possible to reduce defibrillation energy by reducing capacitance to obtain a shorter duration, but at the expense of higher voltage and current [1,11]. However, several clinical studies showed that the combination of biphasic s with smaller capacitance, and the resulting shorter duration, failed to reduce energy despite a significant increase in voltage and current [6 9]. Thus, device size reduction by using smaller capacitance may not be as attractive as once believed, although it still appears to be feasible. Unlike monophasic s, strength-duration relationships for fixed-tilt biphasic s determined in animal models [13] showed that energy requirements are relatively constant between 3 and 1 ms and only increased at durations greater than 2 ms. Peak voltage and peak current decreased continuously over the entire range of pulse durations. Figure 4 illustrates defibrillation strength-duration relationships for both monophasic and biphasic s using data collected from a swine study [13]. Energy for a monophasic almost doubled when pulse duration was increased from 3 to 1 ms, while energy for a biphasic was unchanged over this same range of pulse durations. The biphasic energy requirement was only slightly higher at 2 ms and did not reach a statistically different value until 3 ms. The voltage-duration relationships show that voltage was reduced as pulse durations increased for both monophasic and biphasic s, but that biphasic s yielded a larger reduction in voltage at longer durations. This difference between monophasic and biphasic strength duration relationships explains the lack of energy reduction seen in the human biphasic studies with smaller capacitances. The biphasic strength-duration data also suggests that a device using a single, larger capacitance value might be feasible. Previous studies in humans [18] and in swine [19] showed that a single 45 5 μf capacitor and 6% tilt could be used for effective defibrillation. Peak voltages and currents were reduced by almost 45%, without any change in delivered energy compared with a standard 14 μf, 8% tilt biphasic. While these studies showed that defibrillation with a single larger capacitor was a feasible method to reduce device size, the animal studies [13] (Fig. 4) suggest 45 5 μf capacitance produces a longer than desired (>2 ms) and may store more energy than necessary for future ICDs ( 34 Joules). The present study used a single capacitor with an intermediate capacitance value and investigated the importance of tilt. Importance of tilt These new human data are the first to confirm the importance of tilt when using a larger capacitance. The results agree with the work of several other groups studying the effect of tilt with smaller capacitances. Swartz et al. [15] and Natale et al. [16] both showed lower tilts reduce energy and voltage requirements for defibrillation in humans. Data from a swine model also showed that tilts in the range of 5 6% performed best for a 336 μf [2]. The efficacy of the 8% tilt experimental was less than the

6 Single-capacitor defibrillation 283 Energy (J) Voltage (V) Capacitance for 8% tilt 78 µf 155 µf Monophasic Pulse duration (ms) Monophasic Biphasic Biphasic 311 µf 3 Capacitance for 8% tilt 78 µf 155 µf 311 µf Pulse duration (ms) 6% tilt despite the fact that more energy was delivered to the heart with the higher tilt. This may be due to spontaneous re-induction of ventricular fibrillation by long pulse durations with low trailing edge voltages [12]. While recent studies suggest large tilts (near 1%) may still be effective for certain types of s with small time constants and short pulse durations [21], the re-fibrillation phenomenon is probably still relevant to the larger time constant, longer pulse durations of the present study. These data as well as the cited references support the conclusion that delivered energy is not the most important determinant of defibrillation success, and in fact delivering too much energy, via a large tilt is actually counter-productive. Thus, optimization of tilt, as suggested by Swartz [15], Natale [16] and others [22 24], may lead to further improvements of ICDs, even though lower tilt s deliver less energy. Finally, optimization of tilt may be of particular importance in the case of s delivered from large capacitance since traditional 8% tilt s become much too long. 466 µf 466 µf Figure 4 Energy duration (a) and voltage duration (b) relationships for defibrillation with a 6% tilt monophasic and an 6/4 phase ratio, biphasic from a swine study [13]. The approximate capacitance required to produce an 8% tilt with a given pulse duration is also shown. Energy requirements for the monophasic almost double between 3 and 1 ms while there was no change with the biphasic. Only a 3 ms duration ( 466 μf) had a statistically significant increase in energy with the biphasic. Voltages decrease with increasing pulse duration for both monophasic and biphasic s, but biphasics have a larger voltage reduction that persist out to at least 2 ms pulse durations. 3 Clinical implications The size and shape of ICDs are dependent on the capacitors used to generate the defibrillating shocks since they consume approximately 1/3 of the total volume in an ICD. Traditional ICD capacitor technologies have maximum voltages between 35 and 39 Volts. Thus, to achieve the voltages required for defibrillation, two or more capacitors must be connected in series. However, connecting two capacitors in series reduces their effective capacitance by half, increases ICD size, and limits devices to a rather rectangular shape since that is the most efficient way to fit two capacitors. The rationale for the present study arises from the recognition that a somewhat larger single capacitor consumes less volume and is easier to package efficiently in a more patient-friendly shape than a two-capacitor system. For example, a single 336 μf aluminum electrolytic capacitor is approximately 33% smaller ( 7 cc) than two μf capacitors (1 μf when in series). The single capacitor would also allow a more rounded or elliptical shape for the ICD, rather than the usual rectangular shape that is most efficient for packaging two capacitors. Charge times should not be affected by changes in capacitance when used with the constant power charge circuits found in many modern ICDs. New developments finally made possible by single capacitor techniques must not compromise defibrillation safety. The data obtained in this study showed that future devices should be able to use the larger capacitance, single capacitor technology without jeopardizing defibrillation safety. As an example, one possible device might use a single 336 μf, 39 Volt capacitor. Such a device would store approximately 26 Joules and could be 7 8 cc, or more, smaller than a similar two capacitor device storing 31 Joules. In our study all patients were successfully defibrillated using a stored energy of 15 J or less. Applying the traditional safety margin of 1 Joules, all 54 patients could have received such a 26 Joule device. The success of commercially available devices with slightly lower energy outputs confirms that such a device should be successful for the vast majority of ICD patients []. Alternatively, the results of this and the other studies cited show that a wide range of capacitances with appropriately selected peak voltages could be employed to make an ICD with specifically designed size and energy specifications. Limitations While the results of this study are clear and an appropriate number of patients were tested to support statistically significant conclusions, the study was performed on a limited number of patients. Overall, both groups of patients tended to have a somewhat higher ejection fraction than a typical ICD patient population and no patients had DFTs greater than 15 Joules. The first series had a higher percentage of patients with primary ventricular fibrillation than might be typical and the first

7 284 J. Brugada et al. group had more patients on Sotalol than the second group (15% vs %). None of these factors should have affected the paired defibrillation results, but may have had some effect on the unpaired comparisons between the two series of patients. While no studies have shown a link between patient characteristics and DFTs, patients with lower ejection fractions and higher DFTs exist. Sotalol may also have some effect on DFTs. Thus, caution should be used in generalizing these results to a large ICD population. While the 336 μf, 6% tilt should be expected to perform at least as well as a standard ICD in all types of patients, a few patients may have higher defibrillation requirements that might not always allow implantation of a lower energy device. Conclusions Defibrillation using a single 336 μf capacitor with a 6% tilt required similar amounts of stored energy but significantly reduced delivered energy, peak voltage and peak current compared with the standard 14 μf that uses two capacitors in series. A single 336 μf capacitor and 6% tilt could be used to improve significantly the size and shape of future implantable defibrillators. References [1] Higgins SL, Lee RS, Kramer RL. Stability: an ICD detection criterion for discriminating atrial fibrillation from ventricular tachycardia. J Cardiovasc Electrophysiol 1995; 6: [2] Neuzner J, Pitschner HF, Schlepper M. Programmable VT detection enhancements in implantable cardioventer defibrillator therapy. Pacing Clin Electrophysiol 1995; 18: [3] Schaumann A, Von zur Muhlen F, Gonska, et al. Enhanced detection criteria in implantable cardioverted defibrillators to avoid inappropriate therapy. Am J Cardiol 1996; 78: [4] Klein H, Auricchio A, Huvelle E, et al. Initial experience with a new down-sized implantable cardioverter defibrillator. Am J Cardiol 1996; 78: [5] KenKnight BH, Jones BR, Thomas AC, et al. Technological advances in implantable cardioverter defibrillators before the year 2 and beyond. Am J Cardiol 1996; 78: [6] Poole JE, Kudenchuck PJ, Dolack GL, et al. A prospective randomized comparison in humans of 9 μf and 12 μf biphasic pulse defibrillation using a unipolar defibrillation system. J Cardiovasc Electrophysiol 1995; 6: [7] Bardy GH, Poole JE, Kudenchick PJ, et al. A prospective randomized comparison in humans of biphasic s 6 μf and 12 μf capacitance pulses using a unipolar defibrillation system. Circulation 1995; 91: [8] Block M, Hammel D, Böcker D, et al. Internal defibrillation with smaller capacitors: a prospective randomized cross-over comparison of defibrillation efficacy obtained with 9 μf and 1 μf capacitors in humans. J Cardiovasc Electrophysiol 1995; 6: [9] Swerdlow CD, Kass RM, Chen PS, et al. Effect of capacitor size and pathway resistance on defibrillation threshold for implantable defibrillators. Circulation 1994; 9: [1] Leonelli FM, Kroll MW, Brewer JE. Defibrillation thresholds are lower with smaller storage capacitors. Pacing Clin Electrophysiol 1995; 18: [11] Rist K, Tchou PJ, Mowrey K, et al. Smaller capacitors improve the biphasic. J Cardiovasc Electrophysiol 1994; 5: [12] Irnich W. The fundamental law of electrostimulation and its application to defibrillation. Pacing Clin Electrophysiol 199; 13: [13] Hahn SJ, Heil JE, Lin Y, et al. Defibrillation strengthduration relationships for fixed-tilt biphasic s. Circulation 1994; 9: I 228 (Abstract). [14] Wessale JL, Bourland JD, Tacker WA, et al. Bipolar catheter defibrillation in dogs using trapezoidal s of various tilts. J Electrocardiol 198; 13: [15] Swartz JF, Fletcher RD, Karasik PE. Optimization of biphasic s for nonthoracotomy defibrillation. Circulation 1993; 88: [16] Natale A, Sra J, Krum D, et al. Relative efficacy of different tilts with biphasic defibrillation in humans. Pacing Clin Electrophysiol 1996; 19: [17] Chapman PD, Wetherbee JN, Vetter JW, et al. Strengthduration curves of fixed pulse width variable tilt truncated exponential s for nonthoracotomy internal defibrillation in dogs. Pacing Clin Electrophysiol 1988; 11: [18] Block M, Hammel D, Bocker D, et al. Biphasic Defibrillation using a single capacitor with large capacitance: reduction of peak voltages and ICD device size. Pacing Clin Electrophysiol 1996; 19: [19] Hahn SJ, Heil JE, Lang DJ. Large capacitor defibrillation reduces peak voltage without increasing energies. Pacing Clin Electrophysiol 1995; 16: [2] Hahn SJ, Heil JE, Lang DJ. Defining the optimal tilt of a 336 μf biphasic defibrillation. Pacing Clin Electrophysiol 1997; 2: 1167 (Abstract). [21] Yamanouchi Y, Fishler MG, Mowrey KA, Wilkoff BL, Mazgalev TN, Tchou PJ. New approach to biphasic s for internal defibrillation: fully discharging capacitors. J Cardiovasc Electrophysiol 2; 11(8): [22] Walcott GP, Walker RG, Cates AW, et al. Choosing the optimal monophasic and biphasic s for ventricular defibrillation. J Cardiovasc Electrophysiol 1995; 6: [23] Kroll MW. A minimal mode of the single capacitor biphasic defibrillation. Pacing Clin Electrophysiol 1994; 17: [24] Hahn SJ, Heil JE, Lin Y, et al. Optimization of 9 μf biphasic s for ICD s using a theoretical model and central composite design of experiments. Pacing Clin Electrophysiol 1996; 19: 655 (Abstract). [] Neuzner J. Safety margins: lessons learned from the Low Energy Endotak Trial (LEET). Am J Cardiol 1996; 78:

Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation

Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation 224 Vol. 8, No. 4, December 2003 Comparative Efficacy of Triphasic and Biphasic Internal Defibrillation K. MISCHKE, M. ZARSE, K. BREHMER, C. STELLBRINK, P. HANRATH, P. SCHAUERTE Department of Cardiology,

More information

Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients

Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients Transvenous Biventricular Defibrillation Halves Energy Requirements in Patients Christian Butter, MD; Eckhard Meisel, MD; Juergen Tebbenjohanns, MD; Lothar Engelmann, MD; Eckart Fleck, MD; Bernd Schubert,

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision

Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System Revision EMILE G. DAOUD, K. CHING MAN, ERED MORADY, and S. ADAM STRIGKBERGER From the Division of

More information

Biphasic Clinical Summaries

Biphasic Clinical Summaries Biphasic Clinical Summaries Defibrillation of Ventricular Fibrillation and Ventricular Tachycardia Background Physio-Control conducted a multi-centered, prospective, randomized and blinded clinical trial

More information

ICD: Basics, Programming and Trouble-shooting

ICD: Basics, Programming and Trouble-shooting ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications

More information

Present Understanding of Shock Polarity for Internal Defibrillation: The Obvious and Non-Obvious Clinical Implications

Present Understanding of Shock Polarity for Internal Defibrillation: The Obvious and Non-Obvious Clinical Implications REVIEWS Present Understanding of Shock Polarity for Internal Defibrillation: The Obvious and Non-Obvious Clinical Implications MARK W. KROLL,* IGOR R. EFIMOV, and PATRICK J. TCHOU From the *California

More information

Tech Corner. ATP in the Fast VT zone

Tech Corner. ATP in the Fast VT zone Tech Corner ATP in the Fast VT zone NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE. FOR ADDITIONAL

More information

The Nuts and Bolts of ICD Therapy

The Nuts and Bolts of ICD Therapy Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction

More information

An Experimental Study of Transvenous Defibrillation Using a Coronary Sinus Catheter

An Experimental Study of Transvenous Defibrillation Using a Coronary Sinus Catheter An Experimental Study of Transvenous Defibrillation Using a Coronary Sinus Catheter ALAN H. KADISH, M.D., KEITH CHILDS, B.S., and JOSEPH LEVINE, M.D. From the Division of Cardiology, Department of Internal

More information

Advanced ICD Concepts

Advanced ICD Concepts 1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation

More information

Effect of a Passive Endocardial Electrode on Defibrillation Efficacy of a Nonthoracotomy Lead System

Effect of a Passive Endocardial Electrode on Defibrillation Efficacy of a Nonthoracotomy Lead System JACC Vol. 29, No. 4 March 15, 1997:825 30 825 Effect of a Passive Endocardial Electrode on Defibrillation Efficacy of a Nonthoracotomy Lead System PARWIS C. FOTUHI, MD, BRUCE H. KENKNIGHT, MS,* SHARON

More information

Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra

Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra Defibrillation testing should be routinely performed at the time of implantable cardioverter-defibrillator implantation Contra Fernando Arribas Cardiology Service Hospital 12 de Octubre Madrid Spain D.O.I.

More information

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs

EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for

More information

Atrial Defibrillation With a Transvenous Lead A Randomized Comparison of Active Can Shocking Pathways

Atrial Defibrillation With a Transvenous Lead A Randomized Comparison of Active Can Shocking Pathways Journal of the American College of Cardiology Vol. 34, No. 2, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00214-4 Atrial

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

Shock Reduction Strategies Michael Geist E. Wolfson MC

Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Reduction Strategies Michael Geist E. Wolfson MC Shock Therapy Thanks, I needed that! Why Do We Need To Reduce Shocks Long-term outcome after ICD and CRT implantation and influence of remote device

More information

Dual-Chamber Implantable Cardioverter-Defibrillator

Dual-Chamber Implantable Cardioverter-Defibrillator February 1998 9 Dual-Chamber Implantable Cardioverter-Defibrillator A.SH. REVISHVILI A.N. Bakoulev Research Center for Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow, Russia Summary

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY COGENT-4 FIELD FOLLOWING STUDY

More information

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis

Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Assessment of Defibrillation Threshold upon Implantable Cardioverter-Defibrillator implant in Relation to patient s prognosis Investigator: Keiko Saito, MD Mentor: Yuji Saito, MD, PhD, FACP, FACC Department

More information

Atrial fibrillation or atrial flutter (AF) may occur after

Atrial fibrillation or atrial flutter (AF) may occur after Relation of Atrial Refractoriness to Upper and Lower Limits of Vulnerability for Atrial Fibrillation/Flutter Following Implantable Ventricular Defibrillator Shocks Amos Katz, MD; Robert J. Sweeney, PhD;

More information

Defibrillation threshold testing should no longer be performed: contra

Defibrillation threshold testing should no longer be performed: contra Defibrillation threshold testing should no longer be performed: contra Andreas Goette St. Vincenz-Hospital Paderborn Dept. of Cardiology and Intensive Care Medicine Germany No conflict of interest to disclose

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Experience with pectoral versus abdominal implantation of a small defibrillator

Experience with pectoral versus abdominal implantation of a small defibrillator European Heart Journal (1998) 19, 185 198 Article No. hj9816 Experience with pectoral versus abdominal implantation of a small defibrillator A multicenter comparison in 778 patients E. Hoffmann, G. Steinbeck,

More information

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention

Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Subcutaneous ICD Emerging Role of Sudden Cardiac Death Prevention Dr Ngai-Yin Chan, MBBS, FRCP(Lond), FRCP(Edin), FRCP(Glasg), FACC, FHRS, Consultant Physician, Department of Medicine & Geriatrics, Princess

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement www.ipej.org 49 Case Report Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement Naga VA Kommuri, MD, MRCPCH, Sri Lakshmi S Kollepara, MD,

More information

New generations pacemakers and ICDs: an update

New generations pacemakers and ICDs: an update Advances in Cardiac Arrhythmias and Great Innovations in Cardiology XXVII Giornate Cardiologiche Torinesi New generations pacemakers and ICDs: an update Prof. Fiorenzo Gaita, MD Division of Cardiology

More information

Tachyarrhythmia Suspicion and Detection

Tachyarrhythmia Suspicion and Detection Tech Corner Tachyarrhythmia Suspicion and Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN THE ARTICLE.

More information

Effects of myocardial infarction on catheter defibrillation threshold

Effects of myocardial infarction on catheter defibrillation threshold Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1983 Effects of myocardial infarction on catheter defibrillation threshold

More information

Azygos Vein Coil: Bailout Strategy for High Defibrillation Thresholds

Azygos Vein Coil: Bailout Strategy for High Defibrillation Thresholds The Journal of Innovations in Cardiac Rhythm Management, 3 (2012), 905 910 INNOVATIVE COLLECTIONS COMPLEX CASE STUDY Azygos Vein Coil: Bailout Strategy for High Defibrillation Thresholds MOUYYAD RAHABY,

More information

Single- versus Dual-Coil ICD Leads: Does it Matter?

Single- versus Dual-Coil ICD Leads: Does it Matter? Single- versus Dual-Coil ICD Leads: Does it Matter? C. W. Israel, M.D. Dept. of Cardiology Evangelical Hospital Bielefeld Germany Carsten.Israel@evkb.de Conflict of Interest Biotronik Boston-Scientific

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Triphasic Waveforms Are Superior to Biphasic Waveforms for Transthoracic Defibrillation Experimental Studies

Triphasic Waveforms Are Superior to Biphasic Waveforms for Transthoracic Defibrillation Experimental Studies Journal of the American College of Cardiology Vol. 42, No. 3, 2003 2003 by the American College of Cardiology Foundation ISSN 0735-1097/03/$30.00 Published by Elsevier Inc. doi:10.1016/s0735-1097(03)00656-9

More information

First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator

First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator Europace (1999) 1, 96 102 First worldwide clinical experience with a new dual chamber implantable cardioverter defibrillator Advantages and complications C. Sticherling, A. Schaumann*, T. Klingenheben

More information

PARAD/PARAD+ : P and R Based Arrhythmia Detection

PARAD/PARAD+ : P and R Based Arrhythmia Detection Tech Corner PARAD/PARAD+ : P and R Based Arrhythmia Detection NOTE: PLEASE NOTE THAT THE FOLLOWING INFORMATION IS A GENERAL DESCRIPTION OF THE FUNCTION. DETAILS AND PARTICULAR CASES ARE NOT DESCRIBED IN

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study ANEESH V. TOLAT, M.D.,* MELISSA WOICIECHOWSKI, M.S.N.,* ROSEMARIE KAHR, R.C.I.S.,* JOSEPH

More information

Review Article Defibrillation Testing of the Implantable Cardioverter Defibrillator: When, How, and by Whom?

Review Article Defibrillation Testing of the Implantable Cardioverter Defibrillator: When, How, and by Whom? www.ipej.org 166 Review Article Defibrillation Testing of the Implantable Cardioverter Defibrillator: When, How, and by Whom? Luis A. Pires, MD, FACC Heart Rhythm Center, Division of Cardiovascular Medicine,

More information

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm

Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm April 2000 107 Clinical Results with the Dual-Chamber Cardioverter Defibrillator Phylax AV - Efficacy of the SMART I Discrimination Algorithm B. MERKELY Semmelweis University, Dept. of Cardiovascular Surgery,

More information

Practice Questions.

Practice Questions. IBHRE Prep Practice Questions Question 1 The relative refractory yperiod of the ventricular myocardium corresponds to which of the following phases of the action potential? A. (0) B. (1) C. (2) D. (3)

More information

The Facts about Biphasic Defibrillation

The Facts about Biphasic Defibrillation The Facts about Biphasic Defibrillation Introduction In reference to the SMART Biphasic waveform, Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, published by the American

More information

Patient-Specific Computational Analysis of Transvenous Defibrillation: A Comparison to Clinical Metrics in Humans

Patient-Specific Computational Analysis of Transvenous Defibrillation: A Comparison to Clinical Metrics in Humans Annals of Biomedical Engineering, Vol. 32, No. 6, June 2004 ( 2004) pp. 775 783 Patient-Specific Computational Analysis of Transvenous Defibrillation: A Comparison to Clinical Metrics in Humans DANIEL

More information

Subcutaneous Implantable Cardioverter Defibrillator (S-ICD)

Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) D. D. MANOLATOS, MD, PhD, FESC Electrophysiology and Device Lab General Hospital Evangelismos, Athens The Problem: 300,000 people die each year

More information

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators

Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators 22 March 2002 Interactive Simulator for Evaluating the Detection Algorithms of Implantable Defibrillators F. HINTRINGER, O. PACHINGER Division of Cardiology, Department for Internal Medicine, University

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

Efficacy and safety of the reciprocal pulse defibrillator current waveform

Efficacy and safety of the reciprocal pulse defibrillator current waveform Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1984 Efficacy and safety of the reciprocal pulse defibrillator current

More information

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Implantable cardioverter defibrillators for the treatment of arrhythmias and cardiac resynchronisation therapy for the treatment of heart failure (review

More information

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD.

Panagiotis N. Margos MD, Rolf Schomburg MD, Jorg Kynast MD, Ahmed A. Khattab MD, Gert Richardt MD. www.ipej.org 64 Case Report Termination of Ventricular Tachycardia with Antitachycardia Pacing after Ineffective Shock Therapy in an ICD Recipient with Hypertrophic Cardiomyopathy Panagiotis N. Margos

More information

An Effective and Adaptable Transvenous Defibrillation System Using The Coronary Sinus in Humans

An Effective and Adaptable Transvenous Defibrillation System Using The Coronary Sinus in Humans lacc Vol. 16, No, 4 887 CARDIAC PACING An Effective and Adaptable Transvenous Defibrillation System Using The Coronary Sinus in Humans GUST H. BARDY, MD, FACC, MARGARET D. ALLEN, MD, RAHUL MEHRA, PHD,

More information

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS

OBJECTIVES BACKGROUND METHODS RESULTS CONCLUSIONS Journal of the American College of Cardiology Vol. 34, No. 5, 1999 1999 by the American College of Cardiology ISSN 0735-1097/99/$20.00 Published by Elsevier Science Inc. PII S0735-1097(99)00363-0 Comparison

More information

Effects of Polarity for Monophasic and Biphasic Shocks on Defihrillation Efficacy with an Endocardiai System

Effects of Polarity for Monophasic and Biphasic Shocks on Defihrillation Efficacy with an Endocardiai System Effects of Polarity for Monophasic and Biphasic Shocks on Defihrillation Efficacy with an Endocardiai System MASAHIRO USUI, GREGORY P. WALCOTT, S. ADAM STRICKBERGER,* DENNIS L. ROLLINS, WILLIAM M. SMITH,

More information

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia

The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia The Therapeutic Role of the Implantable Cardioverter Defibrillator in Arrhythmogenic Right Ventricular Dysplasia By Sandeep Joshi, MD and Jonathan S. Steinberg, MD Arrhythmia Service, Division of Cardiology

More information

Journal of the American College of Cardiology Vol. 37, No. 2, by the American College of Cardiology ISSN /01/$20.

Journal of the American College of Cardiology Vol. 37, No. 2, by the American College of Cardiology ISSN /01/$20. Journal of the American College of Cardiology Vol. 37, No. 2, 2001 2001 by the American College of Cardiology ISSN 0735-1097/01/$20.00 Published by Elsevier Science Inc. PII S0735-1097(00)01133-5 Coronary

More information

SMART Biphasic. A Proven, Fixed, Low-Energy Defibrillation Waveform

SMART Biphasic. A Proven, Fixed, Low-Energy Defibrillation Waveform SMART Biphasic A Proven, Fixed, Low-Energy Defibrillation Waveform INTRODUCTION The 1990s heralded a new era of transthoracic defibrillation in which the rules of conventional practice no longer apply.

More information

Complications Associated With Pectoral Cardioverter-Defibrillator Implantation: Comparison of Subcutaneous and Submuscular Approaches

Complications Associated With Pectoral Cardioverter-Defibrillator Implantation: Comparison of Subcutaneous and Submuscular Approaches 1278 JACC Vol. 28, No. 5 Complications Associated With Pectoral Cardioverter-Defibrillator Implantation: Comparison of Subcutaneous and Submuscular Approaches MICHAEL R. GOLD, MD, PHD, FACC, ROBERT W.

More information

Pacemaker-ICD/Drug Interaction

Pacemaker-ICD/Drug Interaction Review Article Pacemaker-ICD/Drug Interaction Tsu-Juey Wu When a drug is prescribed for a patient with a permanent pacemaker or an implantable cardioverter defibrillator (ICD), consideration must be given

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Call Medtronic at 1 (800) to verify the patient s current implanted system

Call Medtronic at 1 (800) to verify the patient s current implanted system MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient

More information

TASER and In-Custody Deaths

TASER and In-Custody Deaths TASER and In-Custody Deaths Cardiac Effects of Conducted Electrical Weapons Patrick Tchou, MD Heart and Vascular Institute What is a stun gun? A device that delivers rapid electrical impulses that stimulates

More information

Nonthoracotomy Internal Defibrillation in Dogs: Threshold Reduction Using a Subcutaneous Chest Wall Electrode With a Transvenous Catheter Electrode

Nonthoracotomy Internal Defibrillation in Dogs: Threshold Reduction Using a Subcutaneous Chest Wall Electrode With a Transvenous Catheter Electrode 406 lacc Vol. 10, No.2 Nonthoracotomy Internal Defibrillation in Dogs: Threshold Reduction Using a Subcutaneous Chest Wall Electrode With a Transvenous Catheter Electrode JULE N. WETHERBEE, MD, PETER D.

More information

Top 5 Things to Know about Pacemakers and ICD s. Jeffrey S. Osborn, M.D., C.C.D.S. March 4, 2017.

Top 5 Things to Know about Pacemakers and ICD s. Jeffrey S. Osborn, M.D., C.C.D.S. March 4, 2017. Top 5 Things to Know about Pacemakers and ICD s Jeffrey S. Osborn, M.D., C.C.D.S. March 4, 2017. Top 5 Things 1. Remote Monitoring leads to better care and outcomes. 2. MRI s CAN be done on device patients.

More information

Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation

Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation Initial Clinical Experience With Ambulatory Use of an Implantable Atrial Defibrillator for Conversion of Atrial Fibrillation Emile G. Daoud, MD; Carl Timmermans, MD; Chris Fellows, MD; Robert Hoyt, MD;

More information

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure

Implantable cardioverter defibrillator, Inappropriate shock, Lead failure Inappropriate Discharges of Intravenous Implantable Cardioverter Defibrillators Owing to Lead Failure Takashi WASHIZUKA, 1 MD, Masaomi CHINUSHI, 1 MD, Ryu KAZAMA, 1 MD, Takashi HIRONO, 1 MD, Hiroshi WATANABE,

More information

Sudden death as co-morbidity in patients following vascular intervention

Sudden death as co-morbidity in patients following vascular intervention Sudden death as co-morbidity in patients following vascular intervention Impact of ICD therapy Seah Nisam Director, Medical Science, Guidant Corporation Advanced Angioplasty Meeting (BCIS) London, 16 Jan,

More information

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can

EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds

More information

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous

More information

Clinical Data Summary: Avoid FFS Study

Clinical Data Summary: Avoid FFS Study Atrial Pacing Lead with 1.1 mm Tip-to-Ring Spacing Clinical Data Summary: Avoid FFS Study A Multi-center, Randomized, Prospective Clinical Study Designed to Evaluate the 1699T Lead Clinical Data Summary:

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

New scientific documents from EHRA Management of patients with defibrillator shocks

New scientific documents from EHRA Management of patients with defibrillator shocks New scientific documents from EHRA Management of patients with defibrillator shocks Frieder Braunschweig MD PhD FESC Karolinska University Hospital Stockholm, Sweden Evolution of ICD therapy Worldwide

More information

GDT1000 SENSING ACUTE STUDY

GDT1000 SENSING ACUTE STUDY CAUTION: Federal law restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. CLINICAL SUMMARY GDT1000 SENSING ACUTE STUDY Boston

More information

Effect of Shock Thning on Efficacy and Safety Cardioverso on for Ventricular Tachycardia

Effect of Shock Thning on Efficacy and Safety Cardioverso on for Ventricular Tachycardia JAM Vol. 24. No. 3 September 1994 : 70 3 -ti 703 ELECTROPHYSIOLOGY Effect of Shock Thning on Efficacy and Safety Cardioverso on for Ventricular Tachycardia of Internal HUAGUI G. LI, MD, PHD, RAYMOND YEE,

More information

Arrhythmia/Electrophysiology

Arrhythmia/Electrophysiology Arrhythmia/Electrophysiology Ascending-Ramp Biphasic Waveform Has a Lower Defibrillation Threshold and Releases Less Troponin I Than a Truncated Exponential Biphasic Waveform Jian Huang, MD, PhD; Gregory

More information

A Closer Look Product Education at a glance

A Closer Look Product Education at a glance A Closer Look Product Education at a glance an ICD or CRT-D When Atrial Information is Not Used SUMMARY Boston Scientific dual-chamber ICDs and multi-chamber CRT-Ds will respond to atrial sensing regardless

More information

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support

A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support 100 April 2000 A New Dual Chamber Cardioverter-Defibrillator with Left Atrial Pacing Support A. SH. REVISHVILI A.N. Bakoulev Institute of Cardiovascular Surgery, Russian Academy of Medical Sciences, Moscow,

More information

A 'shocking' case rectified

A 'shocking' case rectified A 'shocking' case rectified Journal: Pacing and Clinical Electrophysiology Manuscript ID: PACE-13-00.R1 Wiley - Manuscript type: Device Rounds Date Submitted by the Author: n/a Complete List of Authors:

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial Fibrillation

Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial Fibrillation Novel Passive Implantable Atrial Defibrillator Using Transcutaneous Radiofrequency Energy Transmission Successfully Cardioverts Atrial Fibrillation Ganesh Manoharan, MD; Noel Evans, PhD; Bakthiar Kidwai,

More information

Shock-induced termination of cardiac arrhythmias

Shock-induced termination of cardiac arrhythmias Shock-induced termination of cardiac arrhythmias Group members: Baltazar Chavez-Diaz, Chen Jiang, Sarah Schwenck, Weide Wang, and Jinglei Zhang Cardiac arrhythmias, also known as irregular heartbeat, occur

More information

Internal Cardioversion of Atrial Fibrillation in Sheep

Internal Cardioversion of Atrial Fibrillation in Sheep 1673 Internal Cardioversion of Atrial Fibrillation in Sheep Randolph A.S. Cooper, MD; Clif A. Alferness, BSEE; William M. Smith, PhD; and Raymond E. Ideker, MD, PhD Downloaded from http://ahajournals.org

More information

Shock-induced termination of cardiac arrhythmias

Shock-induced termination of cardiac arrhythmias Shock-induced termination of cardiac arrhythmias Group members: Baltazar Chavez-Diaz, Chen Jiang, Sarah Schwenck, Weide Wang, and Jinglei Zhang Abstract: Cardiac arrhythmias occur when blood flow to the

More information

Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded from an Implantable Defibrillator

Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded from an Implantable Defibrillator Journal of Interventional Cardiac Electrophysiology 9, 49 53, 2003 C 2003 Kluwer Academic Publishers. Manufactured in The Netherlands. Diagnostic Value of Single Versus Dual Chamber Electrograms Recorded

More information

Subcutaneous implantable cardioverter-defibrillator (S-ICD)

Subcutaneous implantable cardioverter-defibrillator (S-ICD) NEW DRUGS AND TECHNOLOGIES IN CARDIOLOGY Cardiology Journal 2011, Vol. 18, No. 3, pp. 326 331 Copyright 2011 Via Medica ISSN 1897 5593 Subcutaneous implantable cardioverter-defibrillator (S-ICD) S. Suave

More information

In the last 2 decades, the implantable cardioverterdefibrillator

In the last 2 decades, the implantable cardioverterdefibrillator 226 Transvenous ICDs in Adolescents and Young Adults Mayo Clin Proc, March 2002, Vol 77 Original Article Role of Transvenous Implantable Cardioverter-Defibrillators in Preventing Sudden Cardiac Death in

More information

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure

Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure Implantable Cardioverter Defibrillator Therapy in MADIT II Patients with Signs and Symptoms of Heart Failure Wojciech Zareba Postinfarction patients with left ventricular dysfunction are at increased risk

More information

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm

Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Recurrent Implantable Defibrillator Discharges (ICD) Discharges ICD Storm Guy Amit, MD, MPH Soroka University Medical Center Ben-Gurion University of the Negev Beer-Sheva, Israel Disclosures Consultant:

More information

ICD leads survival and troubles in the last 10 years

ICD leads survival and troubles in the last 10 years XVI Symposium on Progress in Clinical Pacing, Rome, Italy - Dec 2 to 7, 2014 ICD leads survival and troubles in the last 10 years Antonio Raviele, MD, FESC, FHRS President ALFA Alliance to Fight Atrial

More information

Electrophysiologist-Implanted, Nonthoracotomy-Implantable Cardioverter/Defibrillators

Electrophysiologist-Implanted, Nonthoracotomy-Implantable Cardioverter/Defibrillators 2503 Electrophysiological Laboratory, Electrophysiologist-Implanted, Nonthoracotomy-Implantable Cardioverter/Defibrillators Adam P. Fitzpatrick, MD, MRCP; Michael D. Lesh, MD; Laurence M. Epstein, MD;

More information

EBR Systems, Inc. 686 W. Maude Ave., Suite 102 Sunnyvale, CA USA

EBR Systems, Inc. 686 W. Maude Ave., Suite 102 Sunnyvale, CA USA Over 200,000 patients worldwide are estimated to receive a CRT device each year. However, limitations prevent some patients from benefiting. CHALLENGING PROCEDURE 5% implanted patients fail to have coronary

More information

Destructive Device Removal - Sparks and Deletion of Therapy History From an Implantable Cardioverter Defibrillator. Case Reports

Destructive Device Removal - Sparks and Deletion of Therapy History From an Implantable Cardioverter Defibrillator. Case Reports Case Reports Destructive Device Removal - Sparks and Deletion of Therapy History From an Implantable Cardioverter Defibrillator Takashi Kurita, 1 MD, Shigeyuki Ueda, 2 MD, Hideo Okamura, 2 MD, Takashi

More information

Defibrillation Testing should be routinely performed at the time of ICD implantation. Jeff Healey MD, MSc, FHRS McMaster University

Defibrillation Testing should be routinely performed at the time of ICD implantation. Jeff Healey MD, MSc, FHRS McMaster University Defibrillation Testing should be routinely performed at the time of ICD implantation Jeff Healey MD, MSc, FHRS McMaster University Presenter Disclosure Information Jeff S. Healey, MD, MSc, McMaster University,

More information

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul,

856 Unintentional Deactivation of ICDs Mayo Clin Proc, August 2002, Vol 77 Figure 1. Case 3. Printout from Guidant programmer (Guidant Corp, St Paul, Mayo Clin Proc, August 2002, Vol 77 Unintentional Deactivation of ICDs 855 Case Report Unintentional Deactivation of Implantable Cardioverter-Defibrillators in Health Care Settings MARY JANE RASMUSSEN,

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information