Radiostereometric analysis of hip implants: a critical review of methodology and future directions

Size: px
Start display at page:

Download "Radiostereometric analysis of hip implants: a critical review of methodology and future directions"

Transcription

1 Page 1 of 5 Diagnosis Radiostereometric analysis of hip implants: a critical review of methodology and future directions S Sesselmann *, R Forst, F Tschunko Abstract Introduction Early migration of endoprostheses is predictive for aseptic failure in the long term. Radiostereometric analysis (RSA) is able to measure highly accurate the implant s micromotions and can therefore evaluate the quality of endoprostheses or fixation techniques in a short period of time. A systematic search of the entire PubMed database was carried out. It was limited to publications in English or German published since Inclusion of articles was performed by two examiners who initially read the abstracts, then read full text articles selected from the abstracts and critically summarised the most relevant publications. The aim of this review was to discuss RSA of hip implants. Discussion Due to the current lack of standardisation, there is a deficit of comparability between the results of different RSA studies. In case that no trauma or clinical symptoms of loosening appear, RSA examinations can substitute the annual follow-up Roentgen examinations after joint replacements. Thus, RSA can reduce the patient s radiation exposure. In contrast to other techniques, like conventional radiographs or the Einzel Bild Roentgen analyse, RSA is able to detect early migrations of the implant. The potential of RSA is being * Corresponding author stefansesselmann@web.de Department of Orthopaedic Surgery, Friedrich- Alexander-University of Erlangen-Nuremberg, Rathsberger Street 57, Erlangen, Germany recognised by different regulatory organs. At present, a standard protocol for early preclinical studies of joint implants including RSA is in preparation. Model-based RSA simplifies the method, helping to save time and money. Conclusion Despite some shortcomings of RSA, the potential of the method as a screening device for new joint implants in addition to regulatory and technical developments will facilitate wider application in the future. Introduction Osteoarthritis (OA) and trauma are common indications for the implantation of total joint replacements. OA is a severe disabling disease caused by mechanical factors, cell biological phenomena and a dysregulation of tissue homeostasis 1. Especially in the ageing western societies, the number of OA patients grows. Hence the number of joint replacements increases too and the number of revision surgeries is likely to increase as well. On average, fifteen to twenty years after a primary total hip arthroplasty (THA), revision becomes necessary 2. In many cases, subsequent revision surgery is technically demanding and leading to considerable costs to the public health systems 3. As a consequence, it is worthwhile to avoid implantation of joint prostheses of poor quality and to extend lifespan of hip prostheses. Mechanical loosening of the implant starts with micromotions in the range of mm 4,5. It continues and leads to migrations over larger distances. Radiostereometric analysis (RSA) is able to quantify these early migrations 6,7. In the seventies of the last century, Selvik developed this highly accurate method for the in vivo measurement of musculoskeletal kinematics Today the potency of RSA is to investigate micromotions of a joint implant within a relative short follow-up interval, most commonly two years. Prostheses showing an excessive and even more important a continuous migration in these first two postoperative years are likely to fail in the long term 12,13. Short-term RSA studies can predict the long-term clinical success of new joint implants or fixation techniques. This conclusion enables RSA to be considered as a screening device for novel orthopaedic implant components. The intention of this review is to provide a survey of the potentials and shortcomings of this method and an outlook on possible future developments regarding RSA of joint implants. The focus of this review is on RSA of total hip prostheses as these are the most frequently RSAexamined artificial joints. Methodology A systematic literature research using the PubMed database applying the following search items in the title or abstract was carried out: RSA, radiostereometry, radiostereometric analysis and Roentgen stereophotogrammetric analysis. Results were further filtered using Boolean operators (AND, OR, NOT) and combinations of specific keywords as for example, loosening of joint implants and migration analysis of joint prostheses. The search was limited to publications in English or German published from 01 Jan 1980 to 01 Sep Inclusion of articles was

2 Page 2 of 5 performed by two examiners Stefan Sesselmann and Franz Tschunko who initially read the abstracts then read full text articles selected from the abstracts and later equalised their results. The inclusion criteria set for this review were: Case reports, original research papers, review papers, in vitro/in vivo studies and animal studies on RSA. Exclusion criteria consisted of studies that did not meet the inclusion criteria mentioned above. We reviewed the abstracts of these publications and critically summarised the most relevant of them. The primary electronic search produced 309 articles. Applying inclusion and exclusion criteria, 304 articles were retrieved. After reading the abstracts, respectively the full texts, the two examiners compared their results. This left 32 relevant articles from which results and conclusions were drawn. Discussion The authors have referenced some of their own studies in this review. These referenced studies have been conducted in accordance with the Declaration of Helsinki (1964) and the protocols of these studies have been approved by the relevant ethics committees related to the institution in which they were performed. All human subjects, in these referenced studies, gave informed consent to participate in these studies. One of the shortcomings of RSA is the fact that neither the radiation doses nor the length of follow-up intervals for RSA examinations are standardized and they vary between different studies and research centres. The radiation doses of RSA images are lower than those of corresponding conventional radiographs, as they are exposed with higher voltage and lower amperage 14. As often reported, following surgery, several conventional radiographs are taken during the first year and later on annually to follow up the joint implant as long as no trauma or symptoms of loosening appear. Performing RSA studies, radiographs are commonly scheduled up to 1 week after the operation, then at 6 weeks, 3 and 6 months, 1, 2 and possibly 5 and 10 years postoperatively 14. In case that no trauma or clinical symptoms of loosening appear, RSA examinations can substitute the regular follow-up Roentgen examinations. Thus, RSA can reduce the patient s radiation exposure. The lack of comparability of results in between different RSA studies is an important problem of RSA. As mentioned above, the radiation doses and the follow-up intervals are not standardized. Furthermore, there are different numbers and sizes of marker beads or methods of image acquisition in use. In most reports, translations and rotations are measured at the implant s centre of gravity, whereas some study groups use a different technique, measuring migration for each single marker attached to the prosthesis. An important topic is the need of standardisation of statistics. In 2005, Valstar et al. stated that, for every RSA study, accuracy and precision of the arrangement should be presented 14. Precision usually assessed by double examinations is the basis of power calculations. It should be assessed for each model-based RSA study. Due to higher radiation doses per patient, in many cases the local ethics committees often restrict these double examinations. Unfortunately, this lack of information about the study s precision might lead to false conclusions drawn from this examination 14. To determine accuracy, different cadaver models have been used, various phantoms have been designed and several methods for calculating accuracy have been proposed, resulting in divergent values obtained 15,16. There are many RSA reports that quote accuracy values; however, they do not explain how these values were obtained. Thus, it becomes difficult to compare the results of different studies (Figure 1). Therefore, it is important to establish standards for RSA examinations to guarantee comparability of results in between different studies. In 2005, Valstar et al. Published Guidelines for standardisation of radiostereometry Figure 1: Radiostereometric analysis (RSA) arrangement. (A) Positions of patient, cage and Roentgen tubes. The joint prosthesis is focussed by both X-ray tubes. (B) Example of a uniplanar cage. Figure modified after Valstar et al. 14.

3 Page 3 of 5 of implants to which all RSA study groups should adhere to 14. The markers used for RSA usually consist of tantalum. Tantalum is an element with a high atomic number. It is radiopaque and easy to identify on radiographs. In previous studies, no serious tissue reactions were observed 17,18. In the past decades, tantalum has been used as a bone marker in thousands of patients and no side effects were known 9. Nevertheless, further studies of the marker s biocompatibility should be accomplished in in vivo studies to affirm previous observations and to achieve reliable results (Figure 2). As already mentioned, mechanical loosening starts with implant migration of mm relative to the bone. Normally, the prosthesis does not cause any clinical symptoms showing migration values in the mentioned range. In clinical practise, joint prostheses are assessed in standard radiographs. The accuracy of migration measurements in those standard radiographs of total hip arthroplasties ranges between 5 and12 mm 19. Due to different technical improvements in assessing these standard radiographs, the so-called singleimage Roentgen analysis [Einzel Bild Roentgen analyse (EBRA)] reaches an accuracy of up to 1 mm 20. Hence the accuracy of both techniques is not sufficient to detect early stages of implant loosening. With a reported accuracy ranging between 0.05 and 0.5 mm for translational movements and 0.15 to 1.15 for rotational movements, the RSA technique is able to detect these relevant early migrations of the joint prostheses 9. The method s high accuracy depends on several factors, such as the radiographic technique, the RSA analysis software package used or the positioning of the tantalum spheres. Since the association between early migration and longterm loosening of joint replacements has been proven by various studies, RSA is able to evaluate the quality of endoprostheses in a short period of time 12,13. As another consequence of the method s extraordinary accuracy, only a small sample size of patients is required to obtain reliable results 14. Further advantage of RSA is the possibility to measure migrations in three dimensions. This is of particular interest since the loosening process of implants seems to consist of a complex combination of translational and rotational movements 21. In 1990, the bone cement Boneloc (Polymers Reconstructive, Farum, Denmark) was brought onto the market. The lower curing temperature and its decreased release of toxic monomers were mentioned to lead to a decrease of aseptic loosening of implants. During the first years after its introduction, only experimental Figure 2: Marker-based radiostereometric analysis (RSA) radiograph. A total hip prosthesis in situ with Tantalum markers at the collar, tip and shoulder (blue), markers inserted into the femur (red) and cage markers (green and yellow). Centre of gravity (A) 32.

4 Page 4 of 5 studies were published most of them recommending the product. Five years later, radiostereometric migration analyses and registerbased articles were published. Most of them reported inferior results. The average revision rate in sample-based studies exceeded the reference value in the Norwegian Arthroplasty Register 7.35-fold. When Boneloc was taken from the market in April 1995, units of Boneloc bone cement had been in use 22,23. The Capital Hip (3M Health Care Ltd., London borough, United Kingdom) a total hip replacement was released in the market in About 5000 patients received a Capital Hip. Five years after implantation, 20% of these prostheses had to be revised 24,25. The potency of RSA is to measure micromotions of joint prostheses thus also assessing fixation techniques within a short period of time and with low patient numbers. If all newly developed implants and fixation techniques were studied by RSA before definitely bringing them onto the open market, excessively migrating prostheses and inferior fixation techniques could be identified and patients would be prevented from further disasters. In case of Boneloc or the Capital Hip, only a few patients would have been exposed to the hazard of these products in RSA studies. Future directions In 1994, Freeman and Plante-Bordeneuve 26 already considered RSA as a screening method for new joint prostheses. They gave the opinion, that every newly designed femoral prosthesis should be monitored by using RSA to measure its migration rate in a small number of patients for two years. Thus, excessively migrating implants could be detected and abandoned 26. Nelissen et al. 27 compared the fiveyear revision rate of RSA-tested total knee prostheses with that of non- RSA-tested knee prostheses. In 2011, they concluded that RSA studies with a two-year follow-up of new joint implants would lead to a better patient care, a reduction of cases of revision surgery and therefore also a noticeable reduction of costs associated with those revision arthroplasties. In particular, they recommended combining RSA studies with postmarket surveillance in national joint registries to evaluate the quality of implants 27. The potential role of RSA as a method of assessment of joint implants before their release to the open market is being recognised by different regulatory organs. If authorities demanded reliable premarketing examinations of orthopaedic components like hip or knee prostheses, RSA studies with a follow-up interval of two years could be an alternative to long-term clinical studies. For example, the NICE (National Institute for Health and Clinical Excellence, London, United Kingdom) guidelines advise RSA as a method to achieve suitable clinical data 28. At present, the International Organization for Standardisation (ISO) and the European Standards Working Group on Joint Replacement Implants prepare a standard protocol for early preclinical studies. This protocol will include RSA helping to avoid the implantation of failing endoprostheses in large quantities. This would mean a further revaluation of the RSA method 14. Model-based RSA utilises bone markers and a three-dimensional surface model of the implant to compute the in vivo migration of prostheses29. In contrast to markerbased RSA, no tantalum markers have to be placed on the implant. As no special implants have to be manufactured and no separate certification is needed, the model-based technique can help save time and costs. Compared to marker-based RSA, performing model-based methods leads to a negligible lower data quality 29. The use of model-based RSA may facilitate a wider application of the RSA method. Efforts have been made to establish a completely markerless RSA. This would open the possibility of performing migration and wear measurements on larger groups of patients in clinical follow-ups 30. At the moment, accuracy and precision of the completely markerless approach are lower than that of marker-based RSA to a degree which precludes the use of this method for measuring implant migration in its present form 31. To carry out a markerless RSA of an orthopaedic implant, the examined joint has to be CT scanned prior to the implantation of the prosthesis and further CT follow-up examinations have to be performed after surgery. Thus, a markerless RSA study entails a significantly higher radiation exposure compared to marker-based and model-based RSA studies. Performing RSA requires an expensive special RSA-software, a second Roentgen tube and specially trained personnel. Furthermore, carrying out an RSA study is time consuming. These high requirements possibly prevent RSA from being used in clinical routine 32. Conclusion At present, there are still some downsides of RSA, such as a lack of standardisation. Despite some shortcomings, the high accuracy of RSA to measure micromotions of a joint implant in three dimensions and its ability to identify excessively migrating prostheses in a short period of time, using a small sample size are well-known facts. In addition, the efforts to standardize RSA, the prospect that regulatory organs will demand preclinical RSA studies of new orthopaedic implant components, and the progress provided by the current technical developments of the method might lead to a wider use of RSA in the future. Abbreviations list EBRA, Einzel Bild Roentgen analyse; ISO, International Organization for Standardisation; NICE, National

5 Page 5 of 5 Institute for Health and Clinical Excellence; OA, osteoarthritis; RSA, radiostereometric analysis; THA, total hip arthroplasty References 1. Sesselmann S, Söder S, Voigt R, Haag J, Grogan SP, Aigner T. DNA methylation is not responsible for p21waf1/cip1 down-regulation in osteoarthritic chondrocytes. Osteoarthritis Cartil Apr;17(4): Garellick G, Kärrholm J, Rogmark C, Herberts P. Swedish Hip Arthroplasty Register, annual report Sweden: Department of Ortopaedics, Sahlgrenska University Hospital. 3. Lubbeke A, Katz JN, Perneger TV, Hoffmeyer P. Primary and revision hip arthroplasty: 5-year outcomes and influence of age and comorbidity. J Rheumatol. 2007;34(2): Pilliar RM, Lee JM, Maniatopoulos C. Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Jul;208: Soballe K, Hansen ES, Rasmussen HB, Bunger C. Hydroxyapatite coating converts fibrous tissue to bone around loaded implants. J Bone Joint Surg Br. 1993;75-B(2): Sundfeldt M, Carlsson LV, Johansson CB, Thomsen P, Gretzer C. Aseptic loosening, not only a question of wear: a review of different theories. Acta Orthop. 2006;12(2): Valstar ER, Gill HS. Radiostereometric analysis in orthopaedic surgery: editorial comment. Clin Orthop Relat Res Jul;448:2. 8. Selvik G. Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand. 60, 1989(Suppl. 232): Kärrholm J. Roentgen stereophotogrammetry. of orthopedic applications. Acta Orthop Scand. 1989;12(4): Kärrholm J, Gill RH, Valstar ER. The history and future of radiostereometric analysis. Clin Orthop Relat Res Jul;448: Valstar ER, Nelissen RG, Reiber JH, Rozing PM. The use of Roentgen sterophotogrammetry to study micromotion of orthopaedic implants. ISPRS J Photogramm Remote Sens Aug;56(5 6) : Kärrholm J, Borssen B, Löwenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4 7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br. 1994;76(6): Ryd L, Albrektsson BE, Carlsson L, Dansgård F, Herberts P, Lindstrand A, et al. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg Br. 1995;77(3): Valstar ER, Gill R, Ryd L, Flivik G, Börlin N, Kärrholm J. Guidelines for standardisation of radiostereometry (RSA) of implants. Acta Orthop. 2005;76(4): Önsten I, Berzins A, Shott S, Sumner DR. Accuracy and precision of radiostereometric analysis in the measurement of THR femoral component translations: human and canine in vitro models. J Orthop Res Nov;19(6): Bragdon CR, Malchau H, Yuan X, Perinchief R, Kärrholm J, Börlin N, et al. Experimental assessment of precision and accuracy of radiostereometric analysis for the determination of polyethylene wear in a total hip replacement model. J Orthop Res Jul;20(4): Alberius P. Bone reactions to tantalum markers. A scanning electron microscopic study. Acta Anat. (Basel) 1983;115(4): Aronson AS, Jonsson N, Alberius P. Tantalum markers in radiography. An assessment of tissue reactions. Skeletal Radiol. 1985;14(3): Malchau H, Kärrholm J, Wang YX, Herberts P. Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients. Acta Orthop Scand Oct;66(5): Krismer M, Tschupik JP, Bauer R, Mayrhofer P, Stöckl B, Fischer M, et al. Single-image Roentgen analysis for the measurement of hip endoprosthesis migration. Orthopäde Mar;26(3): Kiss J, Murray DW, Turner-Smith AR, Bithell J, Bulstrode CJ. Migration of cemented femoral components after THR. Roentgen stereophotogrammetric analysis. J Bone Joint Surg Br Sep;78(5): Havelin LI, Espehaug B, Vollset SE, Engesaeter LB. The effect of the type of cement on early revision of Charnley total hip prostheses. A review of eight thousand five hundred and seventy-nine primary arthroplasties from the Norwegian Arthroplasty Register. J Bone Joint Surg Am Oct;77(10): Liebensteiner M, Janda W, Williams A, Pawelka W, Labek G. Detection of inferior products in arthroplasty and implementation of findings: a retrospective analysis of the Boneloc incident. Z Orthop Unfall Nov Dec;147(6): Hazard Notice. Medical devices agency. London: MDA; 1998 [MDA9801]. 25. Muirhead-Allwood SK. Lessons of a hip failure. Br Med J Feb; 316(7132): Freeman MA, Plante-Bordeneuve P. Early migration and late aseptic failure of proximal femoral prostheses. J Bone Joint Surg Br. 1994;76(3): Nelissen RG, Pijls BG, Kärrholm J, Malchau H, Nieuwenhuijse MJ, Valstar ER. RSA and registries: the quest for phased introduction of new implants. J Bone Joint Surg Am Dec;93 (Suppl. 3): NHS National Institute for Clinical Excellence. Guidance on the selection of prostheses for primary total hip replacement; nicemedia/pdf/guidance_on_the_selection_of_hip_prostheses.pdf. [last accessed 06 Sep 2013]. 29. Seehaus F, Emmerich J, Kaptein BL, Windhagen H, Hurschler C. Experimental analysis of model-based Roentgen stereophotogrammetric analysis (MBRSA) on four typical prosthesis components. J Biomech Eng Apr;131(4): Börlin N, Röhrl SM, Bragdon CR. RSA wear measurements with or without markers in total hip arthroplasty. J Biomech. 2006;39(9): Seehaus F, Olender GD, Kaptein BL, Ostermeier S, Hurschler C. Markerless Roentgen stereophotogrammetric analysis for in vivo implant migration measurement using three dimensional surface models to represent bone. J Biomech. 2012;12(8): Wierer T, Forst R, Mueller LA, Sesselmann S. Radiostereometric migration analysis of the Lubinus SP II hip stem: 59 hips followed for 2 years. Biomed Tech (Berl) Aug;58(4):

Guidelines for standardization of radiostereometry (RSA) of implants

Guidelines for standardization of radiostereometry (RSA) of implants Acta Orthopaedica ISSN: 1745-3674 (Print) 1745-3682 (Online) Journal homepage: http://www.tandfonline.com/loi/iort20 Guidelines for standardization of radiostereometry (RSA) of implants Edward R Valstar,

More information

LEVERAGING EXISTING C-ARMS FOR RSA ANALYSIS

LEVERAGING EXISTING C-ARMS FOR RSA ANALYSIS LEVERAGING EXISTING C-ARMS FOR RSA ANALYSIS Vivian WJ Chung 1,2 *, Robyn Newell PhD 1,2, Angela Kedgley PhD 3, Bassam A. Masri MD 2,4, Shahram Amiri PhD 1,2,4, Antony Hodgson PhD 1,2 1 * Dept. of Mech.

More information

The optimal timing of baseline radiostereometric analysis of uncemented press fit cups

The optimal timing of baseline radiostereometric analysis of uncemented press fit cups Scandinavian Journal of Surgery 99: 244 249, 2010 The optimal timing of baseline radiostereometric analysis of uncemented press fit cups O. Wolf, J. Milbrink, S. Larsson, P. Mattsson, H. Mallmin Department

More information

Key messages Radiostereometric Analysis (RSA) is validated to predict long-term revision rates within two years post-operation in

Key messages Radiostereometric Analysis (RSA) is validated to predict long-term revision rates within two years post-operation in 55.200BJR0010.1302/2046-3758.56.2000568 research-article2016 Freely available online open Access BJR Hip Surrogate markers of long-term outcome in primary total hip arthroplasty a systematic review T.

More information

ARTICLE IN PRESS. Journal of Biomechanics

ARTICLE IN PRESS. Journal of Biomechanics Journal of Biomechanics 42 (2009) 2387 2393 Contents lists available at ScienceDirect Journal of Biomechanics journal homepage: www.elsevier.com/locate/jbiomech www.jbiomech.com Implementation and validation

More information

Anne Jacobsen 1, Frank Seehaus 1, Yutong Hong 1, Han Cao 1, Alexander Schuh 2, Raimund Forst 1 and Stefan Sesselmann 1,3*

Anne Jacobsen 1, Frank Seehaus 1, Yutong Hong 1, Han Cao 1, Alexander Schuh 2, Raimund Forst 1 and Stefan Sesselmann 1,3* Jacobsen et al. BMC Musculoskeletal Disorders (2018) 19:335 https://doi.org/10.1186/s12891-018-2259-4 RESEARCH ARTICLE Open Access Model-based roentgen stereophotogrammetric analysis using elementary geometrical

More information

Is Model-based Radiostereometric Analysis Suitable for Clinical Trials of a Cementless Tapered Wedge Femoral Stem?

Is Model-based Radiostereometric Analysis Suitable for Clinical Trials of a Cementless Tapered Wedge Femoral Stem? Clin Orthop Relat Res (2016) 474:2246 2253 DOI 10.1007/s11999-016-4930-0 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons BASIC RESEARCH Is Model-based

More information

Early Subsidence Predicts Failure of a Cemented Femoral Stem With Minor Design Changes

Early Subsidence Predicts Failure of a Cemented Femoral Stem With Minor Design Changes Clin Orthop Relat Res (2016) 474:2221 2229 DOI 10.1007/s11999-016-4884-2 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons SYMPOSIUM: 2015 INTERNATIONAL

More information

Early migration of tibial components is associated with late revision

Early migration of tibial components is associated with late revision Acta Orthopaedica 2012; 83 (Id.no 5477) Early migration of tibial components is associated with late revision A systematic review and meta-analysis of 21,000 knee arthroplasties Bart G Pijls 1, Edward

More information

We studied the migration of 58 cemented Hinek

We studied the migration of 58 cemented Hinek MIGRATION OF CEMENTED FEMORAL COMPONENTS AFTER THR ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS J. KISS, D. W. MURRAY, A. R. TURNER-SMITH, J. BITHELL, C. J. BULSTRODE From the Nuffield Orthopaedic Centre, Oxford,

More information

Does a new knee design perform as well as the design it replaces?

Does a new knee design perform as well as the design it replaces? Freely available online KNEE Does a new knee design perform as well as the design it replaces? M. Molt, P. Ljung, S. Toksvig-Larsen From Hässlehom Hospital, Hässleholm, Sweden Objectives The objective

More information

Prospective five-year subsidence analysis of a cementless fully hydroxyapatite-coated femoral hip arthroplasty component

Prospective five-year subsidence analysis of a cementless fully hydroxyapatite-coated femoral hip arthroplasty component Hip Int 2014; 24 ( 1) : 91-97 DOI: 10.5301/hipint.5000082 ORIGINAL ARTICLE open access Prospective five-year subsidence analysis of a cementless fully hydroxyapatite-coated femoral hip arthroplasty component

More information

CLINICAL PAPER / ORTHOPEDIC

CLINICAL PAPER / ORTHOPEDIC HIP LEG LENGTH AND OFFSET Kelley T.C. and Swank M.L. (2009) Using CAS leads to more accurate positioning within the safe zone (inclination between 30 and 50, anteversion between 5 and 25 ) CAS improves

More information

Low revision rate after total hip arthroplasty in patients with pediatric hip diseases

Low revision rate after total hip arthroplasty in patients with pediatric hip diseases 6 Acta Orthopaedica 2012; 83 (5): x x 1 Low revision rate after total hip arthroplasty in patients with pediatric hip diseases Evaluation of 14,403 THAs due to DDH, SCFE, or Perthes disease and 288,435

More information

Evidence based introduction of orthopaedic implants. Evidence based introduction. Bart Pijls. Bart Pijls

Evidence based introduction of orthopaedic implants. Evidence based introduction. Bart Pijls. Bart Pijls Evidence based introduction of orthopaedic implants RSA, implant quality and patient safety Uitnodiging Evidence based introduction Evidence based Evidence introduction basedimplants introduction of orthopaedic

More information

The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplasties

The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplasties Acta Orthopaedica Scandinavica ISSN: 1-647 (Print) (Online) Journal homepage: http://www.tandfonline.com/loi/iort19 The Norwegian Arthroplasty Register: 11 years and 73, arthroplasties Leif I Havelin,

More information

Early Migration Predicts Aseptic Loosening of Cementless Femoral Stems: A Long-term Study

Early Migration Predicts Aseptic Loosening of Cementless Femoral Stems: A Long-term Study Clin Orthop Relat Res (2016) 474:1697 1706 DOI 10.1007/s11999-016-4857-5 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons CLINICAL RESEARCH Early Migration

More information

Hofteregisteret: god samfunnsokonomi [The hip registry: good economy for society] Engesaeter L B, Furnes A, Havelin L I, Lie S A, Vollset S E

Hofteregisteret: god samfunnsokonomi [The hip registry: good economy for society] Engesaeter L B, Furnes A, Havelin L I, Lie S A, Vollset S E Hofteregisteret: god samfunnsokonomi [The hip registry: good economy for society] Engesaeter L B, Furnes A, Havelin L I, Lie S A, Vollset S E Record Status This is a critical abstract of an economic evaluation

More information

Cover Page. The handle holds various files of this Leiden University dissertation.

Cover Page. The handle  holds various files of this Leiden University dissertation. Cover Page The handle http://hdl.handle.net/1887/23022 holds various files of this Leiden University dissertation. Author: Pijls, Bart Godefridus Catharina Wilhelmus Title: Evidence based introduction

More information

Kiernan, Sverrir; Hermann, K L; Wagner, Philippe; Ryd, Leif; Flivik, Gunnar

Kiernan, Sverrir; Hermann, K L; Wagner, Philippe; Ryd, Leif; Flivik, Gunnar The importance of adequate stem anteversion for rotational stability in cemented total hip replacement: A radiostereometric study with ten-year follow-up. Kiernan, Sverrir; Hermann, K L; Wagner, Philippe;

More information

The effect of a lateral flare feature on implant stability

The effect of a lateral flare feature on implant stability International Orthopaedics DOI 10.1007/s00264-002-0355-3 Original Paper The effect of a lateral flare feature on implant stability Alex Leali( ) Joseph Fetto Harvey Insler David Elfenbein A. Leali J. Fetto

More information

The effect of periapatite on the micromotion of total knee arthroplasty tibial components in osteoarthritis

The effect of periapatite on the micromotion of total knee arthroplasty tibial components in osteoarthritis The effect of periapatite on the micromotion of total knee arthroplasty tibial components in osteoarthritis A controlled prospective, randomized RSA study in 90 patients Mathys J.A. van der Linde 1, Eric

More information

A cementless, elastic press-fit socket with and without screws

A cementless, elastic press-fit socket with and without screws Acta Orthopaedica 2012; 83 (5): 481 487 481 A cementless, elastic press-fit socket with and without screws A 2-year randomized controlled radiostereometric analysis of 37 hips Dean Pakvis 1, Joan Luites

More information

Effects of hydroxyapatite coating of cups used in hip revision arthroplasty

Effects of hydroxyapatite coating of cups used in hip revision arthroplasty Acta Orthopaedica 2012; 83 (5): 427 435 427 Effects of hydroxyapatite coating of cups used in hip revision arthroplasty A Swedish Hip Arthroplasty Register study of 1,780 revisions Stergios Lazarinis 1,

More information

Early subsidence of shape-closed hip arthroplasty stems is associated with late revision

Early subsidence of shape-closed hip arthroplasty stems is associated with late revision Acta Orthopaedica ISSN: 1745-3674 (Print) 1745-3682 (Online) Journal homepage: http://www.tandfonline.com/loi/iort20 Early subsidence of shape-closed hip arthroplasty stems is associated with late revision

More information

Journal of Orthopaedic Research

Journal of Orthopaedic Research ELSEVIER Journal of Orthopaedic Research 20 (1002) 688-695 Journal of Orthopaedic Research www.elsevier.corn/locate/orthres Experimental assessment of precision and accuracy of radiostereometric analysis

More information

Total hip replacement with a zirconium oxide ceramic femoral head

Total hip replacement with a zirconium oxide ceramic femoral head Total hip replacement with a zirconium oxide ceramic femoral head A RANDOMISED ROENTGEN STEREOPHOTOGRAMMETRIC STUDY T. von Schewelov, L. Sanzén, I. Önsten, Å. Carlsson, J. Besjakov From Lund University

More information

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty

Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty The Journal of Arthroplasty Vol. 23 No. 2 2008 Developmental Dysplasia of the Hip Good Results of Later Total Hip Arthroplasty 7135 Primary Total Hip Arthroplasties after Developmental Dysplasia of the

More information

Is reverse hybrid hip replacement the solution?

Is reverse hybrid hip replacement the solution? Acta Orthopaedica 2011; 82 (6): 639 645 639 Is reverse hybrid hip replacement the solution? 3,963 primary hip replacements with cemented cup and uncemented stem, from the Norwegian Arthroplasty Register

More information

Preoperative gait patterns and BMI are associated with tibial component migration

Preoperative gait patterns and BMI are associated with tibial component migration 478 Acta Orthopaedica 2010; 81 (4): 478 486 Preoperative gait patterns and BMI are associated with tibial component migration Janie L Astephen Wilson 1, David A J Wilson 1, Michael J Dunbar 1,2, and Kevin

More information

RM Philosophy Innovation based on clinical evidence.

RM Philosophy Innovation based on clinical evidence. RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher

More information

Dual Mobility Cups. Kris Govaers, MD, PhD Dendermonde Belgium

Dual Mobility Cups. Kris Govaers, MD, PhD Dendermonde Belgium Dual Mobility Cups Kris Govaers, MD, PhD Dendermonde Belgium Introduction Indications Limitations Conclusions All good things in life come from France Gilles Bousquet 1979 + Rambert (SERF) Inventions Bousquet

More information

why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS

why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS why bicompartmental? A REVOLUTIONARY ALTERNATIVE TO TOTAL KNEE REPLACEMENTS TKR is not always the answer Today, many patients with medial or lateral disease and patellofemoral involvement receive a Total

More information

Validation of static and dynamic radiostereometric analysis of the knee joint using bone-models from CT data

Validation of static and dynamic radiostereometric analysis of the knee joint using bone-models from CT data Aalborg Universitet Validation of static and dynamic radiostereometric analysis of the knee joint using bone-models from CT data Stentz-Olesen, Kasper; Nielsen, Emil Toft; de Raedt, Sepp; Jørgensen, Peter

More information

Hvilken evidens har vi? - hvad praktiserer vi?

Hvilken evidens har vi? - hvad praktiserer vi? Symposium, DOS Kongressen 25. oktober 2012, København Afspejler vores kliniske praksis den bedste nuværende evidens? Hvilken evidens har vi? - hvad praktiserer vi? Henrik Malchau, MD, PhD Professor at

More information

The Finnish Arthroplasty Register: Report of the hip register

The Finnish Arthroplasty Register: Report of the hip register Acta Orthopaedica Scandinavica ISSN: 0001-64 (Print) (Online) Journal homepage: http://www.tandfonline.com/loi/iort19 The Finnish Arthroplasty Register: Report of the hip register Timo J S Puolakka, K

More information

Clinical Study The Influence of Tribological Pairings and Other Factors on Migration Patterns of Short Stems in Total Hip Arthroplasty

Clinical Study The Influence of Tribological Pairings and Other Factors on Migration Patterns of Short Stems in Total Hip Arthroplasty Hindawi BioMed Research International Volume 7, Article ID 876, 9 pages https://doi.org/./7/876 Clinical Study The Influence of Tribological Pairings and Other Factors on Migration Patterns of Short Stems

More information

MEDICAL POLICY SUBJECT: COMPUTER ASSISTED NAVIGATION FOR KNEE AND HIP ARTHROPLASTY

MEDICAL POLICY SUBJECT: COMPUTER ASSISTED NAVIGATION FOR KNEE AND HIP ARTHROPLASTY MEDICAL POLICY PAGE: 1 OF: 5 If the member's subscriber contract excludes coverage for a specific service it is not covered under that contract. In such cases, medical policy criteria are not applied.

More information

Better survival of hybrid total knee arthroplasty compared to cemented arthroplasty

Better survival of hybrid total knee arthroplasty compared to cemented arthroplasty Acta Orthopaedica 2015; 86 (6): x x 1 Better survival of hybrid total knee arthroplasty compared to cemented arthroplasty A report from the Norwegian Arthroplasty Register 1999 2012 Gunnar PETURSSON 1,2,

More information

Incidence of total hip replacement for primary osteoarthrosis in Iceland

Incidence of total hip replacement for primary osteoarthrosis in Iceland Acra Orthop Scand 999; 70 (3): 229-233 229 Incidence of total hip replacement for primary osteoarthrosis in Iceland 982-996 Thorvaldur Ingvarsson~2, Gunnar Hagglund2, Halldor Jonsson J6 and L Stefan Lohmander2

More information

Haukeland University Hospital, Bergen

Haukeland University Hospital, Bergen This article was downloaded by:[universitetsbiblioteket i Bergen] On: 2 November 2007 Access Details: [subscription number 768372013] Publisher: Informa Healthcare Informa Ltd Registered in England and

More information

Modern Cementing Technique

Modern Cementing Technique Modern Cementing Technique PRESSURIZATION BONE CEMENT BONE BED PREPARATION MIXING AND DELIVERY Mixing and Delivery Mixing and collecting the cement under vacuum reduces both micro and macro pores to a

More information

Primary stability of a shoulderless Zweymüller hip stem: a comparative in vitro micromotion study

Primary stability of a shoulderless Zweymüller hip stem: a comparative in vitro micromotion study Bieger et al. Journal of Orthopaedic Surgery and Research (2016) 11:73 DOI 10.1186/s13018-016-0410-1 RESEARCH ARTICLE Open Access Primary stability of a shoulderless Zweymüller hip stem: a comparative

More information

DePuy CMW Heritage Bone Cements. Product Information

DePuy CMW Heritage Bone Cements. Product Information DePuy CMW Heritage Bone Cements Product Information INTRODUCTION Our experience in Polymethyl methacrylate (PMMA) bone cements is extensive. The long clinical heritage of our DePuy CMW bone cement product

More information

Prosthetic liner wear in total hip replacement: a longitudinal 13-year study with computed tomography

Prosthetic liner wear in total hip replacement: a longitudinal 13-year study with computed tomography Skeletal Radiology (2018) 47:883 887 https://doi.org/10.1007/s00256-018-2878-8 CASE REPORT Prosthetic liner wear in total hip replacement: a longitudinal 13-year study with computed tomography Lars Weidenhielm

More information

Aalborg Universitet. Published in: Bone & Joint Research. DOI (link to publication from Publisher): / BJR

Aalborg Universitet. Published in: Bone & Joint Research. DOI (link to publication from Publisher): / BJR Aalborg Universitet Validation of static and dynamic radiostereometric analysis of the knee joint using bone-models from CT data Stentz-Olesen, Kasper; Nielsen, Emil Toft; de Raedt, Sepp; Jørgensen, Peter

More information

Countrywise results of total hip replacement: An analysis of 438,733 hips based on the Nordic Arthroplasty Register Association database

Countrywise results of total hip replacement: An analysis of 438,733 hips based on the Nordic Arthroplasty Register Association database Countrywise results of total hip replacement: An analysis of 438,733 hips based on the Nordic Arthroplasty Register Association database The Harvard community has made this article openly available. Please

More information

MEDICAL POLICY SUBJECT: COMPUTER ASSISTED NAVIGATION FOR KNEE AND HIP ARTHROPLASTY

MEDICAL POLICY SUBJECT: COMPUTER ASSISTED NAVIGATION FOR KNEE AND HIP ARTHROPLASTY MEDICAL POLICY SUBJECT: COMPUTER ASSISTED PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including an

More information

A further enhanced classic. Wagner SL Revision Hip Stem

A further enhanced classic. Wagner SL Revision Hip Stem A further enhanced classic Wagner SL Revision Hip Stem The original Wagner SL Revision Stem offers a time-proven solution in the treatment of revision hips. While its underlying anchorage philosophy and

More information

Operating time and survival of primary total hip replacements

Operating time and survival of primary total hip replacements 524 Acta Orthop Scand 2004; 75 (5): 524 532 Operating time and survival of primary total hip replacements An analysis of 31 745 primary cemented and uncemented total hip replacements from local hospitals

More information

Outcomes Following Total Hip Arthroplasty: A Review of the Registry Data

Outcomes Following Total Hip Arthroplasty: A Review of the Registry Data Symposium - Total Hip Arthroplasty Outcomes Following Total Hip Arthroplasty: A Review of the Registry Data Abstract While total hip arthroplasty remains one of the most reliable procedures with excellent,

More information

The effect of middle-age body weight and physical activity on the risk of early revision hip arthroplasty: A cohort study of 1,535 individuals

The effect of middle-age body weight and physical activity on the risk of early revision hip arthroplasty: A cohort study of 1,535 individuals Acta Orthopaedica ISSN: 1745-3674 (Print) 1745-3682 (Online) Journal homepage: http://www.tandfonline.com/loi/iort20 The effect of middle-age body weight and physical activity on the risk of early revision

More information

FEP Medical Policy Manual

FEP Medical Policy Manual FEP Medical Policy Manual Effective Date: October 15, 2018 Related Policies: None Three-Dimensional Printed Orthopedic Implants Description This evidence review addresses orthopedic implants that are constructed

More information

Adherence of hip and knee arthroplasty studies to RSA standardization guidelines: A systematic review

Adherence of hip and knee arthroplasty studies to RSA standardization guidelines: A systematic review Adherence of hip and knee arthroplasty studies to RSA standardization guidelines: A systematic review The Harvard community has made this article openly available. Please share how this access benefits

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Increasing risk of prosthetic joint infection after total hip arthroplasty

Increasing risk of prosthetic joint infection after total hip arthroplasty Acta Orthopaedica ISSN: 1745-3674 (Print) 1745-3682 (Online) Journal homepage: http://www.tandfonline.com/loi/iort20 Increasing risk of prosthetic joint infection after total hip arthroplasty Håvard Dale,

More information

MODEL-BASED RADIOSTEREOMETRIC ANALYSIS OF AN UNCEMENTED MOBILE-BEARING TOTAL ANKLE ARTHROPLASTY SYSTEM. Jason W. Fong

MODEL-BASED RADIOSTEREOMETRIC ANALYSIS OF AN UNCEMENTED MOBILE-BEARING TOTAL ANKLE ARTHROPLASTY SYSTEM. Jason W. Fong MODEL-BASED RADIOSTEREOMETRIC ANALYSIS OF AN UNCEMENTED MOBILE-BEARING TOTAL ANKLE ARTHROPLASTY SYSTEM by Jason W. Fong Submitted in partial fulfilment of the requirements for the degree of Master of Applied

More information

Bone Preservation Stem

Bone Preservation Stem TRI-LOCK Bone Preservation Stem Featuring GRIPTION Coating Surgical Technique Implant Geometry Extending the TRI-LOCK Stem heritage The original TRI-LOCK Stem was introduced in 1981. This implant was

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Micromotion of mobile bearing versus posterior stabilized total knee prostheses. Randomized RSA study of 40 knees followed for 2 years

Micromotion of mobile bearing versus posterior stabilized total knee prostheses. Randomized RSA study of 40 knees followed for 2 years Micromotion of mobile bearing versus posterior stabilized total knee prostheses Randomized RSA study of 40 knees followed for 2 years Eric H. Garling 1, Edward R. Valstar 1,2, Rob G.H.H. Nelissen 1 1 Department

More information

Map 48: Rate of metal-on-metal hip resurfacing procedures undertaken per population by PCT

Map 48: Rate of metal-on-metal hip resurfacing procedures undertaken per population by PCT 162 NHS ATLAS OF VARIATION PROBLEMS OF THE MUSCULO-SKELETAL SYSTEM Map 48: Rate of metal-on-metal hip resurfacing procedures undertaken per population by PCT Directly standardised rate 2009/10 Domain 2:

More information

The KineSpring Knee Implant System Product Information

The KineSpring Knee Implant System Product Information The KineSpring Knee Implant System Product Information The Treatment Gap Increasing numbers of young, active OA patients with longer life expectancy and higher activity demands. 1 Large increase in arthroplasty

More information

Early migration of tibial components is associated with late revision

Early migration of tibial components is associated with late revision 614 Acta Orthopaedica 2012; 83 (6): 614 624 Early migration of tibial components is associated with late revision A systematic review and meta-analysis of 21,000 knee arthroplasties Bart G Pijls 1, Edward

More information

Total hip arthroplasty for primary osteoarthrosis in younger patients in the Finnish arthroplasty register

Total hip arthroplasty for primary osteoarthrosis in younger patients in the Finnish arthroplasty register Acta Orthopaedica ISSN: 1745-3674 (Print) 1745-3682 (Online) Journal homepage: http://www.tandfonline.com/loi/iort20 Total hip arthroplasty for primary osteoarthrosis in younger patients in the Finnish

More information

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty

Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, D. W. Murray From

More information

Effect of Surface Conditions of Metallic Pins Simulating Cemented Femoral Stem on Subsidence in Cyclic Fatigue Testing

Effect of Surface Conditions of Metallic Pins Simulating Cemented Femoral Stem on Subsidence in Cyclic Fatigue Testing SOT2/2005 VOL 28 161 Effect of Surface Conditions of Metallic Pins Simulating Cemented Femoral Stem on Subsidence in Cyclic Fatigue Testing H. Kawaji1,2, A. Koistinen1, M. Takagi2, R. Lappalainen1, S.S.

More information

Hemiarthroplasties after hip fractures in Norway and Sweden: a collaboration between the Norwegian and Swedish national registries

Hemiarthroplasties after hip fractures in Norway and Sweden: a collaboration between the Norwegian and Swedish national registries Hip Int 2014; 24 ( 3) : 223-230 DOI: 10.5301/hipint.5000105 ORIGINAL ARTICLE open access Hemiarthroplasties after hip fractures in Norway and Sweden: a collaboration between the Norwegian and Swedish national

More information

Population-Based Rates of Revision of Primary Total Hip Arthroplasty: A Systematic Review

Population-Based Rates of Revision of Primary Total Hip Arthroplasty: A Systematic Review Population-Based Rates of Revision of Primary Total Hip Arthroplasty: A Systematic Review Kelly L. Corbett 1, Elena Losina 1,2,4, Akosua A. Nti 1, Julian J. Z. Prokopetz 1, Jeffrey N. Katz 1,2,3 * 1 Department

More information

SMARTSET Bone Cements. Product Information

SMARTSET Bone Cements. Product Information SMARTSET Bone Cements Product Information INTRODUCTION Our experience in PMMA bone cements is extensive. The long clinical heritage of our DePuy CMW bone cement product range has allowed us to evaluate

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

LIMITED INFLUENCE OF PROSTHETIC POSITION ON ASEPTIC LOOSENING OF THE SOUTER-STRATHCLYDE TOTAL ELBOW PROSTHESIS

LIMITED INFLUENCE OF PROSTHETIC POSITION ON ASEPTIC LOOSENING OF THE SOUTER-STRATHCLYDE TOTAL ELBOW PROSTHESIS 6 LIMITED INFLUENCE OF PROSTHETIC POSITION ON ASEPTIC LOOSENING OF THE SOUTER-STRATHCLYDE TOTAL ELBOW PROSTHESIS J.C.T. van der Lugt R.B. Geskus P.M. Rozing Acta Orthop. 2005;76(5):654-61 Chapter 6 Abstract

More information

EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE.

EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE. EXPERIENCE GPS FOR TOTAL KNEE ARTHROPLASTY DETERMINE YOUR OWN COURSE. The year 2010 marked Exactech s silver aiversary and 25 years of mobility. As a company founded by an orthopaedic surgeon and a biomedical

More information

The Hydrogel barrier to infection

The Hydrogel barrier to infection The Hydrogel barrier to infection The mechanism of infection: The race to the surface In case of contamination of the implant surface, colonisation by pathogens results in formation of a barrier known

More information

Simplifying the Most Clinically Proven 1 Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation

Simplifying the Most Clinically Proven 1 Partial Knee in the World. Oxford Partial Knee with Microplasty Instrumentation Simplifying the Most Clinically Proven 1 Partial Knee in the World Oxford Partial Knee with Microplasty Instrumentation 1 Microplasty Instrumentation Innovative, Accurate, Reproducible Microplasty Instrumentation

More information

Optimum implant geometry

Optimum implant geometry Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years

Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years KNEE Polyethylene wear of mobile-bearing unicompartmental knee replacement at 20 years B. J. L. Kendrick, D. J. Simpson, B. L. Kaptein, E. R. Valstar, H. S. Gill, D. W. Murray, A. J. Price From the Nuffield

More information

KT-1000 records smaller side-to-side differences than radiostereometric analysis before and after an ACL reconstruction

KT-1000 records smaller side-to-side differences than radiostereometric analysis before and after an ACL reconstruction Knee Surg Sports Traumatol Arthrosc (2006) 14: 529 535 KNEE DOI 10.1007/s00167-006-0061-6 Jonas Isberg Eva Faxe` n Sveinbjo rn Brandsson Bengt I. Eriksson Johan Ka rrholm Jon Karlsson KT-1000 records smaller

More information

Balgrist Shoulder Course 2017

Balgrist Shoulder Course 2017 How do we define a glenoid component at risk for clinical failure? Joseph P. Iannotti MD, PhD Maynard Madden Professor and Chairman Orthopaedic and Rheumatologic Institute Cleveland Clinic Conflict of

More information

The effects of hydroxyapatite coating and bone allograft on fixation of loaded experimental primary and revision implants

The effects of hydroxyapatite coating and bone allograft on fixation of loaded experimental primary and revision implants Acta Orthop Scand 2003; 74 (3): 239 247 239 Acta Orthopaedica Scandinavica Award article 2002 The effects of hydroxyapatite coating and bone allograft on fixation of loaded experimental primary and revision

More information

Rotaglide+ TM. Total Knee System Product overview

Rotaglide+ TM. Total Knee System Product overview Rotaglide+ TM Total Knee System Product overview Rotaglide+ TM Originality Stability History Originally implanted in 1988, Rotaglide+ was the first total knee design to adopt a true mobile bearing philosophy.

More information

The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006

The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006 The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006 Overview Every patient moves differently 1 and their total hip replacement should be optimised

More information

CUTTING GUIDE SYSTEM PRODUCT RATIONALE. Custom-Made Patient Instruments for Total Knee Replacement

CUTTING GUIDE SYSTEM PRODUCT RATIONALE. Custom-Made Patient Instruments for Total Knee Replacement CUTTING GUIDE SYSTEM Custom-Made Patient Instruments for Total Knee Replacement This publication is not intended for distribution in the USA. PRODUCT RATIONALE 1 DePuy Synthes Joint Reconstruction TRUMATCH

More information

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole

More information

Knee Replacement Implants

Knee Replacement Implants Knee Replacement Implants During knee replacement surgery, an orthopaedic surgeon will resurface your damaged knee with artificial components, called implants. There are many different types of implants.

More information

Inger Mechlenburg, PhD a, *, Thomas M. Klebe, MD b, Kaj V. Døssing, MD c, Anders Amstrup, MD d, Kjeld Søballe, DMSc a, Maiken Stilling, PhD a

Inger Mechlenburg, PhD a, *, Thomas M. Klebe, MD b, Kaj V. Døssing, MD c, Anders Amstrup, MD d, Kjeld Søballe, DMSc a, Maiken Stilling, PhD a J Shoulder Elbow Surg (2014) 23, 1427-1436 www.elsevier.com/locate/ymse Evaluation of periprosthetic bone mineral density and postoperative migration of humeral head resurfacing implants: two-year results

More information

Socket wall addition device in the treatment of recurrent hip prosthesis dislocation

Socket wall addition device in the treatment of recurrent hip prosthesis dislocation Acta Orthopaedica 2006; 77 (1): 87 91 87 Socket wall addition device in the treatment of recurrent hip prosthesis dislocation Good outcome in 12 patients followed for 4.5 (1 9) years Anders G Enocson 1,

More information

Optimizing function Maximizing survivorship Accelerating recovery

Optimizing function Maximizing survivorship Accelerating recovery Surgical Technique Optimizing Function Maximizing Survivorship Accelerating Recovery The company believes in an approach to patient treatment that places equal importance on: Optimizing function Maximizing

More information

R/F. Can T-smart Tomosynthesis Improve Diagnostic Accuracy on THA Component Stability? 1. Abstract

R/F. Can T-smart Tomosynthesis Improve Diagnostic Accuracy on THA Component Stability? 1. Abstract R/F Can T-smart Tomosynthesis Improve Diagnostic Accuracy on THA Component Stability? Professor and Chair Dept. of Adult Reconstructive Surgery Beijing Jishuitan Hospital, the 4th Clinical College of PKU

More information

Prosthesis survival after total hip arthroplasty does surgical approach matter?

Prosthesis survival after total hip arthroplasty does surgical approach matter? Acta Orthopaedica 2007; 78 (6): 719 729 719 Prosthesis survival after total hip arthroplasty does surgical approach matter? Analysis of 19,304 Charnley and 6,002 Exeter primary total hip arthroplasties

More information

S h o u l d e r Solutions by Tornier C o n v e r T i b l e S h o u l d e r S y S T e m

S h o u l d e r Solutions by Tornier C o n v e r T i b l e S h o u l d e r S y S T e m S h o u l d e r Solutions by Tornier C o n v e r t i b l e s h o u l d e r s y s t e m C o n v e r t i b l e s h o u l d e r s y s t e m A n a t o m i c Aequalis Ascend Flex - UDZF131 One System. Two Solutions.

More information

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination.

Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination. ISPUB.COM The Internet Journal of Orthopedic Surgery Volume 17 Number 1 Increased Rate Of Dislocation Of Total Hip Arthroplasty With Cementless Implants: Accuracy Of Acetabular Inclination. Roberts, Garlick

More information

Audrey Nebergall (Massachusetts General Hospital, Boston), Meridith Greene (Boston), Henrik Palm (Boston), Henrik Malchau (Boston)

Audrey Nebergall (Massachusetts General Hospital, Boston), Meridith Greene (Boston), Henrik Palm (Boston), Henrik Malchau (Boston) Is the precision of RSA measurements of acetabular cup stability and polyethylene wear increased by the addition of tantalum beads to the polyethylene liner? Audrey Nebergall (Massachusetts General Hospital,

More information

Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response relationship

Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response relationship ELSEVER Journal of Orthopaedic Research 23 (2005) 520-525 Journal of Orthopaedic Research www.elsevier.com/locate/orthres Polyethylene wear rate and osteolysis: critical threshold versus continuous dose-response

More information

Zimmer Segmental System

Zimmer Segmental System Zimmer Segmental System Simple solutions for solving complex salvage cases A Step Forward The Zimmer Segmental System is designed to address patients with severe bone loss associated with disease, trauma

More information

A clinical and radiographic 13-year follow-up study of 138 Charnley hip arthroplasties in patients years old

A clinical and radiographic 13-year follow-up study of 138 Charnley hip arthroplasties in patients years old Acta Orthopaedica 2008; 79 (5): x x 1 9 A clinical and radiographic 13-year follow-up study of 138 Charnley hip arthroplasties in patients 50 70 years old Comparison of university hospital data and registry

More information

Higher revision risk for unicompartmental knee arthroplasty in low-volume hospitals

Higher revision risk for unicompartmental knee arthroplasty in low-volume hospitals Acta Orthopaedica 2014; 85 (4): x x 1 Higher revision risk for unicompartmental knee arthroplasty in low-volume hospitals Data from 5,791 cases in the Norwegian Arthroplasty Register Mona Badawy 1, Birgitte

More information

DESIGN RATIONALE AND SURGICAL TECHNIQUE

DESIGN RATIONALE AND SURGICAL TECHNIQUE DESIGN RATIONALE AND SURGICAL TECHNIQUE ANCHOR PEG GLENOID DESIGN RATIONALE In total shoulder arthroplasty, most cases of clinical and radiographic loosening involve failure of the fixation of the glenoid

More information

More re-operations after uncemented than cemented hemiarthroplasty used in the treatment of displaced fractures of the femoral neck

More re-operations after uncemented than cemented hemiarthroplasty used in the treatment of displaced fractures of the femoral neck J-E. Gjertsen, S. A. Lie, T. Vinje, L. B. Engesæter, G. Hallan, K. Matre, O. Furnes From Norwegian Arthroplasty Register, Norway J-E. Gjertsen, MD, PhD, Orthopaedic Surgeon T. Vinje, MD, Orthopaedic Surgeon

More information

ORIGINAL STUDY INTRODUCTION

ORIGINAL STUDY INTRODUCTION Acta Orthop. Belg., 2012, 78, 484-491 ORIGINAL STUDY Hip resurfacing requires larger acetabular cups than conventional hip replacement : A comparative analysis of 100 hips, based on radiographic templating

More information

MARATHON Cross-Linked Polyethylene

MARATHON Cross-Linked Polyethylene MARATHON Cross-Linked Polyethylene Fixation First... 1 Wear Reduction... 2 Oxidative Stability... 3 Mechanical Integrity... 4 Proven Technology...5-6 References... 7 Fixation First FIXATION FIRST Prodigy

More information