Reduction of Pullout Strength Caused by Reinsertion of 3.5-mm Cortical Screws

Size: px
Start display at page:

Download "Reduction of Pullout Strength Caused by Reinsertion of 3.5-mm Cortical Screws"

Transcription

1 ORIGINAL ARTICLE Reduction of Pullout Strength Caused by Reinsertion of 3.5-mm Cortical Screws Amir Matityahu, MD,* Christof Hurschler, PhD, Markus Badenhop, Christina Stukenborg-Colsman, MD, Hazibullah Waizy, MD, Brock Wentz, MD,* Meir Marmor, MD,* and Christian Krettek, MD Introduction: Osteosynthesis of the tibia, tibial plafond, and calcaneus is commonly performed with plates and 3.5-mm selftapping cortical screws. Screw insertion and reinsertion within the same hole in the bone may occur during surgery. Therefore, the purpose of this study was to evaluate the pullout strength of 3.5-mm self-tapping screws with up to 5 re-insertions in the diaphysis of the tibia, metaphysis of the distal tibia, calcaneus, and a polyurethane synthetic bone model. Methods: Screws were inserted into a synthetic bone model and 5 pairs of human cadaveric diaphyseal tibiae, distal tibiae, and calcanei. The bone was predrilled, and then 3.5-mm cortical selftapping 316 L stainless steel screws with a washer were inserted bicortically. Screws were inserted from 1 to 5 times at each location. The screws were grasped and subjected to 5-mm/min tensional force via the biaxial material testing systems machine. Statistical significance was determined using a paired 2-tailed t test. Results: There was a significant difference in the pullout strength of the tibial diaphysis (1710 ± 550 N), tibial metaphysis (471 ± 266 N), and calcaneus (238 ± 90 N; P, 0.01). The tibial diaphysis pullout strength was 1710 ± 550 N for one insertion differing significantly relative to the groups with 4 (average 1030 ± 543 N, P = 0.004) or 5 (average 364 ± 209 N, P, 0.001) insertions. The tibial metaphyseal pullout strength for the single insertion group was 471 ± 266 N and differed significantly relative to the 3 (P = 0.026), 4 (P = 0.044), and 5(P = 0.042) insertion groups. The calcaneal pullout strength for the single insertion group was 238 ± 90 N with a significant difference of the 1, 3, and 4, versus the 5 insertion group (P =0.027,0.040,and 0.033, respectively). The synthetic bone model pullout strength decreased significantly from the one insertion group relative to all other insertion groups (group 1, 1167 ± 263 N; group 2, 768 ± 199 N; group 3, 694 ± 295 N; group 4, 662 ± 356 N; and group 5, 154 ± 183 N; P, 0.02). Accepted for publication March 7, From the *Department of Orthopaedic Surgery, UCSF/SFGH Orthopaedic Trauma Institute, San Francisco, CA; Orthopaedic Department, Laboratory for Biomechanics and Biomaterials, Annastift, Hannover Medical University, Hannover, Germany; and Department of Trauma, Hannover Medical School, Hannover, Germany. The authors have no funding or conflicts of interest to declare. Reprints: Amir Matityahu, MD, Assistant Clinical Professor of Orthopaedic Surgery, Director of Pelvis and Acetabular Trauma and Reconstruction, Department of Orthopaedic Surgery, UCSF/SFGH Orthopaedic Trauma Institute, rd St, 2nd Floor, San Francisco, CA ( matityahua@orthosurg.ucsf.edu). Copyright 2013 by Lippincott Williams & Wilkins Conclusions: There is a significant decrease in relative pullout strength of 3.5-mm self-tapping cortical screws when comparing the tibial diaphysis, tibial metaphysis, and calcaneus. There is also a significant decrease in 3.5-mm self-tapping screw pullout strength after repeated reinsertions in the synthetic bone model, mid-shaft tibia, metaphyseal tibia, and calcaneus. We recommend that during osteosynthesis, careful screw insertion, and minimal reinsertion be performed. Key Words: screw pullout strength, screws, stainless steal screws, self-tapping screws, biomechanics, osteoporosis, cadaveric study, fracture, osteosynthesis (J Orthop Trauma 2013;27: ) INTRODUCTION Osteosynthesis is commonly performed with plates and 3.5-mm self-tapping screws During the insertion of screws, there may be errors in screw placement technique. 17 It has been our experience that screw reinsertion within the same hole in the bone can happen for reasons such as performing a re-reduction of fractured fragments, changing screw lengths, and manipulating plate position. One previous porcine rib study illustrated no difference in the pullout strength of reinserted screws. 18 However, the porcine rib is a flat bone that is not mechanically or clinically similar to human tubular diaphyseal bone such as the tibia or the metaphyseal area of the tibial plafond and calcaneus. Therefore, we proposed a biomechanical study to evaluate the effect of screw reinsertion on pullout strength of 3.5-mm self-tapping screws. Up to 5 reinsertions in the diaphysis of the tibia, metaphysis of the distal tibia, calcaneus, and a polyurethane synthetic bone model were tested. MATERIALS AND METHODS Synthetic Bone Model Screws were inserted into a synthetic bone model (Pacific Research Labs, Vashon, WA) to a thread depth of 39.5 mm, leaving the screw heads proud 10 mm to engage the pulling apparatus and the screw tip at least 2 mm past the opposite side of the polyurethane. The polyurethane foam selected for testing had a density of 0.32 g/cm 3, compressive strength of 8.4 megapascals (MPa) compressive modulus of 210 MPa, tensile strength of 5.6 MPa, tensile modulus of J Orthop Trauma Volume 27, Number 3, March 2013

2 J Orthop Trauma Volume 27, Number 3, March 2013 Reduction of Pullout Strength 284 MPa, shear strength of 4.3 MPa, and shear modulus of 49 MPa, values close to cancellous bones. 17,19 According to ASTM F , standard specification for rigid polyurethane foam for use as a standard material for testing orthpaedic devices and instruments states that: The uniformity and consistent properties of rigid polyurethane foam make it an ideal material for comparative testing of bones screws and other medical devices and instruments. 20 Cadaveric Specimens Five pairs of fresh frozen and then thawed distal tibiae, ankles, and feet were dissected of all soft tissues the same day as biomechanical testing. The specimens were potted in rigid Polymethylmethacrylate at the proximal tibia and held in place on the material testing machine base (material testing systems (MTS) 858 Mini Bionix, MTS GmbH, Berlin, Germany). Screws with washers were inserted into the tibial shaft, tibial metaphysis, and calcaneus. The specimens were rigidly held to prevent motion using clamps on either side of the screw that was being pulled out (Fig. 1). Anatomic Sites There were 3 anatomic sites used for experimentation. All screws were placed bicortically with the tip of the screw at least 2 mm past the opposite cortex. Tibial diaphyseal cortical screws were placed at least 8 cm proximal to the tibial plafond. Screws were inserted in standard arbeitsgemeinschaft für osteosynthesefragen (AO) fashion from lateral to medial parallel to the lateral face of the tibia at the mid-face location. Distal tibia metaphyseal screws were placed anterior to posterior 1 cm proximal to the distal tibia articular surface and not farther then 3 cm from the articular surface. There were 2 rows of screws. One row of 3 screws 1 cm from the ankle joint surface and the second row of screws 8 mm proximal to the distal row. Calcaneal screws were inserted from lateral to medial in hexagonal patterns at least 1 cm from the posterior and distal ends of the calcaneus. The screw FIGURE 1. Photograph depicting a specimen securely potted into a clamp and held by jig to avoid movement. reinsertion number location was rotated so that there were variable but consistent screw locations among specimens. That is, because there were 5 pairs of cadaver specimens (10 specimens in all), each screw location was used twice for each reinsertion group. All screws were placed using standard operative techniques at 8 mm intervals as described previously. 18 Screw Insertion All holes for screw insertion were predrilled with a 2.5-mm drill bit. Screws [3.5 mm cortical self-tapping 316 L stainless steel Synthes (Paoli, PA) with a washer] were inserted using a 2.5-mm hexagonal screwdriver by a fellowship trained orthopaedic traumatologist. Each inserted screw was placed so that the distal tip was proud at least 2 mm. Washers were used to allow for distribution of force on the screw head, self centering, and multiaxial vector pull along the custom made capturing device as described previously. 17 Experimental Groups There were 5 groups in each anatomical site and synthetic bone (Fig. 2). Each group consisted of 10 specimens. Each specimen had 1 screw inserted as described below. All of the groups used 3.5-mm cortical self-tapping screws. Group 1 had screws inserted once and then tested for pullout strength. Group 2 had screws inserted once, removed, and then reinserted and then tested for pullout strength. Group 3 had screws inserted once, removed, and then reinserted 2 times and then tested for pullout strength (total of 3 insertions). Group 4 had screws inserted once, removed, and then reinserted 3 times and then tested for pullout strength (total of 4 insertions). Group 5 had screws inserted once, removed, and then reinserted 4 times and then tested for pullout strength (total of 5 insertions). All screws were placed free hand by a fellowship-trained orthopaedic traumatologist to mimic surgical insertion. The screw lengths were at least 2 mm longer than the total width of the bone or synthetic model. Mechanical Testing The specimens and synthetic bone were secured to the servohydraulic material tester table. The screw washers were grasped and subjected to increased uniaxial tensional force at a constant rate via the biaxial materials testing system machine at a rate of 5 mm/min. 17 Statistical Analysis Means and standard deviation were calculated. Statistical significance was determined using t test (2-tailed, paired). A value of P, 0.05 was considered significant. Each specimen s contralateral side was used as a control. We also compared the pullout strength of the group with initial insertion (no reinsertion) in each anatomical site to the other anatomical sites. A priori power analysis was performed with G*Power software to compute required sample size with given effect size of 1.3, alpha error probability of 0.05, power of 0.95 with 2-tailed t test resulting in actual power of and of total sample size of 10 needed. Ó 2013 Lippincott Williams & Wilkins 171

3 Matityahu et al J Orthop Trauma Volume 27, Number 3, March 2013 FIGURE 2. Flow diagram illustrating the experimental groups (see Methods). RESULTS A total of mm self-tapping cortical screws were utilized for pullout testing. Failure occurred in the screw to bone interface in all cases. There was local failure around the screw with acute loss of fixation in all specimens. Synthetic Bone Model The synthetic bone model had 6 rows of screws with insertion of 1 to 5 times for a total of 30 screws. The screws failed at the screw to synthetic bone model interface. There was local collapse of the polyurethane foam surrounding the screw threads. There was no gross propagation of the screw hole at failure more than 2 mm in any direction. Pullout strength testing of the synthetic bone model of the 3.5 cortical self-tapping cortical screws was variable depending on the number of insertions (Fig. 3). The average pullout strength for the single insertion group was 1167 ± 263 N with coefficient of variation (CV) of There was a significant decrease in pullout strength from the 1 insertion group relative to all of the other insertion groups. There was also a marked increase in the CV (average 154 ± 183 N, CV 1.2) and significant decrease pullout strength between the 5 insertions group and all other groups. Comparing Anatomic Sites There was a significant difference in the pullout strength of the tibia diaphysis (1710 ± 550 N, CV 0.32), tibia metaphysis (471 ± 266 N, CV 0.56), and calcaneus (238 ± 90 N, CV 0.38) with 1, 2, 3, and 4 screw insertions when comparing each of the anatomic locations. There was no significant difference in pullout strength between the tibial metaphysis and calcaneus or the tibial diaphysis and tibial metaphysis after the fifth screw insertion. Tibial Diaphysis There were 5 pairs of tibial diaphyseal bone used for testing screw pullout strength with a total of 50 screws tested. The failure mechanism was catastrophic failure at the bone screw interface with local splintering of the bone. There was propagation of the fractured bone that was no more than 3 mm in any specimen. The average pullout strength of all screws regardless of number of insertions was 1160 ± 643 N. Pullout strength testing of the tibial diaphysis was variable depending on the number of insertions (Fig. 4). The average pullout strength for the single insertion group was 1710 ± 550 N. There was a significant difference between the single screw insertion group versus the groups with 4 (average 1030 ± 543 N, CV = 0.53 P = 0.004) or 5 (average 364 ± 209 N, CV =.574, P, 0.001) insertions. Tibial Metaphysis There were 5 pairs of tibial metaphysis bones used for testing screw pullout strength with a total of 50 screws tested. The failure mechanism was catastrophic failure at the bone screw interface without local splintering of the bone. There was no propagation of the fractured bone more than 2 mm in any specimen. The average pullout strength of all screws regardless of number of insertions was 393 ± 223 N. Pullout strength testing of the tibial metaphysis was variable depending on the number of insertions. The average pullout strength for the single insertion group was 471 ± 266 N (Fig. 5). There was a significant difference of the single insertion group relative to the 3 (P = 0.025), 4 (P = 0.044), and 5 (P = 0.036) insertion groups. There was also a significant difference between the 2 and 5 insertions groups (P = 0.042). However, there was no significant difference between any of the other screw insertion groups. Calcaneus There were 5 pairs of calcaneus bones used for testing screw pullout strength with a total of 50 screws tested. The failure mechanism was catastrophic failure at the bone screw interface without local splintering of the bone. There was no propagation of the fractured bone more than 1 mm in any Ó 2013 Lippincott Williams & Wilkins

4 J Orthop Trauma Volume 27, Number 3, March 2013 Reduction of Pullout Strength FIGURE 3. Pullout strength of 3.5-mm self-tapping screws in asyntheticbonemodelrelativeto the number of screw insertions with table outlining number of screws insertions and their relative P values when compared with each other. specimen. The average pullout strength of all screws regardless of number of insertions was 230 ± 113 N. Pullout strength testing of the calcaneus was variable depending on the number of insertions (Fig. 5). The average pullout strength for the single insertion group was 238 ± 90 N. There was a significant difference between the 1, 3, and 4 screw insertion groups versus the 5 insertion group (P = 0.027, 0.040, and 0.033, respectively). DISCUSSION Reinsertion of screws during fracture fixation is commonly performed in our experience. This study illustrates that significant loss of pullout strength occurs after the second reinsertion (third insertion) of a 3.5-mm cortical screw in the same predrilled hole in metaphyseal bone. Screw reinsertion in diaphyseal bone also showed a significant decrease in pullout strength during the fourth screw insertions. There have been several authors that previously evaluated the pullout strength of screws during different conditions. 17,18,21 25 Only one study evaluated the effect of repeated insertion of one to 3 times. 18 The authors studied 3.5 mm pretapped screws in the porcine rib and found that there were no differences in pullout strength when repeatedly reinserting the screws up to 3 times. 18 This model is in the flat bone porcine model with thin cortices and would have a different geometry than the mid-shaft tibia, metaphyseal tibia, or calcaneus. In another study that investigated the pullout strength of screws in the canine femoral model, the authors assert in the discussion that self-tapping screws inserted into the same hole up to 12 times at 80 percent of their torque out value had the same push out resistance. 25 This study found significant differences when reinserting the 3.5-mm self-tapping screws in the synthetic bone model and in the tibial diaphysis. Because pullout strength of a screw is related to both the bone density and the screw design, we chose to use 3.5-mm self-tapping cortical screw in Ó 2013 Lippincott Williams & Wilkins 173

5 Matityahu et al J Orthop Trauma Volume 27, Number 3, March 2013 FIGURE 4. Pullout strength testing of 3.5-mm cortical screws in the tibial diaphysis relative to number of screw insertions with table of associated P values when comparing the pullout strength of number of screw insertions in each group. all tested locations to be more consistent between groups. Evaluation of the synthetic bone illustrated that there is a significant decrease in pullout strength from the first insertion to all other reinsertions. Because the density of the synthetic bone is uniform, we expected it to emulate homogeneous metaphyseal bone. We chose metaphyseal emulating synthetic material to imitate a density of bone that is between tibial diaphysis and metaphyseal bone. The advantage of using synthetic bone was in having decreased coefficient of variability of the first insertion group. However, the pullout strength variability increased with subsequent insertion to the point of being comparable to the cadaveric specimens. This finding illustrates that the variability in pullout strength is a result of insertion technique coupled with a lower density material. In cadaveric tibial metaphyseal bone, a significant decrease in pullout strength was seen from the third reinsertion and beyond. This study also found that there was a significant difference between the first insertion and fifth insertion in all anatomic locations. However, this finding in not be clinically relevant in most cases because five reinsertions are a rare occurrence. In cases where there are repeated insertions of a 3.5-mm cortical screw into the same hole, we suggest changing to a different location if possible. The difference in results between our study and the previous studies can be explained by the difference in bone models and by the insertion technique. In the previous studies, a special jig was devised to insert the screws. However, in our study, we tried to simulate the clinical situation by inserting each screw in a free-hand manner by a fellowship trained orthopaedic traumatologist. Technical factors that lead to differences in pullout strength can be attributed to slight changes in the angle of insertion and removal of the screw, causing stripping of the bone threads and change in bone hole morphology. Screws inserted in different locations in the body have differences in screw pullout strength due to variable bone density and architecture. We found that the diaphyseal bone of the tibia has more than 3-fold increase in pullout strength relative to the metaphyseal tibia and more than 7-fold increase pullout strength relative to the calcaneus. The metaphysis of the tibia has a 2-fold increase in pullout strength relative to the calcaneus. There are factors that influence the failure of fixation not tested in this study. These factors include insertion torque, bone density, screw length within the bone, and screw design. Previously, screw design has been shown to affect screw insertion torque but not pullout strength. 26 However, the influence of reinsertion of screws on insertion torque has not been studied previously and may be the subject of future research. This study has some limitations. The local bone density within each specimen may not be consistent. We tried to minimize this limitation by utilizing 5 pairs of specimens and a synthetic bone model. We also tried to address this problem by placing screws in the metaphysis of the tibia equidistant from the ankle joint surface. Another limitation of this study is that there are only 10 screws at each anatomic location (5 pairs). When placing screws in the mid-tibia or distal tibial metaphysis, there was enough of a difference in pullout strength to achieve significance. However, in the calcaneus, we could not illustrate differences in pullout strength between the first though fourth reinsertions. This may be due to the fact that the differences in pullout strength are small, the standard deviation is large, and as the bone density decreases, the overall pullout strength decreases. The phenomenon of decreased bone density with increased standard deviation of pullout strength in cadaveric bone has been described previously and is alimitationofthisstudyasappliedtothedistaltibiaand calcaneus. 21 Moreover, although all screws were placed bicortically, different bone diameter and length of screw within the cadaver tissue dictated different screw lengths within the bone. As a result, this may have influenced the ultimate pullout strength and its standard deviation. Another limitation of the study could have been the technique of insertion of the screws. Because the screws were placed by the surgeon and not via a mechanical apparatus, there may have been inconsistency in screw insertion technique. Therefore, we may have had an increase in pullout strength standard deviation from specimen to specimen if the insertion of the screws was performed by hand and not by an apparatus. However, we chose this technique to simulate the intraoperative situation where the surgeon is placing the screw into the bone by hand and without an apparatus. In summary, this study demonstrated significant decrease in 3.5-mm self-tapping screw pullout strength with reinsertions in the same predrilled hole. This effect was more pronounced Ó 2013 Lippincott Williams & Wilkins

6 J Orthop Trauma Volume 27, Number 3, March 2013 Reduction of Pullout Strength FIGURE 5. Pullout strength testing of 3.5-mm cortical screws in the distal tibial metaphysis and calcaneus relative to number of screw insertions with table of associated P values when comparing the pullout strength of number of screw insertions in each group. in high-density bone. This study also illustrates that avoidance of more than 2 reinsertions in the distal tibial metaphysis could minimize an iatrogenic decrease in pullout strength of screws. REFERENCES 1. Amorosa LF, Brown GD, Greisberg J. A surgical approach to posterior pilon fractures. J Orthop Trauma. 2010;24: Blauth M, Bastian L, Krettek C, et al. Surgical options for the treatment of severe tibial pilon fractures: a study of three techniques. J Orthop Trauma. 2001;15: Boraiah S, Kemp TJ, Erwteman A, et al. Outcome following open reduction and internal fixation of open pilon fractures. J Bone Joint Surg Am. 2010;92; Borens O, Kloen P, Richmond J, et al. Minimally invasive treatment of pilon fractures with a low profile plate: preliminary results in 17 cases. Arch Orthop Trauma Surg. 2009;129: Horisberger M, Valderrabano V, Hintermann B. Posttraumatic ankle osteoarthritis after ankle-related fractures. J Orthop Trauma. 2009;23: Kendoff D, Citak M, Gardner M, et al. Three-dimensional fluoroscopy for evaluation of articular reduction and screw placement in calcaneal fractures. Foot Ankle Int. 2007;28: Kline AJ, Gruen GS, Pape HC, et al. Early complications following the operative treatment of pilon fractures with and without diabetes. Foot Ankle Int. 2009;30: Lee YS, Chen SW, Chen SH, et al. Stabilisation of the fractured fibula plays an important role in the treatment of pilon fractures: a retrospective comparison of fibular fixation methods. Int Orthop. 2009;33: Leonard M, Magill P, Khayyat G. Minimally-invasive treatment of high velocity intra-articular fractures of the distal tibia. Int Orthop. 2009;33: McCann PA, Jackson M, Mitchell ST, et al. Complications of definitive open reduction and internal fixation of pilon fractures of the distal tibia. Int Orthop. 2010;35: Niikura T, Miwa M, Sakai Y, et al. Ankle arthrodesis using antegrade intramedullary nail for salvage of nonreconstructable tibial pilon fractures. Orthopedics. 2009;32: Richter M, Droste P, Goesling T, et al. Polyaxially-locked plate screws increase stability of fracture fixation in an experimental model of calcaneal fracture. J Bone Joint Surg Br. 2006;88: Richter M, Gosling T, Zech S, et al. A comparison of plates with and without locking screws in a calcaneal fracture model. Foot Ankle Int. 2005;26: Ó 2013 Lippincott Williams & Wilkins 175

7 Matityahu et al J Orthop Trauma Volume 27, Number 3, March Ruedi T, Allgower M. Late results after operative treatment of fractures of the distal tibia (pilon tibial fractures) (author s transl) [in German]. Unfallheilkunde. 1978;81: Thermann H, Krettek C, Hufner T, et al. Management of calcaneal fractures in adults. Conservative versus operative treatment. Clin Orthop Relat Res. 1998;353: Wang C, Li Y, Huang L, et al. Comparison of two-staged ORIF and limited internal fixation with external fixator for closed tibial plafond fractures. Arch Orthop Trauma Surg. 2010;130: Collinge C, Hartigan B, Lautenschlager EP. Effects of surgical errors on small fragment screw fixation. J Orthop Trauma. 2006;20: Foley WL, Frost DE, Tucker MR. The effect of repetitive screw hole use on the retentive strength of pretapped and self-tapped screws. J Oral Maxillofac Surg. 1990;48: Swiontkowski MF, Harrington RM, Keller TS, et al. Torsion and bending analysis of internal fixation techniques for femoral neck fractures: the role of implant design and bone density. JOrthopRes.1987; 5: Committee F04.21 on Osteosynthesis. Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopedic Devices and Instruments. West Conshohocken, PA: ASTM International; ASTM F e1. Available at: Accessed March 23, Westmoreland GL, McLaurin TM, Hutton WC. Screw pullout strength: a biomechanical comparison of large-fragment and small-fragment fixation in the tibial plateau. J Orthop Trauma. 2002;16: Zdero R, Elfallah K, Olsen M, et al. Cortical screw purchase in synthetic and human femurs. J Biomech Eng. 2009;131: Zdero R, Rose S, Schemitsch EH, et al. Cortical screw pullout strength and effective shear stress in synthetic third generation composite femurs. J Biomech Eng. 2007;129: Zdero R, Schemitsch EH. The effect of screw pullout rate on screw purchase in synthetic cancellous bone. J Biomech Eng. 2009;131; Schatzker J, Sanderson R, Murnaghan JP. The holding power of orthopedic screws in vivo. Clin Orthop Relat Res. 1975;108: Ricci WM, Tornetta P, Petteys MS, et al. A comparison of screw insertion torque and pullout strength. J Orthop Trauma. 2010;24: Ó 2013 Lippincott Williams & Wilkins

Clinical. Solutions. Synthes Solutions. Foot and Ankle.

Clinical. Solutions. Synthes Solutions. Foot and Ankle. Clinical Solutions Foot and Ankle. Foot and Ankle. Fractures of the tibial shaft Fractures of the distal fibula Fractures of the distal tibia Fractures and osteotomies of the calcaneus Arthrodesis Fractures,

More information

Plate Fixation Options

Plate Fixation Options Distal tibia extra-articular fractures can be difficult to treat Tenuous soft tissue Complex fracture patterns Plate Fixation Options Medial plating: Minimally invasive approach Technical ease Anterolateral

More information

LCP Distal Tibia Plate

LCP Distal Tibia Plate Surgical Technique LCP Locking Compression Plate Original Instruments and Implants of the Association for the Study of Internal Fixation AO/ASIF Table of contents Indications 3 Implants/Instruments 5 Surgical

More information

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates

Surgical Technique. Anterolateral and Medial Distal Tibia Locking Plates Surgical Technique Anterolateral and Medial Distal Tibia Locking Plates PERI-LOC Periarticular Locked Plating System Anterolateral and Medial Distal Tibia Locking Plates Surgical Technique Contents Product

More information

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system.

LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system. LCP Medial Proximal Tibial Plate 4.5/5.0. Part of the Synthes LCP periarticular plating system. Technique Guide This publication is not intended for distribution in the USA. Instruments and implants approved

More information

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. Technique Guide LCP Small Fragment System Table of Contents Introduction

More information

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90

Surgical Technique. Cannulated Angled Blade Plate 3.5 and 4.5, 90 Surgical Technique Cannulated Angled Blade Plate 3.5 and 4.5, 90 Cannulated Angled Blade Plate 3.5 and 4.5, 90 Table of contents Indications/Contraindications 2 Implants 3 Surgical technique 5 Implant

More information

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System.

LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System. LCP Anterior Ankle Arthrodesis Plates. Part of the Synthes Locking Compression Plate (LCP) System. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction

More information

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates

Surgical Technique. 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates Surgical Technique 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plates PERI-LOC Periarticular Locked Plating System 3.5mm and 4.5mm Lateral Proximal Tibia Locking Plate Surgical Technique Contents Product

More information

FLT105 12/02.

FLT105 12/02. FLT105 12/02 www.biometmerck.co.uk Disclaimer Biomet Merck Ltd, as the manufacturer of this device, does not practice medicine and does not recommend this or any other surgical technique for use on a

More information

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System A locking plate system that expands a surgeon s options in trauma surgery Zimmer NCB Plating System The Power of Choice The power of having true intraoperative options is at your fingertips. Using standard

More information

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. Technique Guide This publication is not intended for distribution in the USA. Instruments

More information

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

Technique Guide. LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Technique Guide LCP Proximal Femoral Hook Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Table of Contents Introduction Features and Benefits 2 AO ASIF Principles 4 Indications 5 Surgical

More information

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia

Types of Plates 1. New Dynamic Compression Plate: Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia Types of Plates 1. New Dynamic Compression Plate: DCP Diaphyseal fracture: Radius, Ulna, Humerus, Rarely tibia 1. Undercut adjacent to the holes low contact: less stress shield 2. Undercut at the undersurface

More information

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. Technique Guide LCP Small Fragment System Table of Contents Introduction

More information

ROTATIONAL PILON FRACTURES

ROTATIONAL PILON FRACTURES CHAPTER 31 ROTATIONAL PILON FRACTURES George S. Gumann, DPM The opinions and commentary of the author should not be construed as refl ecting offi cial U.S. Army Medical Department policy. Pilon injuries

More information

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability.

LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. LCP Anterolateral Distal Tibia Plate 3.5. The low profile anatomic fixation system with optimal plate placement and angular stability. Technique Guide LCP Small Fragment System Table of Contents Introduction

More information

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Low Bend Medial Distal Tibia Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Low Bend Medial Distal Tibia Plates

More information

Surgical Technique. Targeter Systems Overview

Surgical Technique. Targeter Systems Overview Surgical Technique Targeter Systems Overview PERI-LOC Locked Plating System Targeter Systems Overview Table of contents Product overview... 2 Introduction... 2 Indications... 2 Design features and benefits...

More information

The Vilex FUZETM. Dual Thread Screw & Intramedullary Nail in One Implant. The Ultimate TTC Arthrodesis Internal Fixator

The Vilex FUZETM. Dual Thread Screw & Intramedullary Nail in One Implant. The Ultimate TTC Arthrodesis Internal Fixator The Vilex FUZETM Dual Thread Screw & Intramedullary Nail in One Implant The Ultimate TTC Arthrodesis Internal Fixator Introduction The Vilex FUZE TM TTC Arthrodesis Compression Nail combines the attributes

More information

LCP Extra-articular Distal Humerus Plate.

LCP Extra-articular Distal Humerus Plate. Technique Guide LCP Extra-articular Distal Humerus Plate. The anatomically shaped and angular stable fixation system for extraarticular fractures of the distal humerus. Table of Contents Introduction

More information

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4

Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4 Relationship between the Apex of Flexible Nail and the Level of Fracture: A Biomechanical Study Ahmed N* 1, Gakhar H 2, Cheung G 3, Sharma A 4 Abstract Background Centre of Orthopaedic biomechanics, Bath

More information

Technique Guide. 4.5 mm LCP Proximal Tibia Plates. Part of the Synthes LCP Periarticular Plating System.

Technique Guide. 4.5 mm LCP Proximal Tibia Plates. Part of the Synthes LCP Periarticular Plating System. Technique Guide 4.5 mm LCP Proximal Tibia Plates. Part of the Synthes LCP Periarticular Plating System. Table of Contents Introduction 4.5 mm LCP Proximal Tibia Plates 2 AO Principles 4 Indications 5 Surgical

More information

EVOS MINI with IM Nailing

EVOS MINI with IM Nailing Case Series Dr. John A. Scolaro EVOS MINI with IM Nailing A series of studies Introduction Intramedullary nailing has become the standard for many long bone fractures. Fracture reduction prior to nail

More information

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM

3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM 3.5 MM VA-LCP PROXIMAL TIBIA PLATE SYSTEM Part of the DePuy Synthes Variable Angle Periarticular Plating System SURGICAL TECHNIQUE TABLE OF CONTENTS INTRODUCTION 3.5 mm VA-LCP Proximal Tibial Plate 2 AO

More information

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System.

Technique Guide. 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System. Technique Guide 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments. Part of the 3.5 mm LCP Percutaneous Instrument System. Table of Contents Introduction 3.5 mm LCP Low Bend Medial Distal

More information

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures.

LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures. LCP Low Bend Medial Distal Tibia Plates 3.5 mm. Anatomic plates with low profile head for intra- and extraarticular fractures. Surgical Technique This publication is not intended for distribution in the

More information

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System

A locking plate system that expands a surgeon s options in trauma surgery. Zimmer NCB Plating System A locking plate system that expands a surgeon s options in trauma surgery Zimmer NCB Plating System The Power of Choice The power of having true intraoperative options is at your fingertips. Using standard

More information

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee.

Technique Guide. TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee. Technique Guide TomoFix Osteotomy System. A comprehensive plating system for stable fixation of osteotomies around the knee. Table of Contents Introduction TomoFix Osteotomy System 2 AO Principles 4 Indications

More information

PROXIMAL TIBIAL PLATE

PROXIMAL TIBIAL PLATE SURGICAL NÁSTROJE TECHNIQUE PRO ARTROSKOPII PROXIMAL INSTRUMENTS TIBIAL FOR PLATE ARTHROSCOPY Proximal Tibial Plate Description of medical device The Proximal Tibial Plate is used in epyphyseal and metaphyseal

More information

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia. LISS DF and LISS PLT. Less Invasive Stabilization Systems for Distal Femur and Proximal Lateral Tibia.

More information

Cannulated Screw System. Product Overview

Cannulated Screw System. Product Overview Cannulated Screw System Product Overview Acumed Cannulated Screw System The Acumed Cannulated Screw System consists of screws, washers, and instruments and is intended for fixation of fractures, fusions,

More information

Biomechanical Performance of Headless Compression Screws: Fixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS

Biomechanical Performance of Headless Compression Screws: Fixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS Biomechanical Performance of Headless Compression Screws: ixos Ø4.0, 5.0 & 7.0 mm compared to Synthes Ø4.5 & 6.5 mm HCS M. Vallotton, A. Torcasio PhD Stryker Trauma AG, Selzach, Switzerland Abstract Introduction:

More information

Ordering information. LCP Locking Compression Plate. Combine without compromise.

Ordering information. LCP Locking Compression Plate. Combine without compromise. Ordering information LCP Locking Compression Plate. Combine without compromise. Table of Contents Small Fragment System Large Fragment System LCP Instruments Emergency Instruments Implants Humerus Plates,

More information

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction LCP Distal Fibula Plates 2 AO Principles 4 Indications 5 Surgical Technique

More information

Surgical Technique. Lower Extremity Plates and Straight Plates

Surgical Technique. Lower Extremity Plates and Straight Plates Surgical Technique Lower Extremity Plates and Straight Plates 2 Table of contents Overview...4 Indications... 4 Contraindications... 4 Screw Options... 5 Straight Plate Options... 6 Proximal Tibia Plate

More information

The Flower Straight Fibula Plate

The Flower Straight Fibula Plate The Flower Straight Fibula Plate PROCEDURE GUIDE www.flowerortho.com The Flower Foot & Ankle Application STRAIGHT LOCKING FIBULA PLATE ANTERIOR LATERAL DISTAL TIBIAL PLATE MEDIAL DISTAL TIBIAL PLATE ANATOMIC

More information

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates

Part of the DePuy Synthes Locking Compression Plate (LCP ) System. 3.5 mm LCP Medial Proximal Tibia Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Medial Proximal Tibia Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Medial Proximal Tibia Plates 2 AO

More information

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 2.7 mm/3.5 mm LCP Distal Fibula Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 2.7 mm/3.5 mm LCP Distal Fibula Plates 2 AO Principles

More information

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Periarticular Proximal Humerus Plate. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Proximal Humerus Plate 2 AO Principles

More information

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate

Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate Surgical Technique The Science of the Landscape Zimmer MIS Periarticular 3.5mm Proximal Tibial Locking Plate Surgical Technique 1 Zimmer MIS

More information

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.

OBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. Surgical Technique LCP Small Fragment System This publication

More information

NCB Distal Femur System. Surgical Technique

NCB Distal Femur System. Surgical Technique NCB Distal Femur System Surgical Technique NCB Distal Femur System Surgical Technique 3 Surgical Technique NCB Distal Femur System Table of Contents Introduction 4 Indications 8 Preoperative Planning

More information

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

More information

Ref. Number Stainless Steel

Ref. Number Stainless Steel Section of humerus plates for operation on distal humerus Y - Distal Humerus Plate 4. Y - Distale Humerusplatte 4. Plaque pour humerus distal en Y 4. Indication : Used to treat coinuted distal humerus

More information

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments

3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments Part of the 3.5 mm LCP 3.5 mm LCP Low Bend Medial Distal Tibia Plate Aiming Instruments Surgical Technique TABLE OF CONTENTS INTRODUCTION 3.5 mm LCP Low Bend Medial Distal Tibia Plate 2 Aiming Instruments

More information

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique

3.0/3.5/4.0/4.5/6.5/7.0/7.3. Cannulated Screws. Surgical Technique 3.0/3.5/4.0/4.5/6.5/7.0/7.3 Cannulated Screws Surgical Technique Image intensifier control This description alone does not provide sufficient background for direct use of DePuy Synthes products. Instruction

More information

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius.

LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Technique Guide LCP Proximal Radius Plates 2.4. Plates for radial head rim and for radial head neck address individual fracture patterns of the proximal radius. Table of Contents Introduction LCP Proximal

More information

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large).

Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large). Technique Guide Periarticular Aiming Arm Instruments for LCP Proximal Tibial Plate 4.5/5.0. Part of the LCP Periarticular Aiming Arm Instrument System (large). Image intensifier control Warning This description

More information

NEW INSTRUMENTS FOR INTERNAL FIXATION OF FRACTURES USING MINIMALLY INVASIVE TECHNIQUES

NEW INSTRUMENTS FOR INTERNAL FIXATION OF FRACTURES USING MINIMALLY INVASIVE TECHNIQUES NEW INSTRUMENTS FOR INTERNAL FIXATION OF FRACTURES USING MINIMALLY INVASIVE TECHNIQUES Dr.eng. Comşa Stanca, sing. Gheorghiu Doina, eng. Ciobota Dan National Institute of Research & Development for fine

More information

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System.

LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. LCP Condylar Plate 4.5/5.0. Part of the LCP Periarticular Plating System. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

More information

4.5 mm LCP Medial Proximal Tibia Plates

4.5 mm LCP Medial Proximal Tibia Plates Part of the DePuy Synthes LCP Periarticular Plating System 4.5 mm LCP Medial Proximal Tibia Plates Surgical Technique Table of Contents Introduction 4.5 mm LCP Medial Proximal Tibia Plates 2 AO Principles

More information

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system.

2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the Synthes locking compression plate (LCP) system. 2.7 mm/3.5 mm LCP Distal Fibula Plate System. Part of the locking compression plate (LCP) system. Anatomically contoured Multiple screw options for fixedangle support Coaxial holes minimize screw head

More information

Asnis III Cannulated Screw System. Operative technique

Asnis III Cannulated Screw System. Operative technique Asnis III Cannulated Screw System Operative technique Asnis III Cannulated Screws Operative technique Asnis III Cannulated Screw System Contents 1. Indications and contraindications... 4 2. Technical specifications...

More information

Surgical Technique International Version

Surgical Technique International Version Surgical Technique International Version PERI-LOC VLP Variable-Angle Locked Plating System Surgical Technique Table of Contents Product Overview...2 Introduction...2 Indications and Contraindications...3

More information

SMV Scientific Bone Plate and Screw System Surgical Technique

SMV Scientific Bone Plate and Screw System Surgical Technique SMV Scientific Bone Plate and Screw System Surgical Technique Description: The SMV Scientific Bone Plate and Screw System consists of non-locking plates and bone screw fasteners in a variety of lengths,

More information

Cannulated Angled Blade Plate 3.5 and 4.5, 90.

Cannulated Angled Blade Plate 3.5 and 4.5, 90. Cannulated Angled Blade Plate 3.5 and 4.5, 90. Technique Guide This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Table of Contents Introduction

More information

2.7 mm/3.5 mm Variable Angle LCP. Ankle Trauma System

2.7 mm/3.5 mm Variable Angle LCP. Ankle Trauma System Part of the DePuy Synthes Variable Angle Locking Compression Plate (VA LCP ) System 2.7 mm/3.5 mm Variable Angle LCP Ankle Trauma System Surgical Technique Table of Contents Introduction 2.7 mm/3.5 mm

More information

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system.

LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. LCP Medial Proximal Tibial Plate 3.5. Part of the Synthes small fragment Locking Compression Plate (LCP) system. Surgical Technique This publication is not intended for distribution in the USA. Instruments

More information

3.5 mm LCP Extra-articular Distal Humerus Plate

3.5 mm LCP Extra-articular Distal Humerus Plate Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Extra-articular Distal Humerus Plate Surgical Technique Table of Contents Introduction 3.5 mm LCP Extra-articular Distal Humerus

More information

Increasing surgical freedom Restoring patient function

Increasing surgical freedom Restoring patient function Increasing surgical freedom Restoring patient function Fracture specific plating solutions for the most common tibia and fibula fractures Frequency of fracture occurrences* 66% 61% 36% 36% 28% 14% 20%

More information

Surgical Technique. Calcaneal Locking Plate

Surgical Technique. Calcaneal Locking Plate Surgical Technique Calcaneal Locking Plate PERI-LOC Locked Plating System Calcaneal Locking Plate Surgical TechniqueCatalog Infor Table of Contents Introduction...2 Indications...3 Plate Features...3 Patient

More information

3.5 mm LCP Distal Tibia T-Plates

3.5 mm LCP Distal Tibia T-Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Distal Tibia T-Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Distal Tibia T-Plates 2 AO Principles 4 Indications

More information

LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL

LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL LOCKING TEP LOCKING TITANIUM ELASTIC PIN INTRAMEDULLARY NAIL ... Index -3 3-8 8 9 9 0 7 Introduction Features Indicatiıons Surgical Technique Femoral Surgical Technique Tibial Surgical Technique Ulna Radius

More information

2.7 mm/3.5 mm LCP Distal Fibula Plate

2.7 mm/3.5 mm LCP Distal Fibula Plate Part of the DePuy Synthes Locking Compression Plate (LCP ) System 2.7 mm/3.5 mm LCP Distal Fibula Plate Surgical Technique Table of Contents Introduction 2.7 mm/3.5 mm LCP Distal Fibula Plates 2 AO Principles

More information

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible.

Technique Guide. Compact 2.0 LOCK Mandible. The locking system for the mandible. Technique Guide Compact 2.0 LOCK Mandible. The locking system for the mandible. Table of Contents Introduction Compact 2.0 LOCK Mandible 2 AO Principles 4 Indications and Contraindications 5 Surgical

More information

Zimmer MIS Periarticular Distal Femoral Locking Plate

Zimmer MIS Periarticular Distal Femoral Locking Plate For Clinical Evaluations Zimmer MIS Periarticular Distal Femoral Locking Plate Surgical Technique The Science of the Landscape Zimmer MIS Periarticular Distal Femoral Locking Plate Surgical Technique

More information

Low Bend Distal Tibia Plates

Low Bend Distal Tibia Plates Part of the DePuy Synthes Locking Compression Plate (LCP ) System 3.5 mm LCP Low Bend Medial Distal Tibia Plates Surgical Technique Table of Contents Introduction 3.5 mm LCP Low Bend Medial Distal Tibia

More information

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts.

Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts. Large External Fixator Delta Frame Ankle Bridge. Using pin clamps with outrigger posts. Technique Guide Part of the Large External Fixation System Large External Fixator Delta Frame Ankle Bridge Technique

More information

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium.

Technique Guide. Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium. Technique Guide Small Fragment Locking Compression Plate (LCP) System. Stainless steel and titanium. Table of Contents Introduction Small Fragment Locking Compression Plate (LCP) System 2 AO Principles

More information

COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION

COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION C H A P T E R 3 0 COMPARING KIRSCHNER WIRE FIXATION TO A NEW DEVICE USED FOR PROXIMAL INTERPHALANGEAL FUSION Scott R. Roman, DPM INTRODUCTION The most common fixation for proximal interphalangeal fusion

More information

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system.

Technique Guide. 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system. Technique Guide 3.5 mm LCP Olecranon Plates. Part of the Synthes locking compression plate (LCP) system. Table of Contents Introduction 3.5 mm LCP Olecranon Plates 2 AO Principles 3 Indications 3 Clinical

More information

OptiLock Periarticular Plating System For Distal Femoral Fractures. Pre-Launch Surgical Technique

OptiLock Periarticular Plating System For Distal Femoral Fractures. Pre-Launch Surgical Technique OptiLock Periarticular Plating System For Distal Femoral Fractures Pre-Launch Surgical Technique Contents Introduction... Page 1 Indications And Fracture Classification... Page 6 Design Features... Page

More information

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique

OptiLock Periarticular Plating System For Distal Tibial Fractures. Surgical Technique OptiLock Periarticular Plating System For Distal Tibial Fractures Surgical Technique Contents Introduction... Page 1 Indications & Contraindications... Page 6 System Features... Page 7 Surgical Technique...

More information

Ankle Fractures in the Elderly: How to Deal with Poor Bone Quality

Ankle Fractures in the Elderly: How to Deal with Poor Bone Quality : How to Deal with Poor Bone Quality Richard T. Laughlin, MD Professor of Orthopaedic Surgery University of Cincinnati College of Medicine No disclosures relative to this presentation acknowledgement Some

More information

3.5 mm Locking Attachment Plate

3.5 mm Locking Attachment Plate For Treatment of Periprosthetic Fractures 3.5 mm Locking Attachment Plate Surgical Technique Table of Contents Introduction 3.5 mm Locking Attachment Plate 2 Indications 4 Surgical Technique Preparation

More information

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis

AO / Synthes Proximal Posterior Medial Tibia Plate. Tibial Plateau Fx. Osteosynthesis Tibial Plateau Fx. Osteosynthesis Articular Fractures Osteosynthesis Characteristics of the LCP system LCP 3.5 system LCP 4.5 system Many different traditional plates (small, large) Lag screws Rafting

More information

Conventus CAGE PH Surgical Techniques

Conventus CAGE PH Surgical Techniques Conventus CAGE PH Surgical Techniques Conventus Orthopaedics The Conventus CAGE PH (PH Cage) is a permanent implant comprised of an expandable scaffold, made from nitinol and titanium, which is deployed

More information

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology.

2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Technique Guide 2.4 mm Variable Angle LCP Volar Extra-Articular Distal Radius System. For fragment-specific fracture fixation with variable angle locking technology. Table of Contents Introduction 2.4

More information

Techique. Results. Discussion. Materials & Methods. Vol. 2 - Year 1 - December 2005

Techique. Results. Discussion. Materials & Methods. Vol. 2 - Year 1 - December 2005 to each other. The most distal interlocking hole is 3 mm proximal to distal end of nail, is in anteroposterior direction & proximal distal interlocking hole is in medial to lateral direction i.e. at right

More information

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System.

Technique Guide. 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System. Technique Guide 3.5 mm LCP Proximal Tibia Plate. Part of the Synthes Small Fragment LCP System. Table of Contents AO ASIF Principles of Internal Fixation 4 Indications/Contraindications 5 Surgical Technique

More information

tep Ankle Fracture Plating System

tep Ankle Fracture Plating System tep Response Ortho is a global orthopaedic trauma solutions manufacturer offering premium products created under its founding principles of innovation, excellence by design, and functional superiority.

More information

Large Fragment LCP Instrument and Implant Set

Large Fragment LCP Instrument and Implant Set Part of the DePuy Synthes Locking Compression Plate (LCP ) System Large Fragment LCP Instrument and Implant Set Surgical Technique Table of Contents Introduction Large Fragment LCP Instrument and Implant

More information

AxSOS Locking Plate System

AxSOS Locking Plate System AxSOS Locking Plate System Operative Technique Small Fragment Basic Fragment 1 2 Contents Page 1. Introduction 4 2. Features & Benefits 5 4 and 5mm Compression Plates 5 Reconstruction and 1/3 Tubular Locking

More information

Small Fragment Locking Compression Plate (LCP ) System

Small Fragment Locking Compression Plate (LCP ) System Stainless Steel and Titanium Small Fragment Locking Compression Plate (LCP ) System Surgical Technique Table of Contents Introduction Small Fragment Locking Compression Plate (LCP) System 2 AO Principles

More information

Dieter Marquardt Medizintechnik GmbH 08 Cannulated Screws

Dieter Marquardt Medizintechnik GmbH 08 Cannulated Screws Cannulated Screws Product Name: Cannulated Screw Ø 3.5 Screw diameter: 3.5 mm Core diameter: 2.5 mm Head diameter: 5.0 mm Thread: short or full Cannulation: 1.35 mm Inner hexagon: 2.5 mm Screw Length:

More information

TIPMED EXTERNAL FIXATION SYSTEMS

TIPMED EXTERNAL FIXATION SYSTEMS TIPMED EXTERNAL FIXATION SYSTEMS ANATOMICAL LOCATIONS FOR EXTERNAL FIXATION SYSTEMS Humeral Dynamic Axial Fixator Elbow Fixator Pelvic Dynamic Axial Fixator Pennig Wrist Fixator Hand Fixator Finger Fixator

More information

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.

Surgical Technique. This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. LCP Extra-articular Distal Humerus Plate. The anatomically shaped and angular stable fixation system for extraarticular fractures of the distal humerus. Surgical Technique This publication is not intended

More information

WINSTA-R. Distal Radius System

WINSTA-R. Distal Radius System Distal Radius System Table of Contents Introduction WINSTA-R System 2 Indication 2 Surgical Technique Palmar Access for Radius Plate 3 Dorsal Access for Radius Plate 3 Positioning of the Radius Plate

More information

4.5 mm VA-LCP. Part of the Variable Angle Periarticular Plating System

4.5 mm VA-LCP. Part of the Variable Angle Periarticular Plating System 4.5 mm VA-LCP Curved Condylar Plate Part of the Variable Angle Periarticular Plating System Surgical Technique Table of Contents Introduction 4.5 mm VA-LCP Curved Condylar Plates 2 4.5 mm VA-LCP Curved

More information

Locked Plating: Biomechanics and Biology

Locked Plating: Biomechanics and Biology Techniques in Orthopaedics 22(4):E1 E6 2007 Lippincott Williams & Wilkins, Inc. Locked Plating: Biomechanics and Biology Kyle F. Dickson, M.D., M.B.A., John W. Munz, M.D. Summary: Since the early ideas

More information

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures.

Technique Guide. Locking Attachment Plate. For treatment of periprosthetic fractures. Technique Guide Locking Attachment Plate. For treatment of periprosthetic fractures. Table of Contents Introduction Locking Attachment Plate 2 Indications 4 Surgical Technique Patient Positioning 5 Preparation

More information

MetaFix Ludloff Plate

MetaFix Ludloff Plate Merete MetaFix Ludloff Plate Low Profile Locking Bone Plate System Surgical Technique and Ordering Information - Content - Content 1. Description.................................................. 3 2.

More information

2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1

2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1 2.7 mm/3.5 mm Variable Angle LCP Elbow System DJ9257-B 1 System overview Simply complete: A comprehensive system, consisting of five (5) distal humerus plates and three (3) types of olecranon plates Implant

More information

Locked Plating: Clinical Indications

Locked Plating: Clinical Indications Techniques in Orthopaedics 22(3):181 185 2007 Lippincott Williams & Wilkins, Inc. Locked Plating: Clinical Indications Kyle F. Dickson, M.D., M.B.A., John Munz, M.D. Summary: As shown in the previous article,

More information

TABLE OF CONTENTS. 2 (8144 Rev 2)

TABLE OF CONTENTS. 2 (8144 Rev 2) 1 (8144 Rev 2) TABLE OF CONTENTS Introduction Conventus CAGE TM - Proximal Humerus...3 Indications and Contraindications...4 Surgical Summary...5 Patient Positioning & Approach...6 Surgical Technique Plate

More information

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System.

2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System. 2.4 mm LCP Radial Head Plates. Part of the Synthes LCP Distal Radius Plate System. Technique Guide Instruments and Implants approved by the AO Foundation Table of Contents Introduction 2.4 mm LCP Radial

More information

Technique Guide. DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP).

Technique Guide. DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP). Technique Guide DCP and LC-DCP Systems. Dynamic Compression Plates (DCP) and Dynamic Compression Plates with Limited Bone Contact (LC-DCP). Table of Contents Introduction DCP and LC-DCP Systems 2 Indications

More information

Hoffmann II External Fixation System

Hoffmann II External Fixation System Hoffmann II External Fixation System Modular System for Long Bones Pelvis Introduction In 1938, Raoul Hoffmann, a surgeon from Geneva, Switzerland, designed a revolutionary External Fixation System. The

More information

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING

QUICK REFERENCE GUIDE. The PreFix Fixator (92000 Series) ALWAYS INNOVATING 21 The PreFix Fixator (92000 Series) ALWAYS INNOVATING INTRODUCTION The PreFix fixator is designed to provide temporary external fixation. This may be needed when local facilities or the condition of the

More information