1 : Controversies for Bearing Surfaces in THR Learning Objectives Upon completion of this activity, participants should be able to: 1. Understand the indications and known methods for the use of each of the alternative bearings over conventional polyethylene. 2. Describe the major disadvantages or potential risks with the use of the new alternative bearings when compared with conventional polyethylene. Moderator James A. D Antonio, MD Orthopaedic Surgeon Sewickley Valley Hospital Clinical Associate Professor of Orthopaedic Surgery Department of Orthopaedic Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Highly Cross-linked Polyethylene in THA Mark W. Pagnano, MD Associate Professor of Orthopedics Mayo Clinic College of Medicine Rochester, Minnesota Background: Contemporary total hip arthroplasties with ceramic-on-ceramic, metal-onmetal, and those with highly cross-linked polyethylene all appear to have favorable midterm results in regard to wear. Each bearing surface has a distinct set of advantages and disadvantages and for that reason a consensus on the appropriate use of each is unlikely to emerge in the near future. Some surgeons will undoubtedly choose 1 bearing surface and be comfortable in applying that to all patients. Other surgeons will selectively use each of the bearing surfaces in an effort to balance the pluses and minuses to the clinical circumstance of individual patients. Methods/Clinical results: In both basic science bench testing and early to mid-term clinical follow-up, each of the 3 bearing surfaces demonstrate a substantial reduction in wear as compared to traditional polyethylene. The reduction in wear that is seen with highly cross-linked polyethylene is not quite as dramatic as it is with the hard-on-hard ceramic or metal bearings. Highly cross-linked polyethylene does however present some distinct practical advantages over both ceramic and metal. Highly cross-linked poly is unlikely to experience the catastrophic problem of ceramic fracture nor does it cause the

2 marked elevation in serum metal ion levels caused by metal bearings. Impingement of the hip stem trunnion against a ceramic or metal articulation is cause for substantially more concern than if the impingement occurs against a cross-linked liner. Surgeons have a much wider array of intraoperative choices with cross-linked polyethylene liners in regard to head size (28, 32, 36, 40 etc), orientation (0, 10, 20 elevation), and offset (standard, lateral offset). Finally, as cost pressures once again reemerge it should be recognized that cross-linked polyethylene is substantially less expensive than comparable ceramic or metal bearings. Conclusions/Clinical relevance: For the majority of patients in my total hip practice their needs are met most appropriately with a highly cross-linked bearing surface. For very young, active female patients of child-bearing age, a ceramic-on-ceramic bearing has a role. For a subset of very active male patients, particularly those with posttraumatic arthritis, a metal-on-metal bearing is discussed. 1. Callaghan JJ, Cuckler JM, Huddleston JI, et al. How have alternative bearings (such as metal-on-metal, highly cross-linked polyethylene, and ceramic-onceramic) affected the prevention and treatment of osteolysis? J Am Acad Orthop Surg. 2008;16(suppl 1):S Jacobs CA, Christensen CP, Greenwald AS, et al. Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty. J Bone Joint Surg Am. 2007;89(12): Review. Why I Use Ceramic-Ceramic James D Antonio, MD Sewickley Valley Hospital Sewickley, Pennsylvania University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Background: Despite the great success of conventional ultra high-molecular-weight polyethylene as a bearing surface for THA, wear and resultant osteolysis became one of the leading causes of failure and reoperation. Three alternative bearings to conventional polyethylene have become popular in the past decade: metal-on-metal, metal or ceramicon-highly cross-linked polyethylene, and alumina ceramic-on-alumina ceramic. Alumina ceramic bearings are attractive because of their extreme hardness and scratch resistance, their hydrophilic nature with improved lubrication, the absence of metal ion release, far less volumetric wear debris, and their superior wear resistance compared to all other available bearing surfaces. Methods & materials: In October 1996 a prospective, controlled, randomized, multicenter trial was started comparing ceramic-on-ceramic to metal-on-conventional polyethylene.

3 Five surgeons at 5 independent sites have followed 475 patients divided between 3 cohorts who received ceramic-on-ceramic and 1 cohort who received the control metalon-polyethylene. There was no difference in demographics for the 4 study groups, and the average age of the patients in the 4 study groups ranged from 52 to 55 years at the time of implantation. Clinical results: Survivorship, revision for any reason, for the ceramic implants (96.5%), was significantly higher (P =.012) than for the control metal-on-polyethylene (91.3%). One percent of the ceramic patients report occasional squeaking, none of which are of any major significance or have resulted in revision. The only bearing surface related failures for the ceramic patients (0.5%) occurred for a liner fracture at 6 years and an anterior chip fracture recognized at 9 years. Bearing surface related failures for the polyethylene control occurred for osteolysis (2.1%). Osteolysis for the metal-onpolyethylene control group was seen in 31% and for the ceramic-on-ceramic in 1% (P = <.001). Conclusions & clinical relevance: Increased volumetric wear debris generated at the bearing surface in total hip arthroplasty can lead to osteolysis and a need for reoperation. Ceramic-on-ceramic bearings have superior wear resistance compared to all other available alternative bearing surfaces and provide an excellent alternative for young, active, and high-energy athletic individuals. 1. Clarke IC. Role of ceramic implants. Design and clinical success with total hip prosthetic ceramic-to-ceramic bearings. Clin Orthop Relat Res. 1992;(282): D Antonio JA, Capello W, Manley MT, et al. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;(436): Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty. 2002;17(5): Hamadouche M, Boutin P, Daussange J, et al. Alumina-on-alumina total hip arthroplasty: A minimum 18.5 year follow-up study. J Bone Joint Surg Am. 2002;84-A(1): Harris WH. The problem is osteolysis. Clin Orthop Relat Res. 1995;(311): Sedel L, Kerboull L, Christel P, et al. Alumina-on-alumina hip replacement. Results and survivorship in young patients. J Bone Joint Surg Br. 1990;72(4):

4 Metal-on-Metal Articulations With Large Diameter Heads Adolph V. Lombardi Jr, MD, FACS Joint Implant Surgeons, Inc. The Ohio State University Mount Carmel Health System New Albany, Ohio With the introduction of alternative bearings in the past several years, surgeons have moved from the traditional utilization of 22, 26, 28, and 32-mm heads to larger head diameters in total hip arthroplasty (THA). The reported benefits of large femoral heads are the enhanced stability secondary to the increased range of motion prior to impingement and the increased jump distance required for subluxation of the head from the acetabulum. Numerous studies have documented significant reduction in wear with highly cross-linked polyethylene, even when large femoral heads are used. The Achilles heel of highly cross-linked polyethylene, namely diminution and its mechanical properties, has been reconciled with second-generation highly cross-linked polyethylene. Large-head ceramic components are favored, not only secondary to the afforded enhanced stability, but also because of the increased fracture resistance. To date, there have been no reported incidences of fracture of femoral heads of size 36 mm or larger. However, it is only with metal-on-metal articulations that femoral heads can be offered in sizes that closely replicate normal anatomy. Following our initial experience with 28- and 32-mm metal-on-metal articulations, we performed a series of THA with 38-mm metal-on-metal articulations. We have reported our success with respect to enhanced stability with this particular device. In November 2004, a system of large modular head, monoblock shell metal-on-metal acetabular components was introduced. In this system, head sizes range from 38 mm to 60 mm, with acetabular components being 6 mm larger than the femoral head and therefore ranging from 44 mm to 66 mm. As of October 2008, we have performed 886 primary THAs in 786 patients with this device. In this series the average diameter of the implanted femoral head was 56 mm (range: mm; SD 3.9). In 89 randomly selected THA, the resected femoral head was measured intraoperatively with a caliper by an independent observer. The surgeon was not apprised of the measurement and proceeded to complete the arthroplasty according to standard protocol. In 274 THAs, radiographic measurements of the contralateral native femoral head were made using a calibrated PACS system. In cases measured intraoperatively, the femoral head size was restored to within 6 mm of the resected femoral head, with an average restoration of 0.7 mm increase. In cases measured with the calibrated PACS system, the implanted femoral head was found to be on average 2.4 mm larger than the contralateral femoral head. To date, there have been no dislocations in this series. Nine cups have been revised: 2 because of sepsis, 4 because of aseptic loosening, 1 because of failure of ingrowth, 1 because of metal hypersensitivity, and 1 because of metallosis. Metal Hypersensitivity Testing in Total Joint Arthroplasty Patients: Studies report that lymphocytic reactivity to metal may predict poor implant performance or failure in THA

5 and total knee arthroplasty (TKA). Reports also note an increased incidence of perivascular lymphocytic infiltration (PVLI) in the soft tissues surrounding failed implants in patients with metal hypersensitivity. These findings are of concern with metal-on-metal total hip arthroplasty. Serum from 22 patients with a painful joint or failed total joint arthroplasty, including 13 patients with metal-on-metal THA articulations, 1 metal-on-polyethylene THA, 1 ceramic-on-polyethylene THA, 5 patients with a painful metal-on-polyethylene TKA, 1 with metal-on-polyethylene unicompartmental knee arthroplasty, and 1 preoperative TKA patient, were sent for immune response testing at a national laboratory with previously validated and published methodology. Capsular tissue from 140 revision THA and 137 revision TKA was sent to pathology to rule out perivascular lymphocytic infiltration (PVLI). Serum from 16 of 22 (73%) patients was reactive to 1 or more metals. Ten of 15 patients with painful or failed hips (67%) and 6 of 7 (86%) with painful or failed knees tested positive, which was not statistically different. Eight of 13 (62%) patients with metal-onmetal THA articulations had positive reactions, which was no different statistically in comparison with the patients with metal-on-polyethylene and ceramic-on-polyethylene articulations, both of whom had positive reactions. In the overall cohort, reactivity was to 1 metal in 6 cases, 2 metals in 4, 3 metals in 1, 4 metals in 1, 5 metals in 2, 7 metals in 1, and 8 metals in 1. The most common reaction was to nickel, present in 13 cases (59%). Other reactions noted were to iron in 6, zirconium in 6, cobalt in 5, chromium in 5, molybdenum in 5, vanadium in 3, and aluminum in 3. The preoperative TKA patient tested demonstrated high reactivity to nickel, and mild to zirconium. Surgical pathology results revealed a significantly higher number of failed TKA (65 of 137; 47%) with evidence of PVLI compared with failed THA (46 of 140; 33%; P =.0132 Pearson s chisquare). These findings question the utility of serum metal sensitivity testing in the diagnosis and management of possible metal hypersensitivity with a similar prevalence of positive results as that previously published for osteoarthritic controls. Additionally, findings of PVLI do not appear to hallmark metal hypersensitivity associated with metal-on-metal THA, as the prevalence is higher with failed TKA than failed THA. New testing modalities require development to assist in the evaluation of painful metal-on-metal THA. 1. Smith TM, Berend KR, Lombardi AV Jr, et al. Metal-on-metal total hip arthroplasty with large heads may prevent early dislocation. Clin Orthop Relat Res. 2005;441: Lombardi AV Jr, Mallory TH, Cuckler JM, et al. Mid-term results of a polyethylenefree metal-on-metal articulation. J Arthroplasty. 2004;19(7 suppl 2): Cuckler JM, Moore KD, Lombardi AV Jr, et al. Large versus small femoral heads in metal-on-metal total hip arthroplasty. J Arthroplasty. 2004;19(8 suppl 3): Lombardi AV Jr, Mallory TH, Alexiades MM, et al. Short-term results of the M2a-taper metal-on-metal articulation. J Arthroplasty. 2001;16(8 suppl 1):

6 What Is the Evidence-Based Data for Bearing Surfaces? David Backstein, MD Associate Professor, Department of Surgery - Orthopaedics University of Toronto Mount Sinai Hospital Toronto, Ontario, Canada Modern THA-bearing surfaces include metal on metal (MOM), ceramic, and highly cross-linked polyethylene (XL). The harder the bearing surface, the greater the wear resistance and therefore MOM and ceramic-on-ceramic bearings have the potential for better wear resistance than polyethylene. XL polyethylene studies with greater than 4 years of follow-up have demonstrated reduced proximal head penetration and wear rates. 1,2 At 2-5 years, studies have shown a 42%-72% reduction in wear rates for annealed, cross-linked polyethylene. 3,4 Earlier generations of XL polyethylene suffered from reduced toughness, fatigue, and tensile strength. 5 Liner fractures appear related to high acetabular abduction angles combined with thin polyethylene. 6 A study of alumina ceramic heads on ceramic acetabular components demonstrated a 20- year survival of 85% for cemented and 61.2% for cementless cups. 7 A recent alumina-onalumina cementless design followed for an average of 62 months showed 99.7% of cups were stable, 1.4% developed osteolysis, and there were no fractures. 8 Wear in ceramic designs has been related to positioning of the cup in < 15 of anteversion. 9 Of concern has been the advent of squeaking associated with positions of excessive anteversion and abduction opening. 10 Retrievals studies have shown have rim chip fractures and cracks in the acetabular liners, particularly when the components are malpositioned. 10 A head fracture rate of 1.4% was recently reported with all cases occurring with short femoral necks. 11 MOM designs allow for low wear rates and the use of large heads, which improve stability and range of motion. The 20-year survivorship of the McKee-Farrar design was 84% and 74% at 28 years, with minimal osteolysis. 12 Concerns with MOM include elevated serum ion concentrations, metal allergy, pseudotumors, osteolysis, and runaway wear related to malpositioned acetabular components In summary, all 3 bearing surfaces have demonstrated substantial wear improvements over traditional polyethylene, but with potential complications when sockets are malpositioned. Clinical decision making must take into account not only patient activity level, but tolerance for potential ion elevations, squeaking, and the need for features such as elevated rims or large heads. 1. Digas G, Kärrholm J, Thanner J, et al. The Otto Aufranc Award: Highly crosslinked polyethylene in total hip arthroplasty: randomized evaluation of

7 penetration rate in cemented and uncemented sockets using radiostereometric analysis. Clin Orthop Relat Res. 2004;429: Bragdon CR, Greene ME, Freiberg AA, et al. Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36- vs 28-mm femoral heads. J Arthroplasty 2007;22(6 suppl 2): Martell JM, Verner JJ, Incavo SJ. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: A randomized prospective trial. J Arthroplasty. 2003;18(7 suppl 1): D Antonio JA, Manley MT, Capello WN, et al. Five-year experience with Crossfire highly cross-linked polyethylene. Clin Orthop Relat Res. 2005;441: Bradford L, Baker D, Ries MD, et al. Fatigue crack propagation resistance of highly crosslinked polyethylene. Clin Orthop Relat Res. 2004;429: Tower SS, Currier JH, Currier BH, et al. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89(10): Hannouche D, Hamadouche M, Nizard R, et al. Ceramics in total hip replacement. Clin Orthop Relat Res. 2005;430: D Antonio J, Capello W, Manley M, et al. Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res. 2005;436: Lusty PJ, Watson A, Tuke MA, et al. Wear and acetabular component orientation in third generation alumina-on-alumina ceramic bearings: an analysis of 33 retrievals [corrected]. J Bone Joint Surg Br. 2007;89(9): Walter WL, O toole GC, Walter WK, et al. Squeaking in ceramic-on-ceramic hips: the importance of acetabular component orientation. J Arthroplasty. 2007;22(4): Koo KH, Ha YC, Jung WH, et al. Isolated fracture of the ceramic head after thirdgeneration alumina-on-alumina total hip arthroplasty. J Bone Joint Surg Am. 2008;90(2): Jacobsson SA, Djerf K, Wahlström O. Twenty-year results of McKee-Farrar versus Charnley prosthesis. Clin Orthop Relat Res. 1996;(329 suppl):s60-s Beaulé PE, Campbell P, Mirra J, et al. Osteolysis in a cementless, second generation metal-on-metal hip replacement. Clin Orthop Relat Res. 2001;386: Boardman DR, Middleton FR, Kavanagh TG. A benign psoas mass following metal-on-metal resurfacing of the hip. J Bone Joint Surg Br. 2006;88(3): Willert HG, Buchhorn GH, Fayyazi A, et al. Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Joint Surg Am. 2005;87(1): Case Presentations/Q&A David Backstein, MD; James A. D Antonio, MD; Adolph V. Lombardi Jr, MD, FACS; Mark W. Pagnano, MD; Timothy Wright, PhD

Bone Bangalore

Bone Bangalore Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,

More information

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?

Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).

More information

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574) References 1. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital Cost of Dislocation After Primary Total Hip Arthroplasty. J Bone Joint Surg. 2006;88A:290-294. 2. Brodner W, Grübl A, Jankovsky R, Meisinger

More information

8.3 Tips and Tricks: Fracture of a Ceramic Insert

8.3 Tips and Tricks: Fracture of a Ceramic Insert Tips and Tricks 311 8.3 Tips and Tricks: Fracture of a Ceramic Insert with modern Ceramic Total Hip Replacement B.-W. Min, K.-S. Song, C.-H. Kang, K.-J. Lee, K.-C. Bae, C.-H. Cho and Y.-Y. Won Abstract

More information

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head

Enhancing stability and increasing range of motion. Metasul LDH Large Diameter Head Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that

More information

The Challenge of Durability in Hip Replacement

The Challenge of Durability in Hip Replacement The Challenge of Durability in Hip Replacement Gregory A. Tocks, D.O. Orthopedic Associates of Lancaster Ltd. INTRODUCTION Hip replacement is a very common operation, and the number of long-term postoperative

More information

Painful Metal-on-Metal Total Hip Arthroplasty

Painful Metal-on-Metal Total Hip Arthroplasty The Journal of Arthroplasty Vol. 21 No. 2 2006 Painful Metal-on-Metal Total Hip Arthroplasty Cambize Shahrdar, MD,* Pat Campbell, PhD,y Joseph Mirra, MD,y and Lawrence D. Dorr, MD* Abstract: Two patients

More information

CASE 56 Hip Pain. ? Diagnosis

CASE 56 Hip Pain. ? Diagnosis CSE 56 Hip Pain PIN IN THE HIP: 1. 50-year-old female who had undergone total hip arthroplasty [Metal on metal] of the right hip 3.4 years earlier. Shee had a tingling sensation in the trochanteric region

More information

Product Information. *smith&nephew POLARCUP Dual Mobility System

Product Information. *smith&nephew POLARCUP Dual Mobility System APPLYING SCIENCE TO MOBILITY Combined with VERILAST Technology for unmatched performance Product Information *smith&nephew POLARCUP Dual Mobility System APPLYING SCIENCE TO MOBILITY Combined with VERILAST

More information

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies.

Zimmer TM. Continuum Acetabular System. Bringing together proven technologies. Zimmer TM Continuum Acetabular System Bringing together proven technologies. The power to meet individual patient needs. Zimmer Continuum Acetabular System provides highly flexible solutions for orthopaedic

More information

Impingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C.

Impingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C. Georgi I. Wassilew Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C. Perka Charité - Universitätsmedizin Berlin Orthoload club, Berlin Reasons for revision in THA Reason

More information

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe

1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe 1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe Abstract Introduction: Osteolysis generated by wear debris remains a problem

More information

DIRECT ANTERIOR APPROACH

DIRECT ANTERIOR APPROACH DIRECT ANTERIOR APPROACH JOINT REPLACEMENT PROGRAM 2301 25TH STREET SOUTH FARGO ND 58103 CENTER FOR MINIMAL INVASIVE JOINT SURGERY (p) 701-241-9300 (tf) 866-887-9300 www.jointpain.md FARGO FERGUS FALLS

More information

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases

Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs: Report of Two Cases SAGE-Hindawi Access to Research Advances in Orthopedics Volume 2011, Article ID 519254, 4 pages doi:10.4061/2011/519254 Case Report Anterior Subluxation after Total Hip Replacement Confirmed by Radiographs:

More information

Understanding Hip Implant Options

Understanding Hip Implant Options Understanding Hip Implant Options Cup (Socket) Bearing (Liner) Head Stem (Femur) Modern Hip Implants (Ball) Metal Femoral Head on Marathon Cross-linked Polyethylene Liner Proven materials Polyethylene

More information

Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty

Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty The Journal of Arthroplasty Vol. 19 No. 8 Suppl. 3 2004 Metal-on-Metal Bearings in Cementless Primary Total Hip Arthroplasty Christian P. Delaunay, MD Abstract: One hundred cementless titanium primary

More information

The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006

The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006 The arc of pelvic motion has been shown to be as great as 70 through sit to stand activities 1 DiGioia CORR 2006 Overview Every patient moves differently 1 and their total hip replacement should be optimised

More information

Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty

Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty Wear Analysis of Second-generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty Christopher Samujh, MD; Samrath Bhimani, MS; Langan Smith, BS; Arthur L. Malkani, MD abstract A major

More information

Adverse Local Tissue Reaction in Total Hip Replacement

Adverse Local Tissue Reaction in Total Hip Replacement Adverse Local Tissue Reaction in Total Hip Replacement Authors 1 David C Evans, 2 Ross A Miller, 3 Stephen J Incavo 1 MD Candidate, Baylor College of Medicine Class of 2018 2 Houston Methodist Hospital,

More information

TOTAL HIP REPLACEMENT:

TOTAL HIP REPLACEMENT: THR Prosthesis Design TOTAL HIP REPLACEMENT: PROSTHESIS DESIGN FEATURES JESS JOHNSTON & MELINDA ZIETH History of Hip Prosthesis Joint Replacement Registry Implant Design Technology & Future History and

More information

Prevention of dislocation after hip replacement in elderly patients. Piotr WOJCIECHOWSKI, Damian KUSZ, Mariusz NOWAK, Konrad KOPEĆ

Prevention of dislocation after hip replacement in elderly patients. Piotr WOJCIECHOWSKI, Damian KUSZ, Mariusz NOWAK, Konrad KOPEĆ Prevention of dislocation after hip replacement in elderly patients. Piotr WOJCIECHOWSKI, Damian KUSZ, Mariusz NOWAK, Konrad KOPEĆ Department of Orthopedics and Traumatology Medical University of Silesia

More information

Repeated Ceramic Head Fracture after Ceramic-on-ceramic Total Hip Arthroplasty

Repeated Ceramic Head Fracture after Ceramic-on-ceramic Total Hip Arthroplasty Case Report http://dx.doi.org/10.5371/jkhs.2012.24.1.59 Repeated Ceramic Head Fracture after Ceramic-on-ceramic Total Hip Arthroplasty Myung-Sik Park, MD, Sun-Jung Yoon, MD, Myung-Jae Yoo, MD Department

More information

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear

ARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear The Journal of Arthroplasty Vol. 00 No. 0 2008 Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction Effects on Polyethylene Wear William L. Jaffe, MD,* Eric J. Strauss,

More information

Metal-on-metal articulations generate approximately to particles PSEUDOTUMORS FACTORS PATHOGENESIS

Metal-on-metal articulations generate approximately to particles PSEUDOTUMORS FACTORS PATHOGENESIS PSEUDOTUMORS Vasu Pai ARMD Adverse reaction to metal debris. Langton identified that there is no clear consensus in the literature defining the boundaries of the terms metallosis, aseptic lymphocyte-dominated

More information

Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size

Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size Clin Orthop Relat Res (2016) 474:365 371 DOI 10.1007/s11999-015-4319-5 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons SYMPOSIUM: 2015 HIP SOCIETY

More information

What Are Hip and Knee Replacement Implants Made Of?

What Are Hip and Knee Replacement Implants Made Of? What Are Hip and Knee Replacement Implants Made Of? Hip and knee replacement surgery involve replacing the worn-out bone and cartilage lining your hip or knee joint with new implants that are composed

More information

D E S I G N R A T I O N A L E

D E S I G N R A T I O N A L E D E S I G N R A T I O N A L E FREEDOM TO CHOOSE 1 PROVEN MODULAR SOLUTIONS The Pinnacle Acetabular Cup System is uniquely designed with a range of acetabular cup options, biological and mechanical fixation

More information

Reaching new heights. Comprehensive. Efficient. Simple.

Reaching new heights. Comprehensive. Efficient. Simple. Reaching new heights Comprehensive. Efficient. Simple. Various acetabular cup choices Compatible with the different head and liner options including VERILAST Technology Reach for proven OR efficient Instrumentation

More information

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008)

Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) Pseudo tumor a cystic or solid mass (MRI or ultrasound) 1% of MOM pts develop pseudo tumors by 5 years 17 pts, all female, Pandit et al (JBJS, 2008) 8% at 5yrs FU indicated by preliminary study from same

More information

Active Articulation E1

Active Articulation E1 Active Articulation E1 Dual Mobility Hip System Design Rationale Knees Hips Extremities Cement and Accessories PMI Trauma Technology Active Articulation E1 Dual Mobility Hip System Dual Mobility: Not a

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Hip dislocation after total hip arthroplasty can be classified

Hip dislocation after total hip arthroplasty can be classified Case Report Late Anterior Hip Dislocation after Metasul Metal-on-Metal Total Hip Arthroplasty for 13 Years: A Case Report Larbpaiboonpong V email : virojlarb@gmail.com Viroj Larbpaiboonpong 1 Suthorn Bavonratanavech

More information

Disclosure. Direct Anterior THA via Extension Table. History 11/3/2015

Disclosure. Direct Anterior THA via Extension Table. History 11/3/2015 Direct Anterior THA via Extension Table J. Masonis, M.D. OrthoCarolina Hip & Knee Center Chief Adult Hip & Knee reconstruction Carolinas Medical Center Disclosure Consultant & Design Smith & Nephew Zimmer

More information

Continuing the Tradition. VerSys Heritage Hip System

Continuing the Tradition. VerSys Heritage Hip System Continuing the Tradition VerSys Heritage Hip System Heritage Following the Tradition The low-friction hip prosthesis developed by Sir John Charnley has more than a 20-year history of outstanding results.

More information

FDA Safety Communication: Metal-on- Metal Hip Implants

FDA Safety Communication: Metal-on- Metal Hip Implants FDA Safety Communication: Metal-on- Metal Hip Implants Date Issued: Jan. 17, 2013 Audience: Orthopaedic surgeons Health care providers responsible for the ongoing care of patients with metal-onmetal hip

More information

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital)

Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) 1. Introduction Total hip replacement arthroplasty (THRA) using a metal-on-polyethylene articulation has become one of the most

More information

Peggers Super Summaries: THR and Revision Concepts

Peggers Super Summaries: THR and Revision Concepts Total Hip replacement: ARTICUALTIONS Ceramic on polyethylene Ceramic on ceramic for young Metal on metal (resurfacing) Metal (cobalt chrome) on polyethylene WEAR Wear debris similar to bacteria o < 10

More information

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY

A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY by Timothy McTighe, Exec. Dir., JISRF, Chagrin Falls, Ohio Ying Ko, Ph.D., Cincinnati, Ohio Russell B. Bennett, Ph.D., Cincinnati, Ohio James

More information

Functional safe zone. Safe zone

Functional safe zone. Safe zone Functional safe zone Every patient moves differently 1, and the amount of pelvic rotation through functional activities should be accounted for with total hip replacement. Pelvic tilt can change significantly

More information

Case Report Recurrent Catastrophic Ceramic Femoral Head Failure in Total Hip Arthroplasty

Case Report Recurrent Catastrophic Ceramic Femoral Head Failure in Total Hip Arthroplasty Case Reports in Orthopedics, Article ID 837954, 4 pages http://dx.doi.org/10.1155/2014/837954 Case Report Recurrent Catastrophic Ceramic Femoral Head Failure in Total Hip Arthroplasty S. M. M. Tai, 1 L.

More information

Optimum implant geometry

Optimum implant geometry Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information

Surface Roughness of Femoral Head Prostheses After Dislocation

Surface Roughness of Femoral Head Prostheses After Dislocation An Original Study Surface Roughness of Femoral Head Prostheses After Dislocation Kenny T. Mai, MD, Christopher Verioti, DO, Darryl D Lima, MD, PhD, Clifford W. Colwell, Jr., MD, and Kace A. Ezzet, MD Abstract

More information

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY

ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY E C H E L O N P R I M A R Y H I P S Y S T E M P R O D U C T R A T I O N A L E ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY Echelon Primary Total Hip System HIGH OFFSET STANDARD OFFSET Cementless

More information

Highlights of Hip and Knee Societies/AAHKS Specialty Day 2012

Highlights of Hip and Knee Societies/AAHKS Specialty Day 2012 Highlights of Hip and Knee Societies/AAHKS Specialty Day 2012 James Huddleston, MD Assistant Professor Department of Orthopaedic Surgery Stanford University Medical Center California Orthopaedic Association

More information

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement

National Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement National Joint Replacement Registry Lay Summary 2015 Annual Report Hip and Knee Replacement SUPPLEMENTARY REPORT 2015 TABLE OF CONTENTS Introduction... 1 A brief history of the Registry origins... 1 The

More information

The Leader in Orthopaedic Innovation

The Leader in Orthopaedic Innovation The Leader in Orthopaedic Innovation Wright is a leading international manufacturer and distributor of superior, easy to use, and innovative orthopaedic implants and instrumentation. For over 50 years,

More information

Revision Total Hip Arthroplasty after Ceramic Bearing Fractures in Patients Under 60-years Old; Mid-term Results

Revision Total Hip Arthroplasty after Ceramic Bearing Fractures in Patients Under 60-years Old; Mid-term Results ORIGINAL ARTICLE Hip Pelvis 30(3): 156-161, 2018 http://dx.doi.org/10.5371/hp.2018.30.3.156 Print ISSN 2287-3260 Online ISSN 2287-3279 Revision Total Hip Arthroplasty after Ceramic Bearing Fractures in

More information

CLINICAL PAPER / ORTHOPEDIC

CLINICAL PAPER / ORTHOPEDIC HIP LEG LENGTH AND OFFSET Kelley T.C. and Swank M.L. (2009) Using CAS leads to more accurate positioning within the safe zone (inclination between 30 and 50, anteversion between 5 and 25 ) CAS improves

More information

Two- to 9-Year Clinical Results of Alumina Ceramic-on-Ceramic THA

Two- to 9-Year Clinical Results of Alumina Ceramic-on-Ceramic THA CLINICAL ORTHOPAEDICS AND RELATED RESEARCH Number 453, pp. 97 02 2006 Lippincott Williams & Wilkins Two- to 9-Year Clinical Results of Alumina Ceramic-on-Ceramic THA Stephen B. Murphy, MD; Timo M. Ecker,

More information

CERAMIC-ON-METAL. References: Cat No: version 2

CERAMIC-ON-METAL. References: Cat No: version 2 CERAMIC-ON-METAL References: 1. Beaule, P. E., et al. Jumbo femoral head for the treatment of recurrent dislocation following total hip replacement. J.Bone Joint Surg (2002). 84-A: 256-63. 2. Jolles, B.

More information

Fracture of an Acetabular Component Inserted without Cement: A Case Report

Fracture of an Acetabular Component Inserted without Cement: A Case Report Syracuse University SURFACE Biomedical and Chemical Engineering College of Engineering and Computer Science 1997 Fracture of an Acetabular Component Inserted without Cement: A Case Report R T. Trousdale

More information

Stability without compromise. Epsilon Durasul Constrained Insert

Stability without compromise. Epsilon Durasul Constrained Insert Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common

More information

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5.

Templating and Pre Operative Planning 2. Preparation of the Acetabulum 4. Trial Sizing and Impaction of the Shell 5. Surgical Technique Contents Templating and Pre Operative Planning 2 Preparation of the Acetabulum 4 Trial Sizing and Impaction of the Shell 5 Cup Positioning 6 Joint Stability 7 Trial sizing and Impaction

More information

Metasul LDH Large Diameter Head

Metasul LDH Large Diameter Head Metasul LDH Large Diameter Head Surgical Technique Metasul LDH Large Diameter Head Surgical Technique Enhancing Stability and Increasing Range of Motion Metasul LDH Large Diameter Head Surgical Technique

More information

Are Ceramic-on-ceramic Bearings in Total Hip Arthroplasty Associated With Reduced Revision Risk for Late Dislocation?

Are Ceramic-on-ceramic Bearings in Total Hip Arthroplasty Associated With Reduced Revision Risk for Late Dislocation? Clin Orthop Relat Res (2015) 473:3790 3795 DOI 10.1007/s11999-015-4395-6 Clinical Orthopaedics and Related Research A Publication of The Association of Bone and Joint Surgeons SYMPOSIUM: 2014 INTERNATIONAL

More information

Bi-Polar 22.2mm & 28mm System - Operative technique

Bi-Polar 22.2mm & 28mm System - Operative technique Disclaimer Biomet UK Ltd, as the manufacturer of this device, does not practice medicine and does not recommend any particular surgical technique for use on a specific patient. The surgeon who performs

More information

Early complications after revision total hip arthroplasty with cemented dual-mobility socket and reinforcement ring

Early complications after revision total hip arthroplasty with cemented dual-mobility socket and reinforcement ring Acta Orthop. Belg., 2012, 78, 357-361 ORIGINAL STUDY Early complications after revision total hip arthroplasty with cemented dual-mobility socket and reinforcement ring Christophe PAttyn, Emmanuel AUDEnAErt

More information

What is the Best Bearing Type? 1/11/2011

What is the Best Bearing Type? 1/11/2011 What is the Best Bearing Type? 1/11/2011 A comparison of modern bearing types. Thomas P. Gross, M.D. As a patient there are four reasons you should consider a metal-on-metal bearing total hip replacement

More information

Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner

Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner ORIGINAL ARTICLE Hip Pelvis 27(4): 216-222, 2015 http://dx.doi.org/10.5371/hp.2015.27.4.216 Print ISSN 2287-3260 Online ISSN 2287-3279 Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene

More information

R3think your options. For US distribution only.

R3think your options. For US distribution only. R3think your options For US distribution only. R3think acetabular surgery It is widely accepted that total hip replacement surgery is one of the most successful procedures done today. Typically more thought

More information

LEGION Total Knee System. Exceptional Durability

LEGION Total Knee System. Exceptional Durability LEGION Total Knee System Exceptional Durability How many knee systems have been lab-tested to 30 years of simulated wear? Just one: LEGION Primary Knee System with VERILAST Technology The LEGION Primary

More information

THE CLINICAL PERFORMANCE OF METAL-ON-METAL AS AN ARTICULATION SURFACE IN TOTAL HIP REPLACEMENT

THE CLINICAL PERFORMANCE OF METAL-ON-METAL AS AN ARTICULATION SURFACE IN TOTAL HIP REPLACEMENT THE CLINICAL PERFORMANCE OF METAL-ON-METAL AS AN ARTICULATION SURFACE IN TOTAL HIP REPLACEMENT William T. Long, M.D.* ABSTRACT The metal-on-metal articulations in total hip arthroplasty (THA) were widely

More information

Hip Resurfacing vs. Standard Total Hip Replacement

Hip Resurfacing vs. Standard Total Hip Replacement 1 Hip Resurfacing vs. Standard Total Hip Replacement (My bias is obviously as a hip resurfacing proponent) Written June 6, 2011 HIP RESURFACING: PROS CONS 1. Less bone removal on the femur. 1. Femoral

More information

The power of simplicity

The power of simplicity The power of simplicity The power of simplicity Consider the facts, and take a look at the future: Revision TKA procedures are projected to double between now and 2015. 1 Restoration of a normal joint

More information

Design Rationale. ECHELON Primary Hip System

Design Rationale. ECHELON Primary Hip System Design Rationale ECHELON Primary Hip System ECHELON Primary Total Hip System Addressing clinical issues of cementless hip arthroplasty Cementless total hip arthroplasty has provided a proven method of

More information

Specialty Update What s New in Hip Arthroplasty BY MICHAEL H. HUO, MD, AND NATHAN F. GILBERT, MD

Specialty Update What s New in Hip Arthroplasty BY MICHAEL H. HUO, MD, AND NATHAN F. GILBERT, MD 2133 COPYRIGHT 2005 BY THE JOURNAL OF BONE AND JOINT SURGERY, INCORPORATED Specialty Update What s New in Hip Arthroplasty BY MICHAEL H. HUO, MD, AND NATHAN F. GILBERT, MD Total hip arthroplasty is one

More information

Exeter. Designed for Anatomic Restoration. Clinical Evidence Education

Exeter. Designed for Anatomic Restoration. Clinical Evidence Education Exeter Designed for Anatomic Restoration Clinical Evidence Education Anatomic Restoration Exeter re-creating biomechanics In hip replacement, the smallest adjustment can make a big difference. Stryker

More information

Ceramic Failure After Total Hip Arthroplasty with an Alumina-on-Alumina Bearing

Ceramic Failure After Total Hip Arthroplasty with an Alumina-on-Alumina Bearing This is an enhanced PDF from The Journal of Bone and Joint Surgery The PDF of the article you requested follows this cover page. Ceramic Failure After Total Hip Arthroplasty with an Alumina-on-Alumina

More information

Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty

Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty Case Reports in Orthopedics, Article ID 792781, 5 pages http://dx.doi.org/10.1155/2014/792781 Case Report Inflammatory Pseudotumor Complicated by Recurrent Dislocations after Revision Total Hip Arthroplasty

More information

Current Concepts of Using Large Femoral Heads in Total Hip Arthroplasty

Current Concepts of Using Large Femoral Heads in Total Hip Arthroplasty REVIEW ARTICLE Hip Pelvis 28(3): 134-141, 2016 http://dx.doi.org/10.5371/hp.2016.28.3.134 Print ISSN 2287-3260 Online ISSN 2287-3279 Current Concepts of Using Large Femoral Heads in Total Hip Arthroplasty

More information

Granulomatous Lung Disease: A Novel Complication following Metallosis from Hip Arthroplasty

Granulomatous Lung Disease: A Novel Complication following Metallosis from Hip Arthroplasty CSE REPORT Hip Pelvis 28(4): 249-253, 2016 http://dx.doi.org/10.5371/hp.2016.28.4.249 Print ISSN 2287-3260 Online ISSN 2287-3279 Granulomatous Lung Disease: Novel Complication following Metallosis from

More information

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique

Enhanced Stability Constrained Liners. Design Rationale Surgical Technique Enhanced Stability Constrained Liners Design Rationale Surgical Technique The Pinnacle Acetabular Cup System was designed to maximize the number of options available to the surgeon, and provide those options

More information

Featuring. Technology. Product Rationale

Featuring. Technology. Product Rationale Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge

More information

TITLE: Metal on Metal Total Hip Replacements or Hip Resurfacing for Adults: A Review of Clinical Effectiveness and Cost Effectiveness

TITLE: Metal on Metal Total Hip Replacements or Hip Resurfacing for Adults: A Review of Clinical Effectiveness and Cost Effectiveness TITLE: Metal on Metal Total Hip Replacements or Hip Resurfacing for Adults: A Review of Clinical Effectiveness and Cost Effectiveness DATE: 15 November 2012 CONTEXT AND POLICY ISSUES Total hip replacement

More information

Total Hip Arthroplasty Performed Using Conventional and Computer-Assisted, Tissue- Preserving Techniques 6

Total Hip Arthroplasty Performed Using Conventional and Computer-Assisted, Tissue- Preserving Techniques 6 Total Hip Arthroplasty Performed Using Conventional and Computer-Assisted, Tissue- Preserving Techniques 6 Stephen B. Murphy, MD, Timo M. Ecker, MD and Moritz Tannast, MD Introduction Less invasive techniques

More information

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC

Zimmer Longevity Highly Cross-linked Polyethylene. Clinical Value Dossier - APAC Zimmer Longevity Highly Cross-linked Polyethylene Clinical Value Dossier - APAC Table of Contents 1 Executive Summary...................................................... 1 2 Burden of Illness........................................................

More information

ALUMINA CERAMIC-CERAMIC TOTAL HIP ARTHROPLASTY USING COMPUTER- ASSISTED SURGICAL NAVIGATION AND A NEW MINIMALLY INVASIVE TECHNIQUE

ALUMINA CERAMIC-CERAMIC TOTAL HIP ARTHROPLASTY USING COMPUTER- ASSISTED SURGICAL NAVIGATION AND A NEW MINIMALLY INVASIVE TECHNIQUE In: Bioceramics in Joint Arthroplasty Darmstadt Germany: Steinkopff Verlag Publishers, 2004. ALUMINA CERAMIC-CERAMIC TOTAL HIP ARTHROPLASTY USING COMPUTER- ASSISTED SURGICAL NAVIGATION AND A NEW MINIMALLY

More information

The Spatial Location of Impingement in Total Hip Arthroplasty

The Spatial Location of Impingement in Total Hip Arthroplasty The Journal of Arthroplasty Vol. 15 No. 3 2000 The Spatial Location of Impingement in Total Hip Arthroplasty Motoi Yamaguchi, MD, PhD,* Toshihiro Akisue, MD, PhD,* Thomas W. Bauer, MD, PhD,* and Yasushi

More information

CAUTION: Ceramic liners are not approved for use in the United States.

CAUTION: Ceramic liners are not approved for use in the United States. Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of

More information

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS

SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments

More information

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE

This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE This publication is not intended for distribution in the USA. SURGICAL TECHNIQUE DePuy Synthes DURALOC Surgical Technique CONTENTS Templating and Pre Operative Planning 2 Preparation of the Acetabulum

More information

Informed Consent for HRA

Informed Consent for HRA Updated March 09 Thomas P Gross MD Informed Consent for HRA Dr. Gross has now performed over 5500 Hip Resurfacing Arthroplasty (HRA) procedures over the last 8 years. Most failures occur during the first

More information

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces

Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole

More information

Manza Cup HA SURGICAL TECHNIQUE.

Manza Cup HA SURGICAL TECHNIQUE. 1 PRE-OPERATIVE PLANNING. Preoperative assessment of the appropriate size and position of the acetabular component will provide intraoperative guidance for acetabular reaming. To determine the acetabluar

More information

Dual Mobility System Surgical technique

Dual Mobility System Surgical technique Trinity Dual Mobility System Surgical technique 2 Contents Operative summary 4 Overview 5 Operative technique 6 1. Acetabular reaming 6 Reamer guide 6 2. Acetabular shell trial 7 3. Acetabular shell implantation

More information

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code

BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code Patrick Chaboche, MD patrick.chaboche@wanadoo.fr Former Head of Department of Orthopaedic Surgery. Centre

More information

Dual Mobility System Evaluation surgical technique

Dual Mobility System Evaluation surgical technique Trinity Dual Mobility System Evaluation surgical technique Contents Operative summary 4 Overview 5 Operative technique 6 1. Acetabular reaming 6 Reamer guide 6 2. Acetabular shell trial 6 3. Acetabular

More information

Design Rationale and Clinical Review of a New Compliant Bearing Material for Acetabular Reconstruction

Design Rationale and Clinical Review of a New Compliant Bearing Material for Acetabular Reconstruction Joint Implant Surgery & Research Foundation Chagrin Falls, Ohio, USA Design Rationale and Clinical Review of a New Compliant Bearing Material for Acetabular Reconstruction Timothy McTighe, Dr. H.S. (hc)*;

More information

Metal ion levels in patients with a lumbar metal-on-metal Total Disc Replacement:

Metal ion levels in patients with a lumbar metal-on-metal Total Disc Replacement: Metal ion levels in patients with a lumbar metal-on-metal Total Disc Replacement: Similar concerns compared to metal bearing Total Hip Arthroplasties? EuroSpine 2011, Milan Job L.C. van Susante, MD, PhD

More information

RM Philosophy Innovation based on clinical evidence.

RM Philosophy Innovation based on clinical evidence. RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher

More information

Hip Resurfacing.

Hip Resurfacing. Hip Resurfacing http://www.birminghamhipresurfacing.com/hipresurfacing/technology.cfm The end of the femur is capped somewhat like a tooth cap The hip socket receives a cupped implant to move together

More information

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique

ESC. Enhanced Stability Liners. Design Rationale & Surgical Technique ESC Enhanced Stability Liners Design Rationale & Surgical Technique Choice Without Compromise DePuy Synthes PINNACLE Hip Solutions are designed with a wide range of acetabular cup options, biological and

More information

Scan Bi-Polar 22/28. Operative Technique

Scan Bi-Polar 22/28. Operative Technique Scan Bi-Polar 22/28 Operative Technique Disclaimer This publication and all content, artwork, photographs, names, logos and marks contained in it are protected by copyright, trademarks and other intellectual

More information

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval

Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval J. P. Collier,* M. B. Mayor,+ V. A. Surprenant,* H. P. Surprenant,* and R.E. Jensen* *Dartmouth Biomedical

More information

Comparison of ceramic-on-ceramic to metal-on-polyethylene bearing surfaces in total hip arthroplasty: a meta-analysis of randomized controlled trials

Comparison of ceramic-on-ceramic to metal-on-polyethylene bearing surfaces in total hip arthroplasty: a meta-analysis of randomized controlled trials Hu et al. Journal of Orthopaedic Surgery and Research (2015) 10:22 DOI 10.1186/s13018-015-0163-2 RESEARCH ARTICLE Open Access Comparison of ceramic-on-ceramic to metal-on-polyethylene bearing surfaces

More information

Scandinavian Journal of Surgery 103: 54 59, 2013

Scandinavian Journal of Surgery 103: 54 59, 2013 345SJS103110.1177/1457496913495345Hip resurfacing arthroplasty vs. large headed metal-on-metal total hip arthroplastym. Junnila, et al. ORIGINAL ARTICLE Scandinavian Journal of Surgery 103: 54 59, 2013

More information

Partial hip resurfacing is considered investigational/not medically necessary for all other indications not listed above.

Partial hip resurfacing is considered investigational/not medically necessary for all other indications not listed above. Subject: Policy #: Current Effective Date: 11/13/2006 Status: Revised Last Review Date: 09/14/2006 Description/Scope Hip resurfacing can be categorized in two ways: as a partial (hemi) hip resurfacing

More information

Optimizing function Maximizing survivorship Accelerating recovery

Optimizing function Maximizing survivorship Accelerating recovery Surgical Technique Optimizing Function Maximizing Survivorship Accelerating Recovery The company believes in an approach to patient treatment that places equal importance on: Optimizing function Maximizing

More information

ACETABULAR CUP SURGICAL TECHNIQUE

ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP SURGICAL TECHNIQUE ACETABULAR CUP DEVICE INDICATIONS FOR USE The ICONACY I-Hip total hip replacement is indicated for the following conditions: 1. A severely painful and/or disabled hip

More information

Optimum implant geometry

Optimum implant geometry Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available

More information